1 Report: Intercomparison of calibration of master meters (screw meters) installed in a test rig, by two independent test methods (by BEV-standard capacity tanks and by CMI- Brooks piston prover) Euromet number 831 Authors: Libor Lojek, Gerhard Baubinder Period, in which the intercomparison took place: until Persons in charge: Libor Lojek c/o CMI, CZ Brno/Czech Republic, Okruzni 31 T , E llojek@cmi.cz Gerhard Baubinder c/o BEV, A-1160 Vienna/Austria, Arltgasse 35 T , E g.baubinder@metrologie.at Release date of the report:

2 1 Purpose of the project The error curves of four master meters (subsequently short MMs) (two for gasoline covering a flow rate range of l/min and two for diesel oil for the same flow rate range) are measured by using two different test methods: 1. use of standard capacity tanks (of the BEV test rig). use of a transportable Brooks compact piston prover "BCP" (owner CMI) connected to the master meters for the purpose of this intercomparison. By this, the reliability into the hydrocarbon CMC entries of both countries shall be reinforced. The project is comparable to BCR-project 3476/1/0/03/9/9-BCR-S(30) and anticipating on the agreed CCM.FF-K project on hydrocarbons. The MMs are 4 KRAL screw meters, normally permanently installed in the BEV test rig as secondary working standards. The deviation of the resulting CMI-error curves from the resulting BEV-error curves is a measure of the reproducibility of both test rigs. Brief history of the measurements At the first stage of the intercomparison the MMs were taken from the BEV test rig and installed in line with the BCP at CEPRO cooperation Strelice/Czech Republic. But measurements on the BEV-test rig directly following the tests at CEPRO had shown, that the vertical position of the meters in the BEV test rig with the inlet from below (see table 1 - Test specifications) caused - at certain (low) flow rates -, a pressure of the screw wheels upon the meter case, which led to a singular peak of the error curve. Subsequently, the corresponding pipe of the BEV test rig was changed so that the meters now have their inlet from above. So, a second stage of intercomparisons became necessary: contrary to the first stage the BCP came to Vienna and by this, the test liquid was identical for both the BCP and the BEV-test rig. The test results only represent the measurements of the second stage. Test rigs The CMI-test rig is a mobile Brooks Compact Prover (BCP), mounted on a trailer and used for on-site calibrations of flowmeters e.g. on product pipelines, in refineries, on the crude oil pipeline and in transmission terminals. Serial. Nr , Q = 6,6 l/min 6600 l/min, tube diameter 1 inch, prover volume 60 l. It enables a dynamic measurement under operational conditions without any interruption of the measurement. Connections to the pipework are made by rubber hoses (4 m long, flanges DN100/PN16 or DN150/PN16). The liquid and the hydraulic energy is provided by external sources. The CMI-BCP is calibrated gravimetrically by the water draw method. The BEV has two stationary, equally designed test rigs (one for gasoline and one for diesel oil), each of them being equipped with two of the above mentioned KRAL screw meters as secondary working standards. The primary working standards of the BEV-test rig are static 1000 litres-standard capacity tanks for the flying and start/stop-operation (flying operation by means of a diverter), Q (test rig) = 30 l/min 1500 l/min.

3 The liquid and the hydraulic energy is provided by underground tanks/pumps. The standard capacity tanks are calibrated with water by delivery from a gravimetrically calibrated standard capacity tank. The MMs are calibrated by the standard capacity tanks of the test rig. 3 Meters under test The KRAL screw meters are especially suitable as master meters as well as transfer standards because they are independent of the flow profile at the meter inlet and have a good repeatability and a good long term stability (see also BCR- Kerosene - Report "Calibration Intercomparison on Flowmeters for Kerosene - Synthesis Report Part I & Part II /1/0/03/9/9-BCR-S(30) - SP, Weights and Measures Boras, 1995). Meter output by pulses (K m = 71 Imp/l for the DN40 meters and K m = 17 Imp/l for the DN100 meters); pulses fed into the electronics of the corresponding test rigs. The pulses were transmitted by one channel only, but it was assumed that no disturbances on the lines would occur, because at the comparison no source of interference was present in the vicinity of the installation (no pumps, no controllers); the good repeatability is an evidence that no spurious pulses had occurred. A long term drift is taken into consideration because of the different dates of the calibration of the MMs by the BEV-standard capacity tanks and by the CMI-BCP. As the long term drift of the error curves of the screw meters has not been determined by the BEV yet, the data for the estimation of the drift are taken from the experimental data of Kräutler, the manufacturer of the KRAL-meters (drift KRAL, see table below). screw meter drift KRAL = K/a = u long term u long terrm (% ) DN40 meters: K/a = 0,005 0, type OM5, serial Nr. B1065V (gasoline) und B1064V (diesel), K m = 71 DN100 meters: K/a = 0,00 0, type OM100, serial Nr. B681V (diesel) und B1060V (gasoline), K m = 17 Remark: The short term stability is covered by the uncertainty of repeatability.

4 4 Test specifications Place and date of test, used standards (first stage) first stage: CEPRO/Strelice, June 00 CMI-BCP (first and second stage) first stage: BEV/Vienna, July 00 standard capacity tanks transfer standard position of installation of meters Test liquids second stage: BEV/Vienna, July 003 standard capacity tanks; BEV/Vienna, July 004 CMI-BCP KRAL-screw meter type OM100, serial Nr. B681V, diesel oil, K = 17 Imp/l, Q = l/min, DN100 (pipe thread connection) KRAL-screw meter type OM100, serial Nr. B1060V, gasoline, K = 17 Imp/l, Q = l/min, DN100 KRAL-screw meter type OM5, serial Nr. B1065V, gasoline, K = 71 Imp/l, Q = l/min, DN40 KRAL-screw meter type OM5, serial Nr. B1064V, diesel oil, K = 71 Imp/l, Q = l/min, DN40 horizontal, coupled with outlet of BCP diesel oil ν = 4, mm /s (15 C) (first stage) gasoline ν = 0,57 mm /s (15 C) (first stage) first stage: vertical, meter inlet from below second stage: meter inlet from above diesel oil ν = 5,58 mm /s (15 C) (first stage) ν = 5,6 mm /s (15 C) (second stage) gasoline ν = 0,71 mm /s (15 C) (first stage) ν = 0,71 mm /s (15 C) (second stage) Average liquid 19 C (first stage) C (first stage) temperature * 19 C (second stage) *actual values see protocol Table 1 - Test specifications (one meter pipe thread connection DN40 and one meter pipe thread connection DN100 for gasoline, one meter pipe thread connection DN40 and one meter pipe thread connection DN100 for diesel oil). 5 Test set-up and test procedure 5.1 CEPRO/Strelice/first stage: With the BCP-tests each MM was mounted at the outlet of the BCP. The BCP was installed in a bypass of a pipeline. Prior to all test series each MM was rinsed in the installation for 5 min in order to adapt it to the liquid temperature and to separate eventually entrapped air.

