A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

Size: px
Start display at page:

Download "A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C"

Transcription

1 Int J Thermophys (2017) 38:37 DOI /s TEMPMEKO 2016 A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C R. Bosma 1 A. Peruzzi 1 R. Van Breugel 1 C. Bruin-Barendregt 1 Received: 21 June 2016 / Accepted: 16 December 2016 / Published online: 5 January 2017 Springer Science+Business Media New York 2016 Abstract A new sub-millikelvin calibration facility for the range 0 Cto30 Cis described, that allows calibration of customer thermometers, other than standard platinum resistance thermometers, with an uncertainty lower than 1 millikelvin. The improvements with respect to the traditional calibration facility are reported with particular emphasis on the temperature control (better than 0.2 mk), resistance measurement and calibration procedure. The new facility was validated by using 6 standard platinum resistance thermometers and the calibration uncertainty in the range from 0 Cto30 C amounted to 0.31 mk 0.35 mk. To demonstrate the potentiality of this facility, two oceanographic thermometers, Sea-Bird Electronics SBE 3 and SBE 35, were calibrated with an expanded uncertainty of 0.8 mk (k = 2). Keywords Calibration Deep ocean thermometer Sub-millikelvin 1 Introduction In the range from 0 Cto30 C, the realization and dissemination of the ITS-90 temperature scale are performed using two fixed points, water and gallium, and a standard platinum resistance thermometer, SPRT [1]. The scale between the fixed points is based on the interpolating function for the SPRT, W r (T 90 ), and the deviation function, W (T 90 ) W r (T 90 ). The uncertainty of the SPRT is then well below the millikelvin Selected Papers of the 13th International Symposium on Temperature, Humidity, Moisture and Thermal Measurements in Industry and Science. B R. Bosma rbosma@vsl.nl 1 VSL, Dutch Metrology Institute, Delft, The Netherlands

2 37 Page 2 of 8 Int J Thermophys (2017) 38:37 Fig. 1 Deep ocean thermometers Sea-Bird, Model SBE 35 (a) and Model SBE 3 (b) level and at the fixed points even near tenth of microkelvin. For the dissemination of the ITS-90 scale, other thermometers, such as thermistors, are calibrated by comparison with the SPRT s in liquid baths [2], but the best uncertainties attainable for these services are at the moment typically several millikelvin. In the framework of the European Metrology Research Programme, project MeteoMet 2, VSL needed to calibrate two deep sea ocean thermometers, the Sea- Bird Electronics SBE 3 and SBE 35 (see Fig. 1) in the range from 0 Cto30 C. The uncertainty of the calibration needed to be well within 1 mk to meet the requirements of the oceanografic community. A paper on the Effect of pressure on deep-ocean thermometers is under review in the same journal. While the SBE 35 design allows it to be calibrated at the fixed points, the SBE 3 has only a very thin thermometer stem of 0.8 mm diameter and 60 mm length, which does not suit the fixed points. As both thermometers are thermistor-based instruments, there is not a reference function similar as for a standard platinum resistance thermometer, SPRT. In this paper, the modifications are described made to a conventional calibration bath in order to obtain the sub-millikelvin uncertainty level. Special attention is given to the temperature control of the bath, the temperature stability and uniformity and the non-uniqueness contribution. The new set-up was validated using multiple SPRT s and as an example of sub-millikelvin calibration, two deep ocean thermometers, a SBE 3 and a SBE35, were calibrated. 2 Realization of the Sub-millikelvin Set-up in the Range from 0 Cto 30 C The comparison calibrations in the range from 0 Cto30 C at VSL are realized using an ice point, an ethanol bath (below 10 C) and a water bath (above 10 C). The baths temperatures are controlled with controllers using Pt100 control sensors. The reference SPRT and the customer thermometer(s) are inserted in the wells of a comparison block, which can be submerged from a few centimetre to up to 45 cm (stem length of SPRT s)

