Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Size: px
Start display at page:

Download "Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator"

Transcription

1 Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele, Romania scarlat.f@gmail.com ABSTRACT This paper presents the dosimetric measurements performed at INFLPR 7 MeV Linear accelerator by means of the secondary standard chamber H(10), type. A secondary standard for photon radiation is based on the linear relation between the ionizing current and the conventionally true value of the ambient dose equivalent. The measurements were made for the electron beam accelerator operation mode. Determination of the ambient dose equivalent was made by means of the PTW ionization chamber, type, calibrated by PTW Freiburg. Key Words: Radiological Protection/Secondary Standard Chamber/Ambient Dose quivalent. INTRODUCTION In the lectron Accelerator Department in the National Institute for Laser, Plasma and Radiation Physics (INFLPR) a Secondary Standard Dosimetry Laboratory for high energies it was built. It is outfitted with a secondary standard dosimeter UNIDOS equipped with proper accessories for the absorbed dose and the absorbed dose rate measurement in certain energy domains (between 8 kev and 50 MeV). For the penetration powerful radiations, the operational parameter used for certain area of the field radiation monitor, is the ambient dose equivalent H(d). According to international standards recommendations, the conventional true values of radioprotection operational parameters are based on air kerma free in air, K a, by conversion coefficients application. The mono-energetic conversion coefficients, from air kerma to H(d), very much depend o photon energy. The work is starting with presenting the ambient and directional dose equivalent concepts, photon radiation characteristic parameters specification, followed by STARDOOR laboratory secondary standard ionization chamber description with its main characteristics, the ambient dose equivalent H(d) determination methods, where d is the depth under a specific point on the body or in Ω direction inside the ICRU Sphere. Using TN ionization chamber, the ambient dose equivalent in one of STARDOOR laboratory chamber having a common wall with one wall of the irradiation hall were INFLPR 7 MeV linear accelerator is placed, is determined. The work concludes with the results and conclusions. AMBINT AND DIRCTIONAL DOS QUIVALNT To monitor the environment and area, two concepts which connect the external radiation field, to the effective dose and equivalent dose in skin are introduced. The first one, the 297

2 ambient dose equivalent, H(d), is adequate to the powerful penetration radiation, and the second, the directional dose equivalent H (d,ω), is proper for soft penetration radiation. Ambient dose equivalent, H(d), at a point in a radiation field is the dose equivalent that would be produced by the corresponding expanded and aligned field in the ICRU Sphere at a depth d on the radius opposing the direction of the aligned field. In SI, H(d) is in J/kg, and it has been given the special name Sievert (symbol Sv). In CGS, its unit is the rem (1 Sievert = 100 rem) [ICRU 85]. xpanded Radiation Field is a theoretical construct by which the influence and its angular and energy distribution have the same values throughout the volume of interest as in the actual field at the point of reference. In an aligned and expanded field the fluence and its energy distribution are the same as in the expended field but the fluence is unidirectional [ICRU85]. The directional dose equivalent, H (d, Ω,) at a point in a radiation field is the dose equivalent that would be produced by the corresponding expanded field in the ICRU Sphere at depth d on a radius in a specified direction Ω. The SI unit is J/kg or sievert [ICRU 85]. The CGS unit is the rem. 1 Sv =100 rem. Reference radiation Point of test Reference axis Fig. 1 - General view of radiation geometry for in phantom mesurements: cross section through ICRU Sphere with 300 mm diameter. Fig. 1 presents ICRU Sphere used for ambient equivalent dose measurements. ICRU Sphere is a tissue equivalent Sphere prescribed by the ICRU to have a diameter of 30 cm, a composition by mass of 76.2 % O, 10.1 % H, 11.1% C, 2.6 % N and a density of 1 g/cm 3. PHOTON RADIATION PARAMTRS In accordance with international standards [ISO ], photon radiation is characterized by following parameters: a) mean photon beam, Ē; b) spectral resolution, R ; c) half-value layer, HVL; d) homogeneity coefficient, h. Mean energy photon, Ē, expressed in kev is defined by the formula (1): max Φ d 0 =, (1) max Φ d 0 where Φ is the derivative of the fluency Φ of the primary photons of energy with respect to energies between and +d, defined as (2): 298

