The Thermo Scientific ice 3500 AA Spectrometer - a True Dual Atomizer AAS

Size: px
Start display at page:

Download "The Thermo Scientific ice 3500 AA Spectrometer - a True Dual Atomizer AAS"

Transcription

1 Technical Note: The Thermo Scientific ice 3500 AA Spectrometer - a True Dual Atomizer AAS This technical note provides an insight into the design concepts of the ice 3500 dual atomizer instrument. Key Words Absorption Dual Atomizer FAAS GFAAS Graphite Furnace Quadline D 2 Background Correction Zeeman Background Correction Introduction Dual atomizer atomic absorption spectrometers are, as the name suggests, simply those that include both a flame and graphite furnace atomizer in one instrument. The benefits of such instruments are numerous and this technical article discusses the design merits of the Thermo Scientific ice 3500 dual atomizer spectrometer, and contrasts them against other optical designs. The benefits of dual atomizer instruments in the modern analytical laboratory will be explored, with a simple comparison between a single and a dual atomizer method being presented. The advantages of a dual atomizer instrument over what can be regarded as the more simple approach of buying two separate single atomizer AA instruments is also discussed. Atomic Absorption Spectrometry Atomization Sources The technique of Atomic Absorption Spectrometry requires an atomization source. This is typically either a flame (FAAS) or a graphite furnace (GFAAS) depending on the sensitivity required and the sample matrix. FAAS - Flame Atomization The sample is normally presented to the instrument as a liquid. When it is passed through the nebulizer, it forms a fine mist or spray, which is mixed with the oxidant and fuel gases in the spray chamber before being passed into the flame. Once in the flame, the sample undergoes evaporation of solvent, vaporization of salts and finally atomization. The flame fuel gas is usually acetylene, and either air or nitrous oxide is used as the oxidant gas. Nitrous oxide supported flames burn at much higher temperatures than air supported flames. Flame temperature has a strong influence on the atomization process, because hotter flames are required to vaporize some salts and to decompose certain compounds. Stoichiometry also affects flame temperatures and consequently has to be optimized for each sample and element required. Flame atomization systems are simple and relatively easy to use but do have several disadvantages: They require fairly large sample volumes They have limited sensitivity GFAAS - Electrothermal (Graphite Furnace) Atomization In this technique, an electrically heated graphite furnace is used as the atomizer. GFAAS can be used to measure most of the same elements as the flame but with typically 1,000 times the sensitivity and a fraction of the sample size. A discrete volume, typically a few micro-litres, of the sample is placed inside an small graphite cuvette, and heated using the following general procedure: i. Drying phase, where the sample is warmed to remove the solvent ii. Ashing phase, where as much of the sample matrix as possible is removed iii. Atomization or measurement phase iv. Cleaning phase, where the cuvette is heated to a high temperature to remove any previous sample The analytical signal is measured as a transient peak instead of the continuous signal used in flame work. The residence time of analyte atoms in the light path is increased with this technique and the total volume of the sample used contributes to the final signal. Both of these effects result in much improved analytical sensitivity. A dual atomizer instrument includes both the flame and graphite furnace atomizers in one instrument. Why would a Modern Analytical Laboratory want a Dual Atomizer AAS? Many laboratories are required to carry out analyses that necessitate the application of both FAAS and GFAAS techniques. This requires one of three generic instrument package options (Figure 1), and the optimum solution depends largely on the throughput and frequency of measurements requiring the two techniques. The possible generic solutions are: Option 1: Option 2: Two separate, dedicated instruments - one FAAS instrument one GFAAS instrument (e.g. A Thermo Scientific ice ) A FAAS instrument with a GFAAS atomizer option (Thermo Scientific ice GFS33) Option 3: A Dual Atomizer Instrument (Thermo Scientific ice 3500)

