MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide

Size: px
Start display at page:

Download "MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide"

Transcription

1 MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide /Revision A Copyright Waters Corporation All rights reserved.

2 Copyright notice 2005 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER. The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use. Waters Corporation 34 Maple Street Milford, MA USA Trademarks Micromass and Waters are registered trademarks of Waters Corporation. MassLynx is a trademark of Waters Corporation. Other trademarks or registered trademarks are the sole property of their respective owners. Customer comments Please contact us if you have questions, suggestions for improvements, or find errors in this document. Your comments will help us improve the quality, accuracy, and organization of our documentation. You can reach us at tech_comm@waters.com.

3 Table of Contents 1 Peak Response and Peak Ratio Absolute Response Compound response primary Compound response secondary Compound response both External Response Internal Response Primary/Secondary Peak Ratio Calibration Curve Calculations Weighted Calibration Curves Include Origin Average RF Linear Quadratic and Higher Order Curves Peak Amount Calculations User Specified Response Factor Average RF Calibration Curve Linear Calibration Curve Quadratic and Higher Order Calibration Curves User Parameters Calibration Curve Statistics Coefficient of Determination Curve Correlation Coefficient Table of Contents iii

4 RRF mean, SD, and %RSD for Average RF curves Totals Compounds Peak Response External Response Internal Response Calibration Curve Calculations Peak Amount Calculations Limits of Detection (LOD) and Limits of Quantitation (LOQ) Chromatogram Noise Calculation Manual noise measurement Automatic noise measurement Chromatogram Area Noise Response Value for Noise LOD and LOQ Concentrations LOD and LOQ Flags T-LOD and T-LOQ Concentrations Peak Statistics Peak area Peak height Height / Area Baseline width Width Peak sigma (Standard deviation) Peak skew (Asymmetry) Peak kurtosis (Peak top flatness) iv Table of Contents

5 Peak quality criteria Index... Index-1 Table of Contents v

6 vi Table of Contents

7 1 Peak Response and Peak Ratio There are two main methods of calculating peak response value: External Standard and Internal Standard. Both are based on the Absolute Response calculated for a peak. Contents: Topic Page Absolute Response 1-2 External Response 1-2 Internal Response 1-3 Primary/Secondary Peak Ratio

8 Absolute Response Absolute Response is based on either peak area or height and the selected combination of compound primary and secondary peaks. The examples below are shown for peak area. Compound response primary This is the default if no secondary trace is specified. Abs.Resp. = Primary Area Compound response secondary Abs.Resp. = Secondary Area Compound response both Abs.Resp. = Primary Area + Secondary Area Primary Area is the area of a peak calculated by peak detection on the primary chromatogram trace. Secondary Area is the area of a peak calculated by peak detection on the secondary chromatogram trace. External Response Peak Response = Abs.Resp Where Abs.Resp is the absolute response of a peak as calculated in Absolute Response on page Peak Response and Peak Ratio

9 Internal Response Peak Response Abs.Resp Amount = I Abs.Resp I Abs.Resp is the absolute response of a peak as calculated in Absolute Response on page 1-2. Amount I is the given amount of the Internal Standard in the sample. Abs.Resp I is the area of the internal standard peak as calculated in Absolute Response on page 1-2. Primary/Secondary Peak Ratio If primary and secondary traces are specified, the ratio of peaks detected on each trace is calculated as shown below. Ratio is based on either peak area or height: the example below is shown for peak area. Primary Peak Area Ratio = Secondary Peak Area Primary Peak Area is the area of a peak calculated by peak detection on the primary chromatogram trace. Secondary Peak Area is the area of a peak calculated by peak detection on the secondary chromatogram trace. Absolute Response 1-3

10 1-4 Peak Response and Peak Ratio

11 2 Calibration Curve Calculations MassLynx can fit several types of calibration curve. Contents: Topic Page Weighted Calibration Curves 2-2 Include Origin 2-2 Average RF 2-2 Linear 2-3 Quadratic and Higher Order Curves

12 Weighted Calibration Curves Calibration points used when fitting curves can be given a weighted importance: the larger the weighting the more significant a point is treated when the curve is fitted. Weighting (w i ) of i th calibration point is calculated using one of the following. All w i are set to 1 for no weighting w i w i w i 4. w i = x i y i is Y value (response) of i th calibration point. x i is X value (concentration) of i th calibration point. Include Origin If Include Origin is selected as a calibration curve type, an extra point with zero concentration and response is used in the regression. The extra point has the same weighting as the lowest calibration standard unless Force Origin is set, in which case the origin will always be included as a point directly on the calibration curve. Average RF = = = 1 y i 2 y i 1 x i 2 The calibration curve formed is linear passing through the origin with a gradient equal to the average response values of the calibration points. Average RF = Swy Sw y i x i Swy = w i 2-2 Calibration Curve Calculations

13 Sw = Σw i Linear y i is Y value (response) of i th calibration point. x i is X value (concentration) of i th calibration point. w i is weighting of i th calibration point, all set to 1 for no weighting. The calibration curve is formed by fitting a line using linear regression to a set of calibration points. Gradient = Swxy Swxx Intercept = y w,mean Gradient x w,mean Σy y i w i w,mean = Σw i Σx x i w i w,mean = Σw i Swxy = Σ( x i x w,mean ) ( y i y w,mean ) w i Swxx = Σ( x i x w,mean ) 2 w i y i is Y value (response) of i th calibration point x i is X value (concentration) of i th calibration point w i is weighting of i th calibration point, all set to 1 for no weighting. If Force Origin is selected, a line with zero intercept is fitted. Σx Gradient i y i w = i Σx i 2 w i Weighted Calibration Curves 2-3

