PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

Size: px
Start display at page:

Download "PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY"

Transcription

1 WHITEPAPER ENGLISH PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY PARTICLE MEASUREMENT Particle measurement in cleanrooms. WP EN, V1R0,

2 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY WHITEPAPER Content Content... 2 History Particle measurement in clean room technology What do particle counters measure actually? How do particle counters work? Can I use any particle counter with any monitoring system? Where to place the sampling probe? With or without integrated vacuum pump? Setting the measurement time What else should be taken into account?...6 History Document number: wp en Version / Revision: V1R0, Status: Initial version, released History Version Date Comment / Note V1R Initial Version Table 1: History PAGE 2 OF 6 WP EN, V1R0,

3 WHITEPAPER PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY 1 Particle measurement in clean room technology The central task of cleanroom technology is the protection of products against particulate contamination. While there are millions of tiny particles in normal rooms, which are for the most part below the threshold of perception, the ISO defines various cleanroom classes by permissible limits on the number of particles in size from 0.1μm to 5,0μm. The table below shows the maximum admissible number of particles per cubic meter of air for each cleanroom class. Cleanroom classes according to ISO (particles per m³) Class 0,1µm 0,2µm 0,3µm 0,5µm 1,0m 5,0µm ISO ISO ISO ISO ISO ISO ISO ISO ISO Table 2: Cleanroom classes according to ISO What do particle counters measure actually? In order to capture such small particles with reasonable reliability in real-time, we need a powerful measurement principle which can detect both the size of the individual particles and the number within a determined volume of air. While counting particles generally causes fewer problems, as long as the quantity is not too large, the correct classification of the particle size is actually only possible with idealized spherical particles. Since in practice, particles can, however, take any shape, a unique assignment to a certain size class is de facto impossible. For this type of particle, we therefore use the term "equivalent diameter", which states that the optical properties correspond to a spherical particle having the same diameter. 1.2 How do particle counters work? The measurement procedure practically all particle counters used in clean room technology are based on is the optical scattering of light NIOTRONIC GMBH WP EN, V1R0, PAGE 3 OF 6

4 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY WHITEPAPER In this method of measurement the light reflected by the particles in a stream of air that is passed through a tightly shaped laser beam inside the measuring cell is used to both count the particles and measure their size. When the particles carried by the airflow pass the laser beam, they cause extremely short flashes of light which are focused by corresponding optics to a single point and are detected by a sensitive photodetector. The intensity of the light flash allows conclusions about the particle size, although this is not always unambiguously possible due to the non-monotonicity of the relationship between particle size and scattered light intensity in some size ranges (Mie theory). In addition, it cannot always be guaranteed that all the particles pass through the "active volume" strictly sequentially, i.e. at higher particle concentrations, a certain amount of so-called coincidences is also to be expected - in other words, when two or more particles are simultaneously in the active volume, the particle counter cannot distinguish whether this is one larger, or several smaller particles. Due to the problems described, the count and size classification of real occurring particles is not comparable with the precision of a temperature measurement. Only a small part of the air in the room can be measured, due to the low flow rate (usually 28,3l / min), and neither a statement about the total particle number in the room is made, nor information is obtained about the current particle concentration at the other end of the room. Nevertheless, scattered light particle counters have been established as a reliable measuring device for local recording and detection of particle contamination in clean room technology. 1.3 Can I use any particle counter with any monitoring system? Nowadays, laser optical particle counters are available from different manufacturers with different properties in respect to their counting efficiency, number of evaluable size classes, as well as different equipment and interfaces. Often, the manufacturer also provides associated data acquisition and recording software, which is usually tailored to their particle counter. This allows most of the parameterization of the particle counter and offers various automation functions such as starting and stopping the measurement, etc., which are not realized in connecting the particle counter via analog interfaces (4. 20mA etc.). There are only a few manufacturers of particle counters, which keep the data transfer protocol publicly available to enable third-party monitoring systems to connect to the particle counter. These solutions, however, are to be treated with caution, as third-party suppliers of monitoring systems often do not have all the information necessary to implement the communication protocol fully and correctly. For this reason it is advisable to use the monitoring software of the particle counter manufacturer, since in case of problems it can be ensured that the necessary expertise is available to resolve them. In general, a connection via analog interfaces is discouraged because the measuring cycles of the particle counter run asynchronously to the scan cycle of the monitoring system, i.e. there may be a delay between the occurrence of particulate contamination and the display and alarm in the monitoring system. 1.4 Where to place the sampling probe? In the interest of the shortest possible reaction time, the sampling probe of the particle counter should be placed as directly as possible at the location of critical processes, e.g. in filling, bearing in mind that the length of tubing between the sample probe and the particle counter should also be kept as short as possible. Experiments have shown that deposition of large particles can occur at hose lengths of less than 5 m. Depending on the location of the particle counter with respect to the PAGE 4 OF 6 WP EN, V1R0,

