A compression load cell, with strain gauges, tested as a part of a weighing instrument. Manufacturer : Molen or Precia Molen.
|
|
- Dylan Shelton
- 6 years ago
- Views:
Transcription
1 Test certificate Page 1 of 4 Issued by Hugo de Grootplein EG Dordrecht The Netherlands Notified Body Number 0122 In accordance with Applicant In respect of Paragraph 8.1 of the European Standard on Metrological aspects of non-automatic weighing instruments EN 45501:1992/AC:1993 and by application of the OIML International Recommendation R60 (Edition 1991). The applied error fraction p, meant in the paragraph of the standard is 0.7. Precia Molen B.V. Franse Akker AL Breda The Netherlands A compression load cell, with strain gauges, tested as a part of a weighing instrument. Manufacturer : Molen or Precia Molen Type : CSP-M- t-..-c..-ss/eex(d)/eex(i) Characteristics Maximum capacity (E max ) Accuracy class Maximum number of load cell verification intervals (n max ) Ratio of minimum LC verification interval Y = E max / V min 5, 10, 25, 40, 60 or 100 t C * * * = special type with additional type description MB In the description number TC2194 revision 3 further characteristics are described. Nederlands Meetinstituut Hugo de Grootplein EG Dordrecht Telephone Telefax NMi B.V. (Chamber of Commerce no ) Subsidiary companies: NMi Van Swinden Laboratorium B.V. ( ) ( ) Verispect B.V. ( ) This document is issued under the provision that NMi. B.V. nor its subsidiary companies accept any liability. Reproduction of the complete document is allowed. Parts of the document may only be reproduced after written permission.
2 Test certificate Page 2 of 4 Description and The load cell is described in the description number TC2194 revision 3 and documentation documented in the documentation folder TC2194-3, appertaining to this test certificate. Remarks Summary of the test involved: see Appendix number TC2194 revision 3. This revision test certificate replaces the earlier version(s), including its documentation folder. Dordrecht, 30 June 2008 Ing. C. Oosterman Manager Product Certification
3 Description Page 3 of 4 1 General information about the load cell All properties of the load cell, whether mentioned or not, may not be in conflict with the standard mentioned in the test certificate. 1.1 Essential parts Description Drawing number Rev. Remarks Load cell assembly CSP-M-40-EEx-C Mechanical Outline Dimensions Special CSP-M 10t / 100 t Mechanical Load cell Assy. type CSP Electrical Cable: - The load cell is provided with a 4-wire system; - The cable length has to be approximately 10 meters for capacity of 5 t and 10 t; - The cable length has to be approximately 20 meters for capacity of 25, 40, 60 and 100 t; - The cable should be a shielded cable, the shield is connected to the load cell. 1.2 Essential characteristics Minimum dead load : 0 kg Safe overload : 150 % of E max Rated Output : 2 mv/v Input impedance : 450 Ω ± 4.5 Ω Output impedance : 480 Ω ± 4.8 Ω Recommended excitation : 5-20 V DC/AC Excitation maximum : 25 V DC/AC See also drawing Specifications load cell CSP-M-..t-., drawing number HV Essential shapes The load cell is built according to the drawings: - Load cell assembly CSP-M-40-EEx-C, drawing number ; - Outline Dimensions Special CSP-M 10t / 100 t, drawing number The data plate is secured against removal by sealing or will be destroyed when removed. The data plate mentions at least the information and markings as described in the OIML R60 document. In the countries where it is mandatory the load cell should bear this test certificate number: TC2194. Securing: The connecting cable of the load cell or the junction box is provided with possibility to seal.