3 dφ Φ =. d (2) Table 1 presents some mean energies, corresponding to radiations characterized by certain parameters shows in ISO/FDIS Table 1 Radiation quality according to ISO/FDIS Ē ph (kev) according to ISO/FDIS N-10 8 N N N S-Cs 662 S-Co 1250 Spectral resolution, R, is expressed in percent and is defined by the ratio (3): Δ R = 100, (3) were Δ, is the spectrum width corresponding to half the maximum ordinate of the spectrum. Half value layer, HVL, represents the thinness of the specified material which attenuates the radiation beam to an extent such that the air kerma rate is reduced to half of its original value. Table 2 Homogeneity Name of series Resolution, R coefficient, h Low air-kerma rate 18 to Narrow spectrum 27 to to 1.0 Wide spectrum 48 to to 0.98 High air-kerma rate Not specified 0.64 to 0.86 Fig. 2 shows the theoretic spectral series for soft air kerma rate, corresponding to energies between 10 kev and 240 kev. Counts kev Counts kev , kev Fig. 2 - Theoretic spectra for low air-kerma series [4]. 299

4 Kerma, K = d tr /dm, where d tr is the sum of the initial kinetic energies of all the changed ionizing particles liberated by uncharged ionizing particle in a material of mass dm. In SI, K is in J/kg or Gy. Homogeneity coefficient, h, is given by the ratio of the first half value layer to the second half value layer (4). st 1 HVL h =, (4) nd 2 HVL Table 2 renders the spectral resolution and homogeneity coefficient corresponding to photon radiation versus their spectral series. SCONDARY STANDARD CHAMBR For ambient dose equivalent determination, a standard ionization chamber H(10) TN type, was used. The H(10) chamber characteristics having the calibration factor N H = 2,863x10 4 Sv/C (1), are the followings: - sensible volume: 1l; - the reference point: in the chamber center; nominal answer - 30 μc/sv; - chamber voltage: 400 V; - answer in energy: ± 4 %; - leakage current: ± 10 fa; - central electrode diameter: 50 mm; - exterior diameter of the chamber mm and radiation energy that can be measured is in 25 kev 1.3 MeV domain. The environment conditions in which must be done the measurements with this chamber are: - temperature: ( ) ºC, - humidity: ( ) % and air pressure: ( ) hpa. Fig. 3 - General view of secondary standard ionization chamber used for ambient dose equivalent H (d) and directional dose equivalent, H (d). The homogeneity of the answer for all chamber orientations with Makrolon surface inside the 4π solid angle was investigated by Aukerten for soft photon energies (N-15 and N-20). The leakage current is A. For the ambient dose equivalent with 1 msv/h value, the leakage current was of 1 x A. Fig. 3 presents the H(10) ionization chamber. The answer of H(10) chamber shows a minimal dependence for domain kev and it is in accordance with ISO specifications. The answer is varying under 1% for variations of the photons with energies of ± 1keV. The Spherical TN32002 ionization chamber was connected to a UNIDOS secondary standard dosimeter delivered by PTW, accompanied by its Test Certificate. 300

