The Theory of HPLC. Quantitative and Qualitative HPLC

Size: px
Start display at page:

Download "The Theory of HPLC. Quantitative and Qualitative HPLC"

Transcription

1 The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

2 Aims and Objectives Aims and Objectives Aims To define Quantitative HPLC and explain the information that can be derived from this type of HPLC analysis To examine the use of Peak Height or Area in quantitative calculations and to investigate Integration of chromatographic peaks To explain the use of Calibration and Calibration curves in Quantitative HPLC To outline the principles of Single and Multi-level calibration in Quantitative analysis To investigate practical uses of External Standard, Internal Standard and Normalisation methods of Quantitative analysis To outline example calculations for the various Quantitative methods Objectives At the end of this Section you should be able to: To define Qualitative HPLC and explain the information that can be derived from this type of HPLC analysis To define and explain the principles of peak identification and analyte characterisation from a practical perspective

3 Content Qualitative Analysis Overview 3 Peak Identification and Assignment 4 Sample Spiking 5 Spectral Peak Identification 6 Peak Purity 7 Spectral Characterisation 7 Quantitative Analysis Overview 8 Chromatographic Requirements 9 Peak Integration 10 Integration Events 11 Peak Height or Peak Area 11 Principles of Quantitative Analysis 12 Area %/ Height % (Normalisation) 13 External Standard Quantitation 14 Calibration Curve 16 Statistical Information 17 External Standard Multi-level Calibration 18 Calibration Curve Information 19 Internal Standard Analysis 21 Crawford Scientific 2

4 Qualitative Analysis Overview The aim of qualitative analysis is to answer the question What is in the sample? Options for Qualitative HPLC Analysis Two discrete situations exist for qualitative analysis: the sample components are known and peaks within the chromatogram need to be assigned to the known components the sample is a complete unknown which you are attempting to characterise In the former case, it might be possible to inject standards of the pure compound, and assign the peaks in the chromatogram based on the retention time of the standard. Having a selective detector, such as diode-array UV or Fluorescence detector, that assists in identification by producing spectra or a specific response can assist in peak assignment. In the second case, it may be necessary to employ detectors that can be used to aid in identification, such as mass spectrometers. It may also be necessary to collect the eluent fraction containing the peak of interest for off-line characterisation using Infra-red or Nuclear Magnetic Resonance Spectroscopy (NMR). Crawford Scientific 3

5 Peak Identification and Assignment Perhaps the most straightforward way to assign peaks within the chromatogram of a sample solution is to inject standard solutions under identical analytical conditions. By comparing the retention factor (k) and response of the peak in the chromatogram of the standard solution, with the sample chromatogram, peaks may be tentatively assigned. It is important that the concentration of the standard solution is matched to the sample solution as closely as possible. This avoids peak mis-assignment due to peak shape effects. Using reference standards for Peak Identification For more rigorous peak assignment it is important that the analysis is also carried out under orthogonal (different) conditions. The analysis is carried out in the reversed phase using a different column and mobile phase, and the retention factor and response compared again. If a match is found, then the confidence in the peak identification is increased. Crawford Scientific 4

6 Sample Spiking The technique of spiking a sample involves the addition of a known reference material to a sample matrix, in order to confirm the identity of one of the sample component peaks. Use of sample spiking to increase confidence in Peak Identification In this case, one of the peaks in the sample is suspected to be insulin. The sample is spiked with insulin at approximately the same concentration as the sample components. If any of the peaks within the chromatogram gets larger, then that peak may be insulin. If a new chromatographic peak is seen, or if any of the peaks develops a shoulder, then it is unlikely that any of the peaks in the chromatogram is due to insulin within the sample. Crawford Scientific 5

7 Spectral Peak Identification The identification and assignment of peaks within sample chromatograms using retention time alone can be unreliable. The use of selective detectors and spectrometers can greatly increase the confidence in the peak assignment. Detector systems such as Diode Array UV Spectrometers or Mass Spectrometers are able to record unique spectra for each peak within the sample chromatogram. The spectra may be recorded in real time as the eluent can be directly introduced into the detector system. The use of Spectral Detectors to assist with Peak Identification It is usual that the spectrum obtained from the unknown peak is compared with a spectrum from either a standard solution of pure reference standard run alongside the sample, or with spectra from a library held within the chromatographic data system. There are many ways to compare spectra, and many algorithms exist for producing a Match between the recorded spectra and the sample spectrum. These statistical methods of comparison give a quantitative measure of confidence in the peak assignment. Mass spectra and higher order (derivative) UV spectra can also be used to check for peak purity' whether the peak is due to a single component or is an amalgamation of two or more peaks. Crawford Scientific 6

