CCQM-K45: Sn in tomato paste key comparison

Size: px
Start display at page:

Download "CCQM-K45: Sn in tomato paste key comparison"

Transcription

1 CCQM-K45: Sn in tomato paste key comparison Final report September 2006

2 CCQM-K45: Sn in tomato paste key comparison Final report September 2006 Contact Point: Gill Holcombe Tel: +44 (0) Prepared by: Céline S.J. Wolff Briche, Gill Holcombe, Mike Sargent Approved by: Date: 12 September 2006 LGC Limited 2006

3 Contents 1 Summary 1 2 Introduction 1 3 Rationale of this comparison 1 4 Participation in CCQM-K Instructions to participants 2 6 Methods and instrumentation used 2 7 CCQM-K45 participants results Reported results for Sn Total solids determinations 3 8 Agreement on the key comparison reference value 5 9 Equivalence statements 6 10 Acknowledgements 8 11 References 8 12 Appendix A: Protocol distributed to participants 9 13 Appendix B: Sample preparation and homogeneity Sample preparation Homogeneity of the tin content Homogeneity of the total solids content 13 Page i

4 1 Summary CCQM-K45 key comparison was performed to demonstrate and document the capability of interested National Metrology Institutes in the determination of the mass fraction of Sn in a food matrix. This comparison was an activity of the Inorganic Analysis Working Group of the Comité Consultatif pour la Quantité de Matière and was co-ordinated by LGC (Teddington, UK). The following laboratories participated in this key comparison (alphabetical order). KRISS (Republic of South Korea) LGC (UK) LNE (France) PTB (Germany) UME (Turkey) Good agreement was observed between the reported results of four of the participants. The other participant did not use any HCl during the sample preparation, which can affect the stability of the tin in solution, and its reported value was lower than for the other participants. It was not included in the calculation of the key comparison reference value with the agreement of the NMI concerned. The reference value was calculated as the arithmetic mean of the reported participants results (excluding one), and is equal to mg kg -1, with an expanded uncertainty of 2.4 mg kg -1 (1.1%). The equivalence statements were calculated on this basis. 2 Introduction In October 2004, at the Inorganic Analysis Working Group (IAWG) of the Comité Consultatif de la Quantité de Matière (CCQM) meeting in Mexico, LGC proposed a comparison to analyse three elements in tomato paste. It was agreed to organise two studies in parallel: a key comparison, CCQM-K45, for Sn in tomato paste, and a pilot study, CCQM-P72, for Sn, Pb and Cd in tomato paste. The tomato paste analysed by the CCQM-K45 and CCQM-P72 participants was also used for two studies involving field laboratories. These were organised by LGC as part of the certification as a reference material, and by the Central Science Laboratory (CSL) for a round of the FAPAS PT scheme. 3 Rationale of this comparison This key comparison builds on CCQM-P13 (metals in food digest) pilot study and uses a sample spiked with three toxic metals. The main scope was to underpin Calibration and Measurement Capabilities (CMC) for determination of tin in food, for which there is an EU regulatory limit of 200 mg kg Proficiency testing (FAPAS ) 2,3 has highlighted problems with the measurement of tin in food matrices by field laboratories, leading to inconsistent data between laboratories. In a number of cases, low results were achieved, leading to a low consensus value compared to the National Metrology Institute (NMI) reference value. In addition, the use of multi-element techniques for food applications is increasing rapidly and errors arise because multi-element digestion or extraction procedures are frequently unsuitable for the determination of tin. Key comparison CCQM-K45 (for Sn only) was conducted in parallel with the pilot study CCQM- P72. The pilot study also included lead and cadmium, for which accurate determination at low levels in food is of high importance, and was a test of the capability of NMIs to make accurate measurements at very low concentrations in the presence of interfering species. Key objectives of the study were to establish international comparability for measurement of Sn in food samples and to underpin NMI measurements of toxic metals in a food matrix. Page 1

5 4 Participation in CCQM-K45 The institutes / organisations which registered for CCQM-K45 (as entered in the Key Comparison Database) are listed in Table 1. Table 1. CCQM-K45 participants Institute / Organisation Country Contact KRISS Korea Research Institute of Standards and Science LGC Republic of South Korea United Kingdom E. Hwang R. Hearn LNE Laboratoire National d Essais NMi-VSL NMi Van Swinden Laboratorium B.V PTB Physikalisch-Technische Bundesanstalt UME National Metrology Institute France The Netherlands Germany Turkey G. Labarraque M. van Son D. Schiel I. Akdağ D. Karakas NMi-VSL (The Netherlands) registered to participate in this key comparison but, due to the relocation of its facilities, it was unable to report its results within the deadline for this study. 5 Instructions to participants A protocol (Appendix A) was sent to all participants prior to sample distribution and provided information concerning storage and analysis of the sample. Participants were free to use a method of their choice for tin. They were asked to report results on the sample as received, as this is the usual practice for field laboratories. However, to ensure no changes had occurred in the sample between bottling and analysis, or to explain any differences in tin results, participants were asked to determine the total solids content using a common procedure (see Appendix A). 6 Methods and instrumentation used Participants were free to use a method of their choice for Sn, however, it was recommended that HCl should be used during the digestion, as it stabilises the tin in solution (Appendix A). All participants used microwave digestion, using various reagents, including HNO 3, HCl, and H 2 O 2. However, UME did not use HCl, and LGC did not include H 2 O 2. All participants carried out their measurements by ICP-MS, and four of the five participants applied isotope dilution. One participant (PTB) used a magnetic sector ICP-MS, UME did not detail which instrument was used, and the three other participants used quadrupole ICP-MS. Table 2 gives an overview of the methods applied and the instrumentation used by each CCQM-K45 participant. For the total solids determination, a specific protocol was recommended in order to be able to compare the results directly. This is necessary as excess drying may cause degradation of the sample. Page 2

