PERFORMANCE VERIFICATION: IONSCAN LS with Autosampler Module

Size: px
Start display at page:

Download "PERFORMANCE VERIFICATION: IONSCAN LS with Autosampler Module"

Transcription

1 Version 5.0 Page 1 of 8 PERFORMANCE VERIFICATION: IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ Copyright Smiths Detection 007 All rights reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed by copyright laws. Warranty The information contained in this document is subject to change without notice. Smiths Detection makes no warranty of any kind with regards to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Smiths Detection shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

2 Version 5.0 Page of 8 PRE EXECUTION PROTOCOL APPROVAL SMITHS DETECTION - PROTOCOL APPROVAL I agree that the Performance Verification (PV) procedures, assembled by Smiths Detection, are appropriate for the equipment defined in the table of contents. Name Signature Date CUSTOMER PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date QUALITY PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date POST EXECUTION PROTOCOL APPROVAL SMITHS DETECTION - PROTOCOL APPROVAL I agree that the PV procedures, assembled by Smiths Detection, are appropriate for the equipment defined in the table of contents. Name Signature Date CUSTOMER PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date QUALITY PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date

3 Version 5.0 Page 3 of 8 TABLE OF CONTENTS PRE EXECUTION PROTOCOL APPROVAL... SMITHS DETECTION - PROTOCOL APPROVAL... CUSTOMER PROTOCOL APPROVAL... QUALITY PROTOCOL APPROVAL... POST EXECUTION PROTOCOL APPROVAL... SMITHS DETECTION - PROTOCOL APPROVAL... CUSTOMER PROTOCOL APPROVAL... QUALITY PROTOCOL APPROVAL... TABLE OF CONTENTS...3 GENERAL INSTRUCTIONS...6 SIGNATURE PROTOCOL...6 ERROR CORRECTION...6 NO-ENTRY SPACES...6 USER REQUIREMENTS...6 PARTS AND MATERIALS...7 REQUIRED TEST PARTS AND MATERIALS...7 REQUIRED TEST STANDARDS...7 INSTRUMENT TO BE VERIFIED...8 TEST PROCEDURE...8 TEST LOCATION...9 DESORBER SEAL CHECK...10 PURPOSE...10 BACKGROUND...10 TEST EQUIPMENT LIST...10 PROCEDURE...10 RESULTS...10 AUTOSAMPLER ALIGNMENT CHECK...11 PURPOSE...11 BACKGROUND...11 TEST EQUIPMENT LIST...11 PROCEDURE...11 RESULTS...11 AUTOSAMPLER SYRINGE VOLUME CALIBRATION CHECK...1 PURPOSE...1 BACKGROUND...1 TEST EQUIPMENT LIST...1 PROCEDURE...1 RESULTS...1 AUTOSAMPLER SLIDE TRAY CHECK...13 PURPOSE...13

4 Version 5.0 Page 4 of 8 BACKGROUND...13 TEST EQUIPMENT LIST...13 PROCEDURE...13 RESULTS...13 NOISE CHECK...14 PURPOSE...14 BACKGROUND...14 TEST EQUIPMENT LIST...14 PROCEDURE...14 RESULTS...14 CALIBRANT ION CHECK...15 PURPOSE...15 BACKGROUND...15 PROCEDURE...15 RESULTS...16 REACTANT ION CHECK...17 PURPOSE...17 BACKGROUND...17 PROCEDURE...17 CARRYOVER CHECK...18 PURPOSE...18 BACKGROUND...18 TEST EQUIPMENT LIST...18 PROCEDURE...18 RESULTS...18 LIMIT OF DETECTION CHECK...19 PURPOSE...19 BACKGROUND...19 TEST EQUIPMENT LIST...19 PROCEDURE...19 RESULTS...19 DETECTOR LINEARITY CHECK...0 PURPOSE...0 BACKGROUND...0 TEST EQUIPMENT LIST...0 PROCEDURE...0 RESULTS...0 PEAK PROFILE CHECK...1 PURPOSE...1 BACKGROUND...1 TEST EQUIPMENT LIST...1 PROCEDURE...1 RESULTS...1 INJECTION REPEATABILITY CHECK... PURPOSE... BACKGROUND... TEST EQUIPMENT LIST...

5 Version 5.0 Page 5 of 8 PROCEDURE... RESULTS... SUMMARY OF PERFORMANCE VERIFICATION...3 FINAL CONCLUSION...3 NOISE ANALYSIS...4 REGRESSION ANALYSIS...4 DETERMINATION OF INJECTION REPEATABILITY...4 QUALIFICATION CONTROL PARAMETERS...4 AUTOSAMPLER MODULE...4 HIGH PERFORMANCE INJECTOR, AUTOSAMPLER ONLY...4 CRITICAL IONSCAN PARAMETERS, NEGATIVE AND POSITIVE ION MODES...4

6 Version 5.0 Page 6 of 8 GENERAL INSTRUCTIONS SIGNATURE PROTOCOL TESTED BY: Signature and/or initials of Smiths Detection Field Service Engineer, or other personnel performing tests. Date indicates the date test was executed. ACCEPTED BY: Signature and/or initials of customer representative. QA ACCEPTED: Signature and/or initials of customer QA department responsible for reviewing data. ERROR CORRECTION In the event of an entry error, cross out the incorrect value with a single line and write the correct entry. Initial and date the error. If space does not allow, footnote the error, and correction at the bottom of the page. eg. 0FEB04 NO-ENTRY SPACES Any entry that is not filled in (ie. test not necessary, space for comments not needed, etc.) should be crossed out with one diagonal line. Write a short reason, or N/A on the line along with initials and date. USER REQUIREMENTS The IONSCAN-LS detection system is a trace chemical detection tool that utilizes the analytical technique of Ion Mobility Spectrometry. Ion Mobility Spectrometry measures the movement of ions created via atmospheric pressure chemical ionization. Movement of ions is through a counter flow gas phase. Characteristic mobilities are automatically compared to user-generated libraries. Quantitative reports are generated by a custom 1CFR11 compliant software package. A dedicated liquid autosampler increases reproducibility ofthe compounds analyzed by this analytical tool. Analysis times are typically seconds. The timeliness of the data provides faster overall data turnaround for any analytical laboratory. It requires standard power, and a PC, for automated analysis, 1CFR11 compliance, and quantitation. The features of the IONSCAN-LS system are testable as documented in the Operation Qualification procedures, unless they are functionally assumed as inherent to the system, ie. Ni63 ionization source, or a constraint by design, ie maximum temperature limitations. The IONSCAN-LS analytical system has demonstrated it applicability in the areas of pharmaceutical cleaning validation, raw material analysis, illegal/counterfeit drug detection and interdiction, industrial hygiene and more

