According to NCCLS EP6-A, a quantitative analytical

Size: px
Start display at page:

Download "According to NCCLS EP6-A, a quantitative analytical"

Transcription

1 Evaluation of Linearity in the Clinical Laboratory Jeffrey S. Jhang, MD; Chung-Che Chang, MD, PhD; Daniel J. Fink, MD, MPH; Martin H. Kroll, MD Context. Clinical laboratory assessment of test linearity is often limited to satisfying regulatory requirements rather than integrating this tool into the laboratory quality assurance program. Although an important part of quality control and method validation for clinical laboratories, linearity of clinical tests does not get the attention it deserves. Objective. This article evaluates the concepts and importance of linearity evaluations for clinical tests. Design. We describe the theory and procedural steps of each linearity evaluation. We then evaluate the statistical methods for each procedure. Results. Visual assessment, although simple, is subjective. The lack-of-fit error and the 1986 NCCLS EP6-P G test are sensitive to imprecision and assume that the data are first order. Regression analysis, as developed as the polynomial method, is partly based on the experiences of the College of American Pathologists Instrumentation Resource Committee and has proved to be a robust statistical method. Conclusions. We provide general guidelines for handling non-linear results from a linearity evaluation. Handling linearity data in an objective manner will aid clinical laboratorians whose goal is to improve the quality of the tests they perform. (Arch Pathol Lab Med. 2004;128:44 48) According to NCCLS EP6-A, a quantitative analytical method is said to be linear when the analyte recovery from a series of sample solutions (measured value) is linearly proportional to the actual concentration or content of the analyte (true value) in the sample solutions. 1 The points at the upper and lower limits of the analytic measurement range that acceptably fit a straight line determine the linear range. In some assays, the instrument response versus concentration of sample solutions is not linear; for example, competitive radioimmunoassays have a parabolic-shaped instrument response when plotted against concentration and a sigmoid-shaped curve when the response is plotted against the logarithm of the concentration. The responses may be transformed using a 4- parameter logistic formula or other formula such as logitlog. The test results from this transformation should be linearly proportional to the true value of the analyte in the sample solutions (Figure 1). Therefore, the curve of the instrument response, which can be parabolic or sigmoidshaped, should not be confused with linearity between the measured value and the true value. Some laboratory personnel mistake the requirement for calibration verification with that for evaluating linearity. The Clinical Laboratory Improvement Amendments of define calibration as the process of testing an analytical method to establish a relationship between known concentrations of an analyte (calibrators) and the measured value, but makes no specifications about linearity. Accepted for publication August 21, From the College of Physicians and Surgeons of Columbia University, New York, NY (Drs Jhang and Fink); Medical College of Wisconsin, Milwaukee (Dr Chang); and University of Texas Southwestern Medical Center, Dallas (Dr Kroll). Reprints: Martin H. Kroll, MD, VA Medical Center, Pathology and Laboratory Medicine, SVC (113), 4500 Lancaster Rd, Dallas, TX ( martin.kroll@med.va.gov). Calibration verification ensures accuracy and involves measuring analytes in calibrators or other samples of known values traceable to a reference method to confirm that the established relationship has remained stable. 2 In contrast, one does not need to know the absolute concentration to perform a linearity evaluation, even though knowledge of such is helpful in establishing upper and lower limits. For example, testing serial dilutions of a sample with an elevated value is an acceptable approach. It is the straight-line relationship between points that is the focus of interest in linearity evaluation. As described in NCCLS E6-P2, this linear relationship is important to clinicians who rely on this linear relationship for easy interpolation of results. 1 Prior to the availability of linearity survey programs, such as those provided by the College of American Pathologists (CAP; Northfield, Ill) and Casco (Portland, Me), individual laboratories were forced to undergo the difficult task of saving specimens from patients with elevated results for testing at the upper limits of the analytic measurement range. In the absence of testing at the upper limit of the analytic measurement range, laboratories had to narrow the range with a subsequent increase in the frequency of sample dilutions. 3 In 1988, the CAP Instrumentation Resource Committee (IRC) began offering linearity surveys as a tool for evaluating linearity. The linearity program provides pre-prepared, analyte-spiked human samples (mostly serum and some urine) covering the full, expected operating range for the analytes being tested for linearity. Although lyophilized samples were used initially, analyte-spiked serum and urine samples that do not require dilution are now being used, eliminating imprecision due to manual pipetting. Since survey samples are made to specific analyte target values, the data analysis verifies calibration within preset tolerances for the participants, unlike the use of stored patient samples, which can only be used to evaluate linearity. These sur- 44 Arch Pathol Lab Med Vol 128, January 2004 Linearity in Clinical Laboratory Testing Jhang et al