3 Int J Thermophys (2017) 38:37 Page 3 of 8 37 Table 1 CMC claim in water bath at 10 C for Pt100 with the previous calibration measurement capability, CMC, and the CMC for a SPRT calibrated with the sub-millikelvin facility Facility Previous CMC Sub-millikelvin CMC Temperature / C 10 0u i /mk 28 Reference thermometers(s) Calibration Selfheating Bridge Resistor Customer thermometer Bridge Resistor Repeatability Bath/comparator block Non-uniqueness Combined uncertainty Expended uncertainty (k = 2) in the liquid. The reference SPRT s are measured with a resistance bridge at 1 ma, without selfheat correction, but adding an uncertainty component in the calibration uncertainty (in most commercial calibrations the calibration time is found to be more important than the lowest calibration uncertainty). A typical uncertainty claim for a comparison calibration in this temperature range is 4 mk (see Table 1). To reduce the calibration uncertainty below 1 mk the following actions were needed: Improve the temperature control for stability and uniformity, Minimize the influence of immersion effect of reference and customer thermometers and uncertainty from bath validation, Reduce the uncertainty in the resistance measurement, Reduce the uncertainty component from the selfheat effect on reference thermometer. The Tamson TV7000SP water bath at VSL is normally used in the range from 10 C to 80 C, and the control stability, measured with a SPRT in the comparison block, is between 2 mk and 3 mk (see Fig. 2). To improve the control stability, a control set-up was used similar to the low-temperature cryostat of VSL [3]. A SPRT was used as control sensor, read by 71/2 digit digital multimeter, DMM, and a dc power supply as control actuator. A computer with a PID algorithm controlled the operation of DMM and power supply. No selfheat correction was used for the control sensor, because only the temperature stability is important and not the absolute temperature. As the control power was only 300 W, the total temperature range could not be covered with a cooler at a fixed temperature setpoint. An optimum was found with an offset of 3 C between the setpoint of the bath, t SP, and the temperature of the cooler, t PV,Cooler. Figure 3

4 37 Page 4 of 8 Int J Thermophys (2017) 38:37 Fig. 2 Stability in the comparison block, t 90 t SP, in Tamson water bath using the Hart Scientific 2100 temperature controller shows the stability of the bath temperature, t PV, and the stability of the temperature in the comparison block, t 90. The monitored stability of the bath, shown in Fig. 3a and obtained from the DMM reading, is limited by the settings of the DMM, which are a trade-off between sampling time and filtering. The effective stability in the comparison block is shown in Fig. 3d, which shows the improvement with respect to the old set-up, shown in Fig. 2. The influence of immersion was minimized by performing the measurements only at full immersion, which means between 40 cm and 45 cm, the bath validation uncertainty was also reduced by performing the measurement with 2 SPRT s at opposite positions in the comparison block and measured the SPRT s before and after the customer thermometer, providing in situ information about the stability and uniformity of the temperature in the comparison block. As the bath had two independent heaters, both the new controller and the default controller were used, using the default controller as booster (setpoint 0.5 C below t SP ). The uncertainty in the resistance ratio measurement was reduced from to by using an ASL-F18 instead of an ASL-700, and the uncertainty component for the selfheat was reduced by correcting for the selfheating effect. Consequence of using an ASL-F18, performing selfheat correction and using 2 SPRT s, was the longer calibration time: the measurement time increased from less than 5 min to more than 30 min. 3 Validation of the Facility Using Calibrated SPRT s For the validation of the facility, 6 thermometers (see Table 2) were used and all were calibrated at fixed points in the range Hg to Ga or Ar to Ga. Two thermometers, R3416 and R3194, were used as reference and measured the reference temperature, t 90, and the other four thermometers were regarded as units under test (UUT). The

5 Int J Thermophys (2017) 38:37 Page 5 of 8 37 Fig. 3 Stability, t PV t SP, in the bath (a) over the range from 0 Cto30 C together with the cooling bath temperature and offset, t 90 t PV,Cooler (b) and the control power, P, (c) and the temperature stability, t 90 t SP, in the comparison block (d) Table 2 Overview of the thermometers used for the validation of the new sub-millikelvin set-up with calibration range and uncertainty, u(t) Identification Manufacturer Type Serial number Calibration u(t) range mk R3416 Rosemount 162CE 3416 Hg Ga 0.25 R3194 Rosemount 162CE 3194 Hg Ga 0.20 LN4240 Leeds & Northrup Hg Ga 0.18 LN8869 Leeds & Northrup Hg Ga 0.18 R2649 Rosemount 162CE 2649 Ar Ga 0.42 T Tinsley 5187 SB Ar Ga 0.34

