Alfredo Mirandola a) * Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy

Size: px
Start display at page:

Download "Alfredo Mirandola a) * Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy"

Transcription

1 Characterization of a multilayer ionization chamber prototype for fast verification of relative depth ionization curves and spread-out-bragg-peaks in light ion beam therapy Alfredo Mirandola a) * Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy Giuseppe Magro* Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy Universita degli Studi di Milano, Milano 20100, Italy Marco Lavagno De.Tec.Tor. S.r.l., Torino 10144, Italy Andrea Mairani Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy Heidelberg Ion Beam Therapy Center (HIT), Heidelberg 69121, Germany Silvia Molinelli, Stefania Russo, Edoardo Mastella, and Alessandro Vai Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy Davide Maestri Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy Universita degli Studi di Milano, Milano 20100, Italy Vanessa La Rosa De.Tec.Tor. S.r.l., Torino 10144, Italy Mario Ciocca Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy (Received 17 October 2017; revised 22 February 2018; accepted for publication 3 March 2018; published 6 April 2018) Purpose: To dosimetrically characterize a multilayer ionization chamber (MLIC) prototype for quality assurance (QA) of pristine integral ionization curves (ICs) and spread-out-bragg-peaks (SOBPs) for scanning light ion beams. Methods: QUBE (De.Tec.Tor., Torino, Italy) is a modular detector designed for QA in particle therapy (PT). Its main module is a MLIC detector, able to evaluate particle beam relative depth ionization distributions at different beam energies and modulations. The charge collecting electrodes are made of aluminum, for a nominal water equivalent thickness (WET) of ~75 mm. The detector prototype was calibrated by acquiring the signals in the initial plateau region of a pristine BP and in terms of WET. Successively, it was characterized in terms of repeatability response, linearity, short-term stability and dose rate dependence. Beam-induced measurements of activation in terms of ambient dose equivalent rate were also performed. To increase the detector coarse native spatial resolution (~2.3 mm), several consecutive acquisitions with a set of certified mm-thick PMMA sheets (Goodfellow, Cambridge Limited, UK), placed in front of the QUBE mylar entrance window, were performed. The ICs/SOBPs were achieved as the result of the sum of the set of measurements, made up of a one-by-one PMMA layer acquisition. The newly obtained detector spatial resolution allowed the experimental measurements to be properly comparable against the reference curves acquired in water with the PTW Peakfinder. Furthermore, QUBE detector was modeled in the FLUKA Monte Carlo (MC) code following the technical design details and ICs/SOBPs were calculated. Results: Measurements showed a high repeatability: mean relative standard deviation within 0.5% for all channels and both particle types. Moreover, the detector response was linear with dose (R 2 > 0.998) and independent on the dose rate. The mean deviation over the channel-by-channel readout respect to the reference beam flux (100%) was equal to 0.7% (1.9%) for the 50% (20%) beam flux level. The short-term stability of the gain calibration was very satisfying for both particle types: the channel mean relative standard deviation was within 1% for all the acquisitions performed at different times. The ICs obtained with the MLIC QUBE at improved resolution satisfactorily matched both the MC simulations and the reference curves acquired with Peakfinder. Deviations from the reference values in terms of BP position, peak width and distal fall-off were submillimetric for both particle types in the whole investigated energy range. For modulated SOBPs, a 2266 Med. Phys. 45 (5), May /2018/45(5)/2266/ American Association of Physicists in Medicine 2266

2 2267 Mirandola et al.: MLIC prototype for QA in particle therapy 2267 submillimetric deviation was found when comparing both experimental MLIC QUBE data against the reference values and MC calculations. The relative dose deviations for the experimental MLIC QUBE acquisitions, with respect to Peakfinder data, ranged from ~1% to ~3.5%. Maximum value of 14.1 lsv/h was measured in contact with QUBE entrance window soon after a long irradiation with carbon ions. Conclusion: MLIC QUBE appears to be a promising detector for accurately measuring pristine ICs and SOBPs. A simple procedure to improve the intrinsic spatial resolution of the detector is proposed. Being the detector very accurate, precise, fast responding, and easy to handle, it is therefore well suited for daily checks in PT American Association of Physicists in Medicine [ doi.org/ /mp.12866] Key words: ion beam therapy, MLIC, QA 1. INTRODUCTION Charged particle beams are particularly effective in cancer treatment because of their physical and radiobiological properties. The depth dose distribution of a monoenergetic heavy charged particle beam, entering the patient body, is characterized by a relatively low dose in the entrance region with low linear energy transfer (LET), a sharply prominent dose in the proximity of the penetration range (with higher LET) and a steeply distal fall-off beyond the peak. This so-called inverse depth dose profile, with respect to photon radiation, allows absorbed dose to be minimized within proximal and distal healthy tissues. Moreover, the relative biological effectiveness (RBE) values of higher Z ion beams (carbon ions, in particular) enhance the therapeutic gain, when compared with photons and lower Z particle beams, such as protons. RBE varies with LET; therefore, it depends on the penetration depth along the ionization curve (IC). 1 Since particles deposit the dose in a relatively small volume and their penetration range is highly sensitive to even small variations of material density and stopping power along the beam path, minimizing every possible source of uncertainty is a crucial aspect. Differently from conventional photon radiotherapy, for ion beams, a wrong estimation of in-patient beam penetration might cause dramatic effects in terms of tumor underdosage and/or normal tissues overdosage. 2,3 As a first step, prior to the complex task of in vivo range verification, 4,5 it is crucial to measure the Bragg peak (BP) curves with high precision, accuracy, and reliability, 6 both during the beam commissioning phase and periodic quality assurance (QA). However, a careful check of the beam energy stability, as part of the daily QA program, could be too timeconsuming if carried out with motorized water phantoms. High spatial resolution and sensitivity detectors can be used for IC measurements (such as diodes or diamonds), 7,8 but the acquisition time would be very long for daily QA, particularly if more than one particle type needs to be checked. Radiochromic films in a solid slab phantom can be used for BP position measurements for proton pencil beams, with a very satisfactory spatial resolution. 9,10 This procedure was only validated for proton beams, while procedures based on double-wedge and scanning beams were proposed for carbon ion energy checks. 10,11 However, besides the strong energy dependence of radiochromic film response, 12,13 the use of films is again time-consuming, since they need to be scanned and analyzed offline with a dedicated scanner and software. Multilayer ionization chambers (MLIC) have been therefore proposed for fast range measurements for QA purposes, especially in case of proton beams In this work, we focused on evaluating the MLIC module of the QUBE 32 channels (De.Tec.Tor., Torino, Italy) detector, for range measurements in both scanning proton and carbon ion beams. The full experimental dosimetric characterization of the detector, together with its detailed implementation into a Monte Carlo (MC) code, has been here reported, for the first time, for the two different particle types. 2. MATERIALS AND METHODS All measurements were performed at the Italian National Center for Oncological Hadronterapy (CNAO) synchrotron facility, in one of the treatment rooms equipped with a horizontal fixed beam line and pencil beam scanning (PBS) as treatment delivery modality. 20 Some representative beam energies in the whole clinical available range ( MeV and MeV/u for protons and carbon ions, respectively) were investigated for the detector characterization. The spot dimension, in terms of full width half maximum at the isocenter in air, varies as a function of the beam energy, from 2.2 to 0.7 cm and from 0.8 to 0.4 cm for protons and carbon ions respectively. 9 2.A. The MLIC QUBE detector QUBE is an in air-vented device developed for QA checks in particle beams, including three different modules in the full commercial version. The main module is composed of a MLIC detector, designed to measure the relative depth ionization distribution at different beam energies and modulations in depth. The strip chambers and the pixel chamber, used for beam position and spot dimension checks, are two optional modules that can be added on request. In this study only the MLIC main module was investigated.

