Science. Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems
|
|
- Rachel Anthony
- 5 years ago
- Views:
Transcription
1 Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems Joseph Boneno, MS, SC(ASCP), 1 Michelle Fokakis, MT(ASCP), 2 Dave Armbruster, PhD, DABCC, FACB, C(ASCP) 3 ( 1 Our Lady of the Lake Medical Center, Baton Rouge, LA, 2 Abbott Laboratories, Global Service Organization, Dallas, TX, 3 Abbott Laboratories, Scientific Affairs, Chicago, IL) DOI: /MMLEYHNBY4WA12J6 Received Revisions Received Accepted This study examined the validation of performance of open clinical chemistry analyzers. Specifically, user defined applications on open clinical chemistry analyzers and analytical errors due to reagent carryover/cross contamination are examined. Control of laboratory errors associated with the analytical phase of testing was also examined. The modern practice of clinical chemistry relies ever more heavily on automation. Automated analyzers employ robotic pipetting and cuvette washing systems, designed to continuously sample patient specimens, dispense reagents, and clean specimen and reagent probes and cuvettes. Clinical chemistry systems can hold a large number of different assay reagents and perform a wide variety of tests. Analyzers typically use 1 or 2 probes to dispense reagents, and these probes are exposed in rapid succession to a wide spectrum of different reagents. Likewise, cuvettes are exposed to the same variety of different reagents as they are used over and over. A very real concern with analyzers is reagent carryover, that is, reagent from an initial assay clinging to a reagent probe and contaminating the reaction mixture of the next test immediately following the initial assay. Inefficient washing of cuvettes can also result in residual reagent from one test remaining in a cuvette and contaminating the reaction mixture for the next assay whose reagents are pipetted into the same cuvette. In many cases, reagent carryover has no effect on patient results, or it is clinically inconsequential. However, reagent carryover can result in a significant positive or negative bias, shifting patient test results falsely high or low to the extent that there are clinical consequences and potential adverse patient outcomes. Even if reagent carryover is not clinically significant, any amount of false depression or elevation of results due to this phenomenon represents analytical error. No laboratory analysis is error free and, in fact, a certain amount of error due to imprecision and bias is always expected and is acceptable, as long as the total error does not exceed a medically significant limit. Clinical laboratory professionals should understand the sources of analytical error inherent in their testing systems and assure themselves that the observed error is within acceptable limits. Many IVD manufacturers of clinical chemistry analyzers also supply the reagents to be used with the analyzers, and the combination of analyzer and reagents comprises an analytical system. Manufacturers perform verification and validation testing to ensure that systems meet performance claims. Such testing includes examining reagent carryover. Situations in which reagent pairs demonstrate an unacceptable degree of interaction are identified and the carryover is eliminated, or decreased to an acceptable level, by optimizing reagent probe and cuvette washing. Many modern clinical chemistry analyzers are open systems and they can perform user defined assays, that is, tests using reagents other than those supplied by the manufacturer of the analyzer. The versatility of an open system is an advantage because a laboratory can purchase reagents from many different suppliers and adapt them to an analyzer to meet its unique testing requirements. Common reasons for developing user defined applications are to add tests to the menu that are not available from the analyzer manufacturer or to substitute a third party assay that offers better performance than one that is available from the analyzer manufacturer. It is the prerogative of a laboratory to use whatever reagents it deems appropriate and most suitable. However, analyzer manufacturers are not obligated to support assay applications other than those for the reagents that they provide. It is the responsibility of the laboratory to consider the potential for reagent carryover and to ensure that the third party assays that it adapts to the analyzer do not cause analytical errors, resulting in falsely low or high test results. Here we describe the use of a reagent carryover protocol used to evaluate reagent crosscontamination on the Abbott c8000 analyzer between 32 Abbott clinical chemistry reagents and 15 specific protein and other assays from EQual Diagnostics. Materials and Methods Analyzer: Testing was performed on an Abbott Architect c8000 clinical chemistry analyzer. The c8000 is a multipurpose, open system instrument and it may be programmed to use third party reagents, in addition to Abbott reagents. The reagent carryover evaluation provided in the Architect c8000 System Assay Applications Guide was used. Downloaded from labmedicine.com November 2005 Volume 36 Number 11 LABMEDICINE 705
2 Assays: Thirty-two Abbott routine clinical chemistry assays (eg, electrolytes, enzymes, lipids, metabolites, etc) comprised the original c8000 test menu. To expand the menu, 15 assays (specific protein assays, ammonia, cholinesterase, HDL, and lithium) from EQual Diagnostics, were adapted to the c8000. Abbott c8000 Assays Acid Phosphatase Albumin (Bromcresol Purple) Alkaline Phosphatase ALT Amylase AST Calcium Cholesterol CK CO 2 Cl Creatinine Direct Bilirubin GGT Glucose Hgb A1c EQual Diagnostics Assays Ammonia C3 C4 Cholinesterase CRP Ferritin Haptoglobin Direct HDL Iron K Lactic Acid LD Direct LDL Lipase Na Mg Phosphorus TIBC Total Bilirubin Total Protein Triglycerides Urea Uric Acid Urine Protein IgA IgG IgM Lithium Microalbumin Prealbumin Transferrin Reagent Carryover Protocol: The protocol from the c8000 applications guide allows users to identify reagent carryover and to eliminate or control crosscontamination by programming reagent probe washes. All assays are tested in pairs, one assay designated as the donor and the other as the recipient. A saline assay, a simulated assay used to clean the reagent probes, is also used. The saline assay is configured like a typical test, except that 0.85% or 0.90% saline is used in place of a reagent. To determine if carryover occurs between the third party reagents (donor assays) being adapted to the system and the manufacturer s reagents (recipient assays), the following test order scheme is used: Position Sample Test Replicates Purpose 1 Saline Saline 5 Cleans reagent probes 2 Control Recipient 5 Target value for uncontaminated assay recipient assay 3 Saline Saline 5 Cleans reagent probes 4 Control Donor 5 Potentially contaminates recipient assay assay via carryover 5 Control Recipient 5 Observed value for potentially assay contaminated recipient assay The control material used for donor and recipient assays must be appropriate for each test (eg, microalbumin and urine glucose require urine controls, serum glucose requires a serum control, etc). Controls should contain analytes at normal reference range concentrations. The protocol may be repeated as appropriate for a given assay with different levels of controls to approximate the medical decision levels if appropriate. To determine if the manufacturer s reagents interfere with the third party reagents, the same test order scheme as above is used, but the manufacturers reagents are now the donor assays and the third party reagents are now the recipient assays. In addition, the third party reagents