APMP.QM-S3 Cd in Rice Powder

Size: px
Start display at page:

Download "APMP.QM-S3 Cd in Rice Powder"

Transcription

1 APMP.QM-S3 Cd in Rice Powder (Coordinated by KRISS, NIM and NMIJ) Final Draft Report Sep. 16, 2011 Yong-Hyeon YIM, Euijin HWANG, Jun WANG, Yanbei ZHU

2 Summary The present study, APMP.QM-S3, was performed to demonstrate and document the capability of interested national metrology institutes in the APMP region to determine the mass fraction of Cd in rice. It was carried out as a regional supplementary comparison related with a previous CCQM key comparison, CCQM- K24, and run in parallel with a pilot study, APMP.QM-P12. The comparison is an activity of the APMP-TCQM and was coordinated by KRISS (Korea Research Institute of Standards and Science, Korea) with NIM (National Institute of Metrology, P. R. China) and NMIJ (National Metrology Institute of Japan) as co-coordination laboratories to provide reference value. Seven national metrology institutes (NMIs) participated in the supplementary comparison. The degrees of equivalence for NMIs employing double ID-ICP/MS techniques are excellent while the results obtained using other techniques may need more improvement. 1

3 1. Introduction Rice is the major source of food for a large proportion of the world s population, especially in Asia. Therefore, many countries set regulations limiting the amount of toxic elements, such as cadmium, in rice to protect public health. With ever increasing international trade of food, traceable measurement of hazardous elements in food supported by NMIs has become one of the essential requirements for food safety. The intention of this study is to establish the comparability in the measurement of cadmium in rice. It is expected that participating laboratories, which can demonstrate their capability to measure cadmium in such a matrix, will be able to apply such expertise to a range of other types of sample. The previous regional APMP comparison on cadmium in rice, APMP.QM-P2, was carried out in parallel with CCQM-K24 key comparison. However, many APMP participants failed to show satisfying comparability with the CCQM results. Subsequent APMP.QM-P5 study (Cd content in oyster tissue) also resulted in limited success with small number of participants. The present supplementary study, APMP.QM-S3 (Cd content in rice powder), has been originally proposed and approved by APMP-TCQM in 2007 as APMP.QM-K24 to provide NMIs in the APMP region a chance to demonstrate the degree of equivalence in their measurement and possibly link it to the CCQM-K24 result. The study was run in parallel with the pilot study, APMP.QM-P12. It has been carried out from July to December 2008 and the results were presented at the APMP meeting in Kuala Lumpur, Malaysia, in December Draft A report for the APMP.QM-K24 was circulated to participants in Feb But it was recognized that the RMO comparison should have been discussed and approved by the Inorganic Analysis Working Group (IAWG) and registered in the Key Comparison Data Base (KCDB) in advance. Therefore, the coordinating laboratories asked a retrospective approval of the study and discussions on the report in the IAWG during the CCQM meeting in Paris in April After submission and discussion of draft A report of APMP.QM-K24 in the CCQM meeting, it was decided to register the study as a supplementary comparison of APMP considering the lack of sound scientific background to link the two studies directly. A new comparison number, APMP.QM-S3 was assigned to the previous APMP.QM- K24 in September The decision was reported and the follow-up actions were discussed further in the CCQM-IAWG meeting at Hindas, Sweden on 29 th September

4 2. List of Participants Participants in APMP.QM-S3 are listed in Table 1. Table 1. APMP.QM-S3 participants. Economy Institute/Organization Contributors China National Institute of Metrology (NIM) Dr. Jun WANG Dr. Chao WEI Dr. Qian WANG Hong Kong Government Laboratory Hong Kong (GLHK) Dr. Yiu-chung YIP India National Physics Laboratory-India (NPLI) Mr. Prabhat K. GUPTA Indonesia Indonesian Institute of Science (LIPI) Dr. SUMARDI Japan National Metrology Institute of Japan (NMIJ) Dr. Yanbei ZHU Korea Thailand Korea Research Institute of Standards & Science (KRISS) National Metrology Institute, Thailand (NMIT) Dr. Yong-Hyeon YIM Dr. Kyung-Seok LEE Mr. Charun YAFA Ms. Pranee PHUKPHATTHANACHAI 3. Samples Each participant was sent more than one bottles of rice powder sample containing approximately 30 g of material in each bottle. The sample originated from rice grown in contaminated area. The material was tested at KRISS for stability and homogeneity and was found to be appropriate for the needs of the comparison. The homogeneity test was carried out using isotope dilution ICP/MS (ID-ICP/MS) after microwave digestion of subsamples. Ten bottles were sampled randomly from 1190 bottles with minimum sampling size of 0.5 g. The relative standard deviation of Cd content 3

