Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Size: px
Start display at page:

Download "Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106"

Transcription

1 Turk J Chem 24 (2000), c TÜBİTAK Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Naciye KILIÇ University of Uludağ, Art and Science Faculty, Chemistry Department, Bursa-TURKEY James A. BALLANTINE Monitoring Unit, Chemistry Department, University of Wales, Swansea, SA2 8PP, UK. Received The aim of the research was to investigate the magnitudes of the losses of some volatile organic compounds from the solid adsorbents, Carbotrap and Chromosorb 106 due to back diffusion during long-term diffusive sampling of volatile organic compounds in the environment. Experiments were performed by loading the compounds to be tested on the adsorbent tubes and leaving them in an outdoor environment to simulate the real life sampling conditions over a 14 day period. The losses due to back diffusion were investigated for both the high and the low initial concentrations of the fourteen compounds covering a wide range of volatilies namely acetone, dichloromethane, pentane, hexane, benzene, heptane, toluene, octane, ethylbenzene, m- xylene, o- xylene, nonane, decane and undecane. Introduction Carbotrap, a high purity graphitized carbon black adsorbent has no surface ions or active groups. Hence unlike other common adsorbents the entire surface of Carbotrap is accessible for interactions that depend solely on dispersion (London) forces. The specific surface areas of graphitized carbon blacks range between 20 and 100 m 2 /g. In general porous polymers have a lower surface area and bigger pore sizes than carbon adsorbents. Therefore in comparison with carbon adsorbents they have lower adsorption capacity. Chromosorb porous polymers are available in eight different types. The surface areas and pore sizes of each type differ considerably due to their method of synthesis and/or modification process. Their maximum temperature limits lie between 200 C and 350 C and they can all be used for pre-concentration purposes. Chromosorb 106 is the one with the highest specific surface area and lowest pore size in this series and is classified as a medium strength adsorbent. It has quite high blank values during thermal desorption especially at the heavy ends, which limits the use of it for the sampling of high molecular weight compounds at low concentration levels. The maximum temperature limit of Chromosorb 106 is 250 and hence the thermal desorption of some high boiling compounds may not be complete. 131

2 Diffusive sampling using adsorbent tubes involves the diffusion of the gas or vapour phase compounds from the surrounding environment onto the adsorbent and the adsorbtion of the molecules which come into contact with the adsorbent. The adsoption of the gasses on solid adsorbents occurs as physical adsorption due to the operation of the forces between the solid surface and the adsorbate molecules that are similar to the van der Waals forces between molecules. Since the adsorption of the organic compounds on the solid adsorbents is a reversible physical adsorption, the compounds adsorbed on the adsorbent by diffusion from the surrounding can freely back diffuse from the adsorbent to the surrounding environment. Back diffusion can be a major problem in long-term diffusive sampling causing decreases in uptake rates over time and losses of the adsorbed compounds in the case of fluctuating concentrations, if the adsorbent used does not match the adsorbates sampled. The effect of decreases in uptake rates over time can be optimized by using calculated or experimentally determined effective uptake rates valid for the exposure periods used. The losses of adsorbed material due to back diffusion, after significant decreases in concentrations during the exposure period, causes negative bias in the results obtained. The magnitude of this sampling errors introduced by back diffusion is related to selection of the right adsorbent/ adsorbate combination. If the adsorbent is strong enough to hold the adsorbate, the losses from back diffusion will be low, but if the adsorbent is not strong enough the losses due to back diffusion will cause significant systematic errors in the final concentration calculations. Therefore it is important to measure the likely magnitude of the error introduced by the effect of back diffusion. The effect of exposure to lower or zero concentrations after a high concentration of the pollutants during the sampling period can be examined by exposing the tubes to clean air following a period of exposure to a given pollutant 1. The relative overall uncertainty calculated as precision (measured as twice of coefficient of variation) plus the absolute bias (given by the difference in the results for the tubes exposed to zero air relative to the unexposed tubes) in environmental measurements of benzene have been found to increase with concentration (ranging from 4 ppb to 220 ppb) and to be less than 50 % in each case 2. The losses due to the effect of back diffusion during a seven-day exposure period have been evaluated by exposing samplers loaded with known amounts of vapours to zero air and measuring the fraction retained. This study has shown that losses from Tenax TA were higher than those from Carbotrap for the compounds tested except 1- butanol which retained less on Carbotrap than on Tenax TA which agrees with the adsorbtion characteristics of these adsorbents 3. Experiment To simulate a worst-case scenario of back diffusion effect with extreme changes in the exposed concentrations, the tubes were initially loaded with a certain amount of adsorbates and left in an outdoor environment containing zero or trace levels of the volatile organic compounds. The analytes to be tested were loaded on adsorbent tubes by exposing the tubes to the vapour of the compounds in a sealed glass vessel. The tubes then were left in an outdoor environment for 14 days. Two of the tubes were removed and analysed every other day. The tube loading method and analytical methods for the analysis are given in the following sections. 132

