1 "Innovations in capillary gas chromatography for improving reproducibility, reliability and ease-of-use in the gas chromatographic separation of complex samples" Jaap de Zeeuw Varian, Inc.. Middelburg, The Netherlands
2 Challenges in the field of fragrance analysis Types of samples in fragrance analysis tend to be more complex and deal with a variety of component types; Many methods use retention index for identification.. Looking for lower level of impurities Analysis must become faster (cost per analysis) Need for better reproducibility Reduced investment in knowledge (black box) GC-Systems, software and columns must become more users friendly
3 Areas of innovation to look at Improvement of: Area to optimize Reproducibility Column manufacture & testing Reliability Stationary phase technology Ease of use Design of capillary columns
4 Why reproducibility? Predictability in GC analysis When replacing a capillary column, the NEW capillary column must: produce the similar retention times (phase-ratio) separate the components of interest (plate number) must show similar peak elution sequence (selectivity) acceptable peak asymmetry and response have low background and acceptable life time
5 Types of stationary phases used Non polar 5% phenyl / 95% methyl PDMS or arylene equivalent Good for general screening usually higher activity for polar compounds can be prepared with tight specifications Polar PEG type or stabilized equivalent highest selectivity more issues with selectivity
6 Fragrance/allergenes analysis, non-polar phases Oven : 50 C (1 min), 12 C/min -> 250 C (11 min) Column : 100% PDMS 50m x 0.25 mm x 0.25 um carrier gas : He, 25 psi, 1 ml/min Injection : Split, 1:15, 0.5 ul Detection : MS. Ion trap, Saturn Peak Identification 1 Benzyl alcohol 2 Phenyl acetadehyde 3 Limolene 4 Linalool 5 2,3-dichlorotoluene 6 methyl heptin carbonate 7 methylchavicol 8 citronellol 9 neral 10 unknown 11 cinnamaldehyde 12 hydroxy citronellal 13 methyl octin carbonate 14 cinnamyl alcohol 15 eugenol 16 methyl eugenol 17 coumarine 18 iso-eugenol 19 methyl gamma ionone 20 lilial 21 amyl cinnamyl aldehyde 22 lyral 23 amyl cinnamic alcohol 24 farnesol isomer I + II 25 farmsol isomer III 26 hexyl cannamic aldehyde (trans) 27 benzyl benzoate 28 benzyl salicylate 29 benzyl cinnamate 7 20 [min]
7 Flavor/Aroma compounds CP-Wax 57 CB 60 C, 2min -> 200 C, 2 C/min
8 Flavor/aroma compounds CP-Wax 58 (FFAP) CB 50 C -> 250 C, 7 C/min PEG -nitro-terephthalic acid ester Ideal for acidic compounds..
9 Allergenes Column : CP-Wax 52 CB, 30 x 0.32 mm X 0.2 um Ref: University Eindhoven
10 Optimise GC column manufacture & testing
11 Fused silica tubing Drawn at 2100 C Surface characterization Flexibility test Winding on column cages
12 Capillary Fused Silica Column Production Preform Oven OD - diameter control Primary coatcup
13 Testing the strength of fused silica in -X and -Y direction Every meter is leaded through the bending-test Eliminates presence of micro cracks
14 Surface silanol groups dehydration 150 C >800 C Irreversible < 400 C Reversible
15 Total phase degradation Total Column Bleed = Σ surface Si-OH + phase Si-OH + hydrolization Si-OH surface Si-OH : depend on column length and ID phase Si-OH : depend on total amount of stationary phase hydrolization Si-OH : depend on system & carrier gas For thin films the fused silica surface becomes dominant
16 Column preparation GC capillary columns are prepared by length.. 10, 15, 25, 30, 50, 60,100 or 150 meters.. Shorter columns cannot be made reproducible by cutting long columns in shorter lengths If you cut a long column in smaller parts there will be higher deviation on selectivity, retention and efficiency For highest reproducibility, columns must be made by a defined length..