must also be tested in donor/recipient pairs to check for carryover between them. The carryover study results are interpreted as follows: Position Sample Test Calculation 1 Saline Saline None 2 Control Recipient assay Mean of 5 Uncontaminated Replicates = Target Value 3 Saline Saline None 4 Control Donor assay None 5 Control Recipient assay Calculate % difference from target for each potentially contaminated replicate; Replicate 1 Target x 100 = % Difference Target Reagent carryover causing crosscontamination is potentially demonstrated if the first replicate of the recipient assay replicates areas higher or lower than the recipient assay target value, calculated using uncontaminated recipient assay values. The value observed for the first replicate represents the worst case scenario, that is, the first replicate is most likely to demonstrate the maximum amount of carryover. Even if reagent carryover is not a problem, it is unlikely that recipient assay values will be exactly equal to the target values. In fact, they are expected to demonstrate some small difference consistent with the inherent variability of the assay system. Thus, a laboratory must specify an acceptance limit for each assay based on clinical criteria. If the percent difference for the recipient assay is within less than or equal to the predetermined acceptance limit, reagent carryover is judged to be clinically insignificant. However, if the difference exceeds the limit, the reagent carryover is considered to be potentially clinically significant. If reagent carryover is confirmed and is considered to be unacceptable, the c8000 is configured for a SmartWash to eliminate or reduce the cross contamination of the recipient assay by the donor assay. The reagent probe SmartWash configuration allows 4 wash solution options: water, detergent A, 0.5% acid wash, and detergent B. The wash volume is variable from 20 to 345 µl, and 1 to 5 probe washes may be selected. The reagent carryover protocol is then repeated with various SmartWash options configured for the donor/recipient assay pairs that are affected by cross contamination. The SmartWash configuration that eliminates or reduces carryover below the acceptance limit of the affected recipient assay is chosen for routine use. Results There were 32 standard c8000 assays and 15 new EQual original equipment manufacturer (OEM) assays. Each of the original 32 c8000 assays was tested with each of the 15 EQual assays, each being tested both as the donor and as the recipient assay. Thus, for each c8000 assay there were 64 assay pair combinations (each assay tested both as a donor and as a recipient with each EQual assay), for a total of 960 donor/recipient combinations (64 donor/recipient pairs per each assay x 15 EQual assays). In addition, each of the EQual OEM assays was checked for reagent carryover against each other on the c8000 for a total of 208 donor/recipient pair combinations. Table 1 lists some examples of donor/recipient pairs that did not exhibit reagent carryover exceeding the acceptance limit. For these assay pairs, representing standard c8000 reagents and EQual OEM reagents, no change in assay parameters were 706 LABMEDICINE Volume 36 Number 11 November 2005 labmedicine.com Downloaded from
3 Table 1_Examples Of Some Donor/Recipient Assay Pairs That Did Not Demonstrate Reagent Carryover Exceeding Cross Contamination Limits Donor Assay Recipient Assay Recipient Assay Target Value Recipient Assay Observed Value % Difference (Target Observed) Glucose Ferritin Ferritin Glucose Haptoglobin ALT ALT Haptoglobin Cholinesterase Urea Urea Cholinesterase C4 Creatinine Creatinine C Lithium Cholesterol Cholesterol Lithium Ferritin Haptoglobin Haptoglobin Ferritin Recipient assay target values are the means of 5 replicates. Recipient assay observed values are the means of 5 replicates of samples exposed to potential reagent carryover. necessary. A difference of +/-10% between the recipient assay target value and the observed recipient assay value following the donor assay was chosen as the acceptance limit. This was a practical, first pass decision and it was recognized that, for some analytes, differences less than +/-10% could be clinically significant. In most cases, test values that were increased or decreased by 10% due to reagent carryover were considered to make no difference in patient treatment. All differences were examined and in cases in which the 10% limit was considered to be inappropriate, additional testing with the donor/recipient pair was conducted and SmartWashes were programmed if appropriate. Data was reviewed to determine if a systematic reagent carryover problem truly existed. On a random basis, for 2 donor/recipient assay pairs, 1 replicate of the 5 recipient assay values exhibited a difference exceeding the 10% limit. Repeat testing of these pairs demonstrated that the unacceptably high recipient values were aberrant results and not reproducible. The CRP assay exhibited reagent carryover from 5 donor assays, GGT, LD, total protein, urea, and lactic acid, with target to observed % differences greater than +10%. However, the positive biases demonstrated were not deemed to be clinically significant for this analyte as CRP is a nonspecific marker of inflammation, and therefore it was not necessary to institute additional reagent probe washing. Table 2 lists the donor/recipient pairs that exhibited reagent carryover that exceeded the acceptance limit. If the result for an assay pair exceeds the acceptance limit for the recipient assay and there is a concern that the result may be aberrant, testing can be repeated to confirm that the initial observed degree of reagent carryover is reproducible. For donor/recipient pairs demonstrating unacceptable carryover, SmartWash testing was conducted to decrease the carryover to an acceptable level. Table 3 lists the donor/recipient pairs for which SmartWash programming was required. It lists the original % difference and the % difference after SmartWash and the wash fluid used. Typical normal reference ranges are also listed. In 2 cases, institution of a SmartWash to decrease the effect of reagent carryover was deemed unnecessary. Although carryover of CRP can decrease IgM values by about 15%, the normal reference range of IgM is so wide and the prevalence of CRP testing in this laboratory so infrequent that additional probe washing was not warranted. Cholinesterase carryover can increase total bilirubin by about 15%, which could have a clinical impact. The laboratory performs only about 2 cholinesterase tests per month and cholinesterase assays are segregated and performed separately to avoid the potential for a carryover effect on total bilirubin results. Discussion The practice of clinical chemistry has come a long way since its infancy when specimens were analyzed for individual analytes using manual procedures. The development of the continuous flow autoanalyzer was a significant leap forward as large numbers of specimens could be tested for individual analytes in an automated fashion. By combining autoanalyzer units, each representing a separate analyte channel, many different tests could be performed simultaneously in panels. However, these continuous flow systems were complex and demanded considerable Table 2_Donor/Recipient Assay Pairs That Exhibited Unacceptable Reagent Carryover Donor Assay Recipient Assay Recipient Assay Target Value Recipient Assay Observed Value % Difference (Target Observed) Ferritin Alkaline Phosphatase 354 IU/mL 316 IU/mL Ferritin CK 1007 IU/mL 901 IU/mL Ferritin Fe 234 µg/dl 209 µg/dl CRP IgM 1,300 mg/dl 1,100 mg/dl Cholinesterase Total bilirubin 0.52 mg/dl 0.60 mg/dl Lactic acid Ferritin 72.2 ng/ml 55.7 ng/ml Urea Ferritin 70.4 ng/ml 62.8 ng/ml HDL CO meq/l 15.4 meq/l -7.2 IgA CO meq/l 12.8 meq/l Downloaded labmedicine.com from November 2005 Volume 36 Number 11 LABMEDICINE 707
