Determination of Trace Anions in Concentrated Hydrofluoric Acid

Size: px
Start display at page:

Download "Determination of Trace Anions in Concentrated Hydrofluoric Acid"

Transcription

1 Technical Note Determination of Trace Anions in Concentrated Hydrofluoric Acid INTRODUCTION There is a need for a reliable method to determine trace anions in hydrofluoric acid. The large excess of fluoride ion hampers this analysis. To determine chloride at µg/l (ppb) in.% (v/v) hydrofluoric acid represents a concentration ratio of : (chloride to fluoride). Diluting the concentrated sample overcomes the problem of the large concentration of the interfering matrix ion but results in a lack of sensitivity for the contaminant ions of interest. An improved method for determining trace anions in concentrated hydrofluoric acid has been developed to overcome this problem., Trace inorganic anions of strong acids are separated from the high concentration of fluoride by an ion exclusion separation prior to an ion chromatographic separation. This Technical Note describes the theory, set up, and analytical procedure for the determination of chloride, nitrate, sulfate, and phosphate at sub mg/l (ppm) levels in.% (v/v) hydrofluoric acid. SUMMARY OF THE METHOD An IonPac ICE-AS ion exclusion column is used on-line for sample pretreatment to separate the analyte ions from an excess of fluoride matrix ions. A selected fraction from the ion exclusion separation is cut and sent to a -mm IonPac AS9-HC, anion exchange concentrator column. The concentrated ions are then eluted onto a -mm IonPac AS9-HC column where the anions of interest are separated and detected by suppressed conductivity. EQUIPMENT DX- Ion Chromatography System consisting of: GP Gradient pump, microbore configuration CD Conductivity detector with a temperature controlled conductivity cell (DS) LC Enclosure, () Rheodyne valves, PEEK, rear loading Dionex RP- single piston pump Pressurizable Reservoir Chamber (P/N 7) AC Power Control Accessory CAM (Controlled Air Module) Low Pressure -Way, - fittings (P/N ) () Plastic bottle assemblies, L, [() for external water, () for rinse solution] (P/N 9) () O-ring, Teflon encapsulated (for rinse solution bottle) (P/N ) () O-rings Teflon encapsulated for reservoir chamber (P/N 7) () Air pressure gauge to psi (7 kpa) (for external water) cm ( in.) of green PEEK tubing, diameter of.7- mm (. in.) to connect columns and make a 7-µL sample loop Teflon bottles as sample containers (VWR P/N 7- or Nalge P/N - for ml narrow mouth bottles) Disposable chemical resistant gloves for hydrofluoric acid (Lab Safety P/N 8A-79) PeakNet Chromatography Workstation Technical Note

2 COLUMNS IonPac AG9-HC guard column, mm (P/N 8) IonPac AS9-HC analytical column, mm (P/N ) IonPac AG9-HC concentrator column, mm (P/N 79) IonPac AG as trap column, mm (P/N 9) IonPac ICE-AS pretreatment column (P/N ) Anion Self Regenerating Suppressor (ASRS -ULTRA), mm (P/N 97) REAGENTS AND STANDARDS Deionized water (DI H O) Type I reagent grade, 7.8 MΩ-cm resistance or better. Sodium hydroxide % (w/w) aqueous solution (Fisher Scientific). M Carbonate Anion Eluent Concentrate (Dionex P/N 7) Chloride standard mg/l, ml (Dionex P/N 79) Sulfate standard mg/l, ml (Dionex P/N 7) Nitrate standard mg/l, ml (Ultra Scientific, VWR P/N ULICC-) Phosphate standard mg/l, ml (Ultra Scientific, VWR P/N ULICC-) CONDITIONS: Ion Exclusion : Trap column: Eluent: Flow rate: IonPac ICE-AS IonPac AG, mm Deionized water. ml/min Ion Chromatography Analytical column: IonPac AS9-HC, mm Guard column: IonPac AG9-HC, mm Concentrator column: IonPac AG9-HC, mm Eluent: 8 mm Sodium carbonate. mm Sodium hydroxide Flow rate:. ml/min volume: 7 µl Detection: Suppressed conductivity, ASRS, AutoSuppression external water mode Suppresor Current setting: Expected system Backpressure: ma Expected Background Conductivity:.8 MPa ( psi) (with concentrator column in line) PREPARATION OF SOLUTIONS AND REAGENTS Eluent Solutions 8. mm Sodium Carbonate /. mm Sodium Hydroxide Add.8 g of. M sodium carbonate and. g of % sodium hydroxide directly to 9 g degassed, deionized water (having a specific resistance of 7.8 MΩ-cm or greater) in an eluent bottle. Dilute to g. Transfer this solution to an eluent container and vacuum degas for minutes. mm Sodium Hydroxide (AG trap column regeneration) preparation: Dilute. g of % (w/w) sodium hydroxide with degassed, deionized water (having a specific resistance of 7.8 MΩ-cm or greater) to a final weight of g in an eluent bottle. Avoid the introduction of carbon dioxide from the air into the aliquot of % (w/w) sodium hydroxide and the deionized water being used to make the eluent. Preparation Pour ml of deionized water into a clean Teflon sample container. While in a fume hood, very slowly and carefully add ml (.8 g) of 9% concentrated hydrofluoric acid to the deionized water. Gently mix the acid with the water. Always add acid to water not water to acid. Allow the solution to cool to room temperature. Caution: Use extreme care when handling hydrofluoric acid. Avoid all contact with exposed skin. The use of eye protection, proper gloves, and a fume hood is an absolute necessity. Consult the Material Safety Data Sheet (MSDS) for more specific details about protective equipment reactivity and health effects. STANDARD SOLUTIONS Stock standard solution ( mg/l) Use Dionex or commercially available mg/l ion standard solution. Determination of Trace Anions in Concentrated Hydrofluoric Acid

