Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements
|
|
- Janis Thomasine Byrd
- 5 years ago
- Views:
Transcription
1 EMRP-ENV03 Solar UV WP 3: Improvement of Reference Spectroradiometers Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements Peter Meindl, Christian Monte, Martin Wähmer Physikalisch-Technische Bundesanstalt (PTB) Braunschweig & Berlin, Germany The EMRP is jointly funded by the EMRP participating countries within EURAMET and the European Union.
2 Portable reference spectroradiometer system QASUME Quality Assurance of Spectral UV Measurements in Europe (QASUME): Bentham DM-150 double grating monochromator measures the spectrum sequentially several minutes of scanning time (difficult in the case of fast varying atmospheric conditions) on-site wavelength calibration using spectral emission lines (e.g. mercury) and extraterrestrial spectra Photomultiplier tube nonlinear response instability of spectral responsivity (on-site calibrator necessary) Uncertainty for solar UV irradiance measurements: 4,6 % to 8,8 % (k=2) depending on wavelength and solar zenith angle dominated by radiometric calibration and wavelength uncertainty for l< 310 nm Appl. Opt. 44, 2005,
3 Motivation: Why using an FTS as a reference instrument? High wavenumber accuracy and wavenumbers are inherently traced to SI by built-in HeNe laser with low uncertainty No on-site wavenumber calibration necessary! Low uncertainty contribution to solar UV measurements (esp. for l < 310 nm)! High throughput circular aperture has larger area compared to linear slits no diffraction losses to higher-order spectra Semiconductor detectors usable instead of photomultiplier tube? FTS covers broad spectral ranges with high resolution, and all wavenumbers are measured simultaneously Faster than grating spectroradiometer? Instrumental distortions are often accurately calculable and correctable S.P.Davis, M.C.Abrams and J.W.Brault: Fourier Transform Spectrometry (Academic Press, 2001). Herres, Gronholz: Understanding FT-IR Data Processing, Part 1-3. Comp. Appl. Lab. 2 (1984), p.216.
4 Concept of FTS spectroradiometer Commercially available FTS Bruker Vertex80v Global entrance optics CMS Schreder, Austria Semiconductor detectors GaP diode (Bruker) Si diode (Bruker) Si diode (Hamamatsu S8552) Bruker FTS Vertex80v Global entrance optics for irradiance measurements (CMS Schreder, Austria)
5 Wavelength traceability to SI fixed Mirror movable Mirror 10 I I(X) I(X) X 5 10
6 Determination of interferogram sampling positions fixed Mirror l vac = nm HeNe Laser movable Mirror l vac I HeNe (X) I(X) Detector 2 X interferogram sampling positions
7 Wavelength traceability to SI Deviations of the measured mercury peak wavenumbers from the literature data and the deviation of the measured HeNe wavenumber from the SI value. normal air pressure FTS evacuated Refractive index of standard air (15 C, hpa, 450 ppm CO 2, 0 % humidity) Mercury spectrum: Applied Optics Vol. 35 (1), 1996, Physica Scripta, Vol. 63, 2001,
8 Wavelength uncertainty Wavelength uncertainty of the Bruker VERTEX 80v in case of a calibration using an external HeNe laser and FTS usage at normal air pressure (i.e. standard operation) wavelength uncertainty contribution 500 nm 333 nm 250 nm wavenumber calibration 2.5 pm 1.1 pm 0.6 pm dispersion 2.5 pm 2.8 pm 5.0 pm large aperture 3.0 pm 1.3 pm 0.8 pm combined uncertainty 4.6 pm 3.3 pm 5.1 pm calibration using the internal HeNe laser and normal air pressure Wavelength scale of the FTS: combined uncertainty 5.4 pm 3.8 pm 5.4 pm direct traceable to the SI calibration using an external no HeNe on-site laser recalibration and or wavelength check necessary FTS evacuated wavelength uncertainty 5 pm in the range 250 nm to 500 nm combined uncertainty 3.9 pm 1.7 pm 1.0 pm well below 50 pm (demanded in JRP ENV03)
9 SNR for non-calibrated solar UV radiation measurements Source: Sun, 23.July 2013 Resolution: 10 cm -1 Average of 512 interfrerograms in about 300 s Results: FTS usable from 300 nm to 500 nm Steep increase of solar irradiance around 300 nm is not resolvable GaP detector gives a better SNR compared to Si detectors