5 between bottles was less than 0.41%. Sample bottle containing about 30 g of material was sent to each participant presented in the Table 1 with information and instruction documents on July 8, The sample for NIM, China, once returned due to customs clearance problem and finally delivered on October 20, The deadline for reporting of results was Dec. 31, Introductions to Participants The following information was provided to all participants General instructions. Result report form Detailed sample information, guideline for sample preparation and reporting are described in the general instructions. The information provided to participants is given in Appendix A and B. 5. Methods Participant The participants were free to choose one or more suitable methods of Cd measurement. Methods and instruments used are given in Table 2. Table 2. Analytical methods and instrumental techniques used by APMP.QM-S3 participants for cadmium measurements. Subsample mass Number of subsamples Digestion Calibration Instrument NIM 0.5 g 12 HNO 3 + H 2 O 2 MW digestion Double IDMS ICP/MS GLHK 0.5 g 5 HNO 3 +HF MW digestion Double IDMS ICP/MS NPLI 0.5 g 6 HNO 3 + H 2 O 2 + HF MW digestion External calibration GF-AAS LIPI 0.5 g 4 HNO 3 + H 2 O 2 MW digestion Standard addition GF-AAS NMIJ 0.5 g 4 HNO 3 + H 2 O 2 + HF MW digestion Double IDMS ICP/MS KRISS 0.5 g 5 5 ml HNO ml H 2 O 2 MW digestion Double IDMS ICP/MS NMIT 0.5 g 4 5 ml HNO ml H 2 O 2 MW digestion Double IDMS ICP/MS 4

6 For sample digestion, all laboratories used microwave with varying mixture of HNO 3, H 2 O 2 and HF. Two different instrumental techniques were used for the cadmium measurement in samples. Four laboratories used double isotope dilution mass spectrometry (IDMS), while the other two participants used atomic absorption spectroscopy (AAS) with either external calibration or standard addition. 6. Participants Results The participants results are given in Table 3. Participant Table 3. APMP.QM-S3 participants measurement results for cadmium. Reported value (mg kg -1 ) Reported standard uncertainty (mg/kg) Coverage factor k (95% confidence level) Expanded uncertainty (mg kg -1 ) Relative expanded uncertainty (%) NIM GLHK NPLI LIPI NMIJ KRISS NMIT

7 7. Discussion In the original proposal, it was suggested to use the results of three NMIs (NIM, NMIJ, KRISS) that have also taken part in CCQM-K24 to calculate the reference value (RV) and its standard uncertainty (RU). Although the present study has been reclassified as an independent RMO supplementary comparison, it is proposed to use the same strategy for calculation of RV. It should be noted that only those three NMIs, among APMP.QM-S3 participants, had demonstrated their measurement capability of Cd in similar matrix at the time of the present comparison by participating in CCQM key comparisons. As shown in Table 4, the arithmetic mean, the median and the mixture model median of the three results used for RV calculation are listed with associated uncertainties. Standard deviation of the three results was used as associated uncertainties of the mean and the median. Scaled median of absolute deviation, MAD E, was not used as a statistical estimator of the standard uncertainty for the median value due to its negative bias for small number of data set. The difference between mean, median and mixture model median values are negligible considering the associated uncertainties. Hence, the simplest estimator, the mean value and standard uncertainty of results from the three NMIs (which is ± mg/kg), is proposed as the RV and RU [1]. Based on the homogeneity test, uncertainty contribution from sample homogeneity is expected to be less than a quarter of the proposed uncertainty of the RV. Therefore, the uncertainty of the RV is set satisfactorily for evaluation of degree of equivalence without noticeable contribution from sample homogeneity. The reported values and standard uncertainties of the participants are graphically displayed in Figure 1 together with the RV and its uncertainty, RU. The double IDMS results from most of participants show decent agreement with the RV within the stated uncertainties. However, the two results obtained by calibration with external standard or standard addition have relatively larger bias and/or uncertainties than those from double IDMS. Table 4. Reference value calculation for APMP.QM-S3. Reference value,rv (mg/kg) Standard Uncertainty, RU (mg/kg) Expanded uncertainty (mg/kg) Relative expanded uncertainty (%) k (95% CL) Mean Median MM median* * Mixture model median [2]. 6

8 Figure 1. Participants results for APMP.QM-S3 displayed with error bars representing reported standard uncertainties. The solid horizontal line in red color is the proposed RV of mg/kg; the dashed blue lines show the standard uncertainty of the reference value. The degree of equivalence between a NMI result and the RV is calculated, as given in Table 5, according to the BIPM guidelines. The degree of equivalence, D i, between the NMI result, x i, and the RV, x R, is calculated according to the following equations: D i = x x, 2 ( 2 2 U = u + u ) i R i i R where U i is the expanded uncertainty (k = 2) of the D i calculated by combining the uncertainty of the NMI result, x i, and the standard uncertainty of the RV, x R. Equivalence statements for APMP.QM-S3 results are displayed in Figure 2. The degree of equivalence for NMIs employing double IDMS techniques was excellent, while the results from other techniques need more improvement. 7

