Trace ion chromatography

Size: px
Start display at page:

Download "Trace ion chromatography"

Transcription

1 Trace ion chromatography The R&D Notebook 7 A publication of the Lab Water Division of EMD Millipore Stéphane Mabic, Ichiro Kano, and Daniel Darbouret Research & Development, Lab Water Division, EMD Millipore, Saint-Quentin-en-Yvelines, France. Abstract : Ion chromatography (IC) has been recognized as one of the most competitive techniques for trace analysis, based on results from continuing research and development work into different types of instrumentation. Ultratrace analysis requires the use of extremely clean reagents, including water. The ionic content of water was monitored by IC along a water purification chain, from tap water to ultrapure water. Anion and cation concentrations decrease from ppm levels, in the tap feedwater, to ppt levels, in ultrapure water. This study shows that ultrapure water from a Milli-Q system is suitable for trace analysis by IC. Introduction Due to the increased sensitivity of instrumentation, analytical detection limits can now reach ng/l (ppt) levels on a routine basis, and even pg/l (ppq) levels under certain conditions. Confidence in the data recorded for these trace analyses relies on the control of several conditions and parameters. Besides the technical performance of instruments, careful sample handling and reagent purity are the main parameters that can be controlled. 1 Water is one of the most important reagents because it is used in so many applications. Rinsing of liquid chromatography chains, dilution and preparation of blanks and samples, and final elution are just some examples where water is used. Because of this broad range of applications and the volumes required, extreme care must be taken with water quality. In particular, in trace ion analysis, constant and reproducible high purity water, free of residual ions, is necessary. Several methods are available to analyze ions with varying degrees of sensitivity: Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES), Inductively Coupled Plasma Mass Spectrometry (ICP-MS), 2,3 Flame Atomic Absorption Spectroscopy (FAAS), Electrical Atomic Absorption Spectroscopy (GFAAS) 4 and Ion Chromatography (IC) combined with UV-Vis spectroscopy or electrical conductimetry. 5 Ion chromatography, combined with conductivity measurement, has become a reliable and popular method for analyzing a large variety of anions and cations in dilute samples. 6,7 This method has been developed in several research areas, such as the study of Antarctic ice, 8 in the semi-conductor industry, 9 and in analytical research domains, such as environmental analyses of air, 10 soil, 11 and water. 12,13 The need for qualification and validation of drinking water, in order to meet new standards and regulations, has also encouraged more widespread use of IC, and the development of novel analytical techniques. There is a real challenge involved in developing a water purification system that can produce suitable water quality for trace analysis, and can monitor this water quality on demand, in compliance with Good Laboratory Practices. In this study, concentrations of the major ions detected in tap water were studied by IC at various stages of the water purification chain. EMD Millipore is a division of Merck KGaA, Darmstadt, Germany