6 Table 2. Analytical methods and instrumental techniques used by CCQM-K45 participants Participant Digestion Method Instrumentation KRISS LGC LNE PTB UME Microwave digestion, IDMS Quadrupole ICP-MS with HNO 3 /HCl/H 2 O 2 Dynamic Reaction Cell Microwave digestion, HNO 3 /HCl IDMS Quadrupole ICP-MS with collision cell Microwave digestion, IDMS Quadrupole ICP-MS HNO 3 /HCl/H 2 O 2 Microwave digestion, IDMS Magnetic sector field ICP- HNO 3 /HCl/H 2 O 2 MS Microwave digestion, HNO 3 /H 2 O 2 3 point external calibration ICP-MS 7 CCQM-K45 participants results 7.1 Reported results for Sn The CCQM-K45 participants results, as reported to the co-ordinating institute (LGC), are given in Table 3. All uncertainties are expanded uncertainties. The coverage factor used was k = 2, except for KRISS, which used k = All data are reported on the sample as received. Results are represented pictorially in Figure 2. Table 3. CCQM-K45 participants measurement results for Sn Participant Reported result mg kg -1 Expanded uncertainty mg kg -1 Relative uncertainty KRISS % LGC % LNE % PTB % UME % 7.2 Total solids determinations All participants applied the protocol described in Appendix A, subject to the variations noted below. The average and standard deviation calculated from the replicate results reported by the participants are given in Table 4 and Figure 1. Page 3

7 Table 4. CCQM-K45 participants total solids results Participant Average total solids g kg -1 Standard deviation g kg -1 Pre-drying stage KRISS Vacuum oven at 70 C for 1 hour, 300 mbar Final drying stage Vacuum oven at 70 C for 1 hour, 250 mbar (6 times) LGC Vacuum oven Vacuum oven at 70 C LNE Conventional oven at 70 C Conventional oven at 70 C PTB Conventional oven at 70 C for 12 hours UME Conventional oven with air flow Vacuum oven at 70 C Two periods of 3 hours, Conventional oven at 70 C for 120 hour Conventional oven at 70 C LNE applied the protocol supplied for 2 determinations, and a classical protocol for 2 other determinations (using the same protocol as in Appendix A, but without using sand or a pre-drying stage). Using their in-house classical protocol, the resulting total solids mass fraction was equal to 278 g kg -1. (1.2 % lower than with the prescribed protocol). The points in Figure 1 represent the average of the total solids determinations, with the bars corresponding to the standard deviation of those results KRISS Mass fraction total solids (g kg -1 ) UME LGC 280 LNE PTB 275 Figure 1: CCQM-K45 participants total solids results PTB mentioned that in spite of the long drying time, the sample did not reach a stable mass within the uncertainty limits of weighing. LNE also found that after 1 week, the loss of mass continues (few milligrams). As the tomato paste was dried longer than recommended, it is possible that not only the water was removed but other components as well. The extended drying time may explain the lower total solids content compared with results reported by the other participants. Page 4

8 There does not seem to be a link between results reported for Sn and the total solids of the sample material. The low mass fraction of Sn reported by UME cannot be explained by a low level of total solids. 8 Calculation of the key comparison reference value There is no CCQM rule for the calculation of the key comparison reference value (KCRV). This is left to the discretion of the IAWG of the CCQM. In the past, the mean * has often been used to calculate the KCRV, but other methods can also be used. The use of the median as KCRV is recommended if there are results reported with significantly lower or higher mass fraction compared to the other participants results. The weighted mean is applicable for independent normally distributed results of the same population. Unless the number of degrees of freedom associated with the data is very large (> 30), the weighted mean s uncertainty is usually an underestimate. When the mean is used, the standard deviation of the mean is used to estimate the standard uncertainty of the reference value. The KCRV for CCQM-K45 was discussed and agreed upon during the CCQM-IAWG meeting in October 2005, in Berlin. UME reported using a microwave digestion with only HNO 3 and H 2 O 2, but no HCl (Table 2). The protocol (Appendix A) recommended that the use of hydrochloric acid will help to maintain the stability of tin in solution. As this participant used no HCl, it is possible that some Sn was lost during the procedure, explaining the lower reported value. The pilot study CCQM-P72, which was conducted in parallel with CCQM-K45, included 6 more participants for the determination of tin in the tomato paste. 4 In this study, most participants reported values for tin which are higher than the one reported by UME. UME was contacted and agreed that as they did not use HCl, their value could be omitted from the calculation of the KCRV. As a result, only 4 reported values are available to calculate the KCRV. The different approaches considered for the calculation of the KCRV are given in Table 5. Not all participants gave details of their uncertainty calculations. KRISS used a coverage factor equal to 2.45, which indicates only 6 degrees of freedom. As a result, the weighted mean is not considered the best estimate of the KCRV, as all results do not have sufficient degrees of freedom. As only 4 values can be used to calculate the KCRV, the median is also not considered the best estimate of the KCRV. Finally, the arithmetic mean as the average of all the reported results was agreed to serve as the KCRV for the CCQM-K45. Its associated expanded uncertainty was calculated as the standard deviation of the mean, multiplied by a coverage factor equal to 2. The arithmetic mean is equal to mg kg -1, the standard deviation of the mean is used as an estimate of its associated uncertainty, leading to an expanded uncertainty equal to 2.4 mg kg -1 (k=2). * The terms "mean" or "average" indicate the arithmetic mean of observations unless otherwise stated. Page 5

