Determining Elemental Impurities in Pharmaceutical Materials: How to Choose the Right Technique

Size: px
Start display at page:

Download "Determining Elemental Impurities in Pharmaceutical Materials: How to Choose the Right Technique"

Transcription

1 30 Spectroscopy 30(3) March 2015 Atomic Perspectives Determining Elemental Impurities in Pharmaceutical Materials: How to Choose the Right Technique This installment evaluates the application requirements for the determination of 15 elemental impurities in pharmaceutical materials as described in the new United States Pharmacopeia (USP) Chapters <232> and <233> and offers suggestions about which atomic spectroscopic technique might be the most suitable to use. Robert Thomas Since the introduction of the first commercially available atomic absorption (AA) spectrophotometer in the early 1960s, there has been an increasing demand for better, faster, easier-to-use, and more flexible trace-element instrumentation. A conservative estimate shows that in 2015, the market for atomic spectroscopy instruments will be close to $1 billion in annual revenue and that doesn t include aftermarket sales and service costs, which could increase this number by an additional $ million. As a result of this growth, we have seen a rapid emergence of more-sophisticated equipment and easier-to-use software. Moreover, with an increase in the number of manufacturers of atomic spectroscopy instrumentation and its sampling accessories, together with the availability of newer techniques like microwave-induced plasma optical emission spectroscopy (MIP-OES) and laser-induced breakdown spectroscopy (LIBS), the choice of which technique to use is often unclear. It also becomes even more complicated when budgetary restrictions allow only one analytical technique to be purchased to solve a particular application problem. To select the best technique for a particular analytical problem, it is important to understand exactly what the problem is and how it is going to be solved. For example, if the requirement is to monitor copper at percentage levels in a copper plating bath and it is only going to be done once per shift, flame atomic absorption would adequately fill this role. Alternatively, when selecting an instrument to determine 15 elemental impurities in pharmaceutical products according to United States Pharmacopeia (USP) Chapters <232> and <233>, this application is probably better suited for a multielement technique such as inductively coupled plasma optical emission spectrometry (ICP- OES) or perhaps inductively coupled plasma mass spectrometry (ICP-MS), depending on the materials being analyzed. So, when choosing a technique it is important to understand not only the application problem, but also the strengths and weaknesses of the technology being applied to solve the problem. However, there are many overlapping areas between the major atomic spectroscopy techniques, so it is highly likely that for some applications, more than one technique would be suitable. For that reason, it is important to go through a carefully WELCOME TO THE NEW EDITOR OF ATOMIC PERSPECTIVES Robert Thomas has been a valuable contributor to Spectroscopy for many years. The editors of Spectroscopy would like to welcome him as the new editor of the Atomic Perspectives column. His contributions in this new role will be a valued asset to our readers.

2 32 Spectroscopy 30(3) March 2015 Flame AA Flame AA AA ICP emission - radial Light source or GFAA Monochromator Hydride generation AA GFAA ICP emission - axial or CV/Hydride AA Echelle ICP-MS Detection limit range (µg/l) ICP-OES Figure 2: Typical atomic spectroscopy detection limit ranges. Axial or radial view, simultaneous or sequential ICP-MS Echelle Mass spectrometer GFAA Flame AA ICP emission Figure 1: Simplified schematic of AA, ICP-OES, and ICP-MS. ICP-MS thought out evaluation process when selecting a piece of equipment. The main intent of this installment is to look at the application requirements for the determination of 15 elemental impurities in pharmaceutical materials as described in the new USP Chapters <232> and <233> (1,2). The goal is to offer some insight as to which might be the most suitable atomic spectroscopic approach, particularly for new users who do not have a strong background or training in the real-world application of these techniques. We will not discuss either the MIP-OES or LIBS techniques in this evaluation. They are both very useful techniques for many sample types and have been described at length in the open literature (3). However, for the purpose of space and clarity, we will be focusing on the four most commonly used atomic spectroscopy techniques: flame atomic absorption (FAA), electrothermal atomization (ETA), ICP-OES, and ICP-MS. To set the stage, let s first take a brief look at their fundamental principles. Simplified schematics of each of these techniques are shown in Figure 1. Flame Atomic Absorption This is predominantly a single-element technique for the analysis of liquid samples that uses a flame to generate ground-state atoms. The sample is aspirated into the flame via a nebulizer and a spray chamber. The ground-state atoms of the sample absorb light of a particular wavelength, either from an element-specific, hollow cathode lamp or a continuum source lamp. The amount of light absorbed is measured by a monochromator (optical system) and detected by a photomultiplier tube or solid-state detector, which converts the photons into an electrical signal. As in all atomic spectroscopy techniques, this signal is used to determine the concentration of that element in the sample, by comparing it to calibration or reference standards. FAA typically uses a liquid sample flow of about 2 5 ml/min and is capable of handling in excess of 10% total dissolved solids, although for optimum performance it is best to keep the solids down below 2%. For the majority of elements, its detection capability is ppb with an analytical range up to ppm, depending on the absorption wavelength used. However, it is not really suitable for the determination of the halogens, nonmetals like carbon, sulfur, and phosphorus, and has very poor detection limits for the refractory, rare earth, and transuranic Orders of magnitude Figure 3: Typical atomic spectroscopy analytical working ranges. elements. Sample throughput for 15 elements per sample is typically 10 samples per hour. Electrothermal Atomization This is also mainly a single-element technique, although multielement instrumentation is now available. It works on the same principle as FAA, except that the flame is replaced by a small heated tungsten filament or graphite tube. The other major difference is that in ETA, a very small sample (typically, 50 µl) is injected onto the filament or into the tube, and not aspirated via a nebulizer and a spray chamber. Because the ground-state atoms are concentrated in a smaller area, more absorption takes place. The result is that ETA offers detection capability at the ppb level, with an analytical range up to ppb. The elemental coverage limitations of the technique are similar to the FAA technique. However, because a heated graphite tube is used for atomization in most commercial instruments, it cannot determine