3 Adsorbent Tube Loading A mixture of the analytes, acetone, pentane, dichloromethane, hexane, benzene, heptane, toluene, octane, ethylbenzene, m-xylene, o-xylene, nonane, decane, and undecane were prepared by syringing 20µlofeach these analytes into a vial fitted with a septum seal and weighing on a four decimal place analytical balance. This standard was used for high and low loading experiments. Eight Perkin-Elmer stainless steel diffusion tubes (9 cm in length, with cm 2 cros-sectional area) were packed with Carbotrap mesh (Supelco), 100 m 2 g 1 surface area and another eight tubes were packed with Chromosorb 106, mesh (Supelco). The adsorbent length was 50 mm for each adsorbent. The tubes fitted with a Swagelok end-cap with a PTFE ferrula at the closed end and a diffusion cap with a silicon membrane at the sampling end were placed in an aluminium tube holder. The tubes held in the tube holder were placed inside the 5 litre volume Pyrex glass vessel. A 4 mg aliquot of the standard mixture was syringed into the vessel for the high loading experiment and the tubes were exposed to this mixture for one hour. For the low loading experiment only 0.5 mg of the standard mixture was used with one hour exposure time. The precision of the tube loading method was investigated for both high and low loading on Chromosorb 106 adsorbent by loading 16 tubes and desorbing them to obtain the % RSD between the tubes. The average % RSD for high loading was 8±3 ranging from 5 to 16 and for low loading 9±4 ranging from 5 to 22. Back Diffusion Experiment A total of 16 tubes for each experiment were loaded with the 14 compounds using the method described in the previous section. Two of the tubes were analysed to obtain the initial weight (ng) loaded on the tubes and the remaining fourteen tubes were placed at equal spaces in a rig and left in an outdoor environment. Two of the tubes were removed and analysed every two days during the 14-day back diffusion period. The background levels of volatile organic compounds present in the outdoor environment were also checked during this period by using an identical adsorbent tube which was changed at the same time that the tubes were removed. Instrumentation and Analytical Parameters A Perkin Elmer Model ATD 400 Automatic Thermal Desorption injector system and a Hewlett Packard 5890 II gas chromatograph fitted with a flame ionisation detector were used for the subsequent analysis of the collected samples. The results were plotted on a Hewlett Packard Model HP 3396A printer/plotter integrator and the data stored on an Elonex 286 M-120 PC using Hewlett Packard Peak 96 software. ATD 400 Automatic Thermal Desorber parameters Desorption Time: 10 min, Desorption Temperature: 250 C (for Chromosorb 106) and 320 C (for Carbotrap), Trap hold: 5 min, Trap high: 250 C, Trap low: -30 C, Trap packing: Tenax TA, Split factor: 5.6 (17.85 % to GC). GC Conditions A 25 x 0.3 mm id WCOT column with a 1 µm film tickness of polydimethysiloxane was used with a flow rate of the carrier gas (helium) of 1.8 ml/min. The column temperature was held for 2 min at 40 C, then increased at 10 C /min to 200 C, which was held for 5 min. 133

