Stray Light Rejection in Array Spectrometers

Size: px
Start display at page:

Download "Stray Light Rejection in Array Spectrometers"

Transcription

1 Stray Light Rejection in Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington, Middlesex, UK 1

2 Overview Basic optical design of an array spectrometer In system (heterochromatic) stray light Techniques for quantifying stray light rejection Results: stray light errors in some commercially available spectrometers. Methods for reducing stray light errors Application of a modified array spectrometer: New NPL Goniospectroradiometer Some other important performance parameters for array spectrometers Conclusions and future work. 2

3 Basic optical layout of an array spectrometer Entrance slit Collimating mirror Diffraction grating Detector array Focussing mirror 3

4 In system (heterochromatic) stray light Often the dominant source of uncertainty in measurements made using compact array spectrometers. Rays strike the wrong part of the detector array causing spurious measured signals at the wrong wavelength. Distinguished from ambient (homochromatic) stray light. 4

5 Causes of stray light errors in array spectrometers Scattering from optical surfaces Interreflections between surfaces particularly reflections off detector array Inadequate blocking of other diffracted orders 5

6 Stray light errors are source dependent ojects.html Spectral Total Flux of a Fluorescent Lamp Deuterium lamp spectrum Relative SPD of Four LEDs Spectral Total Flux of a Tungsten Halogen Lamp E E Spectral Total Flux (arb. units) Spectral Irradiance (W/m^2/nm) 7.00E E E E E E E-04 Spectral Total Flux (arb. units) 1.00E E E E E+02 LED1 LED2 LED3 LED4 Spectral Total Flux (arb. units) E E Stray light errors tend to be most critical when measuring a broadband spectrum with an intensity varying over several orders of magnitude. 6

7 Stray light signal observed using a laser line Dark Corrected measured signal (normalised to max) 1.E+00 1.E-01 1.E-02 1.E-03 1.E-04 1.E-05 1.E Pixel no. Measured Spectra Ideal spectra Background due to heterochromatic stray light These results could be used to state that stray light rejection is < 10-5 some distance away from the centre wavelength of the laser line. However this does not tell us what errors to expect when measuring a broadband light source. 7

8 Stray light errors for an incandescent source 0.03 Spectral Total Flux of a Tungsten Halogen Lamp Spectral Total Flux (arb. units) Relatively low spectral flux at shorter visible and UV wavelengths Relatively high spectral flux at longer visible and NIR wavelengths 8

9 Stray light errors for an incandescent source 0.03 Spectral Total Flux of a Tungsten Halogen Lamp Spectral Total Flux (arb. units) W avelength (nm ) Small fraction of radiation inside the spectrometer is measured as heterochromatic stray light 9

10 Stray light errors for broadband light sources Problem is often exacerbated by spectral responsivity of detector array. e.g. Spectral responsivity of a silicon based detectors tends to be higher at longer wavelengths. 10

11 Quantifying stray light errors Measurement of lamp signal, V lamp (λ) Fibre input to spectrometer Background corrected Signal Measured from a Quartz Tungsten Lamp 1.E+07 G. R. Hopkinson, T. M. Goodman and S. R. Prince, A guide to the use and calibration of detector array equipment (SPIE Press Book), SPIE (2004). Background corrected signal (counts) 1.E+06 1.E+05 1.E+04 1.E

12 Quantifying stray light errors Measurement through cut on filter, V filter (λ) Fibre input to spectrometer Background corrected Signal Measured from a Quartz Tungsten Lamp Through a GG435 cut on Filter 1.E+07 Transmittance (%) Nominal transmittance of GG435 (3mm thickness) cut on filter 1.E E-01 1.E-02 1.E-03 1.E-04 Background corrected signal (counts) 1.E+06 1.E+05 1.E+04 1.E

13 Quantifying stray light errors Shutter to block light source from spectrometer field of view Fibre input to spectrometer Background Signal 1.E+07 Measurement of background signal, V bg (λ) signal (counts) 1.E+06 1.E+05 1.E+04 1.E

14 Analysis of stray light data For an ideal spectrometer: T filter ( λ) = V V filter lamp ( λ) V ( λ) V bg bg ( λ) ( λ) Stray light signals cause deviations from ideal behaviour and indicate erroneously high filter transmittance at wavelengths shorter than the cut on. Transmittance (%) 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% Transmittance of GG435 glass filter (3mm thickness) Measured using array spectrometer Nominal Stray light error of > 90%! 14

