APMP supplementary comparison of absorbed dose rate in tissue for beta radiation

Size: px
Start display at page:

Download "APMP supplementary comparison of absorbed dose rate in tissue for beta radiation"

Transcription

1 APMP supplementary comparison of absorbed dose rate in tissue for beta radiation BIPM KCDB: APMP.R(I)-S2 Technical Protocol N. Saito and M. Kato National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki JAPAN 19 October

2 1. Introduction The objective of this supplementary comparison is to establish the degrees of equivalence for the absorbed dose rate at 0.07 mm tissue depth for beta radiation [1] of the participating NMIs and to support the mutual recognition of calibration certificates for these qualities. The beta radiation fields of the radionuclides Sr-90/Y-90, Kr-85 and Pm-147 will be used for the comparison. Two transfer chambers will be calibrated by each of the participating laboratories for some radiation fields selected previously. The NMIJ/AIST is the pilot laboratory. 2. Participants The participants are listed in Table 1. Table 1. Participants in the comparison. Participant Institute Country Contact person ( ) 1 INER Taiwan Chien-Hau Chu (chchu1@iner.gov.tw) 2 KRISS Korea Chul-Young Yi (cyyi@kriss.re.kr) 3 Malaysian Nuclear Agency Malaysia Taiman Bin Kadni (Taiman@nuclearmalaysia.gov.my) 4 NMIJ Japan Masahiro Kato (masahiro-katou@aist.go.jp) Norio Saito (norio.saito@aist.go.jp) 5 OAP Thailand Vithit Pungkun (vithit@oaep.go.th) 3. Comparison procedure 3.1 Transfer chambers Two parallel-plate chambers of different type are to be used as transfer standards for the comparison, as listed in Table 2. The signal connections of the chambers are a tri-axial BNT plug for the Magna chamber and a BNC with a 4-mm banana plug for the CE-II chamber. The reference point of these chambers is the front surface of the detector case (NOT the surface of the thin film entrance window). A collecting voltage listed in Table 2, supplied at each laboratory, should be applied. Both polarities should be used for each chamber and the mean current evaluated. This voltage should be applied at least 30 minutes before starting measurements. A pre-irradiation of at least 30 minutes should also be made before any measurements. The leakage current should be measured before and after each measurement. Photographs of each chamber are presented in appendix B. Plastic cover of each chamber for protecting its entrance window shall be removed during the measurements. The Magna chamber is supported by its stem, 100 mm in length, 19 mm wide and 10 mm deep (in the beam direction). The CE-II is likewise supported by its cylindrical stem, 300 mm long and 12 mm in diameter. All stated dimensions are indicated in the photographs. The chambers are transported in an air tight box of around 500 mm height, 635 mm width, 305 mm depth, and weighing about 8 kg. It is important to note that the suggested connection for the Magna chamber requires the electrometer to operate in floating mode, the collecting electrode being at the polarizing potential. If the electrometer of the participant does not permit this, then the polarizing voltage must be applied to the outer electrode of the tri-axial connector, with the middle electrode connected to ground. In this case, care must be taken to avoid accidental contact with the connector. 2

3 Table 2. Technical data for the transfer chambers. Supplier EXRADIN OYOGIKEN Model Magna CE-II Serial number D Outer diameter (mm) Outer depth (mm) Diameter of the 79 collecting electrode (mm) Depth of the collecting volume (mm) Window foil material Window thickness Conductive Kapton film Al coated polyethyleneterephtalate inch, 25 m, 3.86 mg/cm mg/cm 2 Volume (cm 3 ) Cable length (m) 1.2 Stem length (mm) Stem size or diameter 19 x in diamter (mm) Cable connection Tri-axial (BNT) Signal: BNC HV: banana plug Outer electrode Ground Ground Middle electrode Inner electrode 250 V (both polarities) Current measurement (at polarizing potential) NG Current measurement (Ground) banana plug NG 400 V (both polarities) 3.2. Reference conditions The reference conditions for the chamber calibrations are as follows: 1. Field size at the reference point: larger than 15 cm in diameter. 2. The variation of the dose rate over the field size at the calibration distance should be less than ± 5 % for 90 Sr + 90 Yand 85 Kr, and less than ± 10 % for 147 Pm. To achieve this, the sources should be used with suitable beam-flattening filters. Examples of the constructions of the beam-flattening filters are shown in ISO :2006 [3]. If the variation is larger than ± 5 % for 90 Sr + 90 Yand 85 Kr, and ± 10 % for 147 Pm, corrections should be applied during the calibration of instruments and these factors given in the comparison report. 3