6 37 Page 6 of 8 Int J Thermophys (2017) 38:37 Table 3 Results of the validation of the sub-millikelvin set-up using the R3416 and R3194 as standard to obtain the reference temperature, t 90,and LN4240 as UUT. The uncertainty for the UUT is 0.18 mk t 90 u(t 90 ) t UUT t 90 E n R3416 & R3194 LN4240 C mk K K/K Fig. 4 Uncertainty contribution of the reference temperature including the standard deviation of the measurement and the comparator block stability and uniformity (a) and the E n -value for all UUT thermometers in the validation test (b) measured resistance, R UUT, was then transformed into temperature, t UUT,usingthe fixed point calibration coefficients, and the obtained temperature was compared with the temperature indication of the two reference thermometers. Table 3 shows the results for the measurement with LN4240 as UUT. The difference between the UUT and reference temperature, t UUT t 90, divided by the combined uncertainty of the UUT and the reference thermometers, E n, is smaller than 1 over the whole range. The uncertainty of the reference temperature, u(t 90 ), in Table 3 is a combination of the measurement standard deviation and temperature stability and uniformity in the comparison block, but is exclusive the calibration uncertainty of the reference standards. The contribution

7 Int J Thermophys (2017) 38:37 Page 7 of 8 37 Fig. 5 Sea-Bird SBE 3 (a) and SBE 35 (b) during the calibration in the water bath Fig. 6 Results of the Sea-Bird SBE 35 (a) and SBE 3 (b) calibration. The dotted lines show the uncertainty of the trend line is smaller than 0.2 mk over nearly the whole range. Only at 30 C, the uncertainty contribution is higher. Figure 4b shows that the E n -value for all UUT thermometers is smaller than 1 and that the facility is capable of calibrating thermometers with an uncertainty smaller than 1 mk (see Table 1).

8 37 Page 8 of 8 Int J Thermophys (2017) 38:37 4 Calibration of Deep Ocean Thermometers The two Sea-Bird Electronics thermometers were calibrated by comparison with the same two thermometers used during the validation, R3416 and R3194. Due to the short length of the SBE 3 sensor, this thermometer was completely submerged in the water bath (Fig. 5a) and the head of the SBE 35 was above the water (Fig. 5b), similar to the two SPRT s. The calibration was performed from 0 Cupto30 C in steps of 3 C(seeFig.6). The results for the SBE 35, shown in Fig. 6a, also include the fixed point measurements, which agree very well with the comparison measurements. The uncertainty of the linear interpolation, representing the reproducibility of the sensor, is 0.19 mk for the SBE 35 and 0.24 mk for SBE 3, and the total calibration uncertainty for the thermometers becomes 0.39 mk and 0.42 mk, respectively. 5 Conclusions Using a high-accuracy temperature controller, a facility was built for calibrations with sub-millikelvin uncertainty in a conventional calibration bath. The CMC for the facility is smaller than 0.7 mk (k = 2), which was confirmed through the validation using 6 SPRT s. The facility was used to calibrate two oceanographic thermometers, a Sea-Bird Electronics, SBE 3 and SBE 35, which were calibrated with a expanded uncertainty of 0.8 mk (k =2). Acknowledgements This work was performed in the frame of the European Metrology Research Programme (EMRP) Joint Research Project ENV58 MeteoMet2. The EMRP is jointly funded by the EMRP participating countries within EURAMET and the European Union. References 1. BIPM, techniques for approximating the international temperature scale of 1990, Paris, July (1990) 2. J. Bojkovski, V. Batagelj, J. Drnovšek, V. Žužek, XIX IMEKO World Congress, Fundamental and Applied Metrology, September 6-11, Lisbon, Portugal, pp (2009) 3. R. Bosma, R. Van Breugel, C. Bruin-Barendregt, A. Peruzzi, AIP Conf. Proc 1552, 480 (2013). doi: /

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:

More information

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM 83.8058 K TO 1234.93 K: METHODS AND UNCERTAINTIES Gregory F. Strouse National Institute of Standards and Technology, Gaithersburg, Maryland, USA ABSTRACT

