CHAPTER D3 TOF ION OPTICS

Size: px
Start display at page:

Download "CHAPTER D3 TOF ION OPTICS"

Transcription

1 Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide Summary Background EquationsofMotionofIons Resolution Reflectron Comparison with other mass spectrometers Conclusion Micromass UK Limited Page 397

2 This page is intentionally blank. Micromass UK Limited Page 398

3 Quick Guide In a Time-Of-Flight (TOF) mass spectrometer, ions formed in an ion source are extracted and accelerated to a high velocity by an electric field into an analyser consisting of a long straight drift tube. The ions pass along the tube until they reach a detector. After the initial acceleration phase, the velocity reached by an ion is inversely proportional to its mass (strictly, inversely proportional to the square root of its m/z value). Since the distance from the source to the detector is fixed, the time taken for an ion to traverse the analyser in a straight line is proportional to its velocity and hence its mass (strictly, proportional to the square root of its m/z value). Thus, each m/z value has its characteristic time-of-flight from the source to the detector. In effect, the ions race each other along the drift tube but the winners are always the ions of smallest m/z value since these have the shortest flight times. The last to arrive at the detector are always those of greatest mass which have the longest flight times. However, as in a race, for there to be a separation at the finish line (the detector), the ions must all start from the ion source at the same time (no handicapping allowed!). The times taken for ions of differing m/z values to reach the detector are of the order of a few microseconds and the separation in times of arrival for ions of differing m/z value is less than this. Thus, if ions of different mass are to be separated adequately in a time domain (good resolution), they should all start from the ion source at exactly the same time or, more practically, within a few nanoseconds of each other. Ions for TOF mass spectrometry must be extracted from the ion source in instantaneous pulses. Therefore, either ions are produced continuously but are extracted from the source in pulses or ions are produced directly in pulses. TOF mass spectrometry is ideally suited to those ionization methods that inherently produce ions in pulses, as with pulsed laser desorption or Cf-radionuclide ionization. Micromass UK Limited Page 399

4 This page is intentionally blank. Micromass UK Limited Page 400

5 There is no theoretical upper limit on m/z that can be examined and TOF mass spectrometry is useful for substances having very high molecular mass. In practice, the current upper limit is about 350,000. Unfortunately, ions even of the same m/z value do have a spread of velocities after acceleration and therefore the resolution achievable with TOF is not very high because bunches of ions of one m/z value overlap those at the next m/z value. By use of an electrostatic ion mirror, called a reflectron the velocities of ions of the same m/z value can be made more nearly equal, thereby making their arrival times at the detector more nearly equal and so improving resolution. After reflection in the reflectron, the ions must pass down a second length of analyser set at a small angle to the first so as to reach the detector. The improvement in resolution with the reflectron is achieved at the expense of some loss in overall sensitivity due to loss of ions in the reflectron and in the second length of analyser. For very high mass, when sensitivity is frequently critical, the reflectron is not used and lower resolution is accepted. The mass spectrum gives the abundances of ions for different times of arrival at the detector. Since the times are proportional to the square root of the m/z values, it is simple to convert the arrival times into m/z values. Summary After acceleration through an electric field, ions pass ( drift ) along a straight length of analyser under vacuum and reach a detector after a time which depends on the square root of their m/z values. The mass spectrum is a record of the abundances of ions and the times (converted to m/z) they have taken to traverse the analyser. TOF mass spectrometry is valuable for its fast response time, especially for substances of high mass which have been ionized in pulses. Micromass UK Limited Page 401

6 Ion source Acceleration region Drift region Detector (a) (b) (c) (d) (e) Figure 1 The essentials of time-of-flight optics. In (a), a pulse of ions is formed and then accelerated out of the source (b) into the drift region (c). After a short time (d), the ions have separated along the drift region according to m/z value. In (e), the ions with smallest m/z value (fastest moving) begin arriving first at the detector, to be followed by the ions of gradually increasing m/z. Micromass UK Limited Page 402