8 Peak Purity Peak purity can be established by taking the ratio of two signals (wavelengths) across the peak(s) of interest. If the peak is pure, then the ratio of the two signals should be constant across the peak. If the peak is impure, then the ratio between the two signals (wavelengths) will change across the peak as the spectral differences caused by the interfering peak change the signal ratio. Spectral Characterisation Mass Spectrometric detectors can be configured to produce fragmentation patterns that can be assigned to analyte moieties, so building up a picture of the analyte molecule. The spectral peaks and patterns combined with the molecular (pseudomolecular) weight acquired can be used to characterise the analyte molecule. Determination of Peak Purity using Spectral Detectors Crawford Scientific 7

9 Quantitative Analysis Overview After the peaks have been integrated and identified, the next step in the analysis is quantification. Quantification uses peak areas or heights to determine the concentration of a compound in the sample. Quantitative Analysis in HPLC A quantitative analysis involves many steps that are briefly summarised as follows: Know the compound you are analysing Establish a method for analysing samples containing this compound Analyse a sample or samples (the Standard) containing a known concentration or concentrations of the compound to obtain the response due to that concentration (called Calibration ) You may analyse a number of these samples with different concentrations of the compounds of interest if your detector has a non-linear response (referred to as multi-level calibration ), or if a wide concentration range is to be measured in the samples Analyse the sample containing an unknown concentration of the compound to obtain the response due to the unknown concentration Compare the response of the unknown concentration to the response of the known (standard) concentration to determine how much of the compound is present To obtain a valid comparison for the unknown sample response to that of the known standard, the data must be acquired and processed under identical conditions. Crawford Scientific 8

10 Chromatographic Requirements A requirement for quantitative analysis is that the peaks to be quantified are known (i.e. characterised), so that standard solutions may be prepared for instrument response calibration. There are notable exceptions to these requirements, for example, impurity and degradation product analysis in Pharmaceutical Stability Studies. In this case, surrogate standards must be used which are representative of the general compound class more on this later. The accuracy of the quantitation is strongly influenced by the resolution of the peaks and the state of the baseline surrounding the peaks of interest. Well separated peaks can be easily and reproducibly integrated because other peaks do not influence the height and area. Peaks eluting on sloping or noisy baselines can be difficult to integrate reproducibly. Peak symmetry is also a performance characteristic for a good quantitative result. Peaks with symmetry values exceeding three cannot be properly or reproducibly integrated. Peak shoulders or tailing / fronting may be a result of two improperly separated compounds, and shouldered peaks are difficult to integrate reproducibly. Improve chromatography prior to quantitative analysis if possible. Chromatographic Requirements for Robust Quantitative Analysis in HPLC Crawford Scientific 9

11 Peak Integration Chromatographic peak integration defines an operation in which the area under the chromatographic peak is measured. The measurement is based on the integral technique of splitting the peak into a large number of rectangles, which are then summed to provide an estimate of the total area under the peak a measure of the cumulative absorbance versus time signal. Two events need to be defined in order for the data system to carry out the calculation, these are the peak start and the peak end these will be determined using threshold and peak width settings in the data system. The method of determination has to be reproducible for rugged integration. The baseline is then drawn between the peak start and end points created by the data system. Methods of Integration and Key Integration events Several other integration events exist to give ruggedness and flexibility to peak integration, however these are manufacturer dependant and outside the scope of this discussion. It is important to note that peak height measures the distance between the peak apex to the intersection with the integrated peak baseline. Crawford Scientific 10

12 Integration Events Peak Height or Peak Area For most HPLC analyses, peak areas are used for quantitative calculations, although, in most cases, equivalent results may be achieved with peak height. Peak area is especially useful because HPLC peaks may be tailed. In this case, because peak heights may vary (although area will remain constant), area vales are more repeatable. There are however, instances when peak height calculations may be better. For trace analysis, when the peak of interest is very small, use peak height for calculations, this reduces the error sustained in small changes in peak start and end time variation. The figure shown below illustrates the fact that areas of a chromatographic peak will change if the flow rate changes. Be aware that poorly maintained pump systems will have an unstable flow rate resulting in a loss of peak area precision (reproducibility). For this reason, always perform routine maintenance on the pump seals, check valves and filters. Make certain that the pump has been primed and dissolved gasses removed. Crawford Scientific 11