3 2268 Mirandola et al.: MLIC prototype for QA in particle therapy 2268 The standard version of the MLIC QUBE module is composed of 128 channels; for this study a prototype, composed of 32 parallel plate ionization chambers, stacked in series, was used. In each chamber, 1-mm-thick aluminum plate is used as beam energy absorber and cathode. Each chamber has a sensitive area of mm 2. MLIC detector is intended for measurements of SOPB and pristine peak penetration ranges; it is not suitable for patientspecific field measurements for two main reasons: (a) the lateral dimensions of the sensitive area may not be large enough for collecting the whole ionization produced by the patientspecific field and (b) each integral plate collects the charge deposited on its whole area by all the traversing pencil beams; therefore, the device is not able to distinguish both the position and the penetration range of 2D spots for the delivered and potentially inhomogeneous patient-specific fields. The QUBE detector datasheet states that both the prototype and the standard version properly work when the operating environment ensures the humidity level between 5% and 60%, the temperature between 15 C and 30 C and the pressure range between 700 and 1060 hpa. The small size ( mm 3 ) and weight (8 kg) of the prototype make the device easy to be placed on the treatment couch. The beam passes through its mylar thin entrance window and a set of homogenous absorbing electrodes, each separated by an air gap. The ionization charge is collected along the beam penetration axis. Relative depth ionization distributions can be therefore instantaneously measured, due to the simultaneous readout of all the individual chambers. Further cross-calibration with water phantom measurements is necessary for absolute range evaluation. Both pristine and spread-out-bragg-peaks (SOBP) can be determined with a native nominal water equivalent resolution, provided by the constructor, of ~2.3 mm and therefore giving to this MLIC module a total water equivalent thickness (WET) of ~75 mm. The charge collected in the ionization chambers is converted into counts by the front-end board 21 and then sent via Transistor-Transistor-Logic protocol to a controller, for data communication. Differently from other types of MLICs, built using water equivalent channels (graphite electrodes on PMMA absorbers), 14,19 the electrodes of the MLIC investigated in this work are made of aluminum, which causes a particle scattering effect more similar to water compared to typical solid water materials A.1. Calibrations and short-term gain stability For a full characterization of the detector s materials, the thickness of a sample cathode was measured with a caliper at different positions (at the center and at the four corners of the slab). In order to perform a physical calibration of the detector, measurements of WET using the Peakfinder (PTW, Freiburg, Germany), were also carried out with both proton and carbon ion monoenergetic beams. Results achieved for the analyzed sample have been generalized to all cathodes. The ionization signal intensity for each channel is linear with the air gap through appropriately designed spacers. Hence, the admitted construction tolerance and uniformity of the spacers determine how the ionization signal is going to vary with respect to the expected value, and therefore the gain of each channel. To correct for such a variation, a gain calibration was performed before each measurement session, by acquiring high-energy ICs ( MeV for proton and MeV/u for carbon ions, whose BP depth in water corresponds to 250 and 270 mm, respectively). This ensured the MLIC module to acquire the entrance plateau region of the IC, which is approximately flat in water slightly increasing in depth and showing an initial buildup region, in particular for high-energy proton beams. 23,24 Referring to the nominal IC in water, in fact, the ratio between the value at around 75 mm in depth (i.e., the MLIC QUBE s WET) and the entrance point, is about 1.09 and 1.03 for MeV proton beam and MeV/u carbon ion beam, respectively. These channel-specific calibration factors were applied for all the measurements and MC simulations described in the following. Values of the characteristic depths and lengths (see Section (2.C)) of the ICs are not affected by the presence of minor inaccuracies in the first channels readout, which may be found by neglecting the buildup region of the applied gain calibration curve. The short-term stability of the gain calibration was assessed, for both particle types, by re-delivering the reference beam energies 1 and 3 h after the reference gain calibration acquisition. Short-term stability was assessed even in case of stress test (power supply unplugged/plugged, cables disconnected/connected, high voltage off/on). Regarding the long-term stability, the manufacturer recommends to perform a gain re-calibration at the beginning of each measurements session. Same procedure is suggested for similar commercial equipment like Zebra and Giraffe (IBA Dosimetry GmbH, Schwarzenbruck, Germany). 2.A.2. SOBP plans To assess the dependence of the MLIC QUBE both on particle type and beam energy, the detector was also tested with a set of cubic volumes of known dose, optimized in water, covering the whole available energy range. Three different cubes of cm 3, centered at 6.5, 13.5, and 24.5 cm in water, were generated using our clinical treatment planning system (TPS), Syngo RT V13 (Siemens AG, Healthcare Sector, Erlangen, Germany), for both particle species. However, a flat SOBP profile measured with the MLIC QUBE does not necessarily represent a flat depth dose distribution since the charge collected on the integral planes is not proportional to dose, but rather to dose-area product. Nevertheless, the chosen lateral dimensions (3 cm) represent a reasonable value for having a homogeneous dose distribution (both in lateral and longitudinal directions) for the volumes optimized by our TPS. Moreover, the limited dimensions of the cubic volumes, when compared with the plate area ( mm 2 ), realistically allow the deposited

4 2269 Mirandola et al.: MLIC prototype for QA in particle therapy 2269 ionization to be collected totally. For the shallowest target, a 3-cm-thick solid water (RW3, PTW Freiburg, Germany) range shifter was added. The dose distribution was optimized to be uniform in terms of physical dose for proton radiation and in terms of (local effect model-based as presented in Ref. [25]) biological dose for carbon ion radiation. High dose homogeneity was required both in the lateral and in the longitudinal directions, by adopting the clinical configuration used for patients treatments. In particular, 3 mm spot spacing was used for protons, while 2 mm spot spacing and two ripple filters 9,26 were used for carbon ions. The iso-energy slice (IES) spacing in depth was set to 2 mm for both particle types, so that each cubic volume was made of 16 IESs. 2.B. Experimental setup and measurements The experimental apparatus is shown in Fig. 1. In order to measure both pristine BPs and SOBPs in a single-shot acquisition, 32-channel QUBE prototype was aligned with the entrance window of the MLIC module at the isocenter. The mylar entrance window of the QUBE was aligned at the room isocenter and clinical beam line settings were adopted for the irradiation: two 2-mm-thick ripple filters for carbon ions, while no energy modifiers were used for protons. To explore the whole clinical energy range available by the synchrotron, a representative sample of energies has been delivered: 81.56, , , and MeV for protons while , , , , and MeV/u for carbon ions. The previous listed beam energy corresponds to BP depths in water of 50, 101, 151, and 250 mm for protons and 50, 100, 150, 200, and 250 mm for carbon ions, respectively. About and particles were delivered within five consecutive spills, for protons and carbon ions, respectively. In both cases, the delivery time needed for each measurement was less than 20 s, including the interspill pauses in which the beam is not available. Because of the reduced longitudinal size of the QUBE prototype used for this study (32 channels), a variable thickness of absorbing material (RW3) was needed to properly shift a subset of Bragg curves and SOBPs within the detector water equivalent measuring range (~75 mm). Table I summarizes the experimental setup for each measurement session. FIG. 1. Experimental setup: 32-channel QUBE. [Color figure can be viewed at wileyonlinelibrary.com] Same settings were adopted with the PTW Peakfinder water column detector, 9,11 here used as a reference, because of its superior performance in terms of spatial resolution (0.05 mm around the BP for our work). Once high acquisition repeatability was experimentally assessed (see Section (3.B)), each IC (both BPs and SOBPs) was then obtained by merging the subset of measurements made up of a one-by-one PMMA layer acquisition (each certified PMMA sheet (Goodfellow, Cambridge Limited, UK) was mm thick). 2.B.1. Short-term repeatability The acquisitions repeatability represents an essential prerequisite for the usage of the QUBE in particle beams QA. As described in the above, the procedure of increasing the intrinsic spatial resolution of the detector also requires a test retest reliability. Therefore, to assess the repeatability of the measurements under proton and carbon ion beams, nine consecutive acquisitions of monoenergetic pencil beams were performed. The beam energies investigated for this part of the study were MeV for protons and MeV/u for carbon ions. About and particles were delivered within four spills, for protons and carbon ions, respectively. In both cases, the delivery time needed for each measurement was around 10 s, including the interspill pauses in which the beam is not available. 2.B.2. Linearity with dose and dose rate dependence Given the fact that the QUBE detector is designed for fast daily QA measurements, it is of paramount importance for TABLE I. List of proton (p) and carbon ion (C) beam types and experimental setup used for the characterization of the MLIC QUBE detector prototype. Beam type Monoenergetic pencil beam (p) Monoenergetic pencil beam (C) mm ripple filters Modulated scanning beam, SOBP (p) Modulated scanning beam, SOBP (C) mm ripple filters Nominal BP position/nominal center of the MR in water [mm] Range shifter (RW3) Thickness [mm] SOBP, spread-out-bragg-peak; BP, Bragg peak; MR, modulation region, the nominal extent of the flat physical dose region.

