DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE

Size: px
Start display at page:

Download "DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE"

Transcription

1 DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE Mehran SHARIFI, Peter JOZA, Bill RICKERT 69 th TOBACCO SCIENCE RESEARCH CONFERENCE September 20-23, 2015 Naples, Florida, USA LABSTAT INTERNATIONAL ULC. 262 Manitou Drive Kitchener, Ontario, Canada N2C 1L3 Phone: (519) Fax: (519) Presentation no 28

2 Objective To develop a sensitive and specific method for quantitative analysis of volatiles and hydroxy-nitrosamines in cigarette filler and mainstream smoke. Historical Methods GC-TEA GC-MS Method Development LC-MS/MS Method Approach Challenges Performance Advantage 2

3 Volatile Nitrosamines (Cigarette Filler) Historical Approach - GC/TEA 3 g sample Extract with Citrate-Phosphate Buffer + ISTD LPE (Hydromatrix) Concentrate ISTD 1 (NDiPA) ISTD 2 (NDHA) Std Solution KY ref. 3R4F NEMA NDEA NDPA NDBA NPIP NDMA Stabilwax (30 m x 0.32 mm x 1.0 µm) NPYR NMOR 3

4 20 cigarettes (Citrate/Phosphate Buffer + Filter Pad ) Add ISTD to the Filter Pad Extract (LPE) Buffer with DCM (3 X 50 ml) Volatile Nitrosamines (Mainstream) Historical Approach GC/TEA Extract Filter Pad with DCM extract ISTD 1 (NDiPA) - Filter - Concentrate Alumina Clean-up Concentrate ISTD 2 (NDHA) Fortified Blank NDMA KY ref. 3R4F Stabilwax (30 m x 0.32 mm x 1.0 µm) NEMA NDEA NDPA NDBA NPIP NPYR NMOR 4

5 Cigarette Filler 1 g sample + Internal Std (NDELA-d 8 ) N-Nitrosodialkanolamines Recent Approach - GC/MS Mainstream Smoke 20 cigarettes (NH 4 SO 3 NH 2 /H 2 SO 4 Buffer) Extracted with 50 ml Water Add Internal Standard (NDELA-d 8 ) Acidified with NH 4 SO 3 NH 2 /H 2 SO 4 Saturation with NH 4 SO 4 Derivatization (MSHFBA) Liquid Phase Extraction with Ethylformate:EtOH Solvent Evaporation (complete dryness) Sample Concentration GC-MS/NCI (NH 3 ) 5

6 What are the Drawbacks of Historical Methods? Detector Not as specific as Mass Spectrometric techniques GC-TEA (VNA) Solvent (CH 2 Cl 2 ) Health risk, low vapour pressure Chlorinated waste disposal (environment) Inconvenient for GC polar column (frequent maintenance) GC-MS (HO-NA) Derivatization (MSHFBA) Ionization (NCI/NH 3 ) Silylation step requires an aprotic environment (complete dryness prior to the addition of the reagent). CI requires frequent cleaning of the source. Both Methods Chromatographic attributes Analytes peaks exhibit low S:N ratios in samples Complex and highly variable baselines (difficulties to identify and integrate nitrosamine peaks). 6

7 Method Development Approach / Challenges Ionization Efficiency Mobile Phase Signal Suppression Matrix Effect Interference Reagent Purity 7

8 Acetonitrile : Formic Acid (1%) Acetonitrile : Ammonium Acetate (2mM) Method Development - LC Mobile Phase and Ionization Efficiency Methanol : Ammonium Acetate (2mM) Methanol : Formic Acid (1%) Acetonitrile : Formic Acid NDMA NEMA Methanol : Ammonium Acetate NDEA 6 5e e Intensity (counts) CH 3 CN : Amm. Acet time (minutes) Intensity (counts) Time,min time (minutes) Intensity (counts) Time,min time (minutes) 8

9 Method Development - LC Mobile Phase and Ionization Efficiency Response Factor = fct (mobile phase composition) S:N Ratios NDMA NMOR NPYR NDEA NPIP NEMA CH 3 CN : HCOOH CH 3 CN : CH 3 COO -,NH 4 + CH 3 OH : CH 3 COO -,NH 4 + CH 3 CN : HCOOH combination appeared to be the best compromise for all analytes. 9

10 Method Development Reagent Water Purity Background Level 3.0 NDMA (m/z 75 43) Intensity (counts) NDMA (m/z 75 43) 1% FA solution in type I H 2 O (ASTM D ) Time,min time (minutes) 1% FA solution in distilled H 2 O 10