9 Participant Table 5. Equivalence statements for APMP.QM-S3. Reported value (mg/kg) Standard uncertainty (mg/kg) D i (mg/kg) U i (mg/kg) KRISS NIM GLHK NMIJ NMIT LIPI NPLI Figure 2. Graph of equivalence statements for APMP.QM-S3. Error bars represent the expanded uncertainties interval calculated with a coverage factor k=2. 8

10 The participants for the supplementary comparison are asked to complete core capability forms to document core capabilities being demonstrated in the comparison. Summarized tables of the demonstrated core capabilities for APMP.QM-S3 are also shown in the Appendix C. 8. Conclusion The present APMP supplementary comparison, APMP.QM-S3, has been carried out in parallel with APMP.QM-P12 to demonstrate and document the capability of NMIs in the APMP region for determination of Cd content in rice and related food matrix materials. The degree of equivalence for results obtained with double ID- ICP/MS was satisfactory with less than 5% deviation from the RV, while those from other methods may require more improvements including rigorous control of measurement method and method validation. The participated laboratories directly demonstrated their capability to measure cadmium in rice and will be able to apply such expertise to a range of samples and analytes with similar matrices. 9. Acknowledgement The work described here contains the contributions of many scientists as detailed in Table 1. The coordinating laboratory gratefully acknowledges Luis Mussio from BIPM, Linsey Mackay from NMIA, Mike Sargent and all members of the IAWG for their comments and advices for RV calculation. Special thanks are due to Kenji Kato and Della Sin for their assistance as chairpersons of the APMP-TCQM. 10. References 1. CCQM Guidance note: Estimation of a consensus KCRV and associated Degrees of Equivalence, 1 March David L. Duewer, A robust approach for the determination of CCQM key comparison reference values and uncertainties, Working document CCQM, 04-15, BIPM,

11 Appendix A. Introduction for APMP.QM-K24 and P12 comparison (Cd content in rice powder) 1. Sample information The rice powder was prepared from rice grown in a contaminated area. Rice grains were rinsed four times to remove rice bran and other impurities, and freeze-dried. The freeze-dried material was broken down into smaller pieces, mixed, and sieved to obtain particles with diameter between 50 to 250 μm. It was further homogenized with V-blender and aliquoted into amber bottles of 60 ml in volume. The amber bottles containing rice powder were sterilized by gamma ray irradiation. Each bottle contains about 30 g of material. Cd content in rice powder: mg/kg Homogeneity: ID-ICP/MS analysis using 10 samples (about 0.5 g each) taken from 10 bottles were carried out. Relative standard deviation was 0.41 %, which is satisfactory for the present pilot study. Each participant will receive one bottle containing about 30 g of sample. The participants will be informed of the date of dispatching of samples. Participants are required to confirm the receipt of the samples (including the serial number of the sample, receipt data, any damages found in the sample package) by (yhyim@kriss.re.kr or ehwang@kriss.re.kr) or FAX( ). If any damage is observed, please contact us immediately to get additional bottle of sample. 2. Sample storage The sample should be kept sealed in its original bottle and stored at room temperature. The opening of the sample bottle should be carefully planned to avoid contamination and decomposition of the sample. 3. Guidelines for sample preparation and dry-mass correction Sample handling in a laboratory with well-controlled humidity and temperature is recommended. Any kind of contamination should be avoided. Sample preparation for analysis: Sampling for the analysis should be carried out at the same time with sampling for the dry-mass correction. Digest the rice powder material completely, if the method used requires digestion. Dry-mass correction: Sampling for the digestion should be carried out at the same time with sampling for the dry-mass correction. It is recommended to take aliquots of sample before and after the sampling for the digestion. Take at least 3 separate aliquots of sample with sample size of g for drymass correction. The material should be dried in a desiccator for 7 days (168 hours) using P 2 O 5 as desiccant. Calculate the moisture content from the mass change observed, and use it for dry-mass correction. 4. Measurement method Each laboratory has to choose the method of measurement for itself and more than one method can be used simultaneously. It is recommended to use appropriately validated methods. 5. Time schedule Dead line for registration: K24 (May 15, 2008), P12 (July 30, 2008) Sample distribution: K24 (July 9, 2008), P12 (before Aug., 2008) 10

12 Deadline for reporting: December 31, 2008 Draft A report: March 31, Reporting The report should be sent to the coordinating laboratory by before Dec. 31, For rigorous evaluation of the inter-laboratory comparison, detailed information should be included in the report (final report form will be distributed with the sample) Final result with relevant uncertainty should be reported as mass faction [mg/kg] based on dry mass. If the final result has been calculated from more than one method, the individual results from the contributing methods must also be reported. One final result should be reported in the key comparison. However, more than two results can be reported for the pilot study, when different methods are used. Detailed description of analytical methods used. If more than one method is used, each method should be described in detail. Information about sample digestion and preparation. Information about the reference material used for calibration or other materials used in the analytical procedure. Details of the uncertainty evaluation: uncertainty should be evaluated according to the ISO Guide to the Expression of Uncertainty in Measurement, 1993, which includes - Complete specification of the measurement equation including corrections. - Description of all uncertainty sources and their typical values 7. Contact Person. Drs. Yong-Hyeon Yim & Euijin Hwang Principal Research Scientists Health Metrology Group, KRISS 1 Doryong-dong, Yuseong-gu, Daejeon , Republic of Korea Tel: , 5361 Fax: yhyim@kriss.re.kr, ehwang@kriss.re.kr 11