4 Calibration of analytes Individual analytes in the standard mixture were identified by GC retention times. A four level external calibration of the analytes of interest was carried out by spiking adsorption tubes in a flowing stream of nitrogen with a multi-component standard mixture made up in methanol. The solvent (methanol) was purged with nitrogen flow to avoid overloading the cold trap and column. Since methanol has a low retention volume on Tenax TA, most of it broke through whereas the other compounds were adsorbed. The possible breakthrough of the compounds during nitrogen purging was checked and the optimum flow rate and volume were used. Nitrogen was purged through the adsorbent tube at 33 ml min 1 for 5 min to ensure the adsorption of the compounds of interest and removal of methanol. Results The analysis of two tubes at two day intervals during 14 days gave the ng amounts of the analytes retained on the tubes. The percentage of the analytes retained on the tubes on the day which they were sampled, relative to the initial concentration was calculated. The results of these calculations are given graphically by plotting the percentage of the analytes retained after losses in the amounts of analytes from the tubes during the 14-day period due to back diffusion against time. Figures 1 and 2 show the results obtained for high and low loading on Chromosorb 106. The results obtained for high and low loading on Carbotrap are showninfigures3and4. Figure 1. Percentage of the compounds retained on Chromosorb 106 during the 14-day period for a high loading case. 134

5 Figure 2. Percentage of the compounds retained on Chromosorb 106 during the 14-day period for a low loading case. Figure 3. Percentage of the compounds retained on Carbotrap during the 14-day period for a high loading case. 135

6 Figure 4. Percentage of the compounds retained on Carbotrap during the 14-day period for a low loading case. As can be seen from the graphical results, the losses due to back diffusion are significantly high for low boiling analytes, on both adsorbents and decrease with decreasing volatility of the analytes. The bias (systematic error) that could be introduced due to back diffusion if such extreme concentration changes are seen in a real sampling situation were calculated using the equation below 4. Bias = x x ref. x ref In this equation x is the mean value of a number of repeated measurements whereas x ref isthetrueor accepted reference value for that measurement. In the calculations of the bias for the data obtained in this experiment: x=the weight of the analyte left on the tube after the 14th day (day 14). x ref =the initial weight of the analyte loaded on the tube (day 0). Relative % biases calculated for low and high loadings on Chromosorb 106 are given in Table 1 and the % bias values calculated for high and low loading cases on Carbotrap adsorbent are given in Table 2. The high negative values obtained for low boiling compounds illustrate the significant losses in amounts of analytes adsorbed in this worst-case scenario with a dramatic decrease in concentration. The bias values obtained for Carbotrap are higher than those obtained for Chromosorb 106 for low boiling compounds. The high loading case showed higher negative bias values than the low loading case for Carbotrap, but same was not true for Chromosorb 106. The experimental results are subject to random errors due to a number of reasons, such as between tube variations during loading and the effect of uncontrolled outdoor conditions such as humidity, temperature, rainfall and wind velocity during the 14-day back diffusion period. The precision of the experiments 136

7 were calculated to give percentage relative standard deviations for the two tubes analysed every two days during 14 days. Table 1. Table 1 A comparison of % bias due to back diffusion from Chromosorb 106 during 14 days for low and high loading cases. CHROMOSORB 106 COMPOUND Low Loading (ng) High Loading (ng) Day 0 Day 14 %Bias Day 0 Day 14 %Bias Acetone N-Pentane Dichloromethane N-Hexane Benzene Heptane Toluene N-Octane Ethylbenzene M Xylene O-Xylene N-Nonane N-Decane N-Undecane Table 2. A comparison of % bias due to back diffusion from Carbotrap during 14-days for low and high loading cases. CARBOTRAP COMPOUND Low Loading (ng) High Loading (ng) Day 0 Day 14 %Bias Day 0 Day 14 %Bias Acetone N-Pentane Dichloromethane N-Hexane Benzene Heptane Toluene N-Octane Ethylbenzene M Xylene O-Xylene N-Nonane N-Decane N-Undecane The averaged % RSD values obtained for each analyte during the experiments are given in Table 3. The averaged % RSD values obtained for Chromosorb 106 are low for both low and high loading cases. The relatively high % RSD values for high boiling compounds decane and undecane are found in every experiment during this research. This is perhaps due to incomplete desorption during the thermal desorption step. 137