13 Quantitative analysis can be carried out using either peak height or peak area measurements Both peak height and area measurements will be irreproducible if injection volume varies therefore it is important to also ensure the injection or autosampler system is also well maintained, paying particular attention to the needle and injection valve. Principles of Quantitative Analysis Although there are many different types of detector used for quantitative HPLC analysis, the quantitative premise remains constant. Each detector will produce a response that depends upon the AMOUNT of analyte to which it is responding. The magnitude of the response may be dependent on the analyte itself as well as the matrix from which it comes and therefore standards are usually employed to calibrate the instrument response. There are some general requirements that must be met before a quantitative analysis can be undertaken, and these include: The identity of the component to be analysed should be known The best possible separation of the component should be achieved Standards of known purity should be available (otherwise accuracy may be compromised) Sample preparation and the quality of the separation (governed by the stationary phase) should be reproducible If internal standards are not used, then mobile phase flow rate and injection volume should be reproducible Crawford Scientific 12

14 Primary Methodologies for Quantitative Analysis Of course, quantitative analysis may be possible without meeting all of the above conditions, however, in such cases lower accuracy or precision must be expected. The analytical approach (method), and mathematical model used to calculate the quantity of an analyte can vary and is of great importance. Area %/ Height % (Normalisation) The Area% calculation procedure reports the area of each peak in the chromatogram as a percentage of the total area of all peaks. Area% does not require prior calibration and does not depend upon the amount of sample injected within the limits of the detector. No response factors are used. If all components respond equally in the detector and are eluted, then Area% provides a suitable approximation of the relative amounts of components. Area or Height % Quantitation Crawford Scientific 13

15 It is very important to note restrictions associated with this quantitative methodology which may reduce the accuracy or precision: The method assumes equal detector response for all analytes The method assumes that all components in the sample are seen in the chromatogram (i.e. they all elute from the column and produce a response in the detector) Peaks in the chromatogram due to analytical artefacts (solvents, carry-over etc.), may interfere with quantitation Area% is used routinely where qualitative results are of interest and to produce information used to create the calibration table required for other calibration procedures. The Height% calculation procedure reports the height of each peak in the run as a percentage of the total height of all peaks in the run. External Standard Quantitation The external standard (ESTD) quantitation procedure is the basic quantification procedure in which both calibration and unknown samples are analysed under the same conditions. The results (usually peak height or peak area measured using a data system) from the unknown sample are then related to those of a calibration sample, using a calibration curve, to calculate the amount in the unknown. The ESTD procedure uses absolute response factors. The response factor is normally calculated as amount / area of the analyte in the calibration sample. The response factors are obtained from a calibration and then stored. Response factors are normally viewed as a single point calibration curve, using the origin to determine the second point for the regression line this approach is often called Single Point Calibration. In subsequent sample analyses, component amounts are calculated by applying the response factor to the measured sample area, or by interpolating the result from the calibration curve as shown. Principle of External Standard Calibration Crawford Scientific 14

16 One precaution that must be observed in this type of calculation is that the sample injection size must be reproducible from run to run, since there is no (internal) standard in the sample to correct for variations in injection size or sample preparation. The method type also assumes a linear detector response and that samples do not contain a wide range of analyte concentrations. Worked Example of an External Standard Calculation Crawford Scientific 15

17 Calibration Curve A calibration curve is a graphical representation of the amount and response data for a single analyte (compound) obtained from one or more calibration samples. The curve is usually constructed by injecting an aliquot of the calibration (standard) solution of known concentration and measuring the peak area obtained. Peak height is sometimes used but only in exceptional circumstances. Construction of a calibration curve for quantitative analysis in HPLC A line of best fit (regression line) is used to join the points of the curve obtained. When only one calibration sample is used (i.e. at a single concentration), the origin (X0:Y0) is used as a second point to enable a regression line to be drawn (you can t draw a line through a single point what would the slope be??). With multi-level calibration (using several calibration samples at different analyte concentrations), the origin may be Crawford Scientific 16

