QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS TREVOR THANG 1 Supervisors: Dr.

Size: px
Start display at page:

Download "QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS TREVOR THANG 1 Supervisors: Dr."

Transcription

1 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS TREVOR THANG 1 Supervisors: Dr. Eugene Wong 2, Dr. Rob Bartha 1 Department of Medical Biophysics 1, Department of Physics 2 Western University

2 INTRODUCTION Cancer is one of the leading causes of death in Canada. It is an endogenous disease which arises from the spontaneous mutation of normal tissue cells causing them to divide rapidly. About 40% of Canadians will develop cancer in their lifetime and over half of them will die from it. [1] Thus, cancer detection and treatment research has become very prevalent in today s society and is needed to help fight off this disease. One technique used to treat cancer is the use of Radiation Therapy. The protocol for radiotherapy is first acquiring a Computed Tomography (CT) scan of the tumor area and then creating a treatment plan which irradiates the tumor. Radiation can come in the form of x-rays or gamma rays, and the goal is to have these ionizing rays damage and destroy the DNA of the tumor cells, to prevent them from multiplying further. However, radiation cannot be perfectly targeted so some healthy cells are also irradiated in the process. Therefore, in order to avoid a lethal dose of radiation to healthy tissue, the process is spread out over a few weeks. [1] The long length of treatment allows for many variables to change throughout the treatment process. These factors include: patient position, patient weight, organ movement, tumor movement and tumor aggressiveness. Currently, once a treatment plan has been created by medical physicists and oncologists, the plan rarely get revised. For example, if the patient loses a lot of weight (a side effect of chemotherapy), instead of changing the dose to account for the weight loss, an entirely new treatment plan is create; increasing costs, wasting time and resources. If the tumor moves slightly to the left, the patient is moved slightly to the right in order to compensate. This is the established norm which is designed to maximize patient care and hopefully extend the life of the patient. 2 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

3 One key element missing from the radiation therapy is functional information. Currently, treatment plans only use anatomical and x-ray attenuation data from the CT image. Functional imaging could give insight towards the aggressiveness of certain areas of the tumor, as well as potentially allow for the detection of smaller tumors that can t be seen using traditional CT imaging. Also, this data can be tracked throughout the treatment process. Any changes seen can be reflected in the treatment plan by an appropriate adaption which best treats the immediate needs of the patient. However, one obstacle that all imaging scientist must always consider is whether the changes seen on the image are true biological changes or a result of inconsistent imaging environments. Of the many forms of functional imaging, the one being explored is Magnetic Resonance Imaging (MRI), specifically, Magnetic Resonance Spectroscopy (MRS). MRS can be used to detect in vivo metabolites, some of which may be flags for the aggressiveness of cancerous cells. The drawback to MRS is how many more variables the technology introduces. The biggest factor and the factor to be explored is the homogeneity of the main magnetic field (B 0 ). Furthermore, MRS data requires specialized programs to process and display them properly. Thus, it would be ideal to avoid actually processing and analyzing the MRS before first determining whether or not it is good data. The idea proposed is to use more conventional MRI data as surrogates for the quality of MRS data. Whether this can be done and how it can be done will be explored in this experiment. The overall purpose of the project is to calculate quality assurance metrics which can used as thresholds which govern the validity of the imaging data. This experiment will contribute to this project by determining a surrogate metric which can flag the presence of an 3 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

4 inhomogeneous magnetic field (B 0 ), thus giving insight towards the accuracy of the Magnetic Resonance Spectroscopy data. THEORY Magnetic Resonance Imaging (MRI), involves nuclear resonance within a strong magnetic field. The main magnetic field (B 0 ) generated by the MRI machine can range from 1.5T 7T. This magnetic field is typically generated by a super-cooled solenoid and by convention, the field generated is pointing in the ẑ direction of a Cartesian plane. Magnetic Dipole Physics In order to understand Magnetic Resonance (MR) physics, it is important to understand how a singular dipole exists in a magnetic field, and thus, this report will explain the physics of one dipole in a magnet. Though samples contain multiple dipoles, their effects will not be explained as it does not facilitate further understanding to the report. MR data is acquired by taking advantage of the magnetic dipole of certain molecules, mainly 1 H, 31 P, and 13 C. This magnetic dipole (µ) is produced by electrons orbiting the atomic nucleus. (Figure 1) [2] When the magnetic field (B 0 ) is turned on, the atomic dipole within the machine begins to align with the magnetic field. This is due to the fact that this is the most energetically favorable position for the dipoles to exist. (Figure 2) Figure 1: Atomic Magnetic Dipole created by Orbiting Electron 4 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

5 B 0 Figure 2: Magnetic Dipoles in a Sample Left: No B 0 Right: Large B 0 Once the dipole is aligned with the magnetic field, the MRI machine uses radiofrequency (RF) pulses to excite the dipole into a higher energy state such that it no longer aligns with the magnetic field but instead points at some angle away from B 0. (Figure 3) The most common RF pulse is a 90 flip angle pulse which rotates the dipole 90 or into the xy plane, where the magnetic field is defined as the direction of ẑ. Based on magnetostatics, the magnetic dipole (µ) will begin to precess about the z-axis because of a torque created by the magnetic field. This force and direction of this precession is defined by Figure 3: RF Pulse Rotating Magnetic Dipole (µ) Equation 1. 5 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

6 The frequency at which a dipole precesses (Larmor Frequency) is one of the most fundamental principles of MRI. The Larmor Frequency is dependent on each dipole s ( 1 H, 31 P, and 13 C) gyromagnetic ratio ( ) which relates the precession frequency and the magnetic field. The gyromagnetic ratio of 1 H is MHz/T. Torque of a Magnetic Dipole in a Magnetic Field Larmor Frequency Equation 1 is the torque vector on the dipole is the magnetic dipole vector is the magnetic field vector Equation 2 is the angular frequency (rads/sec) is the precession frequency (cycles/sec) is the gyromagnetic ratio is the magnetic field strength The precession about the z-axis however does not remain in the xy plane. As the dipole emits an electromagnetic signal, there is a loss of energy. As a result,, the magnetic dipole cannot remain in this excited state and thus begins to tilt back towards the main magnetic field. As it tilts back, there is a recovery of longitudinal magnetization and a loss of transverse magnetization. The rates of longitudinal recovery and transverse decay can be expressed using the exponential time constants T1 and T2 respectively. Summary: After a RF pulse (90 flip angle pulse), the magnetic dipole is excited into the xy plane and begins to precess at the Larmor frequency as defined by Equation 1 and 2. As the dipole begins to lose its energy, it must return to its ground state (aligned with the magnetic field). The rate at which the loss of magnetization in the transverse plane (xy plane) and the gain 6 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