2 Figure 1. Examples of 3 patterns of recovery (recovery vs x values): linear (triangles), parabolic (squares), and sigmoid (circles). A first-order fit describes the linear recovery best. A second-order fit describes the parabolic recovery best; note that the slope changes unidirectionally. A third-order fit describes the sigmoid recovery best; note that the slope changes in 2 directions and displays an inflection point. Figure 2. Linear (circles) and loss of linearity (squares) responses. The loss of linearity response has lost its linearity at both the upper and lower ends of the data range. Figure 3. Linear plot of data without bias (circles). Linear plot of data showing proportional bias as indicated by a slope of 1.2 (triangles). Linear plot of data showing constant bias as indicated by a y-intercept of 1.8 (squares). veys are also useful because they allow comparison across laboratories and methods. Participation in a linearity and calibration verification program can add a layer of quality assessment above and beyond that provided by proficiency testing programs. Generally, linearity testing has a narrower range of acceptability than proficiency testing and is more effective in detecting analytical problems. For proficiency testing, either 3 SD from the peer group mean, an absolute percentage, or an absolute percentage deviation from the peer group mean is the usual limit, which is often disparate from the true analytical value for a given sample. Acceptability in linearity testing can have much narrower, absolute limits for error based on medically- or analyticallyrelevant criteria. Also, linearity testing challenges the entire calibration range, including the extremes, and can detect problems such as reagent or spectrophotometer deterioration earlier than quality control or proficiency testing failures. It is also good laboratory practice to periodically demonstrate linearity to detect reagent deterioration, monitor analyzer performance, or re-confirm linearity after a major servicing of equipment. 4 PERFORMANCE OF LINEARITY STUDIES Preparation of Standards The appropriate evaluation of linearity requires 5 different concentrations spanning the analytical range. Five or more samples are necessary because a sigmoid-shaped, nonlinear curve can be missed with a regression using fewer than 5 points. Even though in geometry 2 points define a line, empirical studies require at least 3 points to add an additional degree of freedom for statistical computations. If a parabolic curve is to be captured, then 3 points define the curve, but 4 points are needed to assess goodness of fit ; the 1 additional point is required for statistical calculation. A sigmoid curve can be defined by 4 points if it is symmetric about its axis, but by 5 points if it is asymmetric. The samples should be spaced where analytically relevant. Frequently, equal spacing is sufficient. Spiking a biologic matrix with known amounts of analyte, making serial dilutions, or creating mixtures with different ratios of a high and low standard are all acceptable approaches that can be used to prepare the test samples. 1 Typically, mixtures and serial dilutions have less error than individually prepared solutions. At least 2 replicate samples should then be run to allow for estimation of random error. 1 Samples should be run in random order within the same day after establishing that the instrument is calibrated and in control. Analysis of Linearity Results A wide variety of analytic and statistical methods have been developed to estimate the departure of the linearity experiment from perfect linearity. The methods for interpretation of the data have evolved over the years from simple visual inspection to statistical regression analysis. 3,5,6 The techniques have been extensively reviewed by Tholen 6 and Kroll and colleagues. 5,7 The newer procedures have been developed to determine whether deviation from linearity is significantly relevant to analytic or clinical goals. 5 The more commonly used methods, described by Tholen, 6 are adapted here for completeness, and more recently adopted methods are then described. Arch Pathol Lab Med Vol 128, January 2004 Linearity in Clinical Laboratory Testing Jhang et al 45

3 Data for Analyte Recovery From a Series of Equally Spaced Analyte-Spiked Solutions for a Linear and Nonlinear Experiment* Dilution Analyte Recovery Linear Linear Deltas Analyte Recovery Nonlinear Nonlinear Deltas 46 Arch Pathol Lab Med Vol 128, January 2004 Linearity in Clinical Laboratory Testing Jhang et al * The linear data demonstrate equal successive deltas, while the nonlinear data set shows smaller successive deltas at higher concentrations, indicating loss of linearity at the high end of the measurement range. Visual Review. The most common method used to interpret the results of a linearity experiment is visual review of a plot of the replicate mean of the measured values versus the true value for each level of the sample solutions. 6,8 The desideratum is for y-axis values (measured value) to be as close as possible to the x-axis values (true value). The points are connected, and the evaluation is then based on the degree to which the data follow a straight line (Figure 2). For experiments with known x- values, it may be useful to draw a line with a slope of 1 passing through the origin as a visual reference for deviation from perfect linearity. The best-fit line may be drawn for experiments using equally spaced sample solutions. Visual assessment is a simple and intuitive tool for an expert laboratorian, but it is subjective, unreliable, and poorly reproducible when used without expert understanding of the method. 6,8 Least Squares Linear Regression. The most commonly used method for fitting a line to the data is the least squares linear regression. 6 The true value is plotted on the x-axis. If the solutions are evenly spaced, unitless solution numbers may be assigned to the x-axis. The measured values are plotted on the y-axis. Least squares linear regression fits a straight line to a set of data points such that the sum of the squares of the vertical distance of the points to the fitted line is minimized. This minimization is performed in the vertical direction, since the x-axis represents true values. The equation is of the familiar form y mx b,wheremistheslopeofthelineandbisthe y-intercept. The y-intercept is the point where the regression line crosses the y-axis, that is, the value for y where x equals 0. This value can be either positive or negative. In more common terms, the y-intercept represents the constant systematic error or constant bias (Figure 3). The y- intercept should be as close to 0 as possible. The acceptable value of the y-intercept depends on the analyte being evaluated. Analytes for which the clinical decision points are close to the 0 point of the analytic measurement range require a y-intercept close to 0, while analytes that have clinical decision points in the middle or high end of the analytic measurement range are more tolerant of a larger y-intercept. When the solutions have known values, the ideal value of the slope of the regression line, m, is 1. The deviation of the slope of the regression line from 1 is used as an estimate of the proportional systematic error of the testing system (Figure 3). Proportional error is most often caused by incorrect assignment of the amount of substance in the calibrator. As a result, the error is consistently high or low proportional to the concentration of the analyte. Again, the level of acceptability will depend on the analyte tested. The least squares method is exquisitely sensitive to outliers, which weighs the regression heavily towards the largest values. Thus, the main drawback of least squares linear regression is that a single outlier may pull the regression line steeper or flatter. Mathematically, such problems can be reconciled by appropriately choosing the data range and using the polynomial method (see The Polynomial Method ). Another useful tool when looking at the slope is to look at the successive deltas (ie, the difference between two successive points when the target values are equally spaced or the ratio of deltas when not equally spaced). If the fit is linear, the deltas should be same for each subsequent interval. The successive deltas are the poor man s derivative; when a data set is linear, the first derivative is a constant (Table). Error. 6 The total error around the regression line is equal to the sum of the pure error or random error and the lack-of-fit error. The components of error can be determined using an appropriately constructed linearity evaluation with 2 or more replicates at each level. The pure error is the error of duplicate samples around their common mean and estimates random error. It is the sum of the squared vertical deviations from the mean of replicated samples at all sample levels. The lack-of-fit error estimates the appropriateness of the given model. The lackof-fit error is the sum of the squared vertical distances between the mean of replicates at a given level and the regression line. The total error is the sum of the pure and lack-of-fit error and can also be calculated as the sum of the squared vertical distances between all the measured values and the regression line. If the model is a good fit, the lack-of-fit error will be close to 0 and the total error will be all random error. Error and the G Test. In 1986, the NCCLS EP6-P guidelines incorporated a statistical procedure, the G test, which can be used to determine the appropriateness of a regression model. 6,8 The G statistic is defined as the ratio of the lack-of-fit error to the pure error. This ratio is an F statistic and failure is set at P.05. If the value is less than a critical value, then the fit is linear. One limitation of the G test is that it is too sensitive when precision is good and too insensitive when precision is poor. 3,9 12 Furthermore, the G test assumes that the data are first order, which is often an inappropriate assumption. Therefore, additional methods of statistical evaluation were subsequently developed. The Polynomial Method (The CAP IRC Method). The 2003 proposed revised guideline NCCLS EP6-A 1 has replaced the NCCLS EP6-P guidelines of 1986, 8 which did not provide methods to evaluate nonlinearity with clinically acceptable goals in mind. The approach proposed in 2001 as NCCLS EP6-2 is based in part on the experiences of the CAP IRC. Kroll and Emancipator 7,13 developed a polynomial method to evaluate data for nonlinearity. The CAP IRC developed a computer program incorporating the polynomial approach and also found that first-, second-, and third-order polynomials are commonly observed patterns in participant data, as expected. 5,7,13 The CAP linearity survey provides pre-diluted liquid samples (mostly serum, 1 urine) containing analyte with analyte concentrations at the upper and lower limits of the analytic measurement range. Samples are run in duplicate,