14 Quadratic and Higher Order Curves MassLynx uses a general Least Squares Fit algorithm to regress a polynomial of any order against the calibration points. The method used is outlined below. Polynomial regression can be described as the fitting of m independent variables (Xj, j = 0 to m-1) to a single dependent variable y. In other words: y = Xb + e y is the n x 1 vector containing the n y values (y i ). X is the n x m matrix of x values, (x j i ). b is the m x 1 vector of regression coefficients (b i ). e is the n x 1 vector of residuals from the fit to each y i value. The familiar least squares solution for the regression coefficients is given by: b = ( X X) 1 X y -1 indicates matrix inverse ' indicates matrix transpose The above equation can then be solved using Gauss-Jordan elimination. To implement weighted regression X and y are first multiplied by a diagonal n x n matrix P (in other words, X becomes PX and Y becomes PY), before the above equation is solved. Where each element (p ij ) of P is given by: p ij = w 1/ 2 i for i =j p ij = 0 for i < > j w i is weighting of i th calibration point, all set to 1 for no weighting. 2-4 Calibration Curve Calculations

15 3 Peak Amount Calculations Contents: Topic Page User Specified Response Factor 3-2 Average RF Calibration Curve 3-2 Linear Calibration Curve 3-2 Quadratic and Higher Order Calibration Curves 3-3 User Parameters

16 User Specified Response Factor If a user response factor if selected within the quantitation method, calibration curves are not used. The following calculation is performed to obtain peak amounts: Amount Peak Response Response Factor Peak Response is the response value calculated for a peak. Response Factor is the user-entered response factor for that compound. Average RF Calibration Curve Amounts are calculated using an Average RF calibration as follows: Amount Peak Response is the response value calculated for a peak. Average RF is the average response factor calculated for a set of calibration points. Linear Calibration Curve = = Peak Response Average RF Amounts are calculated using a linear calibration as follows: Peak Response Intercept Amount = Gradient Peak Response is the response value calculated for a peak. Intercept is the intercept calculated for the linear calibration. Gradient is the gradient calculated for the linear calibration. 3-2 Peak Amount Calculations

17 Quadratic and Higher Order Calibration Curves Amounts are calculated by solving the following equation using the Newton-Raphson Method: Peak Response = P(Amount) Peak Response is the response value calculated for a peak. P( ) is the polynomial function calculated for a set of calibration points. User Parameters User parameters can be used to multiply or divide the final quantitation results. These factors are entered per sample in the Sample List. If a factor is not specified, or is zero, it is assumed to be one. Final Amount Amount User Factor 1 User Factor 2 User Factor 3 = User Divisor 1 The User Peak Factor is entered per compound in the Quantify Method. Final Amount = Amount User Peak Factor Rules: User multiplication and divisor factors are not applied to standard samples. In MassLynx 3 onwards, Sample List fields can be renamed to user requirements. When converting from a pre MassLynx V3.0 sample list the following field mappings occur: 'Initial Amount' becomes 'User Divisor 1' 'User Factor' becomes 'User Factor 1' 'Dilution Factor' becomes 'User Factor 2' 'Extract Volume' becomes 'User Factor 3' User Specified Response Factor 3-3

18 3-4 Peak Amount Calculations

19 4 Calibration Curve Statistics Contents: Topic Page Coefficient of Determination 4-2 Curve Correlation Coefficient 4-2 RRF mean, SD, and %RSD for Average RF curves

20 Coefficient of Determination The coefficient of determination is calculated for a regressed calibration curve. In the case of a linear curve it is equivalent to the square of the correlation coefficient and is reported as such. For each data point a value of y (y i,pred ) can be predicted from the calibration curve at the position x i. For each data point a residual between the actual and predicted y value can be calculated as (y i - y i,pred ), and the weighted residual sum of squares (RSS) can be calculated as: RSS = Σ( y i y i,pred ) 2 w i where w i is the weighting of the i th calibration point. The total variation in the data is reflected in the weighted corrected sum of squares (CSS), calculated as: CSS = Σ( y i y w,mean ) 2 w i where y w,mean is the weighted mean value of y Σy y i w i w,mean = Σw i The model sum of squares (MSS) is the portion of the total variation accounted for by the regression: MSS = CSS RSS The coefficient of determination (R 2 ) is the proportion of the variation accounted for by regression, and is given by the ratio of the model sum of squares to the corrected sum of squares: R 2 = MSS / CSS = (CSS RSS) / CSS Curve Correlation Coefficient In the case of linear curves, the square of the correlation coefficient (r) is equivalent to the coefficient of determination (see Coefficient of Determination on page 4-2) and is reported as such. 4-2 Calibration Curve Statistics

21 RRF mean, SD, and %RSD for Average RF curves The calculation of RRF mean, SD, and %RSD are detailed in the following equations: Mean RRF RRF i = N N = Abs.Response i Conc N i N ( RRF i Mean RRF ) 2 SD N RRF = N 1 SD RRF %RSD RRF = % Mean RRF Coefficient of Determination 4-3

22 4-4 Calibration Curve Statistics

23 5 Totals Compounds Totals compounds are formed from groups of peaks. These can include Named peaks that have been identified by other method compounds and Unnamed peaks which have not been identified. Compounds are identified as being a part of a Group by a shared group name. This section details calculations performed specifically for Totals compounds. Contents: Topic Page Peak Response 5-2 External Response 5-2 Internal Response 5-2 Calibration Curve Calculations 5-3 Peak Amount Calculations

24 Peak Response The responses of all peaks that are part of the Totals group are calculated individually. External Response Refer to External Response on page 1-2. Internal Response Calculated Internal Response of an unnamed peak is based upon the average response of the Internal Standards of the named compounds in the Group. Peak Response = Abs.Resp Ave.Named.Resp IS Abs.Resp = Absolute response of Unnamed peak as calculated in Absolute Response on page 1-2. Ave.Named.Resp IS = Average response of group Named peaks Internal Standards. = ( Σ N ( AbsResp ISn Amount ISn )) N N = Number of found Named compounds in group. n = Named compound index. AbsResp ISn = Absolute response of the n th Named Internal Standard peak as calculated in Absolute Response on page 1-2. Amount ISn = Given amount of the n th Named Internal Standard in the sample. 5-2 Totals Compounds

