Ultraviolet characterization of integrating spheres

Size: px
Start display at page:

Download "Ultraviolet characterization of integrating spheres"

Transcription

1 Ultraviolet characterization of integrating spheres Ping-Shine Shaw,* Zhigang Li, Uwe Arp, and Keith R. Lykke Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, Maryland 20899, USA *Corresponding author: Received 20 February 2007; accepted 26 March 2007; posted 10 April 2007 (Doc. ID 80231); published 9 July 2007 We have studied the performance of polytetrafluoroethylene integrating spheres in the ultraviolet (UV) region with wavelengths as short as 200 nm. Two techniques were used for this study; first, the spectral throughput of an integrating sphere irradiated by a deuterium lamp was analyzed by a monochromator. Second, a UV laser beam was directed into an integrating sphere, and spectrally dispersed laser induced fluorescence was studied. Significant absorption and fluorescence features were observed in the UV region and attributed to the contamination in the integrating sphere. We demonstrate that integrating spheres are easily contaminated by environmental pollutants such as polycyclic aromatic hydrocarbons emitted from engine exhaust. Baking of the contaminated integrating sphere can reverse some but not all of the effects caused by contaminants. The implications for using integrating spheres for UV measurement are discussed. OCIS codes: , , , Introduction Integrating spheres play an indispensable role when it comes to diffusing and depolarizing radiation. They are widely used in radiometric applications because of their superior performance in transforming radiation to a nearly ideal Lambertian distribution despite variations in the conditions of the incident radiation. In the visible and near ultraviolet (UV) region, polytetrafluoroethylene (PTFE) has been proven to be the best material for integrating spheres because of high reflectivity and chemical inertness. Integrating spheres made with sintered PTFE are commercially available from many manufacturers. For UV applications, however, concerns over the performance of diffusing materials, such as PTFE, were first raised three decades ago [1] when integrating spheres were used for spectral irradiance measurements of deuterium lamps. When integrating spheres were irradiated by UV radiation from deuterium lamps, spurious effects in the measured spectral irradiance were observed and attributed to fluorescence from diffusing materials induced by UV irradiation. A later study [2] confirmed the fluorescence effect from PTFE powder and suggested that contaminants in the powder were responsible for fluorescence. Since then only a few studies on the UV reflectance and fluorescence of diffusing materials have been reported [3 7] and, to the best knowledge of the authors, no UV work on integrating spheres has been documented. In one study of PTFE as a diffuse reflectance material for space-borne remote sensing application [3,4], contaminants were again shown to be responsible for the degradation in the reflectance of PTFE samples when irradiated with radiation from a Xe lamp. By using several chemical analysis techniques, the nature of the contaminants was identified only as volatile hydrocarbon impurities intrinsic to the commercial material. Furthermore, it was also shown that contaminants in PTFE samples can be cleaned by baking in vacuum although a clean PTFE sample can be contaminated again when exposed to aromatic hydrocarbons. Later, more studies reported measurements of the degradation in the hemispherical reflectance of PTFE material irradiated by Xe or Hg lamps [5 7] with no mention of its implication for PTFE integrating spheres. The correlation between the hemispherical reflectance of PTFE material and the throughput of a PTFE integrating sphere is given by the following expression [8] 1 August 2007 Vol. 46, No. 22 APPLIED OPTICS 5119

2 r 2 A 2 1 A 1 A 2 out in 1 r 1 A 1 A, 2 where in and out are the flux entering and exiting an integrating sphere, A 1 and A 2 are the ratio of the area of the entrance and exit port to the total inner surface area of the sphere, and r is the hemispherical reflectance of the diffusing material. At high reflectance and small port area, as is the case for a typical integrating sphere, the throughput is highly sensitive to small variations in r. Given the reported vulnerability to degradation in the UV hemispherical reflectance of PTFE, it is expected that the degradation in the throughput of a PTFE integrating sphere will be much more pronounced. The instability of the UV throughput of an integrating sphere could result in large errors in long-term measurements. In addition to throughput, fluorescence from an integrating sphere is even more challenging to UV measurement. Fluorescence, generally induced by incident radiation from a wavelength region to the blue, introduces spurious spectral effects and leads to measurement errors that are difficult to identify. A study of the UV properties of integrating spheres in both throughput and florescence will have important implications for radiometric measurements in the UV, such as source calibration, reflectance standards, and remote sensing [9]. Integrating spheres made with PTFE are formed by two major techniques. Most commercially available products are made with sintered PTFE and many laboratories, including the National Institute of Standards and Technology (NIST), also construct their own integrating spheres by pressing PTFE powder onto a spherical surface. Sintered PTFE integrating spheres are more structurally sound but concerns have been raised about the resin used to form the spherical shape [3]. Here, we report the characterization of both sintered and pressed PTFE integrating spheres in the UV to identify the nature of an integrating sphere s absorption and fluorescence properties using two techniques. First, the integrating sphere under study is irradiated by a deuterium lamp, and the relative spectral throughput of the integrating sphere is measured using a monochromator system. Second, the integrating sphere is irradiated with a beam from a tunable UV laser, and the laser induced fluorescence (LIF) is analyzed by a spectrometer. The former technique measures the absorption spectra and total fluorescence while the latter provides fluorescence spectra excited by a fixed wavelength. The combination of absorption and fluorescence spectra provides insight into the chemical nature of any contaminants responsible for fluorescence. Many commercially available sintered PTFE integrating spheres, along with home-made pressed PTFE integrating spheres, were measured using both techniques. We studied the role of contaminants for the observed absorption and fluorescence as well as the degradation of integrating spheres under UV irradiation. We also studied integrating spheres with controlled exposure to environmental pollutants, such as gasoline and diesel exhaust from engines. The effect of baking integrating spheres in both vacuum and air on contaminants is discussed. Detailed results and analysis as well as some recommendations on using integrating spheres for UV work are given. 2. Experimental Setup In characterizing integrating spheres, two UV sources were used to probe the integrating sphere under study; a deuterium lamp and a tunable laser system. The deuterium lamp emits continuous radiation in the UV for the study of spectral throughput while the laser system provides several discrete wavelengths for LIF study. By using these two sources, we investigated both pressed and sintered PTFE integrating spheres, with the former made at NIST using PTFE powders and the latter from several different manufacturers. For better comparison, all of the spheres investigated were chosen with an inner diameter 3.8 cm and 1 cm diameter for both entrance and exit apertures. Pristine integrating spheres were also subject to different treatments and studied using both techniques. These treatments include baking of the integrating sphere, prolonged exposure to a deuterium lamp for stability study, and exposure to engine exhaust to assess the effect of environmental pollutions. A. Spectral Throughput Measurements The spectral throughput of an integrating sphere was measured by utilizing part of a UV irradiance calibration system at the Facility for Irradiance Calibration Using Synchrotrons (FICUS) at NIST [10,11]. As shown in Fig. 1, a deuterium lamp was placed at a typical distance of 30 cm from the entrance port of the integrating sphere under study. The spectral distribution of the exiting radiation from the integrating sphere was measured with a 0.25 m Czerny Turner monochromator coupled to a temperature-stabilized photomultiplier tube. The bandwidth of the monochromator was set at 5 nm and the wavelength range was scanned between 200 and 400 nm for this work. Order-sorting filters were not necessary for this Fig. 1. Schematic diagram of the experimental setup for measurement of integrating sphere spectral throughput APPLIED OPTICS Vol. 46, No August 2007

