Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry

Size: px
Start display at page:

Download "Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry"

Transcription

1 2005 The Japan Society for Analytical Chemistry 143 Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry M. K. TIWARI, A. K. SINGH, and K. J. S. SAWHNEY Synchrotron Utilization Division, Centre for Advanced Technology, Indore , India The influence of analyte mass concentration on determination of detection limits in X-ray fluorescence spectrometry has been investigated experimentally. Both the total reflection X-ray fluorescence (TXRF) and the conventional energydispersive X-ray fluorescence techniques have been used to derive the dependence of analyte mass concentration on the values of detection limits. Results obtained indicate that values of detection limits are optimum, or in other words, they are closer to the true detection limit of the technique, when analyte concentrations are in the range of 10 times of the detection limit. (Received June 30, 2004; Accepted September 10, 2004) Introduction X-ray fluorescence spectrometry is one of the widely used techniques for elemental analysis of materials. Using the energy-dispersive X-ray fluorescence (EDXRF) analysis technique, non-destructive multielement analysis at ppm levels can be performed for elements ranging from sodium to uranium in the periodic table. 1,2 The technique finds several applications to a variety of fields such as archeological, geological and environmental. 3 6 Total reflection X-ray fluorescence (TXRF) technique is a variant of the conventional EDXRF technique, 7 9 where the principle of total external reflection is used to restrict the penetration of the primary beam into the specimen substrate. This yields a substantial reduction of scattered background, which improves the detection limit of the system to subppm/ppb levels. The term detection limit (DL) is an important aspect of any technique because it is an indication of the sensitivity of the technique. As the name itself indicates, the detection limit is a measure of the minimum quantity that can be detected by an instrument or technique under the given experimental conditions (sample geometry, sample matrix, analysis time, etc.). A through description about detection limits can be obtained from publications of Currie et al The detection values for an element are largely influenced by these parameters. For example, in a multielement sample, the presence of several peaks in the fluorescence spectrum can modify the spectral background and hence change the DL value of the element of interest. Such a limitation sometimes results in non-detects i.e. some concentrations that cannot be measured directly. 15 The problem of non-detects often occurs for biological and geological samples where the sample matrix strongly influences the spectral background. The presence of non-detects in XRF and TXRF analysis techniques has been To whom correspondence should be addressed. mktiwari@cat.ernet.in K. J. S. S. present address: Diamond Light Source Ltd., Rutherford Appleton Laboratory, Chilton, Didcto-OX11 0QX, UK. described in detail by Pajek et al. They employed the process of Censoring for determining the detection sensitivities in TXRF. The parameters that improve the detection limits of various X-ray spectrometric techniques, such as PIXE, WDXRF, synchrotron excited XRF and γ-ray spectroscopy, have been discussed thoroughly by several workers Furthermore, it has also shown that in X-ray tube excited XRF, the detection limit can be improved by properly selecting the primary filters. 20 Moreover, a more elaborate theoretical description for detection limit calculation in TXRF with various modes of analysis, viz. grazing incidence (TXRF) and grazing emission (GE-XRF) conditions, has been given by Sanchez. 21 With developments/innovations in the fields of X-ray sources, sample preparation, nuclear electronics, detection systems and data processing procedures in the last decades, the XRF techniques have also been refined to give better and better DLs. Synchrotron radiation based X-ray fluorescence (SR-XRF) and total reflection X-ray fluorescence (SR-TXRF) techniques are the examples of such trends, where parts per trillion (ppt) level or, in absolute mass, femto gram detection sensitivities have been achieved Normally, XRF techniques are comparative in nature and type-standards are generally used to calibrate the spectrometers. Similarly, synthetic standards are employed to determine the DLs. In the literature, no guidelines are available for the selection of analyte concentration to be used for detection limit determination. Our experience, over the years, has been that the DLs tend to depend on the analyte concentration in the synthetic standard employed to determine the DLs. We have undertaken a systematic study to resolve this issue; our results give guidelines for what the analyte mass concentrations should be for achieving the best possible DLs under the given experimental conditions. This study indicates that, to derive the optimum detection limit values, one should locate the analyte mass concentration in the synthetic standards in the range of 10 times the detection limit. Theoretical Background Mathematically, as per the IUPAC recommendation, the