5 WHITEPAPER PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY sampling probe, a more or less large height difference has to be overcome. Particularly large particles are not carried by the air stream through large differences in height (e.g. a vertically rising tubing), but instead follow the gravitational field of the earth, they fall down and therefore never arrive in the measuring cell of the particle counter and, consequently, cannot be measured. Similar problems arise with sagging, horizontally guided hoses. This may in turn lead to a deposition of large particles, as these tend to stick to the lowest point of the hose line. This effect is not as pronounced as the one with vertically guided hoses, however, particles adhering to the tube's inner wall can be removed again at a later time, which may then lead to a false reading or an unwanted alarm. Considering this requirement, it is ideal to plan for the construction of a utility tunnel, as this allows the installation of the particle counter directly on the rear side of the respective cleanroom wall, which facilitates easy access for maintenance and calibration purposes and also reduces hose lengths to a minimum. Unfortunately, this is rarely possible in practice for structural reasons. 1.5 With or without integrated vacuum pump? Moreover, it is necessary to decide whether particle counters are to be used with an integrated vacuum pump, or just a pure counter, which needs an external central vacuum system to operate, usually dimensioned for several counters. The former is offered for very small systems with few or only one particle counter, as this will be significantly cheaper due to the significantly reduced installation effort and the omission of the central vacuum system. Also, existing clean room facilities can usually be upgraded with particle counters with integrated vacuum pumps, since a power supply and a network connection to connect to the monitoring system will be available almost everywhere. The only downside is the increased noise due to the built-in vacuum pump. For larger clean room facilities with more than 2-3 particle counters, cost advantages are significant for counters without vacuum supply in conjunction with an external, usually a central vacuum system accommodated in a separate technical area, despite the increased installation expenses. This option is also much more pleasant for the staff present in the clean room, since the particle counters in-use do not generate any noise. 1.6 Setting the measurement time The particle limits defined in ISO relate to one cubic meter of room air. The typical flow rates of common particle counters, however, amount to 28.3, 50 or 100 l/min. In order to be able to represent the measured value in particles / m³, the measurement result of the particle counter must be converted. The latter in turn depends on the set measuring time. In order to measure one cubic meter with a particle counter whose nominal flow measures 28,3l / min, the measuring time should be set to 35,318min, which in practice, however, would be completely unacceptable, since the first measurement result would be available only after the expiration of 35,318min. A temporal relation to events in the clean room is almost impossible in this case. The situation is not much better for counters with a higher nominal flow rate either. In order to shorten the response time, it is therefore strived for to set the measurement time as short as possible, and to convert the result in particles per cubic meter. In order to convert the measurement result of the particle counter in particles per cubic meter at a measurement time of one minute, it must be multiplied by , since times 28.3l (= 1cf) gives exactly one cubic meter. With increasing shortening of the measurement time and increasing of the conversion factor, however, a problem of a statistical nature appears, as in shorter 2016 NIOTRONIC GMBH WP EN, V1R0, PAGE 5 OF 6

