Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer

Size: px
Start display at page:

Download "Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer"

Transcription

1 Anal. Chem. 2006, 78, Accelerated Articles Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer Alexander Makarov,* Eduard Denisov, Alexander Kholomeev, Wilko Balschun, Oliver Lange, Kerstin Strupat, and Stevan Horning Thermo Electron (Bremen) GmbH, Hanna-Kunath-Strasse 11, Bremen Germany Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/ s) with nominal mass resolving power of and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second. Increasing speed of chromatographic separation and complexity of analyzed mixtures provides a continuous impetus for the development of faster and simultaneously more intelligent and robust mass spectrometric detectors. Reliable separation and reliable identification of complex mixtures with multiple coeluting compounds necessarily require higher resolving power, while structural analysis by MS/MS and accurate mass determination become complementary confirmation tools. * Corresponding author. Tel: +49(0) Fax: +49(0) alexander.makarov@thermo.com. A typical example of such a mass spectrometer is the quadrupole/time-of-flight (TOF) hybrid, 1,2 wherein precursor ions are selected by a quadrupole mass filter and accurate mass determination (including analysis of fragment ions) is carried out in an orthogonal-acceleration TOF. Though very successful for a range of applications, such hybrids nevertheless suffer from low ion transmission (resulting in poor MS/MS sensitivity and detection limits) and a limited intensity range over which accurate mass data can be acquired. These shortcomings were overcome in a next-generation hybrid instrument, a linear ion trap/fourier transform ion cyclotron resonance (FTICR) mass spectrometer. 3 Like earlier FTICR hybrids, 4,5 this mass spectrometer accumulates ions externally to a superconducting magnet, but in addition, it combines high trapping capacity, MS n capabilities, and automatic gain control (AGC) of linear ion trap mass spectrometers 6 with the unsurpassed mass accuracy, dynamic range, and resolving power of FTICR mass spectrometers. The unprecedented specificity and quality of data from this combination necessitates the added complexity of the superconducting magnet. This inspired the quest for a magnetic field-free analyzer of comparable performance, which would be more compatible with the capacities of a typical laboratory. (1) Chernushevich, I. V.; Loboda, A. V.; Thomson, B. A. J. Mass Spectrom. 2001, 36, (2) Morris, H. R.; Paxton, T.; Dell, A.; Langhorne, J.; Berg, M.; Bordoli, R. S.; Hoyes, J.; Bateman, R. H. Rapid Commun. Mass Spectrom. 1996, 10, (3) Syka, J. E. P.; Marto, J. A.; Bai, D. L.; Horning, S.; Senko, M. W.; Schwartz, J. C.; Ueberheide, B.; Garcia, B.; Busby, S.; Muratore, T.; Shabanowitz, J.; Hunt, D. F. J. Proteome Res. 2004, 3, (4) Senko, M. W.; Hendrickson, C. L.; Emmett, M. R.; Shi, S. D.-H.; Marshall, A. G. J. Am. Soc. Mass Spectrom. 1997, 8, (5) Belov, M. E.; Nikolaev, E. N.; Anderson, G. A.; Udseth, H. R.; Conrads, T. P.; Veenstra, T. D.; Masselon, C. D.; Gorshkov, M. V.; Smith, R. D. Anal. Chem. 2001, 73, (6) Schwartz, J. C.; Senko, M. W.; Syka, J. E. P. J. Am. Soc. Mass Spectrom. 2002, 13, /ac CCC: $ American Chemical Society Analytical Chemistry, Vol. 78, No. 7, April 1, Published on Web 02/18/2006

