An Evaluation of the Roche Cobas c 111
|
|
- Victor Lyons
- 6 years ago
- Views:
Transcription
1 An Evaluation of the Roche Cobas c 111 James L. Bowling, PhD, DABCC, Alex Katayev, MD (Laboratory Corporation of America, Department of Science and Technology, Elon, NC) DOI: /LM6T8D1LKQXVNCAC Submitted Revision Received Accepted Abstract Background: The cobas c 111, a new low volume bench-top analyzer from Roche Diagnostics (Indianapolis, IN), was evaluated to determine whether the operational and performance characteristics of the system would make it a good fit for the physicians office laboratories and hospital stat labs in the Laboratory Corporation of America (LabCorp) network. Methods: Twenty-five frequently ordered analytes were evaluated relative to the published specifications of the cobas c 111. Additionally correlation studies, using patient serum samples, were used to determine interinstrument comparability between 2 cobas c 111 instruments and accuracy relative to the standardized assays at LabCorp. Results: Precision, linearity, method comparison to another large Roche system, inter-instrument comparability, potential carryover, and time to result studies met or exceeded expectations. Additionally, sigma metrics analyses graded the overall performance of this analyzer as world class. Conclusions: The analytical performance, compact size, intuitive operation along with the expected low maintenance and reliability make the cobas c 111 a good fit for the intended small laboratory niche environment at LabCorp. Keywords: chemistry analyzer, bench-top, small lab, stat lab The Laboratory Corporation of America (LabCorp) evaluated the cobas c 111 chemistry analyzer for placement in physicians office laboratories and hospital stat labs where it is the potential replacement for the Roche COBAS Mira (Roche Diagnostics, Indianapolis, IN). The cobas c 111 is a bench-top system with a small footprint suitable for low volume laboratories with limited floor space. The comprehensive test menu and easy integration of stat samples support testing of specialized assays, selected clinical panels, and rapid turnaround critical care markers. The intuitive user interface, barcode reader for samples, reagents and applications, and host connectivity option make the system very attractive in situations where operators may have minimal supervision and moderate skill sets. Additionally, the cobas c 111 uses the same bulk reagents as larger Roche systems currently used by LabCorp. This allows standardization of results across the LabCorp network or any other vertically integrated health care system that also selects the larger Roche platforms in their core laboratories. The objective is to provide consistent patient results with the same reference ranges throughout the network regardless of the point of service. Corresponding Author James L. Bowling, PhD, DABCC bowlinj@labcorp.com Abbreviations LabCorp, Laboratory Corporation of America; HIPAA, Health Insurance Portability and Accountability Act; IRB, Institutional Review Board; ALB2, albumin; ALP2S, alkaline phosphatase; ALTL, alanine aminotransferase; AMYL2, amylase; ASTL, aspartate aminotransferase; DBIL, direct bilirubin; BILTS, total bilirubin; CA, calcium; CHO2A, total cholesterol; CKL, creatine kinase; CL-I, chloride; CO2-L, bicarbonate; CRE2, creatinine; GGTS2, gamma glutamyl transferase; GLU2, glucose; HDLC3, high density lipoprotein; LIP, lipase; MG, magnesium; PHOS2, phosphate; K-I, potassium; NA-I, sodium; TP2M, total protein; TRIGL, triglyceride; UREAL, urea; UA2, uric acid; CVs, coefficients of variation; SD, standard deviation; GLU, glucose; K, potassium; TEa, total allowable error Materials and Methods All reagents, calibrators, and controls were FDA approved and used according to the manufacturers recommendations on the appropriate system. Linearity was performed using Validate linearity sets (Maine Standards, Windham, ME), which are US FDA cleared for use as calibration verification/linearity materials. They are available as a liquid; 5 6 bottle sets from Maine Standards. Precision studies used 2 levels of unassayed chemistry controls, Liquichek (Bio-Rad Laboratories, Irvine, CA). Sample Collection: Analyses were performed using fresh, de-identified spent serum samples used in accordance with the Health Insurance Portability and Accountability Act (HIPAA) and Institutional Review Board (IRB) policies in place at LabCorp. Statistical Analyses: Calculations of statistical values were performed using EP Evaluator Release 7 (David G. Rhoads Associates, Kennett Square, PA). Systems: The cobas c 111 bench-top system is part of the cobas family of analyzers from Roche Diagnostics. Method comparisons for all analytes were performed against the Roche/Hitachi MODULAR ANALYTICS P 398 LABMEDICINE Volume 41 Number 7 July 2010 labmedicine.com
2 Module system, also from Roche Diagnostics. Both instruments were operated as specified in Roche s operating manuals, using Roche reagents, standards, and controls as specified in the respective package inserts. Results The cobas c 111 test menu evaluation included the following 25 analytes listed with the Roche kit designation UNCLEAR: albumin (ALB2), alkaline phosphatase (ALP2S), alanine aminotransferase (ALTL), amylase (AMYL2), aspartate aminotransferase (ASTL), direct bilirubin (DBIL), total bilirubin (BILTS), calcium (CA), total cholesterol (CHO2A), creatine kinase (CKL), chloride (CL-I), bicarbonate (CO2-L), creatinine (CRE2), gamma glutamyl transferase (GGTS2), glucose (GLU2), high density lipoprotein (HDLC3), lipase (LIP), magnesium (MG), phosphate (PHOS2), potassium (K-I), sodium (NA-I), total protein (TP2M), triglyceride (TRIGL), urea (UREAL), and uric acid (UA2). The system uses the same calibrators, controls, and reagents available on the larger Roche platforms and performance met or exceeded comparative expectations. 