Methods for the determination of ethylene oxide

Size: px
Start display at page:

Download "Methods for the determination of ethylene oxide"

Transcription

1 Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, Sankt Augustin Expert Committee Chemistry Carcinogenic substances Order number: ZH 1/120.27E Established methods Issue: November 1994 Methods for the determination of ethylene oxide Methods tested and recommended by the Berufsgenossenschaften for the determination of ethylene oxide in working areas after discontinuous sampling. For the assessment of working areas, both personal or stationary sampling are possible: 1 Sampling with a pump and adsorption on activated carbon, head space gas chromatography or gas chromatography after desorption. Operating method A: Head space gas chromatography. Ethylene oxide-1-hsgc Operating method B: Gas chromatography after desorption. Ethylene oxide-1-gc (Issue: May 1985) 2 Sampling with a pump and adsorption on activated carbon, gas chromatography after desorption and derivatization with hydrobromic acid. Ethylene oxide-3-gc (Issue: January 1989, revised November 1994) 3 Sampling with a pump and adsorption on the polymeric resin Amberlite XAD-4, gas chromatography after thermal desorption. Ethylene oxide-3-gc (Issue: May 1993) IUPAC name: CAS No: oxirane, ethylene oxide AirMonitoringMethods,Vol.4

2 Analytical Methods AirMonitoringMethods,Vol Sampling with a pump and adsorption on activated carbon, head space gas chromatography or gas chromatography after desorption Principle: Technical data: With a pump a measured air volume is drawn through activated carbon. The adsorbed ethylene oxide is desorbed with a nonvolatile solvent and detected by head space gas chromatography (Operating method A [1,2]) or determined by gas chromatography after desorption with carbon disulfide (Operating method B [1]). Quantification limit: Operating method A: Under the conditions encountered in practice the relative quantification limit is 0.5 ml/m 3 (ppm) ^= 0.9 mg/m 3 ethylene oxide in 10 L air sample. This corresponds to ^= 9 µg ethylene oxide per each activated carbon tube. Selectivity: Advantages: Disadvantages: Apparatus: Operating method B: Under the conditions encountered in practice the relative quantification limit is 0.5 ml/m 3 (ppm) ^= 0.9 mg/m 3 ethylene oxide in 10 L air sample. The absolute quantification limit is 4.5 ng ethylene oxide. The values may be too high as a result of interfering substances. However, interferences can be eliminated by selecting a different column. Personal sampling and specific measurements are possible. No detection of peak concentrations, very time-consuming. Pump equipped with gas meter or flow meter, activated carbon tubes, head space chromatograph or gas chromatograph equipped with flame ionization detector (FID).

3 117 AirMonitoringMethods,Vol.4 Ethylene oxide 1 Summary This method permits the determination of ethylene oxide concentrations in working areas averaged over the sampling time after personal or stationary sampling. With a personal or a stationary pump a measured air volume is drawn through a glass tube filled with activated carbon. The ethylene oxide is then desorbed with a non-volatile solvent and analysed by head space gas chromatography (HSGC) (Operating method A) or by gas chromatography after desorption with carbon disulfide (Operating method B). Under the conditions encountered in practice the relative quantification limit is 0.5 ml/m 3 (ppm). This corresponds to 0.9 mg/m 3 ethylene oxide for operating method A and a 10 L air sample or to 9 µg ethylene oxide for each activated carbon tube. Under the conditions encountered in practice the relative quantification limit of operating method B is 0.5 ml/m 3 (ppm). This corresponds to 0.9 mg/m 3 ethylene for 10 L air sample. The absolute quantification limit is 4.5 ng ethylene oxide. Detailed description of the method Contents Operating method A: Head space gas chromatography A 2 Equipment, chemicals and solutions A 2.1 Equipment A 2.2 Chemicals A 2.3 Solutions A 3 Sampling and sample preparation A 4 Operating conditions for gas chromatography A 5 Analytical determination A 6 Evaluation A 6.1 Calibration A 6.2 Calculation of the analytical result A 7 Reliability of the method A 7.1 Accuracy A 7.2 Quantification limit A 7.3 Selectivity A 7.4 Recovery A 8 Discussion A 9 Manufacturers A 10 References

4 Analytical Methods AirMonitoringMethods,Vol A 2 Equipment, chemicals and solutions A 2.1 Equipment For sampling and preparation: Pump equipped with gas meter or flow meter Sample vial with polytetrafluoroethylene (PTFE) coated septum and aluminium cap Crimper Activated carbon tubes (standardized, consisting of two sections filled with about 700 and 400 mg activated carbon separated by a porous polymeric material) Caps for sealing the opened activated carbon tubes For analytical determination: Head space gas chromatograph equipped with flame ionization detector, automatic headspace sampler, if necessary with back-flush device Compensation recorder and/or integrator Gastight syringes, if sample charging is carried out manually, instead of using an autosampler A 2.2 Chemicals Ethylene oxide, purity >99% Dimethylphthalate (DMP) for GC Gases for gas chromatography: Nitrogen, hydrogen, purified air or synthetic air A 2.3 Solutions Ethylene oxide standard solution: Solution of about 9 mg ethylene oxide in 20 ml DMP. About 5 ml gaseous ethylene oxide is slowly injected with a gastight syringe through the septum into the liquid phase of a sample vial containing 20 ml DMP. At the tip of the needle ethylene oxide bubbles should not appear so that ethylene oxide does not escape when the needle is withdrawn through the septum. The weight of ethylene oxide is determined by weighing to the nearest ± 0.1 mg. Ethylene oxide calibration solution: Solution of about 4.5 mg ethylene oxide in 100 ml DMP. 10 ml of the ethylene oxide standard solution is transferred to a 100 ml volumetric flask containing 50 ml DMP and diluted to the mark with DMP (cf. Sect. A 8).