5 MASURMNT MTHOD Conventional true value (2) for ambient dose equivalent for R radiation quality is obtained with the relation H(10) = N 60 k(r) Q, (5) where: - N 60 - is the calibration factor at the reference radiation quality N-60; - k(r)- is the correction factor at the reference radiation quality N 60, given by expression: K a N 60 = k K (10, N 60), (6) Q where: - K a - the conventionally true value of the air kerma; - k R (10, N-60) - is the conversion coefficient from K a to H(10) for the quality N-60. For the determination of the calibration factor N 60 and the correction factors k(r), the measurements were carried out for energies ranging kev using the radiation qualities of the ISO, narrow-spectrum series (N 10 N 300 ) and the ISO gamma qualities S-Cs and S-Co. The chamber H(10) answer shows the energetic dependence for all domain kev. The answer r(r) of the chamber for R radiation quality is given by (7): Q Q r( R) = =, (7) H (10) K a hk (10; R) were Q is the measured charge made by the chamber H(10) and represents the conventional true value for ambient dose equivalent H(10), K a is the conventional true value for in air kerma free in air and h K (10; R) is the conversion coefficient from Ka to H(10) for R radiation quality. Conversion coefficient values h K (10; R) for X ray quality was taken from catalogue with characteristics of photon radiation spectra (3). The answer values for each quality of R 3 3 radiation r(r), were corrected by reference conditions (air density ρ0 = g cm, air pressure po= kpa, air temperature T o = K, air relative humidity r o = 0.65). Fig. 4 shows the energy dependence on the relative answer r rel (R) given by (8): r( R) r rel ( R) =, (8) r( N 60) were r(r) is the answer for R quality radiation and r(n-60) is the answer for reference radiation quality N-60 ( = 47.9 kev) which is presented. ph Relative answer r rel (R) for ionization chamber H(10) indicates a variation to 10 % versus photon mean energy from 20 kev si 1250 kev domain for the narrow spectrum N-10 to N-300 and the reference field for S-Cs and S-Co. The big difficulty for a ionization chamber realization is represented by the exact production of the same wall densities in each point of the Sphere and particularly of an thin level (width ~ 44 μm) made of an mixture of different metals on the Sphere ring surface, needed for a large answer for 40 kev energies. Small differences are in the width resulted from different answers, which differ more and more with the decrease of photon energy. The answer homogeneity for all spherical chamber orientations to a 4π solid angle was investigated using all the qualities from N-15 to N-120. Answer displacement except for the domain of stem chamber and open chamber is maxim ~ 2.6 %. Inside the 40 and 1250 kev 301

6 energy domain, (for example for energies for which the chamber H(10) was perfected) the ratio between the maximum and the minimum of the answer is For energy qualities from N-15 (Ē ph =12.7 kev) to N-40 (Ē ph =33.3 kev) this ratio is 2.9. Only the first value is complying with the standard ISO (4) requirements, concerning the energy dependence of secondary standard chamber. 10 r rel (R) kev 1000 Fig. 4 - The relative answer r rel (R) for ionization chamber H(10), Ē is the mean energy for photon beam [2]. Previously taking about the 30 kev mean energy, the maximum - minimum ratio for the instrument answer shall not be grater more than 1.1 over the energy domain for which standard instrument is used. For mean energies between 8 kev and 30 kev, the limit for this ratio shall not be more than 1.2. Any time accessible, the reference radiations used for secondary standard instrument calibration will be the same, like the ones used for the calibrations of radiation protection instruments. RSULTS The room in which the experiments were made, has a common wall with the hall in which the linear electron accelerator is placed and it includes a little room with lead walls in which radioactive checking devices of 20 MBq, respectively 33.3 MBq are stored. Main parameters of 7 MeV linear accelerator of are the followings: Maximum energy of accelerated electros, = hν, 7 MeV Mean beam current, I, 10 μa lectron macro-pulse duration, 2.5 μs lectron fluency at 1Gy dose, 2, e/cm 2 s Macro-pulse frequency, Hz Photon dose rate, 2 Gy/ min lectron dose rate, 100 Gy/min 302

