Primary Standardisation of I-125

Size: px
Start display at page:

Download "Primary Standardisation of I-125"

Transcription

1 Primary Standardisation of S. Pommé, T. Altzitzoglou, R. Van Ammel, G. Sibbens Institute for Reference Materials and Measurements (IRMM) Geel, Belgium Stefaan Pommé, Key Comparison Workshop, BIPM

2 : Seven methods Photon sum-peak counting with ai well detector Photon sum-peak counting with CsI(Tl) spectrometer Total emission counting with CsI(Tl) spectrometer -,γ coincidence counting with ai detectors (fixed or variable distance) electron,γ coincidence counting (PPC+aI well) electron,γ sum counting (PPC+aI well) Liquid scintillation counting

3 IM A544 (005) erratum IM A555 (005) 459 3

4 preparation sources quantitative drop deposition by pycnometer drop of 00 mg/l AgO 3 added to solution to precipitate volatile iodine source drying in laminar flow solid sources covered by 8-35 μg/cm VYS foil LSC: 5 ml Ultima Gold 5 ml Insta-Gel Plus 5 ml Ultima Gold + ml water 4

5 photon sum-peak counting in a ai well ai 6 x6 well 5

6 ratio doubles/singles peak => counting efficiency - and γ-rays in ai(tl) well + A x = + A singles peak doubles peak x =69.3% 6

7 pileup => increases doubles peak - and γ-rays in ai(tl) well singles peak doubles peak 7

8 pile-up correction formula A A ρτ = + e A A corr A, A area of singles and doubles peak ρ=the measured count rate τ is the characteristic (leading edge) pulse width 8

9 pile-up correction vs. count rate.00 photon sum-peak counting with ai well.080 correction factor count rate (cps) 9

10 photon sum-peak counting by ai well : results Act (kbq/g) mass (g) 0

11 CsI(Tl) sandwich spectrometer

12 photon sum-peak counting by CsI(Tl) sandwich spectrometer PM tube Source CsI

13 cover source by plastic caps to stop the electrons CsI(Tl) crystal with semi-spherical cavity source with semi-spherical caps 3

14 photon sum-peak counting,γ K rays and γ-rays (~30keV) are being measured electrons are stopped e - 4

15 photon sum-peak method by CsI(Tl) sandwich spectrometer - and γ-rays in CsI(Tl) x =99.3% singles peak doubles peak 5

16 photon sum-peak counting results Act (kbq/g) source mass (g) 6

17 total emission counting,γ K rays and γ-rays and electrons are being measured e - 7

18 Total emission counting by CsI(Tl) sandwich spectrometer photons only - and γ-rays and electrons in CsI(Tl) singles peak doubles peak 8

19 peak ratio vs. count loss 3% loss 0.339Y+0.045Y % Y=A /A loss % 0% A /A 9

20 results total emission counting Act (kbq/g) count rate (s - ) 0

21 -,γ coincidence by two moveable ai detectors vary distance symmetrically

22 -,γ coincidence by two fixed ai detectors vary efficiency by sandwiching the source between absorber foils

23 at first sight: non-linear? Act (kbq/g) cf. H. Schrader and K.F. Walz, Int. J. Appl. Radiat. Isot. 38 (987) 763

24 4 direct solution instead of extrapolation c c c + + = { } x x 0 / + =

25 direct solution vs. extrapolation Act (kbq/g)

26 residuals dev (kbq/g) bias at large distances 6

27 alternative type of extrapolation + + c as a function of efficiency ratio Act(kBq/g) f= A / B 7

28 escape peak of 35 kev γ-ray in ai number of counts % of spectrum 9% of γ-ray s below threshold error of 0.03% on activity channel number 8

29 Tereshchenko et al., Measurement Techniques 49 (006) 60 9

30 4π Κ,γ coincidence counting by PPC and well ai(tl) γ ai 6 x6 well PPC coinc. c β source 30

31 3 iterative solution { } { } { } + = + = + = β γ e x e x 0 c e e 0 x x 0 E E E E E E E E { } { } e x e c x E E E E E E E E + + = β { } { } e x x c e E E E E + + = γ { } x x 0 / + = γ discriminator energy>0 kev β c x 3. x x e

32 3 x0 5 x0 4 x0 3 choice of threshold PPC L LAuger threshold level K Auger L+ ce channel number number of counts relative activity

