Portable type TXRF analyzer: Ourstex 200TX
|
|
- Hope Holmes
- 5 years ago
- Views:
Transcription
1 Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray Fluorescence Spectrometer Equipped with a Silicon Drift Detector We have adapted a silicon drift detector (SDD) to a portable total reflection X-ray fluorescence (TXRF) spectrometer using polychromatic X-rays from an X-ray source of single-digit watt. Analytical performance was greatly improved in PB ratio and the sensitivity for lighter elements when compared with the former spectrometer equipped with a Si-PIN detector. In this article improved sensitivity of the new spectrometer is described, and additionally the evaluation of the current wave-guide, a newly developed sample preparation device for precise quantitative evaluation and an application to the analysis of suspended particles are reported. Fig. 1 Photograph of portable TXRF spectrometer, OURSTEX 200TX Portable type TXRF analyzer: Ourstex 200TX This analyzer is a compact type, only 8kg weight. It is easy to be movable at on-site measurement, requires AC100V power. Fig. 2 shows its structural photo. For the X-ray tube, transparent compact X-ray tube is used with 50kV and 200µA of maximum voltage and tube current, respectively. Tungsten is used as the target of the X-ray tube. Electronic cooling type Silicon Drift Detector (SDD, light reception area; 7mm 2, Be thickness; 8µm) is used as the detector. The SDD is composed of united structure of Si-PIN with FET, and ring-shaped electrode gradient on the surface. It can be effectively achieved to drift the electron to the centered anode. By this reason, decrease of thermal noise, higher improvement of electron collecting effect and decrease of leak current etc. are attainable. This improvement contributes to obtain higher energy resolution ability, and PB ratio becomes bigger. Hence, the measurement of high calculating ratio is achievable without decreasing the resolution ability. In general, the wider the light reception area is, the lower the resolution ability is in case of the Si-PIN detector, while in case of the SDD, it has a great capability that it is almost no lowering of energy resolution ability. On the contrary, it has a certain demerit that higher energy X-ray may be easy to transparent because the crystal thickness is thin as in 450µm, so the detecting efficiency is not too high at the high energy area. X-ray wave-guide is used for achieving paralleled of incident X-ray illuminating from the X-ray tube. X-ray wave-guide is a photo-element that enables to implement paralleled of incident X-ray by using 2 pieces of silicon wafers as shown in Fig. 3 (a). As a part of incident X-ray is totally reflected by using the mirror-face silicon wafer, stronger X-ray is obtained than using usual slit. The interval between the silicon wafers is designed and made as 10mm horizontal width and 10µm vertical width 3). Fig. 3 (b) shows the outlook of the X-ray wave-guide. 1 / 5 Fig. 2 Photograph of portable TXRF system Fig. 3 (a) Schematic drawing of the wave-guide device (b) Photograph of the wave-guide device
2 Produced wave-guide has a possibility of variation of the slit width by affecting the manufacturing fabrication accuracy and the edge that was cut the silicon wafer. Therefore stainless steel cutter is set at the outlet of the X-ray wave-guide, and by knife-edge scanning method, the evaluation of the slit width of the X-ray wave-guide has implemented. Evaluation of the slit width was determined by obtained data that were differentiated and the full width half maximum (FWHM) value that was calculated by Gaussian fitting. The result is shown in Fig. 4. Calculated slit width was estimated as 16.5µm, which was close to 10µm that was assumed at the initial designing time. Quartz optical flat was used in the sample plate (accuracy of reflection wave surface: λ/20). Fig. 4 A profile of the measured edge-scan, and the evaluated beam profile of the wave-guide device Incident angle was adjusted as setting at ca that is most suitable for transition metals, by the Goniometer that is installed at under the sample plate 6). PB (Peak-Background) ratio improvement by adopting SDD SDD was used in this analyzer. Energy resolution ability of the SDD is 132eV in the FWHM of Mn-K α ray. (Formerly used Si-PIN detector s resolution ability is 180eV in the FWHM of the same ray) Comparison of spectrum between the Si-PIN detector and the SDD is shown in Fig. 5. Spectrum is standardized by the peak intensity of W-L