9 Table 5: Comparison of the different approaches to calculate the KCRV for CCQM-K45, using only 4 values KCRV mg kg -1 KCRV standard uncertainty (u) mg kg -1 Method of calculation of u(kcrv) Mean Standard deviation of the mean Median MADe Weighted mean Participants uncertainties used as weights CCQM-K45 KCRV: ± 2.4 mg kg -1 (U = k u, k=2) The results of the participants are displayed graphically in Figure 2. The diamonds represent the results reported by each participant, with the bars corresponding to their reported expanded uncertainties. The KCRV is represented by the red line together with its expanded uncertainty represented by dotted lines % Mass fraction of Sn (mg kg -1 ) UME LGC KRISS PTB LNE 2% 0% -2% -4% -6% -8% -10% Relative deviation from reference value (%) % Figure 2: CCQM-K45 participants measurement results for tin, with the KCRV (without UME result) 9 Equivalence statements The equivalence statements have been calculated according to the BIPM guidelines. The degree of equivalence (and its uncertainty) between a NMI result and the KCRV is calculated according to the following equations: 2 2 D = x x U = ( u + u ) i i R i 2 i R where D i is the degree of equivalence between the NMI result x i and the KCRV x R, and U i is the expanded uncertainty (k = 2) of the D i calculated by combining the uncertainty (k = 1) of the NMI result u i and the uncertainty (k = 1) of the KCRV u R. The equivalence statements for CCQM-K45 are given in Table 6, and displayed in Figure 3. Page 6

10 Table 6: Equivalence statements D i (mg kg -1 ) U i (mg kg -1 ) KRISS LGC LNE PTB UME The diamonds displayed in Figure 3 represent the degree of equivalence between each participant and the KCRV. The error bars represent the expanded uncertainty associated with the degree of equivalence LNE Di (mg kg -1 ) KRISS LGC PTB UME -25 Figure 3 Graphical display of equivalence statements. The degree of equivalence (and its uncertainty) between two NMI results is calculated according to the following equations: 2 2 D = x x U = ( u + u ) ij i j ij 2 i j Where D ij is the degree of equivalence between the two NMI results x i and x j, and U ij is the expanded uncertainty (k = 2) of the D ij calculated by combining the uncertainties (k = 1) of the two NMI results u i and u j. The equivalence statement between the CCQM-K45 participants is given in Table 7. Page 7

11 Table 7: Matrix of equivalence between the CCQM-K45 participants Lab j KRISS LGC LNE PTB UME Lab i D ij U ij D ij U ij D ij U ij D ij U ij D ij U ij KRISS LGC LNE PTB UME Acknowledgements The work described here contains the contributions of many scientists: Euijin Hwang from KRISS, Ruth Hearn from LGC, Guillaume Labarraque from LNE, Detlef Schiel and Reinhard Jaehrling from PTB, Ibrahim Akdağ and Duran Karakas from UME. Sample preparation was organised by Joanne Croucher (CSL, York), and the homogeneity measurements were performed by John Lewis (CSL, York). Linda Evans of LGC assisted with testing of the protocol for total solids determination. Claire Carter at LGC assisted with sample packaging and distribution. This work was supported under contract with the United Kingdom s Department of Trade and Industry as part of the National Measurement System Valid Analytical Measurement Programme. 11 References Commission Regulation (EC) No 242/2004. FAPAS report 0738, Central Science Laboratory, York, UK (September 2002) FAPAS report 0745, Central Science Laboratory, York, UK (January 2004) C.S.J. Wolff Briche, G. Holcombe, M. Sargent; CCQM-P72: Sn, Pb and Cd in tomato paste pilot study, LGC/2005/025. Page 8