4 Table 3_Donor/Recipient Assay Pairs That Exhibited Significant Carryover Donor Assay Recipient Assay Initial % Carryover % Carryover Between Donor Description of Typical Normal Between Donor and and Recipient Assays after SmartWash Program Reference Range Recipient Assays Programming SmartWash Ferritin Alkaline Phosphatase Water wash, 345 µl IU/mL Ferritin CK Water wash, 345 µl IU/mL Ferritin Fe Water wash, 345 µl µg/dl CRP IgM N/A * mg/l Cholinesterase Total Bilirubin N/A * mg/dl Lactic Acid Ferritin Water wash, 345 µl ng/ml Urea Ferritin % acid wash, 345 µl ng/ml HDL CO Detergent A wash, 345 µl meq/l IgA CO Water wash, 345 µl meq/l * The observed decrease or increase in recipient assay values due to reagent carryover was not deemed to be clinically significant based on the normal reference range for the recipient assay and/or medical decision levels. The table lists the percent difference before and after application of sample probe washes to minimize the observed carryover. laboratory technologist time and effort to maintain them and keep them synchronized. The next generations of analyzers moved away from the continuous flow model to discrete analysis. Discrete analysis, sometimes colorfully known as sip and spit, allows individual specimens to be tested for a single analyte, a few analytes, or a whole panel of analytes. Today, discrete analyzers represent the bulk of systems in use by clinical laboratories. They do not use analyte specific channels but integrate the operations of various subsystems (sample and reagent pipettes, cuvettes or reaction vessels, spectrophotometers, wash assemblies, etc) to process many different specimens (and types of specimens) and reagents very rapidly, one after another, reusing all of the subsystems time and again. A disadvantage of the old continuous flow methodology was that it required duplication of the basic subsystems, or channels. For example, multiple reagent probes, cuvettes (actually, flow cells), spectrophotometric or other signal generating systems were required. Discrete analysis, by the continuous reuse of the subsystems, eliminates the need for duplication and allows for more compact analyzers. Ironically, the duplicative nature of continuous flow systems allowed laboratory personnel to focus on individual channels when performing maintenance, repair, and troubleshooting instead of balancing the operation of the system in toto. A problem with a single channel would impact the results for a single analyte but would not affect all of the others. The advantages of discrete analysis come at the price of increased complexity and require many subsystems to work in harmony. A problem with one subsystem (eg, a partially clogged sample pipettor, a scratched or clouded cuvette, or a malfunctioning photodiode for a given wavelength) has the potential to impact many test results for all assays and may require the entire system to be shutdown for repair. Discrete analyzers tend to be classified as closed or open. Closed systems are restricted to the use of reagents from the manufacturer of the analyzer, as the reagents and the analyzer constitute an integrated system by design. Only tests for which the manufacturer offers reagents may be performed with a closed system. Open systems are more flexible and reagents from a variety of manufacturers may be adapted to them. Manufacturers that provide both analyzers and reagents perform reagent carryover studies to ensure that test results are not affected. Reagent carryover studies are repeated by manufacturers any time a new assay is added to the test menu. Laboratories consider open systems to offer a distinct advantage because of the versatility that they offer. Indeed, being able to add any assay to an analyzer for which suitable reagents are available is an attractive feature. What may go unrecognized is the potential for reagent carryover to cause an unacceptable increase or decrease in test results for some analytes any time a new assay is added. Manufacturers cannot anticipate all possible assay reagent pair combinations or offer any guarantee that unacceptable reagent carryover will not occur. It is the responsibility of every laboratory that modifies an analytical system by adding third party reagents to them to ensure that unacceptable reagent carryover is not a problem. A literature search found relatively few papers discussing specimen or reagent carryover prior to Krouwer, Stewart, and Schlain proposed a multi-factor testing scheme to evaluate random access analyzers for imprecision, agreement with a reference method (slope of regression line), nonlinearity, linear drift, and reagent carryover. 1 These authors recognized that random access instruments differed from batch analyzers in that they are susceptible to reagent cross contamination in addition to sample carryover. They found that an AST donor assay could carryover into a recipient LD assay and falsely elevate the patient LD value because the AST reagent contained LD in high concentration on the Ciba Corning 550 Express analyzer. The analyzer was modified to decrease the reagent carryover to an acceptable level. The authors concluded that a 12 sample multi-factor protocol was a useful screening method but that more involved protocols would allow a better estimate of system performance. They specifically noted that a complete estimate of reagent carryover would require testing all possible reagent pairs available on an analyzer. Haeckel defined carryover as a process by which materials are carried into a reaction mixture to which they do not belong. These materials can be either parts of a specimen, or reagents including the diluent or wash solution. 2 Carryover can be classified by material (eg, specimen, diluent, reagent, reaction mixture, wash solution) or by site of carryover (eg, specimen cup, sample probe, reagent probe, wash station). The following types of carryover have been observed in practice: (1) specimen carryover in a sample probe, (2) carryover from diluent to specimen in a specimen cup, (3) carryover from reaction mixture to reaction mixture, and (4) carryover over from reagent to reaction mixture from a reagent probe. Haeckel proposed various formulae for quantitating carryover and describes various specific scenarios to that should be considered. He concludes by noting that new analytical systems should be designed to avoid any type of carry-over as far as possible. Downloaded 708 from LABMEDICINE Volume Number November 2005 labmedicine.com
