Methods for the determination of vinyl chloride

Size: px
Start display at page:

Download "Methods for the determination of vinyl chloride"

Transcription

1 Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, Sankt Augustin Expert Committee Chemistry Carcinogenic substances Order number: ZH 1/120.12E Established methods Issue: October 1992 Methods for the determination of vinyl chloride Methods tested and recommended by the Berufsgenossenschaften for the determination of vinyl chloride in working areas. Continous sampling 1 Process gas chromatography for stationary measurements 2 Instrument equipped with a galvenic measuring cell for stationary measurements Discontinous sampling For the assessment of working areas, both personal or stationary sampling are possible: 1 Sampling with a pump and adsorption on activated carbon, gas chromatography after desorption. Vinyl chloride-1-gc (Issue: December 1983, revised October 1992) 2 Sampling with a pump and adsorption on activated carbon, headspace gas chromatography after desorption. Vinyl chloride-2-hsgc (Issue: October 1992) IUPAC name: CAS No: vinyl chloride AirMonitoringMethods,Vol.4

2 Analytical Methods AirMonitoringMethods,Vol Continuous sampling For monitoring the concentrations in working areas stationary process gas chromatographs or instruments equipped with a galvanic measuring cell may be used. Portable flame ionization detectors (FID) or photoionization detectors (PID) can be used for the detection of leaks in plants, for control measurements during cleaning processes and repair and also for screening tests. 1 Process gas chromatography for stationary measurements Principle: Technical data: Minimum measuring range: 0 5 ml/m 3 (ppm) Quantification limit: Selectivity: A measured air volume is injected into a carrier gas flow and conducted through a chromatography column. In the column vinyl chloride is separated from accompanying substances in the air sample and determined with a flame ionization detector (FID). 0.1 ml/m 3 (ppm) ^= 0.3 mg/m 3 vinyl chloride in air The method is selective if appropriate columns are used. Measuring time for 1 determination: 1 3 minutes (depending on the accompanying substances). Advantages: Working materials: Manufacturers: Indication of local and temporal concentration changes is possible, rapid and direct response. Pneumatic auxiliary energy, air of highest purity or synthetic air, hydrogen of highest purity, carrier gas. e.g. Asea Brown Boveri, Mannheim, Mannesmann Hartman & Braun, Frankfurt, Siemens AG, Karlsruhe.

3 195 AirMonitoringMethods,Vol.4 Vinyl chloride 2 Instrument equipped with a galvanic measuring cell for stationary measurements Principle: Technical data: Minimum measuring range: 0 10 ml/m 3 (ppm) Quantification limit: Selectivity: Transient response: Advantages: Working material: Manufacturer: Chlorine and hydrogen chloride are released from vinyl chloride at temperatures above 800 C. The components are absorbed in an electrolyte solution and determined potentiometrically with halogen-specific electrodes. 0.2 ml/m 3 (ppm) ^= 0.5 mg/m 3 vinyl chloride in air The method is not selective. (Halogen-specific determination, with selective filter, specific for halogenated hydrocarbons). T 90 2 minutes Indication of local and temporal concentration changes is possible, rapid and direct response. Electrolyte solution. e. g. BASF AG, Ludwigshafen.

4 Analytical Methods AirMonitoringMethods,Vol Discontinuous sampling 1 Sampling with a pump and adsorption on activated carbon, gas chromatography after desorption. Principle: Technical data: With a pump a measured air volume is drawn through activated carbon. After desorption with carbon disulfide the adsorbed vinyl chloride is determined by gas chromatography. Quantification limit: Under the given operating conditions the relative quantification limit is 0.2 ml/m 3 (ppm) ^= 0.5 mg/m 3 vinyl chloride. Selectivity: Advantages: Disadvantages: Apparatus: As a result of interfering substances the concentrations determined may be too high. Interference can generally be eliminated by selecting another column. Personal and selective measurements are possible. No indication of peak concentrations, time-consuming. Pump equipped with gas meter or flow meter, activated carbon tubes, gas chromatograph equipped with flame ionization detector (FID).

5 197 AirMonitoringMethods,Vol.4 Vinyl chloride Detailed description of the method Contents 1 Summary 2 Equipment, chemicals and solutions 2.1 Equipment 2.2 Chemicals 2.3 Solutions 3 Sampling and sample preparation 4 Operating conditions for gas chromatography 5 Analytical determination 6 Calculation of the analytical result 6.1 Calibration curve 6.2 Evaluation of the results by the external standard method 7 Reliability of the method 7.1 Accuracy 7.2 Quantification limit 8 Discussion 9 Manufacturers 1 Summary This method permits the determination of the concentration of vinyl chloride in working areas averaged over the sampling time after personal or stationary sampling. With a pump, which is carried by a person or installed in a stationary position, a measured air volume is drawn through a glass tube containing activated carbon. After the adsorbed vinyl chloride has been desorbed with carbon disulfide it is determined by gas chromatography. Under the conditions found in practice, the absolute quantification limit is 1 ng vinyl chloride. The relative quantification limit is 0.2 ml/m 3. This corresponds to 0.5 mg/m 3 vinyl chloride for a 20 L air sample.