7 Calibration certificate for secondary standard chamber, type TN H(10), specifies the following parameters: calibration factor N H = 2,863x10 4 Sv/C, obtained at a Co source with equivalent dose rate H (10) = (0,5 1000) msv/h and dose (1 50) msv, at DSC = 300 cm, isodose diameter of 97% for 28 cm, incidence angle α = 0 0 and correction factor k Q = Further Table 3 shows the results obtained after the measurements conducted for the absorbed dose equivalent determination, with the accelerator off and on generating an electron radiation with 350 R/min dose rate. The Spherical ionization chamber was oriented to the wall of the hall in the accelerator is placed, at 40 cm distance. Table 3 Measure mode Accelerator on Accelerator off Ambient dose equivalents H(10) [nsv/min] Mean ambient dose equivalent H(10) [nsv/min] / [μsv/h] / / 0.66 From Table 3 one can see that when the accelerator is on, the ambient dose equivalent is with 1.83 nsv/min and 0.11 µsv/h respectively, higher than when the accelerator is off. The preliminary measurements for ambient dose equivalent determination were made with radioactive checking devices for Sr-90 with activity 20 MBq. Table 4 Measure mode Ambient dose equivalent H(10) Without source 0.7 [μsv/h] Open container 0.7 [μsv/h] ; SSD = 100 cm Open container 1.7 [μsv/h] ; SSD = 10 cm Open container 3.2 [μsv/h] ; SSD = 0 cm Four measurements were conducted, in which the radiation Sr-90 source was placed at different distances to the Spherical ionization chamber: first measurement was made without the source followed by other three more with the ionization chamber successively placed at 100, 10 and 0 cm distance the source (Table 4). Throughout the process of obtaining the results, the accelerator was off. CONCLUSIONS Ambient dose equivalent, H(d), in one point in the radiation field, is the dose equivalent that could be produced by the aligned and expanded corresponding field, in ICRU Sphere [5], at an d depth, on a opposite ray with the alienated field. The measurements of ambient dose equivalent made with spherical ionization chamber PTW TN32002 with calibration factor N H = 2,863 x 10 4 Sv/C, lead to the following results: 1) in the room which has a common wall with the hall were the electron accelerator is placed, the ambient dose equivalent was by 1.2 times smaller than 1μSv/h value imposed by radiological standards and 2) near the radioactive checking device container, the dose equivalent was equal to 1 μsv/h at 10 cm distance to the source, value that meets the previsions in the radiological standards. 303

8 RFRNCS (1) Calibration Certificate, no for Ionization Chamber type TN , PTW- Freiburg, Physikalisch-Technische Werkstatten, Germany (2008). (2) Ankerhold, U.; Optimisation of a secondary standard chamber for the measurement of the ambient dose equivalent, H(10), for low photon energies, Radiat. Prot. Dosim. 118 (1), (2006). (3) International Standard ISO , X and Gamma Reference Radiations for Calibrating Dosemeters and Dose rate Meters and for Determining their Response as a Function of Photon nergy Part 3: Calibration of Area and Personal Dosemeters and the Measurement of their Response as a Function of nergy and Angle of Incidence (1997). (4) International Standard ISO , X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy - Part 2: Dosimetry for radiation protection over the energy ranges from 8 kev to 1,3 MeV and 4 MeV to 9 MeV (1997). 304

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS F. SCARLAT 1, A. SCARISOREANU 1, M. OANE 1, E. BADITA 1, E. MITRU 1 1 National Institute for Laser, Plasma and Radiation

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

Quantities and Units for Use in Radiation Protection

Quantities and Units for Use in Radiation Protection ICRU NEWS December 1991 Work in Progress 5 Quantities and Units for Use in Radiation Protection A Draft Report A. Allisy, W.A. Jennings, A.M. Kellerer, J.W. Müller, and H.H. Rossi ICRU Bethesda, MD 20814-3095,

More information

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Leon de Prez and Frans Bader VSL - The Netherlands www.vsl.nl P 1 Contents - Introduction to VSL and Ionizing

More information

11/23/2014 RADIATION AND DOSE MEASUREMENTS. Units of Radioactivity

11/23/2014 RADIATION AND DOSE MEASUREMENTS. Units of Radioactivity CHAPTER 4 RADIATION UNITS RADIATION AND DOSE MEASUREMENTS 1 Units of Radioactivity 2 1 Radiation Units There are specific units for the amount of radiation you receive in a given time and for the total

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma 4.1b - Cavity Theory Lecture 2 Peter R Almond 2011 Overview of Lecture Exposure (W/e) air Exposure and Air Kerma Exposure Exposure is symbolized as X and defined by the ICRU as the quotient of dq by dm,

More information

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

Shielding of Ionising Radiation with the Dosimetry & Shielding Module Shielding of Ionising Radiation with the Dosimetry & Shielding Module J. Magill Overview Biological Effects of Ionising Radiation - Absorber dose, Quality or Weighting Factor, Equivalent Dose Attenuation

More information

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration s of KAERI B.H. Kim 1, J.L. Kim 1, S.Y. Chang 1, J.K. Chang 1, G. Cho 2 1 Korea