α ray that is the excitation source. Measuring sample: Sc, Cr, Co, As and Sr are in 0.5ppm aqueous solution respectively. The solution is dropped on the optical flat as 2mL and it is dried (1ng). As a measuring result, energy resolution ability as well as the PB ratio in the SDD was more improved than those in the Si-PIN detector. Fig. 6 illustrates respective PB ratio. In Co, PB ratio in the Si-PIN detector is 0.921, while on the SDD is detected, almost 2 times higher improved. In the Si-PIN detector, it was impossible to separate the peaks between Ar-K α and Ar-K β ray, and no detect for very little peaks; The SDD enabled to detect the separation of the peaks between Ar-K α and Ar-K β ray, and detect very little peaks effectively. Lower limit of detection (LLD) values estimated from obtained data for Sc, Cr, Co, As and Sr in 1800 seconds measurement are, 29pg, 20pg, 14pg, 51pg and 192pg, respectively. The following theoretical formula was used for the calculation of the LLD value 7). Fig. 5 Total reflection X-ray fluorescence spectra of a dried droplet containing 1ng of Sc, Cr, Co, As and Sr. The data acquisition time was 1800s, and the detector used were a) Si-PIN and b) SDD. : Peak from collimator material of SDD where W: content (ng), t: measuring time (sec), I NET : Net intensity of the peak (cps), I BG : background intensity (cps) Fig. 6 A comparison of PB ratios for various elements obtained with the Si-PIN detector and with the SDD detector 2 / 5
3 The LLD value of Co obtained this time was the same value as in previous reported one 6) by Kunimura et al. For this reason, as the reception light surface area is 7mm 2 in the SDD, X-ray intensity was lower than the Si-PIN detector, which has the reception light surface area of 13mm 2. It depends on the droplet size though, significant detection sensitivity may be obtained by using bigger caliber of the SDD. Quantitative improvement by using the droplet supporting device On the analysis of the total reflection X-ray fluorescence spectrometry, sample preparation is the most important factor in evaluating the analysis quantitatively. In this analyzer, sample solution is dropped on the optical flat and its dries residual is measured. It is quite important that the sample solution should be dropped on the center position of the optical flat precisely, in order not to lower the quantitative accuracy by changing the X-ray intensity significantly by the relationship with the taking-un angle of the detector. We fabricated the droplet supporting device such as in Fig. 7. The device has a small hole that enables to insert the point of the micro-pipette tip and it can insert always at a constant depth. Hence, it enables to drop the solution always at the center part of the optical flat, and the quantitative accuracy improves much higher. The material used in the device is acryl resin with consideration of the contamination, and the device is reusable by washing treatment. To verify the effect of using the droplet supporting device, two conditions were compared: Measuring the change of the X-ray intensity of Cr-K α in Cr standard solution prepared as in 10ppm and drops 5mL: 1) with the droplet supporting device 2) without the device The results are illustrated in Table 1. Compared to the case of without the device, the change of the X-ray intensity shows smaller in the case of using the device. Fig. 7 (a) Schematic drawing of the sample preparation support device. (b) Photograph of the sample preparation support device. In order to evaluate quantitative verification of the analyzer, drawing the calibration curves of Co and Cr was executed. In Fig. 8, the calibration curves obtained from the experiment are shown. Sample solution preparation: Standard solutions for atomic absorption spectrometry with Cr and Co were diluted and adjusted as in 500ppb, 100ppb, 50ppb, 25ppb and 10ppb, respectively. Also, constant amount of Sc solution as the internal standard is added and 10µL of the solution is dropped on the optical flat. The sample is dried on it, and is measured. As a result, good linearity relation is obtained both for Cr and Co, and it can be confirmed that sufficient quantitativity is approved even in the lower concentration range 3 / 5 Fig. 8 Calibration curve of chromium and cobalt