4 and results are submitted to the CAP IRC for statistical comparison with peer group results. The procedure then calculates the regression equation and uses statistical tests to determine if the data are best modeled by a linear, quadratic, or cubic equation. A detailed mathematical treatment of this technique is provided elsewhere. 5,7 The program, after fitting the data to the 3 models, assesses whether the nonlinear coefficients are statistically significant. If the nonlinear coefficients are statistically significant, then the data are nonlinear. If the nonlinear coefficients are not statistically significant, then the data are linear. If the best-fit equation is linear, then the data are called Linear 1 in the CAP IRC survey. If not, the data are nonlinear because a quadratic or cubic equation better models the data, but the data set undergoes a further check to determine if the nonlinearity in the data is clinically significant when tested against clinically relevant allowable error. 5 The process calculates the average distance between the best nonlinear fit and the linear one. The average distance is then compared against an analyte-specific bias adjusted for the random error. 5 If the average deviation is within these predetermined limits, then the deviation from linearity is not clinically important and the CAP IRC surveys designate it Linear 2. It is important to note that Linear 1 and Linear 2 are equally valid demonstrations of linearity. Finally, if the data yield an equation that is higher order than a line (quadratic or cubic) and the difference between the polynomial and the straight-line fall outside preset tolerance limits, the data are determined to be nonlinear. It should be noted that all methods for evaluating linearity must have enough statistical power to detect nonlinearity. This requires that the data have a minimum level of precision. The CAP polynomial method uses a statistically derived, formal approach for assessing whether the data contain sufficient precision to attain appropriate statistical power. 5 Therefore, the data are put through a test of precision prior to initiating the linearity test. If the data are not sufficiently precise (poor repeatability), then linearity is not assessed and the data are labeled imprecise. In other words, the data set has too much variability to assess accurately the departure from linearity, and there is insufficient statistical power to evaluate the data. 5 Calculating the ratio of the SD around the best-fit polynomial to the mean concentration for all assay solutions screens for imprecision. This ratio is compared to a quantity based on the clinically relevant tolerance limit for the analyte, the number of measurements made (number of solutions times the number of replicates), and a constant that depends on the degree of the best-fit polynomial. This statistical approach has gained general acceptance as the best statistical method to evaluate linearity of quantitative tests and has been adopted as an approved guideline (NCCLS EP6-A). 1 TROUBLESHOOTING Laboratories must ensure the reliability of test results when nonlinearities are discovered during a linearity evaluation. The specific actions will depend on the analyte, method, extent of nonlinearity, and the individual laboratory. Since each situation is different, we suggest some general guidelines for evaluating nonlinearity. During the pre-analytic steps, human error is often involved. If the materials are not prepared or stored properly, the amount of analyte may be incorrect or may deteriorate over time. Systematic error can be introduced by incorrect pipette calibration, while an imprecise pipette can result in imprecise results. Samples may be incorrectly labeled or otherwise incorrectly matched for content and testing. Analytic error can arise from various steps in the analytic process. A good starting point is to check maintenance, quality control, and calibration logs during the period prior to the linearity testing to identify measurements that are out of control. This may provide some evidence and direction for finding and correcting the source of poor performance on the linearity evaluation. Potential, common analytic problems include wavelength shift of the spectrophotometer, dirty or aging optics, and dirty or elevated background in scintillation counting wells. 4 Problems with reagents can occur as well. Improperly prepared reagents, expired reagents, or reagents near the end of their shelf life can cause nonlinearities. Transformation formulas, such as those used for immunoassays, can also be inappropriate for the binding characteristics of the antibody system being evaluated, leading to a nonlinear evaluation. One common cause of post-analytic error for this and all other surveys is incorrect transcription of the data onto the submission forms. Transcription should be verified prior to submitting data and should be confirmed if an outlying point is seen on the initial visual inspection. COMMENT Methods for the determination of linearity have evolved over time, from simple visual assessment to the polynomial method. Although visual assessment is an important step in interpretation of a linearity experiment, it is subjective and poorly reproducible. Linear regression techniques were next used to fit the data to a regression line, and the G test was added to determine appropriateness of this model. However, this method is dependent upon precision and assumes that the data set is linear. It was subsequently deemed an inappropriate test for evaluating linearity. Kroll and Emancipator 7,13 developed a method for comparing first-, second-, third-, and higher-ordered polynomials, which the CAP IRC adopted for its linearity survey program. The polynomial method uses clinically relevant goals in assessing linearity. The benefits of the CAP IRC and other linearity surveys are that the assessment picks up problems before quality control or proficiency testing failures occur. Therefore, enrollment in a linearity program can help detect problems, so that corrections can ensure accurate and reliable laboratory results. The linearity assessment can detect reagent deterioration, monitor analyzer performance, or re-confirm linearity after a major servicing. 4 Peer group comparisons allow the clinical laboratory physician to compare the laboratory s instrument to the same instruments of other enrollees. The CAP IRC monitors the survey participant reports to evaluate the adequacy of both the survey materials and the statistical tools. Ongoing considerations include adding new analytes to be tested, refining the statistical methods, deciding how to best supply the materials, and determining how to best process the samples once the product is received in a laboratory. Through proficiency testing, calibration verification, and other means, one can determine the accuracy of a method at specific points. Accuracy within all points in the analytic measurement Arch Pathol Lab Med Vol 128, January 2004 Linearity in Clinical Laboratory Testing Jhang et al 47