7 Version 5.0 Page 7 of 8 PARTS AND MATERIALS The following parts and materials are required for running the installation qualification. REQUIRED TEST PARTS AND MATERIALS Quantity Item 1 10-µL syringe, SGE or compatible 1 Solution standards kit 1 pk Teflon substrates p/n pk Teflon substrates p/n Teflon card, ring assembly 1 Personal computer running Windows 000 or XP 1 Excel software package 100 ml Hexane Pesticide grade or comparable 100 ml Acetone Pesticide grade or comparable 1pk -ml, 1-mm amber autosampler vials 1pk Caps with silicone/teflon septa REQUIRED TEST STANDARDS Quantity Solution 1 ml 5 pg/µl TNT in hexane 1 ml 50 pg/µl TNT in hexane 1 ml 100 pg/µl TNT in hexane 1 ml 150 pg/µl TNT in hexane 1 ml 00 pg/µl TNT in hexane 1 ml 50 pg/µl TNT in hexane 1 ml 500 pg/µl TNT in hexane 1 ml 50 pg/µl diazepam in acetone 1 ml 100 pg/µl diazepam in acetone 1 ml 150 pg/µl diazepam in acetone 1 ml 00 pg/µl diazepam in acetone 1 ml 50 pg/µl diazepam in acetone 1 ml 1000 pg/µl diazepam in acetone Solutions prepared by Smiths Detection using certified analytical standards. Analyte IUPAC Name CAS # Concentration Supplier Lot No.,4,6-TNT Diazepam 1-Methyl-,4,6- trinitrobenzene 1-Methyl-5-phenyl-7- chloro-1,3-dihydro- H-1,4-benzodiazepin- -one mg/mL 1mg/mL Cerilliant Cerilliant

8 Version 5.0 Page 8 of 8 INSTRUMENT TO BE VERIFIED IONSCAN MODEL# IM-Station Software Ver. or Serial Number PROCLAIMS Software Ver. Firmware Version Operational Time Negative Mode Positive Mode Autosampler s/n Sample Count A/S Firmware Version TEST PROCEDURE 1. Verify that the IONSCAN-LS with Autosampler module has been assembled correctly.. Turn instrument on and put into negative ion mode. 3. Verify communication between IONSCAN-LS, Autosampler and operating software. 4. Verify appropriate control parameters for the detection of TNT. Refer to Appendix 5. Allow instrument to come ready and stabilize for approximately 1 hour before proceeding with test. 6. Perform all checks in negative ion mode. 7. Switch instrument to positive ion mode. 8. Verify appropriate control parameters for the detection of diazepam. Refer to Appendix 9. Allow instrument to come ready and stabilize for approximately 1 hour before proceeding with test

9 Version 5.0 Page 9 of 8 TEST LOCATION Instrument underwent Performance Verification as of the date this protocol was performed. Company Name Address Department Building Number Room Date Performance Verification performed Performance Verification Performed by (Name and Signature): Verified PV is Complete by(name and Signature): Date

10 Version 5.0 Page 10 of 8 DESORBER SEAL CHECK PURPOSE To verify that when the IONSCAN-LS performs an analysis and the desorber stage moves up and seals the sample substrate between the desorber and the inlet, a proper seal is obtained. BACKGROUND To analyze a sample, the sample substrate is inserted into the tray on the front of instrument and the slide is moved to the right to introduce the sample into the analyzer. When the analysis cycle starts, the desorber moves up and seals the sample between the desorber heater and the IMS inlet. As the sample is heated the sample carrier gas (clean dry air) sweeps the desorbed vapors into the IMS detector. To provide efficient sample transfer the sample should be properly sealed between the desorber and inlet to ensure that no vapors escape out the side. TEST EQUIPMENT LIST 1. Teflon substrate card assembly with standard Teflon substrate (p/n ) inch calibration tool PROCEDURE This test is performed in negative ion mode only. 1. Remover Inlet cover. CAUTION! INLET IS HOT!. In STANDBY mode, initiate ANVIL CYCLE and select UP position. 3. Using the calibration tool, check gap between desorber and inlet flange. There should be a moderate amount of resistance; however, the bar should still be able to be moved. If an adjustment is needed, turn set screw to the right of desorber, and recycle the ANVIL and place it in the UP position. 4. Repeat until calibration is correct. 5. Allow IONSCAN-LS unit to come ready. 6. Assemble a new Teflon substrate, card and ring. 7. Insert substrate card into IONSCAN-LS sample tray and analyze. 8. Verify that the set screw on the bottom of the transfer line heater block does not interfere with the vertical movement of the sample tray. 9. After the analysis, check for seal on bottom of Teflon substrate. 10. Record below whether the seal on the bottom of Teflon substrate is even and centered on the substrate. If necessary, adjust the mechanical-stop set screw and repeat steps Replace substrate card assembly in tray and run an analysis. During the analysis gently pull on the substrate assembly. It should not be easily removed. RESULTS Is gap correct for inlet/desorber (0.115 )? Is the imprint of seal on the bottom of Teflon substrate (shown in gray below) even and centered on the substrate as shown in the figure below? Was substrate assembly held properly in place during an analysis? Properly Centered Imprint Improperly Centered Imprint