More information

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam Cryostat for Fixed-Point Calibration of Capsule-Type SPRTs I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam International Journal of Thermophysics Journal of Thermophysical Properties and Thermophysics and Its

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon Tohru Nakano and Osamu Tamura National Metrology Institute of Japan (NMIJ), AIST Tsukuba, Japan

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the SMD (Belgium) and the BIPM October 2017 Final

More information

Calibration of temperature sensors within Length Standard Section of NMIJ

Calibration of temperature sensors within Length Standard Section of NMIJ Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1 Contents Temperature measurement system

More information

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators Michael Coleman Fluke Corporation, Temperature Calibration Division, American Fork, Utah, USA mike.coleman@fluke.com

More information

A calibration facility for automatic weather stations

A calibration facility for automatic weather stations METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic

More information

Combined method for establishment and dissemination of the international temperature scale

Combined method for establishment and dissemination of the international temperature scale DISSERTATIONES PHYSICAE UNIVERSITATIS TARTUENSIS 86 RIHO VENDT Combined method for establishment and dissemination of the international temperature scale DISSERTATIONES PHYSICAE UNIVERSITATIS TARTUENSIS

More information

EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-2000 AT NMi

EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-2000 AT NMi EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-000 AT NMi A. Peruzzi and M.J. de Groot Nmi van Swinden Laboratorium, Delft, The Netherlands 1 Abstract The Provisional Low Temperature Scale

More information

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN Michael Stock and Stéphane Solve Bureau International des Poids et Mesures (BIPM) Pavillon de Breteuil,

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Points between PTB and LATU with PTB as Pilot Laboratory O. Robatto, Departamento de Metrología Científica, Laboratorio Tecnológico del Uruguay(LATU),

More information

Measuring Systems for Thermometer Calibration in Low-Temperature Range

Measuring Systems for Thermometer Calibration in Low-Temperature Range Int J Thermophys (2011) 32:2466 2476 DOI 10.1007/s10765-011-1055-2 Measuring Systems for Thermometer Calibration in Low-Temperature Range A. Szmyrka-Grzebyk L. Lipiński H. Manuszkiewicz A. Kowal A. Grykałowska

More information

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 29, Lisbon, Portugal TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS Fumihiro Sakuma NMIJ, National Institute of Advanced

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Overview. TC Thermometry Graham Machin. Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists

Overview. TC Thermometry Graham Machin. Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists TC Thermometry Graham Machin Euramet GA 2 June 2015 Overview Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists Workshops COOMET TC-T links EMPIR project

More information

Towards SI traceable humidity measurements with radiosondes

Towards SI traceable humidity measurements with radiosondes VTT TECHNICAL RESEARCH CENTRE OF FINLAND LTD Centre for Metrology MIKES MIKES METROLOGY Towards SI traceable humidity measurements with radiosondes GRUAN ICM-9, MeteoMet Training Course FMI, Helsinki,

More information

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation to ISO/IEC 17025:2005 United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited

More information

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C DOI 10.1007/s10765-014-1687-0 Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski

More information

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Yun-Kyung Bae and In-Jik Jeong Industrial & Physical Instrument Center, Korea

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the NIMT (Thailand) and the BIPM December 016 Final

More information

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range -38.8344 C to 419.57 C Final report Prepared by. Joung (coordinator) and K. S. Gam Korea Research Institute of Standards and Science

More information

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR Davor Zvizdic, Tomislav Stasic, Lovorka

More information

Traceability of temperature measurements in Estonia

Traceability of temperature measurements in Estonia 116 Proceedings of the Estonian Academy of Sciences, 2013, 62, 2, 116 121 Proceedings of the Estonian Academy of Sciences, 2013, 62, 2, 116 121 doi: 10.3176/proc.2013.2.05 Available online at www.eap.ee/proceedings

More information

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities T. J. Wiandt 1,2 1 Fluke Corporation, Hart Scientific Division, American Fork, Utah United States. 2 E-mail: tom.wiandt@hartscientific.com

More information

Nanoholes for leak metrology

Nanoholes for leak metrology Vacuum Metrology for Industry Nanoholes for leak metrology Università Degli Studi di Genova, Italy OUTLINE INTRODUCTION FABRICATION OF NANOHOLES GEOMETRICAL CHARACTERIZATION LEAK DEVICES RESULTS: PTB INRIM