33 extrapolation to zero gas pressure => eliminate photon detection in PPC Act (kbq/g) PPC gas pressure (0 5 Pa) 33

34 4π,γ sum counting by PPC and well ai(tl) 34

35 summing method aims for 00% efficiency 0 = β + γ c discriminator thresholds as low as possible all sources were covered by thin VYS foil to avoid contamination beta efficiency variation by natural variation of self-absorption two discriminator threshold settings 35

36 classical extrapolation 460 Act (kbq/g) parabolic (- β )/ β 36

37 alternative extrapolation total count loss {[ A ( ) + B( ) + P ]( η ) + P } x e L KL { C ) + D( )} L ( x e M + count loss in PPC = ( β 0) {[ A + B( ) + P ]( η ) + P } e L KL { C + D( )} e L M + quite linear correlation if e x (nearly) constant 37

38 extrapolation (this work) 300 Act (kbq/g) 60 0 linear b / 0 38

39 Liquid Scintillation Spectrometry Wallac Quantulus 0 ultra low-level liquid scintillation counter picture PTB 39

40 Liquid Scintillation Spectrometry I spectrum Energy (kev) Packard 300 TR/AB 40

41 Calculated uclide Efficiency 86 CIEMAT-IST C004 (PTB) 84 efficiency (%) Cocktail = 5mL Ultima Gold + ml H O tsie 4

42 Results 5 I by LSC activity concentration (kbq/g) should be independent of quenching tsie 4

43 Results 5 I KC ±0.6% Act (kbq/g) % 95 CsI(Tl) CsI(Tl) caps b-g sum g sum b-g coinc - coinc LSC 43

Activity measurement of 55 Fe within the scope of the BIPM comparison 2006

Activity measurement of 55 Fe within the scope of the BIPM comparison 2006 Activity measurement of 55 Fe within the scope of the BIPM comparison 2006 Karsten Kossert PTB, Department 6.1 Radioactivity BIPM Workshop 2 on CCRI(II) Activity Uncertainties and Comparisons, 17-18 September

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES 1 BUREAU INTERNATIONAL DES POIDS ET MESURES International comparison of activity measurements of a solution of 3 H (January 2009) Participating laboratory: T ½ = (4 496.862 d; u = 9.131 d)* Ampoule number

More information

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method A Ruzzarin 1, P A L da Cruz 2, A L Ferreira Filho 2, A Iwahara 2 1 Laboratório de Instrumentação Nuclear/Programa

More information

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting Justyna Marganiec-Galazka Ole J. Nähle Karsten Kossert Division 6 Ionizing Radiation Department 6.1 Radioactivity Working Group

More information

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples Mashaba, Machel. 1*, Kotze, Deon. 2, Tshivhase, Victor.M. 1, Faanhof, Arnaud. 1, 2 1

More information

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France The TDCR method in LSC P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France LIQUID SCINTILLATION USERS FORUM 2009 Summary I. Some information on LSC II. LSC in metrology: the free parameter

More information

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua A New Measurement Method For Tritium Rapid Detection by LSC IM.CAEP Xu Qinghua xuqinghuawork@sina.com Outline Background Results and discussion Materials and methods Background FOM value for channel Sample

More information

The metrology of radioactivity. C. Michotte, BIPM

The metrology of radioactivity. C. Michotte, BIPM The metrology of radioactivity C. Michotte, BIPM SUMMARY Introduction Nuclear data of interest in activity measurements Activity measurement methods (primary and secondary) Monte Carlo simulations The

More information

Basics of a/b Discrimination for Liquid Scintillation Counting

Basics of a/b Discrimination for Liquid Scintillation Counting application Note Liquid Scintillation Basics of a/b Discrimination for Liquid Scintillation Counting Author Ronald Edler, PhD and Chuck Passo PerkinElmer, Inc. 940 Winter Street Waltham, MA USA Introduction

More information

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer 1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer PerkinElmer LAS (UK) Ltd, Chalfont Rd, Seer Green, Beaconsfield, Bucks HP9 2FX tel: 0800 896046 www.perkinelmer.com John Davies January

More information

Fundamentals of Radionuclide Metrology

Fundamentals of Radionuclide Metrology Fundamentals of Radionuclide Metrology Brian E. Zimmerman, PhD Physical Measurement Laboratory National Institute of Standards and Technology Gaithersburg, MD USA SIM Metrology Workshop Buenos Aires, Argentina

More information

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER Pall Theodórsson Science Institute, University of Iceland, Dunhaga 3, IS-107 Reykjavík, Iceland. Email: pth@raunvis.hi.is.