3 Fig. 3. Schematic diagram of the experimental setup for the LIF measurements on integrating spheres. Fig. 2. Spectral output with the deuterium lamp irradiating the monochromator entrance directly. This signal was used to normalize spectral outputs measured with integrating spheres to obtain the relative throughput of the integrating spheres as explained in the text. range since wavelengths shorter than 200 nm were eliminated by air and window absorption. The measured spectral distribution using the above setup consists of several spectral effects, such as the output of the deuterium lamp, the throughput of the integrating sphere, the throughput of the monochromator, and the responsivity of the photomultiplier tube. To isolate the throughput of the integrating sphere, a separate measurement was performed by removing the integrating sphere and placing the deuterium lamp directly in front of the entrance slit of the monochromator. The resulting spectral output, as shown in Fig. 2, was used to normalize all spectral distributions measured with an integrating sphere. While this normalization process does not yield absolute throughput of the integrating sphere under study at a specific wavelength, the normalized spectral distribution is proportional to the spectral throughput of the integrating sphere by a constant throughout the wavelength region of measurements. Absorption structures in the integrating spheres can be easily identified from normalized spectral distributions. B. Laser Induced Fluorescence Measurements Tunable lasers at the NIST facility for Spectral Irradiance and Radiance Responsivity Calibrations using Uniform Sources [12] were used for the study of LIF from integrating spheres. The tunable laser, with a range of wavelength from 200 to 400 nm for this work, was generated by frequency doubling, tripling, and quadrupling of a quasi-continuous wave (cw) mode-locked Ti:sapphire laser. The quasi-cw output, at 76 MHz and 2 ps duration, was in the mw power range. The wavelength uncertainty was less than 0.01 nm. The setup used to measure fluorescence from an integrating sphere is very similar to the setup used to measure the spectral throughput using a deuterium lamp discussed previously except that the output radiation from the integrating sphere was measured by a commercial charge coupled device (CCD)-array spectrograph instead of the monochromator as shown in Fig. 3. Because of the high input laser power irradiating the integrating sphere, the CCD-array spectrograph was used to achieve a wider spectral range (from 200 to 800 nm), higher resolution ( 3 nm in bandwidth), and shorter acquisition time. An optical fiber directed the radiation from the exit port of the integrating sphere to the entrance port of the spectrograph. Inside the spectrograph, dispersed radiation was detected by a two-dimensional cooled 10 C CCD array with a total of pixels. The output of the array detector was binned along the column with 128 pixels, and the resolution of the instrument was 15 bits. The measured spectral distribution using the LIF setup with fixed wavelength laser excitation consists of fluorescence from the integrating sphere and the primary exciting laser line on top of two slowly decaying wings around the primary laser peak (i.e., line spread function) contributed by scattered stray light inside the spectrograph [11,13,14]. Shown in Fig. 4 is a typical line spread function measured with a 360 nm laser beam. For better comparison between different measurements, all spectral distributions are normalized to the peak at the wavelength of the excitation laser. Fig. 4. Typical line spread function of the spectrograph with the laser wavelength of 360 nm. 1 August 2007 Vol. 46, No. 22 APPLIED OPTICS 5121

4 3. Results To have a general evaluation on the impact of absorption and fluorescent effects of integrating spheres on the UV radiometry, we characterized a number of newly acquired commercial integrating spheres from several manufacturers. For comparison, we also studied some commercial integrating spheres that have been extensively used in our laboratory. Pressed integrating spheres were not commercially available and were packed at NIST using PTFE powders of Halon [15]. The effect of baking integrating spheres was studied at length. We also measured integrating spheres that were exposed to environmental contamination to shed light on the causes of UV absorption and fluorescence. Finally, we discuss the temporal stability of integrating spheres. Fig. 6. LIF measurements of (a) a pressed PTFE integrating sphere and (b) (d) commercial sintered PTFE integrating spheres with the excitation laser tuned at 220 nm. All curves are normalized to 1 at the laser excitation wavelength of 220 nm. For easy viewing, there is an offset to each curve. The offsets are 0 for (a), for (b), for (c), and for (d). A. Pristine Integrating Spheres Shown in Fig. 5 are the results of spectral throughput measurements performed on several sintered and pressed integrating spheres. Note that all throughput curves were normalized with the throughput at 400 nm to account for the variation in the integrating sphere throughput owing to differences in sphere geometry. All curves in Fig. 5 display a general trend of gradual decrease in throughput toward shorter wavelengths in accordance with the reported decrease in reflectivity from PTFE plaques [2,9]. In addition to the gradual decay, all integrating spheres show the same two-peak-absorption structure below 300 nm. Note that this structure is very similar to the measured reflectance factor of sintered PTFE samples reported previously [9]. For wavelengths above 300 nm, the throughput is generally smooth, but some integrating spheres exhibited a broad peak. Of all integrating spheres shown in Fig. 5, the pressed integrating sphere has the smoothest spectral throughput with the absorption structure barely distinguishable, indicating that the absorption structure is contributed to by contaminants and not an intrinsic property of the PTFE material. In general, we found that pressed PTFE integrating spheres are much less infected with these absorption bands as well as fluorescence, as will be discussed later. This is again in accordance with previous observations [9]. To find the fluorescence properties of the integrating spheres under this study, we conducted LIF measurements for several integrating spheres used in Fig. 5. The measured fluorescence spectra with the excitation laser wavelength set at 220 nm are shown in Fig. 6. As in the case of the throughput spectra, the fluorescence spectra resemble each other in the wavelength region from 250 to 400 nm. In addition, two sintered PTFE integrating spheres exhibit a broad fluorescence band between 400 and 500 nm. In terms Fig. 5. Relative throughput measured with (a) a freshly pressed PTFE integrating sphere, and (b) (d) commercial sintered PTFE integrating spheres with (b) and (c) new from manufacturers when measured, while (d) had been used for UV radiometry prior to this measurement. All curves are normalized at 400 nm to 1. For easy viewing, there is an offset to each curve. The offsets are 0 for (a), 0.6 for (b), 0.8 for (c), and 1.1 for (d). Fig. 7. LIF measurements of a commercial sintered PTFE integrating sphere with the excitation laser tuned at (a) 210, (b) 220, (c) 240, (d) 266, (e) 293, and (f) 360 nm. For easy viewing, the peak of each curve at the wavelength of the excitation laser is normalized to a different value. The values are 10 7 for (a), 10 5 for (b), 10 3 for (c), 10 2 for (d), 10 1 for (e), and 1 for (f) APPLIED OPTICS Vol. 46, No August 2007