10 Detector characteristics com DigiTect_Detectors.pdf
11 Calibration of spectral irradiance responsivity Spectral irradiance?
12 Calibration of spectral irradiance responsivity Gold fixed-point Black Body Radiator (ITS90) Radiation Thermometer LP3 High Temperature Black Body Radiator BB3200pg + aperture Cryogenic radiometer Si-trap detector + aperture Filter Radiometer Blackbody BB3200pg + aperture Spectroradiometer Spectral irradiance standard FTS PTB-FB 4.1 Photometry and Applied Radiometry FTS QASUME
13 SNR for FTS calibration at high-temperature black body Source: HTBB at T =2795 K Resolution: 10 cm -1 Results: Calibration possible from 360 nm to 500 nm GaP detector gives a better SNR compared to Si detectors
14 Spectral irradiance responsivity of FTS + GEO Spectral irradiance responsivity of the FTS with global entrance optics for different types of semiconductor detectors. Calibration against HTBB at T = 2795 K.
15 Relative difference of spectral responsivity Comparison of spectral irradiance calibrations Relative difference of two calibrations against: HTBB at T =2795 K and Secondary spectral irradiance standard (tungsten halogen lamp) Detector: GaP, Resolution: 10 cm -1
16 SNR for calibration measurements Calibration against: HTBB at T =2795 K and Secondary spectral irradiance standard (tungsten halogen lamp) Detector: GaP, Resolution: 10 cm -1
17 Relative difference of spectral responsivity Stability of spectral irradiance calibrations Relative difference of two calibrations against the secondary spectral irradiance standard (tungsten halogen lamp) time interval 1 month Detector: GaP, Resolution: 10 cm -1
18 Solar UV irradiance
19 Results Wavelength scale of the FTS is traceable to SI via built-in HeNe laser wavelength uncertainty: 5 pm (250 nm to 500 nm) FTS + GaP/Si + GEO + measurement times equivalent to QASUME Solar UV measurements possible from 300 nm to 500 nm Calibration of spectral irradiance responsivity with HTBB at 2795 K is limited to 360 nm to 500 nm Solar UV irradiance measurements limited to 360 nm to 500 nm Performance of QASUME is not reached with FTS + GaP/Si detectors + GEO Advantage of stable spectral responsivity of GaP/Si detectors can not be applied
20 Outlook Improvements: Calibration against HTBB at higher temperatures (> 3000 K) Application of a channel photomultiplier with high dynamic range and up to 3 orders of magnitude higher detectivity Wavelength scale traceable to SI, measurements probably faster than QASUME with high dynamic range but: on-site calibrator probably needed as with QASUME Optimization of the global entrance optics Evaluation of uncertainty budgets for the spectral irradiance calibration and for solar UV irradiance measurements with FTS Comparison measurements of the FTS and QASUME
Calibration of instruments measuring broadband and spectral solar (UV) irradiance
Calibration of instruments measuring broadband and spectral solar (UV) irradiance Gregor Hülsen, Julian Gröbner and Luca Egli Physikalisch-Meteorologisches Observatorium Davos, World radiation Center (PMOD/WRC)
More informationBrewer UV measurement and calibration
Brewer UV measurement and calibration Julian Gröbner Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC) 14/10/2014 Brewer UV Measurements and calibration 1 Overview Comparison
More informationThe in-flight calibration system for the airborne imager GLORIA
CALCON Technical Conference Utah State University, Logan, UT, USA August 4, 04 F. Olschewski, A. Ebersoldt, F. Friedl-Vallon, B. Gutschwager 4, J. Hollandt 4, A. Kleinert, C. Piesch, C. Monte 4, P. Preusse
More information2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at
2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various
More informationCharacterisation & Use of Array Spectrometers
Characterisation & Use of Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington Middlesex, UK 1 Overview Basic design and features of
More informationCalibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity
14 th Quantitative InfraRed Thermography Conference Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity by I. Müller*, A. Adibekyan*, B. Gutschwager*, E.
