Nanoscale High-Frequency Contact Mechanics Using an AFM Tip and a Quartz Crystal Resonator

Size: px
Start display at page:

Download "Nanoscale High-Frequency Contact Mechanics Using an AFM Tip and a Quartz Crystal Resonator"

Transcription

1 3698 Langmuir 2004, 20, Nanoscale High-Frequency Contact Mechanics Using an AFM Tip and a Quartz Crystal Resonator JörnF.Lübben and Diethelm Johannsmann*, Max-Planck-Institute for Polymer Research, Ackermannweg 10, D Mainz, Germany Received December 23, In Final Form: February 16, 2004 The transmission of high-frequency shear stress through a microscopic contact between an AFM tip and an oscillating quartz plate was measured as a function of vertical pressure, amplitude, and surface properties by monitoring the MHz component of the tip s deflection. For dry surfaces, the transmission of shear stress is proportional to the vertical load across the contact. This provides a measure of the forces of adhesion between the substrate and the tip. When stretching soft polymeric fibers created by pulling on the surface of a pressure sensitive adhesive, the transmitted shear stress decreased linearly with extension over the entire range of pulling. This contrasts with the static adhesive force, which remained about constant until it discontinuously dropped at the point of rupture. Introduction The atomic force microscope (AFM) has in recent years evolved into a versatile tool to study the mechanical properties of soft surfaces on the nanoscale. 1 Imaging of mechanical stiffness is possible with the tapping mode 2 or the pulsed force mode, 3 which performs a more sophisticated analysis of the cantilever s trajectory and can extract the adhesive properties of the sample. Numerous workers have used AFM technology without lateral scanning to study mechanical properties on the nanoscale. For example, Rief and co-workers have shown that the mechanical properties of single polymer strands can be probed by pulling on the polymer surface. 4 In another context, colloidal spheres have been attached to the tip and the forces between the sphere and a surface have been determined in detail. 5,6 Overney et al. have emphasized that shear force modulation has certain advantages compared to normal force modulation for surface characterization. 7 The displacement pattern is simpler, and the hydrodynamic coupling is weaker. 8 Also, the force-displacement relation does not suffer from the strong nonlinearities associated with the jump into contact. In principle, these measurements can be carried out with spectroscopic resolution: the effective spring constant can be measured as a complex function of frequency. However, this kind of local mechanical dynamical spectroscopy is not easy. One is either limited to frequencies * Corresponding author. Phone: Fax: johannsmann@pc.tu-clausthal.de. Present address: Institute of Physical Chemistry, TU-Clausthal, Arnold-Sommerfeld-Str. 4, Clausthal-Zellerfeld, Germany. Present address: Swiss Federal Laboratories for Materials Testing and Research, Lerchenfeldstrasse 5, CH-9014 St. Gallen, Switzerland. (1) Sheiko, S. In New Developments in Polymer Analytics; Advances in Polymer Science, Vol. 151; Springer: Berlin, (2) Zhong, Q.; Inniss, D.; Kjoller, K.; Elings, V. B. Surf. Sci. 1993, 290, L688. (3) Rosa-Zeiser, A.; Weilandt, E.; Hild, S.; Marti, O. Meas. Sci. Technol. 1997, 8, (4) Rief, M.; Oesterhelt, F.; Heymann, B.; Gaub, H. E. Science 1997, 275, (5) Ducker, W. A.; Senden, T. J.; Pashley, R. M. Nature 1991, 353, 239. (6) Butt, H. J. Biophys. J. 1991, 60, (7) Overney, R. M.; Sills, S. Phys. Rev. Lett. 2003, 9109, (8) Antognozzi, M.; Szczelkun, M. D.; Humphris, A. D. L.; Miles, M. J. Appl. Phys. Lett. 2003, 82, much below the resonance frequency, or the raw data will be affected by the resonance properties of the cantilever and some scheme of correction has to be found. Burnham and co-workers have pointed out that quantitative dynamic measurements are also possible at frequencies much higher than the resonance frequency of the cantilever. 9,10 If the wavelength of sound is much less than the length of the cantilever beam, the stiffness of the beam is of little importance. The force at the tip is governed by inertia, rather than the cantilever s spring constant. There are again simple relations for analysis. Quartz crystal resonators are attractive in this context because they provide a simple and convenient source for shear excitation. Typical frequencies are in the range of a few MHz to a few hundred MHz, which is much above the typical resonance frequencies of cantilevers. The shear amplitude can be anywhere between 0.1 nm (or less) and 100 nm. The behavior of contacts under high-frequency shear is interesting for a number of reasons. High-frequency tribology is an intermediate step between conventional, low-speed, low-frequency contact mechanics, such as typically performed with the AFM or the surface forces apparatus (SFA), 11 and the practical world, where sliding usually occurs at much higher speed. Quartz crystals perform an oscillatory motion, which is different from steady or rotary sliding. On the other hand, the oscillatory motion allows one to investigate a fixed location on the sample, which is impossible in steady sliding. Bureau et al. have recently applied low-frequency oscillatory stress to a multicontact interface between two rough polymer surfaces. 12 Their work focuses on the transition from static friction to sliding friction. It turns out that the transition is not strictly discontinuous. As predicted by Mindlin, 13,14 there is a range of partial slip, where the rim of the contact zone slips, while the center sticks. The in-phase component of the stress measured by Bureau et al. can be well explained with the Mindlin microslip model. The out-ofphase component, on the other hand, is at variance with (9) Oulevey, F.; Gremaud, G.; Semoroz, A.; Kulik, A. J.; Burnham, N. A.; Dupas, E.; Gourdon, D. Rev. Sci. Instrum. 1998, 69, (10) Burnham, N. A.; Kulik, A. J.; Gremaud, G.; Gallo, P.-J.; Oulevey, F. J. Vac. Sci. Technol., B 1996, 14, 794. (11) Yoshizawa, H.; Chen, Y.-L.; Israelachvili, J. J. Phys. Chem. 1993, 97, (12) Bureau, L.; Caroli, C.; Baumberger, T. In press. (13) Deresiewicz, H. Adv. Appl. Mech. 1958, 5, 233. (14) Mindlin, R. Proceedings of the 2nd U.S. National Congress of Applied Mechanics; ASME: Ann Arbor, 1954; p /la CCC: $ American Chemical Society Published on Web 03/19/2004