2 144 ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21 Fig. 1 Variation of Zn detection limits with analyte concentration determined using EDXRF technique for different acquisition times. Excitation source: Cd 109. a, Solid square (s); b, solid circle (s); c, solid triangle (g) show the measurements performed for acquisitions time 1000, 5000 and s, respectively whereas a solid line (m) shows the polynomial fit. detection limit has been defined in terms of standard counting error of the background intensity. 1 If σ P and σ B are the counting errors of the individual peak and the background, respectively, then the standard counting error for a net count can be expressed as, σ = In the limit of detection one can assume σ P σ B. Therefore σ For 95% confidence level, the counting error would be 2σ 3ρ B = 3 N B By taking into account the counting time T and the slope m of fluorescence intensity vs. analyte concentration curve, one can write the expression for minimum detectable concentration for an analyte element as or 2σ B C DL = 3 m σ P2 + σ B 2 C DL = 3 I B I A C A N B T where, I B = background intensity (N B/T); I A = net area intensity; C A = mass concentration of analyte; N B = background counts in a given time T; m = I A/C A, i.e., slope of analyte countsconcentration curve; C DL = minimum detectable concentration (1) of analyte. From Eq. (1), it is clear that, for a specified counting time, the minimum detectable concentration of analyte or detection limit depends mainly on two parameters: the background intensity I B and the net area intensity per unit analyte mass (I A/C A). Experimental Both the EDXRF and TXRF techniques have been used to perform X-ray fluorescence measurements. For EDXRF, an inhouse developed EDXRF spectrometer 25 has been employed that comprises of a Si(Li) detector of energy resolution 155 ev at 5.9 kev and an EG&G ORTEC 92X spectrum master that has a built-in spectroscopy amplifier, a multi channel analyzer and a HV bias supply. A 25 mci Cd 109 annular radioisotope is used as an excitation source. Several synthetic standards, ranging in three decades of analyte concentrations, were prepared. Zn and Cr elements were chosen as the analyte elements and their synthetic standards were prepared in the form of pellets of mass 100 mg/cm 2, by mixing their appropriate salts in cellulose matrix. Moreover, to confirm the reproducibility of system geometry, we prepared synthetic standards of various elements in various concentrations range and we analyzed them separately. The system geometry was found to reproduce within ±0.5%. To perform the same studies using the TXRF technique, researchers have used the indigenously developed TXRF spectrometer. 26 Aqueous residues of U of different concentrations were prepared from liquid solutions and polished Si substrates were used as the sample carriers. Mo X-ray tube (17.4 kev) was used as the excitation source.

3 145 Fig. 2 Variation of relative uncertainty for a signal at different analyte concentrations (expressed in DL units) for two different risk levels, 5% and 0.05%, calculated using Eq. (2). The variation of selfabsorption, A-factor, for Zn-K α signal at different analyte concentrations is also shown in this figure. The figure shows that for analyte mass ranging in times of DL value, both selfabsorption and relative uncertainties are optimum. Results and Discussion Figures 1a, b and c show the variation of detection limits of Zn with various concentrations of Zn for analysis times of 1000 s, 5000 s and s, respectively. Since the DLs depend inversely on the square root of analysis time, the DLs are better for larger analysis times (Fig. 1). The figure shows that the dependency of DL on analyte mass concentration generates broad minima and that the DLs are minimum for analyte concentrations ranging from times of DL. Moreover, one can see that the selection of an analyte concentration about 10 times that of DL is more appropriate for getting the minimum DL value. For example, for 1000 s (5000 s) analysis time, the minimum of DL for Zn is 59 ( 27) ppm which corresponds to analyte mass of 600 (300) ppm. Similarly, for analysis time of s, Zn DL is minimum ( 18 ppm) for the analyte mass of 200 ppm. The appropriateness of analyte concentration corresponding to 10 times that of DL can be explained by the Currie hypothesis of third detection limit. 10 Beside detection limit, which is a characteristic of a measurement, Currie given a concept of third detection limit, determination limit, L Q, above which the precision of a measurement can be made good enough. The relationship between determination limit L Q and detection limit C DL can be approximated by 16 L Q = C DL 2kr where, k defines the risk factor for a signal in terms of a fractile of a Gaussian distribution and r is the required relative uncertainty. If L Q is set to n times of C DL i.e. L Q = nc DL then one can derive a relation, r = 1 (2) 2nk Fig. 3 Influence of analyte concentration on the background intensity and slope m of analyte fluorescence intensity concentration calibration curve. Analysis time, s; analyte, Zn; slope m, (å); background intensity, ( ). In the limit of detection where self-absorption effects 27 for analyte elements are significantly low, L Q simply represents the smallest analyte mass that can be expected to generate a net signal intensity with a relative error less than a selected limit r. This can be understood more clearly from Fig. 2. Here, the relative uncertainty for a signal at different levels of analyte mass (expressed in DL units) has been plotted for two different risk level 5% (k = 1.645) and 0.05% (k = 3.291) along with the self absorption of Zn-K α signal. For analyte masses ranging around times of DL, the relative uncertainties are significantly low, e.g. comparable to acceptable limits of statistical errors ( 0.5 2%) and at the same time selfabsorption effects do not affect severely the analyte signal intensity. Beyond this range, at higher analyte concentrations, though the relative uncertainties improve, the sample selfabsorption effects become dominant. Figure shows that selecting analyte concentration in the range of times of DL would yield a fluorescence signal under acceptable uncertainties. Self-absorption effects for analyte signal can be ignored in this range. Though one can use any analyte mass between times of DL value for detection limit measurement, it is a common practice to use an analyte mass close to the DL value i.e. 10 times of DL. The applicability of analyte mass corresponding to 10 times of DL is shown in Fig. 3. Here, the net area intensity per unit analyte concentration and the background fluorescence intensity have been plotted as a function of Zn analyte concentration. This figure shows that net area intensity of Zn-K α per unit analyte concentration, also referred to as the slope in Eq. (1), show a peaking behavior for Zn analyte concentration of 200 ppm. On the other hand, the background intensity increases monotonously, though this increase is very small. This is because, at higher analyte concentrations, apart from sample absorption effects, the tailing of the fluorescence peak increases, leading to higher background below the fluorescence peak, which in turn reduces the value of slope m. If one lowers the analyte concentration, the background that appears due to increase of tailing width of the fluorescence peak can be reduced significantly. In this case it becomes almost constant and equal to the value that one obtains in case of a blank sample. By optimizing the width of the fluorescence peak within 3σ limit and keeping background line well-known in the area of the peak, the optimum value of m can be obtained for