18 included, excluded or forced depending upon the calibration model and application specifics. The line of best fit is usually given a correlation coefficient which is the square root of the regression coefficient and gives a measure of how well the data points fit a straight line. Of course, the instrument response may be exponential or second (or higher) order polynomial however these relationships are often more difficult to model and wherever possible a linear calibration relationship is preferred. Statistical Information Statistical Information of the nature of a regression line The intercept of the regression equation indicates systematic error a large positive or negative value may indicate an inherent error within the sample preparation or analysis The slope of the line indicates the analytical sensitivity The regression co-effcient is a statistical measure for goodness of fit to a straight line, calculated using the residuals (error) of each data point, an r value of +1 indicates a Crawford Scientific 17

19 straight line with positive slope. Some typical r values are shown it is important to always plot data to check the fit. As can be seen from the examples r values can be misleading! Workers sometimes use an r 2 value which is statistically more rigorous. External Standard Multi-level Calibration Multilevel calibration can be used when it is not sufficiently accurate to assume that a component shows a linear response or to confirm linearity of the calibration range. Each calibration level corresponds to a calibration sample with a particular concentration of components. Calibration samples should be prepared so that the concentration of each component varies across the range of concentrations expected in the unknown samples. In this way, it is possible to allow for a change in detector response with concentration and calculate response factors accordingly. Principle of External Standard Analysis with Multi-Level Calibration and an Example Calculation This multilevel calibration curve has three levels and shows a linear fit through the origin. This method of linear fit through the origin is similar to the single level calibration method. The detector response to concentration is assumed to be linear. The difference between the two calibration types is that, with multi-level calibration, the slope of the detector response can be determined by a best fit through a number of points, one for each level, and the regression coefficients used to substantiate the assumption of linearity. Unknowns are determined in the same way as the single level calibration model the difference is now that the method may be used to determine analyte concentrations over a wider range as the detector response has been calibrated with greater rigor. Most data systems will allow the input of calculation variables to allow a final result to be automatically calculated and printed. Crawford Scientific 18

20 Calibration Curve Information i Important Parameters in Multi-Level Calibration LOD (Limit of detection) The minimum amount of analyte that can be reliably detected. The limit of detection, expressed as the concentration, cl, or the quantity, ql, is derived from the smallest measure, xl, that can be detected with reasonable certainty for a given analytical procedure. The value of xl is given by the equation: xl = xbi - + ksbi where: xbi - is the mean of the blank measures, sbi is the standard deviation of the blank measures, and k is a numerical factor chosen according to the confidence level desired. A sample that contains a complex matrix (e.g. environmental, biological sample) may show response from the matrix. To determine the response a matrix sample without the compounds of interest should be analysed under same condition. The so-called blank chromatogram is treated as a starting point for the determination of LoD. The LoD is often defined as a ratio of S/N or peak area measurement precision. Examples of defining criteria are S/N ratio > 3 - peak height is compared to the noise height of the blank chromatogram to define the signal to noise ratio. Many modern data systems are capable of automatically providing a very accurate signal to noise ratio. LOQ (Limit of quantitation) - Lowest concentration of an analyte in a defined matrix where positive identification and quantitative measurement can be achieved using a specified method. The term limit of quantitation is preferred to limit of determination to differentiate it from LOD. LOQ has been defined as 3 times the LOD (Keith, 1991) or as 50% above the lowest calibration level used to validate the method (US-EPA, 1986). The LoQ is often defined as a ratio of S/N or peak area measurement precision and has much less stringent requirements than for limit of detection. Examples of defining criteria are S/N ratio > 20, or peak area precision better than 10%. The peak height is compared to the noise height of the blank chromatogram to define the signal to noise ratio. Many modern data systems are capable of automatically providing a very accurate signal to noise ratio. Crawford Scientific 19

21 Linear range - Concentration range over which the intensity of the signal obtained is directly proportional to the concentration of the species producing the signal. The linear range of a chromatographic detector represents the range of concentrations or mass flows of a substance in the mobile phase at the detector over which the sensitivity of the detector is constant within a specified variation, usually ±5 percent. The linear range of a detector may be presented as the plot of peak area (height) against concentration or mass flow-rate of the test substance in the column effluent at the detector. This plot may be either linear or log/log. The upper limit of linearity is that concentration (mass flow-rate) at which the deviation from an ideal linearity plot is greater than the specified percentage deviation (± x % window). Numerically, the linear range can be expressed as the ratio of the upper limit of linearity obtained from the linearity plot and the minimum detectability, both measured for the same substance. When presenting the linear range of a detector, either as a plot or as a numerical value, the test substance, the minimum detectability, and the specified deviation must be stated. Linearity is often expressed using the regression coefficient. Linear and Dynamic Range of Calibration Curves Slope the slope of the calibration line is often used to determine the sensitivity of the analytical method. Intercept the intercept indicates the degree of systematic error within the method and is the direct result of background response. Many workers often include the results of a blank analysis (i.e. where no analyte is added and x=0) as a point in the calibration curve from which the regression line and regression co-efficient are calculated. Crawford Scientific 20

