ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW
|
|
- Augustine Cooper
- 6 years ago
- Views:
Transcription
1 ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW
2
3
4
5
6
7
8
9
10
11
12 Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which is seeded with energetic electrons - A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma - A steady-state plasma is produced when the rate at which electrons are released by ionizing collisions equals the rate at which they are lost by recombination. - A bluish-white light is characteristic of Ar ICP plasmas
13 Glass (quartz) torch Sample (aerosol) Plasma Argon gas RF coil (copper)
14 - Why Argon (Ar)? Plasma - It is an inert or noble gas, thus not explosive when subjected to an RF magnetic field or spark - It is relatively cheap to manufacture since the Ar is extracted directly from the atmosphere - The stream of Ar gas is usually between 8 to 20 Litres per minute
15 Plasma Gases (Ar)
16 Plasma - Gases All gas flows are controlled by Mass Flow Controllers
17 Plasma RF (radio frequency) generator RF power directed to the load coil from the RF generator creates an oscillating current in the coil - RF coil operates at a frequency of 27 to 40 MHz (mega-hertz) - Creates magnetic field in the region between the coils
18
19
20 Plasma It s desirable to have the RF generator be able to deliver its maximum power to the coil this is possible if the impedance of the load (the coil) is the same as the impedance of the source (the RF generator) The matching network tries to ensure they are properly matched
21 Plasma Matching network if the coil and RF generator are not matched, then a standing wave will be generated along the line The RF voltage and current will vary along the line and if the mismatch is great enough, the current/voltage can exceed the rated capacity of the line
22 Plasma Watts to sustain plasma All ICPs will have a readout of the forward and reflected power Forward power is ~ the power generated by the RF generator Reflected power is power not used if the load generated by the RF generator is not matched at the coil In ICP s, a high reflected power value indicates that the matching network is not functioning properly and that excess power may be present in the line
23 Plasma Spark initially ionizes some Ar atoms Free electrons are accelerated by the magnetic field This process of energy addition through the use of the RF energy supplied by the coil is the induction part of ICP High-energy electrons collide with other atoms causing further ionization Results in a chain reaction
24 Plasma
25 Components of a Plasma Induction region (IR) - region where most of the energy from the coils is coupled to the plasma. Hottest region ~ 10,000K (PHZ) - pre-heating zone. Region where cool aerosol from nebulizer punches a hole in plasma. Coolest region ~ K Initial radiation zone (IRZ). Region where excitation and ionization take place. ~ 7000K Normal analytical zone (NAZ). Region where ICP-OES wavelengths are measured. NOT necessarily the best place for ICP-MS though!
26 Components of a Plasma
27 Advantages of a Plasma source Mass Spectrometer high temperature good for ionization reduction of interferences stability low background low detection limits inherently multi-element wide calibration range
28
29
30
31
32
33
34 BEYOND THE PLASMA
35 ION PATH Ions move from high pressure to progressively lower pressures through small orifices 1. Plasma = 1000 mbar (atmospheric pressure) 2. Interface region (expansion stage) ~ 2 mbar during pumping of the interface pump (mechanical rotary pump) 3. Intermediate stage (lenses) ~ 2x10-4 mbar 4. Analyzer stage (MS) ~ 2x10-7 mbar
36 Element 2 High Resolution ICP-MS
37 Vacuum readings Element 2 Stage 1: mechanical pumping of expansion chamber Stage 2: mechanical roughing pump (fore pump) for Turbo Pump A - evacuates first part of lens stack Stage 3: Turbo Pump B - evacuates second part of lens stack Stage 4: Turbo Pump C - evacuates flight tube Stage 5: Turbo Pump D evacuates electrostatic analyzer
38
39
40 Vacuum readings Element 2 Stage 1: 2 mbar Stage 2: 3x10-3 mbar Stage 3: 8x10-5 mbar Stage 4: 5x10-7 mbar Stage 5: <1x10-7 mbar
41 ION OPTICS acceleration, focusing the plasma is a hostile environment - hot, atmospheric pressure, lots of ions ICP-MS: how do we get the desirable ions to the MS?
42
43 ION OPTICS acceleration, focusing, detection After being generated in the plasma in front of the sample cone, the ions pass through the small orifices of the cones (sampler and skimmer). In the analyzer housing, the ions get attracted and accelerated by the potential of the Extraction lens and then follow the Transfer optics, which shape and focus the ion beam to the Entrance Slit. The ion beam passes the Magnetic sector for mass separation, and the Electrostatic sector (ESA) for energy separation and energy focusing. Ion detection behind the Exit Slit is realized by a Conversion Dynode and an off-axis Secondary Electron Multiplier (SEM).