4 Sensitivity improvement for light elements In order to confirm analytical sensitivity of the analyzer in the light elements field, P analysis in the solution was implemented. With the formerly used Si-PIN detector, it has lower energy resolution ability that the peak of P-K α ray is hidden in the tailing part of the peak of Si-K α ray that is due to the optical flat. So trace amount of P analysis was not achievable. Now the experiment for the analyzer with the SDD is implemented whether trace amount of P analysis is achievable or not. Standard P solution is diluted as in 1ppm, 0.5ppm and 0.1ppm, respectively. As the measuring sample, 5mL of it is dropped on the optical flat, and after drying measurement is executed. Absolute amount of sample is 5ng, 2.5ng and 0.5ng, respectively. Measured P spectra are shown in Fig. 9. It can be confirmed that even in 0.5ng concentration range, sufficient detection is achievable. The LLD value estimated by 0.5ng peak intensity and the background intensity with the formula (1) becomes 286pg in 1800 second measurement. It can be confirmed that high sensitive measurement is attainable even in the light element field in air atmospheric environment. Fig. 9 Total reflection X-ray fluorescence spectra of phosphorus This achievement could be concluded as: 1) Energy resolution ability improves much higher by the SDD adoption 2) On the structure of total reflection X-ray fluorescence spectrometer, the distance between the sample and detector is extremely short; the detector is located on upper of the sample; and it is no affection of the absorption by the polymer film, etc. that is used as sample holding in usual fluorescence X-ray spectrometer. Application field As the examples of application fields measured by a portable total reflection X-ray fluorescence spectrometer, there are bunch of reports as in measured soil eluding water 8), leachate from tea or toys, etc 9). In this report, the possibility for element analysis in air particulate matter is examined. Optical flat is on the Petri dish and it is left at the sampling place 1 hour remains. Accumulating particles on the optical flat is directly measured. Sampling place is along national route in Neyagawa City, Osaka Pref., that is relatively high traffic-laden street. Measured result spectra are shown in Fig. 10. Compared to the blank, the elements of Al, S, K, Ca, Ti, Cr, Mn, Fe, Zn etc. were detected. Most likely, as the sampling place is high traffic-laden street, elements from automobile exhaust gas and frictional metal by the break pad etc. may be detected presumably. Fig. 10 Total reflection X-ray fluorescence spectra of suspended particulate matter. (Collected place of sample: Neyagawa City, Osaka, Japan) By improving the sampling method, this analyzer might be utilized at the application of environmental monitoring of airborne particulate matter. 4 / 5
5 Conclusion By loading the SDD as the detector of a portable total reflection X-ray fluorescence spectrometer utilizing non-color X-ray generating from small X-ray tube in a few watt, energy resolution ability is improved, and compared to the former used Si-PIN detector, PB ratio improves about 2 times higher. Moreover, analysis sensitivity of P in light element field is improved; it is obtained 286ng as the lower limit detection. Stable sample preparation is achievable by utilizing the droplet sample supporting device; and it enabled to improve quantitative accuracy drastically. This analyzer is compact and light-weighted, so it may be possible to implement measurements in case of having difficulty of sample storage and/or transferring the sample, such as the air particulate matter at site (ex. at Antarctic), as well as by measuring the hair of the cattle at site, cattle s health condition may be determined as routinely. References 1) Y. Yoneda, T. Horiuchi: Rev. Sci. Instrum., 42, 1069 (1971) 2) A. Iida, Y. Gohshi: Jpn. J. Appl. Phys., 23, 1543 (1984) 3) S. Kunimura, J. Kawai: Anal. Chem., 79, 2593 (2007) 4) S. Kunimura, J. Kawai: Chemistry and Industry, 61, 1050 (2008) 5) ditto: Analytical Chemistry, 58, 1041 (2009) 6) ditto: Adv. X-Ray. Chem. Anal., Japan, 41, 29 (2010) 7) I. Nakai: Fluorescence X-ray analysis (2005), Asakura Publishing Co. 8) S. Kunimura, et al.: Adv. X-Ray. Chem. Anal., Japan, 38, 367 (2007) 9) S. Kunimura, et al.: Analytical Chemistry, 57, 135 (2008) 5 / 5
IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE
Copyright JCPDS-International Centre for Diffraction Data 2012 ISSN 1097-0002 281 IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Susumu Imashuku 1, Deh Ping Tee 1, Yasukazu
More informationS2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF
S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?