5 Dixon classified carryover in selective access analyzers as physical (specimen) carryover or chemical (reagent) carryover and proposed formulae for quantitating both types. 3 Brenna and Prencipe described studies to characterize specimen carryover, reagent cross contamination, and cuvette-related carryover on the Bayer Axon for 12 analytes. 4 Specimen carryover was found to range between 1.50% to +0.66%; reagent cross contamination was found to be statistically significant only for CK and cholesterol (negative carryover of 2.5% and 1.9%, respectively); no cuvette-related carryover was noted. Lovric and colleagues examined specimen carryover and reagent cross contamination on the Olympus AU 800 analyzer using 2 approaches: (1) a screening method, and (2) a multifactorial method, but only for ALT and LD. 5 Carryover was not found to be statistically or clinically significant. Ferguson, Mee, and Wong reported the carryover of PSA on the Abbott IMx and AxSYM analyzers was identical at 0.02 ng/ml, or 0.003%. 6 Schlain and colleagues evaluated interassay carryover (reagent crosscontamination) biases on the Abbott AxSYM. 7 They used a 2-stage approach to examine a total of 231 possible assay pair configurations. In the first stage, those assay pairs that did not demonstrate carryover biases were eliminated from further consideration and in the second stage, those assay pairs could potentially result in large carryover biases were studied further. Their statistical model involved constructing 95% confidence intervals for test results from assay pair interactions. If the observed carryover bias was contained within the defined confidence interval, it was considered not to be either statistically or clinically significant. Increasing pipettor probe wash volumes was found to be an effective means of controlling carryover. Sadler, Murray, and Turner searched the literature for carryover papers in 1996 for the period of 1991 through June of 1995 and found only 12 in which the word carryover appeared either in the title or the abstract. 8 Four of the papers concerned testing for drugs of abuse and reported significant carryover problems, a issue that is not unusual in that type of analysis. The other 8 papers report no significant carryover, but few details were given. For example, specimens were described simply as high or low with no quantitative values provided. In other cases, the difference in concentration between high and low specimens was not considered to be great enough to provide a realistic carryover challenge. These authors reported on their own work involving thyroid stimulating hormone (TSH) samples with concentrations of about 30, 12, 3, 0.4, and 0.1 mu/l that were pipetted in different patterns using a fixed, glass capillary pipette for a 2nd generation TSH immunoradiometric assay (IRMA). They estimated sample carryover to be 0.086% for the 0.1 mu/l sample. Several papers discussing sample or reagent carryover appeared in the literature from 2000 to present. Skinner and Waterson reported increased imprecision and a negative bias for urine oxalate on the Roche MIRA S analyzer due to reagent carryover, as demonstrated by external quality assessment scheme performance. 9 The sample probe is used both to pipette specimen and reagent on this analyzer. Improving probe washing was not an option and the problem was resolved by delaying addition of the assay color reagent. Muser and colleagues evaluated the Cobas Integra 400 and examined both sample and reagent carryover. 10 Sample carryover was tested using only albumin, with 3 high concentration samples (50 g/l) followed by 5 low concentration samples (20 g/l), with the low sample values compared to a low reference value. Sample carryover did not exceed the 5% limit specified. Reagent carryover was assessed using 11 potentially interfering assays and 10 potentially sensitive assays in providing and receiving pairs. Carryover was determined using patient samples near the medical decision levels. Replicates (5) of the potentially contaminated assays were compared to the mean control value (n = 10) of the assays and a limit of +/-10% was applied. Reagent carryover occurred for 2 assay pairs: digoxin/antithrombin III and total protein/urine uric acid. An additional automated wash cycle for these assay pairs resolved the carryover problem. Redondo and colleagues performed a similar evaluation of the Cobas Integra Sample carryover was assessed with the same protocol used by Muser and colleagues for the Integra For CK and albumin, carryover was not observed (<5% difference). For reagent carryover, the median control value for recipient assays was determined and donor/recipient pairs were tested in replicate (n = 10). The differences between the median values for the recipient assay control value and the potentially impacted recipient assay were calculated for the triglyceride/lipase assay pair. Reagent carryover was not observed (< 5% difference). Wan and colleagues assessed sample carryover for the Roche E170 immunoanalyzer. 12 High concentration samples for several analytes (ferritin, 67.1 µg/l; AFP, 3,051,200 µg/l; TSH miu/l; free T4, 67.1 nmol/l; and PTH, 2,949 ng/l) were followed by low concentration samples (ferritin, 7.7 µg/l; AFP, 1.6 µg/l; TSH miu/l; free T4, 7.7 nmol/l; and PTH, 6.1 ng/l). The percent carryover was calculated as [(L1 L2)/(H1- L2)] x 100. Carryover was found to be minimal (< 0.2%). In an evaluation of the Architect c8000, Pauli, Seyfarth, and Dibbelt studied specimen carryover by testing 3 replicates of high concentration samples followed by 9 replicates of low concentration samples. 13 No significant carryover was demonstrated for 3 analytes (glucose at 54.4 mmol/l, ALT at 790 U/L, and CK at 40,966 U/L). There are many factors that can affect the accuracy of test results in the clinical laboratory, some more significant than others. The total error associated with a result is usually described as consisting of random error (imprecision) and systematic error (bias). Reagent carryover is an example of an analytical systematic error that may tend to be overlooked. Manufacturers are expected to comprehend reagent carryover and to ensure that it does not exceed reasonable acceptance limits for closed systems and for open systems using the manufacturer s own reagents that have been optimized for a the system. Reagent carryover is an ongoing challenge for manufacturers as carryover studies must be conducted any time that a new reagent is added to a system s test menu, or an existing assay undergoes a significant reformulation. Tempting as it is to argue that some changes to reagent formulations are minor and unlikely to impact results through reagent carryover, this can only be proven by collecting and reviewing data. Likewise, changes to a system s hardware or software, made to improve its performance, could inadvertently affect and change the reagent carryover profile. Again, only by performing reagent carryover studies can a manufacturer know to what extent this phenomenon impacts test results. Understanding reagent carryover is every manufacturer s obligation for its own analyzers and its own reagents. However, for very practical reasons, manufacturers cannot be expected to guarantee the performance of open systems if third party reagents are introduced on them. The onus falls upon each individual laboratory to investigate reagent carryover once it adapts a new third party reagent to a system. The more assays added, the greater the challenge as Downloaded labmedicine.com from November 2005 Volume 36 Number 11 LABMEDICINE 709
6 each new assay must be checked both as the donor and recipient assay with each existing assay and with each new assay as well. With luck, relatively few assay pairs will demonstrate clinically significant reagent carryover and require modification of system parameters to decrease the carryover effect to an acceptable level. That was the case in this study. However, performing carryover studies is a time consuming process. The data must be carefully examined and questionable, possibly aberrant results must be confirmed by retesting. Confirmed, unacceptable reagent carryover situations must then be addressed by appropriate modification of the system s operation, for example, use of additional reagent probe washes with an appropriate wash solution, as reported here. This requires even more testing and data scrutiny. Laboratories must decide on reasonable acceptance limits to judge the significance of changes in result values caused by reagent carryover. In some cases, a change of up to +/-10% may be analytically insignificant, but clinically significant, depending on the specific analyte, the normal reference range, the medical decision level, and other considerations. Conversely, some