4 146 ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21 Fig. 4 X-ray fluorescence spectra recorded for various concentrations of Zn analyte. Analysis time, s; excitation source, Cd 109 radioisotope source. The figure shows that background region/intensity under a fluorescence peak increases non-linearly with increasing analyte concentrations. Fig. 5 Variation of U detection limits with analyte concentration determined using TXRF technique. Excitation source: Mo X-ray tube, 40 kv, 25 ma. Open circles (A) shows the measurements performed for an acquisition time of 500 s whereas the solid line (m) shows the polynomial fit. The inset shows the variation of net area intensity for uranium L α (slope m) and the background intensity at different analyte concentrations. analyte mass 10 times of DL. Figure 4 shows a more elaborated picture of this non-linear behavior of the background intensity, wherein fluorescence spectra for various analyte concentrations have been plotted. The figure shows that, at large analyte concentrations the background region/intensity under a fluorescence peak is higher compared to that at lower analyte concentrations, at the same time the net area intensity behaves linearly at lower concentrations, which is not strictly true at higher analyte concentrations because of sample absorption effects. Since slope m as well as the background intensity I B governs the detection limit value as defined in Eq. (1), the detection limits improve for analyte concentrations ranging in ppm and a minimum is obtained at analyte concentration of 200 ppm. Further reduction of analyte concentration leads to poor statistics of analyte peak, resulting in less than optimum values for DL. It is often true at very low analyte concentrations that the number of counts that appears under a fluorescence peak sometimes may not be sufficient to support the Gaussian approximation i.e. the shape of the fluorescence peak no longer remains Gaussian; rather, it shifts to a Poisson distribution. This not only results in large errors in net area estimation but it also sometimes becomes difficult to define properly a window/roi for a fluorescence peak in a multichannel pulse height analyzer (MCA) spectra. Although the detection limit values determined for two analytes Cr and Zn are found to be minimum for analyte concentrations ranging from times of DLs, the dependency of DL on analyte concentrations still shows that the selection of analyte concentration corresponding to 10 times of DLs would be better for getting optimum values of DL. To further investigate the influence of analyte concentrations on detection limit values, an independent study using TXRF technique has also been performed. The detection limit values in TXRF technique are generally determined by employing aqueous residue generated from synthetic or standard solutions. We used various concentrations of synthetic-standard solution of uranium for determining the detection limit of uranium using this technique. The dependency of detection limit on analyte concentration studied using TXRF technique, Fig. 5, reconfirmed that the best DL is obtained for analyte mass concentration corresponding to 10 times of DL value. For instance, if the true detection limit of the TXRF technique for uranium is 10 ppb, then 100 ppb (0.1 ppm) of analyte concentration would be appropriate for determining this detection limit. In TXRF, as we have used aqueous samples for determination of DLs, apart from the reasons discussed above that influences background intensity, there is a certain probability of increased surface roughness at large volumes of analyte. The inset of Fig. 5 shows the variation of net area intensity and background intensity as a function of analyte (uranium) concentration. It is found that net area intensity varies like it does in cases of conventional XRF measurements, while the increase of background is more rapid at large analyte concentrations. This is mainly due to relatively large surface roughnesses induced at higher analyte mass/volumes. We have also taken into account the total errors contributed in DL measurement as well as in preparation of analyte concentration. Since the detection limit depends linearly on analyte concentration, the net error in these two parameters has been calculated partially. We have assumed that all measurement errors; viz., errors in net area intensity, sample preparation error, statistical error, or geometry reproducibility, influence the DL value and have been determined using synthetic standards of known concentrations. The total error contributed to DL measurement has been found to vary from 2 8%. The error is higher at low analyte concentration and in this case the statistical errors are dominant. On the other hand, to estimate errors in preparation of analyte concentrations, we prepared three sets of synthetic standards and analyzed each standard for each concentration of analyte element. The net error in analyte mass concentration has been determined indirectly by measuring the fluorescence intensity. The total error in analyte concentration has been found to vary from 2 10 %. Conclusions The systematic study presented here gives guidelines for the most suitable analyte mass concentrations for achieving the best possible DLs in XRF techniques. The results obtained indicate that the optimum value of detection limit will correspond to analyte mass ~10 times of the best detection limit. The same