12 12 Appendix A: Protocol distributed to participants Key Comparison CCQM-K45 Determination of Tin in Tomato Paste Protocol Introduction This key comparison builds on CCQM-P13 (metals in food digest) and uses a sample spiked with tin, lead and cadmium. The main scope is to underpin CMCs for determination of tin in food, for which there is an EU regulatory limit of 200 mg kg -1. Proficiency testing (FAPAS) has highlighted problems with the measurement of tin in food matrices by field laboratories, leading to inconsistent data between laboratories. In a number of cases low results are achieved, leading to a low consensus value compared to the NMI reference value. In addition to this, the use of multi-element techniques for food applications is increasing rapidly and errors arise because the multi-element digestion or extraction procedures in wide use are frequently unsuitable for the determination of tin. A key objective of this comparison is to establish equivalence between reference values provided by NMIs for CRMs or PT schemes and hence to aid international traceability for measurement of tin in food samples. Sample Description The sample consists of a glass bottle containing approximately 50 g of frozen moist tomato paste spiked with tin in the mass fraction range of mg kg -1. The sample should be stored in a freezer at 20 C and thawed when required for use. If more sample is needed contact Gill Holcombe at LGC. Analysis At least three replicate analyses should be carried out. Participants are free to use any suitable method but please include a full description of your method of analysis when reporting the results. A full uncertainty budget should also be included with your results, as indicated below. Results should be reported on a moist weight basis, that is, on the thawed sample as received. Normally field laboratories carry out this measurement on the moist sample for regulatory purposes and it is also quite difficult to dry. Participants are, however, requested to determine the total solids content on a portion of the sample as this will assist in discussing the results in the event of any unexpected variability between laboratories. The recommended protocol for total solids determination is given below and for this part of the study participants are requested to adhere to the protocol to ensure consistent data between laboratories. The sample should be thawed to room temperature prior to analysis. The minimum recommended sample size for each determination is 2 g, which is in accordance with the homogeneity testing of the material. Since the sample contains moisture great care must be taken to avoid moisture loss prior to and during weighing. It should never be left uncovered and weighing out of the sample should be carried out quickly to avoid loss of moisture. It should be noted that the use of hydrochloric acid will help to maintain the stability of tin in solution. Page 9

13 Determination of total solids content This method is derived from AOAC Official Method (Total solids in canned vegetables). For each measurement, dry a flat-bottomed metal dish (containing approximately 15 g sand and a short glass rod), at 100 C for at least 1 hour. At the same time dry a metal lid for the dish. Place the lid on the dish and cool in a desiccator to room temperature and weigh (Weight A). Add approximately 2 g well-mixed tomato paste into the dish, place the lid on the dish and reweigh (Weight B). Weigh quickly to avoid moisture loss during weighing. Mix the tomato paste with the sand using the glass rod, adding a little de-ionised water if necessary to facilitate distribution. Take care not to lose any of the sand during the mixing process. The rod should be left in the dish at all stages of the procedure. Carry out a preliminary drying stage using one of the following methods: (a) Dry the dish in vacuum oven, with the lid to the side of the dish, at 70 C with the release valve left partly open to allow a flow of air through the oven at a pressure > 300 mbar (< 700 mbar vacuum). Examine the dish at 30 min intervals and remove when the sample has reached apparent dryness. (b) As for (a) but use a force-draft oven at 70 C with sufficient flow of outside air to remove moisture rapidly. (c) With the lid removed, dry the dish over a boiling water bath with occasional stirring until the sample reaches apparent dryness. Dry the partially dried sample by placing the dish in vacuum oven, with the lid to the side of the dish, at 70 C, 250 mbar pressure (750 mbar vacuum) for one hour. Place the correct lid on the dish and cool in a desiccator to room temperature and weigh. Repeat drying for one-hour periods until constant weight is achieved (Weight C). Weighing should be done as soon as the dish reaches room temperature, as the sample will rapidly absorb moisture if left standing over most desiccants. Note: some laboratories may not have a vacuum oven. We have also determined the solids content using the procedure described above with a conventional oven at 70 C and ambient pressure. No apparent difference in solids content was seen with this sample. Total Solids (g/100g) Uncertainty Evaluation ( Weight C - Weight A) ( Weight B - Weight A) 100 = Equation 1 Each laboratory should make an assessment of the experimental uncertainty according to ISO principles (Guide to the Expression of Uncertainty in Measurement, ISO, Geneva, 1993, ISBN ). Each variable contributing to the uncertainty of the results should be identified and quantified in order to be included in the combined standard uncertainty of the result. A full uncertainty budget must be included, as part of the results. Contributions to the overall uncertainty will arise from the repeatability of the sample preparation, the repeatability of instrumental determination, determination of masses and volumes, concentration of primary and internal standards, and any other parameter specific to each method of analysis chosen by the participant. Results should be submitted using the results report form provided and sent to Gill Holcombe at LGC, by post, or fax, no later than 31 March Page 10