5 The nominal flow rate was set manually (by a valve at the outlet side of the bypass and partially by a valve in the pipeline between inlet and outlet of the bypass) prior to the start of the test series for that definite flow rate. Temperature and pressure were taken from the Pt100 and pressure sensor of the BCP. The viscosity was determined by taking a sample and determining it by the BEV-lab. 5. BEV/Vienna /first stage: The MMs were then installed in the BEV-test rig in their original position. Prior to all test series each MM was rinsed in the installation for 5 min in order to adapt it to the liquid temperature and to separate eventually entrapped air. The nominal flow rate was set in a preliminary run (by a control valve at the outlet of the meters) and the adjustment of that control valve was frozen during the test series for that definite flow rate. Temperature and pressure were taken from the Pt100 and pressure sensor of the test rig. The viscosity was determined by the BEV-lab. 5.3 BEV/Vienna /second stage: The MMs were installed in the BEV-test rig in their original position. Prior to all test series each MM was rinsed in the installation for 5 min in order to adapt it to the liquid temperature and to separate eventually entrapped air. The nominal flow rate was set (by a control valve at the outlet of the meters) and the adjustment of that control valve was frozen during the test series for that definite flow rate. The pressure for the correction of the reference volume of the BCP was taken from the sensor of the BCP; on the contrary the temperature for the correction of the reference volume of the BCP was taken from the test-rig of the BEV because these temperature values seemed to be more plausible: during the tests the Pt100 of the BCP was strongly exposed to sun radiation and the Pt100 is not in direct contact with the liquid but is installed in a well so that heat transmission from outer parts might have corrupted the temperature measurements. The viscosity was determined by the BEV-lab. 6 Assessment of measuring results The test results are the error curves (K-factors at different flow rates Q j ) at a common reference temperature of 15 C. After correction of the K-factor under actual conditions to the K-factor at 15 C (see point 7 - Calculation of correction of viscosity) a least-square-fitting was performed (see point 8 - Fitting of error curve). 7 Calculation of correction of viscosity Method from Calibration Intercomparison on Flowmeters for Kerosene Synthesis Report Part I & Part II 3476/1/0/03/9/9-BCR-S(30), chapter 14 For reasons of better comparability the measurement values are given at reference conditions. These reference conditions might be (e.g. for temperature) 15 C or 0 C, but the reference temperature can also be a quite unusual value (e.g. 17,3 C). In order to keep the measurement uncertainty small by referencing the actual measurement values (by interpolation or by extrapolation) to reference conditions, that value was taken as the reference value which was approximately a mean value or which was lying in the surrounding of most of the measured conditions.

6 So, in the case that the test runs are performed with liquids of different temperatures/viscosities, a K-factor at reference conditions (reference temperature, reference viscosity) can be found by the following algorithm: Step 1: Temperature correction to a reference temperature T 0 (T 0 = 15 C in the present case), where only the volume expansion of the meter shell is applied to the K-factor k(t) (α being the linear expansion coefficient of the meter material). So, for a cubic expansion of the meter shell of screw meters follows: k(t 0 ) = k(t). (1 + 3α. (T T 0 ))= = k(t). (1 + 0, (T T 0 )) Step : Viscosity correction for the error curves which have been adjusted acc. to step 1. When establishing the error curves for different liquids (by taking the temperaturecorrected error curves of step 1), there will be a mutual shift for reasons of the different viscosities. In order to be able to examine the dependency of the K-factors on the viscosity ν, and as a consequence to transfer the actual K-factors into K-factors at a reference viscosity ν r, the K-factors k are plotted separately for each flow rate Q i against viscosity ν. So for every Q j one yields a curve as a relation between k and ν. Based on former measurements, the following simple relations between the K-factor and the viscosity (Tab. ) have been achieved for the screw meters Viskosity ν (mm /s) 0,5 1,5 1,5,5,5 6 Relative slope ( K/1 mm /s) / K m.100 (%) + 0,19 + 0,05-0,01 Tab. Dependence of K-factor on viscosity These relative slopes K/ ν represent a mean value of all four meters. Deviations from the mean are allowed for in the calculation of the measurement uncertainty. In this step the K-factors are referenced to that viscosity which would appear at the reference temperature (in the present case of the second stage measurements 15 C). So the K-factors are valid for that liquid which would have the indicated viscosity at 15 C. 8 Fitting of error curve Method of least square fitting. Prior to the fitting, the error curve was referenced to an average viscosity (see 7 - Calculation of correction of viscosity). Errors caused by the fitting are allowed for in the calculation of uncertainties. As an assumption for the error curve the following polynomial was used: K(Q) = a 0 + a 1. Q + a.q + a 3.Q -1 + a 4.Q -

7 9 Measurement uncertainties MMs calibrated by standard capacity tanks of BEV-test rig (see also CMC-tables of BEV): ( 0, ,06 ) 0,1% U ( k = ) = u + er = ctestrig u cmastermet Because the measurements of the comparison had been performed in the same way as the measurements being the basis of the CMC-tables, the value U < 0,1% is still valid. Nevertheless, only a smaller number of repeated measurements could be performed during the comparison. But it can be shown by former measurements that the repeatability of the meters is extremely good and has little influence on the magnitude of the overall uncertainty. MMs calibrated by CMI-BCP The overall uncertainty U (k=) of the BCP is 0,05 % (see calibration certificate based upon CMC-database CZ8). This value takes into consideration the different ambient conditions under which a mobile prover is used. Under the assumption that the uncertainty of the screw meter being tested by the BCP is approximately the same as by the standard capacity tanks (similar test conditions like pressure, temperature, viscosity, no pressure variations, flying test method, no gaseous formations) the uncertainty of the meter can still be assessed by the same value as above (u cmastermeter = 0,06 %). So, the overall uncertainty U (k=) reads: U ( k ( u cbcp + u ) = ( 0,05 + 0,06 ) 0,08% = ) = cmaster The long term stability (expressed as the change of the K-factor per year K/a) is allowed for in the measurement uncertainty of the meter ( K/a < 0,01 for DN40 and < 0,005 for DN100). 10 Summary of measurement results Remark: The presented measurement results represent only the second stage. The meters are stable enough so that long term variations do not influence the results within the range of the given measurement uncertainty. Installation effects caused by swirls and irregular flow profiles at the meter inlet do not influence the results, because the meters are of positive displacement type, for which the flow profiles at the inlet and outlet side of the meter are irrelevant. Presentation of results 10.1 Presentation by graphical format: The measurement results are given as a fitted curve for each meter together with their uncertainty (see 4 figures below).

