José Benito Quintana, PhD University of Santiago de Compostela Castellón Training Course

Size: px
Start display at page:

Download "José Benito Quintana, PhD University of Santiago de Compostela Castellón Training Course"

Transcription

1 José Benito Quintana, PhD University of Santiago de Compostela Castellón Training Course

2 Overview LOD/LOQs: definition(s) LOD/LOQs: theoretical background Ways of estimating LODs/LOQs Examples Low resolution MS Vs High resolution MS (My) recommendations 2

3 LOD/LOQ - Terminology Limit of detection / Detection limit / Method detection limit / Minimum detectable value : LOD, LoD, LD, L D, MDL Limit of Quantification / Quantification limit : LOQ, LoQ LQ, L Q, MQL Critical value/limit of decision: L c Decision Limit: CC (same as L c ) Detection capability: CCβ (same as LOD) More at: 3

4 LOD - Definition IUPAC The limit of detection, expressed as the concentration, c L, or the quantity, q L, is derived from the smallest measure, x L, that can be detected with reasonable certainty for a given analytical procedure. The value of x L is given by the equation: X l = X blank + k s blank where X blank is the mean of the blank measures, s blank is the standard deviation of the blank measures, and k is a numerical factor chosen according to the confidence level desired. Typically k=3 4

5 LOD/LOQ - Definition In general LOD: concentration/quantity that can be detected with reasonable certainty X l = X blank + k s blank Typically k=3 for LOD LOQ: concentration/quantity that can be quantified with reasonable certainty k=10 or 9 or 6 for LOQ (not a clear consensus) WHAT IS A REASONABLE CERTAINTY? ESTIMATING THE BLANK MEAN AND SD 5

6 LOD - Background Hypothesis testing - H 0 : the signal measured is not significantly different from the blank (the analyte is not present in the sample) - H 1 : the signal measured is different from the blank (the analyte is present in the sample) For a value of (e.g or 5%) the null hypothesis can be rejected or accepted If accepted, then the signal in the sample is accepted to be due to random variations in our determination 6

7 LOD - Background Normal (Gaussian) distribution - Type I ( ) errors: possibility of producing a false positive - Type II (β) errors: possibility of producing a false negative L c : critical level. Value with a probability of producing a false positive Figures from Boqué and Vander Heyden. LCGC Europe 22(2), 2009: 7

8 LOD - Background β=0.5 (50%) Let s imagine =0.05 L C =5 ng/l, LOD=5 ng/l? Figures from Boqué and Vander Heyden. LCGC Europe 22(2), 2009: 8

9 LOD - Background 3.28*S blank =β=0.05 If 3*S blank Then =β=0.07 If LOD is set to10ng/l? Figures from Boqué and Vander Heyden. LCGC Europe 22(2), 2009: 9

10 LOD - Background ISO LOD: the true net concentration (or quantity) of component in the material subject to analysis that will lead, with a probability (1-β), to the conclusion that the concentration (or quantity) of component in the material analyzed is greater than that of a blank sample Similar to an older IUPAC definition of minimum detectable value 10

11 2002/657/EC LOD - Background CC and CCβ Two possibilities: No permitted (regulated) limit Permitted (regulated) limit 11

12 2002/657/EC LOD - Background No permitted (regulated) limit Decision Limit: CC (=L c ) Detection capability: CCβ (=LOD) 12

13 LOD - Background 2002/657/EC Permitted (regulated) limit Decision Limit: CC ( L c ) Detection capability: CCβ ( LOD) Minimum Required Performance Level (MRPL) 13

14 Measuring the blank Blank Matrix similar to the sample NOT containing the analyte Not that easy 1. Sewage will likely contain the analyte: use something simpler 2. Ultrapure water may present a peak of the analyte: real blank signal (and problem) 3. If there is no peak what is the chromatographic signal? 4. We could also use standards and then use a factor to convert that to sample LOD by using some factors 14