2 Materials and methods Water purification chain A water purification chain used to produce water for analytical applications employs a combination of several technologies. The chain can be divided into two basic parts; pretreatment, involving an initial purification system, such as the Elix system, which produces pure water that is fed to a storage reservoir, and a final polishing unit, such as the Milli-Q system, which produces ultrapure water from the reservoir water 14 (Figure 1). The pretreatment system used here, the Elix 10 system, combines sequential water purification by filtration/adsorption, reverse osmosis (RO), and electrodeionization (EDI). The initial purification step in the Elix system employs a Progard pretreatment pack, which is located before the RO membrane, for several important reasons. Activated carbon beads in the Progard cartridge prevent damage to the RO membrane from chlorine (Cl 2 ) present in tap water, and remove some dissolved organics by adsorption. Polyphosphates, polymeric sequestering anti-scaling agents, act as an efficient chelating agent for Ca 2+ and Mg 2+ ions, 15 protecting the RO membrane and the EDI module from scaling. Finally, a 0.5 μm nominal pore size polypropylene depth filter effectively removes large colloids, silt, and particulates. This pretreated water is then processed through a spiral, thin-film, composite RO membrane, which reduces the ion content significantly. RO permeate water is directed to an EDI module, which combines ion-exchange membranes, ion-exchange resins, and a direct electrical current to remove ions from the water. 14 The current continuously regenerates the ion-exchange resins in the system avoiding the need for resin replacement. The Elix unit used in these experiments delivered ten liters per hour of purified water from tap water to a storage reservoir. This reservoir, made of specifically selected polyethylene, 16 fed the final polishing unit to remove traces of residual ions. The Milli-Q Element system was used here as a final polishing system, and delivered ultrapure water via sequential photooxidation, ion exchange, and filtration (Figure 2). 17,18 The photooxidation process employs a UV lamp ( = 185 nm and 254 nm), causing organic chelating molecules to break up, and inducing a release of chelated ions. Breaking up the organics also helps to avoid organic fouling of the resin beads. Ions are then sequentially removed by two mixed-bed resins made of low-leaching materials. A microfiltration membrane was placed at the point-of-use of the polishing system for particulate removal and to prevent reverse contamination from the external environment. The flow schematic depicted in Figure 1 shows the purification chain and the various sampling ports. Samples were collected at each purification step in polyethylene (PE) bottles 19 that were previously rinsed thoroughly with ultrapure water. Sampling ports were selected to monitor the efficiency of each purification step in the following places: from the tap water, the RO reject and permeate water, the EDI concentrate and product water, the reservoir water, and the final ultrapure water. Figure 1 : Water purification chain and sampling ports. Figure 2 : Milli Q Element Ultrapure Water System (UHMWPE filter - Ultra-High Molecular-Weight Polyethylene, PVDF final filter - Polyvinyldiene fluoride) Ion chromatography Ion analysis was performed using a Dionex DX 500 apparatus with a GP50 pump and an AS40 autosampler, connected to a CD20 conductivity meter with peaknet software. Sample volumes of 25 μl were used for a 50 to 5,000 ppb range, and 200 μl for a 5 to 1,000 ppb range. For anion chromatography, AS12A/AG12A columns with 2.7 mm Na 2 CO 3 /0.3 mm NaHCO 3 eluents, and a 1.5 ml/min flow rate, were used. An ASRS Ultra suppressor, at 100 ma, with an external water feed, was chosen. 2

3 For cation analysis, CS12A/CG12A columns, with 20 mm methanesulfonic acid eluents and 1.0 ml/ min flow rates, were used. In this case a CSRS Ultra suppressor, with a 50 ma recycle mode, was used. For trace analysis, the same columns and suppressors were used, but a preconcentration column was installed beforehand. A Dionex TAC-2 preconcentration column was used for trace anion analysis and a Dionex TCC-2 was installed for cation trace analysis. To avoid airborne contamination, samples were kept in PE bottles, inside a chamber under bar N 2 pressure. Samples of 7 ml were then forced, under pressure, to the preconcentration column at a rate of 1 ml/ min. PE bottles were rinsed three times before sample collection, and one liter of water was flushed through the system before collecting samples. Standard solutions for trace analysis were prepared by mixing and diluting standard solutions (from SpexCertiprep Inc, c.a. 1,000 ppm) for each ion. Water from the Milli-Q Element system was used for dilution. Results Pretreatment purification unit: to ppb levels As mentioned above, the initial treatment step of tap water using activated carbon beads and an antiscaling agent does not reduce the total amount of ions present in tap water, but efficiently protects the water purification chain from chlorine and scaling. Reverse osmosis was the first process employed to eliminate ions (Figure 1). Concentrations of the major ions present in tap water (Li +, Na +, K +, Mg 2+, Ca 2+, F -, Cl -, NO 3-, and SO 4 ) and the respective concentrations in the RO permeate are displayed in Table 1. Also reported is the percentage rejection of each ion, i.e. the percentage of each ion eliminated by RO. Clearly, an RO system is very efficient in removing a very large proportion of ions (87% to 99%, depending on the ion) present in tap water. The selective permeability is a function of the charge, the size, and the solvation of a particular ion. Thus, divalent ions were removed more efficiently (> 99% for Mg 2+, Ca 2+, and SO 4 ) than monovalent ions (87% to 93%). In parallel with ion concentrations, the conductivity decreased from 569 μs/cm in tap water to 12.3 μs/cm in RO permeate, corresponding to an overall rejection of 97.8%. Because ion rejection is a percentage of the incoming total ion concentration, it is clear that the ion concentration in the RO permeate is feedwater related. In a second experiment, monitoring of ion concentrations over a 20 min period (ca. 3.3 L) showed that the system reached the final concentration within five minutes, or after 0.8 L (Figures 3A and 3B). More importantly, this ion concentration remained stable for all ions studied, as long as the feedwater quality remained fairly stable. As could be anticipated, the conductivity readings paralleled these results and reached a stable value within five minutes. It can, be concluded, therefore from these results that reverse osmosis is a very efficient means of consistently removing all types of ions. Such a system is particularly suitable as the initial purification treatment in a complete water purification chain. Ions Tap water a RO permeate a % Reject Li Na K Mg Ca F Cl No SO Table 1 : Effect of reverse osmosis on ion concentration. a Ion concentrations in ppb (µg/l) Figure 3 : Reverse osmosis performance for cation (Figure 3A) and anion (Figure 3B) rejection. 3