5 147 inference can also be drawn from the TRXF analysis techniques results, thereby indicating widespread possible applications of this guideline especially in calibration of XRF spectrometers for their DLs under given experimental conditions. Acknowledgements The authors thank R. V. Nandedkar for fruitful discussions and suggestions. References 1. E. P. Bertin, Principles and Practice of X-ray Spectrometric Analysis, 1975, Plenum Press, New York. 2. R. E. Van Grieken, Hand Book of X-ray Spectrometry, 2nd ed., 2002, Marcel Dekker, Inc., New York, K. J. S. Sawhney and G. S. Lodha, Computers and Geosciences, 1989, 15(7), G. Misra, K. J. S. Sawhney, G. S. Lodha, V. K. Mittal, and H. S. Sahota, Appl. Radiat. Isot., 1992, 43(5), S. Piorek, Nucl. Instr. Meth., 1994, A353, V. Zaichick, N. Ovchjarenko, and S. Zaichick, Appl. Radiat. Isot., 1999, 50(2), Y. Yoneda and T. Horiuchi, Rev. Sci. Instrum., 1971, 42, P. Wobrauschek and H. Aiginger, Anal. Chem., 1975, 47, R. Klockenkamper, Total-Reflection X-ray Fluorescence Analysis, 1997, John Wiley and Sons, Inc., New York. 10. L. A. Currie, Anal. Chem., 1968, 40, L. A. Currie, Lower limit of detection: definition and elaboration of a proposed position for radiological effluent and environmental measurements, 1984, Report issued by the US Nuclear Regulatory Commission, NUREG/CR- 4007, L. A. Currie, Anal. Chim. Acta, 1999, 391, L. A. Currie, J. Radioanal. Nucl. Chem., 2000, 245, L. A. Currie, Detection: international update, and some emerging dilemmas involving calibration, the blank, and multiple detection decisions, 1997, Vol. 37, Chemometrics Intell. Lab Systems, M. Pajek and A. Kubala-Kukus, in 10th International Conference on Total Reflection X-ray Fluorescence Analysis and the 39th Annual Conference on X-ray Chemical Analysis (TXRF 2003), 2003, Sept , Hyogo, Japan, Lars-Erik De Geer, Appl. Radiat. Isot., 2004, 61, L. A. Currie, Appl. Radiat. Isot., 2004, 61, G. L. Long and J. D. Winefordner, Anal. Chem., 1983, 55, 712A. 19. C. J. Sparks Jr., Synchrotron Radiation Research, ed. H. Winick and S. Z. Doniach, 1980, Plenum Press, New York, M. N. Ingham and B. A. R. Vrebos, Adv. X-ray Anal., 1994, 37, H. J. Sanchez, Spectrochim. Acta Part B, 2001, 56, K. Baur, S. Brennan, B. Burrow, D. Werho, and P. Pianetta, Spectrochim. Acta, Part B, 2001, 56, N. Awaji et al., Jpn. J. Appl. Phys., 2000, 39, Pt. 2, No. 12A, L C. Streli, P. Wobrauschek, H. Aiginger, W. Ladisich, and R. Rieder, Adv. X-ray Anal., 1994, 37, K. J. S. Sawhney, M. K. Tiwari, A. K Singh, and R. V. Nandedkar, Proc. 6th National Seminar on X-ray Spectroscopy and Allied Areas, ed. S. K. Joshi, B. D. Shrivastava, and A. P. Deshpande, 1998, Narosa, New Delhi, M. K. Tiwari, B. Gowri Sankar, V. K. Raghuvanshi, R. V. Nandedkar, and K. J. S. Sawhney, Bull. Mater. Sci., 2002, 25(5), Sample absorption correction factor, A, in X-ray fluorescence 1 exp µ (E 0) µ + (E i) sinφ sinϕ M can be evaluated from A =, µ (E 0) + µ (E i) sinφ sinϕ M where µ(e 0) and µ(e i) represent sample mass absorption coefficients at incident energy E 0 and characteristic fluorescence line energy of the element of interest. φ and ϕ are incident and exit angles; M represents the mass of the pellet in g/cm 2. To determine the self-absorption correction for analyte signal, normalization of A factor to that in case of blank sample is needed (A = 1 represents no absorption and A = 1.10 means 10% absorption of analyte intensity).

Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis

Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis Bull. Mater. Sci., Vol. 25, No. 5, October 2002, pp. 435 441. Indian Academy of Sciences. Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis M K TIWARI*,

More information

Laboratory and Synchrotron Radiation total. reflection X-ray fluorescence: New Perspectives in. Detection Limits and Data Analysis

Laboratory and Synchrotron Radiation total. reflection X-ray fluorescence: New Perspectives in. Detection Limits and Data Analysis Laboratory and Synchrotron Radiation total reflection X-ray fluorescence: New Perspectives in Detection Limits and Data Analysis K. Baur 1*, S. Brennan 1, B. Burrow 2, D. Werho 3 and P. Pianetta 1 1 Stanford

More information

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater PRAMANA c Indian Academy of Sciences Vol. 76, No. 2 journal of February 2011 physics pp. 361 366 Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

More information

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS 227 IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS Fusheng Li, Weijun Guo, and Robin P. Gardner Center for Engineering Applications of

More information

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Copyright JCPDS-International Centre for Diffraction Data 2012 ISSN 1097-0002 281 IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Susumu Imashuku 1, Deh Ping Tee 1, Yasukazu

More information

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF 2016 Fall Semester MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF Byungha Shin Dept. of MSE, KAIST 1 Course Information Syllabus 1. Overview of various characterization techniques (1 lecture)

More information

SYNCHROTRON RADIATION INDUCED X-RAY EMISSION - SRIXE W.M. KWIATEK

SYNCHROTRON RADIATION INDUCED X-RAY EMISSION - SRIXE W.M. KWIATEK Vol. 82 (1992) -ACTA PHYSICA POLONICA A No 2 Proceedings of the ISSSRNS,92, Jaszowiec 1992 SYNCHROTRON RADIATION INDUCED X-RAY EMISSION - SRIXE W.M. KWIATEK Institute of Nuclear Physics, Department of