22 Internal Standard Analysis The ISTD procedure eliminates the disadvantages of the ESTD method by adding a known amount of a component that serves as a normalizing factor. This component, the internal standard, is added to both calibration and unknown samples and compensates for losses during sample preparation or variability during the analytical determination. General Quantitative Principle of Internal Standard Analysis The compound used as an internal standard should be similar to the calibrated compound (analyte), chemically and in retention time, but it must be chromatographically distinguishable. If the ISTD procedure is used for calibrations with a non-linear characteristic, care must be taken that errors resulting from the calculation principle do not cause systematic error. In multi-level calibrations, the amount of the ISTD compound should be kept constant, i.e. the same for all levels. The internal standard is usually added at a concentration of approximately one third to one half of the expected MAXIMUM analyte concentration. This is especially important when using MS detectors to ensure the internal standard does not interfere with the ionisation of the analyte molecule. For internal standard analysis, the amount of the component of interest is related to the amount of the internal standard component by the ratio of the responses of the two peaks. This is normally achieved by plotting the response ratio against the concentration ratio as shown next. Crawford Scientific 21

23 Principle and Calculation for Single Level Internal Standard Calibration Internal standards are usually well characterised compounds which are not present in the sample and is added in a known (constant) concentration to standard and sample solutions (or matrices) usually at the beginning of the analytical process to compensate for losses and variability throughout the sample preparation and analytical processes. Good internal standards: elute near to, but are well resolved from, the analyte of interest are chemically and physically similar to the analyte are not present in the original sample mixture are unreactive towards any of the sample components are available in highly pure form are added in the concentration range of the expected MAXIMUM analyte concentration It is often fairly difficult to fulfil all of these requirements for HPLC analysis. Often for MS detection, the internal standard will be an isotopically labelled version of the analyte, which can be spectrally rather than chromatographically resolved (i.e. the labelled version has a different mass than that analyte). Care must be taken when using this approach to avoid ion-suppression effects, which can serious affect the reproducibility of the analysis. Crawford Scientific 22

24

Instrumentation of HPLC. Detectors

Instrumentation of HPLC. Detectors Instrumentation of HPLC Detectors i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

Validation of an Analytical Method

Validation of an Analytical Method Validation of an Analytical Method Refer to: ICH Guideline Q2(R1), Validation of Analytical Procedures: Teaxt andmethodology. Introduction All major laboratories eg. in the industry operates with clearly

More information

Varian Galaxie Chromatography Data System for Preparative HPLC

Varian Galaxie Chromatography Data System for Preparative HPLC Varian Galaxie Chromatography Data System for Preparative HPLC By Gary Burce Varian, Inc. 2700 Mitchell Drive, Walnut Creek, CA 95498 USA Abstract Galaxie is an ideal chromatography data system for the

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion chromatography (GPC/SEC)

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).

More information

Polymer analysis by GPC-SEC. Technical Note. Introduction

Polymer analysis by GPC-SEC. Technical Note. Introduction Polymer analysis by GPC-SEC Technical Note Introduction Gel Permeation Chromatography (GPC), also referred to as Size Exclusion Chromatography (SEC) is a mode of liquid chromatography in which the components

More information

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition Understanding Your Spectra Module Agilent OpenLAB CDS ChemStation Edition Notices Agilent Technologies, Inc. 1994-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System G. Maio 1, F. Steiner 1, H. Franz 1, M. Tracy 2, X. Liu 2, F. Arnold 1 1 Dionex Softron GmbH, D-21 Germering,

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

Luminescence transitions. Fluorescence spectroscopy

Luminescence transitions. Fluorescence spectroscopy Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation

More information

High Pressure/Performance Liquid Chromatography (HPLC)