More informationWhy do I chose this topic? By: Medani Sangroula 12/6/2013 1
Characterization of atmospheric aerosols using Synchrotron radiation total reflection X-ray fluorescence(sr-txrf) and Fe K-edge total reflection X-ray fluorescence(txrf)-x-ray absorption near edge structure
More informationApplication of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater
PRAMANA c Indian Academy of Sciences Vol. 76, No. 2 journal of February 2011 physics pp. 361 366 Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater
More informationAEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD
Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto
More informationMT Electron microscopy Scanning electron microscopy and electron probe microanalysis
MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron
More informationMS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF
2016 Fall Semester MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF Byungha Shin Dept. of MSE, KAIST 1 Course Information Syllabus 1. Overview of various characterization techniques (1 lecture)
More informationTRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER
Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 236 ABSTRACT TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Hagen Stosnach Röntec GmbH,
More informationSilicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010
Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis Patrick Camus, PhD Applications Scientist March 18, 2010 EDS Detector Requirements Detect whole energy range of x-rays 50 ev
More informationCALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR
CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR A. C. Neiva 1, J. N. Dron 1, L. B. Lopes 1 1 Escola Politécnica da Universidade de São Paulo
More informationPraktikum zur. Materialanalytik
Praktikum zur Materialanalytik Energy Dispersive X-ray Spectroscopy B513 Stand: 19.10.2016 Contents 1 Introduction... 2 2. Fundamental Physics and Notation... 3 2.1. Alignments of the microscope... 3 2.2.
More informationOBSERVATION OF SURFACE DISTRIBUTION OF PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY DURING D 2 GAS PERMEATION THROUGH PD COMPLEXES
Iwamura, Y., et al. Observation Of Surface Distribution Of Products By X-Ray Fluorescence Spectrometry During D2 Gas Permeation Through Pd Complexes. in The 12th International Conference on Condensed Matter
More informationDevelopment of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis
Bull. Mater. Sci., Vol. 25, No. 5, October 2002, pp. 435 441. Indian Academy of Sciences. Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis M K TIWARI*,
More informationFEASIBILITY OF IN SITU TXRF
FEASIBILITY OF IN SITU TXRF A. ngh 1, P. Goldenzweig 2, K. Baur 1, S. Brennan 1, and P. Pianetta 1 1. Stanford Synchrotron Radiation Laboratory, Stanford, CA 94309, US 2. Binghamton University, New York,
More informationS8 TIGER Series 2 for ASTM D 6443
Lab Report XRF 139 S8 TIGER Series 2 for ASTM D 6443 Standard Test Method for Determination of Ca, Cl, Cu, Mg, P, S and Zn in Unused Lubricating Oils and Additives Introduction Lubricating oils are generally
More informationFUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON
More informationAnalysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy
Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.
More informationDirect Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ
Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction
More informationBENEFITS OF IMPROVED RESOLUTION FOR EDXRF
135 Abstract BENEFITS OF IMPROVED RESOLUTION FOR EDXRF R. Redus 1, T. Pantazis 1, J. Pantazis 1, A. Huber 1, B. Cross 2 1 Amptek, Inc., 14 DeAngelo Dr, Bedford MA 01730, 781-275-2242, www.amptek.com 2
More informationChemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy
Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental
More informationElectron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages
Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by 18-30 kev Applied Voltages Bolortuya DAMDINSUREN and Jun KAWAI Department of Materials Science and Engineering, Kyoto University Sakyo-ku,
More informationDetermination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity
More informationDEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT
Copyright JCPDS - International Centre for Diffraction Data 003, Advances in X-ray Analysis, Volume 46. 6 DEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT
More informationIn-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*
In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer* Anthony Thomas 1, Joachim Heckel 1, and Dirk Wissmann 1 Search and Discovery Article #41836 (2016)
More informationINFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS
111 INFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS C. Horntrich 1, F. Meirer 1, C. Streli 1, P. Kregsamer 1, G. Pepponi 2, N. Zoeger 1, P. Wobrauschek 1 1 Vienna University
More informationChemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:
More informationSEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM
THE UNIVERSITY Chemical Analysis in the SEM Ian Jones Centre for Electron Microscopy OF BIRMINGHAM Elastic and Inelastic scattering Electron interacts with one of the orbital electrons Secondary electrons,
More informationRepeatability of Spectral Intensity Using an Auger Electron Spectroscopy Instrument Equipped with a Cylindrical Mirror Analyzer
A. Kurokawa et al. Repeatability of Spectral Intensity Using an Auger lectron Spectroscopy Instrument quipped with a Cylindrical Mirror Analyzer Paper Repeatability of Spectral Intensity Using an Auger
More informationS2 PICOFOX. Innovation with Integrity. Benchtop TXRF Spectrometer TXRF
S2 PICOFOX Benchtop TXRF Spectrometer Innovation with Integrity TXRF S2 PICOFOX - Fast Trace Element Analysis with XRF The S2 PICOFOX is the world s first portable benchtop spectrometer for fast quantitative
More informationAuger Electron Spectroscopy (AES)
1. Introduction Auger Electron Spectroscopy (AES) Silvia Natividad, Gabriel Gonzalez and Arena Holguin Auger Electron Spectroscopy (Auger spectroscopy or AES) was developed in the late 1960's, deriving
More informationOptics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P
Optics Spectrometer Prism spectrometer LD Physics Leaflets P5.7.1.1 Measuring the line spectra of inert gases and metal vapors using a prism spectrometer Objects of the experiment Adjusting the prism spectrometer.