15 Measuring the blank Case 2. We perform blanks and there are peaks Best scenario Measure the area several blank replicates (how many?) Calculate the mean and standard deviation Apply the formula(s) and stablish your LOD, LOQ, CC or CCβ X l = X blank + k s blank 2002/657/EC: n 20 15

16 Measuring the blank Case 1, 3-4. Peak from natural concentration or no peak Get a matrix similar to your sample containing a (or several) known concentration(s) of analyte The concentration(s) should be close to the LOD Use an indirect approach to estimate the mean and SD of the blank 16

17 The signal-to-noise approach European Pharmacopeia LOD: S/N=3 LOQ: S/N=10 Noise determined in a peak-to-peak basis (maximum amplitude) over a range of 20 times the width of the peak at half-maximum Figure from Boqué and Vander Heyden. LCGC Europe 22(2), 2009: 17

18 S/N Low Res systems (QQQ) Chromatogram Plot File: c:\varianws\data\tania\150317\ _p_2.5ppb.xms Sample: P_2.5ppb Operator: Scan Range: Time Range: min. Date: 17/03/ : kcounts 136.1>91.0 [-14.0V] S/N (PP): 44 S: N: AMP_ _P_2.5ppb.xms 91.0 (136.1>91.0 [-14.0V]) Filtered S/N (PP): 44 S: N: A Seg 1, Time: , Scan Functions: minutes Seg 1, Time: , Scan Functions: Scans In some instruments it is quite unclear how it is calculated Do it yourself! Remember to measure height! In this case noise is ca. 45,000 18

19 S/N Low Res Chromatogram Plots In low res systems you have to measure secondary ions 90 ng/l of amphetamine File: c:\varianws\data\tania\150317\ _s1_f2_3.xms Sample: S1_F2_3 Operator: Scan Range: Time Range: min. Date: 18/03/ :03 MCounts AMP_ _S1_F2_3.xms 91.0 (136.1>91.0 [-14.0V]) Filtered 136.1>91.0 [-14.0V] 1.00 From 1 st MRM: 0.75 S/N: 91 S/N (PP): 91 S: 1.1e+006 N: LOQ: ca. 10 ng/l 0.50 LOD: ca. 3.3 ng/l But 2 nd MRM should be visible: S/N 3 15 ng/l necessary kcounts 136.1>119.0 [-6.0V] AMP _S1_F2_3.xms (136.1>119.0 [-6.0V]) Filtered S/N (PP): 23 S: N: S/N: LOD=LOQ=15 ng/l Seg 1, Time: , Scan Functions: 17 minutes Scans 19

20 100 0 kcounts 343.2>245.0(-) [26.0V] 100 S/N Low Res THCCOOH_ _S1_F1_1.xms (343.2>245.0(-) [26.0V]) Filtered 170 ng/l of carboxy-thc S/N (PP): 450 S: N: 234 S/N: 450 LOQ ca 3 ng/l Is this realistic? Seg 1, Time: , Scan Functions: 6 Is it better to use another matrix with a lower concentration? Does it really matter anyway? minutes Scans 20

21 S/N High res 20 ng/ml standards of nitrosamines in GC-QTOF Noise can be zero! 21

22 S/N High res A wastewater containing 200 ng/l NDMA: m/z NMEA: m/z YES! Noise can be zero! Even with a complex matrix! 22

23 S/N High res The same 20 ng/ml standard in profile mode of acquisition The noise is here!! Why? 23

24 NDMA spectrum S/N High res 24

25 S/N High res Background spectrum 25

26 S/N High res Background spectrum 26

27 Background spectrum S/N High res The reality is not centroid! 27

28 Alternatives to S/N Basically two options: 1. Based on certainty, i.e. acceptable RSD. LOD/LOQ: analyte concentration where the standard deviation should be larger than 10% (IUPAC) or 20% (Eurachem) or variable (2002/657/EC depending on concentration level) RSD calculated from n 20 (IUPAC or 2002/657/EC) or n 10 (Eurachem) 2. Deriving the blank parameters indirectly from the calibration plot 28