1 Within run precision using 2 levels of Bio-Rad Laboratories Liquichek Unassayed Chemistry Controls and between day precision using Roche Diagnostics PreciNorm and PreciPath controls are described in Table 1. Initial inspection of the precision data shows some between run coefficients of variation (CVs) that are unexpectedly less than the within run CV. This divergence is primarily true for low concentrations of ALT, BILT, CO2, HDL, LIP, MG, and UA, is related to the use of different control material, and should not distract from additional review. Intra-assay CVs were largely <3.1% except for the BILTS low control, which was 5.6% at a level of 0.87 mg/dl. However, the BILTS inter-assay precision at a level of 1.0 mg/dl was 0.0%. Interassay precision was performed by running each control level twice per day over 5 days and included 2 to 3 calibrations per analyte. All inter-assay CVs were <2%, except for CO2 and DBIL, which were 3.4% and 4.2% respectively. The effect this large CV has on the DBIL assay performance is demonstrated by the poor performance grade (Sigma 2.36) assigned by the Sigma metric data presented in Table 5. Linearity (Table 2) was performed with ready-to-use 5-level linearity sets from Maine Standards that are loaded at 100% and 75%, 50%, 25%, and 12.5% of the high sample. Also included is a base matrix 0. Each level was analyzed in triplicate, and the mean value was compared to the expected value using regression analyses. Due to the concentrations in the sample sets, some analytes were not challenged up to the limits of their measuring range; however others were challenged above the manufacturer s stated claim, and all analytes showed linear recovery as evidenced by the slope and intercept values. Table 3 lists the linear regression slopes, intercepts, and correlation coefficients for all analytes comparing the recoveries on the cobas c 111 to the Roche/Hitachi MODULAR ANALYTICS P Module system. Analyte-specific sample sets were picked to challenge the measuring range. Except for CA, CL, and DBIL all methods showed acceptable comparative recovery. Both CA and CL show the effects of correlation data generated from specimens from a narrow distribution of values relatively to the expected 0 intercept. The CA slope of is fully compensated by the intercept as results approach 10.0 mg/dl. Likewise Table 1_Precision Test Within-Run Precision Between-Run Precision Test Test Units Mean %CV Mean %CV ALB g/dl ALP U/L ALT U/L AMY U/L AST U/L DBI mg/dl BILT mg/dl CA mg/dl CHO mg/dl CK U/L CL mmol/l CO2 mmol/l CRE mg/dl GGT U/L GLU mg/dl HDL mg/dl K mmol/l LIP u/l MG mg/dl NA mmol/l PHOS mg/dl TP g/dl TRIG mg/dl UA mg/dl UREAL mg/dl labmedicine.com July 2010 Volume 41 Number 7 LABMEDICINE 399
3 Table 2_Linearity Standard Error Test Slope (SD) Intercept (SD) of the Estimate Measured Range Expected Range ALB (0.009) 0.06 (0.03) ALP (0.004) 3.1 (2.1) ALT (0.004) 2.3 (1.6) AMY (0.011) 18.2 (12.3) AST (0.006) 0.0 (2.2) DBIL (0.025) 0.17 (0.25) BILT (0.003) 0.01 (0.04) CA (0.009) 0.10 (0.09) CHO (0.008) 3.7 (3.4) CK (0.003) 2.6 (3.7) CL (0.066) 2.06 (0.77) CO (0.007) 0.8 (0.2) CRE (0.002) 0.07 (0.04) GGT (0.005) 5.3 (3.4) GLU (0.007) 1.4 (2.7) HDL (0.009) 0.9 (0.6) K (0.010) 0.18 (0.15) LIP (0.012) 2.74 (1.61) MG (0.013) 0.07 (0.04) NA (0.014) 1.1 (2.0) PHOS (0.005) 0.04 (0.05) TP (0.010) 0.13 (0.06) TRIG (0.006) 4.5 (2.4) UA (0.016) 0.31 (0.20) UREAL (0.008) 0.7 (0.5) Table 3_Method Comparison Sample # Slope Intercept Corr Analyte (N) (Deming) (Deming) (R) ALB ALP ALT AMY AST DBIL BILT CA CHO CK CL CO CRE GGT GLU HDL K LIP MG NA PHOS TP TRIG UA UREAL Table 4_Cobas c 111 Time Study Standby to Additional First Patient Profiles Profile/Test Capacity Used Profile (min) (min) Comprehensive metabolic panel 14 tests/ (basic panel, liver function) 11 cuvettes Basic metabolic panel (GLU, CA, 8 tests/5 cuvettes electrolytes, kidney function) Liver function (ALB, TP, 6 tests/6 cuvettes ALP, ALT, AST, BILT) Electrolytes (NA, K, CL, CO2) 4 tests/1 cuvette Kidney function 2 tests/2 cuvettes (UREAL, CRE) Glucose 1 test/1 cuvette 12 1 the CL slope of is compensated by the intercept of 6.1 as results approach 100 mmol/l. The DBIL slope of supports the need for method improvement, which is also indicated by the Sigma metric of 2.36 (Table 5). Interference from bilirubin, lipids, or hemolysis was not evaluated as part of this study. LabCorp has already performed these experiments on our larger Roche/Hitachi MODULAR ANA- LYTICS systems and Roche COBAS Integra systems. The cobas c 111 has the same optics as Integra, and the reagents, calibrators, and controls are largely the same across all Roche chemistry platforms, which is an important characteristic for vertically integrated organizations seeking standardization of testing throughout testing locations. 400 LABMEDICINE Volume 41 Number 7 July 2010 labmedicine.com