15 Comparison of Different Array Detectors The cut on filter method provides a way to compare the performance of different array spectrometers for measuring the spectral irradiance from a broadband light source. Transmittance of GG435 measured using a quartz Tungsten lamp and different array detectors 100% Measured transmittance (%) 10% spectrometer A Spectrometer B Spectrometer C Spectrometer E Spectrometer F Spectrometer G Spectrometer H GG435 Nominal 1%

16 How to handle stray light? Spectral Total Flux (arb. units) Live with it Correct for it Reduce it Spectral Total Flux of a Tungsten Halogen Lamp Determine stray light contribution to measurement uncertainty. Minimise the effect of stray light by calibrating the detector under conditions as close as possible to those under which it will be used. Match the F/# of input beam to F/# of spectrometer. Stray light errors are too large for many applications

17 How to handle stray light? Live with it Correct for it Reduce it Input laser radiation at different wavelengths into the array spectrometer to determine amount scattered onto each pixel as a function of wavelength stray light contribution to detector responsivity. Characterise the stray light rejection of the instrument and then correct for it. Dark Corrected measured signal (normalised to max) 1.E+00 Spectrum of HeNe Laser Measured Using Array Spectrometer 1.E-01 1.E-02 1.E-03 1.E-04 1.E-05 1.E Pixel no. S. W. Brown, B. C. Johnson, M. E. Feinholz, M. A. Yarbrough, S. J. Flora, K. R. Lykke, and D. K. Clark, Stray light correction algorithm for spectrographs, Metrologia 40, S81-83 (2003). Y. Zong, S. W. Brown, B. C. Johnson, K. R. Lykke, and Y. Ohno, Simple spectral stray light correction method for array spectroradiometers, Applied Optics, Vol 45 No. 6, 20 Feb

18 How to handle stray light? Live with it Correct for it Reduce it Limitations: This approach can be time consuming and expensive as it requires the use of laser radiation over a large wavelength band. The resulting correction may also be sensitive to changes in the spectrometer. Derive a spectral stray light correction matrix which can be applied to future measurements made with the spectrometer. Y Y meas IB = = [ I A 1 + D] Y Y meas IB = AY Has been shown to reduce some stray light errors by 1-2 orders of magnitude. IB 18

19 How to handle stray light? Live with it Correct for it Reduce it Use additional baffles inside spectrometer to block interreflections difficult to implement and many detectors are sealed. Use stray light blocking filters to limit the wavelengths of light reaching the detector array 19

20 Stray Light Blocking Filters Reduce the bandwidth of radiation reaching the spectrometer using bandpass filters. Spectral Total Flux of a Tungsten Halogen Lamp 0.03 Measure the spectrum over a reduced wavelength range without influence from stray light caused by scattering of other wavelengths. Spectral Total Flux (arb. units)

21 Application of stray light blocking filters: NPL Goniospectroradiometer An instrument to measure the spectral radiant intensity distribution, I e (λ, C, γ), of light sources. Spectral and luminous flux, chromaticity, CCT etc. derived from I e (λ, C, γ) Array spectrometer used primarily for speed of data acquisition and compact size. Shaw M J, Goodman T M, Array based goniospectroradiometer for measurement of spectral radiant intensity and spectral total flux of light sources, Applied Optics, vol 47 No. 13, 01 May

22 Implementation of stray light blocking filters approach at NPL Use data from cut on filter measurements to estimate stray light level through theoretical filters. Stray light signal Transmittance / total measured (%) signal 100% 100.0% 90% 80% 70% 60% 10.0% 50% 40% 30% 1.0% 20% 10% Estimated Gaussian Fractional Transmittance Stray Light Profiles Errors of Through Four Theoretical Four Theoretical Blocking Stray Light Filters Blocking Filters 0% 0.1% filter 1 filter 2 filter 3 filter 4 no filter Filter 1 Filter 2 Filter 3 Filter

23 Choice of stray light blocking filters Look at effect of filter FWHM and CWL on predicted stray light error Gaussian Transmittance Profiles of Four Theoretical Blocking Filters 100% 90% 80% Transmittance (%) 70% 60% 50% 40% 30% 20% 10% Filter 1 Filter 2 Filter 3 Filter 4 0% Theoretical filters with Gaussian transmittance 23

24 Real blocking filters Measured Transmittance of Four Real Blocking Filter Combinations Transmittance (%) 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% T Filter 1 T Filter 2 T Filter 3 T Filter 4 0% Blocking filters fitted into filter wheel behind spectrograph entrance slit 24