25 Calibration Curve Calculations The calibration curve for a Totals compound is formed by averaging the calibrations of the Named compounds in the group. The Totals curve is used to calculate the concentration of the Unnamed peaks in the group. Each coefficient of the Totals curve is calculated as follows: C i = ( Σ N C i,n ) N N is the number of Named compounds in the group. C i is the i th coefficient of the Totals curve. C i,n is the i th coefficient of the n th Named compound in the group. Peak Amount Calculations Totals concentration is the sum of the concentrations of the selected peaks that make up the group. Peaks summed can be specified as only Named peaks, only Unnamed peaks, or both. The concentration of each individual peak in the group is calculated as described in Peak Amount Calculations on page 3-1. Peak Response 5-3

26 5-4 Totals Compounds

27 6 Limits of Detection (LOD) and Limits of Quantitation (LOQ) The calculation of the LOD and LOQ for a compound is dependent upon the noise contained within the chromatogram trace used to quantify that compound. Contents: Topic Page Chromatogram Noise Calculation 6-2 Chromatogram Area Noise 6-3 Response Value for Noise 6-3 LOD and LOQ Concentrations 6-4 LOD and LOQ Flags 6-4 T-LOD and T-LOQ Concentrations

28 Chromatogram Noise Calculation Chromatogram noise is calculated as the standard deviation of the noise in the trace (effectively a height), which is multiplied by a user specified factor. Chromatogram noise may be calculated automatically or manually. Manual noise measurement To select this method, the user must specify an RT range of the trace which contains noise only. The resulting noise value is the standard deviation of the data in this region, multiplied by the User Factor. Tip: An RT entry of 0.0 may be used to signify either the start or end of the trace if input as the start or end RT respectively. If both the start and end RT are set to 0.0, automatic noise measurement (see page 6-2) will be initiated. Noise Height = SD selected data User Factor ( x i Mean x ) SD N x = N 1 Automatic noise measurement This method is initiated by selecting the entire range of the trace for noise measurement (that is, start and end RT equal 0.0). An algorithm is used to determine which regions of the trace consist only of noise, the result being the standard deviation of these regions multiplied by the user specified factor. Rule: The noise detection algorithm relies on a relationship between the median difference between adjacent points and standard deviation which is specific to gaussian deviates. This approximation provides good results when used with raw chromatogram data, but not for smoothed data. As a result, the algorithm measures the noise in the raw trace, to which a scaling factor is then applied to produce the effect of the current smoothing parameters. This smoothing correction is specific to the effect of mean smoothing on gaussian deviates, explained below. 6-2 Limits of Detection (LOD) and Limits of Quantitation (LOQ)

29 The accuracy of this method is affected: when Savitzky Golay smoothing is used to a lesser extent, when the noise regions are not gaussian in nature. SD meansmoothed This equation describes the relationship between the standard deviation of raw and mean smoothed data (for gaussian deviates). Iterations is the number of smooths applied to the raw data and Winsize is the size of the smoothing window. MassLynx requires a half window size to be entered which has the relationship with Winsize described above. A multiplication factor of 3 is commonly used as, for gaussian deviates, this corresponds to a range within which 95% of the data will lie. Chromatogram Area Noise The noise value is effectively a measure of noise height: an area value may be calculated that is equivalent to the calculated noise. The Compound s IS peak is used for this: the ratio of height to area of the corresponding IS peak is calculated, and this is applied to the noise value if the corresponding noise area is required. Response Value for Noise = SD raw data Winsize 0.5 Iterations 0.25 Winsize = ( 2 WinHalfSize) + 1 The absolute noise values are converted into equivalent response values. If compounds are being quantified using Heights, use the original noise values; if using Areas, use the Area noise values. Calculate an absolute noise response for the compounds based upon the method Secondary Parameters Compound response, Primary, Secondary or Both (in which case sum the noise values from both traces). If a Secondary trace is not specified the Primary is always used. Calculate a noise response value from the absolute noise response by applying the IS response, in the same way as if it were a normal peak. Chromatogram Noise Calculation 6-3

30 LOD and LOQ Concentrations The LOD and LOQ concentrations are calculated by applying the compound s calibration curve to the noise response to obtain a value for the concentration, which is then multiplied by user specified factors for LOD and LOQ respectively. The LOD and LOQ factors are specified as part of the QuanLynx or TargetLynx method and are set for all compounds. The default values are 3 and 8 respectively. Resulting LOD and LOQ concentrations should be stored in Peak Record and be available for display in the Summary. If no value is available output should be blank. LOD and LOQ Flags These flags indicate if the calculated concentration of a compound falls outside the LOD or LOQ concentration threshold. If a compound concentration exceeds the threshold the flag is Yes. If it is less than the threshold the flag is No. T-LOD and T-LOQ Concentrations The Toxic LOD concentration is calculated by multiplying the Toxic Equivalence Factor by the Limit Of Detection. The Toxic LOQ concentration is calculated by multiplying the Toxic Equivalence Factor by the Limit Of Quantification. Resulting T-LOD and T-LOQ concentrations should be stored in the Peak Record and be available for display in the Summary. If no value is available output should be blank. 6-4 Limits of Detection (LOD) and Limits of Quantitation (LOQ)

31 7 Peak Statistics Contents: Topic Page Peak height 7-2 Height / Area 7-2 Baseline width 7-3 Width 7-3 Peak sigma (Standard deviation) 7-3 Peak skew (Asymmetry) 7-4 Peak kurtosis (Peak top flatness) 7-4 Peak quality criteria