More informationThermodynamic Temperature of High-Temperature Fixed Points Traceable to Blackbody Radiation and Synchrotron Radiation
Int J Thermophys (217) 38:144 DOI 1.17/s1765-17-2273-z TEMPMEKO 216 Thermodynamic Temperature of High-Temperature Fixed Points Traceable to Blackbody Radiation and Synchrotron Radiation M. Wähmer 1 K.
More informationTraceability of on-machine measurements under a wide range of working conditions
Traceability of on-machine measurements under a wide range of working conditions Frank Keller, Matthias Franke, Norbert Gerwien, Lisa Groos, Christian Held, Klaus Wendt Traceable in-process dimensional
More informationProtocol of the intercomparison at LKO, Arosa, Switzerland on July 21 to 28, 2008 with the travelling reference spectroradiometer QASUME from PMOD/WRC
/09/00 Protocol of the intercomparison at LKO, Arosa, Switzerland on July to, 00 with the travelling reference spectroradiometer QASUME from PMOD/WRC Report prepared by Gregor Hülsen, Julian Gröbner Operator:
More informationIndoor calibration method for UV index meters with a solar simulator and a reference spectroradiometer
Int. J. Metrol. Qual. Eng. 7, 101 (2016) c EDP Sciences 2016 DOI: 10.1051/ijmqe/2015031 Indoor calibration method for UV index meters with a solar simulator and a reference spectroradiometer Helmi Zaini
More informationMeasurement method for the proficiency testing program
APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of
More informationA guide to measuring solar UV spectra using array spectroradiometers
ENV03 solaruv A guide to measuring solar UV spectra using array spectroradiometers Mario Blumthaler Julian Gröbner Luca Egli Saulius Nevas 2011 Version 1.0 1 This guide was compiled within the EMRP ENV03
More informationREFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS
REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS S. G. R. Salim a,, N. P. Fox a, E. R. Woolliams a. R. Winkler a, H. M. Pegrum a, T. Sun b, K. T. V. Grattan b a National Physical
More informationCharacterization of high temperature solar thermal selective absorber coatings at operation temperature
Available online at www.sciencedirect.com Energy Procedia 00 (2013) 000 000 www.elsevier.com/locate/procedia SolarPACES 2013 Characterization of high temperature solar thermal selective absorber coatings
More informationFundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p.
Preface p. xiii Fundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p. 4 Fundamental Definitions p. 7 Lambertian Radiators
More informationEmissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites
Thermographie-Kolloquium 2017 More info about this article: http://www.ndt.net/?id=22484 Emissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites Albert ADIBEKYAN
More informationThe steps to an uncertainty budget
The steps to an uncertainty budget Emma Woolliams 4 April 2017 Prepared by Paul Miller http://www.emceoc.org Uncertainty Where to start? What to do? How to be consistent? Make it easy. At the end of this
More informationSOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION
SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION G. Thuillier 1, D. Bolsee 2, T. Foujols 1 1 LATMOS-CNRS, France 2 Institut d Aéronomie
More informationCalibration of Ocean Colour Sensors
Dr. A. Neumann German Aerospace Centre DLR Remote Sensing Technology Institute Marine Remote Sensing What is Calibration, why do we need it? Sensor Components Definition of Terms Calibration Standards
More informationCourse Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I
SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Course Details September 22 October 10 September 22 November 7 November 17 December 1 Topic Period Introduction to Spectrometric
More informationHigh Resolution Optical Spectroscopy
PHYS 3719 High Resolution Optical Spectroscopy Introduction This experiment will allow you to learn a specific optical technique with applications over a wide variety of phenomena. You will use a commercial