4 Appendix Page 4 of 4 Tests performed for this test certificate: Test Institute type, version, remarks Temperature test and repeatability (20, 40, -10 and 20 C) Temperature effect on minimum dead load output (20, 40, -10 and 20 C) CSP-M-25t-SC-C3.5-MB CSP-M-25t-SC-C3.5-MB Creep (20, 40 and 10 C) CSP-M-25t-SC-C3.5-MB Minimum dead load output return (20, 40 and 10 C) CSP-M-25t-SC-C3.5-MB Barometric pressure effects at room temperature CSP-M-25t-SC-C3.5-MB Humidity test CSP-M-25t-SC-C3.5-MB
5 Documentation folder Number TC Page 1 of 1 Description Drawing number Rev. Remarks Load cell assembly CSP-M-40-EEx-C Mechanical Outline Dimensions Special CSP-M 10t / 100 t Mechanical Load cell Assy. type CSP Electrical Specifications load cell CSP-M-..t-. HV
6
7
E v a l u a t i o n C e r t i f i c a t e
E v a l u a t i o n C e r t i f i c a t e Page 1 of 1 Issued by In accordance with Producer WELMEC guide 8.8 General and Administrative Aspects of the Voluntary System of Modular Evaluation of Measuring
More informationE C t y p e - e x a m i n a t i o n C e r t i f i c a t e
E C t y p e - e x a m i n a t i o n C e r t i f i c a t e Page 1 of 1 Issued by Manufacturer NMi Certin B.V., designated and notified by the Netherlands to perform tasks with respect to conformity modules
More information+ + ( indicator ) + (display, printer, memory)
CERTIFICATION OF STRAIN GAUGE LOAD CELL FAMILIES B. Meissner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory 38116 Braunschweig, Germany Abstract: Testing and certification of load
More informationType UB1 Load Cell. Option Stainless steel cable gland
Type UB1 Load Cell IP68 APPROVED Product Description Type UB1 is a stainless steel universal load cell which allows for tension and compression loading. Its complete hermetic sealing makes it suitable
More informationCertification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60
Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Bernd Meißner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory Braunschweig Abstract
More information9.1 Compatibility of modules for nonautomatic weighing instruments (NAWIs) Common requirements
Übersetzung aus: Verwaltungsvorschriften gesetzliches Messwesen (administrative regulations legal metrology) 9.1 Compatibility of modules for nonautomatic weighing instruments (NAWIs) 9.1.1 Common requirements
More informationLoad Cell Approval Field and Laboratory Evaluation Manual
Load Cell Approval Field and Laboratory Evaluation Category: MASS Revision: c Page: 1 of 9 In 2011 Measurement Canada began phasing in requirements for using approved load cells in selected Heavy Capacity
More informationLevel Sensor PanCake PR 6251 and Mounting Kit PR 6051
Level Sensor PanCake PR 6251 and Mounting Kit PR 6051 The space-saving solution for simple applications The sensors of the PR 6251 series have been specially designed for simple weighing of silos and horizontal
More informationSingle point load cells
SP4M Single point load cells Special features Maximum capacities: kg 200 kg Aluminum High ratio of minimum verification Offcenter load compensation Shielded connection cable Explosion protection and other
More informationNEW VSL Conditioning plate!