3 34 Spectroscopy 30(3) March 2015 Table I: Approximate sample throughput capability of atomic spectroscopy techniques Technique Elements at a Time Duplicate Analysis (min) Samples per Hour (One Element) Samples per Hour (Five Elements) Samples per Hour (15 Elements) FAA ETA ~1 ICP-OES Up to ICP-MS Up to Table II: Elemental impurities for drug products defined in USP Chapter <232>. (Note: These are proposed PDE levels to align with ICH guidelines.) Element Oral Daily Dose PDE (µg/day) Parenteral Daily Dose PDE (µg/day) Inhalation Daily Dose PDE (µg/day) LVP Component Limit (µg/g) Cadmium Lead Arsenic (inorganic) Mercury (inorganic) Iridium Osmium Palladium Platinum Rhodium Ruthenium Chromium * * 2.9 * Molybdenum Nickel Vanadium Copper *Note: Cr is not considered a safety concern for these drug delivery methods the refractory, rare earth, and transuranic elements because they tend to form stable carbides that cannot be readily atomized. One of the added benefits is that ETA can also analyze slurries and some solids because no nebulization process is involved in introducing the sample. This technique is not ideally suited for multielement analysis because it takes 3 4 min to determine one element per sample. As a result, sample throughput for 15 elements is in the order of one sample per hour. Hydride Generation AA Hydride generation (HG) is a very useful analytical technique to determine the hydride forming metals, such as As, Bi, Sb, Se, and Te, usually by AA (although detection by either ICP-OES or ICP-MS can also be used). In this technique, the analytes in the sample matrix are first reacted with a very strong reducing agent, such as sodium borohydride, to release their volatile hydrides, which are then swept into a heated quartz tube for atomization. The tube is heated either by a flame or a small oven that creates the ground state atoms of the element of interest and then measures by atomic absorption. When used with ICP-OES or ICP-MS, the volatile hydrides are passed directly in the plasma for excitation or ionization. By choosing the optimum chemistry, mercury can also be reduced in solution in this way to generate elemental mercury. This is known as the cold vapor (CV) technique. HGAA and CVAA can improve the detection for these elements compared to FAA by up to three orders of magnitude, achieving detection capability of ppb with an analytical range up to ppb, depending on the element of interest. It should also be pointed out that dedicated mercury analyzers (some using gold amalgamation techniques), coupled with atomic absorption or atomic fluorescence are capable of better detection limits. Because of the on-line chemistry involved, these techniques are very time-consuming and are normally used in conjunction with FAA, so they will most likely impact the overall sample throughput. Radial ICP-OES Radially viewed ICP-OES is a multielement technique that uses a traditional radial (side-view) inductively coupled plasma to excite ground-state atoms to the point where they emit wavelengthspecific photons of light that are characteristic of a particular element. The number of photons produced at an element-specific wavelength is measured by high resolution optics and a photonsensitive device such as a photomulti-

4 36 Spectroscopy 30(3) March 2015 Table III: USP J-values compared to limits of quantitation for FAA and ETA Element Concentration Limits for an Oral Drug with a Max Daily Dose of 10 g/ day (μg/g) J-Value with a Sample Dilution of 2g/100 ml AA LOQ (IDL 10) Factor Improvement (J-Value/LOQ) FAA ETA FAA ETA Cadmium Lead Arsenic * * 60 Mercury * 5 300* 6 Iridium NA 0.7 NA Osmium NA NA NA NA Palladium Platinum Rhodium NA 3.3 NA Ruthenium Molybdenum Nickel Vanadium Copper ,000 *Note: These values are based on HGAA for arsenic and CVAA for mercury; NA = not appropriate for the technique plier tube or a solid state detector. This emission signal is directly related to the concentration of that element in the sample. The analytical temperature of an ICP is about K, compared to that of a flame or a graphite furnace, which is typically K. For the majority of elements, a radial ICP instrument can achieve detection capability of ppb with an analytical range up to ppm, depending on the emission wavelength used. The technique can determine a similar number of elements as FAA can, but has the advantage of offering the capability of nonmetals like sulfur and phosphorus, together much better performance for the refractory, rare earth, and transuranic elements. The sample requirement for ICP-OES is approximately 1 ml/min, although lower flow nebulizers are available. Instruments are capable of aspirating samples containing up to 10% total dissolved solids; for optimum performance, the level is usually kept below 2%. ICP- OES is a rapid multielement technique, so sample throughput for 15 elements per sample is in the order of 20 samples an hour. Axial ICP-OES The principle of axial ICP-OES is exactly the same as that of radial ICP- OES, except that in axial ICP-OES, the plasma is viewed horizontally (end-on). The benefit is that more photons are seen by the detector, and for this reason detection limits can be as much as an order of magnitude lower, depending on the design of the instrument. The disadvantage is that the working range is also reduced by an order of magnitude. As a result, for the majority of elements, an axial ICP instrument can achieve detection capability of ppb with an analytical range up to ppm, depending on the emission wavelength used. The other disadvantage of viewing axially is that more severe matrix suppression is observed, which means that the total dissolved solids content of the sample needs to be kept much lower. Sample flow requirements are the same as for radial ICP-OES. Most commercial ICP-OES instruments have both radial and axial capability built in. However, because of the different hardware configurations available, some instruments work by carrying out the analysis either using the radial or the axial view (sequentially). Others can use both the radial and axial view at exactly the same time (simultaneously), which for some applications may be advantageous. Strategic use of both radial and axial view, together with the optimum wavelength selection, can extend the analytical range by 2 3 orders of magnitude. Sample throughput is approximately 20 samples per hour, the same as that of radial ICP-OES. ICP-MS The fundamental difference between ICP-OES and ICP-MS is that in ICP- MS, the plasma is not used to generate photons, but to generate positively charged ions. The ions produced are transported and separated by their atomic mass-to-charge ratio using a mass-filtering device such as a quadrupole or a magnetic sector. The generation of such large numbers of positively charged ions allows ICP-MS to achieve detection limits approximately three orders of magnitude lower than ICP-OES, even for the refractory, rare earth, and transuranic