29 Alternatives to S/N The nitrosamines example (NDEA): 4-replicates of standards in the 0.1/ ng/l range RSD relationship with concentration can be modelled, e.g.*: RSD = a 1 Conc (+b) ng/ml RSD % 0,1 0,3 55,7 0,5 4,1 1 5,7 2 4,0 3 3,4 5 5,4 10 1,9 20 2,0 *Eppe et al. Analytica Chimica Acta 519 (2004)

30 Alternatives to S/N The nitrosamines example (NDEA): R=0.88 ng/ml RSD % 0,1 0,3 55,7 0,5 4,1 1 5,7 2 4,0 3 3,4 5 5,4 10 1,9 20 2,0 RSD=10% LOQ= 1.2 ng/ml RSD=20% LOD= 0.6 ng/ml 30

31 Alternatives to S/N Deriving the blank parameters indirectly from the calibration plot (ISO ): Intercept blank signal S y/x : std. dev. of the estimate blank std. dev. Figure from J. Miller and J. Miller. Statistics and Chemometrics for Analytical Chemistry. 7 th Edition. Pearson. Essex, UK. 31

32 Alternatives to S/N The nitrosamines example (NDEA): ng/ml a=225 b=545 S y/x =269 R=0.997 Blank signal at LOD = *269 = 1032 LOD = 1032/545 = 1.9 ng/l 32

33 Alternatives to S/N The nitrosamines example (NDEA): ng/ml (X 1000,0) 4 Plot of Fitted Model NDEA = 168, ,629*Conc NDEA 3 2 a=169 b=599 S y/x =98 R= Conc Blank signal at LOD = *98 = 463 LOD = 463/599 = 0.77 ng/l Calibration range should be close to LOD and LOQ 33

34 My recommendations Report clearly how LOD/LOQs are calculated If you have blank problems, minimize them and use them for calculation of LOD/LOQ Use the S/N approach with a matrix similar to your samples But at a level closer to LOD Remember your confirmation ion(s) Otherwise use something else (e.g. river water or ultrapure water) and take into account matrix effects (and recovery) Use profile mode in HR-MS (if feasible) If there is no noise, use the RSD approach (with standards) and be pragmatic 34

35 My recommendations Check your calculations Check LODs/LOQs from time to time DO NOT GET MAD!! IT IS AN ESTIMATION! 35

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure

More information

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Ask The Expert Webinar Series EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Richard Burrows, Ph.D. Corporate Technical Director A Revision to the Method Detection Limit EPA

More information

VOTING DRAFT STANDARD

VOTING DRAFT STANDARD page 1 of 7 VOTING DRAFT STANDARD VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 Description This Voting Draft Standard is a proposed revision of the 2009 standard (EL-

More information

Limit of Detection and Its Establishment in Analytical Chemistry

Limit of Detection and Its Establishment in Analytical Chemistry Limit of Detection and Its Establishment in Analytical Chemistry Chang KH Forensic Science Programme, School of Health Sciences, Universiti Sains Malaysia, Kelantan, Malaysia ABSTRACT: Method validation

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

Application of Detection and Quantification Concepts to Chlorine Residual Measurements

Application of Detection and Quantification Concepts to Chlorine Residual Measurements Application of Detection and Quantification Concepts to Chlorine Residual Measurements Charles D. Hertz, Ph.D. PADEP Proposed Disinfection Requirements Rule Stakeholder Meeting Harrisburg, PA March 9,

More information

Estimating limit of detection for mass spectrometric analysis methods

Estimating limit of detection for mass spectrometric analysis methods DISSERTATIONES CHIMICAE UNIVERSITATIS TARTUENSIS 154 HANNO EVARD Estimating limit of detection for mass spectrometric analysis methods 1 DISSERTATIONES CHIMICAE UNIVERSITATIS TARTUENSIS 154 DISSERTATIONES

More information

Basic Statistics. 1. Gross error analyst makes a gross mistake (misread balance or entered wrong value into calculation).