5 119 AirMonitoringMethods,Vol.4 Ethylene oxide A 3 Sampling and sample preparation An activated carbon tube is opened and connected to the pump. Pump and tube are either carried by a person during working hours or stationary sampling is performed. The air sample volume of 10 L per 1 g activated carbon must not be exceeded. The flow rate should not exceed 3 L/h. Each section of the loaded sampling tube is transferred to a 25 ml sample vial, 1 ml DMP is added and the sample vials are sealed tightly with septa and aluminium caps. A 4 Operating conditions for gas chromatography The method was characterized under the following experimental conditions. Apparatus: Head space gas chromatograph Perkin Elmer F 45 with flame ionization detector, autosampler and back-flush device Column: Material: Quartz capillary Length: 50 m Internal diameter: 0.3 mm Stationary phase: Carbowax 20 M Temperatures: Sample thermostat: 70 C Autosampler: 150 C Injector block: 150 C Column: Detector: 250 C Carrier gas: Nitrogen: 2 ml/min Detector gases: Hydrogen: 30 ml/min Synthetic air: 400 ml/min 25 C, isothermal (room temperature) A 5 Analytical determination A gas chromatogram of the contents of an unloaded sampling tube and 1 ml DMP is recorded according to the instructions for the analytical determination given below to ensure that the dimethylphthalate and the activated carbon are free from interfering impurities. The loaded sample vial containing activated carbon from a sampling tube and 1 ml DMP is placed in the sample thermostat at 70 C for 30 min. The same volume as used for calibration (cf. Sect. A 6.1) is then injected from the vapour space into the gas chromatograph and a gas chromatogram is recorded (operating conditions cf. Sect. A 4). Quantitative evaluation is carried out using the area or the height of the ethylene oxide peaks.

6 Analytical Methods AirMonitoringMethods,Vol A 6 Evaluation A 6.1 Calibration In a sample vial the contents of an unloaded activated carbon tube are mixed with 1 ml of the ethylene oxide calibration solution and closed tightly. The sample vial is placed in the thermostat at 70 C for 30 min. A fixed volume is then injected into the gas chromatograph from the vapour space and a gas chromatogram is recorded (cf. Sect. A 4). Quantitative evaluation is carried out using the area or the height of the ethylene oxide peaks (cf. Sect. A 8). A 6.2 Calculation of the analytical result The ethylene oxide concentration in the air sample in mg/m 3 is calculated according to the equation (1): F. w cw F. V (1) 1 The concentration by volume c v in ml/m 3 is calculated from c w at 20 C and 1013 hpa: c v = 0.55 c w Legend: c w Ethylene oxide concentration by weight in the air sample in mg/m 3 c v Ethylene oxide concentration by volume in the air sample in ml/m 3 (ppm) F Peak area (peak height) of ethylene oxide in the sample F 1 Peak area (peak height) of ethylene oxide in the calibration solution w 1 Weight of ethylene oxide in 1 ml of the calibration solution V Air sample volume in L (2) A 7 Reliability of the method A 7.1 Accuracy For ethylene oxide concentrations in the range of 8 ml/m 3 to 33 ml/m 3 the method described yielded a standard deviation of 5.4% and a confidence interval of about 10% (n = 30, P = 95 %) compared to the values measured by a process gas chromatograph. The measuring error of the process gas chromatograph of about 3 % was taken into account.

7 121 AirMonitoringMethods,Vol.4 Ethylene oxide A 7.2 Quantification limit Among other things the quantification limit depends on the type of gas chromatograph and its quality. Under the conditions encountered in practice, amounts as low as 9 µg ethylene oxide per 1 g activated carbon can be detected. This corresponds to ethylene oxide concentrations of 0.5 ml/m 3 (ppm) or 0.9 mg/m 3 for an air sample volume of 10 L. A 7.3 Selectivity The selectivity of the method depends on the type of gas chromatography column. In practice the capillary column described has proved suitable. If interfering substances are present the separation phase must be changed. A 7.4 Recovery Recovery was >90% under the experimental conditions described in Sect. A 7.1. A 8 Discussion Instead of dimethyl phthalate, chlorobenzene, or in the absence of acids and bases, water can be used as a desorbent with no change in the operation conditions. When injecting from the head space, less-volatile compounds, depending on their vapour pressure, especially DMP, are also injected into the gas chromatograph. If the gas chromatograph is not equipped with a back-flush device, interfering peaks or baseline fluctuations may result. Sealed with caps, the loaded activated carbon tubes can be stored for one day at room temperature. However, it is recommended that they are kept cool. If possible, the peaks of the sample and the calibration solutions should be of equal size. In general, deviation by a factor of ± 5 (above and below) is permissible without repeating the gas chromatographic determination with adapted concentrations. Calibration must be carried out at regular intervals. Acid and bases may lead to errors in the measured values and must be taken into account in the complete method. Continuous determination of ethylene oxide using a process gas chromatograph is also possible; however, extensive data obtained with this method is not yet available.

8 Analytical Methods AirMonitoringMethods,Vol A 9 Manufacturers Pump: e.g. Compur Electronic GmbH, München, Du Pont Instruments, Supplier in Germany: DEHA-Haan & Wittmer GmbH, Friolzheim Activated carbon tubes: Type Qazi & Ketcham (SKC Cat. No ), e. g. Compur Electronic GmbH, München, Fleischhacker KG, Schwerte/Ruhr, Günter Karl OHG, Geisenheim, Chrompack Deutschland, Müllheim (Cat. No ) MSA (Charcoal Collection Tube for Organic Hydrocarbons, part. No ), e.g. Auergesellschaft GmbH, Berlin Dräger, Type B, e. g. Drägerwerk AG, Lübeck Head space gas chromatograph: e. g. Bodenseewerk Perkin Elmer & Co GmbH, Überlingen, Carlo Erba, Supplier in Germany: Erba Science, Hofheim/Ts., Siemens AG, Karlsruhe A 10 References [1] Hachenberg H, Schmidt AP (1979) Gas Chromatographic Head Space Analysis. Heyden and Sons, London, Philadelphia, Rheine [2] Qazi, Ketcham (1977) A new method for monitoring personal exposure to ethylene oxide in the occupational environment. Am. Ind. Hyg. Ass. J. 38:

9 123 AirMonitoringMethods,Vol.4 Ethylene oxide Detailed description of the method Contents Operating method B: Gas chromatography after desorption B 2 Equipment, chemicals and solutions B 2.1 Equipment B 2.2 Chemicals B 2.3 Solutions B 3 Sampling and sample preparation B 4 Operating conditions for gas chromatography B 5 Analytical determination B 6 Evaluation B 6.1 Calibration curve B 6.2 Calculation of the analytical result B 7 Reliability of the method B 7.1 Accuracy B 7.2 Quantification limit B 7.3 Selectivity B 8 Discussion B 9 Manufacturers B 10 References B 2 Equipment chemicals and solutions B 2.1 Equipment For sampling and preparation: Pump equipped with gas meter or flow meter Sample vial with polytetrafluoroethylene (PTFE) coated septum and aluminium cap Crimper Activated carbon tubes, (standardized, consisting of two sections filled with about 700 and 400 mg activated carbon separated by a porous polymeric material) Caps for sealing the opened activated carbon tubes For analytical determination: Gas chromatograph equipped with flame ionization detector Compensation recorder and/or integrator

10 Analytical Methods AirMonitoringMethods,Vol B 2.2 Chemicals Ethylene oxide, purity >99% Carbon disulfide for GC, analytical grade Gases for gas chromatography: Nitrogen, hydrogen, purified air or synthetic air B 2.3 Solutions Ethylene oxide standard solution: Solution of about 9 mg ethylene oxide in 20 ml carbon disulfide. About 5 ml gaseous ethylene oxide is slowly injected through the septum with a gastight syringe into the liquid phase of a sample vial containing 20 ml carbon disulfide. Ethylene oxide bubbles should not appear at the tip of the needle, to avoid escape of ethylene oxide when the needle is withdrawn through the septum. The weight of ethylene oxide is determined by weighing to the nearest ±0.1 mg. Ethylene oxide calibration solutions: Solutions of about 0.45, 0.9, 2.25, and 4.50 mg ethylene oxide in 20 ml carbon disulfide. 1.0, 2.0, 5.0 and 10.0 ml of the ethylene oxide standard solution are each transferred to a 20 ml volumetric flask and diluted to the mark with carbon disulfide. B 3 Sampling and sample preparation An activated carbon tube is opened and connected to the pump. Pump and tube are either carried by a person during working hours or stationary sampling is performed. The air sample volume of 10 L per 1 g activated carbon must not be exceeded. The flow rate should not exceed 3 L/h. Each section of the loaded sampling tube is transferred to a 5 ml sample vial, 2 ml carbon disulfide is added and the sample vials are sealed tightly with septum and aluminium cap. The vials are allowed to stand for 15 min and are occasionally shaken (sample solutions). B 4 Operating conditions for gas chromatography The method was characterized under the following experimental conditions: Apparatus: Carlo Erba, Type 2300 equipped with flame ionization detector Column: Material: Stainless steel Length: 2 m Internal diameter: 3 mm Stationary phase: Porapak P

11 125 AirMonitoringMethods,Vol.4 Ethylene oxide Temperatures: Injector block: 150 C Column: 120 C, isothermal Detector: 150 C Carrier gas: Nitrogen: 50 ml/min B 5 Analytical determination A gas chromatogram of the contents of an unloaded sampling tube and 2 ml carbon disulfide is recorded according to the instructions for the analytical determination given below to ensure that the carbon disulfide and the activated carbon are free from interfering impurities. 1 µl of each of the sample solutions (according to Sect. B 3) is injected into the gas chromatograph and a gas chromatogram is recorded. Quantitative evaluation is carried out using the peak areas or the peak heights obtained. B 6 Evaluation B 6.1 Calibration curve 1 µl of each of the calibration solutions is injected into the gas chromatograph. To obtain the calibration curve, the peak areas (peak heights) determined are plotted against the ethylene oxide weights in µg in 2 ml of each calibration solution. B 6.2 Calculation of the analytical result The ethylene weight in µg in 2 ml of the sample solution is taken from the calibration curve using the peak areas (peak heights). The ethylene oxide concentration in the air sample in mg/m 3 is calculated according to the equation (1): c w w V The concentration by volume c v in ml/m 3 is calculated from c w at 20 C and 1013 hpa: (1) c v = 0.55 c w (2) Legend: c w Ethylene oxide concentration by weight in the air sample in mg/m 3 c v Ethylene oxide concentration by volume in the air sample in ml/m 3 (ppm)

12 Analytical Methods AirMonitoringMethods,Vol w V Weight of ethylene oxide in µg in the sample solution taken from the calibration curve Air sample volume in L B 7 Reliability of the method B 7.1 Accuracy For a gas mixture containing 10 ml/m 3 ethylene oxide prepared in the laboratory a relative standard deviation of ± 6% for the complete analytical method was determined from 10 individual measurements. Experience has shown that the accuracy of operating methods A and B are comparable. B 7.2 Quantification limit Under the experimental conditions described the relative quantification limit for ethylene oxide is 0.5 ml/m 3 (ppm) ^= 0.9 mg/m 3 for an air sample volume of 10 L, 2 ml of the desorption solution and a 1 µl injection volume. The absolute quantification limit is 4.5 ng ethylene oxide. This corresponds to 9 µg ethylene oxide for each activated carbon tube. B 7.3 Selectivity The selectivity of the method depends on the type of gas chromatography column. In practice the packed column described has proved suitable. The column packing must be changed if interfering substances are present. B 8 Discussion The loaded activated carbon tubes sealed with caps can be stored at room temperature for one day. However, it is recommended to store them cool. Acids and bases may lead to errors in the measuring values and must be taken into account in the complete method.