7 of magnetization in the longitudinal direction (z) can be expressed as the exponential time constants T2 and T1 respectively. Magnetic Resonance Spatial Imaging [3] Magnetic Resonance Imaging is used to measure a variety of variables concerning magnetic dipoles. Common variables are T1 and T2 values and atomic diffusion. However, these signals from the sample have to be labeled in order to create a three dimensional image and this is done through the use of magnetic gradients. These gradients along the x, y and z axis are used to encode spatial data into the spinning dipole by change the magnetic field strength along any axis. Slice Encoding (Z Coordinate) A magnetic gradient is applied in the direction. This gradient causes slight variations in the magnetic field along the z axis and thus, there is a slight energy difference between the ground and excited dipole states. Thus, when a 90 flip angle RF pulse, with the chosen energy is applied to the sample, only the dipoles with that exact energy gap (between ground and excited) will precess and give off signal. As a result, any signal picked up from the sample must have been emitted from the known xy plane. It is important to note that the magnetic field is always pointing in the z direction; gradients simply change the magnitude of the magnetic vector. Phase Encoding (Y Coordinate) A magnetic gradient is applied along the direction. The slight changes in magnetic field affect the precession frequency of the dipoles as explained by Equation 2. This gradient is only applied for a short period of time so that the dipoles are no longer in phase with one another; instead, only dipoles in the same y co-ordinate will have the same phase. 7 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

8 Frequency Encoding (X Coordinate) The final magnetic gradient is applied along the x direction. This causes the Larmor frequency to change and thus some dipoles will rotate faster than others. When this final gradient is turned on, the signal is acquired thus allowing the frequencies to be encoded. Summary: In order to encode spatial information, MRI machines use 3 magnetic gradient encode a three dimensional Cartesian coordinate into the magnetic dipoles. The z coordinate is selected with the use of a specific RF pulse which only excites one slice of the sample. The y coordinate is encoded in the dipole phase which resulted from a temporary magnetic gradient, dephasing the dipoles. Finally, the x coordinate is encoded in the dipole frequencies in the presence of a third magnetic field. All of this information is summarized by Figure 4. Figure 4: Magnetic Resonance Imaging and Spatial Encoding 8 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

9 Chemical Shifts The examples and discussion used throughout this section will pertain to using 1 H as the dipole. However, the basic concepts remain the same for any dipole. Magnetic Resonance Spectroscopy (MRS) is a technique used to gather metabolic data in-vivo from a sample. It can detect protons within slightly different chemical environments and thus protons belonging to specific molecules can be identified and quantified. Typically, magnetic dipoles do not exist by themselves; instead, they are bound to a molecule. When bound to a molecule, the effects of neighboring atoms and their electrons are especially influential on the dipole and its magnetic environment. Electrons can partially shield nuclei from the magnetic field thus changing the effective magnetic field felt by the dipole. However, this effect is very minute: approximately 0.1% of the B 0. Though small, this change in magnetic field is what causes a slight change in Larmor Frequency which is the key to differentiating nuclei from one another. [2] This slight change in frequency can be quantified and must be experimentally determined. This change in frequency is called the chemical shift and is defined by Equation 3. Chemical Shifts Equation 3 represents the chemical shift (ppm) represents the change in frequency due to altered chemical environment (Hz) represents the frequency of the reference molecule (Hz) (for 1 H, the reference molecule is TMS, trimethylsilane) NOTE: The chemical shift values are independent of the main magnetic field strength. 9 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

10 For example, examine the molecule, choline below. (Figure 5) Figure 5: Choline with Highlighted Hydrogen Atoms All hydrogen atoms colored in with the same color are chemically identical due to identical bonding sequences. Thus, it would be expected that the signal received from the blue protons would by nine times the signal received from the green protons. [2] The chemical shifts for each set of protons have been determined by experimentally, the most important chemical shift is the one for the blue protons. This is because there is significantly more signal available and thus, is more noticeable on the trace. (Figure 6) The chemical shift for the blue protons is approximately 3.24ppm. [4] Figure 6: Data from a Prostate Cancer Patient illustrating the Choline Peak at 3.2ppm 10 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

11 Summary: Based on the peaks seen on an MRS trace, metabolites can be identified based on the frequency shifts due to their chemical environments. These shifts are quantified as chemical shift. METHODS In order to generate thresholds, it is important to understand how the data is currently being acquired and the condition/accuracy of the images taken. From London s Cancer Institutes of Health Research (CIHR), multiple types of MR data (T2w, DW-EPI) were gathered from prostate cancer patients who were imaged just prior to their prostatectomy. These images were acquired at Robart s Research Institute using a 3T General Electric (GE) MRI Machine. Of the many patients data files (~ 45), only a small subset (11) were identified to have had MRS data along with the standard MR images. Two types of images were identified to have the largest variability in accuracy: Diffusion Weight Echo Planar Imaging (DW-EPI) and Magnetic Resonance Spectroscopy (MRS). It was suspected that the two resulted in poor data because of inhomogeneous magnetic fields as supported by the MR Physics described in the Theory section of this report. MRS data is deemed poor if the peaks are not clearly resolved. This can result from having thick peaks that have a large Full Width at Half Max (FWHM) (Figure 7). Large FWHM can arise due to many variations in the magnetic field. The peaks blur together and thus, specific metabolites cannot be differentiated from one another. 11 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

12 Figure 7: Large FWHM & Blurring of Peaks 1 & 2 DW-EPI data is deemed poor when here is warping of the image. This is a result of poor spatial encoding. As mentioned previously, the spatial encoding is coded with the use of magnetic gradients. If the magnetic field is non-uniform, multiple additional gradients may exist within the machine which the computer processes unknowingly. The result is a distorted image, such as Figure 8 below. Figure 8: Large Deformation in DW-EPI Since both imaging modalities (MRS and DW-EPI) are theoretically susceptible to an inhomogeneous magnetic field, it is hypothesized that the quality of the two may be correlated. 12 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

13 The quality of the images can be quantified using the Full Width at Half Max (FWHM) for MRS and Volumetric Distortion for DW-EPI. Gathering the FWHM Water and Lipid The MRS data gathered from the MRI is imported into a computer as a P-File, a specially encoded data format which GE uses to protect their data. Using a program developed by Dr. Rob Bartha called fitman, the P-File is converted in a more appropriate format which fitman is able to open and display. A sample raw data is presented in Figure 9. A noise filter is then applied in order to reduce the noise as shown in Figure 10. The largest peak is from water (4.7ppm) within the prostate. The FWHM of this peak is acquired. Afterwards, in order to gather information about smaller peaks, such as the lipid peak, the water peak must be removed from the spectrum using another filtering tool. (Figure 11) Finally, the FWHM of the Lipid peak (1.5ppm) can also be acquired. Figure 9: Raw MRS data from Prostate Figure 10: Noise Filtered MRS Water FWHM Figure 11: Water and Noise Filtered MRS data Lipid FWHM Quantifying the Magnetic Distortion There are many ways to quantify the amount of distortion within the image but the method used in this experiment is a volumetric percent change. Using the Cancer Treatment 13 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