8 Although the % RSD values obtained for Carbotrap were found to be relatively high, in general the results are reasonable. The % bias values for Carbotrap shown in Table 2 are in agreement with the known low retention of the analytes with a functional group on graphitized carbon adsorbent Carbotrap. Table 3. Averaged % RSD values for the compounds AVERAGED % RSD CHROMOSORB 106 CARBOTRAP COMPOUND Low Loading High Loading Low Loading High Loading Acetone 4±2 7±4 7±5 2± N-Pentane 8±4 9±3 36±30 54±31 Dichloromethane 6±4 7±4 9± 1± N-Hexane 5±3 3±3 12±10 14±10 Benzene 5±3 4±3 18±12 29±19 Heptane 4±2 4±3 7±8 4±5 Toluene 3±2 6±4 7±9 7±9 N-Octane 5±4 5±3 5±5 5±6 Ethylbenzene 4±1 6±4 5±6 7±8 M Xylene 4±1 6±3 5±5 7±7 O-Xylene 4±1 5±3 5±6 7±7 N-Nonane 7±4 4±2 5±4 7±7 N-Decane 11±8 4±2 12±14 9±9 N-Undecane 18±15 8±6 23±21 14±16 Conclusions The adsorption of the gasses on solid adsorbents occurs as physical adsorption due to the operation of the forces between the solid surface and the adsorbate molecules that are similar to the van der Waals forces between molecules. Since the adsorption of the organic compounds on the solid adsorbents is a reversible physical adsorption, the compounds adsorbed on the adsorbent by diffusion from the surrounding can freely back diffuse from the adsorbent to the surrounding environment. This will cause the vapour pressure of the adsorbate to built up on the surface of the adsorbent and, thus decrease in the uptake rates over time. The degree of back diffusion will be depend on the match of the adsorbent adsorbate pair and the uptake rate of each adsorbate on a particular adsorbent will be different. The magnitude of the sampling error introduced by back diffusion is related to the selection of right adsorbent/adsorbate combination. If the adsorbent is strong enough to hold the adsorbate, the losses from back diffusion will be low, but if the adsorbent is not strong enough the losses due to back diffusion will cause significant systematic errors in the final concentration calculations. This study showed that significant losses in the amount of low boiling compounds occurred during long-term diffusive sampling due to back diffusion of the adsorbed compounds in this worst-case scenario after a dramatic decrease in the atmospheric concentration of the analytes. The losses from the Chromosorb 106 are much lower than those from Carbotrap, about 50 % loss for acetone, pentane and dichloromethane in both high and low loading cases. The losses decrease with the decreasing volatility of the compounds as expected. Although the results of these experiments are subject to random errors due to a number of reasons, such as between tube variations during loading and the effect of uncontrolled outdoor conditions such as humidity, temperature, rainfall and wind velocity during 14 day back diffusion period, in general, the results obtained are meaningful and simulate the a real sampling conditions. 138

9 The adsorbent loading method used to load the analytes on the adsorbent tubes was very quick and practical and simulates the adsorption of the compounds onto the adsorbents a in real sampling situation. The between tube variations (%RSD) obtained for trial loading of 16 Chromosorb 106 tubes are 8±3 for high loading and 9±4 for low loading which are slightly higher than normally obtained % RSD (3 to 5) for introduction of the analytes on adsorbent tubes by syringing. Clearly these results show that for the monitoring of environmental air it is more satisfactory to use Chromosorb 106 than Carbotrap in diffusion experiments where the identity of the analytes are unknown. Back diffusion in a worst-case scenario has given a maximum negative bias of 50% using Chromosorb 106, except for the case of dichloromethane where negative bias values of 81 and 69 % were observed. Where the analytes are known, and their adsorbate/adsorbent adsorption isotherm constants are also available, it is possible to calculate the effective uptake rates using the computer program of Nordstrand and Kristensson 5 and hence achieve much better precision in the analysis over long-term environmental low level monitoring. References 1. Protocol for Assessing the Performance of a Diffusive Sampler, MDHS 27, HSE, Occupational Medicine and Hygene Laboratory, February, K. J. Saunders, The Use of Diffusive Samplers for Measuring Benzene at ppb Levels in the Environment, The Diffusive Monitor, HSE/CAR Working Group 5, Issue 6, October Bertoli M. De et. al., Proc. of the 4th Conf. on Indoor Air Quality, , West Berlin, August S. Zloczysti, Clean Air at Work - New Trends in Assessment and Measurement for 1990s, R. H. Brown et. al.(ed)., E. Nordstrand and J. Kristensson, J. of Am. Ind. Hyg. Assoc., V.55(10), ,

