FUTURE CONFIRMATORY CRITERIA
|
|
- Neal Thornton
- 5 years ago
- Views:
Transcription
1 FUTURE CONFIRMATORY CRITERIA Revision CD 2002/657 Marco Blokland, Bjorn Berendsen, Hans Mol, Leen van Ginkel, Saskia Sterk
2 REVISION 2
3 Revision 2002/657 EU Commission has explicitly asked for a revision of 2002/657 Status of new document: not yet decided on EURLs in the area are tasked with making a proposal State of the art criteria for confirmatory analysis Should stand for a long time 3
4 EXAMPLE: Ambient Ionization-MS Ambient ionization methods permit direct ionization of samples outside the mass spectrometer in the open atmosphere There are >70 different types of AIMS described Minimal sample preparation No chromatography 4
5 What would this mean for new criteria? Since no chromatography is involved it will break the retention time criteria Without the use of retention time the number of false positives will increase?? Solution 1: The most selective MS measurement should be used in confirmatory analysis using AIMS Solution 2: Inclusion of another selective orthogonal technique 5
6 Ion Ratio 6
7 LC-MS/MS relative ion intensity In 2002/657 MS/MS criteria are based on expert opinions, at that time no scientific data available Experimental data is now available Relative intensities Allowed Dev. [rel] >50 % ± 20 % >20 % to 50 % ± 25 % >10 % to 20 % ± 30 % <10 % ± 50% Relative Intensity Berendsen et al: Drug Testing and Analysis, 2016, 8,
8 LC-MS/MS relative ion intensity The data shows deviations depend on the intensity of the less abundant ion From: Berendsen et al: Drug Testing and Analysis 2016, 8,
9 LC-MS/MS relative ion intensity The S/N of the lowest product ion is very important in obtaining stable ion ratio s Maximum allowed deviation a compromise between the false positive and false negative rate New guideline proposed ± 40 % (relative) for all relative ion intensity In case the precursor selection has a mass selection window of more than one Dalton (e.g. in case of Data Independent Acquisition) the technique is considered as full-scan confirmatory analysis. 9
10 Retention Time 10
11 Revision: Retention times Almost all current guidelines have a relative requirement (e.g. 2.5%) for the maximum deviation of the Rt between sample and standard Using gradient elution, empirical studies (Mol et al. Berendsen et al.) show deviation is absolute over the whole retention time range 11
12 Revision: Retention times From: Mol et al: Analytica Chimica Acta, 2015, 873, 1-13 From: Berendsen et al: Drug Testing and Analysis 2016, 8,
13 Fast chromatography: one-minute gradiënt 1 minute Oven: 60 C Typical run times seconds on 2-5 mm ID column Flow ml/min 13
14 Revision: Retention times All mass spectrometric analyses are combined with a sufficiently orthogonal separation (e.g. chromatography) or any separation technique that shows enough separation power for the specific application Possible to use of an absolute or relative criterion ± 0.1 minute for LC and GC In case retention time is below 1 minute then 5%, allowing the use of fast chromatography At least twice the retention time corresponding to the void volume of the column If an internal standard is used the relative retention time of the analyte, shall correspond to that of the calibration standard at a tolerance of ± 0.5 % for GC and ± 1 % for LC 14
15 Separation in the first dimension 1 n c 15
16 Coupling of second separation dimension 1 n c 16
17 Fully coupling of two separation techniques When truly orthogonal n c,2d = 1 n c x 2 n c 1 n c 17
18 Example added value of LC as orthogonal technique Kaabia, Z., Laparre, J., Cesbron, N., Le Bizec, B., & Dervilly-Pinel, G. (2018). Comprehensive steroid profiling by liquid chromatography coupled to high resolution mass spectrometry. The Journal of Steroid Biochemistry and Molecular Biology, 183,
19 Ion Mobility 19
20 Ion Mobility Mass spectrometry Ions separated by mass-to-charge ratio Ion mobility Ions separated by size, shape, and charge Size and charge: ions driven through gas filled chamber using electric field Shape: compact ions travel faster, less hindered by drift gas 20
