Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Size: px
Start display at page:

Download "Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry"

Transcription

1 Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories. Users have to study the instrumental conditions appropriate for their own instrumentation.

2 CONTENTS PREMISE SCOPE ACRONYMS AND CONVENTIONS METHOD DESCRIPTION REAGENTS SAMPLE TREATMENT INSTRUMENTAL ANALYSIS Al calibration solutions Matrix modifier: Calibration Standards Sample Dilution Drift Check Spectrometer Settings CALCULATION PARAMETERS OF THE EURL CEFAO METHOD FINAL CONSIDERATIONS... 6 Annex 1: Instrumental Settings... 7 Page 2 of 7

3 PREMISE One of the major problem in Al determination is the control of the contamination. In general, there are two potential sources of external Al contamination in a laboratory: migration from the glassware and environmental contamination during sampling and analysis procedures. The use of glassware has to be minimized and all the equipment checked for this element before the use. 1. SCOPE The method describes the procedure for the direct determination of aluminium in milk by GF- AAS. 2. ACRONYMS AND CONVENTIONS GF-AAS Graphite Furnace Atomic Absorption Spectrometry sp Suprapure grade up Ultrapure grade d Density c Concentration expressed in units of milligrams/micrograms per litre RSD r Relative Standard Deviation under repeatability conditions HCL Hollow Cathode Lamp THGA Transversely Heated Graphite Atomizer Tubes LoD Limit of Quantification as defined in CR 333/2007 LoQ Limit of Quantification as defined in CR 333/ METHOD DESCRIPTION The direct determination of Al is performed by GF-AAS. Instrumental quantification is based on the matrix matched calibration approach. The sample is prepared according to the procedure described at point 5. The calibration curve is prepared as specified in section 6.3. WARNING: the application of the method includes the use of hazardous materials, operations and equipment. The user is responsible for the appropriate use of safety tools. APPARATUS and EQUIPMENT - GF-AAS (AAnalyst 800, Perkin Elmer) - Tube THGA - HCL Lamp (Al) 4. REAGENTS The reagents shall have the required adequate purity. Sp or up grade guarantees that the concentration of the analyte in reagents and water is negligible compared with the level of concentration to be determined. 4.1 Up-water, specific resistance 17 MΩ cm for all samples preparation and dilutions. 4.2 HNO 3 sp, not less than 65% mass fraction, d(hno 3 ) ~1,4 g/ml HNO 3 1% from HNO 3 (4.2) with up-water (4.1) HNO 3 10% from HNO 3 (4.2) with up-water (4.1). 4.3 Certified Element Standard Solutions, c = 1000 µg/ml. Page 3 of 7

4 4.3.1 Al 1 mg/l Pipette 0,1 ml of Al stock solution (4.3) and make up to 100 ml with up-water (4.1) in a certified decontaminated volumetric flask Prepare these standard solutions in triplicate, transfer to a plastic decontaminated storage bottle (e.g. Kartell), thoroughly mix. 4.4 Magnesium nitrate: Mg(NO 3 ) 2 6H 2 O (99.99 Sp) Magnesium nitrate (Mg(NO 3 ) 2 ) 6g/l Example: weigh 0.6g of Mg(NO 3 ) 2 (4.5) and make up to 100 ml with up-water (4.1) in a volumetric flask. 4.5 Triton X-100: TX: reagent for analysis Triton X-100 1% (TX1) in water Even if the reagent is liquid, it is easier to prepare the solution by weighting Triton X % (TX2) in water From TX1 (4.6.1) with up-water (4.1). 5. SAMPLE TREATMENT Samples are directly diluted 1:5 with a solution of Triton X 0.05% (4.5.2) in the cuvettes for analysis. 6. INSTRUMENTAL ANALYSIS a) please note: New tips were used and they were rinsed with HNO 3 1% (4.2.2), and up-water (4.1), before the use. Plastic wares (cuvettes, storage bottles, tubes, etc) shall be decontaminated filling them with a diluted HNO 3 water solution (e.g. 10% (4.2.3)), left stand overnight and rinsed at least 3 times with up-water (4.1) before their storage. Before the analysis, the cuvettes shall be rinsed with HNO 3 1% (4.2.2) and up-water. The micropipette used for mixing shall not be the same used for dilution, nor the one used for sample/standard preparation purposes (e.g. declassified micropipette used only for mixing). 6.1 Al calibration solutions These solutions shall be daily prepared c (Al) = 0,25 µg/l, into a 1ml cuvette for instrumental analysis (example): pipette 1 ml of Up-water (4.1) take off 250 µl add 250 µl of Al 1 mg/l (4.3.1) mix the solution by pipetting several times c (Al) = 0, 50 µg/l, into a 1ml cuvette for instrumental analysis (example): pipette 500 µl of Up-water (4.1) add 500 µl of Al 1 mg/l (4.3.1) mix the solution by pipetting several times 6.2 Matrix modifier: Mg(NO 3 ) 2 3 g/l Example: pipette 10 ml of Mg(NO 3 ) 2 6g/l (4.4.1) and make up to 20 ml with up-water (4.1). Page 4 of 7