5 of fluorescence yield, we again found that the pressed PTFE integrating sphere with the least visible absorption band in throughput has the smallest fluorescent yield over other sintered PTFE integrating spheres. The fluorescence spectra of integrating spheres are also strongly dependent on the wavelength of the exciting laser beam as depicted in Fig. 7 for a sintered PTFE integrating sphere. The fluorescence between 250 and 400 nm decreases as the wavelength of the laser increases. On the other hand, the fluorescence band between 400 and 500 nm shows a distinct structure with laser wavelengths of 240, 266, and 360 nm. These fluorescence peaks have a nearly equal spacing of 1700 cm 1, indicating vibrational progression from different vibrational transitions. B. Effects of Baking Baking of integrating spheres in vacuum has been previously suggested as a way to remove contaminants [3]. With the characterization techniques used for this study, we were able to determine the effectiveness of baking integrating spheres both in vacuum and in air. We found that the effects of baking are easier to quantify using the LIF measurements. Shown in Fig. 8 is the LIF measurement using several exciting laser wavelengths performed on a sintered PTFE integrating sphere before and after baking in vacuum for 40 h at 90 C. It is clear from Fig. 8 that the fluorescence yield is much reduced after baking, but not eliminated for all wavelengths. Notice also that the reduction of fluorescence between 250 and 400 nm is much less than that between 400 and 500 nm. Overall, the remnants of fluorescence after baking still resemble the spectra before baking. Our tests show that longer baking time in vacuum does not completely eliminate the fluorescence. This fact suggests that the contaminants in an integrating sphere are somewhat volatile and baking can remove some, but not all, contaminants. For even longer laser wavelengths than those shown in Fig. 8, the fluorescence is relatively small and, by itself, cannot be easily separated from the line spread function of the monochromator. Nevertheless, the fluorescence can be readily identified by comparing the fluorescence spectra of integrating spheres before and after baking. Figure 9 shows such a comparison for a laser wavelength of 360 nm. The fluorescence between 400 and 500 nm before the bake is evident. Similar results, shown in Fig. 10, were obtained for baking a pressed PTFE integrating sphere. While we observed no discernible fluorescence between 400 and 500 nm, the reduction of fluorescence between 250 Fig. 8. LIF measurements of a sintered PTFE integrating sphere before and after baking the integrating sphere 40 h in vacuum. For each graph, the top curves are before baking and the bottom curves are after baking. All curves are normalized to 1 at the corresponding laser excitation wavelengths. Fig. 9. LIF measurements of a sintered PTFE integrating sphere before and after baking the integrating sphere 40 h in vacuum. The top curve is before baking and the bottom curve is after baking. 1 August 2007 Vol. 46, No. 22 APPLIED OPTICS 5123

6 Fig. 11. Relative throughput measurements of (a) a sintered PTFE integrating sphere before and after 65 h vacuum bake, (b) a sintered PTFE integrating sphere before and after 24 h air bake, and (c) a pressed PTFE integrating sphere before and after 40 h vacuum bake. Solid lines are before the bake, and open circles are after the bake. Fig. 10. LIF measurements of a pressed PTFE integrating sphere before and after baking the integrating sphere 40 h in vacuum. For each graph, the top curves are before baking and the bottom curves are after baking. All curves are normalized to 1 at the corresponding laser excitation wavelengths. and 400 nm due to baking is very similar to that of a sintered integrating sphere shown in Fig. 8. It is also of interest to investigate the throughput change of an integrating sphere after baking. Shown in Fig. 11 is the throughput of two sintered PTFE integrating spheres and a pressed PTFE integrating sphere before and after baking in vacuum and in air. It is evident that baking tends to smooth the throughput spectra from the absorption structure below 300 nm. Baking also improves the throughput in the short wavelength range; while in some regions on the longer wavelength side, throughput decreases. In view of the changes of the fluorescence spectra from baking, we can interpret this phenomenon as follows: the absorption structure below 300 nm is caused by the absorption of UV radiation from contaminants inside the integrating sphere. As the integrating sphere is baked, some of the contaminants leave the integrating sphere, which leads to improved UV throughput with reduced absorption structure. On the other hand, less contaminants cause decreased fluorescence, which predominantly fluoresce with wavelengths longer than 300 nm. Finally, we studied and compared the effects of baking an integrating sphere in vacuum and in air, which as a practical matter could be done much easier and less costly. For this comparison, we packed two integrating spheres with the same batch of Halon PTFE powder under the same procedure. One inte APPLIED OPTICS Vol. 46, No August 2007

7 Fig. 12. Relative changes in throughput for pressed integrating spheres after 24 h vacuum bake and after 105 h air bake. grating sphere was baked under vacuum at 90 C with a base pressure of 10 3 Pa; while the other was baked in an oven in air, also at 90 C. The changes in the spectral throughput for both integrating spheres are shown in Fig. 12 for 24 h vacuum baking and 105 h baking in air. Both spectra show increased throughput for the short wavelength side and decreased throughput for the longer wavelength side. It is clear from Fig. 12 that baking in vacuum is much more effective than in air. In Fig. 13, we show the change for throughput at two wavelengths, one long 396 nm and one short 206 nm, as a function of the baking time in air. Even after more than 4 days of baking, a plateau was not reached. performed. Since exhaust from vehicles is a common source of aromatic compounds and it was reported previously that PTFE integrating spheres are susceptible to contamination by hydrocarbon aromatic compounds [3], it is of interest to examine the behavior of integrating spheres when exposed to such environmental pollutants. This simple but effective experiment started with a fresh well-baked sintered PTFE integrating sphere with its UV fluorescence and absorption measured. The integrating sphere was then placed near the tail pipe of a car or a truck with its engine running. After several seconds of exposure, the integrating sphere was removed and taken into our laboratories for both throughput and LIF measurements again. We repeated the experiment for both a gasoline-powered car and a diesel-powered truck, and the results are shown in Figs. 14 and 15. Note that these measurements were not performed under well controlled exposures and the results provide a clear qualitative characterization on the behavior of integrating spheres when exposed to pollutants. For the LIF measurements in Fig. 14, exposure to gasoline and diesel exhaust increases the fluorescence, remarkably, in exactly the same spectral regions where the integrating spheres fluoresce before the exposure to exhaust. In particular, the increase in fluorescence spectra from diesel exhaust exposure between 250 and 400 nm closely resembles that observed for pristine integrating spheres (shown in C. Effects of Exposure to Environmental Pollutants The stability of an integrating sphere is critical when used in a calibration facility, such as FICUS. In calibrating deuterium lamps at FICUS [11], we observed that an integrating sphere can have dramatic changes in its throughput over a few days. The cause of such a change was traced to diesel truck fumes leaking into the room where the measurement was Fig. 13. Relative change in throughput for pressed integrating spheres as a function of baking time in air for 206 and 396 nm. Fig. 14. LIF measurements of a sintered PTFE integrating sphere exposed to exhaust from a gasoline engine and from a diesel engine. Thin solid lines are before the exposure, thick solid lines are after the exposure. For gasoline exposure, dotted line is 19 h of baking in vacuum after exposure. For diesel exposure, dotted lines are baking for both 19 and 68 h. 1 August 2007 Vol. 46, No. 22 APPLIED OPTICS 5125