5 2270 Mirandola et al.: MLIC prototype for QA in particle therapy 2270 the device to perform linearly with dose and with same accuracy for all beam flux levels (i.e., number of particles per second), provided by the synchrotron and commissioned for clinical use, so not being prone to day by day possible intensity variations. Linearity with dose was investigated for the reference dose rate and beam energy ( MeV/u for carbon ions and MeV for protons) at several clinically relevant beam intensities, ranging from to particles for carbon ions and from to particles for protons. These intensity values were suitable for a fast QA (from about 1 s as to 10 s of delivery time). In order to assess the dependence on the dose rate, a MeV/u monoenergetic carbon ion pencil beam of a fixed beam intensity (total number of particles of ) was delivered with three machine different configurations, as available from our synchrotron: 100%, 50%, and 20% of the full beam. The 100% of beam flux has been taken as a reference for the other beam levels, which are achieved with the insertion of intensity degraders acting as beam passing filters in the extraction line B.3. Range resolution Given the detector specifications, the available number of channels and the water equivalent measuring range, the intrinsic native spatial resolution of the MLIC QUBE is quite coarse (around 2.3 mm). Hence, it is neither enough to precisely determine the position of the BP in depth nor to fulfill the accuracy requested by our QA protocol, 9 particularly for low-energy beams with the narrower BP width. A direct comparison between the IC acquired with the MLIC QUBE detector and those acquired with the Peakfinder cannot be easily performed, both due to a different spatial resolution and medium which the particle beams interact with (mainly aluminum for the MLIC QUBE and water for the Peakfinder). Therefore, some procedures or methods, at least to increase the spatial resolution of the detector under study, should be implemented. Nevertheless, differences of the detectors material may affect the peak-to-entrance ratio of the acquired ionization profiles, being more pronounced for higher energies and heavier ions, for which nuclear interaction processes dominate. The former task was successfully accomplished by performing several acquisitions in the same conditions, both in terms of number of particles delivered and dose rate, with a series of thin PMMA layers (WET of 0.22 mm each) placed in front of the QUBE mylar entrance window. This way potentially misleading identification of the maximum signal channel with the one of reference for the BP position, due to the obvious limitation of the sampling procedure, is prevented. 2.B.4. Radiation-induced device activation In order to give an indication about the potential risks to the operators handling the QUBE detector, an evaluation of the device activation was performed after the irradiations. Measurements were carried out 50 cm away from the detector (roughly representing the distance which the operators work at) and at the entrance window (representing the highest activation area). The induced radioactivity was assessed with a plastic scintillator model AT1123 (ATOMTEX, Belarus) expressed in terms of ambient dose equivalent rate, _H ð10þ, with an uncertainty of around 5%. The measurements were performed for both short (about 110 s with and particles for carbon ions and protons, respectively) and long acquisition of about 20 min with particles delivered with a MeV/u carbon ion beam. The latter was chosen to assess radioactive contamination in a high dose scenario. In addition, for the long acquisition case, a comparative evaluation on _H ð10þ was performed in the same experimental condition with the PTW Peakfinder and the IBA Giraffe detector. 2.C. Data analysis Aiming at a full characterization the MLIC QUBE detector, the BP position here defined as the mean value between the 95% proximal and distal of the maximum dose level has been evaluated. Then, a deeper investigation of the experimental pristine ICs was carried out by evaluating the following quantities for both the MLIC QUBE and the MC simulations: BP full width at 50% (70%) of the maximum (FW50% M, FW70%M), defined as the distance between the 50% (70%) distal and proximal dose levels; Dose distal fall-off (DDF), 28 defined as the distance between the 80% distal and the 20% distal of the IC, normalized to the maximum dose level; The practical range (R p ), 28 defined as the depth, distal to the BP, at which the dose is 10% of its maximum value; Additionally, for carbon ion beams, a further analysis was performed about the fragmentation tail. The integral dose between the 110% of the position of maximum ionization value and the last available data point in depth has been computed and presented in terms of percentage difference with respect to the Peakfinder calculated value. For the analysis of the SOBPs the following parameters have been assessed: Protons: Range equivalent depth 28 difference (DR): the difference between the MLIC QUBE (both experimental and MC) and the Peakfinder measured position of the distal 80% dose of the SOBP normalized at the center of the modulation region (MR, the nominal extent of the flat physical dose region); R p as defined in the above, with the exception of the normalization point, now taken at the center of the MR;

6 2271 Mirandola et al.: MLIC prototype for QA in particle therapy 2271 SOBP width defined as the distance between the distal 90% and the proximal 90% dose normalized at the center of the MR; DDF as the distance between the distal 20% and 80% dose normalized at the center of the MR. Carbon ions: Depth of the distal 50% of the dose normalized at the center of the MR (the nominal extent of the corresponding flat biologically optimized dose region); as an estimate of the particle range; DDF as the distance between the distal 30% and 70% dose normalized at the center of the MR Depth dose profiles, normalized at the center of MR, were further analyzed by taking the average of the absolute values of the relative dose differences, between Peakfinder (reference) and MLIC QUBE/MC data, along the curve. This served as an evaluator of the dosimetric agreement. In the case of carbon ions, a slightly different analysis has been performed. The physical dose related to a LEM-optimized SOBP is obviously not flat as a function of depth in water, and the same analysis proposed for proton SOBPs cannot be easily adapted to this context. Moreover, the relatively high-dose level of the measured fragmentation tails prevented the low-dose points (i.e., distal 10% or 20%) to be considered as good indicators for both the DDF and the particle range, as defined in Ref. [28]. Unless otherwise stated, results have been presented as differences with respect to the Peakfinder measured values, taken as reference. The measured ICs, here investigated for the repeatability, short-term stability of the gain calibration, linearity with dose and dose rate dependence assessment, were all normalized to the reading of the first channel of the detector. A threshold of 3% of the maximum measured value of each acquisition was additionally applied to exclude those channels readings which could be potentially affected by readout fluctuations at very low-dose levels. Considering repeatability and short-term stability analysis, results have been presented in terms of channel mean relative standard deviation (i.e., mean coefficient of variation), defined as the average value of the ratios between the standard deviation and mean readouts for each channel of the MLIC QUBE. The evaluation was performed over nine consecutive measurements for the repeatability assessment and over three acquisitions at different times for the short-term stability assessment. In order to quantify the detector linearity with dose, the channels readouts were analyzed as a function of five beam intensity levels (number of particles) for both protons and carbon ion beams of a fixed energy, as reported in Section (2.B.2). The R 2 index was then used as a goodnessof-fit statistics for the linear fit. For the dose rate dependence, an average value over the channel-by-channel deviation to the reference beam flux (100%) was calculated for both the 50% and the 20% of the full beam. For the sake of completeness, minimum and maximum values of the averaged interval have been reported for the repeatability, short-term gain stability and dose rate dependence measurements. For the SOBPs analysis, a dose threshold criterion (1.5% of the highest dose value) coupled with a gradient-based (0.5 a.u./mm) exclusion criterion have been applied to exclude values potentially affected by readout fluctuations at lowdose levels and points lying on steep dose gradients (i.e., points on the DDF), which may cause major deviations between the two datasets under comparison. 2.C.1. MC validation By following the detector design details, both mylar thin entrance window and each electrode were modeled with all its constituents to cover its nominal thickness of 12 lm and 1.16 mm, respectively. Short preliminary run allowed to slightly adjusting the MC input parameter for the aluminum ionization potential (I) to obtain the average value for the WET as measured experimentally. As representative energies, to retrieve a suitable value for I, a MeV (BP at 151 mm) and a MeV/u (BP at 200 mm) proton and carbon ion pencil beams were used. The null shift of the BP position reported in Table III for these two energies is a proof of the described procedure. As already specified in Section (2.C), all the parameters listed for the analysis of BPs and SOBPs have been also extracted from the MC outcomes. For a complete cross-validation of the MC simulation of the MLIC QUBE detector, a further analysis has been then performed by evaluating the mean of the absolute values of the local dose differences between MC and MLIC QUBE data, normalized to the experimental value. To avoid misleading results, again a dose threshold (1.5% of the maximum dose value) and a dose-gradient (0.5 a.u./mm) exclusion criteria have been applied. The percentage of remaining data (1-g) was used as normalization factor for this mean absolute deviation (Adj. MAD, Eq.(1)) in order to make an un-biased overall evaluation of the agreement between MC and experimental data: Adj:MAD ½%Š ¼ 1 1 g 1 X di MC di DATA (1) N i d DATA i where g is the percentage of data satisfying the exclusion criteria reported in the above, d i.is the dose at i-th point for MC and experimental data and N is the total number of point doses. 3. RESULTS 3.A. Calibrations and short-term gain stability The resulting thickness of the sample cathode was mm, thus confirming that the manufacturing process, the coupling of the material components and the