Calibration of Prover Tanks using a Coriolis Mass Flow Meter as the Master Meter. Is this possible for Pipe Prover calibration too? Erik Smits (fsmits@vsl.nl) 2 nd European Flow Measurement Workshop Sintra,
More informationZ7A. Load cells. Special features. Data sheet. Wiring code
Z7A Load cells Special features - Complies with OIML R60 regulations up to 3000 d - Max. capacities: 500 kg... 10 t - Fulfills EMC requirements in accordance to EN 45 501 - Low transducer height - Robust
More informationLoad Cells. 3/2 Introduction. 3/3 Mounting components 3/3 Introduction
Siemens AG 2018 /2 Introduction / Mounting components / Introduction /4 Single point load cells /4 Overview /5 SIWAREX WL260 SP-S AA /5 - Load cell /6 SIWAREX WL260 SP-S AB /6 - Load cell /7 SIWAREX WL260
More informationBenefits/Application. Very high-cycle fatigue resistant up to 100 % of nominal load. Extremely robust against side forces and bending moments
Data sheet Force transducer Series RF-I (160 kn 4000 kn) Benefits/Application Accuracy class 0,05 For static and dynamic tensile and compressive forces Very high-cycle fatigue resistant up to 100 of nominal
More informationGlossary Innovative Measurement Solutions
Glossary GLOSSARY OF TERMS FOR TRANSDUCERS, LOAD CELLS AND WEIGH MODULES This purpose of this document is to provide a comprehensive, alphabetical list of terms and definitions commonly employed in the
More informationMorehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:
Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration
More informationOIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 50-3 Edition 2014 (E) Continuous totalizing automatic weighing instruments (belt weighers). Part 3: Test report format Instruments de pesage totalisateurs continus à
More informationGuidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018
Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for
More informationType examination and testing of 825 non-automatic weighing indicator according to EN Performed for Cardinal Detecto Scale Manufacturing Company
DELTA Test Report TEST Reg. no. 19 Type examination and testing of 825 non-automatic weighing indicator according to EN 4551 Performed for Cardinal Detecto Scale Manufacturing Company DANAK-19135 Project
More informationTraceable Mass Determination and Uncertainty Calculation
Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT
More informationKINECTRICS NORTH AMERICA INC. TEST REPORT FOR 774-KCMIL 3M TM COMPOSITE CONDUCTOR
KINECTRICS NORTH AMERICA INC. TEST REPORT FOR 774-KCMIL 3M TM COMPOSITE CONDUCTOR Test Name: SUSTAINED LOAD TEST ON COMPRESSION DEADEND FITTING FOR 774-KCMIL 3M TM COMPOSITE CONDUCTOR AT ROOM TEMPERATURE
More informationBenefits/Application. For static and dynamic tensile and compressive forces. Easy assembling, lots of possibilities. Options/Accessories
D a t a s h e e t F o r c e T r a n s d u c e r S e r i e s R F ( 2 5 k N 0 M N ) Benefits/Application For static and dynamic tensile and compressive forces Outstanding overload-tolerance Very high-cycle
More informationRing-Torsion Load Cells RTB
Ring-Torsion Load Cells RTB % PTB & OIML approved as suitable for trade use (up to 6000 d and 7500 d in case of multi-divisional scales) % High accuracy, even for very small utilisation ranges (down to
More informationD a t a s h e e t. F o r c e T r a n s d u c e r. S e r i e s K. ( 4 k N k N ) Benefits/Application. Options/Accessories
D a t a s h e e t F o r c e T r a n s d u c e r S e r i e s K ( k N 6 3 0 k N ) Benefits/Application For static and dynamic tensile and compressive forces Hermetically sealed Very small force application
More informationOIML R 76-2 RECOMMENDATION. Edition 2007 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Non-automatic weighing instruments
INTERNATIONAL RECOMMENDATION OIML R 76-2 Edition 2007 (E) Non-automatic weighing instruments Part 2: Test report format Instruments de pesage à fonctionnement non automatique Partie 2: Format du rapport
More informationINTERNATIONAL OIML R 61-2 RECOMMENDATION
INTERNATIONAL OIML R 61-2 RECOMMENDATION Edition 2004 (E) Automatic gravimetric filling instruments Part 2: Test report format Doseuses pondérales à fonctionnement automatique Partie 2: Format du rapport
More informationED 701 General Industry Pressure Transmitter
ED 701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response
More informationInstallation guide 862 MIT / MIR
Installation guide 862 MIT / MIR in combination with: 864 MTT or 863 MRT or (dual) spot element November 2005 Part no. 4416.232_Rev3 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100
More informationZF Series Force Transducers
Data Sheet Series Force Transducers FEATURES 1 kn to 00 kn High accuracy For tensile or pressure forces Made of stainless steel Enclosure protection IP 67 DESCRIPTION The force transducers of the series
More informationCapaciTorr HV Pumps. making innovation happen,together
CapaciTorr HV Pumps making innovation happen,together CapaciTorr HV HIGHLIGHTS General Features High pumping speed for all active gases High sorption capacity and increased lifetime Constant pumping speed
