TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

Size: px
Start display at page:

Download "TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air"

Transcription

1 Application TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air Summary This Application Note describes validation of a thermal desorption system for on-line monitoring of a complex mix of volatile and very volatile hydrocarbons ( ozone precursors ) in air. Background The presence of volatile hydrocarbons in urban atmospheres is believed to contribute to the formation of ground-level ozone, one of the main constituents of urban smog. The compounds of interest range in volatility from acetylene (ethyne) to trimethylbenzene, and are generally referred to as 'ozone precursors'. Vehicle emissions are thought to be the main source of these compounds. Recent European and US regulations 1,2 require round-the-clock monitoring of target species in all major urban centers to establish and monitor the link between periods of high traffic density and high pollution levels (key compounds include benzene, toluene, xylene and buta-1,3-diene). Continuous real-time monitoring provides an insight into emission episodes from local industry and can be used to monitor the effect of weather conditions such as wind direction, precipitation and temperature inversion. Introduction The UNITY 2 Air Server is a cost-effective system for round-the-clock speciated measurement of multiple trace-level volatile organic compounds (VOCs) in air or pure gases. It combines automated, controlled-flow sampling with cryogen-free concentration technology. The system connects to standard GC and GC/MS technology, and is designed for unattended operation in remote field locations. In summary, sample air is pulled directly onto an electrically-cooled, sorbent-packed focusing trap at a controlled flow rate. No liquid cryogen is required. An optional membrane dryer selectively eliminates water and other low-molecular-weight polar components, reducing the risk of interference from unknown species and facilitating the use of FID rather than MS detection. Sampling flows are controlled by an electronic mass flow controller and pump, located downstream of the focusing trap to avoid contamination. All sampling parameters are selected by the user and monitored by system software as an integral part of the analytical method. After sample collection, the flow path is purged with carrier gas to prevent carryover and eliminate oxygen from the focusing trap. The trap then heats rapidly, at rates approaching 100 C/min, to inject retained analytes into the capillary column as a highly concentrated band of vapour. This transfer may be performed splitless for maximum sensitivity. Once the trap has desorbed, it re-cools, reequilibrates to the trapping temperature and begins collection of the next sample while analysis of the previous sample is ongoing. The UNITY 2 Air Server also offers automatic interchange between a minimum of three sample channels (typically sample, reference and blank) for remote system calibration/validation as per user requirements. An 8-channel version is also available. This Application Note describes validation of the system for on-line monitoring of the 27 ozone precursors specified by European regulators, with the addition of 1,2,4-trimethylbenzene, 1,3,5-trimethylbenzene and isoprene (present in the gas standard employed). Table A1 (see Appendix) provides the full list of 30 compounds. Key factors taken into account when developing this method included the need for a system that could operate round-the-clock unattended, so cryogen-free operation was essential. The system also needed to allow hourly sample collection, with as much of the hour as possible dedicated to sampling. Finally, detection limits needed to be below 0.5 ppb (ideally 0.1 ppb).

2 Page 2 Defining the ideal focusing trap The design and packing of the trap is crucial for this application for many reasons. The narrow-bore design of the UNITY 2 trap (Figure 1) allows true splitless operation, allowing transfer of the whole sample to the GC system, giving increased sensitivity for lowconcentration compounds. The narrow-bore design also gives the best possible peak shape for early-eluting compounds (Figure 2). Quartz wool Weak sorbent Strong sorbent Strongest sorbent Retaining spring Figure 1: Design of the focusing trap. Note the narrow-bore outlet and the long sorbent bed. 1 Ethane 2 Ethene 3 Acetylene 4 Propane Propene 6 2-Methylpropane 7 Butane 8 trans-but-2-ene Figure 2: Chromatogram showing a splitless injection of 500 ml of the calibration standard. 6 8 Experimental (a) Breakthrough tests TD: Prepurge time: 1 min Sampling rate: 25 ml/min Sampling time: Various Cold trap: U-T17O3P-2S (ozone precursor) Cold trap low: 30 C Cold trap high: 325 C Cold trap hold: 5 min Flow path temp.: 80 C GC: Column: Agilent J&W GS-GasPro, 30 m 0.32 mm GC program: 40 C (hold 6 min), then 20 C/min to 200 C (hold 1 min) (b) Two-column Deans switch GC system TD: Prepurge time: 2 min Sampling rate: 25 ml/min Sampling time: Various Cold trap: U-T17O3P-2S (ozone precursor) Cold trap low: 30 C Cold trap high: 325 C Cold trap hold: 5 min Flow path temp.: 80 C GC/FID: Inlet pressure: 44.9 psi Midpoint pressure: psi GC oven program: 30 C (hold 12 min), then 5 C/min to 170 C, then 15 C/min to 200 C Switching time: 17.5 min FID temp.: 250 C Air flow: 400 ml/min H 2 flow: 40 ml/min The sorbents chosen are also crucial, as they permit increased retention of very volatile compounds, thereby increasing the sample volume taken, and allowing lower levels to be detected.