21 Ion Mobility Drift times related to collisional cross sections (CCS) 21 Drift time
22 Collisional Cross Sections (CCS) CCS = Effective area of interaction between individual ion and neutral gas molecule Related to chemical structure and dimensions Can be obtained from experimental ion mobility data, and by molecular modelling Not all types of ion mobility yield CCS data Different protonation sites can yield different CCS values Boschmans et al., Analyst, 2016, 141,
23 Drift tube ion mobility (DT-IMS) Types of ion mobility High field asymmetric IMS (FAIMS) Differential Mobility Spectrometry (DIMS) Traveling wave ion mobility (TWIM) Trapped ion mobility (TIMS) López et al., New J. Chem., 2013, 37, Silveira et al., Analytical Chemistry, 2014, 86,
24 Drift tube ion mobility (DT-IMS) Types of ion mobility Agilent 6560 High field asymmetric IMS (FAIMS) Differential Mobility Spectrometry (DIMS) AB Sciex SelexIon Traveling wave ion mobility (TWIM) Waters Synapt Trapped ion mobility (TIMS) Bruker timstof López et al., New J. Chem., 2013, 37, Silveira et al., Analytical Chemistry, 2014, 86,
25 Separation of isomers: Diethylstilbestrol cis-des DTIMS trans-des Observed CCS difference ~ 5% TWIMS 25
26 Experimental determination of CCS value CCS can be calculated from drift time using Mason-Schamp equation: Ω / z = charge state of the analyteion e= charge of an electron N= number density of the drift gas µ= reduced mass of the ion-neutral pair k B = Boltzmann constant T = gas temperature K 0 = reduced mobility; mobility measured at standard temperature and pressure Only valid for drift tube instruments TWIMS and TIMS instruments need to be calibrated CCS calculation is not possible from DIMS data 26
27 CCS repeatability using TWIMS CCS of eight compounds determined on 7 different days using TWIMS instrument Compound Detected ion m/z Average Experimental CCS (Å 2, n=7) Cimaterol [M+H] % Clenbuterol [M+H] % Isoxsuprine [M+H] % THC [M+H] % RR-p-chloramphenicol [M-H] % THC-COOH [M-H] % Salmeterol [M+H] % Okadaicacid [M-H] % Repeatability is < 1% for all compounds, with the exception of RR-p-CAP (1.1%) RSD No significant effect of matrix observed Literature: <2%, inter-laboratory <5% Article in preparation 27
28 CCS as extra identification point? + Highly reproducible, no effect of matrix + Possibility to compare with literature or database + Reduction of matrix - Large deviation between DTIMS and TWIMS data found for some compounds -High correlation between CCS and m/z, which reduces uniqueness of CCS value Drift time (bins) β-agonists R² = m/z 28
29 Orthogonality of ion mobility Hernández-Mesa, M., Le Bizec, B., Monteau, F., García- Campaña, A. M., & Dervilly- Pinel, G. (2018). Collision Cross Section (CCS) database: An additional measure to characterize steroids. Analytical chemistry, 90(7), % error bars Only [M+H] + 29
30 Evaluation future Ion Mobility 1 cycle DTX-2 Okadaic acid 2 cycle 3 cycle 4 cycle 5 cycles 30
31 Serpentine UltralongPath with Extended Routing (SUPER) High Resolution Traveling Wave Ion Mobility-MS using Structures for Lossless Ion Manipulations (SLIM) 1 pass 13.5 m R= ms peak capacity > pass R=340 31
32 Examples of SLIM 32
33 Ion-mobility MS Cross-section additional confidence in confirmatory analysis Future alternative for chromatography? Criteria proposed: Cross-section is within 2 percent of a reference standard 0.5 identification point Not (yet) in compliance with orthogonality criteria 33
34 High Resolution Mass Spectrometry 34
35 Definition full scan hrms Current Definition of hrms: 657/2002 Resolution > 10,000 10% valley FDA FVM guidance Resolution > 10,000 FWHM SANTE/11945 not defined 35
36 full scan hrms Define resolving power, mass accuracy or both? Resolving power: Sante/11945: use enough resolving power to achieve a mass accuracy <5 ppm Mass accuracy: 657/2002: no mass accuracy requirement SANTE/11945: 5 ppm (< m/z mda) FDA FVM: MS: 5 ppm MS/MS: 10 ppm 36