7 TIME-OF-FLIGHT ION OPTICS Background Whereas understanding the fundamentals of the ion optics for magnetic and electric sector, quadrupole, ion trap and FT ion cyclotron resonance mass spectrometers ranges from fairly simple to difficult, the basic ion optics of Time-Of-Flight (TOF) instruments is very straightforward. Basically, ions are extracted from the ion source in short pulses and directed down an evacuated straight tube to a detector. The time taken to travel the length of the drift or flight tube depends on the mass of the ion and its charge. For singly charged ions (z = 1;m/z= m), the time taken to traverse the distance from the source to the detector is proportional to a function of mass, the greaterthemassoftheionthesloweritisinarrivingatthedetector. Thus, there are no electric or magnetic fields to constrain the ions into curved or complicated trajectories. After initial acceleration, the ions pass in a straight line, at constant speed, to the detector. The arrival of the ions at the detector is recorded in the usual way as a trace of ion abundance against time of arrival, the latter being converted into a mass scale to give the final mass spectrum. Equations of Motion of Ions A short pulse of ions is extracted from the ion source (Figure 1). It is necessary to use a pulse because, using only time to differentiate amongst the masses, it is important that the ions all leave the ion source at the same instant (like runners in a race on hearing a starting pistol). The first step is acceleration through an electric field (E volts). With the usual nomenclature (m = mass, z = number of charges on an ion, e = thechargeonanelectron,v= the final velocity reached on acceleration), the kinetic energy (mv 2 /2) of the ion is given by equation (1). mv 2 /2 = z.e.e (1) Equation (2) follows by simple rearrangement. v = (2z.e.E/m) (2) Micromass UK Limited Page 403

8 Slowest ions Single m/z value (a) Number of ions Fastest ions Spread in velocity Number of ions m/z Overlap (m+1)/z (b) Spead in velocity Number of ions m/z (m+1)/z (c) Increased overlap Spead in velocity (d) m/z (m+200)/z Number of ions Increased overlap Spead in velocity Figure 2 (a) Each set of ions at any one m/z value will have a small spread in speed because they are not all formed at exactly the same place in the ion source. (b) Two sets of ions separated by one mass unit (m, m+1) overlap because the slower ions of smaller mass overlap the faster ions of greater mass. (c) For larger m/z values, this effect leads to almost total disappearance of unit resolution. (d) At still greater m/z values, even mass differences of 200 or more may not be separated. Micromass UK Limited Page 404

9 If the distance from the ion source to the detector is d, then the time (t) taken for an ion to traverse the drift tube is given by equation (3). t = d/v = d/ (2z.e.E/m) = d.[ (m/z)]/ (2e.E) (3) In equation (3), d is fixed, E is held constant in the instrument and e is a universal constant. Thus, the flight time of an ion t is directly proportional to the square root of m/z (equation 4). t = (m/z) x a constant (4) Equation (4) shows that an ion of m/z 100 will take twice as long to reach the detector as an ion of m/z 25: (t 100 /t 25 = 100/ 25 = 10/5 = 2) Resolution Generally, the attainable resolving power of a TOF instrument is limited, particularly at higher mass. There are two major reasons for this, one inherent in the technique, the other a practical problem. First, the flight times are proportional to the square root of m/z. The difference in the flight times (t m and t m+1 ) for two ions separated by unit mass is given by equation (5). t m - t m+1 = t = [( m/z)-( (m+1)/z] x a constant (5) As m increases, t becomes progressively smaller (compare the difference between the square roots of 1 and 2 (= 0.4) with the difference between 100 and 101 (= 0.05). Thus, the difference in arrival times of ions arriving at the detector become smaller and smaller and more difficult to differentiate. This inherent problem is a severe restriction even without the second difficulty which is due to the fact that not all ions of any one given m/z value reach the same velocity after acceleration nor are they all formed at exactly the same point in the ion source. Therefore, even for any one m/z value, the ions reach the detector over an interval of time instead of all at one time. Clearly, where separation of flight times is very short, as with TOF instruments, the spread for individual ion m/z values means there will be overlap in arrival times between ions of closely similar m/z values. This effect (Figure 2) decreases available (theoretical) resolution but it can be ameliorated by modifying the instrument to include a reflectron (see below). Micromass UK Limited Page 405