13 Lab. Identification Name: Institute: Address: Country: Tel. number: Fax. Number: Date: Appendix B. Result report form APMP.QM-K24 Cd Content in Rice Powder Result Report Please report all your results and uncertainties in the unit mg/kg. Cd content (mg/kg) Combined standard uncertainty (mg/kg) Coverage factor k (in 95% level of confidence) Expanded uncertainty (mg/kg) Result from individual aliquout (dry-mass corrected) Aliquot number Cd content (mg/kg) Average Standard deviation Standard deviation of the mean 12

14 Questionnaire 1. Analytical method used: 2. Description of the measurement procedure including sample preparation: 3. Correction for dry mass (% of weighted): 4. Amount & number of sample aliquots taken for dry mass correction: 5. Uncertainty for dry mass correction: 6. Amount & number of sample aliquots taken for Cd analysis: 7. (Isotopic) Reference materials used for calibration purposes: 8. For IDMS, indicate reference & spiked isotopes used: 9. Details of the uncertainty evaluation - Complete specification of the measurement equation including corrections - Description of all uncertainty sources and their typical values 13

15 Appendix C. Tables of demonstrated core-capabilities Method Summary of APMP.QM-S3 (Cd in rice powder) NIM GLHK NPLI LIPI NMIJ KRISS NMIT ID-ICP/MS Cd Cd Cd Cd Cd GF-AAS Cd Cd 14

16 CCQM Study: APMP.QM-S3 Inorganic Core Capabilities Summary Table Institute(s): GLHK, KRISS, NIM, NMIJ, NIMT Method: ID-ICP-MS Analyte(s): Cd Capabilities/Challenges Degree of Difficulty Explanation Low Moderate High Contamination control and correction techniques and procedures employed to reduce potential contamination of samples as well as blank correction procedures. The level of difficulty is greatest for analytes that are environmentally ubiquitous and also present at very low concentrations in the sample. Digestion/dissolution of organic matrices techniques and procedures used to bring a sample that is primarily organic in nature into solution suitable for liquid sample introduction to the ICP. Digestion/dissolution of inorganic matrices techniques and procedures used to bring a sample that is primarily inorganic in nature into solution suitable for liquid sample introduction to the ICP. Volatile element containment techniques and procedures used to prevent the loss of potentially volatile analyte elements during sample treatment and storage. Pre-concentration Techniques and procedures used to increase the concentration of the analyte introduced to the ICP. Includes evaporation, ion-exchange, extraction, precipitation procedures, but not vapor generation procedures. Vapor generation Techniques such as hydride generation and cold vapor generation used to remove the analyte from the sample as a gas for introduction into the ICP. Matrix separation Techniques and procedures used to isolate the analyte(s) from the sample matrix to avoid or reduce interferences caused by the matrix. Includes ion-exchange, extraction, precipitation procedures, but not vapor generation procedures. Techniques and procedures used to isolate the analyte(s) from the sample matrix to avoid or reduce interferences caused by the matrix. Includes ionexchange, extraction, precipitation procedures, but not vapor generation procedures. Spike equilibration with sample The mixing and equilibration of the enriched isotopic spike with the sample. Signal detection The detection and recording of the analyte isotope signals. The degree of difficulty increases for analytes present at low concentrations, of low isotopic abundance, or that are poorly ionized. Memory effect Any techniques used to avoid, remove or reduce the carry-over of analyte between consecutively measured standards and/or samples. 15

17 Capabilities/Challenges Degree of Difficulty Explanation Low Moderate High Correction or removal of isobaric/polyatomic interferences Any techniques used to remove, reduce, or mathematically correct for interferences caused by mass overlap of analyte isotopes with isobaric or polyatomic species. Includes collision cell techniques, high resolution mass spectrometry, or chemical separations. The relative concentrations and sensitivities of the analyte isotopes and the interfering species will affect the degree of difficulty. Detector deadtime correction Measurement of, and correction for, ion detector deadtime. Importance increases in situations where high ion count rates are encountered. Mass bias/fractionation control and correction Techniques used to determine, monitor, and correct for mass bias/fractionation. Spike calibration Techniques used to determine the analyte concentration in the enriched isotopic spike solution. Correction/removal of isobaric/polyatomic interferences (e.g., Sn, MoO) or alternative selection of isotopes is required to prevent isobaric interferences. 16

