Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Size: px
Start display at page:

Download "Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70"

Transcription

1 Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds also have an extremely low odor detection threshold and because of this, many drinking water laboratories require detection levels of below 10ppt. This application note investigates the detection of Geosmin and 2-Methylisoborneol at a one part per trillion level utilizing Purge and Trap (P&T) coupled with a Gas Chromatograph and Mass Spectrometer (GC/MS). Introduction Geosmin and 2-Methylisoborneol have very poor purge efficiencies. In order to detect these compounds down to the 1ppt level it was necessary to optimize the P&T and GC/MS techniques. In this study, the GC/MS was run in Selective Ion Monitoring (SIM) mode while the P&T conditions were modified in order to achieve a 1ppt level of detection. A 10% (w/v) salt solution was utilized along with a 25mL purge volume in order to aid in compound detection. The Stratum PTC was configured with a #1 trap and three different injection techniques were examined in order to determine the optimum method for injection. Experimental-Instrument Conditions The Stratum PTC and Aquatek 70 Autosampler were coupled to an Agilent 7890A GC and a 5975 inert XL MS system for analysis. A #1 trap was the analytical trap used. The GC was configured with a J&W Scientific DB-VRX 30m x 0.250mm x 1.4μm column. The MS scanned in the SIM mode. The Stratum PTC transfer line was configured to the GC column three different ways. The first was through a cryofocusing module, the second was through the Split/Splitless inlet of the GC and the third configuration was a direct column connection. Finally, a 25mL purge volume of 10% (w/v) salt water was used. The GC/MS parameters are outlined in Tables 1 and 2 respectively while Tables 3 and 4 outline the P&T conditions. GC Parameters MSD Parameters GC: Agilent 7890A MSD: Agilent 5975C inert XL Column: J&W Scientific DB-VRX 30m x 0.250mm x1.4um Source: 230 C Oven Program: 40 C for 2.0 min; 16 C/min to 160 C for 0 min; 20 C /min to 240 C for 5.0 min; 18.5 min. runtime Quad: 150 C Inlet: 220 C Solvent Delay: 5.0 min Column Flow: 1.02mL/min SIM Ions 95, 107, 108, 112, 125, 126 Split: (Used for Split/Splitless 10:1 Dwell Time: 100 msec dwell per ion Inlet Injection only) Gas: Helium Pressure: psig Table 1: GC Parameters Table 2: MS Parameters Sales/Support: Main:

2 Stratum PTC and AQUATek 70 Parameters Water Parameters for Cryofocusing Injection Variable Value Variable Value Pressurize Time 0.60 min Purge Time Fill IS Time 0.00 min Purge Temp 0 C Sample Transfer Time 0.75 min Purge Flow 45mL/min Rinse Loop Time 0.75 min Dry Purge Time 5.00 min Purge Loop Time 1.00 min Dry Purge Temp 20 C Bake Rinse On Dry Purge Flow 45mL/min Number of Bake Rinses 3 GC Start End of Desorb Bake Drain Time 1.50 min Desorb Preheat Temp 220 C Bake Drain Flow 250mL/min Desorb Drain On Valve Oven Temp 175 C Desorb Time 6.00 min Transfer Line Temp 175 C Desorb Temp 225 C Sample Mount Temp 60 C Desorb Flow 300mL/min Purge ready Temp 40 C Bake Time min Condenser Ready Temp 40 C Bake Temp 230 C Condenser Purge Temp 20 C Bake Flow 250mL/min Standby Flow 45mL/min Condenser Bake Temp 175 C Pre-Purge Time 0.0 min Focus Temp -100 C Pre-Purge Flow 0.0mL/min Inject Time 2.00 min Sample Heater On Inject Temp 200 C Sample Preheat Time 0.01 min Standby Temp 150 C Sample Temp 40 C Table 3: Stratum PTC and AQUATek Parameters for a Cryofocusing Injection Stratum PTC Parameters are in Blue Cryofocusing Parameters are in Green Stratum PTC and AQUATek 70 Parameters Water Parameters for Split/Splitless Inlet and Direct Column Injections Variable Value Variable Value Pressurize Time 0.60 min Purge Time Fill IS Time 0.00 min Purge Temp 0 C Sample Transfer Time 0.75 min Purge Flow 100mL/min Rinse Loop Time 0.75 min Dry Purge Time 5.00 min Purge Loop Time 1.00 min Dry Purge Temp 20 C Bake Rinse On Dry Purge Flow 45mL/min Number of Bake Rinses 3 GC Start Start of Desorb Bake Drain Time 1.50 min Desorb Preheat Temp 220 C Bake Drain Flow 250mL/min Desorb Drain On