32 Peak area The peak area is calculated by summing the areas of each of the segments that make up the peak. Y2 Y3 Y4 Y1 X0 X1 X2 X3 X4 Xn Xn is the retention time (mins) at which sample n was taken. Yn is the intensity above the baseline for sample n. The area of each individual segment is calculated using the trapezium rule. ( Y Area i Y i ) i = ( X 2 i + 1 X i ) Total Area = ΣArea i n 1 The advantage of using this method is that the calculated peak area is less dependent on the sampling rate across the peak. Peak height The peak top is the retention time with the largest absolute intensity. Peak height is calculated as the distance between this largest intensity and the peak baseline at that retention time. Height / Area The peak height divided by the peak area. 7-2 Peak Statistics

33 Baseline width Width The difference, in seconds, between the end of the peak baseline and the start of it. Peak width measured in seconds at 50% of the peak height above the baseline. The value displayed is an approximation calculated as follows. The peak width at the inflection points is actually calculated this is known as the peak width at 4 σ (or 4 sigma). The majority of peaks are Gaussian in shape above the inflexion points, since the most variability is at the base of the peak. The peak width at half height is calculated from the equation: Width at half height = 2 2ln(2) σ = σ Width at inflection points = 4 σ Peak sigma (Standard deviation) This is the standard deviation of the data, calculated as the square root of the variance. The weighted variance is calculated as: -- 1 d w i ( x i x) 2 x i is the retention time value for observation i. w i is the intensity associated with x i. d = Σw i x is the weighted mean of x whose formula is: Σw i x i Σw i Peak area 7-3

34 The standard deviation is calculated as the square root of the variance: σ( x 1...x n ) = Var( x 1...x n ) The variance and standard deviation both require at least two non-missing observations. Peak skew (Asymmetry) The peak skewness characterizes the degree of asymmetry of a distribution around its mean (that is, a measurement of the tendency of the deviations to be larger in one direction than in the other). It is non-dimensional. It is a pure number that characterizes only the shape of the distribution, defined as follows: 1 = -- n ( ( x i x) σˆ j) 3 1 = -- n w 3 2 i (( x i x) σˆ ) 3 x is the weighted mean of x. x i is the retention time value for observation i. w i is the intensity associated with x i. σ is the distribution s weighted standard deviation defined above. A positive value of skew signifies a distribution with an asymmetric tail extending out towards more positive x; a negative value signifies a distribution whose tail extends out towards more negative x. Skewness requires at least three non-missing observations. Peak kurtosis (Peak top flatness) Peak kurtosis is a non-dimensional quantity. It measures the relative peakedness or flatness of a distribution, relative to a normal distribution. A distribution with positive kurtosis is termed leptokurtic; the outline of a very peaked mountain is an example. 7-4 Peak Statistics

35 A distribution with negative kurtosis is termed platykurtic; the outline is topped like a plateau. Distributions that are neither leptokurtic nor platykurtic are termed mesokurtic. Weighted kurtosis is computed as follows: 1 = -- n ( ( x i x) σˆ i) = -- n w 2 i (( x i x) σˆ ) 4 3 σ 2 i is σ 2 w i x is the weighted mean of x. x i is the retention time value for observation i. w i is the intensity associated with x i. σ is the distribution s weighted standard deviation defined above. The subtraction of 3 in the above expressions normalizes the kurtosis of a Gaussian distribution to zero. Kurtosis requires at least four non-missing observations. Peak quality criteria The peak statistics described above are measured against a selection of peak quality criteria (defined in the Method Editor) to generate Pass or Fail status in Peak Quality and a description of any failure in Peak Quality Description. Peak area 7-5

36 7-6 Peak Statistics

37 Index Symbols %RSD 4-3 A Absolute Response 1-2 asymmetry 7-4 Average RF 2-2, 4-3 calibration 3-2 B baseline width 7-3 C calibration curves 2-1 calculations 5-3 Chromatogram noise 6-2 coefficient of determination 4-2 concentration thresholds 6-4 corrected sum of squares 4-2 E External Response 1-2, 5-2 F Flags LOD and LOQ 6-4 G Gaussian 7-3, 7-5 Gauss-Jordan elimination 2-4 Gradient 3-2 H Higher Order curves 2-4, 3-3 I Include Origin 2-2 Intercept 3-2 Internal Response 1-3, 5-2 Internal Standards 5-2 K kurtosis 7-4 L Least Squares Fit 2-4 Limits of Detection 6-1 Limits of Quantitation 6-1 linear 2-2, 2-3 calibration 3-2 LOD 6-1 LOQ 6-1 M model sum of squares 4-2 N Named peaks 5-1 Newton-Raphson Method 3-3 Noise measurement 6-2 P peak area 7-2 height 7-2 kurtosis 7-4 named 5-1 response 1-1, 3-2, 5-2 skewness 7-4 top flatness 7-4 unnamed 5-1 Peak Amount Calculations 5-3 peak quality criteria 7-5 Primary Peak Area 1-3 Index-1

38 Q Quadratic curves 2-4, 3-3 quantitation method 3-2 results 3-3 R ratio primary and secondary peaks 1-3 regressed calibration curve 4-2 residual sum of squares 4-2 Response Factor 3-2 retention time 7-2 RRF mean 4-3 S Savitzky Golay 6-3 SD 4-3 Secondary Peak Area 1-3 standard deviation 7-3 sum of squares corrected 4-2 model 4-2 residual 4-2 T Totals compounds 5-1 U Unnamed peaks 5-1 User parameters 3-3 V variance weighted 7-3 W Weighted Calibration Curves 2-2 weighted variance 7-3 Index-2

Non-linear least squares

Non-linear least squares Non-linear least squares Concept of non-linear least squares We have extensively studied linear least squares or linear regression. We see that there is a unique regression line that can be determined

More information

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT [ application note ] T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT Keith Worrall, Anthony Newton and Ramesh Rao, Waters Corporation,

More information

-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics).

