Ion traps. Quadrupole (3D) traps. Linear traps

Size: px
Start display at page:

Download "Ion traps. Quadrupole (3D) traps. Linear traps"

Transcription

1 Ion traps Quadrupole (3D) traps Linear traps

2 3D-Ion traps A 3D-ion trap can be considered the tridimensional analogue of the linear quadrupole mass analyzer: B A C D The central ring electrode of the 3D-ion trap can be imagined as generated by rods A and C, connected after being deformed. Rods B and D are ideally modified to generate the end caps.

3 The ring electrode and the end caps are assembled through insulating spacers, avoiding any possible electrical contact between them: The assembled 3D trap is actually less than 10 cm long and less than 5 cm large: End cap Ring electrode End cap

4 Geometry of 3D-ion traps A different co-ordinate system (with respect to quadrupole) needs to be defined to describe ion motion within the 3D-ion trap: The internal diameter of the ring electrode defines the r axis, whereas the axis of the two end caps corresponds to the z axis. The maximum displacements from the trap center permitted along the two axes are r 0 and z 0, respectively. An rf voltage is applied between the ring and end-cap generate an electric quadrupole field. electrodes to

5 The potential at any point (r,z) in the quadrupolar field is given by: where 0 is the potential applied to the ring electrode: As for the quadrupole, U and V are the dc potential and the rf potential amplitude, respectively, is the rf potential angular frequency. Ions of a given m/z value may undergo stable or unstable motion in the trap. The quantitative solution to stability conditions is described by second order differential equations of the Mathieu form. The solutions are readily summarised in the form of a stability diagram.

6 Mathieu stability diagram for a 3D-ion trap The Mathieu diagram for a 3Dion trap is still drawn within a (q,a) plane, but the co-ordinate definitions are slightly different than those for quadrupoles: The stability regions are now labelled as z-stable and r-stable and are not symmetrical. Once again, different z-r stable regions, like those marked as A and B, can be found in the diagram.

7 Region A is the one commonly exploited for 3D-ion trap operation: In the figure, lines corresponding to the same z and r parameters, in turn correlated to the oscillation frequency of ions in the z and r directions (by analogy with x and y oscillations of ions within the quadrupole), are also reported. It is worth noting that the stability region crosses the q axis, in this case.

8 After U, V, (i.e. 2 f, with f = 1.1 Mhz) have been fixed, along with r 0 and z 0, a and q coordinates can be calculated for any m/z ratio at any time and the motion for the relevant ion can be simulated: r z r direction A stable trajectory within the 3D-ion trap implies a continuous oscillation between the end caps and inside the ring electrode aperture. The oscillations describe a three-dimensional 8, which is slightly bended. Small oscillations are super-imposed on the main motion.

9 The ion motion can be inferred also by considering the potential surface in a quadrupole ion trap: During positive excursions of the potential applied to the ring electrode, the ion is pushed towards the end caps (i.e. towards z 0 or z 0 ) but far from the ring surface (r 0 0). During negative excursions, the surface shape changes so that the ion returns to the center (z 0 0) along the z axis but is pushed towards the inner surface of the ring (i.e. towards r 0 or r 0 ).

10 Starting from the a and q parameters for a specific ion under fixed experimental conditions (U, V, ), its oscillation frequencies along the z and r direction, known as secular frequencies, can be calculated: where u is equal to z or r, according to the cases, and u obtained from the Mathieu diagram. can be The fundamental secular frequencies (i.e. those corresponding to n = 0) are usually considered, since higher order frequencies are of little practical significance.

11 Ion trap operation The process leading a 3D-ion trap to a mass scan can be divided into four stages: 1. Ion trapping and ion cooling 2. Ion selection 3. Ion excitation 4. Mass scan

12 During stages 1-4 no further ion can be admitted within the trap, thus a electrostatic ion gate, i.e. a system of electron lenses controlling ion transfer into the trap, has to be present before the trap. Ions can be loaded within the trap only if the ion gate is open. For example, in the case of negative ions, this means applying a positive potential to electron lenses: When ion loading is complete the gate potential is switched to negative values, so that no further ions can enter the trap and the scan procedure can begin. After inverting the potentials the same scheme can be applied to positive ion mode.

13 Ion trapping and cooling Ion entering the ion trap can be trapped, i.e. stored within the device without hitting any of the electrodes or exiting through the end cap apertures, only if their (q, a) co-ordinates are included in a z-r stability region of the Mathieu diagram. Simulations of motion for ions with a specific m/z ratio, under certain U, V and conditions will show if trapping will actually occur.