8 Fig. 15. Relative throughput measurements of a sintered PTFE integrating sphere before exposure, after exposure to gasoline engine exhaust, and after exposure to diesel engine exhaust. Fig. 10). As one might expect, the diesel exhaust exposure caused a much stronger increase in fluorescence than the gasoline exhaust. Figure 14 also shows the results of baking the exposed integrating sphere to remove the contaminants. For gasoline exhaust, after 19 h of baking in vacuum at 90 C, the fluorescence spectrum was almost fully restored to its original spectra before exposure to the exhaust. For diesel exhaust, the fluorescence was only partially reduced even after more than 80 h of baking in vacuum. Other cleaning techniques we tried, such as the use of various solvents, did not restore the integrating sphere to its state before exposure. Currently, we are still experimenting with other options to rid the integrating spheres of contaminants. For the throughput measurements on the exposed integrating spheres shown in Fig. 15, the diesel exposure caused a substantial decrease in throughput over the whole wavelength region of our measurement, indicating the presence of a large amount of contaminants. As in the case of the LIF measurement, baking partially restored the throughput to that before the exposure. Lastly, exposure to gasoline exhaust resulted in the buildup of the two-peakabsorption structure in the shorter wavelengths we discussed previously. For the longer wavelengths, there is a slight increase in the throughput. This is in contrast to the observed decrease in throughput after baking an integrating sphere discussed previously. D. Ultraviolet Stability In radiometric applications aiming for the highest measurement accuracy, the stability of an optical element, such as an integrating sphere, is an essential property [5]. Given the vulnerability of PTFE integrating spheres to contamination as discussed above, it is not surprising that the stability of an integrating sphere is primarily influenced by the level of contamination. We have performed several preliminary tests to check the stability in the throughput of an integrating sphere under UV irradiation from a deuterium lamp. Our preliminary results show that, in general, the throughput of an integrating sphere with relatively small absorption and fluorescence features, such as a baked integrating sphere, is stable to within a few percent in the UV after a few days of continuous irradiation from a deuterium lamp. On the other hand, the throughput of an integrating sphere with large absorption and fluorescence features varies as much as 10% under the same UV irradiation even for a few hours. The exact change in throughput varies from one sphere to another. At this time, it is not clear if the instability is caused by UV radiation interacting with the contaminants or a continued buildup of contaminants from the surrounding environment. Further experiments are under way to shed light on the mechanism for such an instability. 4. Discussion and Conclusions As discussed previously, all PTFE integrating spheres studied for this work showed a consistent but varying degree of absorption and fluorescence features in the UV, and moderate baking can reduce these features. This indicates that the observed fluorescence and the associated absorption in the UV are not intrinsic to the PTFE material. Contamination of integrating spheres by somewhat volatile impurities either during the manufacturing process or exposure to contaminating sources afterward is mostly responsible for the observed effects. This is supported by a previous study where the existence of impurities, identified as organic hydrocarbons, was confirmed [3]. Furthermore, it was shown that PTFE can absorb aromatic hydrocarbons of toluene and naphthalene after exposure to these compounds [3]. Toluene and naphthalene belong to a class of the polycyclic aromatic hydrocarbons (PAHs), and it is worthwhile to compare our results with that of PAHs in general. Known for chemical stability from their conjugated -electron systems, PAHs typically exhibit strong absorption and fluorescence spectra in the UV [16] as a result of the transition of an electron between -bonding to *-antibonding orbitals [17,18]. The transition normally involves a ground vibrational level of a ground singlet electronic state to a vibrationally excited level of an excited singlet electronic state. The resulting fluorescence is typically redshifted from the absorption because of transitions to higher vibrational levels of the ground electronic state. Because the vibrational level spacings in ground and excited electronic states are usually similar, the fluorescence spectrum is often virtually the mirror image of the absorption spectrum [19]. In our observation of integrating spheres, the persistent absorption band between 200 and 300 nm and the fluorescence band between 250 and 400 nm fits very well with that of PAH impurities. For some sintered PTFE integrating spheres, the additional fluorescence band observed between 400 and 500 nm (see Fig. 8) contains a clear vibrational structure ( 1700 cm 1 spacing) and is caused by different PAH impurities. Further support 5126 APPLIED OPTICS Vol. 46, No August 2007

9 for PAH impurities is given by the close resemblance of the fluorescence spectra from 250 to 400 nm between pristine integrating spheres and integrating spheres exposed to engine exhaust, which is known to contain a large number of PAHs. While the exact chemical compounds of the PAH impurities cannot be easily identified by absorption and fluorescence spectra alone, one can still obtain information about the chemical nature of the impurities from the measured spectra using properties that are generally true for PAHs. In general, as the number of conjugated aromatic rings increases, the UV absorption maxima are shifted to longer wavelengths and exhibit a greater complexity of its absorption spectrum [20]. This explains the low response for example, at 254 nm for PAHs with six or more rings [21]. Similarly for fluorescence, the spectrum also redshifts as the size of the conjugated aromatic ring system increases. The redshifting of the absorption and fluorescence spectrum can be explained as a particle (electron) in a box ( -conjugated rings) model the larger the box, the more closely spaced the levels, and the longer the wavelength of absorption and emission. In the case of linear PAHs, such as benzene (one ring), naphthalene (two ring), anthracene (three ring), tetracene (four ring), and pentacene (five ring), a comparison between published fluorescence spectra [22] and our results shows that the measured fluorescence spectrum below 400 nm is caused by PAHs with one to three rings, while fluorescence between 400 and 500 nm is caused by three or four ring PAHs. The lack of fluorescence in the longer wavelength up to 800 nm indicates that only PAHs with a small number of conjugated aromatic rings were present in the integrating spheres we studied as impurities. The measurements on integrating spheres exposed to engine exhaust also shed light on the nature and the origin of the impurities. Emission from vehicle exhaust contains a variety of PAH compounds and their derivatives depending on such conditions as the fuel type and the degree of incomplete combustion of fuel [17]. The emitted PAHs can be in the form of gas phase, or the PAH can be adsorbed or encapsulated in particulates. In general, diesel exhaust contains more particulates than gasoline exhaust. These particulates diesel particulate matter (DPM) include diesel soot and aerosols, such as ash particulates, metallic abrasion particles, sulfates, and silicates. Our LIF measurements on integrating spheres with exhaust exposure show that both gasoline and diesel exhaust deposit PAHs with a low number of aromatic rings, giving the similar fluorescence spectra from the pristine spheres that was identified above as PAH with one to three aromatic rings. The fact that diesel exhaust exposure induced a fluorescence peak in the slightly longer wavelength than gasoline exhaust exposure suggests that contaminants with higher ring-number PAHs resulted from diesel exhaust exposure. In the throughput measurements, as in LIF measurements, exposure to gasoline exhaust caused similar absorption features for the pristine integrating spheres. However, diesel exposure caused a broad structureless absorption. This is most likely due to the deposition of DPM in the sphere that degraded the overall reflectance of PTFE. It is of interest to note that, in common environmental studies, PTFE membrane filters are preferred for the collection of particulates because of their high collection efficiency and inertness [17]. Finally, the fact that baking can restore an integrating sphere exposed to gasoline exhaust is in accordance with the volatility of some lighter PAH compounds, such as phenanthrene or anthracene, due to thermal desorption. On the other hand, it is possible that the difficulty in removing contaminants from a diesel exposed integrating sphere by baking is because either the PAHs involved have a much lower vapor pressure or the PAHs were trapped in the particulates, which are impossible to bake out at these low temperatures 90 C. Given the important role of PAHs as contaminants, there are two possible sources for PAH contamination: (1) the process of manufacturing the integrating sphere, such as the use of resin, and (2) the exposure of integrating spheres to an environment that contains PAHs. As shown above, engine exhaust in the environment will cause significant contamination that affects the performance of an integrating sphere in the UV. Even in normal air, suspended atmospheric particulate matter that in many cases contains a large number of different PAHs could also cause contamination. This clearly suggests that integrating spheres should be protected from contaminated air where possible to ensure longer stability in the UV. Once contaminated with PAH compounds, the PAHs can react in the air or undergo photoreaction with UV irradiation causing more instability. This was observed in our stability measurements where a contaminated integrating sphere tends to have much lower UV stability than a baked integrating sphere with less contamination. From the standpoint of practical UV applications of integrating spheres, the absorption, fluorescence, and stability are of great concern for accurate measurements. The absorption in the UV causes lower throughput and to some extent affects the uniformity of the exiting radiation from an integrating sphere. Fluorescence is more challenging because it alters the spectral correlation between the input and output radiation. One way to correct this effect for an integrating sphere and monochromator system is to treat the fluorescence as part of the monochromator line spread function. With the combined system completely characterized using a tunable laser, contributions from fluorescence and scattering can be mathematically alleviated from the output spectra by using a stray light reduction algorithm as was previously demonstrated [11]. While the absorption and fluorescence problems can be solved in principle, the issue of instability caused by contamination from both the laboratory and field environment is more difficult to resolve. Baking and limiting exposure to 1 August 2007 Vol. 46, No. 22 APPLIED OPTICS 5127