5 March 2015 Spectroscopy 30(3) 37 Table IV: USP J-values compared to limits of quantitation for axially viewed ICP-OES Element Concentration Limits for an Oral Drug with a Max Daily Dose of 10 g/day (μg/g) J-Value with a Sample Dilution of 2 g/100 ml Axial ICP-OES LOQ (IDL 10) Factor Improvement (J-Value/LOQ) Cadmium Lead Arsenic Mercury Iridium Osmium Palladium Platinum Rhodium Ruthenium Molybdenum Nickel Vanadium Copper elements. As a result, for the majority of elements, an ICP-MS instrument can achieve detection capability of ppb with an analytical range up to ppm, using pulse counting measurement, but can be extended even further up to ,000 ppm by using analog counting techniques. However, it should be emphasized that if such large analyte concentrations are being measured, expectations should be realistic about also carrying out ultratrace determinations of the same element in the same sample run. The sample requirement for ICP-MS is approximately 1 ml/min, although lower flow nebulizers are available. ICP-MS systems are capable of aspirating samples Applied Rigaku Technologies, Inc. website: info@rigakuedxrf.com

6 38 Spectroscopy 30(3) March 2015 Table V: USP J-values compared to limits of quantitation for ICP-MS Element Concentration Limits for an Oral Drug with a Max Daily Dose of 10 g/day (μg/g) J-Value with a Sample Dilution of 0.2 g/100 ml ICP-MS LOQ (IDL 10) Factor Improvement (J-Value/LOQ) Cadmium Lead Arsenic Mercury Iridium Osmium Palladium Platinum ,385 Rhodium ,000 Ruthenium ,000 Molybdenum Nickel ,588 Vanadium Copper ,970 containing up to 5% total dissolved solids for short periods with the use of specialized sampling accessories. However, because the sample is being aspirated into the mass spectrometer, for optimum performance, matrix components should ideally be kept below 0.2%. This is particularly relevant for laboratories that experience a high sample workload. Sample throughput will be approximately 20 samples per hour for the determination of 15 elements in a sample. Comparison Highlights There is no question that the multielement techniques like ICP-OES and ICP-MS are better suited for the determination of the 15 elements defined in USP Chapters <232>, especially in laboratories that are expected to be carrying out the analysis in a highthroughput, routine environment. If the best detection limits are required, ICP-MS offers the best choice, followed by graphite furnace AA (ETA). Axial ICP-OES offers very good detection limits for most elements, but generally not as good as ETA. Radial ICP-OES and FAA show approximately the same detection limits performance, except for the refractory, rare earth, and the transuranic elements, for which performance is much better with ICP-OES. For mercury and those elements that form volatile hydrides, such as As, Bi, Sb, Se, and Te, the cold vapor or hydride generation techniques offer exceptional detection limits. Figures 2 and 3 as well as Table I show an overview of the detection capability, working analytical range, and sample throughput of the major atomic spectroscopy approaches three of the metrics most commonly used to select a suitable technique. This information should be considered an approximation and used only for guidance purposes. It is also worth mentioning that the detection limits of quadrupole-based ICP-MS, when used in conjunction with collision-reaction cell or multiple mass separation-selection devices, is now capable of sub-parts-per-trillion (ppt) detection limits, even for elements such as Fe, K, Ca, Se, As, Cr, Mg, V, and Mn, which traditionally suffer from plasma- and solvent-based polyatomic interference. This is not meant to be a detailed description of the fundamental principles and comparisons of each technique, but a basic understanding as to how they differ from each other. For a more detailed comparison of the strengths and weaknesses of the major atomic spectroscopy techniques, please see reference 5. Now, let s turn our attention to selecting the best technique based on the demands of the application. Application Demands Comparing an instrumental technique based on its performance specifications with simple standards is important, but it bears little relevance to how that instrument is going to be used in a real-world situation. For example, instrument detection limits (IDLs) are important to know, but how are they impacted by the matrix and sample preparation procedure? What are the real-world method detection limits (MDLs) and is the technique capable of quantifying the maximum concentration values expected for this analysis? Also, what kind of precision and accuracy can be expected if you re working close to the limit of quantitation (LOQ) for the overall methodology being used? Additionally, on the sample throughput side, how many samples are expected and at what frequency will they be coming into the laboratory? How much time can be spent on sample preparation and how quickly must they be analyzed? Sometimes the level of interferences from the matrix will have a major impact on the selection process or even the amount or volume of sample available for analysis. Understanding the demands of an application is therefore of critical importance when a technique is being purchased, and particularly if there is a minimum amount of expertise or expe-