10 Ion source Acceleration region Drift region Reflection (a) Detector (b) (c) (d) (e) (f) Figure 3 In (a), a pulse of ions is formed but, for illustration purposes, all with the same m/z value. In (b), the ions have been accelerated but, because they were not all formed in the same space, they are separated in time and velocity, some ions having more kinetic energy than others. In (c), the ions approach the ion mirror or reflectron which they then penetrate to different depths, depending on their kinetic energies (d). The ones with greater kinetic energy penetrate furthest. In (e), the ions leave the reflectron in a bunch (i.e., no longer separated in time) and travel on to the detector (f). The path taken by the ions is indicated by the dotted line in (f). Micromass UK Limited Page 406

11 Reflectron A homogeneous electrostatic field is placed at the end of the flight path of the ions and has a polarity the same as that of the ions, viz., positive (negative) ions experience a retarding positive (negative) potential. The ions come to a stop and are then accelerated in the opposite direction. The ions are reflected by the ion mirror or reflectron. The ion mirror is often at a slight angle (Figure 3) to the line of flight of the ions and, when reflected, the ions do not travel back along the same path but along a slightly deflected line (some instruments reflect ions back along the path they took to the reflection). Consider ions of any one given m/z value. The faster ions, having greater kinetic energy, travel further into the electrostatic field before being reflected than do slower ions. As a result, the faster ions spend slightly more time within the reflectron than do the slower ones. This causes the faster and slower ions to bunch up so that they leave the reflectron closer together and travel onto the detector, where the ions arrive close together. There is a similar effect for all other individual m/z values and overall resolution is greatly improved. Instead of a typical TOF resolution of about 1000, resolutions of around 10,000 can be achieved. There is a disadvantage in using a reflectron in that the sensitivity of the instrument is decreased through ion loss by collision and dispersion from the main beam. It is a more severe problem for ions of large mass and, for these, the reflectron is often not used despite the better resolution attainable with its use Comparison with other mass spectrometers TOF mass spectrometers are very robust and usable with a wide variety of ion sources and inlet systems. Having only simple electrostatic and no magnetic fields, their construction, maintenance and calibration are usually straightforward. There is no upper theoretical mass limitation; all ions can be made to proceed from source to detector. In practice, there is a mass limitation in that it becomes increasingly difficult to discriminate between times of arrival at the detector as the m/z value becomes large. This effect coupled with the spread in arrival times for any one m/z value means that discrimination between unit masses becomes difficult at say m/z At m/z 50,000 overlap of 50 mass units is more typical, i.e., mass accuracy is no better than about mass units (Figure 2). Nevertheless, the ability of a TOF instrument to measure routinely and simply masses this large gives it a decided advantage over many other types of mass analyser. Micromass UK Limited Page 407

12 Ion source Acceleration region Drift region Detector (a) (b) (c) (d ) (e) Figure 4 In this mode, ions are formed continuously in the ion source (a) but the electrostatic accelerating potential is applied in pulses (b). Thus, a sample of ions is drawn into the drift region (c) with more ions being formed in the source. As shown in figure 1, the ions separate according to m/z values (d) and arrive at the detector (e), the ions of largest m/z arriving last. Micromass UK Limited Page 408

13 A major advantage of the TOF mass spectrometer is its fast response time and its applicability to ionization methods that produce ions in pulses. As discussed earlier, because all ions follow the same path, all ionsneedtoleavetheionsourceatthesametimeifthereistobeno overlap between m/z values at the detector. In turn, this means that, if ions are produced continuously as in a typical electron ionization source, then samples of these ions must be utilized in pulses by switching the ion extraction field on and off very quickly (Figure 4). On the other hand, there are some ionization techniques that are very useful, particularly at very high mass, but only produce ions in pulses. For these sources, the ion extraction field can be left on continuously. Two prominent examples are Californium radionuclide and laser desorption ionization. In the former, nuclear disintegration occurs within a very short time frame to give a short pulse of ions; the same disintegration is used to start the timer ( stopwatch ) for the race of the ions down the flight tube. Similarly, a laser pulse lasting only a few nanoseconds produces a bunch of ions and acts to start the timer also. Conclusion By accelerating a pulse of ions into a flight or drift tube, the ions achieve different velocities depending on their individual m/z values; the bigger the m/z value, the slower the ion travels down the tube. A detector at the end of the tube records the times of arrival of the ions and hence their flight times. The flight times are easily converted into m/z information by a simple formula. The Time-of-Flight mass spectrometer is particularly useful for ions produced in pulses and where fast response times are needed. It can be used to detect ions of very large m/z values. Micromass UK Limited Page 409