11 Method Development Matrix Effect & Signal Suppression Lab. No Matrix Fortified (n Blank = 4) Lab. Matrix Fortified (n = Matrix 8) Signal Suppression 87% 100% 90% 80% Intensity (counts) % 60% 50% 40% % 7% 9% D6-NDMA 6 DD8-NPYR 8 D10-NDEA D D14-NDPA 10 -NDEA D 14 30% 20% 10% 0% 11

12 Intensity (counts) Fortified KY 3R4F 7. 3 Method Development Signal Suppression - NDMA ) 9.91 Open Scan (m/z ) Intensity (counts) me,mi n Ti Q p ( ( ) ) ( ) Fortified Reagent Blank MS/MS-MRM (m/z 75 43) minutes minutes me,mi n Ti Further clean-up or better LC separation? 12

13 Method Development Reducing Signal Suppression - LC LC C18 (LC) NEMA NDEA Improved separation of polar compounds NDMA NMOR NDiPA NDPA Intensity (counts) me,mi n Ti (p ) p ( ) p ( ) p C18 + HILIC (LC LC) NDMA NEMA NMOR NDiPA n me,mi Ti NDEA NDPA NDBA D 14 -NDBA NDBA Reduced signal suppression

14 Findings from Method Development Background Levels NDMA content in deionized water NPYR found in Ethylformate (removal of NDMA from ethylfomate) NMOR/NDBA in Hydromatrix (removal of NDMA from ethylfomate) Detection (APCI + -MS/MS-MRM) Mobile Phase choice LC X LC separation Scope Direct/simultaneous VNA/HO-NA analysis in Mainstream Emission HO-NA in Cigarette Filler requires additional clean-up step 14

15 Resulting Methodology (Mainstream Emission) Sample Generation 10 Cigarette smoked (Sulfamate buffer traps + glass fiber filter disc) Sample Extraction Extract pad with trap solutions (shake for 30 minutes). Acidify a 12 ml-aliquot of the extract with ammonium sulphamate/h 2 SO 4 Saturate the extract with (NH 4 ) 2 SO 4 Sample Clean-up Clean-up on hydromatrix cartridge (20g) Elute with 70 ml ethyl formate ethanol (98:2, v/v) Evaporate to 1 ml (Rotavap) Sample Analysis LC-APCI + -MS/MS-MRM (40 minutes) 15

16 Resulting Methodology (Cigarette Filler) Cigarette Filler 1.5 g sample + Internal Std mixture N-Nitrosodialkanolamines Volatile Nitrosamines Extract in 50 ml distilled water Acidify to ph~2 (NH 4 SO 3 NH 2 /H 2 SO 4 ) Saturate with Ammonium Sulfate Liquid Phase Extraction (Ethylformate:EtOH) Concentration Florisil clean-up (MeOH ) Concentration VNA Analysis NDELA/NDiPLA Analysis 16

17 Analytical Performance Validation Data Analytical Range Sensitivity Precision/Recovery 17

18 Analytical Calibration range: Analytical Performance Method Linearity Mainstream Smoke : ng/cig Cigarette Filler: ng/g

19 Method Recovery & Precision Recovery (%) Analytical Performances (Cigarette Filler) Precision (%) NDMA 103±9 8.6 NEMA 95±8 8.5 NDEA 102±9 8.7 NDiPA 95±4 4.6 NDPA 100±6 5.6 NPYR 102±5 4.8 NMOR 96±7 7.0 NPIP 103±8 7.8 NDBA 102±7 6.7 NDELA 103± NDiPLA 97± Method Sensitivity: LOQ Values LC-MS (ng/g) GC-MS (ng/g) GC-TEA (ng/g) NDMA NEMA NDEA NDiPA NDPA NPYR NMOR NPIP NDBA NDELA NDiPLA

20 Method Recovery & Precision Recovery Precision (%) (%) NDMA 100±6 5.9 NEMA 103±3 3.1 NDEA 102±5 5.0 NDiPA 98 ±7 6.9 NDPA 111±6 5.5 NPYR 102±5 4.7 NMOR 100± NPIP 107±6 5.9 NDBA 110± NDELA 102±5 4.6 Analytical Performances (Mainstream Emission) Method Sensitivity: LOQ Values LC-MS (ng/cig) GC-MS (ng/cig) GC-TEA (ng/cig) NDMA NEMA NDEA NDiPA (ISTD) NDPA NPYR NMOR NPIP NDBA NDELA