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES Mihriban Yılmaz Civan, Öznur Kuntasal and Gürdal Tuncel 1 1 OUTLINE Introduction Importance and

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Method for the determination of 1,3-butadiene

Method for the determination of 1,3-butadiene Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011 Test Report Tremco-Illbruck GmbH & Co KG Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer June 2011 Client: Tremco-Illbruck GmbH & Co KG Von Der Wettern Str. 27 51149 Köln Germany

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

TDTS.   Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination Thermal Desorption Technical Support Note 75: Liquid standard injection, tube impedance and other factors which may cause discrimination during the calibration of thermal desorption methods Keywords TD

More information

Raw Material Finished product By product Exposure standards Length of shift Physical environment

Raw Material Finished product By product Exposure standards Length of shift Physical environment Measurement &Sampling Part 1 Introduction to Sampling What to sample? - Hazard Raw Material Finished product By product Exposure standards Length of shift Recognition Physical environment To Determine

More information

method for monitoring exposure to gasoline vapour in air revision 2002

method for monitoring exposure to gasoline vapour in air revision 2002 method for monitoring exposure to gasoline vapour in air revision 2002 Prepared by: P. E. Tindle Reviewed for CONCAWE by: A. Bianchi E. Loader C. Money G. Pizzella J. Urbanus (Technical Co-ordinator) Reproduction

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02 Polyseam A/S Ravneveien 7 Linnestad Næringsområde 3174 Revetal Norway Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 09 October 2014

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

Tar & (SPA) tar analysis

Tar & (SPA) tar analysis Tar & (SPA) tar analysis Truls Liliedahl Chemical Technology School of Chemical Science and Engineering Royal Institute of Technology (KTH) Stockholm, Sweden Tar when gasifying biomass Gasification at

More information

Characterising Volatile Organic Compounds from Sewer Emissions by Thermal Desorption coupled with Gas- Chromatography-Mass Spectrometry

Characterising Volatile Organic Compounds from Sewer Emissions by Thermal Desorption coupled with Gas- Chromatography-Mass Spectrometry A publication of CHEMICAL ENGINEERING TRANSACTIONS VOL. 30, 2012 Guest Editor: Renato Del Rosso Copyright 2012, AIDIC Servizi S.r.l., ISBN 978-88-95608-21-1; ISSN 1974-9791 The Italian Association of Chemical

More information

Hexane isomers (except n-hexane) and methylcyclopentane

Hexane isomers (except n-hexane) and methylcyclopentane 1 Hexane isomers (except n-hexane) and methylcyclopentane Method number 1 Application Air analysis Analytical principle Gas chromatography Completed in July 2010 Summary This analytical method permits

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

Methods for the determination of vinyl chloride

Methods for the determination of vinyl chloride Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

GAS CHROMATOGRAPHY (GC)

GAS CHROMATOGRAPHY (GC) GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Method for the determination of dimethyl sulfate

Method for the determination of dimethyl sulfate German Social Accident Insurance Deutsche Gesetzliche Unfallversicherung Analytical Subcommittee of the Chemistry Board of Experts* Carcinogenic substances Order number: BGI 505-7-05 Established methods:

More information

VOC TEST REPORT CDPH

VOC TEST REPORT CDPH Baux AB Östermalmsgatan 26A 114 26 Stockholm SWEDEN Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT CDPH 29 September 2017

