Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co
|
|
- Darcy Bond
- 5 years ago
- Views:
Transcription
1 Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Pirchio Rosana a*, Lindner Carlos a, Molina Laura a and Vallejos Matías a. a Comisión Nacional de Energía Atómica, CAE, Pro. Luis González y Aragón N 15, B180AYA, Buenos Aires, Argentina. Abstract. Radioprotection Laboratory belongs to Dosimetry Reference Regional Centre of Atomic Energy National Commission, C.A.E. This laboratory offers the service of calibrations for radioprotection instruments as Geiger Muller detector, ionisation chamber, probe, proportional counters, electronic personal dosimeters and others used in nuclear medicine, radiotherapy centres, nuclear power plants, industry and in other applications of ionising radiation. Also it offers the service of calibrated irradiations. An gamma Irradiator and a Stabilipam 300 X ray are the equipments of the Radioprotection Laboratory used for calibrations. Hopewell Designs Irradiator was installed to improve the quality of services in 005. The irradiator has a 137 Cs source of 10 Curies and a 60 Co source of 1 Curie, approximately. Theoretical and experimental studies were done to analyse symmetry, flatness, penumbra and secondary radiation of photon beams. For symmetry, flatness and penumbra X-OMAT-V films and termoluminiscent dosimeters (TLDs) were used. Films were placed to 0.7 and 1 meter from collimator exit receiving 10, 40, 80, 100 and 10mGy. TLDs were placed to 1 and 1.80 meters from collimator on a surface higher than 137 Cs beam cross section. Also studies were done to a distance of 1m and 1.80 from exit collimator using LiF powder in capsules. Irradiations were done without attenuators and with a collimator aperture of 8º. Results were compared to those obtained with Monte Carlo simulation (MCNP5c code). For secondary radiation calculation three methods were employed: Shadow-shield, Multiple distance and Monte Carlo simulation. Finally, from theoretical and experimental studies could conclude that the secondary radiation resulted to be lower than 3.5%, total radiation, symmetry and flatness were higher than 90% and penumbra was lower than 13 mm. Those results agree to recommendations ISO 4037 Standard. KEYWORDS: Radioprotection; irradiator; dosimetry; flatness, Monte Carlo; dispersion. 1. Introduction Instruments and personal dosimeters are able to detect and measure directly ionizing or indirectly ionising radiation and to provide information to control radiological hazards. Radiation survey meters should be calibrated to measure different dosimetric magnitudes such as air Kerma, ambient equivalent dose, personal equivalent dose and exposition in a confident way. Radiation protection instruments, such as dosemeters and rate dosemeters are used in nuclear medicine, brachytherapy, nuclear installations and industries. Radiation dose measurements must have international traceability and for that reason the International Atomic Energy Agency (IAEA) and the World Health Organization (WHO) created a Network of Secondary Standards Dosimetry Laboratories (SSDLs) in SSDLs are the link in the traceability chain for radiation metrology users. They provide calibrations at specific radiation qualities appropriate for the use of radiation measuring instruments. The calibration of instruments by the SSDLs mean that these instruments are traceable to the Primary Standards Dosimetry Laboratories (PSDLs) and the Bureau International des Poids et Mesures (BIPM). [1] In Argentina irradiations and calibrations in cobalt and cesium beam are done at Radioprotection Laboratory of Dosimetry Reference Regional Center of Atomic Energy National Commission some years ago. This is a Secondary Standard Dosimetry Laboratory. * Presenting author, pirchio@cae.cnea.gov.ar 1
2 Calibration means the determination of the instrument response in a specified radiation field delivering a known dose (or other magnitude) at the point reference of the instrument. The procedure used to calibrate radiation protection instruments in a radiation field is by substitution [-3]. This method consists in two steps; in the first the reference dosimeter is placed at the calibration point to determine the reference output rate of the beam. In the second step it is replaced by the dosimeter to be calibrated, and reading are taken. The goal to acquire the new gamma irradiator Hopewell Designs, Inc. Model G with 137 cesium and 60 cobalt source and a new calibration bench was mainly to offer a better and more services to the users, and to have a more reliable measurement system. Instead of using sources with different activities, the air Kerma rate may also be varied by means of lead attenuators for collimated beams of cesium and cobalt. The attenuators shall be placed in close vicinity to the diaphragm. A sequence of lead attenuator with different thickness leads to a reduction in air Kerma, or other magnitude by successive orders of magnitude for cesium and cobalt. In this work were studied irradiator performance, beam characteristics and other aspects. The facilities available for the laboratory to calibrate instrument are an irradiator, X ray equipment, reference instruments and others. Ionization cylindrical chamber 600cm 3 Nuclear Enterprises and an electrometer Farmer are secondary standards traceable to a primary laboratory. The irradiator has a 137 Cs source of 10 Curies