14 Planning System, Pinnacle, manual contours of the prostate were made in two sets of MR images: DW-EPI and T2 Weighted Fast Spin Echo (T2w-FSE) (Figure 12). Unlike DW-EPI which is susceptible to an inhomogeneous magnetic field, T2w-FSE is not due to a special pulse sequence known as a 180 Refocusing Pulse. Since the T2w-FSE volume is closest to the truth, it will be used as the gold standard which the DW-EPI will be compared with. The volumetric percent change was calculated using Equation 4. Figure 12: Manual Contours on Pinnacle of DW-EPI (left) and T2w-FSE (right) Volumetric Percent Change Equation 4 represents the volumetric percent change represents the T2w-FSE Volume of the Prostate represents the DW-EPI Volume of the Prostate The Statistics After completing these important appropriate measurements, statistical analysis was performed to determine whether or not a correlation exists. Also, based on the resulting trends, a suitable threshold was determined. 14 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

15 Volumetric Percent Change (%) RESULTS The first correlation was done with the FWHM of the water peak and the Volumetric Distortion. As the water peak is typically the largest peak, it is a good place to start. The correlation is shown below in Figure Volumetric Distortion as a Surrogate for MRS Quality - Water %Δ = 1.03 FWHM R² = FWHM of Water Peak (Hz) Figure 13: Correlating FWHM of the Suppressed Water Peak with the Volumetric Distortion Plotting FWHM of Suppressed Water Peak vs. Volumetric Percent Change shows a positive correlation with an R 2 value of ( N=11) It is important to note that the x-intercept is not 0. This is due to MR Physics which even in a perfectly uniform magnetic field, prevents a infinitely thin peak from existing. The second largest peak, lipid peak was also analyzed. This was done for consistency to ensure that the FWHM dependency was not associated with certain metabolites. The results of the lipid correlation are shown below in Figure QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

16 Volumetric Percent Change (%) Volumetric Distortion as a Surrogate for MRS Quality - Lipid %Δ = 0.52 FWHM Point A R² = FWHM of Lipid Peak (Hz) Figure 14: Correlating FWHM of The Lipid Peak with the Volumetric Distortion Plotting FWHM of Lipid Peak vs Volumetric Percent Change shows a posistive correlation with an R 2 value of ( N=11) One outlier is present amoungst the data. Though the correlation between the 2 variables is slightly weaker (R 2 = 0.63), the correlation is definitely still present. The reason for a weaker correlation is due to an outlier present (Point A). There are many factors which could have resulted in such a large volumetric percent change (ie. patient movement). Thus, in order generate a threshold, outliers need to be ignored. Therefore, Point A was removed from this data set as shown in Figure QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

17 Volumetric Percent Change (%) Volumetric Distortion as a Surrogate for MRS Quality Lipid (Revised) %Δ = 0.67 FWHM R² = FWHM of Lipid Peak (Hz) Figure 15: Correlating FWHM of The Lipid Peak with the Volumetric Distortion - Revised Plotting FWHM of Lipid Peak vs Volumetric Percent Change shows a posistive correlation with an R 2 value of ( N=10) This change results in a much stronger correlation with an R 2 value of However, with only 10 data points, the correlation would be strengthened if the number of patients available were doubled. 17 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

18 DISCUSSION Since there are strong correlations between MRS FWHM and DW-EPI distortions, quality assurance thresholds can be generated. The R 2 values for the water peak (0.82) and for the lipid peak (0.87) suggest that there is likely to be an underlining cause which is damaging both imaging techniques. It is hypothesized that it may be an inhomogeneous magnetic field, but this was not proven during this experiment. Based off of numerous scientific reports, the typical FWHM for a suppressed water peak should be approximately 8 Hz. [5] Thus, based on the line of best fit in Figure 13, an appropriate threshold for Volumetric Percent Change would be 7%. Of the 11 patients (2 data points are identical), 7 were deemed to be acceptable since they were above the threshold. A similar result was seen for the revised lipid peak data. With a threshold of 7% and the corresponding lipid FWHM threshold of 35Hz, the same 4 patients were deemed unacceptable. In regards to omitting the Point A, the goal of this project is to create a set of thresholds that can be used to quickly check the validity of the MR data. Thus, when generating each threshold, it is not necessary for each threshold to individually differentiate all the good data from the bad. It is hopefully, a cumulative effort of all the quality assurance metrics which gives the best complete picture of the image accuracy. Although, it is expected that even though by our threshold, Point A would be deemed unacceptable (greater than 7% change), other quality assurance metrics could help identify it as being good MRS data (lipid FWHM less than 35Hz). There are many revisions that can be made to improve the results, such as how volumetric distortion was quantified. In this experiment, a volumetric percent change was employed but may not reflect a true value for distortion as shown in Figure 16. Alternatives 18 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

19 include Surface Area-to-Volume ratio and Percent Overlap. Correlations for these new metrics should also be done. Figure 16: Equal Volumes (0% Percent Change) with Large Distortion The scale of this project is very large and thus the results presented in this paper are merely the first of many potential quality assurance metrics. In order to validate this threshold, Receiver Operator Curve analysis must be performed, more patient data must be gathered to strengthen the correlation value. Furthermore, this threshold may be replaced in the future as there are special pulses sequences have been developed that can directly measure the homogeneity of the magnetic field. These techniques include Phase Difference Method ( [6] and Bandwidth-Difference ( ) [7] However, at the time of this research, this data was not being acquired at Robart s and thus a surrogate had to be used. Whether or not these new types of data are beneficial is up to further research which must balance the amount of new information gained with the increased patient scanning time. CONCLUSION In this experiment, DW-EP images were used as a surrogate for MRS quality. The reason for this is due to the hypothesized effect of an inhomogeneous magnetic field on the two data types. Using data acquired from the London s CIHR project, images taken at Robart s were processed and metrics were calculated to quantify the quality of both the DW-EPI and MRS images. Volumetric percent change and FWHM metrics were shown to have a strong correlation. An R 2 19 QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