More information

Open Access Repository eprint

Open Access Repository eprint Open Access Repository eprint Terms and Conditions: Users may access, download, store, search and print a hard copy of the article. Copying must be limited to making a single printed copy or electronic

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH Hannelore Bloemink, Jan Bijma Royal Netherlands Meteorological Institute (KNMI) Instrumentation Division P.O. Box 201, 3730 AE De Bilt, The Netherlands

More information

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements Int J Thermophys (2009) 30:257 264 DOI 10.1007/s10765-008-0495-9 Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements L. Rosso N. Koneva V. Fernicola Published online: 12 August

More information

Temperature Measurement

Temperature Measurement MECE 3320 Measurements & Instrumentation Temperature Measurement Dr. Isaac Choutapalli Department of Mechanical Engineering University of Texas Pan American Introduction Temperature is one of the most

More information

Argon Triple-Point Device for Calibration of SPRTs

Argon Triple-Point Device for Calibration of SPRTs Int J Thermophys (2015) 36:229 239 DOI 10.1007/s10765-014-1768-0 Argon Triple-Point Device for Calibration of SPRTs B. Kołodziej H. Manuszkiewicz A. Szmyrka-Grzebyk L. Lipiński A. Kowal P. P. M. Steur

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis Thomas Wiandt Fluke Corporation, Hart Scientific Division NCSLI 2007 Session 8B 1 Introduction Uncertainty quantification is a primary

More information

Analysis of shock force measurements for the model based dynamic calibration

Analysis of shock force measurements for the model based dynamic calibration 8 th Worshop on Analysis of Dynamic Measurements May 5-6, 4 Turin Analysis of shoc force measurements for the model based dynamic calibration Michael Kobusch, Sascha Eichstädt, Leonard Klaus, and Thomas

More information

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES XVII IMEKO World Congress Metrology in the 3rd Millennium June 7, 003, Dubrovnik, Croatia MODEL OR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION O THERMOCOUPLES Georges Bonnier, Eliane Renaot, DavorZvizdic,

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

European Association of National Metrology Institutes

European Association of National Metrology Institutes European Association of National Metrology Institutes EURAMET GUIDELINES ON TEMPERATURE: Extrapolation of SPRT calibrations below the triple point of argon, 83.8058 K, and traceability in baths of liquid

More information

Final Report of COOMET.T-K7: Regional key comparison of water triple point cells (COOMET theme No 395/BY/07)

Final Report of COOMET.T-K7: Regional key comparison of water triple point cells (COOMET theme No 395/BY/07) Final Report of COOMET.T-K7: Regional key comparison of water triple point cells (COOMET theme No 395/BY/07) Belarusian State Institute of Metrology (BelGIM) Starovilensky trakt 93, 220053, Minsk, RB Dikun

More information

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017 Metrology Principles for Earth Observation: the NMI view Emma Woolliams 17 th October 2017 Interoperability Decadal Stability Radiometric Accuracy Identical worldwide Century-long stability Absolute accuracy

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators Journal of Physics: Conference Series PAPER OPEN ACCESS Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators To cite this article: S A Carvajal and C A Sánchez 08 J. Phys.:

More information

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION METROLOGY AND MEASUREMENT SYSTEMS VOL. XV, NUMBER 4 (2008) A. SZMYRKA-GRZEBYK 1, A. KOWAL 1, L. LIPIŃSKI 1, PETER P. M. STEUR 2, R. DEMATTEIS 2 1 Institute of Low Temperature and Structure Research PAN,

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

Scope of Accreditation For ISOLAB Inc. Calibration

Scope of Accreditation For ISOLAB Inc. Calibration Scope of Accreditation For ISOLAB Inc 6260 Hawthorne Dr Windsor, ON N8T 1J9 Pradip Jansari 519-948-8371 In recognition of a successful assessment to ISO/IEC 17025:2005 to the following and Measurement

More information

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT JO Favreau 1, E Georgin 1, B Savanier 1, A. Merlone 2 1 CETIAT, 96100 Villeurbanne, France 2 INRIM, Torino, Italy Abstract. CETIAT is a calibration laboratory