4 3. Air temperature, pressure and relative humidity of T = K, P = kpa and h = 65 %. 4. The calibration coefficients for the transfer chambers should be given in terms of the absorbed dose in tissue per charge, in units of Gy C The values of the half-lives to be used are: ( ) days for Sr-90/Y-90 (3915 3) days for Kr-85 ( ) days for Pm Calibration and Source-to-filter distances are listed in Table 3. Table 3 Calibration distances and source to beam-flattening filter distances. Radionuclide Calibration distance / (cm) Source-to-filter distance / (cm) Sr-90/Y Kr Pm Measurements This comparison is performed for the calibration of an ionization chamber in terms of the absorbed dose rate in tissue, t (0,07; α ). The current measured by the chamber at laboratory ambient conditions should be corrected to the current at reference conditions (pressure: kpa, temperature: K, rel. humidity: 65 %) in the collecting volume. D Each participant should calibrate the transfer chamber in several beta reference fields of the radionuclides Sr-90/Y-90, Kr-85 and Pm-147 for the incidence angle of 0. As a minimum, all participants must take part with Sr-90/Y-90 (Table 4). Table 4 Measurement qualities. All participants must take part with the quality labelled A. The label B denotes optional qualities for the comparison. Radionuclide Incidence angle Label of the quality Sr-90/Y-90 0 A Kr-85 0 B Pm B The calibration coefficient of the transfer chamber N in mgy/(h A) is given as the following equation: N D I c t (0,07; I c α ) / 2, where t (0,07; α ) is the conventional true dose rate at a depth of 0.07 mm in tissue in mgy/h for the angle of incidence α, measured with a primary standard and corrected to the ambient air conditions during the calibration measurements, and I + c and I - c are the ionization current in amperes at the two polarities, corrected to reference conditions. We note that the symbol I tc in the MS-Excel sheet (see section 3.6) is used to denote the ionization current before correction to reference conditions. D The comparison reference value C E (N n ) will be determined at each beta energy E as the weighted mean of the calibration coefficients N n of the participants. The weighting factor for each calibration factor N n will be its reciprocal of the variance, i.e. (1/u Nn ) 2. These reference 4

5 values C E (N n ) will be used to determine the degree of equivalence for each participating laboratory Structure of comparison There will be a star-shaped circulation of the transfer chambers between the NMIJ and the participants. The NMIJ will pay for the transport from NMIJ to the airport of the participant s country and, for the return, from Narita airport in Japan to the NMIJ. Each participant will pay for the transport from the airport of the participant s country to their institute and, for the return, from their institute to Narita airport. The participants should pay for the import custom duties. The chambers should stay at the participants site for no longer than 1 week. After each participant s calibrations, the NMIJ will perform chamber constancy checks by recalibrating both transfer chambers at all of the qualities. These data will form the basis of the uncertainty estimate for chamber stability entering in the data analysis Comparison schedule The comparison is scheduled to commence in July 2011 and the proposed schedule is shown in Table 5. Table 5. Proposed schedule for the comparison. Date of chambers Participant leaving NMIJ for participant Malaysian Nuclear Agency KRISS INER OAP 7-Jan Jun Jul Oct-2013 Measurement period at the laboratory 21- Jan-2013 to 25- Jan Jun-2013 to 21-Jun Jul-2013 to 2-Aug Nov-2013 to 15-Nov-2013 Date of chambers leaving participant for NMIJ 28- Jan Jun Aug Nov-2013 Notes: 1. The time allowed for measurements for each participant is one week. 2. Allowance is made for a transportation time for the chambers of about two weeks. 3.6 Submission of calibration results It is expected that all participating laboratories will submit their calibration results within 6 weeks of calibration. An MS-Excel sheet will be provided by the pilot laboratory in which information on the participants radiation qualities, primary standards and calibration results can be submitted. 3.7 Evaluation of measurement uncertainty All participating laboratories are required to evaluate the uncertainty of their calibration coefficients as Type A and Type B according to the criteria given in the Guide to The Expression of Uncertainty in Measurement issued by the International Organization for Standardization (ISO) in The Type A uncertainty is obtained by the statistical analysis of a series of observations; the Type B uncertainty is obtained by means other than the statistical analysis of series of observations. The participating laboratories submit to the 5