10 environmental contaminants can help with stability. More research on the UV behavior of integrating spheres is currently underway to identify the molecules responsible for the contamination, find ways to clean the contaminants, and protect integrating spheres from further contamination. In summary, we have used two techniques to characterize PTFE integrating spheres in the UV: measuring both the UV throughput spectrum using a deuterium lamp and the fluorescence spectrum induced by a tunable UV laser. All of the integrating spheres studied show significant absorption and fluorescence features. These features, more pronounced for sintered integrating spheres than pressed ones, are remarkably similar for all of the integrating spheres we studied and are consistent with PAH contaminants. Baking for long periods of time in vacuum can reduce the level of contaminants but does not completely eliminate them. In addition, fluorescence induced by the UV laser provided important clues for identifying the chemical nature of contaminants. PAH contaminants from the environment are the main suspect that deteriorate the performance of PTFE integrating spheres. The authors thank Yuqin Zong for lending us the fiber optic spectrograph, and Robert Saunders for packing integrating spheres. We are also grateful to Howard Yoon and Albert Parr for their discussions and support of this project. References 1. R. D. Saunders and W. R. Ott, Spectral irradiance measurements: effect of UV-produced fluorescence in integrating spheres, Appl. Opt. 15, (1976). 2. V. R. Weidner and J. J. Hsia, Reflection properties of pressed polytetrafluoroethylene powder, J. Opt. Soc. Am. 71, (1981). 3. A. E. Stiegman, C. J. Bruegge, and A. W. Springsteen, Ultraviolet stability and contamination analysis of Spectralon diffuse reflectance material, Opt. Eng. 32, (1993). 4. C. J. Bruegge, A. E. Stiegman, R. A. Rainen, and A. W. Springsteen, Use of Spectralon as a diffuse reflectance standard for in-flight calibration of earth-orbiting sensors, Opt. Eng. 32, (1993). 5. D. R. Gibbs, F. J. Duncan, R. P. Lambe, and T. M. Goodman, Ageing of materials under intense UV radiation, Metrologia 32, ( ). 6. P. R. Spyak and C. Lansard, Reflectance properties of pressed Algoflon F6: a replacement reflectance-standard material for Halon, Appl. Opt. 36, (1997). 7. W. Möller, K.-P. Nikolaus, and A. Höpe, Degradation of the diffuse reflectance of Spectralon under low-level irradiation, Metrologia 40, S212 S215 (2003). 8. H. J. Kostkowski, Reliable Spectroradiometry (Spectroradiometry Consulting, 1997), Appendix B. 9. P. Y. Barnes, M. E. Nadal, and E. A. Early, Reflectance standards at ultraviolet wavelength, Proc. SPIE 3818, 9 14 (1999). 10. P. S. Shaw, U. Arp, H. Y. Yoon, R. D. Saunders, A. C. Parr, and K. R. Lykke, A SURF beamline for synchrotron source-based absolute radiometry, Metrologia 40, S124 S127 (2003). 11. P. S. Shaw, U. Arp, R. D. Saunders, D. J. Shin, H. W. Yoon, C. E. Gibson, Z. Li, A. C. Parr, and K. R. Lykke, Synchrotron radiation based irradiance calibration from 200 nm to 400 nm at SURF III, Appl. Opt. 46, (2007). 12. S. W. Brown, G. P. Eppeldauer, and K. R. Lykke, Facility for spectral irradiance and radiance responsivity calibrations using uniform sources, Appl. Opt. 45, (2006). 13. S. W. Brown, B. C. Johnson, M. E. Feinholz, M. A. Yarbrough, S. J. Flora, K. R. Lykke, and D. K. Clark, Stray-light correction algorithm for spectrographs, Metrologia 40, S81 S84 (2003). 14. Y. Zong, S. W. Brown, B. C. Johnson, K. R. Lykke, and Y. Ohno, Simple spectral stray light correction method for array spectroradiometers, Appl. Opt. 45, (2006). 15. Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose. 16. See, for example, D. Cox, J. M. Key, M. S. Lai, J. R. Sutherland, M. R. Zahn, and J. M. Arrington, On-line supercritical fluid extraction synchronous luminescence spectroscopy: a screening method for polycyclic aromatic hydrocarbons, J. Under. Chem. Res. 2, (2005) and references therein. 17. See, for example, M. L. Lee, M. V. Novotny, and K. D. Bartle, Analytical Chemistry of Polycyclic Aromatic Compounds (Academic, 1981). 18. M. L. Lee, M. V. Novotny, and K. D. Bartle, Handbook of Polycyclic Aromatic Hydrocarbons, A. Bjørseth, ed. (Marcel Dekker, 1983). 19. N. J. Turro, Molecular Photochemistry (Benjamin Cummings, 1965). 20. E. Clar, Polycyclic Hydrocarbons (Academic, 1964). 21. R. A. Friedel and M. Orchin, Ultraviolet Spectra of Aromatic Compounds (Wiley, 1951). 22. J. B. Birks, Photophysics of Aromatic Molecules (Wiley, 1970) APPLIED OPTICS Vol. 46, No August 2007

UV Fluorescence from Integrating Spheres Measurement and Theory

UV Fluorescence from Integrating Spheres Measurement and Theory UV Fluorescence from Integrating Spheres Measurement and Theory Ping-Shine Shaw, Zhigang Li, Uwe Arp, and Keith R. Lykke National Institute of Standards and Technology, Gaithersburg, MD, USA Outline: 1.