Basic Statistics. 1. Gross error analyst makes a gross mistake (misread balance or entered wrong value into calculation). Basic Statistics There are three types of error: 1. Gross error analyst makes a gross mistake (misread balance or entered wrong value into calculation). 2. Systematic error - always too high or too low

More information

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4 page1 of 8 TNI Standard VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 January 2016 Description This TNI Standard has been taken through all of the voting stages and has

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

How to Describe Accuracy

How to Describe Accuracy OK, so what s s the speed of dark? When everything is coming your way, you're obviously in the wrong lane MARS 450 Thursday, Feb 14 2008 A) Standard deviation B) Student s t-test - Test of a mean C) Q-test

More information

-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics).

-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics). Chemometrics Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. -However, this definition can be expanded to include: biology (biometrics), environmental

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Validation of an Analytical Method

Validation of an Analytical Method Validation of an Analytical Method Refer to: ICH Guideline Q2(R1), Validation of Analytical Procedures: Teaxt andmethodology. Introduction All major laboratories eg. in the industry operates with clearly

More information

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline Experiment (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with,0- Phenanthroline Background The first step of an analytical spectrophotometric procedure for quantitative determination

More information

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Contact: Oliver el-atma Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Germany (OMCL

More information

Package chemcal. July 17, 2018

Package chemcal. July 17, 2018 Version 0.2.1 Date 2018-07-17 Package chemcal July 17, 2018 Title Calibration Functions for Analytical Chemistry Suggests MASS, knitr, testthat, investr Simple functions for plotting linear calibration

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Resval. Practical tool for the validation of an analytical method and the quantification of the uncertainty of measurement.

Resval. Practical tool for the validation of an analytical method and the quantification of the uncertainty of measurement. Workshop... Resval. Practical tool for the validation of an analytical method and the quantification of the uncertainty of measurement. Henk Herbold Marco Blokland Saskia Sterk General topics. *What s

More information

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert The LOD Procedure has changed Federal Register /Vol. 82, No. 165 /Monday, August 28, 2017 ACTION: Final rule. DATES: This regulation

More information

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of Analytical method validation By Juree Charoenteeraboon, Ph.D Analytical method Goal consistent, reliable and accurate data Validation analytical method serve the goal of analytical lmethod Its data reveals

More information

IMPROVE DETECTION AND QUANTITATION. Richard Burrows

IMPROVE DETECTION AND QUANTITATION. Richard Burrows WORKING WITH THE EPA TO IMPROVE DETECTION AND QUANTITATION Richard Burrows FEDERAL ADVISORY COMMITTEE ON DETECTION AND QUANTITATION ML First, a little history Lead in Albacore: Guide to Lead Pollution

More information

Method Validation and Accreditation

Method Validation and Accreditation SELAMAT Mycotoxins Workshop China, 11-15th December 2006 Method Validation and Accreditation Dr Hamide Z Şenyuva Senior Research Scientist TÜBİTAK-ATAL, TURKEY hamide.senyuva@tubitak.gov.tr SELAMAT Mycotoxins

More information

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp Validation of analytical methods Adrian Covaci Toxicological Center, University of Antwerp Why validate? The objective of validating an analytical method is to demonstrate its suitability for its intended

More information

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method) NRL for Cereals DTU National Food Institute Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method) Mette Erecius Poulsen & Hanne Bjerre Christensen 16 July 21

More information

Proposal: criteria for positive identification in LC-MS analysis

Proposal: criteria for positive identification in LC-MS analysis Proposal: criteria for positive identification in LC-MS analysis Lubertus Bijlsma, Maria Bronze, Emma Gracia Lor, Susana Simoes Any analytical methodology should comply with several quality requirements,