3 Page 3 Results and discussion (a) Breakthrough tests All the C 2 hydrocarbons are highly volatile species. Acetylene (ethyne) has a boiling point of 84 C and is the most difficult to trap, and to quantitatively retain it without liquid cryogen requires careful selection of the cold trap sorbent(s) and focusing temperature. Experiments to determine the limit of quantitative retention (the breakthrough volume) were performed on an ozone precursor cold trap using the ppb-level certified gas standard generated by the National Physical Laboratory (Teddington, UK). Steadily increasing volumes of the gas standard were introduced, and a graph of detector response against sampled volume plotted. Figure 3 is a plot of the peak area against the sampled volume of calibration gas for acetylene, and shows that even at 1.5 L of sampled gas, there is no deviation from linearity, i.e. there is negligible breakthrough. FID 1 Sample in FID 2 Column 3 ~0.5 m, 0.1 mm uncoated deactivated fused silica Column 1 60 m, 0.25 mm, 1 µm dimethylpolysiloxane Column 2 50 m, 0.32 mm, 8 µm alumina PLOT (Na 2 SO 4 wash) Switching point Pressure controller Figure 4: Schematic representation of the two-column ozone precursor analytical system. Acetylene peak area R 2 = Sample volume (ml) Figure 3: Plot of peak area against volume sampled for acetylene. Image courtesy of Ecole des Mines de Douai. (b) Two-column Deans switch GC system The column used in all previous tests was a 30 m 0.32 mm i.d. Agilent J&W GS-GasPro. Unfortunately, this column did not allow full resolution of the C 4 and C 5 components in the standard mixture. For better separation of these components, it was necessary to use a two-column system with a Deans switch (schematically illustrated in Figure 4). All the effluent from the focusing trap of the desorber was directed into the dimethylpolysiloxane column at 40 C. The C 2 C 5 compounds are unresolved on this column at this temperature, and are passed, via the Deans switch, into the alumina PLOT column as they elute from the primary column. For the first 17.5 minutes of the run, the Deans switch between the primary and secondary columns works in this direction. At this point, all the compounds eluting from the primary column are C 6 and higher, and are well-resolved. The Deans switch is thus activated, directing effluent from the primary column to the deactivated, uncoated fused silica link and to the other FID. From this point on, two chromatograms are produced in parallel; the C 6+ compounds from the dimethylpolysiloxane column (labelled FID 1), and the C 2 C 6 compounds from the alumina PLOT column (labelled FID 2). Figure 5 shows a plot for the calibration gas run through the two-column system, and Figure 6 shows the equivalent plot for the 56-component US EPA ozone precursor mix.