7 March 2015 Spectroscopy 30(3) 39 rience in house on how best to use it to solve a particular application problem. This could be the likely scenario in a pharmaceutical production laboratory that is being asked to check the elemental impurities of incoming raw materials used to manufacture a drug compound. They will now have to follow the new USP Chapters <232> and <233>, which recommends the use of a plasma-based spectroscopic technique to carry out the analysis... or any other atomic spectroscopy technique as long as the validation protocols are met. The expertise of the operator should never be underestimated. If ICP-MS is being seriously considered, it generally requires an analyst with a higher skill level to develop rugged methodology free of interferences that can eventually be put in the hands of an inexperienced user to operate on a routine basis. This again is a real concern if the technique is being used by novice users who have limited expertise in running analytical instrumentation, which may be the case in the pharmaceutical industry. So, let s take a closer look at the new USP chapters to understand what atomic spectroscopy techniques might best meet the demands of this application. New USP General Chapters on Elemental Impurities Although the risk factors for heavy metal impurities in pharmaceutical materials have changed dramatically, standard methods for their testing and control have not changed much for more than 100 years, and as a result, most heavy metals limits had little basis in toxicology. For that reason, one of the most significant standards introduced by the USP in the past decade has been new methodology for determining elemental impurities and contaminants in drug products and dietary supplements. The new methods have been summarized in the United States Pharmacopeia 37 National Formulary 32 (USP-NF) General Chapters <232> and <233> for pharmaceutical products and Chapter <2232> for dietary supplements (4). These three methods will be replacing Chapter <231> (6), which is a 100-year old colorimetric test for heavy metals based on precipitation of the metal sulfide in a sample and comparing it to a lead standard. These new methods have been going through a review and approval process for a number of years, but by a combination of pushback from the pharmaceutical industry and discussions with other global pharmacopeias, toxicity levels and implementation timelines have been changed a number of times. However, a recent announcement by the USP (7) has indicated that these new chapters will finally be implemented on January 1, 2018, to coincide with the full approval with the International Conference on Harmonization Q3D Step 4 guidelines, which is expected to be December 16, 2017, for existing pharmaceutical products. (Note: the International Conference on Harmonization [ICH] of Technical Requirements for Registration of Pharmaceuticals for Human Use is an international

8 40 Spectroscopy 30(3) March 2015 body that brings together the regulatory authorities and pharmaceutical industry in Europe, Japan, and the United States to discuss scientific and technical aspects of drug registration) (8). Let s take a closer look at USP Chapters <232> and <233> to get a better understanding of the analytical requirements for this application. USP General Chapter <232> Chapter <232> specifies limits for the amount of elemental impurities in drug products, drug substances, active ingredients, and excipients. These impurities may be present naturally, derived from the production catalysts, or introduced inadvertently through the manufacturing process, or they could be environmental contaminants in the pharmaceutical raw materials. When elemental impurities are known to have the potential to be present, compliance to the specified levels is a requirement. Additionally, because of the ubiquitous nature of arsenic, cadmium, lead, and mercury, those four elements at a very minimum must be monitored. The elemental impurity levels in the drug products, unless otherwise specified in an individual drug product monograph, must show compliance with the limits specified and must be made available to the regulatory agency upon request. A total of 15 elemental impurities (Cd, Pb, As, Hg, In, Os, Pd, Pt, Rh, Ru, Cr, Mo, Ni, V, and Cu) are specified together with their toxicity limits, defined as maximum permitted daily exposure (PDE) levels in micrograms per day for the four major drug delivery categories. The PDE limits, which represent updated levels proposed by USP in an attempt to align itself with ICH guidelines, are shown in Table II. Note: ICH guidelines define impurity levels for an additional nine elements. Of the 15 common elements, the majority of PDE levels are similar in both USP and ICH methods, but some are slightly different, which is the basis of ongoing discussions between the two organizations. For specific details about elemental impurities and their limits, please refer to the cited ICH reference (8). It s also worth pointing out that USP and ICH are going through the alignment process, and until this is finalized, it is difficult to know what the final USP PDE limits will be. For that reason, please refer to the USP elemental impurities web pages (7) for updates. Speciated Forms Chapter <232> also addresses speciation, although it does not specify an analytical procedure. Each of the elemental impurities has the potential to be present in differing oxidation states or species. However, arsenic and mercury are of particular concern because of the differing toxicities of their inorganic and organic forms. The arsenic limits are based on the inorganic form, which is the most toxic. Arsenic can be measured using a total-arsenic procedure under the assumption that all arsenic contained in the material under test is in the inorganic form. When the limit is exceeded using a total-arsenic procedure, it should be demonstrated using a suitable procedure to separate the species and to ensure that the inorganic form meets the specification. The mercury limits are based on the inorganic form because methyl mercury, the most toxic form, is rarely an issue for pharmaceuticals. Therefore, the limit was established assuming that if mercury was present in the drug compound it would exist as the most common inorganic form. However, if there is a known potential for the material to contain methyl mercury (such as drugs, supplements, or compounds derived from fish or kelp), an appropriate speciation procedure is required. USP General Chapter <233> Chapter <233> deals with the analytical procedure, including the sample preparation procedure, instrumental method, and validation protocols for measuring the elemental impurities using one of two plasma-based spectrochemical techniques ICP-OES, ICP-MS, or alternatively any other trace-element technique such as FAA or graphite furnace AA, as long as it meets the data quality objectives of the method defined in the validation protocol section. In addition, before any technique is used, the overall analytical procedure must be confirmed to be appropriate for the instrument being used and the samples being analyzed by meeting the Alternative Procedure Validation protocol (9). The chapter also recommends reading USP General Chapter <730> on plasma spectrochemistry for further guidance (10). Validation Protocol Meeting the validation protocol is critical when selecting the best technique for this application because all aspects of the analytical procedures, including the instrumental technique and dissolution process, must be validated and shown to be acceptable, in accordance with the validation protocol described in Chapter <233>. The requirements for the validation of a procedure for elemental impurities is then determined by following a set of quality protocols, which cover a variety of performance tests, including Detectability Precision Specificity Accuracy Ruggedness Limit of quantification Linear range Meeting the performance requirements defined in these tests must be demonstrated experimentally using an appropriate system suitability procedure and reference material. The suitability of the method must be determined by conducting studies with the material under test supplemented or spiked with known concentrations of each target element of interest at the appropriate acceptance limit concentration. It should also be emphasized that the materials under test must be spiked before any sample preparation steps are performed. Suitability of the Technique To get a better understanding of the suitability of the technique being used and whether its detection capability is appropriate for the analytical task, it s important to know the PDE limit for each target element, and in particular, what the USP calls the J-value. In Chapter <233>, the J-value is defined as the PDE concentration of the element of interest, appropriately diluted to the working