11 Version 5.0 Page 11 of 8 AUTOSAMPLER ALIGNMENT CHECK PURPOSE To verify the accuracy of the autosampler s positional setpoints. BACKGROUND It is critical that an autosampler be able to position itself at various points throughout a run in order for a successful analysis to take place. This test verifies the autosampler s positional setpoints. Once this test is completed, the other qualification tests can be run. TEST EQUIPMENT LIST 1. IONSCAN Autosampler module. 1-mm vial with cap 3. Teflon Substrate 4. Rinse/Waste vials/caps PROCEDURE Note: this test is performed in the negative ion mode only. 1. Place 50 µl of water in a clean, dry autosampler vial, cap the vial and then place the vial in position 1 on the autosampler tray.. Place Rinse and Waste vials with caps in autosampler tray 3. Place Teflon Substrate on sample tray and close 4. Power up autosampler. 5. On autosampler keypad press PROG and select COBRA SETUP. 6. In SETUP Menu, select CONFIGURATION. 7. Select SAMPLE Verify a/s syringe moves to the proper position indicated in the table below. 9. Press ENTER. 10. Repeat steps 6 10 for WASTE, RINSE, INJECT A. Note: If the autosampler is out of position at any of the check points, adjust the setting using the keypad and retest the check point. Be sure that the Right/Left coordinate is the same for all waste and rinse targets. Record any changes to the control parameters in the IM-Station method. RESULTS Target Proper Position Test Result Sample 1 Center of Sample 1 vial Tip of syringe needle must be below the surface of the water, but must not touch the bottom of the vial. Waste Center of Waste 1 vial Vertical Setting = 1000 Rinse Inject A Center of Rinse 1 vial Vertical Setting = 000 Center of Teflon membrane Tip of needle must extend beyond syringe guide, but not contact Teflon substrate

12 Version 5.0 Page 1 of 8 AUTOSAMPLER SYRINGE VOLUME CALIBRATION CHECK PURPOSE To verify the accuracy of the autosampler s syringe volume calibration. BACKGROUND This test verifies the autosampler s syringe volume calibration. It is critical that an autosampler be able to deliver the correct volume of solution in order for a successful analysis to take place. For example, when using a 10-µL syringe, the plunger movement between the zero position and the full-scale position should equal 10 µl, as read from the syringe scale. Once this test is completed, the other qualification tests can be run. TEST EQUIPMENT LIST 1. IONSCAN Autosampler module. 10-µL syringe PROCEDURE Note: This test is performed in the negative ion mode only. With autosampler installed: 1. Power up autosampler.. The gap between bottom of plunger and top of needle(0ul) should be set to smallest possible amount. If an adjustment is needed, loosen the two screws that secure the syringe holder, slide the syringe holder up or down, and re-tighten. Verify Zero position by cycling power on autosampler. 3. On autosampler keypad press PROG and select Cobra Setup. 4. In SETUP Menu, select CONFIGURATION. 5. Select Vol Calib. 6. Press ENTER 7. Verify that the bottom of the plunger is at the 10uL mark on syringe for 100% setting. Press ENTER. 8. Verify that the plunger rises above 10uL mark. Press ENTER then HOLD/STOP when done. 9. Turn autosampler off and on, and verify that the plunger motor can find its home and does not chatter. RESULTS Plunger Position Proper Position Test Result Zero Bottom of the plunger is above 0uL. Smallest gap is desirable. Full Scale (100%) Bottom of the plunger set at 10uL mark. Maximum Bottom of plunger above 10uL line.

13 Version 5.0 Page 13 of 8 AUTOSAMPLER SLIDE TRAY CHECK PURPOSE To verify the autosampler slide drive correctly positions the slide tray above the desorber plate for an analysis. BACKGROUND This test verifies the positioning of the slide tray when the autosampler moves it into position for an analysis. When moved into position, the tray should activate the magnetic switch to initiate an analysis. The tray should also be centered above the desorber plate and below the inlet block. The positioning is determined by the # of slider steps parameter in the A/S general page of the control parameters. This parameter specifies the number of steps made by the stepper motor in moving from the home position to be inserted position. TEST EQUIPMENT LIST 1. IONSCAN Autosampler module PROCEDURE Note: this test is performed in the negative ion mode only. With autosampler installed: 1. Power up autosampler. Place the IONSCAN in Ready mode. 3. Cut a strip of a business card and hold it against the mechanical-stop set screw. 4. Initiate a single injection autosampler run 5. This will bring the slide tray into the analysis position. Verify that: a. The tray does not crash into the mechanical stop. This can be verified by observing no right shift of slide tray upon return to home position after analysis is complete. b. The slide tray gently clamps down on the business card so that there is some resistance when attempting to pull out the card. 6. Record PASS/FAIL in the PV report RESULTS Slide Position Proper Position Test Result Inserted for an analysis About one business card width from the mechanical-stop set screw

14 Version 5.0 Page 14 of 8 NOISE CHECK PURPOSE To verify the IONSCAN-LS detector is operating within the defined acceptance limits for system noise. BACKGROUND Noise is an important characteristic that specifies a detector s performance. Low noise level is especially important when analyzing compounds at low trace levels. IONSCAN detector noise primarily originates from microphonic noise (motion of guard grid relative to collector plate) and preamp noise (resistance noise). The root mean square noise is determined by using a built-im-station software function to measure the root mean square noise in a defined drift time interval. RMS Amp RMS (root mean square) deviation from the mean signal amplitude over all segments of the Seg Range, in the drift time interval specified. P-P Amp Displays the peak-to-peak amplitude of all segments of the range (difference between maximum and minimum of all segments). TEST EQUIPMENT LIST 1. Instrument Manager Software. (IM-Station software). PROCEDURE Negative Ion Mode 1. With slide tray down and no Teflon substrate installed, run three blank samples.. Acquire sample plasmagram with IM-Station software. 3. With built in IM-Station software function, measure RMS and Peak to Peak noise from msec and record in table. Positive Ion Mode 1. With slide tray down and no Teflon substrate installed, run three blank samples.. Acquire sample plasmagram with IM-Station software. 3. With built in IM-Station software function, measure RMS and Peak to Peak noise from msec and record in table. RESULTS Negative Ion Mode Positive Ion Mode Expected Result Actual Result Expected Result Actual Result RMS Amp < 3 du < 3 du P-P Amp < 15 du < 15 du Are noise levels in negative ion mode within specified tolerances? Are noise levels in positive ion mode within specified tolerances?