2 Nanoscale High-Frequency Contact Mechanics Langmuir, Vol. 20, No. 9, Figure 1. Scheme of measurement. In addition to the quasistatic deflection, the MHz component of the motion is picked with a high-frequency lock-in amplifier. This dynamic signal reflects the transmission of shear stress between the sample and the tip. the model. The authors conclude that the macroscopic laws of friction (and Coulomb s law, in particular) do not apply on the local scale. Our work follows a similar concept but differs in two respects: a single-asperity contact (given by the AFM tip contacting a flat surface) is studied, and the frequency is in the MHz range. In principle, it would be desirable to infer the lateral force from the frequency shift of the quartz resonator induced by the contact with the AFM tip. 15 Unfortunately, the lateral force is too small to induce a frequency shift large enough for quantitative analysis. Kim et al. have been able to acquire AFM images of an oscillating quartz surface and record the frequency shift in parallel to the scanning process, thereby providing a map of the frictional properties of the surface. 16 A tribological contrast between patches of gold and polystyrene could be observed. However, the variations in frequency were only of the order of 1 Hz, which makes detailed mechanical studies difficult. The frequency shift can be increased by increasing the contact radius, for instance, by attaching a colloidal sphere to the tip. 17,18 This, on the other hand, implies a loss of lateral resolution. To probe the interaction more sensitively, we have adopted a different approach: rather than detecting the frequency shift or the tip-induced change in the resonance bandwidth, we monitor the MHz component of the cantilever deflection. Even though this procedure is technically demanding, it provides a direct access to the transmission of shear stress between the substrate and the tip. Instrumental Section Figure 1 shows a sketch of the experimental setup. The sample is coated onto the upper electrode of the resonator. A laser beam is deflected from the back of the tip, passing a knife edge on the way to the detector. The signal from the detector is electronically split into a low-frequency component, providing the quasi-static normal force, and a high-frequency component, providing the response of the cantilever to the MHz shear excitation from below. Our analysis builds on the comparison of static and dynamic forces. Figure 2 shows details of the experimental setup. The output of a laser diode (Thorlabs, S1FC635, power adjustable from 0 to 3 mw, λ ) 635 nm) is fed through a single-mode optical fiber (15) Laschitsch, A.; Johannsmann, D. J. Appl. Phys. 1999, 85, (16) Kim, J. M.; Chang, S. M.; Muramatsu, H. Appl. Phys. Lett. 1999, 74, 466. (17) Berg, S.; Prellberg, T.; Johannsmann, D. Rev. Sci. Instrum. 2003, 74, 118. (18) Borovsky, B.; Mason, B. L.; Krim, J. J. Appl. Phys. 2000, 88, (Thorlabs, P FC-5) for the purpose of beam cleaning. The beam exiting from the fiber is focused onto the aluminum-coated back of the cantilever by a focusing collimator (Linos, MB02). The reflected light is directed to a knife edge which cuts out some of the light, depending on the deflection. We used a PIN-diode (SSO-PDQ TO5, Silicon Sensor). With regard to the size of the beam at the knife edge, a compromise between sensitivity, on one hand, and a large linear range, on the other, has to be found. In the reference state, the knife edge is positioned in the center of beam. As the deflection increases, the knife edge intersects the beam at its periphery. The differential sensitivity with regard to deflection then decreases and eventually goes to zero. The working resistance of the driving electronics was chosen such that the high-frequency cutoff was around 10 MHz. The working resistance together with the capacitance of the photodiode forms a low-pass filter. The detection electronics reads the voltage across the working resistor. Therefore, a large working resistor increases the output voltage but slows down the detector s response at the same time. With the current setting, the sensitivity of the detector decreases with increasing overtone order. One can increase the spectral range at the expense of sensitivity by choosing a smaller working resistance. In most experiments, there also was a low-frequency optical chopper (EG&G, model 197) placed between the knife edge and the detector. A high-frequency lock-in amplifier (Stanford Research Systems, SR844) referenced to the quartz oscillation was used to filter and amplify the raw signal. The analogue output of this amplifier was fed into a second slower lock-in amplifier (Scitec, 500MC) referenced to the optical chopper. This second filtering stage proved necessary to eliminate capacitive crosstalk between the quartz plate and the detector. In most cases, the magnitude (R) from the first lock-in amplifier was used for further amplification. For a complex mode of measurement, the real and the imaginary part of the signal (X and Y) can be analyzed, as well. A third, slow lock-in amplifier (EG&G 5109) was used to measure the low-frequency signal from the detector, referenced to the chopper. This is the static signal, proportional to the static deflection. The resonator (5 MHz blanks, Maxtek, CA) with gold electrodes was mounted in a commercial holder (CHT100, Maxtek), which allows for measurements in air or in liquid. A network analyzer (HP4396A, Agilent) is used to acquire the resonance curves. Resonance frequency and resonance bandwidth are determined by fitting resonance curves to the spectra of the complex admittance of the quartz. Note that the resonance parameters usually do not change upon contact because the tip is so small. Therefore, the experiments can be performed at one fixed frequency on the center of the resonance. We have occasionally checked for out-of-plane components of the motion by placing a droplet of water on the quartz surface and comparing the frequency shift and the half-band-half-width. For a pure shear motion in liquids, one expects the increase in the half-band-half-width to be the same as the decrease in frequency. That expectation was fulfilled within about 20%, whenever we checked. The possibility of a vertical contribution to the motion should be kept in mind. The resonator plate can be coarsely positioned in three dimensions with micrometer screws and an x-y-z stage (OWIS, Stauffen, Germany). Fine approach of the quartz surface to the AFM tip in the z-direction occurs by means of a piezo actuator (P517.3CL, PI, Göttingen). Results and Discussion Figure 3 shows the modulus of the dynamic signal aquired at a fixed distance between the cantilever and the tip while the frequency of the resonator is swept across the resonance. Out of contact, the tip does not respond to the frequency sweep, whereas in contact, the dynamic signal has the same shape as the conductance curve. The conductance is a measure of the shear amplitude: the shear strain is proportional to the current through the electrodes. In contact, the motion of the tip is locked to the motion of the quartz plate.

3 3700 Langmuir, Vol. 20, No. 9, 2004 Lübben and Johannsmann Figure 2. Details of the experimental setup. Figure 3. Response of the modulus of the dynamic deflection to a frequency sweep through the resonance. When tip and sample are out of contact, the dynamic signal is within the noise. In contact, there is a dynamic signal roughly proportional to the instantaneous amplitude of oscillation. Figure 4. Dependence of the dynamic signal on the direction of shear. The tip was approached to the sample three times, and the dynamic signal was recorded as a function of time. For a shear motion parallel to the knife edge (4), the dynamic signal is much smaller than for perpendicular deflection (O). At large deflection, the dynamic signal saturates, resulting in a dip. The shear oscillation of the quartz plate occurs along the crystallographic x-direction. The dynamic signal should therefore change when the resonator is rotated in the sample plane. Figure 4 shows that this is indeed the case. Three approaches of the tip to the sample are displayed. The round dots (O) correspond to a situation in which the deflection was perpendicular to the knife edge, whereas the triangles (4) show data for which the displacement is along the knife edge. In the latter case, the dynamic signal is much smaller. The residual variation Figure 5. Dependence of the dynamic signal on amplitude. The data were taken with a cantilever, where beads with a diameter of 10 µm were glued to the tip. The three approaches were done at different drive levels. The magnitude of the signal is about proportional to the driving voltage, which has been converted to an amplitude of lateral motion as explained in the text. is caused by a slight misalignment. When the displacement is perpendicular to the knife edge (O), the dynamic signal actually saturates for large deflection, resulting in a dip. In Figure 5, we show the dependence of the dynamic signal on the amplitude of oscillation. The amplitude of motion at the peak of the resonance, A, was calculated by the relation 19 A ) (1/π)d 26 UQ, where d 26 ) 3.21 pm/v is the piezoelectric coefficient, U is the driving voltage, and Q is the quality factor. The amplitude of oscillation does not depend on whether contact has been achieved. The disturbance exerted by the tip is so small that neither the frequency nor the Q-factor is affected. This particular data set was taken on a sample in which a glass bead with a diameter of 10 µm had been glued to the tip. For other geometries, we find the same result. As long as the shear amplitude is not too high, there is a linear dependence between the dynamic signal and the shear amplitude. At very high shear amplitudes, the limits of the dynamic (19) Bottom, V. E. Introduction to Quartz Crystal Unit Design; Van Nostrand-Reinhold: New York, 1982.