6 pilot laboratory their detailed uncertainty budgets (with relative standard uncertainties, k = 1). An MS-Excel sheet will be provided by the pilot laboratory in which the participants can detail the uncertainty. The sheet should be submitted together with the calibration results. 3.8 Comparison report The pilot laboratory will prepare a draft report for circulation to all participants for comments and discussion of the results. A revised final report will be submitted to the APMP/TCRI Chairman and to the CCRI(I). After the agreement of the APMP and the CCRI(I), the report will be published as the Technical Supplement in Metrologia. In addition, the comparison report will be sent to the BIPM for inclusion in the key comparison database (KCDB). 4. References [1] International Organization for Standardization. ISO , 2004 Nuclear energy -- Reference beta-particle radiation -- Part 2: Calibration fundamentals related to basic quantities characterizing the radiation field [ [2] Schötzig, U. and Schrader, H. Halbwertszeiten und Photonen-Emissionswahrscheinlichkeiten von häufig verwendeten Radionukliden PTB-report Ra-16/6 (2000) [3] International Organization for Standardization. ISO , 2006 Nuclear energy -- Reference beta-particle radiation -- Part 1: Methods of production 4 Guide to the Expression of Uncertainty in Measurement, International Organization of Standards, Switzerland (1995) 6

7 Appendix A Addresses of the participants Pilot laboratory NMIJ National Metrology Institute of Japan NMIJ National Institute of Advanced Industrial Science and Technology, AIST Quantum Radiation Division Umezono, Tsukuba, Ibaraki Japan contact person: Masahiro Kato Tel.: Fax: masahiro-katou@aist.go.jp Norio Saito Tel.: Fax: norio.saito@aist.go.jp Participants INER Institute of Nuclear Energy Research Health Physics Division No.1000, Wunhua Rd., Jiaan Village, Longtan Township,Taoyuan County, 32546, Taiwan(R.O.C.) Contact person : Dr. Chien-Hau Chu Tel: EXT7673 Fax: chchu1@iner.gov.tw KRISS Center for Ionizing Radiation Korea Research Institute of Standards and Science (KRISS) 267 Gajeong-ro, Yuseong-gu, Daejeon Korea Contact person : Dr. Chul-Young Yi Tel: or Fax: cyyi@kriss.re.kr Nuclear Malaysia Malaysian Nuclear Agency (Nuclear Malaysia) Bangi, Kajang, Selangor Darul Ehsan, Malaysia Contact person: Mr. Taiman Bin Kadni Tel.: ext 1260 or (direct) Fax: taiman@nuclearmalaysia.gov.my 7

8 OAP Office of Atoms for Peace 16 Vibhavadi Rangsit Road, Laadyaw, Chatuchak, Bangkok Thailand Contact person: Mr.Thongchai Soodprasert Tel: ext Fax: thongcha@oaep.go.th 8

9 Appendix B Photographs of the chambers EXRADIN Magna 19 mm 100 mm 10 mm Oyogiken CE-II 300 mm 12 mm Center of the chamber 9

Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-192 brachytherapy sources

Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-192 brachytherapy sources Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-19 brachytherapy sources T. Kurosawa and N. Saito National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki 305-8568 JAPAN Revised

More information

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) D V Webb a, J-H Lee b, C T Budiantari c, J Laban d, N Saito e, S Srimanoroth f, N E Khaled g a Australian Radiation Protection

More information

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv EUROMET 545: Final report 25/9/28 page 1 of 48 Comparison of national air kerma standards for ISO 437 narrow spectrum series in the range 3 kv to 3 kv L. Büermann (a), M. O Brien (b), D. Butler (c), I.