15 Version 5.0 Page 15 of 8 CALIBRANT ION CHECK PURPOSE To verify that a calibrant peak is present in the correct position in both positive and negative modes. BACKGROUND After the IONSCAN unit has warmed up and temperatures have stabilized, the unit searches for the presence of the calibrant peak before the unit goes into READY mode. The IONSCAN-LS unit uses the calibrant peak position to determine the drift times of all programmed analyte peaks based on the equation below: t anal d K K o = anal t o cal cal d wheret anal d K cal o is the expected drift time of the analyte ion, t cal d is measured drift time of the calibrant peak, K anal are the programmed reduced ion mobility values of the analyte and calibrant peaks. o and Parameters which effect the calibrant ion drift time are drift tube temperature, inlet temperature, drift flow rate, ambient atmospheric pressure and humidity level of the drift gas. All these parameters are either controlled or compensated for. The instrument controls drift tube and inlet temperature and drift flow. An air purification cartridge provides clean dry air for the instrument to use. An absolute pressure transducer determines the absolute pressure that is used to determine the expected position of calibrant. The IONSCAN-LS is designed to operate only when its internal algorithm has identified the calibrant peak. This test will verify that the calibrant algorithm is properly programmed in and the IONSCAN-LS is detecting it within manufacturer s specifications. PROCEDURE Negative Ion Mode 1. Verify IONSCAN entered READY mode within 1 hour.. Record the calibrant installed in instrument. 3. From control parameters, record the calibrant Ko. 4. Run a blank sample and record the FWHM, maximum amplitude, and delta of the calibrant ion peak. The maximum amplitude and delta are taken from the Channel Data summary table. The FWHM is taken from a Gaussian fit of the first segment of the D plasmagram. Positive Ion Mode 1. Verify IONSCAN entered READY mode within 1 hour.. Record the calibrant installed in instrument. 3. From control parameters, record the calibrant Ko. 4. Run a blank sample and record the FWHM, maximum amplitude, and delta of the calibrant ion peak. The maximum amplitude and delta are taken from the Channel Data summary table. The FWHM is taken from a Gaussian fit of the average of the D plasmagram.

16 Version 5.0 Page 16 of 8 RESULTS READY after 1 hour? Calibrant Calibrant Ko Calibrant FWHM (μs) Calibrant Max Amp (du) Calibrant delta (μs) Negative Ion Mode Positive Ion Mode Ready after 1 hour? Calibrant Ko within expected range (See table)? Calibrant FWHM within expected range (See table)? Calibrant Max Amp greater than threshold listed (See table)? Calibrant delta within + 60 μs? Negative Ion Mode Positive Ion Mode Calibrant Ko Range FWHM Range (μs) Max Amp Threshold (du) IBA (Isobutyramide) Nic (Nicotinamide) NBN (4-nitrobenzonitrile)

17 Version 5.0 Page 17 of 8 REACTANT ION CHECK PURPOSE To verify that reactant peaks are present in the correct position in the negative mode. BACKGROUND In the negative mode, a reactant is added via a diffusion source in the sample path to aid in the ionization process. Since the IONSCAN-LS does not actively monitor the reactant level, as it does the calibrant, it is necessary to verify the level to establish that the instrument is operating within specifications. This is done by evaluating the reactant level at the beginning and end of a series of consecutive runs. The IONSCAN-LS must maintain a minimum level over the course of the analyses. PROCEDURE Negative Ion Mode Only 1. Run a blank sample with no substrate present.. Scroll through segments and using the Gaussian fit function, find the maximum amplitude of the reactant peak nearest 7.0 ms. 3. Run 5 consecutive analyses (starting each analysis as soon as the IONSCAN-LS comes READY) with no substrate present. 4. On the last sample, scroll through segments and using the Gaussian fit function, find the maximum amplitude of the reactant peak nearest 7.0 ms. Initial Reactant Amplitude Reactant Amplitude after Sample series Maximum Reactant Amplitude Amplitude Threshold 600 du 00 du Max Amp above threshold?

18 Version 5.0 Page 18 of 8 CARRYOVER CHECK PURPOSE To test for carryover following an injection of a high-concentration standard. BACKGROUND When a high-concentration sample is analyzed, it has the potential to persist in the instrument and contribute to the instrument response observed for the next sample. This is termed carryover. To test for carryover, an analysis of a high-concentration standard is performed, followed by an analysis of a solvent blank. The standard should not be detected when the blank is analyzed. TEST EQUIPMENT LIST pg/µL TNT solution pg/µL diazepam solution 3. Teflon membrane (p/n ) PROCEDURE Negative Ion Mode 1. Prepare TNT solution and 3 hexane blanks in 1-mm amber vials. Cap and place in autosampler tray.. Verify that suitable parameters for an Autosampler Only run are entered in. Refer to Appendix. 3. Run autosampler method of TNT standard followed by 3 blank injections onto Teflon membrane 3 times. 4. Record cumulative amplitude of TNT after each standard and blank run. Positive Ion Mode 1. Prepare diazepam solution and 3 acetone blanks in 1-mm amber vials. Cap and place in autosampler tray.. Verify that suitable parameters for the Autosampler Only run are entered in. Refer to Appendix. 3. Run autosampler method of diazepam standard followed by 3 blank injections onto Teflon membrane 3 times. 4. Record cumulative amplitude of diazepam after each standard and blank run. Note: If the 1 st blank after the TNT or diazepam standard gives no detection, then the remaining two can be skipped. RESULTS Negative Ion Mode, Teflon 500 pg of Sample TNT (CumAmp) A 1 3 Average Positive Ion Mode, HPI 1000 pg of Sample diazepam (CumAmp) A 1 3 Average Blanks (TNT CumAmp) Blank 1 Blank Blank 3 Blanks (diazepam CumAmp) Blank 1 Blank Blank 3 Sum of Blanks (CumAmp) B Sum of Blanks (CumAmp) B % Carry Over: (B/A) x 100 % Carry Over: (B/A) x 100 Is average % carryover in negative mode < %? Is average % carryover in positive mode < %?