4 Nanoscale High-Frequency Contact Mechanics Langmuir, Vol. 20, No. 9, range of the detection system are reached. In this particular case, one finds a nonzero dynamical signal at a vertical position of the cantilever where the static signal still does not give evidence of contact. This result was only found when a sphere was glued to the tip. Presumably, there is a transmission of shear stress across the small air gap. Having proven that the instrument in principle works well, we justify the approach on a more fundamental level. It is essential to realize that the quartz plate cannot bend the entire cantilever at a frequency of 5 MHz. Rather, the bending distortions have the character of elastic waves. The wavelength can be estimated from the length of the cantilever and the fundamental resonance frequency. At the fundamental, the length of the cantilever equals about a quarter of the wavelength. Regardless of the size and the shape of the cantilever, it can be assumed that the wavelength scales about inversely with frequency. Using a resonance frequency of the cantilever of 10 khz, one finds that a resonance frequency of the quartz plate of 5 MHz corresponds to a wavelength of the bending waves which is 125 times less than the length of the lever. In principle, these waves may be reflected at the support of the lever, giving rise to standing waves or, equivalently, discrete modes of vibration. Note, however, that the Q-factor of a cantilever in air rarely exceeds 100. Applied to traveling waves, this means that the waves decay over a range of about 100 cycles and, as a consequence, do not return to the tip. In the frequency domain, this implies that the modes are overdamped to the extent that discrete resonances can no longer be seen. The spectrum of the mechanical susceptibility is a broad, featureless continuum. The waves launched by the shear excitation therefore just dissipate the energy; they do not contribute to the elastic stiffness. For the purpose of modeling, the tip can be regarded as a rigid mass in a viscous environment, where the dissipation is achieved by the bending waves traveling along the lever. Having said that the elastic waves in most cases do not reach the mount of the cantilever, one may ask whether they reach the mirror. In other words, if the cantilever is too long to be considered a rigid object, is the pyramidal tip short enough? This question again comes down to a comparison of the dimensions of the tip and the wavelength of sound. Using a speed of sound, v, equal to the speed of sound in a typical solid-state material (v 3000 m/s), a quick calculation shows that the size of the pyramid ( 10 µm) is indeed smaller than the wavelength at 5 MHz by a factor of about 60. We now discuss the dependence of the dynamic signal on the vertical force in more detail. Figure 6 displays the dynamic and the static signal observed when approaching a silicon tip (CSC12, MikroMasch) to the resonator s gold electrode. The static signal exhibits a strong adhesion hysteresis, presumably caused by capillary forces. The experiment was conducted at ambient conditions. We show two out of many more cycles of approach and retraction. The data are well reproducible. Changing the point of contact did not make a difference. Not surprisingly, the dynamic signal strongly increases when contact is being made. We attribute the leveling off at high vertical force to the limited dynamic range of the detection system. Upon retraction, the dynamic signal decreases linearly. It intersects the x-axis at the point of detachment. The linear dependence of the dynamic signal on the deflection suggests that the transmission of shear sound is proportional to the true vertical force at the contact, where the true vertical force is the sum of the external force (as Figure 6. Static and dynamic deflection for a contact between a Si tip and a gold surface as a function of piezo displacement. The static signal displays the familiar feature of an adhesive contact. The dynamic signal discontinuously increases when contact is made but gradually decreases on retraction. There is not the slightest sign of a discontinuity connected to the snap-off. given by the spring constant and the deflection of the cantilever) and the forces of adhesion. A lateral stress proportional to a vertical load is reminiscent of Coulomb s law for sliding friction. In the context of sliding friction, it is argued that the force is transmitted across many small asperities. The sliding force is proportional to the true contact area, which in turn is proportional to the load. Bureau et al. have shown that a similar argument can be applied to the elastic stress in oscillatory shear. 12 They apply the Mindlin model of partial slip 13 on the level of single asperities but extend it in order to take the statistical properties of the rough surface into account. Note, however, that Coulomb friction assumes plastic deformation of small asperities. The local stress is assumed to be so high that protrusions flatten out until the local stress is the same as the yield stress of the material. The fact that we do not observe aging gives evidence against plastic deformation. For a sphere-plate contact deforming elastically, one would expect the stiffness to be proportional to the contact radius, r c. 20 In the Hertz model, the contact radius, in turn, is proportional to the cubic root of vertical load and the effective spring constant should therefore also scale as the cubic root of the vertical force. The Hertz model has been extended to account for forces of adhesion 21 and long-range attraction. 22 However, these extensions do not predict a contact radius proportional to the load. In the following, we discuss different explanations of the fact that the transmission of shear sound is proportional to the vertical force. First, the tip and the surface might be so irregular that we in fact have a multicontact interface. This explanation does not seem plausible because the dependence of the dynamic signal on the vertical load should then be variable. Depending on the details of the geometry, a cubic-root dependence should sometimes be observed. On the contrary, the experiments for a silicon-gold contact are rather reproducible. We have never observed a dynamic signal proportional to the cubic root of the vertical load. A second explanation might be that the tip slides on the surface and therefore samples the statistics of the surface (20) The stiffness is not proportional to the contact area, πr c2, because there is a stress concentration in the contact area which scales as r c -1. (21) Johnson, K. L.; Kendall, K.; Roberts, A. D. Proc. R. Soc. London, Ser. A 1971, 324, 301. (22) Pollock, H. M.; Maugis, D.; Barquins, M. Appl. Phys. Lett. 1978, 33, 798.

5 3702 Langmuir, Vol. 20, No. 9, 2004 Lübben and Johannsmann over time. The Hertz model would not describe such a situation. However, an estimation of the ratio of the shear force and the normal force shows that sliding is unlikely. If we set the lateral force, F, equal to the product of the mass of the tip and the acceleration (F ω 2 a 2 m tip, with a being the shear amplitude) and use a mass of 10-9 g, a shear amplitude of 20 nm, and a frequency of 5 MHz, we arrive at a lateral force of 3 pn. (Note that inertia governs the high-frequency dynamics.) Typical normal forces, on the other hand, are in the nanonewton range. Assuming that the static friction coefficient is of the order of unity, it seems unlikely that sliding sets in. Third, one could argue that the transmission of shear stress is affected by partial slip in the rim of the contact area. For a sphere-plate contact under oscillatory shear, there is an annulus close to the contact line, where the two contacting surfaces slide relative to each other. The width of this rim depends on the ratio of the lateral and the normal force. As the normal force increases, the partial slip is reduced and the total stiffness of the contact increases. Mindlin has predicted the compliance, S, under such conditions of partial slip as 13 S ) 2 - ν 4Kr c ( 1 - F -1/3 ff ) (23) Zong, Y. Ph.D. Thesis, (1) Figure 7. A demonstration of sliding and wear. The sample is covered with a Langmuir-Blodgett film of a polyamic acid. The tip was approached to the sample 15 times. The dynamic (top) and the static signal (bottom) are displayed versus time. Initially the tip slides. Even though the static normal force is appreciable, there is essentially no response at MHz frequencies. After a few approach/retraction cycles, the coupling between the surface and the tip sets in. Figure 8. Static and dynamic deflection acquired during a contact between a Si tip and a sticky polymer surface. Upon retraction, fibers are pulled out of the surface. The transmission of shear stress through the fibers gradually decreases with extension. There is no discontinuity when the fibers finally rupture. where ν is Poisson s number, K is the modulus, r c is the contact radius, f is the friction coefficient in the Coulomb sense, and F and F are the lateral and the normal force, respectively. Given that the lateral force is much smaller than the normal force, eq 1 predicts that the compliance scales about as r c -1, which is in accordance with the Hertz model and at variance with our data. Finally, the sphere-plate geometry may not be a good approximation for the contact between the tip and the surface. The Hertz model only applies as long as the contact radius is much smaller than the radius of the sphere. If the shape of the tip under large vertical force is closer to a flat punch than to a sphere, then there is no stress concentration at the point of contact and the transmitted stress is proportional to the contact area (as opposed to the contact radius). If the contact area would increase linearly with vertical load, this would explain our findings. None of the above arguments can rigorously explain why the stiffness of the contact is proportional to the vertical force. Again, the absence of aging provides evidence against an interpretation in the frame of plastic deformation. In the following, we present two applications of the technique to more complex situations. In a first example, we have examined the contact of a silicon tip with a Langmuir-Blodgett film of 19 layers of the polyamic acid PAA6B. The details of the chemical structure are provided in ref 23. Figure 7 shows the static deflection versus time. The tip has been approached 15 times. The curve displays neither a jump into contact nor capillary forces. Presumably, the adhesion is prevented by the fact that the sample is hydrophobic. A dynamic signal is hardly detectable for the first seven cycles of approach. Starting with the eighth cycle, the signal gradually increases. Clearly, there is wear. The static signal remains entirely unchanged. We interpret our finding in the sense that the tip penetrates the film after a few cycles, which increases the friction coefficient. In passing, we note that the friction coefficient must have been very low initially in order to allow for sliding. Possibly, the friction coefficients at high frequencies are smaller than the quasi-static friction coefficients because the fast motion prevents relaxation. Finally, we show the static and the dynamic signal acquired while retracting the tip from the surface of a pressure sensitive adhesive (Figure 8). As we have shown previously, an AFM tip can pull fibers out the surface. 24 The adhesive is a commercial product from BASF Aktiengesellschaft and mainly consists of a polyacrylate latex. A transmission of shear waves does occur through the fibers. It decreases with elongation, suggesting that the fiber s shear stiffness decreases during stretching. The static force-distance curve displays a plateau. Both the static and the dynamic signal provide the rupture length, which varied between 300 and 1500 nm. For the first rupture event, there is a discontinuity in both curves, but for the latter two pulls the dynamic signal continuously approaches zero. This contrasts to the static signal, which displays a plateau and drops discontinuously at the point of rupture. Conclusions The transmission of high-frequency shear stress through nanoscale contacts has been measured by touching an (24) Dimitrova, T. D.; Johannsmann, D.; Willenbacher, N.; Pfau, A. Langmuir 2003, 19, 5748.