4 Page 4 1 Ethane 2 Ethene 3 Propane 4 Propene 5 2-Methylpropane 6 n-butane 7 Acetylene 8 trans-but-2-ene 9 But-1-ene 10 cis-but-2-ene 11 2-Methylbutane 12 Pentane 13 Buta-1,3-diene 14 trans-pent-2-ene 15 Pent-1-ene 16 2-Methylpentane 17 Isoprene 18 n-hexane 19 Benzene 25 m- and p-xylene 20 2,2,4-Trimethylpentane n-heptane 27 o-xylene 22 Toluene 28 1,3,5-Trimethylbenzene 23 Octane 29 1,2,4-Trimethylbenzene 24 Ethylbenzene 30 1,2,3-Trimethylbenzene Alumina PLOT column (FID 2) Dimethylpolysiloxane column (FID 1) Figure 5: Chromatograms for the ppb-level 30-component calibration standard (generated by NPL) on the two-column analytical system. All components had a concentration in the range ppb, with uncertainties for each of ±0.08 ppb. 1 Ethane 2 Ethene 3 Propane 4 Propene 5 2-Methylpropane 6 n-butane 7 Acetylene 8 trans-but-2-ene 9 But-1-ene 10 cis-but-2-ene 11 Cyclopentane 12 2-Methylbutane 13 Pentane 14 trans-pent-2-ene 15 Pent-1-ene 16 cis-pent-2-ene 17 2,2-Dimethylbutane 18 2,3-Dimethylbutane 19 2-Methylpentane 20 3-Methylpentane 21 Isoprene 22 2-Methylpent-1-ene 23 n-hexane 24 Methylcyclopentane 25 2,4-Dimethylpentane 26 Benzene 27 Cyclohexane 28 2-Methylhexane 29 2,3-Dimethylpentane 30 3-Methylhexane 31 2,2,4-Trimethylpentane 32 Heptane 33 Methylcyclohexane 34 2,3,4-Trimethylpentane 35 Toluene 36 2-Methylheptane 37 3-Methylheptane 38 Octane 39 Ethylbenzene 40 m- and p-xylene 41 Styrene 42 o-xylene 43 Nonane 44 Isopropylbenzene 45 Propylbenzene 46 m-ethyltoluene 47 p-ethyltoluene 48 1,3,5-Trimethylbenzene 49 o-ethyltoluene 50 1,2,4-Trimethylbenzene Decane 52 1,2,3-Trimethylbenzene 53 m-diethylbenzene 54 p-diethylbenzene 55 Undecane 56 Dodecane Figure 6: Chromatograms for the ppb-level 56-component calibration standard (from the US EPA) on the two-column analytical system. Data courtesy of ITC, Korea

5 Page 5 Reproducibility of retention time Due to the nature of the two-column set-up, the stability of the retention time can be related to the pressure balance in the system. With the added capability of electronic carrier gas control (ECC) on the UNITY 2 Air Server, stable retention times on both columns are achieved. Tables A1 and A2 (see Appendix) show the relative standard deviations for the compounds in the 30- component standard; those on the dimethylpolysiloxane column (C 6+ compounds) were close to 0.01%, and those in the alumina PLOT column (C 2 C 6 compounds) were in the range %. Reliability study multipoint calibrations In order for a system to be reliable, stringent tests must be carried out. Figure 7 shows five repeat analyses of 500 ml of the ozone precursor calibration standard on the two-column system. Alumina PLOT column (FID 2) Dimethylpolysiloxane column (FID 1) Figure 7: Repeat analysis of a splitless gas injection of a 500 ml sample of gas standard.