19 Version 5.0 Page 19 of 8 LIMIT OF DETECTION CHECK PURPOSE To verify the IONSCAN-LS system is able to detect a minute quantity of TNT and diazepam. BACKGROUND This test verifies the ability of the IONSCAN-LS to detect low levels of a standard. This is accomplished by demonstrating the detection of 5 pg of TNT on Teflon in negative ion mode and 50 pg of diazepam on the HPI in positive ion mode. This is done to test both means of sample introduction. TEST EQUIPMENT LIST 1. Teflon membrane (p/n ). 5-pg/µL TNT solution pg/µL diazepam solution PROCEDURE Negative Ion Mode 1. Place a Teflon membrane in the swipe tray and close tray.. Prepare a 5-pg/µL TNT solution in a 1-mm amber autosampler vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for an Autosampler Only run are entered in. Refer to Appendix. 4. Run a solvent blank. The TNT response must be zero before continuing to the next step. 5. Run autosampler method with 3 replicate injections onto the Teflon membrane. 6. Record if TNT was detected after each standard run. Positive Ion Mode 1. Place a Teflon membrane in the swipe tray and close tray.. Prepare a 50-pg/µL diazepam solution in a 1-mm amber autosampler vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for the Autosampler and HPI run are entered in. Refer to Appendix. 4. Run a solvent blank. The diazepam response must be zero before continuing to the next step. 5. Run autosampler method with 3 replicate injections into the HPI port. 6. Record if diazepam was detected after each standard run. RESULTS Run 1 Run Run 3 TNT signal (Cum Amp) Diazepam signal (Cum Amp) Did the last blank before Run 1 give a response of zero before both tests? Was 5 pg of TNT on Teflon detected in all 3 runs? Was 50 pg of diazepam on the HPI detected in all 3 runs?

20 Version 5.0 Page 0 of 8 DETECTOR LINEARITY CHECK PURPOSE To verify the detector shows a linear response over a working range. BACKGROUND Detectors become non-linear when the amount of the test compound increases above a certain limit. The linearity test is based on a sequence of runs where a TNT solution or diazepam solution is injected several times from vials containing different concentrations using the same injection volume. For the appropriate concentration range, when the instrument response is plotted against the concentration, the points should lie on a straight line. The square of the correlation coefficient, r, of this calibration curve is used to assess detector linearity. Three injections of five standards will be analyzed. TEST EQUIPMENT LIST 1. Teflon membrane (p/n ). 50, 100, 150, 00, and 50 pg/µl TNT solutions 3. 50, 100, 150, 00, and 50 pg/µl diazepam solutions PROCEDURE Negative Ion Mode 1. Place a Teflon membrane in the swipe tray and close tray.. Prepare all TNT solutions in 1-mm amber vials. Cap and place in autosampler tray. 3. Verify that suitable parameters for an Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method with 3 replicate injections of each standard onto the Teflon membrane. 5. Record the cumulative amplitude of TNT after each standard run. Positive Ion Mode 1. Verify a Teflon membrane is in the swipe tray and close tray.. Prepare all diazepam solutions in 1-mm amber vials. Cap and place in autosampler tray. 3. Verify that suitable parameters for the Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method with 3 replicate injections of each standard onto the Teflon membrane. 5. Record the cumulative amplitude of diazepam after each standard run. Note: Use a suitable program to plot average amplitude versus concentration and to calculate the r values based on the equation y = ax + b. RESULTS Sample Conc. pg/µl TNT on Teflon, Cum Amp Run 1 Run Run 3 Avg. Conc. pg/µl Sol Sol Sol Sol Sol Slope = Intercept = r = Is r > 0.98? Diazepam on HPI, Cum Amp Run 1 Run Run 3 Avg.

21 Version 5.0 Page 1 of 8 PEAK PROFILE CHECK PURPOSE To verify that the peak profile for a compound has the proper appearance time and width. BACKGROUND The peak profile of a standard is an important indicator of optimal performance of the IONSCAN-LS/HPI system. A peak at the characteristic ion drift time for the standard should be detected a certain length of time after the start of the IONSCAN analysis, and the time profile of the peak should have a certain width. This is evaluated by analyzing a 00 pg/ul standard of TNT in the negative ion mode and a 00 pg/ul standard of diazepam in the positive mode. The desorption time at maximum amplitude and number of segments in which the peak is detected are recorded and compared to operational specifications. TEST EQUIPMENT LIST 1. Teflon membrane. 00 pg/µl TNT solution pg/µl diazepam solution PROCEDURE Negative Ion Mode 1. Place a Teflon membrane in the swipe tray and close tray.. Prepare a 00 pg/µl TNT solution in a 1-mm amber vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for an Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method that performs three replicate injections using the Teflon membrane. 5. For the third analysis of 00 pg/μl of TNT, record the desorption time at maximum amplitude and the number of segments in which TNT was detected in table below. Positive Ion Mode 1. Verify a Teflon membrane is in the swipe tray and close tray.. Prepare a 00 pg/µl diazepam solution in a 1-mm amber vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for the Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method that performs three replicate injections using the Teflon membrane. 5. For the third analysis of 00 pg/μl of diazepam, record the desorption time at maximum amplitude and the number of segments in which diazepam was detected in table below. RESULTS Negative Mode, TNT via Teflon Desorption Max Amp Specified Tolerance s Number of detected segments 5-6 Positive Mode, diazepam via HPI Desorption Max Amp Specified Tolerance s Number of detected segments pg of TNT Pass Specification? 00 pg of diazepam Pass Specification?

22 Version 5.0 Page of 8 INJECTION REPEATABILITY CHECK PURPOSE To verify that IONSCAN-LS repeatability for the analysis of sample aliquots is within acceptable limits. BACKGROUND The injection volume precision is based on a sequence of runs where a TNT standard is used in negative ion mode and a diazepam standard is used in positive ion mode. Each standard is injected 5 times, always with the same volume. The average cumulative amplitude and relative standard deviation (RSD) are determined for each standard, and the RSDs are compared to acceptance criteria. TEST EQUIPMENT LIST 1. Teflon membrane (p/n ). 50 pg/µl diazepam solution pg/µl TNT solution PROCEDURE Negative Ion Mode 1. Place a Teflon membrane in the swipe tray and close tray.. Prepare a 50 pg/µl TNT solution in a 1-mm amber vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for an Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method with 5 replicate injections onto Teflon membrane. 5. Record cumulative amplitude of each standard in table below. Positive Ion Mode 1. Verify a Teflon membrane is in the swipe tray and close tray.. Prepare a 50 pg/µl diazepam solution in a 1-mm amber vial. Cap and place in autosampler tray. 3. Verify that suitable parameters for the Autosampler Only run are entered in. Refer to Appendix. 4. Run autosampler method with 5 replicate injections onto Teflon membrane. 5. Record cumullative amplitude of each standard run in table below. RESULTS Sample 1 Sample Sample 3 Sample 4 Sample 5 Average Std Dev Relative Standard Deviation (RSD) Is RSD < 8 % 50 pg TNT Teflon membrane (Cum Amp) 50pg Diazepam HPI port (Cum Amp)