6 Analytical Methods AirMonitoringMethods,Vol Equipment, chemicals and solutions 2.1 Equipment For sampling and sample preparation: Pump Adsorption tubes with activated carbon, Type NIOSH, standardised, consisting of two sections each filled with activated carbon weighing about g which are separated by a porous polymeric material. 10 ml Volumetric flasks with ground glass stoppers For analysis: Gas chromatograph with flame ionization detector 2.2 Chemicals Carbon disulfide, analytical grade Vinyl chloride, purity >99% Gases for gas chromatography: Nitrogen, hydrogen, synthetic air 2.3 Solutions Vinyl chloride stock solution: 2.58 mg vinyl chloride dissolved in 10 ml carbon disulfide. 10 ml carbon disulfide in a vial equipped with septum and screw cap are cooled with dry ice. 1 L vinyl chloride is introduced into an evacuated plastic vessel. At 20 C and 1013 hpa 1 ml of this volume is removed with a gastight syringe and injected through the septum into the vial containing the cooled carbon disulfide. Then the vial is shaken thoroughly. Vinyl chloride calibration solutions: From the vinyl chloride stock solution calibration solutions are prepared containing e.g. 1290, 516 and 51.6 µg vinyl chloride in 10 ml carbon disulfide. 3 Sampling and sample preparation An activated carbon tube is opened and connected to the pump. During working hours the pump and tube are carried by a person or used in a stationary position. For two hours air is drawn through the sampling tube at a flow rate of 10 L/h. The loaded activated carbon is transferred to a 10 ml volumetric flask. 10 ml carbon disulfide is added and the solution is allowed to stand with occasional shaking for 30 minutes (sample solution).

7 199 AirMonitoringMethods,Vol.4 Vinyl chloride 4 Operating conditions for gas chromatography The method was characterized under the following experimental conditions: Apparatus: Perkin-Elmer gas chromatograph F 22, equipped with flame ionization detector Column: Material: Length: Glass column 3 m Internal diameter: 2.2 mm Stationary phase: 10 % DC 200 on Chromosorb W/HP µm or 25 m Marlophen glass capillary column Temperatures: Injector block: 200 C Column: Detector: 75 C isothermal 200 C Carrier gas: Nitrogen: 20 ml/min 5 Analytical determination After half an hour 1 µl of the sample solution (cf. Sect. 3) is injected into the injector block of the gas chromatograph. The area or the height of the vinyl chloride peak is determined. 6 Calculation of the analytical result 6.1 Calibration curve Depending on the concentration of the vinyl chloride calibration solution (cf. Sect. 2.3) 1 4 µl is injected into the gas chromatograph. Using the baseline method the peak areas or the peak heights are taken from the chromatogram and plotted against the vinyl chloride weights. 6.2 Evaluation of the results by the external standard method By means of the calibration curve the vinyl chloride weight in the sample solution is determined from the peak areas or the peak heights taking the blank value into consideration.

8 Analytical Methods AirMonitoringMethods,Vol The vinyl chloride concentration in the air sample c w in mg/m 3 is calculated as follows: c w w V For the calculation of the concentration by volume c v in ml/m 3 (ppm) at 20 C and 1013 hpa from c w : c v = 0.38 c w (2) Legend: c w Vinyl chloride concentration by weight in the air sample in mg/m 3 c v Vinyl chloride concentration by volume in the air sample in ml/m 3 w Vinyl chloride weight in the sample solution in µg, taken from the calibration curve and corrected by the blank value V Air sample volume in L (1) 7 Reliability of the method 7.1 Accuracy The relative standard deviation is ± 4%, determined from calibration measurements by loading of activated carbon tubes with vinyl chloride calibration solutions. 7.2 Quantification limit Under the experimental conditions the absolute quantification limit is 1 ng vinyl chloride. The relative quantification limit is 0.2 ml/m 3 (ppm). This corresponds to 0.5 mg/m 3 vinyl chloride for 20 L air sample, 10 ml sample solution and 1 µl injection volume. 8 Discussion The selection of the chromatography column depends on any additional measuring and interfering components.