5 6.3 Calibration Standards Standards for calibration curve and scheme of preparation: Al Standard Milk Al (6.1.1) 250 µg/l (µl) Al (6.1.2) 500 µg/l (µl) Triton X 0.05% (4.5.2) (µl) Blank Base Std 1 (2.5 µg/l) Std 2 (5 µg/l) Std 3 (10 µg/l) Std 4 (15 µg/l) Std 5 (20 µg/l) Sample Dilution Dilution 1:5: into a 1 ml cuvette for instrumental analysis (example): pipette 200µl of milk add 800 µl of triton X 0.05% (4.5.2) and mix thoroughly by pipetting (see 6.1 a). 6.5 Drift Check Drift is checked by analyzing the second and the fourth standard: at the end of the calibration curve (zero point) at the end of an analysis, if the number of samples analyzed is less than 8; after each series of 8 samples; at the end of an analysis, if more than 3 samples have been run after the last check. If the instrumental drift is higher than the pre-fixed value (e.g. 7%), the calibration curve shall be repeated and the previous samples read up to the last acceptable check. 6.6 Spectrometer Settings Optimize the instrument in accordance with the relevant manual and the internal procedures. Set ashing and atomization temperature (step 4 and 5) taking into account the instrument used and the ageing of the spare parts (tubes, contacts, furnace) Furnace programme set in the EURL-CEFAO method: Step T ( C) Ramp Time Hold Time (s) Internal Flow Gas Type (s) Normal Normal Normal Normal Normal Normal Compared to the instrument default method, a further drying step (450 C) has been added in the furnace programme. In this way, the process shows a negligible sample spattering during the ashing step resulting in a scarce production of residual combustion products and therefore in a better method repeatability. Samples and standards volume injected 20 µl; matrix modifier injected 10 µl. Page 5 of 7

6 Two autosampler tip extra-washes and two furnace extra clean-outs are performed. Autosampler tip washing solution HNO 3 1.0% (4.2.1). 7. CALCULATION Calculate the content, w, of the chemical element as mass fraction mg/kg as follows: w = D a V 1000 where: D = dilution of the test solution before the instrumental determination a = is the concentration of the element in the test solution in μg/l V = final volume of the solution (1ml). Considering the dilution, the density of the sample can be approximated to 1. Please note: the concentration of procedure blank is not considered in the final calculation (see 10). 8. PARAMETERS OF THE EURL CEFAO METHOD Commercially available skimmed milk was spiked with a standard solution as to have a sample of about 0.06 mg/kg. LoD and LoQ were determined using 20 aliquots of skimmed milk purchased at the market spiked with a solution containing a small amount (~ 2 µg/l) of the analyte. The parameters obtained during the method validation are summarized as follows: Parameter Al Valitation level (mg/kg) 0.06 s r RSD r (%) 1.6 r (mg/kg) LoD(=3xSD) (mg/kg) LoQ(=10xSD) (mg/kg) FINAL CONSIDERATIONS Care must be taken in the decontaminations of the lab wares. The procedure applied allowed to obtain negligible blank values during the instrumental runs; for this reason these values are not included in the calculation. Page 6 of 7