20 value of 0.82 for the water peak and 0.87 for the lipid peak (0.87) allowed a threshold of 7% volumetric percent change to be calculated. Of the 11 patients, 4 were deemed to have unacceptable data. Further studies for this project include alternative quantification of distortion, analyzing the FWHM of functional metabolites such as citrate and choline, and exploring the effectiveness of new pulse sequence which can directly measure the magnetic field. As more of these quality assurance metrics are determined, medical physicist and oncologist will have more flexibility in their treatment plan. As long as their treatment plan fits within these metrics, the physician is free to adapt the plan giving more power and meaning to the world of Adaptive Radiotherapy. REFERENCES [1] Canadian Cancer Statistics 2011: Featuring Colorectal Cancer, Canadian Cancer Society. Accessed: Feb [2] Dr. Neil Gelman. MEDBIO 3505F Lecture Notes: Fall Western University [3] Functional Magnetic Resonance Imaging. Scott A. Huettel. Sinauer Associates, Inc. Sunderland, Massachusetts, USA. [4] MR Spectroscopy Quick Guide. Siemens Medical. [5] Prostate cancer metabolite quantification relative to water in 1 H-MRSI in vivo at 3 Tesla. Mclean. [6] Magnetic Resonance Imaging Quality Control Manual (American College of Radiology), Weinreb, 1991 [7] Routine Testing of Magnetic Field Homogeneity on clinical MRI Systems. Hua-Hsuan Chen, QUALITY ASSURANCE OF MAGNETIC RESONANCE IMAGING FOR ADAPTIVE RADIOTHERAPY: PRELIMINARY INVESTIGATIONS

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution. Introduction to MRI Spin & Magnetic Moments Relaxation (T1, T2) Spin Echoes 2DFT Imaging Selective excitation, phase & frequency encoding K-space & Spatial Resolution Contrast (T1, T2) Acknowledgement:

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu MRI advantages Superior soft-tissue contrast Depends on among

More information

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam (closed book, 1 sheets of notes double sided allowed, no calculator or other electronic devices allowed) 1. Ultrasound Physics (15 pt) A) (9

More information

Magnetic Resonance Imaging (MRI)

Magnetic Resonance Imaging (MRI) Magnetic Resonance Imaging Introduction The Components The Technology (MRI) Physics behind MR Most slides taken from http:// www.slideworld.org/ viewslides.aspx/magnetic- Resonance-Imaging- %28MRI%29-MR-Imaging-

More information

MRS: IN VIVO SPECTROSCOPIC IMAGING MAIN POINTS

MRS: IN VIVO SPECTROSCOPIC IMAGING MAIN POINTS MRS: IN VIVO SPECTROSCOPIC IMAGING MAIN POINTS 1. A MR spectrum can identify many metabolites other than water by: Locating the peak(s) determined by a characteristic chemical shift (ppm) resulting from

More information

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics Magnetic Resonance Imaging Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics pal.e.goa@ntnu.no 1 Why MRI? X-ray/CT: Great for bone structures and high spatial resolution Not so great

More information

Magnetic resonance imaging MRI

Magnetic resonance imaging MRI Magnetic resonance imaging MRI Introduction What is MRI MRI is an imaging technique used primarily in medical settings that uses a strong magnetic field and radio waves to produce very clear and detailed

More information

Fundamental MRI Principles Module Two

Fundamental MRI Principles Module Two Fundamental MRI Principles Module Two 1 Nuclear Magnetic Resonance There are three main subatomic particles: protons neutrons electrons positively charged no significant charge negatively charged Protons

More information

MRI Physics I: Spins, Excitation, Relaxation

MRI Physics I: Spins, Excitation, Relaxation MRI Physics I: Spins, Excitation, Relaxation Douglas C. Noll Biomedical Engineering University of Michigan Michigan Functional MRI Laboratory Outline Introduction to Nuclear Magnetic Resonance Imaging

More information

Field trip: Tuesday, Feb 5th

Field trip: Tuesday, Feb 5th Pulse Sequences Field trip: Tuesday, Feb 5th Hardware tour of VUIIIS Philips 3T Meet here at regular class time (11.15) Complete MRI screening form! Chuck Nockowski Philips Service Engineer Reminder: Project/Presentation

More information

Nuclei, Excitation, Relaxation

Nuclei, Excitation, Relaxation Outline 4.1 Principles of MRI uclei, Excitation, Relaxation Carolyn Kaut Roth, RT (R)(MR)(CT)(M)(CV) FSMRT CEO Imaging Education Associates www.imaginged.com candi@imaginged.com What nuclei are MR active?

More information

MRI in Review: Simple Steps to Cutting Edge Part I

MRI in Review: Simple Steps to Cutting Edge Part I MRI in Review: Simple Steps to Cutting Edge Part I DWI is now 2 years old... Mike Moseley Radiology Stanford DWI, b = 1413 T2wt, 28/16 ASN 21 San Francisco + Disclosures: Funding NINDS, NCRR, NCI 45 minutes

More information

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons Fundamental MRI Principles Module 2 N S 1 Nuclear Magnetic Resonance There are three main subatomic particles: protons positively charged neutrons no significant charge electrons negatively charged Protons

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 11 pages Written test, 9 December 2010 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: none. "Weighting": All problems weight equally.

More information

The NMR Inverse Imaging Problem

The NMR Inverse Imaging Problem The NMR Inverse Imaging Problem Nuclear Magnetic Resonance Protons and Neutrons have intrinsic angular momentum Atoms with an odd number of proton and/or odd number of neutrons have a net magnetic moment=>

More information

NMR/MRI examination (8N080 / 3F240)

NMR/MRI examination (8N080 / 3F240) NMR/MRI examination (8N080 / 3F240) Remarks: 1. This test consists of 3 problems with at total of 26 sub-questions. 2. Questions are in English. You are allowed to answer them in English or Dutch. 3. Please

More information

Nuclear Magnetic Resonance Imaging

Nuclear Magnetic Resonance Imaging Nuclear Magnetic Resonance Imaging Simon Lacoste-Julien Electromagnetic Theory Project 198-562B Department of Physics McGill University April 21 2003 Abstract This paper gives an elementary introduction

More information

The Basics of Magnetic Resonance Imaging

The Basics of Magnetic Resonance Imaging The Basics of Magnetic Resonance Imaging Nathalie JUST, PhD nathalie.just@epfl.ch CIBM-AIT, EPFL Course 2013-2014-Chemistry 1 Course 2013-2014-Chemistry 2 MRI: Many different contrasts Proton density T1

More information

NMR and MRI : an introduction

NMR and MRI : an introduction Intensive Programme 2011 Design, Synthesis and Validation of Imaging Probes NMR and MRI : an introduction Walter Dastrù Università di Torino walter.dastru@unito.it \ Introduction Magnetic Resonance Imaging