7 2272 Mirandola et al.: MLIC prototype for QA in particle therapy 2272 gluing has been done very accurately. As far as WET measuring is concerned, slight differences have been found between proton and carbon ions when using the same set of energies as listed in Section (2.C). In particular, the relative water equivalent path length (rwepl, 9 ) of the sample cathode ranged from 1.96 to 2.02 for protons, with the increasing energy. Higher values of rwepl were measured for carbon ions, ranging from 2.10 to The WET of the thin mylar entrance window was 30 lm for both particle types. As a result of the gain calibration process (see subsection (2.A.1)), the detector readout was verified to be in full agreement with reference calibration IC in water, channel-by-channel, within 1%. The short-term stability of the gain calibration was assessed for both particle types by comparing, channel by channel, three gain calibration acquisitions, delivered at different times (see subsection (2.A.1)). Results show a very good stability: the mean relative standard deviation of the channels readout was less than 1% for both particle types. As an indication, in case of the proton gain calibration, the relative standard deviation ranged from 0.2% to 0.7%, with a mean of 0.4%. 3.B. Repeatability, linearity with dose, and dose rate dependence As reported in Table II, the agreement among the nine runs performed is very satisfactory for both particle species, being the mean relative standard deviation of the channels readout less than 0.5%. Moreover, this value is ranging from about 0% to 1%, which can be a further assessment that all the channels well perform independently of their position relative to the Bragg curve. The linearity with dose was assessed by delivering, at a fixed dose rate, the same beam energy at five different beam intensities (number of particles delivered) for both proton and carbon ion beams. As an example, in case of carbon ions, Fig. 2 shows how the response of each channel, normalized to the first channel reading, is independent on the investigated beam intensity range. The channel reading as a function of number of particles is reported in the upper left panel of the Fig. 2, for a representative channel (#15). A good linear behavior was TABLE II. Channel mean, minimum, and maximum relative deviations for a proton (p) and a carbon ion (C) monoenergetic beam, specified by their nominal Bragg peak position in water. Beam type Nominal BP position [mm] Channel mean relative standard deviation (min.; max.) [%] Monoenergetic pencil beam (p) (0.1; 0.9) Monoenergetic pencil beam (C) (0.0; 1.2) mm ripple filters Abbreviations as defined in Table I. observed, with the R 2 of the linear best fit >0.998 for all the 32 channels of the prototype, for any particle type. The MLIC QUBE reading has been shown to be independent also on the dose rate, since the mean dose deviation with respect to the reference 100% beam flux level was 0.7% (ranging from 0.2% to 2.6%) and 1.9% (ranging from 0.3% to 6.0%) for the 50% and the 20% of the full beam, respectively. A deeper analysis on the electronics of the detector showed how the largest deviations (i.e., 2.6% and 6.0%) were attributed, for both cases, to a single defective channel. By dropping out these data, the final mean deviations remain substantially unchanged. 3.C. Procedure for range resolution improvement As introduced in Section (2.B.3), the spatial resolution of the detector has been increased up to about 0.5 mm by adopting thin plastic layers of well-known WET. As an example, Fig. 3 shows a convolution of four ICs (corresponding to the insertion of none, 1, 3 and 5 PMMA layers) as measured by the MLIC QUBE, for which the range resolution has been dramatically increased with respect to a single acquisition. The resulting superposition well reproduces the main features (for instance, BP position and DDF) of the extremely resolved Peakfinder-based acquisition (see the dotted line). At this stage, each layer was manually inserted, which explains the choice of four experimental configurations (see Fig. 3 and followings sections) as a good compromise between the number of acquisitions (i.e., detector s runs) and an acceptable spatial resolution. Since an optimization of the detector is ongoing, the range resolution could still be possibly improved with a QUBE-specific add-on, which automatically inserts the thin PMMA layers in front of the entrance window. 3.C.1. Monoenergetic pencil beams Figure 4 shows the comparison among the Peakfinder reference curves, MLIC QUBE data and the MC simulation, reproducing the same experimental setup. As reported in Table III, all parameters characterizing the ICs show a very good agreement with the reference values for both particle types in the full range of the investigated energies. For protons at MeV a discrepancy, in terms of plateau to peak ratio, is observed between Peakfinder and QUBE/MC data. This is probably due to the presence of a thick range shifter (140 mm thick for this specific case) that degrades the beam energy and considerably increases the scatter effect. Except for the FW50%M of carbon ions at 100 mm in water, only submillimeter differences can be noticed for all the peak widths. On average, a mm shift and a mm shift was registered for experimental nominal BP positions and both DDF and practical range, respectively. Similar results were found for MC. Even the integral fragmentation tails (see Section (2.C)), acquired either with the MLIC

8 2273 Mirandola et al.: MLIC prototype for QA in particle therapy 2273 FIG. 2. Monoenergetic MeV/u carbon ion pencil beam passing through mm ripple filters. Five different Bragg curves are shown with different markers depending on the beam intensity as displayed in the legend. Relative ionization for each channel is displayed for the five acquisitions with the MLIC QUBE. Each dataset has been normalized to the first channel reading. The upper left panel shows the linear behavior of one representative channel (#15). [Color figure can be viewed at wileyonlinelibrary.com] FIG. 3. Monoenergetic MeV/u carbon ion pencil beam passing through mm ripple filters +140-mm-thick RW3 range shifter. The Bragg curve is shown with a dotted line, acquired by the Peakfinder. The dots indicate four different MLIC QUBE configurations: none, one, three, and five plastic layers in front of the mylar entrance window. Each dataset has been shifted backwards by the water equivalent thickness of the added PMMA layers. The upper right panel shows a magnification of the peak region. The curves are normalized at the lowest depth achievable with the Peakfinder (13.4 mm). [Color figure can be viewed at wileyonlinelibrary.com] QUBE or simulated with MC, satisfactorily match the reference data: deviations were less than 1%. For the highest carbon ion beam energy, a slight larger deviation was found in the measured fragmentation tail: this may be due to the presence of a thick range shifter which worsens the beam quality. simulations against the Peakfinder reference values. Dose absolute deviations for the experimental MLIC QUBE acquisitions were within 1% to 3.5%. Nevertheless, they are still in close agreement with the MC, thus leading to conclude that MC outcomes may be independent of the modeling accuracy. 3.C.2. Modulated scanning beams (SOBPs) Figure 5 shows the comparison among the Peakfinder reference data, MLIC QUBE acquisitions and MC simulations reproducing the same experimental setup. On the whole, the MLIC QUBE measured values match the overall SOBP shapes and agree with the reference data satisfactorily, as well as the MC predicted values. The results concerning the parameters characterizing the SOBPs are listed in Table IV. On average, a submillimetric deviation was found when comparing both experimental MLIC QUBE data and MC 3.D. MC simulations As a further validation of the MC code, a specific analysis was performed between MLIC QUBE data and MC predictions. The adjusted MAD (see Eq. (1)) has been computed as described in subsection (2.C.1) and results are reported in Table V. The MC predicted values for the IC, which were verified channel by channel, show discrepancy lower than 5.6%. This is an acceptable value considering how the adjusted MAD may overestimate the potential disagreement between the