2 FAAS GFAAS FAAS GFAAS FAAS GFAAS Dual Atomizer Figure 1: The Three Options for Flame and Furnace Atomizer Capability The benefits and disadvantages of each option are explained as follows: Option 1: Two separate, dedicated instruments The first package is predominately used by extremely high throughput laboratories with roughly equal, constant and very high numbers of FAAS and GFAAS samples. If it is established that the throughput of the laboratory is higher than a single dual atomizer instrument can cope with (then the laboratory really does run a lot of samples!), Thermo Scientific can of course provide two AA systems. It is sometimes suggested that cost can be saved, and convenience improved, by running both instruments with a single Data Station. However, Thermo Fisher Scientific does not recommend this approach. Running two instruments from a single Data Station is relatively easy to accomplish, but we know that the problems introduced for the analyst far outweigh the perceived benefits of such an approach. There are a number of problems that can be encountered when using two AAS instruments running off one Data Station. The user can find that rather than doubling their productivity they actually cause themselves 'double the trouble'. The potential exists that a user wishing to run the same element by both FAAS and GFAAS in the same analysis would require two sets of hollow cathode lamps (HCL) - doubling the cost Two separate instruments running off one PC will require two independent instances of the instrument software to be running on one PC. Imagine trying to find your way around two versions running separately and with no doubt cluttered screens - very confusing, how can you tell which is the flame version and which is the furnace? Two databases on one PC Two sets of instrument spares Two separate service contracts - one for each instrument would be necessary Two AA's have twice the footprint, bench space is precious in most labs, is there room for 2 AA's? The potential exists for 2 operators to be using the software but only one can be securely logged on, surely there are compliance issues with this approach? Option 2: A single atomizer instrument with a GFAAS accessory The second option, is to buy a FAAS with a separate GFAAS accessory. This is sometimes chosen as a cost effective solution when the number of flame samples to be analyzed greatly exceeds the number of furnace samples, so that GFAAS capability is required only rarely. For example, for infrequent, unknown or emergency samples. Initially only a flame system may be necessary but over time the analytical needs of the laboratory can grow, so that a GFAAS capability is required. Hence when making the purchasing decision on flame-only systems it is important to ensure the chosen system is adequately 'futureproof' i.e. the system is capable of GFAAS upgrade. The disadvantage of this approach is the time and complexity involved in changing over the atomizers in the instrument. Different vendors have alternate solutions for this operation but the generic process of changing over atomizers is illustrated in Figure 2. At least two critical alignment procedures are required; if these are not performed correctly for any reason the quality of the data generated will deteriorate. Switch on HCL s and allow stabilzation Light flame Load Flame/Dual method Make flame measurements Extinguish the flame Dual Atomizer Make furnace measurements Single Atomizer Load furnace method Allow burner to cool Align furnace autosampler Remove flame autosampler Remove burner assembly Fit furnace assembly and autosampler Align furnace assembly Figure 2: Flowchart showing the additional steps required if using a single Option 3: A dual atomizer instrument A dual atomizer instrument will have both atomizers permanently mounted so there is no requirement to spend time realigning atomizers prior to an analysis. Both flame and furnace auto-samplers can be left in place, and the furnace auto-sampler in particular should retain its alignment. With properly designed software, a dual atomizer instrument can be easily controlled from a single Data Station without any of the problems encountered with the two AA's approach. An additional benefit is that flexible combinations of atomizers are possible on dual atomizer instruments which provide further flexibility to enable the optimum configuration to be set-up. Examples of typical combinations: FAAS and GFAAS (ice 3500) FAAS and Electrically Heated Vapor Generation (ice EC90 + VP100)