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

Comparison of V and 10 V DC Voltage References

Comparison of V and 10 V DC Voltage References Comparison of 1.018 V and 10 V DC Voltage References Technical protocol for BIPM.EM-K11.a and.b comparisons 1. Introduction The Mutual Recognition Arrangement among National Metrology Institutes (NMIs)

More information

International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation

International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation EUROMET project No. 739 BIPM KCDB: EUROMET.RI(I)-S2 Final Report R. Behrens, K. Helmstädter,

More information

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Norio Saito, Tadahiro Kurosawa, Masahiro Kato, Yuichiro Morishita, Takahiro Tanaka, and Morihito

More information

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard

More information

Ion-Chamber Survey Meter OD-02

Ion-Chamber Survey Meter OD-02 Ion-Chamber Survey Meter OD-02 Dose and dose rate meter for measuring the ambient dose equivalent H*(10) and dose rate equivalent dh*(10)/dt as well as the directional dose equivalent H (0,07) and dose

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

Report on APMP Supplementary Comparison High precision roundness measurement APMP.L-S4. Final report

Report on APMP Supplementary Comparison High precision roundness measurement APMP.L-S4. Final report National Institute of Metrology (Thailand) Report on APMP Supplementary Comparison APMP.L-S4 J. Buajarern, National Institute of Metrology (Thailand) (NIMT), Thailand K. Naoi, National Metrology Institute

More information

Instructions for Use for the transfer chamber and the electronic measuring device Revision 1

Instructions for Use for the transfer chamber and the electronic measuring device Revision 1 EURAMET supplementary comparison of the personal dose equivalent at 0.07 mm and 3 mm depth, H p (0.07) and H p (3), for beta radiation EURAMET Project No. 1398 BIPM KCDB: EURAMET.R(I)-S16 Instructions

More information

Survey Meter OD-01 Address: Phone: Fax: URL:

Survey Meter OD-01 Address: Phone: Fax:   URL: Survey Meter OD-01 Dose meter and dose rate meter for the measurement of the ambient dose and dose rate equivalent H*(10), dh*(10)/dt and the directional dose and dose rate equivalent H'(0.07), dh'(0.07)/dt

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688 Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio

More information

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Ho-166m activity measurements to include the IRA and the NPL and a re-evaluation of the degrees of equivalence for the APMP.RI(II)-K2.Ho-166m comparison C.

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 6980-2 Première edition 2004-10-15 Nuclear energy Reference beta-particle radiation Part 2: Calibration fundamentals related to basic quantities characterizing the radiation

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force < Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force Kazunaga Ueda*, Toshiyuki Hayashi*, Hiroshi Maejima*, Rolf Kumme**, Dirk Röske**, Mark Seidel** * National Metrology of

More information

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1

More information

Abstract An indirect comparison of the standards for reference air kerma rate for

Abstract An indirect comparison of the standards for reference air kerma rate for Comparison BIPM.RI(I)-K8 of high dose-rate Ir-19 brachyerapy standards for reference air kerma rate of e VSL and e BIPM J.T. Alvarez 1, J. A. de Pooter, C. Andersen 3, A.H.L. Aalbers, P.J. Allisy-Roberts

More information

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter Christopher A Perks *a and Stephane Faugoin a a LANDAUER EUROPE, 33 Avenue du General Leclerc, F-92266, Fontenay-aux-Roses

More information

FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3

FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3 FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3 29 November, 2011 Dr. Hyu-Sang Kwon, Dr. Sang-Joon Suh and Jae-Gap Suh Korea Research Institute of Standards and Science (KRISS) Metrologia Tech. Suppl. 49

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays D T Burns 1, C Kessler 1, E Mainegra-Hing 2, H Shen 2 and M R McEwen 2 1 Bureau International

More information

Final Report on Key Comparison APMP.M.P-K13 in Hydraulic Gauge Pressure from 50 MPa to 500 MPa

Final Report on Key Comparison APMP.M.P-K13 in Hydraulic Gauge Pressure from 50 MPa to 500 MPa ASIA-PACIFIC METROLOGY PROGRAMME 5 MPa HYDRAULIC PRESSURE INTERLABORATORY COMPARISON Comparison Identifier: APMP.M.P-K13 Final Report on Key Comparison APMP.M.P-K13 in Hydraulic Gauge Pressure from 5 MPa

More information

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Leon de Prez and Frans Bader VSL - The Netherlands www.vsl.nl P 1 Contents - Introduction to VSL and Ionizing

More information

A bilateral comparison of a 1 kg platinum standard

A bilateral comparison of a 1 kg platinum standard DANSK FUNDAMENTAL METROLOGI DANISH FUNDAMENTAL METROLOGY Report on A bilateral comparison of a 1 kg platinum standard by Lars Nielsen Danish Fundamental Metrology B 307, Matematiktorvet, DK-800 Kgs. Lyngby