More information

Time-Resolved Liquid Scintillation Counting

Time-Resolved Liquid Scintillation Counting CHAPTER 44 Time-Resolved Liquid Scintillation Counting Norbert Roessler, Robert J. Valenta, and Stat van Cauter ABSTRACT A comparison is made between standard, two-tube coincidence liquid scintillation

More information

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

LIQUID SCINTILLATION COUNTERS. {Beta Counters} LIQUID SCINTILLATION COUNTERS {Beta Counters} We offer a range of Alpha, Beta & Gama counters, from Hidex Oy, Finland to meet your specific Liquid Scintillation counting requirements. Triathler LSC Sense

More information

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY K Kossert 1 O Ott O Nähle Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116

More information

SCINTILLATION DETECTORS AND PM TUBES

SCINTILLATION DETECTORS AND PM TUBES SCINTILLATION DETECTORS AND PM TUBES General Characteristics Introduction Luminescence Light emission without heat generation Scintillation Luminescence by radiation Scintillation detector Radiation detector

More information

Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples

Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples a new procedure for Alpha/Beta determination in water samples Bouchra Aboudou, Hector Bagán, Alex Tarancón Sanz, Gemma Rauret, Jose Francisco García Departmento de Química Analitica de la Universidad de

More information

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS International Conference Nuclear Energy in Central Europe 2001 Hoteli Bernardin, Portorož, Slovenia, September 10-13, 2001 www: http://www.drustvo-js.si/port2001/ e-mail: PORT2001@ijs.si tel.:+ 386 1 588

More information

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices Week 7: Ch. 0 Spec. w/ Scintillation Ctrs. multiplier Devices Spectroscopy with Scint. Counters -- gamma-ray interactions, reprise -- observed spectra --- spectral components, backscatter --- summing --

More information

Applied Radiation and Isotopes

Applied Radiation and Isotopes Applied Radiation and Isotopes 70 (2012) 2091 2096 Contents lists available at SciVerse ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Disintegration rate

More information

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters Validation of detection efficiency of alpha particles in commercial liquid scintillation counters Liquid Scintillation Users' Forum 14th October 2008 Andy Pearce Radionuclide and Neutron Metrology Group

More information

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Unit 2. Instrumentation. Experts Teaching from Practical Experience Unit 2 Instrumentation Experts Teaching from Practical Experience Gas-Filled Detectors Gas-filled detectors measure the charge released when radiation interacts with the gas Three types: Ion Chambers,

More information

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM CHAPTER 43 Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM Lauri Kaihola ABSTRACT Low-potassium glass vials can obtain reduced background count rates for beta

More information

Euromet 907 124 Sb Activity measurement and determination of photon emission intensities Part A Euromet 907 Results 1. Introduction 2. Decay scheme of antimony 124 3. Programme and procedure 3.1. General

More information

Čerenkov counting and liquid scintillation counting of 36 Cl

Čerenkov counting and liquid scintillation counting of 36 Cl Čerenkov counting and liquid scintillation counting of 36 Cl Karsten Kossert, Ole Nähle Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany and Agustín Grau Carles Instituto de Física Fundamental

More information

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome AGENDA FOR 5th VERMI YOUNG RESEARCHERS WORKSHOP ON STANDARDISATION OF S in the frame of IPA Turkey* 1 6 November 2009, TAEK, Ankara, Turkey Sunday, 1 November 2009 19:00 21:30 Registration and reception

More information

Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute

Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute Jost Eikenberg, Maya Jäggi, Max Rüthi Paul Scherrer Institute, CH-5232 Villigen, Switzerland Content

More information

ARTICLE IN PRESS. Applied Radiation and Isotopes

ARTICLE IN PRESS. Applied Radiation and Isotopes Applied Radiation and Isotopes 68 (21) 1349 1353 Contents lists available at ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Standardization and measurement

More information

LSUF Update on Organically Bound Tritium Standard project. Arvic Harms. 5 May 2004

LSUF Update on Organically Bound Tritium Standard project. Arvic Harms. 5 May 2004 LSUF 2004 Update on Organically Bound Tritium Standard project Arvic Harms 5 May 2004 Contents What is tritium and what is organically bound tritium (OBT) and why is it important? Why do we need an OBT