More informationSHORT TERM SCIENTIFIC MISSION (STSM) SCIENTIFIC REPORT
SHORT TERM SCIENTIFIC MISSION (STSM) SCIENTIFIC REPORT This report is submitted for approval by the STSM applicant to the STSM coordinator Action number: ES1309 STSM title: Laboratory based radiometric
More informationEuropean Project Metrology for Radioactive Waste Management
European Project Metrology for Radioactive Waste Management Petr Kovar Czech Metrology Institute Okruzni 31 638 00, Brno, Czech republic pkovar@cmi.cz Jiri Suran Czech Metrology Institute Okruzni 31 638
More informationCorrection of Angular Response Error in Brewer UV Irradiance Measurements
2018 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 25 Correction of Angular Response Error in Brewer UV Irradiance Measurements MANUEL ANTÓN, ANTONIO SERRANO, AND
More informationCalibration demonstration system for an imaging spectrometer to provide climatequality
Calibration demonstration system for an imaging spectrometer to provide climatequality reflectance K. Thome, J. McCorkel, J. Hair, B. McAndrew, A. Daw, D. Jennings, D. Rabin NASA/GSFC 1 Calibration Demonstration
More information5.7. Summit, Greenland (01/22/12 11/22/12)
CHAPTER 5: QUALITY CONTROL AND CALIBRATION STANDARDS 5.7. Summit, Greenland (0/22/2 /22/2) This section describes quality control of Volume 22 solar data recorded by the SUV-50B spectroradiometer at Summit
More informationStray Light Rejection in Array Spectrometers
Stray Light Rejection in Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington, Middlesex, UK 1 Overview Basic optical design of an array
More informationNIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.
NIRSpec Performance Report NPR-2013-011 / ESA-JWST-RP-19649 Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.2013 Revision: 1 estec European Space Research and Technology Centre Keplerlaan
More informationTHE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS
THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately
More informationIntroduction to Fourier Transform Infrared Spectroscopy
molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method
More informationFourier Transform Infrared. Spectrometry
Fourier Transform Infrared. Spectrometry Second Editio n PETER R. GRIFFITH S JAMES A. de HASETH PREFACE x v CHAPTER 1 INTRODUCTION TO VIBRATIONAL SPECTROSCOPY 1 1.1. Introduction 1 1.2. Molecular Vibrations
More informationChemistry 524--Final Exam--Keiderling Dec. 12, pm SES
Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet
More informationNovel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007
Novel Flux Calibration Source CORM 2007 David C. Gross May 10, 2007 LED Calorimetry Absolute Radiometry Design Goals Proof of Concept - Design and Results Total Spectral Flux Calibration Absolute Radiometry
More informationPAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 7: Instrumentation for IR spectroscopy
KNOW MORE Web links https://en.wikipedia.org/wiki/infrared_ http://hiq.lindegas.com/en/analytical_methods/infrared_/non_dispersive_infrared.html http://blamp.sites.truman.edu/files/2012/11/322-ir-and-ftir.pdf
More informationCOMPREHENSIVE ANALYSIS OF A PULSED SOLAR SIMULATOR TO DETERMINE MEASUREMENT UNCERTAINTY COMPONENTS
COMPREHENSIVE ANALYSIS OF A PULSED SOLAR SIMULATOR TO DETERMINE MEASUREMENT UNCERTAINTY COMPONENTS F. Plag* 1, F. Haas 2, K. Ramspeck 2, H. Nagel 3, H. Albert 2, S. Nevas 1, P. Sperfeld 1, F. Witt 1 and
More informationMethods of total spectral radiant flux realization at VNIIOFI
Journal of Physics: Conference Series PAPER OPEN ACCESS Methods of total spectral radiant flux realization at VNIIOFI To cite this article: Evgeniy Ivashin et al 2018 J. Phys.: Conf. Ser. 972 012011 View
More informationThermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement
Thermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement Appendix A Outlines of GOSAT and TANSO Sensor GOSAT (Greenhouse
More informationSBUV(/2) and SSBUV Solar Irradiance Measurements Matthew DeLand, Richard Cebula, Liang-Kang Huang Science Systems and Applications, Inc.