21 Advantages of Current LC-MS/MS over Historical Methods Unequivocal Mass Spectral Identification (higher accuracy) The LC-MS/MS technique is a quantitative, sensitive, and reliable method for analysis of volatiles and hydroxy-nitrosamines Higher Recovery ( absolute ) in tobacco, cigarette smoke and new generation of tobacco products. Higher Precision (use of isotopically-labelled analogues as ISTD) Higher Sensitivity (higher S:N ratios in samples) Significantly cleaner chromatograms : providing ease of peak integration minimizing the risk of peak misidentification. Safer Lab./personnel Safety (CH 2 Cl 2 ) 21

22 Acknowledgment Thank you for Your Attention!

Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT

Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT Determination of Ethylene Oxide (ETO) in Mainstream Cigarette Smoke Using Hydrobromic Acid Derivatization and Gas Chromatography-Mass Spectrometry Method Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT

More information

TSNAs in Tobacco Filler and Cigarette Smoke

TSNAs in Tobacco Filler and Cigarette Smoke TSNAs in Tobacco Filler and Cigarette Smoke Chris Wright Carol Goss Magdalena Minnion British American Tobacco Group R&D UK US FDA Center for Tobacco Products Scientific Workshop 30-31 July 2013 1 A. Describe

More information

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Contact: Oliver el-atma Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Germany (OMCL

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

CORESTA Recommended Method No. 86

CORESTA Recommended Method No. 86 Cooperation Centre for Scientific Research Relative to Tobacco Tobacco and Tobacco Products Analytes Sub-Group CORESTA Recommended Method No. 86 DETERMINATION OF SELECT CARBONYLS IN TOBACCO AND TOBACCO

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples Technical Note 101 Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples This technical note includes by specific information on the extraction

More information

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

NEW LC-MS/MS METHOD FOR THE DETERMINATION OF EIGHT NITROSAMINES IN DRINKING WATER

NEW LC-MS/MS METHOD FOR THE DETERMINATION OF EIGHT NITROSAMINES IN DRINKING WATER Journal of Environmental Protection and Ecology 17, No 1, 74 82 (2016) Water pollution NEW LC-MS/MS METHOD FOR THE DETERMINATION OF EIGHT NITROSAMINES IN DRINKING WATER T. GALAON*, L. CRUCERU, J. PETRE,

More information

TÜV NORD Umweltschutz Sector: Hazardous substances

TÜV NORD Umweltschutz Sector: Hazardous substances TÜV NORD Umweltschutz Sector: Hazardous substances Umweltschutz S U M M AR I S E D R E P O R T on the analysis of a test liquid for the presence of harmful substances Client: Analytical laboratory: Happy

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Column Xin Zhang, Jonathan P. Danaceau, and Erin E. Chambers Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Semi-validated

More information

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry Shijun Lua, Buu N. Trana, Jamie L. Nelsenb, Kenneth M. Aldousa. Journal of Chromatography

More information

CORESTA Recommended Method No. 83

CORESTA Recommended Method No. 83 Cooperation Centre for Scientific Research Relative to Tobacco Special Analytes Sub-Group CORESTA Recommended Method No. 83 DETERMINATION OF AMMONIA IN MAINSTREAM CIGARETTE SMOKE BY ION CHROMATOGRAPHY

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

Method - Determination of aromatic amines in mainstream cigarette smoke

Method - Determination of aromatic amines in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of aromatic amines in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicaple to quantitative determination

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Application Note AN772 Introduction This application note describes the extraction of cocaine and a full range

More information

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS 1.0 Purpose - This procedure specifies the required elements for the extraction and quantitation of THC,

More information

METHANOLYSIS OF ACETAL

METHANOLYSIS OF ACETAL Chem 367-2/ Methanolysis of Acetal 25 METHANOLYSIS OF ACETAL 1. Purpose Diethyl acetal undergoes an acid-catalyzed reaction with methanol in two steps, as follows: k 1 CH 3 CH(OEt) 2 + MeOH CH 3 CHOEtOMe

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV* Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV* EPA Method 8315A UCT Part Number: EUC1812M15 (Unendcapped C18-2000 mg/15 ml cartridge) March 2013 Method

More information

HPLC solvents and mobile phase additives

HPLC solvents and mobile phase additives HPLC solvents and mobile phase additives The following is intended to be a guide for the choice of solvent and mobile phase additives. Of course the choice of solvents for LC will be dictated primarily

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Smoke Analytes Sub-Group (SMA) Report 2018

Smoke Analytes Sub-Group (SMA) Report 2018 Smoke Analytes Sub-Group (SMA) Report 2018 CORESTA Congress Science and Innovation: Addressing the Needs Kunming, October 25, 2018 SMA SG Objectives To propose and maintain CORESTA Recommended Methods

More information

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent

More information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydroquinone in cosmetic products. 2. PRINCIPLE Hydroquinone is identified by thin layer chromatography

More information

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC.