8 B1060V/DN100 =>fliegend=>bev.cmi, Juli 004, Benzin (gasoline) K-Faktor 16,840 16,835 16,830 16,85 16,80 16,815 16,810 16,805 16,800 16,795 16,790 16,785 16,780 16,775 16,770 16,765 16,760 16,755 0,1% En= 0, Durchflussstärke BEV CMI B1065V/DN40 =>fliegend=>bev-cmi, Juli 004, Benzin (gasoline) K-Faktor 71,00 71,180 71,160 71,140 71,10 71,100 71,080 71,060 71,040 71,00 71,000 70,980 70,960 70,940 70,90 70,900 70,880 70,860 70,840 0,1% En= 0, Durchflussstärke BEV CMI

9 B681V/DN100 =>fliegend => BEV-CMI, Juli 004, Diesel K-Faktor 16,840 16,835 16,830 16,85 16,80 16,815 16,810 16,805 16,800 16,795 16,790 16,785 16,780 16,775 16,770 16,765 16,760 16,755 0,1% En= 0, Durchflussstärke BEV CMI B1064V/DN40 =>fliegend=>bev-cmi, Juli 004, Diesel K-Faktor 71,00 71,180 71,160 71,140 71,10 71,100 71,080 71,060 71,040 71,00 71,000 70,980 70,960 70,940 70,90 70,900 70,880 70,860 70,840 0,1% En= 0, Durchflussstärke BEV CMI

10 10. Presentation according to ISO 575 (repeatability under alternated measurement conditions) In order to make the measurement results more comparable each curve is expressed by a representative K-factor, together with the repeatability and reproducibility. The differences between the error curves and between the K-factors represent the differences between the measuring results, which would be provided to a customer by using the one or the other test rig. The reproducibility of the representative K-factors is defined in accordance with ISO 575 and gives the repeatability under alternated measurement conditions. Therefore the reproducibility is the measure to assess the measurements of this comparison. The calculation of the reproducibility is based upon the following definitions:

11 Indices used by the definitions: I J XXX k i (Q j ) k s j j To indicate a single measurement within a series at one flowrate To distinguish a selected flowrate To identify the lab (in the present case CMI, BEV) K-factor for a single measurement i at the flowrate Q j Mean value for a series at flowrate j ki ( Q j ) k j =, 1 < i < n n n number of repeated measurements at Q j remark: n 10 (BEV), n 5 (CMI) Variance of k j from repeated measurements at flowrate Q j s j = 1/(n-1). Σ (k i - k j ) Units: (Imp/l) k(r) Representative K-factor for the present error curve 1 Q j. k j k( r) = m, 1 Q j m 1 < j < m Units: (Imp/l) m...number of selected flowrates Q j over the given flow range k(r) XXX Representative K-factor for lab XXX < k(r)> (Global) mean value of the representative K-factors of all p labs (p = in our case) < k(r)> = 1/p. Σ k(r) XXX = ½.(k(r) CMI + k(r) BEV ) Units: (Imp/l) remark: (k(r) CMI mean value from CMI Protocol, reduced to 15 C s Average calibration variance of participant (CMI / BEV), i.e. mean value for all m CMI flowrates Q s j BEV s XXXj s XXX =, 1 < j < m Units: (Imp/l) m Global mean calibration variance, i.e. average over the variances from CMI and s BEV r ec s C R ec s = 1/.( s CMI + s BEV ) Units: (Imp/l) remark : only s BEV was calculated and then set equal to CMI because the repeatability of the meters is not changed by the different standards. (inter)comparison repeatability (stating the statistical range of variances at different labs (the possible variance of any test series will lie within r ec with a 95 %-probability) 95 n rec = t 1 s Units: (Imp/l) Inter-laboratory variance, i.e. from the inter-comparison of representative K- factors between labs 1 sc = ( k( r) XXX < k( r) > ) Units: (Imp/l) 1 Reproducibility for the total (inter)comparison defining the probable maximum difference between any two laboratory concerning the representative K-factor (states the probable maximum difference of the representative K-factor between any two labs) R 95 ec tn sc + = 1 s Units: (Imp/l) s

12 With the above definitions the reproducibility R ec for the comparison (states the probable max. difference of the representative K-factors between CMI and BEV) is: Master Meter s B1060V k=16,8 DN100 gasoline B1065V k=71,0 DN40 gasoline B681V k=16,8 DN100 diesel oil B1064V k=71,0 DN40 diesel oil s C 95 = tn rec 1 s ReC = tn 1 sc + s (Imp/l) (Imp/l) t=3,18 for (n-1)=4 t=3,18 for (n-1)=4 Imp/l Imp/l 0, , ,01 0,017 0, , ,09 0,057 0, , ,003 0,005 0, , ,059 0,06 95