25 Stray light tests using blocking filters Transmittance of GG435 (3mm) Measured Without Blocking Filters Transmittance (%) 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% Nominal No blocking filter -10%

26 Stray light tests using blocking filters Significant reduction in 100% stray light signals 90% at short wavelengths. 80% 70% Transmittance of GG435 (3mm) Measured With Blocking Filters Transmittance (%) 60% 50% 40% 30% 20% 10% 0% Filt1 Filt2 Filt3 Filt4 Nominal -10% Increased noise at shorter wavelengths. 26

27 Limitations of using stray light blocking filters Increased measurement time. If using N blocking filters in a filter wheel then N different exposures are necessary + time to move filter wheel. Slightly reduced detector sensitivity (not significant if filters are well chosen). Temperature effects (need to be aware of temperature sensitivity of filter transmittance). 27

28 Implementation of stray light blocking filters Another implementation of the blocking filters is to coat them onto corresponding areas of the detector array, effectively blinding each pixel to radiation at wavelengths other than those which it should see. Introducing another optical component into the system will change its overall stray light characteristics. Model the optical system with the filters to determine their effect. 28

29 Results compact fluorescent lamp Luminous intensity distribution 29

30 Results compact fluorescent lamp Spectral total flux Scatter plot of chromaticity coordinates 30

31 Results - white LED cluster Spatially varying correlated colour temperature and chromaticity 31

32 Some other important performance characteristics of array spectrometers Wavelength accuracy Spectral resolution Linearity (Spectral) responsivity There are also many other important performance parameters to consider, including those relating to the detector array itself such as uniformity, well capacity, noise, etc. Apparatus for measurement of detector linearity. 32

33 Conclusions Array spectrometers often suffer from poor stray light rejection which can make them unsuitable for applications requiring a low measurement uncertainty. NPL have modified an array spectrometer to incorporate a series of custom designed stray light blocking filters. This instrument has been used to measure the spectral and spatial output characteristics of a variety of different light sources. 33

34 Future work Better understanding of uncertainties arising from stray light errors. Investigate use of stray light blocking filters further into the UV. Investigation into feasibility of monochromator based SL correction measurements at NPL. 34

35 Acknowledgements Thanks to colleagues in the optical technologies and scientific computing team at NPL and Teresa Goodman in particular for her help and advice. 35

36 Questions? Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington, Middlesex, UK Tel

Characterisation & Use of Array Spectrometers

Characterisation & Use of Array Spectrometers Characterisation & Use of Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington Middlesex, UK 1 Overview Basic design and features of

More information

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products White Paper Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products 1. Introduction The LED industry is growing rapidly and this naturally brings up an important need for reliable

More information

REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS

REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS S. G. R. Salim a,, N. P. Fox a, E. R. Woolliams a. R. Winkler a, H. M. Pegrum a, T. Sun b, K. T. V. Grattan b a National Physical

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately

More information

High Resolution Optical Spectroscopy

High Resolution Optical Spectroscopy PHYS 3719 High Resolution Optical Spectroscopy Introduction This experiment will allow you to learn a specific optical technique with applications over a wide variety of phenomena. You will use a commercial

More information

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Spectroscopy. Page 1 of 8 L.Pillay (2012) Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has

More information

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,

More information

Estimating the spectral response function of the CASI-2

Estimating the spectral response function of the CASI-2 Estimating the spectral response function of the CASI-2 K. -Y. CHOI and E. J. MILTON School of Geography, University of Southampton, UK, SO17 1BJ Abstract The spectral response function (SRF) of an instrument

More information

Spectral and photopic studies for high intensity discharge (HID) lamps

Spectral and photopic studies for high intensity discharge (HID) lamps Int. J. Metrol. Qual. Eng. 6, 105 (2015) c EDP Sciences 2015 DOI: 10.1051/ijmqe/2015005 Spectral and photopic studies for high intensity discharge (HID) lamps A.-E.A. Abd-Elmageed and E.M. El-Moghazy National

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

HARP Assessment of Uncertainty

HARP Assessment of Uncertainty HARP Assessment of Uncertainty The HIAPER Airborne Radiation Package (HARP) was designed to produce accurate measurements of actinic flux and irradiance. The Atmospheric Radiation Group (ARG) at the University

More information

Basics of UV-Visible Spectroscopy *

Basics of UV-Visible Spectroscopy * OpenStax-CNX module: m34525 1 Basics of UV-Visible Spectroscopy * Brittany L. Oliva-Chatelain Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution

More information

Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong

Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong CORM 2007 Annual Conference, Gaithersburg, Maryland, May 8-11, 2007 Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong Outline 1) Introduction 2) Principle of spatial

More information

Brewer UV measurement and calibration

Brewer UV measurement and calibration Brewer UV measurement and calibration Julian Gröbner Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC) 14/10/2014 Brewer UV Measurements and calibration 1 Overview Comparison

More information

UV Fluorescence from Integrating Spheres Measurement and Theory

UV Fluorescence from Integrating Spheres Measurement and Theory UV Fluorescence from Integrating Spheres Measurement and Theory Ping-Shine Shaw, Zhigang Li, Uwe Arp, and Keith R. Lykke National Institute of Standards and Technology, Gaithersburg, MD, USA Outline: 1.

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

Calibration of Ocean Colour Sensors

Calibration of Ocean Colour Sensors Dr. A. Neumann German Aerospace Centre DLR Remote Sensing Technology Institute Marine Remote Sensing What is Calibration, why do we need it? Sensor Components Definition of Terms Calibration Standards

More information

Uncertainty determination of correlated color temperature for high intensity discharge lamps

Uncertainty determination of correlated color temperature for high intensity discharge lamps Uncertainty determination of correlated color temperature for high intensity discharge lamps A.B. El-Bialy 1, M.M. El-Ganainy 2 and E.M. El-Moghazy 3 1 University College for Woman for Art, science and

More information

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages ) Spectroscopy: Introduction Required reading Chapter 18 (pages 378-397) Chapter 20 (pages 424-449) Spectrophotometry is any procedure that uses light to measure chemical concentrations Properties of Light

More information

application note LED measurement radiant power measurement

application note LED measurement radiant power measurement application note LED measurement radiant power measurement Contents 1 Introduction... 3 2 Setup... 4 2.1 Introduction... 4 2.2 Measurement setup... 4 2.3 Integrating spheres... 5 3 LED measurement procedure...

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Optical Properties of Binoculars and Telescopes Relevant to Laser Safety

Optical Properties of Binoculars and Telescopes Relevant to Laser Safety ILSC 2001 Conference Proceedings Optical Properties of Binoculars and Telescopes Relevant to Laser Safety Karl Schulmeister, Herbert Hödlmoser, Helmut Schön and Volker Stübler Please register to receive

More information

Analytical Spectroscopy Review

Analytical Spectroscopy Review Analytical Spectroscopy Review λ = wavelength ν = frequency V = velocity = ν x λ = 2.998 x 10 8 m/sec = c (in a vacuum) ν is determined by source and does not change as wave propogates, but V can change

More information

Uncertainty determination of correlated color temperature for high intensity discharge lamps.

Uncertainty determination of correlated color temperature for high intensity discharge lamps. Uncertainty determination of correlated color temperature for high intensity discharge lamps A.B. El-Bialy 1, M.M. El-Ganainy 2 and E.M. El-Moghazy 3 1 University College for Woman for Art, science and

More information

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007 Novel Flux Calibration Source CORM 2007 David C. Gross May 10, 2007 LED Calorimetry Absolute Radiometry Design Goals Proof of Concept - Design and Results Total Spectral Flux Calibration Absolute Radiometry

More information

MERIS US Workshop. Instrument Characterization Overview. Steven Delwart

MERIS US Workshop. Instrument Characterization Overview. Steven Delwart MERIS US Workshop Instrument Characterization Overview Steven Delwart Presentation Overview On-Ground Characterisation 1. Diffuser characterisation 2. Polarization sensitivity 3. Optical Transmission 4.

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

Analysis of Stray Light in a Brewer Spectrophotometer Brewers are not Perfect!

Analysis of Stray Light in a Brewer Spectrophotometer Brewers are not Perfect! Analysis of Stray Light in a Brewer Spectrophotometer Brewers are not Perfect! C. A. McLinden, D. I. Wardle, C. T. McElroy, &V. Savastiouk Environment Canada Stolen Stuff: David Wardle many slides Tom

More information

Introduction to FT-IR Spectroscopy

Introduction to FT-IR Spectroscopy Introduction to FT-IR Spectroscopy An FT-IR Spectrometer is an instrument which acquires broadband NIR to FIR spectra. Unlike a dispersive instrument, i.e. grating monochromator or spectrograph, an FT-IR

More information

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal. L A B O R A T O R Y C O M P E T E N C E Compact Knowledge: Absorbance Spectrophotometry Flexible. Reliable. Personal. The interaction of light with molecules is an essential and well accepted technique

More information

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation.