37 Proposal: full scan hrms 2002/657 revision In high-resolution mass spectrometry (HRMS), the resolution shall typically be greater than 10,000 for the entire mass range at 10 % valley or 20,000 at FHWM. Only diagnostic ions with a relative intensity of more than 10 % in the reference spectrum of the calibration standard or MMS are suitable For confirmation of identity the mass deviation of all diagnostic ions should be below 5 ppm (or in case M/z <200 below 1mDa). Adducts and isotopes of selected diagnostic ions are excluded. In full scan mode, the isotopic pattern should match the molecular formula. 37
38 Suggestions for identification points Technique IPs earned Chromatography (GC or LC) 1 Ion mobility 0.5 LR-MS ion 1 Precursor ion selection at <±0.5 Da mass range 1 (indirect) LR-MS n product ion 1.5 HR-MS ion 1.5 HR-MS n product ion 2.5 Non MS techniques 1 38
39 Examples Technique(s) Separation Number of ions IP GC-MS (EI or CI) GC n GC-MS (EI and CI) GC 2 (EI) + 2 (CI) GC-MS (EI or CI) 2 derivates GC 2 (Derivate A) + 2 (Derivate B) LC-MS LC n (MS) GC-or LC-MS/MS GC or LC 1 precursor + 2 products GC-or LC-MS/MS GC or LC 2 precursor + 2 products GC-or LC-MS 3 GC or LC 1 precursor + 1 MS 2 product + 1 MS 3 product GC-or LC-HRMS GC or LC n IM-HRMS IM 3 GC-or LC-HRMS/MS GC or LC 1 precursor (<±0.5 Da mass range) + 2 products GC-or LC-HRMS and HRMS/MS GC or LC 1 full scan ion + 1 HRMS product ion a GC-and LC-MS GC and LC 2 ions (GCMS) + 1 ion (LCMS) 39
40 Examples Technique(s) Separation Number of ions IP GC-MS (EI or CI) GC n 1+n GC-MS (EI and CI) GC 2 (EI) + 2 (CI) 1+4 GC-MS (EI or CI) 2 derivates GC 2 (Derivate A) + 2 (Derivate B) 1+4 LC-MS LC n (MS) 1+n GC-or LC-MS/MS GC or LC 1 precursor + 2 products *1.5 = 5 GC-or LC-MS/MS GC or LC 2 precursor + 2 products *1.5 = 6 GC-or LC-MS 3 GC or LC 1 precursor + 1 MS 2 product + 1 MS 3 product = 5 GC-or LC-HRMS GC or LC n 1 + n*1.5 IM-HRMS IM *1.5 = 5 GC-or LC-HRMS/MS GC or LC 1 precursor (<±0.5 Da mass range) + 2 products *2.5 = 7 GC-or LC-HRMS and HRMS/MS GC or LC 1 full scan ion + 1 HRMS product ion a = 5 GC-and LC-MS GC and LC 2 ions (GCMS) + 1 ion (LCMS) = 5 40
41
WADA Technical Document TD2003IDCR
IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must
More informationWADA Technical Document TD2015IDCR
MINIMUM CRITERIA FOR CHROMATOGRAPHIC-MASS SPECTROMETRIC CONFIRMATION OF THE IDENTITY OF ANALYTES FOR DOPING CONTROL PURPOSES. The ability of a method to identify an analyte is a function of the entire
More informationPesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS
PesticideScreener Comprehensive Pesticide Screening and Quantitation Innovation with Integrity UHR-TOF MS The Challenge of Comprehensive Pesticide Residue Analysis The use of pesticides to reduce crop
More informationProposal: criteria for positive identification in LC-MS analysis
Proposal: criteria for positive identification in LC-MS analysis Lubertus Bijlsma, Maria Bronze, Emma Gracia Lor, Susana Simoes Any analytical methodology should comply with several quality requirements,
More informationMulti-residue analysis of pesticides by GC-HRMS
An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict
More informationIntroduction to GC/MS
Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules
More informationThe Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications
The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications Doug Stevens Waters Corp. Introduction The range of GC/MS instrument types has expanded
More informationAnalytical Technologies and Compound Identification. Daniel L. Norwood, MSPH, PhD SCĪO Analytical Consulting, LLC.
Analytical Technologies and Compound Identification Daniel L. Norwood, MSPH, PhD SCĪ Analytical Consulting, LLC. Extractables/Leachables Characterization (USP) Characterization is the discovery,
More informationMassHunter TOF/QTOF Users Meeting
MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks
More informationAnalyst Software. Peptide and Protein Quantitation Tutorial
This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document
More informationLC-MS Based Metabolomics
LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile
More informationAB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences
AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor
More informationTheory and Practical Consequences.