9 201 AirMonitoringMethods,Vol.4 Vinyl chloride 9 Manufacturers Pump: e.g. Compur Electronic GmbH, München, Du Pont Instruments, Supplier in Germany: DEHA-Haan & Wittmer GmbH, Friolzheim Activated carbon tubes: e. g Drägerwerk AG, Lübeck Gas chromatograph: e.g. Carlo Erba Supplier in Germany: Erba Science, Hofheim/Ts., Bodenseewerk Perkin-Elmer & Co GmbH, Überlingen, Siemens AG, Karlsruhe, Varian GmbH, Darmstadt

10 Analytical Methods AirMonitoringMethods,Vol Sampling with a pump and adsorption on activated carbon, headspace gas chromatography after desorption This method permits the determination of vinyl chloride concentrations in working areas averaged over the sampling time after personal or stationary sampling. Principle: Technical data: With a pump a measured air volume is drawn through activated carbon. The adsorbed vinyl chloride is desorbed with dimethylacetamide/water (3:1) and analysed by headspace gas chromatography (HSGC). Evaluation is carried out according to the internal standard method (diethyl ether) using a flame ionization detector (FID). Quantification limit: relative: 0.03 mg/m 3 vinyl chloride for 7 L air sample. This corresponds to 0.2 µg vinyl chloride per activated carbon tube. Selectivity: Advantages: Disadvantages: Apparatus: As a result of interfering substances the concentrations may be too high. Interference can generally be eliminated by selecting another column. Personal and selective measurements are possible. No indication of peak concentrations, only one measurement possible. Pump equipped with gas meter or flow meter, activated carbon tubes, headspace gas chromatograph equipped with flame ionization detector (FID).

11 203 AirMonitoringMethods,Vol.4 Vinyl chloride Detailed description of the method Contents 1 Equipment, chemicals and solutions 1.1 Equipment 1.2 Chemicals 1.3 Solutions 2 Sampling 3 Analytical determination 3.1 Sample preparation 3.2 Operating conditions for gas chromatography 4 Evaluation 4.1 Calibration Implementation Determination of the amount of vinyl chloride in the calibration gas 4.2 Calculation of the analytical result 5 Reliability of the method 5.1 Accuracy 5.2 Quantification limit 5.3 Selectivity 5.4 Recovery 6 Discussion 7 References 1 Equipment, chemicals and solutions 1.1 Equipment For sampling and sample preparation: Pump with gas meter or flow meter (e. g. Compur 4903 or Du Pont S 30) Gas collection vessel (gas sampling tube containing some pieces of polytetrafluoroethylene (PTFE) tubing) about 300 ml with PTFE screw taps and side ports Manometer 20 ml Injection bottles (effective volume about 23 ml) with PTFE coated septa and aluminium caps Adsorption tubes with activated carbon, (standardized, consisting of two sections containing about 100 and 50 mg activated carbon which are separated by a porous polymeric material) Caps for the opened activated carbon tubes

12 Analytical Methods AirMonitoringMethods,Vol For analysis: Headspace gas chromatograph with flame ionization detector and sample thermostat with constant mode setting, autosampling from the headspace and back flush (e. g. Sigma 2000 with HS 100 from Perkin Elmer) Continuous-line recorder and/or laboratory data system (e.g A from Hewlett- Packard) Microliter syringes, gastight (e.g. Hamilton 1001 LTN, 1750 N and 1725 N) 1.2 Chemicals Vinyl chloride, purity > 99,95% N,N-Dimethylacetamide, analytical grade Diethyl ether, analytical grade (internal standard) De-ionized water Gases for gas chromatography: Nitrogen, hydrogen, synthetic air 1.3 Solutions Desorption solution: Dimethylacetamide/water mixture (3:1). 250 ml water is added to a 1 L volumetric flask, the flask is filled to the mark with dimethylacetamide and shaken. Diethyl ether stock solution: Solution of 100 mg diethyl ether in 100 ml desorption solution. 5 ml of the desorption solution is added to a 25 ml volumetric flask and about 25 mg diethyl ether weighed to the nearest 0.1 mg is added. The flask is filled to the mark with desorption solution. Diethyl ether standard solution: Solution of 0.5 mg diethyl ether in 100 ml desorption solution. 500 µl of the stock solution is added to a 100 ml volumetric flask which is filled to the mark with desorption solution. Vinyl chloride-diethyl ether stock solution: Solution of 250 mg vinyl chloride and 250 mg diethyl ether in 100 ml desorption solution. 20 ml of the desorption solution are pipetted into a 20 ml injection bottle. 50 mg diethyl ether are weighed to the nearest 0.1 mg and added. Then the injection bottle is closed with septum and crimp cap. With a gastight syringe 20 ml vinyl chloride (about 50 mg) are injected through the septum and weighed to the nearest 0.1 mg. Note! The tip of the needle must not dip into the desorption solution. Vinyl chloride-diethyl ether calibration solution: Solution of 500 µg vinyl chloride and 500 µg diethyl ether in 100 ml desorption solution. 100 µl of the stock solution is added to a 50 ml volumetric flask which is filled to the mark with desorption solution.