6 Nanoscale High-Frequency Contact Mechanics Langmuir, Vol. 20, No. 9, oscillating quartz plate with an AFM tip and monitoring the MHz component of the tip s deflection. The transmission of shear stress is proportional to the true vertical load at the contact, which is the sum of the external load and the forces of adhesion. We have shown two applications of the instrument to measurements of sliding friction and adhesion. The transmission of shear stress depends much more sensitively on the details of the contact than the normal force. Acknowledgment. This work was supported by the Multifunctional Materials and Miniaturized Devices Center (BMBF 03N 6500). We also acknowledge help by T. Dimitrova, F. Benmouna, W. Zhehui, Y. Zong, and S. Berg. The Langmuir-Blodgett film was provided by Y. Zong. The latex dispersions were kindly provided by BASF AG. LA

Chapter 2 Correlation Force Spectroscopy

Chapter 2 Correlation Force Spectroscopy Chapter 2 Correlation Force Spectroscopy Correlation Force Spectroscopy: Rationale In principle, the main advantage of correlation force spectroscopy (CFS) over onecantilever atomic force microscopy (AFM)

More information

A simple equation predicting the amplitude of motion of quartz crystal resonators

A simple equation predicting the amplitude of motion of quartz crystal resonators JOURNAL OF APPLIED PHYSICS 100, 094505 006 A simple equation predicting the amplitude of motion of quartz crystal resonators Diethelm Johannsmann a Institute of Physical Chemistry, Clausthal University

More information

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM)

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM) Basic Laboratory Materials Science and Engineering Atomic Force Microscopy (AFM) M108 Stand: 20.10.2015 Aim: Presentation of an application of the AFM for studying surface morphology. Inhalt 1.Introduction...

More information

Instrumentation and Operation

Instrumentation and Operation Instrumentation and Operation 1 STM Instrumentation COMPONENTS sharp metal tip scanning system and control electronics feedback electronics (keeps tunneling current constant) image processing system data

More information

Scanning Force Microscopy

Scanning Force Microscopy Scanning Force Microscopy Roland Bennewitz Rutherford Physics Building 405 Phone 398-3058 roland.bennewitz@mcgill.ca Scanning Probe is moved along scan lines over a sample surface 1 Force Microscopy Data

More information

A General Equation for Fitting Contact Area and Friction vs Load Measurements

A General Equation for Fitting Contact Area and Friction vs Load Measurements Journal of Colloid and Interface Science 211, 395 400 (1999) Article ID jcis.1998.6027, available online at http://www.idealibrary.com on A General Equation for Fitting Contact Area and Friction vs Load

More information

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity MECH 373 Instrumentation and Measurements Lecture 19 Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity Measuring Accepleration and

More information

Atomic Force Microscopy imaging and beyond

Atomic Force Microscopy imaging and beyond Atomic Force Microscopy imaging and beyond Arif Mumtaz Magnetism and Magnetic Materials Group Department of Physics, QAU Coworkers: Prof. Dr. S.K.Hasanain M. Tariq Khan Alam Imaging and beyond Scanning

More information

Is Partial Slip Under Transverse Oscillatory Loading a Transient Event?

Is Partial Slip Under Transverse Oscillatory Loading a Transient Event? Is Partial Slip Under Transverse Oscillatory Loading a Transient Event? Frederick Meyer, 1 Daniel Forchheimer, 2 Arne Langhoff, 1 Diethelm Johannsmann* 1 1 Institute of Physical Chemistry, TU Clausthal,

More information

FEM-SIMULATIONS OF VIBRATIONS AND RESONANCES OF STIFF AFM CANTILEVERS

FEM-SIMULATIONS OF VIBRATIONS AND RESONANCES OF STIFF AFM CANTILEVERS FEM-SIMULATIONS OF VIBRATIONS AND RESONANCES OF STIFF AFM CANTILEVERS Kai GENG, Ute RABE, Sigrun HIRSEKORN Fraunhofer Institute for Nondestructive Testing (IZFP); Saarbrücken, Germany Phone: +49 681 9302

More information

Enrico Gnecco Department of Physics. University of Basel, Switzerland

Enrico Gnecco Department of Physics. University of Basel, Switzerland AOSCIECES AD AOTECHOOGIES anotribology - Enrico Gnecco AOTRIBOOGY Enrico Gnecco Department of Physics. University of Basel, Switzerland Keywords: Atomic stick-slip, Friction force microscopy, oad dependence

More information

Contents. Preface XI Symbols and Abbreviations XIII. 1 Introduction 1

Contents. Preface XI Symbols and Abbreviations XIII. 1 Introduction 1 V Contents Preface XI Symbols and Abbreviations XIII 1 Introduction 1 2 Van der Waals Forces 5 2.1 Van der Waals Forces Between Molecules 5 2.1.1 Coulomb Interaction 5 2.1.2 Monopole Dipole Interaction

More information

Lecture 4 Scanning Probe Microscopy (SPM)

Lecture 4 Scanning Probe Microscopy (SPM) Lecture 4 Scanning Probe Microscopy (SPM) General components of SPM; Tip --- the probe; Cantilever --- the indicator of the tip; Tip-sample interaction --- the feedback system; Scanner --- piezoelectric

More information

Scanning Probe Microscopy. Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010

Scanning Probe Microscopy. Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010 Scanning Probe Microscopy Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010 Scanning Probe Microscopy High-Resolution Surface Analysis