3 Electrically Heated Vapor Generation and Zeeman GFAAS (ice EC90 + VP100) Flame heated vapor generation and Zeeman/D2 only GFAAS (ice VP100) Since the switch between analyzers is automatic, dual analysis sequences can be set up such that the flame analyses are completed then the instrument switches analyzers and runs the furnace analyses. The benefits and disadvantages of each of these approaches are summarized in Table 1. For the vast majority of laboratories where sample throughput is medium to high and the use of GFAAS in addition to FAAS is regular and/or frequent the table indicates that the third option of a dual atomizer is the optimum configuration. The ice 3500 dual atomizer instrument has been designed based on these considerations. FEATURE OPTION 1. OPTION 2. OPTION 3. TWO SEPARATE SINGLE ATOMIZER DUAL ATOMIZER DEDICATED AA S WITH GFAAS AAS ACCESSORY Flame High Medium/High High Throughput Furnace Medium/High Low Medium/High Throughput Frequency High Low High of GFAAS Samples Hollow Cathode 2 Sets 1 Set 1 Set Lamps Software 2 Instances 1 Instance 1 Instance Atomizer Change Manual Software changeover Instruments Changeover controlled, automatic Ease of Atomizer switch N/A Poor Excellent Instrument Service Contracts Spares 2 Sets 1 Set 1 Set Table 1. Instrument Option Comparison. Do all Dual Atomizer Instruments use the Same Arrangement of Atomizers? There are a number of possible options for arranging the dual atomizers. An understanding of the other optomechanical arrangements for the switch between atomizers is useful to put into perspective the elegance and simplicity of the ice 3500 design. Such options are illustrated in Figure 3 include: Fixed atomizers, where the two atomizers are positioned in a serial arrangement, with the light beam always passing through both atomizers. The major disadvantage of this approach is that the instrument suffers from greatly reduced light throughput because extra interatomizer optics are required. This results in degraded detection limits. The physical positioning of the atomizers also impacts on the size of the instrument required to house the optics, leading to large, long instruments. Additionally, perfect alignment of two serial atomizers is very difficult to achieve. a) Burner b) Burner c) Burner Burner Furnace Furnace Furnace Figure 3: Other Opto-mechanical arrangements for dual atomizer instruments. a) Serial b) Parallel c) Moving. Moving atomizers, which shuttle or swing into position in a single sample compartment create various problems. The major disadvantage of this approach is that the instruments rely on positioning complex precision mechanisms in chemically vulnerable areas prone to corrosion. Atomizer changeover can be quite slow in operation, as the movement of one atomizer out of position must take place before the new one can be moved in. Poor repeatability of the repositioning can result in inadequate optical reproducibility, degrading the quality of the analytical results. The mechanics and engineering required with such designs result in large instrument footprint sizes. Designs based on this principle cannot be considered to be truly automatic as the changeover usually involves some sort of manual intervention. For example, it is often necessary to remove and re-fit auto-samplers and burner heads. This means that the changeover is not truly automated and cannot be carried out automatically under software control.

4 How was the ice 3500 Dual Atomizer Instrument Designed? The ice 3500 dual atomizer instrument was designed with very specific optical requirements. The following factors were considered crucial to the instrument design: Dual-atomization - with the atomizers permanently in position, such that they would require no re-alignment The switch between atomizers should be instantaneous and the changeover capable of occurring whilst the instrument is unattended No compromise in instrument performance should be sacrificed to fulfil the dual atomizer increased functionality The instrument must present continuum source background correction as standard and should encompass a Zeeman furnace option The lamp carousel should house six data-coded hollow cathode lamps all being capable of auto-alignment and with separate power supplies The optics of the instrument should incorporate Stockdale double beam optics, and include the option for a Furnace Vision System' The instrument should provide full accessory compatibility A footprint of less 0.5 m 2 was desired as a response to the reality that in today's extremely cost-conscious, commercial, environment laboratory space is a valuable commodity The first major conceptual breakthrough with the design of the ice 3500 was the realization that in a double beam system the function of reference and sample beam are interchangeable. Therefore an atomizer could be placed in each beam and the function of the beams changed by a simple software command. The rest of the instrument design was based around optical components that optimize the optical signal, combined with highly functional and ergonomic user access. The ice 3500 Optical Concept In the ice 3500 optical system the lamp turret is mounted at the front of the instrument in the centre, enabling easy user access. The turret is mounted vertically, optimizing bench-space and ensures that no forces exist that would act to misalign the lamps. Radiation from the HCL passes vertically upwards towards a motorized deflecting mirror which in tandem with microrotation of the HCL carousel achieves fast, automatic lamp selection and alignment. From this first lamp selector mirror the optical beam passes through a speckled beam combiner. This component combines the light beam from the HCL with the light emitted by the QuadLine deuterium arc lamp, which is required to implement the QuadLine background correction system. The next step is that both HCL and D 2 beams meet the front beam selector mirror, which directs the combined beam either to the left, to the flame atomizer or to the right to the furnace atomizer. After passing through the atomizer, a plane mirror then deflects each beam to the rear of the instrument. Toroidal focusing mirrors at the rear corners of the instrument deflect and focus the beams onto the rear beam selector mirror, which works in concert with the front beam selector mirror, and determines whether the beam which has passed through the flame or the furnace atomizer is directed to the monochromator (see Figure 4). It is inevitable that the design of a dual atomizer will involve a small increase in reflective surfaces in comparison with a simple single beam system. It is important that these small losses are more than compensated for by the quality of the remaining optics. It was determined that the optical system should be designed around the requirements for a graphite furnace analysis and the furnace cuvette profile since this is the more demanding application. Longer light paths necessitate the use of a pencil beam optical path (Figure 5) through the instrument, to reduce divergence. This has the additional benefit that the measurement beam only passes through the hottest region of the atomizer, resulting in reduced tube wall emission effects (breakthrough) and improved sensitivity for furnace measurements and less curvature and improved sensitivity for flame measurements. Figure 4: The Optical design of the ice 3500 Dual Atomizer AAS. a) b) Figure 5: 'Pencil beam' optics passing through a) furnace cuvette b) flame atomizer.