More information

Characterisation & Use of Array Spectrometers

Characterisation & Use of Array Spectrometers Characterisation & Use of Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington Middlesex, UK 1 Overview Basic design and features of

More information

Stray Light Rejection in Array Spectrometers

Stray Light Rejection in Array Spectrometers Stray Light Rejection in Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington, Middlesex, UK 1 Overview Basic optical design of an array

More information

Resonance Raman measurements utilizing a deep UV source

Resonance Raman measurements utilizing a deep UV source Resonance Raman measurements utilizing a deep UV source Adam Willitsford a, C. Todd Chadwick b, Hans Hallen b, and C. Russell Philbrick a a The Pennsylvania State University, Department of Electrical Engineering,

More information

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Spectroscopy. Page 1 of 8 L.Pillay (2012) Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has

More information

Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong

Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong CORM 2007 Annual Conference, Gaithersburg, Maryland, May 8-11, 2007 Spatial Stray Light Correction for Imaging Instruments Using Matrix Method Yuqin Zong Outline 1) Introduction 2) Principle of spatial

More information

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 1. Examples of advantages and disadvantages with laser-based combustion diagnostic techniques: + Nonintrusive + High

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)

More information

Laser Dissociation of Protonated PAHs

Laser Dissociation of Protonated PAHs 100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

High Resolution Optical Spectroscopy

High Resolution Optical Spectroscopy PHYS 3719 High Resolution Optical Spectroscopy Introduction This experiment will allow you to learn a specific optical technique with applications over a wide variety of phenomena. You will use a commercial

More information

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately

More information

Chapter 15 Molecular Luminescence Spectrometry

Chapter 15 Molecular Luminescence Spectrometry Chapter 15 Molecular Luminescence Spectrometry Two types of Luminescence methods are: 1) Photoluminescence, Light is directed onto a sample, where it is absorbed and imparts excess energy into the material

More information

Chapter 4 Nadir looking UV measurement. Part-I: Theory and algorithm

Chapter 4 Nadir looking UV measurement. Part-I: Theory and algorithm Chapter 4 Nadir looking UV measurement. Part-I: Theory and algorithm -Aerosol and tropospheric ozone retrieval method using continuous UV spectra- Atmospheric composition measurements from satellites are

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

two slits and 5 slits

two slits and 5 slits Electronic Spectroscopy 2015January19 1 1. UV-vis spectrometer 1.1. Grating spectrometer 1.2. Single slit: 1.2.1. I diffracted intensity at relative to un-diffracted beam 1.2.2. I - intensity of light

More information

FLS980 Series Reference Guide

FLS980 Series Reference Guide FLS980 Series Reference Guide Integrating Sphere for Measurements of Fluorescence Quantum Yields and Spectral Reflectance Revision 1 Copyrights Copyright 2016 Edinburgh Instruments Ltd. All rights reserved.

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source 3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi

More information

Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths

Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths Progress Towards an Absolute Calibration of Lunar Irradiance at Reflected Solar Wavelengths Claire Cramer, Steve Brown, Keith Lykke, John Woodward (NIST) Tom Stone (USGS) Motivation for using the Moon

More information

Calibration of Ocean Colour Sensors

Calibration of Ocean Colour Sensors Dr. A. Neumann German Aerospace Centre DLR Remote Sensing Technology Institute Marine Remote Sensing What is Calibration, why do we need it? Sensor Components Definition of Terms Calibration Standards

More information

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes M. W. Teague*, Tonya Felix*, M. K. Hudson, and R. Shanks *Department of Chemistry, Hendrix College, Conway, AR

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

Characterization of high temperature solar thermal selective absorber coatings at operation temperature Available online at www.sciencedirect.com Energy Procedia 00 (2013) 000 000 www.elsevier.com/locate/procedia SolarPACES 2013 Characterization of high temperature solar thermal selective absorber coatings

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

MOLECULAR AND ATOMIC SPECTROSCOPY

MOLECULAR AND ATOMIC SPECTROSCOPY MOLECULAR AND ATOMIC SPECTROSCOPY 1. General Background on Molecular Spectroscopy 3 1.1. Introduction 3 1.2. Beer s Law 5 1.3. Instrumental Setup of a Spectrophotometer 12 1.3.1. Radiation Sources 13 1.3.2.

More information

HARP Assessment of Uncertainty

HARP Assessment of Uncertainty HARP Assessment of Uncertainty The HIAPER Airborne Radiation Package (HARP) was designed to produce accurate measurements of actinic flux and irradiance. The Atmospheric Radiation Group (ARG) at the University

More information

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. CHEM 322 Name Exam 3 Spring 2013 Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. Warm-up (3 points each). 1. In Raman Spectroscopy, molecules are promoted

More information

Chem Homework Set Answers

Chem Homework Set Answers Chem 310 th 4 Homework Set Answers 1. Cyclohexanone has a strong infrared absorption peak at a wavelength of 5.86 µm. (a) Convert the wavelength to wavenumber.!6!1 8* = 1/8 = (1/5.86 µm)(1 µm/10 m)(1 m/100

More information

Analytical Spectroscopy Review

Analytical Spectroscopy Review Analytical Spectroscopy Review λ = wavelength ν = frequency V = velocity = ν x λ = 2.998 x 10 8 m/sec = c (in a vacuum) ν is determined by source and does not change as wave propogates, but V can change

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

Advanced Spectroscopy Laboratory

Advanced Spectroscopy Laboratory Advanced Spectroscopy Laboratory - Raman Spectroscopy - Emission Spectroscopy - Absorption Spectroscopy - Raman Microscopy - Hyperspectral Imaging Spectroscopy FERGIELAB TM Raman Spectroscopy Absorption

More information

Singlet. Fluorescence Spectroscopy * LUMO

Singlet. Fluorescence Spectroscopy * LUMO Fluorescence Spectroscopy Light can be absorbed and re-emitted by matter luminescence (photo-luminescence). There are two types of luminescence, in this discussion: fluorescence and phosphorescence. A

More information

The wavelength reconstruction from toroidal spectrometer image data

The wavelength reconstruction from toroidal spectrometer image data Computational Methods and Experimental Measurements XIII 165 The wavelength reconstruction from toroidal spectrometer image data J. Blazej 1, M. Tamas 1, L. Pina 1, A. Jancarek 1, S. Palinek 1, P. Vrba

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 6 Spectroscopic Techniques Lecture - 2 UV-Visible Spectroscopy

More information

C101-E145 TALK LETTER. Vol. 17

C101-E145 TALK LETTER. Vol. 17 C101-E145 TALK LETTER Vol. 17 UV-VIS Spectroscopy and Fluorescence Spectroscopy (Part 1 of 2) ------- 02 Applications: Spectrofluorophotometers Used in a Variety of Fields ------- 06 Q&A: Is there a way

More information

C101-E111. Talk Letter. Vol.2 February 2009

C101-E111. Talk Letter. Vol.2 February 2009 C101-E111 UV Talk Letter Vol.2 February 2009 UV Talk Letter UV Talk Letter The Structure of a Spectrophotometer Vol.2 February 2009 1.The Measurement Principle Used by a Spectrophotometer The basic measurement

More information

Spectroscopic Investigation of Polycyclic Aromatic Hydrocarbons Trapped in Liquid Helium Clusters