14 This page is intentionally blank. Micromass UK Limited Page 410

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Chemistry Instrumental Analysis Lecture 35. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

Time-of-Flight Mass Analyzers

Time-of-Flight Mass Analyzers Time-of-Flight Mass Analyzers Jonathan Karty C613 lecture 1 March 6, 8 (Section 4. in Gross, pages 115-18) TOF Overview Time-of-flight (TOF) is the least complex mass analyzer in terms of its theory Ions

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI Back to Basics Section A: Ionization Processes CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI TABLE OF CONTENTS Quick Guide...27 Summary...29 The Ionization Process...31 Other Considerations on Laser

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced I. Mass spectrometry: capable of providing both quantitative and qualitative information about samples as small as 100 pg (!) and with molar masses in the 10 4-10 5 kdalton range A. The mass spectrometer

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY international UNION OF PURE AND APPLIED CHEMISTRY ANALYTICAL CHEMISTRY DIVISION COMMISSION ON ANALYTICAL NOMENCLATURE RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY RULES APPROVED 1973 LONDON BUTTER

More information

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky 1 Secondary Ion Mass Spectrometry (SIMS) Thomas Sky Depth (µm) 2 Characterization of solar cells 0,0 1E16 1E17 1E18 1E19 1E20 0,2 0,4 0,6 0,8 1,0 1,2 P Concentration (cm -3 ) Characterization Optimization

More information

CHM 424 EXAM 4 CRIB - COVER PAGE FALL

CHM 424 EXAM 4 CRIB - COVER PAGE FALL CHM 44 EXAM 4 CRIB - COVER PAGE FALL 007 There are six numbered pages with five questions. Answer the questions on the exam. Exams done in ink are eligible for regrade, those done in pencil will not be

More information

Theory English (Official)

Theory English (Official) Q3-1 Large Hadron Collider (10 points) Please read the general instructions in the separate envelope before you start this problem. In this task, the physics of the particle accelerator LHC (Large Hadron

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

v t 2 2t 8. Fig. 7 (i) Write down the velocity of the insect when t 0. (ii) Show that the insect is instantaneously at rest when t 2and when t 4.

v t 2 2t 8. Fig. 7 (i) Write down the velocity of the insect when t 0. (ii) Show that the insect is instantaneously at rest when t 2and when t 4. 1 Fig. 7 is a sketch of part of the velocity-time graph for the motion of an insect walking in a straight line. Its velocity, v ms 1, at time t seconds for the time interval 3 t 5 is given by v ms -1 v

More information

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby j5 1 Mass Analysis Gene Hart-Smith and Stephen J. Blanksby 1.1 Introduction Modern day mass analyzer technologies have, together with soft ionization techniques, opened powerful new avenues by which insights

More information

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION Columbia Physics: Lab -6 (ver. 10) 1 EXPERMENT -6 e/m OF THE ELECTRON GENERAL DSCUSSON The "discovery" of the electron by J. J. Thomson in 1897 refers to the experiment in which it was shown that "cathode

More information

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Spectrometry. Fundamental LC-MS. Mass Analysers Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

PARTICLE ACCELERATORS

PARTICLE ACCELERATORS VISUAL PHYSICS ONLINE PARTICLE ACCELERATORS Particle accelerators are used to accelerate elementary particles to very high energies for: Production of radioisotopes Probing the structure of matter There

More information

P Q 2 = -3.0 x 10-6 C

P Q 2 = -3.0 x 10-6 C 1. Which one of the following represents correct units for electric field strength? A. T B. N/C C. J / C D. N m 2 /C 2 2. The diagram below shows two positive charges of magnitude Q and 2Q. P Q 2Q Which

More information

(Total for Question = 5 marks) PhysicsAndMathsTutor.com

(Total for Question = 5 marks) PhysicsAndMathsTutor.com 1 Rutherford designed an experiment to see what happened when alpha particles were directed at a piece of gold foil. Summarise the observations and state the conclusions Rutherford reached about the structure

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

1.1 Atomic structure

1.1 Atomic structure 1.1 Atomic structure History of the atom The model of the atom has changed as our observations of its behavior and properties have increased. A model is used to explain observations. The model changes

More information

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS)

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS) 5.16 Incident Ion Techniques for Surface Composition Analysis 5.16.1 Ion Scattering Spectroscopy (ISS) At moderate kinetic energies (few hundred ev to few kev) ion scattered from a surface in simple kinematic

More information

1. Draw in the magnetic field inside each box that would be capable of deflecting the particle along the path shown in each diagram.