17 Pretreatment of fused silica tubing Fused silica tubing often contains acid HNO3 and HCl are formed during drawing process Flushing with gas removes the acid Note: Flushing of non-deactivated fused silica surfaces can cause microcrack-formation
18 Fused silica surface characterization Intermediate test decanoic acid-methyl ester C10 25 m fused silica, untreated 130 C ca 0.2 ng on the column 2,6-DMA nonanal 1-octanol Surface = silica = adsorbent Retention factors on the surface reveil silanol interactions
19 Deactivation of silica surface Chemical process:reaction of Si-OH with reagent Octamethyl-tetrasiloxane (D4); Diphenyl-tetramethyl-disilane(DPTMDS), Hexamethyldisilane(HMDS), Hydrosilane Physical process: Carbowax Tri Ethanol Amine (TEA) covering activity of surface (shielding)
20 Static coating Vacuum Thermostrated Water-bath End-seal Static coating allows you to know exactly the amount stationary phase brought into the capillary..
21 Steps after coating the film of stationary phase.. Flushing/drying with high purity gas Polymerization (thermal, ideally 99+ %) Column rinsing (VERY clean solvents) Flushing/drying with high purity gas Conditioning at Tmax for > 10 hours, ultra pure gas Testing
22 Column testing Critical test mixture Isothermal analysis Efficiency Nth Inertness As Selectivity RI Retention k Stability bleed Permeability dp (PLOT) All these parameters must be specified
23 Column testing A testmixture is run isothermally 1 decane 2 octanol 3 dimethyl phenol 4 octanoic acid me-ester 5 2,6-dimethyl phenol 6 2,6-dimethyl aniline 7 dodecane 8 decanol 9 tridecane 10 decanoic acid me-ester Permeability Inertness Selectivity Retention Temperature setting is VERY important.. Efficiency
24 Test laboratory The test temperature must be adjusted very accurately Measurement of temperature in different ovens must be done at the same position in the oven (near the actual position of the test column) See chimneys on each GC.. An absolute necessity to calibrate each oven for absolute temperature using Pt-100 sensor
25 Retention Index non-polar phases octanoic acid me-ester RI values standard phase (CP-Sil 8 CB, Dimensions: 30 m x 0.25 mm x 0.25 µm RI values using new technologies (VF-5ms, 3500 columns) Deviation in selectivity has been decreased by factor 2
26 Test for PEG columns Peak identification 1 5-nonanone 2 1-octanol 3 n-hexadecane 4 n-heptadecane 5 naphtalene 6 n-octadecane 7 2,6-dimethylaniline Problem with PEG columns is that RI measurement is difficult..
27 Inertness Inertness asymmetrie As = b/a For good inertness measurement: needs to be a crital component (alcohol, amine, acid); component must have some retention (k >1); absolute amount injected < 2 nanograms; Columns: 30 m x 0.25 mm df = 0.25 µm 100 C ; 60 kpa H2 VF-200ms As=1.42 X -200ms As=2.94 Y -200 As=3.59 Asymmetry factor for 1-decanol is the lowest for the VF-200ms
28 Improve reliability.. Optimise stabilization of Stationary phases
29 Steps to optimize 1 Reduce Stationary Phase breakdown process by using polymer technology: Stabilizing blocks Ladder technology Purification methods Long polymer chains Surface bonding Non aggressive initiators 2 Optimization of the Fused silica surface using defined conditions for drawing 3 Optimizing the conditioning steps high temperature conditioning using high purity carrier gas
30 The stationary phase breakdown process Bleed product formation and ions found in MS spectrum Back - eating mechanism Spect 1 BP 207 (2075=100%) Col CP 0049.SMS % min. Scan: 1954 Chan: 1 Ion: us RIC: % 50% % % m/z
31 In house synthesis of all stationary phases.. Apply newest stabilization technology High viscosity polymers for low-bleed and stable stationary phases
32 Stabilization of today s generation of ms phases Name Composition CH 3 VF-1ms 100 % methyl Si O CH 3 100% VF-5ms arylene/methyl modified O CH 3 Si CH 3 CH 3 Si O Si CH 3 CH 3 CH 3 X Y VF-Xms arylene/methyl modified CH 3 CH 3 CH 3 VF-35ms arylene/methyl modified O Si CH 3 O Si Si CH 3 CH 3 Si O 35% X Y VF-17ms arylene/methyl modified O CF 3 CH 2 CH 2 Si O CH 3 Si CH 3 CH 3 VF-200ms trifluoropropyl methyl modified C N (CH 2 ) 3 O Si VF-23ms high cyano modified (CH 2 ) 3 C N X
33 C Cyanopropyl/phenyl type phases have now also been developed showing improved stability.. O Si N (CH 2 ) 3 O CH 3 Si CH 2 86% Type 1701 Type %
34 Future development on column coating technology In-Situ deposition / preparation of phases Sol-gel technology: liquid Adsorbents: porous polymers, silica, carbon
35 Improving Ease of Use.. Simplify column Installation, operation and storage
36 Practical concerns with present GC columns Mounting of a pre-column, a retention gap or a transfer line is difficult Column coupling is difficult Column connectors cannot be fixated in the oven Inlet and outlet of column after usage, is difficult to recognize Multiple columns in a oven take a lot of space with risk of breakage The column dimensions and information is difficult to read Installation of column in injector/detector must be simplified
37 Simplification of Column Installation with EZ-GRIP Both ends are fixed I Aluminium EZ-grip allows writing with a pencil D Injector side Injector and detector sides fully identified Detector side 21