5 The ice 3500 Monochromator These design considerations led to the specification of a monochromator capable of covering the normal range of AA wavelengths, from 180 to 900 nm, with high light throughput, a small optical aperture, high dispersion, and a small focal length, and able to provide conventional AA bandpasses. The option of a conventional monochromator was ruled out since a 40 x 60 nm grating with 1800 l/mm and a 500 nm focal length would be required to provide the dispersion needed. The long focal length could cause more thermal/mechanical drift and would not meet the desired footprint requirement. Increasing the number of lines on the grating and/or the size of the grating would restrict the wavelength range, and reducing the slit widths would lead to reduced light throughput and poorer detection limits. The only real alternative was to move to the use of an Echelle monochromator. An Echelle monochromator contains two dispersing elements (typically a grating and a prism), and provides excellent optical performance combined with very small space requirements. Other benefits of this design include: High dispersion even with short focal lengths Relatively small physical size Very rapid wavelength selection - selected by moving grating & order sorter simultaneously Monochromator Very low stray light, as a result of the double monochromator design, giving improved AA sensitivity and less calibration Camera curvature Greater stability - less demanding mechanics and assembly lead to greater reliability Physically wider slits can be used for a given bandpass, giving greater light throughput, and better detection limits HCL The Echelle system used on the ice 3500 has a reciprocal linear dispersion (RLD) of ~0.5 nm/mm at 200 nm. It is desirable to have as small a reciprocal linear dispersion as possible since this improves the amount of radiation passing the monochromator and reaching the detector even for small bandpass widths, improving sensitivity. To put the performance of the echelle into perspective it is approximately as good as a half metre focal length traditional monochromator. Field lens Additional Design Features The unique Thermo Fisher Scientific Stockdale double beam optics are effectively included for free in the ice 3500 optical system. The traditional 'leaping' mirrors are now replaced by the front and rear beam selectors. This makes all systems automatically double beam, with the added benefit that the Stockdale method ensures that during the measurement period, the system effectively operates in the optimum single beam mode, whilst reference bracketing measurements before and after the key measuring period ensure perfect drift correction. The back of the rear beam selector has also been designed to be reflective and is cleverly used to deflect the furnace optical beam below the optical base and into a small additional optical system that terminates in the optional 'Furnace Vision' GFTV camera (Figure 6). GFTV provides high definition images of events inside the furnace cuvette, this allows monitoring of the sample injection, dry and ash phases of the furnace program (see Figure 7). Another additional component to the furnace optical system is the polarizer, which is only fitted to Zeeman systems. This component is motorized in order to simply flip into the beam when the Zeeman background correction is being employed and to flip out of the beam when a Quadline only analysis is being carried out. Leaping mirror Cuvette GFTV Off Monochromator Camera HCL Field lens Leaping mirror Cuvette GFTV On Figure 6: Schematic showing the deflection of the furnace optical beam into the GFTV camera The entire optical system is mounted in an aluminum precision casting to provide it with mechanical and thermal stability and the system is built into a dust sealed structure to ensure long-term optical cleanliness and stability. The ice 3500 was designed such that the instrument depth fits a standard laboratory benchtop, with adequate space for access, ventilation, extraction and supply lines, the width was then minimized as much as the instrument optics would allow resulting in the ice 3500 final dimensions being the smallest footprint of any dual atomizer instrument on the market.