change may seem to be analytically significant (eg, < +/-10%), but ultimately may be clinically insignificant. When making a judgment call about the importance to recipient assays of the demonstrated differences between target and observed values, laboratories should remember that the differences are not totally due to the reagent carryover phenomenon but are dampened or exacerbated by the inherent random variability (imprecision) of the test. The use of 5 replicates of donor and recipient assays is a practical attempt to capture the mean influence of reagent carryover. However, for any given analytical event, the actual impact of reagent carryover may be either greater or less than that predicted by a reagent carryover study. Laboratories cannot realistically expect to eliminate reagent carryover (ie, percent difference between target and observed values equals zero). They can only try to objectively measure reagent carryover and control it. The judicious application of laboratory experience and clinical insight by the laboratory director is called for when making decisions about the amount of reagent carryover that can be tolerated. Manufacturers are faced with a balancing act when designing automated systems. While it may be possible to engineer the probe sampling and washing systems to virtually eliminate the possibility of carryover, such efforts invariably affect other aspects of sample handling, for example, the unwelcome consequence of decreasing specimen throughput. Juggling the practical needs for efficient, high volume sample throughput with minimization of specimen and reagent carryover, means that the potential for carryover is unavoidable. Manufacturers must test the performance envelope of their systems and understand their capabilities and limitations. Users of these systems must take into account the performance information provided by manufacturers when judging the acceptability of patient results and conducting investigations of suspect results. Users of open systems who adapt reagents from third party vendors to them should consider the effect of reagent carryover on results and perform testing to measure it. The c8000 is an example of a general purpose open system clinical chemistry analyzer that allows the user to optimize probe washing to eliminate or control reagent carryover, reducing laboratory analytical phase error and improving patient safety. LM Acknowledgements: Michelle Fokakis of Abbott Laboratories provided technical assistance to Our Lady of the Lake Medical Center to perform reagent carryover studies for EQual specific protein assays (third party reagents) that were adapted to Abbott c8000 clinical chemistry analyzer. 1. Krouwer JS, Stewart WN, Schlain B. A multi-factor experimental design for evaluating random-access analyzers. Clin Chem. 1988;34: Haeckel R. Proposals for the description and measurement of carry-over effects in clinical chemistry. Pure Appl Chem. 1991;63: Dixon K. A theoretical study of carryover in selective access analyzers. Ann Clin Biochem. 1990;27: Brenna S, Prencipe L. Axon clinical chemistry analyzer evaluated according to ECCLS protocol. Clin Chem. 1992;38: Lovric M, Cvoriscec D, Petrovecki M, et al. Evaluation of the Olympus AU 800 clinical chemistry analyzer. Clin Lab. 1995;41: Ferguson RA, Mee VA, Wong PY. Comparative evaluation of serum prostate specific antigen analysis by the Abbott AxSYM and IMx analyzers. J Clin Lig Assay. 1995;18: Schlain B, Frush H, Pennington C, et al. Two-stage procedure for evaluating interassay carryover on random-access instruments. Clin Chem. 1996;42: Sadler WA, Murray LM, Turner JG. Influence of specimen carryover on sensitive thyrotropin (TSH) assays: is there a problem? Clin Chem. 1996;42: Skinner E, Waterson M. Improvement of a standard MIRA method for urinary oxalate by elimination of reagent carry-over. Ann Clin Biochem. 2001;38: Muser J, Bienvenu J, Blanckaert N, et al. Inter-Laboratory evaluation of the COBAS INTEGRA 400 analytical system. Clin Chem Lab Med. 2001;39: Redondo FL, Bermudez P, Cocco C, et al. Evaluation of Cobas Integra 800 under simulated routine conditions in six laboratories. Clin Chem Lab Med. 2003;41: Wan B, Augustin R, Chan MK, et al. Analytical performance and workflow evaluation of the Roche E170 modular immunoassay analyzer in a pediatric setting. Clin Biochem. 2005;38: Pauli D, Seyfarth M, Dibbelt L. The Abbott Architect c8000: analytical performance and productivity characteristics of a new analyzer applied to general chemistry testing. Clin Lab. 2005;51: LABMEDICINE Volume 36 Number 11 November 2005 labmedicine.com Downloaded from
The Clinical Chemistry System that offers more so you do less!
The Clinical Chemistry System that offers more so you do less! A unique level of automation in a compact system Safe and flexible sample management Fully automatic cuvette management Cuvette management
More informationMiura 120 photometric tests per hour, up to 180 tests per hour with integrated ise module. Miura 200
F A M I L Y R A N D O M A C C E S S A N A L Y Z E R S D E D I C AT E D R E A G E N T S P U R E R A N D O M A C C E S S Miura One Miura 200 Miura 120 photometric tests per hour, up to 180 tests per hour
More informationVALID ONLY FOR C AND SC(ASCP) AND C AND SC(ASCP i ) TESTING DATES UP TO AND INCLUDING JUNE 30, 2019
TECHNOLOGIST AND INTERNATIONAL TECHNOLOGIST IN CHEMISTRY, C(ASCP) AND C(ASCP i ) SPECIALIST AND INTERNATIONAL SPECIALIST IN CHEMISTRY, SC(ASCP) AND SC(ASCP i ) EXAMINATION CONTENT GUIDELINE EXAMINATION
More informationProficiency Testing Evaluation
Facility/Contact Phone: GENC GAMA HASTANESI LABORATUVARI / Fax: GENC GAMA HASTANESI LABORATUVARI Email (primary): Email (secondary): (Please add pt@aafp.org to your electronic address book) ATATURK MAH.
More informationCOMBI has been developed to be easily integrated in all types of laboratories and hospitals for routine, emergency as well as specific analysis.
L A B O R A T O R Y Clinical chemistry analyzers and reagents for laboratories needs. BIOCHEMICAL SYSTEMS INTERNATIONAL, as a diagnostic device manufacturer, has been researching for 25 years, with the
More informationAn Evaluation of the Roche Cobas c 111
An Evaluation of the Roche Cobas c 111 James L. Bowling, PhD, DABCC, Alex Katayev, MD (Laboratory Corporation of America, Department of Science and Technology, Elon, NC) DOI: 10.1309/LM6T8D1LKQXVNCAC Submitted
More informationELISA QUALITY ASSURANCE Analytical Phase
LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-
More informationHumaStar 600. So easy to handle. CoreLab DX. Random-Access Clinical Chemistry System
CoreLab DX HumaStar 600 Random-Access Clinical Chemistry System So easy to handle > High throughput > Reliable performance > Workflow-optimized handling Clinical Clinical Chemistry Chemistry Movie Clinical
More informationI.S.E. S.r.l Clinical Chemistry Solutions
Fully-automated Clinical Chemistry Analyzers up to 180 tests/hour I.S.E. S.r.l Clinical Chemistry Solutions CUSTOMIZED SOLUTIONS FOR YOUR LAB check our website at www.logotech-ise.com for news and updates
More informationProviding Vital Solutions for Your Laboratory
Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine
More informationJAPANESE CHEMISTRY AND IMMUNODIAGNOSTICS MARKETS. Table of Contents
JAPANESE CHEMISTRY AND IMMUNODIAGNOSTICS MARKETS Introduction Table of Contents Worldwide Market and Technology Overview A. Major Routine Chemistry Tests 1. Albumin 2. Alkaline Phosphatase 3. ALT/SGPT
More informationProviding Vital Solutions for Your Laboratory
Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine
More informationFuture US Clinical Chemistry and Immunodiagnostics Market: Growth Opportunities and Business Expansion Strategies.