2 From its first appearance as a proof-of-principle device, 7 the orbitrap mass analyzer was considered as a potential alternative for FTICR because of its high resolution and good mass accuracy. Using the principle of orbital trapping in electrostatic fields, 8 the orbitrap consists of an inner and an outer electrode, which are shaped to create a quadro-logarithmic electrostatic potential. Ions rotate about the inner electrode and oscillate harmonically along its axis (the z-direction) with a frequency characteristic of their m/z values. An image current transient of these oscillations is converted to a frequency spectrum using a Fourier transform similar to the approach used in FTICR. For a given spread of initial parameters of ions, the orbitrap provides the best performance within the smallest dimensions, relative to other types of electrostatic traps, 8 thus making it the most suitable candidate for image current detection and Fourier transform mass spectrometry. Initially, high resolving power, internal mass accuracy, and high space charge capacity were demonstrated with the orbitrap mass analyzer using pulsed ion sources. Because of high ion velocities and the absence of any collisional cooling inside the orbitrap, injection must occur in a short burst, on a time scale well below 1 ms (actually, only a few microseconds if coherence of ion packets is to be provided by the injection, thereby eliminating the need for additional excitation 9 ). For continuous ion sources, this is possible only when additional ion storage is introduced to store ions and then inject them rapidly into the orbitrap. Such implementation for injection of electrosprayed ions into the orbitrap was initially based on axial ejection of ions from a linear rf-only quadrupole As the space charge capacity of such a trap is compromised by the need to extract ions within hundreds of nanoseconds, use of large ion numbers resulted in great variation of ion kinetic energies, together with angular and spatial spreads that limited mass range, transmission, dynamic range, and mass accuracy over a wide mass range. This work involves a new approach to ion storage that is based on orthogonal, rather than axial, ion ejection from an rf-only quadrupole. 11 This implementation alleviates the above problems and paves the way for a hybrid mass spectrometer combining the tandem mass spectrometry capability of the linear ion trap mass spectrometer with the high resolution and mass accuracy capability of the orbitrap. This combination allows high-quality accurate mass MS n spectra to be acquired using brief ion accumulation periods and relatively small ion populations. Here we characterize the analytical parameters and features of this hybrid instrument. EXPERIMENTAL SECTION Figure 1 presents a schematic diagram of the LTQ orbitrap hybrid mass spectrometer (Thermo Electron, Bremen, Germany). The design, operation, and control of the front-end LTQ mass spectrometer has been described elsewhere. 6 Briefly, ions from the electrospray ion source are admitted via rf-only multipoles into the linear trap of the LTQ, wherein ions are analyzed and radially ejected to a pair of secondary electron multipliers. 6 An important feature of the instrument is the procedure of AGC, 12 wherein a short prescan is used to determine the ion current within the mass range of interest, hence enabling storage of a defined number of ions ( AGC target value ) in the subsequent analytical scan. As in the LTQ FT mass spectrometer, all modes of LTQ operation remain available, but they are complemented by the ability to analyze ions in an additional mass detector (FTICR or, as in this work, the orbitrap). In the LTQ orbitrap, a transfer octapole (300 mm long, 400 V p-p rf, 5.7 mm inscribed diameter) delivers ions into a curved rf-only quadrupole whose central axis follows a C-shaped arc (hence the name C-trap). The C-trap uses rods with hyperbolic surfaces and is enclosed by two flat lenses with apertures for ion transport through them. The plate between the octapole and C-trap forms the gate electrode and the other plate forms the trap electrode. The enclosed volume is filled with nitrogen bath gas at 1 mtorr via an open-split interface connected to the nitrogen line of electrospray ion source. Nitrogen has been chosen as a bath gas in favor of helium due to better collisional damping and lower gas carryover toward the orbitrap. The C-trap assembly is pumped by the split-flow turbopump of the LTQ mass spectrometer, as indicated in Figure 1. Ions start from the back section of the linear trap held at a dc offset of 6-10 V (here and below voltages are given for positive ions). After acceleration into the transfer octapole by a 2-10 V potential difference, ions are transferred into the C-trap held at a dc offset of 0 V with an applied rf voltage of V p-p and are reflected by the trap electrode (12-15 V). The gate electrode is always kept at a positive offset of 3-6 V. On entry into the C-trap, ions lose energy in collisions with nitrogen bath gas, these collisions being mild enough to avoid any fragmentation. Due to the relatively low gas pressure and short length of the C-trap, ions need more than one pass through the entire system to be trapped. Nevertheless, ions finally come to rest at the place where gas collisions occur with the lowest dc offset along their path, i.e., in the C-trap. As the result of collisional cooling, the ions form a thin, long thread along the curved axis of the C-trap. This thread is compressed axially by applying 200 V to both the gate and the trap electrodes. After that, the rf voltage on the quadrupolar electrodes of the C-trap is rapidly ramped down (over ns) and dc pulses are applied to the electrodes as follows: 1200 V to the push-out electrode (i.e., the electrode furthest from the center of C-trap curvature), 1000 V to the pull-out electrode (the electrode closest to the center of curvature), and 1100 V to both the upper and lower electrodes. This voltage distribution forces ions orthogonally to the axis of the C-trap (center of curvature of the C-trap) where they leave via a slot in the pull-out electrode. Unlike axial ejection, 9,10 fast and uniform extraction is provided for large ion populations. After leaving the C-trap, the ions pass through appropriately curved ion optics, are accelerated to high kinetic energies, and converge into a tight cloud, which is able to pass through a small entrance aperture and enter the orbitrap tangentially. On their way from the C-trap, ions pass through three stages of differential pumping until they reach the ultrahigh vacuum compartment of (7) Makarov, A. A., Anal. Chem. 2000, 72, (8) Kingdon, K. H. Phys. Rev. 1923, 21, (9) Hardman, M. E.; Makarov, A. A. Anal. Chem. 2003, 75, (10) Hu, Q.; Noll, R.; Li, H.; Makarov, A. A.; Hardman, M. E.; Cooks, R. G. J. Mass Spectrom. 2005, 40, (11) Makarov, A. A.; Denisov, E.; Lange, O.; Kholomeev, A.; Horning, S. Proc. 53rd Conf. Am. Soc. Mass Spectrom., San Antonio, TX, June 5-9, 2005; Poster (12) Schwartz, J. C.; Zhou, X. G.; Bier, M. E. U.S. Patent 5,572, Analytical Chemistry, Vol. 78, No. 7, April 1, 2006