5 range requires that one be able to show that the mathematical relationship between input and output (concentration, activity, etc) is continuous and acceptably linear. If that relationship is nonlinear, then one must know it exactly, which requires empirical study. If it is linear, then with 2 determined points, one can generate the rest of the response curve and report a consistent, reliable, and clinically meaningful value from the entire analytical range. The authors extend their thanks to William J. Castellani, MD (Department of Pathology, Truman Medical Center, Kansas City, Mo), and the members and staff of the College of American Pathologists Standards and Instrumentation Committee for their review of this manuscript. References 1. NCCLS. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. Approved guideline NCCLS document EP6-A (ISBN ). Wayne, Pa: NCCLS; Department of Health and Human Services, Health Care Financing Administration. Clinical Laboratory Improvement Amendments of 1998; Final Rule. Federal Register (1992) (codified at 42 CFR 493). 3. Floering D. College of American Pathologists experience with the linearity surveys, Arch Pathol Lab Med. 1992;116: Kroll M, Gilstad C, Gochman G, et al, eds. Laboratory Instrument Evaluation, Verification and Maintenance Manual. 5th ed. Northfield, Ill: College of American Pathologists; 1999:1. 5. Kroll MH, Praestgaard J, Michaliszyn E, Styer PE. Evaluation of the extent of nonlinearity in reportable range studies. Arch Pathol Lab Med. 2000;124: Tholen DW. Alternative statistical techniques to evaluate linearity. Arch Pathol Lab Med. 1992;116: Kroll MH, Emancipator K. A theoretical evaluation of linearity. Clin Chem. 1993;39: Passey RB, Bee DE, Caffo A, Erikson JM. Evaluation of Linearity of Quantitative Analytical Methods. Proposed guideline EP6-P. Wayne, Pa: NCCLS; Tetrault G. Evaluation of assay linearity [letter]. Clin Chem. 1990;36: Passey RB. Evaluation of assay linearity [response to letter]. Clin Chem. 1990;36: Redondo FL. G test and evaluation of assay linearity [letter]. Clin Chem. 1990;36: Tetrault G. G test and evaluation of assay linearity [response to letter]. Clin Chem. 1990;36: Emancipator K, Kroll MH. A quantitative measure of nonlinearity. Clin Chem. 1993;39: Arch Pathol Lab Med Vol 128, January 2004 Linearity in Clinical Laboratory Testing Jhang et al

Linearity Program Users Guide

Linearity Program Users Guide U Linearity Program Users Guide New for 2016 Two New CVL Programs LN43 Lamellar Body Count CVL LN44 Fibrinogen CVL s r me e C ea e Pro e o s e L C ea ti e rotein CVL ur ey i no an in trument e i rogram

More information

ELISA QUALITY ASSURANCE Analytical Phase

ELISA QUALITY ASSURANCE Analytical Phase LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-

More information

Objectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to:

Objectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to: Concepts in Quality Control Data Management Bio-Rad Laboratories Quality System Specialist Thomasine Newby Objectives At the conclusion, the participant should be able to: Describe how laboratory quality

More information

Uncertainty of Measurement

Uncertainty of Measurement Uncertainty of Measurement Contents Uncertainty of Measurement.... 3 Uncertainty of Measurement and Measurement Error.... 4 Measuring Uncertainty.... 5 Coverage factor k...6 Factors Affecting Uncertainty....