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Supplementary Information for Manuscript: Nanoscale wear as a stress-assisted chemical reaction Supplementary Methods For each wear increment, the diamond indenter was slid laterally relative to the silicon

More information

! Importance of Particle Adhesion! History of Particle Adhesion! Method of measurement of Adhesion! Adhesion Induced Deformation

! Importance of Particle Adhesion! History of Particle Adhesion! Method of measurement of Adhesion! Adhesion Induced Deformation ! Importance of Particle Adhesion! History of Particle Adhesion! Method of measurement of Adhesion! Adhesion Induced Deformation! JKR and non-jkr Theory! Role of Electrostatic Forces! Conclusions Books:

More information

Lecture 12: Biomaterials Characterization in Aqueous Environments

Lecture 12: Biomaterials Characterization in Aqueous Environments 3.051J/20.340J 1 Lecture 12: Biomaterials Characterization in Aqueous Environments High vacuum techniques are important tools for characterizing surface composition, but do not yield information on surface

More information

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements MECH 373 Instrumentation and Measurements Lecture 20 Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature 1 Measuring Acceleration and Vibration Accelerometers using

More information

Features of static and dynamic friction profiles in one and two dimensions on polymer and atomically flat surfaces using atomic force microscopy

Features of static and dynamic friction profiles in one and two dimensions on polymer and atomically flat surfaces using atomic force microscopy Features of static and dynamic friction profiles in one and two dimensions on polymer and atomically flat surfaces using atomic force microscopy Author Watson, Gregory, Watson, Jolanta Published 008 Journal

More information

Finite Element Analysis of Piezoelectric Cantilever

Finite Element Analysis of Piezoelectric Cantilever Finite Element Analysis of Piezoelectric Cantilever Nitin N More Department of Mechanical Engineering K.L.E S College of Engineering and Technology, Belgaum, Karnataka, India. Abstract- Energy (or power)

More information

CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS

CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS 61 CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS 4.1 INTRODUCTION The analysis of cantilever beams of small dimensions taking into the effect of fringing fields is studied and

More information

COURSE OUTLINE. Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors. Sensors, Signals and Noise 1

COURSE OUTLINE. Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors. Sensors, Signals and Noise 1 Sensors, Signals and Noise 1 COURSE OUTLINE Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors Piezoelectric Force Sensors 2 Piezoelectric Effect and Materials Piezoelectric

More information

Introductory guide to measuring the mechanical properties of nanoobjects/particles

Introductory guide to measuring the mechanical properties of nanoobjects/particles Jeremias Seppä MIKES Metrology, VTT Technical Research Centre of Finland Ltd P.O. Box 1000, FI-02044 VTT, Finland Contents: AFM Cantilever calibration F-d curves and cantilever bending Hitting the particles

More information

Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors

Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors Frank Ceballos 1, Ming-Gang Ju 2 Samuel D. Lane 1, Xiao Cheng Zeng 2 & Hui Zhao 1 1 Department of Physics and Astronomy,

More information

Imaging Methods: Scanning Force Microscopy (SFM / AFM)

Imaging Methods: Scanning Force Microscopy (SFM / AFM) Imaging Methods: Scanning Force Microscopy (SFM / AFM) The atomic force microscope (AFM) probes the surface of a sample with a sharp tip, a couple of microns long and often less than 100 Å in diameter.

More information

Lecture Note October 1, 2009 Nanostructure characterization techniques

Lecture Note October 1, 2009 Nanostructure characterization techniques Lecture Note October 1, 29 Nanostructure characterization techniques UT-Austin PHYS 392 T, unique # 5977 ME 397 unique # 1979 CHE 384, unique # 151 Instructor: Professor C.K. Shih Subjects: Applications

More information

Supplementary Figure 1: SAW transducer equivalent circuit

Supplementary Figure 1: SAW transducer equivalent circuit Supplementary Figure : SAW transducer equivalent circuit Supplementary Figure : Radiation conductance and susceptance of.6um IDT, experiment & calculation Supplementary Figure 3: Calculated z-displacement

More information

Chapter 2 Surface Acoustic Wave Motor Modeling and Motion Control

Chapter 2 Surface Acoustic Wave Motor Modeling and Motion Control Chapter 2 Surface Acoustic Wave Motor Modeling and Motion Control 1 Abstract For miniaturization of ultrasonic transducers, a surface acoustic wave device has an advantage in rigid mounting and high-power-density

More information

Friction Drive Simulation of a SAW Motor with Slider Surface Texture Variation

Friction Drive Simulation of a SAW Motor with Slider Surface Texture Variation Advances in Science and Technology Vol. 54 (28) pp 366-371 online at http://www.scientific.net (28) Trans Tech Publications, Switzerland Online available since 28/Sep/2 Friction Drive Simulation of a SAW

More information

Institute for Electron Microscopy and Nanoanalysis Graz Centre for Electron Microscopy

Institute for Electron Microscopy and Nanoanalysis Graz Centre for Electron Microscopy Institute for Electron Microscopy and Nanoanalysis Graz Centre for Electron Microscopy Micromechanics Ass.Prof. Priv.-Doz. DI Dr. Harald Plank a,b a Institute of Electron Microscopy and Nanoanalysis, Graz

More information

STM: Scanning Tunneling Microscope

STM: Scanning Tunneling Microscope STM: Scanning Tunneling Microscope Basic idea STM working principle Schematic representation of the sample-tip tunnel barrier Assume tip and sample described by two infinite plate electrodes Φ t +Φ s =

More information

1 Force Sensing. Lecture Notes. 1.1 Load Cell. 1.2 Stress and Strain

1 Force Sensing. Lecture Notes. 1.1 Load Cell. 1.2 Stress and Strain Lecture Notes 1 Force Sensing 1.1 Load Cell A Load Cell is a structure which supports the load and deflects a known amount in response to applied forces and torques. The deflections are measured to characterize

More information

Measuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other methods

Measuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other methods INSTITUTE OF PHYSICS PUBLISHING Nanotechnology 3 (2002) 33 37 NANOTECHNOLOGY PII: S0957-4484(02)27598-4 Measuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other

More information

3.052 Nanomechanics of Materials and Biomaterials Thursday 02/15/07 Prof. C. Ortiz, MIT-DMSE I LECTURE 4: FORCE-DISTANCE CURVES

3.052 Nanomechanics of Materials and Biomaterials Thursday 02/15/07 Prof. C. Ortiz, MIT-DMSE I LECTURE 4: FORCE-DISTANCE CURVES I LECTURE 4: FORCE-DISTANCE CURVES Outline : LAST TIME : ADDITIONAL NANOMECHANICS INSTRUMENTATION COMPONENTS... 2 PIEZOS TUBES : X/Y SCANNING... 3 GENERAL COMPONENTS OF A NANOMECHANICAL DEVICE... 4 HIGH

More information

Foundations of Ultraprecision Mechanism Design

Foundations of Ultraprecision Mechanism Design Foundations of Ultraprecision Mechanism Design S.T. Smith University of North Carolina at Charlotte, USA and D.G. Chetwynd University of Warwick, UK GORDON AND BREACH SCIENCE PUBLISHERS Switzerland Australia

More information

3 Mechanical Diode-Based Ultrasonic Atomic Force Microscopies

3 Mechanical Diode-Based Ultrasonic Atomic Force Microscopies 3 Mechanical Diode-Based Ultrasonic Atomic Force Microscopies M. Teresa Cuberes Abstract. Recent advances in mechanical diode-based ultrasonic force microscopy techniques are reviewed. The potential of

More information

EE 5344 Introduction to MEMS CHAPTER 6 Mechanical Sensors. 1. Position Displacement x, θ 2. Velocity, speed Kinematic