6 TDTS 16 Page 6 As can be seen, the repeat analyses appear identical, but to confirm this numerically, a 5-point calibration curve was generated, and five replicates at each level were analysed (see Table A3 in the Appendix). Nearly all the relative standard deviations were below 1%, demonstrating the reliability of the method. Each of the repeat calibrations was then plotted and a linear relationship derived; the corresponding R 2 values are provided in Table A4 (see Appendix). R 2 values of 0.99 or above were seen for more than 98% of the data, confirming the reliability of the calibration. French norm NF T specifies a method for testing calibration data and systems in accordance with a stringent goodness-of-fit test. We used this F-test method to compare a theoretical calibration against the five 5-point experimental calibrations, assuming the homogeneity of the variances. The critical value (F crit ) corresponds to the value that F must exceed for this assumption to be incorrect, and in this case it is at a risk level (α) of 1%. Table 1 shows the F-values for each of the compounds, and they are all less than F crit, so the calibration is acceptable. Detection/quantitation limits Signal-to-noise ratios for the 500 ml volume of standard using the two-column system are in the order of 250:1 for the light hydrocarbons at ~4 ppb, and 350:1 for compounds from n-butane upwards at ~4 ppb. Assuming a minimum detection signal of 3:1, approximate detection/quantitation limits for the 500 ml air samples were calculated as follows: C 2 C 3 hydrocarbons: Detection: 0.05 ppb; quantification: 0.1 ppb C 4+ hydrocarbons: Detection: 0.03 ppb; quantification: 0.06 ppb Compound Calculated F-value Ethane Ethene Propane Propene Methylpropane n-butane Acetylene trans-but-2-ene But-1-ene cis-but-2-ene Methylbutane Pentane Buta-1,3-diene trans-pent-2-ene Pent-1-ene Methylpentane Isoprene n-hexane n-heptane Benzene ,2,4-Trimethylpentane Octane Toluene Ethylbenzene m-xylene p-xylene o-xylene ,3,5-Trimethylbenzene ,2,4-Trimethylbenzene ,2,3-Trimethylbenzene Table 1: F-test values.

7 Page Concentration (ppb) Benzene Toluene Propane 2-Methylpropane n-butane Acetylene 2-Methylbutane Buta-1,3-diene :00 20:00 00:00 04:00 08:00 12:00 Time (GMT) 16:00 Figure 8: Diurnal profile of selected hydrocarbons, showing distinct differences between traffic-free and traffic-heavy periods. Suburban air monitoring Analyser performance for unattended field operation was evaluated in a forested semi-rural area, with high volumes of traffic during the working week and very few cars at weekends. Figure 8 shows a diurnal profile of various compounds from 16:40 on a Sunday through to a Monday evening the difference between the two periods is distinct. Most of the hydrocarbons were at low levels during the traffic-free periods, whilst rapid increases were seen during peak commuting hours. These data were collected at the end of the trial, and show that good analytical performance was maintained. Importantly, no user intervention was required during the trial, while detection limits, even in this relatively clean environment, were also satisfactory. Conclusions A reliable, semi-continuous, cryogen-free sampling and analysis system has been demonstrated for this challenging and complex mix of volatile and very volatile hydrocarbons, confirming Markes UNITY 2 Air Server system as an excellent choice for those wishing to quantify ozone precursors in ambient air. References 1. Council Directive 96/62/EC of 27 September 1996 on ambient air quality assessment and management, Council of the European Parliament, 1996; and its daughter directive, Directive 2000/69/EC of the European Parliament and of the Council of 16 November 2000 relating to limit values for benzene and carbon monoxide in ambient air, Council of the European Parliament, See for more details. 2. Provisions for attainment and maintenance of national ambient air quality standards (US Clean Air Act Amendments of 1990). See for more details. 3. Revision de la norme NF T90-210: Qualite de l eau. Protocole d évaluation initiale des performances d une méthode dans un laboratoire, Association Francaise de Normalisation, Available from gne.aspx?&nivctx=nelznelz1a10a101a107&ts= &CLE_ART=FA Applications were performed under the stated analytical conditions. Operation under different conditions, or with incompatible sample matrices, may impact the performance shown.

8 TDTS 16 Page 8 Appendix Compound Retention time RSD (%) (n = 15) Benzene ,2,4-Trimethylpentane n-heptane Toluene Octane Ethylbenzene m- and p-xylene o-xylene ,3,5-Trimethylbenzene ,2,4-Trimethylbenzene ,2,3-Trimethylbenzene Table A1: Retention time stability on the dimethylpolysiloxane column (C 6+ compounds). Compound Retention time RSD (%) (n = 15) Ethane Ethene Propane Propene Methylpropane n-butane Acetylene trans-but-2-ene But-1-ene cis-but-2-ene Methylbutane Pentane Buta-1,3-diene trans-pent-2-ene Pent-1-ene Methylpentane Isoprene n-hexane Table A2: Retention time stability on the alumina PLOT column (C 2 C 6 compounds).