9 2274 Mirandola et al.: MLIC prototype for QA in particle therapy 2274 FIG. 4. Monoenergetic ionization curves as measured by the Peakfinder (dashed lines) compared against MLIC QUBE data (squares) and MC simulation (circles). Proton and carbon ion data are shown in the upper and lower panels, respectively. [Color figure can be viewed at wileyonlinelibrary.com] TABLE III. Outcomes of the analysis performed on the monoenergetic proton (p) and carbon ion (C) beams, specified by their nominal Bragg peak position in water. Results for MLIC QUBE data (A) and MC simulation (B) about the Bragg peak (BP) position, the full width at 50% and 70% of the maximum (FW50% M, FW70%M), the dose distal fall-off (DDF), the practical range a (R p, 28 ) are shown as differences (D) with respect to the Peakfinder measured values, taken as reference. Mean (l) and standard deviation (r) is reported for each column. In addition, for carbon ion beams, a further analysis has been performed on the fragmentation tail, as explained in Section (2.C). Particle type Nominal BP position [mm] DBP [mm] DFW50%M [mm] DFW70%M [mm] DDDF [mm] DR p [mm] Frag. tail dose relative dev.[%] A B A B A B A B A B A B p p p p C C C C a 0.3 a C a 0.2 a l r Abbreviations as defined in Table I and Section (2.C). a The relatively high-dose level of the measured fragmentation tails prevented the 10% of the maximum dose level to be adopted as indicators for R p. 28 The depth corresponding to the 20% of the maximum dose level was used instead. computed and the experimental data. The limiting factors to a better overall agreement (see Figs. 4 and 5) are given by the absolute value of the evaluated differences and the normalization factor taking into account the number points not matching the exclusion criteria described in subsection (2.C.1). As a single and representative unbiased parameter, the average value for the adjusted MAD has been then computed and found to be %, which is not far from the average statistical uncertainty affecting MC results (from 1.2% to 2.5% at the channel level). The agreement is therefore satisfactory. 3.E. Device activation For the beam intensities delivered during each of the short acquisition, _H ð10þ was found to be <130 nsv/h 50 cm away from the detector and <2.5 lsv/h at the entrance window. At the end of the long acquisition, the detected _H ð10þ was

10 2275 Mirandola et al.: MLIC prototype for QA in particle therapy 2275 FIG. 5. SOBPs as measured by the Peakfinder (black stars with dashed lines for visual guide only), compared against MLIC QUBE data (squares) and MC simulation (circles). Proton and carbon ion data are shown in the upper and lower panels, respectively, and normalized to the entrance point. The centers of the modulation regions are reported in terms of their nominal depth in water. [Color figure can be viewed at wileyonlinelibrary.com] TABLE IV. Outcomes of the analysis performed on the modulated scanning proton (p) and carbon ion (C) beams, specified by the nominal center of the modulation region in water. Results for MLIC QUBE data (A) and MC simulation (B) about the SOBP width, the range equivalent depth a (R, 28 ), the practical range (R p, 28 ), and the dose distal fall-off (DDF) are shown as differences (D) with respect to the Peakfinder measured values, taken as reference. Mean (l) and standard deviation (r) is reported for each column. In addition, a further analysis has been performed point-by-point to quantitatively assess the global agreement, on the whole acquired dose profile, as explained in Section (2.C). Particle type Nominal center of MR in water [mm] Dwidth [mm] DR [mm] DR p [mm] DDDF [mm] Absolute mean local charge percentage dev. [%] A B A B A B A B A B p p p C a 0.2 a C a 0.2 a C a 0.1 a l r Abbreviations as defined in Table I and Section (2.C). a The relatively high-dose level of the measured fragmentation tails prevented the 10% of the dose at the center of the MR to be assessed as indicators for R p. 28 The depth corresponding to the 50% of the dose at the center of the MR was used instead lsv/h at the entrance window and 0.3 lsv/h 50 cm away from the detector. As a method of comparison, the same long acquisition was performed with the Peakfinder and the Giraffe detectors. _H ð10þ values were 10.8 and 27.3 lsv/h at the entrance window, while an _H ð10þ of 0.5 and 0.4 lsv/h was detected 50 cm away for Peakfinder and Giraffe, respectively. For the long acquisition with the QUBE, some measurements were also performed at different times: 25 min after irradiation, _H ð10þ decreased to 3.5 lsv/h (entrance) and 150 nsv/h (50 cm away); after 20 h, _H ð10þ dropped everywhere to the natural background radiation level (~8090 nsv/h in the treatment room). 4. DISCUSSION For a full characterization of the detector, both the physical and gain calibration are mandatory prerequisites: details on these procedures are reported in subsection (2.A.1). For a

11 2276 Mirandola et al.: MLIC prototype for QA in particle therapy 2276 TABLE V. Results of the analysis performed on the pristine and modulated scanning proton (p) and carbon ion (C) beams, specified by the nominal Bragg peak position or the center of the modulation region in water. As explained in Section (2.C.1), the adjusted MAD has been computed. Mean (l) and standard deviation (r) is reported for each column. Beam type Monoenergetic pencil beam (p) Monoenergetic pencil beam (C) mm ripple filters Modulated scanning beam, SOBP (p) Modulated scanning beam, SOBP (C) mm ripple filters Nominal BP position/nominal center of MR in water [mm] Adj. MAD [%] (Eq. (1)) l 3.5 r 1.2 Adj. MAD, adjusted mean absolute deviation (see Eq. (1)), other abbreviations as defined in Table I. longer version of the QUBE and beam energies not high enough to traverse at the same time all the available channels with the entrance plateau region, a two-steps gain calibration procedure could be adopted. The detector, in fact, can be flipped 180, so even those channels not hit by the beam plateau during the first acquisition will be calibrated at the second run. Moreover, the detector is a modular device, so the number of channels can be adapted to the customer s needs. The first part of this study was dedicated to test the shortterm repeatability, linearity with dose and the dose rate dependence of the detector. To this end, several ICs have been acquired with the MLIC QUBE detector prototype and a high repeatability, linearity and dose rate independence have been proved. Once repeatability was assessed, an important part of this work was addressed to increase the coarse spatial resolution of the native detector. Both low-energy proton and carbon ion beams show a very narrow pristine Bragg peak, even less than 1.5 mm width. 9 For these critical conditions, in particular for carbon ions, the commonly used fitting approximation 29 could not perform accurately enough. This entails that increasing the native spatial resolution is highly recommended. Alternative to the more sophisticated method recently presented by, 30 based on a MC-generated lookup function, the procedure adopted in this study is direct and practical, being based on the insertion of thin PMMA sheets, acting as range modifiers, in front of the mylar entrance window of the detector. In principle, the spatial resolution can be increased depending on the number and the thickness of adopted layers. At this stage, each layer was manually inserted, which explains the choice of four experimental configurations as a good compromise among number of acquisitions, acceptable spatial resolution and delivery time. The convolution of the measured curves showed, even for a limited number of PMMA sheets, a spatial resolution around 0.5 mm, enough for the energy checks. Due to the limited size of the prototype, the insertion of thick range shifters (up to 20 cm) in front of the detector was needed to collect the full ICs. Although passive absorbers slightly perturbs the shape of the IC, particularly at the distal end, all the parameters characterizing the IC (both pristine BP and SOBPs), including the amount of dose delivered in the carbon ion fragmentation tails, were found to be in agreement with the reference values and/or within the tolerance levels established in our QA protocol. Furthermore, the full detector geometry and the elemental composition of all its constituents have been implemented within the FLUKA MC code. 31,32 Results about the comparison of the MC predictions and experimental measurements were successfully used as a validation of the detector modeling. The latter will be used for further developments of the QUBE detector. The acquisition time of the MLIC QUBE detector was short compared to the Peakfinder s one: a couple of seconds are enough to obtain each pristine IC with a satisfying signal to noise ratio (~1000). Several minutes are instead required for highly resolved data acquisition with the Peakfinder. The measurements of QUBE activation showed that no specific precaution has to be adopted by the operators when handling the QUBE device after irradiation sessions. However, as a general rule for all devices irradiated with ion beams, the exposure time due to handling should be minimized and the device itself properly stored after usage. In particular, at the working distance, _H ð10þ is slightly higher than the natural background radiation level, even immediately after the irradiation. In the unusual case of a very long acquisition, few minutes are enough to make the _H ð10þ values dropping rapidly to very low activation levels. In the latter case the induced radioactivity is about 50% less for QUBE when compared with a similar detector like the IBA Giraffe. Conversely, slightly lower exposure levels were found with PTW Peakfinder (at least at the entrance window) respect to the QUBE. Nevertheless, the Peakfinder is designed for much longer acquisitions, compared to the QUBE, being the former a scanning single-channel detector type. In addition, QUBE is easy to handle and to place on the treatment couch and few connection cables are needed. The full characterization of the MLIC detector prototype QUBE De.Tec.Tor. for the acquisition of ICs and SOBPs was carried out, for the first time, for both proton and carbon ion scanning beams from a fixed horizontal beam line. The dosimetric characterization here presented can be extended to the full commercial version of the device, already available on the market. Furthermore, this detector could be used for different beam line geometries (oblique or vertical) and for passive delivery technique. 33 Nevertheless, in case of

