Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration
|
|
- Andrea Dawson
- 6 years ago
- Views:
Transcription
1 Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration Speaker/Author: Frank Liebmann, Fluke Calibration, 799 Utah Valley Dr., American Fork, Utah, 84003, USA , FAX Author: Tom Kolat, Fluke Calibration Abstract Industrial level infrared thermometers are being increasingly designed to measure temperatures above 500 C. A thermal radiation source is needed to calibrate these instruments. The infrared thermometers designed to measure these temperatures generally measure with a smaller field-ofview. This means there is a possibility of using a blackbody cavity as the thermal radiation source. A previous attempt was made to use a cavity mounted inside of a furnace meant for thermocouple calibration for a thermal radiation source. However, when the cavity s emissivity was measured, it was found the emissivity was not constant for different wavelengths and was very dependent on cavity position. This paper discuses a new attempt to mount a cavity inside a thermocouple furnace with much better temperature uniformity. It discusses the measurements made to verify the emissivity of the cavity using industrial radiation thermometers. It also talks about measurements made with industrial handheld infrared thermometers and compares these measurements to measurements made using other thermal radiation sources and the associated measurement uncertainty. Learning Objectives Learn about IR thermometer calibration Be informed about available calibration sources for radiation thermometer Learn about attributes of a blackbody cavity Learn how to qualify a calibration source 1 Introduction In 2008, Fluke Calibration released a series of flat plate infrared calibrators for the calibration of infrared thermometers. The temperature range of the flat plates is -15 C to 500 C. Since the release of these products, there has been increased demand for a calibrator to cover a higher temperature range. Previously, a blackbody cavity was used inside a thermocouple calibration furnace for evaluation purposes. This solution was found to be marginal. This was due to the poor temperature uniformity along the cavity walls. Since then, Fluke Calibration has come out with a new thermocouple calibration furnace with much better uniformity on its walls.
2 2 Background To understand the developments for this new design, one must understand the work that lead up to the research presented in this paper. The aforementioned need for a source was principle in doing this research. There are several commercial solutions on the market that did not meet the need due to issues with traceability and emissivity in the 8 14 µm band. The availability of a more uniform furnace would solve many of these problems. Such a furnace became available, and testing was done to verify that it would be a good solution for the customer. 2.1 Calibration Sources for Radiation Thermometry There are a number of calibration sources for infrared (IR) thermometers. These sources take on one of two forms, a flat-plate thermal radiation source or a cavity thermal radiation source [1]. These sources have their own advantages and disadvantages. The cavity source theoretically provides an emissivity of close to unity throughout the entire electromagnetic spectrum [2]. However, there are two factors that can diminish the effective emissivity. First, this emissivity can be diminished when there is poor temperature uniformity, especially poor temperature uniformity on the cavity walls. Second, this effective emissivity is also diminished when the temperature of the cavity bottom is poorly estimated [3]. This is especially the case when using Scheme I traceability which means traceability comes through contact thermometry [4]. The solution for many laboratories is to calibrate the cavity with a radiation thermometer, otherwise known as Scheme II traceability [4]. This is only valid if the radiation thermometer has the same spectral bandwidth as the IR thermometers to be calibrated using the cavity. In other words, the units under test must have the same spectral bandwidth as the transfer standard. The other limitation with using a cavity is difficulty due to the wide field-of-view of less expensive IR thermometers. The depth of the cavity is often so deep that the IR thermometer field-of-view is larger than the cavity bottom. For these types of measurements a flat plate calibrator is often used [1]. Flat plate calibrators often have a range of up to 500 C. Flat plates with higher temperature ranges are costly. 2.2 Previous attempts Previously a cavity was made using a Fluke Calibration Model 9112B thermocouple calibration furnace [5]. This attempt of using a cavity inside a thermocouple furnace gave marginal results. The poor uniformity of the furnace caused two effects that caused large uncertainty. First, the heat flow between the cavity bottom and the reference probe was large enough that there was a considerable temperature drop between these two locations. Second, the temperature on the cavity walls was poor enough that the effective emissivity of the cavity was not well estimated. This was evident when measuring the cavity at two different wavelengths. This is a problem common to many industrial level cavities. After doing this research, it was decided that this was not a good solution for IR thermometer calibrations above 500 C. 2.3 Availably of a more uniform furnace In 2014, Fluke Calibration released the Model 9118A Thermocouple Calibrator furnace. The furnace has improved axial uniformity over the Model 9112B. It was hoped that this improvement in uniformity would result in a better solution for a blackbody cavity. The one drawback with the 9118A was the diameter of the tube, 40 mm. This was thought not to be large enough for work
3 with IR thermometers. As a result, the 9118A used for this investigation was modified to accommodate a cavity of 50 mm in diameter. 2.4 Description of cavity The cavity is shown in Figure 1 as it is installed in the furnace. The cavity used is a cylindroconical design with an apex angle of 120 [2]. The cavity walls were designed to extend well beyond the cavity bottom. In Figure 1, the cavity is shown with the reference probe in place. The intent was to place the sensor of the reference probe as close as possible to the conical apex of the cavity bottom. The reason for this is to minimize temperature measurement error due to heat transfer between the probe s sensor and the cavity bottom surface [3]. The cavity may be positioned at various axial depths. These axial depths are measured from the reflector plate by the cavity opening to the opening of the cavity walls. These depths are referred to as cavity positions for the rest of this paper. 3 Contact measurements 3.1 Uniformity Figure 1: 9118A blackbody cavity diagram. In order to calculate data needed for the blackbody emissivity model and to estimate uncertainty due to cavity bottom heat exchange, temperature gradient was measured along the cavity axis at cavity positions of 5, 10, and 15 cm and at three cavity temperatures. These results are shown in Figure 2.