EE 5344 Introduction to MEMS CHAPTER 6 Mechanical Sensors. 1. Position Displacement x, θ 2. Velocity, speed Kinematic I. Mechanical Measurands: 1. Classification of main types: EE 5344 Introduction MEMS CHAPTER 6 Mechanical Sensors 1. Position Displacement x, θ. Velocity, speed Kinematic dx dθ v =, = ω 3. Acceleration

More information

Measurement of liquid surface properties by laser-induced surface deformation spectroscopy

Measurement of liquid surface properties by laser-induced surface deformation spectroscopy PHYSICAL REVIEW E, VOLUME 63, 046302 Measurement of liquid surface properties by laser-induced surface deformation spectroscopy K. Sakai, D. Mizuno, and K. Takagi Institute of Industrial Science, University

More information

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

Lorentz Contact Resonance for viscoelastic measurements of polymer blends The nanoscale spectroscopy company The world leader in nanoscale IR spectroscopy Lorentz Contact Resonance for viscoelastic measurements of polymer blends Lorentz Contact Resonance (LCR) reliably compares

More information

DESIGN AND APPLICATION

DESIGN AND APPLICATION III. 3.1 INTRODUCTION. From the foregoing sections on contact theory and material properties we can make a list of what properties an ideal contact material would possess. (1) High electrical conductivity

More information

MAS.836 PROBLEM SET THREE

MAS.836 PROBLEM SET THREE MAS.836 PROBLEM SET THREE FSR, Strain Gauge, and Piezo Circuits: The purpose of this problem set is to familiarize yourself with the most common forms of pressure and force measurement. The circuits you

More information

Stator/Rotor Interface Analysis for Piezoelectric Motors

Stator/Rotor Interface Analysis for Piezoelectric Motors Stator/Rotor Interface Analysis for Piezoelectric Motors K Harmouch, Yves Bernard, Laurent Daniel To cite this version: K Harmouch, Yves Bernard, Laurent Daniel. Stator/Rotor Interface Analysis for Piezoelectric

More information

NIS: what can it be used for?

NIS: what can it be used for? AFM @ NIS: what can it be used for? Chiara Manfredotti 011 670 8382/8388/7879 chiara.manfredotti@to.infn.it Skype: khiaram 1 AFM: block scheme In an Atomic Force Microscope (AFM) a micrometric tip attached

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION 1. Supplementary Methods Characterization of AFM resolution We employed amplitude-modulation AFM in non-contact mode to characterize the topography of the graphene samples. The measurements were performed

More information

Active elastomer components based on dielectric elastomers

Active elastomer components based on dielectric elastomers Gummi Fasern Kunststoffe, 68, No. 6, 2015, pp. 412 415 Active elastomer components based on dielectric elastomers W. Kaal and S. Herold Fraunhofer Institute for Structural Durability and System Reliability

More information

Fine adhesive particles A contact model including viscous damping

Fine adhesive particles A contact model including viscous damping Fine adhesive particles A contact model including viscous damping CHoPS 2012 - Friedrichshafen 7 th International Conference for Conveying and Handling of Particulate Solids Friedrichshafen, 12 th September

More information

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

Lorentz Contact Resonance for viscoelastic measurements of polymer blends The world leader in nanoscale IR spectroscopy for viscoelastic measurements of polymer blends (LCR) reliably compares viscoleastic properties with nanoscale spatial resolution With no moving parts in the

More information

Characterization of MEMS Devices

Characterization of MEMS Devices MEMS: Characterization Characterization of MEMS Devices Prasanna S. Gandhi Assistant Professor, Department of Mechanical Engineering, Indian Institute of Technology, Bombay, Recap Characterization of MEMS

More information

AFM Imaging In Liquids. W. Travis Johnson PhD Agilent Technologies Nanomeasurements Division

AFM Imaging In Liquids. W. Travis Johnson PhD Agilent Technologies Nanomeasurements Division AFM Imaging In Liquids W. Travis Johnson PhD Agilent Technologies Nanomeasurements Division Imaging Techniques: Scales Proteins 10 nm Bacteria 1μm Red Blood Cell 5μm Human Hair 75μm Si Atom Spacing 0.4nm

More information

ACOUSTIC EMISSION CHARACTERISTICS OF SURFACE FRICTION IN BIO-MEDICAL APPLICATION

ACOUSTIC EMISSION CHARACTERISTICS OF SURFACE FRICTION IN BIO-MEDICAL APPLICATION ACOUSTIC EMISSION CHARACTERISTICS OF SURFACE FRICTION IN BIO-MEDICAL APPLICATION D. PREVOROVSKY 1, Z. PREVOROVSKY 1, J. ASSERIN 2, D. VARCHON 3 1 Institute of Thermomechanics AS CR, Czech Republic; 2 EVIC

More information

Material Anisotropy Revealed by Phase Contrast in Intermittent Contact Atomic Force Microscopy

Material Anisotropy Revealed by Phase Contrast in Intermittent Contact Atomic Force Microscopy University of Pennsylvania ScholarlyCommons Departmental Papers (MEAM) Department of Mechanical Engineering & Applied Mechanics 5-17-2002 Material Anisotropy Revealed by Phase Contrast in Intermittent

More information

Effect of sliding velocity on capillary condensation and friction force in a nanoscopic contact

Effect of sliding velocity on capillary condensation and friction force in a nanoscopic contact Materials Science and Engineering C 26 (2006) 751 755 www.elsevier.com/locate/msec Effect of sliding velocity on capillary condensation and friction force in a nanoscopic contact Pierre-Emmanuel Mazeran

More information

DYNAMIC ROTARY TORQUE MEASUREMENT USING SURFACE ACOUSTIC WAVES

DYNAMIC ROTARY TORQUE MEASUREMENT USING SURFACE ACOUSTIC WAVES DYNAMIC ROTARY TORQUE MEASUREMENT USING SURFACE ACOUSTIC WAVES Abstract A. Lonsdale Technical Director Sensor Technology Ltd The subject of torque measurement has previously been addressed extensively.

More information

Outline. 4 Mechanical Sensors Introduction General Mechanical properties Piezoresistivity Piezoresistive Sensors Capacitive sensors Applications

Outline. 4 Mechanical Sensors Introduction General Mechanical properties Piezoresistivity Piezoresistive Sensors Capacitive sensors Applications Sensor devices Outline 4 Mechanical Sensors Introduction General Mechanical properties Piezoresistivity Piezoresistive Sensors Capacitive sensors Applications Introduction Two Major classes of mechanical

More information

Atomic and molecular interactions. Scanning probe microscopy.

Atomic and molecular interactions. Scanning probe microscopy. Atomic and molecular interactions. Scanning probe microscopy. Balázs Kiss Nanobiotechnology and Single Molecule Research Group, Department of Biophysics and Radiation Biology 27. November 2013. 2 Atomic

More information

7.Piezoelectric, Accelerometer and Laser Sensors

7.Piezoelectric, Accelerometer and Laser Sensors 7.Piezoelectric, Accelerometer and Laser Sensors 7.1 Piezoelectric sensors: (Silva p.253) Piezoelectric materials such as lead-zirconate-titanate (PZT) can generate electrical charge and potential difference

More information

The molecules that will be studied with this device will have an overall charge of

The molecules that will be studied with this device will have an overall charge of The Basics of the Rotation of Polarized Light The molecules that will be studied with this device will have an overall charge of zero but will have localized polarities that can be used to orient the molecule.