38 Improving Ease of Use.. Simplify Column coupling..
39 Problem areas of column coupling Guard columns/transfer lines have to be coupled Challenges: Do the right column cutting To get a leak-free connection To get this in an acceptable time Get the guard column on and off the column cage The connection must stay leak-free while in operation
40 Popular coupling system Quick-Seal/Press-fit Column Connector
41 Alternative connectors metal body & ferrule
42 Can we avoid Column Coupling? Yes.. The guard column and transfer lines can be integrated with the capillary using segment coating...
43 Segment Coating Segment Coating can be applied away from: The inlet section : Integrated Guard column The outlet section : Integrated Transfer- line There are different techniques to manufacture a segment coated capillary
44 Segment coated columns Removing stationary phase from existing coated capillary Only for non-bonded phases For CB phases must be done before polymerization Big risk for glogging impossible for thick films Very labour intensive Deposit stationary phase only in that part where we need it Can be done using existing coating technologies Principally for all phases and film thicknesses Relative fast to do Need to have good eyes (..)
45 Segment coating Static coating Vacuum Water-bath End-seal When coat front has reached this point, residue is pushed out.. Fill column and coating process starts here
46 Configuration of a segment-coated system Integrated guard or EZ-Guard, typically 5-10 meters Integrated transfer line outlet, approx. 50 cm
47 Having an (un)coated Inlet The integrated (or EZ-)guard column is realized by not-coating the stationary phase on the first meters of the capillary.. Advantages: There is no coupling involved and column cutting is not critical; No issues with dead volume, thermal capacity and fragility; No hassle in getting the guard column on and off the column cage; Absolute no risk of leakage;
48 Properties of EZ-Guard columns Low retention for optimal refocusing effect in the analytical column High inertness for good peak shape and recovery Organic solvent resistant Inertness can be an issue..
49 Comparison of Integrated-Guard Activity Mixture, 4 ng per compound on the column
50 Comparison of columns Traces Pesticide analysis
51 Segment coating: uncoated transfer lines That part of the capillary column that is positioned in the transfer line and/or in the detector, will not contribute to the chromatography as it s always at high temperature.. Stationary phase in this part of the capillary can only decompose and produce breakdown products: a potential source for increased background..
52 Having an (un)coated Outlet The segment coating also allows the possibility to leave the last meters of the column without stationary phase: Integrated transfer line Advantages: There is no coupling involved and column cutting is not critical; Its fast operate; No issues with dead volume, thermal capacity and fragility; Absolute no risk of leakage; Faster stabilization in ms and other detection systems Lower background
53 Uncoated transfer line Especially important in: Sensitive detection systems (MS, PDD, ECD,..) Higher detector temperatures Oxygen & moisture sensitive phases When applying THICKER films with GC and GC/MS
54 Impact of detector temperature Oven : 100ºC iso thermally Column : VF-1ms with integrated transfer line Detector : FID, Temperature increased with steps of 30 ºC 360 ºC 330 ºC Without integrated transfer line 300 ºC With integrated transfer line 270 ºC Especially > 300C the impact becomes visible; As velocity at the end of the column is high, stabilization time (conditioning) will be relatively short..