9 March 2015 Spectroscopy 30(3) 41 range of the instrument, after completion of the sample preparation procedure to get the sample into solution. To understand the J-value, let s use Pb as an example. The PDE limit for Pb in an oral medication defined in Chapter <232> is 5 µg/day. Based on a suggested dosage of 10 g of the drug product per day, that is equivalent to 0.5 µg/g Pb. If 2 g of sample is digested or dissolved and made up to 100 ml, that s a 50-fold dilution, which is equivalent to 10 µg/l. So the J-value for Pb in this example is equal to 10 µg/l. (Note: Further sample dilutions can be made if required, based on the technique being used.) The method then suggests using a calibration made up of two standards: standard 1 = 2.0 J, standard 2 = 0.5 J. So for Pb, this is equivalent to 20 µg/l for standard 1 and 5 µg/l for standard 2. The suitability of a technique is then determined by measuring the calibration drift and comparing results for standard 1 before and after the analysis of all the sample solutions under test. This calibration drift should be <20% for each target element. However, after the suitability of the technique has been determined, further validation protocols described previously must be carried out to show compliance to the regulatory agency if required. It should also be pointed out that no specific instrumental parameters are suggested in this section, but only to perform the analysis according to the manufacturer s suggested conditions and to calculate and report results based on the original sample size. However, it does say that appropriate measures must be taken to correct for interferences, such as matrix-induced wavelength overlaps in ICP-OES and argon-based polyatomic interference with ICP-MS. So with this information, let s take as an example the oral drug delivery method and calculate the J-values for each elemental impurity and compare them with the limits of quantitation (LOQ) for each technique to give us an assessment of their suitability. For this analytical scenario, we ll take the LOQ for the method as 10 the IDL. These LOQs were calculated by taking the average of published IDLs from three instrument vendors application material and multiplying them by 10 to get an approximation of the LOQ. In practice, a method LOQ is typically determined by processing the matrix blank through the entire sample preparation procedure and taking 10 replicate measurements. The method LOQ, sometimes referred to as the method detection limit (MDL), is then calculated as 3 the standard deviation of these 10 measurements. To make this comparison valid, the sample weight was adjusted for each technique, based on the detection limit and analytical working range. For AA and ICP-OES, we used a sample dilution of 2 g/100 ml, whereas for ICP-MS we used 0.2 g/100 ml. AA and ICP-OES could definitely use larger sample weights, but for high-throughput routine analysis, we are probably at the optimum dilution for ICP-MS. Tables III, IV, and V show the comparison of AA (FAA and ETA), ICP-OES, and ICP-MS, respectively. The important data to consider are in the final column, labeled Factor Improvement, which is the J-value divided by the LOQ. Generally speaking, the higher this number, the more suitable the technique is. Discussion It should be emphasized again that LOQ in these examples is just a guideline as to the real-world detection capability of the technique for this method. However, it does offer a very good approximation as to whether the technique is suitable based on the factor improvement number compared to the J-values for each elemental impurity. Clearly, if this improvement number is close to or less than one, as it is with the majority of elements by FAA, the technique is just not going to be suitable, particularly for the big four heavy metals, which are the most critical. On the other hand, ETA would be suitable for the majority of the impurities (including the heavy metals), except for a few of the catalyst-based elements, which are not ideally suited to the technique. However, the ETA technique is very time-consuming and Get to Know Metrohm Laboratory Titration Ion Chromatography Electrochemistry Spectroscopy Process Find out more at