High Pressure/Performance Liquid Chromatography (HPLC) High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the

More information

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography CHEM 329 Professor Vogt TA: Fahad Hasan Allison Poget Date Performed: April 5, 2016 Date Submitted: April 12,

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

PDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION

PDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION 1 2 3 4 5 INTRODUCTION G-20 CHROMATOGRAPHY 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 Chromatographic separation techniques are multi-stage separation methods in which the components

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Training Courses. Chromatography training from the Crawford Scientific experts

Training Courses. Chromatography training from the Crawford Scientific experts Training Courses Chromatography training from the Crawford Scientific experts HPLC Fundamentals This one-day course introduces the fundamentally important concepts associated with HPLC analysis including

More information

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective SHULAMIT LEVIN HPLC in Pharmaceutics Σ Stability Indicating Methods Extra

More information

Chromatography & instrumentation in Organic Chemistry

Chromatography & instrumentation in Organic Chemistry Chromatography & instrumentation in Organic Chemistry What is Chromatography? Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Rosemary extract liquid

Rosemary extract liquid EUROPEAN COMMISSION DIRECTORATE-GENERAL JOINT RESEARCH CENTRE Institute for Reference Materials and Measurements Community Reference Laboratory - Feed Additives Authorisation Evaluation Report of the Community

More information

Theory and Instrumentation of GC. Chromatographic Parameters

Theory and Instrumentation of GC. Chromatographic Parameters Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this

More information

Instrumental Chemical Analysis

Instrumental Chemical Analysis L1 Page1 Instrumental Chemical Analysis Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester, 2016/2017 L1 Page2 Course Syllabus Course title:

More information

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp Validation of analytical methods Adrian Covaci Toxicological Center, University of Antwerp Why validate? The objective of validating an analytical method is to demonstrate its suitability for its intended

More information

Spectroscopy and Chromatography

Spectroscopy and Chromatography Spectroscopy and Chromatography Introduction Visible light is one very small part of the electromagnetic spectrum. The different properties of the various types of radiation depend upon their wavelength.

More information

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure. High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.

More information

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CONFIRMATORY TESTS CAC/GL 56-2005 When analyses are

More information

COLA Mass Spec Criteria

COLA Mass Spec Criteria Wednesday April 5, 2017 2pm COLA Mass Spec Criteria Kathy Nucifora, MPH, MT(ASCP) COLA, Accreditation Division Director Columbia, MD DESCRIPTION: Kathy Nucifora, COLA Accreditation Division Director, will

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

TR ACE ORGANIC. lgcstandards.com/foodandenvironment

TR ACE ORGANIC. lgcstandards.com/foodandenvironment U S E O F S TA B L E I S OTO P E I NTE R N A L S TA N D A R D S F O R TR ACE ORGANIC A N A LY S I S lgcstandards.com/foodandenvironment dr.ehrenstorfer@lgcgroup.com This paper explains why stable isotope

More information

Introduction to Pharmaceutical Chemical Analysis

Introduction to Pharmaceutical Chemical Analysis Introduction to Pharmaceutical Chemical Analysis Hansen, Steen ISBN-13: 9780470661222 Table of Contents Preface xv 1 Introduction to Pharmaceutical Analysis 1 1.1 Applications and Definitions 1 1.2 The

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

Work plan & Methodology: HPLC Method Development

Work plan & Methodology: HPLC Method Development Work plan & Methodology: HPLC Method Development The HPLC analytical Method developed on the basis of it s chemical structure, Therapeutic category, Molecular weight formula, pka value of molecule, nature,

More information

Analytical Chemistry

Analytical Chemistry Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography

More information

Volume 6, Issue 2, January February 2011; Article-015

Volume 6, Issue 2, January February 2011; Article-015 Research Article DEVELOPMENT AND VALIDATION OF A RP-HPLC METHOD FOR THE DETERMINATION OF DAPOXETINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATION USING AN EXPERIMENTAL DESIGN Pratik Mehta*, Ujjwal Sahoo,

More information

3) In CE separation is based on what two properties of the solutes? (3 pts)

3) In CE separation is based on what two properties of the solutes? (3 pts) Final Exam Chem 311 Fall 2002 December 16 Name 1) (3 pts) In GC separation is based on the following two properties of the solutes a) polarity and size b) vapor pressure and molecular weight c) vapor pressure