44
45 Interface Sampler & Skimmer cones Cones are made of nickel (Ni) most of the time, but they could also be made of Pt, Cu or Al. The metal must be characterized by high thermal conductivity, otherwise it will melt! A high melting point is therefore important and it should also be as hard as possible. However thermal conductivity seems to be the best criteria with regards to performance/price ratio
46 Sampler & Skimmer Cones Material Thermal Conductivity (Wm -1 K -1 ) Melting Point (degrees C) Hardness Al soft Cu soft Ni hard Pt hard
- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma
THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which
More informationENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques
ENVG 60500 FALL 2013 ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques HISTORY In the 1940s, arc and high-voltage spark spectrometry became widely utilized for metal analysis
More informationInterface (backside) & Extraction Lens
Plasma Interface Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control
More informationPRINCIPLE OF ICP- AES
INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively
More informationThe Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)
The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6
More informationSpectrometric Methods of Analysis. OCN 633 Fall 2013
Spectrometric Methods of Analysis OCN 633 Fall 2013 Plasma Emission and Plasma Mass Spectroscopy Two fields of elemental analysis undergoing the most study Myriad analytical applications Three categories
More informationSuzanne Bell Second Edition
Forensic hemistry Suzanne Bell Second Edition Pearson Education Limited Edinburgh Gate Harlow Essex M20 2JE England and Associated ompanies throughout the world Visit us on the World Wide Web at: www.pearsoned.co.uk
More informationINDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS
More informationThe 30-Minute Guide to ICP-MS
TECHNICAL NOTE The 30-Minute Guide to ICP-MS ICP - Mass Spectrometry A Worthy Member of the Atomic Spectroscopy Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More informationCHAPTER 4: ANALYTICAL INSTRUMENTATION
CHAPTER 4: ANALYTICAL INSTRUMENTATION 4.1 INTRODUCTION In this section, a review of the analytical instrumentation used during sample preparation and analysis is presented which includes an overview of
More informationIntroduction to the Q Trap LC/MS/MS System
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap
More informationICP-MS. High Resolution ICP-MS.
ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively
More informationChapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques
Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are
More informationIntroduction to Elemental Analysis
Introduction to Elemental Analysis A dust storm charges across the Sahara desert, whipping up sand and blowing it west, towards the ocean. Meanwhile, in the middle of the Atlantic Ocean, a team of scientist
More informationHands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment
: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it
More informationIonization Techniques Part IV
Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector
More informationUse of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark
Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES
More informationDEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD
Chapter 4 DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD 4.1 INTRODUCTION Sputter deposition process is another old technique being used in modern semiconductor industries. Sputtering
More informationApplication Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP
Application Note MBMS for Preformed Ions, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster Presented at 45th ASMS Conference on Mass Spectrometry, June 1-5, 1997) In order to accurately characterize a plasma
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationICPMS Doherty Lecture 1
ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms
More informationOES - Optical Emission Spectrometer 2000
OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the
More informationAtomic Emission Spectroscopy
Atomic Emission Spectroscopy Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:
More informationAnalysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS
TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically
More informationApplications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry
Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS
More informationOptical Atomic Spectroscopy
Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state
More informationMass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.
Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted
More informationMulti-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS
Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction
More informationa. An emission line as close as possible to the analyte resonance line
Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal
More informationAtomic Spectroscopy AA/ICP/ICPMS:
Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationELECTROMAGNETIC WAVES
VISUAL PHYSICS ONLINE MODULE 7 NATURE OF LIGHT ELECTROMAGNETIC WAVES SPECTRA PRODUCED BY DISCHARGE TUBES CATHODE RAYS (electron beams) Streams of electrons (negatively charged particles) observed in vacuum
More informationHydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES
Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with
More informationInductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Introduction What is ICP-MS? Inductively Coupled Plasma Mass Spectrometry Mass spectrometry
More informationVacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular
Vacuum Technology Vacuum Pumps Two general classes exist: Gas transfer physical removal of matter Mechanical, diffusion, turbomolecular Adsorption entrapment of matter Cryo, sublimation, ion Mechanical
More informationThermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 for Certified, Refurbished Lab Equipment Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER TECHNICAL
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationDiscovered by German scientist Johann Hittorf in 1869 and in 1876 named by Eugen Goldstein.