13 205 AirMonitoringMethods,Vol.4 Vinyl chloride Vinyl chloride calibration gas: Gas mixture with 4 to 5 µg vinyl chloride in 1 ml gas (at 100 kpa). In an evacuated 300 ml gas sampling tube connected to a manometer, 1 ml vinyl chloride is injected with a microliter syringe via the side ports. Then an overpressure of about 100 kpa is produced with nitrogen. The gas phase is mixed with the pieces of PTFE tubing contained in the gas sampling tube. When the calibration gas is removed, the pressure in the syringe returns to atmospheric pressure. 2 Sampling An activated carbon tube is opened and connected to the pump. During working hours the pump and tube are carried by a person or used in a stationary position. The flow rate can be varied between 0.6 L/h and 0.9 L/h. The air sample volume should not exceed 7 L. 3 Analytical determination 3.1 Sample preparation The contents of the loaded activated carbon tube (separated in measuring and safety zones) are each transferred to a 20 ml injection bottle and 1 ml diethyl ether standard solution is added. Then the injection bottle is sealed with septum and aluminium cap and thermostatted at 70 C in the sample thermostat for 30 minutes. Then samples are automatically taken from the headspace and a gas chromatogram is recorded (operating conditions cf. Sect. 3.2). To ensure that the desorption solution used, the diethyl ether and activated carbon do not contain any interfering impurities, a chromatogram is recorded with the contents of an unloaded activated carbon tube and 1 ml of the diethyl ether standard solution as described for the analytical determination. 3.2 Operating conditions for gas chromatography The method was characterized under the following experimental conditions: Apparatus: Perkin-Elmer Sigma 2000 gas chromatograph with temperature programmer and flame ionization detector, HS 100 headspace autosampler, back-flush device. Column: Material: Stainless steel tube Length: 3 m

14 Analytical Methods AirMonitoringMethods,Vol Internal diameter: Stationary phase: 2.4 mm Carbopack B (60 80 mesh) coated with 1% SP 1000 To apply the back-flush mode the column is divided into sections of 1.0 and 2.0 m. The shorter piece is on the injection side. Carrier gas: Nitrogen: 40 ml/min, 380 kpa Back flush: 270 kpa Detector gases: Hydrogen: 42 ml/min, 250 kpa Synthetic air: 250 ml/min, 270 kpa Temperatures: Column: Temperature controlled from 80 C to 90 C at 3 C/min, then from 90 C to 170 C at 5 C/min Detector: 220 C Sample thermostat: 70 C Transfer line: 100 C Times: Sample thermostatting: 30 min Injection: 0.15 min Pressure stabilization: 0.5 min Back-flush after: 10 min Analysis cycle: 21 min 4 Evaluation Quantitative evaluation is carried out according to the internal standard method via the peak areas. 4.1 Calibration Two calibration factors are determined for quantification and evaluation of the method. Calibration factor f c : Vinyl chloride-diethyl ether and desorption solution without activated carbon (2-phase equilibrium). The calibration factor f c serves to determine the exact amount of vinyl chloride in the calibration gas; f c is determined using the vinyl chloride/diethyl ether calibration solution. Calibration factor f a : Vinyl chloride-diethyl ether with 100 mg activated carbon and desorption solution (3- phase equilibrium). The calibration factor f a is needed to determine the weight of vinyl chloride on the activated carbon. A known weight of vinyl chloride from the calibration gas (cf. Sect. 1.3) is adsorbed on the activated carbon and then desorbed with the diethyl ether standard solution.

15 207 AirMonitoringMethods,Vol.4 Vinyl chloride Implementation Calibration factor f c (2-phase equilibrium): An injection bottle is sealed with septum and aluminium cap, then 1 ml of the vinyl chloride-diethyl ether calibration solution is added and the solution analysed as described in Sect Using the vinyl chloride and diethyl ether peak areas, the calibration factor f c is calculated according to the formula (1): A. st w1 fc (1) A. w 1 st Legend: f c Calibration factor of vinyl chloride A 1 Peak area of vinyl chloride from the calibration solution A st Peak area of diethyl ether from the calibration solution w 1 Weight of vinyl chloride in mg in 1 ml of the calibration solution w st Weight of diethyl ether in mg in 1 ml of the calibration solution Calibration factor f a (3-phase equilibrium): An injection bottle is filled with the 100 mg section of an activated carbon tube. This is sealed with septum and aluminium cap and 1 ml of the vinyl chloride calibration solution and 1 ml of the diethyl ether standard solution are added. The solution is then analysed according to Sect 3.2. Using the vinyl chloride and diethyl ether peak areas, the calibration factor f a is calculated according to the formula (2): A. st w2 fa (2) A. w 2 st Legend: f a Calibration factor of vinyl chloride in 3-phase equilibrium A 2 Peak area of vinyl chloride from the calibration gas A st Peak area of diethyl ether from the standard solution w 2 Weight of vinyl chloride in mg in 1 ml of the calibration gas w st Weight of diethyl ether in mg in 1 ml of the diethyl ether standard solution Determination of the amount of vinyl chloride in the calibration gas An injection bottle is sealed with septum and aluminium cap. 1 ml of the vinyl chloride calibration gas and 1 ml of the diethyl ether standard solution are added. The solution is then analysed according to Sect The vinyl chloride weight in µg in 1 ml of the calibration gas is calculated using diethyl ether as internal standard taking into consideration the calibration factor f c according to the formula (3):