More information

Response to Questions Raised by Tom Phillips Concerning the Enforcement Analyitical Method for Avail and Nutrisphere Products. Verdesain Life Sciences

Response to Questions Raised by Tom Phillips Concerning the Enforcement Analyitical Method for Avail and Nutrisphere Products. Verdesain Life Sciences Response to Questions Raised by Tom Phillips Concerning the Enforcement Analyitical Method for Avail and Nutrisphere Products Verdesain Life Sciences December 12, 2018 Philip Davidson of Maryland read

More information

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk International Journal of Engineering Science Invention (IJESI) ISSN (Online): 2319 6734, ISSN (Print): 2319 6726 www.ijesi.org PP. 36-41 Development And Validation Of Rp-Hplc Method For Determination Of

More information

Definitive EtG/EtS LC-MS/MS Analysis:

Definitive EtG/EtS LC-MS/MS Analysis: Definitive / LC- Analysis: A Rugged 4-Min Method for High-Throughput Labs By Justin Steimling and Frances Carroll Abstract Methods for monitoring alcohol consumption biomarkers and are generally limited

More information

Chromatography and other Separation Methods

Chromatography and other Separation Methods Chromatography and other Separation Methods Probably the most powerful class of modern analytical methods for analyzing mixture of components---and even for detecting a single component in a complex mixture!

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of Analytical method validation By Juree Charoenteeraboon, Ph.D Analytical method Goal consistent, reliable and accurate data Validation analytical method serve the goal of analytical lmethod Its data reveals

More information

CHMC 16 Instrumental Analysis (Syllabus)

CHMC 16 Instrumental Analysis (Syllabus) CHMC 16 Instrumental Analysis (Syllabus) Course Description CHMC16 is a laboratory course designed to compliment CHMC11 (Principles of Analytical Instrumentation). This course will provide a practical

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

Determination of Caffeine by HPLC

Determination of Caffeine by HPLC Determination of Caffeine by HPLC Introduction It was a long history before real high performance liquid chromatography (HPLC) had evolved. The very first indication of a chromatographic separation was

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Liquid chromatography was developed by Tswett in early 1900 s and was shown to be a powerful separation

More information

Substance Characterisation for REACH. Dr Emma Miller Senior Chemist

Substance Characterisation for REACH. Dr Emma Miller Senior Chemist Substance Characterisation for REACH Dr Emma Miller Senior Chemist Submission of a REACH registration dossier, either as a lead registrant or a joint registrant Submission of an Inquiry to ECHA Submission

More information

Analytical Methods Validation

Analytical Methods Validation Analytical Methods Validation In-Process Control Methods for the Manufacture of Active Pharmaceutical Ingredients Jose Zayas*, Victor Sanchez, and Michelle Talley The authors propose a strategy for classifying

More information

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Ind. J. Anal. Chem Vol. 7 11. 2008 SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Authors for correspondence : R. R. Singh1*, M. V. Rathnam,

More information

MassHunter TOF/QTOF Users Meeting

MassHunter TOF/QTOF Users Meeting MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks

More information

Analytical Method Validation: An Updated Review

Analytical Method Validation: An Updated Review INTERNATIONAL JOURNAL OF ADVANCES IN PHARMACY, BIOLOGY AND CHEMISTRY Analytical Method Validation: An Updated Review G. Geetha, Karanam Naga Ganika Raju, B. Vignesh Kumar and M. Gnana Raja* Sankaralingam

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION Research Article ISSN:2230-7346 Journal of Global Trends in Pharmaceutical Sciences Vol.3, Issue 2, pp -619-627, April June 2012 DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

The Claviature of Gas Analysis

The Claviature of Gas Analysis The Claviature of Gas Analysis Mass spectrometric gas analysis for quality assurance of gases and gas mixtures High purity and special gases are becoming increasingly important in the development of modern

More information

METHOD 3600C CLEANUP

METHOD 3600C CLEANUP METHOD 3600C CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

HPLC analysis of VFAs, Furfural and HMF

HPLC analysis of VFAs, Furfural and HMF HPLC analysis of VFAs, Furfural and HMF Introduction The following chromatographic analysis is developed in order to analyse volatile fatty acids and carbohydrate degradation products; furfural, hydroxymethylfurfural

More information

Ion Chromatography (IC)

Ion Chromatography (IC) Ion Chromatography (IC) Purpose: This module provides an introduction to Ion Chromatography (IC). In this module the basic theory and applications of IC will be presented at a level that assumes a basic