More informationOverview of X-Ray Fluorescence Analysis
Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic
More informationPHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy
PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy The very basic theory of XPS XPS theroy Surface Analysis Ultra High Vacuum (UHV) XPS Theory XPS = X-ray Photo-electron Spectroscopy X-ray
More informationX-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu
X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging
More informationSecondary Ion Mass Spectrometry (SIMS) Thomas Sky
1 Secondary Ion Mass Spectrometry (SIMS) Thomas Sky Depth (µm) 2 Characterization of solar cells 0,0 1E16 1E17 1E18 1E19 1E20 0,2 0,4 0,6 0,8 1,0 1,2 P Concentration (cm -3 ) Characterization Optimization
More informationElemental analysis by X-ray f luorescence. High performance EDXRF elemental analyzer
Elemental analysis by X-ray f luorescence High performance EDXRF elemental analyzer Simple elemental analysis 1. Prepare samples No digestion; minimal sample preparation. Direct non-destructive measurement
More informationDevelopment of a new MeV gamma-ray camera
Development of a new MeV gamma-ray camera ICEPP Symposium February 16, 2004 Hakuba, Nagano, Japan Kyoto University Atsushi Takeda H. Kubo, K. Miuchi, T. Nagayoshi, Y. Okada, R. Orito, A. Takada, T. Tanimori,
More informationSecondary ion mass spectrometry (SIMS)
Secondary ion mass spectrometry (SIMS) Lasse Vines 1 Secondary ion mass spectrometry O Zn 10000 O 2 Counts/sec 1000 100 Li Na K Cr ZnO 10 ZnO 2 1 0 20 40 60 80 100 Mass (AMU) 10 21 10 20 Si 07 Ge 0.3 Atomic
More informationApplication of the GD-Profiler 2 to the PV domain
Application of the GD-Profiler 2 to the PV domain GD Profiler 2 RF GDOES permits to follow the distribution of the elements as function of depth. This is an ultra fast characterisation technique capable
More informationCHEM-E5225 :Electron Microscopy X-Ray Spectrometry
CHEM-E5225 :Electron Microscopy X-Ray Spectrometry 2016.11 Yanling Ge Outline X-ray Spectrometry X-ray Spectra and Images Qualitative and Quantitative X-ray Analysis and Imaging Discussion of homework
More informationFINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS
FINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS Brand N. W 1 and Brand CJ 1 1 Portable XRF Services Pty Ltd. Suite 1, 5 Colin Street, West Perth,
More informationStructural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence
Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence vnaoki Awaji (Manuscript received December 13, 21) We have developed a grazing incidence
More informationXRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:
1 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: XRF spectroscopy 1 exam question. Notes on: www.nuigalway.ie/nanoscale/3rdspectroscopy.html XRF books: Analytical Chemistry,
More informationDevelopment of a Compact XAFS Measurement Chamber under Atmospheric Pressure in the Soft X-ray Region
Development of a Compact XAFS Measurement Chamber under Atmospheric Pressure in the Soft X-ray Region Koji Nakanishi, Toshiaki Ohta Abstract We have developed a compact experimental set-up for X-ray absorption
More information2D XRD Imaging by Projection-Type X-Ray Microscope
0/25 National Institute for Materials Science,Tsukuba, Japan 2D XRD Imaging by Projection-Type X-Ray Microscope 1. Introduction - What s projection-type X-ray microscope? 2. Examples for inhomogeneous/patterned
More informationHow Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source
How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source What Makes Up The Canadian Light Source? 4. Storage Ring 5. Synchrotron Light 6. Beamline 1. Electron Gun 2. Linear Accelerator
More informationSecondary ion mass spectrometry (SIMS)
Secondary ion mass spectrometry (SIMS) ELEC-L3211 Postgraduate Course in Micro and Nanosciences Department of Micro and Nanosciences Personal motivation and experience on SIMS Offers the possibility to
More informationX-ray fluorescence analysis - calibration of the X-ray energy detector
X-ray fluorescence analysis - TEP Related topics Bremsstrahlung, characteristic X-radiation, energy levels, fluorescence radiation, conduction processes in semiconductors, doping of semiconductors, pin-diodes,