4 In anticipation of multiple cobas c 111 placements throughout the LabCorp network, an inter-instrument comparison between 2 systems was performed. A serum pool was aliquoted into 5 cups for each system and 5 values obtained for each analyte from which a mean and standard deviation (SD) were derived. Acceptable interinstrument recovery was a mean bias of 5% and an individual instrument SD less than the 2 SD values of the combined data. Inter-instrument recovery passed for all analytes (data not shown). Both cobas c 111 systems were additionally challenged for potential carryover using the analytes glucose (GLU) and potassium (K). Two serum samples for GLU with high (H1, H2) and low (L1, L2) values of 358 mg/dl and 11 mg/dl respectively, and 2 serum samples for K with high and low values of 13 mmol/l and 2 mmol/l respectively, were run alternately in duplicate 10 times using the following sequence: (H1, H2, L1, L2) 10. Percent carryover determined using the formula % Carryover = [(L1-L2)/(H2-L2)] 100 gave values for both GLU and K of <0.1% on both systems (data not shown). Additionally the performance of the cobas c 111 was evaluated using sigma metrics. The calculations were performed using James O. Westgard methodology. 2 This approach uses an assay s total allowable error (TEa), accuracy (bias), and precision (SD) to calculate a Sigma metric to characterize the performance of that assay. Sigma = (TEa bias)/sd Table 5_Cobas c 111 Sigma Metrics Sigma values of 6 or greater are classified as world class and suggest the opportunity to use single rule QC with wide control limits. Decreasing sigma values of 5, 4, and <3 are classified as excellent, good, and poor respectively and reflect decreasing assay performance. A total of 23 analytes have available criteria for allowable total error (%TEa) as outlined in CLIA-88 criteria for acceptable performance. For 2 methods (CO2 and GGT) the %TEa was not available at the time of the study. Table 5 shows the cobas c 111 is very close to complete world-class performance. Only 1 method (DBIL) demonstrates some challenges (sigma 2.36 graded as a poor performance). This Sigma grade identifies DBIL as an assay in need of improvement. If imprecision could be reduced for DBIL, it is very possible that it could reach world-class performance. Seventeen out of 23 tested methods performed at the world-class level, 2 demonstrated excellent performance, and 3 demonstrated good performance. Overall, the average sigma for 21 methods (excluding ones that provided infinity sigma results due to 0 imprecision) is 16.09, which grades it as world class. While the cobas c 111 has the test menu one would expect from a larger system, one must not expect an equally Test Units Mean %CV %TEa %Bias Sigma Grade ALB g/dl World Class ALP U/L World Class ALT U/L World Class AMYL U/L World Class AST U/L World Class BILT mg/dl Infinity World Class CA mg/dl World Class CHO mg/dl World Class CK U/L World Class CL mmol/l Excellent CRE mg/dl Infinity World Class DBIL mg/dl Poor GLU mg/dl Good HDL mg/dl World Class K mmol/l World Class LIP U/L World Class MG mg/dl World Class NA mmol/l Good PHOS mg/dl World Class TP g/dl Good TRIG mg/dl World Class UA mg/dl World Class UREAL mg/dl Excellent Average: World Class large throughput. Table 4 shows the results of a time study on the cobas c 111 using the following profile or test options: comprehensive metabolic panel (GLU, CA, ALB, TP, NA, K, CO2, CL, UREAL, CRE, ALP, ALT, AST, BILT), basic metabolic panel (GLU, CA, NA, K, CO2, CL, URE, CRE), liver function tests (ALB, TP, ALP, ALT, AST, BILT), electrolytes (NA, K, CO2, CL), kidney function tests (UREAL, CRE), and GLU alone. From standby mode to results for the first, 14-test, comprehensive metabolic panel was 20 minutes, with subsequent profiles available in 10 to 11 minutes each. Examination of Table 4 shows a test throughput of approximately 1 test per minute. The patient testing rate is therefore a function of the number of tests ordered. As with all instrumentation, one must carefully manage the scheduling of patient testing on the analyzer to maximize efficiency. The cobas c 111 walk away time is controlled by the instrument s incubation cuvette capacity of 60. This translates to 60 non-ise tests that can be completed before the instrument must be placed in standby to replace disposable cuvettes. As an example, if 5 Comprehensive Metabolic Panels are scheduled on the instrument only 5 cuvettes will be available for the introduction of stat testing, via the stat interrupt process. However, if only 3 panels are scheduled at 1 time, the stat interrupt mode has 27 cuvettes available to maintain stat capability while running a large panel. While the time requirement for the pre-analytical portion is not always predictable, the analytical phase for any of the scenarios tested makes a vein to brain turnaround time of 1 hour or less achievable. Discussion The cobas c 111 is a random access analyzer with stat capability offering a mix of enzymes, substrates, serum proteins, labmedicine.com July 2010 Volume 41 Number 7 LABMEDICINE 401
5 and electrolytes suitable for a small hospital or physician office laboratory. With dimensions of 29 inches wide, 22 inches deep, and 19 inches high this bench-top system is compact, robust, intuitive, and requires minimal, system-prompted maintenance. Daily start-up and system checks are prompted by the user interface and involve: 1) a simple fluid s check (water and cleaner bottles full, waste empty), 2) less than 10 minutes of daily maintenance involving an automated probe cleaning and priming of the fluid system, 3) replacing the incubation cuvette sections as indicated by the system screen, 4) loading the reagent disk and performing system-prompted calibration and QC as needed. Weekly maintenance takes about 15 minutes, and there are system-generated prompts for all maintenance interval requirements. In addition to 3 ISE assays, 14 reagents can be loaded on each reagent disk, which is stored refrigerated in a reagent disk container when not in use. Barcoded or non-barcoded patient samples can be continuously loaded into 8 positions, and a variety of primary tubes and sample cups can be used. If the cobas c 111 is linked to a host system, results that are not flagged can be automatically accepted, and data can be downloaded to a USB stick or the laboratory information system. Standby to operation takes approximately 3 minutes. Although not evaluated