SBUV(/2) and SSBUV Solar Irradiance Measurements Matthew DeLand, Richard Cebula, Liang-Kang Huang Science Systems and Applications, Inc. (SSAI) Solar Spectral Irradiance Variations Workshop NIST, Gaithersburg,
More informationIntroduction to FT-IR Spectroscopy
Introduction to FT-IR Spectroscopy An FT-IR Spectrometer is an instrument which acquires broadband NIR to FIR spectra. Unlike a dispersive instrument, i.e. grating monochromator or spectrograph, an FT-IR
More informationTwo-electron systems
Two-electron systems Laboratory exercise for FYSC11 Instructor: Hampus Nilsson hampus.nilsson@astro.lu.se Lund Observatory Lund University September 12, 2016 Goal In this laboration we will make use of
More informationOptical Frequency Comb Fourier Transform Spectroscopy with Resolution beyond the Path Difference Limit
Optical Frequency Comb Fourier Transform Spectroscopy with Resolution beyond the Path Difference Limit Aleksandra Foltynowicz, Alexandra C. Johansson, Amir Khodabakhsh, Lucile Rutkowski Department of Physics,
More informationTable 1: Participating Instruments. 11/01/2010
Electronic Supplementary Material (ESI) for Photochemical & Photobiological Sciences This journal is The Royal Society of Chemistry and Owner Societies // Protocol of the intercomparison at Ny Ålesund,
More informationChemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES
Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or
More informationHARP Assessment of Uncertainty
HARP Assessment of Uncertainty The HIAPER Airborne Radiation Package (HARP) was designed to produce accurate measurements of actinic flux and irradiance. The Atmospheric Radiation Group (ARG) at the University
More informationABB Remote Sensing Atmospheric Emitted Radiance Interferometer AERI system overview. Applications
The ABB Atmospheric Emitted Radiance Interferometer AERI provides thermodynamic profiling, trace gas detection, atmospheric cloud aerosol study, air quality monitoring, and more. AERI high level overview
More informationPRINCIPLES OF PHYSICAL OPTICS
PRINCIPLES OF PHYSICAL OPTICS C. A. Bennett University of North Carolina At Asheville WILEY- INTERSCIENCE A JOHN WILEY & SONS, INC., PUBLICATION CONTENTS Preface 1 The Physics of Waves 1 1.1 Introduction
More informationFUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON
More informationIR spectral characterization of customer blackbody sources: first calibration results
IR spectral characterization of customer blackbody sources: first calibration results S. Mekhontsev, M. Noorma, A. Prokhorov and L. Hanssen Optical Technology Division National Institute of Standards and
More informationIntroduction to Fourier Transform Infrared Spectroscopy
Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR
More informationSolar irradiance measurement up to 2500nm with the Arcoptix FT-NIR
Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,
More information25 Instruments for Optical Spectrometry
25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.
More informationThe Compact SIM (CSIM) and Compact TIM (CTIM) Instruments
The Compact SIM (CSIM) and Compact TIM (CTIM) Instruments Dave Harber, Zach Castleman, Ginger Drake, Nat Farber, Melanie Fisher, Maxwell Fowle, Karl Heuerman, Joel Rutkowski, Matt Smith, Paul Smith, Jacob
More informationSpectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )
Spectroscopy: Introduction Required reading Chapter 18 (pages 378-397) Chapter 20 (pages 424-449) Spectrophotometry is any procedure that uses light to measure chemical concentrations Properties of Light
More informationFinal Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu
Bilateral Comparison of between NMIA (Australia) and SPRING (Singapore) (KCDB reference No. CCPR-K1.a.1) Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Contents 1. Introduction..2 2.
More informationFull solar rotations observed by the SOLAR payload on the ISS in December 2012 and June 2013.