More information

Method Validation Essentials, Limit of Blank, Limit of Detection and Limit of Quantitation

Method Validation Essentials, Limit of Blank, Limit of Detection and Limit of Quantitation Method Validation Essentials, Limit of Blank, Limit of Detection and Limit of Quantitation Thomas A. Little PhD 3/14/2015 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003

More information

Performance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies

Performance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies Clinical Chemistry 44:5 1024 1031 (1998) Laboratory Management Performance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies Kristian Linnet Application

More information

And how to do them. Denise L Seman City of Youngstown

And how to do them. Denise L Seman City of Youngstown And how to do them Denise L Seman City of Youngstown Quality Control (QC) is defined as the process of detecting analytical errors to ensure both reliability and accuracy of the data generated QC can be

More information

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010 Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use

More information

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant SWGDRUG GLOSSARY These definitions were developed and adopted by the SWGDRUG Core Committee from a variety of sources including The United Nations Glossary of Terms for Quality Assurance and Good Laboratory

More information

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative

More information

Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey

Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey Presentation Topics & Objectives Calibration Verification Key Definitions Why do I need to perform CV?

More information

Setting Performance Goals and Evaluating Total Analytical Error for Diagnostic Assays

Setting Performance Goals and Evaluating Total Analytical Error for Diagnostic Assays Clinical Chemistry 48:6 919 927 (2002) Laboratory Management Setting Performance Goals and Evaluating Total Analytical Error for Diagnostic Assays Jan S. Krouwer Background: Total analytical error has

More information

Clinical Chemistry Quality Control Policies, Processes, & Procedures

Clinical Chemistry Quality Control Policies, Processes, & Procedures Clinical Chemistry Quality Control Policies, Processes, & Procedures Prepared By: Brendan Jones Clinical Chemistry Quality Supervisor Laverty Pathology What is Quality Control? The statistical process

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Directly testing the linearity assumption for assay validation. Steven Novick / Harry Yang May, 2013

Directly testing the linearity assumption for assay validation. Steven Novick / Harry Yang May, 2013 Directly testing the linearity assumption or assay validation Steven Novick / Harry Yang May, 2013 Manuscript accepted March 2013 or publication in Journal o Chemometrics 1 Steven Novick Associate Director

More information

Statistical methods for evaluating the linearity in assay validation y,z

Statistical methods for evaluating the linearity in assay validation y,z Research Article Received: 28 February 2008, Revised: 21 June 2008, Accepted: 20 August 2008, Published online in Wiley InterScience: 6 October 2008 (www.interscience.wiley.com) DOI: 10.1002/cem.1194 Statistical

More information

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015 Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES STANDARD OPERATING PROCEDURES for Spectrophotometric Measurement Chl a, Chl b, and Bulk Carotenoids using Methanol Solvent Prepared by EcoAnalysts, Inc. 1420 South Blaine Street, Suite 14 Moscow, ID 83843

More information

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s): SOP No: ICLN-Validation Levels Between Networks.003.01 Page 1 of 19 1. Purpose: This document establishes guidelines for assigning validation levels to methods based on national and international standards.

More information

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results.

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results. Page 1 of 8 Home Science & Research Field Science and Laboratories Laboratory Manual Science & Research Volume II - Assuring the Quality of Test Results ORA-LAB.5.9 ORA Laboratory Procedure Food and Drug

More information

Part IVB Quality Assurance/Validation of Analytical Methods

Part IVB Quality Assurance/Validation of Analytical Methods 1. Introduction Part IVB Quality Assurance/Validation of Analytical Methods 1.1 Definition and purpose of validation Validation is the confirmation by examination and the provision of objective evidence

More information

Developers Guide. Development of new instrument applications for D-dimer

Developers Guide. Development of new instrument applications for D-dimer Developers Guide Development of new instrument applications for D-dimer medirox d dimer application guide.doc MediRox AB www.medirox.se Phone: +46 (0)155 45 44 10 SE-611 82 Nyköping info@medirox.se Fax:

More information

CHGA (Human) ELISA Kit

CHGA (Human) ELISA Kit CHGA (Human) ELISA Kit Catalog Number KA1884 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Incubation time too short Incubate samples overnight at 4 C or follow the manufacturer guidelines.