9 TDTS 16 Page 9 Response RSD (%) (n = 5) Compound 250 ml 375 ml 500 ml 625 ml 750 ml Ethane Ethene Propane Propene Methylpropane n-butane Acetylene trans-but-2-ene But-1-ene cis-but-2-ene Methylbutane Pentane Buta-1,3-diene trans-pent-2-ene Pent-1-ene Methylpentane Isoprene n-hexane Benzene ,2,4-Trimethylpentane n-heptane Toluene Octane Ethylbenzene m- and p-xylene o-xylene ,3,5-Trimethylbenzene ,2,4-Trimethylbenzene ,2,3-Trimethylbenzene Table A3: Reproducibility of the 30-component gas standard for five sample volumes.

10 Page 10 Compound Calibration 1 Calibration 2 Calibration 3 Calibration 4 Calibration 5 Ethane Ethene Propane Propene Methylpropane n-butane Acetylene trans-but-2-ene But-1-ene cis-but-2-ene Methylbutane Pentane Buta-1,3-diene trans-pent-2-ene Pent-1-ene Methylpentane Isoprene n-hexane Benzene ,2,4-Trimethylpentane n-heptane Toluene Octane Ethylbenzene m- and p-xylene o-xylene ,3,5-Trimethylbenzene ,2,4-Trimethylbenzene ,2,3-Trimethylbenzene Table A4. R 2 values for five repeats of the five-point calibration. Version 12 June 2012

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

Site Specific Conditional Sampler Garfield County, Colorado. VOC Data Summaries. Prepared for

Site Specific Conditional Sampler Garfield County, Colorado. VOC Data Summaries. Prepared for Site Specific Conditional Sampler Garfield County, Colorado VOC Data Summaries Prepared for Garfield County Public Health 195 West 14 th Street Rifle, Colorado 81650 Prepared by 1901 Sharp Point Dr., Suite

More information

J. Li et al. Correspondence to: S. D. Xie

J. Li et al. Correspondence to: S. D. Xie Supplement of Atmos. Chem. Phys., 15, 7945 7959, 2015 http://www.atmos-chem-phys.net/15/7945/2015/ doi:10.5194/acp-15-7945-2015-supplement Author(s) 2015. CC Attribution 3.0 License. Supplement of Characterization

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors by FID from WATER or AIR 0 / 325 µg/m 3 or higher

CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors by FID from WATER or AIR 0 / 325 µg/m 3 or higher CASE STUDY : VOC analysis in ambient air by FID dectection - Complete system airmozone Page 1 of 13 E X P E R T S I N G A S A N A L Y S I S CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors

More information

PERCH Air Quality Study. Quarterly Report. February 7, 2004

PERCH Air Quality Study. Quarterly Report. February 7, 2004 PERCH Air Quality Study Quarterly Report February 7, 2004 Submitted to: Professor Ranga Rao Center for Environmental Diagnostics and Bioremediation University of West Florida 11000 University Parkway Pensacola,

More information

The Analysis of Ozone Precursors by AutoGC

The Analysis of Ozone Precursors by AutoGC The Analysis of Ozone Precursors by AutoGC The Role of Calibration and Quality Control Strategies in Data Management for Fully Automated Thermal Desorption GC FID Systems Orsat, LLC 1416 Southmore Ave

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

Identification of Hydrocarbons in Ambient Air

Identification of Hydrocarbons in Ambient Air Identification of Hydrocarbons in Ambient Air Randall Bramston-Cook and Edward Bramston-Cook Lotus Consulting 5781 Campo Walk, Long Beach, Ca 90803 310/569-0128 ebramstoncook@msn.com Presented at 2010

More information

Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM

Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM To: Lorinda Gardner, Director, R15 Date: Carlos Rubinstein, Texas Border Area Director From: Valerie E. Meyers, Ph.D. Toxicology Section,

More information

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International

More information

GAW Expert Workshop VOC, C. Plass-Dülmer et al., FEHp-B 1

GAW Expert Workshop VOC, C. Plass-Dülmer et al., FEHp-B 1 ACTRIS-GAW: Uncertainty Evaluation in VOC Analysis Draft VOC Measurement Guidelines C. Plass-Duelmer, A. Werner, S. Reimann, S. Sauvage, R. Steinbrecher and the ACTRIS VOC workshop participants comparable