14 Actually, optimum trap operation requires that the ions have favorable initial conditions, in particular, a proper distribution of kinetic energies. This adjustment is achieved by using low pressure helium gas to remove excess kinetic energy from the ions and cause them to occupy the central region of the trap (ion cooling):

15 The importance of ion cooling by helium buffer gas is clearly evidenced by these examples: MRFA

16 Ion trap loading: Automatic Gain Control (AGC) Typically the ion trap can hold up to ions before mutual coulombic repulsions significantly affect their trajectories (space-charge effects) and greatly reduce mass resolution. In most modern 3D-ion trap instruments an automatic control of ion trap loading (Automatic Gain Control) is available to minimize such effects. The main stages of a AGC process are: 1) Loading of the ion trap for a definite time (10 ms) 2) Pre-scan of trapped ions 3) Evaluation of the signal obtained at the detector (AGC prescan signal) 4) Comparison between the obtained signal and a user-defined target value (depending on the MS experiment: full scan, SIM, SRM, CRM, etc.) 5) Adjustment of ion injection time to get compliance with target values.

17 The AGC Prescan Signal can be expressed by the relationship: where the number of ions is related to the loading time. The injection time (ion time) can be calculated, as a multiple of the presccan loading time, by the following simple equation:

18 Effect on mass resolution of the ion population loaded inside a 3D-trap; MRFA peptide:

19 Ion scanning based on mass-selective instability Differently from the quadrupole, in a 3D-ion trap a ion is required to become unstable, along the z-axis, to be transferred to the detector and thus provide a mass signal. The main approach to ion scanning with a 3D-trap is thus based on the mass-selective instability: In this case, the dc component of the 0 potential is equal to 0 and only a RF voltage is applied to the ring electrode: The rf voltage amplitude is raised to a certain value during ion trapping, then slightly lowered, to promote ion cooling. Afterwards, a linear increase is effected, to promote destabilization of ions with increasing m/z ratios.

20 According to the Mathieu co-ordinates for a 3D-ion trap: working with U = 0 means that the mass scan line is the q z axis: As the rf voltage amplitude, V, is increased, ions move to higher q values. Eventually, ions with the lowest m/z ratios reach the stabiliity boundary (q = 0.908).

21 Simulations show that a ion reaching the condition q > becomes unstable in the z direction: As the rf voltage V is increased, ions with increasing mass will reach the critical co-ordinate q z and will be ejected from the ion trap.

22 Stability diagram and mass range for a 3D-ion trap The m/z ratio of the highest mass ejected ion, for a given rf voltage V, can be easily calculated using the equation: In this case q z has to be replaced with the ejection value, For typical ion trap geometries, like the one based on the equation r 02 = 2 z 02, with r 0 = 1 cm, a maximum rf voltage amplitude, V max, equal to 7500 V at = 1.1. MHz will determine a maximum ejectable m/z ratio equal to 650. The mass range can be extended: by increasing V max reducing the trap dimensions (r 0 and /or z 0 ) decreasing reducing the q z ejection value

23 The initial (and also minimum) amplitude of the rf voltage, V min, has a similar influence on the low mass limit, usually called exclusion limit or lowmass cut off (LMCO) of the ion trap. Variation of the low mass limit with the rf voltage can be easily seen by comparing the relative positions of ions with three different mass-to-charge ratios along the mass-selective instability line (a z = 0):

24 q z, Low Mass Cut Off (LMCO) and secular frequency calculation Let us consider a butylbenzene ion (m/z 134) in an ion trap with r 0 =1.00 cm and with z 0 = cm and under the following conditions: The corresponding q z value can be easily calculated: As < the ion will follow stable trajectories inside the ion trap.

25 As far as the LCMO value is concerned, it is worth noting that the m q z product is constant at constant V: As a consequence, the follow relationship can be written: With a rf potential of 757 V (0-P) applied to the ring electrode, only ions with m/z > 66.4 will be stored in the ion trap.

26 The general equation for q z can be re-written in a further form: Consequently, also the (m/e)/v ratio is constant, for a given q z. The potential V to be applied to the ring electrode to get a given LMCO can be calculated by the following equation:

27 Calculation of secular frequencies can be made starting from the already shown equation: and considering that z (and, equivalently, r ) is given approximately by the relation: For the butylbenzene ion (m/z 134) at a rf voltage of 757 V, q z = 0.450, thus z = and the fundamental secular frequency for the axial direction, z,0, corresponds to rad s -1 or, equivalently, 167 khz. Similarly, a fundamental secular frequency for the radial direction, r,0, can be estimated.