Incubation time too short Incubate samples overnight at 4 C or follow the manufacturer guidelines. Poor standard curve Improper standard solution Standard improperly reconstituted Standard degraded Curve doesn't fit scale Pipetting error Confirm dilutions are made correctly. Briefly spin vial before

More information

Saleh Nasiri (Ph.D.) IBTO-Research Center

Saleh Nasiri (Ph.D.) IBTO-Research Center Quality and safety improvement of blood component products in blood transfusion by using statistical methods and control charts in screening test results Saleh Nasiri (Ph.D.) IBTO-Research Center What

More information

Canine Erythropoietin,EPO ELISA Kit

Canine Erythropoietin,EPO ELISA Kit Canine Erythropoietin,EPO ELISA Kit Catalog No: E0028c 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

SPECTROPHOTOMETRY/BEER S LAW LECTURE HONORS CHEMISTRY NAME

SPECTROPHOTOMETRY/BEER S LAW LECTURE HONORS CHEMISTRY NAME SPECTROPHOTOMETRY/BEER S LAW LECTURE HONORS CHEMISTRY NAME Overview: Spectroscopy will be a tool that you will use as you continue in your chemistry, biology and physics courses. Background: We will begin

More information

Introduction to Uncertainty and Treatment of Data

Introduction to Uncertainty and Treatment of Data Introduction to Uncertainty and Treatment of Data Introduction The purpose of this experiment is to familiarize the student with some of the instruments used in making measurements in the physics laboratory,

More information

BÜHLMANN fcal turbo. Reagent Kit. Calprotectin turbidimetric assay for professional use B-KCAL-RSET-U. Intended use

BÜHLMANN fcal turbo. Reagent Kit. Calprotectin turbidimetric assay for professional use B-KCAL-RSET-U. Intended use BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET-U For research use only. Not for use in diagnostic procedures. BÜHLMANN Laboratories AG Baselstrasse 55

More information

11.1 Uncertainty and error in measurement (1 Hour) 11.2 Uncertainties in calculated results (0.5 Hour) 11.3 Graphical techniques (0.

11.1 Uncertainty and error in measurement (1 Hour) 11.2 Uncertainties in calculated results (0.5 Hour) 11.3 Graphical techniques (0. Chapter 11 Measurement and Data Processing Page 1 Students are to read and complete any part that requires answers and will submit this assignment on the first day of class. You may use internet sources

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

ECE521 week 3: 23/26 January 2017

ECE521 week 3: 23/26 January 2017 ECE521 week 3: 23/26 January 2017 Outline Probabilistic interpretation of linear regression - Maximum likelihood estimation (MLE) - Maximum a posteriori (MAP) estimation Bias-variance trade-off Linear

More information

Human papillomavirus,hpv ELISA Kit

Human papillomavirus,hpv ELISA Kit Human papillomavirus,hpv ELISA Kit Catalog No: E0787h 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2014 Beckman Coulter, Inc. All rights reserved. SALY Salicylate 378194 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date

More information

Instructions for use. Chromogranin A ELISA TM E-9000

Instructions for use. Chromogranin A ELISA TM E-9000 Instructions for use Chromogranin A ELISA TM E-9000 Chromogranin A ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Chromogranin A in serum and plasma.

More information

Date. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

Date. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. SYNCHRON CX 3 DELTA System Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Urea REF 443350 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

Draft Guideline on Bioanalytical Method (Ligand Binding Assay) Validation in Pharmaceutical Development. (24 January, 2014, MHLW, Japan)

Draft Guideline on Bioanalytical Method (Ligand Binding Assay) Validation in Pharmaceutical Development. (24 January, 2014, MHLW, Japan) Draft Guideline on Bioanalytical Method (Ligand Binding Assay) Validation in Pharmaceutical Development (24 January, 2014, MHLW, Japan) Table of Contents 1. Introduction 2. Scope 3. Reference Standard

More information

i-stat ctni Performance Verification

i-stat ctni Performance Verification Report #1 i-stat ctni Performance Verification vs. IMMULITE TURBO DPC i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for all ctni data 0.32 Critical

More information

Measurements and Data Analysis

Measurements and Data Analysis Measurements and Data Analysis 1 Introduction The central point in experimental physical science is the measurement of physical quantities. Experience has shown that all measurements, no matter how carefully

More information

Explain and apply the use of calibrators and controls in immunoassay Criteria for assay run acceptance/rejection, including Westgard rules

Explain and apply the use of calibrators and controls in immunoassay Criteria for assay run acceptance/rejection, including Westgard rules RCPA Part II Exam Study Group WEEK 8 Objectives Explain and apply the use of calibrators and controls in immunoassay 2.1.1.1. Standardisation 2.1.1.2. Traceability of calibrators 2.1.1.3. Choice of control

More information

STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS

STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS & STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS Edina Vranić¹*, Alija Uzunović² ¹ Department of Pharmaceutical Technology, Faculty of

More information

Use of Coefficient of Variation in Assessing Variability of Quantitative Assays

Use of Coefficient of Variation in Assessing Variability of Quantitative Assays CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, Nov. 2002, p. 1235 1239 Vol. 9, No. 6 1071-412X/02/$04.00 0 DOI: 10.1128/CDLI.9.6.1235 1239.2002 Use of Coefficient of Variation in Assessing Variability

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

Update on ASTM and Standard Methods method development activities

Update on ASTM and Standard Methods method development activities Update on ASTM and Standard Methods method development activities William Lipps, Eurofins Eaton Analytical Monrovia South Bend www.eatonanalytical.com Full Service Laboratories Eurofins Lancaster Laboratories

More information

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure

More information

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY PURPOSE This procedure provides instructions for developing, maintaining, and documenting, radiation safety procedures conducted at the Cyclotron Radiochemistry

More information

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit Catalog No: E0311h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

i-stat ctni Performance Verification vs. ABBOTT AxSYM

i-stat ctni Performance Verification vs. ABBOTT AxSYM Report #1 i-stat ctni Performance Verification vs. ABBOTT AxSYM i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]

More information

Hydrogen Peroxide Colorimetric Detection Kit

Hydrogen Peroxide Colorimetric Detection Kit Hydrogen Peroxide Colorimetric Detection Kit Catalog Number KA1017 96 assays Version: 06 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the

More information

MyBioSource.com. This package insert must be read in its entirety before using this product.