7 Annex 1: Instrumental Settings The annex is the instrumental report in paper form Page 7 of 7

8 Method Name: Al Milk DD 15 Define Element Method Description : Al Spectrometer Element : Al Wavelength (nm) : Slit Width (nm) : 0.7L Signal Type : AA - BG Measurement : Peak Area Settings Read Parameters Time (sec) : 3.0 Delay Time (sec) : 0.0 BOC Time (sec) : 2 Replicates... Same for All Samples : 2 Lamp Current Use value entered in Lamp Setup window Furnace Program Step Temp( C) Ramp Time Hold Time Internal Flow Gas Type Normal Normal Normal Normal Normal Normal Read Step : 5 Injection Temperature( C) : 20 Extra Furnace Cleanout Steps Run furnace step 6, 2 times: Furnace Autosampler Sample Volume : 20 ul Diluent Volume : 0 ul Diluent Location : 1 Matrix Modifiers #1 Volume : 10 ul Location : 20 Add to calibration blank and standards. Add to reagent blank and samples. #2 Volume : 0 ul Location : 1 Add to calibration blank and standards. Add to reagent blank and samples. Autosampler and Furnace Sequence Step Actions and Parameters A Pipet modifier1 + sample/std Page 1 06/12/ :32:37

9 Method Name: Al Milk DD 15 B C Run furnace steps 1 to end Wash sampler 2 extra times Pipet Speed... Take up : 100% Dispensing : 100% Calibration Equation and Units Equation : Method of Additions Calibrate Maximum Decimal Places : 3 Maximum Significant Figures : 4 Analyst-prepared additions. Calibration Units : ug/l Sample Units : ug/l Calibration Standard Concentrations ID Concentration A/S Loc Stock(uL) Diluent(uL) Calib Blank Bianco 1 20 Addition 1 Addition Addition 2 Addition Addition 3 Addition Addition 4 Addition Addition 5 Addition Initial Calibration Options When opening this method manually: Clear calibration curve(s) and construct a new calib curve. When using this method in a multimethod sequence: Start by constructing new calibration curves. Calibration Check Correlation coefficient checking is not enabled in this method. Recalibration Periodic recalibration is not enabled in this method. Analyze standards at end of analysis is not enabled in this method. Precision Checks Precision checking is not enabled in this method. Beyond Calibration Range Beyond calibration range checking is not enabled in this method. Page 2 06/12/ :32:37

10 Method Name: Al Milk DD 15 Matrix Recovery Matrix recovery checking is not enabled in this method. Automatic Recovery Automatic recovery checking is not enabled in this method. QC Sample Definition A/S Count as Subtract QC Sample ID Loc Sample Reagent Blank QC Sample Concentrations and Limits Schedule for QC Analyses After After QC Sample ID Init Cal Recal Periodic Freq At End Periodic Timing of Analyses Frequency... Same for all QC's : 1 Count : Samples Failure Actions for After-Calibration QC's Times to When All Additional QC Sample ID Retry QC Tries Fail Message Failure Actions for Periodic QC's Times to When All Additional QC Sample ID Retry QC Tries Fail Message Failure Actions for At-End QC's Times to When All Additional QC Sample ID Retry QC Tries Fail Message Maximum Retries After QC Failure After a group of standards or unknowns has been reanalyzed 1 times, then Continue Options Include in Results Display and Printed Log: Page 3 06/12/ :32:37

11 Method Name: Al Milk DD 15 Headers: * Analytical Header * Method Header (Short) * Sample Header (Short) Start each sample on a new page Sample Data Items: Summary Items: * Replicate Data Analysis List * Means and Statistics * Matrix Test Reports Transient Peak Plots (Last) * Calibration Summary Calibration Curves Save with Results: * Transient Peak Profiles Remarks: MM: Mg(NO3)2 3 g/l Sample diluted in Triton X 0,05% Page 4 06/12/ :32:37

Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction

Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction NOTICE: This method: Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction - has to be considered only as guideline

More information

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation. Determination of Cadmium and Lead in powdered infant formula by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction NOTICE: This method: - has to be considered only

More information

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS 1. SCOPE AND FIELD OF APPLICATION The method describes the determination of heavy metals (arsenic, cadmium, lead and mercury) in cosmetic products. 2. PRINCIPLE Organic matter in sample is digested by

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Quality Control Procedures for Graphite Furnace AAS using Avanta Software Application Note Quality Control Procedures for Graphite Furnace AAS using Avanta Software Introduction With the arrival of quality standards in the analytical laboratory, an ever increasing emphasis is

More information

DRAFT EAST AFRICAN STANDARD

DRAFT EAST AFRICAN STANDARD ICS 71.100.70 DRAFT EAST AFRICAN STANDARD Oils for cosmetic industry Methods of test Part 16 : Determination of Heavy metal Content EAST AFRICAN COMMUNITY EAS 2015 First Edition 2015 Copyright notice