3 Working standard solution ( µg/l) To prepare a mixed working standard solution, combine. ml of each anion stock solution with deionized water and dilute to a final volume of ml. CALIBRATION Prepare at least three different calibration standards at different concentrations of the working standard. Select a range similar to the expected concentrations in the samples. The method of standard additions (adding one or more increments of a standard solution to sample aliquots of the same size) should be used to minimize the effect of the concentrated acid matrix on the measured conductivity of the analytes of interest. Calibration standards are added to the deionized water used to dilute the concentrated hydrofluoric acid from 9% (w/w) to.% (v/v). The following formula can be used to calculate concentrations for dilutions: (Conc. of stock solution, mg/l) (Vol. of stock solution, ml) = (Conc. of dilute standard, mg/l) (Vol. of dilute standard, ml) The procedure for preparing standards in ml of.% (v/v) HF is as follows:. Prepare aqueous standards that are twice as concentrated (X) as desired by diluting µg/l anion stock standard solutions using the volumes listed in Table.. Carefully add. ml 9% (w/w) HF to. ml with each of the aqueous standards. This will generate.% (v/v) HF standards containing,, and µg/l anions. Table Method of Standard Additions for a Final Volume of.-ml Aliquots of.% (v/v) HF IONPAC AG TRAP COLUMN REGENERATION The AG trap column must first be regenerated. Monitoring the blank will indicate when regeneration is necessary. Typically, monthly regeneration is necessary, but it will depend upon the quality of the deionized water and usage rate of the instrument. creased contamination in the water blank indicates that the IonPac AG needs to be regenerated. The procedure is as follows:. Pump mm sodium hydroxide through the AG at. ml/min for minutes.. Follow with a rinse of deionized water at the same flow rate for minutes. DISCUSSION OF THE METHOD This method addresses the challenge of determining trace concentrations of contaminant ions such as chloride, nitrate, sulfate, and phosphate in a matrix composed of a high concentration of fluoride ion. This is accomplished in two steps: an ion exclusion (ICE) preseparation followed by injecting a portion of the ICE separation to an ion chromatographic (IC) separation. The ion exclusion mechanism separates ionized species from nonionized or weakly ionized species. This occurs because of a negatively charged hydration shell on the stationary phase surface called the Donnan membrane. Figure illustrates the application of this ICE mechanism to the separation of mg/l of chloride from mg/l : IonPac ICE-AS Eluent: Deionized water Flow Rate:. ml/min Volume: 7 µl Detection: Conductivity Peaks:. Chloride mg/l. Fluoride Concentration of X anion stock (µg/l) Volume of µg/l anion stock (ml) Aqueous Volume (ml) To To Concentrator To 9 Figure. Ion exclusion separation of mg/l chloride from mg/l fluoride. Technical Note

4 fluoride using an ICE-AS ion exclusion column. This chromatogram is a measurement only of the unsuppressed conductivity response for the ICE separation. The strong acid ions, such as chloride and sulfate, are excluded and elute as a small peak at approximately nine minutes. The weakly ionized fluoride matrix ions are retained and elute as a large peak. A series of schematics (Figures ) illustrates the operation of the chromatographic hardware. The concentrated hydrofluoric acid sample is loaded via a pressurized reservoir into the 7-µL sample loop (Figure ). Helium or nitrogen at. kpa ( psi) is used to push sample from the sample container into the sample loop at a flow rate of ~ ml/min (.8 g/min). This technique ensures that a representative sample of the concentrated hydrofluoric acid sample is loaded into the sample loop. It is important to pass at least four loop volumes through the sample loop to ensure reproducible sampling. The concentrated hydrofluoric acid in the sample loop is then delivered with the high purity water carrier stream to the IonPac ICE-AS. An AG is placed after the RP- pump to act as an anion trap column for the deionized water. Any contaminants present in this water will impact the quality of the blank. The first portion of the ICE separation from. to 7. min is sent to waste, (Figure ). Then, the concentrator column is placed in line with the ICE column and the portion from 7. to. min is captured on the concentrator column (Figure ). After. minutes the -mm AG-9HC concentrator column is placed in-line with the -mm AS9-HC analytical column set and the concentrated ions are separated (Figure ). This time window should ensure the preconcentration of all the chloride, nitrate, sulfate, and phosphate with a minimal amount of fluoride. The IC separation uses an AS9-HC with an isocratic eluent of 8 mm sodium carbonate. mm sodium hydroxide. The high capacity of the AS9-HC column allows injection of these concentrated hydrofluoric acid samples without overloading. Figure shows a separation of the common anions in deionized water with the -mm IonPac AS9-HC column. These chromatographic conditions enable the separation of chloride from fluoride as well as carbonate. A -mm microbore analytical column set was chosen because it has a four fold increase in mass sensitivity over the standard bore column. This facilitates faster loop loading because smaller sample volumes are required. The microbore format also offers low eluent consumption as well as less waste generation. An IonPac AG9-HC ion exhange column was used as the concentrator column in the x mm format instead of x mm because the -mm column had four times more capacity than the -mm column and lower back pressure at the microbore flow rate. No significant degradation in separation efficiency was observed when coupling a -mm concentrator column with a -mm analytical column set. Pump DI H O AG Trap Eluent Loading Helium Out Helium Loop Reservoir ICE-AS Concentrator AG9-HC, mm jection AG9-HC AS9-HC, mm Analytical s 9 Figure. Loading the sample loop. Determination of Trace Anions in Concentrated Hydrofluoric Acid