More information

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DEVELOPMENT

More information

Portable type TXRF analyzer: Ourstex 200TX

Portable type TXRF analyzer: Ourstex 200TX Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp. 115-123 (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray

More information

INFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS

INFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS 111 INFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS C. Horntrich 1, F. Meirer 1, C. Streli 1, P. Kregsamer 1, G. Pepponi 2, N. Zoeger 1, P. Wobrauschek 1 1 Vienna University

More information

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON

More information

Use of Feedback to Maximize Photon Count Rate in XRF Spectroscopy

Use of Feedback to Maximize Photon Count Rate in XRF Spectroscopy Use of Feedback to Maximize Photon Count Rate in XRF Spectroscopy Benjamin A Lucas Abstract: The effective bandwidth of an energy dispersive x-ray fluorescence spectroscopy system is limited by the timing

More information

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials 296 Time-Resolved μ-xrf and Elemental Mapping of Biological Materials K. Tsuji 1,2), K. Tsutsumimoto 1), K. Nakano 1,2), K. Tanaka 1), A. Okhrimovskyy 1), Y. Konishi 1), and X. Ding 3) 1) Department of

More information

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS 45 ABSTRACT FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS W. T. Elam, Robert B. Shen, Bruce Scruggs, and Joseph A. Nicolosi EDAX, Inc. Mahwah, NJ 70430 European Community Directive 2002/95/EC

More information

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory Altitude influence of elemental distribution in grass from Rila mountain Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory I. Introduction The application of modern instrumental

More information

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer* In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer* Anthony Thomas 1, Joachim Heckel 1, and Dirk Wissmann 1 Search and Discovery Article #41836 (2016)

More information

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 505 QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.

More information

Advances in Field-Portable XRF

Advances in Field-Portable XRF Advances in Field-Portable XRF Volker Thomsen and Debbie Schatzlein Field-portable x-ray fluorescence (XRF) allows us to take the laboratory to the sample. The latest generation of such handheld x-ray

More information

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF 135 Abstract BENEFITS OF IMPROVED RESOLUTION FOR EDXRF R. Redus 1, T. Pantazis 1, J. Pantazis 1, A. Huber 1, B. Cross 2 1 Amptek, Inc., 14 DeAngelo Dr, Bedford MA 01730, 781-275-2242, www.amptek.com 2

More information

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS ANTOANETA ENE 1, I. V. POPESCU 2, T. BÃDICÃ 3, C. BEªLIU 4 1 Department of Physics, Faculty

More information

print first name print last name print student id grade

print first name print last name print student id grade print first name print last name print student id grade Experiment 2 X-ray fluorescence X-ray fluorescence (XRF) and X-ray diffraction (XRD) may be used to determine the constituent elements and the crystalline

More information

EDS User School. Principles of Electron Beam Microanalysis

EDS User School. Principles of Electron Beam Microanalysis EDS User School Principles of Electron Beam Microanalysis Outline 1.) Beam-specimen interactions 2.) EDS spectra: Origin of Bremsstrahlung and characteristic peaks 3.) Moseley s law 4.) Characteristic

More information

Summer Students lectures

Summer Students lectures Summer Students lectures XRF: X-ray fluorescence spectrometry Matthias Alfeld XRF: X-ray fluorescence spectrometry Hamburg, 13.08.13 > What is XRF? X-Ray Fluorescence spectrometry > What can it do? Detect

More information

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1) Project Memorandum To : Jebediah Q. Dingus, Gamma Products Inc. From : Patrick R. LeClair, Material Characterization Associates, Inc. Re : 662 kev Gamma ray shielding Date : January 5, 2010 PH255 S10 LeClair

More information

Within the vast field of atomic physics, collisions of heavy ions with atoms define

Within the vast field of atomic physics, collisions of heavy ions with atoms define Chapter 1 Introduction Within the vast field of atomic physics, collisions of heavy ions with atoms define one of the most active areas of research. In the last decades, the design and construction of

More information

XUV 773: X-Ray Fluorescence Analysis of Gemstones

XUV 773: X-Ray Fluorescence Analysis of Gemstones Fischer Application report vr118 HELM UT FISCHER GMBH + CO. KG Institut für Elektronik und Messtechnik Industriestrasse 21-7169 Sindelfingen, Germany Tel.: (+49) 731 33- - Fax: (+49) 731 33-79 E-Mail:

More information

Semiconductor Detector

Semiconductor Detector Semiconductor Detector General 1. Introduction A semiconductor detector is used for a. checking the radionuclidic purity of radiopharmaceuticals and calibration sources; b. the quantitative analysis of

More information

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data Indian Journal of Pure & Applied Physics Vol. 54, Februray 2016, pp. 137-143 Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental

More information

Overview of X-Ray Fluorescence Analysis

Overview of X-Ray Fluorescence Analysis Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic

More information

Determination of the activity of radionuclides

Determination of the activity of radionuclides BUREAU NATIONAL DE MÉTROLOGIE COMMISSARIAT À L'ÉNERGIE ATOMIQUE LABORATOIRE NATIONAL HENRI BECQUEREL Note technique LNHB/04-33 Determination of the activity of radionuclides contained in volume samples

More information

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS 255 Yoshiyuki Kataoka 1, Naoki Kawahara 1, Shinya Hara 1, Yasujiro Yamada 1, Takashi Matsuo 1, Michael Mantler 2 1 Rigaku

More information

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS 302 ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS M. Song 1, K. Mitsuishi 1, M. Takeguchi 1, K. Furuya 1, R. C. Birtcher 2 1 High Voltage Electron Microscopy Station, National

More information

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR A. C. Neiva 1, J. N. Dron 1, L. B. Lopes 1 1 Escola Politécnica da Universidade de São Paulo

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Quantitative XRF Analysis. algorithms and their practical use

Quantitative XRF Analysis. algorithms and their practical use Joint ICTP-IAEA School on Novel Experimental Methodologies for Synchrotron Radiation Applications in Nano-science and Environmental Monitoring Quantitative XRF Analysis algorithms and their practical use

More information

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto

More information

Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32

Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32 Armenian Journal of Physics, 2016, vol. 9, issue 1, pp. 29-33 Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32 A. Gyurjinyan 1,2, R. Avetisyan 1,2 1 Experimental Physics Division, Yerevan

More information

Spectroscopy on Mars!

Spectroscopy on Mars! Spectroscopy on Mars! Pathfinder Spirit and Opportunity Real World Friday H2A The Mars Pathfinder: Geological Elemental Analysis On December 4th, 1996, the Mars Pathfinder was launched from earth to begin

More information

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: 1 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: XRF spectroscopy 1 exam question. Notes on: www.nuigalway.ie/nanoscale/3rdspectroscopy.html XRF books: Analytical Chemistry,

More information

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS 390 ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS J. C. Rao 1, 2 *, M. Song 2, K. Mitsuishi 2, M. Takeguchi 2, K. Furuya 2 1 Department

More information

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes Nuclear Instruments and Methods in Physics Research B 213 (24) 267 271 www.elsevier.com/locate/nimb X-ray spectrometry with Peltier-cooled large area avalanche photodiodes L.M.P. Fernandes, J.A.M. Lopes,

More information

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Alison V. Goodsell, William S. Charlton alisong@tamu.edu, charlton@ne.tamu.edu Nuclear Security Science & Policy Institute Texas A&M University,

More information

Measurements of K- shell production cross-section and fluorescence yield for Y element

Measurements of K- shell production cross-section and fluorescence yield for Y element American Journal of Physics and Applications 2015; 3(1): 1-5 Published online January 29, 2015 (http://www.sciencepublishinggroup.com/j/ajpa) doi: 10.11648/j.ajpa.20150301.11 ISSN: 2330-4286 (Print); ISSN:

More information

Atomic and Nuclear Analytical Methods

Atomic and Nuclear Analytical Methods H.R. Verma Atomic and Nuclear Analytical Methods XRF, Mössbauer, XPS, NAA and Ion-Beam Spectroscopic Techniques With 128 Figures and 24 Tables Springer Contents 1 X-ray Fluorescence (XRF) and Particle-Induced

More information

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 266 NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

More information

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract Organized and hosted by the Canadian Nuclear Society. Vancouver, BC, Canada. 2006 September 10-14 Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor

More information

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 236 ABSTRACT TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Hagen Stosnach Röntec GmbH,

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

INTERNATIONAL JOURNAL OF SCIENTIFIC & TECHNOLOGY RESEARCH VOLUME 3, ISSUE 2, FEBRUARY 2014 ISSN

INTERNATIONAL JOURNAL OF SCIENTIFIC & TECHNOLOGY RESEARCH VOLUME 3, ISSUE 2, FEBRUARY 2014 ISSN Implication Of X-Ray Path, Region Of Interest, Tube Current And Voltage In Calibration Of X- Ray Fluorescence Instrument: A Case Study Of X-Supreme 8000 Amuda, A.K., Okoh, S., Ekwuribe, S., Bashir, M.

More information

arxiv:nucl-ex/ v2 21 Jul 2005

arxiv:nucl-ex/ v2 21 Jul 2005 Gamma-spectrometric uranium age-dating using intrinsic efficiency calibration arxiv:nucl-ex/0506029v2 21 Jul 2005 Cong Tam Nguyen and József Zsigrai Institute of Isotopes of the Hungarian Academy of Sciences

More information

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production Experiment N2: Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production References: 1. Experiments in Nuclear

More information

The Best Gamma-Ray Detector

The Best Gamma-Ray Detector /sec The Best Gamma-Ray Detector What is the best detector for gamma-ray spectroscopy? Amptek, Inc. provides signal processing electronics which can be used with a variety of detectors (some available

More information

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS) Activities of the neutron standardization at the Korea Research Institute of Standards and Science (KRISS) I. Introduction The activities of neutron standardization in KRISS have been continued for last

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron

More information

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA 90 RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA Christopher G. Worley Los Alamos National Laboratory, MS G740, Los Alamos, NM 87545 ABSTRACT Energy dispersive X-ray fluorescence (EDXRF) is a rapid, straightforward

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production Experiment N2: Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production References: 1. Experiments in Nuclear

More information

Neutron activation analysis. Contents. Introduction

Neutron activation analysis. Contents. Introduction Neutron activation analysis Contents Neutron activation analysis... 1 Introduction... 1 Principle of method... 2 Detection of radionuclides... 3 Kinetics of activation... 4 Choosing the appropriate procedure...