6 Figure 7: Clear, high-resolution 'Furnace Vision' (GFTV) images of the interior of the cuvette. Indicating the position of capillary tip as: too high, too low, optimum and with a platform cuvette. Deuterium Background Correction The unique QuadLine D 2 arc source that provides the broad band radiation used to implement the QuadLine background correction system is much brighter than most HCL's. The 'speckle ratio' of the HCL/D 2 beam combiner can therefore be increased to an 80/20 split in favor of the HCL to allow more energy from the measuring HCL to reach the detector. The Deuterium arc lamp behaves as a point source, while the HCL s sources produce a broader, diffuse patch of light. The QuadLine background correction approach demands that the two optical beams have the same energy profile and image size. To achieve this, the ice 3500 optically magnifies the HCL beam by a factor of 2 at the Atomizer, and the D 2 beam by a factor of 3. The D 2 beam is deliberately de-focused, so that the resultant light patch at the atomizer matches the energy profile of the HCL beam. The imbalance between the intensities of the QuadLine and HCL lamps increases as the wavelength decreases. The QuadLine beam is therefore focused with a silica lens rather than a mirror. The focal length of a lens is dependent on the wavelength of the light passing throughout, so that the effect of this component is to progressively de-focus the QuadLine beam as the wavelength decreases, thus reducing the excessive energy of the QuadLine beam and enabling a better match with the HCL intensity to be achieved. This design, together with other unique attributes described in detail in reference 1, provide confidence in the accuracy of the QuadLine background correction, and enables Thermo Fisher Scientific to guarantee that the system can correct up to 2A of background signal with less than 2 % error. Zeeman Background Correction The AC inverse Zeeman system used on the ice 3500 and ice 3400 instruments is the only background correction technique available that corrects at the exact analyte wavelength, gives accurate correction of structured backgrounds and spectral overlaps (See Figure 8). This is vital if the sample background signal has significant fine structure or if there is a spectral interference present. The Thermo Scientific Zeeman system employs a very high field strength magnet (0.9 Tesla), which improves the magnitude of the Zeeman effect, improving sensitivity and detection limits, and reducing the calibration curvature traditionally associated with Zeeman effect based background correction. The magnet drive current is modulated at Hz with a trapezoidal waveform, so that fast furnace signals are accurately tracked, and the complete background correction measurement is taken at full field strength. Uniquely the ice 3500 allows a combined QuadLine/Zeeman mode to be run for method development work. Zeeman measurements are only made at the time of measurement, but QuadLine can be used to monitor background or interfering species in the dry and ash phases of a furnace measurement.

7 Absorption + π Absorption profile split into π and components by magnetic field around atomizer Emission Normal hollow cathode lamp radiation Magnet off: Normal AA system Signal measured = AA + Background Static Polarizer π: component removed by polarizer HCL measures background only Background Figure 8: The AC Inverse Zeeman Background Correction Magnet on: Signal measured = Background only Conclusions On the ice 3500 the switch between a flame analysis and a furnace analysis is simply achieved by the flip of a mirror, therefore, flame to furnace changeover can be totally automatic - no manual intervention is required. Alignment of both atomizers is perfect, productivity is significantly enhanced and performance guaranteed. No energy is wasted and single beam performance is achieved with double beam stability. The design of our dual atomizer means that there is no energy loss due to extra relay optics. References 1) Design Considerations for High Performance Background Correction Systems in Atomic Absorption Spectrometry - Thermo Scientific Technical Brief

8 Thermo Scientific ice 3000 Series Product Range In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. Thermo Scientific ice 3300 Single flame atomizer AAS with fully automatic gas box. Complete solution for laboratories with a main need to perform flame analysis but with occasional furnace samples. Simple flame system but with incredible versatility Six lamp auto-aligning carousel Double beam optics and self-calibrating Ebert monochromator Thermo Scientific ice 3400 Single furnace atomizer AAS with Zeeman and D2 background correction. When challenging detection limits are critical. Six lamp auto-aligning carousel Furnace vision Echelle dual prism and and grating monochromator Vapour system and electrically heated cell can be utilised in this instrument Thermo Scientific ice 3500 Dual flame and furnace AAS with standard or Zeeman furnace option. Essential furnace vision tool included as standard. Ideal for high throughput environments with a requirement for quick and regular flame and furnace analysis. Software-controlled changeover from flame to furnace analysis without the operator even being in the room! Six lamp auto-aligning carousel for maximum light throughput D2 background correction for flame and furnace analysis Zeeman background correction option available for furnace work Double beam optics with a dual monochromator consisting of an echelle prism and a grating Africa Australia Austria Belgium Canada China Denmark Europe-Other France Germany India Italy Japan Latin America Middle East Netherlands South Africa Spain Sweden / Norway / Finland Switzerland UK USA FM Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Electron Manufacturing Ltd (Cambridge) is ISO Certified. TN40942_E 02/08C Part of Thermo Fisher Scientific