18 CCQM Study: APMP.QM-S3 Institute(s): NPLI, LIPI Method: ETA-AAS (or GF-AAS) Analyte(s): Cd Inorganic Core Capabilities Summary Table Capabilities/Challenges Degree of Difficulty Explanation Low Moderate High Contamination control and correction techniques and procedures employed to reduce potential contamination of samples as well as blank correction procedures. The level of difficulty is greatest for analytes that are environmentally ubiquitous and also present at very low concentrations in the sample. Digestion/dissolution of organic matrices techniques and procedures used to bring a sample that is primarily organic in nature into solution suitable for liquid sample introduction to the ETA-AAS. Digestion/dissolution of inorganic matrices techniques and procedures used to bring a sample that is primarily inorganic in nature into solution suitable for liquid sample introduction to the ETA-AAS. Volatile element containment techniques and procedures used to prevent the loss of potentially volatile analyte elements during sample treatment and storage. Pre-concentration Techniques and procedures used to increase the concentration of the analyte introduced to the ETA-AAS. Includes evaporation, ion-exchange, extraction, precipitation procedures, but not vapor generation procedures. Matrix separation Techniques and procedures used to isolate the analyte(s) from the sample matrix to avoid or reduce interferences caused by the matrix. Includes ion-exchange, extraction, precipitation procedures, but not vapor generation procedures. Hydride preconcentration/matrix separation of volatile species. Coupling of a hydride system to the ETA-AAS and optimization of conditions. Calibration of analyte concentration The preparation of calibration standards and the strategy for instrument calibration. Includes external calibration and standard additions procedures. Also use of matrix-matched standards to minimize effect of interferences. Signal detection The detection and recording of the absorption signals of analytes. The degree of difficulty increases for analytes present at low concentrations, of low atomic absorption coefficient. Requires selection of operating conditions such as light source, absorption line, Zeeman background correction conditions. Includes selection of signal processing conditions (peak area or height). Memory effect Any techniques used to avoid, remove or reduce the carry-over of analyte between consecutively measured standards and/or samples. Optimization of the furnace temperature 17 Careful matrix matching for external calibration, or checking linearity of calibration by standard addition is required to achieve reliable calibration.

19 Capabilities/Challenges Degree of Difficulty Explanation Low Moderate High program Optimization of temperature and duration of steps for sample drying, pyrolysis to remove (residual) organics, and atomization. Furnace temperature program to minimize analyte loss in the drying/pyrolysis steps, while maximizing analyte vaporization in the atomization step. Correction or removal of matrix effects or interferences Chemical or instrumental procedures used to avoid or correct for spectral and non-spectral interferences. Includes effects of differences in viscosity and chemical equilibrium states of analyte between the standard and sample. Selection of matrix modifier to adjust volatility of analyte and/or matrix to eliminate these effects is also included. Addition of reactive gases (eg oxygen) to the carrier gas to improve matrix separation. Also included is Zeeman or other background correction techniques to remove interference due to absorption and scattering from coexisting molecules/atoms in the sample. Correction for additive interferences in standard addition method is difficult. Use of matrix modifier, optimized temperature program for selective volatilization, background correction and removal of matrix interferences (inhibiting formation of free analyte atoms in the atomization step) are required to get right results in external calibration. 18

Analysis of copper in ethanol. Final Report

Analysis of copper in ethanol. Final Report Key comparison CCQM-K100 Analysis of copper in ethanol Final Report Authors: Tao ZHOU () 1, Elias Kakoulides () 2, Yanbei ZHU () 3, Reinhard Jaehrling, Olaf Rienitz () 4, David Saxby () 5, Pranee Phukphatthanachai,

More information

CCQM-K45: Sn in tomato paste key comparison

CCQM-K45: Sn in tomato paste key comparison CCQM-K45: Sn in tomato paste key comparison Final report September 2006 CCQM-K45: Sn in tomato paste key comparison Final report September 2006 Contact Point: Gill Holcombe Tel: +44 (0)20 8943 7442 Prepared

More information

Report of the key comparison CCQM-K108 Determination of arsenic species, total arsenic and cadmium in brown rice flour

Report of the key comparison CCQM-K108 Determination of arsenic species, total arsenic and cadmium in brown rice flour Final Report of CCQM-K108 1 Report of the key comparison CCQM-K108 Determination of arsenic species, total arsenic and cadmium in brown rice flour (Final Report) Authors: Akiharu Hioki, Tomohiro Narukawa,

More information

CCQM-K123. Trace Elements in Biodiesel Fuel Final Report

CCQM-K123. Trace Elements in Biodiesel Fuel Final Report Trace Elements in Biodiesel Fuel Final Report Authors: Takayoshi Kuroiwa, Yanbei Zhu, Kazumi Inagaki (NMIJ) 1, Stephen Long, Steven Christopher (NIST) 2, Mabel Puelles, Monica Borinsky, Nadia Hatamleh

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August Pilot: KRISS, Republic of Korea Yon-Kyu Park

Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August Pilot: KRISS, Republic of Korea Yon-Kyu Park Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August 2018 Pilot: KRISS, Republic of Korea Yon-Kyu Park Muktar Sawi (NMIM) Seif M. Osman (NIS) S.S.K.Titus (NPLI)

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Examples of Method Validation Studies Conducted in Different Economies