Future US Clinical Chemistry and Immunodiagnostics Market: Growth Opportunities and Business Expansion Strategies Introduction Table of Contents Worldwide Market and Technology Overview A. Major Routine
More informationCME/SAM. An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory
Clinical Chemistry / Repeat Testing Practices in Clinical Chemistry An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory Carl O. Deetz, MD, PhD,
More informationApplication Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010
Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Biochemistry Department Macclesfield District General Hospital Victoria Road Macclesfield SK10 3BL United Kingdom Contact: Jane E Clarke Tel:
More informationPositive patient identification Reagent barcode scanning Primary tubes STAT handling Reagent level sensing Test incompatibility
Performance. Value. The Selectra XL delivers what your laboratory needs - an economical solution for your chemistry testing that doesn t sacrifi ce performance and reliability. The Selectra XL is built
More informationcobas c 111 analyzer Big analyzer performance in a small package
For more information, contact your Roche Diagnostics representative or call 1-800-852-8766 (option 7). cobas c 111 analyzer Big analyzer performance in a small package www.poc.roche.com COBAS, COBAS B,
More informationStatistical validation of reagent lot change in the clinical chemistry laboratory can confer insights on good clinical laboratory practice
Original Article Annals of Clinical Biochemistry 2014, Vol. 51(6) 688 694! The Author(s) 2014 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563214520749 acb.sagepub.com
More informationRoles and responsibilities in verification of traceability of in vitro medical diagnostics (IVD)
Roles and responsibilities in verification of traceability of in vitro medical diagnostics (IVD) Federica Braga University of Milan Medical School Centre for Metrological Traceability in Laboratory Medicine
More informationValue of Clinical IVD Systems
Value of Clinical IVD Systems EAC, Enterprise Analysis Corporation, Stamford, CT Summary Having the flexibility to adapt as the healthcare landscape continues to change is critical to your success. ARCHITECT
More informationTable of Contents. Introduction. Executive Summary. Rationale for Testing Decentralization
2013 Analysis of the Point-of-Care Clinical Chemistry Market: US, Germany, Spain Physician Offices, Emergency Rooms, Operating/Recovery Suites, ICUs/CCUs, Cancer Clinics, Ambulatory Care Centers, Surgery
More informationBLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE)
PRODUCT LIST 01 02 03 04 05 BLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE) QUALITY CONTROLS ELECTRODES TABLE OF CONTENTS BLOOD GAS / ISE / CO-OXIMETRY ANALYZERS Siemens-Bayer (Ciba-Corning / Chiron) 238,248,348
More informationACID PHOSPHATASE (ACP) Continuous-spectrophotometric SFBC
Acid phosphatase (ACP) catalyzes in acid medium the hydrolysis of the phosphate group from α-naphtyl phosphate. The α-naphtol formed reacts with a diazonium salt (Fast Red TR) originating a chromogen.
More informationProduct Catalogue 2016
Product Catalogue 2016 CLINICHEM Ltd H-1117 Budafoki street 111-113 Budapest, Hungary Content Immunochemistry reagents Standards, controls Concentrated reagents (NEW) Supplying quality for clinical laboratories
More informationHumaStar 100 HumaStar 200
HumaStar 100 HumaStar 200 Key advantages Competition Product features Rev. 08d 07.09.2015 Scope of supply The slides provided in this section are concerned with availability, scope of supply and registration
More informationHuman anti-deoxyribonuclease B, anti-dnase B ELISA Kit
Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit Catalog No: E0311h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ
More informationTotal Protein Assay Kit
Total Protein Assay Kit Catalog Number KA3946 1000 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...
More informationSpeed and flexibility
Speed and flexibility Automate your environmental, water, detergent, food and beverage analysis Thermo Scientific Gallery and Gallery Plus discrete industrial analyzers Improve productivity, reproducible
More informationi-stat ctni Performance Verification vs. ABBOTT AxSYM
Report #1 i-stat ctni Performance Verification vs. ABBOTT AxSYM i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]
More informationcobas 6000 analyzer series Flexibility you can build on
cobas 6000 analyzer series Flexibility you can build on cobas 6000 analyzer series Flexibility you can build on Many of today s laboratories are challenged with delivering high standards of laboratory
More informationStatistical Report: Quantitative Urine Clinical Chemistry Mail out date: October 26, 2015
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Statistical Report: Quantitative Urine Clinical Chemistry Mail out date: October
More informationMouse Tri-iodothyronine, T3 ELISA Kit
Mouse Tri-iodothyronine, T3 ELISA Kit Catalog No: E0453m 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE
More informationAccording to NCCLS EP6-A, a quantitative analytical
Evaluation of Linearity in the Clinical Laboratory Jeffrey S. Jhang, MD; Chung-Che Chang, MD, PhD; Daniel J. Fink, MD, MPH; Martin H. Kroll, MD Context. Clinical laboratory assessment of test linearity
More information2. Reagent Store at 2-8ºC. 3. Standard - - Calcium 10 mg/dl - Store at 2-30ºC. CALCIUM Liquiform Insert. 01 English - Ref.: 90. Ref.
CALCIUM Liquiform Insert Ref.:90 Intended use. End point reaction system for determination of calcium in blood and urine. Professional Use. [For in vitro diagnostic only] Test principle. Calcium reacts
More informationRegulatory and Alternative Analytical Procedures are defined as follows 2 :
Title: Alternative Analytical Method Validation in Pharmaceuticals: Replacing a Regulatory Analytical Method in Cleaning Validation Authors: Stephen Lawson, Will McHale and Brian Wallace This re titled
More informationBS-330E Chemistry Analyzer
BS-330E Chemistry Analyzer BS-330E Chemistry Analyzer Intelligent collision protection Vertical & horizontal collision Audible alarm Ensure operator safety Smart, Versat Smart probe function Effective
More informationHuman Creatinine Serum Detection Kit
Human Creatinine Serum Detection Kit Catalog No: IRAAKT2508 Lot No: SAMPLE INTENDED USE The Serum Creatinine kit is designed to quantitatively measure creatinine present in serum samples. Please read the
More informationCreatinine Colorimetric Detection Kit
Creatinine Colorimetric Detection Kit For the quantitative determination of creatinine in urine. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 74-CRNHU-E02 Size: 2x96 Wells
More informationMouse Creatinine Urinary Detection Kit
Mouse Creatinine Urinary Detection Kit Catalog No: IRAAKT2549 Lot No: SAMPLE INTENDED USE BACKGROUND Creatinine (2 amino 1 methyl 5H imadazol 4 one) is a metabolite of phosphocreatine (p creatine), a molecule
More informationSTANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER
STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER Estimation of Dietary Intake of Sodium (DIS) in 24h Urine on electrolyte
More informationCu-Creatinine- Metol system
Quantification of Creatinine in Human Serum using Metol as a Chromogenic Probe Materials and methods 6.1. Reagents 6.1.1. N-methyl-p-aminophenol sulfate N-methyl-p-aminophenol sulfate also denoted as Metol
More informationALLOWAY METHOD OUTLINE
ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:
More informationGeneral. Power requirements: 110/220 Volts 50/60 VAC. Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm. Weight: 57 Kg.
General Power requirements: 110/220 Volts 50/60 VAC Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm Weight: 57 Kg. Features Reagents cooling Cuvettes washer Continuous input of samples Auto re-run
More informationEUROLYSER CCA 180 MARKETING MANUAL
EUROLYSER CCA 180 MARKETING MANUAL Notice Every effort has been made to avoid errors in text and diagrams, however, Eurolyser Diagnostica GmbH, assumes no responsibility for any errors which may appear
More informationTURBIDIMETRY & NEPHELOMETRY. Mohsin Al-Saleh, Senior Biomedical Scientist (BSc. MLS-2004/MSc Clini. Chem. 2013) ID#6837
TURBIDIMETRY & NEPHELOMETRY Mohsin Al-Saleh, Senior Biomedical Scientist (BSc. MLS-2004/MSc Clini. Chem. 2013) ID#6837 Lecture Objectives: Introduction (An Overview about laboratory techniques) Definition
More informationcobas e 411 analyzer Confidence in every result.