23 Version 5.0 Page 3 of 8 SUMMARY OF PERFORMANCE VERIFICATION Check DESORBER SEAL CHECK AUTOSAMPLER ALIGNMENT CHECK AUTOSAMPLER SYRINGE VOLUME CALIBRATION CHECK AUTOSAMPLER SLIDE TRAY CHECK Negative Ion Mode Positive Ion Mode NOISE CHECK CALIBRANT ION CHECK REACTANT ION CHECK CARRYOVER CHECK LIMIT OF DETECTION CHECK DETECTOR LINEARITY CHECK PEAK PROFILE CHECK INJECTION REPEATABILITY CHECK FINAL CONCLUSION Did Instrument pass all of the tests in the above table? Yes (PASSED QUALIFICATION) No (FAILED QUALIFICATION) Protocol Deviations and/or Comments:

24 Version 5.0 Page 4 of 8 APPENDIX 1 In this appendix the documentation of mathematical or statistical formulas used in the tests described in the previous chapters are discussed. NOISE ANALYSIS During an analysis, multiple scans are co-added to define a segment of analysis and the total number of segments defined by the analysis time. For example a typical scan period is 0 msec, 0 scans are co-added to define segment of analysis (0.4 sec) and the analysis time is 8 seconds giving 0 segments of analysis. The IM-Station software noise analysis feature calculates signal and peak-to-peak amplitude data in the specified drift time interval. A drift time interval at the beginning of the analysis is selected which is free of peaks. The following data is then displayed in the top left corner of the plasmagram: Seg Range: RMS Amp: P-P Amp: Max P-P Seg: Max P-P Amp: Range of segments used in the calculations. RMS (root mean square) deviation from the mean signal amplitude over all segments of the Seg Range, in the drift time interval specified. Displays the peak-to-peak amplitude of all segments of the range (difference between maximum and minimum of all segments). Specifies the segment with the highest peak-to-peak amplitude. Shows the peak-to-peak amplitude of the Max P-P Seg. 3D Plasm agram 40 Peak Height (du) Drift Time (ms) IN THE ABOVE FIGURE NOISE WOULD BE ANALYZED OVER THE DRIFT TIME INTERVAL FROM 5.0 TO 5.3 MSEC. EACH OF THE LINES IN THE FIGURE IS A SEGMENT OF ANALYSIS. DATA SET OVER WHICH NOISE ANALYSIS IS PERFORMED: aij signal (du) for segment j at point i in drift time s j Number of scans per segment for segment j n, n 1 Range of indices ( i ) over which RMS and Peak-to-Peak noise is to be calculated k, k 1 Range of segments ( j ) over which RMS and Peak-to-Peak noise is to be calculated Define:

25 Version 5.0 Page 5 of 8 u b b j = ( n n + 1) u i= u1 a j, min ij j 1 ij = Min( a u ) For n1 i n = Max( aij u ) For n1 i n j, max j σ j = u [ aij u j ] i= u1 ( n n + 1) 1 Therefore, the RMS Amp is σ = k j= k σ j s j 1 k j= k 1 s j Max P-P Amp: Maximum Peak-to-Peak Amplitude: b = Max b [ b ] P P j, max j,max For j= k 1, k Max P-P Seg: Segment associated with Max P-P amplitude kmax = Value of j on which b p-p is based Further: min Min( b j, min ) max Max( b j, max b = b = Therefore, P-P Amp is b abs = ( b max bmin ) )

26 Version 5.0 Page 6 of 8 REGRESSION ANALYSIS The linearity test uses regression analysis during the evaluation. Linearity is derived from the linear function: Let: N = number of discrete observations X i = independent variable, i th observation Y i = dependent variable, i th observation Determination of coefficients: X Y b N = = 1 X i N N = i = 1 N Y i i N { ( X i X )( Yi Y )} i= 1 = N ( X i X ) i= 1 a = Y bx Regression Coefficient: r = Y X = a + bx N { ( X i X )( Yi Y )} N N ( X ) ( ) i X Yi Y i= 1 i= 1 i= 1 DETERMINATION OF INJECTION REPEATABILITY N = number of replicate samples A = Measured value associated with i th sample i Average A Standard Deviation SD = N i= = 1 N A i N ( Ai A) i= 1 N 1 (Results are a small sampling) Relative Standard Deviation RSD SD 100 = A

27 Version 5.0 Page 7 of 8 APPENDIX QUALIFICATION CONTROL PARAMETERS AUTOSAMPLER MODULE Autosampler Only Mode, Negative and Positive Ion Modes (Values listed are for Cobra II Autosampler, with values for Cobra I Autosampler in parentheses) Start Sample Vial: As placed #Sample Rinses 0 Pre-Dispense Delay (s) 0 End Sample Vial: As placed #Sample Pumps 4 Post-Disp. Delay (s) 7 (5) #Repetitions/Sample As needed Sample Fill Rate % 40 (85) Retract Delay (s) 0 #Rinses/Solv #1 5 Sample Dispense Rate % 100 Start Delay (s) 0 #Rinses/Solv# 0 Syringe Offset (μl) 0 Dispense to Target A Rinse Volume (μl) 5 Pre-Fill Air (μl) 0 Min. Sub. Cool Time (s) 10 Rinse Fill Rate % 50 (85) Mid-Fill Air (μl) 0 Oper. Mode No HPI Spike Volume (μl) 0 Post-Fill Air (μl) 0.5 (Continuous) Sample Volume (μl) 1.0 Post-Fill Delay (s) 0 HIGH PERFORMANCE INJECTOR, Autosampler Only UseHPI be set to OFF, which places the HPI in StandBy CRITICAL IONSCAN PARAMETERS, Negative and Positive Ion Modes Analysis delay following start of desorption 0.05 s Scan Period 0 ms Maximum analysis duration 1.00 s No. of co-added scans/analysis 10 Detection Algorithm: Channel Ko Vblty Blok Th Amp Th FWHM Seq Hits TNT μs 1.5 du 30 du 330 μs 1 diazepam μs 1.5 du 30 du 410 μs 1