Examples of Method Validation Studies Conducted in Different Economies Examples of Method Validation Studies Conducted in Different Economies Template for group discussion National Metrology Laboratory, Malaysia 8th APMP TCQM-DEC MiC Workshop, Kobe, Japan December 2011 Description

More information

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

NPL, India activity related to water as per the need of the country

NPL, India activity related to water as per the need of the country NPL, India activity related to water as per the need of the country, Ph.D. (Sr. Scientist) Chemical & Food BND Group Indian Reference Materials (BND) Division CSIR-National Physical Laboratory, New Delhi,

More information

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3. AFRIMETS Secretariat Private Bag X34 Lynnwood Ridge 0040 AFRIMETS Supplementary Comparison Programme Calibration of Gauge Blocks by Mechanical Comparison Method AFRIMETS.L S3 Final Report Giza, Egypt,

More information

ASIA PACIFIC METROLOGY PROGRAME

ASIA PACIFIC METROLOGY PROGRAME APMP.M.M-K6 (SUB) multiples mass key comparison ASIA PACIFIC METROLOGY PROGRAME Final Report of APMP Comparison of Mass standards APMP.M.M-K6 (SUB) multiples mass key comparison Coordinated by: Vietnam

More information

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method Preparation of nitrous oxide ( O) in air standard being traceable to SI by gravimetric method 15th WMO Meeting of Experts on Carbon Dioxide, Other Greenhouse Gases, and Related Tracer Measurement Techniques

More information

Key comparison CCQM-K34 Assay of potassium hydrogen phthalate

Key comparison CCQM-K34 Assay of potassium hydrogen phthalate Slovak Institute of Metrology Karloveská 63, SK-842 55 Bratislava 4, Slovakia Key comparison CCQM-K34 Assay of potassium hydrogen phthalate Final report Michal Máriássy With participation of: Martin Breitenbach

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Report of the key comparison APMP.QM-K19 APMP comparison on ph measurement of borate buffer

Report of the key comparison APMP.QM-K19 APMP comparison on ph measurement of borate buffer Report of the key comparison APMP.QM-K19 APMP comparison on ph measurement of borate buffer (Final Report) Participants: Akiharu Hioki, Toshiaki Asakai, Igor Maksimov, Toshihiro Suzuki and Tsutomu Miura

More information

APMP supplementary comparison of absorbed dose rate in tissue for beta radiation

APMP supplementary comparison of absorbed dose rate in tissue for beta radiation APMP supplementary comparison of absorbed dose rate in tissue for beta radiation BIPM KCDB: APMP.R(I)-S2 Technical Protocol N. Saito and M. Kato National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki

More information

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium, 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

CCQM-K128: Measurement of Heavy Metals and Organo-Tin in Leather Powder. Final Report

CCQM-K128: Measurement of Heavy Metals and Organo-Tin in Leather Powder. Final Report CCQM-K128: Measurement of Heavy Metals and Organo-Tin in Leather Powder Final Report Authors and Contributors: Liandi Ma 1, QianWang 1, Chao Wei 1, WONG Yee-lok 2, TANG Po-on 2, LEE Ho-man 2, SZE Kwan-lok

More information

Lab i x i u i Lab i x i u i / (mg/kg) / (mg/kg) / (mg/kg) / (mg/kg)

Lab i x i u i Lab i x i u i / (mg/kg) / (mg/kg) / (mg/kg) / (mg/kg) MEASURAND : Mass fraction of total Arsenic in brown rice flour NOMINAL VALUE : 0.05 mg/kg to 0.5 mg/kg MEASURAND : Mass fraction of Cadmium in brown rice flour NOMINAL VALUE : 0.1 mg/kg to 0.5 mg/kg x

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 8th APMP/TCQM Gas CRM Workshop June 10 th, 2010 Tukuba, Ibaraki, Japan Nobuyuki Aoki, Takuya Shimosaka, Nobuhiro Matsumoto, Kenji

More information

GMA STATISTICS Countries. Achievements of metrology HRD service for Foreign customers in 2018

GMA STATISTICS Countries. Achievements of metrology HRD service for Foreign customers in 2018 GMA STATISTICS 018 Achievements of metrology HRD service for Foreign customers in 018 70 people from 1 countries have enjoyed opportunities of studying at KRISS. Group courses Individual courses Graduate

More information

COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY

COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY Final Report COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY List of authors: a) Coordinator of comparison: Alexey Demyanov, D. I.