MyBioSource.com. This package insert must be read in its entirety before using this product. Tetracyclines ELISA Kit Catalog Number. MBS940077 This immunoassay kit allows for the in vitro quantitative determination of Tetracyclines concentrations in honey, tissue(chicken, pork). This package insert

More information

Experimental Uncertainty (Error) and Data Analysis

Experimental Uncertainty (Error) and Data Analysis Experimental Uncertainty (Error) and Data Analysis Advance Study Assignment Please contact Dr. Reuven at yreuven@mhrd.org if you have any questions Read the Theory part of the experiment (pages 2-14) and

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Really, A Revised MDL Procedure

Really, A Revised MDL Procedure NEMC 2016 Really, A Revised MDL Procedure Richard Burrows, Ph.D. TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure in the Federal

More information

Human Creatinine Serum Detection Kit

Human Creatinine Serum Detection Kit Human Creatinine Serum Detection Kit Catalog No: IRAAKT2508 Lot No: SAMPLE INTENDED USE The Serum Creatinine kit is designed to quantitatively measure creatinine present in serum samples. Please read the

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18 Draft PS 18 APPENDIX A Dynamic Spiking Procedure (4-4-2013) A1. Scope and Application 1.1 This appendix to Performance Specification 18 describes the procedure and performance requirements for dynamic

More information

camp Direct Immunoassay Kit

camp Direct Immunoassay Kit camp Direct Immunoassay Kit Catalog Number KA0886 100 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS 1.0 Purpose - This procedure specifies the required elements for the extraction and quantitation of THC,

More information

i-stat ctni Performance Verification vs. J&J Vitros ECi

i-stat ctni Performance Verification vs. J&J Vitros ECi Report #1 i-stat ctni Performance Verification vs. J&J Vitros ECi i-stat Corporation - 1 - Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]

More information

HACCP Concept. Overview of HACCP Principles. HACCP Principles. Preliminary Tasks Development of the HACCP Plan

HACCP Concept. Overview of HACCP Principles. HACCP Principles. Preliminary Tasks Development of the HACCP Plan HACCP Concept Overview of HACCP Principles A systematic approach to be used in food production as a means to assure food safety. National Advisory Committee on Microbiological Criteria for Foods (NACMCF)

More information

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4 page1 of 8 TNI Standard VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 January 2016 Description This TNI Standard has been taken through all of the voting stages and has

More information

Overview of HACCP Principles. OFFICE OF THE TEXAS STATE CHEMIST Texas Feed and Fertilizer Control Service Agriculture Analytical Service

Overview of HACCP Principles. OFFICE OF THE TEXAS STATE CHEMIST Texas Feed and Fertilizer Control Service Agriculture Analytical Service Overview of HACCP Principles OFFICE OF THE TEXAS STATE CHEMIST Texas Feed and Fertilizer Control Service Agriculture Analytical Service OFFICE OF THE TEXAS STATE CHEMIST 2 HACCP Concept A systematic approach

More information

CALC PRINCIPLE REF & REF A28937 & A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

CALC PRINCIPLE REF & REF A28937 & A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Calcium REF 467935 & 467915 REF A28937 & A28945 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Ask The Expert Webinar Series EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure Richard Burrows, Ph.D. Corporate Technical Director A Revision to the Method Detection Limit EPA

More information

Prealbumin (Mouse) ELISA KitI

Prealbumin (Mouse) ELISA KitI Prealbumin (Mouse) ELISA KitI Catalog Number KA2070 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of

More information

White Paper. Moisture Analyzer Routine Testing

White Paper. Moisture Analyzer Routine Testing Moisture Analyzer Routine Testing This white paper describes the influences and sources of error which may be present when conducting moisture analyses. It discusses the routine tests which are necessary

More information

Lack-of-fit Tests to Indicate Material Model Improvement or Experimental Data Noise Reduction

Lack-of-fit Tests to Indicate Material Model Improvement or Experimental Data Noise Reduction Lack-of-fit Tests to Indicate Material Model Improvement or Experimental Data Noise Reduction Charles F. Jekel and Raphael T. Haftka University of Florida, Gainesville, FL, 32611, USA Gerhard Venter and

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

International Journal of Pharma and Bio Sciences

International Journal of Pharma and Bio Sciences Research Article Biochemistry International Journal of Pharma and Bio Sciences ISSN 0975-6299 CONCORDANCE OF SERUM ELECTROLYTES ON TWO DIFFERENT LAB EQUIPMENTS DRY CHEMISTRY VS. ABG AND ITS IMPORTANCE

More information

Practical Statistics for the Analytical Scientist Table of Contents

Practical Statistics for the Analytical Scientist Table of Contents Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning

More information

Introducing New Technology for Liquid Handling Quality Assurance

Introducing New Technology for Liquid Handling Quality Assurance Introducing New Technology for Liquid Handling Quality Assurance Bill Gigante and John Bradshaw Ph.D. Tutorial Topics Product overview Real world example Compound Management Technology Operation Maintenance

More information

Mouse Creatinine Urinary Detection Kit

Mouse Creatinine Urinary Detection Kit Mouse Creatinine Urinary Detection Kit Catalog No: IRAAKT2549 Lot No: SAMPLE INTENDED USE BACKGROUND Creatinine (2 amino 1 methyl 5H imadazol 4 one) is a metabolite of phosphocreatine (p creatine), a molecule

More information

C3 (Mouse) ELISA Kit. Catalog Number KA assays Version: 15. Intended for research use only.