55 To show the impact of the transfer line visible we tested this also with ms and took a capillary with a thick film: Column : 60 m x 0.32mm VF-5ms, df= 3 µm Detector: Saturn 2000 ion trap One column was WITH - and one column WITHOUT integrated transferline.. Both columns were treated exactly the same way and tested in the same instrumentation..
56 Effect of uncoated transferline [kcounts] 500 s Column : 60m X 0.32mm VF-5ms, df=3.0um Column flow : 1.5 ml/min, helium Temp Injector: 250 C Detector : Saturn 2000; Trap: 150 C; Manifold: 40 C Transferline : 180 C Temp prog : 50 C(5min) to 250 C(5min), 25 C/min RIC all sms 400 WITHOUT Integrated transferline Lower offset after temperature ramp s R I C a ll s m s WITH Integrated transferline 100 Lower offset Lower background
57 Ionization times Ionization time in ion trap ms is a measure for the number of ions present High value few ions (clean) Low value many ions (dirty/overload)
58 Ionization Times 50 C 25 C/min 250 C 50 C
59 Stabilization profile Oven Column : 50 ºC 250 ºC, 10 ºC/min : 30 m x 0.32mm VF-1ms, 3.0 um Without integrated transfer line With integrated transfer line Stable in approx. 40 minutes Still stabilizing [min]
60 Transfer line in MS Transfer lines can be activated by exposure to oxygen and water: Upon maintenance in inlet section (vacuum sucks air through) when injecting sample and introducing an air plug with the sample headspace liquid injection How much stationary phase is present in this small section of the capillary?
61 Example A capillary column of 30 m x 0.25 mm x 0.25 µm Transfer line for ms = 20 cm long In this transfer line we have approx. 40 µg of stationary phase.. If all this phase is decomposed due to oxydation/hydrolization it will bleed into the detector, resulting in background, drift and contamination.. How long will this process take and what signal can such a small amount of stationary phase produce?
62 Reference values are available using calibrated FID as this detector is used for bleed measurements.. Decomposing 40 µg PDMS will produce a signal of 1 pa for 1400 hours Decomposing 40 µg PDMS will produce a signal of 10 pa for 140 hours Decomposing 40 µg PDMS will produce a signal of 100 pa for 14 hours The impact of a coated transfer line can therefore be significant Uncoated transfer line is preferred
63 Factors that potentially can initiate stationary phase breakdown at higher temperature: water oxygen aggressive compounds derivatization agents acidic / basic compounds reactive chemicals
64 Impact of water in carrier gas on phase degradation Carrier gas is contaminated with 200 ppm water; Column repeatedly programmed from: 50 C (10min), 20 C/min to 325 C(60min). After water exposure it takes a long time before the signal stabilizes.. 4 hours!!
65 Limitations of segment coating Diameter of uncoated section is the same as the analytical column No different deactivations can be used Upon cutting a part of the EZ-guard column, retention times will change and integration window has to be adjusted
66 Segment coating technologies will become more important for future generation of optimized GCcolumns: using thicker films transfer lines for direct heating hardware designs making miniaturized & integrated solutions, like wafers
67 Conclusions New technologies has resulted in more stable stationary phases that also show high reproducibility Reproducible GC columns can only be realized using accurate testing The use of segment coating allows integration of guard columns/transferlines offering many advantages No Leaks (no coupling) Protects analytical column EZ to install
68 Thank you for your attention
Selection of a Capillary
Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector
More informationSelection of a Capillary GC Column
Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer
More informationIntroduction and Principles of Gas Chromatography
Introduction and Principles of Gas Chromatography Jaap de Zeeuw Restek, Middelburg, The Netherlands Jaap.dezeeuw@restek.com Definition and Uses of Gas Chromatography GC Components and Types of Columns
More informationAgilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings
Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings Technical Overview Introduction Modern GC and GC/MS instruments are important analytical tools for accurate and
More informationUnderstanding the Capillary GC Column: How to Choose the Correct Type and Dimension
Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension Simon Jones Application Engineer Things to Consider Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?