5 Pump DI H O AG Trap Eluent Loading Helium Out Helium Loop Reservoir ICE-AS Concentrator jection AG9-HC AS9-HC, mm Analytical s AG9-HC, mm 97 Figure. First stage of ICE separation (time. 7. min). Pump DI H O AG Trap Eluent Loading Helium Out Helium Loop Reservoir ICE-AS Load Concentrator AG9-HC, mm jection AG9-HC AS9-HC, mm Analytical s 98 Figure. Concentrating the "cut" portion from the ICE separation (time 7.. min). Pump DI H O AG Trap Eluent Loading Helium Out Helium Loop Reservoir ICE-AS Concentrator jection AG9-HC AS9-HC, mm Analytical s AG9-HC, mm 99 Figure. Separating the retained ions. Technical Note

6 During the IC separation, the deionized water rinses the ICE-AS and associated tubing to ensure that there is no contamination from the previous sample. To prevent sample depletion, the PeakNet method stops the flow of helium pressure at.9 minutes to the reagent reservoir after enough sample has been loaded. SYSTEM PREPARATION AND TEST Refer to the system configuration schematics in Figures - and Table, that summarize the types and lengths of tubing required for system configuration. The chromatographic hardware is divided into two parts: the ion exclusion pretreatment portion with the ICE-AS and the IC analysis portion with the AS9-HC. IC system. Prepare the ASRS by following the Quickstart instructions (Dionex Document 8-) included with the structions and Troubleshooting Guide for the ASRS.. stall the -mm IonPac AG9-HC and AS9-HC column set using the red.-mm (. in.) tubing. To minimize dead volume, use the smallest possible lengths of tubing and ensure that the ends of the tubing are smooth and level.. Construct a -µl sample loop by cutting a 9.9-cm (.9 in.) portion of the black.-mm (. in.) PEEK tubing.. stall this sample loop between ports and of the injection valve in the IC analysis system.. stall the ASRS and configure it in the external water mode as described in the ASRS manual.. Establish eluent flow through the -mm IonPac AG9-HC and AS9-HC analytical column set. The expected background conductivity is ~. (Note: for trace analysis it will take at least five hours for the system to achieve a stable background conductivity.) 7. Verify proper operation of the IC portion of the system by injecting a low level ppm standard to replicate the column test chromatogram. 8. Remove the -µl sample loop and install the -mm IonPac AG9-HC column. Make sure that the arrow indicating flow on the column is pointed from port to port and that the tubing length connecting the outlet of this column and port is as short as possible. 9. Configure the IC valve so that the -mm IonPac AG9-HC column is in-line with the -mm IonPac AG9-HC and AS9-HC analytical column set. Check for leaks. The expected system back pressure for these three columns at. ml/min is ~.8 MPa (, psi). Ion Exclusion Pretreatment System This section describes the preparation of the pretreatment portion of the system. It is important that the same type and length of tubing as described in Table be used to Table Details of Tubing Configuration for Trace Anions in Hydrofluoric Acid Connection Tubing Length Remarks Points Description (cm) ICE outlet => Port Green.7-mm (. in.) ICE inlet => Port Green.7-mm 7 (. in.) Port => Port Green.7-mm 7-µL (. in.) sample loop AG9-HC, mm => Port Red.-mm Should be as (. in.) short as possible AG9-HC, mm => Port Black.-mm (. in.) Port => Analytical Red.-mm Should be as (. in.) short as possible Pretreatment : IonPac ICE-AS ICE Eluent: Deionized water ICE Flow Rate:. ml/min Volume: 7 µl : IonPac AG9-HC, AS9-HC, -mm Concentrator: IonPac AG9-HC, mm Eluent: 8. mm Sodium carbonate. mm Sodium hydroxide IC Flow Rate:. ml/min Detection; Suppressed conductivity, ASRS-ULTRA, AutoSuppression, external water mode Peaks:. Fluoride µg/l (ppb). Chloride. Carbonate. Nitrate. Unidentified. Sulfate Phosphate 7 Figure. Trace anion standard in deionized water. Determination of Trace Anions in Concentrated Hydrofluoric Acid

7 successfully perform this analysis. Changes in tubing length will result in a different cut from the ICE-AS column being delivered to the AS9-HC concentrator column.. Cut a -cm ( in.) portion of the.7-mm (. in.) i.d. green PEEK tubing to make a 7-µL sample loop and install this loop between ports and of the sample valve.. Prepare the AG trap column according to the directions in the section entitled IonPac AG Trap Regeneration. (Caution: Before the AG is installed in the system it is important that the sodium hydroxide solution used for storage or cleaning be completely rinsed away. The ICE-AS is not compatible with sodium hydroxide eluents.). The entire pathway from the RP- pump to port of the IC valve is plumbed with the green PEEK.7-mm (. in.) tubing. stall the AG at the exit port of the RP- pump.. stall the ICE-AS column using a -cm piece of the green PEEK.7-mm (. in.) tubing between the outlet of the ICE-AS and port of the injection valve. Use a 7-cm portion of green tubing between port of the sample valve and the inlet of the ICE-AS.. Check to see that there is about. kpa ( psi) head pressure on the incoming deionized water that feeds the RP- pump.. Connect the exit port of the reagent reservoir to port of the sample valve with.7-mm (. in.) green tubing. 7. Configure a waste line from port of the sample valve with the green PEEK tubing. 8. Connect the reagent reservoir to helium pressure of about. kpa ( psi). 9. Begin with a container filled with deionized water as a sample to rinse the sample lines of any trace contamination.. Set a flow rate of. ±. ml/min for the RP- pump by adjusting the dial on the pump. This should be measured with the -mm AG9-HC concentrator column out of line. Measure the flow rate by collecting the waste coming out of port of the IC valve. It is critical for the success of this method that this flow rate be consistent.. Pump deionized water at. ml/min through the ICE-AS to waste without the -mm AG9-HC in line for h. This will remove the. mm heptafluorobutyric acid storage solution. It is typical with this method to detect trace amounts of sulfate in the deionized water blank.. The ICE-AS can be further conditioned by rinsing it with mm HF for h at. ml/min followed by a -h rinse with 7.8 MΩ-cm deionized water. This will reduce the sulfate blank to µg/l or less. Continue to monitor the blank, especially when starting up the system after it has been idle for more than two days.. Connect the wires between the GP and the AC according to the wiring diagram in Figure 7. This wiring scheme will ensure that in the event of a power outage, that the units will be in the OFF position when power is restored.. Plug the RP- pump into outlet of the AC and the Controlled Air Module (CAM) into outlet.. Connect the color-coded air lines of the CAM (see Figure 7). The CAM directs 89 kpa ( psi) air pressure to the low pressure slider valve to regulate the flow of. kpa ( psi) helium to the reagent reservoir. a. Red tubing to a source of air pressure b. Yellow tubing to the barbed fitting at the top of the slider valve (orientation groove marks the top of the valve) c. Green tubing to the barbed fitting at the bottom of the slider valve. Configure the valve ports on the slider valve as follows: a. Port - Helium Out to the reagent reservoir b. Port - Helium. kpa ( psi) c. Port - Closed d. Port - Open 7. Confirm that by actuating Relay that the RP- pump can be turned on and off. Likewise, confirm that by actuating Relay that the flow of helium to the reagent reservoir can be turned on and off. 8. To reduce the sulfate blank continue pumping deionized water through the ICE-AS overnight. Technical Note 7