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Brooklyn College Department of Chemistry

Brooklyn College Department of Chemistry Brooklyn College Department of Chemistry Instrumental Analysis (Chem 42/790) Atomic Absorption Spectroscopy An atomic absorption spectrometer is used in this experiment to analyze a copper-base alloy for

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS Determination of Total Bromine and Iodine Page 2 of 5 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes an inductively coupled plasma optical emission spectrometry method

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA GB 5009.15-2014 National Food Safety Standard Determination

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

A Spectrophotometric Analysis of Calcium in Cereal

A Spectrophotometric Analysis of Calcium in Cereal CHEM 311L Quantitative Analysis Laboratory Revision 1.2 A Spectrophotometric Analysis of Calcium in Cereal In this laboratory exercise, we will determine the amount of Calium in a serving of cereal. We

More information

Example method validation and MU approach for Cadmium in seafood test method

Example method validation and MU approach for Cadmium in seafood test method Example method validation and MU approach for Cadmium in seafood test method Charun Yafa National Institute of Metrology (Thailand) (NIMT) 8th APMP TCQM-DEC MiC Workshop, Kobe, Japan December 011 QA Measures

More information

Cadmium and lead content

Cadmium and lead content Accepted 1997 Pulps, papers and boards Cadmium and lead content Atomic absption spectrometry - graphite furnace technique 1 Scope This SCAN-test Method specifies a method f the determination of traces

More information

National Research Council Institute for Ecosystem Study Verbania Pallanza - Italy

National Research Council Institute for Ecosystem Study Verbania Pallanza - Italy Tot-N UV220 pag. 1 National Research Council Institute for Ecosystem Study Verbania Pallanza - Italy Water Chemistry Laboratory Analytical Methods for internal use - http://www.idrolab.ise.cnr.it Gabriele

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples A P P L I C A T I O N N ot e Atomic Absorption Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples Introduction The Flow Injection Mercury System (FIMS) is a dedicated system

More information

UNITED NATIONS. ENVIRONMENT PROGRAMME MEDITERRANEAN ACTION PLAN 25 May 2015 Original: English. UNEP(DEPI)/ MED WG.417/Inf.14

UNITED NATIONS. ENVIRONMENT PROGRAMME MEDITERRANEAN ACTION PLAN 25 May 2015 Original: English. UNEP(DEPI)/ MED WG.417/Inf.14 UNITED NATIONS UNEP(DEPI)/ MED WG.417/Inf.14 UNITED NATIONS ENVIRONMENT PROGRAMME MEDITERRANEAN ACTION PLAN 25 May 2015 Original: English MED POL Focal Points Meeting Malta, 16-19 June 2015 Joint Session

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct 1. Application Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct Parameter and Code: Mercury, total, I-6463-86 (µg/g as Hg): none assigned 1.1 This method is used to analyze

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer Maintenance with OQ Operational Qualification OQ within the scope of a maintenance for the Atomic Absorption Spectrometer ZEEnit 700 Contents 1. GENERAL INFORMATION...4 2. DEFINITIONS...4 2.1 Validation...4

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

[Unless otherwise stated all tolerances are ±5%] SAMPLES

[Unless otherwise stated all tolerances are ±5%] SAMPLES PHOSPHATASE TEST - FLUOROPHOS ALP TEST SYSTEM IMS #28 [Unless otherwise stated all tolerances are ±5%] SAMPLES 1. Laboratory Requirements (see Cultural Procedures [CP] items 33 & 34) [See current version

More information

CHEMISTRY DEPARTMENT

CHEMISTRY DEPARTMENT POINT LEPREAU GENERATING STATION GADOLINIUM ARSENAZO III Issued By: Date: APPROVED BY: of 6 REVISION RECORD AUTHOR REV DESCRIPTION DATE 0 Initial Issue W. Mawhinney Not Tracked Oct./9 W. T. Underhill 2

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Application field / Industry branch:

Application field / Industry branch: Application field / Industry branch: Chemistry / Polymer Industry Electronics Energy Nutrition / Agriculture Geology / Mining Semiconductor Technology Clinical Chemistry / Medicine / Sanitation / Health

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of

More information

Laboratory Measurements and Procedures

Laboratory Measurements and Procedures 18 Introduction Measurements of masses, volumes, and preparation of chemical solutions of known composition are essential laboratory skills. The goal of this exercise is to gain familiarity with these

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

The analysis of organic acid content of additives, premix, feed, and water.