C3 (Mouse) ELISA Kit. Catalog Number KA assays Version: 15. Intended for research use only. C3 (Mouse) ELISA Kit Catalog Number KA1926 96 assays Version: 15 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

cobas c 111 analyzer Big analyzer performance in a small package

cobas c 111 analyzer Big analyzer performance in a small package For more information, contact your Roche Diagnostics representative or call 1-800-852-8766 (option 7). cobas c 111 analyzer Big analyzer performance in a small package www.poc.roche.com COBAS, COBAS B,

More information

Aflatoxin M1 (AFM1) ELISA Kit

Aflatoxin M1 (AFM1) ELISA Kit Aflatoxin M1 (AFM1) ELISA Kit Catalog Number. CSB-EL027236 This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. This package insert

More information

Design of D-dimer Applications

Design of D-dimer Applications Design of D-dimer Applications Introduction For us, an instrument application is a number of settings in the software of a coagulation- or clinicalchemistry analyzer, to make the performance of a reagent

More information

FUNDAMENTAL CONCEPTS IN MEASUREMENT & EXPERIMENTATION (continued) Measurement Errors and Uncertainty:

FUNDAMENTAL CONCEPTS IN MEASUREMENT & EXPERIMENTATION (continued) Measurement Errors and Uncertainty: FUNDAMENTAL CNCEPTS N MEASUREMENT & EXPERMENTATN (continued) Measurement Errors and Uncertainty: The Error in a measurement is the difference between the Measured Value and the True Value of the Measurand.

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Assessing Test Quality from EQA/PT Surveys. James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc.

Assessing Test Quality from EQA/PT Surveys. James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc. Assessing Test Quality from EQA/PT Surveys James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc. Madison, WI Issues to consider Is analytical quality better than needed

More information

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA For the quantitative determination of pig VEGF-A in serum, plasma and other biological fluids. Cat. No. KT-32008 For Research Use Only. Not

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

EFFICACY OF INTERNAL AND EXTERNAL QUALITY CONTROL IN CHEMICAL PATHOLOGY

EFFICACY OF INTERNAL AND EXTERNAL QUALITY CONTROL IN CHEMICAL PATHOLOGY EFFICACY OF INTERNAL AND EXTERNAL QUALITY CONTROL IN CHEMICAL PATHOLOGY Pages with reference to book, From 255 To 259 Rukhsana Lalani, Mirza Naqi Zafar, Mohammad Khurshid ( Clinical Laboratories, Department

More information

IgE (Dog) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only.

IgE (Dog) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only. IgE (Dog) ELISA Kit Catalog Number KA1937 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General Information...

More information

Correlating Federal Reference Method and Continuous PM 2.5 Monitors in the MARAMA Region

Correlating Federal Reference Method and Continuous PM 2.5 Monitors in the MARAMA Region Correlating Federal Reference Method and Continuous PM 2.5 Monitors in the MARAMA Region The following paragraphs have been extracted from the final draft report titled Correlating Federal Reference Method

More information

Experimental Uncertainty (Error) and Data Analysis

Experimental Uncertainty (Error) and Data Analysis E X P E R I M E N T 1 Experimental Uncertainty (Error) and Data Analysis INTRODUCTION AND OBJECTIVES Laboratory investigations involve taking measurements of physical quantities, and the process of taking

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Potassium REF & 467915 467935 REF & A28945 A28937 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of

More information

Lecture Outline. Biost 518 Applied Biostatistics II. Choice of Model for Analysis. Choice of Model. Choice of Model. Lecture 10: Multiple Regression:

Lecture Outline. Biost 518 Applied Biostatistics II. Choice of Model for Analysis. Choice of Model. Choice of Model. Lecture 10: Multiple Regression: Biost 518 Applied Biostatistics II Scott S. Emerson, M.D., Ph.D. Professor of Biostatistics University of Washington Lecture utline Choice of Model Alternative Models Effect of data driven selection of

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Immunoassay Kit (Colorimetric)

Immunoassay Kit (Colorimetric) RayBio cgmp Direct Immunoassay Kit (Colorimetric) User Manual Version 1.0 May 25, 2014 RayBio cgmp Direct Immunoassay Kit (Colorimetric) Protocol (Cat#: 68AT-cGMP-S100) RayBiotech, Inc. We Provide You

More information

Regulatory and Alternative Analytical Procedures are defined as follows 2 :

Regulatory and Alternative Analytical Procedures are defined as follows 2 : Title: Alternative Analytical Method Validation in Pharmaceuticals: Replacing a Regulatory Analytical Method in Cleaning Validation Authors: Stephen Lawson, Will McHale and Brian Wallace This re titled

More information

Standard Practice for Heat Aging of Plastics Without Load 1

Standard Practice for Heat Aging of Plastics Without Load 1 Designation: D 3045 92 (Reapproved 2003) Standard Practice for Heat Aging of Plastics Without Load 1 This standard is issued under the fixed designation D 3045; the number immediately following the designation

More information

Technical Notes & Troubleshooting in ELISA

Technical Notes & Troubleshooting in ELISA PISHTAZ TEB DIAGNOSTICS Technical Notes & Troubleshooting in ELISA DR. MEHDI BOUTORABI DR. ALI MIRJALILI CONTENTS Kit Selection Introduction to Brochure and related points Test provision : Equipment calibration

More information