More informationReal-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System
Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Presented at the BevTech Conference, Albuquerque, NM 2018 by Charles M. Phillips Ph.D., Max Analytical Technologies Mark Taylor, Vice
More informationThe Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements
The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements Uwe Scheithauer, 82008 Unterhaching, Germany E-Mail: scht.uhg@googlemail.com Internet: orcid.org/0000-0002-4776-0678;
More informationSemiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH
Semiconductor X-Ray Detectors Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Part A Principles of Semiconductor Detectors 1. Basic Principles 2. Typical Applications 3. Planar Technology 4. Read-out
More informationDefining quality standards for the analysis of solid samples
Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples
More informationFEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada.
FEATURE ARTICLE The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process Takaaki Yada Cleaning Cleaning Cleaning Cleaning Wet process Wet Etching process SPM Organic
More informationOES - Optical Emission Spectrometer 2000
OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the
More informationDevelopment of Laser Thickness Gauge in Steel Plate Shearing Line
JFE TECHNICAL REPORT No. 21 (Mar. 2016) Development of Laser Thickness Gauge in Steel Plate Shearing Line TEZUKA Koichi *1 Abstract: JFE Steel has developed a laser thickness gauge for the purpose of guaranteeing
More informationElectrochemical Cell for in-situ XAFS Measurements
Electrochemical Cell for in-situ XAFS Measurements Ryota Miyahara, Kazuhiro Hayashi, Misaki Katayama, and Yasuhiro Inada Applied Chemistry Course, Graduate School of Life Sciences, Ritsumeikan University,
More informationChemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence
Chemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence Shinji OZAKI, Matsushita Technoresearch Inc. ozaki.s@jp.panasonic.com The chemical states of a SiO 2 /Si govern the conductivity
More informationExtrel Application Note
Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical
More informationAdvances in Field-Portable XRF
Advances in Field-Portable XRF Volker Thomsen and Debbie Schatzlein Field-portable x-ray fluorescence (XRF) allows us to take the laboratory to the sample. The latest generation of such handheld x-ray
More informationAltitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory
Altitude influence of elemental distribution in grass from Rila mountain Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory I. Introduction The application of modern instrumental
More informationA New Cross-Shaped Graphite Furnace with Ballast Body for Reduction of Interferences in Atomic Absorption Spectrometry
http://www.e-journals.net ISSN: 0973-4945; CODEN ECJHAO E- Chemistry 2010, 7(S1), S127-S130 A New Cross-Shaped Graphite Furnace with Ballast Body for Reduction of Interferences in Atomic Absorption Spectrometry
More informationX-Ray Fluorescence and Natural History
X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,
More informationIndustrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application
FFeature Article Article Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application Yoko NAKAI HORIBA s CS-Series chemical concentration monitors that use ultraviolet
More informationCHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE
CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE 218 Chris M. Sparks 1, Elizabeth P. Hastings 2, George J. Havrilla 2, and Michael Beckstead 2 1. ATDF,
More informationPROCESS MONITORING OF PLASMA ELECTROLYTIC OXIDATION J.-W. Liaw, C.-C. Hsiao, Clinton Fong, Y.-L. Tsai, S.-C. Chung, Oleg Demin Materials Research
PROCESS MONITORING OF PLASMA ELECTROLYTIC OXIDATION J.-W. Liaw, C.-C. Hsiao, Clinton Fong, Y.-L. Tsai, S.-C. Chung, Oleg Demin Materials Research Laboratories, Industrial Technology Research Institute,
More informationXylan Adsorption on Cellulose Model Films & Influence on Bond Strength