3 Figure 1. (A) Schematic layout of the LTQ orbitrap mass spectrometer: (a) Transfer octapole; (b) curved rf-only quadrupole (C-trap); (c) gate electrode; (d) trap electrode; (e) ion optics; (f) inner orbitrap electrode; (g) outer orbitrap electrodes. (B) Simplest operation sequence of the LTQ orbitrap mass spectrometer (not shown are the following: optional additional injection of internal calibrant; additional MS or MS n scans of linear trap during the orbitrap detection) the orbitrap sustained at mbar. The ion optics assembly is pumped by two 70 L/s TMH-71P turbopumps and the orbitrap by a 250 L/s TMU-262 turbopump (Pfeiffer Vakuum GmbH, Asslar, Germany). To avoid direct carryover of gas from the C-trap to the orbitrap, ions are displaced in the vertical direction using a dual electrostatic deflector. 9 The short transfer distance reduces deleterious time-of-flight separation and thus minimizes differences in intensity distributions between mass spectra acquired with the linear trap and orbitrap mass analyzers. Ions are captured in the orbitrap by rapidly increasing the electric field 7,9,10 and gradually spread into rotating thin rings oscillating axially along the inner electrode. The inner diameter of the orbitrap outer electrodes is 30 mm, and the maximum outer diameter of the inner electrode is 12 mm. Axial oscillations are initiated by injecting ions at an offset of 7.5 mm relative to the equator of the orbitrap, eliminating the necessity of any excitation. Ion packets are injected into the orbitrap with a duration much shorter than one axial oscillation, ensuring coherent motion. A deflector electrode near the entrance slit of the orbitrap is used to compensate fringing fields. Detection of image current from coherent ion packets takes place after the voltage on the inner electrode has been stabilized at 3.5 kv. Signals from each of the outer electrodes are amplified by a differential amplifier and transformed into a frequency spectrum by fast Fourier transformation. These frequencies relate to axial oscillations of ions along the orbitrap, which are independent of the energy and spatial spread of the ions. The typical frequency for m/z ) 524 is 300 khz. Single zero-filling and Kaiser-Bessel apodization 13 is used to improve the peak shape. The frequency spectrum is converted into a mass spectrum using a two-point calibration and processed (13) Goodner, K. L.; Milgram, K. E.; Williams, K. R.; Watson, C. H.; Eyler, J. R. J. Am. Soc. Mass Spectrom. 1998, 9, Analytical Chemistry, Vol. 78, No. 7, April 1,