More information

Performance characteristics of analytical tests

Performance characteristics of analytical tests Performance characteristics of analytical tests Jaap-Willem Hutter 31-3-2011 1 CONTENT Background Performance characteristics Bottlenecks Detection limits, definitions procedures in various countries Some

More information

Estimation of LOQ for the Analysis of Persistent Organic Pollutants, in particular PCDD/Fs and PCBs

Estimation of LOQ for the Analysis of Persistent Organic Pollutants, in particular PCDD/Fs and PCBs 38th International Symposium on Halogenated Persistent Organic Pollutants (POPs) Thermo User Meeting - DFS Magnetic Sector GC-HRMS Krakow, August 26 31, 2018 Estimation of LOQ for the Analysis of Persistent

More information

Measurement Uncertainty: A practical guide to understanding what your results really mean.

Measurement Uncertainty: A practical guide to understanding what your results really mean. Measurement Uncertainty: A practical guide to understanding what your results really mean. Overview General Factors Influencing Data Variability Measurement Uncertainty as an Indicator of Data Variability

More information

Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT

Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT Determination of Ethylene Oxide (ETO) in Mainstream Cigarette Smoke Using Hydrobromic Acid Derivatization and Gas Chromatography-Mass Spectrometry Method Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT

More information

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 646 Triple Quadrupole LC/MS system with Triggered MRM Application Note Authors Thomas Glauner Agilent Technologies,

More information

Analytical method development and validation of gabapentin in bulk and tablet dosage form by using UV spectroscopic method

Analytical method development and validation of gabapentin in bulk and tablet dosage form by using UV spectroscopic method IJPAR Vol.6 Issue 2 April - June -2017 Journal Home page: www.ijpar.com ISSN:2320-2831 Research article Open Access Analytical method development and validation of gabapentin in bulk and tablet dosage

More information

Tom O Haver said, The quality of a signal is often

Tom O Haver said, The quality of a signal is often Electronically reprinted from October 2 What s the Most Meaningful Standard for Mass Spectrometry: Instrument Detection Limit or Signal-to-Noise Ratio? When evaluating the performance of mass spectrometers,

More information

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT Peter Hancock 1, Mike Hetmanski 2 and Richard J. Fussell 2 1 Waters Corporation, Atlas Park, Simonsway, Manchester M22 5PP, UK 2 Central

More information

MULTIVARIATE PATTERN RECOGNITION FOR CHEMOMETRICS. Richard Brereton

MULTIVARIATE PATTERN RECOGNITION FOR CHEMOMETRICS. Richard Brereton MULTIVARIATE PATTERN RECOGNITION FOR CHEMOMETRICS Richard Brereton r.g.brereton@bris.ac.uk Pattern Recognition Book Chemometrics for Pattern Recognition, Wiley, 2009 Pattern Recognition Pattern Recognition

More information

Highly Sensitive and Rugged GC/MS/MS Tool

Highly Sensitive and Rugged GC/MS/MS Tool Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction

More information

Detection and quantification capabilities

Detection and quantification capabilities 18.4.3.7 Detection and quantification capabilities Among the most important Performance Characteristics of the Chemical Measurement Process (CMP) are those that can serve as measures of the underlying

More information

DRAFT. The NELAC Institute Presents VERIFICATION OPTIONS FOR

DRAFT. The NELAC Institute Presents VERIFICATION OPTIONS FOR DRAFT The NELAC Institute Presents VERIFICATION OPTIONS FOR TNI Limit of Detection A laboratory's estimate of the minimum amount of an analyte in a given matrix that an analytical process can reliably

More information

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Proposed Procedures for Determining the Method Detection Limit and Minimum Level Proposed Procedures for Determining the Method Detection Limit and Minimum Level Published by: ACIL Environmental Services Section Technical Committee Revision 3.0 3/8/006 PROCEDURES These procedures set