The analysis of organic acid content of additives, premix, feed, and water. The analysis of organic acid content of additives, premix, feed, and water. Contents Foreword Introduction Warnings 1. Scope 2 1.1 LOD and LOQ 3 2. Normative References 3 3. Definitions 3 3.1 Feed (or

More information

Method for Nitrite determination on Low-Range Samples

Method for Nitrite determination on Low-Range Samples Method for Nitrite determination on Low-Range Samples 1. PRINCIPLE This method is designed for the determination of nitrite in soil, water and other forms of samples. The sample is prepared offline if

More information

POLYDEXTROSES. SYNONYMS Modified polydextroses; INS No DEFINITION DESCRIPTION. FUNCTIONAL USES Bulking agent, humectant, stabilizer, thickener

POLYDEXTROSES. SYNONYMS Modified polydextroses; INS No DEFINITION DESCRIPTION. FUNCTIONAL USES Bulking agent, humectant, stabilizer, thickener POLYDEXTROSES Prepared at the 51st JECFA (1998) and published in FNP 52 Add 6 (1998) superseding specifications prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995). An ADI "not specified"

More information

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Introduction: Capillary electrophoresis (CE) is a relatively new, but rapidly growing separation

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Chemistry Calibration of a Pipet and Acid Titration

Chemistry Calibration of a Pipet and Acid Titration Chemistry 3200 Today you are given a chance to brush up on some of the techniques that you will be using during the remainder of the semester. Lab grades will be based on obtaining the correct answer in

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

GB Translated English of Chinese Standard: GB

GB Translated English of Chinese Standard: GB Translated English of Chinese Standard: GB31604.34-2016 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA GB 31604.34-2016 National Food Safety Standard

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

Spectrophotometric Determination of Ferrocyanide in Effluents

Spectrophotometric Determination of Ferrocyanide in Effluents Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

METHANOL REFERENCE SPECIFICATIONS

METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION Avenue de Tervuren 149 1150 Brussels, Belgium Tel : (32/2) 7418683 Fax : (32/2) 7418684 e-mail : impca@epca.be IMPCA METHANOL REFERENCE SPECIFICATIONS

More information

By Donna L. Damrau U.S. GEOLOGICAL SURVEY. Open-File Report Denver, Colorado

By Donna L. Damrau U.S. GEOLOGICAL SURVEY. Open-File Report Denver, Colorado METHODS OF ANALYSIS BY THE U.S. GEOLOGICAL SURVEY NATIONAL WATER QUALITY LABORATORY-- DETERMINATION OF LOW-LEVEL SILVER BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROPHOTOMETRY By Donna L. Damrau U.S. GEOLOGICAL

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7 Troy, NY 12180 Date 01/25/18 Page 1 of 7 USGS Water Science Center Laboratory, Troy, NY Ion Chromatography Standard Operating Procedure 1. Scope and Application 1.1 Analytes Chloride, Nitrate, and Sulfate

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Use objectives: state purpose for which results will be used. Specifications: How good should the numbers be

More information

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow 1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II) : Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption

More information

Corn Syrup Analysis E-51-1 PHOSPHORUS

Corn Syrup Analysis E-51-1 PHOSPHORUS Corn Syrup Analysis E-51-1 PRINCIPLE SCOPE The sample is ignited in the presence of a fixative to destroy organic matter and convert phosphorus to inorganic phosphates which are not volatilized during

More information

On the use of Deming and quadratic calibrations in analytical chemistry

On the use of Deming and quadratic calibrations in analytical chemistry On the use of Deming and quadratic calibrations in analytical chemistry Henry Foden Rand Water October 09-2013 PO Box 3526, Vereeniging, 1930, Republic of South Africa e-mail: hfoden@randwater.co.za Phone:

More information

The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40

The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40 The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40 Application Note Atomic Absorption Author Keith Brodie The widespread use of metals in modern industry has meant that traces

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC) Page 1 of 15 Standard Operating Procedures Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC) Analytical Sciences Department Innovation, Technology, and Development

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Calculator 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Chloride by Flow Injection Analysis (FIA)