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

1. RADIOACTIVITY AND RADIATION PROTECTION

1. RADIOACTIVITY AND RADIATION PROTECTION 1. Radioactivity and radiation protection 1 1. RADIOACTIVITY AND RADIATION PROTECTION Revised August 2011 by S. Roesler and M. Silari (CERN). 1.1. Definitions [1,2] 1.1.1. Physical quantities: Fluence,

More information

Survey Meter OD-01 Address: Phone: Fax: URL:

Survey Meter OD-01 Address: Phone: Fax:   URL: Survey Meter OD-01 Dose meter and dose rate meter for the measurement of the ambient dose and dose rate equivalent H*(10), dh*(10)/dt and the directional dose and dose rate equivalent H'(0.07), dh'(0.07)/dt

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Akira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency

Akira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency The 3rd International Symposium on the System of Radiological Protection Seoul, Korea October 20-22, 2015 Akira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency ICRP Publication

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

Ionization Chamber. Pocket dosimeter

Ionization Chamber. Pocket dosimeter Ionization Chamber Pocket dosimeter 1 Ionization Chamber Pocket dosimeter 2 Radiation Quantities and Units Radiation measurements require specification of the radiation field at various points At the source

More information

LET! (de / dx) 1 Gy= 1 J/kG 1Gy=100 rad. m(kg) dose rate

LET! (de / dx) 1 Gy= 1 J/kG 1Gy=100 rad. m(kg) dose rate Basics of Radiation Dosimetry for the Physicist http://en.wikipedia.org/wiki/ionizing_radiation I. Ionizing radiation consists of subatomic particles or electromagnetic waves that ionize electrons along

More information

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi Accuracy, Calibration, Type Testing and Traceability General [Chapter 6 & 7 of RP 160] Peter Ambrosi EURADOS Training Course Lisbon, Portugal, 18-22 Mai Content Introduction - Aim of Measurement & type

More information

Quantities, Units and Definitions

Quantities, Units and Definitions Quantities, Units and Definitions ICRU 51 - Quantities and Units in Radiation Protection Dosimetry (1993) A. FLUENCE dν Φ da a. dn number of particles incident on a sphere of cross sectional area da b.

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Gy can be used for any type of radiation. Gy does not describe the biological effects of the different radiations.

Gy can be used for any type of radiation. Gy does not describe the biological effects of the different radiations. Absorbed Dose Dose is a measure of the amount of energy from an ionizing radiation deposited in a mass of some material. SI unit used to measure absorbed dose is the gray (Gy). 1J 1 Gy kg Gy can be used

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Radiation Quantities and Units

Radiation Quantities and Units Radiation Quantities and Units George Starkschall, Ph.D. Lecture Objectives Define and identify units for the following: Exposure Kerma Absorbed dose Dose equivalent Relative biological effectiveness Activity

More information

Ion-Chamber Survey Meter OD-02

Ion-Chamber Survey Meter OD-02 Ion-Chamber Survey Meter OD-02 Dose and dose rate meter for measuring the ambient dose equivalent H*(10) and dose rate equivalent dh*(10)/dt as well as the directional dose equivalent H (0,07) and dose

More information

Technical basis of occupational dosimetrygeneral

Technical basis of occupational dosimetrygeneral Technical basis of occupational dosimetrygeneral review Antti Kosunen, Hannu Järvinen STUK 2017 NACP-RPC COURSE Occupational dosimetry in interventional radiology, - cardiology and nuclear medicine 27-29

More information

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3) PROPOSAL FOR EYE-LENS DOSEMETER CALIBRATION AND TYPE TESTING ORAMED WP2 J.-M. Bordy (1), J. Daures (1), M. Denozière (1), G. Gualdrini (2), M. Guijaume (3), E. Carinou (4),F. Vanhavere (5) (1) Laboratoire

More information

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),

More information

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen Principles of applied dosimetry - illustrated by ionometry Lesson FYSKJM4710 Eirik Malinen Ionometry Ionometry: the art of measuring ionizations Number of ionizations proportional to dose Air filled ionization