More information

APMP Key Comparison of DC Voltage at V and 10 V

APMP Key Comparison of DC Voltage at V and 10 V Final Report APMP.EM.BIPM-K11.3 APMP Key Comparison of DC Voltage at 1.018 V and 10 V Kyu-Tae Kim* (Korea Research Institute of Standards and Science, KRISS) Michitaka Maruyama (National Metrology Institute

More information

Basic Digestion Principles

Basic Digestion Principles Basic Digestion Principles 1 From Samples to Solutions Direct Analytical Method Solid Sample Problems: Mech. Sample Preparation (Grinding, Sieving, Weighing, Pressing, Polishing,...) Solid Sample Autosampler

More information

Chemical Standard Department, 1600, Shimo-Takano, Sugito-machi, Kitakatsushikagun, Saitama , JAPAN

Chemical Standard Department, 1600, Shimo-Takano, Sugito-machi, Kitakatsushikagun, Saitama , JAPAN International Key Comparison APMP.QM-K4.1 Final report 24/10/2007 Kenji Kato 1, Masaaki Maruyama 2, Jin Seog Kim 3, Oh, Sang Hyub 3, Jin-Chun Woo 3, Yongdoo Kim 3, Hyunkil Bae 3, Qiao Han 4, Zeyi Zhou

More information

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič H1: Metrological support Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič Basic requirement How to ensure appropriate measurements and monitor changes in time and space? Analytical

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3

FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3 FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3 29 November, 2011 Dr. Hyu-Sang Kwon, Dr. Sang-Joon Suh and Jae-Gap Suh Korea Research Institute of Standards and Science (KRISS) Metrologia Tech. Suppl. 49

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test UKWIR SS 17 PAH in surface water with suspended solids Anthracene, Fluoranthene, Naphtalene, Benzo[a]pyrene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13 Determination of Major, Minor and Trace Elements in a Clay Sample IAEA Laboratories, Seibersdorf November 2017 CONTENTS

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Final Report for the APMP.T-K4 (Draft B on October 27, 2011)

Final Report for the APMP.T-K4 (Draft B on October 27, 2011) Final Report for the APMP.T-K4 (Draft B on October 27, 2011) Comparison of Realizations of Aluminum Freezing-Point Temperatures Prepared by K. S. Gam (coordinator) and W. Joung Korea Research Institute

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

International Atomic Energy Agency

International Atomic Energy Agency International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ANAB Cert# AR-1470 ISO/IEC 17025 ANAB Cert# AT-1467 ATROPINE SULFATE CERTIFIED REFERENCE MATERIAL CERTIFIED PURITY: 99.9%, U crm = ±0.2% k = 2 (Mass Balance/anhydrous

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Compact Orifice FlowMeter

Compact Orifice FlowMeter Steam Data Reference Tables IM/OM/SDT Issue 1 Compact Orifice FlowMeter ABB The Company We are an established world force in the design and manufacture of instrumentation for industrial process control,

More information

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures RADIOLOGICAL CHARACTERIZATION Laboratory Procedures LORNA JEAN H. PALAD Health Physics Research Unit Philippine Nuclear Research Institute Commonwealth Avenue, Quezon city Philippines 3-7 December 2007

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ACLASS Cert# AR-1470 ISO/IEC 17025 ACLASS Cert# AT-1467 AMOXICILLIN TRIHYDRATE CERTIFIED REFERENCE MATERIAL HO O H H N H S CH 3 3H 2 O NH 2 N CH 3 COOH CERTIFIED PURITY:

More information

ANALYSIS CERTIFICATE OF

ANALYSIS CERTIFICATE OF CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 INORGANIC VENTURES is an

More information

Final report on CCQM-K36.1 with erratum

Final report on CCQM-K36.1 with erratum Final report on CCQM-K36.1 with erratum 5 October 2010 Ref 1107-03 HDJ Report no. DFM-2008-R07 Summary A follow-up comparison to CCQM-K36 has been carried out November 2007 to February 2008, designated

More information

Comparison protocol EURAMET project

Comparison protocol EURAMET project Comparison protocol EURAMET project Bilateral three-phase AC and energy comparison at frequency 1. Introduction This bilateral comparison was triggered by MIKES need to get support for their planned extension

More information

Prof. Dr. Biljana Škrbić, Jelena Živančev

Prof. Dr. Biljana Škrbić, Jelena Živančev 5 th CEFSER Training Course Analysis of chemical contaminants in food and the environment Faculty of Technology, University of Novi Sad, Novi Sad, Republic of Serbia 7-11 May 2012 Analysis of heavy elements

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Report on APMP Supplementary Comparison High precision roundness measurement APMP.L-S4. Final report

Report on APMP Supplementary Comparison High precision roundness measurement APMP.L-S4. Final report National Institute of Metrology (Thailand) Report on APMP Supplementary Comparison APMP.L-S4 J. Buajarern, National Institute of Metrology (Thailand) (NIMT), Thailand K. Naoi, National Metrology Institute

More information

Isotope Dilution Mass Spectrometry

Isotope Dilution Mass Spectrometry Isotope Dilution Mass Spectrometry J. Ignacio Garcia Alonso and Pablo Rodriguez-Gonzalez Faculty of Chemistry, University of Oviedo, Oviedo, Spain E-mail: jiga@uniovi.es, rodriguezpablo@uniovi.es RSC Publishing

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

International Standardization for Measurement and Characterization of Nanomaterials