1. Draw in the magnetic field inside each box that would be capable of deflecting the particle along the path shown in each diagram. Charged Particles in Magnetic Fields 1. Draw in the magnetic field inside each box that would be capable of deflecting the particle along the path shown in each diagram. a b c d 2. a. Three particles with

More information

The Analyzer. We will deal mainly with magnetic sector instruments here although we will talk briefly about other separation methods later.

The Analyzer. We will deal mainly with magnetic sector instruments here although we will talk briefly about other separation methods later. The Analyzer The analyzer is where the beam of ions generated in the ion source is separated into multiple beams each representing a single charge to mass ratio (ideally). We will deal mainly with magnetic

More information

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality?

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Laure Joly & Vincent Hanot WIV-ISP, Juliette Wytsmanstraat 14, 15 Brussel The passengers pesticides are requested......

More information

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i Gel Permeation Chromatography (GPC) : Size Exclusion Chromatography GPC : 1. Chromatogram (V R vs H) H i Detector response Baseline N i M i 130 135 140 145 150 155 160 165 Elution volume (V R ) (counts)

More information

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap 1 Last Time Mass Analyzers; CAD and TOF mass analyzers: 2 Fourier Transforms A transform is when you change your analytical space without

More information

AQA Chemistry A-Level : Atomic Structure

AQA Chemistry A-Level : Atomic Structure AQA Chemistry A-Level 3.1.1: Atomic Structure Detailed Notes 3.1.1.1 - Fundamental Particles The model for atomic structure has evolved over time as knowledge and scientific understanding changes. Plum

More information

Photoelectron Spectroscopy using High Order Harmonic Generation

Photoelectron Spectroscopy using High Order Harmonic Generation Photoelectron Spectroscopy using High Order Harmonic Generation Alana Ogata Yamanouchi Lab, University of Tokyo ABSTRACT The analysis of photochemical processes has been previously limited by the short

More information

This is the total charge on an ion divided by the elementary charge (e).

This is the total charge on an ion divided by the elementary charge (e). 12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+. Mass spectrometry is the study of systems causing the formation of gaseous ions, with or without fragmentation, which are then characteried by their mass to charge ratios (m/) and relative abundances.

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

PhysicsAndMathsTutor.com 1

PhysicsAndMathsTutor.com 1 PhysicsAndMathsTutor.com 1 1. Millikan determined the charge on individual oil droplets using an arrangement as represented in the diagram. The plate voltage necessary to hold a charged droplet stationary

More information

Secondary ion mass spectrometry (SIMS)

Secondary ion mass spectrometry (SIMS) Secondary ion mass spectrometry (SIMS) Lasse Vines 1 Secondary ion mass spectrometry O Zn 10000 O 2 Counts/sec 1000 100 Li Na K Cr ZnO 10 ZnO 2 1 0 20 40 60 80 100 Mass (AMU) 10 21 10 20 Si 07 Ge 0.3 Atomic

More information

Laser Dissociation of Protonated PAHs

Laser Dissociation of Protonated PAHs 100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3

More information

ANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION

ANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION ANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION Conventional Centrifugal Methods Centrifugal sedimentation of particles suspended in a fluid is a well known method (1, 2)

More information

Force on a Moving Charge in a Magnetic Field: Examples and Applications

Force on a Moving Charge in a Magnetic Field: Examples and Applications Force on a Moving Charge in a Magnetic Field: Examples and Applications Bởi: OpenStaxCollege Magnetic force can cause a charged particle to move in a circular or spiral path. Cosmic rays are energetic

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

PhysicsAndMathsTutor.com 1

PhysicsAndMathsTutor.com 1 PhysicsAndMathsTutor.com 1 Q1. (a) The diagram below shows a narrow beam of electrons produced by attracting electrons emitted from a filament wire to a metal plate which has a small hole in it. (i) Why