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation. Problem Solving picture θ f = 10 m s =1 cm equation rearrange numbers with units θ factors to change units s θ = = f sinθ fθ = s / cm 10 m f 1 m 100 cm check dimensions 1 3 π 180 radians = 10 60 arcmin

More information

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION G. Thuillier 1, D. Bolsee 2, T. Foujols 1 1 LATMOS-CNRS, France 2 Institut d Aéronomie

More information

A very versatile, large A-omega, fibre-fed spectrograph design. Ian Parry IoA, Cambridge

A very versatile, large A-omega, fibre-fed spectrograph design. Ian Parry IoA, Cambridge A very versatile, large A-omega, fibre-fed spectrograph design Ian Parry IoA, Cambridge 1 But first a quick diversion to support Alvio s case NIR multi-object spectroscopy with fibres works! CIRPASS was

More information

USE OF A LOW-COST SILICON DIODE-ARRAY SPECTRORADIOMETER TO MEASURE SOLAR UV-B

USE OF A LOW-COST SILICON DIODE-ARRAY SPECTRORADIOMETER TO MEASURE SOLAR UV-B USE OF A LOW-COST SILICON DIODE-ARRAY SPECTRORADIOMETER TO MEASURE SOLAR UV-B Terence M. Murphy, Section of Plant Biology, University of California, Davis, CA 95616 Presented at an NIST Conference on Critical

More information

Laser induced fluorescence

Laser induced fluorescence Report experiment 1 by 4-7 February 22, id#471917 Summary Several measurements are made around laser induced fluorescence (LIF). First the system was calibrated. The resolution of the spectrograph was

More information

Display Calibrations and Uncertainties. Christine Wall National Physical Laboratory

Display Calibrations and Uncertainties. Christine Wall National Physical Laboratory Display Calibrations and Uncertainties Christine Wall National Physical Laboratory Overview Why measure displays? What is measurement uncertainty? Measurement traceability NPL and UKAS What is an uncertainty,

More information

Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer. Jeff Bourne Majid Badiei (Advisor)

Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer. Jeff Bourne Majid Badiei (Advisor) Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer Jeff Bourne Majid Badiei (Advisor) What is a spectrometer? Any device that converts different input wavelengths into different

More information

Ultraviolet characterization of integrating spheres

Ultraviolet characterization of integrating spheres Ultraviolet characterization of integrating spheres Ping-Shine Shaw,* Zhigang Li, Uwe Arp, and Keith R. Lykke Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, Maryland

More information

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13) An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13) Beer s Law: A = -log T = -logp 0 / P = e x b x C See Table 13-1 for terms. In measuring absorbance or transmittance, one should

More information

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Course Details September 22 October 10 September 22 November 7 November 17 December 1 Topic Period Introduction to Spectrometric

More information

Instrumental Analysis: Spectrophotometric Methods

Instrumental Analysis: Spectrophotometric Methods Instrumental Analysis: Spectrophotometric Methods 2007 By the end of this part of the course, you should be able to: Understand interaction between light and matter (absorbance, excitation, emission, luminescence,fluorescence,

More information

Reduction procedure of long-slit optical spectra. Astrophysical observatory of Asiago

Reduction procedure of long-slit optical spectra. Astrophysical observatory of Asiago Reduction procedure of long-slit optical spectra Astrophysical observatory of Asiago Spectrograph: slit + dispersion grating + detector (CCD) It produces two-dimension data: Spatial direction (x) along

More information

Spectrophotometric calibration system for DECam

Spectrophotometric calibration system for DECam Spectrophotometric calibration system for DECam J.-P. Rheault*, D. L. DePoy, J. L. Marshall, T. Prochaska, R. Allen, J. Wise, E. Martin, P. Williams Department of Physics and Astronomy, Texas A&M University,

More information

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Bilateral Comparison of between NMIA (Australia) and SPRING (Singapore) (KCDB reference No. CCPR-K1.a.1) Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Contents 1. Introduction..2 2.