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

REQUEST FOR QUOTATION

REQUEST FOR QUOTATION REQUEST FOR QUOTATION March 5, 2010 PROJECT: Analysis of Non-Methane Organic Compounds QUOTES DUE BY: 5:00 PM on Friday, March 19, 2010 OVERVIEW The San Joaquin Valley Unified Air Pollution Control District

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist Innovative Thermal Desorption From Agilent Technologies Vivek Dhyani Application Chemist What is thermal desorption (TD)? How does thermal desorption compare to Head Space? Thermal desorption (TD) is the

More information

WP3: In-situ chemical, physical and optical properties of aerosols. WP4: Trace gases networking: Volatile organic carbon and nitrogen oxides

WP3: In-situ chemical, physical and optical properties of aerosols. WP4: Trace gases networking: Volatile organic carbon and nitrogen oxides WP3: In-situ chemical, physical and optical properties of aerosols WP4: Trace gases networking: Volatile organic carbon and nitrogen oxides WP3: In-situ chemical, physical and optical properties of aerosols

More information

Determination of Hydrocarbon Components in Petroleum Naphthas

Determination of Hydrocarbon Components in Petroleum Naphthas Determination of Hydrocarbon Components in Petroleum Naphthas LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS, Petrochemical, Naphtha, Deconvolution, Retention Index 1. Introduction The

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated Lee Marotta, Field Application Scientist, PerkinElmer Miles Snow, Research Scientist, PerkinElmer

More information

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

TDTS.   Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination Thermal Desorption Technical Support Note 75: Liquid standard injection, tube impedance and other factors which may cause discrimination during the calibration of thermal desorption methods Keywords TD

More information

Supplement of Secondary formation of nitrated phenols: insights from observations during the Uintah Basin Winter Ozone Study (UBWOS) 2014

Supplement of Secondary formation of nitrated phenols: insights from observations during the Uintah Basin Winter Ozone Study (UBWOS) 2014 Supplement of Atmos. Chem. Phys., 16, 2139 2153, 2016 http://www.atmos-chem-phys.net/16/2139/2016/ doi:10.5194/acp-16-2139-2016-supplement Author(s) 2016. CC Attribution 3.0 License. Supplement of Secondary

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Series 2 Air Server Automated canister and on-line air/gas analysis

Series 2 Air Server Automated canister and on-line air/gas analysis Series 2 Air Server Thermal Desorption Options Automated canister and on-line air/gas analysis Series 2 Air Server A range of accessories for series 2 UNITY thermal desorbers offering automated canister

More information

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Turk J Chem 24 (2000), 131 139. c TÜBİTAK Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106 Naciye KILIÇ University of Uludağ,

More information

Activity in the FID Detection Port: A Big Problem if Underestimated

Activity in the FID Detection Port: A Big Problem if Underestimated Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be

More information

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278 Application Note Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method Application Note Abstract In this application note, we demonstrate the suitability of Markes

More information

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns Brochure Brief Porous Layer Open Tubular (PLOT) columns have been replacing traditional packed columns

More information

Table of Physical Properties for Hydrocarbons and Other Compounds of Interest to the Natural Gas and Natural Gas Liquids Industries

Table of Physical Properties for Hydrocarbons and Other Compounds of Interest to the Natural Gas and Natural Gas Liquids Industries GPA Midstream Standard 2145-16 Table of Physical Properties for Hydrocarbons and Other Compounds of Interest to the Natural Gas and Natural Gas Liquids Industries (Includes data previously published in

More information

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis Name Manraj Gill (Partner: Tanner Adams, Lab Section: 102) Introduction In this experiment, we use chromatography and mass spectrometry

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

Study of Residual Solvents in Various Matrices by Static Headspace

Study of Residual Solvents in Various Matrices by Static Headspace Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.