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectroscopy (SIMS) Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special

More information

A New, Universal Frame Of Reference

A New, Universal Frame Of Reference Rothwell Bronrowan physbron@t-online.de This paper begins with an example chosen to indicate the difficulties associated with determining which of several plausible, alternative frames of reference - if

More information

Force Due to Magnetic Field You will use

Force Due to Magnetic Field You will use Force Due to Magnetic Field You will use Units: 1 N = 1C(m/s) (T) A magnetic field of one tesla is very powerful magnetic field. Sometimes it may be convenient to use the gauss, which is equal to 1/10,000

More information

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER UNIT G485 Module 1 5.1.2 Magnetic Fields 11 Thus, in order for the particle to suffer NO DEFLECTION and so exit the device at Y : From which : MAGNETIC FORCE UP = ELECTRIC FORCE DOWN BQv = EQ THE MASS

More information

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 11, 1467 1477 (1997) Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass

More information

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: I. BASIC ELEMENTS OF A MASS SPECTROMETER Inlet System or Chromatograph Ion Source Mass Analyzer Detector Computer II. ION SOURCES A. Electron

More information

Accelerator Mass Spectroscopy

Accelerator Mass Spectroscopy Accelerator Mass Spectroscopy Accelerator Mass Spectroscopy (AMS) is a highly sensitive technique that is useful in isotopic analysis of specific elements in small samples (1mg or less of sample containing

More information

Today s lecture: Motion in a Uniform Magnetic Field continued Force on a Current Carrying Conductor Introduction to the Biot-Savart Law

Today s lecture: Motion in a Uniform Magnetic Field continued Force on a Current Carrying Conductor Introduction to the Biot-Savart Law PHYSICS 1B Today s lecture: Motion in a Uniform Magnetic Field continued Force on a Current Carrying Conductor Introduction to the Biot-Savart Law Electricity & Magnetism A Charged Particle in a Magnetic

More information

Appendix A2. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France.

Appendix A2. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France. Appendix A. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France. Prepared by: Arash Akbari-Sharbaf Why Build Accelerators? Probe deeper From

More information

Abstract. 1. Introduction. 2. Methodology

Abstract. 1. Introduction. 2. Methodology Abstract When a transverse electric field between a pair of closely spaced slits is used to deflect an ion beam, the length of the field region along the beam axis directly affects the length of time required

More information

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest

More information

Effect of Magnet Geometry on the Magnetic Component of the Lorentz Force Equation

Effect of Magnet Geometry on the Magnetic Component of the Lorentz Force Equation Effect of Magnet Geometry on the Magnetic Component of the Lorentz Force Equation Author: Singer, Michael Date: 1st May 2017 email: singer43212@gmail.com Page 1 Abstract All forces in the universe are

More information

Pumping Speed in the Drydown Zone

Pumping Speed in the Drydown Zone A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Pumping Speed in the Drydown Zone The extended pumpdown time through the water vapor-dominated drydown zone is a complex process

More information

Cesium Dynamics and H - Density in the Extended Boundary Layer of Negative Hydrogen Ion Sources for Fusion

Cesium Dynamics and H - Density in the Extended Boundary Layer of Negative Hydrogen Ion Sources for Fusion Cesium Dynamics and H - Density in the Extended Boundary Layer of Negative Hydrogen Ion Sources for Fusion C. Wimmer a, U. Fantz a,b and the NNBI-Team a a Max-Planck-Institut für Plasmaphysik, EURATOM

More information

1. The diagram shows the electric field lines produced by an electrostatic focussing device.

1. The diagram shows the electric field lines produced by an electrostatic focussing device. 1. The diagram shows the electric field lines produced by an electrostatic focussing device. Which one of the following diagrams best shows the corresponding equipotential lines? The electric field lines

More information

Dynamics of Ions in an Electrostatic Ion Beam Trap

Dynamics of Ions in an Electrostatic Ion Beam Trap Dynamics of Ions in an Electrostatic Ion Beam Trap Daniel Zajfman Dept. of Particle Physics Weizmann Institute of Science Israel and Max-Planck Institute for Nuclear Physics Heidelberg, Germany Charles