More information

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Pirchio Rosana a*, Lindner Carlos a, Molina Laura a and Vallejos Matías a. a Comisión

More information

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017 Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018

More information

City University of Hong Kong

City University of Hong Kong City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and

More information

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter Energy response for high-energy neutrons of multi-functional electronic personal dosemeter T. Nunomiya 1, T. Ishikura 1, O. Ueda 1, N. Tsujimura 2,, M. Sasaki 2,, T. Nakamura 1,2 1 Fuji Electric Systems

More information

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of a pencil ionization chamber with and without preamplifier Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

Chapiter VII: Ionization chamber

Chapiter VII: Ionization chamber Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry F. Gutermuth *, S. Beceiro, H. Emling, G. Fehrenbacher, E. Kozlova, T. Radon, T. Aumann, T. Le Bleis, K. Boretzky, H. Johansson,

More information

Chapter 2 Dose Quantities and Units for Radiation Protection

Chapter 2 Dose Quantities and Units for Radiation Protection Chapter 2 Dose Quantities and Units for Radiation Protection S oren Mattsson and Marcus S oderberg 2.1 Introduction In all fields where there is a need for quantitative measurements, it is necessary to

More information

Radiation exposure of personnel during IORT: radiation protection aspects.

Radiation exposure of personnel during IORT: radiation protection aspects. Radiation exposure of personnel during IORT: radiation protection aspects. L. Strigari 1, V. Bruzzaniti 1, V. Landoni 1, A. Soriani 1, S.Teodoli 1, M. Benassi 1 1 Lab. Fisica Medica e Sistemi Esperti,

More information

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter Christopher A Perks *a and Stephane Faugoin a a LANDAUER EUROPE, 33 Avenue du General Leclerc, F-92266, Fontenay-aux-Roses

More information

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005 Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp. 233 237 Advance Access publication 6 October 2005 doi:10.1093/rpd/nci353 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS IN HIGH-ENERGY BREMSSTRAHLUNG

More information

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency Facility of Radiation Standards It is important that radiation measuring instruments are calibrated by

More information

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Recent Activities on Neutron Calibration Fields at FRS of JAERI Recent Activities on Neutron Calibration Fields at FRS of JAERI Michio Yoshizawa, Yoshihiko Tanimura, Jun Saegusa and Makoto Yoshida Department of Health Physics, Japan Atomic Energy Research Institute

More information

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,

More information

11 Gamma Ray Energy and Absorption

11 Gamma Ray Energy and Absorption 11 Gamma Ray Energy and Absorption Before starting this laboratory, we must review the physiological effects and the proper use of the radioactive samples you will be using during the experiment. Physiological

More information

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT) 1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,

More information

Lab #13: MEASUREMENT OF RADIATION DOSES

Lab #13: MEASUREMENT OF RADIATION DOSES Lab #13: MEASUREMENT OF RADIATION DOSES THEORETICAL BACKGROUND In order to estimate the radiation exposure it is essential to determine the quantity of radiation. If we identify the dose of radiation to

More information

5) Measurement of Nuclear Radiation (1)

5) Measurement of Nuclear Radiation (1) 5) Measurement of Nuclear Radiation (1) Registration of interactions between nuclear radiation and matter Universal principle: Measurement of the ionisation Measurement of the ionisation measurement of

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

A new neutron monitor for pulsed fields at high-energy accelerators

A new neutron monitor for pulsed fields at high-energy accelerators A new neutron monitor for pulsed fields at high-energy accelerators Marlies Luszik-Bhadra *, Eike Hohmann Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116, Braunschweig, Germany. Abstract.

More information

Personal Dose Monitoring System

Personal Dose Monitoring System Personal Dose Monitoring System Kei Aoyama Osamu Ueda Takeshi Kawamura 1. Introduction The level of radiation safety control in the nuclear power field and at facilities that utilize radiation has become

More information

Influence of the PMMA and the ISO Slab Phantom for Calibrating Personal Dosemeters

Influence of the PMMA and the ISO Slab Phantom for Calibrating Personal Dosemeters Influence of the PMMA and the ISO Slab Phantom for Calibrating Personal Dosemeters M. Ginjaume, X. Ortega, A. Barbosa Institut de Tècniques nergètiques Universitat Politècnica de Catalunya, Barcelona -

More information

INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5)

INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5) INTRODUCTION TO IONIZING RADIATION (Attix Chapter 1 p. 1-5) Ionizing radiation: Particle or electromagnetic radiation that is capable of ionizing matter. IR interacts through different types of collision

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta Master of Science Thesis Supervisor: Mats Isaksson Contents Abstract...