More informationPerformance characteristics of metallic resistance strain gauges
INTERNATIONAL OIML R 62 RECOMMENDATION Edition 1985 (E) Performance characteristics of metallic resistance strain gauges Caractéristiques de performance des extensomètres métalliques à résistance OIML
More informationHumidity Calibration Solutions
Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified
More informationA comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs
IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. A comparison of primary platinum-iridium kilogram mass
More informationSVENSK STANDARD SS-EN ISO 376. Metalliska material Kalibrering av kraftmätdon avsedda för kontroll av maskiner för enaxlig provning (ISO 376:1999)
SVENSK STANDARD SS-EN ISO 376 Fastställd 2002-02-01 Utgåva 1 Metalliska material Kalibrering av kraftmätdon avsedda för kontroll av maskiner för enaxlig provning (ISO 376:1999) Metallic materials Calibration
More informationHumidity Calibration Solutions
Humidity Calibration Solutions www.michell.com Humidity Calibration Solutions The Importance of Regular Calibration The reliable operation of a hygrometer and indeed any measuring instrument, can only
More informationT h e rm i s t o r s
Data Pack E Issued March 00 - T h e rm i s t o r s NTC thermistors The R S range of NTC thermistors includes standard tolerance negative temperature coefficient thermistors, a range of small close tolerance
More informationINTERNATIONAL OIML R 15 RECOMMENDATION
INTERNATIONAL OIML R 15 RECOMMENDATION Edition 1974 (E) Instruments for measuring the hectolitre mass of cereals Instruments de mesure de la masse à l hectolitre des céréales OIML R 15 Edition 1974 (E)
More informationUnderstanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator
Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment
More informationSpecifications. Performance. Mechanical Properties. Optional Accessories SR50. Saddle (CA 10B) (Option) Mount Base. LCR-B-5KNS7 to 50KNS7
LR--S7 Tension Meter Load ell 5 to 0 kn TRNSURS igh safe overload rating of 0% max. Mechanical stopper activating against overloads of 1% to 0% ermetically sealed structure with inert gas filled in ighly
More informationPC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS
XIX IMEKO World Congress Fundamental and Applied Metrology September 611, 009, Lisbon, Portugal PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS Adriana Vâlcu 1, Sterică Baicu 1 National Institute
More informationSHIVAJI UNIVERSITY, KOLHAPUR
s Seat GROUP III 1. Transistor Series Voltage Regulator With a given rectifier, prepare a regulated power supply, using transistor. Vary output currents changing load resistance. Hence measure output voltages
More informationLoadcell Calibration - Evaluation of Uncertainties
Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of
More informationACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES
ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Asesoría Industrial en Mantenimiento y Calibración
More informationOIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air
INTERNATIONAL DOCUMENT OIML D 28 Edition 2004 (E) Conventional value of the result of weighing in air Valeur conventionnelle du résultat des pesées dans l'air OIML D 28 Edition 2004 (E) ORGANISATION INTERNATIONALE
More informationInfrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
More informationGuidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01
More informationOIML R RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Instruments for measuring the areas of leathers
INTERNATIONAL RECOMMENDATION OIML R 136-2 Edition 2006 (E) Instruments for measuring the areas of leathers Part 2: Test Report Format Instruments pour la mesure de la surface des cuirs Partie 2 : Format
More informationStandard burettes for verification officers
INTERNATIONAL OIML R 41 RECOMMENDATION Edition 1981 (E) Standard burettes for verification officers Burettes étalons pour agents de vérification OIML R 41 Edition 1981 (E) ORGANISATION INTERNATIONALE DE
More informationCheck Homework. Reading Quiz Applications Equations of Equilibrium Example Problems Concept Questions Group Problem Solving Attention Quiz
THREE-DIMENSIONAL FORCE SYSTEMS Today s Objectives: Students will be able to solve 3-D particle equilibrium problems by a) Drawing a 3-D free body diagram, and, b) Applying the three scalar equations (based
More informationAn Introduction to the Differences Between the Two Most Recognized Force Standards
An Introduction to the Differences Between the Two Most Recognized Force Standards Morehouse Instrument Company Introduction Morehouse has been performing both ASTM E74 and ISO 376 calibrations for more
More informationOIML R 51-2 RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Automatic catchweighing instruments
INTERNATIONAL RECOMMENDATION OIML R 51-2 Edition 2006 (E) Automatic catchweighing instruments Part 2: Test report format Instruments de pesage trieurs-étiqueteurs à fonctionnement automatique Partie 2:
More informationThis specification covers the requirements for product performance, test methods and quality assurance provisions of FG Terminal.