More information

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc. Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical

More information

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Real Time On-Site Odor and VOC Emission Measurements Using a znose Real Time On-Site Odor and VOC Emission Measurements Using a znose Edward J. Staples, Electronic Sensor Technology, EST@ESTCAL.COM Remediation Site Description Remediation of contaminated soil from where

More information

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES Mihriban Yılmaz Civan, Öznur Kuntasal and Gürdal Tuncel 1 1 OUTLINE Introduction Importance and

More information

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES JPACSM 127 AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES Trace Analytical Inc. Menlo Park, CA ABSTRACT GC based gas analyzers with Reduction Gas Detector (RGD) and Flame Ionization

More information

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length

More information

Combustible Gas Catalytic Bead (0-100 %LEL) Part No FM Performance Certified 1,4 FM Performance Certified 1

Combustible Gas Catalytic Bead (0-100 %LEL) Part No FM Performance Certified 1,4 FM Performance Certified 1 Sensor Data Sheet Document No. 365-2211-31 (Rev D) Combustible Gas Catalytic Bead (0-100 %LEL) Part No. 823-0211-31 FM Performance Certified 1,4 FM Performance Certified 1 Minimum Indicated Concentration...

More information

Highly Sensitive and Rugged GC/MS/MS Tool

Highly Sensitive and Rugged GC/MS/MS Tool Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Determination of Volatile Organic Compounds in Air

Determination of Volatile Organic Compounds in Air JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.

More information

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline Fast : Cycles in

More information

NIST gas standards containing volatile organic compounds in support of ambient air pollution measurements

NIST gas standards containing volatile organic compounds in support of ambient air pollution measurements Air Pollution XVI 357 NIST gas standards containing volatile organic compounds in support of ambient air pollution measurements G. C. Rhoderick Analytical Chemistry Division, Chemical Science and Technology

More information

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne

More information

VOLATILE ORGANIC COMPOUNDS (VOCs) IN THE AIR, THEIR IMPORTANCE AND MEASUREMENTS

VOLATILE ORGANIC COMPOUNDS (VOCs) IN THE AIR, THEIR IMPORTANCE AND MEASUREMENTS VOLATILE ORGANIC COMPOUNDS (VOCs) IN THE AIR, THEIR IMPORTANCE AND MEASUREMENTS Chaithanya D. Jain, Harish S. Gadhavi, Lokesh K. Sahu and A. Jayaraman What are VOCs? Volatile Organic Compounds (VOCs) are

More information

Automated Characterization of Compounds in Fire Debris Samples

Automated Characterization of Compounds in Fire Debris Samples 125 Sandy Drive, Newark, Delaware 19713 USA tel: 302-737-4297 fax: 302-737-7781 www.midi-inc.com Automated Characterization of Compounds in Fire Debris Samples Application Note Forensics Fire Debris Analysis

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Analysis of BTEX in Natural Water with SPME

Analysis of BTEX in Natural Water with SPME Analysis of BTEX in Natural Water with SPME Application Note Environmental Author Giordano Vassalli Sezione Protezione Aria Introduction Benzene, toluene, ethylbenzene and xylene (BTEX) isomers are monocyclic

More information

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Series 2. Continuous monitoring of trace chemical vapours

Series 2. Continuous monitoring of trace chemical vapours Series 2 TT24-7 Continuous monitoring of trace chemical vapours Series 2 TT24-7 TT24-7 Series 2 Continuous monitoring of trace chemical vapours The TT24-7 Series 2 is an advanced thermal desorption (TD)

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing.

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing. Thermal Desorption Technical Support Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing Keywords certification protocols, (T)(V)(S)VOCs, ULTRA-UNITY, field/laboratory

More information

VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations

VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations Environmental Chemistry group, water Science Forum and the Separation Science Group

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column Analyze Hydrocarbon Impurities in,3-butadiene with an Agilent J&W GS-Alua PT Column Application Note Energy and Chemicals Authors Yun Zou and Chunxiao Wang Agilent Technologies Shanghai Ltd Abstract The

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY AppNote 6/2004 Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons Edward A. Pfannkoch, Jacqueline A. Whitecavage, and Jeffrey Christenson Gerstel, Inc., 701 Digital Drive,