4 approximately a s scan cycle time, wherein ion filling of the linear trap and ion transfer take a significant share of this time. Measured values of resolving power are always higher than the nominal value. Typical AGC target values N are for the linear trap detector and for the orbitrap detector. With the use of an electrometer, it was determined that the number N is actually quite close (within factor of 2) to the real number of ions stored in the linear trap. Mass calibration coefficients are determined for four different AGC target values and interpolated for intermediate values. No magnitude-dependent corrections of m/z are made for data processing. Initial calibration of the instrument is performed using the standard LTQ calibration mixture with caffeine, the peptide MRFA, and Ultramark 1600 dissolved in 50:50 v/v. water/acetonitrile solution. Bradykinin, bovine serum albumin, horse heart apomyoglobin, horse cytochrome c, and carbonic anhydrase proteins were obtained from Sigma-Aldrich Chemie GmbH (Munich, Germany) and used without further purification. A Finnigan Surveyor Plus liquid chromatograph (Thermo Electron, San Jose, CA) fitted with a Hypersil C18 column (2.1-mm i.d., 5-µm particles, Thermo Electron) was used for LC/MS runs. Nanospray emitters from Proxeon (Odense, Denmark) were used for static nanospray experiments. PicoFrit columns (75-µm i.d., New Objective, Woburn, MA) were used for nano-lc/ms. During experiments, special care was taken handling caffeine, TFA, and acetonitrile according to the manufacturer s guidelines. Figure 2. Wide-mass range, single-scan mass spectra of calibration mixture in the linear trap (a) and in the orbitrap at nominal resolving powers of (b) 7500 and (c) at AGC target value N ) with external calibration. Inset shows a minor doublet of two isotopomers ( 34 S and 13 C 2) of MRFA peptide separated by a theoretical mass difference Da. with Xcalibur software. Mass spectral data can be stored in fullprofile or reduced-profile format. In the latter, data below the threshold of detection (i.e., close to the thermal noise of the preamplifier) are removed to reduce the size of the dataset. The resolving power of the orbitrap (full width half-height) is switched in discrete steps between the following nominal values at m/z ) 400: (1.9-s scan cycle time), (1-s scan cycle time), , and The latter corresponds to RESULTS AND DISCUSSION Resolving Power. Figure 2 demonstrates wide-mass range, single-scan mass spectra acquired with different resolving powers. Intensity distributions are similar at the two resolving powers (R ) 7500 and R ) ) and absolute abundances differ by <20%. This reflects the fact that resolving power is determined by the acquisition time and not by orbitrap imperfections such as inaccuracy of manufacturing or insufficient vacuum. It should be noted that the resolving power is also unaffected by the AGC target value. Because trapping in the orbitrap is electrostatic (i.e., no magnetic field and no rf potentials), the frequency of the axial oscillations is inversely proportional to the square root of m/z, 7 in contrast to the cyclotron frequency in FTICR, which is inversely proportional to m/z. As a result, for a fixed acquisition time, the resolving power of the orbitrap mass analyzer diminishes as the square root of m/z, i.e., slower than in FTICR (though it should be noted that the resolving power of the orbitrap remains lower than that of 7-T FTICR until m/z 900 and to m/z 2500 of 12-T FTICR). As previously observed, 9 the resolving power of the orbitrap diminishes with an increase of ion mass, even if m/z is unchanged. This effect has been attributed to collisions with background gas that lead to fragmentation of ions and formation of noncoherent packs of fragment ions. For a given m/z, the center-of-mass collision energy with residual gas remains the same while the collision cross section increases with mass, thus leading to faster scattering, fragmentation, and transient decay. This process is more pronounced in the orbitrap than in FTICR because the ion energy is independent of m/z in the orbitrap, while in FTICR it decreases as (m/z) -1. Thus, ultrahigh vacuum is critical for high Analytical Chemistry, Vol. 78, No. 7, April 1, 2006