and a 60 Co source of 1 Curie, approximately. Theoretical and experimental studies were done to analyse symmetry, flatness, penumbra and secondary radiation of photon beams as ISO 4037 suggests it [3-4].. Materials and methods.1 Irradiator description Hopewell Designs, Inc. Model G Gamma Beam Irradiator produces gamma beam that is used to calibrate instruments for radiation detection and to irradiate personal dosimeters [5]. Irradiator is cylindrical and it contains two gamma sources shielded by lead. External diameter is 17.78cm and height is 71.1cm, the thickness lead assure that the value of exposition rate on the surface be lower than 5mR/h. Sources movement is possible by an automatic system based on vacuum. Base Irradiator is made of iron and height is 68.5cm. Beam centerline is 110cm on the floor. Collimator it has tungsten disks specially design to absorb scatter radiation according to ISO , 1999 [4]. Aperture collimator is equal to 8º. In Figure 1 is showed a schematic diagram of the irradiator. Figure 1: Schematic diagram of the irradiator
3 Cylindrical sources have an active nucleus of 0.53cm and 1.84cm, diameter and length. They have a double encapsulation and sealed hermetically. External encapsulation is made of stainless type 304 with 1.8cm and 1.9cm of external diameter and height, respectively. In Figure is showed the irradiator belongs to the Radioprotection Laboratory. Figure : Irradiator of the Radioprotection Laboratory of Atomic Energy Commission. Penumbra, flatness and symmetry..1 TLDs measurements External Dosimetry Section of Constituyentes Atomic Center provided termoluminiscent dosimeters (TLDs). They were used to study penumbra and homogeneity of the photon beam. TLDs LiF 700 Bicron/Harshaw with dimension 3x3x0.89mm 3 were selected with a sensibility factor lower than 10%. They were inside a capsule. A total of 10 dosimeters were fixed on backside of acrylic slab of 5mm thickness and surface 50x50cm. They were distributed on Z and Y axis, perpendicular to beam axis, each 5cm on the entire surface. This slab was used to provide electronic equilibrium; its centre was positioned to 1.80 meters from exit collimator. Irradiation beam was 137 Cs and the time was the necessary to give an air Kerma of mgy, approximately. A total of 5 dosimeters were used to repeat the irradiation but to 1 meters from exit collimator. External Dosimetry Section read the TLDs in a Harshaw 3500 Reader following procedures written at Laboratory. Dose received by dosimeters were normalized to the maximum value and a dispersion value (max-min) and the uncertainty were calculated. Other studies were done to a distance of 1m from exit collimator using an ICRU slab phantom. LiF- 100 powder in capsules was fixed each 5 cm to the thinner and greater surface of the phantom and on the Z and Y axis. They were read in Harshaw Automatic reading. It was done for cesium and cobalt and the personal equivalent dose received was msv... Films measurements Dosimetric films X-Omat-V were positioned on a perpendicular direction of beam axis to 0.7 and 1m from exit collimator to film. For 0.7 meters only a film was used and for 1m 5 films were used to cover the radiation field, they were fixed to a slab acrylic of surface 50x50cm and thickness of 5mm. 3
4 They film to 0.70m received 10, 40, 80, 100 and 10 mgy to built calibration curve Optical density Vs Air Kerma. Films were developed and read at Spanish Hospital..3 Measurements out axis radiation beam According to the suggested in ISO measurements out axis radiation beam were done. Cylindrical reference chamber of 600cm 3 NE TYPE 575C Model 467 and FARMER DOSEMETER Model 70/1 Series Nº 1163 electrometer were used. Accumulated charges were collected to 0.70 and 1m distances from collimator exit-reference point of the chamber and to 0, 15, 30, 45, 60 and 75cm along an axis perpendicular to radiation beam. Those distances are the maximum possible due to the proximity of a wall. A percentage was calculated respect to the maximum value. Before using calibration equipment radioactive check and leakage control were done to assure the good performance..4 Scattering radiation measurement Scattering radiation was evaluated by 3 methods: Shadow-shield, Multiple distance and Monte Carlo simulation Shadow-shield Method. In this method air Kerma rate ( K & ) due to scattering radiation is assumed constant on all distances exit collimator-point reference. In this method a cone of a high atomic number (Z) is located between the source and the chamber to avoid that primary photon reach the chamber. The chamber is completely shielded of primary radiation and only reaches the chamber scattering radiation. Dispersion value is obtained by the quotient between charge without and with shielding. Shielding dimension should be he minimum required to reduce the primary radiation intensity in detector position less than % of value without shielding. Height cone is enough to provide attenuation and it is not so close of chamber to avoid scattering from the cone. A cylindrical ionisation chamber of 600cm 3, NE 551, Series Nº 490 with build up, an electrometer PTW-UNIDOS Model Series Nº 0119 were used. Dimension of cylinder made in lead were 8cm and 0cm, of diameter and length, respectively. Charge accumulated was collected at 0.7, 1, 1., 1.4, 1.5,,.5, 3, 3.5, 5 and 5m from exit collimator and reference point of chamber. For each position were taken 