-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics). Chemometrics Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. -However, this definition can be expanded to include: biology (biometrics), environmental

More information

Descriptive Statistics C H A P T E R 5 P P

Descriptive Statistics C H A P T E R 5 P P Descriptive Statistics C H A P T E R 5 P P 1 1 0-130 Graphing data Frequency distributions Bar graphs Qualitative variable (categories) Bars don t touch Histograms Frequency polygons Quantitative variable

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Solving Systems of Linear Equations Using Matrices

Solving Systems of Linear Equations Using Matrices Solving Systems of Linear Equations Using Matrices What is a Matrix? A matrix is a compact grid or array of numbers. It can be created from a system of equations and used to solve the system of equations.

More information

DISPLAYING THE POISSON REGRESSION ANALYSIS

DISPLAYING THE POISSON REGRESSION ANALYSIS Chapter 17 Poisson Regression Chapter Table of Contents DISPLAYING THE POISSON REGRESSION ANALYSIS...264 ModelInformation...269 SummaryofFit...269 AnalysisofDeviance...269 TypeIII(Wald)Tests...269 MODIFYING

More information

3.3 Dividing Polynomials. Copyright Cengage Learning. All rights reserved.

3.3 Dividing Polynomials. Copyright Cengage Learning. All rights reserved. 3.3 Dividing Polynomials Copyright Cengage Learning. All rights reserved. Objectives Long Division of Polynomials Synthetic Division The Remainder and Factor Theorems 2 Dividing Polynomials In this section

More information

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition Understanding Your Spectra Module Agilent OpenLAB CDS ChemStation Edition Notices Agilent Technologies, Inc. 1994-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

Analyst Software. Peptide and Protein Quantitation Tutorial

Analyst Software. Peptide and Protein Quantitation Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

Prep for the CSU ELM

Prep for the CSU ELM Prep for the CSU ELM This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet curricular

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch and Method Setup Outliers, Data Review, Reporting 1 MassHunter Quantitative Analysis

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Prep for College Algebra

Prep for College Algebra Prep for College Algebra This course covers the topics outlined below. You can customize the scope and sequence of this course to meet your curricular needs. Curriculum (219 topics + 85 additional topics)

More information

Prep for College Algebra with Trigonometry

Prep for College Algebra with Trigonometry Prep for College Algebra with Trigonometry This course covers the topics outlined below. You can customize the scope and sequence of this course to meet your curricular needs. Curriculum (246 topics +

More information

Lecture 44. Better and successive approximations x2, x3,, xn to the root are obtained from

Lecture 44. Better and successive approximations x2, x3,, xn to the root are obtained from Lecture 44 Solution of Non-Linear Equations Regula-Falsi Method Method of iteration Newton - Raphson Method Muller s Method Graeffe s Root Squaring Method Newton -Raphson Method An approximation to the

More information

8/28/2017. PSY 5101: Advanced Statistics for Psychological and Behavioral Research 1

8/28/2017. PSY 5101: Advanced Statistics for Psychological and Behavioral Research 1 PSY 5101: Advanced Statistics for Psychological and Behavioral Research 1 Aspects or characteristics of data that we can describe are Central Tendency (or Middle) Dispersion (or Spread) ness Kurtosis Statistics

More information

Review of Multiple Regression

Review of Multiple Regression Ronald H. Heck 1 Let s begin with a little review of multiple regression this week. Linear models [e.g., correlation, t-tests, analysis of variance (ANOVA), multiple regression, path analysis, multivariate

More information

Using a graphic display calculator

Using a graphic display calculator 12 Using a graphic display calculator CHAPTER OBJECTIVES: This chapter shows you how to use your graphic display calculator (GDC) to solve the different types of problems that you will meet in your course.

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

PDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION

PDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION 1 2 3 4 5 INTRODUCTION G-20 CHROMATOGRAPHY 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 Chromatographic separation techniques are multi-stage separation methods in which the components

More information

Subject: Cycle Modification on the ABI 3900 DNA Synthesizer using software v1.2

Subject: Cycle Modification on the ABI 3900 DNA Synthesizer using software v1.2 Subject: on the ABI 3900 DNA Synthesizer using software v1.2 Overview: This bulletin describes how to modify the cycle files on the ABI 3900 DNA Synthesizer. In this User Bulletin: Topic Page Cycle File

More information

Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc.

Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc. Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc. greghudson@envirocompliance.com www.envirocompliance.com Abstract It should come as no surprise that the correlation coefficient is

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch Table, Compound Information Setup, Calibration Curve and Globals Settings 1 MassHunter

More information

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction Testing for GHB in hair by GC tandem quadrupole MS Marie Bresson, Vincent Cirimele, Pascal Kintz, Marion Villain; Laboratoire Chemtox, Illkirch, France Timothy Jenkins, Waters Corporation, Manchester,

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Statistics for Managers using Microsoft Excel 6 th Edition

Statistics for Managers using Microsoft Excel 6 th Edition Statistics for Managers using Microsoft Excel 6 th Edition Chapter 3 Numerical Descriptive Measures 3-1 Learning Objectives In this chapter, you learn: To describe the properties of central tendency, variation,

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

An Automated Application Template for Dissolution Studies

An Automated Application Template for Dissolution Studies An Automated Application Template for Dissolution Studies in SDMS Vision Publisher Wolfgang Lemmerz, Chris L. Stumpf, Thomas Schmidt, and Phil Kilby Waters Corporation, Milford, MA U.S.A. OVERVIEW Dissolution

More information

XGAUPROF. A GIPSY program to fit parameters of a Gauss function, the Gauss-Hermite series or a Voigt line-shape to data from GIPSY sets or ASCII files