DO PHYSICS ONLINE CATHODE RAYS CATHODE RAYS (electron beams) Streams of electrons (negatively charged particles) observed in vacuum tubes - evacuated glass tubes that are equipped with at least two metal
More informationDirect Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ
Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction
More informationLaser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements
Practicum Spectroscopy Fall 2010 LIM Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements Jorge Ferreiro, study degree Chemistry, 5 th semester,
More informationAN INTRODUCTION TO ATOMIC SPECTROSCOPY
AN INTRODUCTION TO ATOMIC SPECTROSCOPY Atomic spectroscopy deals with the absorption, emission, or fluorescence by atom or elementary ions. Two regions of the spectrum yield atomic information- the UV-visible
More informationExtrel Application Note
Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical
More informationMass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers
More informationICP-MS. Inductively Coupled Plasma Mass Spectrometry. A Primer
ICP-MS Inductively Coupled Plasma Mass Spectrometry A Primer Table of Contents ICP-MS Primer Section 1 Introduction to ICP-MS 1 History and Development of ICP-MS 2 Agilent Technologies - History in ICP-MS
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationBRIEFING 1730 THEORY AND PRACTICE OUTLINE
40(6) In-Process Revision: PLASMA SPECTROCHEMISTRY-THEORY AND... Page 1 of 15 BRIEFING 1730 Plasma Spectrochemistry Theory and Practice. The Chemical Analysis Expert Committee proposes this new
More informationHuashun Zhang. Ion Sources. With 187 Figures and 26 Tables Э SCIENCE PRESS. Springer
Huashun Zhang Ion Sources With 187 Figures and 26 Tables Э SCIENCE PRESS Springer XI Contents 1 INTRODUCTION 1 1.1 Major Applications and Requirements 1 1.2 Performances and Research Subjects 1 1.3 Historical
More informationLecture 10. Vacuum Technology and Plasmas Reading: Chapter 10. ECE Dr. Alan Doolittle
Lecture 10 Vacuum Technology and Plasmas Reading: Chapter 10 Vacuum Science and Plasmas In order to understand deposition techniques such as evaporation, sputtering,, plasma processing, chemical vapor
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationDETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS
7 th INTERNATIONAL MULTIDISCIPLINARY CONFERENCE Baia Mare, Romania, May 17-18, 2007 ISSN-1224-3264 DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS Mariana Dobra, Vasile Viman,
More informationhigh temp ( K) Chapter 20: Atomic Spectroscopy
high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment
More informationChemistry Instrumental Analysis Lecture 17. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 17 Introduction to Optical Atomic Spectrometry From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry
More informationTrace elemental analysis solutions for your application. June 6, 2018
Trace elemental analysis solutions for your application June 6, 2018 Understanding how each technique works Components of instrument Selection Criteria Application Fields OUTLINE All these techniques can
More informationIon Beam Sources for Neutral Beam Injectors: studies and design for components active cooling and caesium ovens
Ion Beam Sources for Neutral Beam Injectors: studies and design for components active cooling and caesium ovens Andrea Rizzolo Consorzio RFX, Padova, Italy andrea.rizzolo@igi.cnr.it Advanced Physics Courses,
More informationAgilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent
More informationof mass spectrometry
Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules
More informationLecture 4. Ultrahigh Vacuum Science and Technology
Lecture 4 Ultrahigh Vacuum Science and Technology Why do we need UHV? 1 Atmosphere = 760 torr; 1 torr = 133 Pa; N ~ 2.5 10 19 molecules/cm 3 Hertz-Knudsen equation p ZW 1/ 2 ( 2mk T) At p = 10-6 Torr it
More informationHarris: Quantitative Chemical Analysis, Eight Edition
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of
More informationICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer
Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full
More informationDetermination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity
More informationAtomic Absorption & Atomic Fluorescence Spectrometry
Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,
More informationPrevious Lecture. Electron beam lithoghraphy e - Electrons are generated in vacuum. Electron beams propagate in vacuum
Previous Lecture Electron beam lithoghraphy e - Electrons are generated in vacuum Electron beams propagate in vacuum Lecture 6: Vacuum & plasmas Objectives From this vacuum lecture you will learn: What
More information6.5 Optical-Coating-Deposition Technologies
92 Chapter 6 6.5 Optical-Coating-Deposition Technologies The coating process takes place in an evaporation chamber with a fully controlled system for the specified requirements. Typical systems are depicted
More informationTable of Content. Mechanical Removing Techniques. Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB)
Table of Content Mechanical Removing Techniques Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB) Ultrasonic Machining In ultrasonic machining (USM), also called ultrasonic grinding,
More informationMultiple charge states of titanium ions in laser produced plasma
PRAMANA cfl Indian Academy of Sciences Vol. 55, Nos 5 & 6 journal of Nov. & Dec. 2000 physics pp. 781 787 Multiple charge states of titanium ions in laser produced plasma M SHUKLA, S BANDHYOPADHYAY, V
More informationInstrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai
303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in
More informationLecture 16 Instrumentation for ICP AES-VIII-Instruments
Inductive Couple Plasma Atomic Emission Spectrometry (ICP-AES) for Pollution Monitoring Dr. J R Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Lecture 16 Instrumentation
More informationOptics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P
Optics Spectrometer Prism spectrometer LD Physics Leaflets P5.7.1.1 Measuring the line spectra of inert gases and metal vapors using a prism spectrometer Objects of the experiment Adjusting the prism spectrometer.