in this study, the menu has since been extended to include pancreatic α-amylase, c-reactive protein, high sensitive c-reactive protein, CKMB, CRE2 enzymatic, LDL-cholesterol, lactate, lactate dehydrogenase, iron, ethanol, urine ALB2, whole blood HbA1c, and ammonia. D-dimer and homocysteine are planned for later release. The results of this study support the use of the cobas c 111 in the small laboratory niche environment of less than 60 patients per day. Management of this limitation is critical as the large test menu provides the temptation to exceed the instrument s throughput capacity. The objective is to provide timely availability of a large laboratory s menu to a limited number of patients whose medical care requires near point of service testing. Although the cobas c 111 can complete the testing of panels like the 14 assay Comprehensive Metabolic Panel, the throughput of the instrument indicates workflow for such testing should be carefully managed to allow instrument time to serve more patients who need stat testing of much fewer analytes. LM Acknowledgements: This work was carried out in close cooperation with Dr. Andrea Rose from Roche Diagnostics. We thank Daniel Whisenhunt for performing the technical and laboratory part of this study. 1. Chapman M, Padden E, Pogue S, et al. Multicenter performance evaluation of the cobas c 111 analyzer at three physician office laboratories. Clin Chem. 2007;53:A Westgard JO, Ehrmeyer SS, Darcy TP. Doing the right QC. In: CLIA Final Rules for Quality Systems. Madison WI: Westgard QC; 2004: LABMEDICINE Volume 41 Number 7 July 2010 labmedicine.com
The Clinical Chemistry System that offers more so you do less!
The Clinical Chemistry System that offers more so you do less! A unique level of automation in a compact system Safe and flexible sample management Fully automatic cuvette management Cuvette management
More informationProficiency Testing Evaluation
Facility/Contact Phone: GENC GAMA HASTANESI LABORATUVARI / Fax: GENC GAMA HASTANESI LABORATUVARI Email (primary): Email (secondary): (Please add pt@aafp.org to your electronic address book) ATATURK MAH.
More informationPositive patient identification Reagent barcode scanning Primary tubes STAT handling Reagent level sensing Test incompatibility
Performance. Value. The Selectra XL delivers what your laboratory needs - an economical solution for your chemistry testing that doesn t sacrifi ce performance and reliability. The Selectra XL is built
More informationcobas c 111 analyzer Big analyzer performance in a small package
For more information, contact your Roche Diagnostics representative or call 1-800-852-8766 (option 7). cobas c 111 analyzer Big analyzer performance in a small package www.poc.roche.com COBAS, COBAS B,
More informationProviding Vital Solutions for Your Laboratory
Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine
More informationProviding Vital Solutions for Your Laboratory
Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine
More informationCOMBI has been developed to be easily integrated in all types of laboratories and hospitals for routine, emergency as well as specific analysis.
L A B O R A T O R Y Clinical chemistry analyzers and reagents for laboratories needs. BIOCHEMICAL SYSTEMS INTERNATIONAL, as a diagnostic device manufacturer, has been researching for 25 years, with the
More informationACID PHOSPHATASE (ACP) Continuous-spectrophotometric SFBC
Acid phosphatase (ACP) catalyzes in acid medium the hydrolysis of the phosphate group from α-naphtyl phosphate. The α-naphtol formed reacts with a diazonium salt (Fast Red TR) originating a chromogen.
More informationcobas 8000 modular analyzer series Intelligent LabPower
cobas 8000 modular analyzer series Introducing the cobas 8000 modular analyzer series Because demanding jobs require intelligent and powerful solutions The cobas 8000 modular analyzer series is the latest
More informationYampa Valley Medical Center
Yampa Valley Medical Center PROCEDURE TITLE: Chemistry - Piccolo Version Number: 2 Next Review Date: 01/28/2018 Procedure for: Laboratory Originating Source: McCartin, Rachel (Laboratory Section Head (BB/Chem/Hemo/Micro))
More informationSpeed and flexibility
Speed and flexibility Automate your environmental, water, detergent, food and beverage analysis Thermo Scientific Gallery and Gallery Plus discrete industrial analyzers Improve productivity, reproducible
More informationABX Pentra 400. Clinical Chemistry System. Up to 420 tests/hour 55 Assays Validation Station
ABX Pentra 400 Clinical Chemistry System Up to 420 tests/hour 55 Assays Validation Station ABX Pentra400 From routine to esoteric testing ABX Pentra 400 is an extremely compact Clinical Chemistry benchtop
More informationIdeal and powerful solution
Thermo Scientific Indiko Plus Clinical and specialty chemistry system Ideal and powerful solution for operational excellence Enzymes Substrates Electrolytes Specific Proteins Drugs of Abuse Analysis Therapeutic
More informationHumaStar 600. So easy to handle. CoreLab DX. Random-Access Clinical Chemistry System
CoreLab DX HumaStar 600 Random-Access Clinical Chemistry System So easy to handle > High throughput > Reliable performance > Workflow-optimized handling Clinical Clinical Chemistry Chemistry Movie Clinical
More informationMiura 120 photometric tests per hour, up to 180 tests per hour with integrated ise module. Miura 200
F A M I L Y R A N D O M A C C E S S A N A L Y Z E R S D E D I C AT E D R E A G E N T S P U R E R A N D O M A C C E S S Miura One Miura 200 Miura 120 photometric tests per hour, up to 180 tests per hour
More informationIDEXX VetTest Chemistry Analyzer. Lawrence A. Dwyer, Ph.D. January 2000 Senior Scientist