Xylan Adsorption on Cellulose Model Films & Influence on Bond Strength Experiments Conducted in Context of My Master Thesis at the Solid State Physics Institute of TU Graz Siegfried Zöhrer szoehrer@sbox.tugraz.at
More informationICP-Mass Spectrometer
ICP-Mass Spectrometer New Mass Spectrometers The main issue: sequential vs. simultaneous Scanning, peak hopping are sequential Like viewing a photo through a peephole One pixel at a time Other parts of
More informationInternational Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN
316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science
More informationPRINCIPLE OF ICP- AES
INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively
More informationX-ray Fluorescence Imaging Following Synchrotron Beam Excitation
Conference on Applied Digital Imaging Techniques for Understanding the Palimpsest X-ray Fluorescence Imaging Following Synchrotron Beam Excitation Uwe Bergmann Stanford Synchrotron Radiation Laboratory
More informationDesign and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications
Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,
More informationDiffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves
Diffraction & Interference Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves Diffraction in Nature What is Interference? The resultant
More informationGround Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2
Ground Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2 Beverly LaMarr a, Mark Bautz a, Steve Kissel a, Gregory Prigozhin a Kiyoshi Hayashida b Takeshi
More information1+2 on GHD (5 µl) Volume 1+2 (µl) 1 on GHD 1+2 on GHD
1+2 on GHD (20 µl) 1+2 on GHD (15 µl) 1+2 on GHD (10 µl) 1+2 on GHD (5 µl) Volume 1+2 (µl) 1 on GHD 1+2 on GHD Supplementary Figure 1 UV-Vis measurements a. UV-Vis spectroscopy of drop-casted volume of
More informationSample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry
2005 The Japan Society for Analytical Chemistry 143 Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry M. K. TIWARI, A. K. SINGH, and K. J. S. SAWHNEY Synchrotron
More informationTRITIUM PRODUCTION IN PALLADIUM DEUTERIDE/HYDRIDE IN EVACUATED CHAMBER
Yamada, H., et al. Tritium Production in Palladium Deuteride/Hydride in Evacuated Chamber. in 8th International Conference on Cold Fusion. 2000. Lerici (La Spezia), Italy: Italian Physical Society, Bologna,
More informationPeter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.
783 SCOPE AND LIMITATIONS XRF ANALYSIS FOR SEMI-QUANTITATIVE Introduction Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. Historically x-ray fluorescence spectrometry has
More informationUNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.
UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering
More informationA COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 320 A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS Hiroyoshi SOEJIMA and
More informationDetection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR
Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Ensuring ultimate cleanliness for maximum adhesion Application Note Author John Seelenbinder Agilent Technologies,
More informationReference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )
September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum
More informationShorten your Time-to-Result
Shorten your Time-to-Result Effective quality and process control requires the shortest time-to-result possible. This is the time needed from sampling to the final quantitative result. Any advantage results
More informationAS 101: Day Lab #2 Summer Spectroscopy
Spectroscopy Goals To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are related To see spectral lines from different elements in emission and
More informationFINE PATTERN ETCHING OF SILICON USING SR-ASSISTED IONIZATION OF CF4 GAS
Technical Paper Journal of Photopolymer Science and Technology Volume 6, Number 4(1993) 617-624 1993TAPJ FINE PATTERN ETCHING OF SILICON USING SR-ASSISTED IONIZATION OF CF4 GAS YASUO NARA, YosHIHto SUGITA,
More informationPulse Height Analysis System (PHA) designed for W7-X Presented by Monika KUBKOWSKA