8 GP CPU/RLY PCB Rly Outputs Rly TTL TTL TTL puts TTL AC TTL put Relay controls AC TTL Relay controls AC TTL RP- Pump AC Outlet CAM Compressed Air Out { RED YELLOW GREEN Slider Compressed Air Open Barbed Fitting Top of Closed To Reservoir Helium TTL TTL + + Barbed Fitting Bottom of Loading Figure 7. Wiring diagram for AC and CAM. SYSTEM OPERATION After all aspects of the instrumentation have been prepared and the system is ready for sample analysis.. Load the PeakNet method shown in Table.. Fill the 7-µL sample loop with deionized water. Use helium gas pressure to push sample using the reagent reservoir (Figure ).. Analyze a blank by loading deionized water as the sample. It may take several runs until the system has been rinsed of contamination.. After an acceptable blank has been established the system is ready for analysis.. Caution: Exercise extreme care when handling concentrated hydrofluoric acid. Consult the applicable Material Safety Data Sheet (MSDS) for specific details about protective equipment, reactivity, storage, disposal and health effects.. Concentrated hydrofluoric acid, 9.% (w/w) is diluted to.% (v/v) with the procedure described in the Preparation section. Table PeakNet Method for the Analysis of Concentrated Hydrofluoric Acid Total Time (min) ICE Time (min) IC Time (min) jection Relay %A Figure Comments it ject A Open. ject A Closed Begin loading the sample loop.9 ject A Open End loading the sample loop.. ject B Open Begin ICE separation. 7. Load B Open Send cut portion from ICE separation to Concentrator column... ject B Open Begin IC separation. Concentrator column in-line.. ject B Open End IC Separation A = ject, B = Load 8 Determination of Trace Anions in Concentrated Hydrofluoric Acid

9 7. The diluted.% (v/v) hydrofluoric acid can be loaded directly into the 7-µL loop by the reagent delivery module as discussed earlier. Ensure that the loop has had enough sample pass through by collecting the liquid that exits port of the sample pretreatment valve. A good practice is to load at least four loop volumes. The method is set up so that the reagent reservoir pushes sample into the sample loop prior to the IC separation. By applying helium pressure at. kpa ( psi) for.9 minutes ~ ml will have been passed through the sample loop. 8. Ensure that the RP- pump is delivering at a consistent rate of. ±. ml/min. Figure 8 illustrates what happens when the flow rate is faster or slower. At the slower flow rate, not enough of the sample is cut from the ICE separation, resulting in lower recovery of the analytes. Changing the flow rate from. ml/min to. ml/min in this example significantly reduced chloride response. At the faster flow rate, however, more of the ICE separation was cut, resulting in a less separation between chloride and carbonate. 9. Other factors will also affect the quality and consistency of the ICE preseparation. Changing the cut time window (7. to. minutes) specified in the method will impact the amount of analyte and matrix ions that are delivered to the concentrator column. Varying the sample volume will also affect the character of the ICE separation. Method development will be needed to ascertain the impact of any changes from the specified method.. Quantitating the levels of anions in hydrofluoric acid is best accomplished by the method of standard additions. This involves adding one or more increments of a standard solution to sample aliquots of the same size. (See Calibration section.) RESULTS AND DISCUSSION: A blank was determined by using deionized water as a sample. This is used to establish a background level of contamination present from the chromatographic pathway. The only anion present in significant concentration from the initial series of deionized water blanks was sulfate at approximately µg/l, quantified based on sulfate standards in deionized water. A rinse of the column with deionized water followed by replicate injections of concentrated.% HF brought the blank down to µg/l sulfate as shown in Figure 9. This deionized water blank was subtracted from the levels found in the concentrated hydrofluoric acid samples. A chromatogram for the analysis of trace anions in semiconductor grade.% (v/v) hydrofluoric acid is shown in Figure. The large fluoride peak beginning at minutes is well separated from chloride. Using a 8. mm carbonate /. mm sodium hydroxide eluent allows chloride to be well resolved from carbonate. Using an eluent gradient on a hydroxide selective column such as the IonPac AS or AS could have enhanced this separation between fluoride and chloride. The drawback of this approach is that the analysis time increases to 7 minutes. Thus the use of the AS9-HC represents the best compromise in separation efficiency and analysis time. To verify proper quantification of analytes in the hydrofluoric acid matrix, increasing concentrations of chloride, nitrate, phosphate and sulfate were added into the deionized water used to dilute the 9% hydrofluoric acid to.%. Spikes of,, and µg/l of chloride, nitrate, and phosphate and,, and µg/l of sulfate yielded coefficients of determination (r ) values of greater than.99. Based on this calibration curve, values of µg/l nitrate, µg/l phosphate, µg/l chloride, and µg/l sulfate were obtained. These values are spikes at % of the specified levels for the Semiconductor Equipment and Materials ternational (SEMI) guidelines established for the best grade of concentrated hydrofluoric acid. Recoveries were found to be in the target range specified by SEMI between 7% and %, as shown in Table. To determine precision for the method, a sample of semiconductor purity grade hydrofluoric acid diluted to.% (v/v) was analyzed by this method. For n = 7 replicate injections of the same sample, a relative standard deviation (RSD) of less than % was obtained for an average value of 8. µg/l chloride,.9 µg/l nitrate and. µg/l phosphate. Sulfate levels were determined to Table Spike Recovery of Trace Anions in.% (v/v) Hydrofluoric Acid Anion * Spike Found-blank Recovery (µg.l ± S.D.) (µg/l) µg/l ± S.D.) (%) Chloride 8. ±.7 ±. 8 Nitrate.9 ±.9.7 ±. 87 Sulfate. ±..9 ±.9 Phosphate. ±..7 ±.9 87 * Corrected for system blank n = 7 Technical Note 9