16 Analytical Methods AirMonitoringMethods,Vol f c. A 3. w 3 A st w st (3) Legend: f c Calibration factor of vinyl chloride w 3 Weight of vinyl chloride in µg in 1 ml of the calibration gas w st Weight of diethyl ether in µg in 1 ml of the standard solution A 3 Peak area of vinyl chloride A st Peak area of diethyl ether 4.2 Calculation of the analytical result The concentration by weight of vinyl chloride in the sample air in mg/m 3 is calculated according to the formula (4): c w f. A. w A. V a i st st (4) For the calculation of the concentration by volume c v in ml/m 3 (ppm) at 20 C and 1013 hpa from c w : c v = c w (5) Legend: c w Vinyl chloride concentration by weight in the air sample in mg/m 3 c v Vinyl chloride concentration by volume in the air sample in ml/m 3 f a Calibration factor of vinyl chloride in 3-phase equilibrium A i A st Peak area of vinyl chloride Peak area of diethyl ether w st V Weight of diethyl ether in µg in 1 ml of the standard solution Air sample volume in L 5 Reliability of the method 5.1 Accuracy The relative standard deviation of the whole measuring procedure was determined from 14 measurements to be about 3% for the weights 0.2 µg to 4.5 µg vinyl chloride per activated carbon tube (100 mg).

17 209 AirMonitoringMethods,Vol.4 Vinyl chloride 5.2 Quantification limit The relative quantification limit is 0.03 mg/m 3 vinyl chloride for a 7 L air sample; this corresponds to about 0.2 µg vinyl chloride per activated carbon tube using 100 mg activated carbon and 1 ml of the diethyl ether standard solution. 5.3 Selectivity The selectivity of the procedure depends above all on the type of chromatography column used. In practice the described column has proved suitable. When there are interfering components a column with other separation characteristics must be selected. 5.4 Recovery Determination of the recovery rate was carried out according to the operating conditions named in Sect Mixtures of calibration gases were produced according to Sect. 1.3 and known volumes injected directly into the activated carbon tubes with a gastight syringe. Then 3 to 4 L synthetic air was drawn through the activated carbon tubes with a pump. In the range from 0.2 µg to 4.5 µg vinyl chloride recovery was found to be > 95 %. 6 Discussion The whole procedure has been tested under laboratory conditions up to vinyl chloride concentrations of max. 140 mg/m 3. The shelf life of adsorbed vinyl chloride is limited. After 2 to 3 days migration begins from the measuring section to the safety section. With separate preparation of the safety section it must be borne in mind that the calibration factor for the safety section is slightly smaller than that for the measuring section. The calibration factor for the measuring section can, however, be used to evaluate the results. If more than 25 % of the vinyl chloride is found in the safety section, sampling must be repeated. For the shelf life of loaded tubes cf. e.g. [1]. If the safety section of the activated carbon tube is quantitatively evaluated, it must be borne in mind that this only contains 50 mg activated carbon and the calibration factor f a cannot be used. If needed this calibration factor must be determined as described in Section 4.1. Under the operating conditions described it is possible to determine 1,2-dichloroethane (procedure ZH 1/ E) simultaneously with vinyl chloride.

18 Analytical Methods AirMonitoringMethods,Vol References [1] Vandendriessche S, Griepink B (1989) The certification of benzene, toluene and m-xylene sorbed on tenax in tubes-crm 112. Community Bureau of Reference, Commission of the European Communities, EUR

Method for the determination of 1,3-butadiene

Method for the determination of 1,3-butadiene Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Methods for the determination of ethylene oxide

Methods for the determination of ethylene oxide Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Method for the determination of dimethyl sulfate

Method for the determination of dimethyl sulfate German Social Accident Insurance Deutsche Gesetzliche Unfallversicherung Analytical Subcommittee of the Chemistry Board of Experts* Carcinogenic substances Order number: BGI 505-7-05 Established methods:

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory. VINYL CHLORIDE Method no.: 04 Matrix: Air Target concentration: 3 1 ppm (2.5 mg/m ) (OSHA PEL) Procedure: Collection on charcoal (two-tubes in-series), desorption with carbon disulfide, analysis by gas