More information

He-3 Neutron Detectors

He-3 Neutron Detectors Application He-3 Neutron Detectors General Considerations, Applications: He-3 filled proportional counters are standard neutron detectors and are most suitable for the detection of thermal neutrons. Larger

More information

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France TDCR in a nutshell P. Cassette, Laboratoire National Henri Becquerel, France Summary LSC in radionuclide metrology, free parameter model The TDCR model Examples of relations between efficiency and TDCR

More information

APPLICATION OF A FREE PARAMETER MODEL TO PLASTIC SCINTILLATION SAMPLES

APPLICATION OF A FREE PARAMETER MODEL TO PLASTIC SCINTILLATION SAMPLES APPLICATION OF A FR PARAMTR MODL TO PLASTIC SCINTILLATION SAMPLS Alex Tarancón Sanz 1, Hector Bagan 1, Karsten Kossert 2, Ole Nähle 2 1 Departmento de Química Analitica de la Universidad de Barcelona.

More information

Standardization of 18 F by Digital β(ls)-γ Coincidence Counting

Standardization of 18 F by Digital β(ls)-γ Coincidence Counting Standardization of 18 F by Digital β(ls)-γ Coincidence Counting Rodrigues D. 1, Balpardo C. 1, Cassette P. 2, Arenillas P. 1, Capoulat M. E. 1, Ceruti G. 1, García-Toraño E. 3 1 Laboratorio de Metrología

More information

Monte Carlo simulations as a tool for the development of a new reference ionisation chamber

Monte Carlo simulations as a tool for the development of a new reference ionisation chamber Monte Carlo simulations as a tool for the development of a new reference ionisation chamber Johan Camps, Jan Paepen, Stefaan Pommé and Uwe Wätjen Institute for Reference Materials and Measurements (IRMM)

More information

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second Contributors p. xxix Acronyms, Abbreviations, and Symbols p. xxxi Foreword to the First Edition p. xliii Foreword to the Second Edition p. xlv Preface to the First Edition p. xlvii Preface to the Second

More information

CCRI(II) activity comparison of 241 Pu CCRI(II)-K2.Pu-241

CCRI(II) activity comparison of 241 Pu CCRI(II)-K2.Pu-241 Metrologia 49 (2012) Tech. Suppl. 06012 CCRI(II) activity comparison of 241 Pu CCRI(II)-K2.Pu-241 C. Michotte 1, L. Johansson 2 1 BIPM and 2 NPL Abstract In 2010, seven laboratories took part in the CCRI(II)

More information

Radioactivity standardization in South Africa

Radioactivity standardization in South Africa Applied Radiation and Isotopes 56 (2002) 301 305 Radioactivity standardization in South Africa B.R.S. Simpson* Radioactivity Standards Laboratory, CSIR NML, 15 Lower Hope Road, Rosebank 7700, Cape Town,

More information

TECHNETIUM-99 IN SOIL

TECHNETIUM-99 IN SOIL Analytical Procedure TECHNETIUM-99 IN SOIL 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in soil. 1.2. This method does not address all aspects of safety, quality

More information

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Radiation Detection. 15 th Annual OSC Readiness Training Program. Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all

More information

Results from the 237 Np exercise. EUROMET action 416

Results from the 237 Np exercise. EUROMET action 416 Results from the 237 Np exercise EUROMET action 416 D. Smith, M. J. Woods, D. H. Woods (based on contributions from the participants to Euromet action 416) Centre for Ionising Radiation Metrology National

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (TEVA DISC METHOD) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all aspects

More information

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING W. Hennig 1, H. Tan 1, A. Fallu-Labruyere 1, W. K. Warburton 1, J. I. McIntyre 2, A. Gleyzer 3 XIA, LLC 1, Pacific Northwest National Laboratory

More information

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson RADIOCARBON, Vol 45, Nr 1, 2003, p 113 122 2003 by the Arizona Board of Regents on behalf of the University of Arizona BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING P TheodÛrsson 1 S Ingvarsdottir

More information

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON W. Hennig 1, H. Tan 1, W.K. Warburton 1, and J.I. McIntyre 2 XIA LLC 1, Pacific Northwest

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Determination of 210 Pb and 210 Po in Water Samples