More information

Introductory MRI Physics

Introductory MRI Physics C HAPR 18 Introductory MRI Physics Aaron Sodickson EXRNAL MAGNETIC FIELD, PROTONS AND EQUILIBRIUM MAGNETIZATION Much of the bulk of the magnetic resonance imaging (MRI) scanner apparatus is dedicated to

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Spin Echo Imaging Hahn Spin Echo

More information

Contrast Mechanisms in MRI. Michael Jay Schillaci

Contrast Mechanisms in MRI. Michael Jay Schillaci Contrast Mechanisms in MRI Michael Jay Schillaci Overview Image Acquisition Basic Pulse Sequences Unwrapping K-Space Image Optimization Contrast Mechanisms Static and Motion Contrasts T1 & T2 Weighting,

More information

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging G16.4426 Medical Imaging Physics of Magnetic Resonance Imaging Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department of Electrical and Computer Engineering,

More information

Basis of MRI Contrast

Basis of MRI Contrast Basis of MRI Contrast MARK A. HORSFIELD Department of Cardiovascular Sciences University of Leicester Leicester LE1 5WW UK Tel: +44-116-2585080 Fax: +44-870-7053111 e-mail: mah5@le.ac.uk 1 1.1 The Magnetic

More information

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96 p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96 B 1,94-96 M,94-96 B oro!' 94-96 BIro!' 94-96 I/r, 79 2D linear system, 56 2D FFT, 119 2D Fourier transform, 1, 12, 18,91 2D sinc, 107, 112

More information

Principles of Magnetic Resonance Imaging

Principles of Magnetic Resonance Imaging Principles of Magnetic Resonance Imaging Hi Klaus Scheffler, PhD Radiological Physics University of 1 Biomedical Magnetic Resonance: 1 Introduction Magnetic Resonance Imaging Contents: Hi 1 Introduction

More information

Exam 8N080 - Introduction to MRI

Exam 8N080 - Introduction to MRI Exam 8N080 - Introduction to MRI Friday April 10 2015, 18.00-21.00 h For this exam you may use an ordinary calculator (not a graphical one). In total there are 5 assignments and a total of 50 points can

More information

New developments in Magnetic Resonance Spectrocopy and Diffusion MRI. Els Fieremans Steven Delputte Mahir Ozdemir

New developments in Magnetic Resonance Spectrocopy and Diffusion MRI. Els Fieremans Steven Delputte Mahir Ozdemir New developments in Magnetic Resonance Spectrocopy and Diffusion MRI Els Fieremans Steven Delputte Mahir Ozdemir Overview Magnetic Resonance Spectroscopy (MRS) Basic physics of MRS Quantitative MRS Pitfalls

More information

Electron Spin Resonance, Basic principle of NMR, Application of NMR in the study of Biomolecules, NMR imaging and in vivo NMR spectromicroscopy

Electron Spin Resonance, Basic principle of NMR, Application of NMR in the study of Biomolecules, NMR imaging and in vivo NMR spectromicroscopy Electron Spin Resonance, Basic principle of NMR, Application of NMR in the study of Biomolecules, NMR imaging and in vivo NMR spectromicroscopy Mitesh Shrestha Electron Spin Resonance Electron paramagnetic

More information

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES 1. What are potential consequences to patients and personnel should there be a release of gaseous

More information

Magnetic Resonance Imaging in a Nutshell

Magnetic Resonance Imaging in a Nutshell Magnetic Resonance Imaging in a Nutshell Oliver Bieri, PhD Department of Radiology, Division of Radiological Physics, University Hospital Basel Department of Biomedical Engineering, University of Basel,

More information

Introduction to MRI Acquisition

Introduction to MRI Acquisition Introduction to MRI Acquisition James Meakin FMRIB Physics Group FSL Course, Bristol, September 2012 1 What are we trying to achieve? 2 What are we trying to achieve? Informed decision making: Protocols

More information

Linear and nonlinear spectroscopy

Linear and nonlinear spectroscopy Linear and nonlinear spectroscopy We ve seen that we can determine molecular frequencies and dephasing rates (for electronic, vibrational, or spin degrees of freedom) from frequency-domain or timedomain

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging http://www.qldxray.com.au/filelibrary/mri_cardiovascular_system_ca_0005.jpg Magnetic Resonance Imaging 1 Overview 1. The magnetic properties of nuclei, and how they behave in strong magnetic fields. 2.

More information

Physical Background Of Nuclear Magnetic Resonance Spectroscopy

Physical Background Of Nuclear Magnetic Resonance Spectroscopy Physical Background Of Nuclear Magnetic Resonance Spectroscopy Michael McClellan Spring 2009 Department of Physics and Physical Oceanography University of North Carolina Wilmington What is Spectroscopy?

More information

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR MRI in Practice Third edition Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK and Carolyn Kaut RothRT(R) (MR) (CT) (M) (CV) Fellow SMRT (Section for Magnetic

More information

16.1 Introduction to NMR Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy 4/11/2013

16.1 Introduction to NMR Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy 4/11/2013 What is spectroscopy? NUCLEAR MAGNETIC RESONANCE (NMR) spectroscopy may be the most powerful method of gaining structural information about organic compounds. NMR involves an interaction between electromagnetic

More information

Nuclear Magnetic Resonance Imaging

Nuclear Magnetic Resonance Imaging Nuclear Magnetic Resonance Imaging Jeffrey A. Fessler EECS Department The University of Michigan NSS-MIC: Fundamentals of Medical Imaging Oct. 20, 2003 NMR-0 Background Basic physics 4 magnetic fields

More information

Physics of MR Image Acquisition

Physics of MR Image Acquisition Physics of MR Image Acquisition HST-583, Fall 2002 Review: -MRI: Overview - MRI: Spatial Encoding MRI Contrast: Basic sequences - Gradient Echo - Spin Echo - Inversion Recovery : Functional Magnetic Resonance

More information

K-space. Spin-Warp Pulse Sequence. At each point in time, the received signal is the Fourier transform of the object s(t) = M( k x

K-space. Spin-Warp Pulse Sequence. At each point in time, the received signal is the Fourier transform of the object s(t) = M( k x Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2015 MRI Lecture 4 k (t) = γ 2π k y (t) = γ 2π K-space At each point in time, the received signal is the Fourier transform of the object

More information

Biophysical Chemistry: NMR Spectroscopy

Biophysical Chemistry: NMR Spectroscopy Relaxation & Multidimensional Spectrocopy Vrije Universiteit Brussel 9th December 2011 Outline 1 Relaxation 2 Principles 3 Outline 1 Relaxation 2 Principles 3 Establishment of Thermal Equilibrium As previously