13 10.3 Presentation by the E n -criterion (according to EAL-P7 "EAL Interlaboratory Comparisons") As a criterion on the compatibility of two different measurements (two test rigs) of the same measurand (KRAL master meters) not only the results of measurements but also the associated uncertainties must be taken into account. The "normalized error" E n serves as a compatibility criterion with (for both labs A and B) the same fixed expansion factor k (valid for not correlated measurements as it is in the present case: two independent test rigs): x A xb 1 x A xb 1 d En = = = k U + U u + u k u( d) A B A B x A, x B results of measurement (in the present case representative K-factor k(r), see 10.) U A, U B associated expanded uncertainty, no correlation between x A and x B, x A and x B normally distributed The smaller u(d) the more reliable is the estimate d (of the true difference D = X A - X B ); but for compatibility reasons, d must also be close to zero. For k = the criterion: E 1 n is often proposed. Master Meter k(r) BEV (Imp/l) k(r) CMI (Imp/l) U BEV U CMI E n B1060V k=16,8 DN100 gasoline 16,799 16, ,10 % + 0,0168 Imp/l + 0,08 % + 0,0134 Imp/l 0,3 B1065V k=71,0 DN40 Gasolina B681V k=16,8 DN100 diesel oil B1064V k=71,0 DN40 diesel oil 71,08 71, ,10 % + 0,071 Imp/l 16,87 16,85 + 0,10 % + 0,0168 Imp/l 71,15 71, ,10 % + 0,071 Imp/l + 0,08 % + 0,057 Imp/l + 0,08 % + 0,0134 Imp/l + 0,08 % + 0,057 Imp/l 0,4 0,08 0,09

14:1400 (0206) KRAL Volumeter - OMG

14:1400 (0206) KRAL Volumeter - OMG 14:1400 (0206) KRAL Volumeter - OMG OMG - Universal OMG The flowmeter for a wide range of applications Operating Conditions and Materials Flow Range: 0.1 to 7,500 l/min Max. Pressure: 250 bar Temperature

More information

INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) BIPM Pavillon de Breteuil F Sevres France

INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) BIPM Pavillon de Breteuil F Sevres France INTERNATIONAL KEY COMPARISON OF LIQUID HYDROCARBON FLOW FACILITIES CCM-FF-K2 (Final Report) A Report for BIPM Pavillon de Breteuil F-92312 Sevres France Project No: FFIN56 October 2008 Report No: XXXX/XXX

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8426 Second edition 2008-02-01 Hydraulic fluid power Positive displacement pumps and motors Determination of derived capacity Transmissions hydrauliques Pompes et moteurs volumétriques

More information

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook Reference Manual 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com Reference Manual 405 and 1595 405 Compact Orifice Series and 1595 Conditioning

More information

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com 405 and 1595 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test

More information

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book NOTICE Read

More information

ALASTAIR MCLACHLAN Applications Cameron

ALASTAIR MCLACHLAN Applications Cameron ALASTAIR MCLACHLAN Applications Cameron Have worked in the oil and gas flow measurement industry for 27 years primarily in custody transfer measurement using ultrasonic meters. Joined Cameron Caldon Ultrasonics

More information

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement

More information

SITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181

SITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181 Siemens AG 09 SITRANS F flowmeters System information Overview Siemens offers two types of ultrasonic flowmeters, in-line flowmeters and clamp-on flowmeters. This offers the end user the maximum flexibility

More information

Flow Management Devices, LLC An ISO 9000:2008 Certified Company

Flow Management Devices, LLC An ISO 9000:2008 Certified Company Flow Management Devices, LLC An ISO 9000:2008 Certified Company Liquid Flow Prover US Patent # 7650775 UNCERTAINTY ANALYSIS Model FMD-060 For the proving of volumetric pulse output flow meters Flow Management

More information

Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level

Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level Overview Function s are for measuring the liquid levels in wells, tanks, channels and dams. Sensor Diaphragm p U const. U +2 (brown) I EM I - (blue) 0, U B s are submersible sensors for hydrostatic level

More information

Euramet project 1075 Bilateral comparison

Euramet project 1075 Bilateral comparison Bilateral comparison Between the gas flow standards of the national Metrology Institutes of Bulgaria (BIM-NCM) and The Netherlands (VSL) Copyright VSL B.V. Dutch Metrology Institute Page 2 of 19 VSL Thijsseweg

More information

INTERNATIONAL OIML D 25 DOCUMENT. Vortex meters used in measuring systems for fluids ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

INTERNATIONAL OIML D 25 DOCUMENT. Vortex meters used in measuring systems for fluids ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL OIML D 25 DOCUMENT Edition 1996 (E) Vortex meters used in measuring systems for fluids Compteurs à vortex utilisés dans les ensembles de mesurage de fluides OIML D 25 Edition 1996 (E) ORGANISATION

More information

Calibration Models and Uncertainty Analysis of the Euroloop Liquid Hydrocarbon Flow Facility

Calibration Models and Uncertainty Analysis of the Euroloop Liquid Hydrocarbon Flow Facility Calibration Models and Uncertainty Analysis of the Euroloop Liquid Hydrocarbon Flow Facility Jos van der Grinten 1 and Bart van der Stap 2 1 NMi EuroLoop, Petroleumweg 36, 3196 KD Rotterdam-Vondelingenplaat,

More information

NEW VSL Conditioning plate!

NEW VSL Conditioning plate! Calibration of Prover Tanks using a Coriolis Mass Flow Meter as the Master Meter. Is this possible for Pipe Prover calibration too? Erik Smits (fsmits@vsl.nl) 2 nd European Flow Measurement Workshop Sintra,

More information

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG Introduction CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG A mass flowmeter is a system that provides a measurement of

More information

Pressure Measurement. Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level 2/27

Pressure Measurement. Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level 2/27 Overview Siemens AG 011 Function s are for measuring the liquid levels in wells, tanks, channels and dams. Diaphragm p Sensor U const. U + (brown) I EM I -1 (blue) 0, U B s are submersible sensors for

More information

PRIMARY STANDARD AND TRACEABILITY CHAIN FOR MICROFLOW OF LIQUIDS

PRIMARY STANDARD AND TRACEABILITY CHAIN FOR MICROFLOW OF LIQUIDS PRIMARY STANDARD AND TRACEABILITY CHAIN FOR MICROFLOW OF LIQUIDS Miroslava Benkova 1 and Ivan Mikulecky 1 Czech Institute of Metrology, Okruzni 31, Brno, Czech Republic Slovak Institute of Metrology, Karloveska

More information

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers The Uncertainty of a Waterdraw Calibration vs. Gravimetric Written by: Gary Cohrs Introduction The focus of the API Petroleum Measurement Manuals is to reduce the uncertainty and improve the accuracy of

More information

Standard Practices for Air Speed Calibration Testing

Standard Practices for Air Speed Calibration Testing Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument

More information

Flow Measurement. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/22 Siemens FI

Flow Measurement. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/22 Siemens FI Function All are based on Faraday s law of induction: U M = B v d k U M = Measured voltage induced in the medium perpendicular to the magnetic field and the flow direction. The voltage is tapped at two

More information

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components: SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine

More information

E 094 E 103 E 143. Tank top mounting Connection up to G1 Nominal flow rate up to 135 l/min e

E 094 E 103 E 143. Tank top mounting Connection up to G1 Nominal flow rate up to 135 l/min e R e t u r n F i l t e rs E 094 E 103 E 143 Tank top mounting Connection up to G1 Nominal flow rate up to 135 l/min 20.30-5e D e s c r i p t i o n Application In the return line circuits of hydraulic systems.