6 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY WHITEPAPER measurement intervals with the same particle concentration, fewer particles per measuring cycle are counted, thus increasing the statistical uncertainty. In addition, multiplication with a large conversion factor means that when particles are measured, their numbers are vastly exaggerated. At too short measurement times, a single particle may in extreme cases cause a threshold exceedance, although far fewer particles are present in the cubic meters that is to be measured. For this reason, the reduction of the measurement time of the particle counter is severely limited. One way out of this situation is the accumulation of measured values in conjunction with a short measurement time. The last N measured results of the particle counter are always stored and added. Thus, a measurement result can be obtained related to one cubic meter in spite of a short measurement time and without a conversion factor. This process is only supported by a few particle counters on the market. 1.7 What else should be taken into account? Since all measuring devices must be recurrently maintained and calibrated, it is of considerable importance that the manufacturer of the particle counter or its partners should be able to offer support in this regard quickly and easily. For smaller installations with few particle counters, shipping for calibration is an option, while for larger systems, onsite calibration can also be performed. Another useful feature is a replacement service that guarantees always fully functional and currently calibrated particle counters. In this case, the manufacturer sends up-to-date calibrated replacement equipment and in return receives the particle counters that are to be calibrated. Often this presents itself as an excellent solution because it de facto does not come to any downtime of the plant due to the annual calibration. Niotronic Hard- & Software GmbH office@niotronic.com, PAGE 6 OF 6 WP EN, V1R0,

Fundamentals of Particle Counting

Fundamentals of Particle Counting Fundamentals of Particle Counting 1 Particle Counting: Remains the most significant technique for determining the cleanliness level of a fluid Useful as a tool for qualification and monitoring cleanroom

More information

Determination of particle size distribution Single particle light interaction methods. Part 4:

Determination of particle size distribution Single particle light interaction methods. Part 4: Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne

More information

ISO STANDARD INTERNATIONAL. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness

ISO STANDARD INTERNATIONAL. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness INTERNATIONAL ISO STANDARD 4644- First edition 999-05-0 Cleanrooms and associated controlled environments Part : Classification of air cleanliness Salles propres et environnements maîtrisés apparentés

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 21501-2 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 2: Light scattering liquid-borne particle counter

More information

Cleanrooms and associated controlled environments. Part 1: Classification of air cleanliness by particle concentration

Cleanrooms and associated controlled environments. Part 1: Classification of air cleanliness by particle concentration INTERNATIONAL STANDARD ISO 14644-1 Second edition 2015-12-15 Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness by particle concentration Salles propres et environnements

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Mercury Stack Monitor SM-4

Mercury Stack Monitor SM-4 Member of the envea Group Mercury Stack Monitor SM-4 Automatic continuous emission monitor (CEM) for mercury Continuous operation Sample dilution directly at stack - works with virtually any sample matrix

More information

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date ATS-SOI-3660 Page: 1 of 6 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013

More information

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids October 26 Ottawa, Canada Summary The introduction of a growing number of targeted protein-based drug formulations

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials Thermo Scientific ICP-MS solutions for the semiconductor industry Maximize wafer yields with ultralow elemental detection in chemicals and materials Trace impurity analysis in chemicals an used in the

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 21501-4 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne particle counter for

More information

Certificate: / 9 September 2014

Certificate: / 9 September 2014 Test report: 936/21219977/A of 26 March 214 Initial certification: 5 August 214 Date of expiry: 4 August 219 Publication: BAnz AT 5 August 214 B11, chapter III, no. 2.1 Approved application The certified

More information

CONTINUOUS FINE DUST MEASURING SYSTEM

CONTINUOUS FINE DUST MEASURING SYSTEM CONTINUOUS FINE DUST MEASURING SYSTEM PM 10 - PM 2,5 - PM 1 Particle count distribution DustMonit ENGINEERING SRL Via Paolo Diacono, 1-20133 Milano (MI) - ITALIA Tel ++39.02.26684830 - E-mail : info@conteng.it