Theory and Practical Consequences. Number of hits depending on reachable mass accuracy 250 hits @ ±10 ppm 137 hits @ ±5 ppm 78 hits @ ±3 ppm 29 hits @ ±1 ppm 3 hits @ ±0.1 ppm 1 hit @ ± 0.05 ppm Intens.
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationGUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL
CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CONFIRMATORY TESTS CAC/GL 56-2005 When analyses are
More informationIdentification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate
Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and
More informationMaking Sense of Differences in LCMS Data: Integrated Tools
Making Sense of Differences in LCMS Data: Integrated Tools David A. Weil Agilent Technologies MassHunter Overview Page 1 March 2008 How Clean is our Water?... Page 2 Chemical Residue Analysis.... From
More informationTargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
TargetScreener A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS TargetScreener Get the Complete Picture Forensic laboratories are frequently required to perform
More informationAgilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012
1 Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012 Review: Technology for Accurate Mass Analysis: TOF LC/MS Mass measurements accurate to several
More informationQuantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system
Quantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system Kevin Mchale Thermo Fisher Scientific, San Jose CA Presentation Overview Anabolic androgenic
More information4-Methyldiethcathinone (4-MDEC)
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 4-Methyldiethcathinone (4-MDEC) Sample Type: Seized Material Latest Revision: May 18 th, 2018 Date Received: December 12 th, 2017 Date of Report: February
More informationIntroduction. Chapter 1. Learning Objectives
Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography
More informationQuantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays
: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability
More information5Cl-AKB48 and 5Br-AKB48
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 5Cl-AKB48 and 5Br-AKB48 5Cl-AKB48 Sample Type: Seized Material Latest Revision: May 18 th, 2018 Date Received: February 16 th, 2018 Date of Report: April
More information2-[[2-(4-ethyl-2,5-dimethoxy-phenyl)ethylamino]methyl]phenol
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 25E-NBOH Sample Type: Seized Material Latest Revision: May 18 th, 2018 Date Received: January 12 th, 2018 Date of Report: February 27 th, 2018 1. GENERAL
More informationQuantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.
Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationA Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance
A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,
More informationEPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS
EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of
More informationMass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS
Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from
More informationOverview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)
Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationProfiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry
Supporting Information for: Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid Chromatography-Tandem Mass Spectrometry Ramin Vismeh a,b, Fachuang Lu c,d, Shishir
More informationMassHunter Software Overview
MassHunter Software Overview 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks A
More informationLC-HRMS: Challenges for Routine Implementation
Joint EURL pesticides meeting October 23 rd -25th, 2013 Almeria Spain LC-HRMS: Challenges for Routine Implementation Richard J. Fussell Food and Environment Research Agency (York, UK) www.fera.defra.gov.uk
More informationChoosing the metabolomics platform
GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation
More informationAnalyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.
Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments Application Environmental Authors Jim Lau, Chin-Kai Meng, and Jennifer Gushue Agilent Technologies, Inc. 285 Centerville
More informationAssay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer
38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved
More informationN-(1-carbamoyl-2-methyl-propyl)-1-(5-chloropentyl)indazole-3- carboxamide
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 5Cl-AB-PINACA Sample Type: Seized Material Latest Revision: October 5, 2018 Date Received: June 6, 2018 Date of Report: October 5, 2018 1. GENERAL INFORMATION
More information- Supporting Information -
Protomers of Benzocaine: Solvent and Permittivity Dependence Stephan Warnke, Jongcheol Seo, Jasper Boschmans, Frank Sobott, James H. Scrivens, Christian Bleiholder,,? Michael T. Bowers, Sandy Gewinner,
More informationApplying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis
PO-CON1768E Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide ASMS 2017 TP-194 David Baker 1, Christopher Titman 1, Neil Loftus 1, Jonathan Horner
More informationMethyl 1-[(4-fluorophenyl)methyl]indazole-3-carboxylate
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 MFUBINAC Sample Type: Seized Material Latest Revision: October 30, 2018 Date Received: August 17, 2018 Date of Report: October 30, 2018 1. GENERAL INFORMATION
More informationConfirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System
Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System Claire Bramwell-German, Elliott Jones and Daniel Lebre AB SCIEX, Foster City, California
More informationOverview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada
LC-MS/MS Based Strategy for the Non-Targeted Screening of an Unlimited Number of Contaminants in Food Using the AB SCIEX TripleTOF 5600 System and Advanced Software Tools André Schreiber 1 and Yun Yun
More informationEPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS
Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the
More informationTHE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS
THE IMPLEMENTATION O A SCREENING WORKLOW OR ION MOBILITY QUADRUOPOLE TIME-O-LIGHT MASS SPECTROMETRIC ANALYSIS O POS ISOMERS Ken Rosnack ken_rosnack@waters.com Ingrid Ericson Jogsten and Bert van Bavel
More informationGAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions
GAS CHROMATOGRAPHY MASS SPECTROMETRY Pre-Lab Questions Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions are
More informationAtomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4
High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic
More informationAgilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis
Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis Technical Overview Introduction Metabolomics studies measure the relative abundance of metabolites
More informationChoosing Fit-For-Purpose Food Safety Methods
Choosing Fit-For-Purpose Food Safety Methods Outline Method selection Screening & quantification methods Extraction, cleanup and chromatography Detection methods Method criteria validation Conclusions
More informationThe Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System
The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System Bertram ieland 1 and Stephen Lock 2 1 AB SCIEX ieuwerkerk aan den Ijssel, The etherlands;
More informationHighly Sensitive and Rugged GC/MS/MS Tool
Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction
More informationApplication Note. Edgar Naegele. Abstract
Fast identification of main drug metabolites by quadrupole time-of-flight LC/MS Measuring accurate MS and MS/MS data with the Agilent 651 Q-TOF LC/MS and identification of main meta-bolites by comparison
More informationMaximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System
Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and
More informationDe Novo Metabolite Chemical Structure Determination. Paul R. West Ph.D. Stemina Biomarker Discovery, Inc.
De Novo Metabolite Chemical Structure Determination Paul R. West Ph.D. Stemina Biomarker Discovery, Inc. As a part of non targeted metabolomic analysis, when peaks in the mass spectra are thought to be
More informationChemistry Instrumental Analysis Lecture 37. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at
More informationNew Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses
New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller
More informationBruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward
Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry MAGLAB Summer School 2015 Ryan P. Rodgers Electrospray Ionization Source Drying Gas Capillary Evaporation Chamber High Voltage Supply Charged
More informationAll Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems
All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution
More information1-(1,3-benzodioxol-5-yl)-2-(ethylamino)hexan-1-one
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 N-ethyl Hexylone Sample Type: Seized Material Latest Revision: May 18 th, 2018 Date Received: February 12 th, 2018 Date of Report: April 19 th, 2018 1.
More informationApplication Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics
Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics Introduction Determining the structure of secondary metabolites
More information[ instrument specifications ]
Designed for leading researchers working at the limits of conventional mass spectrometry capabilities who need to further characterize and define their samples the Waters SYNAPT High Definition MS (HDMS
More informationINTRODUCTION OVERVIEW
COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies
More informationDetermination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry
PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole
More informationIncreasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer
Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance
More information4-methyl-1-phenyl-2-pyrrolidin-1-yl-pentan-1-one
NMS Labs 2300 Stratford Ave Willow Grove, PA 19090 alpha-pihp Sample Type: Seized Material Latest Revision: October 30, 2018 Date Received: August 17, 2018 Date of Report: October 30, 2018 1. GENERAL INFORMATION
More informationMS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies
MS n Analysis With Fast Polarity Switching in the Agilent Series LC/MSD Trap SL Application Note Christine Miller Introduction During mass spectrometric analysis, many compounds respond better to a particular
More informationSimplified Approaches to Impurity Identification using Accurate Mass UPLC/MS
Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION
More informationThe Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products
The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products Bruno Veyrand, Aline Brosseaud, Philippe Marchand, Fabrice Monteau, François Andre,
More informationFeatures or compounds? A data reduction strategy for untargeted metabolomics to generate meaningful data
TECHNICAL NOTE 65204 Features or compounds? A data reduction strategy for untargeted metabolomics to generate meaningful data Authors Amanda Souza and Ralf Tautenhahn Thermo Fisher Scientific Keywords
More informationIdentifying Disinfection Byproducts in Treated Water
Identifying Disinfection Byproducts in Treated Water Jonathan D. Byer, LEC Corporation; Saint Joseph, Michigan USA Susan D. Richardson, University of South Carolina; Columbia, South Carolina USA 1. Introduction
More informationThe Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity
The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High
More informationMethods for proteome analysis of obesity (Adipose tissue)