More information

Overview of Spectroscopy

Overview of Spectroscopy Overview of Spectroscopy A. Definition: Interaction of EM Radiation with Matter We see objects because they remit some part of the light falling on them from a source. We function as reflection/ transmission

More information

Short Term Scientific Mission Report COST Action OPTIMISE: ES1309

Short Term Scientific Mission Report COST Action OPTIMISE: ES1309 Short Term Scientific Mission Report COST Action OPTIMISE: ES1309 STSM Applicant: Dr. Laura MIHAI, laura.mihai@inflpr.ro National Institute for Laser, Plasma and Radiation Physics / CETAL-PhIL, Atomistilor

More information

EXPOSURE TIME ESTIMATION

EXPOSURE TIME ESTIMATION ASTR 511/O Connell Lec 12 1 EXPOSURE TIME ESTIMATION An essential part of planning any observation is to estimate the total exposure time needed to satisfy your scientific goal. General considerations

More information

FLS980 Series Reference Guide

FLS980 Series Reference Guide FLS980 Series Reference Guide Integrating Sphere for Measurements of Fluorescence Quantum Yields and Spectral Reflectance Revision 1 Copyrights Copyright 2016 Edinburgh Instruments Ltd. All rights reserved.

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

2.01 INFRARED ANALYZER

2.01 INFRARED ANALYZER NIR INFORMATION PAGE 1 OF 5 What does it do? First, it allows for better control and knowledge of incoming ingredients. If the ability exists to analyze a truck or rail car immediately upon arrival, the

More information

Report No.: HZ p. (Lumens) Stabilization Time CRI (K)

Report No.: HZ p. (Lumens) Stabilization Time CRI (K) Test Summary Sample Tested: ML2LA23LTRBNI827 Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 73.6 1737.0 23.61 0.9872 CCT Stabilization Time CRI (K) (Light & Power)

More information

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is Optical Spectroscopy Lennon O Naraigh, 0000 Date of Submission: 0 th May 004 Abstract: This experiment is an exercise in the principles and practice of optical spectroscopy. The continuous emission spectrum

More information

Spectroscopic studies of impurities in the LHD plasmas

Spectroscopic studies of impurities in the LHD plasmas Spectroscopic studies of impurities in the LHD plasmas Visitor: Zhenwei Wu (the institute of plasma physics, CAS -ASIPP) Host: Shigeru Morita (the national institute for fusion science -NIFS) Content 1.

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 1. Examples of advantages and disadvantages with laser-based combustion diagnostic techniques: + Nonintrusive + High

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5. NIRSpec Performance Report NPR-2013-011 / ESA-JWST-RP-19649 Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.2013 Revision: 1 estec European Space Research and Technology Centre Keplerlaan

More information

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements EMRP-ENV03 Solar UV WP 3: Improvement of Reference Spectroradiometers Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements Peter Meindl, Christian

More information

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps Practical 1P4 Energy Levels and Band Gaps What you should learn from this practical Science This practical illustrates some of the points from the lecture course on Elementary Quantum Mechanics and Bonding

More information

Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths

Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths Claire Cramer, Steve Brown, Keith Lykke, John Woodward (NIST) Tom Stone (USGS) Motivation for using the Moon

More information

Quality Manual of Reference Spectrometer Laboratory

Quality Manual of Reference Spectrometer Laboratory Aalto University School of Electrical Engineering Metrology Research Institute Farshid Manoocheri Dmitri Lanevski Saulius Nevas Maksim Shpak Quality Manual of Reference Version 2.9 7/12/2017 Page 2 (33)

More information

Hanle Echelle Spectrograph (HESP)

Hanle Echelle Spectrograph (HESP) Hanle Echelle Spectrograph (HESP) Bench mounted High resolution echelle spectrograph fed by Optical Fiber Second generation instrument for HCT The project is a technical collaboration between Indian Institute

More information

Astronomy 203 practice final examination

Astronomy 203 practice final examination Astronomy 203 practice final examination Fall 1999 If this were a real, in-class examination, you would be reminded here of the exam rules, which are as follows: You may consult only one page of formulas

More information

NIST Role in Supporting Solid State Lighting Initiative

NIST Role in Supporting Solid State Lighting Initiative CORM 2007 Annual Meeting, Gaithersburg, May 8-11, 2007 NIST Role in Supporting Solid State Lighting Initiative Yoshi Ohno, Ph.D. Optical Technology Division National Institute of Standards and Technology

More information

LDLS Laser-Driven Light Source

LDLS Laser-Driven Light Source LDLS Laser-Driven Light Source From The Innovators in Light ENERGETIQ 7/8/2011 1 Peter J. Dwyer, Ph.D. pdwyer@energetiq.com +1 781 939 0763 x 141 The LDLS Product Range EQ-99FC LDLS System EQ-1500 LDLS

More information

Spectral changes of solar simulator xenon flash bulbs over lifetime and its effects on measurement uncertainty