DESIGN OBJECTIVES The product described in this document has not been fully tested to ensure conformance to the requirements outlined below. Therefore AMP Ltd makes no representation or warranty, express
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 140-6 Second edition 1998-08-15 Acoustics Measurement of sound insulation in buildings and of building elements Part 6: Laboratory measurements of impact sound insulation of
More informationVX25 Enclosure System. Technical documentation PE conductor connection, current carrying capacity
VX Enclosure System Technical documentation PE conductor connection, current carrying capacity Enclosure system VX Contents Contents. General remarks. Introduction. Notes on the design of the earthing
More information16 Appendix 16 Electrical Research Association Report
Guide to the Wiring Regulations 16 Appendix 16 Electrical Research Association Report (ERA) report on armoured cables with external CPCs The attached report was commissioned by the ECA and is copyright
More informationRenolit stretch ceilings
Müller-BBM GmbH Robert-Koch-Str. 11 82152 Planegg bei München Telephone +49(89)85602 0 Telefax +49(89)85602 111 www.muellerbbm.de Dipl.-Ing. (FH) Eva Müller Telephone +49(89)85602 3206 Eva.Mueller@mbbm.com
More informationIntegrated Circuits Thermal Test Method Environment Conditions - Natural Convection (Still Air)
EIA/JEDEC STANDARD Integrated Circuits Thermal Test Method Environment Conditions - Natural Convection (Still Air) EIA/JESD51-2 DECEMBER 1995 ELECTRONIC INDUSTRIES ASSOCIATION ENGINEERING DEPARTMENT NOTICE
More informationOIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 141 Edition 2008 (E) Procedure for calibration and verification of the main characteristics of thermographic instruments Procédure pour l'étalonnage et la vérification
More informationCapaciTorr HV 200 HIGHLIGHTS
CapaciTorr HV 200 General Features High pumping speed for all active gases High sorption capacity and increased lifetime Costant pumping speed in HV, UHV and XHV Reversible pumping of hydrogen and its
More informationSVENSK STANDARD SS-EN ISO
SVENSK STANDARD SS-EN ISO 376:2005 Fastställd 2005-02-04 Utgåva 2 Metalliska material Kalibrering av kraftmätdon avsedda för kontroll av maskiner för enaxlig provning (ISO 376:2004) Metallic materials
More informationAutomated volume measurement for weihts using acoustic volumeter
IMEKO 20th TC3, 3rd TC16 and 1st TC22 International Conference Cultivating metrological knowledge 27th to 30th November, 2007. Merida, Mexico. Automated volume measurement for weihts using acoustic volumeter
More informationTest Report EN (2008)
Test Report EN 62233 (2008) Product Name and address of the applicant Name and address of the manufacturer Model Rating Brand name Heating elements for building into floors Advanced Heating Technologies
More informationBlower Motor 3P Connector
1. Scope Blower Motor P Connector 1.1 Content This specification covers the requirements for product performance, test methods and quality assurance provisions of Blower Motor P connector. Applicable product