More information

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS APPLICATION NOTE CC-119 (A4) This application note works through the As Found Calibration Certificate of the AeroTrak Counters

More information

PARTICLE SIZE ANALYTICAL RANGES AND APPLICABILITY. m mm (10-6 m) nm (10-9 m)

PARTICLE SIZE ANALYTICAL RANGES AND APPLICABILITY. m mm (10-6 m) nm (10-9 m) P A R T I C L E S I Z E A N A L Y S I S PARTICLE SIZE Accurately determining particle size has become essential in many industries, as it is a fundamental physical characteristic that must be selected,

More information

Dewpoint Meter Data Sheet / Instructions

Dewpoint Meter Data Sheet / Instructions Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment

More information

Data Sheet / Instructions. Dewpoint MeterO

Data Sheet / Instructions. Dewpoint MeterO Data Sheet / Instructions Dewpoint MeterO Dewpoint Meter ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment of surface cleanliness.

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-1 NATIONAL SECURITY COMPLEX J. A. Mullens, J. K. Mattingly, L. G. Chiang, R. B. Oberer, J. T. Mihalczo ABSTRACT This paper describes a template matching

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

Standard Practices for Air Speed Calibration Testing

Standard Practices for Air Speed Calibration Testing Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument

More information

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3 Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3 1 CANBERRA Industries, Meriden, CT 06450 2 Los Alamos National Laboratory,

More information

Latest Technology and Standardization Trends for Liquid-borne Particle Counters

Latest Technology and Standardization Trends for Liquid-borne Particle Counters Latest Technology and Standardization Trends for Liquid-borne Particle Counters Kaoru Kondo Rion Co., Ltd. (Original paper is K.Kondo, Bunseki,Vol.9, p499 (2012)) Abstract Fine particles contaminating

More information

Review of Anemometer Calibration Standards

Review of Anemometer Calibration Standards Review of Anemometer Calibration Standards Rachael V. Coquilla rvcoquilla@otechwind.com Otech Engineering, Inc., Davis, CA Anemometer calibration defines a relationship between the measured signals from

More information

FRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement. 1 The Inventor of "Reverse Fourier Optics"

FRITSCH analysette 22 NanoTec The Result of 30 Years of Experience in Particle Size Measurement. 1 The Inventor of Reverse Fourier Optics FRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement Wolfgang Mutter FRITSCH GmbH Manufacturers of Laboratory Instruments Industriestrasse 8 55743 Idar-Oberstein

More information

Performance Characteristics of Deterministic Leak Detection Systems

Performance Characteristics of Deterministic Leak Detection Systems Performance Characteristics of Deterministic Leak Detection Systems Michael Kasch 3S Consult München GmbH Germany Abstract The aim of leak detection and leak location is the limitation of losses and possible

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

Surveying Prof. Bharat Lohani Department of Civil Engineering Indian Institute of Technology, Kanpur. Module - 3 Lecture - 4 Linear Measurements

Surveying Prof. Bharat Lohani Department of Civil Engineering Indian Institute of Technology, Kanpur. Module - 3 Lecture - 4 Linear Measurements Surveying Prof. Bharat Lohani Department of Civil Engineering Indian Institute of Technology, Kanpur Module - 3 Lecture - 4 Linear Measurements Welcome again to this another video lecture on basic surveying.

More information

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY PURPOSE This procedure provides instructions for developing, maintaining, and documenting, radiation safety procedures conducted at the Cyclotron Radiochemistry

More information

Controlled Environment Testing Report

Controlled Environment Testing Report Client: Controlled Environment Testing Report Services Conducted: Product Testing Project Contact: Stefan Hocom Project Manager: Stefan E. Hocom 323 Spreckels Drive, Ste. B Aptos, CA 95003 The final approval

More information

Accurate Measurement of Transmittance and Reflectance for Engineering Applications

Accurate Measurement of Transmittance and Reflectance for Engineering Applications Accurate Measurement of Transmittance and Reflectance for Engineering Applications Dr. Chen Fangzhi Laboratory Manager Façade & Roof Materials Testing Laboratory OTM Solutions Pte Ltd PerkinElmer INTour