10 A Peaks:. Fluoride. Chloride. Carbonate. Sulfate. ml/min Pretreatment : IonPac ICE-AS ICE Eluent: Deionized water ICE Flow Rate:. ml/min Volume: 7 µl : IonPac AG9-HC, AS9-HC, -mm Concentrator: IonPac AG9-HC, mm IC Eluent: 8. mm Sodium carbonate. mm Sodium hydroxide IC Flow Rate:. ml/min Detection; Suppressed conductivity, ASRS-ULTRA, AutoSuppression, External water mode Peaks:. Fluoride µg/l (ppb). Carbonate. Unidentified. Sulfate B. ml/min Figure 9. Representative system blank. C. ml/min have an average concentration of. µg/l with an RSD of.7. This higher RSD is due to the level of sulfate present in the blank. Method detection limits (MDLs) were calculated using the standard deviation of the anions of interest found in seven replicate injections of unspiked semiconductor grade.% HF multiplied by the Student s t value for the 99.% confidence level. MDLs for chloride, sulfate, and nitrate ranged from. to. µg/l (ppb). The MDLs for this method compare favorably with the maximum limit of impurity guidelines for hydrofluoric acid established by SEMI for the purest grade of hydrofluoric acid, as shown in Table. D. ml/min Table Method Detection Limits and SEMI Specifications for Trace Anions in Concentrated High Purity Hydrofluoric Acid Anion Method SEMI Method detection limits Limits (µg/l) C8.-9 Specification (µg/l) HF concentration.% (v/v) 9% (w/w) Chloride. Nitrate. Sulfate. Phosphate. Method Detection Limit = (SD) x (t s ) 99.% where (t s ) is for a single sided Student s t-test distribution for n = 7. Figure 8. Effect of flow rate on ICE separation. Determination of Trace Anions in Concentrated Hydrofluoric Acid

11 Pretreatment : IonPac ICE-AS ICE Eluent: Deionized water ICE Flow Rate:. ml/min Volume: 7 µl : IonPac AG9-HC, AS9-HC, mm Concentrator: IonPac AG9-HC, -mm ICE Eluent: 8. mm Sodium carbonate. mm Sodium hydroxide IC Flow Rate:. ml/min Detection; Suppressed conductivity, ASRS-ULTRA, AutoSuppression, external water mode Peaks:. Fluoride µg/l (ppb). Chloride 7.9. Carbonate. Nitrate.89. Unidentified. Sulfate. 7. Phosphate. Figure. Determination of trace anions in high purity.% hydrofluoric acid. PRECAUTIONS Exercise extreme caution when handling concentrated hydrofluoric acid. Use only the highest quality deionized water for the preparation of standards and eluents. Any ionic contamination present in the deionized water will adversely affect results. Teflon containers are recommended for holding the concentrated acid samples for delivery to the sample loop. Containers should be soaked for at least h with 7.8 MΩ-cm deionized water prior to use. It is good practice to dedicate all containers for trace analysis and keep them filled with deionized water when not in use. The level of sulfate in the deionized water blank should be monitored on a regular basis. Method success depends on maintaining a consistent flow rate of deionized water from the RP- so that the proper fraction is cut from the ICE-AS. Verify that the flow is. ±. ml/min. If the deionized water container feeding the RP- pump is not pressurized to at least. kpa ( psi) the pump will be prone to losing prime. Periodically rinse the Rheodyne valve used for loading sample. Residual hydrofluoric acid can crystallize and block injector passages. Also, impurities present in the concentrated acid samples can lead to a loss of column capacity. This causes an increase in back pressure and shorter retention times. Consult the IonPac AS9-HC column stallation and Troubleshooting Guide for more information. Do not leave concentrated HF in the sample loop and sample inlet lines for more than hours. The PEEK tubing can degrade after extended contact time with the concentrated acidic sample. 7 REFERENCES. Watanabe, K. Presented at the ternational Ion Chromatography Symposium, Dallas, TX, October 99; Poster.. Wu, M.; Chen, J. Micro 997, (), 7.. Bader, M. J. Chem Educ. 98, 7, 7.. Weiss, J. Ion Chromatography, nd ed., VCH, Weinheim, Germany, 99, 9.. Troubleshooting Guide for HPLC jection Problems, Rheodyne: Cotati, CA, 99.. SEMI ternational Standards: Semiconductor Equipment and Materials ternational, Mountain View, CA, Chemical/Reagents Volume, 997. LIST OF SUPPLIERS Fisher Scientific, 7 Forbes Ave., Pittsburgh, PA 9-78, USA. Tel: (8) 7-7 VWR Scientific, P.O. Box 79, San Francisco, CA 9, USA. Tel: (8) 9- Lab Safety Supply c., PO Box 8, Janesville, WI 7-8, USA. Tel: (8) -78 Technical Note