More information

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading: Assigned Reading: Nier mass spectrometry paper Extra Reading: 1) Introduction a) We are now on to the heart of the mass spectrometer. How do you separate ions of different mass, so that you can detect/quantify

More information

Test Review FQ3eso_U5_4_Electric field_test_review

Test Review FQ3eso_U5_4_Electric field_test_review Test Review FQ3eso_U5_4_Electric field_test_review Identify the letter of the choice that best completes the statement or answers the question. 1.- In which diagram do the field lines best represent the

More information

Section 4 : Accelerators

Section 4 : Accelerators Section 4 : Accelerators In addition to their critical role in the evolution of nuclear science, nuclear particle accelerators have become an essential tool in both industry and medicine. Table 4.1 summarizes

More information

Secondary Ion Mass Spectrometry (SIMS)

Secondary Ion Mass Spectrometry (SIMS) CHEM53200: Lecture 10 Secondary Ion Mass Spectrometry (SIMS) Major reference: Surface Analysis Edited by J. C. Vickerman (1997). 1 Primary particles may be: Secondary particles can be e s, neutral species

More information

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement

More information

Chapter 1 The discovery of the electron 1.1 Thermionic emission of electrons

Chapter 1 The discovery of the electron 1.1 Thermionic emission of electrons Chapter 1 The discovery of the electron 1.1 Thermionic emission of electrons Learning objectives: What are cathode rays and how were they discovered? Why does the gas in a discharge tube emit light of

More information

The distortion observed in the bottom channel of Figure 1 can be predicted from the full transport equation, C t + u C. y D C. z, (1) x D C.

The distortion observed in the bottom channel of Figure 1 can be predicted from the full transport equation, C t + u C. y D C. z, (1) x D C. 1 8. Shear Dispersion. The transport models and concentration field solutions developed in previous sections assume that currents are spatially uniform, i.e. u f(,y,). However, spatial gradients of velocity,

More information

Time of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields

Time of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields Time of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields Mingtong Han 5 th August, 013 Abstract Photoionization of acetylene, dimethylacetylene and methylacetylene in an

More information

Particle detection 1

Particle detection 1 Particle detection 1 Recall Particle detectors Detectors usually specialize in: Tracking: measuring positions / trajectories / momenta of charged particles, e.g.: Silicon detectors Drift chambers Calorimetry:

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

Particle physics experiments

Particle physics experiments Particle physics experiments Particle physics experiments: collide particles to produce new particles reveal their internal structure and laws of their interactions by observing regularities, measuring

More information

Temperature-Programmed Desorption

Temperature-Programmed Desorption Temperature-Programmed Desorption Alternative terms: TPD - Temperature Programmed Desorption TDS - Thermal Desorption Spectroscopy TPRS - Temperature Programmed Reaction Spectroscopy TPR/D - Temperature

More information

1.1 Atomic Structure Details of the three Sub-atomic (fundamental) Particles

1.1 Atomic Structure Details of the three Sub-atomic (fundamental) Particles 1.1 Atomic Structure Details of the three Sub-atomic (fundamental) Particles Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/1840-1 An atom of

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

Interface (backside) & Extraction Lens

Interface (backside) & Extraction Lens Plasma Interface Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control

More information

Atomic Spectra. d sin θ = mλ (1)

Atomic Spectra. d sin θ = mλ (1) Atomic Spectra Objectives: To measure the wavelengths of visible light emitted by atomic hydrogen and verify that the measured wavelengths obey the empirical Rydberg formula. To observe emission spectra

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

LECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of

LECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of LECTURE-13 Peptide Mass Fingerprinting HANDOUT PREAMBLE Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of proteins, drugs and many biological moieties to elucidate

More information

MR. HOLL S PHYSICS FACTS MECHANICS. 1) Velocity is a vector quantity that has both magnitude and direction.

MR. HOLL S PHYSICS FACTS MECHANICS. 1) Velocity is a vector quantity that has both magnitude and direction. MR. HOLL S PHYSICS FACTS MECHANICS 1) Velocity is a vector quantity that has both magnitude and direction. 2) Speed is a scalar quantity that has ONLY magnitude. 3) Distance is a scalar and represents

More information