More information

Fluorescence and Nuclear Magnetic Resonance (NMR) Spectroscopy

Fluorescence and Nuclear Magnetic Resonance (NMR) Spectroscopy Fluorescence and Nuclear Magnetic Resonance (NMR) Spectroscopy Murphy, B. (2017). Fluorescence and Nuclear Magnetic Resonance Spectroscopy: Lecture 3. Lecture presented at PHAR 423 Lecture in UIC College

More information

The Theory of Nuclear Magnetic Resonance Behind Magnetic Resonance Imaging. Catherine Wasko Physics 304 Physics of the Human Body May 3, 2005

The Theory of Nuclear Magnetic Resonance Behind Magnetic Resonance Imaging. Catherine Wasko Physics 304 Physics of the Human Body May 3, 2005 The Theory of Nuclear Magnetic Resonance Behind Magnetic Resonance Imaging Catherine Wasko Physics 304 Physics of the Human Body May 3, 2005 Magnetic resonance imaging (MRI) is a tool utilized in the medical

More information

Measuring Spin-Lattice Relaxation Time

Measuring Spin-Lattice Relaxation Time WJP, PHY381 (2009) Wabash Journal of Physics v4.0, p.1 Measuring Spin-Lattice Relaxation Time L.W. Lupinski, R. Paudel, and M.J. Madsen Department of Physics, Wabash College, Crawfordsville, IN 47933 (Dated:

More information

The physics of medical imaging US, CT, MRI. Prof. Peter Bogner

The physics of medical imaging US, CT, MRI. Prof. Peter Bogner The physics of medical imaging US, CT, MRI Prof. Peter Bogner Clinical radiology curriculum blocks of lectures and clinical practice (7x2) Physics of medical imaging Neuroradiology Head and neck I. Head

More information

Physical fundamentals of magnetic resonance imaging

Physical fundamentals of magnetic resonance imaging Physical fundamentals of magnetic resonance imaging Stepan Sereda University of Bonn 1 / 26 Why? Figure 1 : Full body MRI scan (Source: [4]) 2 / 26 Overview Spin angular momentum Rotating frame and interaction

More information

Chemistry 431. Lecture 23

Chemistry 431. Lecture 23 Chemistry 431 Lecture 23 Introduction The Larmor Frequency The Bloch Equations Measuring T 1 : Inversion Recovery Measuring T 2 : the Spin Echo NC State University NMR spectroscopy The Nuclear Magnetic

More information

Chapter 13 Nuclear Magnetic Resonance Spectroscopy

Chapter 13 Nuclear Magnetic Resonance Spectroscopy Organic Chemistry, 6 th Edition L. G. Wade, Jr. Chapter 13 Nuclear Magnetic Resonance Spectroscopy Jo Blackburn Richland College, Dallas, TX Dallas County Community College District 2006, Prentice Hall

More information

7.3.A. The expression for signal recovery is similar to that derived under exercise 7.2 and is given by:

7.3.A. The expression for signal recovery is similar to that derived under exercise 7.2 and is given by: 7..A. Chemical shift difference 3..0. ppm, which equals 54.5 Hz at 3.0 T. Spatial displacement 54.5/00 0.87, which equals.03 cm along the 8 cm side and 0.77 cm along the 6 cm. The cm slice does not have

More information

FREQUENCY SELECTIVE EXCITATION

FREQUENCY SELECTIVE EXCITATION PULSE SEQUENCES FREQUENCY SELECTIVE EXCITATION RF Grad 0 Sir Peter Mansfield A 1D IMAGE Field Strength / Frequency Position FOURIER PROJECTIONS MR Image Raw Data FFT of Raw Data BACK PROJECTION Image Domain

More information

Biomedical Imaging Magnetic Resonance Imaging

Biomedical Imaging Magnetic Resonance Imaging Biomedical Imaging Magnetic Resonance Imaging Charles A. DiMarzio & Eric Kercher EECE 4649 Northeastern University May 2018 Background and History Measurement of Nuclear Spins Widely used in physics/chemistry

More information

Introduction to Medical Imaging. Medical Imaging

Introduction to Medical Imaging. Medical Imaging Introduction to Medical Imaging BME/EECS 516 Douglas C. Noll Medical Imaging Non-invasive visualization of internal organs, tissue, etc. I typically don t include endoscopy as an imaging modality Image

More information

June 16, Signal generation and gradient fields in MRI. Maximilian Oehm. Summary of physical fundamentals. Motivation. Complex representation

June 16, Signal generation and gradient fields in MRI. Maximilian Oehm. Summary of physical fundamentals. Motivation. Complex representation in MRI of Signal in MRI June 16, 2015 in MRI Contents of 1 of 2 3 4 5 6 7 in MRI of of Magnetic field B e z (few T) Splits up energy levels N+ N N ++N 1ppm M = m V B No measurement in z-direction possible

More information

Spectroscopy and braintumours

Spectroscopy and braintumours Spectroscopy and braintumours jatta.berberat@ksa.ch Introduction Background Metabolites Shimming Sequences Summary 1 Spectroscopy Nuclear magnetic resonance spectroscopy (MRS) is an analytical tool, based

More information

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar. 1 Doppler echocardiography & Magnetic Resonance Imaging History: - Langevin developed sonar. - 1940s development of pulse-echo. - 1950s development of mode A and B. - 1957 development of continuous wave

More information

Initial Certification

Initial Certification Initial Certification Medical Physics Part 1 Content Guide Part 1 Content Guides and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 1 Physics Exam The ABR provides

More information

MRI Physics II: Gradients, Imaging. Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor

MRI Physics II: Gradients, Imaging. Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Basic p rinciples COPYRIGHTED MATERIAL. Introduction. Atomic s tructure

Basic p rinciples COPYRIGHTED MATERIAL. Introduction. Atomic s tructure 1 Basic p rinciples Introduction 1 Atomic structure 1 Motion in the atom 2 MR active nuclei 2 The hydrogen nucleus 4 Alignment 4 Precession 8 The Larmor equation 9 Introduction The basic principles of

More information

Principles of MRI. Vinyl Record. Last time: Today: Homework Due tonight! EE225E / BIO265. Transforms a temporal signal to a spatial signal

Principles of MRI. Vinyl Record. Last time: Today: Homework Due tonight! EE225E / BIO265. Transforms a temporal signal to a spatial signal What is this? ` Principles of MRI Lecture 05 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS The first NMR spectrum of ethanol 1951. 1 2 Today Last time: Linear systems, Fourier Transforms, Sampling