More information

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty October 21, 2013 Summary The Working Group for Fluid Flow (WGFF) defines Calibration and Measurement Capability (CMC) uncertainty

More information

Applied Fluid Mechanics

Applied Fluid Mechanics Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and

More information

Cooler-Bypass Thermostat Valve, Size 10

Cooler-Bypass Thermostat Valve, Size 10 Cooler-Bypass Thermostat Valve, Size Q max = l/min, p max = bar Temperature-controlled, integral relief function Description These direct acting bypass thermostat valves are size, temperature controlled,

More information

TankExampleNov2016. Table of contents. Layout

TankExampleNov2016. Table of contents. Layout Table of contents Task... 2 Calculation of heat loss of storage tanks... 3 Properties ambient air Properties of air... 7 Heat transfer outside, roof Heat transfer in flow past a plane wall... 8 Properties

More information

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS Gianalberto Grego, Fabio F. Muciaccia W.E.S.T. Srl, Milan, Italy west-hydro@libero.it ABSTRACT The

More information

Installation guide 862 MIT / MIR

Installation guide 862 MIT / MIR Installation guide 862 MIT / MIR in combination with: 864 MTT or 863 MRT or (dual) spot element November 2005 Part no. 4416.232_Rev3 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100

More information

The new Closed Loop pigsar and its projected CMC. Jos van der Grinten, Detlef Vieth, Bodo Mickan

The new Closed Loop pigsar and its projected CMC. Jos van der Grinten, Detlef Vieth, Bodo Mickan The new Closed Loop pigsar and its projected CMC Jos van der Grinten, Detlef Vieth, Bodo Mickan Introduction pigsar Calibration facilities Closed Loop pigsar (CLP) Calibration model Uncertainty analysis

More information

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION R. Kumme

More information

Differential Mobility Particle Sizer (Aerosol measurements)

Differential Mobility Particle Sizer (Aerosol measurements) Institute for Atmospheric and Climate Science - IACETH Atmospheric Physics Lab Work Differential Mobility Particle Sizer (Aerosol measurements) Abstract A differential mobility particle sizer (DMPS) is

More information

MCE380: Measurements and Instrumentation Lab

MCE380: Measurements and Instrumentation Lab MCE380: Measurements and Instrumentation Lab Chapter 8: Flow Measurements Topics: Basic Flow Equations Flow Obstruction Meters Positive Displacement Flowmeters Other Methods Holman, Ch. 7 Cleveland State

More information

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection

More information

Chemical Engineering 3P04 Process Control Tutorial # 1 Learning goals

Chemical Engineering 3P04 Process Control Tutorial # 1 Learning goals Chemical Engineering 3P04 Process Control Tutorial # 1 Learning goals 1. Sensor Principles with the flow sensor example 2. The typical manipulated variable: flow through a conduit Sensors: We need them

More information

CALIBRATION CERTIFICATE # 503

CALIBRATION CERTIFICATE # 503 CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion

More information

Experiment (4): Flow measurement

Experiment (4): Flow measurement Experiment (4): Flow measurement Introduction: The flow measuring apparatus is used to familiarize the students with typical methods of flow measurement of an incompressible fluid and, at the same time

More information

UNCERTAINTIES FOR PRESSURE-TIME EFFICIENCY MEASUREMENTS

UNCERTAINTIES FOR PRESSURE-TIME EFFICIENCY MEASUREMENTS UNCERTAINTIES FOR PRESSURE-TIME EFFICIENCY MEASUREMENTS ABSTRACT Jørgen Ramdal Norwegian University of Science and Technology Pontus P. Jonsson Luleå University of Technology Ole Gunnar Dahlhaug Norwegian

More information

TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 2017

TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 2017 TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 017 111 Wilson Blvd, Suite 500 Arlington, Virginia 01 USA +001 (703) 54-8800 Published by: TABLE OF CONTENTS SECTION

More information

Turbine flowmeter and viscosity effects of liquid hydrocarbons

Turbine flowmeter and viscosity effects of liquid hydrocarbons Turbine flowmeter and viscosity effects of liquid hydrocarbons Valmir Ruiz, Marcos Tadeu Pereira,2, Nilson Massami Taira Institute for Technological Research IPT, 2 Universidade de São Paulo Avenida Prof.

More information

Direct-operated 2/2-way solenoid valves Type EV210B

Direct-operated 2/2-way solenoid valves Type EV210B Data sheet Direct-operated 2/2-way solenoid valves Type EV20B EV20B covers a wide range of direct-operated 2/2-way solenoid valves for universal use. EV20B is a very robust valve program with high performance

More information

Laboratory work No 2: Calibration of Orifice Flow Meter

Laboratory work No 2: Calibration of Orifice Flow Meter Laboratory work No : Calibration of Orifice Flow Meter 1. Objective Calibrate the orifice flow meter and draw the graphs p = f 1 (Q) and C d = f (Re ).. Necessary equipment 1. Orifice flow meter. Measuring

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

Cartridge Valves Technical Information. Quick reference

Cartridge Valves Technical Information. Quick reference Quick reference Contents CPF0... 8. CP600-5... 8. CP60-5... 8. 066 handle kit... 8.5 MP06... 8.6 MP... 8.6 X05-FD0 Traction manifold... 8.9 X05-FD6 Traction manifold... 8.0 X05-FD0 Traction manifold...