More informationHow To Select the Correct GC Column. Simon Jones Application Engineer
How To Select the Correct GC Column Simon Jones Application Engineer Things to Consider Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What
More informationUltra-Inert chemistry for Trace Level Analysis
Ultra-Inert chemistry for Trace Level Analysis Cikui Liang, Ph.D. Challenges and Needs of Today s Laboratories Challenges Qualification/quantification of trace samples Keep instrument up and running Needs
More informationActivity in the FID Detection Port: A Big Problem if Underestimated
Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be
More informationAn Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at
More informationWater Injections in GC - Does Water Cause Bleed?
Water Injections in GC - Does Water Cause Bleed? Eberhardt Kuhn Applications Chemist April 4, 2001 Practical Advice and Useful Tips for the Analysis of Semivolatile Organics by GC and GC/MS 11:00 a.m.
More informationAnalysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)
PO-CON9E Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer) ASMS 0 TP7 Sanket Chiplunkar, Prashant Hase, Dheeraj Handique, Ankush Bhone, Durvesh
More informationGAS CHROMATOGRAPHY (GC)
GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and
More informationNew ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column
New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column The World's First Non-metal 5% Phenyl Phase GC Column Rated to 430 C * Specially processed for thermal stability up to 430 C A true
More informationChoosing the Correct GC Column Dimensions and Stationary Phase
Choosing the Correct GC Column Dimensions and Stationary Phase Daron Decker Chromatography Technical Specialist Page 1 Nothing is useless it can always serve as a bad example Custom Column: 150 m x 250
More informationUnderstanding Gas Chromatography
Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector
More informationTrajan SGE GC Columns
Trajan Scientific and Medical Trajan SGE GC Columns Trajan Scientific and Medical Our focus is on developing and commercializing technologies that enable analytical systems to be more selective, sensitive
More informationA New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column
A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column Competitive Comparison Authors Ngoc-A Dang and Allen K. Vickers Agilent
More informationGUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY
Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid
More informationSecrets of GC Column Dimensions
Secrets of GC Column Dimensions GC Columns and Consumables Simon Jones Application Engineer May 20, 2008 Slide 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution Equation Changes
More informationSpeakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine
Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge the slide window at any time by clicking on the Enlarge Slides button, located below the presentation
More informationGet Selective. By Jaap de Zeeuw
34 Get Selective Modern narrow bore columns have made chromatographers lazy when it comes to stationary phase selection. Here s how getting back to basics in gas chromatography by using selectivity can
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask
More informationGAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas
GAS CHROMATOGRAPHY Mobile phase is a gas! Stationary phase could be anything but a gas Gas Chromatography (GC) GC is currently one of the most popular methods for separating and analyzing compounds. This
More informationColumn Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010
Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 Page 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution
More informationIntroduction to Gas Chromatography
Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition
More informationSo Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist
So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist GC Columns Wall Coated Open Tubulars Liquid phase coated capillaries Internal Diameter 0.05 0.53mm Length 5m 100m Porous
More informationUltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation
UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation Technical Overview Introduction Inert flow path technology Modern GC and GC/MS instrumentation is an important analytical tool
More informationCharacterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation
Application Released: January 214 Application Note 522 Characterisation of allergens in cosmetics by GC GC TF MS with Select-eV variable-energy electron ionisation Summary This Application Note shows that
More informationChemistry Instrumental Analysis Lecture 28. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 28 Two types in general use: -packed (stationary phase) -open tubular or capillary determine selectivity and efficiency of the sample. Column Materials Column
More informationGC Instruments. GC Instruments - Sample Introduction
GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes
More informationThe Suite for Environmental GC Analysis
The Suite for Environmental GC Analysis SGE Environmental GC Columns Performance Selectivity Delivery Promise www.sge.com SGE Environmental GC Columns The Suite for Environmental GC Analysis SGE GC Columns
More informationEvaluation of Capillary Columns for General Performance Parameters
Evaluation of Capillary s for General Performance Parameters Application Authors Mitch Hastings, Eberhardt R. Kuhn and Allen K. Vickers Agilent Technologies, Inc 91 Blue Ravine Road Folsom, CA 95630 USA
More informationIntroduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length
More informationColumn Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation
Column Selection there is more to life than a boiling point column Jaap de Zeeuw Restek Corporation, The Netherlands Stationary Phase Selectivity Boiling Point versus Solubility Retention of a compound
More informationChapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications
Chapter 27: Gas Chromatography Principles Instrumentation Detectors Columns and Stationary Phases Applications GC-MS Schematic Interface less critical for capillary columns Several types of Mass Specs
More informationCH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian
CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603
More informationAnalyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column
Analyze Hydrocarbon Impurities in,3-butadiene with an Agilent J&W GS-Alua PT Column Application Note Energy and Chemicals Authors Yun Zou and Chunxiao Wang Agilent Technologies Shanghai Ltd Abstract The
More informationTheory and Instrumentation of GC. Chromatographic Parameters
Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this
More informationChapter 27: Gas Chromatography
Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid
More informationIntroduction to Capillary GC
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 3 November 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #14 Chromatography: Theory (Skoog, Chapt. 26, pp.674-693) (Harris, Chapt. 23) (641-664) David Reckhow
More informationPartitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.