More information

HYDROCARBONS, AROMATIC 1501

HYDROCARBONS, AROMATIC 1501 HYDROCARBONS, AROMATIC 1501 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1501, Issue 3 EVALUATION: Full Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2

More information

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003 ALCOHOLS IV 1403 (1) HOCH 2 CH 2 OCH 3 MW: (1) 76.09 CAS: (1) 109-86-4 RTECS: (1) KL5775000 (2) HOCH 2 CH 2 OCH 2 CH 3 (2) 90.12 (2) 110-80-5 (2) KK8050000 (3) HOCH 2 CH 2 O(CH 2 ) 3 CH 3 (3) 118.17 (3)

More information

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin OIVMABS15 Anethole. Gas chromatograhpy determination of transanethole in Type II method SCOPE This method is suitable for the determination of transanethole in aniseedflavoured spirit drinks using capillary

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

Hexane isomers (except n-hexane) and methylcyclopentane

Hexane isomers (except n-hexane) and methylcyclopentane 1 Hexane isomers (except n-hexane) and methylcyclopentane Method number 1 Application Air analysis Analytical principle Gas chromatography Completed in July 2010 Summary This analytical method permits

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO) Date 02/24/96 Page 1 1.1. This method is designed to measure the concentration

More information

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Turk J Chem 24 (2000), 131 139. c TÜBİTAK Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Naciye KILIÇ University of Uludağ,

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

BRIEFING. (EM2: K. Moore.) RTS C Add the following: Methylcellulose

BRIEFING. (EM2: K. Moore.) RTS C Add the following: Methylcellulose BRIEFING Methylcellulose. The Japanese Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the compendial standards for the Methylcellulose monograph, as part of the process

More information

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC Aysel T. ATIMTAY Middle East Technical University, Environmental Engineering Department, 06531 Ankara Turkey (aatimtay@metu.edu.tr)

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances BRIEFING Saccharin, NF 22 page 2825 and page 1711 of PF 29(5) [Sept. Oct. 2003]. The United States Pharmacopeia is the coordinating pharmacopeia for the international harmonization of the compendial standards

More information

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

1. A gas chromatograph equipped with flame ionization detection and a computing integrator. HYDRO.02-1 HYDROXYETHYL SUBSTITUTION LEVEL PRINCIPLE SCOPE The hydroxyethyl group is cleaved from the starch by hydrogen iodide in the presence of adipic acid and an organic solvent containing a known

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Analysis of BTEX in Natural Water with SPME

Analysis of BTEX in Natural Water with SPME Analysis of BTEX in Natural Water with SPME Application Note Environmental Author Giordano Vassalli Sezione Protezione Aria Introduction Benzene, toluene, ethylbenzene and xylene (BTEX) isomers are monocyclic

More information

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of phenols in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicable to the quantitative determination of

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

METHYLTIN CHLORIDES 5526

METHYLTIN CHLORIDES 5526 METHYLTIN CHLORIDES 5526 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5526, Issue 1 EVALUATION: FULL Issue 1: 15 March 2003 OSHA : See Table 1 NIOSH: See Table 1 ACGIH: See Table 1

More information

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description WHO/SIT/19.R4 TEMEPHOS TECHNICAL TECHNICAL TEMEPHOS 1. Specification 1.1 Description Full specification WHO/SIT/19.R4 Revised 10 December 1999 The material shall consist of temephos together with related

More information

PYRIPROXYFEN TECHNICAL

PYRIPROXYFEN TECHNICAL WHO/IS/TC/715/2001 TECHNICAL TECHNICAL 1. Specification 1.1 Description Interim specification WHO/IS/TC/715/2001 The material shall consist of pyriproxyfen together with related manufacturing impurities.

More information

TEST METHOD. Ethoxyl Substitution in ETHOCEL TM Ethylcellulose Polymers. 1. Scope. 2. Principle. 3. Safety

TEST METHOD. Ethoxyl Substitution in ETHOCEL TM Ethylcellulose Polymers. 1. Scope. 2. Principle. 3. Safety TEST METHOD Effective: 15 July 2010 Supersedes: DOWM 100931-ME91A DOWM 100931-E10B Ethoxyl Substitution in ETHOCEL TM Ethylcellulose Polymers 1. Scope This method is applicable to the determination of

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

FORMALDEHYDE by GC 2541

FORMALDEHYDE by GC 2541 FORMALDEHYDE by GC 2541 H 2 C=0 MW: 30.03 CAS: 50-00-0 RTECS: LP8925000 METHOD: 2541, Issue 2 EVALUATION: PARTIAL Issue 1: 15 May 1989 Issue 2: 15 August 1994 OSHA : NIOSH: ACGIH: 0.75 ppm; 2 ppm STEL