3 readings of 60 seconds each one, and an average value was calculated. Those reading corresponded to total radiation, primary plus secondary. Air Kerma rate was measurements in Then accumulated charge was collected for the same distances but interposing a cylinder, to approximately middle distance exit collimator-reference point of the chamber. For each position were taken 3 readings of 300 seconds each one, and an average value was calculated. In this way those values correspond to secondary radiation. Current values were corrected by a factor of pressure and temperature and corrected by drift. The difference between total and secondary radiation gave the primary radiation. The quotients between primary and secondary radiation and total and secondary radiation were calculated for different distances..4..analytical Method of multiple distances Readings to different distances show the inverse quadratic law and it assumes a constant dispersion. It is essential in this method that distances be precise and correct to derivate the correction c that gives the true distance exit collimator (center source)-chamber (d ). Where d = d + c d is the true distance center chamber center source. 4
5 d is the apparent distance center chamber center source. c reading error. The contribution of scattering radiation to K& total, K& S, is included in K& measured ( K & (d ) ). The value of K& due only primary photons K& P (d ) K & P ( d ) = K& ( d ) K& (1) total S & P ( d ) d = ( K& total ( d ) K S ) ( d + c) () Constant = K & Clearing gives Cte & total ( d ) = K& S + or y a + K ( d + c) b = (3) ( d + c) Air Kerma rate was calculated starting from the values got in the before point for 0.7, 1, 1., 1.4, 1.5,,.5, 3, 3.5, 5 and 5m from exit collimator and reference point of chamber. The curve obtained for Current vs distance for each beam was adjusted by a function which equation gave the value constant of the secondary radiation. Mathematic software and Curve Table software were used to obtain a, b, c values..4.3 Theoretical calculation with MCNP5c. Field size of radiation field, flatness and penumbra were calculated by Monte Carlo simulation. Photon flux was calculated on different point of radium radio cm in air using MCNP5c code [6]. Tally f5, flux at a point detector (particles/cm), was used for different distances from collimator exit, Z- axis, and on the transversal plane of the axis beam. The center of the point detectors were at 0, 5, 10, 13, 15, 17, 0, 3, 5 y 30 cm from the center of the beam for combinations using the same positions for Y axis. All was surrounded by an air sphere of radium 6 meters and outside was vacuum. The same was repeated for 0.70, 1 and 1.80 m from collimator exit for cesium and cobalt. For 1.80 m was considered more points on transversal plane because the size field was greater (0, 5, 10,13, 15,17, 0, 3,5, 30, 33, 34, 36, 37, 38, 39, 40, 4, 44, 45 and 50cm). Irradiator s geometry was obtained from manual given by manufacturer. Densities considered were g.cm -3, 19.3 g.cm -3 and x 10-3 g.cm -3 for lead, tungsten and air, respectively. Air composition by mass was 75.50, 3.176, 1.88, and % nitrogen, oxygen, argon, and carbon respectively. No variance reduction techniques were used. Enough photon histories were run to pass statistics test proposed for the code. The appropriate photon (MCNPLIB04) and electron (EL3) cross-section libraries were chosen from the DLC-0 library ENDF/B-VI Release 8 [7]. The ENDF/B-VI.8 photo atomic and atomic relaxation data are in turn based on the EPDL97 library, RSICC Data Library DLC Results 3.1 Penumbra, flatness and symmetry TLDs measurements Personal equivalent dose, Hp(10) was evaluated in units of milisievert (msv) and normalized to the value of maximum dose. Dispersion was calculated and resulted 19.4%. Uncertainty was 10% (k=). In Table 1 and is showed the percentage of photon flux (kerm or dose) for cesium and for 180cm, 100cm from collimator exit, respectively, using TLDs rods. 5
6 Table 1: Percentage of photon flux (kerm or dose) for cesium and 180cm from collimator exit using TLDs rods. Y/Z Table : Percentage of photon flux (kerm or dose) for cesium and 100cm from collimator exit using TLDs rods. Y/Z According to the table 1 all dosimeters were inside radiation field and penumbra could not be measured. Beam homogeneity on a surface of 50x50cm to 1.80 meters from exit collimator 80.6%, size field is 5cm of radium and less of 15cm for 1meter. Penumbra is less than 3 cm. Difference between the value of the air kerm given and the read it is less than 10%. Respect to measurements made with powder LiF, the most of them showed standard deviation greater than 5% so they could not be taken into account Films measurements In Figure 3 is showed several imaging s, the first is for cesium to 1 meter from collimator exit and it is more diffuser than others. Second and third figure are for cobalt to 0.7m and 1 meter from exit collimator. Radiation field had a radium of 19cm for cobalt, cesium and penumbra resulted of 0.7cm. For 1 meter the results were similar for cesium and cobalt, 16cm of radium and penumbra of 0.4cm. Symmetry and flatness were greater than 90%. 6