More information

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs Vapor Intrusion Sampling Options: Performance Data for s, Badges, and Sorbent Tubes for VOCs Linda S. Coyne SKC Inc., 863 Valley View Road, Eighty Four, PA 1533 George Havalias, Maria C. Echarte American

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe L Isolante K-flex Srl Via Leonardi da Vinci 36 20877 Roncello (MB) Italy Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 29 August 2013

More information

Trajan SGE GC Columns

Trajan SGE GC Columns Trajan Scientific and Medical Trajan SGE GC Columns Trajan Scientific and Medical Our focus is on developing and commercializing technologies that enable analytical systems to be more selective, sensitive

More information

Tar analysis by Solid Phase Adsorption (SPA) associated with Thermal Desorption (TD) and Gas Chromatography (GC) analysis

Tar analysis by Solid Phase Adsorption (SPA) associated with Thermal Desorption (TD) and Gas Chromatography (GC) analysis Tar analysis by Solid Phase Adsorption (SPA) associated with Thermal Desorption (TD) and Gas Chromatography (GC) analysis E. Masson (CRITT Bois, Epinal, France) A. Dufour (CNRS, Nancy, France) International

More information

Recent advances in Inverse Gas Chromatography - IGC - measurements of HSP. RALF DUEMPELMANN, ERIC BRENDLE and STEVEN ABBOTT

Recent advances in Inverse Gas Chromatography - IGC - measurements of HSP. RALF DUEMPELMANN, ERIC BRENDLE and STEVEN ABBOTT Recent advances in Inverse Gas Chromatography - IGC - measurements of HSP RALF DUEMPELMANN, ERIC BRENDLE and STEVEN ABBOTT Content Introduction Inverse Chromatography 4Some principles How to determine

More information

SURVEY OF CHEMICALS EMISSION FACTORS FROM BUILDING MATERIALS FOR INTERIORS

SURVEY OF CHEMICALS EMISSION FACTORS FROM BUILDING MATERIALS FOR INTERIORS SURVEY OF CHEMICALS EMISSION FACTORS FROM BUILDING MATERIALS FOR INTERIORS Hitomi Yoshida Japan Testing Center for Construction Materials, Japan Abstract Indoor air pollution in buildings is caused by

More information

VOC EMISSION TEST REPORT ISO 16000

VOC EMISSION TEST REPORT ISO 16000 Proxy A/S Gothersgade 12, 1.th 1123 København K DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark CustomerSupport@eurofins.com www.eurofins.com/voc-testing 1 Sample Information VOC

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

Analysis of BTEX in Natural Water with SPME

Analysis of BTEX in Natural Water with SPME Analysis of BTEX in Natural Water with SPME Application Note Environmental Author Giordano Vassalli Sezione Protezione Aria Introduction Benzene, toluene, ethylbenzene and xylene (BTEX) isomers are monocyclic

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory. VINYL CHLORIDE Method no.: 04 Matrix: Air Target concentration: 3 1 ppm (2.5 mg/m ) (OSHA PEL) Procedure: Collection on charcoal (two-tubes in-series), desorption with carbon disulfide, analysis by gas

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Fabriksvej 19 6000 Kolding DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 13 August 2015 Test Report- VOC emission regulations

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

VOC EMISSION TEST REPORT BREEAM NOR

VOC EMISSION TEST REPORT BREEAM NOR Hey di AS Tretjerndalsveien 68 Pb 13 2017 Frogner NORWAY Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark CustomerSupport@eurofins.com www.eurofins.com/voc-testing 1 Sample Information

More information

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.

More information

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing.