4 The second ion removal step occurred via electrodeionization (EDI). 14,18 Results displayed in Table 2 show a clear decrease in all ion concentrations from the RO permeate water to the EDI product water. Ions are concentrated and then rejected in the concentrating channel. Concentrations obtained in the product water were below one ppb (1 μg/l) for every ion studied, and the resulting conductivity was below 1 μs/cm. In contrast to RO purification, removal of ions by EDI is much less dependent on the nature of the ions. Due to the electrical current applied in the module, migration of a particular ion is not dependent on its size. The electric current also reduces ion solvation, so that both monovalent and divalent ions respond well to the current. The ion content of the product water is therefore not directly related to the ion content of the feedwater. Typical chromatograms (Figures 4A and 4B) show the decrease in ion concentrations from tap water to RO water to Elix water. Under the experimental conditions used here, no more ions were detected at the ppb level. Estimates of the remaining ionic contamination was obtained in a separate study using ICP-MS analysis. 18 Ions RO permeate a EDI concentrate a EDI product water Li < 1 Na < 1 K < 1 Mg < 1 Ca < 1 F < 1 Cl < 1 No < 1 So < 1 Table 2 : Effect of electrodeionization on ion concentration. a Ion concentrations in ppb (μg/l) Figure 4 : Typical ion chromatograms showing the concentration of cations (Figure 4A) and anions (Figure 4B) in tap water, RO permeate, and EDI dilute. Polishing system: to ppt levels Whilst an EDI system fed by RO permeate water delivers water with ionic concentrations at the ppb level, a higher water purity is needed for ion chromatography. To reach sub-ppb purification levels, a specific water polishing system is necessary. A Milli-Q Element system, employing a range of purification technologies, 18 specifically designed for ultratrace analysis, was set up for this experiment. The IC system was fitted with an ion-exchange preconcentration column, as described above. Elix (RO-EDI)-fed reservoir water (Figure 1) and Milli-Q water were compared to standard solutions. Solutions of cations, Li +, Na +, K +, Mg 2+ and Ca 2+, and anions, F -, Cl -, Br -, NO, NO 3-, SO 4 and PO 4, were prepared, with concentrations of 1,000 ppt for each ion. Results are reported in Figure 5 (cations) and Figure 6 (anions). For comparison purposes, chromatograms were superimposed, and tracings were arbitrarily set at 0, 0.1, 0.2, 0.3 and 0.4 μs/cm. Comparison of the IC traces indicated a reduction of ion concentrations from Elix water to Milli-Q water. Integrated peak values are reported in Table 3. Anion concentrations in Milli-Q water were all below detection limits (< 1 ppt). Experiments are currently in progress to improve detection limits for early eluting ions, using KOH as an eluent, and then an eluent generator to suppress the water deep. Cation concentrations were also very low, with the exception of Ca 2+ and NH 4+. Whilst the presence of NH 4+, most likely arose from airborne NH 3 contamination, the Ca 2+ contamination probably originated from the tubing and the PE bottle used to collect the samples. Based on these results, Milli-Q water, used in standard IC conditions, is suitable for dilution and preparation of blanks. Specific ion concentrations can be improved by using specifically designed purification resins, by carrying out particular rinses of the tubing (such as HNO 3 for cation analysis), or by special chromatographic configurations and columns. 4

5 Ions Reservoir water Milli-Q water Li + < 1 < 1 Na K + 50 < 2 Mg Ca Figure 5 : Ion chromatography of Elix (RO-EDI)-fed reservoir water, Milli-Q water and a standard (1,000 ppt) cation solution. F - < 1 < 1 Cl < 1 NO < 1 SO < 1 Table 3: Concentrations (ppt) of ions in Elix -fed reservoir water and Milli-Q Element water Figure 6 : Ion chromatography of Elix (RO-EDI)-fed reservoir water, Milli-Q water and a standard (1,000 ppt) anion solution. Conclusion Water delivered by an ultrapure water system, such as the Milli-Q, is suitable for IC analysis. However, it is important to use good quality feedwater for the polishing system, by employing a pretreatment system in the water purification chain, such as an Elix system, to ensure optimum efficiency of the polishing packs. As sampling techniques and instrumentation constantly improve, detection limits are being pushed ever lower. High quality ultrapure water, delivered by a Milli-Q system, is necessary to meet all the requirements of ultratrace analysis, as required for critical cleaning and the preparation of blanks and standard solutions. 5