More information

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219 REX Evaluation Guide American Micro Detection Systems Inc. 2800 March Lane, Suite 200 Stockton, CA 95219 I. INTRODUCTION REX (Real-time Elemental X-ray Fluorescence System) is the only instrument capable

More information

Laser induced release of encapsulated noble gas in SCHOTT receiver

Laser induced release of encapsulated noble gas in SCHOTT receiver Available online at www.sciencedirect.com Energy Procedia 00 (2013) 000 000 www.elsevier.com/locate/procedia SolarPACES 2013 Laser induced release of encapsulated noble gas in SCHOTT receiver O. Sohr a,

More information

Photoluminescence Spectrometer (FLS980)

Photoluminescence Spectrometer (FLS980) Photoluminescence Spectrometer (FLS980) Instrument Summary: The Edinburgh Instruments FLS980 photoluminescence spectrometer offers both steady state and time resolved (lifetime) fluorescence spectroscopy

More information

Direct reading aerosol monitors

Direct reading aerosol monitors Health and Safety Executive Direct reading aerosol monitors 27 th June 2018 Andrew Thorpe Introduction Direct reading Aerosol Monitors (DRAMs) are often used as a complementary approach to conventional

More information

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement :

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement : Application Report #001 Measuring Lysozyme Monomer at 0.1 mg/ml Concentration Equipment used : ALV-NIBS / HPPS High Sensitivity Version, Lysozyme (MERCK), 0.1 molar Sodium-Acetate buffer (ph 4.25), syringe

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

ENVR 416 Aerosol Technology - Laboratory Session Fall 2007

ENVR 416 Aerosol Technology - Laboratory Session Fall 2007 ENVR 416 Aerosol Technology - Laboratory Session Fall 2007 CALIBRATION AND USE OF THE OPTICAL MICROSCOPE The objective of this lab is for you to become familiar with use of the optical microscope to measure

More information

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification Technical Paper 3 Verifying Particle Counter Calibration: Size Verification By D. John Hunt, Pacific Scientific Instruments. Originally printed in Pollution Equipment News Publication, June 1996. Volume

More information

Measurement & Analytics Measurement made easy. MB3600-CH70 FT-NIR polyol analyzer Pre-calibrated for OH value determination

Measurement & Analytics Measurement made easy. MB3600-CH70 FT-NIR polyol analyzer Pre-calibrated for OH value determination Measurement & Analytics Measurement made easy MB3600-CH70 FT-NIR polyol analyzer Pre-calibrated for OH value determination Adapted for polyols and polyurethanes applications The MB3600-CH70 Polyol Analyzer

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable

More information

Certificate: / 29 April 2014

Certificate: / 29 April 2014 Test report: 936/21218896/A of 20 September 2013 Initial certification: 01 April 2014 Date of expiry: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter IV, No. 5.1 Approved application The

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator. Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator Hiromu Sakurai hiromu.sakurai@aist.go.jp Particle Measurement Research Group National

More information

Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports?

Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports? Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports? A. Geach 1 1 CINRG Systems Inc., Burlington, Ontario, Canada Abstract In 2015 NIST released

More information

Mining Analyzer XRF-3360

Mining Analyzer XRF-3360 Turnkey Laboratories Solutions Mining Analyzer XRF-3360 Mining Analyzer Intrument is a professional manufacturer of high-performance X-ray Fluorescence Spectrometer. The high-performance desktop XRF geology

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory Bülent ÖZGÜR 1,*, Okhan GANİOĞLU 1, Nasser Al-Qahtani 2, Faisal Al-Qahtani 2 1 TÜBİTAK, National

More information

Ultrasonic Gas Flow Meter

Ultrasonic Gas Flow Meter P r o d u c t i n f o r m at i o n FLOWSIC600 Ultrasonic Gas Flow Meter For custody transfer and process applications Analyzers and Process Instrumentation FLOWSIC600 For Custody Transfer and Process Applications

More information

The Coulter Principle for Outlier Detection in Highly Concentrated Solutions

The Coulter Principle for Outlier Detection in Highly Concentrated Solutions The Coulter Principle for Outlier Detection in Highly Concentrated Solutions Particle Characterization Application Note Abstract This application note will benefit chemical and material manufacturers concerned

More information

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL.