Performance of IDEXX VetTest Chemistry Analyzer Lawrence A. Dwyer, Ph.D. January 2 Senior Scientist IDEXX Laboratories, Inc. The Evaluation Objective: To correlate accuracy and precision of the IDEXX VetTest
More informationValue of Clinical IVD Systems
Value of Clinical IVD Systems EAC, Enterprise Analysis Corporation, Stamford, CT Summary Having the flexibility to adapt as the healthcare landscape continues to change is critical to your success. ARCHITECT
More informationPotassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL
Note: Serum specimens and all materials coming in contact with them should be handled and disposed as if capable of transmitting infection. Avoid contact with skin by wearing gloves and proper laboratory
More informationELISA QUALITY ASSURANCE Analytical Phase
LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-
More informationcobas 6000 analyzer series Flexibility you can build on
cobas 6000 analyzer series Flexibility you can build on cobas 6000 analyzer series Flexibility you can build on Many of today s laboratories are challenged with delivering high standards of laboratory
More informationTable of Contents. Introduction. Executive Summary. Rationale for Testing Decentralization
2013 Analysis of the Point-of-Care Clinical Chemistry Market: US, Germany, Spain Physician Offices, Emergency Rooms, Operating/Recovery Suites, ICUs/CCUs, Cancer Clinics, Ambulatory Care Centers, Surgery
More informationBLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE)
PRODUCT LIST 01 02 03 04 05 BLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE) QUALITY CONTROLS ELECTRODES TABLE OF CONTENTS BLOOD GAS / ISE / CO-OXIMETRY ANALYZERS Siemens-Bayer (Ciba-Corning / Chiron) 238,248,348
More informationBS-330E Chemistry Analyzer
BS-330E Chemistry Analyzer BS-330E Chemistry Analyzer Intelligent collision protection Vertical & horizontal collision Audible alarm Ensure operator safety Smart, Versat Smart probe function Effective
More informationDrug Testing Solutions INDIKO DRUG TESTING ANALYZER. 1 P F
Drug Testing Solutions INDIKO DRUG TESTING ANALYZER 1 www.dtpm.ws P. 256-845-1261 F. 888-317-4178 sales@dtpm.ws Overview Indiko Sample oriented random access system for clinical and specialty testing,
More informationepoc Blood Analysis System: Summary of Analytical Methods and Performance
White Paper epoc Blood Analysis System epoc Blood Analysis System: Summary of Analytical Methods and Performance siemens.com/poc White Paper epoc Blood Analysis System epoc Blood Analysis System White
More informationSimple Use Compact and Robust Fast and Efficient
INFECTIOUS DISEASES Simple Use Minimal hands on time Intuitive operator interface and touchscreen Simpli fies training and day-to-day operation Primary sample tubes, aliquot tubes and cups No reagent handling:
More informationJAPAN QUALITY. Automated Clinical Analyzer
JAPAN QUALITY Automated Clinical Analyzer Higher test performance and operability with new functions The clinical analyzer in the next stage 1 The best to the requirements for functionality and usability
More informationAutomated Clinical Analyzer
Automated Clinical Analyzer Improved ease of operation and better test efficiency, and provides the optimal solution for routine operations, STAT operation and a broad range of clinical analysis requirements.
More informationcobas pro integrated solutions Simplicity meets Excellence
cobas pro integrated solutions Simplicity meets Excellence Introducing cobas pro integrated solutions Simplicity meets Excellence 01 02 03 2 01 Sample supply unit Up to 1,000 samples per hour processing
More informationJAPAN QUALITY. Automated Clinical Analyzer
JAPAN QUALITY Automated Clinical Analyzer Higher test performance and operability with new functions The clinical analyzer in the next stage 1 The best to the requirements for functionality and usability
More informationGeneral. Power requirements: 110/220 Volts 50/60 VAC. Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm. Weight: 57 Kg.
General Power requirements: 110/220 Volts 50/60 VAC Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm Weight: 57 Kg. Features Reagents cooling Cuvettes washer Continuous input of samples Auto re-run
More informationProduct Catalogue 2016
Product Catalogue 2016 CLINICHEM Ltd H-1117 Budafoki street 111-113 Budapest, Hungary Content Immunochemistry reagents Standards, controls Concentrated reagents (NEW) Supplying quality for clinical laboratories
More informationApplication Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010
Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use
More informationScience. Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems
Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems Joseph Boneno, MS, SC(ASCP), 1 Michelle Fokakis, MT(ASCP), 2 Dave Armbruster, PhD, DABCC, FACB, C(ASCP) 3 ( 1
More informationVerification of out-of-control situations detected by "average of normal" approach.
Bond University epublications@bond Faculty of Health Sciences & Medicine Publications Faculty of Health Sciences & Medicine 11-1-2016 Verification of out-of-control situations detected by "average of normal"
More informationEvaluation of the Olympus Demand random
Automatic Chemistry of Clinical Laboratory Automation, Vol. 9, No. 3 (July-September 1987), pp. 113-118 Evaluation of the Olympus Demand random access chemistry analyser Susan B. Schotters, James H. McBride,
More informationVALID ONLY FOR C AND SC(ASCP) AND C AND SC(ASCP i ) TESTING DATES UP TO AND INCLUDING JUNE 30, 2019
TECHNOLOGIST AND INTERNATIONAL TECHNOLOGIST IN CHEMISTRY, C(ASCP) AND C(ASCP i ) SPECIALIST AND INTERNATIONAL SPECIALIST IN CHEMISTRY, SC(ASCP) AND SC(ASCP i ) EXAMINATION CONTENT GUIDELINE EXAMINATION
More informationcobas e 411 analyzer Confidence in every result.