14 Gill Holcombe Study Co-ordinator LGC Limited Queens Road Teddington Middlesex TW11 0LY United Kingdom Fax: Page 11

15 13 Appendix B: Sample preparation and homogeneity 13.1 Sample preparation Forty-nine cans from the same manufacturing batch of tomato paste, each containing 800 g, were used in the preparation of the test material. The natural levels of lead, cadmium and tin were determined in the paste prior to spiking with the metallic contaminants and were found to be ~ 11µg kg -1 of lead, ~ 21 µg kg -1 of cadmium and ~ 1 mg kg -1 of tin. The paste was spiked with appropriate quantities of lead (12 mg), cadmium (4 mg) and tin (8624 mg) using standard solutions and mixed for 1 hour 30 minutes. Sub-samples (50 g) of the bulked material were weighed into glass bottles. Each bottle was individually numbered and stored at 20 C. The bottles were irradiated at a dose between 14 and 18.3 kgy, and refrozen Homogeneity of the tin content Twenty-two bottles were analysed in duplicate in order to assess the homogeneity of the batch for tin content. Aliquots of 2 g were taken and microwave digested in a mixture of nitric and hydrochloric acid. The measurements were carried out by ICP-MS. The results of these analyses are shown in Figure 4. The points represent the average of the two determinations performed on each bottle, and the bars represent the corresponding standard deviation Mass fraction of Sn (mg kg -1 ) Bottle Figure 4: Results obtained for 22 bottles analysed in duplicate for tin for homogeneity determination The average of all the results is equal to mg kg -1, with a standard deviation equal to 3.5 mg kg -1 (1.6 %). An analysis of variances was performed on these results, and it was concluded that there was no significant differences between the bottles. As a conclusion, this material was fit for the intercomparison. Page 12

16 13.3 Homogeneity of the total solids content The total solids content was also checked, in 10 bottles, in order to check homogeneous distribution of the material. The recommended procedure described in the protocol (Appendix A) was used. The result of these determinations is given in Figure 5. The points represent the average of the two determinations performed on each bottle, and the bars represent the corresponding standard deviation. The red line represents the overall mean. Analysis of these results revealed that there was no significant difference between bottles. The average total solids content, calculated from these 20 data (red line in Figure 5), is equal to g kg -1. An uncertainty contribution was calculated and is equal to 1.8 g kg -1 (0.61 %, with 9 associated degrees of freedom) Total solids (g/kg) Bottles Figure 5: Determination of total solids content in 10 bottles Page 13

APMP.QM-S3 Cd in Rice Powder

APMP.QM-S3 Cd in Rice Powder APMP.QM-S3 Cd in Rice Powder (Coordinated by KRISS, NIM and NMIJ) Final Draft Report Sep. 16, 2011 Yong-Hyeon YIM, Euijin HWANG, Jun WANG, Yanbei ZHU Summary The present study, APMP.QM-S3, was performed

More information

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Application note Environmental Author Sarah Hill and Rebeca Santamaria-Fernandez

More information

CCQM-K14 key comparison Calcium amount content in serum. Final version. L. Van Nevel, Y. Aregbe, P.D.P. Taylor

CCQM-K14 key comparison Calcium amount content in serum. Final version. L. Van Nevel, Y. Aregbe, P.D.P. Taylor h EUROPEAN COMMISSION DIRECTORATE GENERAL JRC JOINT RESEARCH CENTRE IRMM Institute for Reference Materials and Measurements IRMM GE/R/IM/17/03 Isotope Measurements March 2004 CCQM-K14 key comparison Calcium

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

CCQM-K28: Tributyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK

CCQM-K28: Tributyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK CCQM-K28: Tributyltin in sediment Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK a Institute for National Measurement Standards, National Research Council

More information

High pressure comparison among seven European national laboratories

High pressure comparison among seven European national laboratories IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. High pressure comparison among seven European national

More information

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2)

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2) 1 Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K) 1. Introduction L.O. Becerra 1, W. Bich, N. Bignell 3, G.D. Chapman 4, J.W. Chung 5 S. Davidson 6, M. Gläser

More information

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium, 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm NPL REPORT COM 1 NPL REPORT OP 12 Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm - 2500 nm TERESA GOODMAN 1, WILLIAM SERVANTES 1, EMMA WOOLLIAMS 1, PETER SPERFELD 2,

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Final report on CCQM-K36.1 with erratum

Final report on CCQM-K36.1 with erratum Final report on CCQM-K36.1 with erratum 5 October 2010 Ref 1107-03 HDJ Report no. DFM-2008-R07 Summary A follow-up comparison to CCQM-K36 has been carried out November 2007 to February 2008, designated

More information

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW NPL REPORT ENG 17 Report on EURAMET key comparison of multiples and submultiples of the kilogram (EURAMET.M.M-K2.1) M Perkin NOT RESTRICTED July 2009 National Physical Laboratory Hampton Road Teddington

More information

Examples of Method Validation Studies Conducted in Different Economies

Examples of Method Validation Studies Conducted in Different Economies Examples of Method Validation Studies Conducted in Different Economies Template for group discussion National Metrology Laboratory, Malaysia 8th APMP TCQM-DEC MiC Workshop, Kobe, Japan December 2011 Description

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS CERTIFICATE OF ANALYSIS Processed meat Proximates, chloride, nitrate and hydroxyproline Constituent Number of laboratories Certified value 2,3 Uncertainty 4 Moisture 1 (g/kg) 18 618 7 Nitrogen 1 (g/kg)

More information

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

Key Comparisons, the MRA and CMCs: An International Measurement Infrastructure. Robert Wielgosz. Bureau International des Poids et Mesures

Key Comparisons, the MRA and CMCs: An International Measurement Infrastructure. Robert Wielgosz. Bureau International des Poids et Mesures Key Comparisons, the MRA and CMCs: An International Measurement Infrastructure Robert Wielgosz The task of the BIPM is to ensure world-wide uniformity of measurements and their traceability to the International