5 Figure 3. Measured and theoretical isotope distributions of proteins at setting R ) (100 scans averaged, external calibration): (a) horse heart apomyoglobin, m ) Da (monoisotopic, neutral), z ) +10; (b) carbonic anhydrase, m ) Da (monoisotopic, neutral), z )+21. resolution measurements of proteins. Figure 3 presents examples of high-resolution spectra of higher-mass proteins acquired at the same settings as Figure 2b. Although the resolution of these proteins is inferior to that for singly charged ions of the same m/z (Figure 2), it remains sufficient to correctly resolve the isotopic envelope and to provide good mass accuracy for both of the proteins ( ppm). Mass Accuracy. A two-point calibration is used to calculate mass in the orbitrap, even though it is only an approximation for a more complex calibration law. This can be seen in Figure 4, where mass traces for different ions (those in Figure 2) are drawn as a function of AGC target number N with MRFA at m/z ) 524 serving as an internal calibrant. It is clear that despite widely varying peak intensities within and between mass spectra, internal calibration mass accuracy stays well within 2 ppm even for extremely large numbers of ions (i.e., N > 10 6 ). With even larger ion populations, space charge effects in the C-trap distort energy and spatial distributions to such an extent that this starts to affect ion distribution and movement inside the orbitrap. Figure 5 shows the long-term stability with external calibration, which demonstrates that mass accuracy stays well within 5 ppm over 15 h. Variability is due to shot noise and thermal sensitivity of the inner electrode voltage, which is greater than mass shifts due to space charge effects. To optimize external calibration stability, the orbitrap and associated power supplies are thermally regulated. Transmission and Detection Limit. Signals measured with the linear trap and orbitrap detectors were compared in order to estimate the transmission. This comparison was carried out at a very low AGC target value N ) 200 using spectra shown in Figure 6. The full-profile orbitrap spectrum shows the noise band produced by thermal noise of the image current preamplifier. It should be emphasized that linear trap detection employs electron multipliers capable of single-ion detection while image current detection of the orbitrap has an inherent noise band equivalent to 20 ions (for 1-s acquisition length), as estimated using a previously published method. 14 Such a relatively small noise band was achieved by optimizing the preamplifier design as well as minimizing the length of signal wires. As linear trap noise peaks are clearly single ions of chemical background, this could be used to estimate the total number of ions at m/z ) in Figure 6a: ions total in 10 scans (or ions in a single scan). Given that the ion transmission from the axis of the linear trap to its detectors is better than 50%, 6 this brings the original number of ions inside the linear trap to for N ) 200. For the mass peak at mass m/z ) 524, the signal-to-noise ratio on the orbitrap is S/N ) 13 for 10 averaged scans, which results in S/N ) 4 for a single spectrum. The average S/N ) 4inthe orbitrap at noise band equivalent of 20 ions yields an estimation for an average number of detected ions of (taking into account isotopes). This indicates a transfer efficiency from the linear trap to the C-trap and then to the orbitrap of 30-50% (with the first number corresponding to m/z < 200 and the second to m/z > 1000), which is consistent with independent direct electrometer measurements at higher target numbers N (data not shown). To estimate detection limits, direct measurements of a lowconcentration solution of bradykinin were carried out in static (14) Limbach, P. A.; Grosshans, P. B.; Marshall, A. G. Anal. Chem. 1993, 65, Analytical Chemistry, Vol. 78, No. 7, April 1,

6 Figure 6. Comparison of 10 averaged spectra for (a) the LTQ and (b) the orbitrap detectors for the same very low AGC target N ) 200 and a narrow mass range containing MRFA peptide (m/z ) ) (R ) in the orbitrap, full-profile mode, external mass calibration). Figure 4. Mass errors plotted for different m/z as a function of AGC target value N with the mass peak of MRFA peptide (m/z ) ) used as an internal mass calibrant at R ) : (a) m/z ) , (b) m/z ) , and (c) m/z ) Figure 5. Long-term mass stability for MRFA (m/z ) ) using external mass calibration (R ) , N ) , single scan acquired every 6 s over 15 h). Figure 7. Mass spectrum of low-concentration bradykinin (charge state +2, theoretical monoisotopic m/z ) ) on the LTQ orbitrap (3 amol consumed, 3 nm sample concentration, R ) , N ) 5000, 1.16-s injection time and 2-s total scan cycle time, external mass calibration). nanospray mode using a nanospray ion source at 30 ( 10 nl/ min flow rate. Freshly prepared stock solutions of bradykinin at 100 fmol/µl were serially diluted to a concentration 3 fmol/µl. Figure 7 shows a spectrum with 3 amol of bradykinin consumed over the duration of a 2-s scan (R ) ). Dynamic Range. The dynamic range over which accurate mass measurements can be made is a key figure of merit as it actually determines the utility of accurate-mass capability for reallife applications much more than other parameters (e.g., resolving power). While detailed consideration is to be published elsewhere, Figure 8 provides an example of the dynamic range of mass accuracy of the LTQ orbitrap for real-life HPLC-MS analysis and shows that mass accuracy is achieved in a single scan even for mass peaks with up to 5000 times difference in abundances (peak at m/z ) 260 vs peak at m/z ) 610). The spectrum in Figure 8 was acquired in the reduced profile mode Analytical Chemistry, Vol. 78, No. 7, April 1, 2006