17591_Brochure.qxd 11/1/06 10:46 AM Page 2 The cobas e 411 analyzer offers... Accurate Results ECL technology delivers excellent low-end sensitivity and broad dynamic ranges Quality Results Assay-specific
More informationI.S.E. S.r.l Clinical Chemistry Solutions
Fully-automated Clinical Chemistry Analyzers up to 180 tests/hour I.S.E. S.r.l Clinical Chemistry Solutions CUSTOMIZED SOLUTIONS FOR YOUR LAB check our website at www.logotech-ise.com for news and updates
More informationMouse Creatinine Urinary Detection Kit
Mouse Creatinine Urinary Detection Kit Catalog No: IRAAKT2549 Lot No: SAMPLE INTENDED USE BACKGROUND Creatinine (2 amino 1 methyl 5H imadazol 4 one) is a metabolite of phosphocreatine (p creatine), a molecule
More informationCHEMISTRY WORKSTATION. The power of proven quality GIVING YOU PEACE OF MIND
CHEMISTRY WORKSTATION The power of proven quality GIVING YOU PEACE OF MIND The new Vitalab Flexor XL extends the excellent reputation of the Flexor family. With over five thousand Flexor installations
More informationA new chapter. medicacorp.com
A new chapter. \EDIC@ medicacorp.com Same operational reliability. Reliable performance, test after test. The photometer s xenon fl ash lamp typically has a 5-year use life and is maintenance-free. No
More informationSIMPLICITY AND TECHNOLOGY THE LIASYS CLINICAL CHEMISTRY SYSTEMS FOR PHYSICIAN OFFICE LABORATORIES
SIMPLICITY AND TECHNOLOGY THE LIASYS CLINICAL CHEMISTRY SYSTEMS FOR PHYSICIAN OFFICE LABORATORIES INNOVATIVE AND COST-SAVING TECHNOLOGIES AT YOUR SERVICE DISTRIBUTED BY LIASYS MODELS LIASYS 330 - for POL
More informationHumaStar 100 HumaStar 200
HumaStar 100 HumaStar 200 Key advantages Competition Product features Rev. 08d 07.09.2015 Scope of supply The slides provided in this section are concerned with availability, scope of supply and registration
More informationHuman Creatinine Serum Detection Kit
Human Creatinine Serum Detection Kit Catalog No: IRAAKT2508 Lot No: SAMPLE INTENDED USE The Serum Creatinine kit is designed to quantitatively measure creatinine present in serum samples. Please read the
More informationCME/SAM. An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory
Clinical Chemistry / Repeat Testing Practices in Clinical Chemistry An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory Carl O. Deetz, MD, PhD,
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Biochemistry Department Macclesfield District General Hospital Victoria Road Macclesfield SK10 3BL United Kingdom Contact: Jane E Clarke Tel:
More informationDiagnostic Equipments
Diagnostic Equipments Diagnostic Equipments FACA-200 Fully Automatic Chemistry Analyzer Labomed introduces an automatic chemistry analyzer with varying testing capacities. 200 tests/hour CABINET IS OPTIONAL
More informationauto-creatinine liquicolor
Design Verification auto-creatinine liquicolor Contents 1 Introduction...2 2 Imprecision...2 3 Linearity and Detection Limit...2 3.1 Linearity...2 3.2 Sensitivity...4 4 Comparison of Methods...4 5 Stability...5
More informationCreatinine Colorimetric Detection Kit
Creatinine Colorimetric Detection Kit For the quantitative determination of creatinine in urine. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 74-CRNHU-E02 Size: 2x96 Wells
More informationAssessing Test Quality from EQA/PT Surveys. James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc.
Assessing Test Quality from EQA/PT Surveys James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc. Madison, WI Issues to consider Is analytical quality better than needed
More informationSEQC EXTERNAL QUALITY ASSURANCE PROGRAMS
SPANISH SOCIETY OF CLINICAL BIOCHEMISTRY AND MOLECULAR PATHOLOGY SEQC EXTERNAL QUALITY ASSURANCE PROGRAMS Commutable Serum with Reference Values (SCR) Impact on standardization Beatriz Boned (1) / Emma
More informationEUROLYSER CCA 180 MARKETING MANUAL
EUROLYSER CCA 180 MARKETING MANUAL Notice Every effort has been made to avoid errors in text and diagrams, however, Eurolyser Diagnostica GmbH, assumes no responsibility for any errors which may appear
More informationCP3000. Automated Coagulation System POWE R F U L LY CO M PAC T STREAM LI NED EF F I CI ENCY SI M PLE AND INTUITIVE CH OO SE T RA NS FO R MATIO N
DI AG N OST I C S CP3000 Automated Coagulation System POWE R F U L LY CO M PAC T STREAM LI NED EF F I CI ENCY SI M PLE AND INTUITIVE CH OO SE T RA NS FO R MATIO N CP3000 AUTOMATED COAGUL ATION SYSTEM The
More informationFully Automated Random Access Chemistry Analyser!