More information

High pressure comparisons between seven European National Laboratories

High pressure comparisons between seven European National Laboratories J4/2007 High pressure comparisons between seven European National Laboratories Range 50 MPa to 500 MPa Report on EUROMET Project 881 Markku Rantanen, Sari Semenoja, Måns Ackerholm, Antoine Condereys, Zdenek

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

EUROMET Project 702 EUROMET.M.D-K4

EUROMET Project 702 EUROMET.M.D-K4 EUROMET TC-M Meeting 2007 Thursday, 01 March 2007 EUROMET Project 702 EUROMET.M.D-K4 Comparison of the calibrations of high resolution hydrometers for liquid density determinations Salvatore Lorefice INRIM,

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report Michele M. Schantz, Reenie M. Parris, and Willie E. May National Institute of Standards and Technology (NIST)

More information

Guest Forum. ph Measurement using the Harned Cell. The Screening Committee Lecture for the First Dr. Masao Horiba s Award

Guest Forum. ph Measurement using the Harned Cell. The Screening Committee Lecture for the First Dr. Masao Horiba s Award Guest Forum The Screening Committee Lecture for the First Dr. Masao Horiba s Award ph Measurement using the Harned Cell -ph Values Acceptable for the International Society- Susumu Nakamura National Institute

More information

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

VAM Project Development and Harmonisation of Measurement Uncertainty Principles VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088

More information

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3. AFRIMETS Secretariat Private Bag X34 Lynnwood Ridge 0040 AFRIMETS Supplementary Comparison Programme Calibration of Gauge Blocks by Mechanical Comparison Method AFRIMETS.L S3 Final Report Giza, Egypt,

More information

EURAMET.QM-K12. EURAMET Key Comparison on the Determination of the Mass Fraction of Creatinine in Human Serum. Final report

EURAMET.QM-K12. EURAMET Key Comparison on the Determination of the Mass Fraction of Creatinine in Human Serum. Final report EURAMET.QM-K1 EURAMET Key Comparison on the Determination of the Mass Fraction of Creatinine in Human Serum Final report November 01. D. Bell, C. Hopley, S. Ellison & G. O Connor LGC, Teddington, UK With

More information

NPL, India activity related to water as per the need of the country

NPL, India activity related to water as per the need of the country NPL, India activity related to water as per the need of the country, Ph.D. (Sr. Scientist) Chemical & Food BND Group Indian Reference Materials (BND) Division CSIR-National Physical Laboratory, New Delhi,

More information

CCQM-P43: Tributyltin and dibutyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK

CCQM-P43: Tributyltin and dibutyltin in sediment. Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK CCQM-P43: Tributyltin and dibutyltin in sediment Céline S.J. Wolff Briche, Raimund Wahlen and Ralph E. Sturgeon a LGC, Teddington, Middlesex, UK a Institute for National Measurement Standards, National

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials A. Verbruggen, R. Eykens, F. Kehoe, H. Kühn, S. Richter, Y. Aregbe EUR 22944 EN - 2007 The mission of the IRMM

More information

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method Preparation of nitrous oxide ( O) in air standard being traceable to SI by gravimetric method 15th WMO Meeting of Experts on Carbon Dioxide, Other Greenhouse Gases, and Related Tracer Measurement Techniques

More information

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn Aimo Pusa 09.02.2009 MIKES Finland 2 Content Page Content 3 Foreword 4 Chapter 1 1.1 General 6 1.2 Characteristics of the transducers 8 1.3

More information

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. A comparison of primary platinum-iridium kilogram mass

More information

Analysis of copper in ethanol. Final Report

Analysis of copper in ethanol. Final Report Key comparison CCQM-K100 Analysis of copper in ethanol Final Report Authors: Tao ZHOU () 1, Elias Kakoulides () 2, Yanbei ZHU () 3, Reinhard Jaehrling, Olaf Rienitz () 4, David Saxby () 5, Pranee Phukphatthanachai,

More information

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT Michael Borys, Michael Gläser, Michael Mecke Physikalisch-Technische Bundesanstalt (PTB), Germany ABSTRACT Spheres made of a silicon

More information

ANALYSIS CERTIFICATE OF

ANALYSIS CERTIFICATE OF CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 INORGANIC VENTURES is an

More information

Key comparison CCQM-K34 Assay of potassium hydrogen phthalate

Key comparison CCQM-K34 Assay of potassium hydrogen phthalate Slovak Institute of Metrology Karloveská 63, SK-842 55 Bratislava 4, Slovakia Key comparison CCQM-K34 Assay of potassium hydrogen phthalate Final report Michal Máriássy With participation of: Martin Breitenbach

More information

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688 Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211

More information

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards NPL REPORT CMAM 36(a) A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards This Report Supersedes NPL Report CMAM 36 I Severn SEPTEMBER 2000

More information

DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1)

DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1) 1 DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1) C. Aupetit 1, L.O. Becerra, N. Bignell 3, W. Bich 4, G.D. Chapman 5, J.W. Chung 6, J. Coarasa 7, S. Davidson