6 References 1 Kocherlakota, N.; Lamothe, D.; Obenauf, R.; Noonan, J. Standard requirements for ppt analysis of trace metals Amer. Environ. Lab., (11), 26-27, (1995). 2 Analysis of major, minor and trace elements in NIST biological reference materials Optima 3000, ICP Application Study N 62, (1992). 3 Lanza, F.; Trincherini, P.R. The determination of cadmium, lead and vanadium by high resolution ICP-MS in Antarctic snow samples Annali di Chim. 90, 61-71, (1999). 4 Kojima, I.; Katsuzaki, M. Selective extraction and one-drop flame atomic absorption spectrometric determination of silver in biological standard reference materials Anal. Sciences, 13, , (1997). 5 Liu, C.-Y.; Lee, N.-M.; Wang, T.-H. Chelation ion chromatography as a technique for trace elemental analysis in complex matrix samples Analytica Chimica Acta, 337, (2), , (1997). 6 López-Ruiz, B. Advances in the determination of inorganic anions by ion chromatography J. Chromatogr. A, 881, , (2000). 7 Nesterenko, P.N. Simultaneous separation and detection of anions and cations in ion chromatography Trends in Anal. Chem., 20, , (2001). 8 Curran, M.A.J.; Palmer, A.S. Suppressed ion chromatography methods for the routine determination of ultra-low level anions and cations in ice cores J. Chromatogr. A, 919, , (2001). 9 Vanatta, L.E. Application of ion chromatography in the semiconductor industry Trends in Anal. Chem., 20, , (2001). 10 Rahmalan, A.; Abdullah, M.Z.; Sanagi, M.M.; Rashid, M. Determination of heavy metals in air particulate matter by ion chromatography J. Chromatogr. A, 739, , (1996). 11 Woods, C.; Rowland, A.P. Applications of anion chromatography in terrestrial environmental research J. Chromatogr. A, 789, , (1997). 12 Gros, N.; Gorenc, B. Performance of ion chromatography in the determination of anions and cations in various natural waters with elevated mineralization J. Chromatogr. A, 770, , (1997). 13 Singh, R.P.; Abbas, N.M.; Smesko, S.A. Suppressed ion chromatographic analysis of anions in environmental waters containing high salt concentrations J. Chromatogr. A, 733, 73-91, (1996). 14 Darbouret, D.; Stewart, B.M. Advancement in the production of ultrapure water for ICP-MS metals analysis Amer. Laboratory News, 30, (9), (1998). 15 Cheng, P.T.; Pritzker, K.P. Pyrophosphate, phosphate ion interaction: effects on calcium pyrophosphate and calcium hydroxyapatite crystal formation in aqueous solutions J Rheumatol., (10), , (1983). 16 Darbouret, D.; Kano, I.; Youf, E.; Stewart, B. Optimizing storage of purified water for laboratory applications R&D Notebook, EMD Millipore, RD001, (1999). 17 Darbouret, D.; Kano, I. Ultrapure water for elemental analysis down to ppt levels R&D Notebook, EMD Millipore, RD002, (1999). 18 Darbouret, D.; Kano, I. Ultrapure water for boron and silica sensitive laboratory applications R&D Notebook, EMD Millipore, RD003, (1999). 19 U. S. Environmental Protection Agency. Sample preservation in Methods for Chemical Analysis of Water and Wastes, EPA-600/ U.S.E.P.A., Cincinnati, Ohio, USA, (1983). 6 Lit. No. RD007ENUS. 01/12. Copyright 2012, EMD Millipore Corporation, Billerica, MA, U.S.A. EMD Millipore, Milli-Q, Progard and Elix are registered trademarks of Merck KGaA, Darmstadt, Germany. The M mark and EMD Millipore are trademarks of Merck KGaA. Dionex is a registered trademark of Dionex Corporation. All rights reserved.