13 127 AirMonitoringMethods,Vol.4 Ethylene oxide B 9 Manufacturers Pump: e. g. Compur Electronic GmbH, München, Du Pont Instruments, Supplier in Germany: DEHA-Haan & Wittmer GmbH, Friolzheim Activated carbon tubes: Type Qazi & Ketcham (SKC Cat. No ), e. g. Compur Electronic GmbH, München, Fleischhacker KG, Schwerte/Ruhr, Günter Karl OHG, Geisenheim, Chrompack Deutschland, Müllheim (Cat. No ) MSA (Charcoal Collection Tube for Organic Hydrocarbons, part. No ), e. g. Auergesellschaft GmbH, Berlin Dräger, Type B, e.g. Drägerwerk AG, Lübeck Gas chromatograph: e. g. Bodenseewerk Perkin Elmer & Co GmbH, Überlingen, Carlo Erba, Supplier in Germany: Erba Science, Hofheim/Ts., Siemens AG, Karlsruhe, Varian GmbH, Darmstadt B 10 References [1] Qazi, Ketcham (1977) A new method for monitoring personal exposure to ethylene oxide in the occupational environment. Am. Ind. Hyg. Ass. J. 38:

14 Analytical Methods AirMonitoringMethods,Vol Sampling with a pump and adsorption on activated carbon, gas chromatography after desorption and derivatization with hydrobromic acid This method permits the determination of ethylene oxide concentrations in working areas averaged over the sampling time after personal or stationary sampling. Principle: Technical data: With a pump a measured air volume is drawn through an activated carbon tube. The adsorbed ethylene oxide is desorbed with a mixture of toluene and carbon disulfide, reacted with hydrobromic acid to form bromoethanol and analysed by gas chromatography. Quantification limit: absolute: 0.71 ng bromoethanol ^= 0.25 ng ethylene oxide, relative: 0.1 mg/m 3 ethylene oxide for 12 L air sample, 5 ml desorption solution and 1 µg injection volume. Selectivity: Advantages: Disadvantages: Apparatus: The selectivity must be checked in each individual case. Personal and selective measurements are possible. Peak concentrations are not detected. Pump equipped with gas meter or flow meter, activated carbon tubes, gas chromatograph equipped with electron capture detector (ECD).

15 129 AirMonitoringMethods,Vol.4 Ethylene oxide Detailed description of the method Contents 1 Equipment, chemicals and solutions Equipment Chemicals Solutions Sampling 3 Analytical determination 3.1 Sample preparation and analysis 3.2 Operating conditions for gas chromatography 4 Evaluation 4.1 Calibration 4.2 Calculation of the analytical result Reliability of the method Accuracy Quantification limit Selectivity Recovery Discussion 1 Equipment, chemicals and solutions 1.1 Equipment For sampling: Pump equipped with flow meter or gas meter and holder for the activated carbon tubes (e. g. Compur 4903 from Hartmann & Braun AG, Compur Monitors, München) Activated carbon adsorption tubes (standardized, consisting of two sections filled with about 400 and 200 mg activated carbon separated by a porous polymeric material) (e.g activated carbon tubes from MSA, Pittsburgh, supplier in Germany: Auer, Berlin) For sample preparation and analytical determination: Flat-bed shaker 10 ml and 100 ml Volumetric flasks 10 ml Beaded rim glass Sample vials with polytetrafluoroethylene (PTFE) coated septum and aluminium cap Glass cutter 1 ml Pipette Glass capillaries for dosing the hydrobromic acid Gas chromatograph equipped with ECD Recording and/or computing unit: Compensation recorder and/or computing integrator

16 Analytical Methods AirMonitoringMethods,Vol Chemicals Carbon disulfide, analytical grade Toluene, analytical grade Ethylene oxide, purity > 99 % Hydrobromic acid, 47%, analytical grade Potassium carbonate, analytical grade 2-Bromoethanol, about 95 % (GC) (e. g. Aldrich, Steinheim) 1.3 Solutions Desorption solution: Mixture of toluene and carbon disulfide in a 99:1 volume ratio. 2-Bromoethanol stock solution: Solution of 3.53 mg/ml bromoethanol in elution solution. 200 µl ( ^= 353 mg) 2-bromoethanol is transferred to a 100 ml volumetric flask and diluted to the mark with the desorption solution and occasionally shaken. 2-Bromoethanol calibration solutions: 2-Bromoethanol solutions of 0.71, 1.77, 3.53, and 35.3 µg/ml. 2, 5, 10, 50, and 100 µl of the 2-bromoethanol stock solution are transferred to five separate 10 ml volumetric flasks and diluted to the mark with the desorption solution. Using these solutions and with a sample volume of 12 L a concentration range of mg/m 3 ethylene oxide is covered. 2 Sampling An activated carbon tube is opened and connected to the pump. Pump and sampling tube are either carried by a person during working hours or they are used in a stationary position. The flow rate must not exceed 2.4 L/h. The total air sample volume must not exceed 12 L. 3 Analytical determination 3.1 Sample preparation and analysis The contents of the loaded sampling tube are transferred to a sample vial. For desorption of the ethylene oxide, 5 ml of the desorption solution is added. The vial is occasionally shaken. If necessary, the activated carbon sections are prepared separately.

17 131 AirMonitoringMethods,Vol.4 Ethylene oxide After 1 hour of desorption, 2 ml of the liquid phase is removed with a pipette, transferred to a sample vial and 30 µl hydrobromic acid is added. The vial is shaken on a flat-bed shaker for 30 min. The surplus hydrobromic acid is then neutralized with 75 mg potassium carbonate. The supernatant solution (sample solution) is transferred to a sample vial and the bromoethanol formed in the derivatization of ethylene oxide with hydrobromic acid is determined using gas chromatography (cf. Sect. 3.2). For this purpose 1 µl of the sample solution is injected. Quantitative determination is carried out according to the external standard method using the peak areas or peak heights. Derivatization must not be carried out in the presence activated carbon. Otherwise the formation of bromoethanol is not quantitative. 3.2 Operating conditions for gas chromatography The method was characterized under the following instrumental conditions: Apparatus: Varian 3400 equipped with ECD and autosampler Varian 8100 Column: Material: Quartz capillary Length: 25 m Internal diameter: 0.25 mm Stationary phase: FFAP (modified polyethylene glycol) Film thickness: 1 µm Detector: Electron capture detector Temperatures: Injector block: 210 C Detector: 250 C Column temperature programme: Starting temperature: 80 C Heating rate: 10 C/min Final temperature: 180 C, 5 min isothermal Injection: Splitless Carrier gas: Helium 4.6 Make-up gas: Nitrogen 4 Evaluation 4.1 Calibration 1 µl of each of the calibration solutions (cf. Sect. 1.3) is injected into the gas chromatograph. The calibration curve is obtained by plotting the peak areas (peak heights) against the corresponding weights of ethylene oxide obtained after the conversion of the 2-bromoethanol weights (conversion factor 0.352).