International Standardization for Measurement and Characterization of Nanomaterials 2012/SOM2/SCSC/CON/008 International Standardization for Measurement and Characterization of Nanomaterials Submitted by: Japan 9 th Conference on Standards and Conformance in the Field of Innovations and

More information

Improvement of traceability in the field of mass concentration measurements of particles in aerodisperse media carried out in the Russian Federation

Improvement of traceability in the field of mass concentration measurements of particles in aerodisperse media carried out in the Russian Federation International Bureau of Weights and Measures Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology Working Group on Gas Analysis PARTICULATE WORKSHOP Improvement of traceability

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ACLASS Cert# AR-1470 ISO/IEC 17025 ACLASS Cert# AT-1467 CAFFEIE MELTIG POIT STADARD CERTIFIED REFERECE MATERIAL O CH 3 H 3 C O CH 3 OMIAL PACKAGE SIZE: 1g CATALOG #:

More information

Laboratory Performance Assessment. Analysis of Analytes in Dried Apple Chips. Report

Laboratory Performance Assessment. Analysis of Analytes in Dried Apple Chips. Report Laboratory Performance Assessment Analysis of Analytes in Dried Apple Chips Report September 2016 Summary The laboratory performance assessment related to analytes in dried apple chips was designed and

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2. JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS Deliverable 1.2.32 (June 2015) Author : Guillaume Labarraque 1 Introduction High purity

More information

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

CCM short note on the dissemination process after the proposed redefinition of the kilogram

CCM short note on the dissemination process after the proposed redefinition of the kilogram CCM short note on the dissemination process after the proposed redefinition of the kilogram Consultative Committee for Mass and Related Quantities 1. Introduction This note proposes how the mise en pratique

More information

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

GULF ASSOCIATION FOR METROLOGY. Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1

GULF ASSOCIATION FOR METROLOGY. Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1 GULF ASSOCIATION FOR METROLOGY Calibration of Gauge Blocks by Mechanical Comparison Method GULFMET.L-S1 Instructions and Technical Protocols GULFMET.L-S1: Calibration of Gauge Blocks by Mechanical Comparison

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for

More information

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report Michele M. Schantz, Reenie M. Parris, and Willie E. May National Institute of Standards and Technology (NIST)

More information

Nanocrystalline Cellulose:

Nanocrystalline Cellulose: Nanocrystalline Cellulose: International Standardization & Metrological Approaches Alan Steele National Research Council Institute for National Measurement Standards TAPPI International Conference on Nano

More information

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Bilateral Comparison of between NMIA (Australia) and SPRING (Singapore) (KCDB reference No. CCPR-K1.a.1) Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Contents 1. Introduction..2 2.

More information

Sample Analysis Design Internal Standard

Sample Analysis Design Internal Standard Internal Standard Advantages: Fluctuations are monitored in each sample/ calibration / blank Disadvantages: Assume that behavior of IS is the same as the analyte Instrumental Drift Correct for instrument

More information

Certificate of Analysis

Certificate of Analysis The Japan Society for Analytical Chemistry Certificate of Analysis Certified Reference Material JSAC 0601-3 Plastics (tip form) for Chemical Analysis of Hazardous Metals This certified reference material

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688 Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation. Determination of Cadmium and Lead in powdered infant formula by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction NOTICE: This method: - has to be considered only

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

APMP TCTF Working Group on the Mutual Recognition Agreement

APMP TCTF Working Group on the Mutual Recognition Agreement Asia Pacific Metrology Programme APMP TCTF Working Group on the Mutual Recognition Agreement Louis Marais, NMIA Michael Wouters, NMIA APMP, Beijing, November 2015 Role of the WG MRA Co-ordinate both inter-

More information

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Technical Overview Abstract Inductively coupled plasma mass spectrometry

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com CERTIFICATE OF ANALYSIS tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4. CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS 1. SCOPE AND FIELD OF APPLICATION The method describes the determination of heavy metals (arsenic, cadmium, lead and mercury) in cosmetic products. 2. PRINCIPLE Organic matter in sample is digested by

More information

Sample Analysis Design PART II

Sample Analysis Design PART II Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com CERTIFICATE OF ANALYSIS tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Ask The Expert Webinar Series EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Richard Burrows, Ph.D. Corporate Technical Director A Revision to the Method Detection Limit EPA

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

á233ñ ELEMENTAL IMPURITIES PROCEDURES

á233ñ ELEMENTAL IMPURITIES PROCEDURES Second Supplement to USP 38 NF 33 Chemical Tests / á233ñ Elemental Impurities Procedures 1 á233ñ ELEMENTAL IMPURITIES PROCEDURES INTRODUCTION This chapter describes two analytical procedures (Procedures

More information

Report of the Inter-laboratory Comparison Project 2004 on Inland Aquatic Environment

Report of the Inter-laboratory Comparison Project 2004 on Inland Aquatic Environment The Network Center for the Acid Deposition Monitoring Network in East Asia Report of the Inter-laboratory Comparison Project 2004 on Inland Aquatic Environment 5 th Attempt November 2005 Acid Deposition

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information