XGAUPROF. A GIPSY program to fit parameters of a Gauss function, the Gauss-Hermite series or a Voigt line-shape to data from GIPSY sets or ASCII files XGAUPROF A GIPSY program to fit parameters of a Gauss function, the Gauss-Hermite series or a Voigt line-shape to data from GIPSY sets or ASCII files M.G.R. Vogelaar May 7, 1999, revision Mar 25, 2012

More information

High School Preparation for Algebra 1

High School Preparation for Algebra 1 High School Preparation for Algebra 1 This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence

More information

Introduction to Statistics

Introduction to Statistics Introduction to Statistics By A.V. Vedpuriswar October 2, 2016 Introduction The word Statistics is derived from the Italian word stato, which means state. Statista refers to a person involved with the

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

PERFORMANCE VERIFICATION IONSCAN LS

PERFORMANCE VERIFICATION IONSCAN LS Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--900 Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without

More information

Excel SOLVER for Fitting Bioanalytical Data to Weighted Polynomial Equations

Excel SOLVER for Fitting Bioanalytical Data to Weighted Polynomial Equations Solver Regression Page 1 of 14 SQA 2005 Poster (POSTER # 13) Excel SOLVER for Fitting Bioanalytical Data to Weighted Polynomial Equations Richard, Ph.D.; RQAP-GLP Biotechnical Services, Inc.; North Little

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

Chapter Four. Numerical Descriptive Techniques. Range, Standard Deviation, Variance, Coefficient of Variation

Chapter Four. Numerical Descriptive Techniques. Range, Standard Deviation, Variance, Coefficient of Variation Chapter Four Numerical Descriptive Techniques 4.1 Numerical Descriptive Techniques Measures of Central Location Mean, Median, Mode Measures of Variability Range, Standard Deviation, Variance, Coefficient

More information

RiskAMP Reference Manual

RiskAMP Reference Manual RiskAMP Reference Manual RiskAMP Version Structured Data, LLC. RiskAMP is a trademark of Strucuted Data, LLC. Excel is a registered trademark of Microsoft Corporation. Other trademarks are held by their

More information

Review of Statistics

Review of Statistics Review of Statistics Topics Descriptive Statistics Mean, Variance Probability Union event, joint event Random Variables Discrete and Continuous Distributions, Moments Two Random Variables Covariance and

More information

AES Features in CasaXPS

AES Features in CasaXPS Copyright 2004, Casa Software Ltd. All Rights Reserved. 1 of 6 AES Features in CasaXPS Converting AES Direct Spectra to Differential Form Spectra acquired in direct form must be processed into the differential

More information

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays : Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Enhancement of Linearity and Response in Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: This work

More information

Algebra 2. Curriculum (384 topics additional topics)

Algebra 2. Curriculum (384 topics additional topics) Algebra 2 This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet curricular needs.

More information

Demand Forecasting. for. Microsoft Dynamics 365 for Operations. User Guide. Release 7.1. April 2018

Demand Forecasting. for. Microsoft Dynamics 365 for Operations. User Guide. Release 7.1. April 2018 Demand Forecasting for Microsoft Dynamics 365 for Operations User Guide Release 7.1 April 2018 2018 Farsight Solutions Limited All Rights Reserved. Portions copyright Business Forecast Systems, Inc. This

More information

8-15. Stop by or call (630)

8-15. Stop by or call (630) To review the basics Matrices, what they represent, and how to find sum, scalar product, product, inverse, and determinant of matrices, watch the following set of YouTube videos. They are followed by several

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

Math Prep for College Physics

Math Prep for College Physics Math Prep for College Physics This course covers the topics outlined below. You can customize the scope and sequence of this course to meet your curricular needs. Curriculum (190 topics + 52 additional

More information

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT Introduction PCBs (polychlorinated biphenyls) are a class of man-made chemicals first manufactured commercially in the late 1920's and were used primarily as a dielectric fluid in electrical equipment.

More information

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,

More information

Preliminary Statistics course. Lecture 1: Descriptive Statistics

Preliminary Statistics course. Lecture 1: Descriptive Statistics Preliminary Statistics course Lecture 1: Descriptive Statistics Rory Macqueen (rm43@soas.ac.uk), September 2015 Organisational Sessions: 16-21 Sep. 10.00-13.00, V111 22-23 Sep. 15.00-18.00, V111 24 Sep.

More information

Introduction Linear system Nonlinear equation Interpolation

Introduction Linear system Nonlinear equation Interpolation Interpolation Interpolation is the process of estimating an intermediate value from a set of discrete or tabulated values. Suppose we have the following tabulated values: y y 0 y 1 y 2?? y 3 y 4 y 5 x

More information

Simplifying the Process: Automated USP 643 / EP

Simplifying the Process: Automated USP 643 / EP Simplifying the Process: Automated USP 643 / EP 2.2.44 Purified Water and Water For Injection Testing Using A Next Generation TOC Analyzer Application Note By: Brian Wallace and Stephen Lawson Background

More information

Practical Statistics for the Analytical Scientist Table of Contents

Practical Statistics for the Analytical Scientist Table of Contents Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning

More information

Final Review. Yang Feng. Yang Feng (Columbia University) Final Review 1 / 58

Final Review. Yang Feng.   Yang Feng (Columbia University) Final Review 1 / 58 Final Review Yang Feng http://www.stat.columbia.edu/~yangfeng Yang Feng (Columbia University) Final Review 1 / 58 Outline 1 Multiple Linear Regression (Estimation, Inference) 2 Special Topics for Multiple

More information

OBJECTIVE DATA PACKAGE DELIVERABLES

OBJECTIVE DATA PACKAGE DELIVERABLES PAGE: 1 of 10 OBJECTIVE This Standard Operating Procedure (SOP) describes the requirements for the analytical data packages that will be generated in association with the Stage 2 Delaware River Estuary