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC. Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC. Abstract Chetna Mittal, PhD, Asha Oroskar, PhD,, Anil Oroskar, PhD Orochem Technologies Inc. Lombard, IL,

More information

Fast Analysis of Chlorinated Phenoxy Herbicides in Environmental Samples by Negative ESI LC-MS/MS 14337

Fast Analysis of Chlorinated Phenoxy Herbicides in Environmental Samples by Negative ESI LC-MS/MS 14337 Fast Analysis of Chlorinated Phenoxy Herbicides in Environmental Samples by Negative ESI LC-MS/MS 14337 Wenjun Han, Ben Hicks, Elaine Wild TestAmerica Laboratories Inc. 13715 Rider Trail North, Earth City,

More information

Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns

Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns Name of Procedure: Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns Suggested Uses: This is an extraction procedure for base drugs. The

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all

More information

Choosing Fit-For-Purpose Food Safety Methods

Choosing Fit-For-Purpose Food Safety Methods Choosing Fit-For-Purpose Food Safety Methods Outline Method selection Screening & quantification methods Extraction, cleanup and chromatography Detection methods Method criteria validation Conclusions

More information

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA

2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA An Ion Chromatographic Method for the Quantitative Determination of Hydrogen Cyanide in Cigarette Smoke using Pulsed Amperometric Detection Dr W Guthery and Dr M J Taylor Filtrona Technology Centre CORESTA

More information

16 Malachite green 17 Leucomalachite green

16 Malachite green 17 Leucomalachite green 16 17 Leucomalachite green hydrochloride Leucomalachite green N N N + Cl - N N-[4-[[4-(dimethylamino)-phenyl] phenylmethylene]-2,5-cyclohexadien-1-ylidene ]-N-methylmethanaminium chloride C 23 H 25 ClN

More information

Validation of TNCO, TSNAs, and PAHs

Validation of TNCO, TSNAs, and PAHs Validation of TNCO, TSNAs, and PAHs Chris Wright British American Tobacco Group R&D UK US FDA Center for Tobacco Products Scientific Workshop 30-31 July 2013 1 General Method Testing for TNCO, TSNAs and

More information

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensic Toxicology Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner,

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application Environmental Authors Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville Road Wilmington,

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis Keywords: Automated GX-271 GPC Clean-up System, Contaminants,

More information

NICKEL-63/59 IN WATER

NICKEL-63/59 IN WATER Analytical Procedure NICKEL-63/59 IN WATER 1. SCOPE 1.1. This is a method for the separation and measurement of nickel- 63/59 in water samples. 1.2. This method does not address all aspects of safety,

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of

More information

The Analysis of Organophosphate Pesticides by LC/MS Application

The Analysis of Organophosphate Pesticides by LC/MS Application The Analysis of Organophosphate Pesticides by LC/MS Application LC-MS Authors Paul Zavitsanos Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Paul Yang MOE Ontario Canada

More information

*Correspondence to:

*Correspondence to: Supporting Information for Carbonate-promoted hydrogenation of carbon dioxide to multi-carbon carboxylates Aanindeeta Banerjee 1 and Matthew W. Kanan 1 * 1 Department of Chemistry, Stanford University,

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB31604.35-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB 31604.35-2016

More information

Physical Separations and Chromatography

Physical Separations and Chromatography Lab #5A & B: Physical Separations and Chromatography Individual Objectives: At the end of these experiments you should be able to: Ø Distinguish between Rf and tr; chromatograph and chromatogram; adsorption

More information

This method describes the identification of the following prohibited colorants in cosmetic products:

This method describes the identification of the following prohibited colorants in cosmetic products: A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION This method describes the identification of the following prohibited colorants in cosmetic products: Names C I number Pigment Orange 5 12075 Metanil

More information

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges Application ote Food Testing & Agriculture Author Andy Zhai Agilent Technologies Shanghai Co. Ltd. Abstract This application

More information

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD)

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD) ISO 2001 All rights reserved HORIZONTAL 13.0 April 2006 Second draft Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection

More information

Basics in Solid Phase Extraction (SPE)

Basics in Solid Phase Extraction (SPE) Basics in Solid Phase Extraction (SPE) Claudia Thiele, Sigma Aldrich Taufkirchen Euroanalysis 2011 Belgrade Lunch Symposium 13.09.2011 Agenda The Importance of Sample Prep Overview of SPE Technology SPE