More information

KENYA BUREAU OF STANDARDS (KEBS)

KENYA BUREAU OF STANDARDS (KEBS) CD13/SERV/ 2018 DRAFT KENYA STANDARD Dial thermometers - Specification Public Review Draft March 2018 KENYA BUREAU OF STANDARDS (KEBS) WORKING GROUP REPRESENTATION The following organizations were represented

More information

CVE 372 HYDROMECHANICS EXERCISE PROBLEMS

CVE 372 HYDROMECHANICS EXERCISE PROBLEMS VE 37 HYDROMEHNIS EXERISE PROLEMS 1. pump that has the characteristic curve shown in the accompanying graph is to be installed in the system shown. What will be the discharge of water in the system? Take

More information

Variable Displacement Plug-In Motor A6VE. Series 6, for open and closed circuits Axial tapered piston, bent axis design

Variable Displacement Plug-In Motor A6VE. Series 6, for open and closed circuits Axial tapered piston, bent axis design rueninghaus Hydromatik Series 6, for open and closed circuits xial tapered piston, bent axis design Size 28...20 Nom. pressure up to 400 bar Peak pressure up to 40 bar RE 9606/0.97 RE 9606/0.97 replaces

More information

Linearity of Coriolis Flowmeters at Full Scale Value. Dr. Jesús Aguilera - Endress + Hauser Paul Ceglia - Endress + Hauser

Linearity of Coriolis Flowmeters at Full Scale Value. Dr. Jesús Aguilera - Endress + Hauser Paul Ceglia - Endress + Hauser Linearity of Coriolis Flowmeters at Full Scale Value Dr. Jesús Aguilera - Endress + Hauser Paul Ceglia - Endress + Hauser Products Solutions Services Linearity of Coriolis flowmeters at full scale value

More information

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560 FLOW MEASUREMENT INC 102 Fundamental of Instrumentation and Process Control 2/2560 TABLE OF CONTENTS A. INTRODUCTION B. LOCAL FLOW MEASUREMENT B.1 Particle Image Velocimetry (PIV) B.2 Laser doppler anemometry

More information

Application of Density Meters to Liquid Hydrocarbon Measurement

Application of Density Meters to Liquid Hydrocarbon Measurement 02/2018 Application of Density Meters to Liquid Hydrocarbon Measurement Written by Dean Minehart, Micro Motion, Inc. www.emerson.com/micromotion 1 Introduction The American Petroleum Institute (API) Manual

More information

Final Results of Bilateral Comparison between NIST and PTB for Flows of High Pressure Natural Gas

Final Results of Bilateral Comparison between NIST and PTB for Flows of High Pressure Natural Gas Final Results of Bilateral Comparison between NIST and for Flows of High Pressure Natural Gas B. Mickan (, Germany, Bodo.Mickan@ptb.de) H. Toebben (pigsar TM, Germany) A. Johnson (NIST,USA, Aaron.Johnnson@nist.gov)

More information

Safety valve with unit approval type MV..X.., SV..X..

Safety valve with unit approval type MV..X.., SV..X.. Safety valve with unit approval type MV..X.., SV..X.. Operating pressure p max = 0 bar Flow Q max = 0 lpm Safety valve type CMVX 2.. with unit approval see D 77 TÜV These safety valves are intended for

More information

Applied Fluid Mechanics

Applied Fluid Mechanics Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and

More information

Applied Fluid Mechanics

Applied Fluid Mechanics Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and

More information

Review of Anemometer Calibration Standards

Review of Anemometer Calibration Standards Review of Anemometer Calibration Standards Rachael V. Coquilla rvcoquilla@otechwind.com Otech Engineering, Inc., Davis, CA Anemometer calibration defines a relationship between the measured signals from

More information

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions Products Solutions Services Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions Dr. Alfred Rieder, Slide 1 Overview Introduction Process meter Features Working

More information

Installation effects on an ultrasonic flow meter with implications for self diagnostics

Installation effects on an ultrasonic flow meter with implications for self diagnostics Flow Measurement and Instrumentation 11 (2000) 109 122 www.elsevier.com/locate/flowmeasinst Installation effects on an ultrasonic flow meter with implications for self diagnostics Carl Carlander *, Jerker

More information

Extending Transducer Calibration Range by Extrapolation

Extending Transducer Calibration Range by Extrapolation Introduction Extending Transducer Calibration by olation by LaVar Clegg Force and torque transducers must be calibrated in a laboratory in order to be useful in their intended application. Applications

More information

Corrections related to operation and calibration of liquid flow meters NFOGM Temadag 2016 Morten Marstein, FMC Kongsberg Metering

Corrections related to operation and calibration of liquid flow meters NFOGM Temadag 2016 Morten Marstein, FMC Kongsberg Metering Corrections related to operation and calibration of liquid flow meters NFOGM Temadag 2016 Morten Marstein, FMC Kongsberg Metering 29.02.2016 FMC Kongsberg Metering General formulas Volumetric flow meter

More information

Cone differential pressure flow meter. Operating manual WIDE PLUS PRECISION INSTRUMENTS CO., LTD.

Cone differential pressure flow meter. Operating manual WIDE PLUS PRECISION INSTRUMENTS CO., LTD. Cone differential pressure flow meter Operating manual WIDE PLUS PRECISION INSTRUMENTS CO., LTD. INDEX Introduction 2 Principle Application The characteristics of flow meter Specification Technical features

More information

UNI EN ISO 9001:2008 UNI EN ISO 14001:2004 BS OHSAS 18001:2007 EC DIRECTIVE 2014/34/EU (ATEX) CERTIFIED MANAGEMENT SYSTEM

UNI EN ISO 9001:2008 UNI EN ISO 14001:2004 BS OHSAS 18001:2007 EC DIRECTIVE 2014/34/EU (ATEX) CERTIFIED MANAGEMENT SYSTEM EN ISM-BSM CERTIFIED Technology Made in Italy Since 1955 the Varvel Group has been making speed reducers and variators for light industry applications. Reliable partner in power transmission equipment

More information

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT 62 CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT 5.1 INTRODUCTION The primary objective of this work is to investigate the convective heat transfer characteristics of silver/water nanofluid. In order

More information

Thermodynamics. Joule-Thomson effect Ideal and Real Gases. What you need: Complete Equipment Set, Manual on CD-ROM included

Thermodynamics. Joule-Thomson effect Ideal and Real Gases. What you need: Complete Equipment Set, Manual on CD-ROM included Ideal Real Gases Thermodynamics 30600 What you can learn about Real gas Intrinsic energy Gay-Lussac theory Throttling Van der Waals equation Van der Waals force Inverse Inversion temperature Principle:

