New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis
|
|
- Bryan Skinner
- 6 years ago
- Views:
Transcription
1 New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis Alessandro Santato, 1 Doug Hamilton, 1 Jan Wijbrans, 2 Claudia Bouman 1 1 Thermo Fisher Scientific, Bremen, Germany 2 VU University Amsterdam, Faculty of Earth and Life Sciences, Amsterdam, The Netherlands
2 Overview Purpose: We test the performance of new high gain amplifiers equipped with Ω feedback resistors. The tests are carried out on the multi-collector Thermo Scientific HELIX MC Plus noble gas MS. Methods: Circa 2.4 x mole of argon was analyzed using three different multi-collector modes and 40 Ar/ 36 Ar and 38 Ar/ 36 Ar was measured. Results: The new Ω amplifier show a decisive improvement in the measurement of small samples. Introduction On the Earth, noble gases are present as rare elements and in most of the cases their concentration within samples is extremely low. Therefore their analysis requires a high detection efficiency which implies ultra high vacuum systems, mass spectrometers able to operate in static mode, and detectors capable of reading small signals. Here we present a comparison between the latest amplifiers equipped with Ω resistors and the established state-of-art Thermo Scientific detector technology, represented by: amplifiers and the latest electron multiplier: Compact Discrete Dynode (CDD). Methods Sample Preparation and Analysis Argon from atmospheric air was prepared with the Thermo Scientific NG Prep System and used as standard sample for all the tests. Circa 2.4 x mol of argon were repeatedly analyzed using three different collector configurations. Collector Configurations Argon mass 40, 38 and 36 were analyzed using three different multi-collector modes, see figure 1. FIGURE 1. Scheme showing the three different multicollector settings used in this work. 2 New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis
3 40 Ar and 38 Ar were always analyzed, respectively: on a Faraday cup coupled with a Ω amplifier and on an electron multiplier CDD. The 36 Ar was analyzed with three different detector configurations: a Faraday cup coupled with the Ω amplifier, an electron multiplier CDD and a Faraday cup coupled with the latest Ω amplifier. Mass Spectrometry The Thermo Scientific HELIX MC Plus multi-collector static vacuum mass spectrometer is the ultimate solution for noble gas analysis. Equipped with a high-resolution magnetic sector analyzer and with a variable multi collector array, which incorporates 5 Combined Faraday Multiplier (CFM) detectors, and is capable of analyzing up to five isotopes of neon, argon, krypton and xenon simultaneously, at new levels of resolution. Data Analysis Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) was used for data acquisition and control software. Plots were made using Microsoft Office Excel Results The isotopic ratio of 40 Ar/ 36 Ar, using the Ω amplifier for 36 Ar, gave a value of 308 ± 0.90 and the standard error among the samples ranged from 1.3 to 3.09, whereas using the CDD configuration (for 36 Ar) the value of 40 Ar/ 36 Ar was ± 0.56 and a standard error among the samples ranged from 0.21 to With the new Ω amplifier, used for 36 Ar, we were able to achieve the following results: 40 Ar/ 36 Ar = ± 0.55 and a standard error among the samples ranged from 0.40 to (For this analysis the new Ω was not calibrated to the Ω). 38 Ar/ 36 Ar ratios were ± , ± and ± ; respectively for 36 Ar measured on: the Faraday cup with the Ω amplifiers, the CDD and the Faraday cup with the Ω amplifiers. Thermo Scientific Poster Note PN-SANTATO-GOLDSCHMIDT-2014-EN-0614S 3
4 FIGURE 2. Thermo Scientific HELIX MC Plus noble gas MS (right) and NG Prep System (left). FIGURE Ar/ 36 Ar AIR. Comparison among CDD/CDD, CDD/Faraday Ω amplifier and CDD/Faraday Ω amplifier. The Ω amplifier was not gain calibrated. This is why there is a bias in the absolute ratio. 4 New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis
5 Calibration of the New Amplifier with a Ω Resistor The Ω amplifier is by design 10 times more sensitive than the corresponding Ω amplifier. Given this increased sensitivity, at this time, it is not possible to calibrate the new Ω amplifier with the existing cross calibration system, as is routinely done on the Ω and Ω amplifiers fitted to the Helix MC Plus. A reference / standard should be used to calibrate the Ω amplifier. FIGURE Ar/ 36 Ar AIR. Comparison among CDD/CDD, CDD/Faraday Ω amplifier and CDD/Faraday Ω amplifier. The Ω amplifier was not gain calibrated. This is why there is a bias in the absolute ratio. Current amplifier Noise Decay Ohm 0.2 fta (20 4 s 10 ppm in 2 s Ohm 50 ata (5 4 s 100 ppm in 8 s FIGURE 5. Specifications of the Ω amplifier compared with the Ω amplifier. Thermo Scientific Poster Note PN-SANTATO-GOLDSCHMIDT-2014-EN-0614S 5
6 FIGURE 6. Complete resolution of 36 Ar from the hydrocarbon interference 12 C 3 and partial resolution from the H 35 Cl. High Resolution Mass Spectrometer One of the key features of the HELIX MC Plus noble gas MS is its high resolution capability. The instrument utilizes a large radius geometry and is capable to achieve a resolution > 1500 and a resolving power > 5000 (routinely >7000) when using the high resolution entrance slit. FIGURE 7. Resolution of the three interferences of 20 Ne (top) and resolution of 21 Ne from 20 NeH (bottom). 6 New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis
7 Conclusion The higher sensitivity of the new amplifiers equipped with a Ω feedback resistor represents a decisive improvement in small sample analysis. In the 40 Ar/ 36 Ar measurement the new Ω amplifier is capable of achieving a relative standard error between 0.13 and 0.44%, which represents a significant progress if compared with Ω amplifier (relative standard error is between 0.41 and 1%). The electron multiplier is the best detector in terms of sensitivity but its drift due to the variation of the yield is still a limiting factor. Future Work Further important steps in small sample analysis of noble gases will involve the development of an appropriate gain calibration technique for the new amplifiers as well as the improvement of the yield calibration for the electron multipliers during multi-collector analysis. Acknowledgements This project has received funding from the European Union s Seventh Framework Programme for research, technological development and demonstration under grant agreement no We gratefully thank Prof. Masahiko Honda from ANU Canberra for the two high resolution Neon scans. We would also like to thank the following person for their constant support: Soehnke Rathmann (test field engineer), Andreas Trint (test field engineer), Alexander Duhr (support noble gas MS), Peter Komander (electronic engineer), Michael Deerberg (R&D department) Thermo Fisher Scientific Inc. All rights reserved. Microsoft and Excel are trademarks of Microsoft Corp. All other trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa Australia Austria Belgium Canada China (free call domestic) Denmark Europe-Other Finland France Germany India Italy Japan Latin America Middle East Netherlands New Zealand Norway Russia/CIS Singapore Spain Sweden Switzerland UK USA PN-Santato-Goldschmidt-2014-EN-10/16S
of mass spectrometry
Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules
More informationARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision
ARGUS VI Static Vacuum Mass Spectrometer Static Vacuum ARGUS VI Multicollection Low Volume Precision Based on more than 20 years of experience in noble gas mass spectrometry instrumentation, we have developed
More informationHELIX MC Plus. Static Vacuum. ARGUS VI Thermo Scientific HELIX MC Plus Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer
ARGUS VI Thermo Scientific HELIX MC Plus Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX MC Plus Dynamic range Multi collection High mass resolution Based on more than
More informationHELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer
ARGUS VI HELIX SFT Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX SFT Low Volume Multicollection High Mass Resolution 3 Based on more than 20 years of experience in
More informationThermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples
Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples
More informationInsights Into the Nanoworld Analysis of Nanoparticles with ICP-MS
Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2
More informationA Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS
A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo
More informationImproved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer
Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher
More informationA Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry
A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1
More informationfor XPS surface analysis
Thermo Scientific Avantage XPS Software Powerful instrument operation and data processing for XPS surface analysis Avantage Software Atomic Concentration (%) 100 The premier software for surface analysis
More informationCharacterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer
Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy
More informationIntroduction to Fourier Transform Infrared Spectroscopy
Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR
More informationA Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry
A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,
More informationFitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap
Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Dirk Nolting and Andreas Wieghaus Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Characterize the rate
More informationIncreasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer
Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance
More informationExploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis
Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,
More informationComplementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites
Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:
More informationThermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS
Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier
More informationHR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput
HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer
More informationKey Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery
Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan
More informationEvaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use
Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,
More informationThermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3
Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques
More informationThermo Scientific 253 Plus 10 kv Isotope Ratio MS. Utilize the Gold Standard. in isotope ratio MS. Robustness Sensitivity Precision
Thermo Scientific 253 Plus 1 kv Isotope Ratio MS Utilize the Gold Standard in isotope ratio MS Robustness Sensitivity Precision Utilize the Gold Standard in Isotope Ratio MS 2 The Thermo Scientific 253
More informationIntroduction to Fourier Transform Infrared Spectroscopy
molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method
More informationHigh-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems
High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview
More informationActive Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency
Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Anthony Edge, 1 Luisa Pereira, 1 Dafydd Milton 1 and Andrew Shalliker 2 1 Thermo Fisher Scientific,
More informationThe Raman Spectroscopy of Graphene and the Determination of Layer Thickness
Application Note: 52252 The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Mark Wall, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words DXR Raman Microscope 2D Band D
More informationRapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis
Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Tim Stratton 1, Yingying Huang 1, Katianna Pihakari 1, Ian Acworth 2, and Michael Weber 2 1 Thermo Fisher Scientific,
More informationFast, Effective XPS Point Analysis of Metal Components
Application Note: 52297 Fast, Effective XPS Point Analysis of Metal Components Chris Baily and Tim Nunney, Thermo Fisher Scientific, East Grinstead, West Sussex, UK Key Words K-Alpha Auto-Analysis Multi-Spectrum
More informationMultiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer
Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,
More informationHILIC Method Development in a Few Simple Steps
HILIC Method Development in a Few Simple Steps Monica Dolci, Luisa Pereira, Dafydd Milton and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Overview This poster presents a systematic approach
More informationExploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications
Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology
More informationKey Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS
Nanoparticle Characterization Via Single Particle Inductively Coupled Plasma Mass Spectrometry (spicp-ms) Using a Dedicated Plug-in for Qtegra ISDS Software Technical Note 43279 Daniel Kutscher, Julian
More informationMetWorks Metabolite Identification Software
m a s s s p e c t r o m e t r y MetWorks Metabolite Identification Software Enabling Confident Analysis of Metabolism Data Part of Thermo Fisher Scientific MetWorks Software for the Confident Analysis
More informationAnalyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling
Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key
More informationPlasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method
Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,
More informationAccelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil
Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)
More informationXPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces
Application Note: 52287 XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces Brian R. Strohmeier, Thermo Fisher Scientific, Madison, WI, USA
More informationImproved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer
Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,
More informationThermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization
Thermo Scientific K-Alpha + XPS Spectrometer Fast, powerful and accessible chemical analysis for surface and thin film characterization X-ray Photoelectron Spectroscopy Quantitative, chemical identification
More informationDefining quality standards for the analysis of solid samples
Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples
More informationThermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis
Thermo Scientific Pesticide Explorer Collection Start-to-finish workflows for pesticide analysis Comprehensive Pesticide Analysis Solutions Pesticide Explorer Collection Selection table Lab Profile Routine
More informationUtility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery
Application Note: 3 Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Yoko Yamagishi, Thermo Fisher Scientific, C-2F 3-9
More informationImprovement of bulk analysis of environmental samples by using a multiple collector ICP-MS
Improvement of bulk analysis of environmental samples by using a multiple collector ICP-MS IAEA Safeguards Symposium - Vienna Amélie Hubert, Anne-Claire Pottin and Fabien Pointurier 20-24 OCTOBER 2014
More informationMonitoring Protein PEGylation with Ion Exchange Chromatography
Monitoring Protein PEGylation with Ion Exchange Chromatography Peter Yu, Deanna Hurum, Jinyuan (Leo) Wang, 2 Terry Zhang, 2 and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA; 2 Thermo Fisher Scientific,
More informationKey Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia
Variable Data-Independent Acquisition (vdia) Delivers High Selectivity and Sensitivity in Combined Targeted and Untargeted Analyses for Small Molecules Olaf Scheibner 1, Markus Kellmann 1, Charles Yang
More informationDirect Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology
Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology Maria C. Prieto Conaway, 1 Nicholas E. Manicke, 2 Marta Kozak 1
More informationDetermination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration
Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration Zhang Xuebin 1, Yu Chongtian 1, Liang Lina 1, Hans-Joachim Huebschmann 2, Thermo Fisher Scientific Shanghai, 1 China,
More informationIN QUALITATIVE ANALYSIS,
IN QUALITATIVE ANALYSIS, YOU VE ALWAYS HAD TO CHOOSE BETWEEN ACCURACY AND THROUGHPUT. NOW THERE S NO REASON TO ever have to. WATERS QUALITATIVE ANALYSIS solutions Waters ACQUITY UPLC System with the Waters
More informationApplicationNOTE Abstract
Abstract It has long been known that a range of non-native, stable, saturated hydrocarbon biomarkers present in crude oil extracts could be used to identify the original facie from which the oil was formed.
More informationAccurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS
APPLICATION NOTE 434 Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS Authors Introduction Marcus Manecki, Daniel Kutscher, Shona McSheehy
More informationThermo Scientific ELEMENT GD Glow Discharge Mass Spectrometer
m a s s s p e c t r o m e t r y Thermo Scientific ELEMENT GD Glow Discharge Mass Spectrometer Reset Your Expectations Part of Thermo Fisher Scientific Thermo Scientific ELEMENT GD Glow Discharge Mass Spectrometer
More informationThermo Scientific GasBench II
MASS SPECTROMETRY Thermo Scientific GasBench II Universal On-Line Gas Preparation and Introduction System for Isotope Ratio MS Product Specifications The Thermo Scientific GasBench II is an innovative
More informationFinnigan TC/EA. High Temperature Conversion Elemental Analyzer
m a s s s p e c t r o m e t r y Finnigan TC/EA High Temperature Conversion Elemental Analyzer Analyze Detect Measure Control Finnigan TC/EA High Temperature Conversion Elemental Analyzer Finnigan TC/EA
More informationMethod Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD
Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD Massimo Santoro, 1 Sergio Guazzotti, 1 Danilo Pierone, 2 Alexandre Souza, 3 Jaqueline Lorena,
More informationSimultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS
Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang
More informationThermo Scientific FLASH HT Plus Elemental Analyzer for IRMS
m a s s s p e c t r o m e t r y Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS Fully Automated Multi-Element Isotope Analysis for C, N, S, O, H Food Testing Forensics Environmental Analysis
More informationQuantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer
Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer Charles Yang and Dipankar Ghosh, Thermo Fisher Scientific, San Jose, CA Olaf
More informationImproving Intact Antibody Characterization by Orbitrap Mass Spectrometry
Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry Kai Scheffler, 1 Eugen Damoc, 2 Mathias Müller, 2 Martin Zeller, 2 Thomas Moehring 2 Thermo Fisher Scientific, Dreieich 1 and Bremen,
More informationNoble Gas Mass Spectrometer
Next Generation Excellence in Mass Spectrometry 600 Noble Gas Mass Spectrometer Full Multicollection, High Precision, High Resolution Next Generation Noble Gas Mass Spectrometry Noble Gas Mass Spectrometry
More informationComplete Materials Deformulation Using TGA-IR
Application Note: 51694 Complete Materials Deformulation Using TGA-IR Michael Bradley, Thermo Fisher Scientific, Madison, WI, USA Key Words Feedstock Supplies Finished Products Multi-component Search OMNIC
More information2017 Source of Foreign Income Earned By Fund
2017 Source of Foreign Income Earned By Fund Putnam Emerging Markets Equity Fund EIN: 26-2670607 FYE: 08/31/2017 Statement Pursuant to 1.853-4: The fund is hereby electing to apply code section 853 for
More informationSemi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS
Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS Kelly Munro, 1 Anthony Edge, 2 Claudia Martins, 3 David Cowan 4 and Leon Barron 1 1 Analytical & Environmental
More informationLow level Os isotopic measurements using multiple ion counting
APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité,
More informationThermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance
CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic
More informationQuantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection
Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,
More informationAgilent EEsof EDA.
Agilent EEsof EDA This document is owned by Agilent Technologies, but is no longer kept current and may contain obsolete or inaccurate references. We regret any inconvenience this may cause. For the latest
More informationHigh-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization
igh-resolution Accurate-Mass (RAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization Catharina Crone, 1 Markus Kellmann, 1 Yue Xuan, 1 Elizabeth Crawford 2 1 Thermo Fisher
More informationComparison of Solid Core HPLC Column Performance: Effect of Particle Diameter
Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Luisa Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20755 Key Words Solid core, fused core, superficially
More informationOvercoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES
Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Technical Note 43332 Key Words Overcoming Interferences, Element Finder plug-in Introduction Interferences are common in
More informationDetermination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production
Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo
More informationQuantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography
Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview
More informationQuick Guide QUICK GUIDE. Activity 1: Determine the Reaction Rate in the Presence or Absence of an Enzyme
Activity 1: Determine the Reaction Rate in the Presence or Absence of an 1. Find your 15 ml conical tubes labeled Stop Solution, 1.5 mm Substrate, and Buffer. Write your initials on each tube. 2. Label
More informationEnhancement of Linearity and Response in Charged Aerosol Detection
Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: This work
More informationTargeted protein quantification
Targeted Quantitative Proteomics Targeted protein quantification with high-resolution, accurate-mass MS Highly selective Very sensitive Complex samples HR/AM A more complete quantitative proteomics picture
More informationAn Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening
An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening Jonathan McNally, 1 Nicholas Duczak Jr., 1 Patrick Bennett, 1 Francois Espourteille, and Maciej