Novel Electrodeionization Devices: Applications in Inorganic Analysis

Novel Electrodeionization Devices: Applications in Inorganic Analysis Novel Electrodeionization Devices: Applications in Inorganic Analysis John M. Riviello and Archava Siriraks ABSTRACT Electrodeionization (EDI) is a well established technique for the production of ultrapure

More information

Ultrapure Water for Boron and Silica-Sensitive Laboratory Applications

Ultrapure Water for Boron and Silica-Sensitive Laboratory Applications Ultrapure Water for Boron and Silica-Sensitive Laboratory Applications The R&D Notebook 3 A publication of the Lab Water Division of Merck Millipore Ichiro Kano and Daniel Darbouret Research & Development,

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

columns IonPac AS17-C Anion Exchange Column

columns IonPac AS17-C Anion Exchange Column columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column

More information

07/03/18. Laboratory Water and Water Purification Systems. Water in the Laboratory. Sources of Water and Water Contamination. Contaminants in Water

07/03/18. Laboratory Water and Water Purification Systems. Water in the Laboratory. Sources of Water and Water Contamination. Contaminants in Water Laboratory Water and Water Purification Systems KMU 392 Chemical Process Industries March 2018 Water in the Laboratory Water is the most commonly used solvent in laboratories and constitutes often more

More information

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample... I J P F A International Science Press Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample from Different Area of North Gujarat Region by Ion Chromatography Bharat

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography APPLICATION NOTE 113 Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography Author Edward Kaiser Thermo Fisher Scientific, Sunnyvale, CA Keywords Trace ionic

More information

The Determination of Trace Anions in Concentrated Phosphoric Acid

The Determination of Trace Anions in Concentrated Phosphoric Acid Technical Note The Determination of Trace Anions in Concentrated Phosphoric Acid INTRODUCTION The determination of trace anions in phosphoric acid is hampered by a large excess of phosphate ion. Chloride

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

Expanded capabilities in ammonium and amine detection

Expanded capabilities in ammonium and amine detection PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium

More information

anion-exchange column

anion-exchange column columnsionpac ASA Anion-Exchange Column The IonPac ASA is a carbonateselective anion-exchange column designed for the fast separation of inorganic anions, including fluoride, chlorite, bromate, chloride,

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath APPLICATIN NTE 9 Determination of an anionic fluorochemical surfactant in a semiconductor etch bath Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography,

More information

columns IonPac AS18 Anion-Exchange Columns

columns IonPac AS18 Anion-Exchange Columns columns IonPac AS Anion-Exchange Columns The IonPac AS is a hydroxideselective anion-exchange column designed for the determination of inorganic anions and low-molecularweight organic acids including fluoride,

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,

More information

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Harpreet Dhillon and John Statler Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 11 Introduction

More information

The determination of trace anions in concentrated phosphoric acid

The determination of trace anions in concentrated phosphoric acid TECHNICAL NOTE The determination of trace anions in concentrated phosphoric acid Authors Edward Kaiser and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Ion chromatography, IC, IonPac

More information

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,

More information

Determination of Trace Anions in Organic Solvents

Determination of Trace Anions in Organic Solvents Application Note 8 Determination of Trace Anions in Organic Solvents INTRODUCTION In the manufacture of semiconductor materials, a great deal of attention is focused on minimizing sources of contamination.

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

Air Monitoring. Semi-continuous determination of ambient air quality

Air Monitoring. Semi-continuous determination of ambient air quality Air Monitoring Semi-continuous determination of ambient air quality The Particle Into Liquid Sampler a simple solution for the determination of ions in aerosol particles 02 Combustion of fossil fuels for

More information

columns IonPac SCS 1 Silica Cation Separator

columns IonPac SCS 1 Silica Cation Separator columns IonPac SCS Silica Cation Separator The SCS Silica Cation Separator is designed for use with nonsuppressed conductivity detection, or single-column ion chromatography (SCIC). This column is particularly

More information

arium Water Purification Terms

arium Water Purification Terms arium Water Purification Terms A.C.S. American Chemical Society Absorption The surface retention of a substance (absorbent, e.g., CO 2 (carbon dioxide) by a material, such as water (absorbing medium).