18 Analytical Methods AirMonitoringMethods,Vol Calculation of the analytical result The ethylene oxide concentration in the air sample in mg/m 3 is calculated according to the equation (1): w c w V (1) To calculate the concentration by volume c v in ml/m 3 at 20 C and 1013 hpa: c v = 0.55 c w Legend: c w Ethylene oxide concentration by weight in the air sample in mg/m 3 c v Ethylene oxide concentration by volume in the air sample in ml/m 3 (ppm) w Weight of ethylene oxide in the sample solution in µg V Air sample volume in L (2) 5 Reliability of the method 5.1 Accuracy To determine the accuracy of the method 2, 5, 10, 25 and 35 µl of a solution (1.8 mg ethylene oxide per ml in the desorption solution) were each injected into a glass tube, six times. Laboratory air was drawn through each glass tube connected serially with two activated carbon tubes with a flow rate of 2.4 L/h for 5 hours. For an air sample volume of 12 L, the quantities injected correspond to average concentrations of mg/m 3. The relative standard deviations were between 6 and 12%. 5.2 Quantification limit The absolute quantification limit is 0.71 ng bromoethanol. This corresponds to 0.25 ng ethylene oxide. For 12 L air sample, 5 ml of the sample solution and 1 µl injection volume the relative quantification limit is 0.1 mg/m 3. This corresponds to 0.05 ml/m 3 (ppm) of ethylene oxide. 5.3 Selectivity The selectivity must be checked in each individual case. Compared to that of the ethylene oxide method with flame ionization detection, the selectivity is significantly increased by the derivatization of the ethylene oxide and the use of an electron capture detector.

19 133 AirMonitoringMethods,Vol.4 Ethylene oxide 5.4 Recovery For an air sample volume of 12 L/h and a flow rate of 2.4 L/h recovery was >95% under the conditions described in Sect Recovery was not influenced by a relative humidity of the calibration gas of up to 75 %. 6 Discussion The loaded activated carbon tubes should be analysed immediately after sampling. The shelf life of pure ethylene oxide in the adsorbed state is 7 days.

20 Analytical Methods AirMonitoringMethods,Vol Sampling with a pump and adsorption on the polymeric resin Amberlite XAD-4, gas chromatography after thermal desorption This method permits the determination of ethylene oxide concentrations in working areas averaged over the sampling time after personal or stationary sampling. Principle: Technical data: With a pump a measured air volume is drawn through a metal tube filled with XAD-4. The adsorbed ethylene oxide is desorbed by heating in a thermal desorber and analysed by gas chromatography. Quantification limit: absolute: 20 ng ethylene oxide, relative: 0.1 mg/m 3 ethylene oxide in 200 ml air sample. Selectivity: Advantages: Disadvantages: Apparatus: The selectivity must be checked in each individual case. Personal and selective measurements are possible. Peak concentrations are not detected, generally only one measurement for each adsorption tube possible. Pump for 1 10 ml/min, flow meter for 1 10 ml/min, XAD-4 adsorption tubes, thermal desorber, gas chromatograph equipped with flame ionization detector (FID), Permeation oven and permeation tube filled with ethylene oxide for the preparation of calibration gases (e.g. MCZ, Rosbach).

21 135 AirMonitoringMethods,Vol.4 Ethylene oxide Detailed description of the method Contents 1 Equipment and chemicals 1.1 Equipment 1.2 Chemicals 2 Sampling 3 Analytical determination 3.1 Sample preparation and analysis 3.2 Instrumental operating conditions 4 Evaluation 4.1 Calibration 4.2 Calculation of the analytical result 5 Reliability of the method 5.1 Accuracy 5.2 Quantification limit 5.3 Selectivity 6 Discussion 7 References 1 Equipment and chemicals 1.1 Equipment For sampling: Pump, flow rate 1 10 ml/min (e.g. SKC-PCEx224 from Analyt, Müllheim) Adsorption tubes with XAD-4 as collection phase The adsorption tubes are metal tubes filled with XAD-4. The external diameter and length are adapted to the thermal desorber used. The length of the adsorption section which is enclosed by metal grids depends on the heating zone of the desorption apparatus. After the tube is filled it has to be prepurified in a stream of helium several times in the desorber at 150 C. It is then tightly sealed. Caps for sealing the adsorption tubes For sample preparation and analytical determination: Thermal desorber Gas chromatograph equipped with flame ionization detector Recording and/or computing unit: Compensation recorder and/or computing integrator Permeation furnace and permeation tubes filled with ethylene oxide for the preparation of calibration gases (e. g. MCZ, Rosbach)

22 Analytical Methods AirMonitoringMethods,Vol Chemicals Polymer resin, Amberlite XAD-4, mesh (e.g Serva, Heidelberg) Tenax TA, mesh (e.g. Chrompack, Frankfurt) Gases for gas chromatography: Helium (4.6), purity > % Hydrogen (5.0), purity > % Purified or synthetic air 2 Sampling Prior to sampling the prepurified adsorption tube has to be heated again in the thermal desorber for 15 min at 100 C in a stream of helium and closed tightly. For sampling the adsorption tube is opened and connected to the pump. During sampling the pump and tube are carried by a person or they are used in a stationary position. The flow rate and air sample volume must be adjusted to ensure that 2/3 of the breakthrough volume is not exceeded [1]. These conditions are fulfilled for a tube content of 450 mg XAD-4 at 20 C and 500 ml air sample volume. After sampling the tube is again sealed tightly. 3 Analytical determination 3.1 Sample preparation and analysis The loaded adsorption tube is opened and placed into the thermal desorber. The adsorbed ethylene oxide is desorbed by heating in a stream of helium, trapped in a cooling trap packed with Tenax TA, transferred to the analytical column of the gas chromatograph by flash-heating and then chromatographed. Quantitative evaluation is performed according to the external standard method using the peak areas or the peak heights. 3.2 Instrumental operating conditions The method was characterized under the following instrumental conditions: Apparatus: Thermal desorber ATD-400 and gaschromatograph 2000 equipped with FID from Perkin Elmer GmbH, Überlingen Column: Material: Quartz capillary Length: 25 m Internal diameter: Stationary phase: 0.32 mm Poraplot Q from Chrompack, Frankfurt