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

Supporting Information. Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and

Supporting Information. Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and 1 2 3 Supporting Information Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and Estrogens in Drinking Water of China: Comparison to the Parent Compounds 4 5 6 7 8 1 Laboratory

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.35-2016 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard Determination

More information

Impurities Sources, Detection and Measurement HPLC-MS Analyses. Dr Courtney Milner Chemical Analysis

Impurities Sources, Detection and Measurement HPLC-MS Analyses. Dr Courtney Milner Chemical Analysis Impurities Sources, Detection and Measurement HPLC-MS Analyses Dr Courtney Milner Chemical Analysis Chemical Analysis Integrated Analytical Resource (Physical, Chemical and Microbiological) that has facilities

More information

Definitive EtG/EtS LC-MS/MS Analysis:

Definitive EtG/EtS LC-MS/MS Analysis: Definitive / LC- Analysis: A Rugged 4-Min Method for High-Throughput Labs By Justin Steimling and Frances Carroll Abstract Methods for monitoring alcohol consumption biomarkers and are generally limited

More information

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract. Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC) Document A. THIN LAYER CHROMATOGRAPHY (TLC) 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydrocortisone acetate, dexamethasone, betamethasone, betamethasone 17-valerate

More information

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009) February 2009. NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009) This monograph was adopted at the Forty-third WHO Expert Committee on Specifications

More information

2017 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide

2017 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide 217 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide O O Cl NH 3 NH 2 C 9 H 7 ClO (166.6) (17.) C 9 H 9 NO (147.2) Classification Reaction types and substance classes reaction of

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Experiment 1: Thin Layer Chromatography

Experiment 1: Thin Layer Chromatography Experiment 1: Thin Layer Chromatography Part A: understanding R f values Part B: R f values & solvent polarity Part C: R f values & compound functionality Part D: identification of commercial food dye

More information

METHOD 3600B CLEANUP

METHOD 3600B CLEANUP METHOD 3600B CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

Determination of Phenoxyacid Herbicides in Water by Solid Phase Extraction and LC-MS/MS Detection

Determination of Phenoxyacid Herbicides in Water by Solid Phase Extraction and LC-MS/MS Detection Determination of Phenoxyacid Herbicides in Water by Solid Phase Extraction and LC-MS/MS Detection UCT Part Numbers: ECHLD156-P (Enviro Clean HL DVB 5mg/6mL, PE Frits) VMFSTFR12 (Sample Transfer Tubes)

More information

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized

More information

A Simple SPE Method for the Determination of Malachite Green, Crystal Violet and Other Synthetic Dyes in Seafood

A Simple SPE Method for the Determination of Malachite Green, Crystal Violet and Other Synthetic Dyes in Seafood A Simple SPE Method for the Determination of Malachite Green, Crystal Violet and Other Synthetic Dyes in Seafood UCT Part Numbers CSDAU206 Clean Screen DAU 200 mg/6ml Column SLC-18100ID21-3UM Selectra

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma White Paper A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma Lars Neudert, MSc, Senior Scientist Method Development

More information

Archived Content. This content was archived on June 24, 2013.

Archived Content. This content was archived on June 24, 2013. This content was archived on June 24, 2013. Archived Content Information identified as archived on the Web is for reference, research or recordkeeping purposes. It has not been altered or updated after

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System Application Note Authors Wenwen Wang, Zhe Cao Agilent Technologies, Inc. No.3, Wang Jing Bei Lu, Chao Yang District,

More information

METHOD 3600C CLEANUP

METHOD 3600C CLEANUP METHOD 3600C CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

TECHNETIUM-99 IN SOIL

TECHNETIUM-99 IN SOIL Analytical Procedure TECHNETIUM-99 IN SOIL 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in soil. 1.2. This method does not address all aspects of safety, quality

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 129 Infi nity LC, and 646 Triple Quadrupole LC/MS Application Note Clinical Research Authors Linda Côté and Christophe

More information

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS Liquid Chromatography Mass Spectrometry SSI-LCMS-096 Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS LCMS-8045 Summary: A short, robust quantitative method for the analysis

More information

METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION

METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION 1.0 SCOPE AND APPLICATION 1.1 This method describes a procedure for isolating organic compounds from aqueous samples. The method also describes concentration

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System Bertram ieland 1 and Stephen Lock 2 1 AB SCIEX ieuwerkerk aan den Ijssel, The etherlands;

More information