More information

Ion Chromatography (IC)

Ion Chromatography (IC) Ion Chromatography (IC) Purpose: This module provides an introduction to Ion Chromatography (IC). In this module the basic theory and applications of IC will be presented at a level that assumes a basic

More information

Overcoming Challenging Matrices in Ion Chromatography

Overcoming Challenging Matrices in Ion Chromatography Overcoming Challenging Matrices in Ion Chromatography Presented by: Kirk Chassaniol Gulf Coast Conference 2014 October 14, 2014 Thermo Fisher Scientific NA IC Tech Support 1 The world leader in serving

More information

IonPac Trace Cation Concentrator

IonPac Trace Cation Concentrator for the IonPac Trace Cation Concentrator (TCC-LP1, TCC-ULP1 and TCC-XLP1) Page 1 of 17 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR (TCC) TCC-LP1 Column Low Pressure, 4 x 35 mm (P/N 046027) TCC -ULP1 Column

More information

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Determination of Trace Anions in Concentrated Hydrofluoric Acid Technical Note Determination of Trace Anions in Concentrated Hydrofluoric Acid INTRODUCTION There is a need for a reliable method to determine trace anions in hydrofluoric acid. The large excess of fluoride

More information

Research & Development, Laboratory Water Division, Millipore S.A., Saint-Quentin-en- Yvelines, France

Research & Development, Laboratory Water Division, Millipore S.A., Saint-Quentin-en- Yvelines, France Ultrapure water blank for boron trace analysis Daniel Darbouret* and Ichiro Kano Research & Development, Laboratory Water Division, Millipore S.A., Saint-Quentin-en- Yvelines, France Interlaboratory Note

More information

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm) columns and Trap Columns IonPac AG, AS, mm Eluent Source: EG Temperature: C IonPac AC ( mm) Flow Rate:. m/min Eluent: Potassium hydroxide, Inj. Volume: m, preconcentrated mm from to min, mm mm from min

More information

Determination of Trace Anions in Concentrated Glycolic Acid

Determination of Trace Anions in Concentrated Glycolic Acid Technical Note Determination of Trace Anions in Concentrated Glycolic Acid INTRODUCTION There is a need for a reliable method to determine trace chloride and sulfate in glycolic acid. The presence of these

More information

Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments

Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments Environ. Chem. 2014, 11, 624-631 CSIRO 2014 Supplementary material Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments Kerstin Hockmann, A,C Susan

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation EPA Method 549.2 Revision 1.0 Determination of Diquat and Paraquat in Drinking Water by Liquid-Solid Extraction and High Performance Liquid Chromatography with Ultraviolet Detection* UCT Products: ENVIRO-CLEAN

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

Determination of trace anions in concentrated glycolic acid

Determination of trace anions in concentrated glycolic acid TECHNICAL NOTE Determination of trace anions in concentrated glycolic acid Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography, IC, Dionex ICS- +,

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la

More information

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY As originally published in the SMTA Proceedings. DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY Christopher J. Pontius,

More information

Ion chromatography. Ion chromatography

Ion chromatography. Ion chromatography Ion chromatography Ion chromatography - Contents Professional IC Systems 850 Professional IC Professional IC 850 introduction IC. 11 Professional IC ProfIC Systems How to choose the right ProfIC system

More information

High-Performance Chromatography Superior isocratic separation of

High-Performance Chromatography Superior isocratic separation of columns I o n P a c C S A C a t i o n - E x c h a n g e C o l u m n The IonPac CSA cation-exchange column provides fast analysis of mono- and divalent cations, ammonium, and aliphatic amines simultaneously

More information

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep pplication Note 70 Determination of Silicate in High Purity Water Using Ion Chromatography and utoprep Introduction The water used in the manufacture of semiconductors and other modern electronic components

More information

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 This method provides procedures for

More information

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Suppression in anion chromatography. More sensitive analysis of anions and organic acids Suppression in anion chromatography More sensitive analysis of anions and organic acids What is suppression? 02 Suppression plays a key role in the analysis of anions and organic acids using ion-exchange

More information

Determination of Adsorbable Organic Halogen in Wastewater

Determination of Adsorbable Organic Halogen in Wastewater Determination of Adsorbable Organic Halogen in Wastewater Jingli Hu, Jeff Rohrer, and Kirk Chassaniol Thermo Fisher Scientific, Sunnyvale, CA The world leader in serving science Outline AOX (Adsorbable

More information

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of silicate in high-purity water using ion chromatography and online sample preparation PPLICTION NOTE 70 Determination of silicate in high-purity water using ion chromatography and online sample preparation uthors Weerapong Worawirunwong and Jeffrey Rohrer Thermo Fisher Scientific, Bangkok,