More information

Method for Characterization of Gum Rosin by Capillary Gas Chromatography

Method for Characterization of Gum Rosin by Capillary Gas Chromatography PCA TEST METHOD PCTM 27 Published 1/1 2004 1. Scope Method for Characterization of Gum Rosin by Capillary Gas Chromatography 1.1 These test methods cover the determination of the amounts of certain rosin

More information

637. Thiamethoxam. HPLC method

637. Thiamethoxam. HPLC method 637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB5009.28-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA GB 5009.28-2016

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Application Note. Abstract. Authors. Introduction

Application Note. Abstract. Authors. Introduction Multiple Headspace Extraction for the Quantitative Determination of Residual Monomer and Solvents in Polystyrene Pellets Using the Agilent 7697A Headspace Sampler Application Note Authors Wenwen Shen and

More information

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO)

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO) Page 1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY 1. SCOPE AND APPLICATION: METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO) 1.1. This method is designed to measure the concentration of diesel

More information

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Chemistry Instrumental Analysis Lecture 31. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery

More information

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD Kamarruddin ASRI 1, Robert A. ANDERSON 2 1 Department of Chemistry, Jalan

More information

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations OUTLINE OF METHOD This gas chromatography/mass spectrometry (GC/MS) method provides

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

ORGANOPHOSPHORUS PESTICIDES 5600

ORGANOPHOSPHORUS PESTICIDES 5600 ORGANOPHOSPHORUS PESTICIDES 5600 Formula: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 5600, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : Table 2 NIOSH: Table 2 ACGIH: Table 2 PROPERTIES:

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

http://www.sis.se http://www.sis.se http://www.sis.se http://www.sis.se http://www.sis.se Provläsningsexemplar / Preview SVENSK STANDARD SS-ISO 22155:2006 Fastställd 2006-11-06 Utgåva 1 Markundersökingar

More information

HYDROGEN CYANIDE 6017

HYDROGEN CYANIDE 6017 HYDROGEN CYANIDE 6017 HCN MW: 27.03 CAS: 74-90-8 RTECS: MW6825000 METHOD: 6017, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA: 10 ppm (skim) NIOSH: STEL 4.7 ppm ACGIH: C 10 ppm (skin) (1 ppm

More information

IMPCA METHANOL REFERENCE SPECIFICATIONS

IMPCA METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION IMPCA METHANOL REFERENCE SPECIFICATIONS a.i.s.b.l. International Methanol Producers & Consumers Association i.v.z.w. Avenue de Tervueren 270

More information

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11890-2 Third edition 2013-03-15 Paints and varnishes Determination of volatile organic compound (VOC) content Part 2: Gas-chromatographic method

More information

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS 1. Scope This method is applicable for determining the presence

More information

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous

More information

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009) Method OIV-MA-AS315-16 Type IV method Determination of releasable in wine by cork stoppers (Resolution OIV-Oeno 296/2009) 1 SCOPE: The method of determination of releasable (TCA) by cork stoppers measures

More information

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION. APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Kira. Yang PerkinElmer, Inc. Shanghai, China Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS Introduction

More information

ChromTech options for PAL systems

ChromTech options for PAL systems ChromTech options for PAL systems SPDE features full automation for dynamic sample extraction, desorption and analysis using the CTC Combi PAL. Most of our customers use SPDE for food and flavor research,

More information

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) 683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and

More information

H 3 CO H 3 CO S CH 3

H 3 CO H 3 CO S CH 3 FENITROTHION 35 H 3 CO P H 3 CO S O CH 3 NO 2 ISO common name Chemical name Fenitrothion O,O-Dimethyl O-4-nitro-m-tolyl phosphorothioate (IUPAC) O,O-Dimethyl O-(3-methyl-4-nitrophenyl)- phosphorothioate

More information

METHANOL REFERENCE SPECIFICATIONS

METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION Avenue de Tervuren 149 1150 Brussels, Belgium Tel : (32/2) 7418683 Fax : (32/2) 7418684 e-mail : impca@epca.be IMPCA METHANOL REFERENCE SPECIFICATIONS

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

ACETIC AND FORMIC ACIDS IN WORKPLACE ATMOSPHERES ppm for Acetic Acid, CH 3 COOH

ACETIC AND FORMIC ACIDS IN WORKPLACE ATMOSPHERES ppm for Acetic Acid, CH 3 COOH ACETIC AND FORMIC ACIDS IN WORKPLACE ATMOSPHERES Method no: ID-186SG Matrix: Air OSHA standard: 5.0 ppm for Formic Acid, HCOOH 10.0 ppm for Acetic Acid, CH 3 COOH Collection procedure: A known volume of

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

IMPCA METHANOL REFERENCE SPECIFICATIONS

IMPCA METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION IMPCA METHANOL REFERENCE SPECIFICATIONS a.i.s.b.l. International Methanol Producers & Consumers Association i.v.z.w. Avenue de Tervueren 270