4 Stability / K Heat Exchange / K / cm C 660 C 960 C Cavity Position / cm 3.2 Short Term Stability Figure 2: 9118A axial uniformity in blackbody cavity configuration. The short term stability of the 9118A blackbody cavity measured by a radiation thermometer was observed and recorded. The stability measurements took place for a 5 minute period. The stability reported is 2 standard deviations of the samples over this 5 minute period. In Figure 3, the results of the 9118A stability as measured by a radiation thermometer are compared to the Fluke Calibration IR Laboratory (AFL) in American Fork s HT cavity s stability as measured by a radiation thermometer [6][7] Temperature / C µm µm HT 8-14 µm HT 3.9 µm Figure 3: Cavity stability (2σ) as measured by a radiation thermometer.
5 3.3 Traceability of contact measurements The traceability of the contact measurements comes through the American Fork Primary Temperature Laboratory, through NIST, to the Système International d'unités (International System of Units or SI). The traceability of each of the measurements is covered more in the next section. 4 Non-contact measurements In order to test the cavity design, measurements were made on the cavity. These measurements involved comparing the traceable readout temperature of a contact probe located close to the conical apex at the cavity bottom to that of a radiation thermometer measuring the cavity bottom. The radiation thermometers used for the comparison ranged from hand held instruments with thermopile detectors to precision radiation thermometers with pyroelectric detectors [8]. In all measurements two chains of traceability to the SI were maintained. These are the traceability by contact thermometry and traceability by radiation thermometry. In all cases the two measurements were compared using normal equivalence (En) [9][10]. The traceability for each of the measurements and the comparison made is covered in Figure 4. Scheme A Scheme B Scheme C Scheme D Figure 4: Traceability schemes. The measurements were divided into six sets for organizational purposes. Only a portion of the results are reported in this paper due to the sheer number of tests. The data was taken at cavity positions of 5, 10, and 15 cm. Data for each of these positions was consistent. In all but Set 6, the 15 cm cavity position data is reported. In the case of Set 6, the 5 cm cavity position data is reported, since the 10 and 15 cm positions were found to be unusable with one of the infrared thermometers. A summary of the sets of measurements performed with their traceability is shown in Table 1. Details on these measurements are discussed in the following sections.
6 εeff Table 1: Sets of measurements with their traceability scheme. Data Set RT λ / µm Scheme Ref Source 9118A Ref Cav Pos. 1 KT B HT Cav F cm 2 TRTII 3.9 B HT Cav F cm 3 TRTII 8 14, 3.9 C PTB F cm 4 KT B HT Cav F cm 5 F568, 561, D F4181 F cm 6 F62M D F4181 F cm 4.1 Initial measurements Initial measurements were made using an IR thermometer model Fluke 568 compared to a type K thermocouple using traceability Scheme A [11]. Two cavity designs were measured. These are termed as a thin cavity and a thick cavity for the sake of discussion. The thick cavity is the one shown in Figure 1. The measured emissivity of the thin cavity showed a large dependence to its axial position. This was believed to be a result of heat flow in and around the mass of the thin cavity. This design was abandoned and not considered any further. The thick cavity was measured at various axial positions. The results of one of these tests is shown in Figure 5. This testing showed that the design was promising, since the emissivity was maintained close to unity for a wide range of cavity axial positions. This meant that the furnace s temperature uniformity did not adversely influence the emissivity of the cavity as was observed in the previous study Thin PRT No I Thin PRT I Thin TC No I Thin TC I Thick TC No I d / cm Figure 5: Results of initial measurements.