28 Resonance Excitation As the motion of ions confined in a quadrupole 3D-ion trap is characterized by two secular frequencies, axial and radial, ion motion can be excited upon resonant excitation at either or both of these frequencies. Excitation along the the z axis can be effected, after ion cooling, by applying a small (a few hundred mv) supplementary potential, oscillating at the secular frequency, across the end-cap electrodes, thus working in dipolar mode. Under the effect of the exciting potential ions are forced to move away from the trap center and acquire higher kinetic energies (up to tens of ev).

29 A simulation of ion motion shows the destabilization occuring along the axial direction:

30 Resonant excitation can be used for several purposes: (i) to remove unwanted ions generated during ionization, thus isolating a narrow range of m/z ratios; in this case, wavebands of frequencies are applied to the end-cap electrodes to excite and eject many ion species simultaneously, thus leaving a single ion species (or a small range of m/z ratios) isolated within the ion trap; (ii) to increase ion kinetic energy, thus promoting endothermic ionmolecule reactions; (iii) to deposit internal energy in ions through momentum-exchange collisions with helium atoms, a fundamental process to promote fragmentation; (iv) to eject ions while the amplitude V of the main r.f. potential is being ramped up, a procedure known as axial modulation.

31 Space-charge effects and axial modulation One of the inherent features of ion trap is that while ions with lower m/z ratio (red) are being scanned out of the trap, those with higher m/z ratios (orange, blue) are still inside the trap. a q lim = q The motion of higher m/z ratio ions generates an interference (spacecharge effects) on the trapping potential, causing a broadening of peaks related to lighter ions. The phenomenon, deleterious for mass resolution, can be reduced significantly by the so-called axial modulation, consisting in applying, between the end caps, a supplementary oscillating field of 6 V (p-p) at a frequency about one half that of the rf voltage applied to the ring electrode.

32 Under these conditions the secular motion of ions enters into resonance with the supplementary field just before they are ejected from the trap. The ions are energized and come into step, i.e. the energy broadening due to space-charge effects is recovered. Interestingly, axial modulation can be exploited also to extend the m/z range over the limit imposed by the maximum value attainable for the RF voltage on the ring electrode. As shown by the following calculations, supplemental RF voltages in the khz range result in extension of the higher limit of the ion trap m/z range from 650 units (attainable with a 7500 V rf voltage amplitude, at 1.1 MHz, in a ion trap with r 02 = 2 z 02, with r 0 = 1 cm) to 6500 units: q lim q lim

33 Resonance ejection A supplemental RF voltage applied to the end-caps can be exploited for a systematic ejection of unwanted ions from the ion trap. In this case also the supplemental RF voltage can be scanned, like the fundamental RF voltage:

34 In a figurative sense the application of a supplemental voltage at a specific frequency results in opening a hole (actually a resonance or instability point) in the stability diagram: The q z value at which the hole is located can be easily calculated from the z value corresponding to the secular frequency:

35 Ion isolation using resonance points A careful use of resonant ejection and RF voltage scan at the ring electrode can lead to isolation of ions having a narrow m/z range in the ion trap. At 1000 V none of the ions has a q z value approaching that of the resonance point so the ions remain inside the trap. Resonance (or instability) point q = At 3000 V the ion with m/z 500 has been ejected and the one with m/z 1,000 is about to be ejected. At 6000 V the q z values for all the ions are greater than 0.227, the value of the resonance point.

36 A combination of forward and reverse resonance ejection ramps may be used to isolate ions in the trap: (i) reverse scanning resonantly ejects ions from high to low m/z (ii) forward scanning resonantly ejects ions from low to high m/z (iii) isolation of ions with a single m/z value is achieved

37 Further approaches to ion isolation Upper apex isolation Supposing that the orange ion, located in the A point of the stability diagram, is to be isolated, the following steps are followed: a the ion is moved to point B by increasing the RF voltage of the ring electrode: several (lighter) ions along the q axis will be ejected; q lim = a dc voltage U is also applied to the ring electrode: the orange ion is lifted to the C point, the upper apex of the diagram. All the other ions are istantaneously placed outside the stability diagram. q the orange ion is brought back to the q axis.