Determination of 210 Pb and 210 Po in Water Samples 1 Determination of 210 Pb and 210 Po in Water Samples Marin Ayranov 1, Zornitza Tosheva 2, Antoine Kies 2 1 Institute for Nuclear Research and Nuclear Energy, 72 Tzarigradsko chaussee, BG-1784 Sofia, Bulgaria

More information

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER V.A. MOROZOV 1, N.V. MOROZOVA 1, T. BĂDICĂ 2, D. DELEANU 2,3, D. GHIŢĂ 2, S. PASCU 2,3 1 Joint

More information

Liquid Scintillation at NPL: Practical Applications

Liquid Scintillation at NPL: Practical Applications 2003 LS User s Forum Liquid Scintillation at NPL: Practical Applications Andy Pearce CAIR, NPL 2 nd April 2003 Introduction Overview of applications Current projects Neutron-activated thioacetamide discs

More information

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

TIME-RESOLVED LIQUID SCINTILLATION COUNTING [RADOCARBON, VOL 32, No. 3, 1990, P 381-386] TME-RESOLVED LQUD SCNTLLATON COUNTNG MCHAEL KESSLER Packard nstrument Company, One State Street, Meriden, Connecticut 06450 ABSTRACT. Historically, scientists

More information

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator 4, May, 2017 Jun Woo Bae, Ukjae Lee, Hee Reyoung Kim 1 Outline Introduction Material

More information

05 - Scintillation detectors

05 - Scintillation detectors 05 - Scintillation detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_05, Scintillation detectors Version 2 1 / 39 Scintillation detector principles

More information

Radioactivity and Ionizing Radiation

Radioactivity and Ionizing Radiation Radioactivity and Ionizing Radiation QuarkNet summer workshop June 24-28, 2013 1 Recent History Most natural phenomena can be explained by a small number of simple rules. You can determine what these rules

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies CHARACTERIZATION OF PHOSWICH WELL DETECTORS FOR RADIOXENON MONITORING Wolfgang Hennig 1, Hui Tan 1, William K. Warburton 1, Anthony Fallu-Labruyere 1, Konstantin Sabourov 1, Justin I. McIntyre 2, Matthew

More information

Development of an active scintillating target for fission studies. J. Aupiais, G. Bélier CEA, DAM, DIF, F Arpajon France

Development of an active scintillating target for fission studies. J. Aupiais, G. Bélier CEA, DAM, DIF, F Arpajon France Development of an active scintillating target for fission studies J. Aupiais, G. Bélier CEA, DAM, DIF, F-91297 Arpajon France 1 Motivations Fast neutron induced reactions are important for future nuclear

More information

2010 Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

2010 Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN AND MODELING OF A COMPTON-SUPPRESSED PHOSWICH DETECTOR FOR RADIOXENON MONITORING Abi T. Farsoni and David M. Hamby Oregon State University Sponsored by the National Nuclear Security Administration

More information

GEMMA RAURET, J S MESTRES and J F GARCIA

GEMMA RAURET, J S MESTRES and J F GARCIA [RADIOCARBON, VOL31, No. 3, 1989, P 380-3861 OPTIMIZATION OF LIQUID SCINTILLATION COUNTING CONDITIONS WITH TWO KINDS OF VIALS AND DETECTOR SHIELDS FOR LOW-ACTIVITY RADIOCARBON MEASUREMENTS GEMMA RAURET,

More information

Standardization of Tritium by CIEMAT/NIST Method and TDCR Method

Standardization of Tritium by CIEMAT/NIST Method and TDCR Method Standardization of Tritium by CIEMAT/NIST Method and TDCR Method Wu Yongle 1,3 ; Liu Haoran 1,2 ; Liang Juncheng 2 ; Liu Jiacheng 2 ; Yue Huiguo 3 ; Liu Senlin 1 ; Yang Yuandi 2 ; Yuan Daqing 1 ; 1.China

More information

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy Thomas Niedermayr, I. D. Hau, S. Terracol, T. Miyazaki, S. E. Labov and S. Friedrich Former colleagues: M. F. Cunningham, J. N.

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods Journal of Physics: Conference Series PAPER OPEN ACCESS Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods To cite this article: A Ruzzarin et al 2016 J. Phys.: Conf.