More information

Determination of trace anions in high-nitrate matrices by ion chromatography

Determination of trace anions in high-nitrate matrices by ion chromatography APPLICATION NOTE 7 Determination of trace anions in high-nitrate matrices by ion chromatography Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Contamination,

More information

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection columns IonPac AS Anion-Exchange Column The IonPac AS is a highcapacity, hydroxide-selective anion exchange column designed for separation of haloacetic acids (HAAs) and bromate in drinking water prior

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

columns IonPac AS22 Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column columns IonPac AS Anion-Exchange Column The IonPac AS is a carbonate based anion-exchange column designed for the determination of inorganic anions and low-molecular-weight organic acids including fluoride,

More information

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7 Troy, NY 12180 Date 01/25/18 Page 1 of 7 USGS Water Science Center Laboratory, Troy, NY Ion Chromatography Standard Operating Procedure 1. Scope and Application 1.1 Analytes Chloride, Nitrate, and Sulfate

More information

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0 Metrosep A Supp 10-100 /4.0 Branch Environment, Food, Beverages Keywords IC; 850; 858; Metrosep A Supp 10-100/4.0; Cyanide; Sulfide; 2.850.9110; DC Mode Summary The determination of sulfide and cyanide

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Automated Determination of PPQ Levels of Thorium in High Purity Copper PPQ Levels of Th in High Purity Cu TRUFAST ICPMS Automated Determination of PPQ Levels of Thorium in High Purity Copper Using the ESI TRUFAST System and ICPMS Detection By Nathan Saetveit, PhD and Dan

More information

Feed. Figure 1. The above image depicts the construction of a typical spiral wound element.

Feed. Figure 1. The above image depicts the construction of a typical spiral wound element. In a reverse osmosis (RO) process, pressure is applied to the saline side of a semi-permeable membrane to produce low salinity water. Upon application of the feed pressure, water molecules pass through

More information

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000 Evaluation of Various Anion-Exchange Chemistries for the Analysis of Haloacetic Acids in Drinking Water Using -D Matrix Elimination Ion Chromatography and Suppressed Conductivity Detection White Paper

More information

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,

More information

IonPac SCS 1 IonPac SCG 1

IonPac SCS 1 IonPac SCG 1 IonPac SCS 1 IonPac SCG 1 Document No. 031948-05 IonPac SCG1 and SCS 1 Page 1 of 33 PRODUCT MANUAL for the IONPAC SCG 1 Guard Column 2 x 50 mm, P/N 061522 4 x 50 mm, P/N 061523 IONPAC SCS 1 Analytical

More information

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Deionization. resistivities can not be achieved MIXED BED REACTION

Deionization. resistivities can not be achieved MIXED BED REACTION Deionization W H A T I S D E I O N I Z A T I O N? WHAT ARE IONS? Ions are divided into twogroups; cations and anions. Cations have a positive charge and include sodium (Na + ), calcium (Ca ++ ), and magnesium

More information

Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts

Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts Technical Note 26 Determination of in Water, Waste Water, and Solid Waste Extracts INTRODUCTION Chromium, while not unique in its properties, is commonly used in various industries because of the characteristics

More information

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for

More information

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column Application Note 4 Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS6 Column INTRODUCTION The common alkali and alkaline earth cations are

More information

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures. User Manual IonPac NS2 Columns 065501 Revision 02 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for IonPac NS2 Columns IonPac NS2, 5µm, 4x250mm Analytical Column

More information

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,

More information

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device Determination of Bromate in Bottled Mineral Water Using the CRD 00 Carbonate Removal Device Thunyarat Phesatcha, Weerapong Worawirunwong, and Jeff Rohrer Thermo Fisher Scientific, Bangkok, Thailand; Thermo

More information

INSTALLATION INSTRUCTIONS and TROUBLESHOOTING GUIDE. for the OnGuard- CARTRIDGE

INSTALLATION INSTRUCTIONS and TROUBLESHOOTING GUIDE. for the OnGuard- CARTRIDGE INSTALLATION INSTRUCTIONS and TROUBLESHOOTING GUIDE for the OnGuard- CARTRIDGE (OnGuard-RP, P/N 039595) (OnGuard-H, P/N 039596) (OnGuard-P, P/N 039597) (OnGuard-Ag, P/N 039637) (OnGuard-A, P/N 042102)