38 Two-stage isolation The orange ion is moved from the A to the B point by reducing the RF voltage at the ring electrode; a it is then moved to the C point by applying a negative dc voltage to the ring; all the heavier ions (like the light blue ion) will be outside the stability diagram; q lim = the dc voltage is put to 0 (C B), the RF voltage is increased (B D), then a positive dc voltage is applied to the ring: the orange ion is lifted to the E point. At this stage, all the other ions (lighter) are destabilised; the orange ion is taken back to the q axis (E D) q

39 A MS/MS experiment in a 3D-ion trap After isolation in the 3D-trap, an ion can be dissociated through collisions with helium atoms. Before proceeding with fragmentation, a stabilization is performed by lowering the RF voltage applied to the ring electrode: The procedure leads to a decrease of the q co-ordinate for the isolated ion:

40 Collisional Induced Dissociation (CID) is started by applying a Supplementary AC Voltage pulse between the end caps: Product ions, with lower m/z ratios, are generated inside the trap. Due to preliminary stabilization of the precursor ion (green), most new ions are also stable, as their q co-ordinates are lower than the critical value:

41 During the collisional stage the excitation voltage amplitude (usually referred to as Collisional Energy, CE) plays a key role in determining the relative intensity of the precursor and product ions in the MS/MS spectrum that will be subsequently acquired: Precursor ion (Product ions) It is worth noting that high excitation voltages result in loss of ions and thus of total intensity. This is due to ejection of the precursor and parent ions through the end-caps apertures before actual MS/MS scanning.

42 Product ions and residual precursor ions are progressively ejected outside the trap by raising the Main RF voltage applied to the ring electrode. Axial modulation is usually applied in this stage: a supplementary AC voltage is applied between the endcaps, at a frequency corresponding to a q value slightly lower than the critical one. A hole is then opened in the stability diagram. In this snap-shot of the Mathieu stability diagram, red ions are just falling into the hole : Conversion dynode Electron multiplier

43 While proceeding with Main RF and Supplementary AC voltages scan also yellow ions are ejected. In this snap-shot of the ion trapdetector assembly, the last electrons generated on the conversion dinode by red ions are visible while yellow ions are just being ejected from the trap: Conversion dynode At this stage, blue product ions and residual green precursor ions are still inside the trap: Electron multiplier

44 When ions hit the conversion dynode electrons are generated and subsequently accelerated towards the electron multiplier. The current signal is then registered as a function of the m/z ratio: a MS/MS spectrum is obtained. Conversion dynode Electron multiplier

45 A 3D-ion trap mass spectrometer: Thermo LCQ Skimmer 3D-ion trap Heated Capillary Interoctapole Lens Electrospray Source Tube Lens First Octapole Second Octapole Detector

46 Pros and Cons of 3D-Ion Trap Mass Analysers Compactness and mechanical simplicity Enhanced specificity and sensitivity for MS/MS experiments MS n (n up to 9-10) capability (simply by the use of additional ion isolations/fragmentations, performed sequentially in time) Poor dynamic range, due to space-charge effects (reduced by procedures like AGC) Relatively low resolution (about at normal scan rates) and mass accuracy (usually higher than 100 ppm) Poor trapping efficiency (1-10% for externally generated ions)

47 Linear ion traps Since 1988 linear ion traps have been studied by the researchers at Finnigan as a possible alternative to 3D-ion traps. In 1995 Mark Bier and John Syka patented several types of ion traps with increased ion capacity, including the linear ion trap, which could be operated in the mass selective instability mode, the same method used in all commercial 3D traps for producing mass spectra.

48 Among traps patented in 1995 one of the most successful (and then applied to commercial instrumentation) is based on a quadrupole with hyperbolic rods cut into three sections, insulated electrically by quartz spacers: 37 mm 12 mm 12 mm Y Z X Front Section Center Section Back Section Slot: 30 mm 0.25 mm The length of front and back sections, identical, is 1/3 of that of the center section, whose peculiarity is having slots cut in the x rods.

49 In order to correct non-linearities in the quadrupole fields generated by the slots in the center section, x rods are slightly more distant than y ones from the quadrupole center: Y 0 = 4 mm X 0 = 4.76 mm The slots cut into the x rods are exploited for ion radial ejection towards two detectors, placed at the sides of the linear trap, as shown in the upper view of the device: Detector 1 Center Section Front Lens Front Section Detector 2 Back Section Back Lens

50 The division of quadrupole into three sections has a remarkable effect on the electric field operating within the center section. Field simulations obtained with the SIMION software Three-section trap show that fringe fields, clearly observed in a single quadrupole, are absent in the center section, thus enabling ion radial ejection to occur along the entire length of the section: One-section trap

51 Ions generated in the source are transferred into the linear ion trap by complex ion optics, subject to differential pumping. In the case of the Thermo Electron LTQ XL spectrometer, having a ESI interface with orthogonal geometry, the general scheme is the following: Electrospray Source Ion Transfer Ion Transfer Tube Tube Skimmer Square Quadrupole Skimmer Inter-multipole Lens 1 Split Gate Lens Detector Torr He Center Section Tube Lens Square Quadrupole Octopole Tube Lens Front Lens Detector 2 Back Lens Front Section Back Section 8.3 L/s 25 L/s 300 L/s 400 L/s

52 A snapshot of ion transfer from the ESI source to the linear ion trap of the LTQ XL spectrometer: The picture shows ions with different m/z ratios (different colors) while being transferred through the Octopole preceding the linear ion trap. Helium atoms (white particles) are present inside the trap.