10 42 Spectroscopy 30(3) March 2015 labor intensive, so it probably wouldn t be a practical solution in a highthroughput pharmaceutical production laboratory. Table IV shows that axial ICP-OES offers some possibilities for monitoring oral drugs because the vast majority of the improvement factors are higher than one. These numbers could be further improved, especially for the heavy metals, by using a much higher sample weight in the sample preparation procedure without compromising the method. (Note: Because most commercial ICP-OES instrumentation offers both axial and radial capability, it was felt that the axial performance was the most appropriate for this comparison.) However, it can be seen in Table V that ICP-MS shows significant improvement factors for all impurities, which are not offered by any other technique. Even for the four heavy metals, there appears to be ample improvement to monitor them with good accuracy and precision. The added benefit of using ICP-MS is that it would also be suitable for the other methods of pharmaceutical delivery, such as intravenous feeding, injection, or inhalation, where the PDE levels are typically an order of magnitude lower. Additionally, if arsenic or mercury levels were found to be higher than the PDE levels, it would be relatively straightforward to couple ICP-MS with high performance liquid chromatography (HPLC) to monitor the speciated forms of these elements. Final Thoughts There are many factors to consider when selecting a trace-element technique that is the most suited to the demands of your application. Sometimes, one technique stands out as being the clear choice, whereas other times it is not quite so obvious. And, as is true with many applications, more than one technique is often suitable. However, the current methodology described in the new USP Chapters <232>, <233>, and <2232> presents unique challenges, not only from a perspective of performance capability, but also because of the validation protocols that have to be met to show suitability of the technique for the analytical procedure being used. From a practical perspective, there is no question that to meet all of the PDE limits in all pharmaceutical delivery methods, ICP- MS is probably the most appropriate technique. However, for oral delivery medications, especially those that can be easily brought into solution with a suitable aqueous or organic solvent, axial ICP-OES could offer a less costly approach. Also, if the sample workload requirements are not so demanding, ETA could provide a solution. It will be interesting to eventually see how USP limits and methodology will be aligned with ICH directives, but the recent announcement delaying implementation should not discourage global pharmaceutical and nutraceutical manufacturers to have the appropriate analytical capability in place as soon as possible to show compliance to regulators that their products are free from elemental impurities and contaminants. This is reinforced by the statement in the frequently asked questions that went along with the most recent USP announcement (7): Given that General Chapters <232> and <2232> provide significant improvements over existing approaches in the control of elemental impurities, USP encourages users to implement the new methods as soon as reasonably possible. So it s clear, the USP takeaway message is that even though full implementation has been delayed to coincide with the completion of the ICH Q3D approval process, the pharmaceutical industry should continue to move forward with the adoption of the new methodology. References (1) General Chapter <232> Elemental Impurities in Pharmaceutical Materials Limits, in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). (2) General Chapter <233> Elemental Impurities in Pharmaceutical Materials Procedures, in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). (3) R.J. Thomas, Spectroscopy 3(29), (2014). (4) General Chapter <2232> Elemental Contamination in Dietary Supplement, in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). (5) R.J. Thomas, How to Select an ICP- MS: Some Important Considerations, Practical Guide to ICP-MS: A Tutorial for Beginners, 3rd Edition, (CRC Press, Boca Raton, Florida, 2013) ISBN (6) General Chapter <231> Heavy Metals Test in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). (7) United States Pharmacopeia Update Communication, January 14, 2015: (8) International Conference on Harmonization, ICH Q3D Step 4 - Guideline for Elemental Impurities (ICH, Geneva, Switzerland, 2014). (9) General Chapter <1225> Validation of Compendial Procedures in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). (10) General Chapter <730> Guidelines on Plasma Spectrochemistry in second supplement to USP37 NF32, 2014 (United States Pharmacopeial Convention, Rockville, Maryland). Robert Thomas has more than 40 years of experience in trace-element analysis. He is the principal of his own freelance writing and scientific consulting company, Scientific Solutions, based in Gaithersburg, Maryland. For more information on this topic, please visit our homepage at:

Elemental Scientific. Stability. Elemental Scientific

Elemental Scientific. Stability. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline

More information

á233ñ ELEMENTAL IMPURITIES PROCEDURES

á233ñ ELEMENTAL IMPURITIES PROCEDURES Second Supplement to USP 38 NF 33 Chemical Tests / á233ñ Elemental Impurities Procedures 1 á233ñ ELEMENTAL IMPURITIES PROCEDURES INTRODUCTION This chapter describes two analytical procedures (Procedures

More information

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific. Spike Recovery. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast

More information

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Validating performance of an Agilent 7800 ICP MS for USP / & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Authors Lindsey Whitecotton, Ed McCurdy, Craig Jones and Amir Liba

More information

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009] 2 Vol. 36(1) [Jan. Feb. 2009] BRIEFING h233i Elemental Impurities Procedures. This proposed new general test chapter is the second of two being developed to replace the general test chapter Heavy Metals

More information

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products APPLICATION NOTE Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products AN43381 Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany Key Words al