7 Figure 3: Developed films for cesium to a distance of 1meter, cobalt to 0.70m and 1 meter from collimator exit, respectively. 3. Measurements out axis radiation beam Air Kerma was calculated for 0.7 and 1m from exit collimator to reference point of chamber and values were normalized to the maximum. It was realized for cesium and cobalt. In Table 3 is showed the percentage of dose for cesium and cobalt for 0.7m and 1m from collimator exit. Its showed that the size field to 1m has a radium of 15cm and to 0.7m is less than 10cm. Table 3: Percentage of dose for cesium and cobalt for 0.7m and 1m from collimator exit. 60 Co (1m) 137 Cs (1m) 60 Co (0.7m) 137 Cs (0.7m) Distances (cm) % % % % Scattering radiation measurement Shadow-shield method Results of that part showed that the scattering radiation was lower than 4%, below that 5% suggested for ISO Analytical Method of multiple distances In Table is showed of a, b, c and regression coefficient values for equation (3) for cesium and cobalt sources. Table 4: a, b, c and regression coefficient values for equation (3) for cesium and cobalt sources. a b c r 1-Cesio Cesio Cesio Cobalto Cobalto
8 Remembering that y is the total current, a is the scattering current (secondary), c is the error from distance and b is a constant. Secondary intensity varies from 0.13% to 4.7% of total intensity for 60 Co and for 0.7 and 5 meters, respectively. For the case of 137 Cs the values varies from 0.1% to 4.9% for 0.7 and 5 meters. Differences in Table 4 are due that different distances between 0.7 and 5 meters were considered, in some cases they were from 1 to 4 meters Monte Carlo Simulation Results of that part showed that the scattering radiation was lower than 4.%, below that 5% suggested for ISO Results for 137 cesium, 60 cobalt showed a radium for radiation field of 0cm and 17cm for 100cm from collimator exit. For 180cm the results were 33cm and 30cm respectively. For 0.7m the results were 15cm and 13cm respectively. Penumbra was of.5cm approximately and beam homogeneity was 85%. Size field results agree to the ones from films. For penumbra the differences are greater. 4. Conclusion An exhausted study was done to the irradiator and this showed the good performance of itself. Differences from theoretical and experimental studies of radiation field size and penumbra, specifically with TLDs, are going analysed. Values of secondary radiation are acceptable as well as homogeneity beam. In a future work the secondary radiation is analysed which part is due to the floor, walls and ceiling using Monte Carlo simulation. Respect to the future of the Radioprotection Laboratory, this is going to have an intern audit to intent to pass next year and audit of Accreditation Argentinean Organism. For that reason instructions and procedures are being written at this time. Acknowledgements The authors wish to thank Mrs. Rosana Sansogne from Spanish Hospital and Mr. Horacio Sanchez for their collaboration. REFERENCES [1] INTERNATIONAL ATOMIC ENERGY AGENCY, Calibrations of Radiation Protection Monitoring Instruments, Safety Reports Series No. 16, Vienna (000). [] NACIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS, Calibrations of Survey Instruments used in Radiation Protection for the Asessment of Ionizing Radiation Fields and Radioactive Surface Contamination, NCRP Report No. 11, Bethesda, (1991). [3] INTERNATIONAL ORGANIZATION FOR STANDARIZATION, X and gamma reference radiation for calibrating dosimeters and dose rate meters and for determining their response as a function of photon energy. Part : Dosimetry for radiation protection over the energy ranges 8keV to 1,3MeV and 4MeV to 9MeV. ISO 4037-, Geneva, (1997). [4] INTERNATIONAL ORGANIZATION FOR STANDARIZATION, X and gamma reference radiation for calibrating dosimeters and dose rate meters and for determining their response as a function of photon energy. Part 1: Radiation characteristics and production methods. ISO Geneva, (1997). 8
9 [5] FERNANDES, E., et al., The radiation field characteristics of a 137 Cs source used for calibration of radiation protection instruments, Applied Radiation and Isotopes 61 (004) [6] MCNP A General Monte Carlo N-Particle Transport Code, Versión 5, X-5 Monte Carlo Team, Los Alamos National Laboratory, Los Alamos, NM (003). [7] RADIATION SAFETY INFORMATION COMPUTATIONAL, RSICC Data Library DLC- 0, Oak Ridge National Laboratory (005), 0.html. 9
Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved
Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah
More informationDesign, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors
6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator
More informationM [scale units/s] of the system
APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers
More informationTITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137
TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;
More informationImplementing the ISO Reference Radiations at a Brazilian Calibration Laboratory
Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry
More informationSpecific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements
Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected
More informationCommissioning of the Beta Secondary Standard (BSS2)
Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,
More informationDetermination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator
Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,
More informationStudy of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
More informationCalibration of a pencil ionization chamber with and without preamplifier
Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.
More informationLaboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)
PROPOSAL FOR EYE-LENS DOSEMETER CALIBRATION AND TYPE TESTING ORAMED WP2 J.-M. Bordy (1), J. Daures (1), M. Denozière (1), G. Gualdrini (2), M. Guijaume (3), E. Carinou (4),F. Vanhavere (5) (1) Laboratoire
More informationNeutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace
Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace Gregori, B.N.; Carelli, J.L; Cruzate, J.A.; Papadópulos, S. and Kunst, J.J. Presentado en: Second European of IRPA (International
More informationIonizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI
Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio
More informationDose Rate Levels around Industrial Gamma Sources
Dose Rate Levels around Industrial Gamma Sources Héctor René Vega-Carrillo *, Ricardo Rodríguez-Juárez, Eduardo Manzanares- Acuña, Rubén Hernández-Villasana, Jaime Ramírez-González Universidad Autónoma
More informationAn extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter
An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter Christopher A Perks *a and Stephane Faugoin a a LANDAUER EUROPE, 33 Avenue du General Leclerc, F-92266, Fontenay-aux-Roses
More informationTRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES
TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES S. MORATÓ, A.BERNAL, A. QUEROL, A. ABARCA, C. GÓMEZ-ZARZUELA, R.MIRÓ, G.VERDÚ Institute for Industrial, Radiophysical and Environmental
More informationPersonal Dose Monitoring System
Personal Dose Monitoring System Kei Aoyama Osamu Ueda Takeshi Kawamura 1. Introduction The level of radiation safety control in the nuclear power field and at facilities that utilize radiation has become
More informationKCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck
Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV
More informationHp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015
Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Leon de Prez and Frans Bader VSL - The Netherlands www.vsl.nl P 1 Contents - Introduction to VSL and Ionizing
More informationComparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2
Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,
More informationCity University of Hong Kong
City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and
More informationMonte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data
Pramana J. Phys. (2017) 88: 49 DOI 10.1007/s12043-016-1344-1 c Indian Academy of Sciences Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data M
More informationWork Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures
Work Programme of the International Bureau of Weights and Measures for the four years 2016-2019 Comité international des poids et mesures Section II : BIPM Work Programme for 2016-2019 35 Priority activities
More informationTA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1
TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order
More informationTechnical basis of occupational dosimetrygeneral
Technical basis of occupational dosimetrygeneral review Antti Kosunen, Hannu Järvinen STUK 2017 NACP-RPC COURSE Occupational dosimetry in interventional radiology, - cardiology and nuclear medicine 27-29
More informationCHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS
CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,
More informationReferensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden
Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties
More informationUse of a radioactive check device for redundancy check of ionization chambers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman
More informationACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS
ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS Hernán G. Meier, Martín Brizuela, Alexis R. A. Maître and Felipe Albornoz INVAP S.E. Comandante Luis Piedra Buena 4950, 8400 San Carlos
More informationRecent Activities on Neutron Calibration Fields at FRS of JAERI
Recent Activities on Neutron Calibration Fields at FRS of JAERI Michio Yoshizawa, Yoshihiko Tanimura, Jun Saegusa and Makoto Yoshida Department of Health Physics, Japan Atomic Energy Research Institute
More informationEvaluate Shielding Design of the Brachytherapy Unit by Using Monte Carlo Simulation Code
Journal of Modern Physics, 15, 6, 9-911 Published Online June 15 in SciRes. http://www.scirp.org/journal/jmp http://dx.doi.org/1.436/jmp.15.6794 Evaluate Shielding Design of the Brachytherapy Unit by Using
More informationA Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators
A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,
More informationProgress in Nuclear Science and Technology, Volume 6,
DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi
More informationNeutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom
Neutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom Problem P4 Rick J Tanner National Radiological Protection Board Chilton, Didcot, Oxon OX11 0RQ, United Kingdom Intercomparison
More informationABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER
ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),
More informationDosimetry. Sanja Dolanski Babić May, 2018.