53 The introduction of octapoles (like in the LCQ instrument) and square quadrupoles (like in LTQ XL), extends ion transmission efficiency over wider mass ranges:

54 Linear ion trap voltages During linear ion trap operation three types of voltage can be applied to its sections. DC 1 DC 2 DC 3 y The LTQ uses three dc axial trapping voltages, one for each rod section, to establish axial (Z-axis) trapping. DC 1 DC 2 DC 3 z For positively charged ions the operating potentials for the three sections during ion storage and mass analysis are: Section Front (DC 1) Center (DC 2) Rear (DC 3) Ion Storage - 9 V - 14 V - 12 V Mass analysis + 20 V - 14 V + 20 V

55 Ion storage is effected by applying only an AC voltage (V) between the rod opposite couples. This means working on the q axis ot the Mathieu stability diagram for quadrupole: q y Once ions have been stored, a AC voltage (at a 1.2 MHz frequency), with amplitudes up to 10 kv peakto-peak, is applied between couples of opposite rods: RF - RF + RF + RF - y x

56 A comparison of potential surfaces obtained with a 3D and a linear ion trap shows that the trapping space is about 20 times larger in the case of the linear trap: 3D ion trap linear ion trap

57 Mass scan is performed by increasing linearly the AC voltage (Main RF Voltage): The limit q-coordinate is progressively reached by higher m/z ratios, which are then ejected through the slots cut in the x rods: q y

58 Supplementary voltages in the linear ion trap Ion isolation, excitation and ejection, useful for MS/MS and MS n operation, can be performed by applying an AC voltage only to two the exit rods (x), while the y rods are grounded (GND): AC+ GND GND AC - y x Three voltage types are usually adopted: ion isolation waveform voltage, i.e. an AC voltage with a distribution of frequencies between 5 and 500 khz, containing all resonance frequencies except those corresponding to the ions to be trapped; resonance excitation RF voltage, applied to excite a previously isolated ion and promote its dissociation induced by collisions with helium atoms; resonance ejection RF voltage, useful to promote resonance in ions that are about to be ejected, to improve mass resolution.

59 A combination between the main RF voltage and a supplementary AC voltage, applied between the x rods, can provide isolation of ions with a single m/z ratio: In this case, once grey and green-colored ions have been ejected using the Main RF voltage, a square pulse in the waveform voltage opens a hole in the stability diagram, where blue ions fall:

60 A MS/MS experiment with the linear ion trap Once isolated in the linear ion trap, red ions, representing precursor ions for a subsequent MS/MS experiment, can be stabilized by lowering the Main RF Voltage, so that their q co-ordinate is decreased: q y As a result, the oscillations of red ions within the trap are minimized:

61 Excitation of red ions is then promoted by a pulse of Supplementary AC voltage at the same frequency as their secular frequency. q y Product ions (blue, green and grey) are generated by collisions with helium atoms (Collision Induced/Activated Dissociation, CID/CAD):

62 In order to obtain a MS/MS spectrum, fast, contemporary scanning of the Main RF Voltage and of the Supplementary AC Voltage is effected: Product ions (and residual parent ion) are thus progressively destabilised, ejected through the x-rod slots and finally detected by one of the detectors placed outside the trap. In the snap-shot on the right green ions are being ejected from the trap and transferred towards the detectors, while red ions are still stable inside the trap:

63 Ion ejection towards the two side detectors occurs along most of the trap length, through the slot in the x-rods: A significant increase in sensitivity is achieved through this approach. Actually, due to high sensitivity, linear ion trap instruments can be operated in MS n mode with n up to 15!

64 Comparison between LTQ Linear Ion Trap and 3D Ion traps Trapping efficiency: 10 Detection efficiency: 2 Global efficiency: 10 Ionic capacity 20 Scan speed (u/s): 4 Zoom Scan scan speed: 4 Ultra Zoom scan speed: 27 u/s

65 Comparisons between LTQ Linear Ion Trap and 3D Ion traps Single scan 0.5 s Myoglobin spectrum and deconvolution Deconvolution 3D- trap Single scan 0.41 s Deconvolution Linear trap

66 Isotope distributions for caffeine (195), MRFA (524) and Ultramark (1522) in a LTQ MS spectrum obtained with a scan rate of u/s: A similar resolution can be obtained by LCQ only at a quite lower scan rate (5500 u/s).