Spectral changes of solar simulator xenon flash bulbs over lifetime and its effects on measurement uncertainty Loughborough University Institutional Repository Spectral changes of solar simulator xenon flash bulbs over lifetime and its effects on measurement uncertainty This item was submitted to Loughborough University's

More information

The Performance of the EUV Spectroscope (EXCEED) Onboard the SPRINT-A Mission

The Performance of the EUV Spectroscope (EXCEED) Onboard the SPRINT-A Mission The Performance of the EUV Spectroscope (EXCEED) Onboard the SPRINT-A Mission K. Yoshioka, G. Murakami, A. Yamazaki, K. Uemizu, T. Kimura (ISAS/JAXA), I. Yoshikawa, K. Uji (Univ. Tokyo) F. Tsuchiya, and

More information

UV Degradation of Polycarbonate

UV Degradation of Polycarbonate Utah State University DigitalCommons@USU Senior Theses and Projects Materials Physics 5-2017 UV Degradation of Polycarbonate Katie Gamaunt Utah State University Follow this and additional works at: https://digitalcommons.usu.edu/mp_seniorthesesprojects

More information

Transvision: a Light Transmission Measurement System for Greenhouse Covering Materials

Transvision: a Light Transmission Measurement System for Greenhouse Covering Materials Transvision: a Light Transmission Measurement System for Greenhouse Covering Materials G.L.A.M. Swinkels Wageningen UR Greenhouse Horticulture Wageningen The Netherlands Keywords: hemispherical, light

More information

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)

More information

Photobiological Safety of Luminaires: Refining the New Approach

Photobiological Safety of Luminaires: Refining the New Approach Photobiological Safety of Luminaires: Refining the New Approach Leslie Lyons Bentham Instruments Limited Reading, UK llyons@bentham.co.uk We are all familiar with the visual characteristics of lighting

More information

Improved Process Measurement & Control. OD 600 vs. AU Measurements in Fermentation Cell Density

Improved Process Measurement & Control. OD 600 vs. AU Measurements in Fermentation Cell Density Finesse, LLC 3350 Scott Boulevard #1 Santa Clara, CA 95054 9351 Irvine Boulevard Irvine, CA 92618 800-598-9515 www.finesse-inc.com P01 OD 600 vs. AU Measurements in Fermentation Cell Density This technical

More information

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Spectrophotometry Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Content Introduction Beer-Lambert law Instrument Applications Introduction 3 Body fluids such as blood, csf and urine contain organic and

More information

EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES

EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES CISBAT 2005, Proceedings, EPFL 2005, p. 441-446 EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES R. Steiner, P. Oelhafen, G. Reber and A. Romanyuk Institute

More information

Thermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement

Thermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement Thermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement Appendix A Outlines of GOSAT and TANSO Sensor GOSAT (Greenhouse

More information

Correcting for bandwidth in monochromator measurements

Correcting for bandwidth in monochromator measurements Correcting for bandwidth in monochromator measurements Emma R. Woolliams, Maurice G. Cox, Peter M. Harris Presentation to: ORM Club 8 th June 6 Yep, maths again ( ) ( ) I( λ) = I ( λ) Δ λβi ( λ) + Δλ β

More information

Chem 310 rd. 3 Homework Set Answers

Chem 310 rd. 3 Homework Set Answers -1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption

More information

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps Practical 1P4 Energy Levels and Band Gaps What you should learn from this practical Science This practical illustrates some of the points from the lecture course on Elementary Quantum Mechanics and Bonding

More information

C101-E111. Talk Letter. Vol.2 February 2009

C101-E111. Talk Letter. Vol.2 February 2009 C101-E111 UV Talk Letter Vol.2 February 2009 UV Talk Letter UV Talk Letter The Structure of a Spectrophotometer Vol.2 February 2009 1.The Measurement Principle Used by a Spectrophotometer The basic measurement

More information

Integrating Spheres in Molecular Spectrophotometry

Integrating Spheres in Molecular Spectrophotometry Integrating Spheres in Molecular Spectrophotometry Theory and Practice 2012 Perkin Elmer 2012 Perkin Elmer General Sphere Theory 3 Integrating Spheres Types of Sphere Measurements Total Reflectance (Specular

More information

CALIBRATION OF SPECTRORADIOMETERS FOR OUTDOOR DIRECT SOLAR SPECTRAL IRRADIANCE MEASUREMENTS

CALIBRATION OF SPECTRORADIOMETERS FOR OUTDOOR DIRECT SOLAR SPECTRAL IRRADIANCE MEASUREMENTS CALIBRATION OF SPECTRORADIOMETERS FOR OUTDOOR DIRECT SOLAR SPECTRAL IRRADIANCE MEASUREMENTS Paraskeva Vasiliki 1*, Matthew Norton 1,2, Maria Hadjipanayi 1, George E. Georghiou 1 1 Photovoltaic Technology,