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 2 Radioisotopes Techniques Lecture - 3 GM Counting and

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC)

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC) Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC) Bob Read, Ph.D. Director, Environmental Chemistry Laboratory Tennessee Department of Health Division of Laboratory Services

More information

Preparatory experiments for cold-neutron induced fission studies at IKI

Preparatory experiments for cold-neutron induced fission studies at IKI Preparatory experiments for cold-neutron induced fission studies at IKI A. Oberstedt 1, S. Oberstedt 2, R. Billnert 1, J. Karlsson 1, X. Ledoux 3, J.-G. Marmouget 3 and F.-J. Hambsch 2 1 School of Science

More information

Researchers at the University of Missouri-Columbia have designed a triple crystal

Researchers at the University of Missouri-Columbia have designed a triple crystal Childress, N. L. and W. H. Miller, MCNP Analysis and Optimization of a Triple Crystal Phoswich Detector, Nuclear Instruments and Methods, Section A, 490(1-2), 263-270 (Sept 1, 2002). Abstract Researchers

More information

Dario Barberis Evaluation of GEANT4 Electromagnetic and Hadronic Physics in ATLAS

Dario Barberis Evaluation of GEANT4 Electromagnetic and Hadronic Physics in ATLAS Dario Barberis Evaluation of GEANT4 Electromagnetic and Hadronic Physics in ATLAS LC Workshop, CERN, 15 Nov 2001 Dario Barberis Genova University/INFN 1 The ATLAS detector LC Workshop, CERN, 15 Nov 2001

More information

Application of positrons in materials research

Application of positrons in materials research Application of positrons in materials research Trapping of positrons at vacancy defects Using positrons, one can get defect information. R. Krause-Rehberg and H. S. Leipner, Positron annihilation in Semiconductors,

More information

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator 88 The 1 st NPRU Academic Conference SCI-O11 Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator P. Limkitjaroenporn and W.Chewpraditkul Radiation Physics Laboratory,

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

Waveform Analysis for DM-Ice17. Zachary Pierpoint University of Wisconsin - Madison October 21, 2013 Yale Weak Interactions Discussions Group

Waveform Analysis for DM-Ice17. Zachary Pierpoint University of Wisconsin - Madison October 21, 2013 Yale Weak Interactions Discussions Group Waveform Analysis for DM-Ice17 Zachary Pierpoint University of Wisconsin - Madison October 21, 213 Yale Weak Interactions Discussions Group DM-Ice17 Energy Spectrum counts / day / kev / kg 1 28 Tl+ 214

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO Wolfango Plastino Lauri Kaihola Paolo Bartolomei 3 Francesco

More information

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES V. MAIOROV, A. CIURAPINSKI, W. RAAB and V. JANSTA Safeguards Analytical Laboratory, International Atomic Energy

More information

Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer

Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer NUKLEONIKA 2010;55(2):155 161 ORIGINAL PAPER Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer Andrzej Komosa, Katarzyna Ślepecka Abstract.

More information

Detectors for the measurement of ionizing radiation

Detectors for the measurement of ionizing radiation For the measurement of radiation, the following reactions during the irradiation of matter are predominantly utilized: Ionization in gases (Ionization chamber, proportional flow counter, release counter)

More information

Dario Barberis Evaluation of GEANT4 electromagnetic physics in ATLAS

Dario Barberis Evaluation of GEANT4 electromagnetic physics in ATLAS Dario Barberis Evaluation of GEANT4 electromagnetic physics in ATLAS G4 Workshop, Genova, 5 July 2001 Dario Barberis Genova University/INFN 1 The ATLAS detector G4 Workshop, Genova, 5 July 2001 Dario Barberis

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

Large area scintillation detector for dosimetric stand with improved light collection

Large area scintillation detector for dosimetric stand with improved light collection NUKLEONIKA 2011;56(4):317 321 ORIGINAL PAPER Large area scintillation detector for dosimetric stand with improved light collection Bronisław Machaj, Jan Mirowicz, Ewa Kowalska Abstract. In order to improve

More information

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa.