Full solar rotations observed by the SOLAR payload on the ISS in December 2012 and June 2013. C. Muller and the SOLAR operation and science teams, SOLAR instruments: SOVIM Total sun irradiance measurement
More informationn ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is
Optical Spectroscopy Lennon O Naraigh, 0000 Date of Submission: 0 th May 004 Abstract: This experiment is an exercise in the principles and practice of optical spectroscopy. The continuous emission spectrum
More informationTemperature and Synchrotron Radiation
info sheet PHYSIKALISCH-TECHNISCHE BUNDESANSTALT Division 7 Temperature and Synchrotron Radiation Infoblatt_7_Temperatur und Synchrotrons_en_wa.indd 1 14.08.2012 13:49:09 Temperature and Synchrotron Radiation
More informationSpectral and photopic studies for high intensity discharge (HID) lamps
Int. J. Metrol. Qual. Eng. 6, 105 (2015) c EDP Sciences 2015 DOI: 10.1051/ijmqe/2015005 Spectral and photopic studies for high intensity discharge (HID) lamps A.-E.A. Abd-Elmageed and E.M. El-Moghazy National
More informationProgress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths
Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths Claire Cramer, Steve Brown, Keith Lykke, John Woodward (NIST) Tom Stone (USGS) Motivation for using the Moon
More informationTRUTHS-Lite: A Microsatellite Based Climate Benchmark Mission
IAA-B10-0404 TRUTHS-Lite: A Microsatellite Based Climate Benchmark Mission Craig Underwood 1, Nigel Fox 2, Javier Gorroño 2 1 Surrey Space Centre, University of Surrey, Guildford, Surrey, GU2 7XH, UK 2
More informationSpectroscopy. Page 1 of 8 L.Pillay (2012)
Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has
More informationRecent advances in the detection efficiency calibration of Silicon Single-Photon Avalanche Diodes using double attenuator technique
Recent advances in the detection efficiency calibration of Silicon Single-Photon Avalanche Diodes using double attenuator technique S. Kück 1, M. López 1, K. Dhoska 2, T. Kübarsepp 2,3, H. Hofer 1, B.
More informationThe official newsletter of the Thematic Network for Ultraviolet Measurements
The official newsletter of the Thematic Network for Ultraviolet Measurements Issue 12 / March 2017 Thematic Network for Ultraviolet Measurements Contents Editorial 3 EMRP ENV59 Traceability for atmospheric
More informationPRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS
PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used
More informationAdvanced Spectroscopy Laboratory
Advanced Spectroscopy Laboratory - Raman Spectroscopy - Emission Spectroscopy - Absorption Spectroscopy - Raman Microscopy - Hyperspectral Imaging Spectroscopy FERGIELAB TM Raman Spectroscopy Absorption
More informationAbsolute Radiance Re-Calibration of FIRST
Utah State University DigitalCommons@USU Space Dynamics Lab Publications Space Dynamics Lab 1-1-2012 Absolute Radiance Re-Calibration of FIRST Harri Latvakoski Utah State University Marty Mylncak Utah
More informationSI Broader Scope - Projects
European Metrology Programme for Innovation and Research SI Broader Scope - Projects An overview of the funded projects from the Targeted Programme SI Broader Scope. The aim of these projects is to underpin
More informationElectro-Optical System. Analysis and Design. A Radiometry Perspective. Cornelius J. Willers SPIE PRESS. Bellingham, Washington USA
Electro-Optical System Analysis and Design A Radiometry Perspective Cornelius J Willers SPIE PRESS Bellingham, Washington USA Nomenclature xvii Preface xxiii 1 Electro-Optical System Design 1 11 Introduction
More informationInstruments to measure solar ultraviolet radiation
WORLD METEOROLOGICAL ORGANIZATION GLOBAL ATMOSPHERE WATCH No. 125 Instruments to measure solar ultraviolet radiation Part 1: Spectral instruments Lead Author: G. Seckmeyer Co-authors: A. Bais, G. Bernhard,
More informationUSE OF A LOW-COST SILICON DIODE-ARRAY SPECTRORADIOMETER TO MEASURE SOLAR UV-B
USE OF A LOW-COST SILICON DIODE-ARRAY SPECTRORADIOMETER TO MEASURE SOLAR UV-B Terence M. Murphy, Section of Plant Biology, University of California, Davis, CA 95616 Presented at an NIST Conference on Critical
More informationReal-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System
Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Presented at the BevTech Conference, Albuquerque, NM 2018 by Charles M. Phillips Ph.D., Max Analytical Technologies Mark Taylor, Vice
More information5. Quality Control and Calibration Standards
5. Quality Control and Calibration Standards Successful operation of a network of complex instruments, such as scanning spectroradiometers, depends upon a well-defined approach to quality assurance and
More informationAccuracy and Precision Requirements for Climate-Level Data Sets
Accuracy and Precision Requirements for Climate-Level Data Sets K. Thome NASA/GSFC Libya-4 Workshop Paris, France October 4-5, 2012 Accuracy requirements Commercial imagers Precision and SNR drive calibration
More informationEXPOSURE TIME ESTIMATION
ASTR 511/O Connell Lec 12 1 EXPOSURE TIME ESTIMATION An essential part of planning any observation is to estimate the total exposure time needed to satisfy your scientific goal. General considerations
More informationBecause light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.
Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light
More informationLDLS Laser-Driven Light Source
LDLS Laser-Driven Light Source From The Innovators in Light ENERGETIQ 7/8/2011 1 Peter J. Dwyer, Ph.D. pdwyer@energetiq.com +1 781 939 0763 x 141 The LDLS Product Range EQ-99FC LDLS System EQ-1500 LDLS
More informationFabry-Perot Interferometer for atmospheric monitoring useful for EAS detection E.Fokitis 1, K. Patrinos 1, Z. Nikitaki 1
Fabry-Perot Interferometer for atmospheric monitoring useful for EAS detection E.Fokitis 1, K. Patrinos 1, Z. Nikitaki 1 ABSTRACT A piezotunable Fabry-Perot interferometer is studied as a candidate Doppler
More informationA TES Bolometer for THz FT-Spectroscopy
A TES Bolometer for THz FT-Spectroscopy M. Kehrt, J. Beyer, C. Monte, J. Hollandt Physikalisch-Technische Bundesanstalt Abbestraße 2-12, Berlin, Germany E-Mail: Mathias.Kehrt@PTB.de Abstract - We recently
More informationESA Ground-Based Air-Quality Spectrometer Validation Network and Uncertainties Study
LuftBlick Report 2014009 ESA Ground-Based Air-Quality Spectrometer Validation Network and Uncertainties Study Laboratory Report Name Company Date prepared by Martin Tiefengraber LuftBlick 30th January
More informationAstronomy. Optics and Telescopes
Astronomy A. Dayle Hancock adhancock@wm.edu Small 239 Office hours: MTWR 10-11am Optics and Telescopes - Refraction, lenses and refracting telescopes - Mirrors and reflecting telescopes - Diffraction limit,
More informationA Calibration Procedure Which Accounts for Non-linearity in Singlemonochromator Brewer Ozone Spectrophotometer Measurements
Atmos. Meas. Tech. Discuss., https://doi.org/.194/amt-18-17 A Calibration Procedure Which Accounts for Non-linearity in Singlemonochromator Brewer Ozone Spectrophotometer Measurements Zahra Vaziri Zanjani
More informationRemote Sensing Systems Overview
Remote Sensing Systems Overview Remote Sensing = Measuring without touching Class objectives: Learn principles for system-level understanding and analysis of electro-magnetic remote sensing instruments
More informationMERIS US Workshop. Instrument Characterization Overview. Steven Delwart
MERIS US Workshop Instrument Characterization Overview Steven Delwart Presentation Overview On-Ground Characterisation 1. Diffuser characterisation 2. Polarization sensitivity 3. Optical Transmission 4.