Dosimetry Sanja Dolanski Babić May, 2018. What s the difference between radiation and radioactivity? Radiation - the process of emitting energy as waves or particles, and the radiated energy Radioactivity
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 6980-2 Première edition 2004-10-15 Nuclear energy Reference beta-particle radiation Part 2: Calibration fundamentals related to basic quantities characterizing the radiation
More informationRadionuclide Imaging MII Detection of Nuclear Emission
Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor
More informationMeasurements with the new PHE Neutron Survey Instrument
Measurements with the new PHE Neutron Survey Instrument Neutron Users Club Meeting National Physical Laboratory 16 th October 2013 Jon Eakins, Rick Tanner and Luke Hager Centre for Radiation, Chemicals
More informationHigh-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
More informationCalculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*
SLAC-PUB-70 Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* J. C. Liu, K. R. Kase, X. S. Mao, W. R. Nelson, J. H. Kleck, and S. Johnson ) Stanford Linear
More informationCOMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD
COMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD Kirill SKOVORODKO, Birute GRICIENE, Milda PETKELYTE Radiation Protection Division, Vilnius University Hospital Santaros
More informationAccuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi
Accuracy, Calibration, Type Testing and Traceability General [Chapter 6 & 7 of RP 160] Peter Ambrosi EURADOS Training Course Lisbon, Portugal, 18-22 Mai Content Introduction - Aim of Measurement & type
More informationRadiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007
Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp. 229 233 Advance Access publication 11 May 2007 doi:10.1093/rpd/ncm047 CHARACTERIZATION AND UTILIZATION OF A BONNER SPHERE SET BASED ON GOLD
More informationAir Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira
Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation
More informationNORM Monitor-IS. Performance data
NORM Monitor-IS Performance data Contents 1. Introduction... 3 Scintillator probe... 3 GM probe... 3 2. Energy response... 4 Scintillator probe gamma energy response... 4 Scintillator probe alpha/beta
More informationProgress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima
Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Norio Saito, Tadahiro Kurosawa, Masahiro Kato, Yuichiro Morishita, Takahiro Tanaka, and Morihito
More informationUpdated Contact Dose Rate Conversion Factors for Encapsulated Gamma Sources Including Secondary Electron Emission
Updated Contact Dose Rate Conversion Factors for Encapsulated Gamma Sources Including Secondary Electron Emission Eric Heritage Supervisor: Dr. Ed Waller Outline Background Objectives PHITS Simulations
More informationDosimetry: Electron Beams
ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann
More informationEnergy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI
Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration s of KAERI B.H. Kim 1, J.L. Kim 1, S.Y. Chang 1, J.K. Chang 1, G. Cho 2 1 Korea
More informationCOMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*
Original Article COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Roseli T. Bulla 1, Linda V.E. Caldas 2 * Study developed at Instituto de Pesquisas Energéticas
More informationComparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017
Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018
More informationChapter 10 Acceptance Tests and Commissioning Measurements
Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:
More informationGeneral characteristics of radiation dosimeters
General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationDetermination of the tissue attenuation factor along two major axes of a high dose rate HDR 192 Ir source
Determination of the tissue attenuation factor along two major axes of a high dose rate HDR 192 Ir source Sang Hyun Cho a) Department of Radiation Physics, The University of Texas M. D. Anderson Cancer
More informationMonte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng
SLAC-PUB-7785 November, 1998 Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng Stanford Linear Accelerator Center MS 48, P.O.
More informationRadiation exposure of personnel during IORT: radiation protection aspects.
Radiation exposure of personnel during IORT: radiation protection aspects. L. Strigari 1, V. Bruzzaniti 1, V. Landoni 1, A. Soriani 1, S.Teodoli 1, M. Benassi 1 1 Lab. Fisica Medica e Sistemi Esperti,
More informationReliability and Traceability in Radiation Calibration
Reliability and Traceability in Calibration Sadao Nakashima Teruo Ebato 1. Introduction Fuji Electric has a successful history of delivering radiation measuring equipment used in the controlled areas of
More informationCHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR
ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan
More informationOccupational Radiation Protection at Accelerator Facilities: Challenges
Occupational Radiation Protection at Accelerator Facilities: Challenges Haridas.G Health Physics Division Bhabha Atomic Research Centre INDIA Int. Conf. on Occupational Radiation Protection: Enhancing
More informationCalibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.
Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience. Osvaldo Piñones O., Sylvia Sanhueza M. Radio medicine Section, Chilean Commission
More informationDetermination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5
Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5 P. Deatanyah 1, C.C. Arwui 1, S. Wotorchi- Gordon 1, H. Lawluvi
More informationAtoms, Radiation, and Radiation Protection
James E. Turner Atoms, Radiation, and Radiation Protection Third, Completely Revised and Enlarged Edition BICENTENNIAL J 0 1 8 0 Q 71 z m z CAVILEY 2007 1 ;Z z ü ; m r B10ENTENNIAL WILEY-VCH Verlag GmbH
More informationSurvey Meter OD-01 Address: Phone: Fax: URL:
Survey Meter OD-01 Dose meter and dose rate meter for the measurement of the ambient dose and dose rate equivalent H*(10), dh*(10)/dt and the directional dose and dose rate equivalent H'(0.07), dh'(0.07)/dt
More informationMCNP CALCULATION OF NEUTRON SHIELDING FOR RBMK-1500 SPENT NUCLEAR FUEL CONTAINERS SAFETY ASSESMENT
MCNP CALCULATION OF NEUTRON SHIELDING FOR RBMK-15 SPENT NUCLEAR FUEL CONTAINERS SAFETY ASSESMENT R. Plukienė 1), A. Plukis 1), V. Remeikis 1) and D. Ridikas 2) 1) Institute of Physics, Savanorių 231, LT-23
More informationInduced photonuclear interaction by Rhodotron-TT MeV electron beam
PRAMANA c Indian Academy of Sciences Vol. 78, No. 2 journal of February 2012 physics pp. 257 264 Induced photonuclear interaction by Rhodotron-TT200 10 MeV electron beam FARSHID TABBAKH 1,, MOJTABA MOSTAJAB
More informationA new neutron monitor for pulsed fields at high-energy accelerators
A new neutron monitor for pulsed fields at high-energy accelerators Marlies Luszik-Bhadra *, Eike Hohmann Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116, Braunschweig, Germany. Abstract.
More informationRadiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005
Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp. 233 237 Advance Access publication 6 October 2005 doi:10.1093/rpd/nci353 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS IN HIGH-ENERGY BREMSSTRAHLUNG
More informationInvestigation Of The Effects Of Variation Of Neutron Source-Detector Distance On The Emitted Neutron Dose Equivalent
ISSN: 9- Vol. Issue, June - Investigation Of The Effects Of Variation Of Neutron Source-Detector Distance On The Emitted Equivalent Igwesi, D. I. Physics and Industrial Physics Department, Faculty of Physical
More informationFast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry
Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry F. Gutermuth *, S. Beceiro, H. Emling, G. Fehrenbacher, E. Kozlova, T. Radon, T. Aumann, T. Le Bleis, K. Boretzky, H. Johansson,
More informationPERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER
SLAC-PUB-95-6749 (March 1995) PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER J. C. Liu, D. Busick 1, K. R. Kase, R. C. McCall 2, R. Sit and H. Tran 3 Stanford Linear
More informationGamma Radiation Absorption Characteristics of Concrete with Components of Different Type Materials
Vol. 117 (2010) ACTA PHYSICA POLONICA A No. 5 Selected papers presented at the Eleventh Annual Conference of the Materials Research Society of Serbia, YUCOMAT 2009 Gamma Radiation Absorption Characteristics
More informationA Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields
Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard
More informationMonte Carlo Simulation concerning Particle Therapy
Monte Carlo Simulation concerning Particle Therapy Masaaki Takashina Graduate School of Medicine, Osaka University, Osaka 565-0871, Japan INTRODUCTION It is well known that the particle therapy has some
More informationABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationSTANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS
STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,
More informationESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT
Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and
More informationRadiation Shielding of Extraction Absorbers for a Fermilab Photoinjector
Fermilab FERMILAB-TM-2220 August 2003 Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector I.L. Rakhno Fermilab, P.O. Box 500, Batavia, IL 60510, USA August 12, 2003 Abstract Results
More informationMONTE CARLO SIMULATIONS USED TO ESTABLISH A FRENCH NATIONAL HDR BRACHYTERAPY REFERENCE
Mathematics and Computation, Supercomputing, Reactor Physics and Nuclear and Biological Applications Palais des Papes, Avignon, France, September 12-15, 2005, on CD-ROM, American Nuclear Society, LaGrange
More informationRecent Activities on Neutron Standardization at the Electrotechnical Laboratory
Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)
More informationCalibration of the GNU and HSREM neutron survey instruments
Calibration of the GNU and HSREM neutron survey instruments Neutron Users Club Meeting National Physical Laboratory 20 th October 2015 J. S. Eakins 1, L. G. Hager 1, J. W. Leake 2, R. S. Mason 2 and R.