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Comparison protocol EURAMET project

Comparison protocol EURAMET project Comparison protocol EURAMET project Bilateral three-phase AC and energy comparison at frequency 1. Introduction This bilateral comparison was triggered by MIKES need to get support for their planned extension

More information

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays C Kessler 1, D T Burns 1, P Roger 1, V Sochor 2 1 Bureau International des Poids et

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the

More information

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, F. Delaunay 2, M. Donois 2 1 Bureau International des Poids

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 6980-3 First edition 2006-10-01 Nuclear energy Reference beta-particle radiation Part 3: Calibration of area and personal dosemeters and the determination of their response as

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Rapport BIPM 02/01 Bureau International des Poids et Mesures Comparison of the standards of air kerma of the Yugoslavia and the BIPM for 60 Co γ rays by P.J. Allisy-Roberts, D.T. Burns, C Kessler BIPM

More information

Calibration of gauge blocks by interferometry. Final Report Results

Calibration of gauge blocks by interferometry. Final Report Results APMP Key comparison L-K1.1: Calibration of gauge blocks by interferometry Page 1 of 17 Asia-Pacific Metrology Programme APMP Key Comparison APMP.L-K1.1 Calibration of gauge blocks by interferometry Final

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range to 10 k June 016 Author: Co-author: Danuta Dobrowolska (GUM, pilot) Alexander Kosterov

More information

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation ey comparison BIPM.RI(I)-1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation C. essler 1, D. Burns 1,. Wang 2,Y. Fan 2,S. Jin 2, X. Yang 2 1 Bureau International des Poids

More information

Report on EURAMET.M.M-S9

Report on EURAMET.M.M-S9 FINAL REPORT Report on EURAMET.M.M-S9 Supplementary comparison of 500 microgram, 200 microgram, 100 microgram and 50 microgram weights (EURAMET project number 1310) Stuart DAVIDSON 1, Adriana VALCU 2,

More information

Appendix B1. Reports of SMU

Appendix B1. Reports of SMU Appendix B1 Reports of SMU Euromet Project 600 Comparison of Surface Roughness Standards /13 EUROMET SUPPLEMENTARY COMPARISON SURFACE TEXTURE Project No. 600 Final Report Elaborated by: M. Szmicskova Bratislava,

More information

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3. AFRIMETS Secretariat Private Bag X34 Lynnwood Ridge 0040 AFRIMETS Supplementary Comparison Programme Calibration of Gauge Blocks by Mechanical Comparison Method AFRIMETS.L S3 Final Report Giza, Egypt,

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 G. Ratel*, C. Michotte*, M. Reinhard, D. Alexiev, L. Mo *BIPM, ANSTO, Australia

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

Final Report. APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf

Final Report. APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf Final Report APMP.EM-K4.1 APMP Key Comparison of Capacitance at 10 pf H. L. Johnson National Measurement Institute, Australia February 09 1 Contents 1 Introduction...5 Participants and organisation of

More information

GULF ASSOCIATION FOR METROLOGY. Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1

GULF ASSOCIATION FOR METROLOGY. Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1 GULF ASSOCIATION FOR METROLOGY Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1 Instructions and Technical Protocols GULFMET.L-S1: Calibration of Gauge Blocks by Mechanical Comparison

More information

GMA STATISTICS Countries. Achievements of metrology HRD service for Foreign customers in 2018

GMA STATISTICS Countries. Achievements of metrology HRD service for Foreign customers in 2018 GMA STATISTICS 018 Achievements of metrology HRD service for Foreign customers in 018 70 people from 1 countries have enjoyed opportunities of studying at KRISS. Group courses Individual courses Graduate

More information

Final Report on the APMP Air Speed Key Comparison (APMP.M.FF-K3)

Final Report on the APMP Air Speed Key Comparison (APMP.M.FF-K3) Final Report on the APMP Air Speed Key Comparison (APMP.M.FF-K3) July, 00 (Revised) Yoshiya Terao, Yong Moon Choi, Mikhail Gutkin 3, Wu Jian 4, Iosif Shinder 5 and Cheng-Tsair Yang 6 NMIJ, AIST, Japan

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

Measurement of Tritium in Helium

Measurement of Tritium in Helium detect and identify Measurement of Tritium in Helium Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 22 nd Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas,