3 Valve Oven Temp 175 C Desorb Time 2.00 min Transfer Line Temp 175 C Desorb Temp 225 C Sample Mount Temp 60 C Desorb Flow 300mL/min Purge ready Temp 40 C Bake Time min Condenser Ready Temp 40 C Bake Temp 230 C Condenser Purge Temp 20 C Bake Flow 250mL/min Standby Flow 45mL/min Condenser Bake Temp 175 C Pre-Purge Time Pre-Purge Flow Sample Heater Sample Preheat Time 0.0 min 0.0mL/min On 0.01 min Sample Temp 40 C Table 4: Stratum PTC and Aquatek Parameters for Split/Splitless Inlet and Direct Column Injections Stratum PTC Parameters are in Blue A 50ppb working calibration standard was prepared in methanol. standards were prepared in a 50mL volumetric flask and filled to volume with 10% (w/v) de-ionized salt water solution. The calibration range was ppt. The standards were transferred to headspace free 40mL vials for analysis. The calibration data was analyzed using Agilent Chemstation software. The calculated linear regression and the of each compound and their respective injection techniques are outlined in Tables 5, 6 and 7. Method Detection Limit (), Carryover, and and Accuracy Study A statistical determination of the s was determined for both of the compounds with each injection technique by analyzing seven replicate standards of a 1ppt calibration standard. Furthermore, seven replicate standards of a 10ppt calibration standard were analyzed for each injection method in order to determine the precision and accuracy of the experimental conditions. Finally, three blanks were run after a high, 100ppt calibration standard in order to determine percent carryover for each technique. The detection limits, precision and accuracy data and percent carryover for each compound and their respective injection method are provided in Tables 5, 6, and 7. Compound % Carryover Geosmin Methylisoborneol Table 5: Cryofucusing Injection Experimental Results Summary Compound %Carryover Geosmin Methylisoborneol Table 6: Split/Splitless Inlet Injection Experimental Results Summary

4 Compound %Carryover Geosmin Methylisoborneol Table 7: Direct Column Injection Experimental Results Summary Figure 1: Total Ion Chromatogram of a 50ppt Geosmin and 2-Methylisoborneol Standard Using Cryofocusing Injection Figure 2: Total Ion Chromatogram of a 50ppt Geosmin and 2-Methylisoborneol Standard using Split/Splitless Inlet Injection

5 Figure 3: Total Ion Chromatogram of a 50ppt Geosmin and 2-Methylisoborneol Standard Using a Direct Column Injection Conclusions The Stratum PTC and AQUATek 70 configured with an Agilent GC/MS system operating in SIM mode performed very well in detecting Geosmin and 2-Methylisoborneol with all three injection techniques. The linearity of all of the curves was for both compounds with a of below 10%. The system displayed excellent accuracy and precision results and the detection limits for the compounds were below 0.250ppt for all three injection methods. The 25mL sample volume and 10% (w/v) salt solution aided in increasing the purge efficiency of both the Geosmin and the 2-Methylisoborneol compounds. Further analysis of the compound responses for all three techniques displayed that the direct column injection configuration had the optimum compound response, thus although the three techniques all showed consistent analytical results and comparable detection limits, I would recommend the direct column injection for Geosmin and 2-Methylisoborneol detection down to a 1ppt level.