More information

Lecture 19: Building Atoms and Molecules

Lecture 19: Building Atoms and Molecules Lecture 19: Building Atoms and Molecules +e r n = 3 n = 2 n = 1 +e +e r y even Lecture 19, p 1 Today Nuclear Magnetic Resonance Using RF photons to drive transitions between nuclear spin orientations in

More information

Introduction to the Course and the Techniques. Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology

Introduction to the Course and the Techniques. Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology Introduction to the Course and the Techniques Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology (jralger@ucla.edu) CTSI Neuroimaging April 2013 Rationale for the Course

More information

Lecture 12 February 11, 2016

Lecture 12 February 11, 2016 MATH 262/CME 372: Applied Fourier Analysis and Winter 2016 Elements of Modern Signal Processing Lecture 12 February 11, 2016 Prof. Emmanuel Candes Scribe: Carlos A. Sing-Long, Edited by E. Bates 1 Outline

More information

Lecture PowerPoints. Chapter 31 Physics: Principles with Applications, 7th edition Giancoli

Lecture PowerPoints. Chapter 31 Physics: Principles with Applications, 7th edition Giancoli Lecture PowerPoints Chapter 31 Physics: Principles with Applications, 7th edition Giancoli This work is protected by United States copyright laws and is provided solely for the use of instructors in teaching

More information

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle L 36 Modern Physics [2] X-rays & gamma rays How lasers work Medical applications of lasers Applications of high power lasers Medical imaging techniques CAT scans MRI s Modern physics ideas are strange!

More information

Lecture 19: Building Atoms and Molecules

Lecture 19: Building Atoms and Molecules Lecture 19: Building Atoms and Molecules +e r n = 3 n = 2 n = 1 +e +e r ψ even Lecture 19, p 1 Today Nuclear Magnetic Resonance Using RF photons to drive transitions between nuclear spin orientations in

More information

Relaxation times in nuclear magnetic resonance

Relaxation times in nuclear magnetic resonance Relaxation times in TEP Related topics Nuclear spins, atomic nuclei with a magnetic moment, precession movement of the nuclear spins, Landau-Lifshitz equation, Bloch equation, magnetisation, resonance

More information

Tissue Characteristics Module Three

Tissue Characteristics Module Three Tissue Characteristics Module Three 1 Equilibrium State Equilibrium State At equilibrium, the hydrogen vector is oriented in a direction parallel to the main magnetic field. Hydrogen atoms within the vector

More information

Principles of MRI EE225E / BIO265. Name That Artifact. RF Interference During Readout. RF Interference During Readout. Lecture 19

Principles of MRI EE225E / BIO265. Name That Artifact. RF Interference During Readout. RF Interference During Readout. Lecture 19 Name That Artifact Principles of MRI EE225E / BIO265 Lecture 19 Instructor: Miki Lustig UC Berkeley, EECS 1 http://mri-info.net 2 RF Interference During Readout RF Interference During Readout 1D FFT 1D

More information

12/1/17 OUTLINE KEY POINTS ELEMENTS WITH UNSTABLE NUCLEI Radioisotopes and Nuclear Reactions 16.2 Biological Effects of Nuclear Radiation

12/1/17 OUTLINE KEY POINTS ELEMENTS WITH UNSTABLE NUCLEI Radioisotopes and Nuclear Reactions 16.2 Biological Effects of Nuclear Radiation OUTLINE 16.1 Radioisotopes and Nuclear Reactions 16.2 Biological Effects of Nuclear Radiation PET scan X-ray technology CT scan 2009 W.H. Freeman KEY POINTS Radioactivity is the consequence of an unstable

More information

Nuclear Physics and Astrophysics

Nuclear Physics and Astrophysics Nuclear Physics and Astrophysics PHY-302 Dr. E. Rizvi Lecture 24 Medical Imaging Effects of Radiation We now know what radiation is But what does it mean for our bodies? Radioactivity is quantified in

More information

AQA Physics /7408

AQA Physics /7408 AQA Physics - 7407/7408 Module 10: Medical physics You should be able to demonstrate and show your understanding of: 10.1 Physics of the eye 10.1.1 Physics of vision The eye as an optical refracting system,

More information

BMB/Bi/Ch 173 Winter 2018

BMB/Bi/Ch 173 Winter 2018 BMB/Bi/Ch 173 Winter 2018 Homework Set 8.1 (100 Points) Assigned 2-27-18, due 3-6-18 by 10:30 a.m. TA: Rachael Kuintzle. Office hours: SFL 220, Friday 3/2 4:00-5:00pm and SFL 229, Monday 3/5 4:00-5:30pm.

More information

Principles of Nuclear Magnetic Resonance Microscopy

Principles of Nuclear Magnetic Resonance Microscopy Principles of Nuclear Magnetic Resonance Microscopy Paul T. Callaghan Department of Physics and Biophysics Massey University New Zealand CLARENDON PRESS OXFORD CONTENTS 1 PRINCIPLES OF IMAGING 1 1.1 Introduction

More information

ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY

ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY 1. AIM OF THE EXPERIMENT This is a model experiment for electron spin resonance, for clear demonstration of interaction between the magnetic moment

More information

Basic MRI physics and Functional MRI

Basic MRI physics and Functional MRI Basic MRI physics and Functional MRI Gregory R. Lee, Ph.D Assistant Professor, Department of Radiology June 24, 2013 Pediatric Neuroimaging Research Consortium Objectives Neuroimaging Overview MR Physics

More information

Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination

Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination Nuclear Magnetic Resonance Spectroscopy: Tools for Structure Determination Chung-Ming Sun Department of Applied Chemistry National Chiao Tung University Hualien 300, Taiwan Introduction NMR (Nuclear Magnetic

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":

More information

Lecture PowerPoint. Chapter 31 Physics: Principles with Applications, 6 th edition Giancoli

Lecture PowerPoint. Chapter 31 Physics: Principles with Applications, 6 th edition Giancoli Lecture PowerPoint Chapter 31 Physics: Principles with Applications, 6 th edition Giancoli 2005 Pearson Prentice Hall This work is protected by United States copyright laws and is provided solely for the

More information

Chapter 14:Physics of Magnetic Resonance

Chapter 14:Physics of Magnetic Resonance Chapter 14:Physics of Magnetic Resonance Slide set of 141 slides based on the chapter authored by Hee Kwon Song of the publication (ISBN 978-92-0-131010-1): Diagnostic Radiology Physics: A Handbook for

More information

The physics US and MRI. Prof. Peter Bogner

The physics US and MRI. Prof. Peter Bogner The physics US and MRI Prof. Peter Bogner Sound waves mechanical disturbance, a pressure wave moves along longitudinal wave compression rarefaction zones c = nl, (c: velocity, n: frequency, l: wavelength