More information

THREADED CLOSURES. For access to Filter Vessels, Strainers, Pig Launchers & Receivers. Tube Turns Products

THREADED CLOSURES. For access to Filter Vessels, Strainers, Pig Launchers & Receivers. Tube Turns Products THREADED CLOSURES For access to Filter Vessels, Strainers, Pig Launchers & Receivers Tube Turns Products HISTORY Tube Turns was founded in 927 in Louisville, Kentucky, USA. We were the f i r s t A m e

More information

VOLUMEC. Valve Position Indicator 5

VOLUMEC. Valve Position Indicator 5 VOLUMEC Valve Position Indicator 5 2 KRACHT CORP. 8600 S Wilkinson Way Unit A Perrysburg, OH 43551 USA P +1 419 874 1000 F +1 419 874 1006 flowmeters@krachtcorp.com www.krachtcorp.com VOLUMEC Valve Position

More information

Volume flow rate sensors

Volume flow rate sensors Volume flow rate sensors Kalibrierdienst K - 00-0 DIN EN 00 DIN EN ISO 900 00 79 Test with confidence Messen mit System English Hydrotechnik gear flow meters type GFM Gear flow meters are used for the

More information

EURAMET Project no Inter-comparison of a 1000 L proving tank

EURAMET Project no Inter-comparison of a 1000 L proving tank EURAMET Project no. 1157 Inter-comparison of a 1000 L proving tank Final Report IPQ (Elsa Batista) Coordinator of the comparison VSL (Erik Smits) Pilot and coordinator of the comparison September 2013

More information

Two mark questions and answers UNIT I BASIC CONCEPT AND FIRST LAW SVCET

Two mark questions and answers UNIT I BASIC CONCEPT AND FIRST LAW SVCET Two mark questions and answers UNIT I BASIC CONCEPT AND FIRST LAW 1. What do you understand by pure substance? A pure substance is defined as one that is homogeneous and invariable in chemical composition

More information

Check-Q-meter. Table of contents. Features. Functions. RE 27551/ /10 Replaces: Type FD

Check-Q-meter. Table of contents. Features. Functions. RE 27551/ /10 Replaces: Type FD Check-Q-meter RE /0.0 /0 Replaces: 09.9 Type FD Nominal size... Series ax. Operating pressure 0 bar ax. Flow 0 l/min K9/ Table of contents Contents Page Features Functions Ordering details Symbols Functional

More information

High pressure comparisons between seven European National Laboratories

High pressure comparisons between seven European National Laboratories J4/2007 High pressure comparisons between seven European National Laboratories Range 50 MPa to 500 MPa Report on EUROMET Project 881 Markku Rantanen, Sari Semenoja, Måns Ackerholm, Antoine Condereys, Zdenek

More information

close to performances of the turbine gas meters designed for custody transfer measurements.

close to performances of the turbine gas meters designed for custody transfer measurements. QUANTOMETER DESIGN AND FUNCTION CPT Quantometers have been designed in order to provide our customers with reliable and inexpensive measuring instruments for secondary flow measurements. Our great experience

More information

CHAPTER THREE FLUID MECHANICS

CHAPTER THREE FLUID MECHANICS CHAPTER THREE FLUID MECHANICS 3.1. Measurement of Pressure Drop for Flow through Different Geometries 3.. Determination of Operating Characteristics of a Centrifugal Pump 3.3. Energy Losses in Pipes under

More information

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH Submitted By: ABDULLAH IBN ABDULRAHMAN ID: 13456789 GROUP A EXPERIMENT PERFORMED

More information

Comparison between Numerical and Experimental for UVP Measurement in Double Bent Pipe with Out-of-Plane Angle

Comparison between Numerical and Experimental for UVP Measurement in Double Bent Pipe with Out-of-Plane Angle Journal of Flow Control, Measurement & Visualization, 24, 2, 54-64 Published Online October 24 in SciRes. http://www.scirp.org/journal/jfcmv http://dx.doi.org/.4236/jfcmv.24.247 Comparison between Numerical

More information

Reprinted from February Hydrocarbon

Reprinted from February Hydrocarbon February2012 When speed matters Loek van Eijck, Yokogawa, The Netherlands, questions whether rapid analysis of gases and liquids can be better achieved through use of a gas chromatograph or near infrared

More information

SonTek IQ - Principles of Operation

SonTek IQ - Principles of Operation SonTek IQ - Principles of Operation SonTek IQ - Principles of Operation This doument introduces the operating principles of the SonTek IQ acoustic Doppler current meter. It does not attempt to provide

More information

FLUID MECHANICS D203 SAE SOLUTIONS TUTORIAL 2 APPLICATIONS OF BERNOULLI SELF ASSESSMENT EXERCISE 1

FLUID MECHANICS D203 SAE SOLUTIONS TUTORIAL 2 APPLICATIONS OF BERNOULLI SELF ASSESSMENT EXERCISE 1 FLUID MECHANICS D203 SAE SOLUTIONS TUTORIAL 2 APPLICATIONS OF BERNOULLI SELF ASSESSMENT EXERCISE 1 1. A pipe 100 mm bore diameter carries oil of density 900 kg/m3 at a rate of 4 kg/s. The pipe reduces

More information

Comparison of Noise Test Codes when Applied to air Compressors

Comparison of Noise Test Codes when Applied to air Compressors VDI-Berichte Nr. 1932, 2006 A 6 83 Comparison of Noise Test Codes when Applied to air Compressors Michael J. Lucas, INCE Bd. Cert., Ingersoll-Rand Company, Davidson/NC INTRODUCTION Beginning January 2004,

More information

Made in Italy HOSE REELS CATALOGUE TECHNOLOGY FOR FLUIDYNAMICS. No. E14/A1

Made in Italy HOSE REELS CATALOGUE TECHNOLOGY FOR FLUIDYNAMICS. No. E14/A1 Made in Italy HOSE REELS CATALOGUE TECHNOLOGY FOR FLUIDYNAMICS No. E14/A1 The complete range of our products is presented in the specific catalogues: HOSE REELS CATALOGUE No. E14/A2 HOSE REELS CATALOGUE

More information

High pressure comparison among seven European national laboratories

High pressure comparison among seven European national laboratories IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. High pressure comparison among seven European national