More informationDeterminations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection
Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview
More informationKeysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note
Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement Application Note Table of Contents Introduction... 3 S-parameter measurement uncertainty of VNA... 4 Simplification of systematic
More informationCharged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance
Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance David Thomas, Bruce Bailey, Marc Plante and Ian Acworth Thermo Fisher Scientific,
More informationFORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION
FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION A purpose-built collection of the best-inclass components for forensic toxicology Whether you re challenged to present reliable and secure forensic sample
More informationˆ GDP t = GDP t SCAN t (1) t stat : (3.71) (5.53) (3.27) AdjustedR 2 : 0.652
We study the relationship between SCAN index and GDP growth for some major countries. The SCAN data is taken as of Oct. The SCAN is of monthly frequency so we first compute its three-month average as quarterly
More informationHigh-Field Orbitrap Creating new possibilities
Thermo Scientific Orbitrap Elite Hybrid Mass Spectrometer High-Field Orbitrap Creating new possibilities Ultrahigh resolution Faster scanning Higher sensitivity Complementary fragmentation The highest
More informationPerformance and Control of the Agilent Nano Indenter DCM
Performance and Control of the Agilent Nano Indenter DCM Application Note Introduction With new materials and material applications come new test challenges. As these new challenges emerge in materials
More information2012 OCEAN DRILLING CITATION REPORT
2012 OCEAN DRILLING CITATION REPORT Covering Citations Related to the Deep Sea Drilling Project, Ocean Drilling Program, and Integrated Ocean Drilling Program from GeoRef Citations Indexed by the American
More informationThermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3
Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, and Improved Robustness of Data Christopher
More informationExport Destinations and Input Prices. Appendix A
Export Destinations and Input Prices Paulo Bastos Joana Silva Eric Verhoogen Jan. 2016 Appendix A For Online Publication Figure A1. Real Exchange Rate, Selected Richer Export Destinations UK USA Sweden
More informationKeysight Technologies Oxygen-Free High-Resolution Electrochemical SPM. Application Note
Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM Application Note Introduction For two decades, scanning probe microscopy (SPM) has provided scientists a unique tool to study in situ
More informationThermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials
Thermo Scientific ICP-MS solutions for the semiconductor industry Maximize wafer yields with ultralow elemental detection in chemicals and materials Trace impurity analysis in chemicals an used in the
More informationLC-MS/MS Method for the Determination of Diclofenac in Human Plasma
LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core
More informationTotal Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS
Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos
More informationCyclone Click to go to the page. Cyclone
Cyclone.0311.2 Click to go to the page Cyclone The Cyclone (1,000-8,000 CFM) Suitable for collection of virtually any type of dust Applications: Heavy Moulding Grinding Planing Cabinet Shops Dust Transfer
More informationSPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging
SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging Katerina Bousova, 1 Michal Godula, 2 Michele Suman 3 1 Thermo Fisher Scientific, Special Solutions
More informationA Platform Method for Pharmaceutical Counterion Analysis by HPLC
A Platform Method for Pharmaceutical Counterion Analysis by HPLC Xiaodong Liu, Mark Tracy, and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Results and Simple, fast, economical
More informationEnvironmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP
Application Note: 40986 Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP Karen Harper, Applications Group Leader, Thermo Scientific, Cambridge,
More informationScreening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS
Application Note: 514 Screening Method for 3 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS Yang Shi, Catherine Lafontaine, Thermo Fisher Scientific, Franklin,
More information2006 Supplemental Tax Information for JennisonDryden and Strategic Partners Funds
2006 Supplemental Information for JennisonDryden and Strategic Partners s We have compiled the following information to help you prepare your 2006 federal and state tax returns: Percentage of income from
More informationThermo Scientific Neptune XT MC-ICP-MS
Thermo Scientific Neptune XT MC-ICP-MS Neptune XT MC-ICP-MS A new edition of the market leading Thermo Scientific Neptune Series MC-ICP-MS, capturing the best of technology for high-precision isotope ratio
More informationA Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library
A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library Junhua Wang, 1 David A. Peake, 1 Robert Mistrik, 2 Yingying Huang 1 1 Thermo Fisher Scientific
More informationThermo Scientific icap TQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively
More informationThermo Scientific TRITON Thermal Ionization MS. Thermo Scientific NEPTUNE Multicollector ICP-MS. High Resolution Multicollector Mass Spectrometers
m a s s s p e c t r o m e t r y Thermo Scientific TRITON Thermal Ionization MS Thermo Scientific NEPTUNE Multicollector ICP-MS High Resolution Multicollector Mass Spectrometers Part of Thermo Fisher Scientific
More information