12 IonPac, ASRS, and AutoSuppression are registered trademarks of Dionex Corporation. Printed on recycled & recyclable paper with soy-based inks. Dionex Corporation Dionex Corporation Dionex U.S. Regional Offices Dionex ternational Subsidiaries 8 Titan Way Salt Lake City Technical Center Sunnyvale, CA (8) 77-8 Austria () Belgium () 8 Canada (9) 8-9 France 9 8 Germany -99- P.O. Box West South, Suite A Westmont, IL () 789- Italy () Japan () 88- The Netherlands () Switzerland () 99 United Kingdom (7) 97 Sunnyvale, CA Salt Lake City, UT Houston, TX (8) 87- * Designed, developed, and manufactured under an NSAI registered ISO 9 Quality System Atlanta, GA (77) -8 Determination of Trace Anions in Concentrated Hydrofluoric Acid (8) 77-7 (8) Marlton, NJ (8) 9- LPN -.M / Dionex Corporation

Determination of Trace Anions in Concentrated Glycolic Acid

Determination of Trace Anions in Concentrated Glycolic Acid Technical Note Determination of Trace Anions in Concentrated Glycolic Acid INTRODUCTION There is a need for a reliable method to determine trace chloride and sulfate in glycolic acid. The presence of these

More information

The Determination of Trace Anions in Concentrated Phosphoric Acid

The Determination of Trace Anions in Concentrated Phosphoric Acid Technical Note The Determination of Trace Anions in Concentrated Phosphoric Acid INTRODUCTION The determination of trace anions in phosphoric acid is hampered by a large excess of phosphate ion. Chloride

More information

Determination of trace anions in concentrated glycolic acid

Determination of trace anions in concentrated glycolic acid TECHNICAL NOTE Determination of trace anions in concentrated glycolic acid Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography, IC, Dionex ICS- +,

More information

The determination of trace anions in concentrated phosphoric acid

The determination of trace anions in concentrated phosphoric acid TECHNICAL NOTE The determination of trace anions in concentrated phosphoric acid Authors Edward Kaiser and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Ion chromatography, IC, IonPac

More information

Determination of Trace Anions in Organic Solvents

Determination of Trace Anions in Organic Solvents Application Note 8 Determination of Trace Anions in Organic Solvents INTRODUCTION In the manufacture of semiconductor materials, a great deal of attention is focused on minimizing sources of contamination.

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

Determination of trace anions in high-nitrate matrices by ion chromatography

Determination of trace anions in high-nitrate matrices by ion chromatography APPLICATION NOTE 7 Determination of trace anions in high-nitrate matrices by ion chromatography Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Contamination,

More information

anion-exchange column

anion-exchange column columnsionpac ASA Anion-Exchange Column The IonPac ASA is a carbonateselective anion-exchange column designed for the fast separation of inorganic anions, including fluoride, chlorite, bromate, chloride,

More information

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography APPLICATION NOTE 113 Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography Author Edward Kaiser Thermo Fisher Scientific, Sunnyvale, CA Keywords Trace ionic

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath APPLICATIN NTE 9 Determination of an anionic fluorochemical surfactant in a semiconductor etch bath Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography,

More information

columns IonPac SCS 1 Silica Cation Separator

columns IonPac SCS 1 Silica Cation Separator columns IonPac SCS Silica Cation Separator The SCS Silica Cation Separator is designed for use with nonsuppressed conductivity detection, or single-column ion chromatography (SCIC). This column is particularly

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

columns IonPac AS18 Anion-Exchange Columns

columns IonPac AS18 Anion-Exchange Columns columns IonPac AS Anion-Exchange Columns The IonPac AS is a hydroxideselective anion-exchange column designed for the determination of inorganic anions and low-molecularweight organic acids including fluoride,

More information

columns IonPac AS17-C Anion Exchange Column

columns IonPac AS17-C Anion Exchange Column columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column

More information

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Application Note 75 Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Ethanol used as a blending agent in gasoline can be contaminated with chloride and sulfate that form plugging

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Application of Eluent Generation for Trace Anion Analysis of Borated Waters Application of Eluent Generation for Trace Anion Analysis of Borated Waters Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 166 Introduction Boric acid is used

More information

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography Application Update 144 Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography INTRODUCTION Hexavalent chromium, Cr(VI), is the most toxic form of the metal chromium, a primary

More information

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,

More information

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,

More information

IonPac Trace Cation Concentrator

IonPac Trace Cation Concentrator for the IonPac Trace Cation Concentrator (TCC-LP1, TCC-ULP1 and TCC-XLP1) Page 1 of 17 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR (TCC) TCC-LP1 Column Low Pressure, 4 x 35 mm (P/N 046027) TCC -ULP1 Column

More information

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,

More information

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters Application Update 5 Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters INTRODUCTION For nuclear power plants using a pressurized water

More information

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Inorganic Anions in Wastewater by Ion Chromatography Application Note 5 Determination of Inorganic Anions in Wastewater INTRODUCTION The determination of common inorganic anions in environmental waters, such as wastewater and drinking, ground, and surface

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep pplication Note 70 Determination of Silicate in High Purity Water Using Ion Chromatography and utoprep Introduction The water used in the manufacture of semiconductors and other modern electronic components

More information

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Harpreet Dhillon and John Statler Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 11 Introduction