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist;

More information

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount

More information

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring Application Note Abstract The United States Environmental Agency (USEPA) has established the Unregulated

More information

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is

More information

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).

More information

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note Analysis of Epichlorohydrin in Drinking Water Using the Lumin P & T Concentrator Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract Epichlorohydrin (ECH) is a versatile

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

Volatile Organic Compounds in Every Day Food

Volatile Organic Compounds in Every Day Food Volatile Organic Compounds in Every Day Food Application Note Abstract Volatile Organic Compounds (VOCs) have been analyzed in drinking water and pharmaceuticals extensively in recent years. Now, VOCs

More information

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration

More information

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds

More information

Methanol Extraction of high level soil samples by USEPA Method 8260C

Methanol Extraction of high level soil samples by USEPA Method 8260C Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed

More information

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA

More information

Optimizing. Abstract: is standardd. procedures. altered to

Optimizing. Abstract: is standardd. procedures. altered to Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications

More information

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction

More information

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol. Determination of Volatile Petroleum Hydrocarbons ANNE JUREK Abstract: Due to current events, the importance of determining volatile petroleum hydrocarbons in both soils and waters has become an issue.

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract

More information

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to. Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection

More information

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative

More information

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to. ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental

More information

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators

More information

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD) Application Note 16181100 Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD) Keywords Beer PFPD Purge and Trap Sulfur Introduction The presence or absence

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods

More information

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application Pharmaceuticals Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville

More information

Optimizing Standard Preparation

Optimizing Standard Preparation JSB is an authorised partner of Optimizing Standard Preparation Anne Jurek #106 Abstract: Standard preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s JSB is an authorised partner of Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s Anne Jurek #120 Introduction: The USEPA developed

More information

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8 application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the

More information

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application Environmental Author Philip L. Wylie Agilent Technologies, Inc. 2850 Centerville Road

More information

Determination of Volatile Organic Compounds in Air

Determination of Volatile Organic Compounds in Air JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Study of Residual Solvents in Various Matrices by Static Headspace

Study of Residual Solvents in Various Matrices by Static Headspace Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.

More information

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of phenols in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicable to the quantitative determination of

More information

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Method of determination of phtalates in spirituous beverages by gaschromatography/mass OIV-MA-BS-33 Method of determination of phtalates in spirituous beverages by gas-chromatography/mass Type IV method 1. SCOPE OF APPLICATION This method applies to the detection and assay of some phthalates

More information

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography Introduction ASTM International publishes a number of different methods for the determination of volatile compounds in water-based coatings.

More information

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies,

More information

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road

More information

ICBA Guidance Document to Mitigate the Potential for Benzene Formation in Beverages Adopted by the ICBA Council on 29 April 2006

ICBA Guidance Document to Mitigate the Potential for Benzene Formation in Beverages Adopted by the ICBA Council on 29 April 2006 ICBA Guidance Document to Mitigate the Potential for Benzene Formation in Beverages Adopted by the ICBA Council on 29 April 2006 Secretariat 2005-2006 UNESDA - Union of European Beverages Associations

More information

USP<467> residual solvents

USP<467> residual solvents Application Note Pharmaceuticals USP residual solvents Applying the Agilent 977A MSD with the Agilent 797A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies,

More information

Analyzing Semi-Volatiles by GC-MS

Analyzing Semi-Volatiles by GC-MS Analyzing Semi-Volatiles by GC-MS Jonathan A. Karty July 31, 2008 Overview GC-MS Instrumentation Description GC-MS Method Sections In Detail Sequence Programming Data Analysis Tips and Tricks 1 Quick Description

More information

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput JSB is an authorised partner of Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput Application Note - Environmental #129 Author Anne Jurek Applications Chemist

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES ROBOKROM 1 TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview ROBOKROM 2 +8 OPERATIONAL MODES HRGC+HRGC-MS HRGC+HPLC-MS STATIC HEAD SPACE PURGE & TRAP SMPE