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Tex-620-J, Determining Chloride and Sulfate Contents in Soil Contents in Soil Contents: Section 1 Overview...2 Section 2 Sample Preparation...3 Section 3 Ion Chromatography Method...5 Section 4 Wet Chemical Method...9 Section 5 Archived Versions...15 Texas Department

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Foods, Flavors & Fragrances Applications A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates By Corby Hilliard; Amanda Rigdon; William Schroeder*,

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T18204.26-2000 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA GB/T 18204.26-2000 Methods for determination

More information

Ondansetron Hydrochloride Tablets

Ondansetron Hydrochloride Tablets Ondansetron Hydrochloride Tablets Dissolution Perform the test with 1 tablet of Ondansetron Hydrochloride Tablets at 50 revolutions per minute according to the Paddle method, using 900 ml of water

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

METHOTREXATE mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for

METHOTREXATE mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for METHOTREXATE Matrix: Air Control no.: T-PV2146-01-8804-CH Target Concentration: 0.04 mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for methotrexate. Procedure: Samples are collected by drawing

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Method of determination of phtalates in spirituous beverages by gaschromatography/mass OIV-MA-BS-33 Method of determination of phtalates in spirituous beverages by gas-chromatography/mass Type IV method 1. SCOPE OF APPLICATION This method applies to the detection and assay of some phthalates

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

Method - Determination of aromatic amines in mainstream cigarette smoke

Method - Determination of aromatic amines in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of aromatic amines in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicaple to quantitative determination

More information

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25) Gas Chromatography NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25) The Non-Methane Organic Compounds (NMOC) Analyzer is a gas chromatograph configured for analyzing gaseous samples for total organic

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2), ISSN: 0973-4945; CODEN ECJHAO E-Journal of Chemistry http://www.e-journals.net 2010, 7(2), 629-635 Low-level Determination of Residual Methyl Methane Sulfonate and Ethyl Methane Sulfonate in Pharmaceuticals

More information

Supporting Information

Supporting Information Supporting Information A Teflon microreactor with integrated piezoelectric actuator to handle solid forming reactions Simon Kuhn, a Timothy oёl, b Lei Gu, a Patrick L. eider a and Klavs F. Jensen a* a

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP 1.0 SCOPE AND APPLICATION 1.1 This method is suitable for the rigorous cleanup of sample extracts prior to analysis for polychlorinated biphenyls. This method

More information

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1 Designation: D 4735 02 Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1 This standard is issued under the fixed designation D 4735; the number immediately

More information

2401 Gas (liquid) Chromatography

2401 Gas (liquid) Chromatography 2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic

More information

NCASI METHOD CI/SG/PULP-94.02

NCASI METHOD CI/SG/PULP-94.02 NCASI METHOD CI/SG/PULP-94.02 CHILLED IMPINGER/SILICA GEL TUBE TEST METHOD AT PULP MILL SOURCES FOR METHANOL, ACETONE, ACETALDEHYDE, METHYL ETHYL KETONE AND FORMALDEHYDE NCASI Southern Regional Center

More information

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Contact: Oliver el-atma Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Germany (OMCL

More information

CIPAC. CIPAC Free relevant impurities methods:

CIPAC. CIPAC Free relevant impurities methods: CIPAC COLLABORATIVE INTERNATIONAL PESTICIDES ANALYTICAL COUNCIL LIMITED Commission Internationale des Méthodes d'analyse des Pesticides (CIMAP) CIPAC Free relevant impurities methods: Methods for relevant

More information

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization

More information

*Correspondence to:

*Correspondence to: Supporting Information for Carbonate-promoted hydrogenation of carbon dioxide to multi-carbon carboxylates Aanindeeta Banerjee 1 and Matthew W. Kanan 1 * 1 Department of Chemistry, Stanford University,

More information

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5 Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 66 No. 6 pp. 611ñ615, 2009 ISSN 0001-6837 Polish Pharmaceutical Society DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE

More information

4023 Synthesis of cyclopentanone-2-carboxylic acid ethyl ester from adipic acid diethyl ester

4023 Synthesis of cyclopentanone-2-carboxylic acid ethyl ester from adipic acid diethyl ester NP 4023 Synthesis of cyclopentanone-2-carboxylic acid ethyl ester from adipic acid diethyl ester NaEt C 10 H 18 4 Na C 2 H 6 C 8 H 12 3 (202.2) (23.0) (46.1) (156.2) Classification Reaction types and substance

More information

Chapter 11 Conventional Gas Chromatography

Chapter 11 Conventional Gas Chromatography Chapter 11 Conventional Gas Chromatography Gas Chromatography GC is the first instrumental chromatographic method developed commercially It is relatively easy to introduce a stable flow and pressure for

More information