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1 Vol. 36(1) [Jan. Feb. 2010] 1 Page 1 of 23 Time:10:03 Date:10/14/09 Instance: g:/pf/production/final PF36(1)/m5193.xml Template:s:/Pf/Template/PFRedesign/Pf353/PFR-pf-server-2009.3f BRIEFING h233i Elemental

More information

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific. Spike Recovery. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Sources of Errors in Trace Element and Speciation Analysis

Sources of Errors in Trace Element and Speciation Analysis Sources of Errors in Trace Element and Speciation Analysis Zoltan Mester, National Research Council of Canada, Institute for National Measurement Standards Outline Definitions Sources of errors in the

More information

Preparation of Food and Feed Samples for Metals Analysis

Preparation of Food and Feed Samples for Metals Analysis White Paper Preparation of Food and Feed Samples for Metals Analysis Alan Cross Technical Specialist, Metals Laboratory Preparation of Food and Feed Samples for Metals Analysis The relative virtues of

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Prof. Dr. Biljana Škrbić, Jelena Živančev

Prof. Dr. Biljana Škrbić, Jelena Živančev 5 th CEFSER Training Course Analysis of chemical contaminants in food and the environment Faculty of Technology, University of Novi Sad, Novi Sad, Republic of Serbia 7-11 May 2012 Analysis of heavy elements

More information

Proposed new USP general chapters <232> and <233> for elemental impurities: The application of ICP-MS for pharmaceutical analysis

Proposed new USP general chapters <232> and <233> for elemental impurities: The application of ICP-MS for pharmaceutical analysis Proposed new USP general chapters and for elemental impurities: The application of ICP-MS for pharmaceutical analysis White paper Authors Amir Liba and Ed McCurdy Agilent Technologies USA Abstract

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP <232><233> AND EP 5.20

RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP <232><233> AND EP 5.20 LIFE SCIENCE I TECHNICAL BULLETIN JUNE 2015 RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP AND EP 5.20 AUTHOR: THOMAS DEVADDER, TEAM LEADER QC AAS/ICP/PARTICLES/TGA

More information

ATOMIC SPECROSCOPY (AS)

ATOMIC SPECROSCOPY (AS) ATOMIC ABSORPTION ANALYTICAL CHEMISTRY ATOMIC SPECROSCOPY (AS) Atomic Absorption Spectroscopy 1- Flame Atomic Absorption Spectreoscopy (FAAS) 2- Electrothermal ( Flame-less ) Atomic Absorption Spectroscopy

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Atomic Spectroscopy AA/ICP/ICPMS:

Atomic Spectroscopy AA/ICP/ICPMS: Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy. UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

a. An emission line as close as possible to the analyte resonance line

a. An emission line as close as possible to the analyte resonance line Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal

More information

Atomic Emission Spectroscopy

Atomic Emission Spectroscopy Atomic Emission Spectroscopy Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

2101 Atomic Spectroscopy

2101 Atomic Spectroscopy 2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Lecture 7: Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the

More information

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES APPLICATION NOTE 43145 Analysis of trace elements in polymer s to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES Authors Introduction Sanja Asendorf, Application

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES TECHNICAL NOTE 43252 Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES Keywords Auto-dilution, Intelligent dilution,

More information

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES)

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES) The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES) Part I: The 1970s Since its commercial inception in 1974, ICP-OES has seen significant technological advancements over

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal

More information

Pros and Cons of Water Analysis Methods

Pros and Cons of Water Analysis Methods Water Lens, LLC 4265 San Felipe, Suite 1100 Houston, Texas 77027 Office: (844) 987-5367 www.waterlensusa.com Pros and Cons of Water Analysis Methods Prepared by: Adam Garland, CTO Water Lens, LLC ICP-MS/OES

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms that have been excited to higher energy levels by absorption

More information

Atomic Absorption & Atomic Fluorescence Spectrometry

Atomic Absorption & Atomic Fluorescence Spectrometry Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,

More information

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS 1. SCOPE AND FIELD OF APPLICATION The method describes the determination of heavy metals (arsenic, cadmium, lead and mercury) in cosmetic products. 2. PRINCIPLE Organic matter in sample is digested by

More information

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS FOR ERRORS AND UPDATES, PLEASE VISIT WWW.TSFX.COM.AU/VCE-UPDATES QUESTION 45 QUESTION 46 Answer is A QUESTION 47 The number of protons in the element.

More information

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace_Rev3 Introducing

More information

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY C1_C0011 1 INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY Janya Buanuam,* Thareerut Woratanmanon, Temporn Sookawee Regional Failure Analysis and Reliability Laboratory,

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

Simultaneous ICP-MS in the Pharmaceutical Industry

Simultaneous ICP-MS in the Pharmaceutical Industry A Powerful New Tool for Elemental Analysis Mass spectrometry has become a popular and widely used technique in the pharmaceutical industry for the analysis and identification of organic compounds. The

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

ISO INTERNATIONAL STANDARD. Water quality Determination of trace elements using atomic absorption spectrometry with graphite furnace

ISO INTERNATIONAL STANDARD. Water quality Determination of trace elements using atomic absorption spectrometry with graphite furnace INTERNATIONAL STANDARD ISO 15586 First edition 2003-10-01 Water quality Determination of trace elements using atomic absorption spectrometry with graphite furnace Qualité de l'eau Dosage des éléments traces

More information

Atomic Absorption Spectroscopy

Atomic Absorption Spectroscopy CH 2252 Instrumental Methods of Analysis Unit IV Atomic Absorption Spectroscopy Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration.