7 Figure 8. Illustration of dynamic range of mass accuracy of the LTQ orbitrap (siloxane impurities in propanolol sample) in a single 1-s scan (R ) , N ) , external mass calibration, reduced profile mode). Figure 9. Example of data-dependent acquisition with external mass calibration for a sample containing small molecules, with one highresolution mass spectrum recorded of the precursors at R ) and N ) (a) followed by three data-dependent MS/MS spectra at R ) 7500, N ) , (b) for precursor at m/z ) 260, (c) for precursor at m/z ) 310, and (d) for precursor at m/z ) 386. Data-Dependent Acquisition. Figure 9 is an example of rapid data-dependent acquisition for a sample containing small molecules using a fast gradient (3 min from 100% A to 100% B, where A is water with 0.1% TFA and B is acetonitrile with 0.1% TFA) LC separation on a Hypersil C18 column (2.1-mm i.d., 5-µm particles). A mass spectrum at nominal resolving power of (scan cycle time 1 s) is followed by three rapid data-dependent MS/MS acquisitions at nominal R ) 7500 at a rate of 2.5 spectra/s. Fast acquisition of MS/MS spectra is possible because of the high transmission between the linear trap and the orbitrap, which allows shorter ion accumulation times than in FTICR and TOF. Despite lower resolving power settings in MS/MS spectra, mass accuracy is not sacrificed so elemental composition can be unambiguously determined in all cases. CONCLUSION The main performance parameters of a novel linear ion trap/ orbitrap hybrid mass spectrometer have been characterized. A novel approach to ion storage and injection into the orbitrap allows high resolving power, mass accuracy, and transmission over a wide dynamic range and forms the basis for a hybrid mass spectrometer combining these analytical parameters with the MS n capability of the linear ion trap mass spectrometer. Utilizing short fill times of the linear trap and relatively low number of ions for Analytical Chemistry, Vol. 78, No. 7, April 1,

8 analysis in the orbitrap, this hybrid is capable of providing rapid, accurate-mass MS n analysis of complex mixtures with external calibration. ACKNOWLEDGMENT The authors express their deep gratitude to colleagues who made invaluable contribution to the development of this instrument: Dr. Reinhold Pesch, Dr. Gerhard Jung, Frank Czemper, Oliver Hengelbrock, Silke Strube, Dr. Hans Pfaff, Dr. Torsten Ueckert, Ralf-Achim Purrmann, Juergen Srega of Thermo Electron (Bremen) GmbH, Dr. Mike Senko, Mark Hardman, Mike Antonczak, Dr. Eric Hemenway of Thermo (San Jose), Dr. Steve Davis, Robert Lawther, and Andrew Hoffmann of HD Technologies Ltd. and later Thermo Masslab Ltd. The authors thank Dr. Mike Senko for valuable comments on the manuscript. Received for review October 20, Accepted February 1, AC Analytical Chemistry, Vol. 78, No. 7, April 1, 2006

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Yuichiro Hashimoto, Hideki Hasegawa, Takashi Baba, and Izumi Waki Central Research Laboratory, Hitachi, Ld., Tokyo, Japan We

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Tendenze nell innovazione della strumentazione in spettrometria di massa: Tendenze nell innovazione della strumentazione in spettrometria di massa: Trappola lineare, orbitrap, ion mobility e nuova strumentazione René Magritte La Condition Humaine 1 Ion Trap Mass Spectrometry