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL. The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL. Weather Data Benefit For Society The four most important things about weather prediction are quality, reliability,

More information

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER Technical Description Version 2 BGRIMM AUTOMATION, South 4th Ring Road West, Beijing, China TEL: +86(10)59069762 Fax: +86(10)68360101 Contents BGRIMM AUTOMATION

More information

v = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes

v = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Examples of Application Flow measurements also in hazardous locations X zone 0 or zone 1 in clean rooms in air

More information

IEC Tritium Standard

IEC Tritium Standard detect and identify IEC 62303 Tritium Standard Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 21 st Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas, Nevada,

More information

CHAPTER D3 TOF ION OPTICS

CHAPTER D3 TOF ION OPTICS Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide...399 Summary...401 Background...403 EquationsofMotionofIons...403 Resolution...405 Reflectron...407 Comparison

More information

XR Analog Clock - Manual Setting Model Troubleshooting Guide

XR Analog Clock - Manual Setting Model Troubleshooting Guide Primex XR 72MHz Synchronized Time Solution XR Analog Clock - Manual Setting Model Troubleshooting Guide 2018 Primex. All Rights Reserved. The Primex logo is a registered trademark of Primex. All other

More information

Membrane Performance Forecast

Membrane Performance Forecast Membrane Performance Forecast Interested in Membranes? Liquid Selectivity in s ea cr Zeta potential analysis with SurPASS 3 from Anton Paar opens up new possibilities in the characterization of membranes

More information

NEUTRON TECHNOLOGIES (PAPER II)

NEUTRON TECHNOLOGIES (PAPER II) Appendix O NEUTRON TECHNOLOGIES (PAPER II) David A. Sparrow, Institute for Defense Analyses 1 EXECUTIVE SUMMARY Neutrons pass relatively easily through matter and have been used successfully to interrogate

More information

Dr. OligoTM DNA / RNA / OLIGO SYNTHESIZERS

Dr. OligoTM DNA / RNA / OLIGO SYNTHESIZERS Dr. OligoTM DNA / RNA / OLIGO SYNTHESIZERS High Throughput Oligo Synthesis Synthesize Cleave Deprotect Desalt Elute Dr. Oligo TM The Dr. Oligo TM High Throughput Oligo Synthesizer is available in four

More information

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

White Paper. Perform Conductivity Measurements In Compliance with USP <645> Perform Conductivity Measurements In Compliance with USP Water is the most widely used substance, raw material, or ingredient in the production, processing and formulation of compendial articles.

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

FEDERAL STANDARD 209E. Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones

FEDERAL STANDARD 209E. Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones FEDERAL STANDARD 209E Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones Revised 1992 by the Institute of Environmental Sciences 940 E. Northwest Highway Mount Prospect, Illinois 60056

More information

A Case for Periodic Calibration or Verification of RTDs

A Case for Periodic Calibration or Verification of RTDs A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone 650-479-3208 Access code 668 676 434 Your Host and Presenter

More information

THE SPEAR3 UPGRADE PROJECT AT SLAC *

THE SPEAR3 UPGRADE PROJECT AT SLAC * IWAA2004, CERN, Geneva, 4-7 October 2004 THE SPEAR3 UPGRADE PROJECT AT SLAC * Catherine Le Cocq, Brian Fuss, Michael Gaydosh Stanford Linear Accelerator Center, Stanford, CA, USA 1. INTRODUCTION The Stanford

More information

PTG-NIR Powder Characterisation System (not yet released for selling to end users)