More information

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Hans-Georg Menzel International Commission on Radiation Units and Measurements (ICRU) The principal objective of ICRU

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Secondary Particles Produced by Hadron Therapy

Secondary Particles Produced by Hadron Therapy Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad

More information

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Georgia Institute of Technology. Radiation Detection & Protection (Day 3) Georgia Institute of Technology The George W. Woodruff School of Mechanical Engineering Nuclear & Radiological Engineering/Medical Physics Program Ph.D. Qualifier Exam Spring Semester 2009 Your ID Code

More information

Physics of Radiography

Physics of Radiography EL-GY 6813 / BE-GY 6203 / G16.4426 Medical Imaging Physics of Radiography Jonathan Mamou and Yao Wang Polytechnic School of Engineering New York University, Brooklyn, NY 11201 Based on Prince and Links,

More information

Neutron Users Club Meeting

Neutron Users Club Meeting Neutron Users Club Meeting 20 October 2015 www.npl.co.uk Neutron Fluence and Dose Standards: Thermal www.npl.co.uk XXWell characterised thermal neutron fields are available at NPL for the calibration

More information

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Reference Dosimetry for Megavoltage Therapy Beams: Electrons Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy

More information

Nuclear Fusion and Radiation

Nuclear Fusion and Radiation Nuclear Fusion and Radiation Lecture 9 (Meetings 23 & 24) Eugenio Schuster schuster@lehigh.edu Mechanical Engineering and Mechanics Lehigh University Nuclear Fusion and Radiation p. 1/42 Radiation Interactions

More information

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from Extensions to the Beta Secondary Standard BSS 2 Consolidated version from www.ptb.de R. Behrens 1 and G. Buchholz 2 Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany

More information

Occupational Radiation Protection at Accelerator Facilities: Challenges

Occupational Radiation Protection at Accelerator Facilities: Challenges Occupational Radiation Protection at Accelerator Facilities: Challenges Haridas.G Health Physics Division Bhabha Atomic Research Centre INDIA Int. Conf. on Occupational Radiation Protection: Enhancing

More information

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES THE PUBLISHING HOUSE PROCEEDINGS OF THE ROMANIAN ACADEMY, Series A, OF THE ROMANIAN ACADEMY Volume 18, Number 1/2017, pp. 42 49 MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES Ion GRUIA University

More information

Measurements with the new PHE Neutron Survey Instrument

Measurements with the new PHE Neutron Survey Instrument Measurements with the new PHE Neutron Survey Instrument Neutron Users Club Meeting National Physical Laboratory 16 th October 2013 Jon Eakins, Rick Tanner and Luke Hager Centre for Radiation, Chemicals

More information

Reliability and Traceability in Radiation Calibration

Reliability and Traceability in Radiation Calibration Reliability and Traceability in Calibration Sadao Nakashima Teruo Ebato 1. Introduction Fuji Electric has a successful history of delivering radiation measuring equipment used in the controlled areas of

More information

a). Air measurements : In-air measurements were carried out at a distance of 82cm from the surface of the light beam

a). Air measurements : In-air measurements were carried out at a distance of 82cm from the surface of the light beam Determination of Backscatter Factors For Diagnostic X-Ray Beams By Experimental And Monte Carlo Methods And Determination Of Air Kerma To Dose Equivalent Conversion Factors For The Calibration Of Personal

More information

2015 Ph.D. Comprehensive Examination III. Radiological Sciences - Medical Physics

2015 Ph.D. Comprehensive Examination III. Radiological Sciences - Medical Physics January 2015 2015 Ph.D. Comprehensive Examination III Radiological Sciences - Medical Physics In this three-hour exam, you are required to answer all of the questions in Part A and any two (2) out of the