Monte Carlo simulations of ripple filters designed for proton and carbon ion beams in hadrontherapy with active scanning technique

Monte Carlo simulations of ripple filters designed for proton and carbon ion beams in hadrontherapy with active scanning technique Journal of Physics: Conference Series Monte Carlo simulations of ripple filters designed for proton and carbon ion beams in hadrontherapy with active scanning technique To cite this article: F Bourhaleb

More information

Secondary Particles Produced by Hadron Therapy

Secondary Particles Produced by Hadron Therapy Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad

More information

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.162 pissn 2508-4445, eissn 2508-4453 Secondary Neutron Dose Measurement for Proton Line Scanning

More information

Outline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy

Outline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy Outline Proton Therapy I: Basic Proton Therapy Bijan Arjomandy, Ph.D. Narayan Sahoo, Ph.D. Mark Pankuch, Ph.D. Physics of charge particle motion Particle accelerators Proton interaction with matter Delivery

More information

Estimating proton beam energy spread using Bragg peak measurement

Estimating proton beam energy spread using Bragg peak measurement Estimating proton beam energy spread using Bragg peak measurement V. Anferov 1, V. Derenchuk, R. Moore, A. Schreuder ProNova Solutions LLC, Knoxville, TN. 1. Introduction ProNova is installing and commissioning

More information

Instrumentation for Verification of Dose

Instrumentation for Verification of Dose Instrumentation for Verification of Dose MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 52 Essen, Germany, May 3 5, 2013 Consistent and harmonized dosimetry guidelines

More information

Measurements of secondary particles emitted by 12 C, 4 He and 16 O ion beams in view of innovative dose profiling technique in Particle Therapy

Measurements of secondary particles emitted by 12 C, 4 He and 16 O ion beams in view of innovative dose profiling technique in Particle Therapy Measurements of secondary particles emitted by C, 4 He and 16 O ion beams in view of innovative dose profiling technique in Particle Therapy A. Rucinski b,d, G. Battistoni g, F. Collamati a,b, E. De Lucia

More information

Overview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz

Overview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz Overview and Status of the Austrian Particle Therapy Facility MedAustron Peter Urschütz MedAustron Centre for ion beam therapy and non-clinical research Treatment of 1200 patients/year in full operation

More information

Geant4 studies of the CNAO facility system for hadrontherapy treatment of uveal melanomas

Geant4 studies of the CNAO facility system for hadrontherapy treatment of uveal melanomas th International Conference on Computing in High Energy and Nuclear Physics (CHEP13) IOP Publishing Journal of Physics: Conference Series 13 (1) 8 doi:1.188/17-9/13//8 Geant studies of the CNAO facility

More information

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in. A Monte Carlo Study of the Relationship between the Time Structures of Prompt Gammas and in vivo Radiation Dose in Proton Therapy Wook-Geun Shin and Chul Hee Min* Department of Radiation Convergence Engineering,

More information

Instrumentation for Verification of Dose

Instrumentation for Verification of Dose Instrumentation for Verification of Dose S. Vatnitsky MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 53 Shanghai, China, June 9-11, 2014 Consistent and harmonized dosimetry

More information

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego Physics of Novel Radiation Modalities Particles and Isotopes Todd Pawlicki, Ph.D. UC San Diego Disclosure I have no conflicts of interest to disclose. Learning Objectives Understand the physics of proton

More information

THE mono-energetic hadron beam such as heavy-ions or

THE mono-energetic hadron beam such as heavy-ions or Verification of the Dose Distributions with GEANT4 Simulation for Proton Therapy T.Aso, A.Kimura, S.Tanaka, H.Yoshida, N.Kanematsu, T.Sasaki, T.Akagi Abstract The GEANT4 based simulation of an irradiation

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Optimization of hadron therapy proton beam using Monte Carlo code on GPU

Optimization of hadron therapy proton beam using Monte Carlo code on GPU Dottorato in Fisica degli Acceleratori, XXIX ciclo Optimization of hadron therapy proton beam using Monte Carlo code on GPU Candidata: Martina Senzacqua N matricola: 1163436 Supervisor: Prof. Vincenzo

More information

FLUKA simulations of selected topics regarding proton pencil beam scanning

FLUKA simulations of selected topics regarding proton pencil beam scanning FLUKA simulations of selected topics regarding proton pencil beam scanning C. Bäumer, J. Farr, J. Lambert and B. Mukherjee Westdeutsches Protonentherapiezentrum Essen, Germany T. Mertens, and B. Marchand

More information

8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania

8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania Bijan Arjomandy, Ph.D. Mclaren Proton Therapy Center Mark Pankuch, Ph.D. Cadence Health Proton Center Chia-Ho, Hua St. Jude Children's Research Hospital Kevin Teo The Hospital of the University of Pennsylvania

More information

Laser-Accelerated protons for radiation therapy

Laser-Accelerated protons for radiation therapy Laser-Accelerated protons for radiation therapy E Fourkal, I Velchev,, J Fan, J Li, T Lin, C Ma Fox Chase Cancer Center, Philadelphia, PA Motivation Proton beams provide better conformity to the treatment

More information

Radiation protection issues in proton therapy

Radiation protection issues in proton therapy Protons IMRT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. Proton therapy: An overview 2. Radiation protection issues: Staff 3. Radiation

More information

Measurement of secondary particle production induced by particle therapy ion beams impinging on a PMMA target

Measurement of secondary particle production induced by particle therapy ion beams impinging on a PMMA target EPJ Web of Conferences 117, Measurement of secondary particle production induced by particle therapy ion beams impinging on a PMMA target M. Toppi 1,G.Battistoni 2, F. Bellini 3,4, F. Collamati 3,4, E.

More information

Response characteristics of an imaging plate to clinical proton beams

Response characteristics of an imaging plate to clinical proton beams Nuclear Instruments and Methods in Physics Research A 424 (1999) 569 574 Response characteristics of an imaging plate to clinical proton beams A. Nohtomi *, T. Terunuma, R. Kohno, Y. Takada, Y. Hayakawa,

More information

arxiv: v2 [physics.med-ph] 29 May 2015

arxiv: v2 [physics.med-ph] 29 May 2015 The Proton Therapy Nozzles at Samsung Medical Center: A Monte Carlo Simulation Study using TOPAS Kwangzoo Chung, Jinsung Kim, Dae-Hyun Kim, Sunghwan Ahn, and Youngyih Han Department of Radiation Oncology,

More information

Monte Carlo Simulation concerning Particle Therapy

Monte Carlo Simulation concerning Particle Therapy Monte Carlo Simulation concerning Particle Therapy Masaaki Takashina Graduate School of Medicine, Osaka University, Osaka 565-0871, Japan INTRODUCTION It is well known that the particle therapy has some

More information

Chapter 10 Acceptance Tests and Commissioning Measurements

Chapter 10 Acceptance Tests and Commissioning Measurements Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:

More information

Il picco di Bragg. G. Battistoni INFN Milano. 08/06/2015 G. Battistoni

Il picco di Bragg. G. Battistoni INFN Milano. 08/06/2015 G. Battistoni Il picco di Bragg G. Battistoni INFN Milano 08/06/015 G. Battistoni 1 Φ(z) The physics of Bragg Peak 180 MeV proton in water Longitudinal profile: Transversel profile: Φ(z,x) dominated by interaction with

More information

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Reference Dosimetry for Megavoltage Therapy Beams: Electrons Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy

More information

PHYSICAL AND DOSIMETRIC BEAM CHARACTERISATION PROTOCOL AT CENTRO NAZIONALE DI ADROTERAPIA ONCOLOGICA (CNAO)

PHYSICAL AND DOSIMETRIC BEAM CHARACTERISATION PROTOCOL AT CENTRO NAZIONALE DI ADROTERAPIA ONCOLOGICA (CNAO) Project co-funded by the European Commission within the FP7 (2007 2013) ULICE Union of Light Centres in Europe Page 1 of 27 PHYSICAL AND DOSIMETRIC BEAM CHARACTERISATION PROTOCOL AT CENTRO NAZIONALE DI