17591_Brochure.qxd 11/1/06 10:46 AM Page 2 The cobas e 411 analyzer offers... Accurate Results ECL technology delivers excellent low-end sensitivity and broad dynamic ranges Quality Results Assay-specific
More informationEvaluation of Survey results for regulatory grading
Important information regarding grading of specimen C -02 Specim en C-02 w as a carefully prepared fresh frozen hum an serum sam ple that w as sent to all participants in the Survey. Your laboratory s
More informationUniversity Wet Chemistry Analyzer Packages
University Wet Chemistry Analyzer Packages University laboratories and public research institutions have a special place in the scientific community. Often they are tasked with undertaking basic research
More informationMonkey Kidney injury molecule 1,Kim-1 ELISA Kit
Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationcobas pro integrated solutions Simplicity meets Excellence
cobas pro integrated solutions Simplicity meets Excellence Introducing cobas pro integrated solutions Simplicity meets Excellence 01 02 03 2 01 Sample supply unit Up to 1,000 samples per hour processing
More informationYampa Valley Medical Center
Yampa Valley Medical Center PROCEDURE TITLE: Chemistry - Piccolo Version Number: 2 Next Review Date: 01/28/2018 Procedure for: Laboratory Originating Source: McCartin, Rachel (Laboratory Section Head (BB/Chem/Hemo/Micro))
More informationGentian Canine CRP Immunoassay Application Note for ABX Pentra * 400
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative
More informationAutomated Clinical Analyzer
Automated Clinical Analyzer Improved ease of operation and better test efficiency, and provides the optimal solution for routine operations, STAT operation and a broad range of clinical analysis requirements.
More information[Unless otherwise stated all tolerances are ±5%] SAMPLES
PHOSPHATASE TEST - FLUOROPHOS ALP TEST SYSTEM IMS #28 [Unless otherwise stated all tolerances are ±5%] SAMPLES 1. Laboratory Requirements (see Cultural Procedures [CP] items 33 & 34) [See current version
More informationDEVELOP YOUR LAB, YOUR WAY
Technical brochure DEVELOP YOUR LAB, YOUR WAY The FLOW Solution: Expand your potential today Redesign your lab with the FLOW Solution The FLOW Solution is a highly flexible modular, semi-automated data
More informationALA 2008 Workshop Abstract
ALA 2008 Workshop Abstract Assays are usually performed on the benchtop using handheld pipettes before they graduate, or transfer, to an automated liquid handler. Automating a manual method may take time
More information01 English - Ref.: 95. Intended use. End point reaction system for determination of calcium in blood and urine.
Intended use. End point reaction system for determination of calcium in blood and urine. Test principle. Calcium reacts with arsenazo III in a solution a little acid yielding the blue colored complex calcium-arsenazo
More informationQuality-control (QC) performance measures and the QC planning process
Clinical Chemistry 43:4 602 607 (1997) Laboratory Management Quality-control (QC) performance measures and the QC planning process Curtis A. Parvin Numerous outcome measures can be used to characterize
More informationPig immunoglobulin E (IgE) ELISA Kit
Pig immunoglobulin E (IgE) ELISA Kit Catalog Number. CSB-E06803p For the quantitative determination of pig immunoglobulin E (IgE) concentrations in serum, plasma. This package insert must be read in its
More informationSIMPLICITY AND TECHNOLOGY THE LIASYS CLINICAL CHEMISTRY SYSTEMS FOR PHYSICIAN OFFICE LABORATORIES
SIMPLICITY AND TECHNOLOGY THE LIASYS CLINICAL CHEMISTRY SYSTEMS FOR PHYSICIAN OFFICE LABORATORIES INNOVATIVE AND COST-SAVING TECHNOLOGIES AT YOUR SERVICE DISTRIBUTED BY LIASYS MODELS LIASYS 330 - for POL
More informationObjectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to:
Concepts in Quality Control Data Management Bio-Rad Laboratories Quality System Specialist Thomasine Newby Objectives At the conclusion, the participant should be able to: Describe how laboratory quality
More informationAutomation of Routine Microplate Pipetting and Dispensing
Automation of Routine Microplate Pipetting and Dispensing Quantitative assays of all sorts are common to many applications in basic science and clinical research. Most of these assays require the movement,
More informationDevelopers Guide. Development of new instrument applications for D-dimer
Developers Guide Development of new instrument applications for D-dimer medirox d dimer application guide.doc MediRox AB www.medirox.se Phone: +46 (0)155 45 44 10 SE-611 82 Nyköping info@medirox.se Fax:
More informationS-UV (Sentinel Universal Vials) Reagents 6 Calibrators 16 Controls 17
LAB CATALOGUE The mission of Sentinel Diagnostics is the constant commitment to the research and development of innovative in-vitro diagnostics medical devices aimed at rendering clinical diagnostics more
More informationBy Mohsin Al-Saleh Clinical Biochemistry Department (SQUH)
Laboratory Automation By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH) Objectives: Define automation and lab automation Advantages of laboratory automation History of laboratory automation Fluidic
More informationHuman antidiuretic hormone/vasopressin/arginine vasopressin, ADH/VP/AVP ELISA Kit
Human antidiuretic hormone/vasopressin/arginine vasopressin, ADH/VP/AVP ELISA Kit Catalog No: E1139h 96 Tests Operating instruction www.eiaab.cn FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC
More informationCanine Erythropoietin,EPO ELISA Kit
Canine Erythropoietin,EPO ELISA Kit Catalog No: E0028c 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE
More informationCalcium Assay Kit. Catalog Number KA assays Version: 04. Intended for research use only.
Calcium Assay Kit Catalog Number KA1644 500 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...
More informationHuman papillomavirus,hpv ELISA Kit
Human papillomavirus,hpv ELISA Kit Catalog No: E0787h 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE
More informationThe NGAL Test Reagent Kit
PERFORMANCE DATA AND APPLICATION NOTE FOR SIEMENS DIMENSION VISTA The NGAL Test Reagent Kit REF/Cat. No. ST00CA ST00CA ST00CA Product name The NGAL Test Reagent Kit The NGAL Test Calibrator Kit The NGAL
More informationHuman rheumatoid factor (RF) antibody (IgM) ELISA Kit
Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Catalog Number. CSB-EQ027654HU For the qualitative determination of human rheumatoid factor (RF) antibody (IgM) concentrations in serum, plasma. This
More informationState of the Art for Measurement of Urine Albumin: Comparison of Routine. Measurement Procedures to Isotope Dilution Tandem Mass Spectrometry
1 of 27 State of the Art for Measurement of Urine Albumin: Comparison of Routine Measurement Procedures to Isotope Dilution Tandem Mass Spectrometry Supplemental Information, Tables and Data 2 of 27 Supplemental
More informationPotassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL
Note: Serum specimens and all materials coming in contact with them should be handled and disposed as if capable of transmitting infection. Avoid contact with skin by wearing gloves and proper laboratory
More informationAMYLASE CNPG Liquiform
AMYLASE CNPG Liquiform Insert Ref.:14 Purpose. System for determination of α-amylase in blood, urine and other biological fluids. [Only for in vitro diagnostic use] Principle. The α-amylase hydrolyzes
More informationUncertainty sources of reference measurement procedures for enzymes
Medical School Hannover Institute for Clinical Chemistry Director: Prof. Dr. K. Brand Uncertainty sources of reference measurement procedures for enzymes R. Klauke and G. Schumann Renate Strache Rainer
More informationIntended Use: Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.