More information

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia 7 th Workshop on European Collaboration for Higher Education and Research in Nuclear Engineering & Radiological Protection Bruxelles, Belgique 30 May - 1 June 2011 Activities at the Laboratory of the Nuclear

More information

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000 Advanced Laboratory Experiments Universität Siegen Prof. Dr. I. Fleck Beta Spectroscopy Abstract The experiment on beta spectroscopy introduces the student into the field of special relativity and weak

More information

FEASIBILITY OF IN SITU TXRF

FEASIBILITY OF IN SITU TXRF FEASIBILITY OF IN SITU TXRF A. ngh 1, P. Goldenzweig 2, K. Baur 1, S. Brennan 1, and P. Pianetta 1 1. Stanford Synchrotron Radiation Laboratory, Stanford, CA 94309, US 2. Binghamton University, New York,

More information

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution Chikkappa Udagani Abstract: This paper presents the precise method to estimate the dependence of linear and mass

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong Introduction: X-ray fluorescence (XRF) spectrometer is an x-ray instrument used for routine, relatively

More information

Spectral analysis of K-shell X-ray emission of magnesium plasma produced by ultrashort high-intensity laser pulse irradiation

Spectral analysis of K-shell X-ray emission of magnesium plasma produced by ultrashort high-intensity laser pulse irradiation PRAMANA c Indian Academy of Sciences Vol. 82, No. 2 journal of February 2014 physics pp. 365 371 Spectral analysis of K-shell X-ray emission of magnesium plasma produced by ultrashort high-intensity laser

More information

Mass spectrometric determination of the surface compositions of ethanol water mixtures

Mass spectrometric determination of the surface compositions of ethanol water mixtures International Journal of Mass Spectrometry 212 (2001) 267 271 www.elsevier.com/locate/ijms Cluster/kinetic method Mass spectrometric determination of the surface compositions of ethanol water mixtures

More information

Generation of X-Rays in the SEM specimen

Generation of X-Rays in the SEM specimen Generation of X-Rays in the SEM specimen The electron beam generates X-ray photons in the beam-specimen interaction volume beneath the specimen surface. Some X-ray photons emerging from the specimen have

More information

Ion-beam techniques. Ion beam. Electrostatic Accelerators. Van de Graaff accelerator Pelletron Tandem Van de Graaff

Ion-beam techniques. Ion beam. Electrostatic Accelerators. Van de Graaff accelerator Pelletron Tandem Van de Graaff Ion-beam techniques RBS Target nucleus Ion beam STIM RBS: Rutherford backscattering ERD: Elastic recoil detection PIXE: Particle induced x-ray emission PIGE: Particle induced gamma emission NRA: Nuclear

More information

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES Doru Stanga*, Pierino De Felice** *National Institute of R&D for Physics and Nuclear Engineering-Horia Hulubei(IFIN-HH)

More information

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS 298 299 ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS M. Krämer 1), R. Dietsch 1), Th. Holz 1), D. Weißbach 1), G. Falkenberg 2), R. Simon 3), U. Fittschen

More information

Laser heating of noble gas droplet sprays: EUV source efficiency considerations

Laser heating of noble gas droplet sprays: EUV source efficiency considerations Laser heating of noble gas droplet sprays: EUV source efficiency considerations S.J. McNaught, J. Fan, E. Parra and H.M. Milchberg Institute for Physical Science and Technology University of Maryland College

More information

Liquid Waste Analysis

Liquid Waste Analysis Liquid Waste Analysis Garry Smith, XRF Application Specialist SciMed XRF, a division of Scientific and Medical Products Ltd About Us SciMed represent Rigaku (RESE and ART) and Seiko (SIINT) XRF ranges

More information

Valence-selective XAFS spectroscopy using EuL 4 emission

Valence-selective XAFS spectroscopy using EuL 4 emission Valence-selective XAFS spectroscopy using EuL 4 emission H Hayashi 1, N Kawamura 2, M Mizumaki 2, T Takabatake 3, H Imura 4, K Okamoto 5, and T Akai 4 1 Japan Women's University, 2-8-1, Mejirodai, Bunkyo-ku,

More information

Beta Spectrum. T β,max = kev kev 2.5 ms. Eγ = kev

Beta Spectrum. T β,max = kev kev 2.5 ms. Eγ = kev HOM, 1/14/05; DVB 014-Jan-9, 01-Dec-17, 013-Oct-16 Beta Spectrum Goal: to investigate the spectrum of β rays emitted by a 137 Cs source. The instrument used is a so-called 180 o magnetic spectrometer that

More information

Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence

Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence vnaoki Awaji (Manuscript received December 13, 21) We have developed a grazing incidence

More information

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator 88 The 1 st NPRU Academic Conference SCI-O11 Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator P. Limkitjaroenporn and W.Chewpraditkul Radiation Physics Laboratory,