23 137 AirMonitoringMethods,Vol.4 Ethylene oxide Temperatures: Thermal desorber: Desorption furnace: 100 C Transfer line: 70 C Column: 150 C isothermal Detector: 200 C Cooling trap: Filling: 40 mg Tenax TA (60-80 mesh) Temperatures: 30 C (adsorption) 300 C (injection) Desorption time: 5 min Carrier gas: Helium 200 kpa Desorption rate: 45 ml/min Split: 10 ml/min Detector gases: Hydrogen: about 30 ml/min Synthetic air: about 300 ml/min 4 Evaluation 4.1 Calibration A calibration gas is prepared according to the method of permeation [3]. For this purpose a permeation tube filled with ethylene oxide is kept at 35 C in the permeation furnace. With a permeation tube of 5 cm in length a permeation rate of 6 µg/min is obtained. Concentrations of mg/m 3 are prepared by dilution with synthetic air. From this calibration gas known volumes are drawn onto adsorption tubes as described in Sect. 3 and analysed [1]. The calibration curve is obtained by plotting the measured peak areas or peak heights against the ethylene oxide weights in ng. The curve is linear in the range of mg/m Calculation of the analytical result The ethylene oxide weight corresponding to the peak area or peak height is taken from the calibration curve. The ethylene oxide concentration in the air sample in mg/m 3 is calculated according to the equation (1): c w w V Legend: c w Ethylene oxide concentration by weight in the air sample in mg/m 3 w V Weight of ethylene oxide in the sample in ng Air sample volume in ml (1)

24 Analytical Methods AirMonitoringMethods,Vol To calculate the concentration by volume c v in ml/m 3 at 20 C and 1013 hpa: c v = 0.55 c w (2) 5 Reliability of the method 5.1 Accuracy The standard deviation of the method was checked in the range of mg/m 3 for three calibration gas concentrations (preparation according to [3]) with a relative air humidity of 40 60%. Table 1. Characteristic data for the determination of ethylene oxide (n = 6). Concentration mg/m 3 Standard deviation mg/m 3 Standard deviation (rel.) s % Quantification limit The quantification limit is 20 ng ethylene oxide per adsorption tube. This corresponds to a relative quantification limit of 0.1 mg/m 3 in a 200 ml air sample. 5.3 Selectivity The selectivity of the method depends on the analytical column used. In practice the column described has proved suitable. A different column must be selected if interfering components are present. 6 Discussion This method was tested using XAD-4 tubes but other adsorbents may also be used. If other adsorbents or tubes of different dimensions are used the breakthrough volume has to be determined again. If stored for longer periods, the adsorption tubes must be heated immediately before sampling in the thermal desorber as interfering components may be transferred to the adsorption phase from the material of the cap washers. The water vapour in humid air (>90% relative humidity) hardly displaces ethylene oxide from the surface of the adsorption phase.

25 139 AirMonitoringMethods,Vol.4 Ethylene oxide In the temperature range of 20 to 25 C no losses from adsorption tubes loaded with ethylene oxide were found one day after sampling. This was established by determination with a calibration gas of 50% relative humidity at a concentration of 2.1 mg/m 3. Determination of 1,2-epoxypropane is also possible using this method. Determination of ethylene oxide in alveolar air is also possible with two serially connected adsorption tubes using the respiratory gas sampler BSC-1 (Perkin Elmer GmbH, Überlingen) equipped with a holder for adsorption tubes [2]. 7 References [1] Tschickardt M (1990) Routineeinsatz des Thermodesorbers ATD-50 in der Gefahrstoffanalytik. In: Perkin Elmer GmbH (Ed) Angewandte Chromatographie, Vol 51. [2] Tschickardt M, Petersen R (1990) Bestimmung organischer Luftschadstoffe in Alveolarluft. In: Perkin Elmer GmbH (Ed) Angewandte Chromatographie, Vol 51. [3] Association of Engineers (VDI) (1980) VDI Guidline 3490, Sheet 9, Measurement of gases Preparation of calibration gas mixtures by permeation procedures. Beuth Verlag, Berlin.

Method for the determination of 1,3-butadiene

Method for the determination of 1,3-butadiene Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Methods for the determination of vinyl chloride

Methods for the determination of vinyl chloride Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Method for the determination of dimethyl sulfate

Method for the determination of dimethyl sulfate German Social Accident Insurance Deutsche Gesetzliche Unfallversicherung Analytical Subcommittee of the Chemistry Board of Experts* Carcinogenic substances Order number: BGI 505-7-05 Established methods:

More information

Hexane isomers (except n-hexane) and methylcyclopentane

Hexane isomers (except n-hexane) and methylcyclopentane 1 Hexane isomers (except n-hexane) and methylcyclopentane Method number 1 Application Air analysis Analytical principle Gas chromatography Completed in July 2010 Summary This analytical method permits

More information

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Turk J Chem 24 (2000), 131 139. c TÜBİTAK Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Naciye KILIÇ University of Uludağ,

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory. VINYL CHLORIDE Method no.: 04 Matrix: Air Target concentration: 3 1 ppm (2.5 mg/m ) (OSHA PEL) Procedure: Collection on charcoal (two-tubes in-series), desorption with carbon disulfide, analysis by gas

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

HYDROCARBONS, AROMATIC 1501

HYDROCARBONS, AROMATIC 1501 HYDROCARBONS, AROMATIC 1501 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1501, Issue 3 EVALUATION: Full Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2