PTG-NIR Powder Characterisation System (not yet released for selling to end users) Powder Characterisation System - PTG-NIR PTG-NIR Powder Characterisation System (not yet released for selling to end users) The PTG-NIR automatic powder and granule inspection system with integrated NIR

More information

Introduction to Blackbody Sources

Introduction to Blackbody Sources Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with

More information

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ Design and Construction of a High Energy X-Ray R&D Facility, and the Development and Optimization of Real Time Radioisotopic Characterization of Remote Handled Waste at MeV Energies. S. Halliwell, V.J.Technologies

More information

Supporting Information

Supporting Information Supporting Information Analysis Method for Quantifying the Morphology of Nanotube Networks Dusan Vobornik*, Shan Zou and Gregory P. Lopinski Measurement Science and Standards, National Research Council

More information

2πσpred. The probability of exceedance, or the probability of the actual ground motions a exceeding the threshold a thresh, is given by.

2πσpred. The probability of exceedance, or the probability of the actual ground motions a exceeding the threshold a thresh, is given by. 299 Chapter 9 information How to use early warning Ultimately, the goal of seismic early warning is to provide users with information that will help them in deciding the optimal course of action in the

More information

Certificate: / 30. April 2015

Certificate: / 30. April 2015 Test report: 936/21226418/A of 29 September 2014 Initial certification: 2. April 2015 Date of expiry: 1. April 2020 Publication: BAnz AT 02 April 2015 B5, chapter III number 3.1 Approved application The

More information

Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEASUREMENT PROCEDURES...3! III. ANALYSIS PROCEDURES...7!

Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEASUREMENT PROCEDURES...3! III. ANALYSIS PROCEDURES...7! Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEAUREMENT PROCEDURE...3! III. ANALYI PROCEDURE...7! Revised July 2011 I. Introduction In this experiment you will measure the

More information

1 Introduction. Station Type No. Synoptic/GTS 17 Principal 172 Ordinary 546 Precipitation

1 Introduction. Station Type No. Synoptic/GTS 17 Principal 172 Ordinary 546 Precipitation Use of Automatic Weather Stations in Ethiopia Dula Shanko National Meteorological Agency(NMA), Addis Ababa, Ethiopia Phone: +251116639662, Mob +251911208024 Fax +251116625292, Email: Du_shanko@yahoo.com

More information

CMM Uncertainty Budget

CMM Uncertainty Budget Table of Contents Purpose...3 Scope...3 Measurement Example...3 Uncertainty Groups...4 Probe Group...5 Calibration Artifact Group...6 Environment Group...7 CMM Group...8 Part Group...10 Conversion of Potential

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method

More information

Alpha-Energies of different sources with Multi Channel Analyzer

Alpha-Energies of different sources with Multi Channel Analyzer Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy

More information

Measurement of Tritium in Helium

Measurement of Tritium in Helium detect and identify Measurement of Tritium in Helium Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 22 nd Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas,

More information

Update on the BIPM ensemble of mass standards. Estefanía de Mirandés Mass Department BIPM

Update on the BIPM ensemble of mass standards. Estefanía de Mirandés Mass Department BIPM Update on the BIPM ensemble of mass standards Estefanía de Mirandés Mass Department BIPM 1 Acknowledgments Aldo Dupire Stéphane Ségura Faraz Idrees BIPM mechanical workshop 2 Present Configuration: reminder

More information

Compact Orifice FlowMeter

Compact Orifice FlowMeter Steam Data Reference Tables IM/OM/SDT Issue 1 Compact Orifice FlowMeter ABB The Company We are an established world force in the design and manufacture of instrumentation for industrial process control,

More information

Acal BFi supply bespoke photonics solution to Department of Chemistry at Cambridge University

Acal BFi supply bespoke photonics solution to Department of Chemistry at Cambridge University Acal BFi supply bespoke photonics solution to Department of Chemistry at Cambridge University About the project The Department of Chemistry at Cambridge University needed a suite of products to aid a significant

More information

Field study of the latest transmissometers at Hong Kong International Airport

Field study of the latest transmissometers at Hong Kong International Airport Field study of the latest transmissometers at Hong Kong International Airport P. W. Chan Hong Kong Observatory 134A Nathan Road, Kowloon, Hong Kong, China Tel:+852 2926 8435, Fax: +852 2311 9448, Email:

More information

SECTION 8 Part I Typical Questions

SECTION 8 Part I Typical Questions SECTION 8 Part I Typical Questions 1. For a narrow beam of photons, the relaxation length is that thickness of absorber that will result in a reduction of in the initial beam intensity. 1. 1/10. 2. 1/2.