More informationTemperature Compensation for MEAS Pressure Sensors
INTRODUCTION Advancements in microelectronic technology have pushed silicon sensors not only toward greater sophistication and lower functional cost but also in the direction of higher performance. The
More informationORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL RECOMMENDATION
ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL RECOMMENDATION Continuous totalizing automatic weighing instruments (belt weighers) Part 2: Test report format Instruments de pesage totalisateurs
More informationDESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE
DESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE Andy Knott National Physical Laboratory, United Kingdom ABSTRACT This paper describes the development of a 120 kn deadweight
More informationNanometrology and its role in the development of nanotechnology
Nanometrology and its role in the development of nanotechnology Rob Bergmans Nederlands Meetinstituut Van Swinden Laboratorium 1 NMi Van Swinden Laboratorium The Art of Measurement Dutch national metrology
More informationCALYS 80P. Field pressure calibrator with high accuracy 0.01%
Field pressure calibrator with high accuracy 0.01% CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. Description CALYS 80P is a field pressure calibrator within CALYS
More informationPhysics Assessment Unit AS 1
Centre Number 71 Candidate Number ADVANCED SUBSIDIARY (AS) General Certificate of Education January 2013 Physics Assessment Unit AS 1 assessing Module 1: Forces, Energy and Electricity AY111 [AY111] FRIDAY
More information(Round up to the nearest inch.)
Assignment 10 Problem 5.46 LRFD First, select the lightest weight W14 column. Use the recommended design value for K for the pinned-fixed support condition specified (ref. Commentary, Appendix 7, AISC
More informationINVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE
IMEKO 23 rd TC3, 13 th TC5 and 4 th TC22 International Conference 30 May to 1 June, 2017, Helsinki, Finland INVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE
More informationElectrical safety testing Checklist
ety al Check testing Check - Checklist EN) ctrical Electrical safety testing Checklist SOP - Electrical Check Copyright 2013 Eppendorf AG. All rights reserved, including graphics and images. No part of
More informationTEMPERATURE COMPENSATION FOR MEAS PRESSURE SENSORS APPLICATION NOTE
TEMPERATURE COMPENSATION FOR MEAS PRESSURE SENSORS INTRODUCTION Advancements in microelectronic technology have pushed silicon sensors not only toward greater sophistication and lower functional cost but
More informationPRODUCT DATA SHEET. Eagle Eye Outdoor LED Module 5700K. RoHS Compliant. CUSTOMER :. DATE : REV : Rev. 2.0.
CUSTOMER :. DATE : 213. 4. 23. REV : Rev. 2.. PRODUCT DATA SHEET Eagle Eye Outdoor LED Module 57K MODEL NAME : LLDMWW-15K71A (Type II Lens - H:16 V:7 ) LLDMWW-15K72A (Type III Lens - H:135 V:6 ) LLDMWW-15K73A
More informationPRODUCT DATA SHEET. Eagle Eye Outdoor LED Module 3000K. RoHS Compliant. CUSTOMER :. DATE : REV : Rev. 1.0.