7 4.2 Intercomparison with cavities The next set of tests compared the cavity s temperature as measured by a radiation thermometer with a pyroelectric detector to that of a thermocouple Fluke Model 5650 shown as Sets 1 and 2 in Table 1. Additional measurements were made using a Fluke Model 5628 PRT as a reference. These measurements are shown as Set 4. It should be noted that the measurements using a PRT showed larger uncertainty. This was due mainly the cavity bottom heat exchange uncertainty [3]. This was due to the length of the PRT sensor and the heat flow mentioned previously. An additional set of data, shown as Set 3, was taken using a radiation thermometer calibrated at PTB. For these sets of data, measurements were made both in the 8 14 µm and 3.9 µm bands. The radiation thermometers used were Heitronics models KT19.82II and TRTII. The traceability for these measurement is shown as Schemes B and C in Figure Intercomparison with flat-plates Tests involving handheld IR thermometers were performed. These are shown in data Sets 5 and 6. The IR thermometers were measured using calibrated 4181 flat plates and simultaneously measured using the 9118A cavity with a contact thermometer reference probe. These measurements were done with several different models of IR thermometers. The traceability for these tests is shown in Figure 4 under Scheme D. Measurements were made above 500 C on the cavity only. These measurements were done to compere these consistency of these results with the results at lower temperatures. The experimental uncertainties for the infrared thermometer measurements are shown in Table 2. Table 2: IR thermometer experimental measurement uncertainty. Thermal Radiation Source: 9118A Cavity U / K (k = 2) IRT 300 C 390 C 480 C 540 C 600 C 680 C Fluke Fluke NA NA NA Fluke 62 Max NA Fluke NA NA NA Thermal Radiation Source: Fluke Calibration 4181 U / K (k = 2) IRT 300 C 390 C 480 C 540 C 600 C 680 C Fluke NA NA NA Fluke NA NA NA Fluke 62 Max NA NA NA Fluke NA NA NA 4.4 Modeling of the cavity The cavity was modeled for emissivity using STEEP3 [12]. The results of this modeling are shown in Table 3 for the cavity positions 5 and 15 cm. It should be noted that the paint used on the cavity walls and cavity bottom have a significantly lower emissivity in the 3.9 µm band than in the 8 14 µm band [7].
8 Table 3. Results of STEEP3 modeling. Cavity Position Temperature Isothermal Non-Isothermal 3.9 µm 8 14 µm cm cm Results Table 4 gives the results of comparison of the AFL cavity with the 9118A cavity. In addition the comparisons with the PTB calibration are shown. Results are given for both the 8 14 µm band and the 3.9 µm band. Table 5 gives the results of the comparisons between the 9118A cavity and the 4181s using IR thermometers. All values in these tables were normalized to the nominal temperature. Table 4: Results of radiation thermometer comparisons. λ / µm LAB CAV / C 9118A / C Diff / C U CAV / K U 9118A / K En Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set
9 Table 5: Results of infrared thermometer comparisons. IRT 4181 / C 9118A / C Diff / C U CAV / K U 9118A / K En Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 5 F Set 6 F62M Set 6 F62M Set 6 F62M Set 6 F62M Set 6 F62M Set 6 F62M Tables 4 and 5 show that all of the measurements passed the normal equivalence test of being below unity [9] [10]. Another look at the tests is shown in Figure 6 showing the amount of normal equivalencies above and below µm RT 3.9 µm RT IRT < > 1.00 < > 1.00 < > 1.00 Figure 6: Normal equivalence distributions. Additional measurements were made above 500 C with the IR thermometer models Fluke 568 and 62 Max+. These measurements were not compared to another traceable measurement. However, they were plotted to determine their consistency. The results are shown in Figure 7 with a 2 nd order polynomial fit of their data. The uncertainty of the measurements are shown with a coverage factor of 2 (k = 2) by the error bars.
10 measurement error / K measurement error / K measurement error / K measurement error / K F568 #1 F568 # Temperature / C Temperature / C F62M+ #1 F62M+ # Temperature / C Temperature / C Figure 7: Self consistency of measurements above 500 C. 5 Conclusion The testing showed that the Fluke 9118A modified for use with a cavity is a viable option for IR thermometer calibrations over 300 C. The extensive testing completed and presented in this paper proved this out. The test uncertainties were favorable and would meet the needs of IR thermometer calibrations. In addition, dependence on spectral wavelength was not observed. It is possible for a calibration laboratory to reach similar uncertainties as presented in this paper with use of the proper contact thermometer and procedure. References 1. ASTM Standard E , Standard Guide for Selection and Use of Wideband, Low Temperature Infrared Thermometers, ASTM International, West Conshohocken, Pennsylvania, 2010, DOI: /E DeWitt, D., Nutter, G., Editors, Theory and Practice of Radiation Thermometry, 1988, pp , , Fischer, J., Saunders, P., Sadli, M., Battuello, M., Park, C., Yuan, Z., Yoon, H., Li, W., van der Ham, E., Sakuma, F., Yamada, Y., Ballico, M., Machin, G., Fox, N., Hollandt, J., Ugur, S., Matveyev, M., Bloembergen, P., Uncertainty budgets for calibration of radiation thermometers below the silver point, CCT-WG5 working document CCT-WG508-03, BIPM, Sèvres, France, May 2008.