67 Zoom scan spectra obtained by LTQ at 1100 u/s for the Melittin peptide (GIGAVLKVLTTGLPALISWIKRKRQQ, the principal active component of bee venom) at different charge states: A similar Zoom Scan resolution can be obtained by LCQ only at a 280 u/s scan rate.

68 Ultra-zoom scan is a peculiar, very slow mass scan (27 u/s) performed by the linear ion trap, providing very high resolutions, up to 30000:

69 Multiple-event MS experiments with linear ion traps In the MS jargon, an event is a specific MS, MS/MS or MS n acquisition, performed sequentially with others during a MS experiment. High scan rate MS instruments, like those based on ion traps, can perform several MS events in very short times. In the case of linear ion trap instruments this characteristic is pushed to astonishing results: without with Sequence of MS events acquired by LTQ within 4 s!

70 RP-LC-ESI-MS chromatogram, in Base Peak mode, for a mixture of 7 drugs, obtained by LTQ: ketoconazole buspirone Relative Abundance timolol bifonazole nicardipine ampicillin glyburide Time (min)

71 Multiple-event scan for ketoconazole: MS and MS/MS [M+H] + [M+2H] 2+ Wide band activation is a special MS/MS experiment in which both the parent ion and potential product ions arising from H 2 O or NH 3 losses are fragmented contemporarily.

72 Multiple-event scan for ketoconazole: MS 3 and MS 4 The high sensitivity of the linear ion trap provides high S/N ratio spectra even in MS 4 mode.

73 It is interesting to point out that noise is decreased more than signal when subsequent MS stages are considered, thus the S/N ratio will be higher for MS n events than for MS or MS/MS ones:

74 The fragmentation pattern rapidly obtained for each analyte using MS, MS/MS and MS n acquisitions can be used for structural elucidations: MS 2 (event 3) + N O O O N N Cl MS 3 (event 13) O + N N Cl MS 4 (event 14) + O Cl Cl m/z 446 Cl Cl m/z 281 m/z 173 N O N N O Ketoconazole [M+H] + m/z 531 N O O Cl Cl MS 2 (event 3) -CH 2 =CO (H migration) + N N O O O N N Cl MS 3 (event 4) O + N N Cl MS 4 (event 5) O + Cl m/z 489 Cl Cl m/z 255 Cl m/z 187

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April Ion trap 3D Quadrupole Trap -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April 011 1 The trajectories of the ions can be given: They form Mathieu differential equations: with r + Ion

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results A. Kürten 1, J. Curtius 1 and S. Borrmann 1,2 1 Johannes Gutenberg-University Mainz, Germany 2 Max Planck

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Chemistry Instrumental Analysis Lecture 35. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion

More information

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Spectrometry. Fundamental LC-MS. Mass Analysers Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Tendenze nell innovazione della strumentazione in spettrometria di massa: Tendenze nell innovazione della strumentazione in spettrometria di massa: Trappola lineare, orbitrap, ion mobility e nuova strumentazione René Magritte La Condition Humaine 1 Ion Trap Mass Spectrometry

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Yuichiro Hashimoto, Hideki Hasegawa, Takashi Baba, and Izumi Waki Central Research Laboratory, Hitachi, Ld., Tokyo, Japan We

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Space Charge Effects in Linear. Quadrupole Ion Traps

Space Charge Effects in Linear. Quadrupole Ion Traps Space Charge Effects in Linear Quadrupole Ion Traps by Cong Gao B.Sc., Peking University, 2010 A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY in The

More information

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP Application Note MBMS for Preformed Ions, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster Presented at 45th ASMS Conference on Mass Spectrometry, June 1-5, 1997) In order to accurately characterize a plasma

More information

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS. ESS The RGA freenotes Theory page 1 of 14 RGA Theory Notes Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

More information

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) Bruker Daltonics Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) During the last decade, RF ion traps have become a workhorse of the analytical laboratory

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

All of us engaged in science face exciting opportunities

All of us engaged in science face exciting opportunities Ion Trap Mass Spectrometry: A Personal Perspective FOCUS: QUADRUPOLE ION TRAPS ACCOUNT AND PERSPECTIVE George Stafford, Jr. LC and LC/MS Division, Thermo Finnigan, San Jose, California, USA This paper

More information

Quadrupole Storage Mass Spectrometry

Quadrupole Storage Mass Spectrometry Quadrupole Storage Mass Spectrometry RAYMOND E. MARCH AND RICHARD J. HUGHES Trent University Peterborough, Ontario, Canada with a historical review by John F. J. Tbdd University of Kent Canterbury, Kent,

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Quattro Micro - How does it work?