More information

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device Determination of Bromate in Bottled Mineral Water Using the CRD 00 Carbonate Removal Device Thunyarat Phesatcha, Weerapong Worawirunwong, and Jeff Rohrer Thermo Fisher Scientific, Bangkok, Thailand; Thermo

More information

Inline sample preparation for the determination of anions in sodium hydroxide

Inline sample preparation for the determination of anions in sodium hydroxide APPLICATION UPDATE 72331 Inline sample preparation for the determination of anions in sodium hydroxide Authors Hua Yang and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Sample preparation,

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,

More information

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for

More information

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Errata Product Manual for Dionex IonPac UTAC2 and AC-ER 065376-0 For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Part Old Part Number in

More information

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column Application Note 4 Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS6 Column INTRODUCTION The common alkali and alkaline earth cations are

More information

Thermo Scientific. Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS9-µm Anion-Exchange Column Product Specifications The Themo Scientific Dionex IonPac AS9-µm high-capacity, hydroxide-selective, anion-exchange column is

More information

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of silicate in high-purity water using ion chromatography and online sample preparation PPLICTION NOTE 70 Determination of silicate in high-purity water using ion chromatography and online sample preparation uthors Weerapong Worawirunwong and Jeffrey Rohrer Thermo Fisher Scientific, Bangkok,

More information

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes columns Thermo Scientific Dionex IonPac AS Anion-Exchange Column Isocratic Separation of Sulfur Species and Inorganic Anions Using the Dionex IonPac AS Column. Fluoride. Bromate. Chloride. Nitrite. Bromide.

More information

Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts

Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts Technical Note 26 Determination of in Water, Waste Water, and Solid Waste Extracts INTRODUCTION Chromium, while not unique in its properties, is commonly used in various industries because of the characteristics

More information

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography Application Note 9 Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography Introduction The computer, semiconductor, and food industries need analytical

More information

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Errata Product Manual for Dionex IonPac AS and AG Columns 03479-2 For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Part Old Part Number

More information

for IonPac AG11-HC IonPac AS11-HC

for IonPac AG11-HC IonPac AS11-HC for IonPac AG-HC IonPac AS-HC IonPac AS-HC Manual Document No. 3333-7 Page of 45 PRODUCT MANUAL for the IONPAC AG-HC GUARD COLUMN (4 x 5 mm, P/N 59) ( x 5 mm, P/N 593) IONPAC AS-HC ANALYTICAL COLUMN (4

More information

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System G. Maio 1, F. Steiner 1, H. Franz 1, M. Tracy 2, X. Liu 2, F. Arnold 1 1 Dionex Softron GmbH, D-21 Germering,

More information

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography Application Note 9 Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography INTRODUCTION Determination of cations in concentrated acid is important

More information

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Determination of Tartaric cid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Suparerk Tukkeeree, Chanita Chantarasukon, and Jeff Rohrer Thermo Fisher Scientific, angkok,

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Inorganic Anions in Drinking Water by Ion Chromatography Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

IonPac SCS 1 IonPac SCG 1

IonPac SCS 1 IonPac SCG 1 IonPac SCS 1 IonPac SCG 1 Document No. 031948-05 IonPac SCG1 and SCS 1 Page 1 of 33 PRODUCT MANUAL for the IONPAC SCG 1 Guard Column 2 x 50 mm, P/N 061522 4 x 50 mm, P/N 061523 IONPAC SCS 1 Analytical

More information

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography APPLIATION NOTE 78 Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography Authors Soon Fatt Lee, Aaron Rose, Terri hristison, and Jeff Rohrer Thermo Fisher

More information

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA Determination of Sub-μg/L Bromate in Municipal and Natural Mineral Waters Using Preconcentration with Two-Dimensional Ion Chromatography and Suppressed Conductivity Detection Brian DeBorba and Jeff Rohrer

More information

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Determination of trace anions in organic solvents using matrix elimination and preconcentration APPLICATION UPDATE Determination of trace anions in organic solvents using matrix elimination and preconcentration Authors Terri Christison and Jeff Rohrer Thermo Fisher Scientific Sunnyvale, CA Keywords

More information

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Direct Determination of Small Organic Acids in Sea Water by IC-MS Direct Determination of Small Organic Acids in Sea Water by IC-MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Application Note Key Words MSQ Plus Single Quadrupole Mass

More information

Expanded capabilities in ammonium and amine detection

Expanded capabilities in ammonium and amine detection PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium

More information

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000 Evaluation of Various Anion-Exchange Chemistries for the Analysis of Haloacetic Acids in Drinking Water Using -D Matrix Elimination Ion Chromatography and Suppressed Conductivity Detection White Paper

More information

Determination of Glyphosate by Cation-Exchange Chromatography with Postcolumn Derivatization

Determination of Glyphosate by Cation-Exchange Chromatography with Postcolumn Derivatization Application Note 09 Determination of Glyphosate by Cation-Exchange Chromatography with Postcolumn Derivatization INTRODUCTION Glyphosate is a broad-spectrum herbicide with low mammalian toxicity. The herbicide

More information

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts By Ion Chromatography

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts By Ion Chromatography Application Note 44 Determination of Perchlorate in High Ionic Strength Fertilizer Extracts By Ion Chromatography INTRODUCTION Perchlorate anion (ClO 4- ) is a water soluble, mobile, and persistent environmental

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS column is a high-capacity, hydroxide-selective, anion-exchange column

More information

columns IonPac AS22 Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column columns IonPac AS Anion-Exchange Column The IonPac AS is a carbonate based anion-exchange column designed for the determination of inorganic anions and low-molecular-weight organic acids including fluoride,

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures. User Manual IonPac NS2 Columns 065501 Revision 02 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for IonPac NS2 Columns IonPac NS2, 5µm, 4x250mm Analytical Column