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing. Thermal Desorption Technical Support Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing Keywords certification protocols, (T)(V)(S)VOCs, ULTRA-UNITY, field/laboratory

More information

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278 Application Note Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method Application Note Abstract In this application note, we demonstrate the suitability of Markes

More information

Selection of a Capillary

Selection of a Capillary Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Aps Fabriksvej 19 6000 Kolding Denmark Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 14 November 2014 Test Report- VOC emission

More information

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin OIVMABS15 Anethole. Gas chromatograhpy determination of transanethole in Type II method SCOPE This method is suitable for the determination of transanethole in aniseedflavoured spirit drinks using capillary

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

VOC TEST REPORT AgBB

VOC TEST REPORT AgBB Bostik SA 253 rue du Président Wilson 93211 La Plaine St Denis Cedex FRANCE Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT

More information

AIR. Ambient, Indoor, Workplace Air and Stack Emissions Proficiency Testing Scheme. Sample Preparation Instructions Round 1

AIR. Ambient, Indoor, Workplace Air and Stack Emissions Proficiency Testing Scheme. Sample Preparation Instructions Round 1 General Instructions Ambient, Indoor, Workplace Air and Sample Instructions Round 1 Sample Storage All samples should be stored in accordance with the instructions provided on the sample labels from the

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol Dana Lim A/S Københavnsvej 220 4600 Køge DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 6 April 2017 1 Sample

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air Application TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air Summary This Application Note describes validation of a thermal

More information

VOC TEST REPORT CDPH

VOC TEST REPORT CDPH Adolf Würth GmbH und Co. KG Reinhold-Würth Str. 12-17 74653 Künzelsau GERMANY Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

GB Translated English of Chinese Standard: GB NATIONAL STANDARD Translated English of Chinese Standard: GB18581-2009 www.chinesestandard.net Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA ICS 87.040 G 51 GB 18581-2009 Replacing GB

More information

VOC TEST REPORT EMICODE

VOC TEST REPORT EMICODE Polyseam Ltd Shaw Park, Silver Street Huddersfield, West Yorkshire, HD5 9AF UK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe CONICA AG Industriestr. 26 CH-8207 Schauffhausen Switzerland Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 31 July 2014 Test Report-

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol SGS-Doors A/S Thrigesvej 8-16 7430 Ikast DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 16 August 2016 1 Sample

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Selection of a Capillary GC Column

Selection of a Capillary GC Column Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer

More information

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

CH 2252 Instrumental Methods of Analysis Unit V  Gas Chromatography.  M. Subramanian CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603

More information

Methods for the determination of ethylene oxide

Methods for the determination of ethylene oxide Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous

More information

EPA TO-17 Volatile Organic Compounds

EPA TO-17 Volatile Organic Compounds EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary

More information

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc. Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical

More information

HYDROCARBONS, AROMATIC 1501

HYDROCARBONS, AROMATIC 1501 HYDROCARBONS, AROMATIC 1501 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1501, Issue 3 EVALUATION: Full Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2

More information

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

1. A gas chromatograph equipped with flame ionization detection and a computing integrator. HYDRO.02-1 HYDROXYETHYL SUBSTITUTION LEVEL PRINCIPLE SCOPE The hydroxyethyl group is cleaved from the starch by hydrogen iodide in the presence of adipic acid and an organic solvent containing a known

More information

Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography

Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography 2003 The Japan Society for Analytical Chemistry 1365 Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography Mohammad A. MOTTALEB,* Mohammad

More information

Determination of Solubility Parameters using Inverse Gas Chromatography

Determination of Solubility Parameters using Inverse Gas Chromatography Determination of Solubility Parameters using Inverse Gas Chromatography Anett Kondor PhD IGC-SEA Product Manager and igc Product Specialist 7 th April 017 Overview Introduction Inverse Gas Chromatography

More information

VOC TEST REPORT CDPH

VOC TEST REPORT CDPH Euro Trade Flooring SL. Pol. Ind. Can Estella c/galileo no. 11 08635 Sant Esteve Sesrovires BARCELONA SPAIN Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

Rapid Catalyst Screening Reactors

Rapid Catalyst Screening Reactors Rapid Catalyst Screening Reactors Rapid screening of catalysts Analysis of a variety of sample types Multi-modes of operation Rx-3050TR Single µ-reactor Rx-3050SR Overview Two types of rapid catalyst screening

More information

Real-Time Detection: From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953)

Real-Time Detection: From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953) Real-Time Detection: 1953 From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953) Sampling Gases and Vapors Gas: A state of matter characterized by very low density and