More information

Really, A Revised MDL Procedure

Really, A Revised MDL Procedure NEMC 2016 Really, A Revised MDL Procedure Richard Burrows, Ph.D. TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure in the Federal

More information

Making the Transition From a Quantitation Lab to a QuantInformation Lab

Making the Transition From a Quantitation Lab to a QuantInformation Lab Making the Transition From a Quantitation Lab to a QuantInformation Lab Chris Holliman Global Small-Molecule Discovery Bioanalytical Phamacokinetics, Dynamics & Metabolism Acknowledgements Andre Negahban

More information

DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE

DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE Mehran SHARIFI, Peter JOZA, Bill RICKERT 69 th TOBACCO SCIENCE RESEARCH CONFERENCE

More information

Panayot Petrov, John Entwisle, Heidi Goenaga-Infante GC-ICP-MS reference methodology to quantify polybrominated flame retardants in environmental

Panayot Petrov, John Entwisle, Heidi Goenaga-Infante GC-ICP-MS reference methodology to quantify polybrominated flame retardants in environmental Panayot Petrov, John Entwisle, Heidi Goenaga-Infante GC-ICP-MS reference methodology to quantify polybrominated flame retardants in environmental waters relevant to the European Water Frame Directive.

More information

Measurement uncertainty and legal limits in analytical measurements

Measurement uncertainty and legal limits in analytical measurements UNCERTAINTY Measurement uncertainty and legal limits in analytical measurements MIRELLA BUZOIANU, Reference Materials Laboratory, National Institute of Metrology (INM), Bucharest, Romania 1 Introduction

More information

Instrumental methods of analysis

Instrumental methods of analysis Instrumental methods of analysis By Dr Hisham Ezzat Abdellatef Prof. of Analytical Chemistry Background: Analytical Chemistry: The Science of Chemical Measurements. Analyte: The compound or chemical species

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

OER implementation at UNS. Prof. Dr Ivana Ivančev-Tumbas University of Novi Sad, Faculty of Sciences

OER implementation at UNS. Prof. Dr Ivana Ivančev-Tumbas University of Novi Sad, Faculty of Sciences OER implementation at UNS Prof. Dr Ivana Ivančev-Tumbas University of Novi Sad, Faculty of Sciences Overview 1. Planned development of OER 2. What do we need from EU partners? 3. Draft OER example Overview

More information

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility Ultra-Fast Separation Ultra-Fast and Quantification Separation and of isobaric Barbiturates in Serum using LDTD- Quantification of isobaric MS/MS combined with differential mobility spectrometry Barbiturates

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch and Method Setup Outliers, Data Review, Reporting 1 MassHunter Quantitative Analysis

More information

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies,

More information

Business Statistics. Lecture 10: Course Review

Business Statistics. Lecture 10: Course Review Business Statistics Lecture 10: Course Review 1 Descriptive Statistics for Continuous Data Numerical Summaries Location: mean, median Spread or variability: variance, standard deviation, range, percentiles,

More information

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS PO-CON1518E A Method for Simultaneous Analysis of 174 Pesticides Pittcon 2015 2220-12P Feifei Tian, Xizhi Wang, Shiheng Luo, Guixiang Yang, Taohong Huang, Shin-ichi Kawano, Yuki Hashi, Shimadzu Global

More information

ENVIRONMENTAL analysis

ENVIRONMENTAL analysis ENVIRONMENTAL analysis Analyzing Wastewater Effluents for PAH s and PBDE s Using the Agilent 7000 Triple Quadrupole GC/MS Solutions for Your Analytical Business Markets and Applications Programs Authors

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry PO-CON1619E Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry Pittcon 2016 480-25 Yan Wang, Feifei Tian, Yueqi Li, Guixiang Yang, Taohong

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information.

Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. -However, this definition can be expanded to include: biology (biometrics), environmental science

More information

ORF 245 Fundamentals of Statistics Chapter 9 Hypothesis Testing

ORF 245 Fundamentals of Statistics Chapter 9 Hypothesis Testing ORF 245 Fundamentals of Statistics Chapter 9 Hypothesis Testing Robert Vanderbei Fall 2014 Slides last edited on November 24, 2014 http://www.princeton.edu/ rvdb Coin Tossing Example Consider two coins.

More information

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Column Xin Zhang, Jonathan P. Danaceau, and Erin E. Chambers Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Semi-validated

More information

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry APPLICATION NOTE Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry No. 10530 Authors: Cristian Cojocariu, 1 Maria José Farré,

More information

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017.

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017. GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG 90-2017 Adopted in 2017. CAC/GL 90-2017 2 OBJECTIVE 1. The purpose of these guidelines

More information

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Supporting Information. Sweetened Swimming Pools and Hot Tubs Supporting Information Sweetened Swimming Pools and Hot Tubs Lindsay K Jmaiff Blackstock, Wei Wang, Sai Vemula, Benjamin T Jaeger and Xing-Fang Li * Division of Analytical and Environmental Toxicology,

More information

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products Bruno Veyrand, Aline Brosseaud, Philippe Marchand, Fabrice Monteau, François Andre,

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

PERFORMANCE VERIFICATION IONSCAN LS

PERFORMANCE VERIFICATION IONSCAN LS Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--900 Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

FUTURE CONFIRMATORY CRITERIA

FUTURE CONFIRMATORY CRITERIA FUTURE CONFIRMATORY CRITERIA Revision CD 2002/657 Marco Blokland, Bjorn Berendsen, Hans Mol, Leen van Ginkel, Saskia Sterk REVISION 2 Revision 2002/657 EU Commission has explicitly asked for a revision

More information

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE Use objectives: state purpose for which results will be used. Specifications: How good should the numbers be

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

TNI V1M Standard Update Guidance on Detection and Quantitation

TNI V1M Standard Update Guidance on Detection and Quantitation TNI V1M4 2016 Standard Update Guidance on Detection and Quantitation GUID-3-109-Rev0 January 30, 2019 This material represents the opinion of its authors. It is intended solely as guidance and does not

More information

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE Agilent 49 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE AGILENT 49 TRIPLE QUADRUPOLE LC/MS SYSTEM WORK SMARTER, DO MORE ifunnel new Q1 Optics new Ion Detector IMPROVED SENSITIVITY

More information

REVIEW. Comparison of various international guidelines for analytical method validation

REVIEW. Comparison of various international guidelines for analytical method validation Formulation Development and Pharmacokinetic Laboratory, Pharmacy Group, Birla Institute of Technology and Science, Pilani, India Comparison of various international guidelines for analytical method validation

More information

CHEM 3420 /7420G Instrumental Analysis

CHEM 3420 /7420G Instrumental Analysis CHEM 3420 /7420G Instrumental Analysis Prof. Brian Gibney 2411 Ingersoll bgibney@brooklyn.cuny.edu Course Introduction Prof. Brian R. Gibney B.S. Chemistry (ACS Certified) Ph.D. Chemistry Brooklyn College,

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Chapter 7: Hypothesis Testing - Solutions

Chapter 7: Hypothesis Testing - Solutions Chapter 7: Hypothesis Testing - Solutions 7.1 Introduction to Hypothesis Testing The problem with applying the techniques learned in Chapter 5 is that typically, the population mean (µ) and standard deviation

More information

PSY 305. Module 3. Page Title. Introduction to Hypothesis Testing Z-tests. Five steps in hypothesis testing

PSY 305. Module 3. Page Title. Introduction to Hypothesis Testing Z-tests. Five steps in hypothesis testing Page Title PSY 305 Module 3 Introduction to Hypothesis Testing Z-tests Five steps in hypothesis testing State the research and null hypothesis Determine characteristics of comparison distribution Five

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

Keppel, G. & Wickens, T.D. Design and Analysis Chapter 2: Sources of Variability and Sums of Squares

Keppel, G. & Wickens, T.D. Design and Analysis Chapter 2: Sources of Variability and Sums of Squares Keppel, G. & Wickens, T.D. Design and Analysis Chapter 2: Sources of Variability and Sums of Squares K&W introduce the notion of a simple experiment with two conditions. Note that the raw data (p. 16)

More information

ST505/S697R: Fall Homework 2 Solution.