More information

Neutron Dosimetry with Ion Chamber-Based DIS System

Neutron Dosimetry with Ion Chamber-Based DIS System Neutron Dosimetry with Ion Chamber-Based DIS System Christian Wernli 1, Annette Fiechtner 1, Jukka Kahilainen 2 1 Paul Scherrer Institute, Villigen, Switzerland 2 RADOS Technology Oy, Turku, Finland INTRODUCTION

More information

2 Units of Radiation Protection

2 Units of Radiation Protection 4 2 Units of Radiation Protection All composed things tend to decay. Buddha 563 483 B. C. A large number of units has been proposed and used in the course of historical development and research in the

More information

Basic physics Questions

Basic physics Questions Chapter1 Basic physics Questions S. Ilyas 1. Which of the following statements regarding protons are correct? a. They have a negative charge b. They are equal to the number of electrons in a non-ionized

More information

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen Non-radioactive radiation sources Lesson FYSKJM4710 Eirik Malinen X-ray tube Electrons are released from the cathode (negative electrode) by thermionic emission accelerated in an evacuated tube hit the

More information

Chapter 9: Calibration of Photon and Electron Beams

Chapter 9: Calibration of Photon and Electron Beams Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

Chapter VIII: Photographic films

Chapter VIII: Photographic films Chapter VIII: Photographic films 1 Photographic films First (integrating) dosimeters to be developed based on photographic techniques dosifilms Dosifilms were more and more replaced by dosimeters using

More information

ORGANISATION EUROPEENNE POUR LA RECHERCHE NUCLEAIRE EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH

ORGANISATION EUROPEENNE POUR LA RECHERCHE NUCLEAIRE EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH ORGANISATION EUROPEENNE POUR LA RECHERCHE NUCLEAIRE EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH Laboratoire Européen pour la Physique des Particules European Laboratory for Particle Physics TECHNICAL NOTE

More information

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio

More information

Radiation Protection & Radiation Therapy

Radiation Protection & Radiation Therapy Radiation Protection & Radiation Therapy For Medical Students Professor of Medical Physics Radiation Units Activity Number disintegrations per second (Curie, Becquerel) Exposure (Roentgen, C/kg) Absorbed

More information

Dose Rate Levels around Industrial Gamma Sources

Dose Rate Levels around Industrial Gamma Sources Dose Rate Levels around Industrial Gamma Sources Héctor René Vega-Carrillo *, Ricardo Rodríguez-Juárez, Eduardo Manzanares- Acuña, Rubén Hernández-Villasana, Jaime Ramírez-González Universidad Autónoma

More information

PECULIARITIES OF FORMING THE RADIATION SITUATION AT AN AREA OF NSC KIPT ACCELERATORS LOCATION

PECULIARITIES OF FORMING THE RADIATION SITUATION AT AN AREA OF NSC KIPT ACCELERATORS LOCATION PECULIARITIES OF FORMING THE RADIATION SITUATION AT AN AREA OF NSC KIPT ACCELERATORS LOCATION A.N. Dovbnya, A.V. Mazilov, M.V. Sosipatrov National Science Center Kharkov Institute of Physics and Technology,

More information

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer SLAC-PUB-8099 July 1999 Air Kerma Calibration Factors and k ch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL) N. E. Ipe*, K. E. Rosser,

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

Accelerators and radiation spectra

Accelerators and radiation spectra X-ray tube Accelerators and radiation spectra Electrons are released from the cathode (negative electrode) by thermionic emission accelerated in an evacuated tube hit the anode (target, positive electrode)

More information

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER SLAC-PUB-95-6749 (March 1995) PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER J. C. Liu, D. Busick 1, K. R. Kase, R. C. McCall 2, R. Sit and H. Tran 3 Stanford Linear

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

Neutron Metrology Activities at CIAE (2009~2010)

Neutron Metrology Activities at CIAE (2009~2010) Neutron Metrology Activities at CIAE (2009~2010) Ionizing Radiation Metrology Division China Institute of Atomic Energy (P.O.Box 275(20), Beijing 102413, China) 1. Neutron calibration fields So far the

More information