11 4. ASTM Standard E e1, Standard Test Method for Calibration and Accuracy Verification of Wideband Infrared Thermometers, ASTM International, West Conshohocken, Pennsylvania, 2010, DOI: /E Liebmann, F., Kolat, T., Use of a Furnace for a Thermal Radiation Source, Proceedings of NCSLI Liebmann, F., Kolat, T., Traceability and Quality Control in a Radiation Thermometry Laboratory, NCSLI Measure, vol. 7, no. 1, pp 72 77, Liebmann, F., Infrared calibration development at Fluke Corporation Hart Scientific Division, Proceedings of SPIE Thermosense XXX, 6939, 5, VDI/VDE Guideline 3511 Blatt 4: Temperature Measurement in Industry Radiation Thermometry, ISO/IEC 17043:2010 Conformity assessment General requirements for proficiency testing, ILAC-G22:2004 Use of proficiency testing as a tool for accreditation in testing, ASTM Standard E230/E230M 12, Standard Specification and Temperature-Electromotive Force (emf) Tables for Standardized Thermocouples, ASTM International, West Conshohocken, Pennsylvania, 2012, DOI: /E0230_E0230M Prokhorov, A., Monte Carlo Method in Optical Radiometry, Metrologia, Vol. 35, pp , 1998.
Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration
Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration Gavin McQuillan NLA T&M Conference Guateng, September 26-28, 2016 2011 Fluke Calibration 1 Outline Introduction
More informationRADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY
RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY Frank E. Liebmann, Tom Kolat, Michael J. Coleman and Thomas J. Wiandt Fluke Calibration 799 Utah Valley Drive, American
More informationStandards for Radiation Thermometry
Standards for Radiation Thermometry Speaker/Author: Frank Liebmann Fluke Calibration 799 E. Utah Valley Dr. American Fork, Utah 84003 Phone: +1 801-763-1700 Fax: +1 801-763-1010 E-mail: frank.liebmann@flukecal.com
More informationTraceability and Quality Control in a Radiation Thermometry Laboratory
Traceability and Quality Control in a Radiation Thermometry Laboratory Speaker: Frank Liebmann; Fluke Calibration; 799 E. Utah Valley Dr., American Fork, Utah, USA; Phone: +1 801-763-1700; Fax: +1 801-763-1010;
More informationInfrared Calibration Development at Fluke Corporation Hart Scientific Division
Infrared Calibration Development at Fluke Corporation Hart Scientific Division Frank E. Liebmann, a Marco A. Cabrera Carrasco b a Fluke Corporation 799 Utah Valley Dr., American Fork, Utah, Estados Unidos
More informationTraceability and Quality Control in a Radiation Thermometry Laboratory
Traceability and Quality Control in a Radiation Thermometry Laboratory Frank Liebmann and Tom Kolat Abstract: In radiation thermometry, a number of steps have been taken to improve calibration quality
More informationTraceability and Quality Control in a Radiation Thermometry Laboratory
Traceability and Quality Control in a Radiation Thermometry Laboratory Speaker / Author: Frank Liebmann Co-author(s): Tom Kolat Fluke Calibration 799 E Utah Valley Dr., American Fork, Utah, 84003, USA
More informationA Study on the Effects of Bandwidth of IR Thermometry Measurements
A Study on the ffects of Bandwidth of IR Thermometry Measurements Speaker/Author: Frank Liebmann Fluke-Hart Scientific 799 Utah Valley Dr. American Fork, Utah 84003 Phone: 801-763-1600 Fax: 801-763-1010
More informationDETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE
DETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE Frank Liebmann Fluke Corporation Hart Scientific Division 799 Utah Valley Dr. American Fork, Utah 84003 801-763-1600 Frank.Liebmann@hartscientific.com
More informationInfrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
More informationIR spectral characterization of customer blackbody sources: first calibration results
IR spectral characterization of customer blackbody sources: first calibration results S. Mekhontsev, M. Noorma, A. Prokhorov and L. Hanssen Optical Technology Division National Institute of Standards and
More informationThe Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators
The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators Michael Coleman Fluke Corporation, Temperature Calibration Division, American Fork, Utah, USA mike.coleman@fluke.com
More informationCalibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity
14 th Quantitative InfraRed Thermography Conference Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity by I. Müller*, A. Adibekyan*, B. Gutschwager*, E.
More informationNIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES
NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM 83.8058 K TO 1234.93 K: METHODS AND UNCERTAINTIES Gregory F. Strouse National Institute of Standards and Technology, Gaithersburg, Maryland, USA ABSTRACT
More informationDemonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities
Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities T. J. Wiandt 1,2 1 Fluke Corporation, Hart Scientific Division, American Fork, Utah United States. 2 E-mail: tom.wiandt@hartscientific.com
More informationCalibrating the Thermal Camera
1 of 5 4/19/2012 5:33 AM from photonics.com: 12/01/2009 http://www.photonics.com/article.aspx?aid=40679 Calibrating the Thermal Camera As thermal cameras gain ground in the commercial market, testing becomes
More informationIntroduction to Blackbody Sources
Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with
More informationInfrared Temperature Calibration 101 Using the right tool means better work and more productivity
Infrared Temperature Calibration 101 Using the right tool means better work and more productivity Application Note Infrared thermometers let you measure a target s surface temperature from a distance without
More informationSensors - April A Certified-Emissivity Blackbody for Calibrating Infrared Thermometers
Page 1 of 9 www.sensorsmag.com APRIL 2002 SENSOR TECHNOLOGY AND DESIGN A Certified-Emissivity Blackbody for Calibrating Infrared Thermometers As a practical matter, most commercial infrared thermometers
More informationASSET INTEGRITY INTELLIGENCE. Featured Article. ACHIEVING A COMPREHENSIVE FIRED HEATER HEALTH MONITORING PROGRAM By Tim Hill, Quest Integrity Group
ASSET INTEGRITY INTELLIGENCE Featured Article ACHIEVING A COMPREHENSIVE FIRED HEATER HEALTH MONITORING PROGRAM By Tim Hill, Quest Integrity Group VOLUME 20, ISSUE 5 SEPTEMBER OCTOBER 2014 ACHIEVING A COMPREHENSIVE
More informationValidation of NIST s Low Temperature Infrared Spectral Radiance Scale
Validation of NIST s Low Temperature Infrared Spectral Radiance Scale Leonard Hanssen Vladimir Khromchenko Sergey Mekhontsev CALCON Logan, Utah, August 24, 2017 Outline I. Introduction II. NIST IR Radiance
More informationNew Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018
New Thermometer Guidance Document from AASHTO Accreditation Program Maria Knake Asphalt Binder ETG Meeting May 10, 2018 Background Mercury reduction initiatives have hastened the need to find suitable
More information2.1. Accuracy, n- how close the indication of the thermometer is to the true value.
AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria
More informationAbstract. 1. Introduction
The importance of establishing a very-high temperature radiation thermometry measurement capability at the National Metrology Institute of South Africa (NMISA) Speaker: Dr Efrem K Ejigu National Metrology
More informationFinal Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu
Bilateral Comparison of between NMIA (Australia) and SPRING (Singapore) (KCDB reference No. CCPR-K1.a.1) Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Contents 1. Introduction..2 2.
More informationGuide to the Realization of the ITS-90
Bureau International des Poids et Mesures Guide to the Realization of the ITS-9 Consultative Committee for Thermometry under the auspices of the International Committee for Weights and Measures CONTENTS
More informationDIRECT RADIOMETRIC TECHNIQUES
EMISSIVITY AND OTHER INFRARED-OPTICAL PROPERTIES MEASUREMENT METHODS DIRECT RADIOMETRIC TECHNIQUES Measuring principle The principle of direct radiometric techniques is shown schematically in the figure
More informationABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting
Whitepaper_ WP_T_Non-Contacting_Temperature_Measurement ABB temperature measurement Radiation thermometry Measurement made easy Process temperature measurement practice--non-contacting By Gary Freeman,
More informationCertificate of Accreditation to ISO/IEC 17025:2005
United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited
More informationA Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders
A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders Abstract Speaker/Author: Michael Bair Fluke Calibration 4765 East Beautiful Lane Phoenix, AZ 85042 Ph: 602 773
More informationUncertainty of temperature measurements by infrared thermography for metal cutting applications
IOP PUBLISHING Metrologia 50 (2013) 637 653 METROLOGIA doi:10.1088/0026-1394/50/6/637 Uncertainty of temperature measurements by infrared thermography for metal cutting applications B Lane 1, E Whitenton
More informationRecent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar
Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) Dr. Steven D. Phillips Precision Engineering Division National Institute
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Palenque 1 A, Bello Horizonte,. (Hereinafter called
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 PROTEMP MECHANICAL, INC. 3350 Scott Blvd, Bldg. #3 Santa Clara, CA 95054 Martin L. Reich Phone: 408 244 9821 CALIBRATION Valid To: August
More informationOperator's Manual. optris BR 400. Calibration source
Operator's Manual optris BR 400 Calibration source Optris GmbH Ferdinand-Buisson-Str. 14 13127 Berlin Germany Tel.: +49 30 500 197-0 Fax: +49 30 500 197-10 E-mail: info@optris.de Internet: www.optris.de
More informationEmissivity: Understanding the difference between apparent and actual infrared temperatures
Emissivity: Understanding the difference between apparent and actual infrared temperatures By L. Terry Clausing, P.E. ASNT Certified NDT Level III T/IR, for Fluke Corporation Application Note Taking infrared
More informationSection 7. Temperature Measurement
Section 7 Temperature Measurement 7/25/2017 Engineering Measurements 7 1 Working Definition Temperature is a measure of the average kinetic energy of the molecules that make of a substance. After time,
More information1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration
1551A Ex/155A Ex Intrinsically Safe Stik Thermometer The new gold standard of industrial temperature calibration Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers!