ST505/S697R: Fall Homework 2 Solution. ST505/S69R: Fall 2012. Homework 2 Solution. 1. 1a; problem 1.22 Below is the summary information (edited) from the regression (using R output); code at end of solution as is code and output for SAS. a)

More information

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications Doug Stevens Waters Corp. Introduction The range of GC/MS instrument types has expanded

More information

ELISA QUALITY ASSURANCE Analytical Phase

ELISA QUALITY ASSURANCE Analytical Phase LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-

More information

(ii) Scan your answer sheets INTO ONE FILE only, and submit it in the drop-box.

(ii) Scan your answer sheets INTO ONE FILE only, and submit it in the drop-box. FINAL EXAM ** Two different ways to submit your answer sheet (i) Use MS-Word and place it in a drop-box. (ii) Scan your answer sheets INTO ONE FILE only, and submit it in the drop-box. Deadline: December

More information

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) SOP-316 391 S. Vidal St., Sarnia, Ontario, N7T 7L1 Phone: (519) 344-4747 Fax: (519) 344-2350 E-Mail: info@lambtonscientific.com Method

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Chapter 10 Verification and Validation of Simulation Models. Banks, Carson, Nelson & Nicol Discrete-Event System Simulation

Chapter 10 Verification and Validation of Simulation Models. Banks, Carson, Nelson & Nicol Discrete-Event System Simulation Chapter 10 Verification and Validation of Simulation Models Banks, Carson, Nelson & Nicol Discrete-Event System Simulation The Black Box [Bank Example: Validate I-O Transformation] A model was developed

More information

Choosing Fit-For-Purpose Food Safety Methods

Choosing Fit-For-Purpose Food Safety Methods Choosing Fit-For-Purpose Food Safety Methods Outline Method selection Screening & quantification methods Extraction, cleanup and chromatography Detection methods Method criteria validation Conclusions

More information

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging Katerina Bousova, 1 Michal Godula, 2 Michele Suman 3 1 Thermo Fisher Scientific, Special Solutions

More information

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods Abstract The development of a GC/MS/MS method for the analysis of brominated flame retardants will be presented. Standard methods require the use of high resolution magnetic sector instrumentation, which

More information

Estimation of the detection and quantification limits of a method of analysis (Resolution Oeno 7/2000)

Estimation of the detection and quantification limits of a method of analysis (Resolution Oeno 7/2000) Estimation of the detection and quantification limits of a method of analysis (Resolution Oeno 7/2000) 1 - Purpose: to establish the detection and quantification limits of a method N.B. : The proposed

More information

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller

More information

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Application Note Forensic and Toxicology - Criminalistics Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Using the Agilent 126 Infinity II SFC and Detection by Coupled

More information

U.S. EPA Method 8270 for multicomponent analyte determination

U.S. EPA Method 8270 for multicomponent analyte determination ENVIRONMENTAL application note U.S. EPA Method 8270 for multicomponent analyte determination Elaine A. LeMoine and Herman Hoberecht Introduction Multicomponent analytes are compounds that yield several

More information

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350)

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350) Ann Strahl Technical Support Remediation Division TCEQ 512-239-2500 astrahl@tceq.state.tx.us 1 Data

More information

V. LAB REPORT. PART I. ICP-AES (section IVA)

V. LAB REPORT. PART I. ICP-AES (section IVA) V. LAB REPORT The lab report should include an abstract and responses to the following items. All materials should be submitted by each individual, not one copy for the group. The goal for this part of

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information