Methods for proteome analysis of obesity (Adipose tissue) I. Sample preparation and liquid chromatography-tandem mass spectrometric analysis Instruments, softwares, and materials AB SCIEX Triple TOF 5600
More informationAnalysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF
Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,
More informationSimplified Workflow and Improved Confidence in Screening Methods on Time of Flight Mass Spectrometry
Simplified Workflow and Improved Confidence in Screening Methods on Time of Flight Mass Spectrometry Steve Preece MS Business Manager Waters Wilmslow 2016 Waters Corporation 1 Overview Use of high resolution
More informationCerno Application Note Extending the Limits of Mass Spectrometry
Confident Compound Identification with High Mass Accuracy and Elemental Composition Specific NIST Library Search Identification of unknown compounds by unit mass resolution GC-MS systems essentially relies
More informationDesigned for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS
Designed for Accuracy High resolution quantitative proteomics Innovation with Integrity LC-MS Setting New Standards in Accuracy The development of mass spectrometry based proteomics approaches has dramatically
More informationImplementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis
An Executive Summary Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis The Exactive GC Orbitrap MS system can offer the specificity and sensitivity missing from
More informationAn Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining
An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining Dave Weil, Ph.D. and Jim Lau, Ph.D. Typical Method Conditions: 1260 UHPLC
More informationImproved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18
Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Column Xin Zhang, Jonathan P. Danaceau, and Erin E. Chambers Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Semi-validated
More informationUsing TOF for Screening and Quantitation of Sudan Red Colorants in Food Application
Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application Food Authors Yanyan Fang Agilent Technologies, Inc. 412 Ying Lun Road Pu Dong, Shanghai 200131 China Michael Zumwalt
More informationmaterials analysis Solutions for Your Analytical Business Markets and Applications Programs
materials analysis DETERMINATION OF DIISOCYANATES IN PLASTICS USED IN TEXTILE PRODUCTS USING THE AGILENT 6410B LC/MS/MS Solutions for Your Analytical Business Markets and Applications Programs Author Dr.
More informationSignal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationMethyl 2-[[1-(5-fluoropentyl)indole-3-carbonyl]amino]-3,3- dimethyl-butanoate
5F-MDMB-PICA Sample Type: Seized Material & Biological Fluid Latest Revision: July 31, 2018 Date of Report: July 31, 2018 Earliest Identification: November 2017 (Seized Material) 1. GENERAL INFORMATION
More informationMASS SPECTROMETRY. Topics
MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:
More informationUNIFI - Targeted and Untargeted Screening MS Software
UNIFI - Targeted and Untargeted Screening MS Software Dr David Hardy Senior Product Manager Mass Spectrometry Software 2011 Waters Corporation 1 Waters System Solutions - Available with UNIFI Biopharmaceutical
More informationFereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier
Chapter 2 Electrospray Ionization Traveling Wave Ion Mobility Spectrometry Mass Spectrometry for the Analysis of Plant Phenolics: An Approach for Separation of Regioisomers Fereshteh Zandkarimi, Samanthi
More informationAutomated and accurate component detection using reference mass spectra
TECHNICAL NOTE 72703 Automated and accurate component detection using reference mass spectra Authors Barbara van Cann 1 and Amit Gujar 2 1 Thermo Fisher Scientific, Breda, NL 2 Thermo Fisher Scientific,
More informationKey Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery
Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan
More informationAgilent ESI and APCI sources: for polar to non-polar compounds
1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas
More informationYun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION
Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark
More informationImpurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry
Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry Cristian Cojocariu and Paul Silcock Thermo Fisher Scientific, Runcorn,
More informationAccurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS
Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Application Note Authors E. Michael Thurman and Imma Ferrer Center for Environmental
More informationPerfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM
Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),
More informationHplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation
Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved We have made it easy for you to find a PDF Ebooks without any digging. And by having access
More informationHplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation
Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved We have made it easy for you to find a PDF Ebooks without any digging. And by having access
More informationRapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS
Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Application Note Forensic Toxicology Authors Martin Josefsson, and Markus Roman National Board of Forensic Medicine Linköping,
More informationAnalysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS
Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA
More informationTraditional Herbal Medicine Structural Elucidation using SYNAPT HDMS
Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS with Time-Aligned Parallel (TAP) Fragmentation Kate Yu, Jose Castro-Perez, and John Shockcor Waters Corporation, Milford, MA, U.S. INTRODUCTION
More information