More information

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

Available online at  ScienceDirect. Andrej Godina*, Bojan Acko Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 014 ) 191 198 4th DAAAM International Symposium on Intelligent Manufacturing and Automation, 013 Measurement Uncertainty

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Welcome to the Isotech Journal of Thermometry Fundamentals of Thermometry Part 1 Temperature Scales Practical Calibration of Thermometers

More information

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia International Temperature Scale of 1990 From Wikipedia, the free encyclopedia The International Temperature Scale of 1990 (ITS-90) published by the Consultative Committee for Thermometry (CCT) of the International

More information

How do you really know what the temperature is? Michael de Podesta

How do you really know what the temperature is? Michael de Podesta How do you really know what the temperature is? Michael de Podesta TECO: Madrid September 2016 GOLDEN RULE OF TALKS One talk: one thing This Talk: TWO THINGS! Thing 1: The definitions of the SI units of

More information

PRECISION MEASUREMENT OF SELF-HEATING EFFECT OF RESISTANCE THERMOMETERS INSTALLED ON MATERIAL ARTIFACTS

PRECISION MEASUREMENT OF SELF-HEATING EFFECT OF RESISTANCE THERMOMETERS INSTALLED ON MATERIAL ARTIFACTS XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 7 22, 2006, Rio de Janeiro, Brazil PRECISION MEASUREMENT OF SELF-HEATING EFFECT OF RESISTANCE THERMOMETERS INSTALLED ON MATERIAL

More information

Analysis of a national low dc resistance inter-laboratories comparison

Analysis of a national low dc resistance inter-laboratories comparison 16 th International Congress of Metrology, 04008 (2013) DOI: 10.1051/ metrology/201304008 C Owned by the authors, published by EDP Sciences, 2013 Analysis of a national low dc resistance inter-laboratories

More information

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE VALUATION OF TH ACCURACY OF A THRMOMTRIC RITANC BRIDG TO DIFFRNT VALU OF ATMOPHRIC PRUR Diego Almeida cuadorian tandards Institute, dalmeida@inen.gob.ec Abstract: The accuracy of a commercial thermometric

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Metrology and Environment

Metrology and Environment Bridging the Gap Responding to Environmental Change: from Words to Deeds 14-16 May 2008, Portorož Metrology and Environment Prof Janko Drnovšek and Laboratory group Janko.Drnovsek@fe.uni-lj.si Basic dilemma

More information

Bureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya

Bureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya Bilateral Comparison of 1.018 V and 10 V Standards between the KEBS (Kenya) and the BIPM January to March 2018 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by S. Solve +, R. Chayramy +,

More information

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Copyright 2013 by ABCM DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Amaury Roteger de Oliveira Renato Nunes Teixeira Instituto Nacional de Metrologia, Qualidade e Tecnologia - Inmetro Diretoria

More information

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to 273.16 K K. D. Hill 1, T. Nakano 2, P. Steur 3 1 Institute for National Measurement Standards,

More information

Maximum Performance From Dry-Well Thermometer Calibrators. Mark Finch Product Manager EMEA Fluke Calibration

Maximum Performance From Dry-Well Thermometer Calibrators. Mark Finch Product Manager EMEA Fluke Calibration Common Problems in Achieving Maximum Performance From Dry-Well Thermometer Calibrators Mark Finch Product Manager EMEA Fluke Calibration mark.finch@fluke.com Dry-well Thermal Uncertainties Temperature

More information

Measurements & Instrumentation. Module 3: Temperature Sensors

Measurements & Instrumentation. Module 3: Temperature Sensors Measurements & Instrumentation PREPARED BY Academic Services Unit August 2013 Institute of Applied Technology, 2013 Module Objectives Upon successful completion of this module, students should be able

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

Euramet project 1075 Bilateral comparison

Euramet project 1075 Bilateral comparison Bilateral comparison Between the gas flow standards of the national Metrology Institutes of Bulgaria (BIM-NCM) and The Netherlands (VSL) Copyright VSL B.V. Dutch Metrology Institute Page 2 of 19 VSL Thijsseweg