More information

Vertical bounce of two vertically aligned balls

Vertical bounce of two vertically aligned balls Vertical bounce of two vertically aligned balls Rod Cross a Department of Physics, University of Sydney, Sydney NSW 2006, Australia Received 26 April 2007; accepted 14 July 2007 When a tennis ball rests

More information

2.76/2.760 Multiscale Systems Design & Manufacturing

2.76/2.760 Multiscale Systems Design & Manufacturing 2.76/2.760 Multiscale Systems Design & Manufacturing Fall 2004 MOEMS Devices for Optical communications system Switches and micromirror for Add/drops Diagrams removed for copyright reasons. MOEMS MEMS

More information

Measurement of hardness, surface potential, and charge distribution with dynamic contact mode electrostatic force microscope

Measurement of hardness, surface potential, and charge distribution with dynamic contact mode electrostatic force microscope REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 70, NUMBER 3 MARCH 1999 Measurement of hardness, surface potential, and charge distribution with dynamic contact mode electrostatic force microscope J. W. Hong,

More information

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY SECOND EDITION C. JULIAN CHEN Department of Applied Physics and Applied Mathematics, Columbia University, New York OXFORD UNIVERSITY PRESS Contents Preface

More information

Piezoelectric Resonators ME 2082

Piezoelectric Resonators ME 2082 Piezoelectric Resonators ME 2082 Introduction K T : relative dielectric constant of the material ε o : relative permittivity of free space (8.854*10-12 F/m) h: distance between electrodes (m - material

More information

AFM for Measuring Surface Topography and Forces

AFM for Measuring Surface Topography and Forces ENB 2007 07.03.2007 AFM for Measuring Surface Topography and Forces Andreas Fery Scanning Probe : What is it and why do we need it? AFM as a versatile tool for local analysis and manipulation Dates Course

More information

Determining thermal noise limiting properties of thin films

Determining thermal noise limiting properties of thin films Determining thermal noise limiting properties of thin films Courtney Linn Institute for Gravitational Research University of Glasgow Summer 2011 Abstract In order to make thermally stable mirrors to be

More information

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM Module 26: Atomic Force Microscopy Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM 1 The AFM apart from generating the information about the topography of the sample features can be used

More information

Supporting Information

Supporting Information Supporting Information Analysis Method for Quantifying the Morphology of Nanotube Networks Dusan Vobornik*, Shan Zou and Gregory P. Lopinski Measurement Science and Standards, National Research Council

More information

Measuring nanomechanical properties of a dynamic contact using an indenter probe and quartz crystal microbalance

Measuring nanomechanical properties of a dynamic contact using an indenter probe and quartz crystal microbalance JOURNAL OF APPLIED PHYSICS VOLUME 90, NUMBER 1 15 DECEMBER 001 Measuring nanomechanical properties of a dynamic contact using an indenter probe and quartz crystal microbalance B. Borovsky a) and J. Krim

More information

Observing the Doppler Absorption of Rubidium Using a Tunable Laser Diode System

Observing the Doppler Absorption of Rubidium Using a Tunable Laser Diode System Observing the Doppler Absorption of Rubidium Using a Tunable Laser Diode System Ryan Prenger 5/5/00 Final Submission Purdue University Physics Department Abstract Using a tunable laser diode, Doppler absorption

More information

Arbitrary Normal and Tangential Loading Sequences for Circular Hertzian Contact

Arbitrary Normal and Tangential Loading Sequences for Circular Hertzian Contact Arbitrary Normal and Tangential Loading Sequences for Circular Hertzian Contact Philip P. Garland 1 and Robert J. Rogers 2 1 School of Biomedical Engineering, Dalhousie University, Canada 2 Department

More information

CHAPTER 6 FRICTION AND WEAR ANALYSIS FOR BUSHING

CHAPTER 6 FRICTION AND WEAR ANALYSIS FOR BUSHING CHAPTER 6 FRICTION AND WEAR ANALYSIS FOR BUSHING 6.1 TEST RIG SETUP FOR THE FRICTION AND WEAR ANALYSIS Knowing the frictional coefficient is important for the determination of wear loss and power loss

More information

Atomic force microscopy study of polypropylene surfaces treated by UV and ozone exposure: modification of morphology and adhesion force

Atomic force microscopy study of polypropylene surfaces treated by UV and ozone exposure: modification of morphology and adhesion force Ž. Applied Surface Science 144 145 1999 627 632 Atomic force microscopy study of polypropylene surfaces treated by UV and ozone exposure: modification of morphology and adhesion force H.-Y. Nie ), M.J.

More information

Measurements of interaction forces in (biological) model systems

Measurements of interaction forces in (biological) model systems Measurements of interaction forces in (biological) model systems Marina Ruths Department of Chemistry, UMass Lowell What can force measurements tell us about a system? Depending on the technique, we might

More information

RP 2.7. SEG/Houston 2005 Annual Meeting 1525

RP 2.7. SEG/Houston 2005 Annual Meeting 1525 Manika Prasad, Ronny Hofmann, Mike Batzle, Colorado School of Mines; M. Kopycinska-Müller, U. Rabe, and W. Arnold, Fraunhofer Institute for Nondestructive Testing, IZFP, Saarbrücken Summary Seismic wave

More information

LIGHT CONTROLLED PHOTON TUNNELING. University of Maryland, College Park, MD Phone: , Fax: ,

LIGHT CONTROLLED PHOTON TUNNELING. University of Maryland, College Park, MD Phone: , Fax: , LIGHT CONTROLLED PHOTON TUNNELING Igor I. Smolyaninov 1), Anatoliy V. Zayats 2), and Christopher C. Davis 1) 1) Department of Electrical and Computer Engineering University of Maryland, College Park, MD

More information

2d-Laser Cantilever Anemometer

2d-Laser Cantilever Anemometer 2d-Laser Cantilever Anemometer Introduction Measuring principle Calibration Design Comparative measurement Contact: Jaroslaw Puczylowski University of Oldenburg jaroslaw.puczylowski@forwind.de Introduction

More information

Investigation of Dynamical and Static Operation Modes of Atomic Force Microscope

Investigation of Dynamical and Static Operation Modes of Atomic Force Microscope University of Szeged Department of Optics and Quantum Electronics Investigation of Dynamical and Static Operation Modes of Atomic Force Microscope PhD Thesis written by: János Kokavecz Supervisors: Dr.

More information

AFM Studies of Pristine PCBM Changes Under Light Exposure. Erin Chambers

AFM Studies of Pristine PCBM Changes Under Light Exposure. Erin Chambers AFM Studies of Pristine PCBM Changes Under Light Exposure Erin Chambers Faculty of health, science, and technology Department of engineering and physics 15 cr Krister Svensson Lars Johansson 28 March 2013

More information

Quality Factor Thickness (nm) Quality Factor Thickness (nm) Quality Factor 10

Quality Factor Thickness (nm) Quality Factor Thickness (nm) Quality Factor 10 Oxygen-Terminated Fluorine-Terminated Length = 24 mm Length = 2 mm Length = 1 mm Length = 12 mm Length = 8 mm 8 mm Width, K 12 mm Width, K 1 mm Width, K 8 mm Width, K 12 mm Width, K 1 mm Width, K Supplementary

More information

EFFECT OF STRAIN HARDENING ON ELASTIC-PLASTIC CONTACT BEHAVIOUR OF A SPHERE AGAINST A RIGID FLAT A FINITE ELEMENT STUDY

EFFECT OF STRAIN HARDENING ON ELASTIC-PLASTIC CONTACT BEHAVIOUR OF A SPHERE AGAINST A RIGID FLAT A FINITE ELEMENT STUDY Proceedings of the International Conference on Mechanical Engineering 2009 (ICME2009) 26-28 December 2009, Dhaka, Bangladesh ICME09- EFFECT OF STRAIN HARDENING ON ELASTIC-PLASTIC CONTACT BEHAVIOUR OF A

More information

PIEZOELECTRIC TECHNOLOGY PRIMER

PIEZOELECTRIC TECHNOLOGY PRIMER PIEZOELECTRIC TECHNOLOGY PRIMER James R. Phillips Sr. Member of Technical Staff CTS Wireless Components 4800 Alameda Blvd. N.E. Albuquerque, New Mexico 87113 Piezoelectricity The piezoelectric effect is

More information

Improving the accuracy of Atomic Force Microscope based nanomechanical measurements. Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA

Improving the accuracy of Atomic Force Microscope based nanomechanical measurements. Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA Improving the accuracy of Atomic Force Microscope based nanomechanical measurements Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA How can we improve accuracy in our nanomechanical measurements?