More information

Preparing Spatial Data

Preparing Spatial Data 13 CHAPTER 2 Preparing Spatial Data Assessing Your Spatial Data Needs 13 Assessing Your Attribute Data 13 Determining Your Spatial Data Requirements 14 Locating a Source of Spatial Data 14 Performing Common

More information

Elementary Statistics in Social Research Essentials Jack Levin James Alan Fox Third Edition

Elementary Statistics in Social Research Essentials Jack Levin James Alan Fox Third Edition Elementary Statistics in Social Research Essentials Jack Levin James Alan Fox Third Edition Pearson Education Limited Edinburgh Gate Harlow Essex CM20 2JE England and Associated Companies throughout the

More information

ncounter PlexSet Data Analysis Guidelines

ncounter PlexSet Data Analysis Guidelines ncounter PlexSet Data Analysis Guidelines NanoString Technologies, Inc. 530 airview Ave North Seattle, Washington 98109 USA Telephone: 206.378.6266 888.358.6266 E-mail: info@nanostring.com Molecules That

More information

1.0 Continuous Distributions. 5.0 Shapes of Distributions. 6.0 The Normal Curve. 7.0 Discrete Distributions. 8.0 Tolerances. 11.

1.0 Continuous Distributions. 5.0 Shapes of Distributions. 6.0 The Normal Curve. 7.0 Discrete Distributions. 8.0 Tolerances. 11. Chapter 4 Statistics 45 CHAPTER 4 BASIC QUALITY CONCEPTS 1.0 Continuous Distributions.0 Measures of Central Tendency 3.0 Measures of Spread or Dispersion 4.0 Histograms and Frequency Distributions 5.0

More information

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7B Triple Quadrupole GC/MS Application Note Authors Dale R. Walker and Fred Feyerherm Agilent Technologies, Inc.

More information

Developmental Algebra Beginning and Intermediate - Preparing for College Mathematics

Developmental Algebra Beginning and Intermediate - Preparing for College Mathematics Developmental Algebra Beginning and Intermediate - Preparing for College Mathematics By Paul Pierce Included in this preview: Copyright Page Table of Contents Excerpt of Chapter 1 For additional information

More information

Technical note on seasonal adjustment for M0

Technical note on seasonal adjustment for M0 Technical note on seasonal adjustment for M0 July 1, 2013 Contents 1 M0 2 2 Steps in the seasonal adjustment procedure 3 2.1 Pre-adjustment analysis............................... 3 2.2 Seasonal adjustment.................................

More information

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD) Application Note 19010803 A Newly Approved ASTM Standard For Analysis of in Using a Pulsed Flame Photometric Detector (PFPD) Keywords ASTM Standard D4735-96 FPD Flame Photometric Detector Model 5380 PFPD

More information

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Yun Wang Alelyunas, Mark D. Wrona, and Nick Tomczyk Waters Corporation, Milford, MA, USA GOAL

More information

Basic Math. Curriculum (358 topics additional topics)

Basic Math. Curriculum (358 topics additional topics) Basic Math This course covers the topics outlined below and is available for use with integrated, interactive ebooks. You can customize the scope and sequence of this course to meet your curricular needs.

More information

Y = ax + b. Numerical Applications Least-squares. Start with Self-test 10-1/459. Linear equation. Error function: E = D 2 = (Y - (ax+b)) 2

Y = ax + b. Numerical Applications Least-squares. Start with Self-test 10-1/459. Linear equation. Error function: E = D 2 = (Y - (ax+b)) 2 Ch.10 Numerical Applications 10-1 Least-squares Start with Self-test 10-1/459. Linear equation Y = ax + b Error function: E = D 2 = (Y - (ax+b)) 2 Regression Formula: Slope a = (N ΣXY - (ΣX)(ΣY)) / (N

More information

Calculation of the Current Transformer Accuracy Limit Factor. Application Note

Calculation of the Current Transformer Accuracy Limit Factor. Application Note Calculation of the Current Transformer Application Note 1MRS 755481 Issued: 09.11.2004 Version: Al09.11.2004 Calculation of the Current Transformer Application Note Contents: 1. Scope... 4 2. Introduction...

More information

National 5 Mathematics Assessment Practice Topic 4: The straight line, solving equations and inequations and simultaneous equations

National 5 Mathematics Assessment Practice Topic 4: The straight line, solving equations and inequations and simultaneous equations SCHOLAR Study Guide National 5 Mathematics Assessment Practice Topic 4: The straight line, solving equations and inequations and simultaneous equations Authored by: Margaret Ferguson Heriot-Watt University

More information

7.5 Operations with Matrices. Copyright Cengage Learning. All rights reserved.

7.5 Operations with Matrices. Copyright Cengage Learning. All rights reserved. 7.5 Operations with Matrices Copyright Cengage Learning. All rights reserved. What You Should Learn Decide whether two matrices are equal. Add and subtract matrices and multiply matrices by scalars. Multiply

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.

More information

Pre Algebra. Curriculum (634 topics)

Pre Algebra. Curriculum (634 topics) Pre Algebra This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet curricular needs.

More information

TNI V1M Standard Update Guidance on Detection and Quantitation

TNI V1M Standard Update Guidance on Detection and Quantitation TNI V1M4 2016 Standard Update Guidance on Detection and Quantitation GUID-3-109-Rev0 January 30, 2019 This material represents the opinion of its authors. It is intended solely as guidance and does not

More information

STATISTICAL ANALYSIS WITH MISSING DATA

STATISTICAL ANALYSIS WITH MISSING DATA STATISTICAL ANALYSIS WITH MISSING DATA SECOND EDITION Roderick J.A. Little & Donald B. Rubin WILEY SERIES IN PROBABILITY AND STATISTICS Statistical Analysis with Missing Data Second Edition WILEY SERIES

More information

GC Application Note. Dilution/standard addition: Efficient and traceable.