More information

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results P EGAU 4D GCxGC TFM Delivering the Right Results LEC s P EGAU 4D GCxGC-TFM The Ultimate Analytical Instrument Pioneering GCxGC technology, combined with the Pegasus TFM and ChromaTF software provides today

More information

The Theory of HPLC. Ion Pair Chromatography

The Theory of HPLC. Ion Pair Chromatography The Theory of HPLC Ion Pair Chromatography i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

WADA Technical Document TD2003IDCR

WADA Technical Document TD2003IDCR IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must

More information

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS 1. Given the HPLC chromatogram below and the fact that compounds A and B, eluting at 17.1 and 22.0 min, respectively, were present at concentrations

More information

Chromatographic Analysis

Chromatographic Analysis Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity

More information

883 Basic IC plus. The smallest Metrohm ion chromatograph

883 Basic IC plus. The smallest Metrohm ion chromatograph 883 Basic IC plus The smallest Metrohm ion chromatograph 883 Basic IC plus small but powerful 02 The 883 Basic IC plus completes the family of intelligent Metrohm ion chromatographs. Although it is the

More information

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page: ISSN: 2320-2831 Research article Open Access Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

More information

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017.

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017. GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG 90-2017 Adopted in 2017. CAC/GL 90-2017 2 OBJECTIVE 1. The purpose of these guidelines

More information

Introduction to E&Ls 1

Introduction to E&Ls 1 Introduction to E&Ls 1 Overview What industries need to determine E&Ls Define extractables and leachables Basic overview of an E&L study Regulatory landscape 2 A leader in plastics analysis Jordi Labs

More information

Quantification of Furocoumarins

Quantification of Furocoumarins Version 8 (12/04/2013) Quantification of Furocoumarins HPLC/DAD Procedure Scope The European Cosmetics Directive 76/768/EEC, which is about to be adopted in the European Cosmetic Regulation, in Annex II

More information

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace_Rev3 Introducing

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS Issn No: 976-39 RESEARCH ARTICLE STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS CHETLAPALLI SATYA SRINIVAS 1, P.RENUKA DEVI 2 and GAMPA VIJAYAKUMAR*

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

MassHunter Software Overview

MassHunter Software Overview MassHunter Software Overview 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks A

More information

CHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives:

CHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives: CHEM 3281 Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract Objectives: The objectives of this experiment are: 1. to evaluate the fundamental parameters of gas chromatography; 2. to identify

More information

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING QUESTION BANK V SEMESTER EI6501 Analytical Instruments Regulation 2013 Academic

More information

All measurement has a limit of precision and accuracy, and this must be taken into account when evaluating experimental results.

All measurement has a limit of precision and accuracy, and this must be taken into account when evaluating experimental results. Chapter 11: Measurement and data processing and analysis 11.1 Uncertainty and error in measurement and results All measurement has a limit of precision and accuracy, and this must be taken into account

More information

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten Test Report Glava AS Emission test of Ecophon Venus according to M1 classification October / November 2006 Client: Glava AS Nybråtveien 2 1832 Askim Norge Date: 8 th of November 2006 Testing Laboratory:

More information

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system Technical Note Introduction The HP 1090 Series HPLC system, introduced in 1983, is regarded to be an outstanding

More information

Physical Separations and Chromatography

Physical Separations and Chromatography Lab #5A & B: Physical Separations and Chromatography Individual Objectives: At the end of these experiments you should be able to: Ø Distinguish between Rf and tr; chromatograph and chromatogram; adsorption

More information

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller

More information

Isotope Dilution Mass Spectrometry

Isotope Dilution Mass Spectrometry Isotope Dilution Mass Spectrometry J. Ignacio Garcia Alonso and Pablo Rodriguez-Gonzalez Faculty of Chemistry, University of Oviedo, Oviedo, Spain E-mail: jiga@uniovi.es, rodriguezpablo@uniovi.es RSC Publishing

More information

WADA Technical Document TD2015IDCR

WADA Technical Document TD2015IDCR MINIMUM CRITERIA FOR CHROMATOGRAPHIC-MASS SPECTROMETRIC CONFIRMATION OF THE IDENTITY OF ANALYTES FOR DOPING CONTROL PURPOSES. The ability of a method to identify an analyte is a function of the entire

More information