Quattro Micro - How does it work? Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One

More information

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap 1 Last Time Mass Analyzers; CAD and TOF mass analyzers: 2 Fourier Transforms A transform is when you change your analytical space without

More information

Design considerations for linear Paul trap mass spectrometer under development

Design considerations for linear Paul trap mass spectrometer under development Design considerations for linear Paul trap mass spectrometer under development S.Sevugarajan and A.G.Menon Department of Instrumentation, Indian Institute of Science, Bangalore 561, India. Abstract This

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

This is the total charge on an ion divided by the elementary charge (e).

This is the total charge on an ion divided by the elementary charge (e). 12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectroscopy (SIMS) Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351)

VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351) VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351) The trap consists of a ring electrode (internal radius is r 0 ) and two end caps separated by z 0.; all internal

More information

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry Graduate Theses and Dissertations Iowa State University Capstones, Theses and Dissertations 2008 Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 11, 1467 1477 (1997) Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass

More information

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement

More information

Ions trapped within a linear quadrupole ion trap can

Ions trapped within a linear quadrupole ion trap can Mass Selective Axial Ion Ejection from a Linear Quadrupole Ion Trap F. A. Londry and James W. Hager MDS SCIEX, Concord, Ontario, Canada The electric fields responsible for mass-selective axial ejection

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer 38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved

More information

Practical Quadrupole Theory: Peak Shapes at Various Ion Energies

Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Randall E. Pedder ABB Inc., Analytical-QMS Extrel Quadrupole Mass Spectrometry, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster presented

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

Interface (backside) & Extraction Lens

Interface (backside) & Extraction Lens Plasma Interface Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control

More information

MICROSCREENED ENDCAP HOLES AND BUFFER GAS PRESSURE TAILORING FOR QUADRUPOLE ION TRAP MASS SPECTROMETRY

MICROSCREENED ENDCAP HOLES AND BUFFER GAS PRESSURE TAILORING FOR QUADRUPOLE ION TRAP MASS SPECTROMETRY MICROSCREENED ENDCAP HOLES AND BUFFER GAS PRESSURE TAILORING FOR QUADRUPOLE ION TRAP MASS SPECTROMETRY By DODGE L. BALUYA A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Powerful Scan Modes of QTRAP System Technology

Powerful Scan Modes of QTRAP System Technology Powerful Scan Modes of QTRAP System Technology Unique Hybrid Triple Quadrupole Linear Ion Trap Technology Provides Powerful Workflows to Answer Complex Questions with No Compromises While there are many

More information

Meng Yu and William J. Chappell School of Electrical and Computer Engineering, Purdue University, West Lafayette, Indiana, USA

Meng Yu and William J. Chappell School of Electrical and Computer Engineering, Purdue University, West Lafayette, Indiana, USA Ion Trajectory Simulation for Electrode Configurations with Arbitrary Geometries Guangxiang Wu, R. Graham Cooks, and Zheng Ouyang Department of Chemistry, Purdue University, West Lafayette, Indiana, USA,

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification

More information

Principles of Quadrupole and Ion Trap Mass Analyzers

Principles of Quadrupole and Ion Trap Mass Analyzers Principles of Quadrupole and Ion Trap Mass Analyzers 2-1 Rev20040526 Bioanalytical LC-MS 2-1 The Mass Peak in the Quadrupole Mass Spectrum 2-2 A common phrase in texts on MS says, Examine the peaks in

More information

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008 Mass spectrometry of proteins, peptides and other analytes: principles and principal methods Matt Renfrow January 11, 2008 Objectives of the Lecture 1. Make ions 2. Separate/Analyze/Detect ions 3. What

More information

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow LCMS n Dhanashri Bagal *, Eddie Kast, Ping Cao Discovery Attribute Sciences Amgen, South San Francisco, California, United

More information

Three dimensional Paul traps may benefit from. Linear Quadrupoles with Added Hexapole Fields

Three dimensional Paul traps may benefit from. Linear Quadrupoles with Added Hexapole Fields Linear Quadrupoles with Added Hexapole Fields Nikolai Konenkov Department of General Physics, Ryazan State Pedagogical University, Ryazan, Russia Frank Londry Pan Galactic Scientific, Omemee, Ontario,

More information

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading: Assigned Reading: Nier mass spectrometry paper Extra Reading: 1) Introduction a) We are now on to the heart of the mass spectrometer. How do you separate ions of different mass, so that you can detect/quantify

More information

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality?