28 Version 5.0 Page 8 of 8 REVISION HISTORY 1. Version 1.0, Effective Date February 19, Based on a rework of Version.1, Installation and Operation Qualification for IONSCAN-LS With Autosampler Module, Effective Date December 13, 00.. Version.0, Effective Date July 7, Removed Calibrant and Reactant Ion Check. - Removed Data Acquisition tables from Test Procedure - Added option to use Teflon membrane filters 3. Version 3.0, Effective Date, February 0, Based on a rework of Version 5.0, Installation and Operation Qualification for IONSCAN-LS With Autosampler Module, Effective Date March 4, Version 3.1, Effective Date, July 1, Adjust Ranges for Calibrant Ion Check. 5. Version 3., Effective Date, August 19, Based on rework of Version 5.1, Installation and Operation Qualification for IONSCAN-LS With Autosampler Module, Effective Date August, Version 4.0, Effective Date: September 10, Based on a rework of Version 1.50, Manufacturing Operational Qualification: IONSCAN-LS with High Performance Injector (HPI) and Autosampler Module, Effective Date August 5, Version 4.1, Effective November, Changed parameters and specifications for Calibrant Ion Check, Limit of Detection Check, and Injection Repeatability Check. 8. Version 4., Effective October 1, Corrected errors in Injection Repeatability Check 9. Version 5.0, Effective October 16, Based on rework of Version.0 Performance Verification: IONSCAN-LS with High Performance Injector(HPI) and Autosampler Module, Effective Date October 3, 007.

PERFORMANCE VERIFICATION IONSCAN LS

PERFORMANCE VERIFICATION IONSCAN LS Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--900 Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without

More information

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter

More information

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing Automated Preparation of Simulated Distillation Standards and Samples for ASTM Methods D2887, D7213, D7398 and D6352 using the 7693A System with Easy SamplePrep Software Author Roger L. Firor Agilent Technologies,

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection

More information

In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide

In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide This guide is for users who have been trained in the proper use of the AssayMAP Bravo Platform and understand the safety guidelines in the Bravo

More information

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry The end of mass-speculation Pre-cleaned and certified vials for mass spectrometry Bonded, ultra-pure closures LC/MS and GC/MS Manufacturing the Ultimate Vial for Mass Spectrometry More than a quarter century

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies,

More information

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process CHROMTECH Application Note PB 401 Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process By LP Raman* Technical Marketing Manager Whatman, Inc. 9 Bridewell

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) 1.0 Equipment: Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) Thermo Exactive Orbitrap Mass Spectrometer Direct Analysis

More information

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010 TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010 BIOCHEMISTRY The Tetrachlorethylene or perchlorethylene, is an organic halide. Its structure is similar to that of a molecule of ethene which

More information

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49109AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510 METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510 BIOCHEMISTRY The methyl isobutyl ketone is a colorless and flammable liquid that is used as organic solvent in the laboratory

More information

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG Quantification of residual Protein A ligands in the presence of excess amounts of IgG Product number P0020456 P0020457 Product Name Gyrolab Protein A Kit for MabSelect SuRe Ligand Gyrolab Protein A Kit

More information

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

ph electrode Instruction Manual For use with the RAH-210 & RPH-250 ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

Analyst Software. Automatic Optimization Tutorial

Analyst Software. Automatic Optimization Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

ChromTech options for PAL systems

ChromTech options for PAL systems ChromTech options for PAL systems SPDE features full automation for dynamic sample extraction, desorption and analysis using the CTC Combi PAL. Most of our customers use SPDE for food and flavor research,

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.

More information

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94. Ethanol Project Chemical Analysis and Testing Task Laboratory Analytical Procedure Procedure Title: LAP-013 HPLC Analysis of Liquid Fractions of Process Samples for Monomeric Sugars and Cellobiose Author:

More information

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP 1.0 SCOPE AND APPLICATION 1.1 This method is suitable for the rigorous cleanup of sample extracts prior to analysis for polychlorinated biphenyls. This method

More information

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput JSB is an authorised partner of Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput Application Note - Environmental #129 Author Anne Jurek Applications Chemist

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Application Note. Automated pipetting and analysis of polymer suspensions for process control in chemical production

Application Note.   Automated pipetting and analysis of polymer suspensions for process control in chemical production Application Note www.palsystem.com Automated pipetting and analysis of polymer suspensions for process control in chemical production Pipetting suspensions for process control in chemical production Anders

More information

FOR RESEARCH USE ONLY.

FOR RESEARCH USE ONLY. MAN-10039-04 Vantage 3D DNA SNV Qualification Kit Vantage 3D DNA SNV Qualification Kit The ncounter Vantage 3D DNA SNV Qualification Kit is designed to assess whether the ncounter MAX, FLEX, or SPRINT

More information

GC/MS Application Note

GC/MS Application Note GC/MS Application Note Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass Spectrometry www.palsystem.com Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer

More information

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities

More information

Conductometric Titration & Gravimetric Determination of a Precipitate

Conductometric Titration & Gravimetric Determination of a Precipitate Conductometric Titration & Gravimetric Determination of a Precipitate Experiment 9 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H2SO4, and barium hydroxide,

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Micro Volume QuEChERS kit

Micro Volume QuEChERS kit 225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page

More information

SC125MS. Data Sheet and Instruction Manual. ! Warning! Salem Controls Inc. Stepper Motor Driver. Last Updated 12/14/2004

SC125MS. Data Sheet and Instruction Manual. ! Warning! Salem Controls Inc. Stepper Motor Driver.   Last Updated 12/14/2004 SC125MS Stepper Motor Driver Salem Controls Inc. Last Updated 12/14/2004! Warning! Stepper motors and drivers use high current and voltages capable of causing severe injury. Do not operate this product

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

GC Application Note. Dilution/standard addition: Efficient and traceable.