2010/03/05 A93A01237BEN A11A01641 60 ml 15 ml Pentra C200 Intended Use: Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.
More informationRHEUMALATEX. 01 English - Ref.: 114. Ref.:114. Insert. Rheumatoid Factor. Precautions and warnings. Intended use. Test principle. Summary.
RHEUMALATEX Insert Ref.:114 Rheumatoid Factor Intended use. System for qualitative and semi-qualitative determination of Rheumatoid Factor (RF) on glass slide. [For in vitro diagnostic use.] Test principle.
More informationS-UV (Sentinel Universal Vials) Reagents 6 Calibrators 16 Controls 17
LAB CATALOGUE The mission of Sentinel Diagnostics is the constant commitment to the research and development of innovative in-vitro diagnostics medical devices aimed at rendering clinical diagnostics more
More informationCalibrators 16 Controls 17. Rheumatology (Latex Turbidimetry) 24 Other Proteins (Latex Turbidimetry) 25 Calibrators 27 Controls 29
LAB CATALOGUE The mission of Sentinel Diagnostics is the constant commitment to the research and development of innovative in-vitro diagnostics medical devices aimed at rendering clinical diagnostics more
More informationEvaluation of the Olympus Demand random
Automatic Chemistry of Clinical Laboratory Automation, Vol. 9, No. 3 (July-September 1987), pp. 113-118 Evaluation of the Olympus Demand random access chemistry analyser Susan B. Schotters, James H. McBride,
More informationActivity. This procedure is to be used in conjunction with the package insert for PLAC Test for Lp-PLA 2
Test for Lp-PLA 2 Abbott ARCHITECT c8000 and ci8200 Chemistry Analyzer Application Sheet This procedure is to be used in conjunction with the package insert for Test for Lp-PLA 2. Read and follow the package
More informationHigh Volume Modular Analyzer For Drug Testing and General Chemistries
High Volume Modular Analyzer For Drug Testing and General Chemistries High Volume Modular Analyzer For Drug Testing and General Chemistries Large Capacity Sample Loader and Track Auto-track loading for
More informationDesign of D-dimer Applications
Design of D-dimer Applications Introduction For us, an instrument application is a number of settings in the software of a coagulation- or clinicalchemistry analyzer, to make the performance of a reagent
More informationMouse Cholecystokinin (CCK) ELISA Kit
Mouse Cholecystokinin (CCK) ELISA Kit Catalog No. CSB-E08115m (96T) This immunoassay kit allows for the in vitro quantitative determination of mouse CCK concentrations in serum, plasma and other biological
More informationASTORIA PACIFIC. Automated Wet-Chemistry Analyzers and Diagnostic Reagents. rapid-t
ASTORIA PACIFIC Automated Wet-Chemistry Analyzers and Diagnostic Reagents rapid-t rapid-t System ASTORIA-PACIFIC PROUDLY OFFERS OUR rapid-t SYSTEM: Durable. Accurate. Precise. Economical. Open software
More informationi-stat ctni Performance Verification
Report #1 i-stat ctni Performance Verification vs. IMMULITE TURBO DPC i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for all ctni data 0.32 Critical
More information1. Biuret - Reagent (ready to use) - Store at ºC. 2. Standard g/dl - Store at ºC. TOTAL PROTEINS Insert. 01 English - Ref.
TOTAL PROTEINS Insert Ref.:99 Intended use. End point reaction system for colorimetric measurement of total proteins in blood, pleural, synovial, and ascitic fluids. Professional use. [For in vitro diagnostic
More informationBÜHLMANN fcal turbo. Reagent Kit. Calprotectin turbidimetric assay for professional use B-KCAL-RSET-U. Intended use
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET-U For research use only. Not for use in diagnostic procedures. BÜHLMANN Laboratories AG Baselstrasse 55
More information1. Substrate - - Store at 2-8ºC. AMYLASE CNPG Liquiform Insert. 01 English - Ref.: 142. Ref.:142
AMYLASE CNPG Liquiform Insert Ref.:142 Purpose. System for determination of -amylasother biological in blood, urine and fluids. Professional use. [For in vitro diagnostic use.] Principle. The α-amylase
More information2. Reagent Store at 2-8ºC. 3. Standard - - Store at 2-30ºC.
UREA UV - Liquiform Insert Ref.:104 Intended use. Enzymatic system for urea determination in serum, plasma and urine by photometry in ultraviolet using two-point, fixed-time kinetic method. Professional
More informationVerification of out-of-control situations detected by "average of normal" approach.
Bond University epublications@bond Faculty of Health Sciences & Medicine Publications Faculty of Health Sciences & Medicine 11-1-2016 Verification of out-of-control situations detected by "average of normal"
More informationauto-creatinine liquicolor
Design Verification auto-creatinine liquicolor Contents 1 Introduction...2 2 Imprecision...2 3 Linearity and Detection Limit...2 3.1 Linearity...2 3.2 Sensitivity...4 4 Comparison of Methods...4 5 Stability...5
More informationLETTER TO THE EDITOR: UNLIKELINESS LIMITS ESTIMATION
Gemma Solé Enrech, Xavier Fuentes Arderiu Letter to the Editor: Unlikeliness limits estimation LETTER TO THE EDITOR: UNLIKELINESS LIMITS ESTIMATION Gemma Solé Enrech, Xavier Fuentes Arderiu Laboratory
More informationi-stat ctni Performance Verification vs. J&J Vitros ECi
Report #1 i-stat ctni Performance Verification vs. J&J Vitros ECi i-stat Corporation - 1 - Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]
More informationHuman Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit
Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit Catalog No. CSB-E12756h (96T) This immunoassay kit allows for the in vitro quantitative determination of human MIS/AMH concentrations
More informationHuman 25-hydroxy vitamin D3 (25HVD3) ELISA Kit
Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit Catalog Number. CSB-E08097h For the quantitative determination of human 25-hydroxy vitamin D3 (25HVD3) concentrations in serum. This package insert must be
More informationBench-top system for Immunoassay & Clinical Chemistry
Bench-top system for Immunoassay & Clinical Chemistry Eclectica is a hybrid analyzer addressing the growing need for consolidation of automation platforms in clinical labs. Eclectica. which derives from
More informationDI 3500 Discrete Aoalvzer
DI 3500 Discrete Aoalvzer Automated Chemistry and Ion Analysis ':t O+Analytical:!, A World ofsolutions. - Discrete Analysis Efficient Microscale Chemistry Increasingly laboratories are being challenged
More informationPerseverance. Experimentation. Knowledge.
2410 Intuition. Perseverance. Experimentation. Knowledge. All are critical elements of the formula leading to breakthroughs in chemical development. Today s process chemists face increasing pressure to
More information