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Laure Joly & Vincent Hanot WIV-ISP, Juliette Wytsmanstraat 14, 15 Brussel The passengers pesticides are requested......

More information

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) 1.0 Equipment: Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) Thermo Exactive Orbitrap Mass Spectrometer Direct Analysis

More information

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

WADA Technical Document TD2003IDCR

WADA Technical Document TD2003IDCR IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must

More information

arxiv:physics/ v1 [physics.plasm-ph] 5 Nov 2004

arxiv:physics/ v1 [physics.plasm-ph] 5 Nov 2004 Ion Resonance Instability in the ELTRAP electron plasma G. Bettega, 1 F. Cavaliere, 2 M. Cavenago, 3 A. Illiberi, 1 R. Pozzoli, 1 and M. Romé 1 1 INFM Milano Università, INFN Sezione di Milano, Dipartimento

More information

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Jon Hao, Rong Ye, and Mason Tao Poochon Scientific, Frederick, Maryland 21701 Abstract Background:

More information

Time-of-Flight Mass Analyzers

Time-of-Flight Mass Analyzers Time-of-Flight Mass Analyzers Jonathan Karty C613 lecture 1 March 6, 8 (Section 4. in Gross, pages 115-18) TOF Overview Time-of-flight (TOF) is the least complex mass analyzer in terms of its theory Ions

More information

SUPPORTING INFORMATION Antimicrobial activity of chlorinated amino acids and peptides

SUPPORTING INFORMATION Antimicrobial activity of chlorinated amino acids and peptides SUPPORTING INFORMATION Antimicrobial activity of chlorinated amino acids and peptides Melanie S. A. Coker*, Wan-Ping Hu, Senti T Senthilmohan and Anthony J. Kettle Free Radical Research Group, University

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Manipulating Short-Lived Isotopes with Inhomogeneous RF-Fields

Manipulating Short-Lived Isotopes with Inhomogeneous RF-Fields Manipulating Short-Lived Isotopes with Inhomogeneous RF-Fields T. Li and H. A. Schuessler Texas A&M University, College Station, Texas, USA Abstract: Online isotopes separators (ISOL-systems) and projectile

More information

Atomic Spectra HISTORY AND THEORY

Atomic Spectra HISTORY AND THEORY Atomic Spectra HISTORY AND THEORY When atoms of a gas are excited (by high voltage, for instance) they will give off light. Each element (in fact, each isotope) gives off a characteristic atomic spectrum,

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby j5 1 Mass Analysis Gene Hart-Smith and Stephen J. Blanksby 1.1 Introduction Modern day mass analyzer technologies have, together with soft ionization techniques, opened powerful new avenues by which insights

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi. via Aldo Moro Siena

Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi. via Aldo Moro Siena Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi Università degli Studi di Siena Dipartimento di Biotecnologie, Chimica e Farmacia via Aldo Moro 53100 Siena e-mail:

More information

MONOCHROMATIZATION AND POLARIZATION OF THE NEON SPECTRAL LINES IN CONSTANT/VARIABLE MAGNETIC FIELD

MONOCHROMATIZATION AND POLARIZATION OF THE NEON SPECTRAL LINES IN CONSTANT/VARIABLE MAGNETIC FIELD Romanian Reports in Physics 69, 49 (217) MONOCHROMATIZATION AND POLARIZATION OF THE NEON SPECTRAL LINES IN CONSTANT/VARIABLE MAGNETIC FIELD I. GRUIA, L.C. CIOBOTARU* University of Bucharest, Faculty of

More information

vacuum analysis plasma diagnostics surface science gas analysis

vacuum analysis plasma diagnostics surface science gas analysis Hiden EQP Systems High Sensitivity Mass and Energy Analysers for Monitoring, Control and Characterisation of Ions, Neutrals and Radicals in Plasma. vacuum analysis surface science gas analysis plasma diagnostics

More information

Physics of and in Ion Traps

Physics of and in Ion Traps Physics of and in Ion Traps Proposed Topics: TRIUMF, Vancouver June 01 Basics of Paul- and Penning-traps (equ. of motion, trap geometries, influence of trap imperfections,) Ion detection and cooling (Buffer

More information

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19 Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. Instructor s Manual CHAPTER 19 19-1. In a continuous wave NMR experiment, the intensity of the absorption signal is monitored

More information