More information

APMP Key Comparison of DC Voltage at V and 10 V

APMP Key Comparison of DC Voltage at V and 10 V Final Report APMP.EM.BIPM-K11.3 APMP Key Comparison of DC Voltage at 1.018 V and 10 V Kyu-Tae Kim* (Korea Research Institute of Standards and Science, KRISS) Michitaka Maruyama (National Metrology Institute

More information

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from Extensions to the Beta Secondary Standard BSS 2 Consolidated version from www.ptb.de R. Behrens 1 and G. Buchholz 2 Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany

More information

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan Update of the BIPM comparison BIPM.RI(II)-K1.Ce-139 of activity measurements of the radionuclide 139 Ce to include the 2004 NMIJ result and links for the 2004 regional comparison APMP.RI(II)-K2.Ce-139

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III)

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III) Air Density Determination Using 1 kg Buoyancy Mass Comparison(III) Jin-Wan CHUNG, Kyung-Ho CHANG, Woo-Gab LEE, Kwang-Pyo KIM Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES 1 BUREAU INTERNATIONAL DES POIDS ET MESURES International comparison of activity measurements of a solution of 3 H (January 2009) Participating laboratory: T ½ = (4 496.862 d; u = 9.131 d)* Ampoule number

More information

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017 Metrology Principles for Earth Observation: the NMI view Emma Woolliams 17 th October 2017 Interoperability Decadal Stability Radiometric Accuracy Identical worldwide Century-long stability Absolute accuracy

More information

ASIA PACIFIC METROLOGY PROGRAME

ASIA PACIFIC METROLOGY PROGRAME APMP.M.M-K6 (SUB) multiples mass key comparison ASIA PACIFIC METROLOGY PROGRAME Final Report of APMP Comparison of Mass standards APMP.M.M-K6 (SUB) multiples mass key comparison Coordinated by: Vietnam

More information

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order

More information

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1 Comparison of the air-kerma x-ray standards of the NRPA, the STUK, the SSM and the LNE-LNHB in the ISO 4037 narrow spectrum series in the range 40 kv to 300 kv H Bjerke 1, J Plagnard 2, J-M Bordy 2, A

More information

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT BY NATIONAL METROLOGY CENTRE (A DIVISION OF SPRING SINGAPORE) COORDINATOR OF THE COMPARISON September

More information

ISO Radiological protection Sealed radioactive sources General requirements and classification

ISO Radiological protection Sealed radioactive sources General requirements and classification INTERNATIONAL STANDARD ISO 2919 Third edition 2012-02-15 Radiological protection Sealed radioactive sources General requirements and classification Radioprotection Sources radioactives scellées Exigences

More information

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:

More information

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon Tohru Nakano and Osamu Tamura National Metrology Institute of Japan (NMIJ), AIST Tsukuba, Japan

More information

Final Report on COOMET Key Comparison of Capacitance at 10 pf (COOMET.EM-K4)

Final Report on COOMET Key Comparison of Capacitance at 10 pf (COOMET.EM-K4) State Enterprise All-Ukrainian State Scientific and Production Center of Standardization, Metrology, Certification and Protection of Consumer (SE Ukrmetrteststandard ) Approved by the chairman of TC 1.3

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August Pilot: KRISS, Republic of Korea Yon-Kyu Park

Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August Pilot: KRISS, Republic of Korea Yon-Kyu Park Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August 2018 Pilot: KRISS, Republic of Korea Yon-Kyu Park Muktar Sawi (NMIM) Seif M. Osman (NIS) S.S.K.Titus (NPLI)

More information

R SANAS Page 1 of 7

R SANAS Page 1 of 7 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:

More information

Introduction of RIC Tsukuba (Japan, RAII)

Introduction of RIC Tsukuba (Japan, RAII) Introduction of RIC Tsukuba (Japan, RAII) 20 March, 2018 Kouichi NAKASHIMA Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. Overview of Organization

More information

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Recent Activities on Neutron Calibration Fields at FRS of JAERI Recent Activities on Neutron Calibration Fields at FRS of JAERI Michio Yoshizawa, Yoshihiko Tanimura, Jun Saegusa and Makoto Yoshida Department of Health Physics, Japan Atomic Energy Research Institute