Chloride by Flow Injection Analysis (FIA) Methodology Chloride by Flow Injection Analysis (FIA) (Cartridge Part #A001568) 1.0 Scope and Application 1.1 This method is used for the determination of chloride in drinking water, surface water, saline

More information

Hexavalent Chromium by Flow Injection Analysis (FIA)

Hexavalent Chromium by Flow Injection Analysis (FIA) Methodology Hexavalent Chromium by Flow Injection Analysis (FIA) (Cartridge Part #A002831) 1.0 Scope and Application 1.1 This method is used for the determination of hexavalent chromium in water, including

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ ) DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Chemistry 102 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

INFLUENCE OF THE MEASUREMENT MEDIUM AND MATRIX MODIFIERS ON THE DETERMINATION OF SILICON IN WATERS BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY

INFLUENCE OF THE MEASUREMENT MEDIUM AND MATRIX MODIFIERS ON THE DETERMINATION OF SILICON IN WATERS BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY oujda University Morocco From the SelectedWorks of elhimri2@hotmail.com Spring April 1, 2013 INFLUENCE OF THE MEASUREMENT MEDIUM AND MATRIX MODIFIERS ON THE DETERMINATION OF SILICON IN WATERS BY GRAPHITE

More information

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer Application Note - AN1502 Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer Page 1 Jeff Forsberg, Product Manager, John McQuatters, Mercury Product Specialist,

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

DR/4000 PROCEDURE. HARDNESS, Total

DR/4000 PROCEDURE. HARDNESS, Total Method 8374 DR/4000 PROCEDURE Calcium and Magnesium; Chlorophosphonazo Colorimetric Method ULR (0 to 1,000 µg/l Ca & Mg as CaCO 3 ) Scope and Application: For ultra pure water. The estimated detection

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

SOUTH AFRICAN NATIONAL STANDARD. Standard test methods for lead in water

SOUTH AFRICAN NATIONAL STANDARD. Standard test methods for lead in water ISBN 978-0-626-19882-4 Edition 1 SOUTH AFRICAN NATIONAL STANDARD Standard test methods for lead in water This national standard is the identical implementation of ASTM D 559:200 and is adopted with the

More information

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method Rating Guide Appendix for the CCME Reference Method for the Canada-Wide Standard for Petroleum (PHC) in Soil - Tier 1 Method (Note: Checklist incorporates requirements from Dec 2000 version + Addendum

More information

Automated determination of the uniformity of dosage in quinine sulfate tablets using a fiber optics autosampler

Automated determination of the uniformity of dosage in quinine sulfate tablets using a fiber optics autosampler Automated determination of the uniformity of dosage in quinine sulfate tablets using a fiber optics autosampler Application Note Author John Sanders. Agilent Technologies, Inc. Mulgrave, Victoria 3170,

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2. Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES 701.0 Version 2.3 July 2018 1. SCOPE AND APPLICATION This method is used for the determination

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS 1.0 Purpose - This procedure specifies the required elements for the extraction and quantitation of THC,

More information

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017. Final Report Title Characterisation of Job No 2016/11/12-34 AS No. 1234A Client Contact Sample Author report Signed Date 2017 Easy Reach Report 2017 v2.docx 1 of 33 Contents 1. Study Summary Page 3 2.

More information

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis Tobacco and Tobacco Products Analytes Sub-Group Technical Report Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis 2016 Collaborative

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

California Environmental Protection Agency Air Resources Board

California Environmental Protection Agency Air Resources Board California Environmental Protection Agency Air Resources Board PROCEDURE FOR THE ANALYSIS OF PARTICULATE ANIONS AND CATIONS IN MOTOR VEHICLE EXHAUST BY ION CHROMATOGRAPHY Standard Operating Procedure No.

More information

METHOD 9022 TOTAL ORGANIC HALIDES (TOX) BY NEUTRON ACTIVATION ANALYSIS

METHOD 9022 TOTAL ORGANIC HALIDES (TOX) BY NEUTRON ACTIVATION ANALYSIS METHOD 9022 TOTAL ORGANIC HALIDES (TOX) BY NEUTRON ACTIVATION ANALYSIS 1.0 SCOPE AND APPLICATION 1.1 Method 9022 determines Total Organic Halides (TOX) in aqueous samples. The method uses a carbon adsorption

More information

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are

More information