More information

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003 ALCOHOLS IV 1403 (1) HOCH 2 CH 2 OCH 3 MW: (1) 76.09 CAS: (1) 109-86-4 RTECS: (1) KL5775000 (2) HOCH 2 CH 2 OCH 2 CH 3 (2) 90.12 (2) 110-80-5 (2) KK8050000 (3) HOCH 2 CH 2 O(CH 2 ) 3 CH 3 (3) 118.17 (3)

More information

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin OIVMABS15 Anethole. Gas chromatograhpy determination of transanethole in Type II method SCOPE This method is suitable for the determination of transanethole in aniseedflavoured spirit drinks using capillary

More information

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous

More information

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC Aysel T. ATIMTAY Middle East Technical University, Environmental Engineering Department, 06531 Ankara Turkey (aatimtay@metu.edu.tr)

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11890-2 Third edition 2013-03-15 Paints and varnishes Determination of volatile organic compound (VOC) content Part 2: Gas-chromatographic method

More information

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

CH 2252 Instrumental Methods of Analysis Unit V  Gas Chromatography.  M. Subramanian CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

FORMALDEHYDE by GC 2541

FORMALDEHYDE by GC 2541 FORMALDEHYDE by GC 2541 H 2 C=0 MW: 30.03 CAS: 50-00-0 RTECS: LP8925000 METHOD: 2541, Issue 2 EVALUATION: PARTIAL Issue 1: 15 May 1989 Issue 2: 15 August 1994 OSHA : NIOSH: ACGIH: 0.75 ppm; 2 ppm STEL

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

ORGANOPHOSPHORUS PESTICIDES 5600

ORGANOPHOSPHORUS PESTICIDES 5600 ORGANOPHOSPHORUS PESTICIDES 5600 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5600, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : Table 2 NIOSH: Table 2 ACGIH: Table 2 PROPERTIES:

More information

METHYLTIN CHLORIDES 5526

METHYLTIN CHLORIDES 5526 METHYLTIN CHLORIDES 5526 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5526, Issue 1 EVALUATION: FULL Issue 1: 15 March 2003 OSHA : See Table 1 NIOSH: See Table 1 ACGIH: See Table 1

More information

2401 Gas (liquid) Chromatography

2401 Gas (liquid) Chromatography 2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic

More information

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

TDTS.   Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination Thermal Desorption Technical Support Note 75: Liquid standard injection, tube impedance and other factors which may cause discrimination during the calibration of thermal desorption methods Keywords TD

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

Raw Material Finished product By product Exposure standards Length of shift Physical environment

Raw Material Finished product By product Exposure standards Length of shift Physical environment Measurement &Sampling Part 1 Introduction to Sampling What to sample? - Hazard Raw Material Finished product By product Exposure standards Length of shift Recognition Physical environment To Determine

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

Method for Characterization of Gum Rosin by Capillary Gas Chromatography

Method for Characterization of Gum Rosin by Capillary Gas Chromatography PCA TEST METHOD PCTM 27 Published 1/1 2004 1. Scope Method for Characterization of Gum Rosin by Capillary Gas Chromatography 1.1 These test methods cover the determination of the amounts of certain rosin

More information

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

GAS CHROMATOGRAPHY (GC)

GAS CHROMATOGRAPHY (GC) GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and

More information

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid

More information

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Gas Chromatography (GC) In gas chromatography, the sample is vaporized and injected onto the head of a chromatographic

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Principles of Gas- Chromatography (GC)

Principles of Gas- Chromatography (GC) Principles of Gas- Chromatography (GC) Mohammed N. Sabir January 2017 10-Jan-17 1 GC is a chromatographic technique utilizes gas as the mobile phase which is usually an inert gas (Hydrogen, Helium, Nitrogen

More information

Introduction to Gas Chromatography

Introduction to Gas Chromatography Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition

More information

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5 Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Chemistry Instrumental Analysis Lecture 31. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery

More information

637. Thiamethoxam. HPLC method

637. Thiamethoxam. HPLC method 637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Foods, Flavors & Fragrances Applications A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates By Corby Hilliard; Amanda Rigdon; William Schroeder*,

More information

Chapter 3: Source Measurement Techniques

Chapter 3: Source Measurement Techniques Chapter 3 Source Measurement Techniques Measurement Methods Method 18, Measurement of Gaseous Organic Compound Emissions by Gas Chromatography Method 25, Determination of Total Gaseous Non-Methane Organic

More information

H 3 CO H 3 CO S CH 3

H 3 CO H 3 CO S CH 3 FENITROTHION 35 H 3 CO P H 3 CO S O CH 3 NO 2 ISO common name Chemical name Fenitrothion O,O-Dimethyl O-4-nitro-m-tolyl phosphorothioate (IUPAC) O,O-Dimethyl O-(3-methyl-4-nitrophenyl)- phosphorothioate

More information

Gas Chromatography. Introduction

Gas Chromatography. Introduction Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be

More information

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances BRIEFING Saccharin, NF 22 page 2825 and page 1711 of PF 29(5) [Sept. Oct. 2003]. The United States Pharmacopeia is the coordinating pharmacopeia for the international harmonization of the compendial standards

More information

Chapter 11 Conventional Gas Chromatography

Chapter 11 Conventional Gas Chromatography Chapter 11 Conventional Gas Chromatography Gas Chromatography GC is the first instrumental chromatographic method developed commercially It is relatively easy to introduce a stable flow and pressure for

More information

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION. APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Kira. Yang PerkinElmer, Inc. Shanghai, China Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS Introduction

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

*Correspondence to:

*Correspondence to: Supporting Information for Carbonate-promoted hydrogenation of carbon dioxide to multi-carbon carboxylates Aanindeeta Banerjee 1 and Matthew W. Kanan 1 * 1 Department of Chemistry, Stanford University,

More information

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS 1. Scope This method is applicable for determining the presence