More informationWORLD METEOROLOGICAL ORGANIZATION GLOBAL ATMOSPHERE WATCH
WORLD METEOROLOGICAL ORGANIZATION GLOBAL ATMOSPHERE WATCH No. 164 Instruments to Measure Solar Ultraviolet Radiation Part 2: Broadband Instruments Measuring Erythemally Weighted Solar Irradiance Lead Author:
More informationAnalytical Spectroscopy Review
Analytical Spectroscopy Review λ = wavelength ν = frequency V = velocity = ν x λ = 2.998 x 10 8 m/sec = c (in a vacuum) ν is determined by source and does not change as wave propogates, but V can change
More informationAccurate Measurement of Transmittance and Reflectance for Engineering Applications
Accurate Measurement of Transmittance and Reflectance for Engineering Applications Dr. Chen Fangzhi Laboratory Manager Façade & Roof Materials Testing Laboratory OTM Solutions Pte Ltd PerkinElmer INTour
More informationMonitoring of UV spectral irradiance at Thessaloniki ( ): data re-evaluation and quality control
Ann. Geophys., 24, 3215 3228, 26 www.ann-geophys.net/24/3215/26/ European Geosciences Union 26 Annales Geophysicae Monitoring of UV spectral irradiance at Thessaloniki (199 25): data re-evaluation and
More informationInfluence of directivity and spectral shape on the measured sound power level
Influence of directivity and spectral shape on the measured sound power level K. Voelkel 1, V. Wittstock Physikalisch Technische Bundesanstalt, Braunschweig, Germany ABSTRACT The sound power is the main
More informationElectronic Excitation by UV/Vis Spectroscopy :
SPECTROSCOPY Light interacting with matter as an analytical tool III Pharm.D Department of Pharmaceutical Analysis SRM College Of Pharmacy,Katankulathur Electronic Excitation by UV/Vis Spectroscopy : X-ray:
More informationNumerical modeling of the primary source in a hemi-anechoic room
Numerical modeling of the primary source in a hemi-anechoic room R. Arina 1, K. Völkel 2 1 Politecnico di Torino, Torino, Italy 2 Physikalisch Technische Bundesanstalt, Braunschweig, Germany ABSTRACT An
More informationChem 310 rd. 3 Homework Set Answers
-1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption
More informationChapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry
Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation
More informationGAS CAMERA MOBILE IMAGING SYSTEM FOR VISUALISING METHANE PLUMES AT DISTANCES BETWEEN 0 M AND 100 M AND MORE
23rd World Gas Conference, Amsterdam 2006 GAS CAMERA MOBILE IMAGING SYSTEM FOR VISUALISING METHANE PLUMES AT DISTANCES BETWEEN 0 M AND 100 M AND MORE Main Author Peter Schwengler, E.ON Ruhrgas AG, Pipelines
More informationCharacterization of the 300 K and 700 K Calibration Sources for Space Application with the Bepicolombo Mission to Mercury
Int J Thermophys (2011) 32:1429 1439 DOI 10.1007/s10765-011-0979-x Characterization of the 300 K and 700 K Calibration Sources for Space Application with the Bepicolombo Mission to Mercury B. Gutschwager
More informationČerenkov counting and liquid scintillation counting of 36 Cl
Čerenkov counting and liquid scintillation counting of 36 Cl Karsten Kossert, Ole Nähle Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany and Agustín Grau Carles Instituto de Física Fundamental
More informationBilateral NIST-PTB Comparison of Spectral Responsivity in
Bilateral NIST-PTB Comparison of Spectral Responsivity in the VUV A Gottwald 1, M Richter 1, P-S Shaw 2, Z Li 2 and U Arp 2 1 Physikalisch-Technische Bundesanstalt, Abbestraße 2-12, 10587 Berlin, Germany
More informationFT-IR Spectroscopy. An introduction in measurement techniques and interpretation
FT-IR Spectroscopy An introduction in measurement techniques and interpretation History Albert Abraham Michelson (1852-1931) Devised Michelson Interferometer with Edward Morley in 1880 (Michelson-Morley
More informationInstrumental Analysis: Spectrophotometric Methods
Instrumental Analysis: Spectrophotometric Methods 2007 By the end of this part of the course, you should be able to: Understand interaction between light and matter (absorbance, excitation, emission, luminescence,fluorescence,
More informationLecture 0. NC State University
Chemistry 736 Lecture 0 Overview NC State University Overview of Spectroscopy Electronic states and energies Transitions between states Absorption and emission Electronic spectroscopy Instrumentation Concepts
More informationNanoholes for leak metrology
Vacuum Metrology for Industry Nanoholes for leak metrology Università Degli Studi di Genova, Italy OUTLINE INTRODUCTION FABRICATION OF NANOHOLES GEOMETRICAL CHARACTERIZATION LEAK DEVICES RESULTS: PTB INRIM
More information