More information

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol Würth Oy Würthintie 1 11710 Riihimäki Finland Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 23 October 2017 1 Sample

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Ch24. Gas Chromatography (GC)

Ch24. Gas Chromatography (GC) Ch24. Gas Chromatography (GC) 24.1 What did they eat in the year 1000? From 13 C content of cholesterol in ancient bone 13 C : 1.1%, 12 C: 98.9% 13 C/ 12 C ratio types of plants Bones of 50 people in Barton-on-Humber

More information

Test Report. Muovihaka Oy. Product Emissions in accordance with the Swedish Building Material Assessment Ultima graphite.

Test Report. Muovihaka Oy. Product Emissions in accordance with the Swedish Building Material Assessment Ultima graphite. Test Report Muovihaka Oy Product Emissions in accordance with the Swedish Building Material Assessment July 2012 Client: Muovihaka Oy Kuusimäentie 12 01900 Nurmijärvi Finland Date: 26 July 2012 Testing

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

CHEM340 Tutorial 4: Chromatography

CHEM340 Tutorial 4: Chromatography CHEM340 Tutorial 4: Chromatography 1. The data in the table below was obtained from a chromatogram obtained with a 10 cm liquid chromatography column. Under the conditions used, the compound uracil is

More information

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten Test Report Glava AS Emission test of Ecophon Venus according to M1 classification October / November 2006 Client: Glava AS Nybråtveien 2 1832 Askim Norge Date: 8 th of November 2006 Testing Laboratory:

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

Test Report in accordance CDPH-IAQ

Test Report in accordance CDPH-IAQ Hilti Entwicklungsgesellschaft GmbH Hiltistraße 6 86916 Kaufering Germany Test Report in accordance CDPH-IAQ Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Emission measurements according to, M1 (3 appendices)

Emission measurements according to, M1 (3 appendices) issued by an Accredited Testing Laboratory Contact person Ulrika Johansson 2017-01-30 6F024225-1 1 (6) Chemistry, Materials and Surfaces +46 10 516 53 22 ulrika.johansson@sp.se Accred. No. 1002 Testing

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

VOC TEST REPORT EMICODE

VOC TEST REPORT EMICODE Bostik SA 253 rue du Président Wilson 93211 La Plaine St Denis Cedex FRANCE Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT

More information

INTRINSICALLY VALID VOLATILE ORGANIC COMPOUND MEASUREMENTS

INTRINSICALLY VALID VOLATILE ORGANIC COMPOUND MEASUREMENTS JPACSM 107 INTRINSICALLY VALID VOLATILE ORGANIC COMPOUND MEASUREMENTS Michael Huxham and C.L. Paul Thomas Department of Instrumentation and Analytical Science UMIST Manchester, UK INTRODUCTION Active adsorption

More information

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Brochure Brief Porous Layer Open Tubular (PLOT) columns have been replacing traditional packed columns

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Air Monitoring Sorbent Pens. For Environmental and IH Applications

Air Monitoring Sorbent Pens. For Environmental and IH Applications Air Monitoring Sorbent Pens For Environmental and IH Applications Sorbent Pens for Quantitative Air Monitoring Sorbent Pens mark the latest innovation in quantitative air monitoring from Entech Instruments.

More information

Determination of Normal Hydrocarbons C 5 -C 18 Using Silanized Algerian Bentonite (B 1100 ) as Support in Gas Chromatographic Analysis

Determination of Normal Hydrocarbons C 5 -C 18 Using Silanized Algerian Bentonite (B 1100 ) as Support in Gas Chromatographic Analysis Asian Journal of Chemistry Vol. 20, No. 1 (2008), 66-74 Determination of Normal Hydrocarbons C 5 -C 18 Using Silanized Algerian Bentonite (B 1100 ) as Support in Gas Chromatographic Analysis M. LAHMEK

More information

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES ROBOKROM 1 TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview ROBOKROM 2 +8 OPERATIONAL MODES HRGC+HRGC-MS HRGC+HPLC-MS STATIC HEAD SPACE PURGE & TRAP SMPE

More information