More information

U.S. EPA Method 8270 for multicomponent analyte determination

U.S. EPA Method 8270 for multicomponent analyte determination ENVIRONMENTAL application note U.S. EPA Method 8270 for multicomponent analyte determination Elaine A. LeMoine and Herman Hoberecht Introduction Multicomponent analytes are compounds that yield several

More information

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations OUTLINE OF METHOD This gas chromatography/mass spectrometry (GC/MS) method provides

More information

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA

More information

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques Application Note No. 085 The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques Diane Nicholas Multiple techniques using one instrument Simple

More information

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 JSB is an authorised partner of A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek #101 Introduction: The United States Environmental Protection Agency (USEPA)

More information

GC/MS Application Note

GC/MS Application Note GC/MS Application Note Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass Spectrometry www.palsystem.com Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass

More information

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Application Note Introduction Environmental laboratories have utilized the Purge & Trap sample concentration technique for over

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

Detection of Chemical Warfare Agents by Transportable GC/MS

Detection of Chemical Warfare Agents by Transportable GC/MS Detection of Chemical Warfare Agents by Transportable GC/MS Application Note Gas Chromatography/Mass Spectrometry Author Robert V. Mustacich Agilent Technologies, Inc. 5301 Stevens Creek Boulevard Santa

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) Date 02/24/96 Page 1 1.1. This method is designed to measure the concentration

More information

Method - Determination of aromatic amines in mainstream cigarette smoke

Method - Determination of aromatic amines in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of aromatic amines in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicaple to quantitative determination

More information

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection) Limit-test of NDMA and NDEA in Sartans 1 Purpose and scope by GC-MS (Liquid-direct-injection) The method is used to test the permissible limits of N-nitrosodimethylamine and N- nitrosodiethylamine in various

More information

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations Ellen Vercruyssen Supervisors: Prof. dr. Jo Schaubroeck, dr. Jan Van Biesen Master's dissertation

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC Applying the Agilent 5977A MSD to the Analysis of USP Residual Solvents with the 7697A Headspace Sampler and 7890B GC Application Note Pharmaceuticals Authors Roger L. Firor and Mike Szelewski Agilent

More information

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing Automated Preparation of Simulated Distillation Standards and Samples for ASTM Methods D2887, D7213, D7398 and D6352 using the 7693A System with Easy SamplePrep Software Author Roger L. Firor Agilent Technologies,

More information

A New Way to Extract Off-Flavor Compounds in the Aquatic Environment. Stir Bar Sorptive Extraction (SBSE)

A New Way to Extract Off-Flavor Compounds in the Aquatic Environment. Stir Bar Sorptive Extraction (SBSE) A New Way to Extract Off-Flavor Compounds in the Aquatic Environment Stir Bar Sorptive Extraction (SBSE) Young In Scientific Co., Ltd. 34-1 Content Introduction Purpose Experimental Results Experimental

More information

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System Analysis of Terpenes in Cannabis Using the Agilent 7697A/789B/977B Headspace GC-MSD System Faster Analysis Time = Greater Productivity Application Note Cannabis, Food Authors Ronald Honnold, Robert Kubas,

More information

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C APPLICATION NOTE 10441 Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C Authors Andrea Caruso, 1 Tommaso Albertini, 1 Jacob A. Rebholz 2 ; 1 Thermo Fisher Scientific,

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the experiment, I expect that you will

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask

More information

ACETONE IN URINE BY GC/MS Headspace Code GC10110

ACETONE IN URINE BY GC/MS Headspace Code GC10110 ACETONE IN URINE BY GC/MS Headspace Code GC10110 BIOCHEMISTRY The acetone or dimethyl ketone is a product which falls into the biochemical physiological small amounts are formed as a product of the metabolism

More information

Optimizing Conditions for Volatiles Analysis

Optimizing Conditions for Volatiles Analysis July 2016 SPEXCertiPrep.com/ECS Optimizing Conditions for Volatiles Analysis by Mark A. Ferry GC/MS Technical Consultant MFerry@SPEX.com A publication by Environmental Consulting & Supplies a Division

More information