GC Application Note. Dilution/standard addition: Efficient and traceable. GC Application Note Dilution/standard addition: Efficient and traceable. www.palsystem.com Dilution / standard addition: Efficient and traceable. Philippe Mottay Brechbuehler AG Steinwiessenstrasse 3 CH-8952

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Keywords: Hydrodynamic radius, diffusion coefficient, sizing, Taylor dispersion analysis, protein, antibodies,

More information

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components: SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine

More information

ICP-OES DETERMINATION OF IRON. Introduction:

ICP-OES DETERMINATION OF IRON. Introduction: ICP-OES DETERMINATION OF IRON. Introduction: Silver impregnated cloth has bene used in bandages and dressings to help kill bacteria associated with a wound. Cloth samples are to anaysed for silver and

More information

TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110

TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110 TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110 BIOCHEMISTRY Trichlorethylene, marketed under the name of thinner, is used as a degreaser of metals, preparation of paints and cleaning of fabrics.

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2 Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2 P/N VALTOL-KT2 Revision 2.0 October 7, 2014 Prepared by: Quality Lab Accessories, LLC 100 Emlen Way, Suite 108 Telford,

More information

XR Analog Clock - Manual Setting Model Troubleshooting Guide

XR Analog Clock - Manual Setting Model Troubleshooting Guide Primex XR 72MHz Synchronized Time Solution XR Analog Clock - Manual Setting Model Troubleshooting Guide 2018 Primex. All Rights Reserved. The Primex logo is a registered trademark of Primex. All other

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

10 - Celestron Telescope II: Operation

10 - Celestron Telescope II: Operation 10 - Celestron Telescope II: Operation Purpose: Gain more experience setting up a 6 Celestron telescope, familiarize yourself with the software interface, and acquire an image with the CCD camera. Due:

More information

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219 REX Evaluation Guide American Micro Detection Systems Inc. 2800 March Lane, Suite 200 Stockton, CA 95219 I. INTRODUCTION REX (Real-time Elemental X-ray Fluorescence System) is the only instrument capable

More information

Technology that counts

Technology that counts Application Note No. 100 Version 1.3 Determination of Sperm Density in Ejaculates from Boars Application This Application Note describes how sperm cell density in ejaculates from boars can be determined

More information

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies MS n Analysis With Fast Polarity Switching in the Agilent Series LC/MSD Trap SL Application Note Christine Miller Introduction During mass spectrometric analysis, many compounds respond better to a particular

More information

Title: ASML Stepper Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 12/21/2010

Title: ASML Stepper Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 12/21/2010 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the ASML PAS 5500 Stepper. All users are expected to have read and understood this

More information

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion: Once approved, the new or revised document or the notice of deleted document(s) should be distributed to all

More information

Ideal Gas Law and Absolute Zero

Ideal Gas Law and Absolute Zero Experiment IX Ideal Gas Law and Absolute Zero I. Purpose The purpose of this lab is to examine the relationship between the pressure, volume and temperature of air in a closed chamber. To do this, you

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

Natural Rock Sample Manual

Natural Rock Sample Manual Natural Rock Sample Manual Revision 2014-06-12 DURRIDGE Company Inc. 524 Boston Road Billerica, MA 01821 Tel: (978) 667-9556 Fax: (978) 667-9557 service@durridge.com www.durridge.com 2014, DURRIDGE Company

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only.

Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only. Please use only the valid version of the package insert provided with the kit This kit is intended for Research Use Only. Not for use in diagnostic procedures. INTENDED USE The Vitamin Niacin test is a

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX

More information

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

Testing for particles in injectable products

Testing for particles in injectable products Testing for particles in injectable products 20-Nov-2018 MONITORING PHARMACEUTICALS Liquid sample technology that is compliant with USP must be part of any testing strategy into winning the war on

More information

Level. Automate your processes Automate your lab Innovate with. sensing. EasyPREP. sample handler. Automated Sample Preparation.

Level. Automate your processes Automate your lab Innovate with. sensing. EasyPREP. sample handler. Automated Sample Preparation. Automate your processes Automate your lab Innovate with EasyPREP sample handler Automated Sample Preparation Dispense up to 7 reagents Dilute Samples Aliquote Samples Transfer Samples Normalize Samples

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

cobas e 411 analyzer Confidence in every result.

cobas e 411 analyzer Confidence in every result. 17591_Brochure.qxd 11/1/06 10:46 AM Page 2 The cobas e 411 analyzer offers... Accurate Results ECL technology delivers excellent low-end sensitivity and broad dynamic ranges Quality Results Assay-specific

More information

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS METHADONE and in urine by GC-MS Code GC48010 Method of Confirmation by GC-MS INTRODUCTION is a synthetic opioid used as a drug of choice in the detoxification of heroin (and / or opiates). is metabolized

More information

[Unless otherwise stated all tolerances are ±5%] SAMPLES

[Unless otherwise stated all tolerances are ±5%] SAMPLES PHOSPHATASE TEST - FLUOROPHOS ALP TEST SYSTEM IMS #28 [Unless otherwise stated all tolerances are ±5%] SAMPLES 1. Laboratory Requirements (see Cultural Procedures [CP] items 33 & 34) [See current version

More information

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation...

More information

WeatherHub2 Quick Start Guide

WeatherHub2 Quick Start Guide WeatherHub2 Quick Start Guide Table of Contents 1 Introduction... 1 2 Packing List... 1 3 Connections... 1 4 IP Addressing... 2 5 Browser Access... 3 6 System Info... 3 7 Weather Station Settings... 4

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

HEAVY DUTY SWING-OUT VALVE

HEAVY DUTY SWING-OUT VALVE ISO 900 REGISTERED COMPANY Akron Brass Company. 20 All rights reserved. No portion of this can be reproduced without the express written consent of Akron Brass Company. We will not be responsible for:

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate oxidation allowing superior carbon liberation from even the most challenging

More information

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS Application Note Food Authors Nobuo Ochiai and Kikuo Sasamoto GERSTEL K.K. 2-13-18 Nakane

More information

Standard Operating Procedure PCM SOP 1-1. Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM)

Standard Operating Procedure PCM SOP 1-1. Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM) Standard Operating Procedure PCM SOP 1-1 Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM) Scope and Purpose: The scope and purpose of this standard operating

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information