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

Fast, Effective XPS Point Analysis of Metal Components

Fast, Effective XPS Point Analysis of Metal Components Application Note: 52297 Fast, Effective XPS Point Analysis of Metal Components Chris Baily and Tim Nunney, Thermo Fisher Scientific, East Grinstead, West Sussex, UK Key Words K-Alpha Auto-Analysis Multi-Spectrum

More information

of mass spectrometry

of mass spectrometry Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Application Note: 52252 The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Mark Wall, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words DXR Raman Microscope 2D Band D

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier

More information

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy

More information

for XPS surface analysis

for XPS surface analysis Thermo Scientific Avantage XPS Software Powerful instrument operation and data processing for XPS surface analysis Avantage Software Atomic Concentration (%) 100 The premier software for surface analysis

More information

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,

More information

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization Thermo Scientific K-Alpha + XPS Spectrometer Fast, powerful and accessible chemical analysis for surface and thin film characterization X-ray Photoelectron Spectroscopy Quantitative, chemical identification

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Defining quality standards for the analysis of solid samples

Defining quality standards for the analysis of solid samples Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples

More information

COMPARISON OF ATOMIZERS

COMPARISON OF ATOMIZERS COMPARISON OF ATOMIZERS FOR ATOMIC ABSORPTION SPECTROSCOPY Introduction Atomic spectroscopic methods are all based on the interaction of light and analyte atoms in the gas phase. Thus, a common component

More information

2101 Atomic Spectroscopy

2101 Atomic Spectroscopy 2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.

More information

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry 9.1 9A. Atomization Most fundamental for both techniques. Typical types 1. flame - burner type 2. Electrothermal graphite furnace 3. Specialized

More information

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2

More information

Optical Atomic Spectroscopy

Optical Atomic Spectroscopy Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state

More information

Atomic Spectroscopy AA/ICP/ICPMS:

Atomic Spectroscopy AA/ICP/ICPMS: Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

Atomic Absorption & Atomic Fluorescence Spectrometry

Atomic Absorption & Atomic Fluorescence Spectrometry Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,

More information

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan

More information

MetWorks Metabolite Identification Software

MetWorks Metabolite Identification Software m a s s s p e c t r o m e t r y MetWorks Metabolite Identification Software Enabling Confident Analysis of Metabolism Data Part of Thermo Fisher Scientific MetWorks Software for the Confident Analysis

More information

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis Alessandro Santato, 1 Doug Hamilton, 1 Jan Wijbrans, 2 Claudia Bouman 1 1 Thermo Fisher Scientific, Bremen, Germany 2 VU University Amsterdam,

More information

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces Application Note: 52287 XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces Brian R. Strohmeier, Thermo Fisher Scientific, Madison, WI, USA

More information

Emission spectrum of H

Emission spectrum of H Atomic Spectroscopy Atomic spectroscopy measures the spectra of elements in their atomic/ionized states. Atomic spectrometry, exploits quantized electronic transitions characteristic of each individual

More information

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS m a s s s p e c t r o m e t r y Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS Fully Automated Multi-Element Isotope Analysis for C, N, S, O, H Food Testing Forensics Environmental Analysis

More information

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision ARGUS VI Static Vacuum Mass Spectrometer Static Vacuum ARGUS VI Multicollection Low Volume Precision Based on more than 20 years of experience in noble gas mass spectrometry instrumentation, we have developed

More information

Complete Materials Deformulation Using TGA-IR

Complete Materials Deformulation Using TGA-IR Application Note: 51694 Complete Materials Deformulation Using TGA-IR Michael Bradley, Thermo Fisher Scientific, Madison, WI, USA Key Words Feedstock Supplies Finished Products Multi-component Search OMNIC

More information

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

IN QUALITATIVE ANALYSIS,

IN QUALITATIVE ANALYSIS, IN QUALITATIVE ANALYSIS, YOU VE ALWAYS HAD TO CHOOSE BETWEEN ACCURACY AND THROUGHPUT. NOW THERE S NO REASON TO ever have to. WATERS QUALITATIVE ANALYSIS solutions Waters ACQUITY UPLC System with the Waters