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

Integral V-Cone Flowmeter VD series

Integral V-Cone Flowmeter VD series TRANSMITTER DIRECTLY MOUNTED! SIMPLE AND EASY INSTALLATION Integral V-Cone Flowmeter VD series OUTLINE VD series Integral V-Cone Flowmeter is a differential pressure type flowmeter. Unlikely to existing

More information

Suction fi lters. MPA - MPM series. Flow rate up to 875 l/min. Suction fi lters

Suction fi lters. MPA - MPM series. Flow rate up to 875 l/min. Suction fi lters Suction fi lters & MPA MPM series Flow rate up to 87 l/min 7 Suction fi lters FILTER SIZING The correct filter sizing have to be based on the variable pressure drop depending by the application. For example,

More information

Process Control, 3P4 Assignment 6

Process Control, 3P4 Assignment 6 Process Control, 3P4 Assignment 6 Kevin Dunn, kevin.dunn@mcmaster.ca Due date: 28 March 204 This assignment gives you practice with cascade control and feedforward control. Question [0 = 6 + 4] The outlet

More information

Certificate: / 9 September 2014

Certificate: / 9 September 2014 Test report: 936/21219977/A of 26 March 214 Initial certification: 5 August 214 Date of expiry: 4 August 219 Publication: BAnz AT 5 August 214 B11, chapter III, no. 2.1 Approved application The certified

More information

FLOW SENSOR FOR LIQUID MEDIA Serie 210

FLOW SENSOR FOR LIQUID MEDIA Serie 210 FLOW SENSOR FOR LIQUID MEDIA Serie 0 Flow range 0.... 0 l/min Nominal diameters DN 6 / 8 / 0 / / 0 / Temperature measurement -0... + ÀC In comparison to the OEM flow sensor (type 00), the type 0 is available

More information

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere Acta Polytechnica Vol. 52 No. 3/202 Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere Petr Kracík,JiříPospíšil, Ladislav Šnajdárek Brno University of

More information

CENTRIFUGAL PUMP SELECTION, SIZING, AND INTERPRETATION OF PERFORMANCE CURVES

CENTRIFUGAL PUMP SELECTION, SIZING, AND INTERPRETATION OF PERFORMANCE CURVES CENTRIFUGAL PUMP SELECTION, SIZING, AND INTERPRETATION OF PERFORMANCE CURVES 4.0 PUMP CLASSES Pumps may be classified in two general types, dynamic and positive displacement. Positive displacement pumps

More information

FLUID MECHANICS PROF. DR. METİN GÜNER COMPILER

FLUID MECHANICS PROF. DR. METİN GÜNER COMPILER FLUID MECHANICS PROF. DR. METİN GÜNER COMPILER ANKARA UNIVERSITY FACULTY OF AGRICULTURE DEPARTMENT OF AGRICULTURAL MACHINERY AND TECHNOLOGIES ENGINEERING 1 5. FLOW IN PIPES Liquid or gas flow through pipes

More information

1/54 Circulation pump, safety valve, expansion vessel

1/54 Circulation pump, safety valve, expansion vessel 1/54 Circulation pump, safety valve, expansion vessel pressure loss efficiency of pump secured heat output safety valve sizing expansion vessel sizing Circulation pump 2/54 similar principle as for heating

More information

Suction fi lters. SF series. Flow rate up to 160 l/min. Suction fi lters

Suction fi lters. SF series. Flow rate up to 160 l/min. Suction fi lters Suction fi lters SF 00 series Flow rate up to 60 l/min Suction fi lters FILTER SIZING The correct filter sizing have to be based on the variable pressure drop depending by the application. For example,

More information

ISO 9906 INTERNATIONAL STANDARD. Rotodynamic pumps Hydraulic performance acceptance tests Grades 1 and 2

ISO 9906 INTERNATIONAL STANDARD. Rotodynamic pumps Hydraulic performance acceptance tests Grades 1 and 2 INTERNATIONAL STANDARD ISO 9906 First edition 1999-1-15 Rotodynamic pumps Hydraulic performance acceptance tests Grades 1 and Pompes rotodynamiques Essais de fonctionnement hydraulique pour la réception

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed

More information

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers Intercomparison of Mobility Particle Size Spectrometers Project No.: MPSS-2015-1-1 Basic information: Location of the quality assurance: TROPOS, lab: 118 Delivery date: October 05, 2015 Setup in the laboratory:

More information

V-Cone Flowmeter DATASHEET

V-Cone Flowmeter DATASHEET V-Cone Flowmeter ATASHEET JUNHO 013 IntraCone Cone Flow Meter Type: ICM etail: Cone-Element Technical Information 05/011 FLOW THE SPECIALIST IN LEVEL AN FLOW Cone Flow Meter Technical Information Intra-Automation

More information

Full Metal Variable Area Flow Meter and Counter

Full Metal Variable Area Flow Meter and Counter Full Metal Variable rea Flow Meter and Counter for horizontal and vertical mounting measuring monitoring analysing Special versions up to 600 bar Measuring range: 10-100...00-000 L/h water 0.3-3.0...-0

More information

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688 Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211

More information

METHOD OF SUM OF POWER LOSSES AS A WAY FOR DETERMINING THE K I

METHOD OF SUM OF POWER LOSSES AS A WAY FOR DETERMINING THE K I POLISH MARITIME RESEARCH 2 (90) 2016 Vol 23; pp 57-63 101515/pomr-2016-0021 METHOD OF SUM OF POWER LOSSES AS A WAY FOR DETERMINING THE K I COEFFICIENTS OF ENERGY LOSSES IN HYDRAULIC MOTOR A Maczyszyn,

More information

THE PRESSURE-TIME MEASUREMENTS PROJECT AT LTU AND NTNU

THE PRESSURE-TIME MEASUREMENTS PROJECT AT LTU AND NTNU THE PRESSURE-TIME MEASUREMENTS PROJECT AT LTU AND NTNU ABSTRACT Pontus P. Jonsson, Luleå University of Technology Jørgen Ramdal, Norwegian University of Science and Technology Michel J. Cervantes, Luleå

More information

IBV Series C Carbon Steel Inverted Bucket Vertical Steam Trap

IBV Series C Carbon Steel Inverted Bucket Vertical Steam Trap Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without

More information