More information

Scanning Force Microscopy II

Scanning Force Microscopy II Scanning Force Microscopy II Measurement modes Magnetic force microscopy Artifacts Lars Johansson 1 SFM - Forces Chemical forces (short range) Van der Waals forces Electrostatic forces (long range) Capillary

More information

HOMOGENEOUS ELECTRORHEOLOGICAL FLUIDS APPLIED TO VIBRATION CONTROL

HOMOGENEOUS ELECTRORHEOLOGICAL FLUIDS APPLIED TO VIBRATION CONTROL HOMOGENEOUS ELECTRORHEOLOGICAL FLUIDS APPLIED TO VIBRATION CONTROL A.K. El Wahed Division of Mechanical Engineering & Mechatronics, Faculty of Engineering and Physical Sciences, University of Dundee, Dundee.

More information

Lecture 11 Friction Lubrication, and Wear

Lecture 11 Friction Lubrication, and Wear Lecture 11 Friction, Lubrication and Wear Definitions Friction force between the interacting surfaces that resists or hinders their relative movement Static friction force to overcome to start movement

More information

AME COMPUTATIONAL MULTIBODY DYNAMICS. Friction and Contact-Impact

AME COMPUTATIONAL MULTIBODY DYNAMICS. Friction and Contact-Impact 1 Friction AME553 -- COMPUTATIONAL MULTIBODY DYNAMICS Friction and Contact-Impact Friction force may be imposed between contacting bodies to oppose their relative motion. Friction force can be a function

More information

Design and Analysis of Various Microcantilever Shapes for MEMS Based Sensing

Design and Analysis of Various Microcantilever Shapes for MEMS Based Sensing ScieTech 014 Journal of Physics: Conference Series 495 (014) 01045 doi:10.1088/174-6596/495/1/01045 Design and Analysis of Various Microcantilever Shapes for MEMS Based Sensing H. F. Hawari, Y. Wahab,

More information

Pacs numbers ( v, Fc, d, Uv, p)

Pacs numbers ( v, Fc, d, Uv, p) Dynamics of the tip-sample interaction in near-field scanning optical microscopy and the implications for shear-force as an accurate distance measure C. L. Jahncke Department of Physics, St. Lawrence University,

More information

Sound Propagation through Media. Nachiketa Tiwari Indian Institute of Technology Kanpur

Sound Propagation through Media. Nachiketa Tiwari Indian Institute of Technology Kanpur Sound Propagation through Media Nachiketa Tiwari Indian Institute of Technology Kanpur LECTURE-13 WAVE PROPAGATION IN SOLIDS Longitudinal Vibrations In Thin Plates Unlike 3-D solids, thin plates have surfaces

More information

(Refer Slide Time: 1: 19)

(Refer Slide Time: 1: 19) Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras Module - 4 Lecture - 46 Force Measurement So this will be lecture

More information

Scanning Tunneling Microscopy

Scanning Tunneling Microscopy Scanning Tunneling Microscopy References: 1. G. Binnig, H. Rohrer, C. Gerber, and Weibel, Phys. Rev. Lett. 49, 57 (1982); and ibid 50, 120 (1983). 2. J. Chen, Introduction to Scanning Tunneling Microscopy,

More information

Notes on Rubber Friction

Notes on Rubber Friction Notes on Rubber Friction 2011 A G Plint Laws of Friction: In dry sliding between a given pair of materials under steady conditions, the coefficient of friction may be almost constant. This is the basis

More information

VIBRATION ANALYSIS OF E-GLASS FIBRE RESIN MONO LEAF SPRING USED IN LMV

VIBRATION ANALYSIS OF E-GLASS FIBRE RESIN MONO LEAF SPRING USED IN LMV VIBRATION ANALYSIS OF E-GLASS FIBRE RESIN MONO LEAF SPRING USED IN LMV Mohansing R. Pardeshi 1, Dr. (Prof.) P. K. Sharma 2, Prof. Amit Singh 1 M.tech Research Scholar, 2 Guide & Head, 3 Co-guide & Assistant

More information

Micro-Rheology Measurements with the NanoTracker

Micro-Rheology Measurements with the NanoTracker Micro-Rheology Measurements with the NanoTracker JPK s NanoTracker optical tweezers system is a versatile high resolution force measurement tool. It is based on the principle of optical trapping and uses

More information

Vibration Studying of AFM Piezoelectric Microcantilever Subjected to Tip-Nanoparticle Interaction

Vibration Studying of AFM Piezoelectric Microcantilever Subjected to Tip-Nanoparticle Interaction Journal of Novel Applied Sciences Available online at www.jnasci.org 2013 JNAS Journal-2013-2-S/806-811 ISSN 2322-5149 2013 JNAS Vibration Studying of AFM Piezoelectric Microcantilever Subjected to Tip-Nanoparticle

More information

Carbonized Electrospun Nanofiber Sheets for Thermophones

Carbonized Electrospun Nanofiber Sheets for Thermophones Supporting Information Carbonized Electrospun Nanofiber Sheets for Thermophones Ali E. Aliev 1 *, Sahila Perananthan 2, John P. Ferraris 1,2 1 A. G. MacDiarmid NanoTech Institute, University of Texas at

More information

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level 3 Basic Phenomenon in Effect in Sensor Operation Sensors Prof. Dr. M. Zahurul Haq zahurul@me.buet.ac.bd http://teacher.buet.ac.bd/zahurul/ Department of Mechanical Engineering Bangladesh University of

More information

Micromechanical Instruments for Ferromagnetic Measurements

Micromechanical Instruments for Ferromagnetic Measurements Micromechanical Instruments for Ferromagnetic Measurements John Moreland NIST 325 Broadway, Boulder, CO, 80305 Phone:+1-303-497-3641 FAX: +1-303-497-3725 E-mail: moreland@boulder.nist.gov Presented at

More information

Mesoscale Science and Technology

Mesoscale Science and Technology Mesoscale Science and Technology Classical Sciences Atomistic & Molecular Sciences Applications: - Lubrication - Photonics - Fuel Cells - Data Storage The realm of the Mesoscale fosters new perceptions

More information

RHK Technology Brief

RHK Technology Brief The Atomic Force Microscope as a Critical Tool for Research in Nanotribology Rachel Cannara and Robert W. Carpick Nanomechanics Laboratory, University of Wisconsin Madison Department of Engineering Physics,

More information

Optimal Design and Evaluation of Cantilever Probe for Multifrequency Atomic Force Microscopy

Optimal Design and Evaluation of Cantilever Probe for Multifrequency Atomic Force Microscopy 11 th World Congress on Structural and Multidisciplinary Optimisation 07 th -12 th, June 2015, Sydney Australia Optimal Design and Evaluation of Cantilever Probe for Multifrequency Atomic Force Microscopy

More information