More information

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440 Jazan University College of Science Physics Department جاهعة جازان كلية العل وم قسن الفيزياء Lab Manual Nuclear Physics (2) 462 Phys 8 th Level Academic Year: 1439/1440 1 Contents No. Name of the Experiment

More information

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by 18-30 kev Applied Voltages Bolortuya DAMDINSUREN and Jun KAWAI Department of Materials Science and Engineering, Kyoto University Sakyo-ku,

More information

A measurement of the air fluorescence yield

A measurement of the air fluorescence yield Nuclear Instruments and Methods in Physics Research A 372 (1996) 527-533 A measurement of the air fluorescence yield F. Kakimoto a, E.C. Loh b, M. Nagano c.*, H. Okuno d, M. Teshima c, S. Ueno a a Department

More information

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Course Details September 22 October 10 September 22 November 7 November 17 December 1 Topic Period Introduction to Spectrometric

More information

X-Ray Fluorescence and Natural History

X-Ray Fluorescence and Natural History X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,

More information

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES 187 188 ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES ABSTRACT W. T. Elam, B. Scruggs, F. Eggert, and J. A. Nicolosi EDAX, a unit of Ametek Inc., 91 McKee Drive, Mahwah,

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

PFC/JA Precision Measurements of the Wavelengths of Emission Lines of Mg-like and Na-like Kv in Alcator C Plasmas

PFC/JA Precision Measurements of the Wavelengths of Emission Lines of Mg-like and Na-like Kv in Alcator C Plasmas PFC/JA-86-43 Precision Measurements of the Wavelengths of Emission Lines of Mg-like and Na-like Kv in Alcator C Plasmas K. Kondo, J.L. Terry, J. E. Rice, E. S. Marmar Plasma Fusion Center Massachusetts

More information

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor World Journal of Nuclear Science and Technology, 2014, 4, 96-102 Published Online April 2014 in SciRes. http://www.scirp.org/journal/wjnst http://dx.doi.org/10.4236/wjnst.2014.42015 Characteristics of

More information

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13 Determination of Major, Minor and Trace Elements in a Clay Sample IAEA Laboratories, Seibersdorf November 2017 CONTENTS

More information

Energy Spectroscopy. Ex.: Fe/MgO

Energy Spectroscopy. Ex.: Fe/MgO Energy Spectroscopy Spectroscopy gives access to the electronic properties (and thus chemistry, magnetism,..) of the investigated system with thickness dependence Ex.: Fe/MgO Fe O Mg Control of the oxidation

More information

ORTEC AN34 Experiment 10 Compton Scattering

ORTEC AN34 Experiment 10 Compton Scattering EQUIPMENT NEEDED FROM ORTEC 113 Preamplifier (2 ea.) TRUMP-PCI-2K MCA System including suitable PC operating Windows 98/2000/XP (other ORTEC MCAs may be used) 266 Photomultiplier Tube Base (2 ea.) 4001A/4002D

More information

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer Elemental analysis by X-ray f luorescence Sequential benchtop WDXRF spectrometer Elemental analysis is one of the most important fundamental measurements made for industrial quality control and research

More information

Mass of the electron m 0

Mass of the electron m 0 Mass of the electron m 0 1 Objective To determine the rest mass of the electron, m e, via γ-ray interactions (mainly Compton scattering and photoeffect) in a NaI scintillation detector. Based on the enclosed

More information

Measurement of linear and mass attenuation coefficient of alcohol soluble compound for gamma rays at energy MeV

Measurement of linear and mass attenuation coefficient of alcohol soluble compound for gamma rays at energy MeV Available online at www.scholarsresearchlibrary.com Archives of Applied Science Research, 2012, 4 (4):1748-1752 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-508X CODEN (USA) AASRC9 Measurement

More information

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE 218 Chris M. Sparks 1, Elizabeth P. Hastings 2, George J. Havrilla 2, and Michael Beckstead 2 1. ATDF,

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

Micro-XRF excitation in an SEM

Micro-XRF excitation in an SEM X-RAY SPECTROMETRY X-Ray Spectrom. 2007; 36: 254 259 Published online 8 May 2007 in Wiley InterScience (www.interscience.wiley.com).974 Micro-XRF excitation in an SEM M. Haschke, 1 F. Eggert 2 andw.t.elam

More information

RICE UNIVERSITY GAMMA-RAY SPECTRA FROM BE 3 AND N 1^ WITH C 12. Hsin-Min Kuan A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE

RICE UNIVERSITY GAMMA-RAY SPECTRA FROM BE 3 AND N 1^ WITH C 12. Hsin-Min Kuan A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE RICE UNIVERSITY GAMMA-RAY SPECTRA FROM BE 3 REACTIONS WITH C 12 AND N 1^ by Hsin-Min Kuan A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF MASTER OF ARTS Houston, Texas April

More information

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS NUCLEAR PHYSICS ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS A. DUMITRESCU 1, GH. CǍTA-DANIL 1, 2 1 Physics Department, University Politehnica of Bucharest, Splaiul Independenţei

More information

Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L.

Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L. Efficient Algorithms for Pulse Parameter Estimation, Pulse Peak Localization And Pileup Reduction in Gamma Ray Spectroscopy M.W.Raad 1, L. Cheded 2 1 Computer Engineering Department, 2 Systems Engineering

More information