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Ion traps. Quadrupole (3D) traps. Linear traps

Ion traps. Quadrupole (3D) traps. Linear traps Ion traps Quadrupole (3D) traps Linear traps 3D-Ion traps A 3D-ion trap can be considered the tridimensional analogue of the linear quadrupole mass analyzer: B A C D The central ring electrode of the 3D-ion

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

Electrospray ionization mass spectrometry (ESI-

Electrospray ionization mass spectrometry (ESI- Automated Charge State Determination of Complex Isotope-Resolved Mass Spectra by Peak-Target Fourier Transform Li Chen a and Yee Leng Yap b a Bioinformatics Institute, 30 Biopolis Street, Singapore b Davos

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap 1 Last Time Mass Analyzers; CAD and TOF mass analyzers: 2 Fourier Transforms A transform is when you change your analytical space without

More information

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) Bruker Daltonics Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) During the last decade, RF ion traps have become a workhorse of the analytical laboratory

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

All of us engaged in science face exciting opportunities

All of us engaged in science face exciting opportunities Ion Trap Mass Spectrometry: A Personal Perspective FOCUS: QUADRUPOLE ION TRAPS ACCOUNT AND PERSPECTIVE George Stafford, Jr. LC and LC/MS Division, Thermo Finnigan, San Jose, California, USA This paper

More information

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4 High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic

More information

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF Application Note Donghui Yi and Christine Miller Jon D. Williams, GlaxoSmithKline Introduction Mass spectrometry (MS) has become a core technology

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

This is the total charge on an ion divided by the elementary charge (e).

This is the total charge on an ion divided by the elementary charge (e). 12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS Designed for Accuracy High resolution quantitative proteomics Innovation with Integrity LC-MS Setting New Standards in Accuracy The development of mass spectrometry based proteomics approaches has dramatically

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques

More information

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April Ion trap 3D Quadrupole Trap -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April 011 1 The trajectories of the ions can be given: They form Mathieu differential equations: with r + Ion

More information

Powerful Scan Modes of QTRAP System Technology

Powerful Scan Modes of QTRAP System Technology Powerful Scan Modes of QTRAP System Technology Unique Hybrid Triple Quadrupole Linear Ion Trap Technology Provides Powerful Workflows to Answer Complex Questions with No Compromises While there are many

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Spectrometry. Fundamental LC-MS. Mass Analysers Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

High-Field Orbitrap Creating new possibilities

High-Field Orbitrap Creating new possibilities Thermo Scientific Orbitrap Elite Hybrid Mass Spectrometer High-Field Orbitrap Creating new possibilities Ultrahigh resolution Faster scanning Higher sensitivity Complementary fragmentation The highest

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall, Christopher L.

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer 38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved

More information

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification

More information

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Chemistry Instrumental Analysis Lecture 35. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion

More information

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Jon Hao, Rong Ye, and Mason Tao Poochon Scientific, Frederick, Maryland 21701 Abstract Background:

More information

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Two major approaches, known as bottom-up

Two major approaches, known as bottom-up Calibration Function for the Orbitrap FTMS Accounting for the Space Charge Effect Mikhail V. Gorshkov, a David M. Good, b Yaroslav Lyutvinskiy, b Hongqian Yang, b and Roman A. Zubarev b a Institute for

More information

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 11, 1467 1477 (1997) Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass

More information

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest

More information

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008 Mass spectrometry of proteins, peptides and other analytes: principles and principal methods Matt Renfrow January 11, 2008 Objectives of the Lecture 1. Make ions 2. Separate/Analyze/Detect ions 3. What

More information

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,

More information

PC 235: Mass Spectrometry and Proteomics

PC 235: Mass Spectrometry and Proteomics PC 235: Mass Spectrometry and Proteomics Lecture 1 May 11 th, 2009 Shenheng Guan NIH NCRR Mass Spectrometry Facility, UCSF sguan@cgl.ucsf.edu Course Outline Lectures: 10hrs: 10am-12 in Genentech Hall Room

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers Fourier Transform Ion Cyclotron Resonance Mass Spectrometry MAGLAB Summer School 2015 Ryan P. Rodgers Electrospray Ionization Source Drying Gas Capillary Evaporation Chamber High Voltage Supply Charged