CUSTOMER :. DATE : 213. 3. 29. REV : Rev. 1.. PRODUCT DATA SHEET Eagle Eye Outdoor LED Module 3K MODEL NAME : LLDMWW-15K21A (Type II Lens - H:16 V:7 ) LLDMWW-15K22A (Type III Lens - H:135 V:6 ) LLDMWW-15K23A
More informationTEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong
IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES
More informationCalibration Traceability Guidelines
Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation www.a2la.org ts.nist.gov/standards/accreditation/index.cfm
More informationHigh pressure comparison among seven European national laboratories
IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. High pressure comparison among seven European national
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)
More informationMOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER
Centre for Basic, Thermal and Length Metrology National Physical Laboratory MOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER A Watts Precision Clinometer fitted with a circular glass scale
More informationMultiphysics Simulations for the design of a Superconducting magnet for proton therapy
WIR SCHAFFEN WISSEN HEUTE FÜR MORGEN Paul Scherrer Institut Multiphysics Simulations for the design of a Superconducting magnet for proton therapy Ciro Calzolaio October 19, 2017 OUTLINE Superconducting
More informationObtaining a Converged Solution with Abaqus
Obtaining a Converged Solution with Abaqus Day 1 Lecture 1 Workshop 1 Lecture 2 Lecture 3 Workshop 2 Lecture 4 Workshop 3 Lecture 5 Workshop 4 Introduction to Nonlinear FEA Nonlinear Spring Nonlinear FEA
More informationSingle Phase Motors Technical Datasheets
Single Phase Motors Technical Datasheets EN A dynamic, strong and ambitious Group: Orange1 Holding is an international renown Group, one of the most important European manufacturers of single-phase and
More informationAPPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY
APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationPascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure
DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and
More informationPSMN002-25P; PSMN002-25B
PSMN2-25P; PSMN2-25B Rev. 1 22 October 21 Product data 1. Description N-channel logic level field-effect power transistor in a plastic package using TrenchMOS 1 technology. Product availability: PSMN2-25P
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 10848-3 First edition 2006-04-01 Acoustics Laboratory measurement of the flanking transmission of airborne and impact sound between adjoining rooms Part 3: Application to light
More informationEA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE
Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration
More informationCALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%
ATEX documenting multifunction calibrator with high accuracy 0.006% is the top instrument of our range of ATEX documenting multifunction calibrators. Description is the top instrument of our range of ATEX
More informationSHANGHAI MYLCH WINDOWS & DOORS THERMAL PERFORMANCE TEST REPORT
SHANGHAI MYLCH WINDOWS & DOORS THERMAL PERFORMANCE TEST REPORT SCOPE OF WORK DR65 FIXED REPORT NUMBER H3108.01-301-46 R0 TEST DATE 11/09/17 ISSUE DATE 11/14/17 RECORD RETENTION END DATE 11/09/22 PAGES
More informationTOP Transfer. Data Sheet. Force transfer standard. Special characteristics. D 20.TOPTRANS.10 en
TOP Transfer Force transfer standard Data Sheet Special characteristics Ultimate precision force transducer for international comparative measurements Tensile/compressive force transducer Nominal forces
More informationUncertainty Workshop: Overview of uncertainty factors in HTGHPs
Uncertainty Workshop: Overview of uncertainty factors in HTGHPs EMRP Final Stakeholder Meeting NPL, Teddington, United Kingdom 1 DIN/EU Standards for GHPs ISO 830:1991 Thermal insulation Determination
More informationValkPitched - Trapezoidal
ValkPitched - Trapezoidal Installation manual Fixation: Thin sheet screw Use in combination with the Project Report of the ValkPVplanner Portrait configuration: page 2 to 6 Landscape configuration: page
More informationIn data sheets and application notes which still contain NXP or Philips Semiconductors references, use the references to Nexperia, as shown below.
Important notice Dear Customer, On 7 February 27 the former NXP Standard Product business became a new company with the tradename Nexperia. Nexperia is an industry leading supplier of Discrete, Logic and
More information473-SHX Dew Point Mirror
Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable
More informationLoad Washers. Force. for Forces of 7, kn. F z. Type 9001A A 9081B, 9091B
Force Load Washers for Forces of 7,5... 1 200 kn Type 9001A... 9071A 9081B, 9091B 1-component force sensor for measuring dynamic and quasistatic forces in z direction. F z Calibrated measuring range 100
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Compania Nacional de Metrologia S.A.S - Conamet
More informationEA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement
Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating
More informationVerification for Non-Automatic Weighing Instruments (NAWI)
Verification for Non-Automatic Weighing Instruments (NAWI) Every instrument should be accurate and correct to be permitted for use in any commercial transaction, particularly at the Point Of Sale. An instrument
More information