Pulse Height Analysis System (PHA) designed for W7-X Presented by Monika KUBKOWSKA This scientific work has been partly supported by Polish Ministry of Science and Higher Education within the framework
More informationIodine-Mediated Chemical Vapor Deposition Growth of Metastable Transition Metal
Supporting Information Iodine-Mediated Chemical Vapor Deposition Growth of Metastable Transition Metal Dichalcogenides Qiqi Zhang,, Yao Xiao, #, Tao Zhang,, Zheng Weng, Mengqi Zeng, Shuanglin Yue, ± Rafael
More informationSupplementary Information for. Silver Nanoparticles Embedded Anti-microbial Paints Based on Vegetable Oil
Supplementary Information for Silver Nanoparticles Embedded Anti-microbial Paints Based on Vegetable Oil Ashavani Kumar #, Praveen Kumar Vemula #, Pulickel M. Ajayan, George John * Department of Chemistry,
More informationA Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II
J. Mass Spectrom. Soc. Jpn. Vol. 51, No. 2, 23 REGULAR PAPER A Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II Daisuke O@JBJG6, a) Michisato TDND96, a) Morio IH=>=6G6, a) and Itsuo K6I6@JH:
More informationA DIVISION OF ULVAC-PHI
A DIVISION OF ULVAC-PHI X-ray photoelectron spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and research applications. XPS provides
More informationMulti-element process analyzer
Multi-element process analyzer Elemental analysis by X-ray fluorescence Compact multi-element process analyzer for liquid streams or f ixed position web applications Featuring advanced third generation
More informationIn-Situ FTIR Spectroscopy and Metrology of a Tungsten CVD Process
In-Situ FTIR Spectroscopy and Metrology of a Tungsten CVD Process A. Singhal, L. Henn-Lecordier and J. N. Kidder Jr. University of Maryland, College Park, MD C.A. Gogol, J.F. Kushneir Inficon, Inc. East
More informationDiffractometer. Geometry Optics Detectors
Diffractometer Geometry Optics Detectors Diffractometers Debye Scherrer Camera V.K. Pecharsky and P.Y. Zavalij Fundamentals of Powder Diffraction and Structural Characterization of Materials. Diffractometers
More informationSupporting Information
Electronic Supplementary Material (ESI) for Journal of Materials Chemistry A. This journal is The Royal Society of Chemistry 2018 Supporting Information A minimal non-radiative recombination loss for efficient
More informationIntroduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960
Introduction to X-ray Photoelectron Spectroscopy (XPS) X-ray Photoelectron Spectroscopy (XPS), also known as Electron Spectroscopy for Chemical Analysis (ESCA) is a widely used technique to investigate
More informationARC SPECTRUM OF IRON /COPPER / BRASS
ARC PECTRUM OF IRON /COPPER / BRA Aim : To determine the wavelength of prominent lines in the emission spectrum of iron/ copper/ brass. Apparatus : Constant deviation spectrometer, dc voltage source, metal
More informationphoto-mineralization of 2-propanol under visible light irradiation
Electronic Supplementary Information for WO 3 modified titanate network film: highly efficient photo-mineralization of 2-propanol under visible light irradiation Experimental Preparation of STN, and WO
More informationThe scanning microbeam PIXE analysis facility at NIRS
Nuclear Instruments and Methods in Physics Research B 210 (2003) 42 47 www.elsevier.com/locate/nimb The scanning microbeam PIXE analysis facility at NIRS Hitoshi Imaseki a, *, Masae Yukawa a, Frank Watt
More informationSurface Analysis. Dr. Lynn Fuller Dr. Fuller s Webpage:
ROCHESTER INSTITUTE OF TECHNOLOGY MICROELECTRONIC ENGINEERING Surface Analysis Dr. Lynn Fuller Dr. Fuller s Webpage: http://people.rit.edu/lffeee 82 Lomb Memorial Drive Rochester, NY 14623-5604 Tel (585)
More informationAuger Electron Spectroscopy (AES) Prof. Paul K. Chu
Auger Electron Spectroscopy (AES) Prof. Paul K. Chu Auger Electron Spectroscopy Introduction Principles Instrumentation Qualitative analysis Quantitative analysis Depth profiling Mapping Examples The Auger
More informationMetal-organic frameworks (MOFs) as precursors towards TiO x /C. composites for photodegradation of organic dye
Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2014 Supplementary Information Metal-organic frameworks (MOFs) as precursors towards TiO x /C composites
More information