Chromatography Various techniques for the separation of complex mixtures that rely on the differential affinities of substances for a gas or liquid mobile medium and for a stationary adsorbing medium through
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationSimultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS
Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Application Note Multi-detector Splitter Authors Charlie Spanjers and Andrew Jones Activated Research Company 7561
More informationBP1 PONA BPX1. GC Capillary Columns BP1 PONA and BPX1
GC Capillary Columns 100% Dimethyl Polysiloxane GC Columns and Applications ID (mm) Film Thickness (µm) Length (m) Temperature Limits ( C) Part No. 0.32 0.5 60-60 to 320/340 054069 0.32 1 60-60 to 320/340
More informationA New Web-Based Application for Modeling Gas Chromatographic Separations
A New Web-Based Application for Modeling Gas Chromatographic Separations Jaap de Zeeuw**, Rebecca Stevens*, Amanda Rigdon, Linx Waclaski*and Dan Li* *Restek Corporation, Bellefonte, PA, USA **Restek Corporation,
More informationSimultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID
Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.
More informationPacked Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents
No. 089 SEPARATION REPORT Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS Table of Contents 1. Introduction 1 2. Column Specification 1 3. Features of Packing Materials
More informationAnalysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application
Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies
More informationTHE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS
International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationGas Chromatography (GC)
Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53
More informationIntroduction to Capillary GC
Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2
More informationAnalytical Instrumentation
Gas Chromatography and Mass Spectrophotometry Environmental, food safety, toxicology, and forensic gas chromatography applications are more demanding than ever. Our portfolio of GC, GC- MS, GC-MS/MS and
More informationThe Importance of Area and Retention Time Precision in Gas Chromatography Technical Note
The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which
More informationVolatile organic compounds (VOCs):
Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined
More informationChromatography and Functional Group Analysis
Chromatography Chromatography separates individual substances from a mixture. - to find out how many components there are - to match the components with known reference materials - to use additional analytical
More informationAgilent J&W GC Column Selection Guide
Agilent J&W GC Column Selection Guide Rely on unsurpassed reproducibility, efficiency, and inertness. Speed your selection with this one-stop resource. Agilent J&W GC Column Selection Guide Table of Contents
More informationCHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives:
CHEM 3281 Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract Objectives: The objectives of this experiment are: 1. to evaluate the fundamental parameters of gas chromatography; 2. to identify
More information2401 Gas (liquid) Chromatography
2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic
More informationGC Instruments. GC Instruments - Columns
GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes
More informationGas Chromatography. Introduction
Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationPackings for HPLC. Packings for HPLC
Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.