More information

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

1. A gas chromatograph equipped with flame ionization detection and a computing integrator. HYDRO.02-1 HYDROXYETHYL SUBSTITUTION LEVEL PRINCIPLE SCOPE The hydroxyethyl group is cleaved from the starch by hydrogen iodide in the presence of adipic acid and an organic solvent containing a known

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask

More information

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010 FORMIC ACID in URINE by GC/MS-headspace - Code GC74010 BIOCHEMISTRY The Methanol is used in the production of cellulose, synthetic resins, formaldehyde, acetic acid and fuel. The route of exposure is more

More information

Chapter 31 Gas Chromatography. Carrier Gas System

Chapter 31 Gas Chromatography. Carrier Gas System Chapter 31 Gas Chromatography GAS-LIQUID CHROMATOGRAPHY In gas chromatography, the components of a vaporized sample are fractionated as a consequence of being partitioned between a mobile gaseous phase

More information

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at

More information

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) Date 02/24/96 Page 1 1.1. This method is designed to measure the concentration

More information

Report AFK0242/18 TABLE OF CONTENTS

Report AFK0242/18 TABLE OF CONTENTS Client: Blue Dent Dental OS: 0161/0211-18 Contact: Rafael Gomes E-mail: rafael@bluedent.com.br Phone: (19) 3563-2222 Address: Rua Joaquim Jorge Port, 1272 City/State: Pirassununga/SP Zip Code: 13636-142

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(6):392-399 Estimation of epichlorohydrin content in pharmaceutical

More information

Structural characterization begins with a purity check!

Structural characterization begins with a purity check! Structural characterization begins with a purity check! Refractive Index Melting Point Elemental Analysis (EA) Thin Layer Chromatography (TLC) High Performance Liquid Chromatography (HPLC) Course: 59-320

More information

Volatile organic compounds (VOCs):

Volatile organic compounds (VOCs): Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

ChromTech options for PAL systems

ChromTech options for PAL systems ChromTech options for PAL systems SPDE features full automation for dynamic sample extraction, desorption and analysis using the CTC Combi PAL. Most of our customers use SPDE for food and flavor research,

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD Kamarruddin ASRI 1, Robert A. ANDERSON 2 1 Department of Chemistry, Jalan

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS METHADONE and in urine by GC-MS Code GC48010 Method of Confirmation by GC-MS INTRODUCTION is a synthetic opioid used as a drug of choice in the detoxification of heroin (and / or opiates). is metabolized

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Due to rapid growth in the medical cannabis industry, demand is increasing for analysis of residual solvents in cannabis concentrates in order to protect consumer safety. This application note

More information

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011 Test Report Tremco-Illbruck GmbH & Co KG Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer June 2011 Client: Tremco-Illbruck GmbH & Co KG Von Der Wettern Str. 27 51149 Köln Germany

More information

GC Instruments. GC Instruments - Sample Introduction

GC Instruments. GC Instruments - Sample Introduction GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes

More information

Rapid Catalyst Screening Reactors

Rapid Catalyst Screening Reactors Rapid Catalyst Screening Reactors Rapid screening of catalysts Analysis of a variety of sample types Multi-modes of operation Rx-3050TR Single µ-reactor Rx-3050SR Overview Two types of rapid catalyst screening

More information

CIPAC. CIPAC Free relevant impurities methods:

CIPAC. CIPAC Free relevant impurities methods: CIPAC COLLABORATIVE INTERNATIONAL PESTICIDES ANALYTICAL COUNCIL LIMITED Commission Internationale des Méthodes d'analyse des Pesticides (CIMAP) CIPAC Free relevant impurities methods: Methods for relevant

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2), ISSN: 0973-4945; CODEN ECJHAO E-Journal of Chemistry http://www.e-journals.net 2010, 7(2), 629-635 Low-level Determination of Residual Methyl Methane Sulfonate and Ethyl Methane Sulfonate in Pharmaceuticals

More information

Instrumental Chemical Analysis

Instrumental Chemical Analysis L2 Page1 Instrumental Chemical Analysis Chromatography (General aspects of chromatography) Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester,

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010 METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010 BIOCHEMISTRY The MEK is presented as a colorless liquid odor similar to that of acetone, is volatile and potentially explosive. The

More information

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 66 No. 6 pp. 611ñ615, 2009 ISSN 0001-6837 Polish Pharmaceutical Society DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE

More information

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams Methods of pollution control and waste management - laboratory Adsorptive removal of volatile organic compounds from gases streams Manual for experiment 17 dr Hanna Wilczura-Wachnik and dr inż. Jadwiga

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

Chromatography. Chromatography is a combination of two words; * Chromo Meaning color * Graphy representation of something on paper (writing)

Chromatography. Chromatography is a combination of two words; * Chromo Meaning color * Graphy representation of something on paper (writing) Chromatography Chromatography is a combination of two words; * Chromo Meaning color * Graphy representation of something on paper (writing) Invention of Chromatography Mikhail Tswett invented chromatography

More information

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Method of determination of phtalates in spirituous beverages by gaschromatography/mass OIV-MA-BS-33 Method of determination of phtalates in spirituous beverages by gas-chromatography/mass Type IV method 1. SCOPE OF APPLICATION This method applies to the detection and assay of some phthalates

More information

TABLE 2 MW: TABLE 1 CAS: TABLE 2 RTECS: TABLE 2. METHOD: 1450, Issue 2 EVALUATION: PARTIAL Issue 1: 15 February 1984 Issue 2: 15 August 1994

TABLE 2 MW: TABLE 1 CAS: TABLE 2 RTECS: TABLE 2. METHOD: 1450, Issue 2 EVALUATION: PARTIAL Issue 1: 15 February 1984 Issue 2: 15 August 1994 ESTERS I 1450 TABLE 2 MW: TABLE 1 CAS: TABLE 2 RTECS: TABLE 2 METHOD: 1450, Issue 2 EVALUATION: PARTIAL Issue 1: 15 February 1984 Issue 2: 15 August 1994 OSHA : Table 2 PROPERTIES: Table 2 NIOSH: Table

More information

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION The GENERAL ASSEMBLY, Considering Article paragraph iv of the agreement establishing the International organisation of vine and wine Upon

More information