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Lecture 7: Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Anthony Edge, 1 Luisa Pereira, 1 Dafydd Milton 1 and Andrew Shalliker 2 1 Thermo Fisher Scientific,

More information

a. An emission line as close as possible to the analyte resonance line

a. An emission line as close as possible to the analyte resonance line Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal

More information

Atomic Absorption Spectroscopy (AAS)

Atomic Absorption Spectroscopy (AAS) Atomic Absorption Spectroscopy (AAS) Alex Miller ABC s of Electrochemistry 3/8/2012 Contents What is Atomic Absorption Spectroscopy? Basic Anatomy of an AAS system Theory of Operation Practical Operation

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

Performance and Control of the Agilent Nano Indenter DCM

Performance and Control of the Agilent Nano Indenter DCM Performance and Control of the Agilent Nano Indenter DCM Application Note Introduction With new materials and material applications come new test challenges. As these new challenges emerge in materials

More information

Atomization. In Flame Emission

Atomization. In Flame Emission FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic by its monatomic particles in gaseous state in the

More information

ATOMIC SPECROSCOPY (AS)

ATOMIC SPECROSCOPY (AS) ATOMIC ABSORPTION ANALYTICAL CHEMISTRY ATOMIC SPECROSCOPY (AS) Atomic Absorption Spectroscopy 1- Flame Atomic Absorption Spectreoscopy (FAAS) 2- Electrothermal ( Flame-less ) Atomic Absorption Spectroscopy

More information

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Dirk Nolting and Andreas Wieghaus Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Characterize the rate

More information

Atomic Emission Spectroscopy

Atomic Emission Spectroscopy Atomic Emission Spectroscopy Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately

More information

Chemistry Instrumental Analysis Lecture 18. Chem 4631

Chemistry Instrumental Analysis Lecture 18. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 18 Instrumentation Radiation sources Hollow cathode lamp Most common source Consist of W anode and a cathode sealed in a glass tube filled with Ne or Ar. Hollow

More information

very high temperature for excitation not necessary generally no plasma/arc/spark AAS

very high temperature for excitation not necessary generally no plasma/arc/spark AAS Atomic Absorption Spectrometry (Chapter 9) AAS intrinsically more sensitive than AES similar atomization techniques to AES addition of radiation source high temperature for atomization necessary flame

More information

GTEK Laboratory Atomic Absorption Spectrometer AAS6000 Brochure

GTEK Laboratory Atomic Absorption Spectrometer AAS6000 Brochure GTEK Laboratory Atomic Absorption Spectrometer AAS6000 Brochure Description AAS6000 Series are Single Beam Atomic Absorption Spectrometers controlled and data processed by external computer and internal

More information

2. OPERATIONAL CONDITIONS

2. OPERATIONAL CONDITIONS 1. INTRODUCTION This device was designed for modern physics labs of colleges and graduate schools. It demonstrates the influence of a magnetic field on light, known as Zeeman Effect, and reveals the behavior

More information

110 Lowell Rd Hudson, NH USA Tel:

110 Lowell Rd Hudson, NH USA Tel: 110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental

More information

9/13/10. Each spectral line is characteristic of an individual energy transition

9/13/10. Each spectral line is characteristic of an individual energy transition Sensitive and selective determination of (primarily) metals at low concentrations Each spectral line is characteristic of an individual energy transition 1 Atomic Line Widths Why do atomic spectra have

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

Traceable UV-Vis Reference Materials

Traceable UV-Vis Reference Materials UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof

More information

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Technical Note 43332 Key Words Overcoming Interferences, Element Finder plug-in Introduction Interferences are common in

More information

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Application Note: 3 Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Yoko Yamagishi, Thermo Fisher Scientific, C-2F 3-9

More information

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,

More information

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer ARGUS VI HELIX SFT Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX SFT Low Volume Multicollection High Mass Resolution 3 Based on more than 20 years of experience in

More information

The Plus for your lab!