More information

Measurement ofdepth-dose distribution ofprotons by an imaging plate

Measurement ofdepth-dose distribution ofprotons by an imaging plate Nuclear Instruments and Methods in Physics Research A 511 (2003) 382 387 Measurement ofdepth-dose distribution ofprotons by an imaging plate A. Nohtomi a, *, T. Sakae a, T. Terunuma a, Y. Tsunashima a,

More information

Detectors in Nuclear Physics: Monte Carlo Methods. Dr. Andrea Mairani. Lectures I-II

Detectors in Nuclear Physics: Monte Carlo Methods. Dr. Andrea Mairani. Lectures I-II Detectors in Nuclear Physics: Monte Carlo Methods Dr. Andrea Mairani Lectures I-II INTRODUCTION Sampling from a probability distribution Sampling from a probability distribution X λ Sampling from a probability

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

A study of monitoring performances with the INSIDE system. Francesco Pennazio on behalf of the INSIDE collaboration

A study of monitoring performances with the INSIDE system. Francesco Pennazio on behalf of the INSIDE collaboration A study of monitoring performances with the INSIDE system Francesco Pennazio on behalf of the INSIDE collaboration INSIDE Hadrontherapy monitoring system TRACKER PET heads Courtesy of G. Giraudo 2 INSIDE

More information

A prompt gamma camera for real-time range control in Proton Therapy

A prompt gamma camera for real-time range control in Proton Therapy POLITECNICO DI MILANO Dipartimento di Elettronica e Informazione DOTTORATO DI RICERCA IN INGEGNERIA DELL INFORMAZIONE A prompt gamma camera for real-time range control in Proton Therapy Doctoral Dissertation

More information

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA GROUP I SPRING 2014 KEVIN JORDAN GRADUATE STUDENT, RADIOLOGICAL PHYSICS KARMANOS CANCER CENTER WAYNE STATE UNIVERSITY

More information

MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY *

MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY * Romanian Reports in Physics, Vol. 66, No. 1, P. 148 156, 2014 MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY * DARIUSH SARDARI, EHSAN SALIMI Department of Medical Radiation

More information

Initial Studies in Proton Computed Tomography

Initial Studies in Proton Computed Tomography SCIPP Initial Studies in Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz,

More information

Ion- and proton-beams: Experience with Monte Carlo Simulation

Ion- and proton-beams: Experience with Monte Carlo Simulation Ion- and proton-beams: Experience with Monte Carlo Simulation Katia Parodi, Ph.D. Heidelberg Ion Therapy Centre, Heidelberg, Germany (Previously: Massachusetts General Hospital, Boston, USA) Workshop on

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

PHITS calculation of the radiation field in HIMAC BIO

PHITS calculation of the radiation field in HIMAC BIO PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,

More information

Biological Dose Calculations for Particle Therapy in FLUKA

Biological Dose Calculations for Particle Therapy in FLUKA U N I V E R S I T Y O F B E R G E N Department of Physics and Technology Biological Dose Calculations for Particle Therapy in FLUKA Tordis J. Dahle May 2016 Introduction About half of all cancer patients

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

The EDM Polarimeter Development at COSY-Jülich

The EDM Polarimeter Development at COSY-Jülich Noname manuscript No. (will be inserted by the editor) The EDM Polarimeter Development at COSY-Jülich Fabian Müller for the JEDI Collaboration Received: date / Accepted: date Abstract The JEDI (Jülich

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

A comparison of methods for monitoring photon beam energy constancy

A comparison of methods for monitoring photon beam energy constancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

Beam Optics for a Scanned Proton Beam at Loma Linda University Medical Center

Beam Optics for a Scanned Proton Beam at Loma Linda University Medical Center Beam Optics for a Scanned Proton Beam at Loma Linda University Medical Center George Coutrakon, Jeff Hubbard, Peter Koss, Ed Sanders, Mona Panchal Loma Linda University Medical Center 11234 Anderson Street

More information

Towards efficient and accurate particle transport simulation in medical applications

Towards efficient and accurate particle transport simulation in medical applications Towards efficient and accurate particle transport simulation in medical applications L. Grzanka1,2, M. Kłodowska1, N. Mojżeszek1, N. Bassler3 1 Cyclotron Centre Bronowice, Institute of Nuclear Physics

More information

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

David Gascón. Daniel Peralta. Universitat de Barcelona, ECM department. E Diagonal 647 (Barcelona) IFAE, Universitat Aut onoma de Barcelona

David Gascón. Daniel Peralta. Universitat de Barcelona, ECM department. E Diagonal 647 (Barcelona) IFAE, Universitat Aut onoma de Barcelona LHCb 2000-32, CALO 9 June 2000 Results of a tagged photon test beam for the Scintillator Pad Detector. Llu s Garrido David Gascón Ramon Miquel Daniel Peralta Universitat de Barcelona, ECM department E-08028

More information

The heavy ion irradiation facility at KVI-CART

The heavy ion irradiation facility at KVI-CART The heavy ion irradiation facility at KVI-CART Brian N. Jones 1, Marc-Jan van Goethem 1,2, Rob Kremers 1, Harry Kiewiet 1, Emiel van der Graaf 1, Sytze Brandenburg 1 1 University of Groningen, KVI-Center

More information

Comparison of two commercial detector arrays for IMRT quality assurance

Comparison of two commercial detector arrays for IMRT quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 2, SPRING 2009 Comparison of two commercial detector arrays for IMRT quality assurance Jonathan G. Li, 1,a Guanghua Yan, 1,2 and Chihray Liu

More information

Equalisation of the PMT response to charge particles for the Lucid detector of the ATLAS experiment

Equalisation of the PMT response to charge particles for the Lucid detector of the ATLAS experiment Equalisation of the PMT response to charge particles for the Lucid detector of the ATLAS experiment Camilla Vittori Department of Physics, University of Bologna, Italy Summer Student Program 2014 Supervisor

More information

Validity of the Analysis of Radiochromic Film Using Matlab Code

Validity of the Analysis of Radiochromic Film Using Matlab Code Validity of the Analysis of Radiochromic Film Using Matlab Code Contact Samuel.millington@stfc.ac.uk S.J.Millington J.S.Green D.C.Carroll Introduction Radiochromic film is typically used in experiments

More information

Proton radiography with a range telescope and its use in proton therapy

Proton radiography with a range telescope and its use in proton therapy Proton radiography with a range telescope and its use in proton therapy Juan L. Romero Department of Physics, University of California, Davis, CA Invited talk at Massachusetts General Hospital, Boston,

More information

GEANT4 simulation of the testbeam set-up for the ALFA detector

GEANT4 simulation of the testbeam set-up for the ALFA detector GEANT4 simulation of the testbeam set-up for the detector V. Vorobel a and H. Stenzel b a Faculty of Mathematics and Physics, Charles University in Prague, Czech Republic b II. Physikalisches Institut,

More information

Development of beam delivery systems for proton (ion) therapy

Development of beam delivery systems for proton (ion) therapy 7th 28th of July 213, JINR Dubna, Russia Development of beam delivery systems for proton (ion) therapy S t u d e n t : J o z e f B o k o r S u p e r v i s o r : D r. A l e x a n d e r M o l o k a n o v

More information

Hadron Therapy Medical Applications

Hadron Therapy Medical Applications Hadron Therapy Medical Applications G.A. Pablo Cirrone On behalf of the CATANA GEANT4 Collaboration Qualified Medical Physicist and PhD Student University of Catania and Laboratori Nazionali del Sud -

More information

Accelerators for Hadrontherapy -- Present & Future --

Accelerators for Hadrontherapy -- Present & Future -- IVICFA Institut Valencià d Investigació Cooperativa en Física Avançada Miniworkshop on Medical Physics Accelerators for Hadrontherapy -- Present & Future -- Silvia Verdú-Andrés TERA / IFIC (CSIC-UV) Valencia,

More information

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),

More information

New irradiation zones at the CERN-PS

New irradiation zones at the CERN-PS Nuclear Instruments and Methods in Physics Research A 426 (1999) 72 77 New irradiation zones at the CERN-PS M. Glaser, L. Durieu, F. Lemeilleur *, M. Tavlet, C. Leroy, P. Roy ROSE/RD48 Collaboration CERN,

More information

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.