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES APPLICATION NOTE 43145 Analysis of trace elements in polymer s to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES Authors Introduction Sanja Asendorf, Application

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Proficiency Test Report

Proficiency Test Report Smokeless Tobacco Sub-Group Proficiency Test Report The Determination of Arsenic, Beryllium, Cadmium, Chromium, Cobalt, Nickel, Lead, and Selenium in Reference Materials September 2015 Author and Sub-Group

More information

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS The Fifth International Proficiency Testing Conference Timisoara, Romania () th 8 th September, 0 RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS Adriana Vâlcu, Sterică Baicu National Institute of Metrology,

More information

CCQM KEY COMPARISON K75 DETERMINATION OF TOXIC METALS IN ALGAE

CCQM KEY COMPARISON K75 DETERMINATION OF TOXIC METALS IN ALGAE CCQM KEY COMPARISON K75 DETERMINATION OF TOXIC METALS IN ALGAE Prepared by A. Shakhashiro 1, A. Toervenyi 1, S. Gaudino 2, S. Rosamilia 2, M. Belli 2 G. C. Turk 3 1 IAEA Environment Laboratories, Austria.

More information

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011 Proficiency testing: Aqueous ethanol Test and measurement Workshop Marcellé Archer 20 September 2011 NMISA Ethanol PT schemes Purpose of a PT scheme Consistent good performance is the goal One bad result

More information

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany Guidelines on the Calibration of Temperature Block Calibrators Authorship and Imprint This document was developed by the EURAMET e.v., Technical Committee for Thermometry. Authors: Yves Hermier (LNE-INM,

More information

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction Determination of Nutrients Determination of total phosphorus Extraktion with aqua regia, reflux method Introduction This document is developed in the project Horizontal. It is the result of desk studies

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

Introduction to the evaluation of uncertainty

Introduction to the evaluation of uncertainty Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials SECTION 1 Introduction to the evaluation of uncertainty F A Kandil National Physical Laboratory

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

Annex 2 Formaldehyde

Annex 2 Formaldehyde Annex 2 Formaldehyde The 2 methods are analytical procedures for the determination of formaldehyde CH2O, PM/Ref. No 17260, and hexamethylenetetramine (HMTA) C6H12N4, PM/Ref. No 18670, in food simulants

More information

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič H1: Metrological support Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič Basic requirement How to ensure appropriate measurements and monitor changes in time and space? Analytical

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

ANNEXE 8. Technical protocols for the interlaboratory comparisons

ANNEXE 8. Technical protocols for the interlaboratory comparisons ANNEXE 8 Technical protocols for the interlaboratory comparisons TECHNICAL PROTOCOL FOR BILATERAL COMPARISON (Magnitude of the complex charge sensitivity of accelerometers) EURAMET.AUV.V-S1 1. INTRODUCTION

More information

Introduction. Martin J. T. Milton. Director of the BIPM. Monday 27th June 2016

Introduction. Martin J. T. Milton. Director of the BIPM. Monday 27th June 2016 Introduction Martin J. T. Milton Director of the BIPM Monday 27th June 2016 Bureau International des Poids et Mesures Established in 1875 by the Metre Convention Based in Paris and financed by 57 Member

More information

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf

More information

Certificate of analysis Certified reference material code: CRM-00-GTX1&4

Certificate of analysis Certified reference material code: CRM-00-GTX1&4 Certificate of analysis Certified reference material code: CRM-00-GTX1&4 Name: Gonyautoxin-1 and Gonyautoxin-4 in 0.003 M HCl aq. Batch: 15-001 Description: This Certified Reference Material (CRM) is a

More information

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68

Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 Preparation of nitrous oxide in air standard (320 nmol/mol) for CCQM-K68 8th APMP/TCQM Gas CRM Workshop June 10 th, 2010 Tukuba, Ibaraki, Japan Nobuyuki Aoki, Takuya Shimosaka, Nobuhiro Matsumoto, Kenji

More information

Comparison Euramet 898

Comparison Euramet 898 Euramet 898 Electrolytic at pure water level final report 2010-02-05 Euramet Electrochemical Analysis WG Comparison Euramet 898 Electrolytic at pure water level Petra Spitzer 1, Hans Jensen 2, Bertil Magnusson

More information

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN Michael Stock and Stéphane Solve Bureau International des Poids et Mesures (BIPM) Pavillon de Breteuil,

More information

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of

More information

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015 Uncertainty of Measurement (Analytical) Maré Linsky 14 October 015 Introduction Uncertainty of Measurement GUM (Guide to the Expression of Uncertainty in Measurement ) (Analytical) Method Validation and

More information

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air INTERNATIONAL DOCUMENT OIML D 28 Edition 2004 (E) Conventional value of the result of weighing in air Valeur conventionnelle du résultat des pesées dans l'air OIML D 28 Edition 2004 (E) ORGANISATION INTERNATIONALE

More information

Preparation of Food and Feed Samples for Metals Analysis

Preparation of Food and Feed Samples for Metals Analysis White Paper Preparation of Food and Feed Samples for Metals Analysis Alan Cross Technical Specialist, Metals Laboratory Preparation of Food and Feed Samples for Metals Analysis The relative virtues of