More information

Heat Transfer in Polymers

Heat Transfer in Polymers Heat Transfer in Polymers Angela Dawson, Martin Rides, Crispin Allen 18 April 2007 Heat Transfer in Polymers - Summary Introduction Heat Transfer Coefficients Standards for Thermal Properties Measurement

More information

Model 2300XP PSL & Process-Particle Wafer Deposition System

Model 2300XP PSL & Process-Particle Wafer Deposition System Model 2300XP PSL & Process-Particle Wafer Deposition System Deposit PSL spheres on wafers to create NISTtraceable PSL size standards for - calibrating wafer inspection systems - providing fab-wide and

More information

Quality Assurance Plan. March 30, Chemical Crystallography Laboratory

Quality Assurance Plan. March 30, Chemical Crystallography Laboratory Quality Assurance Plan Page 1 of 7 Quality Assurance Plan March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Quality Assurance Plan Page 2 of 7 Distribution Douglas R. Powell,

More information

Introduction to Differential Sedimentation

Introduction to Differential Sedimentation Introduction to Differential Sedimentation Differential Centrifugal Sedimentation, or DCS (sometimes also called "two-layer" sedimentation) is a widely used analysis method that produces extremely high

More information

Extrel is widely respected for the quality of mass spectrometer systems that are

Extrel is widely respected for the quality of mass spectrometer systems that are Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built

More information

Acceleration/Velocity/Displacement VIBRATION METER

Acceleration/Velocity/Displacement VIBRATION METER Acceleration/Velocity/Displacement VIBRATION METER Model : VB-8220 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is a complex

More information

Characterization of Large Structures & Components

Characterization of Large Structures & Components Structures & Components KEY BENEFITS Key Drivers: Lack of good knowledge about the position, the identification and the radiological specification of contamination on or inside large components. Significant

More information

Measuring Laser Diode Optical Power with an Integrating Sphere

Measuring Laser Diode Optical Power with an Integrating Sphere Measuring Laser Diode Optical Power with an Integrating Sphere Introduction Characterizing radiant sources like laser diodes accurately depends on the ability to measure their optical power output accurately.

More information

Experimental Design and Data Collection

Experimental Design and Data Collection A. Checks to run to make sure the instrument is in good working condition B. Decide which experiment type is most appropriate, equilibrium or velocity C. Buffer considerations D. Speed selection and length

More information

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P ) Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:

More information

Diffraction Limited Size and DOF Estimates

Diffraction Limited Size and DOF Estimates Diffraction Limited Size and DOF Estimates Abstract In this manuscript we describe the process by which we use laboratory measurements together with diffraction theory to improve estimates of depth of

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

v = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes

v = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes with thin-film sensor element Probes Probe guides Measuring tubes Examples of Application Flow measurements also in hazardous locations Ex zone 0 or zone 1 in clean rooms in air conditioning and ventilating

More information

Natural Rock Sample Manual

Natural Rock Sample Manual Natural Rock Sample Manual Revision 2014-06-12 DURRIDGE Company Inc. 524 Boston Road Billerica, MA 01821 Tel: (978) 667-9556 Fax: (978) 667-9557 service@durridge.com www.durridge.com 2014, DURRIDGE Company

More information

Our Product is our Service. Hammer Division. English

Our Product is our Service. Hammer Division. English Our Product is our Service Hammer Division English Oil & Gas applications The Hydrohammer Technology has found application on the local market, mainly in the Oil & Gas industry where pile conductors are

More information