GC Application Note. Dilution/standard addition: Efficient and traceable. GC Application Note Dilution/standard addition: Efficient and traceable. www.palsystem.com Dilution / standard addition: Efficient and traceable. Philippe Mottay Brechbuehler AG Steinwiessenstrasse 3 CH-8952

More information

Copyright 2018 UC Regents and ALEKS Corporation. ALEKS is a registered trademark of ALEKS Corporation. 2/10

Copyright 2018 UC Regents and ALEKS Corporation. ALEKS is a registered trademark of ALEKS Corporation. 2/10 Prep for Calculus This course covers the topics outlined below. You can customize the scope and sequence of this course to meet your curricular needs. Curriculum (281 topics + 125 additional topics) Real

More information

BASICS OF ANALYTICAL CHEMISTRY AND CHEMICAL EQUILIBRIA

BASICS OF ANALYTICAL CHEMISTRY AND CHEMICAL EQUILIBRIA BASICS OF ANALYTICAL CHEMISTRY AND CHEMICAL EQUILIBRIA BASICS OF ANALYTICAL CHEMISTRY AND CHEMICAL EQUILIBRIA BRIAN M. TISSUE Virginia Tech Department of Chemistry Blacksburg, VA Cover Design: Wiley Cover

More information

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS 1. Given the HPLC chromatogram below and the fact that compounds A and B, eluting at 17.1 and 22.0 min, respectively, were present at concentrations

More information

Basics of Experimental Design. Review of Statistics. Basic Study. Experimental Design. When an Experiment is Not Possible. Studying Relations

Basics of Experimental Design. Review of Statistics. Basic Study. Experimental Design. When an Experiment is Not Possible. Studying Relations Basics of Experimental Design Review of Statistics And Experimental Design Scientists study relation between variables In the context of experiments these variables are called independent and dependent

More information

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION INTRODUCTION Full characterization of the physicochemical properties of new chemical entities (NCE s) typically takes place in pharmaceutical development laboratories. However, in an effort to minimize

More information

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Dominic Roberts, 1 Sara Stead, 1 and Jennifer A. Burgess 2 1 Waters Corporation, Manchester, UK 2 Waters Corporation,

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

Math Review for AP Calculus

Math Review for AP Calculus Math Review for AP Calculus This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet

More information

Math Prep for Statics

Math Prep for Statics Math Prep for Statics This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet curricular

More information

CfE Higher Mathematics Assessment Practice 4: Polynomials and quadratics

CfE Higher Mathematics Assessment Practice 4: Polynomials and quadratics SCHOLAR Study Guide CfE Higher Mathematics Assessment Practice 4: Polynomials and quadratics Authored by: Margaret Ferguson Reviewed by: Jillian Hornby Previously authored by: Jane S Paterson Dorothy A

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

Anomaly Density Estimation from Strip Transect Data: Pueblo of Isleta Example

Anomaly Density Estimation from Strip Transect Data: Pueblo of Isleta Example Anomaly Density Estimation from Strip Transect Data: Pueblo of Isleta Example Sean A. McKenna, Sandia National Laboratories Brent Pulsipher, Pacific Northwest National Laboratory May 5 Distribution Statement

More information

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC Online Reaction Monitoring of In- Manufacturing Samples by UPLC Tanya Tollifson Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Better throughput, yield, and process understanding are possible

More information

Scales of Measuement Dr. Sudip Chaudhuri

Scales of Measuement Dr. Sudip Chaudhuri Scales of Measuement Dr. Sudip Chaudhuri M. Sc., M. Tech., Ph.D., M. Ed. Assistant Professor, G.C.B.T. College, Habra, India, Honorary Researcher, Saha Institute of Nuclear Physics, Life Member, Indian

More information

PreCalculus. Curriculum (447 topics additional topics)

PreCalculus. Curriculum (447 topics additional topics) PreCalculus This course covers the topics shown below. Students navigate learning paths based on their level of readiness. Institutional users may customize the scope and sequence to meet curricular needs.

More information

Matrices. A matrix is a method of writing a set of numbers using rows and columns. Cells in a matrix can be referenced in the form.

Matrices. A matrix is a method of writing a set of numbers using rows and columns. Cells in a matrix can be referenced in the form. Matrices A matrix is a method of writing a set of numbers using rows and columns. 1 2 3 4 3 2 1 5 7 2 5 4 2 0 5 10 12 8 4 9 25 30 1 1 Reading Information from a Matrix Cells in a matrix can be referenced

More information

Written as per the revised G Scheme syllabus prescribed by the Maharashtra State Board of Technical Education (MSBTE) w.e.f. academic year

Written as per the revised G Scheme syllabus prescribed by the Maharashtra State Board of Technical Education (MSBTE) w.e.f. academic year Written as per the revised G Scheme syllabus prescribed by the Maharashtra State Board of Technical Education (MSBTE) w.e.f. academic year 2012-2013 Basic MATHEMATICS First Year Diploma Semester - I First

More information

DonnishJournals

DonnishJournals DonnishJournals 204-89 Donnish Journal of Educational Research and Reviews. Vol 2() pp. 008-07 January, 205. http:///djerr Copyright 205 Donnish Journals Original Research Paper Solution of Linear Systems

More information

[ instrument specifications ]

[ instrument specifications ] Designed for leading researchers working at the limits of conventional mass spectrometry capabilities who need to further characterize and define their samples the Waters SYNAPT High Definition MS (HDMS

More information

Pre Algebra and Introductory Algebra

Pre Algebra and Introductory Algebra Pre Algebra and Introductory Algebra This course covers the topics outlined below and is available for use with integrated, interactive ebooks. You can customize the scope and sequence of this course to

More information