Fully Automated Random Access Chemistry Analyser! Sapphire 800 is a powerful and Intelligent Chemistry Analyser recently launched by Audit Diagnostics. The higher throughput of 800 tests/hour is designed
More informationSummary of Analytical Methods and Performance The epoc Blood Analysis System. Data Compilation provided by Alere
Summary of Analytical Methods and Performance The epoc Blood Analysis System Data Compilation provided by Alere Table of Contents Page Introduction.... 4 Glossary... 5 Methodologies.... 6 ph.... 8 pco
More informationHuman Aspartate aminotransferase, AST ELISA Kit
Human Aspartate aminotransferase, AST ELISA Kit Catalog No: E1214h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationguide [ tech ] Chemistry Analyzers Abbott Park, Ill (847) Brea, Calif (800)
Chemistry Analyzers Abbott Beckman Coulter Beckman Coulter Abbott Park, Ill (847) 937-6100 www.abbottdiagnositcs.com Name of product Abbott ARCHITECT immunochemistry analyzers AU480 AU680 1) What does
More informationCEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
CEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A31849 The CEDIA Cyclosporine PLUS assay is for the in vitro quantitative determination
More informationGentian Canine CRP Immunoassay Application Note for ABX Pentra * 400
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative
More informationCuBiAn XC C u l t u r e B i o c h e m i s t r y a n a l y z e r extra compact
CuBiAn XC C u l t u r e B i o c h e m i s t r y a n a l y z e r extra compact P h o t o m e t r i c biochemistry analyzer Photometric biochemistry analyzer CuBiAn XC P r o c e s s Development In process
More informationObjectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to:
Concepts in Quality Control Data Management Bio-Rad Laboratories Quality System Specialist Thomasine Newby Objectives At the conclusion, the participant should be able to: Describe how laboratory quality
More informationStatistical Report: Quantitative Urine Clinical Chemistry Mail out date: October 26, 2015
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Statistical Report: Quantitative Urine Clinical Chemistry Mail out date: October
More informationReviewed by. Joseph V. Bonventre. Document History Date Comment Reviewer Signature 12/20/17 Updated SOP Venkata Sabbisetti
BWH: Center for Kidney Diseases HIM 550, Harvard Institutes of Medicine 77 Louis Pasteur Avenue, Boston, MA 02115 Phone # 617-525-5988 Date: March 5, 2018 Title: human urinary 2 Plex using Randox Daytona
More informationDate. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
SYNCHRON CX 3 DELTA System Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Urea REF 443350 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED
More informationThe Effect of Substitution of Plasma for Serum on Chemistry Stat Turnaround Time
CLNCAL CHEMSTRY The Effect of Substitution of Plasma for Serum on Chemistry Stat Turnaround Time Jerome S. Nosanchuk, MD; Robert Stull, MS; and Robert Keefner, MT(ASCP) Substitution of plasma for serum
More informationTony Badrick and Graham Jones (RCPAQAP)
Session 3: Traceability in external quality assessment. Why traceability is important to EQA providers? Tony Badrick and Graham Jones (RCPAQAP) EQA the more you look the more you see! Sturgeon C. Best
More informationCLINICAL CHEMISTRY ANALYSER. The economical solution for all your needs GIVING YOU PEACE OF MIND
CLINICAL CHEMISTRY ANALYSER The economical solution for all your needs GIVING YOU PEACE OF MIND Proven and modern technology meeting all your needs: reliability, flexibility, convenience and economy. The
More informationFind out what best practice means. Immediate test results better health treatments. Closer to Care.
Find out what best practice means Immediate test results better health treatments Closer to Care. .0 21 cm 7.2 (8 ) in. 14.9 cm (5.87 in.) 32.5 cm (12.80 in.) Discover the great features of a small analyzer
More informationCu-Creatinine- Metol system
Quantification of Creatinine in Human Serum using Metol as a Chromogenic Probe Materials and methods 6.1. Reagents 6.1.1. N-methyl-p-aminophenol sulfate N-methyl-p-aminophenol sulfate also denoted as Metol
More informationHuman hepatitis B virus surface antigen(hbsag) ELISA Kit
Human hepatitis B virus surface antigen(hbsag) ELISA Kit Catalog Number. CSB-E10089h For the qualitative determination of human hepatitis B virus surface antigen (HBsAg) concentrations in serum, plasma.
More informationDEVELOP YOUR LAB, YOUR WAY
Technical brochure DEVELOP YOUR LAB, YOUR WAY The FLOW Solution: Expand your potential today Redesign your lab with the FLOW Solution The FLOW Solution is a highly flexible modular, semi-automated data
More information1. Biuret - Reagent (ready to use) - Store at ºC. 2. Standard g/dl - Store at ºC. TOTAL PROTEINS Insert. 01 English - Ref.
TOTAL PROTEINS Insert Ref.:99 Intended use. End point reaction system for colorimetric measurement of total proteins in blood, pleural, synovial, and ascitic fluids. Professional use. [For in vitro diagnostic
More informationHigh Volume Modular Analyzer For Drug Testing and General Chemistries
High Volume Modular Analyzer For Drug Testing and General Chemistries High Volume Modular Analyzer For Drug Testing and General Chemistries Large Capacity Sample Loader and Track Auto-track loading for
More informationSTA Compact The reliable analysis
STA Compact The reliable analysis Te c h n i c a l s p e c i f i c a t i o n s Dimensions: Height: Width: Depth: Weight: Space required: Height: Width: Depth: Power supply: Voltage: Frequency: 640 mm (25.2
More informationHuman rheumatoid factor (RF) antibody (IgM) ELISA Kit
Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Catalog Number. CSB-EQ027654HU For the qualitative determination of human rheumatoid factor (RF) antibody (IgM) concentrations in serum, plasma. This
More informationSTANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER
STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER Estimation of Dietary Intake of Sodium (DIS) in 24h Urine on electrolyte
More information2. Reagent Store at 2-8ºC. 3. Standard - - Store at 2-30ºC.