More information

Comparison of primary standards for liquid micro flow rates

Comparison of primary standards for liquid micro flow rates Comparison of primary standards for liquid micro flow rates EURAMET project 1291 EURAMET Regional Supplementary Comparison EURAMET.M.FF-S7 Pilot VSL, Netherlands Peter Lucas Co-pilots LNE-CETIAT, France

More information

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010 Int J Thermophys (2010) 31:1360 1370 DOI 10.1007/s10765-010-0728-6 Design and Capabilities of the Temperature Control System for the Italian Experiment Based on Precision Laser Spectroscopy for a New Determination

More information

General Accreditation Guidance. Liquid-in-glass thermometers selection, use and calibration

General Accreditation Guidance. Liquid-in-glass thermometers selection, use and calibration General Accreditation Guidance Liquid-in-glass thermometers selection, use and calibration January 2018 Copyright National Association of Testing Authorities, Australia 2013 This publication is protected

More information

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany Guidelines on the Calibration of Temperature Block Calibrators Authorship and Imprint This document was developed by the EURAMET e.v., Technical Committee for Thermometry. Authors: Yves Hermier (LNE-INM,

More information

AN INHERENTLY LINEAR TRANSDUCER USING THERMAL SIGMA-DELTA MODULATOR

AN INHERENTLY LINEAR TRANSDUCER USING THERMAL SIGMA-DELTA MODULATOR XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal AN INHERENTLY LINEAR TRANSDUCER USING THERMAL SIGMA-DELTA MODULATOR Valter C. Rosa, Amauri Oliveira, Lígia

More information

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement http://dx.doi.org/10.5755/j01.eee.21.1.6834 ELEKTRONIKA IR ELEKTROTECHNIKA ISSN 1392-1215 VOL. 21 NO. 1 2015 Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

More information

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the

More information

1 Phone , fax , 2 Phone , fax ,

1 Phone , fax , 2 Phone , fax , 1 10 k high Precision transportable Setup to calibrate Multifunction Electrical instruments P.P. Capra 1 and F. Galliana 2 National Institute of Metrological Research, (INRIM) str. Cacce, 91 10135 (TURIN

More information

Introduction to metrology

Introduction to metrology Introduction to metrology Emma Woolliams 4 April 2017 http://www.emceoc.org About NPL Founded in 1900 World leading National Metrology Institute ~750 staff; 550+ specialists in Measurement Science plus

More information

World Meteorological Organization

World Meteorological Organization World Meteorological Organization CIMO/WIGOS EXPLORATORY WORKSHOP: IMPROVING SURFACE-BASED DATA QUALITY THOUGH IMPROVED STANDARDIZATION OF PRACTICES AND PROCEDURES Langen, Germany, 5-3 December 2014 Regional

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Ciudad Madero, Tamaulipas, México. C.P. 89520 (Hereinafter

More information

The in-flight calibration system for the airborne imager GLORIA

The in-flight calibration system for the airborne imager GLORIA CALCON Technical Conference Utah State University, Logan, UT, USA August 4, 04 F. Olschewski, A. Ebersoldt, F. Friedl-Vallon, B. Gutschwager 4, J. Hollandt 4, A. Kleinert, C. Piesch, C. Monte 4, P. Preusse

More information

Abstract. 1. Introduction

Abstract. 1. Introduction The importance of establishing a very-high temperature radiation thermometry measurement capability at the National Metrology Institute of South Africa (NMISA) Speaker: Dr Efrem K Ejigu National Metrology

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator Int J Thermophys (2012) 33:1568 1582 DOI 10.1007/s10765-011-1005-z Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator L. Lages Martins A. Silva Ribeiro J. Alves e Sousa

More information

From MeteoMet to MeteoMet2. and more. Andrea Merlone

From MeteoMet to MeteoMet2. and more. Andrea Merlone From MeteoMet to MeteoMet2 and more Presented by Tom Gardiner WP1 WP2 Upper air measurements: sensors and techniques Novel methods, instruments and measurements WP3 Traceable measurement and protocols

More information

Traceability of on-machine measurements under a wide range of working conditions

Traceability of on-machine measurements under a wide range of working conditions Traceability of on-machine measurements under a wide range of working conditions Frank Keller, Matthias Franke, Norbert Gerwien, Lisa Groos, Christian Held, Klaus Wendt Traceable in-process dimensional