More informationA novel sputtering technique: Inductively Coupled Impulse Sputtering (ICIS)
IOP Conference Series: Materials Science and Engineering A novel sputtering technique: Inductively Coupled Impulse Sputtering (ICIS) To cite this article: D A L Loch and A P Ehiasarian 2012 IOP Conf. Ser.:
More informationPlasma Interface Element2
Plasma Interface Element2 Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS IF this positive potential is great enough, a discharge from the plasma into the orifice is observed
More informationInterference-free determination of ultra-trace levels of Arsenic and Selenium using methyl fluoride as reaction gas in ICP-MS/MS
Department of Analytical Chemistry Research group Atomic and Mass Spectrometry Interference-free determination of ultra-trace levels of Arsenic and Selenium using methyl fluoride as reaction gas in ICP-MS/MS
More informationARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision
ARGUS VI Static Vacuum Mass Spectrometer Static Vacuum ARGUS VI Multicollection Low Volume Precision Based on more than 20 years of experience in noble gas mass spectrometry instrumentation, we have developed
More informationHELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer
ARGUS VI HELIX SFT Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX SFT Low Volume Multicollection High Mass Resolution 3 Based on more than 20 years of experience in
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More information2101 Atomic Spectroscopy
2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.
More information10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber
CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described
More informationOptimizing Analytical Performance in ICP-OES Applications
application note Optimizing Analytical Performance in ICP-OES Applications Introduction To obtain the best possible performance from an analytical instrument, it is necessary to optimize the operating
More informationChemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:
More informationChemistry Instrumental Analysis Lecture 34. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry
More informationChemistry 311: Topic 3 - Mass Spectrometry
Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These
More informationElectron beam scanning
Electron beam scanning The Electron beam scanning operates through an electro-optical system which has the task of deflecting the beam Synchronously with cathode ray tube which create the image, beam moves
More informationCEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationDetermination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS
APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords
More informationDetermination of Trace Metals in Human Urine Using the NexION 300 ICP-MS
application Note ICP Mass Spectrometry Authors: David Bass Daniel Jones PerkinElmer, Inc. Shelton, CT USA Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS Abstract This application
More informationICP-MS 基本原理. 陳麗真 Jennifer Chen Inorganic Products Specialist
ICP-MS 基本原理 陳麗真 Jennifer Chen Inorganic Products Specialist Email:Jennifer.Chen@perkinelmer.com What is the ICP-MS Inductively Coupled Plasma Mass Spectrometry can analyze almost all elements in the periodic
More information1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level)
1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level) Electromagnetic induction (Chapter 23): For a straight wire, the induced current or e.m.f. depends on: The magnitude of the magnetic
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The
More information20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n
20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer
More informationLiberty Series II. Liberty. Liberty Series II. Analytical Methods book. Publication No October (i) Publication date: 10/99
Liberty Series II Liberty Liberty Series II Analytical Methods book Publication No. 85 101500 00 October 1999 Publication date: 10/99 (i) Liberty Series II Varian offices Varian has offices in most countries.
More informationThe ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode
The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro
More informationCharacterization of the operation of RITs with iodine
Characterization of the operation of RITs with iodine IEPC-2017-368 Presented at the 35th International Electric Propulsion Conference Georgia Institute of Technology Atlanta, Georgia USA Waldemar Gärtner
More informationTrace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell
Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle
More informationThermo Scientific icap RQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry
More informationPartial Energy Level Diagrams
Partial Energy Level Diagrams 460 nm 323 nm 610 nm 330 nm 819 nm 404 nm 694 nm 671 nm 589 / 590 nm 767 / 769 nm Lithium Sodium Potassium Gas Mixtures Maximum Temperatures, C Air-Coal Gas 1825 Air-Propane
More informationSpeed of Light in Air
Speed of Light in Air Electromagnetic waves represent energy in the form of oscillating electric and magnetic fields which propagate through vacuum with a speed c = 2.9979246x10 8 m/s. Electromagnetic
More information