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test UKWIR SS 17 PAH in surface water with suspended solids Anthracene, Fluoranthene, Naphtalene, Benzo[a]pyrene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,

More information

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012) Euramet project 87 Comparison of Instrument Current Transformers up to 0 ka Technical protocol (March 0). Introduction AC current ratio is one of two basic parameters in the area of metrology of instrument

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

PT-Providers: Providers: Key Partners of

PT-Providers: Providers: Key Partners of PT-Providers: Providers: Key Partners of European Metrology for Dissemination i of Traceability in Analytical Chemistry Detlef Schiel (1), Olaf Rienitz (1), Reinhard Jährling (1), Bernd Güttler (1), Holger

More information

Schedule for a proficiency test

Schedule for a proficiency test Schedule for a proficiency test Once a laboratory haegistered the year s programme the interaction between the laboratory and Eurofins Miljø A/S will be according to the schedule described below. 1 Reminder

More information

INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) BIPM Pavillon de Breteuil F Sevres France

INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) BIPM Pavillon de Breteuil F Sevres France INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) A Report for BIPM Pavillon de Breteuil F-92312 Sevres France Project No: FFIN56 October 2008 Report No: XXXX/XXX

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017 Page 1 of 3 Abstract The increasing use of consumer and electronic device components worldwide has drawn increased attention to their impact on the environment. The correct disposal of these materials

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

ASIA PACIFIC METROLOGY PROGRAME

ASIA PACIFIC METROLOGY PROGRAME APMP.M.M-K6 (SUB) multiples mass key comparison ASIA PACIFIC METROLOGY PROGRAME Final Report of APMP Comparison of Mass standards APMP.M.M-K6 (SUB) multiples mass key comparison Coordinated by: Vietnam

More information

Certificate of analysis Certified reference material code: CRM-00-NEO

Certificate of analysis Certified reference material code: CRM-00-NEO Nº 2/PMR3 Certificate of analysis Certified reference material code: CRM--NEO Name: Neosaxitoxin dihydrochloride in.3 M HCl aq. Batch: 14-1 Description: This Certified Reference Material (CRM) is a solution

More information

International interlaboratory study of forensic ethanol standards

International interlaboratory study of forensic ethanol standards International interlaboratory study of forensic ethanol standards Bernard King and Richard Lawn LGC, Queens Road, Teddington, Middlesex, UK TW11 0LY. E-mail: bmjk@lgc.co.uk Received 18th January 1999,

More information

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS D. E. McCurdy, Duke Engineering & Services, Bolton, MA and J. S. Morton, Department of Energy, Office of Environmental

More information

Final Report CCEM.RF-K8.CL COMPARISON CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report CCEM.RF-K8.CL COMPARISON CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report CCEM.RF-K8.CL COMPARISON CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft,

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Report on EURAMET.M.M-S9

Report on EURAMET.M.M-S9 FINAL REPORT Report on EURAMET.M.M-S9 Supplementary comparison of 500 microgram, 200 microgram, 100 microgram and 50 microgram weights (EURAMET project number 1310) Stuart DAVIDSON 1, Adriana VALCU 2,

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

Physikalisch-Technische Bundesanstalt

Physikalisch-Technische Bundesanstalt Physikalisch-Technische Bundesanstalt COOMET Supplementary Comparison COOMET.L-S18 (COOMET project: 673/UA-a/15) Involute gear measurement standards Final report September 2017 Coordinator: Physikalisch-Technische

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Comparability and Traceability Point of view from a metrological institute

Comparability and Traceability Point of view from a metrological institute Comparability and Traceability Point of view from a metrological institute Ed W.B. de Leer NMi Van Swinden Laboratory Delft, The Netherlands 1 My Discussions in Metrologyland 2 The CCQM in 1996 3 Traceability

More information

MASS AND VOLUME COMPARISONS AT MIKES

MASS AND VOLUME COMPARISONS AT MIKES MITTATEKNIIKAN KESKUS CENTRE FOR METROLOGY AND ACCREDITATION Julkaisu J4/2000 MASS AND VOLUME COMPARISONS AT MIKES Additional results to the EA intercomparison of weights 1 mg 100 g (Ma1) and to the EUROMET

More information

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS 1. SCOPE AND FIELD OF APPLICATION The method describes the determination of heavy metals (arsenic, cadmium, lead and mercury) in cosmetic products. 2. PRINCIPLE Organic matter in sample is digested by

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

centro tecnológico Development of reference materials: a case study on packaging material as a source of taints in foods

centro tecnológico Development of reference materials: a case study on packaging material as a source of taints in foods centro tecnológico ESN CONFERENCE: SENSORY ANALYSIS. MORE THAN JUST FOOD Session 5: Proficiency testing in sensory analysis. Development of reference materials: a case study on packaging material as a

More information

Laboratory Performance Assessment. Analysis of Analytes in Dried Apple Chips. Report

Laboratory Performance Assessment. Analysis of Analytes in Dried Apple Chips. Report Laboratory Performance Assessment Analysis of Analytes in Dried Apple Chips Report September 2016 Summary The laboratory performance assessment related to analytes in dried apple chips was designed and

More information