Spectroscopic Investigation of Polycyclic Aromatic Hydrocarbons Trapped in Liquid Helium Clusters Spectroscopic Investigation of Polycyclic Aromatic Hydrocarbons Trapped in Liquid Helium Clusters Friedrich Huisken and Serge Krasnokutski Max-Planck-Institut für Strömungsforschung, Bunsenstr. 10, D-37073

More information

1901 Application of Spectrophotometry

1901 Application of Spectrophotometry 1901 Application of Spectrophotometry Chemical Analysis Problem: 1 Application of Spectroscopy Organic Compounds Organic compounds with single bonds absorb in the UV region because electrons from single

More information

Electronic Excitation by UV/Vis Spectroscopy :

Electronic Excitation by UV/Vis Spectroscopy : SPECTROSCOPY Light interacting with matter as an analytical tool III Pharm.D Department of Pharmaceutical Analysis SRM College Of Pharmacy,Katankulathur Electronic Excitation by UV/Vis Spectroscopy : X-ray:

More information

Experiment 24: Spectroscopy

Experiment 24: Spectroscopy Experiment 24: Spectroscopy Figure 24.1: Spectroscopy EQUIPMENT High Voltage Power Supply Incandescent Light Source (3) Gas Discharge Tubes: 1. Helium 2. Hydrogen 3. Unknown Element Spectrometer Felt (1)

More information

Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma

Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma Sebastijan BREZINSEK, Albrecht POSPIESZCZYK, Gennadij SERGIENKO, Philippe MERTENS and Ulrich SAMM Institut

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed

More information

A New Hyperspectral Spherical-Cavity Absorption Meter

A New Hyperspectral Spherical-Cavity Absorption Meter A New Hyperspectral Spherical-Cavity Absorption Meter David R. Dana and Robert A. Maffione HOBI Labs, Inc. Adapted from the poster presentation at Ocean Sciences 2006, Honolulu, Hawaii INTRODUCTION How

More information

Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs)

Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs) Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs) Shiv K. Sharma* Hawaii Institute of Geophysics & Planetology, University of Hawaii (UH) at Manoa, Honolulu, Hawaii,

More information

Regular Reflectance and Transmittance Measurements of Transmissive Materials Using a STAR GEM Optical Accessory

Regular Reflectance and Transmittance Measurements of Transmissive Materials Using a STAR GEM Optical Accessory Regular Reflectance and Transmittance Measurements of Transmissive Materials Using a STAR GEM Optical Accessory 1,3 E.Kawate, 1,2 M.Hain 1 AIST, 1-1-1, Central 2, Umezono, Tsukuba, Ibaraki 305-8568, Japan

More information

Diagnósticos em Plasmas

Diagnósticos em Plasmas Tecnologia a Plasma para o Processamento de Materiais Diagnósticos em Plasmas Diagnósticos Ópticos João Santos Sousa, nº50901 Semestre Inverno 2004/2005 21 de Janeiro de 2005, 9h-10h, sala F8 Contents

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

FTIR Spectrometer. Basic Theory of Infrared Spectrometer. FTIR Spectrometer. FTIR Accessories

FTIR Spectrometer. Basic Theory of Infrared Spectrometer. FTIR Spectrometer. FTIR Accessories FTIR Spectrometer Basic Theory of Infrared Spectrometer FTIR Spectrometer FTIR Accessories What is Infrared? Infrared radiation lies between the visible and microwave portions of the electromagnetic spectrum.

More information

Lecture- 08 Emission and absorption spectra

Lecture- 08 Emission and absorption spectra Atomic and Molecular Absorption Spectrometry for Pollution Monitoring Dr. J R Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Lecture- 08 Emission and absorption spectra

More information

LASER MICROPROBE MASS SPECTROMETRY MICROANALYSIS OF POLYCYCLIC AROMATIC HYDROCARBONS IN FLAMES, IN DIESEL FUELS AND IN DIESEL EMISSIONS

LASER MICROPROBE MASS SPECTROMETRY MICROANALYSIS OF POLYCYCLIC AROMATIC HYDROCARBONS IN FLAMES, IN DIESEL FUELS AND IN DIESEL EMISSIONS LASER MICROPROBE MASS SPECTROMETRY MICROANALYSIS OF POLYCYCLIC AROMATIC HYDROCARBONS IN FLAMES, IN DIESEL FUELS AND IN DIESEL EMISSIONS Robert A. Fletcher, 1 Richard A. Dobbins, 2 Bruce A. Benner, Jr.

More information

Advanced Analytical Chemistry

Advanced Analytical Chemistry 84.514 Advanced Analytical Chemistry Part III Molecular Spectroscopy (continued) Website http://faculty.uml.edu/david_ryan/84.514 http://www.cem.msu.edu/~reusch/virtualtext/ Spectrpy/UV-Vis/spectrum.htm

More information

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION G. Thuillier 1, D. Bolsee 2, T. Foujols 1 1 LATMOS-CNRS, France 2 Institut d Aéronomie

More information

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements EMRP-ENV03 Solar UV WP 3: Improvement of Reference Spectroradiometers Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements Peter Meindl, Christian

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

arxiv: v1 [physics.ins-det] 24 Jan 2010

arxiv: v1 [physics.ins-det] 24 Jan 2010 A Study of the Fluorescence Response of Tetraphenyl-butadiene R. Jerry arxiv:1001.4214v1 [physics.ins-det] 24 Jan 2010 Abstract Physics Dept., Howard University, Washington, DC 20059 L. Winslow, L. Bugel

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements in Diesel jets

Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements in Diesel jets Renewable energies Eco-friendly production Innovative transport Eco-efficient processes Sustainable resources Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements

More information

Molecular Luminescence Spectroscopy

Molecular Luminescence Spectroscopy Molecular Luminescence Spectroscopy In Molecular Luminescence Spectrometry ( MLS ), molecules of the analyte in solution are excited to give a species whose emission spectrum provides information for qualitative

More information

ATOMIC SPECROSCOPY (AS)

ATOMIC SPECROSCOPY (AS) ATOMIC ABSORPTION ANALYTICAL CHEMISTRY ATOMIC SPECROSCOPY (AS) Atomic Absorption Spectroscopy 1- Flame Atomic Absorption Spectreoscopy (FAAS) 2- Electrothermal ( Flame-less ) Atomic Absorption Spectroscopy

More information

Spectroscopy Problem Set February 22, 2018

Spectroscopy Problem Set February 22, 2018 Spectroscopy Problem Set February, 018 4 3 5 1 6 7 8 1. In the diagram above which of the following represent vibrational relaxations? 1. Which of the following represent an absorbance? 3. Which of following

More information

Skoog Chapter 6 Introduction to Spectrometric Methods

Skoog Chapter 6 Introduction to Spectrometric Methods Skoog Chapter 6 Introduction to Spectrometric Methods General Properties of Electromagnetic Radiation (EM) Wave Properties of EM Quantum Mechanical Properties of EM Quantitative Aspects of Spectrochemical

More information

IR Spectrography - Absorption. Raman Spectrography - Scattering. n 0 n M - Raman n 0 - Rayleigh

IR Spectrography - Absorption. Raman Spectrography - Scattering. n 0 n M - Raman n 0 - Rayleigh RAMAN SPECTROSCOPY Scattering Mid-IR and NIR require absorption of radiation from a ground level to an excited state, requires matching of radiation from source with difference in energy states. Raman

More information

AS 101: Day Lab #2 Summer Spectroscopy

AS 101: Day Lab #2 Summer Spectroscopy Spectroscopy Goals To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are related To see spectral lines from different elements in emission and

More information

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte 1 In this chapter, absorption by molecules, rather than atoms, is considered. Absorption in the ultraviolet and visible regions occurs due to electronic transitions from the ground state to excited state.