More information

JABLONSKI DIAGRAM 2/15/16

JABLONSKI DIAGRAM 2/15/16 INDICATE THE EXCITED AND GROUND SINGLET AND TRIPLET STATES. INDICATE THE FOLLOWING TRANSITIONS: ABSORPTION, FLUORESCENCE, PHOSPHORESCENCE, NONRADIATIVE DECAY, INTERNAL CONVERSION AND INTERSYSTEM CROSSING.

More information

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors ISSN: 2347-325 Volume 3 Number 6 (June-205) pp. 264-274 www.ijcrar.com Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors Manal A. Haridy* Photometry and

More information

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Electromagnetic Spectrum - Molecular transitions Widely used in chemistry. Perhaps the most widely used in Biological Chemistry.

More information

COLORIMETER AND LAMBERT S-BEER S LAW. Shingala vaishali Sandha prafulla Tiwari Kuldeep

COLORIMETER AND LAMBERT S-BEER S LAW. Shingala vaishali Sandha prafulla Tiwari Kuldeep COLORIMETER AND LAMBERT S-BEER S LAW Shingala vaishali Sandha prafulla Tiwari Kuldeep TOPIC What is colorimeter? Use of colorimeter. Component & It s function. Function of colorimeter. The principle of

More information

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision Clinical Chemistry (CHE221) Professor Hicks Week 1 Statistics Made Slightly Less Boring and Introduction to Spectrophotometry 3 Accuracy vs Precision Precision is the consistency of a measurement made

More information

Emissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites

Emissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites Thermographie-Kolloquium 2017 More info about this article: http://www.ndt.net/?id=22484 Emissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites Albert ADIBEKYAN

More information

Astronomical frequency comb for calibration of low and medium resolution spectrographs

Astronomical frequency comb for calibration of low and medium resolution spectrographs Astronomical frequency comb for calibration of low and medium resolution spectrographs innofspec at AIP has several years expertise in astronomical instrumentation. innofspec succesfully developed a new

More information

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer C101-E107B UV-3600 Shimadzu UV-VIS-NIR Spectrophotometer High sensitivity, high resolution, and an ultra-low stray-light level achieved with the latest technology lead the way to new solutions. UV-3600

More information

New Method for Realizing a Luminous Flux Scale using an Integrating Sphere with an External Source

New Method for Realizing a Luminous Flux Scale using an Integrating Sphere with an External Source Paper for IESNA Annual Conference, Miami, FL, Aug. 994, and published in J. Illum. Eng. Soc. 24(), 06-5 (995). New Method for Realizing a Luminous Flux Scale using an Integrating Sphere with an External

More information

Absorption photometry

Absorption photometry The light Absorption photometry Szilvia Barkó University of Pécs, Faculty of Medicines, Dept. Biophysics February 2011 Transversal wave E Electromagnetic wave electric gradient vector wavelength The dual

More information

Impact of New SSL Standards on Goniospectroradiometric Measurements

Impact of New SSL Standards on Goniospectroradiometric Measurements Impact of New SSL Standards on Goniospectroradiometric Measurements By Dr. Ðenan Konjhodžić, Application Engineer Presented by Justin Blanke, International Sales Manager Outline Goniophotometry and goniophotometer

More information

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract:

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract: NIRSpec Technical Note NTN-2013-014 Authors: P. Ferruit & G. Giardino Date of Issue: 29.11.2013 Version: 1 estec European Space Research and Technology Centre Keplerlaan 1 2201 AZ Noordwijk The Netherlands

More information

Overview of NRC Photometry and Radiometry Measurement Services

Overview of NRC Photometry and Radiometry Measurement Services Overview of NRC Photometry and Radiometry Measurement Services Nelson Rowell Measurement Science and Standards National Research Council of Canada Personalize with title, slogan or I/B/P name in master

More information

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Phys316 Exploration 2: Verifying Stefan-Boltzmann Relationship Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Where A is the effective radiating area,

More information

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A LM-79-08 Test Report for MLS CO., LTD NO.1 MULINSEN RD XIAOLAN TOWNSHIP ZHONGSHAN City, GUANGDONG Linear Ambient Luminaires Model: MT8N20-A Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0

More information