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry Fourier Transform Ion Cyclotron Resonance Mass Spectrometry What is Mass Spectrometry? Gravity F= G m M/r 2 r M m Newton Newton s second law F = m a Mass-Energy equivalence E = m c 2 1 ev = ~10-9 amu Einstein

More information

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results A. Kürten 1, J. Curtius 1 and S. Borrmann 1,2 1 Johannes Gutenberg-University Mainz, Germany 2 Max Planck

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

Quantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system

Quantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system Quantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system Kevin Mchale Thermo Fisher Scientific, San Jose CA Presentation Overview Anabolic androgenic

More information

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Introduction The oil industry requires detailed information on the composition of

More information

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS. ESS The RGA freenotes Theory page 1 of 14 RGA Theory Notes Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

A DIVISION OF ULVAC-PHI. Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification

A DIVISION OF ULVAC-PHI. Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification A DIVISION OF ULVAC-PHI Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification Designed for Confident Molecular Identification and Superior Imaging

More information

CHAPTER D3 TOF ION OPTICS

CHAPTER D3 TOF ION OPTICS Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide...399 Summary...401 Background...403 EquationsofMotionofIons...403 Resolution...405 Reflectron...407 Comparison

More information

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry Graduate Theses and Dissertations Iowa State University Capstones, Theses and Dissertations 2008 Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

More information

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * 1 Department of Chemistry, Pomona College, Claremont, California

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

INTRODUCTION OVERVIEW

INTRODUCTION OVERVIEW COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Mass spectrometric determination of the surface compositions of ethanol water mixtures

Mass spectrometric determination of the surface compositions of ethanol water mixtures International Journal of Mass Spectrometry 212 (2001) 267 271 www.elsevier.com/locate/ijms Cluster/kinetic method Mass spectrometric determination of the surface compositions of ethanol water mixtures

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Catalysis CAPABILITIES

Catalysis   CAPABILITIES Catalysis www.extrel.com CAPABILITIES Contents Extrel instruments have been recognized for their exceptional performance by the world s leading researchers for more than 50 years. Reliability and flexibility

More information

ABI 3200 Q TRAP LC/MS/MS System

ABI 3200 Q TRAP LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 ABI 3200 Q TRAP LC/MS/MS System The advantages of an ion trap and the performance of a triple quad. All in one system.

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP Application Note MBMS for Preformed Ions, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster Presented at 45th ASMS Conference on Mass Spectrometry, June 1-5, 1997) In order to accurately characterize a plasma

More information

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan

More information

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality?

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Laure Joly & Vincent Hanot WIV-ISP, Juliette Wytsmanstraat 14, 15 Brussel The passengers pesticides are requested......

More information

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller

More information

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides Abstract Here we describe a two-step QC procedure for synthetic peptides. In the first step, the

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby j5 1 Mass Analysis Gene Hart-Smith and Stephen J. Blanksby 1.1 Introduction Modern day mass analyzer technologies have, together with soft ionization techniques, opened powerful new avenues by which insights

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

The Orbitrap: a new mass spectrometer

The Orbitrap: a new mass spectrometer JOURNAL OF MASS SPECTROMETRY J. Mass Spectrom. 2005; 40: 430 443 Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/jms.856 SPECIAL FEATURE: PERSPECTIVE The Orbitrap: a new

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

Design considerations for linear Paul trap mass spectrometer under development

Design considerations for linear Paul trap mass spectrometer under development Design considerations for linear Paul trap mass spectrometer under development S.Sevugarajan and A.G.Menon Department of Instrumentation, Indian Institute of Science, Bangalore 561, India. Abstract This

More information

Asymmetrical features of mass spectral peaks produced by quadrupole mass filters

Asymmetrical features of mass spectral peaks produced by quadrupole mass filters RAPID COMMUNICATIONS IN MASS SPECTROMETRY Rapid Commun. Mass Spectrom. 2003; 17: 1051 1055 Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.1017 Asymmetrical features

More information

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS PesticideScreener Comprehensive Pesticide Screening and Quantitation Innovation with Integrity UHR-TOF MS The Challenge of Comprehensive Pesticide Residue Analysis The use of pesticides to reduce crop

More information