More information

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom Force Key Comparison EUROMET.M.F-K2 (50 kn and 100 kn) EURAMET Project No 518 Final Report 14 July 2014 Pilot: NPL, United Kingdom Co-authors: Renato Reis Machado (INMETRO, Brazil), Petr Kašpar (CMI, Czech

More information

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures Work Programme of the International Bureau of Weights and Measures for the four years 2016-2019 Comité international des poids et mesures Section II : BIPM Work Programme for 2016-2019 35 Priority activities

More information

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency Facility of Radiation Standards It is important that radiation measuring instruments are calibrated by

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Laboratory locations: Schedule of Accreditation 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 1G Kilroot Business Park Carrickfergus Co Antrim BT38 7PR Contact: Mr R E Hughes Tel:

More information

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2)

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2) 1 Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K) 1. Introduction L.O. Becerra 1, W. Bich, N. Bignell 3, G.D. Chapman 4, J.W. Chung 5 S. Davidson 6, M. Gläser

More information

Chemical Standard Department, 1600, Shimo-Takano, Sugito-machi, Kitakatsushikagun, Saitama , JAPAN

Chemical Standard Department, 1600, Shimo-Takano, Sugito-machi, Kitakatsushikagun, Saitama , JAPAN International Key Comparison APMP.QM-K4.1 Final report 24/10/2007 Kenji Kato 1, Masaaki Maruyama 2, Jin Seog Kim 3, Oh, Sang Hyub 3, Jin-Chun Woo 3, Yongdoo Kim 3, Hyunkil Bae 3, Qiao Han 4, Zeyi Zhou

More information

TEST-LOCATION SPECIFICATION BY MEANS OF HARDNESS MAPPING ON VICKERS BLOCK SURFACE

TEST-LOCATION SPECIFICATION BY MEANS OF HARDNESS MAPPING ON VICKERS BLOCK SURFACE HARDMEKO 2007 Recent Advancement of Theory and Practice in Hardness Measurement 19-21 November, 2007, Tsukuba, Japan TEST-LOCATION SPECIFICATION BY MEANS OF HARDNESS MAPPING ON VICKERS BLOCK SURFACE Rugkanawan

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

CCM short note on the dissemination process after the proposed redefinition of the kilogram

CCM short note on the dissemination process after the proposed redefinition of the kilogram CCM short note on the dissemination process after the proposed redefinition of the kilogram Consultative Committee for Mass and Related Quantities 1. Introduction This note proposes how the mise en pratique

More information

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017 Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

CCL-K5. CMM 1D: Step Gauge and Ball Bars. Final Report

CCL-K5. CMM 1D: Step Gauge and Ball Bars. Final Report Physikalisch-Technische Bundesanstalt CIPM Key Comparison CCL-K5 CMM 1D: Step Gauge and Ball Bars Final Report V3.2 Otto Jusko, PTB, Braunschweig, September 2009 CCL-K5_Final_Report_V3.2(corr).doc Page

More information

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN Problem P7 Stéphanie Ménard Dosimetry Department 92262 Fontenay-aux aux-roses FRANCE What are the applications of Gamma-Ray Spectrometry in Radiological Protection and in Safety? In the environment: after

More information

APMP.QM-S3 Cd in Rice Powder

APMP.QM-S3 Cd in Rice Powder APMP.QM-S3 Cd in Rice Powder (Coordinated by KRISS, NIM and NMIJ) Final Draft Report Sep. 16, 2011 Yong-Hyeon YIM, Euijin HWANG, Jun WANG, Yanbei ZHU Summary The present study, APMP.QM-S3, was performed

More information

Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards.

Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards. Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards. Pilot laboratory: Node laboratory: (EURAMET, COOMET) Node laboratory: (APMP) Node laboratory: (EURAMET) Node

More information

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf

More information

CCT Working Group for Humidity

CCT Working Group for Humidity CCT/17-41 CCT Working Group for Humidity Report to the CCT 2017 Stephanie Bell CCT Working Group for Humidity Terms of reference: to advise the CCT on matters relating to humidity; to pursue harmonization

More information

Study on a thermal diffusivity standard for the laser flash method measurements 1

Study on a thermal diffusivity standard for the laser flash method measurements 1 Study on a thermal diffusivity standard for the laser flash method measurements 1 M. Akoshima 2,3 and T. Baba 2 1 Paper presented at the Seventeenth European conference on Thermophysical Properties, September

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information