More informationCharacterization of high temperature solar thermal selective absorber coatings at operation temperature
Available online at www.sciencedirect.com Energy Procedia 00 (2013) 000 000 www.elsevier.com/locate/procedia SolarPACES 2013 Characterization of high temperature solar thermal selective absorber coatings
More informationThe in-flight calibration system for the airborne imager GLORIA
CALCON Technical Conference Utah State University, Logan, UT, USA August 4, 04 F. Olschewski, A. Ebersoldt, F. Friedl-Vallon, B. Gutschwager 4, J. Hollandt 4, A. Kleinert, C. Piesch, C. Monte 4, P. Preusse
More informationCharacterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements
EMRP-ENV03 Solar UV WP 3: Improvement of Reference Spectroradiometers Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements Peter Meindl, Christian
More informationg/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density
High-precision calibration for temperature and density g/cm³ F K C DAkkS Calibration Laboratory for Temperature and Density Accredited DAkkS laboratory (D-K-15223-01-00) for temperature and density measurements
More informationD501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect
Microscanner D-Series The Only Certified Accurate Surface Instruments in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS Common Surface Measurement Errors D501 and D501-RS Microscanner
More informationOn the effects of temperature dependence of spectral emissivity in industrial radiation thermometry
High Temperatures ^ High Pressures, 21, volume 33, pages 599 ^ 61 DOI:1.168/htjr16 On the effects of temperature dependence of spectral emissivity in industrial radiation thermometry Peter Saunders Measurement
More informationA Case for Periodic Calibration or Verification of RTDs
A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone 650-479-3208 Access code 668 676 434 Your Host and Presenter
More informationTotal radiation measurements of thermodynamic temperature
Total radiation measurements of thermodynamic temperature metrologia T. J. Quinn and J. E. Martin Abstract. The principles of total radiation thermometry as a method of primary thermometry are presented.
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ALABAMA SCALE & INSTRUMENT MARSHALL, INC. d.b.a. ASI Calibration Labs - Texas 1006 East End Blvd North Marshall, TX 75672 Lawson Hill
More informationHot solutions for temperature measurement. Contact and non-contact thermometers from Fluke
Hot solutions for temperature measurement Contact and non-contact thermometers from Fluke Infrared thermometry explained With the 60 Series and 570 Series, Fluke offers you a wide choice of non-contact
More informationCalibration of instruments measuring broadband and spectral solar (UV) irradiance
Calibration of instruments measuring broadband and spectral solar (UV) irradiance Gregor Hülsen, Julian Gröbner and Luca Egli Physikalisch-Meteorologisches Observatorium Davos, World radiation Center (PMOD/WRC)
More informationVocabulary of Metrology
Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration
More informationPotential for improved radiation thermometry measurement uncertainty through implementing a primary scale in an industrial laboratory
Measurement Science and Technology PAPER OPEN ACCESS Potential for improved radiation thermometry measurement uncertainty through implementing a primary scale in an industrial laboratory To cite this article:
More informationMeasurement method for the proficiency testing program
APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of
More informationNIST Radiance Temperature and Infrared Spectral Radiance Scales at Near-Ambient Temperatures
Int J Thermophys (2008) 29:1026 1040 DOI 10.1007/s10765-008-0384-2 NIST Radiance Temperature and Infrared Spectral Radiance Scales at Near-Ambient Temperatures S. N. Mekhontsev V. B. Khromchenko L. M.