More information

10.4 Continuous Wave NMR Instrumentation

10.4 Continuous Wave NMR Instrumentation 10.4 Continuous Wave NMR Instrumentation coherent detection bulk magnetization the rotating frame, and effective magnetic field generating a rotating frame, and precession in the laboratory frame spin-lattice

More information

Apodization. Gibbs Artifact. Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2013 MRI Lecture 5. rect(k x )

Apodization. Gibbs Artifact. Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2013 MRI Lecture 5. rect(k x ) Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2013 MRI Lecture 5 GE Medical Systems 2003 Gibbs Artifact Apodization rect(k ) Hanning Window h(k )=1/2(1+cos(2πk ) 256256 image 256128

More information

Magnetic Resonance Spectroscopy EPR and NMR

Magnetic Resonance Spectroscopy EPR and NMR Magnetic Resonance Spectroscopy EPR and NMR A brief review of the relevant bits of quantum mechanics 1. Electrons have spin, - rotation of the charge about its axis generates a magnetic field at each electron.

More information

Principles of Molecular Spectroscopy: Electromagnetic Radiation and Molecular structure. Nuclear Magnetic Resonance (NMR)

Principles of Molecular Spectroscopy: Electromagnetic Radiation and Molecular structure. Nuclear Magnetic Resonance (NMR) Principles of Molecular Spectroscopy: Electromagnetic Radiation and Molecular structure Nuclear Magnetic Resonance (NMR) !E = h" Electromagnetic radiation is absorbed when the energy of photon corresponds

More information

Part II: Magnetic Resonance Imaging (MRI)

Part II: Magnetic Resonance Imaging (MRI) Part II: Magnetic Resonance Imaging (MRI) Contents Magnetic Field Gradients Selective Excitation Spatially Resolved Reception k-space Gradient Echo Sequence Spin Echo Sequence Magnetic Resonance Imaging

More information

Advanced Topics and Diffusion MRI

Advanced Topics and Diffusion MRI Advanced Topics and Diffusion MRI Slides originally by Karla Miller, FMRIB Centre Modified by Mark Chiew (mark.chiew@ndcn.ox.ac.uk) Slides available at: http://users.fmrib.ox.ac.uk/~mchiew/teaching/ MRI

More information

Lecture Outlines Chapter 32. Physics, 3 rd Edition James S. Walker

Lecture Outlines Chapter 32. Physics, 3 rd Edition James S. Walker Lecture Outlines Chapter 32 Physics, 3 rd Edition James S. Walker 2007 Pearson Prentice Hall This work is protected by United States copyright laws and is provided solely for the use of instructors in

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

Ferdowsi University of Mashhad

Ferdowsi University of Mashhad Spectroscopy in Inorganic Chemistry Nuclear Magnetic Resonance Spectroscopy spin deuterium 2 helium 3 The neutron has 2 quarks with a -e/3 charge and one quark with a +2e/3 charge resulting in a total

More information

Part III: Sequences and Contrast

Part III: Sequences and Contrast Part III: Sequences and Contrast Contents T1 and T2/T2* Relaxation Contrast of Imaging Sequences T1 weighting T2/T2* weighting Contrast Agents Saturation Inversion Recovery JUST WATER? (i.e., proton density

More information

NMRis the most valuable spectroscopic technique for organic chemists because it maps the carbon-hydrogen framework of a molecule.

NMRis the most valuable spectroscopic technique for organic chemists because it maps the carbon-hydrogen framework of a molecule. Chapter 13: Nuclear magnetic resonance spectroscopy NMRis the most valuable spectroscopic technique for organic chemists because it maps the carbon-hydrogen framework of a molecule. 13.2 The nature of

More information

NMR Spectroscopy: A Quantum Phenomena

NMR Spectroscopy: A Quantum Phenomena NMR Spectroscopy: A Quantum Phenomena Pascale Legault Département de Biochimie Université de Montréal Outline 1) Energy Diagrams and Vector Diagrams 2) Simple 1D Spectra 3) Beyond Simple 1D Spectra 4)

More information

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2006 MRI Lecture 1 Topics The concept of spin Precession of magnetic spin Relaxation Bloch Equation 1 Spin Intrinsic angular momentum of

More information

Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori

Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori Enseignant-chercheur Equipe IMAGES - Télécom ParisTech pietro.gori@telecom-paristech.fr September 20, 2017 P. Gori BIOMED 20/09/2017 1 / 76

More information

Chapter 10. Table of Contents. Section 1 What Is Radioactivity? Section 2 Nuclear Fission and Fusion. Section 3 Nuclear Radiation Today

Chapter 10. Table of Contents. Section 1 What Is Radioactivity? Section 2 Nuclear Fission and Fusion. Section 3 Nuclear Radiation Today Nuclear Chemistry Table of Contents Section 1 What Is Radioactivity? Section 2 Nuclear Fission and Fusion Section 3 Nuclear Radiation Today Section 1 What Is Radioactivity? Bellringer Before studying about

More information

Magnetic Resonance Spectroscopy: Basic Principles and Selected Applications

Magnetic Resonance Spectroscopy: Basic Principles and Selected Applications Magnetic Resonance Spectroscopy: Basic Principles and Selected Applications Sridar Narayanan, PhD Magnetic Resonance Spectroscopy Unit McConnell Brain Imaging Centre Dept. of Neurology and Neurosurgery

More information

16.1 Introduction to NMR. Spectroscopy

16.1 Introduction to NMR. Spectroscopy 16.1 Introduction to NMR What is spectroscopy? Spectroscopy NUCLEAR MAGNETIC RESNANCE (NMR) spectroscopy may be the most powerful method of gaining structural information about organic compounds. NMR involves

More information

INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5)

INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5) INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5) Ionizing radiation: Particle or electromagnetic radiation that is capable of ionizing matter. IR interacts through different types of collision

More information

RADIOCHEMICAL METHODS OF ANALYSIS

RADIOCHEMICAL METHODS OF ANALYSIS RADIOCHEMICAL METHODS OF ANALYSIS 1 Early Pioneers in Radioactivity Rutherfo rd: Discoverer Alpha and Beta rays 1897 Roentge n: Discoverer of X- rays 1895 The Curies: Discoverers of Radium and Polonium

More information

Part III Minor Option in Medical Physics 2018 Examples Sheet

Part III Minor Option in Medical Physics 2018 Examples Sheet Part III Minor Option in Medical Physics 2018 Examples Sheet Any errors or comments should be addressed sent to: seb53@cam.ac.uk URLs that may be useful: Stanford Event Generation Simulator: http://tinyurl.com/pkg476r

More information