More information

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin Ion Exchange Chromatography Chromatographic Process BA Mobile phase Stationary Phase A Shula Levin Bioforum B Distribution: K = C s/c m B shulal@zahav.net.il http://shulalc.co.il/ A Elution through the

More information

Tap Water Analysis Lockwood 1 TAP WATER ANALYSIS BY ATOMIC ABSORPTION SPECTROSCOPY AND ION CHROMATOGRAPHY. Susan Lockwood

Tap Water Analysis Lockwood 1 TAP WATER ANALYSIS BY ATOMIC ABSORPTION SPECTROSCOPY AND ION CHROMATOGRAPHY. Susan Lockwood Tap Water Analysis Lockwood 1 TAP WATER ANALYSIS BY ATOMIC ABSORPTION SPECTROSCOPY AND ION CHROMATOGRAPHY Susan Lockwood Inorganic Quantitative Analysis - Section 6 California State University, Sacramento

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography STANDARDBASE techniques: High Performance Liquid Chromatography Drenthe College, The Netherlands 1. Introduction HPLC. High Performance Liquid Chromatography High Performance Liquid Chromatography (HPLC)

More information

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. 1 Scope This document describes a chromatographic method which allows the identification as well as the determination of

More information

Department of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States

Department of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States Supporting Information for Capture and Reductive Transformation of Halogenated Pesticides by an Activated Carbon-Based Electrolysis System for Treatment of Runoff Yuanqing Li 1 and William A. Mitch 1,

More information

Sample Preparation for Ion Chromatography

Sample Preparation for Ion Chromatography Sample Preparation for Ion Chromatography Comprehensive Solutions to Sample Preparation Challenges Now sold under the Thermo Scientific brand Choose from a Broad Range of Solutions No matter what your

More information

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples A P P L I C A T I O N N ot e Atomic Absorption Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples Introduction The Flow Injection Mercury System (FIMS) is a dedicated system

More information

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Application Note Forensic Toxicology Authors Bellah. Pule,

More information

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS4A Anion-Exchange Column Product Specifications Thermo Scientific Dionex IonPac AS4A anion-exchange column is designed for the separation of haloacetic

More information

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS column is a high-capacity, hydroxide-selective, anion-exchange column

More information

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA Determination of Sub-μg/L Bromate in Municipal and Natural Mineral Waters Using Preconcentration with Two-Dimensional Ion Chromatography and Suppressed Conductivity Detection Brian DeBorba and Jeff Rohrer

More information

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Kirk Chassaniol Manager of IC Technical Sales Support Thermo Fisher Scientific 1 The world leader in serving

More information

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan

More information

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Inorganic Anions in Wastewater by Ion Chromatography Application Note 5 Determination of Inorganic Anions in Wastewater INTRODUCTION The determination of common inorganic anions in environmental waters, such as wastewater and drinking, ground, and surface

More information

Determination of Iron by Dispersive Liquid-Liquid Microextraction Procedure in Environmental Samples

Determination of Iron by Dispersive Liquid-Liquid Microextraction Procedure in Environmental Samples American Journal of Chemistry 2012, 2(1): 28-32 DOI: 10.5923/j.chemistry.20120201.07 Determination of Iron by Dispersive Liquid-Liquid Microextraction Procedure in Environmental Samples F. Sánchez Rojas

More information

Removal of suspended and dissolved organic solids

Removal of suspended and dissolved organic solids Removal of suspended and dissolved organic solids Types of dissolved solids The dissolved solids are of both organic and inorganic types. A number of methods have been investigated for the removal of inorganic

More information

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography APPLIATION NOTE 78 Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography Authors Soon Fatt Lee, Aaron Rose, Terri hristison, and Jeff Rohrer Thermo Fisher

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Thermo Scientific. Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS9-µm Anion-Exchange Column Product Specifications The Themo Scientific Dionex IonPac AS9-µm high-capacity, hydroxide-selective, anion-exchange column is

More information

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Inorganic Anions in Drinking Water by Ion Chromatography Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography Author : Li Guoping Yu Rong Shui Rong Institute for the Control of Agrochemicals, Ministry of Agriculture, P.

More information

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM) Analytical Procedure RADIUM IN WATER (CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This is a method for separation and measurement of radium-226 and radium-228 in water. This method

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14911 First edition 1998-10-01 Water quality Determination of dissolved Li +, Na +, NH 4 +, K +, Mn 2+, Ca 2+, Mg 2+, Sr 2+ and Ba 2+ using ion chromatography Method for water

More information