More information

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Carl Fisher and Linda Lopez Thermo Fisher Scientific, Sunnyvale,

More information

suppressors SRS -ULTRA Self-Regenerating Suppressor

suppressors SRS -ULTRA Self-Regenerating Suppressor suppressors SRS -ULTRA Self-Regenerating Suppressor The Self- Regenerating Suppressor utilizes AutoSuppression technology to enhance analyte conductivity while suppressing eluent conductivity. This results

More information

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS4A Anion-Exchange Column Product Specifications Thermo Scientific Dionex IonPac AS4A anion-exchange column is designed for the separation of haloacetic

More information

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters PPLICTION NOTE 7229 Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters uthors Jingli Hu and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, C, US Keywords

More information

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System arbara Shao, Frank Hoefler, Maria Rey, and Linda Lopez, Thermo Fisher Scientific,

More information

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography Dave Thomas and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 44 Introduction Perchlorate

More information

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 This method provides procedures for

More information

Determination of Methylamine in Drug Products

Determination of Methylamine in Drug Products Determination of Methylamine in Drug Products Thunyarat Phesatcha, Suparerk Tukkeeree, and Jeff Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application

More information

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Kirk Chassaniol Manager of IC Technical Sales Support Thermo Fisher Scientific 1 The world leader in serving

More information

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent APPLICATION UPDATE 72588 Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent Authors Manali Aggrawal and Jeffrey Rohrer Thermo

More information

columns I o n P a c AS11 and AS11-H C Anion-E x c h a n g e C o l u m n s

columns I o n P a c AS11 and AS11-H C Anion-E x c h a n g e C o l u m n s columns I o n P a c AS and AS-H C Anion-E x c h a n g e C o l u m n s The IonPac AS anion-exchange column provides fast profiling of inorganic anions and organic acid anions using sodium hydroxide or potassium

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Novel Electrodeionization Devices: Applications in Inorganic Analysis

Novel Electrodeionization Devices: Applications in Inorganic Analysis Novel Electrodeionization Devices: Applications in Inorganic Analysis John M. Riviello and Archava Siriraks ABSTRACT Electrodeionization (EDI) is a well established technique for the production of ultrapure

More information

Determination of Lanthanide Metals in Digested Rock Samples by Chelation Ion Chromatography

Determination of Lanthanide Metals in Digested Rock Samples by Chelation Ion Chromatography Technical Note 27 Determination of Lanthanide Metals in Digested Rock Samples by Chelation Ion Chromatography INTRODUCTION The naturally occurring high concentrations of certain metals such as iron, aluminum,

More information

PRODUCT MANUAL. for the. IONPAC CG14 GUARD COLUMN (2 x 50 mm, P/N ) (4 x 50 mm, P/N )

PRODUCT MANUAL. for the. IONPAC CG14 GUARD COLUMN (2 x 50 mm, P/N ) (4 x 50 mm, P/N ) IonPac CS IonPac CS Manual Doc. No. 88-9 Page of PRODUCT MANUAL for the IONPAC CG GUARD COLUMN ( x mm, P/N ) ( x mm, P/N ) IONPAC CS ANALYTICAL COLUMN ( x mm, P/N ) ( x mm, P/N ) Dionex Corporation, Document

More information

California Environmental Protection Agency Air Resources Board

California Environmental Protection Agency Air Resources Board California Environmental Protection Agency Air Resources Board PROCEDURE FOR THE ANALYSIS OF PARTICULATE ANIONS AND CATIONS IN MOTOR VEHICLE EXHAUST BY ION CHROMATOGRAPHY Standard Operating Procedure No.

More information

PRODUCT MANUAL OMNIPAC PCX-500 GUARD COLUMN (PCX-500 GUARD, P/N ) OMNIPAC PCX-500 ANALYTICAL COLUMN (PCX-500 ANALYTICAL, P/N )

PRODUCT MANUAL OMNIPAC PCX-500 GUARD COLUMN (PCX-500 GUARD, P/N ) OMNIPAC PCX-500 ANALYTICAL COLUMN (PCX-500 ANALYTICAL, P/N ) PRODUCT MANUAL OMNIPAC PCX-500 GUARD COLUMN (PCX-500 GUARD, P/N 042195) OMNIPAC PCX-500 ANALYTICAL COLUMN (PCX-500 ANALYTICAL, P/N 042191) QUICKSTART STEPS AND LINKS Click blue text below to get started.

More information

Waters. Ion Chromatography

Waters. Ion Chromatography 808q Waters Ion Chromatography Method Anion Analysis Using Hydroxide Eluent and Indirect Conductivity Detection and EPA Method B-1011 for Nitrite and Nitrate Using UV Detection 2000 Required Instrumentation:

More information

suppressors Eluent Suppressors for Ion Chromatography

suppressors Eluent Suppressors for Ion Chromatography suppressors Eluent Suppressors for Ion Chromatography The Suppressor Advantage Dionex introduced suppression in 97, thereby bringing ion chromatography (IC) to the forefront of modern analytical techniques

More information

Product Manual for Dionex IonPac CS12A and CG12A Columns

Product Manual for Dionex IonPac CS12A and CG12A Columns Errata Product Manual for Dionex IonPac CSA and CGA Columns -9 For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Part Old Part Number in

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Chemistry Instrumental Analysis Lecture 31. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column Chromatography Thermo Scientific Anion-Exchange Column Product Specifications Attain the optimal resolution of organic acids and inorganic anions using the new Thermo Scientific Dionex IonPac AS11-HC-4µm

More information

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of

More information

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY As originally published in the SMTA Proceedings. DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY Christopher J. Pontius,

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection columns IonPac AS Anion-Exchange Column The IonPac AS is a highcapacity, hydroxide-selective anion exchange column designed for separation of haloacetic acids (HAAs) and bromate in drinking water prior

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information