More informationGood practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors
Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples
More informationDesigned to go where you work
Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 C over its full range, the 55A/55A Stik Thermometer is the new gold
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ARCONIC POWER AND PROPULSION RESEARCH CENTER 1500 S. Warner St. Whitehall, MI 49461-1895 Michelle McDonald Phone: 231 894 7586 CALIBRATION
More informationCCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards
CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards Prepared by M.J. de Groot, E.W.M. van der Ham and R. Monshouwer
More informationTemperature and Synchrotron Radiation
info sheet PHYSIKALISCH-TECHNISCHE BUNDESANSTALT Division 7 Temperature and Synchrotron Radiation Infoblatt_7_Temperatur und Synchrotrons_en_wa.indd 1 14.08.2012 13:49:09 Temperature and Synchrotron Radiation
More informationMaximum Performance From Dry-Well Thermometer Calibrators. Mark Finch Product Manager EMEA Fluke Calibration
Common Problems in Achieving Maximum Performance From Dry-Well Thermometer Calibrators Mark Finch Product Manager EMEA Fluke Calibration mark.finch@fluke.com Dry-well Thermal Uncertainties Temperature
More informationStudy on a thermal diffusivity standard for the laser flash method measurements 1
Study on a thermal diffusivity standard for the laser flash method measurements 1 M. Akoshima 2,3 and T. Baba 2 1 Paper presented at the Seventeenth European conference on Thermophysical Properties, September
More informationCALIBRATION REPORT FOR THERMOHYDROMETER
CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited
More informationCharacterization of the VIIRS Blackbody Emittance
Characterization of the VIIRS Blackbody Emittance Jeremy Kloepfer, Chris Taylor, and Vijay Murgai jeremiah_kloepfer@raytheon.com Conference on Characterization and Radiometric Calibration for Remote Sensing,
More informationPROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS
PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS Christian Ott Kreuzwertheim, Germany christian.ott@seho.de Heike Schlessmann heike.schlessmann@seho.de Reiner Zoch reiner.zoch@seho.de
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)
More informationEA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement
Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating
More informationREFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS
REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS S. G. R. Salim a,, N. P. Fox a, E. R. Woolliams a. R. Winkler a, H. M. Pegrum a, T. Sun b, K. T. V. Grattan b a National Physical
More informationNational Voluntary Laboratory Accreditation Program
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 Sapphire Technical Solutions, L.L.C. 10230 Rodney St. Pineville, NC 28134 Mr. Ron Wathan Phone: 704-561-3100 x106 Fax: 866-829-1502 E-mail: rwathan@sapphirests.com
More informationProper Platinum Resistance Thermometer Calibration Uncertainty Analysis
Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis Thomas Wiandt Fluke Corporation, Hart Scientific Division NCSLI 2007 Session 8B 1 Introduction Uncertainty quantification is a primary
More informationRadiation properties of IR calibrators with V-grooved surfaces
Radiation properties of IR calibrators with V-grooved surfaces Alexander V. Prokhorov, Leonard M. Hanssen *, Sergey N. Mekhontsev National Institute of Standards and Technology, 00 Bureau Dr., Gaithersburg,
More informationTemperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale
Temperature Scales The Celsius, Fahrenheit, and Kelvin Temperature Scales: Temperature, and Temperature Dependent on Physical Properties Physics Enhancement Programme Dr. M.H. CHAN, HKBU 9 T F T 5 T T
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V Av. México # 329 Col. Valle De
More informationMISE EN PRATIQUE OF THE REALIZATION OF THE KELVIN
MISE EN PRATIQUE OF THE REALIZATION OF THE KELVIN Adopted by the Consultative Committee for Thermometry (CCT) in XXX 1. Scope The text of and supplementary information on the International Temperature
More informationI. Yang, C. H. Song, Y.-G. Kim & K. S. Gam
Cryostat for Fixed-Point Calibration of Capsule-Type SPRTs I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam International Journal of Thermophysics Journal of Thermophysical Properties and Thermophysics and Its
More informationThermal Image Resolution on Angular Emissivity Measurements using Infrared Thermography
, March 18-20, 2015, Hong Kong Thermal Image Resolution on Angular Emissivity Measurements using Infrared Thermography T. Nunak, K. Rakrueangdet, N. Nunak, and T. Suesut Abstract This paper reports the
More informationChapter 1. Blackbody Radiation. Theory
Chapter 1 Blackbody Radiation Experiment objectives: explore radiation from objects at certain temperatures, commonly known as blackbody radiation ; make measurements testing the Stefan-Boltzmann law in
More informationHigh Accuracy High Speed in Your Process