More informationNew irradiation zones at the CERN-PS
Nuclear Instruments and Methods in Physics Research A 426 (1999) 72 77 New irradiation zones at the CERN-PS M. Glaser, L. Durieu, F. Lemeilleur *, M. Tavlet, C. Leroy, P. Roy ROSE/RD48 Collaboration CERN,
More informationRadiation Protection & Radiation Therapy
Radiation Protection & Radiation Therapy For Medical Students Professor of Medical Physics Radiation Units Activity Number disintegrations per second (Curie, Becquerel) Exposure (Roentgen, C/kg) Absorbed
More informationOn-Site Calibration Method of Dosimeter Based on X-Ray Source
World Journal of Nuclear Science and Technology, 2017, 7, 93-102 http://www.scirp.org/journal/wjnst ISSN Online: 2161-6809 ISSN Print: 2161-6795 On-Site Calibration Method of Dosimeter Based on X-Ray Source
More informationInt. Coaf. Physics of Nuclear Science and Technology
BNL-6 5 2 43 Int. Coaf. Physics of Nuclear Science and Technology October S-8, 1998 RADIATION DOSIMETRY AT THE BNL HIGH FLUX BEAM REACTOR* NE Holden', J-P Hu', RN Reciniello* 1. Reactor Division, Brookhaven
More informationCode of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams
Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry
More information1. RADIOACTIVITY AND RADIATION PROTECTION
1. Radioactivity and radiation protection 1 1. RADIOACTIVITY AND RADIATION PROTECTION Revised August 2011 by S. Roesler and M. Silari (CERN). 1.1. Definitions [1,2] 1.1.1. Physical quantities: Fluence,
More informationABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS
ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento
More informationBureau International des Poids et Mesures
Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September
More informationCALIBRATION OF INSTRUMENTS MEASURING RADON OVER A LARGE ACTIVITY RANGE. N. Michielsen, V. Voisin
Radon in the Living Environment, 011 CALIBRATION OF INSTRUMENTS MEASURING RADON OVER A LARGE ACTIVITY RANGE N. Michielsen, V. Voisin Institut de Protection et de Sûreté Nucléaire, Département de Prévention
More informationPRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM
Rapport BIPM-09/02 PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM D T Burns, C Kessler, P Roger BIPM A R El-Sersy NIS (Egypt) July 2009 BUREAU INTERNATIONAL DES POIDS ET
More informationChapter V: Cavity theories
Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly
More informationRadiological impact of rutile covered welding electrodes
LABORATORIO DE MEDIDAS DE BAJA ACTIVIDAD Universidad del País Vasco / Euskal Herriko Unibertsitatea Radiological impact of rutile covered welding electrodes Margarita Herranz 1, Saroa Rozas 1, Carmen Pérez
More informationChapter 9: Calibration of Photon and Electron Beams
Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation
More informationEvaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam
Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam Radhakrishnan B Nair 1*, Ramakrishnan G 2, Chandralingam S 3 and Kurup PGG 1 1 Apollo Speciality Hospital, Chennai, India
More informationMECHANICAL ENGINEERING. Five-year M.Sc. studies School of Mechanical Engineering National Technical University of Athens
MECHANICAL ENGINEERING Five-year M.Sc. studies School of Mechanical Engineering National Technical University of Athens Athens, September 8, 2010 Semester 1: Mathematics Ia (52h lectures and exercises)
More informationRadhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****
Photon Interaction Cross Section of Materials in Heterogeneous Energy Spectrum of Medical Diagnostic X-Ray Beam Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** *Senior Medical Physicist,
More informationPHYS 3650L - Modern Physics Laboratory
PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma
More informationREALISTIC NEUTRON SPECTRA FOR RADIATION PROTECTION AND OTHER APPLICATIONS AT AERI, BUDAPEST
1 Reference number: 067 REALISTIC NEUTRON SPECTRA FOR RADIATION PROTECTION AND OTHER APPLICATIONS AT AERI, BUDAPEST J. Pálfalvi 1), L. Sajó -Bohus 2) and M. Balaskó 1) 1) Atomic Energy Research Inst. P.O.B.
More informationSince the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar
Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.255-263 Beam Dump for High Current Electron Beam at JNC H. Takei and Y. Takeda 1 Japan
More information