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa. Gross Alpha and Beta Measurements of Water Samples from the Wonderfonteinspruit Catchment Area in the Gauteng Province (South Africa), using Liquid Scintillation Counting Machel Mashaba 1, Deon Kotze 2,

More information

Jerde, Peterson and Stein (1967) indicated that high-energy radiation

Jerde, Peterson and Stein (1967) indicated that high-energy radiation [RADIOCARBON, VOL 31, No. 3, 1989, P 332-341] LOW BACKGROUND LIQUID SCINTILLATION COUNTING USING AN ACTIVE SAMPLE HOLDER AND PULSE DISCRIMINATION ELECTRONICS JOHN E NOAKES Center for Applied Isotope Studies

More information

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA Determination of Components of Fuel Matrix in Water and in Bottom Slimes in the MR Reactor Ponds in NRC Kurchatov Institute 14038 Alexey Stepanov *, Iurii Simirskii *, Ilya Semin *, Anatoly Volkovich *

More information

Scintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003

Scintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003 Scintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003 Takayuki Yanagida, Hiromitsu Takahashi, Daisuke Kasama, Takeshi Ito, Hisako Niko, Motohide Kokubun, and

More information

Development and test of the high-precision -beam monitor at the ESFRI facility ELI-NP

Development and test of the high-precision -beam monitor at the ESFRI facility ELI-NP Development and test of the high-precision -beam monitor at the ESFRI facility ELI-NP Göök A., Oberstedt S., Geerts W., Hambsch F.-J. Public 2017 EUR 29002 EN This publication is a Technical report by

More information

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007 XENON Dark Matter Search Juliette Alimena Columbia University REU August 2 nd 2007 Evidence of Dark Matter Missing mass in Coma galaxy cluster (Fritz Zwicky) Flat rotation curves of spiral galaxies (Vera

More information

Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m. C. Michotte 1, R.

Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m. C. Michotte 1, R. Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m C. Michotte 1, R. Fitzgerald 2 1 Bureau International des Poids et Mesures (BIPM), 2 National

More information

Phase stability and lithium loading capacity in a liquid scintillation cocktail

Phase stability and lithium loading capacity in a liquid scintillation cocktail Phase stability and lithium loading capacity in a liquid scintillation cocktail Denis E. Bergeron, H. Pieter Mumm, Mark Tyra Physical Measurement Laboratory, National Institute of Standards and Technology,

More information

RESULTS YOU CAN COUNT ON

RESULTS YOU CAN COUNT ON RESULTS YOU CAN COUNT ON Liquid Scintillation Counting Principle and the Application in Biological Research 1 2009 PerkinElmer liquid scintillation counting β particle Tri-Carb family of liquid scintillation

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber CYRIC Annual Report 2001 V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber Kawata N. Baba M. Kato M.*, Miura T.**, and Yamadera A.***, Cyclotron and Radioisotope Center, Tohoku

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

08 - Miscellaneous and historical detectors

08 - Miscellaneous and historical detectors 08 - Miscellaneous and historical detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_08, Miscellaneous and historical detectors Version 2 1 / 25 Streamer

More information

Application Note. Abstract. Introduction. Experimental

Application Note. Abstract. Introduction. Experimental Application Note Alpha/Beta ABA-004 Quantitation of Transuranium Elements in High Activity Waste (HAW) Using Alpha/Beta Liquid Scintillation Counting and Extractive Scintillators by Jim Floeckher Abstract

More information

arxiv:physics/ v1 3 Aug 2006

arxiv:physics/ v1 3 Aug 2006 Gamma Ray Spectroscopy with Scintillation Light in Liquid Xenon arxiv:physics/6834 v1 3 Aug 26 K. Ni, E. Aprile, K.L. Giboni, P. Majewski, M. Yamashita Physics Department and Columbia Astrophysics Laboratory

More information

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals Wolfgang Hennig, Hui Tan, William K Warburton, and Justin I McIntyre

More information

Simulation of 4π HPGe Compton-Suppression spectrometer

Simulation of 4π HPGe Compton-Suppression spectrometer Vol. 9(2), pp. 13-19, 30 January, 2014 DOI: 10.5897/IJPS2013.4075 ISSN 1992-1950 2014 Academic Journals http://www.academicjournals.org/ijps International Journal of Physical Sciences Full Length Research

More information

Liquid Scintillation Counter

Liquid Scintillation Counter Instrumentation & Methods: s & Tritium Richard Sheibley Pennsylvania Dept of Env Protection Principle Beta particle emission Energy transferred to Solute Energy released as UV Pulse Intensity proportional

More information

Measurement of the carbon 14 activity at natural level in air samples

Measurement of the carbon 14 activity at natural level in air samples Radioprotection, Suppl. 1, vol. 40 (2005) S791-S796 EDP Sciences, 2005 DOI: 10.1051/radiopro:2005s1-116 Measurement of the carbon 14 activity at natural level in air samples A. Olivier 1, L. Tenailleau

More information