DX-Series Handheld Precision IR Thermometers EXERGEN C O R P O R A T I O N The Only Certified Accurate NIST Traceable Infrared Temperature Instruments in the World* High Accuracy High Speed in Your Process
More informationThe Fundamentals of Moisture Calibration
The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity
More informationFor the National Voluntary Laboratory Accreditation Program
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 Additel Corporation 2900 Saturn Street Brea, CA 92821 Eric Chavier Phone: 714-998-6899 Fax: 714-998-6999 E-mail: eric.chavier@additel.com URL: http://additel.com
More informationEuropean Association of National Metrology Institutes
European Association of National Metrology Institutes EURAMET GUIDELINES ON TEMPERATURE: Extrapolation of SPRT calibrations below the triple point of argon, 83.8058 K, and traceability in baths of liquid
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLORIDA METROLOGY, LLC 645 NW Enterprise Dr. Port St. Lucie FL 34986 Ron Kupper Phone: (772)212-7158 CALIBRATION Valid To: October 31,
More informationREDUCTION OF UNCERTAINTY IN CALIBRATION OF GAUGE BLOCKS
Vienna, AUSTRIA, 000, September 5-8 RDUCTION OF UNCRTAINTY IN CALIBRATION OF GAUG BLOCKS B. Acko, A. Sostar and A. Gusel Laboratory for Production Measurement Faculty of Mechanical ngineering University
More informationTRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS
TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,
More informationA2LA. G118: Guidance for Defining the Scope of Accreditation for Calibration Laboratories
Page 1 of 35 Accreditation for Calibration Laboratories October 2018 2018 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written
More informationSIB 52 - THERMO Stakeholder meeting May 16
SIB 52 - THERMO Stakeholder meeting May 16 Metrology for thermal protection materials Challenges in thermal conductivity measurements of thin (multi-layered) thermal insulation materials Laboratoire national
More informationEmissivity modeling of thermal radiation sources with concentric grooves
High Temperatures ^ High Pressures, 2003/2004, volume 35/36, pages 199 ^ 207 DOI:10.1068/htjr093 Emissivity modeling of thermal radiation sources with concentric grooves Alexander V Prokhorov, Sergey N
More informationVersion 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany
Guidelines on the Calibration of Temperature Block Calibrators Authorship and Imprint This document was developed by the EURAMET e.v., Technical Committee for Thermometry. Authors: Yves Hermier (LNE-INM,
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSolar irradiance measurement up to 2500nm with the Arcoptix FT-NIR
Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,
More informationAnnex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005
Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-K-15133-02-00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 24.01.2018 to 23.01.2023 Holder of certificate: Hexagon
More informationThis annex is valid from: to Replaces annex dated: Location(s) where activities are performed under accreditation
Wiltonstraat 25 3905 KW Veenendaal The Netherlands Location(s) where activities are performed under accreditation Head Office Location Abbreviation/ location Main Location Wiltonstraat 25 3905 KW Veenendaal
More informationMethod to Remove the Effect of Atmosphere and Ambient Radiation on Colorimetric Temperature Measurement
937 A publication of VOL. 46, 2015 CHEMICAL ENGINEERING TRANSACTIONS Guest Editors: Peiyu Ren, Yancang Li, Huiping Song Copyright 2015, AIDIC Servizi S.r.l., ISBN 978-88-95608-37-2; ISSN 2283-9216 The
More informationTraceable UV-Vis Reference Materials
UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof
More informationCALIBRATION CERTIFICATE # 503
CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion
More informationIntroduction to Infrared Thermometry
TS-104 Introduction to Infrared Thermometry Fig. 1 - Blackbody Radiation Characteristics General Infrared thermometers have the ability to measure temperature without physical contact. The ability to accomplish
More informationAccreditation of radiochemical analyses, from NAMAS to ISO 17025:2005 and beyond
Accreditation of radiochemical analyses, from NAMAS to ISO 17025:2005 and beyond George Ham Centre for Radiation, Chemicals and Environmental Hazards Health Protection Agency The Analysts Dilemma: Maintaining
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 JEM PRECISION 9615 56 Avenue NW Edmonton, Alberta T6E 0B2 Canada Robert Korthuis Phone: 780-440 1933 CALIBRATION
More informationThe steps to an uncertainty budget
The steps to an uncertainty budget Emma Woolliams 4 April 2017 Prepared by Paul Miller http://www.emceoc.org Uncertainty Where to start? What to do? How to be consistent? Make it easy. At the end of this
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 MTS FIELD SERVICE 14000 Technology Drive Eden Prairie, MN 55344 Hugh Casper Phone: 518 852 3977 CALIBRATION Valid To: September 30,
More information