More information

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask

More information

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement

More information

4. A sample of 0.200g of an organic compound was subjected to combustion analysis and

4. A sample of 0.200g of an organic compound was subjected to combustion analysis and 1. A sample of 0.205g of an organic compound was subjected to combustion analysis and produced 0.660g of carbon dioxide, 0.225g of water and nothing else. Its RMM is 82. 2. A sample of 0.400g of an organic

More information

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration

More information

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

DEVELOPMENT OF OZONE REACTIVITY SCALES FOR VOLATILE ORGANIC COMPOUNDS

DEVELOPMENT OF OZONE REACTIVITY SCALES FOR VOLATILE ORGANIC COMPOUNDS DEVELOPMENT OF OZONE REACTIVITY SCALES FOR VOLATILE ORGANIC COMPOUNDS by William P. L. Carter Published in the Journal of the Air and Waste Management Association Vol 44, pages 881-899 January 20, 1994

More information

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to. ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination

More information

Watercare Air Quality Group March 2002 APPENDIX A. Monitoring Site Location Diagrams

Watercare Air Quality Group March 2002 APPENDIX A. Monitoring Site Location Diagrams APPENDIX A Monitoring Site Location Diagrams Appendix A contains 4 pages including cover Watercare Services Ltd: AQG-A;3. MfE GEMS Annual Report Appendicies A1 of A55 Mt Eden Site 4:65 Figure 1 Mt Eden,

More information

Nomenclature. 133 minutes. 130 marks. Page 1 of 22

Nomenclature. 133 minutes. 130 marks. Page 1 of 22 3.1.5.1 Nomenclature 133 minutes 130 marks Page 1 of 22 Q1. (a) Write an equation for the formation of epoxyethane from ethene, showing the structure of the product. Explain why the epoxyethane molecule

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

method for monitoring exposure to gasoline vapour in air revision 2002

method for monitoring exposure to gasoline vapour in air revision 2002 method for monitoring exposure to gasoline vapour in air revision 2002 Prepared by: P. E. Tindle Reviewed for CONCAWE by: A. Bianchi E. Loader C. Money G. Pizzella J. Urbanus (Technical Co-ordinator) Reproduction

More information

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods

More information

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative.

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative. We don t have anything against mass spectrometry. We just think it s time for a worthy alternative. gas chromatography detector New directions in science are launched by new tools much more often than

More information

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note. Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID Application Note Volatile Organic Compounds (VOCs) Author Andrew Jones Activated

More information

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION. APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Kira. Yang PerkinElmer, Inc. Shanghai, China Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS Introduction

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Aps Fabriksvej 19 6000 Kolding Denmark Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 14 November 2014 Test Report- VOC emission

More information

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD) Application Note 16181100 Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD) Keywords Beer PFPD Purge and Trap Sulfur Introduction The presence or absence

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract

More information

GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation

GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation International Journal of Analytical Mass Spectrometry and Chromatography, 2015, 3, 54-62 Published Online September 2015 in SciRes. http://www.scirp.org/journal/ijamsc http://dx.doi.org/10.4236/ijamsc.2015.33007

More information

[ a ppl ic at ion no t e ]

[ a ppl ic at ion no t e ] [ a ppl ic at ion no t e ] Fast A nalysis of A ldehydes and K etones by A C Q U I T Y U P L C Mark E. Benvenuti Waters Corporation, Milford, MA, USA INT RO DUC T ION Aldehydes and ketones are products

More information

Methanol Extraction of high level soil samples by USEPA Method 8260C

Methanol Extraction of high level soil samples by USEPA Method 8260C Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed

More information

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd. 5/26/2017 Ms. Anna Kelley Hamilton Co. Environmental Services 250 Wm. Howard Taft Cincinnati OH 45219 Project Name: Sun Coke Project #: Workorder #: 1705303 Dear Ms. Anna Kelley The following report includes

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Fabriksvej 19 6000 Kolding DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 13 August 2015 Test Report- VOC emission regulations

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information