UREA UV - Liquiform Insert Ref.:104 Intended use. Enzymatic system for urea determination in serum, plasma and urine by photometry in ultraviolet using two-point, fixed-time kinetic method. Professional
More informationCEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems Catalog No. 10008656
More informationHuman anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit
Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit Catalog Number. MBS700087 For the qualitative determination of human anti-myelin associated glycoprotein antibody (MAG) concentrations
More informationAstoria-Pacific proudly offers our rapid-t software for the ChemWell-T an easy-to-use, compact automated chemistry analyzer:
ASTORIA-PACIFIC rapid-t with Astoria-Pacific proudly offers our rapid-t software for the ChemWell-T an easy-to-use, compact automated chemistry analyzer: Durable. Accurate. Precise. Economical Open software
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Potassium REF & 467915 467935 REF & A28945 A28937 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of
More informationPRINCIPLE. Kit Reorder # Kit Reorder # Kit Reorder # A17652 Kit Reorder # A ANNUAL REVIEW Reviewed by: Date.
SYNCHRON CX 3 DELTA System Chemistry Information Sheet Copyright 2007 Beckman Coulter, Inc. For In Vitro Diagnostic Use Chloride Kit Reorder # 450214 Kit Reorder # 472095 Kit Reorder # A17652 Kit Reorder
More informationMulticentre Evaluation of the Boehringer Mannheim/Hitachi 717 Analysis System 1 )
Baadenhuijsen et al.: Multicentre evaluation of the Hitachi 717 261 J. Clin. Chem. Clin. Biochem. Vol. 28, 1990, pp. 261-269 1990 Walter de Gruyter & Co. Berlin New York Multicentre Evaluation of the Boehringer
More informationVario Photometer DP 300 Operating Manual. Version 5.11 / 5.11 SI
1 Vario Photometer DP 300 Operating Manual Version 5.11 / 5.11 SI 2 Dear customer, Thank you for selecting Diaglobal s Vario Photometer and for the confidence you have placed in us. The Vario Photometer
More information1. Substrate - - Store at 2-8ºC. AMYLASE CNPG Liquiform Insert. 01 English - Ref.: 142. Ref.:142
AMYLASE CNPG Liquiform Insert Ref.:142 Purpose. System for determination of -amylasother biological in blood, urine and fluids. Professional use. [For in vitro diagnostic use.] Principle. The α-amylase
More informationFast and effective. quality monitoring for breweries. Thermo Scientific Gallery Plus Beermaster
Thermo Scientific Gallery Plus Beermaster Fast and effective quality monitoring for breweries Colorimetric beer, wort and malt analysis Automated bitterness analysis Thermo Scientific Gallery Plus Beermaster
More informationMonkey Kidney injury molecule 1,Kim-1 ELISA Kit
Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationAMYLASE CNPG Liquiform
AMYLASE CNPG Liquiform Insert Ref.:14 Purpose. System for determination of α-amylase in blood, urine and other biological fluids. [Only for in vitro diagnostic use] Principle. The α-amylase hydrolyzes
More informationMeasurement Uncertainty - How to Calculate It In The Medical Laboratory
Measurement Uncertainty - How to Calculate It In The Medical Laboratory Godfrey C. Moses, PhD, FCACB National Director, Gamma-Dynacare Medical Laboratories 11/18/2009 Moses, GC; GDML 1 Introduction Disclaimers
More informationHuman 25-hydroxy vitamin D3 (25HVD3) ELISA Kit
Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit Catalog Number. CSB-E08097h For the quantitative determination of human 25-hydroxy vitamin D3 (25HVD3) concentrations in serum. This package insert must be
More informationSaleh Nasiri (Ph.D.) IBTO-Research Center
Quality and safety improvement of blood component products in blood transfusion by using statistical methods and control charts in screening test results Saleh Nasiri (Ph.D.) IBTO-Research Center What
More informationASTORIA PACIFIC. Automated Wet-Chemistry Analyzers and Diagnostic Reagents. rapid-t
ASTORIA PACIFIC Automated Wet-Chemistry Analyzers and Diagnostic Reagents rapid-t rapid-t System ASTORIA-PACIFIC PROUDLY OFFERS OUR rapid-t SYSTEM: Durable. Accurate. Precise. Economical. Open software
More informationBeta-2 M Beta-2-Microglobulin
AU US Instructions For Use 2015 Beckman Coulter, Inc. All rights reserved. Beta-2 M Beta-2-Microglobulin OSR6151 4 x 10 ml R1, 4 x 8 ml R2 ANNUAL REVIEW For in vitro diagnostic use only. Reviewed by Date
More informationEvaluation of Survey results for regulatory grading
Important information regarding grading of specimen C -02 Specim en C-02 w as a carefully prepared fresh frozen hum an serum sam ple that w as sent to all participants in the Survey. Your laboratory s
More information2. Reagent Store at 2-8ºC. 3. Standard - - Calcium 10 mg/dl - Store at 2-30ºC. CALCIUM Liquiform Insert. 01 English - Ref.: 90. Ref.
CALCIUM Liquiform Insert Ref.:90 Intended use. End point reaction system for determination of calcium in blood and urine. Professional Use. [For in vitro diagnostic only] Test principle. Calcium reacts
More information01 English - Ref.: 95. Intended use. End point reaction system for determination of calcium in blood and urine.
Intended use. End point reaction system for determination of calcium in blood and urine. Test principle. Calcium reacts with arsenazo III in a solution a little acid yielding the blue colored complex calcium-arsenazo
More informationVMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual
Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3
More informationCERTIFICATE OF ANALYSIS Nitric Oxide Assay Kit
Ordering Code: EMSNO CERTIFICATE OF ANALYSIS Nitric Oxide Assay Kit Lot Number: MG157087 Product: A complete kit for the quantitative determination of nitrite and nitrate in biological fluids (2 Plate,
More informationMicrodetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats
Application Note 24 Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats INTRODUCTION Most spectrophotometric
More information