More information

Harnessing the Power of Arduino for the Advanced Lab

Harnessing the Power of Arduino for the Advanced Lab P P Herbert Jaeger + Harnessing the Power of Arduino for the Advanced Lab (Final Version) ALPhA Immersion Workshop July 27 29, 2017 Department of Physics Indiana University Purdue University Ft. Wayne,

More information

Experimental Assessment of Thermal Conductivity of a Brick Block with Internal Cavities Using a Semi-scale Experiment

Experimental Assessment of Thermal Conductivity of a Brick Block with Internal Cavities Using a Semi-scale Experiment Int J Thermophys (2013) 34:909 915 DOI 10.1007/s10765-012-1332-8 Experimental Assessment of Thermal Conductivity of a Brick Block with Internal Cavities Using a Semi-scale Experiment Zbyšek Pavlík Lukáš

More information

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS Final Report version 4 M. Heinonen, M. Anagnostou, J. Bartolo, S. Bell, R. Benyon, R. A. Bergerud, J. Bojkovski, N. Böse, V. Fernicola, K. Flakiewicz,

More information

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin Franco Pavese INRIM, Thermodynamic Division, Torino, Italy f.pavese@inrim.it F.Pavese 1 Temperature scales

More information

1 THE CONCEPT OF TEMPERATURE

1 THE CONCEPT OF TEMPERATURE 1 THE CONCEPT OF TEMPERATURE 1 1.1 Historical Perspective, 2 1.2 Early Definitions of Temperature, 9 1.3 A Simple Qualitative Definition of Temperature, 10 1.4 Units of Temperature for Various Temperature

More information

CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-90

CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-90 CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-9 Presenter: Thomas J. Wiandt Southern California Edison 73 Fenwick Lane Westminster, CA 92683 Telephone: (714) 895-443 Fax: (714) 895-686

More information

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity 14 th Quantitative InfraRed Thermography Conference Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity by I. Müller*, A. Adibekyan*, B. Gutschwager*, E.

More information

Develop and characterize a flow metering standard down to 1 nl/min

Develop and characterize a flow metering standard down to 1 nl/min Develop and characterize a flow metering standard down to 1 nl/min Martin Ahrens FH Lüebeck, Germany EMRP Researcher Grant HLT07- REG2 1 Project aims HLT07- REG2 develop and characterize of a flow metering

More information

ERROR SOURCE IDENTIFICATION AND STABILITY TEST OF A PRECISION CAPACITANCE MEASUREMENT SYSTEM

ERROR SOURCE IDENTIFICATION AND STABILITY TEST OF A PRECISION CAPACITANCE MEASUREMENT SYSTEM 106 SOUTH AFRICAN INSTITUTE OF ELECTRICAL ENGINEERS Vol.101(3) September 2010 ERROR SOURCE IDENTIFICATION AND STABILITY TEST OF A PRECISION CAPACITANCE MEASUREMENT SYSTEM S. Nihtianov* and X. Guo* # *

More information

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil A COMPARISON OF THE INITIAL UNCERTAINTY BUDGET AND AN ESTIMATE OF THE POSTERIOR UNCERTAINTIES,

More information

Realisation and study of a CO 2 triple point thermometric cell. Réalisation et étude d une cellule thermométrique au point triple du CO 2

Realisation and study of a CO 2 triple point thermometric cell. Réalisation et étude d une cellule thermométrique au point triple du CO 2 M.G. AHMED Realisation and study of a CO 2 triple point thermometric cell Réalisation et étude d une cellule thermométrique au point triple du CO 2 Mohamed Gamal AHMED 1, Khalid ALI 1 and Yves HERMIER

More information

The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results

The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results Online Supplementary Information for the paper: The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results Franco Pavese

More information

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples

More information

NEW VSL Conditioning plate!

NEW VSL Conditioning plate! Calibration of Prover Tanks using a Coriolis Mass Flow Meter as the Master Meter. Is this possible for Pipe Prover calibration too? Erik Smits (fsmits@vsl.nl) 2 nd European Flow Measurement Workshop Sintra,

More information

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard PTB has developed a test facility to characterize humidity sensors

More information