More information

CHAPTER 4: ANALYTICAL INSTRUMENTATION

CHAPTER 4: ANALYTICAL INSTRUMENTATION CHAPTER 4: ANALYTICAL INSTRUMENTATION 4.1 INTRODUCTION In this section, a review of the analytical instrumentation used during sample preparation and analysis is presented which includes an overview of

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?

More information

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

NaturalFacts. Introducing our team. New product announcements, specials and information from New Roots Herbal. April 2009

NaturalFacts. Introducing our team. New product announcements, specials and information from New Roots Herbal. April 2009 NaturalFacts New product announcements, specials and information from New Roots Herbal April 2009 Introducing our team Introducing Our Science Team Dr. ABZAL HOSSAIN Ph.D. Analytical Chemistry Dr. Hossain

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing

Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing W. Keurentjes, F. Maris, T. Wang, M. van Tiel Merck & Co. Content Introduction How did we start Straight

More information

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Application note Agriculture Authors Annie Guerin INRA, Laboratoire d Analyses

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

Handbook of Inductively Coupled Plasma Spectrometry

Handbook of Inductively Coupled Plasma Spectrometry Handbook of Inductively Coupled Plasma Spectrometry Second Edition MICHAEL THOMPSON, BSc, PhD, ARCS, CChem., FRSC Department of Chemistry Birkbeck College University of London J. NICHOLAS WALSH, BSc, PhD

More information

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry 9.1 9A. Atomization Most fundamental for both techniques. Typical types 1. flame - burner type 2. Electrothermal graphite furnace 3. Specialized

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

Optical Atomic Spectroscopy

Optical Atomic Spectroscopy Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state

More information

Laboratory 101: A Guide to Understanding your Testing Laboratory

Laboratory 101: A Guide to Understanding your Testing Laboratory Laboratory 101: A Guide to Understanding your Testing Laboratory Jim Cronin Product Ecology Manager Environmental Monitoring and Technologies, Inc. (EMT) Morton Grove, Illinois Abstract: Laboratory 101:

More information

CONTINUOUS FLOW CHEMISTRY (PROCESSING) FOR INTERMEDIATES AND APIs

CONTINUOUS FLOW CHEMISTRY (PROCESSING) FOR INTERMEDIATES AND APIs CONTINUOUS FLOW CHEMISTRY (PROCESSING) FOR INTERMEDIATES AND APIs Sripathy Venkatraman, Section Head at AMRI Abstract Many contract manufacturing organizations (CMOs) offer continuous flow chemistry, but

More information

110 Lowell Rd Hudson, NH USA Tel:

110 Lowell Rd Hudson, NH USA Tel: 110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental

More information

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

AN INTRODUCTION TO ATOMIC SPECTROSCOPY AN INTRODUCTION TO ATOMIC SPECTROSCOPY Atomic spectroscopy deals with the absorption, emission, or fluorescence by atom or elementary ions. Two regions of the spectrum yield atomic information- the UV-visible

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

PRINCIPLES OF AAS atomization flame furnace atomization absorption

PRINCIPLES OF AAS atomization flame furnace atomization absorption INTRODUCTION Atomic absorption spectroscopy (AAS) currently is the most widely used of atomic spectroscopic techniques. AAS is a quantitative method of elemental analysis that is applicable to many metals

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Implementation of Methods Translation between Liquid Chromatography Instrumentation Implementation of Methods Translation between Liquid Chromatography Instrumentation Michael D. Jones, Peter Alden, Kenneth J. Fountain, and Andrew Aubin Waters Corporation, Milford, MA, U.S. APPLICATION

More information

COMPARISON OF ATOMIZERS

COMPARISON OF ATOMIZERS COMPARISON OF ATOMIZERS FOR ATOMIC ABSORPTION SPECTROSCOPY Introduction Atomic spectroscopic methods are all based on the interaction of light and analyte atoms in the gas phase. Thus, a common component

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

CH. 21 Atomic Spectroscopy

CH. 21 Atomic Spectroscopy CH. 21 Atomic Spectroscopy 21.1 Anthropology Puzzle? What did ancient people eat for a living? Laser Ablation-plasma ionization-mass spectrometry CH. 21 Atomic Spectroscopy 21.2 plasma In Atomic Spectroscopy

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

9/13/10. Each spectral line is characteristic of an individual energy transition

9/13/10. Each spectral line is characteristic of an individual energy transition Sensitive and selective determination of (primarily) metals at low concentrations Each spectral line is characteristic of an individual energy transition 1 Atomic Line Widths Why do atomic spectra have

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Chemistry Instrumental Analysis Lecture 18. Chem 4631

Chemistry Instrumental Analysis Lecture 18. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 18 Instrumentation Radiation sources Hollow cathode lamp Most common source Consist of W anode and a cathode sealed in a glass tube filled with Ne or Ar. Hollow

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

BRIEFING 1730 THEORY AND PRACTICE OUTLINE

BRIEFING 1730 THEORY AND PRACTICE OUTLINE 40(6) In-Process Revision: PLASMA SPECTROCHEMISTRY-THEORY AND... Page 1 of 15 BRIEFING 1730 Plasma Spectrochemistry Theory and Practice. The Chemical Analysis Expert Committee proposes this new

More information