More information

Investigation of Coherent Emission from the NSLS VUV Ring

Investigation of Coherent Emission from the NSLS VUV Ring SPIE Accelerator Based Infrared Sources and Spectroscopic Applications Proc. 3775, 88 94 (1999) Investigation of Coherent Emission from the NSLS VUV Ring G.L. Carr, R.P.S.M. Lobo, J.D. LaVeigne, D.H. Reitze,

More information

Low-Frequency Raman Spectra of Carbon Nanotubes Measured with an Astigmatism-Free Schmidt-Czerny-Turner Spectrograph

Low-Frequency Raman Spectra of Carbon Nanotubes Measured with an Astigmatism-Free Schmidt-Czerny-Turner Spectrograph Low-Frequency Raman Spectra of Carbon Nanotubes Measured with an Astigmatism-Free Schmidt-Czerny-Turner Spectrograph Abstract Traditional Czerny-Turner (CT) spectrographs suffer from the optical aberration

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

Atomic Spectra. d sin θ = mλ (1)

Atomic Spectra. d sin θ = mλ (1) Atomic Spectra Objectives: To measure the wavelengths of visible light emitted by atomic hydrogen and verify that the measured wavelengths obey the empirical Rydberg formula. To observe emission spectra

More information

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Course Details September 22 October 10 September 22 November 7 November 17 December 1 Topic Period Introduction to Spectrometric

More information

Chem 310 rd. 3 Homework Set Answers

Chem 310 rd. 3 Homework Set Answers -1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption

More information

Atomization. In Flame Emission

Atomization. In Flame Emission FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic by its monatomic particles in gaseous state in the

More information

SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E.

SOME ASPECTS OF PYROMETRY OF WHITE OBJECTS V.E. SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E. Mosharov, V.N. Radchenko, I.V. Senyuev Central Aerohydrodynamic Institute (TsAGI) 140180 Zhukovsky, Moscow Region, Russia Introduction Model surface temperature

More information

Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll.

Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll. Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll. Introduction to Fluorescence: Fluorescence is one of the possible mechanisms for emission of light by

More information

LABORATORY OF ELEMENTARY BIOPHYSICS

LABORATORY OF ELEMENTARY BIOPHYSICS LABORATORY OF ELEMENTARY BIOPHYSICS Experimental exercises for III year of the First cycle studies Field: Applications of physics in biology and medicine Specialization: Molecular Biophysics Fluorescence

More information

DAY LABORATORY EXERCISE: SPECTROSCOPY

DAY LABORATORY EXERCISE: SPECTROSCOPY AS101 - Day Laboratory: Spectroscopy Page 1 DAY LABORATORY EXERCISE: SPECTROSCOPY Goals: To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are

More information

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry 9.1 9A. Atomization Most fundamental for both techniques. Typical types 1. flame - burner type 2. Electrothermal graphite furnace 3. Specialized

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

Laser induced fluorescence

Laser induced fluorescence Report experiment 1 by 4-7 February 22, id#471917 Summary Several measurements are made around laser induced fluorescence (LIF). First the system was calibrated. The resolution of the spectrograph was

More information

09/05/40 MOLECULAR ABSORPTION METHODS

09/05/40 MOLECULAR ABSORPTION METHODS MOLECULAR ABSORPTION METHODS Absorption spectroscopy refers to spectroscopic techniques that measure the absorption of radiation, as a function of wavelength ( absorption spectrum ), due to its interaction

More information

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms that have been excited to higher energy levels by absorption

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Vibrational Spectroscopies. C-874 University of Delaware

Vibrational Spectroscopies. C-874 University of Delaware Vibrational Spectroscopies C-874 University of Delaware Vibrational Spectroscopies..everything that living things do can be understood in terms of the jigglings and wigglings of atoms.. R. P. Feymann Vibrational

More information

Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds. Kyung-Min Lee

Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds. Kyung-Min Lee Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds Kyung-Min Lee Office of the Texas State Chemist, Texas AgriLife Research January 24, 2012 OTSC Seminar OFFICE OF THE TEXAS

More information

Raman Albedo and Deep-UV Resonance Raman Signatures of Explosives

Raman Albedo and Deep-UV Resonance Raman Signatures of Explosives Raman Albedo and Deep-UV Resonance Raman Signatures of Explosives Balakishore Yellampalle and Brian Lemoff WVHTC Foundation, 1000 Technology Drive, Suite 1000, Fairmont, WV, USA 26554 ABSTRACT Deep-ultraviolet

More information

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer.

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer. Experiment 9 Emission Spectra 9.1 Objectives By the end of this experiment, you will be able to: measure the emission spectrum of a source of light using the digital spectrometer. find the wavelength of

More information

Chapter 6 Photoluminescence Spectroscopy

Chapter 6 Photoluminescence Spectroscopy Chapter 6 Photoluminescence Spectroscopy Course Code: SSCP 4473 Course Name: Spectroscopy & Materials Analysis Sib Krishna Ghoshal (PhD) Advanced Optical Materials Research Group Physics Department, Faculty

More information

UV Degradation of Polycarbonate

UV Degradation of Polycarbonate Utah State University DigitalCommons@USU Senior Theses and Projects Materials Physics 5-2017 UV Degradation of Polycarbonate Katie Gamaunt Utah State University Follow this and additional works at: https://digitalcommons.usu.edu/mp_seniorthesesprojects

More information

Ch 313 FINAL EXAM OUTLINE Spring 2010

Ch 313 FINAL EXAM OUTLINE Spring 2010 Ch 313 FINAL EXAM OUTLINE Spring 2010 NOTE: Use this outline at your own risk sometimes a topic is omitted that you are still responsible for. It is meant to be a study aid and is not meant to be a replacement

More information

Basics of UV-Visible Spectroscopy *

Basics of UV-Visible Spectroscopy * OpenStax-CNX module: m34525 1 Basics of UV-Visible Spectroscopy * Brittany L. Oliva-Chatelain Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution

More information

2B Technologies, Inc. An InDevR Company

2B Technologies, Inc. An InDevR Company 2B Technologies, Inc. An InDevR Company Technical Note No. 40 UV-Absorbing Interferences in Ozone Monitors Date: 22 April 2015 Author: John Birks Background Ozone measurements by absorbance of the 253.7-nm

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

A very brief history of the study of light

A very brief history of the study of light 1. Sir Isaac Newton 1672: A very brief history of the study of light Showed that the component colors of the visible portion of white light can be separated through a prism, which acts to bend the light

More information

1 WHAT IS SPECTROSCOPY?

1 WHAT IS SPECTROSCOPY? 1 WHAT IS SPECTROSCOPY? 1.1 The Nature Of Electromagnetic Radiation Anyone who has been sunburnt will know that light packs a punch: in scientific terms, it contains considerable amounts of energy. All

More information