More information

Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy

Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2012 Monte Carlo study of the potential reduction in out-of-field dose

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance Michael P. Barnes 1,2 a and Peter B. Greer 1,3

More information

Muon reconstruction performance in ATLAS at Run-2

Muon reconstruction performance in ATLAS at Run-2 2 Muon reconstruction performance in ATLAS at Run-2 Hannah Herde on behalf of the ATLAS Collaboration Brandeis University (US) E-mail: hannah.herde@cern.ch ATL-PHYS-PROC-205-2 5 October 205 The ATLAS muon

More information

Characterization of heavy charged particle fields using fluorescent nuclear track detectors

Characterization of heavy charged particle fields using fluorescent nuclear track detectors PTCOG 53, Shanghai June 12, 2014 Characterization of heavy charged particle fields using fluorescent nuclear track detectors Grischa M. Klimpki 1, P. Incardona 2, H. Mescher 1, T. Pfeiler 1, M.S. Akselrod

More information

Absorption and Backscattering of β-rays

Absorption and Backscattering of β-rays Experiment #54 Absorption and Backscattering of β-rays References 1. B. Brown, Experimental Nucleonics 2. I. Kaplan, Nuclear Physics 3. E. Segre, Experimental Nuclear Physics 4. R.D. Evans, The Atomic

More information

Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams

Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams Severine Rossomme a) Center of Molecular Imaging, Radiotherapy and Oncology, Institut de Recherche Experimentale et Clinique,

More information

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard

More information

Status of the Siemens IONTRIS Particle Therapy Systems: Marburg, Kiel and Shanghai

Status of the Siemens IONTRIS Particle Therapy Systems: Marburg, Kiel and Shanghai IONTRIS Particle Therapy Systems Status of the Siemens IONTRIS Particle Therapy Systems: Marburg, Kiel and Shanghai Peter Urschütz SIEMENS Healthcare Particle Therapy Siemens AG 2011. All rights reserved.

More information

Outline. Indrin J. Chetty, AAPM 2006 Monte Carlo CE course. Indrin J. Chetty Henry Ford Hospital. David W. O. Rogers Carleton University

Outline. Indrin J. Chetty, AAPM 2006 Monte Carlo CE course. Indrin J. Chetty Henry Ford Hospital. David W. O. Rogers Carleton University AAPM Task Group Report No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning Indrin J. Chetty Henry Ford Hospital David W. O.

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

1 Introduction. A Monte Carlo study

1 Introduction. A Monte Carlo study Current Directions in Biomedical Engineering 2017; 3(2): 281 285 Sebastian Richter*, Stefan Pojtinger, David Mönnich, Oliver S. Dohm, and Daniela Thorwarth Influence of a transverse magnetic field on the

More information

The volume effect of detectors in the dosimetry of small fields used in IMRT

The volume effect of detectors in the dosimetry of small fields used in IMRT The volume effect of detectors in the dosimetry of small fields used in IMRT Wolfram U. Laub a) Department of Medical Physics, University of Tübingen, Germany Tony Wong b) Department of Radiation Oncology,

More information

Physics of Particle Beams. Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School

Physics of Particle Beams. Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School Physics of Particle Beams Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School PTCOG 53 Education Session, Shanghai, 2014 Dose External

More information

Nucleus-Nucleus Interaction Modelling and Applications in Ion Therapy Treatment Planning

Nucleus-Nucleus Interaction Modelling and Applications in Ion Therapy Treatment Planning UNIVERSITÀ DEGLI STUDI DI PAVIA DOTTORATO DI RICERCA IN FISICA XX CICLO Nucleus-Nucleus Interaction Modelling and Applications in Ion Therapy Treatment Planning Andrea Mairani Tesi per il conseguimento

More information

Dose detectors, sensors, and their applications

Dose detectors, sensors, and their applications Dose detectors, sensors, and their applications Simona Giordanengo a) Istituto Nazionale di Fisica Nucleare, Section of Torino, Via Giuria 1, 10125 Torino, Italy Hugo Palmans National Physical Laboratory,

More information

City University of Hong Kong

City University of Hong Kong City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and

More information

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16530 Quantitative Assessment of Scattering

More information

Simulations in Radiation Therapy

Simulations in Radiation Therapy Simulations in Radiation Therapy D. Sarrut Directeur de recherche CNRS Université de Lyon, France CREATIS-CNRS ; IPNL-CNRS ; Centre Léon Bérard Numerical simulations and random processes treat cancer 2

More information

HEATHER. HElium ion Acceleration for radiotherapy. Jordan Taylor, Rob Edgecock University of Huddersfield Carol Johnstone, Fermilab

HEATHER. HElium ion Acceleration for radiotherapy. Jordan Taylor, Rob Edgecock University of Huddersfield Carol Johnstone, Fermilab HEATHER HElium ion Acceleration for radiotherapy Jordan Taylor, Rob Edgecock University of Huddersfield Carol Johnstone, Fermilab PPRIG workshop 1 st -2 nd Dec 2016 Scope Current particle therapy situation

More information

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES *

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES * Romanian Reports in Physics, Vol. 66, No. 1, P. 142 147, 2014 COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES * D. SARDARI, M. HAMEDINEJAD Islamic Azad University,

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY WHITEPAPER ENGLISH PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY PARTICLE MEASUREMENT Particle measurement in cleanrooms. WP1508006-0100-EN, V1R0, 2015-08 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

Control of the fabrication process for the sensors of the CMS Silicon Strip Tracker. Anna Macchiolo. CMS Collaboration

Control of the fabrication process for the sensors of the CMS Silicon Strip Tracker. Anna Macchiolo. CMS Collaboration Control of the fabrication process for the sensors of the CMS Silicon Strip Tracker Anna Macchiolo Universita di Firenze- INFN Firenze on behalf of the CMS Collaboration 6 th International Conference on

More information

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up G. Arico1,2, B. Hartmann2, J. Jakubek3, S. Pospisil3, O. Jäkel1,2,4, M. Martisikova1,2 1 University Clinic

More information

TERA CONTRIBUTIONS TO PARTNER

TERA CONTRIBUTIONS TO PARTNER TERA CONTRIBUTIONS TO PARTNER Ugo Amaldi University of Milano Bicocca and TERA Foundation 1 CNAO status 2 The CNAO Foundation builds with INFN in Pavia the Centre designed by TERA on the basis of PIMMS.

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

Academy of Sciences, Moscow , Russia

Academy of Sciences, Moscow , Russia Analytical expressions for water-to-air stopping-power ratios relevant for accurate dosimetry in particle therapy arxiv:1010.5356v3 [physics.med-ph] 22 Dec 2010 Abstract Armin Lühr 1,2, David C. Hansen

More information

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans The effect of dose calculation uncertainty on the evaluation of radiotherapy plans P. J. Keall a) and J. V. Siebers Department of Radiation Oncology, Medical College of Virginia Hospitals, Virginia Commonwealth

More information

High Energy Electron Radiation Exposure Facility at PSI

High Energy Electron Radiation Exposure Facility at PSI Journal of Applied Mathematics and Physics, 2014, 2, 910-917 Published Online August 2014 in SciRes. http://www.scirp.org/journal/jamp http://dx.doi.org/10.4236/jamp.2014.29103 High Energy Electron Radiation

More information

Spencer-Attix Cavity Theory

Spencer-Attix Cavity Theory Institutionen för icin och hälsa Avdelningen för radiologiska vetenskaper Medicinsk radiofysik Hälsouniversitetet Spencer-Attix Cavity heory Gudrun Alm Carlsson epartment of Medical and Health Science

More information

Electron therapy Class 2: Review questions

Electron therapy Class 2: Review questions Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a

More information

Chapter 8 Electron Beams: Physical and Clinical Aspects

Chapter 8 Electron Beams: Physical and Clinical Aspects 1 Chapter 8 Electron Beams: Physical and Clinical Aspects This set of 91 slides is based on Chapter 8 authored by W. Strydom, W. Parker, and M. Olivares of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

assuming continuous slowing down approximation , full width at half maximum (FWHM), W 80 20

assuming continuous slowing down approximation , full width at half maximum (FWHM), W 80 20 Special Edition 408 Depth Dose Characteristics of Proton Beams within Therapeutic Energy Range Using the Particle Therapy Simulation Framework (PTSim) Monte Carlo Technique Siou Yin Cai 1, Tsi Chain Chao

More information

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.83-87 (2) ARTICLE Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Toshioh FUJIBUCHI,2,*, Satoshi

More information

Optimization of microionization chambers for small-field reference dosimetry

Optimization of microionization chambers for small-field reference dosimetry Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall

More information