The Plus for your lab! BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the

More information

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer

More information

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology

More information

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products APPLICATION NOTE Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products AN43381 Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany Key Words al

More information

Ideal and powerful solution

Ideal and powerful solution Thermo Scientific Indiko Plus Clinical and specialty chemistry system Ideal and powerful solution for operational excellence Enzymes Substrates Electrolytes Specific Proteins Drugs of Abuse Analysis Therapeutic

More information

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Application note Agriculture Authors Annie Guerin INRA, Laboratoire d Analyses

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Oasis HLB DISKS I. Introduction Oasis HLB is a Hydrophilic-Lipophilic-Balanced, water-wettable, reversed-phase sorbent for all your SPE needs. It is made from a specific ratio of two monomers, the hydrophilic

More information

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration.

More information

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)

More information

Atomic Absorption Spectroscopy

Atomic Absorption Spectroscopy CH 2252 Instrumental Methods of Analysis Unit IV Atomic Absorption Spectroscopy Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering

More information

LIBSlab ANALYZERS ANALYZERS

LIBSlab ANALYZERS ANALYZERS ANALYZERS ANALYZERS Chemical multi-elemental analysis with LIBS in modular benchtop design LIBSlab LIBSpector compact sample chamber for the LIBS analysis of solid, liquid and gaseous samples. Sample chamber

More information

SmartNotes. What is meant by the term interference?

SmartNotes. What is meant by the term interference? Interference Removal on ICP-OES icap 7000 Plus Series ICP-OES SmartNotes What is meant by the term interference? In ICP-OES we speak of interference when a result is biased either by other components in

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering

More information

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION A purpose-built collection of the best-inclass components for forensic toxicology Whether you re challenged to present reliable and secure forensic sample

More information

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer m a s s s p e c t r o m e t r y Finnigan TC/EA High Temperature Conversion Elemental Analyzer Analyze Detect Measure Control Finnigan TC/EA High Temperature Conversion Elemental Analyzer Finnigan TC/EA

More information

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER Technical Description Version 2 BGRIMM AUTOMATION, South 4th Ring Road West, Beijing, China TEL: +86(10)59069762 Fax: +86(10)68360101 Contents BGRIMM AUTOMATION

More information

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP Application Note: 40986 Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP Karen Harper, Applications Group Leader, Thermo Scientific, Cambridge,

More information

Product Descriptions and Offerings

Product Descriptions and Offerings Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental analysis. Laboratories in industries ranging from environmental science, oil and

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

RH7-D & RH10-D CAPILLARY RHEOMETERS

RH7-D & RH10-D CAPILLARY RHEOMETERS RH7-D & RH10-D CAPILLARY RHEOMETERS RH7-D and RH10-D Advanced, pow suitable for rese Since its launch, the Rosand RH7 has set new standards in research level capillary rheometry. Today, the Rosand RH7

More information

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Tim Stratton 1, Yingying Huang 1, Katianna Pihakari 1, Ian Acworth 2, and Michael Weber 2 1 Thermo Fisher Scientific,

More information

HILIC Method Development in a Few Simple Steps

HILIC Method Development in a Few Simple Steps HILIC Method Development in a Few Simple Steps Monica Dolci, Luisa Pereira, Dafydd Milton and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Overview This poster presents a systematic approach

More information

Lecture No. 4. i 0 TETA I RY:

Lecture No. 4. i 0 TETA I RY: Lecture No. 4 i 0 TETA I RY: ATOMIC ABSORPTION SPECTROMETRY - PRESENT AND FUTURE Dr. Chandra Sekhar K Scientist Analytical Chemistry Division National Metallurgical Laboratory Jamshedpur 831007 INTRODUCTION

More information

Vapourtec UV-150 Photochemical Reactor. Application notes 37-39, 41-44, 47, 48

Vapourtec UV-150 Photochemical Reactor. Application notes 37-39, 41-44, 47, 48 Vapourtec UV-150 Photochemical Reactor Application notes 37-39, 41-44, 47, 48 Bringing photochemistry to the bench - Photochemistry has a number of important applications in nature and industry. - Traditional

More information

INTRODUCTION TO OPTICAL ATOMIC SPECTROSCOPY (Chapter 8)

INTRODUCTION TO OPTICAL ATOMIC SPECTROSCOPY (Chapter 8) INTRODUCTION TO OPTICAL ATOMIC SPECTROSCOPY (Chapter 8) Atomic spectroscopy techniques: Optical spectrometry Mass spectrometry X-Ray spectrometry Optical spectrometry: Elements in the sample are atomized

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information