More information637. Thiamethoxam. HPLC method
637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1
More informationGas chromatography. Advantages of GC. Disadvantages of GC
Advantages of GC Gas chromatography Fast analysis, typically minutes Effi cient, providing high resolution Sensitive, easily detecting ppm and often ppb Nondestructive, making possible on - line coupling;
More informationAN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION
AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION Paper # 164-8P Pittsburgh Conference 24 T. Wampler, C. Zawodny, L. Mancini CDS Analytical, Inc 465 Limestone Road, Oxford,
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from
More informationOptimize Your Gas Chromatography Analysis Through the Use of a Retention Gap and Integrated Guard Column
WHITE PAPER Optimize Your Gas Chromatography Analysis Through the Use of a Retention Gap and Guard Column Authors Michel Pampaloni France Kyle Oberndorfer Shelton, CT Guy Gaspar Consultant France Introduction
More informationGas Chromatography (Chapter 2 and 3 in The essence of chromatography)
Gas Chromatography 1. Introduction. Stationary phases 3. Retention in Gas-Liquid Chromatography 4. Capillary gas-chromatography 5. Sample preparation and injection 6. Detectors (Chapter and 3 in The essence
More informationRapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction
Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns Application Food, Flavors and Fragrances Authors Mark Sinnott and Simon Jones Agilent Technologies, Inc. 9 Blue Raving Road
More informationCharacterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation
JSB is an authorised partner of Application Note 522 Application Released: January 214 Characterisation of allergens in cosmetics by GC GC TF MS with Select-eV variable-energy electron ionisation Summary
More informationChromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed
Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Gas Chromatography (GC) In gas chromatography, the sample is vaporized and injected onto the head of a chromatographic
More informationGas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5
Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible
More informationGC-CI-MS analysis of TMS derivatives
GC-CI-MS analysis of TMS derivatives This method describes analysis of TMS derivatives with methane chemical ionisation (CI) rather than the more normal electron impact (EI) ionisation. Methane CI is a
More informationGas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid
Gas Chromatography Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Principles Instrumentation Applications 18-1 Retention Volumes Volumes rather than times
More informationApplication Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS
Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure
More informationIndustrial Applications with a New Polyethylene Glycol-Based GC Column
Industrial Applications with a New Polyethylene Glycol-Based GC Column Application Note Authors Taylor Hayward, Yujuan Hua, Ronda Gras,, and Jim Luong, Dow Chemical Canada ULC PO BAG 6, Highway, Fort Saskatchewan,
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 25157 First edition 2013-04-15 Essential oil of rose, Chinese Kushui type (Rosa sertata Rosa rugosa) Huile essentielle de rose, type chinois Kushui (Rosa sertata Rosa rugosa)
More informationMETHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010
METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010 BIOCHEMISTRY The MEK is presented as a colorless liquid odor similar to that of acetone, is volatile and potentially explosive. The
More informationNew Product Update. HPLC Columns
New Product Update HPLC Columns ZORBAX MicroBore HPLC Columns ZORBAX Extend-C18 Columns ZORBAX Bonus-RP Columns Inertsil ODS-2 Cartridge Columns ZORBAX Carbohydrate Analysis Columns ZORBAX Eclipse dsdna
More informationChem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS
Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable
More informationTheory and Instrumentation of GC. GC Columns
Theory and Instrumentation of GC GC Columns i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.
More informationSlightly Polar Columns
9000 Варна, ул.поп Харитон 47 e-mail : gea99@abv.bg www.geya99.com Slightly Polar Columns With the partnership of Teknokroma SLIGHTLY POLAR COLUMNS TRB-5 It is the most universal and versatile stationary
More informationAnalysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System
Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road
More informationBiochemistry. Biochemical Techniques HPLC
Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know
More informationTRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110
TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110 BIOCHEMISTRY Trichlorethylene, marketed under the name of thinner, is used as a degreaser of metals, preparation of paints and cleaning of fabrics.
More informationProduct Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas
Agilent Porous Polymer PLOT Columns: New Products, Expanded Uses, Prices Cut in Half! Product Brief Need improved resolution of small volatile compounds? Didn't try a PLOT column due to high price, short
More informationMethod for the determination of 1,3-butadiene
Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt
More informationCHAPTER 6 GAS CHROMATOGRAPHY
CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1
More informationTETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010
TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010 BIOCHEMISTRY The Tetrachlorethylene or perchlorethylene, is an organic halide. Its structure is similar to that of a molecule of ethene which
More informationAnalysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System
Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System Application Note Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 Abstract
More informationOptimizing GC Parameters for Faster Separations with Conventional Instrumentation
Optimizing GC Parameters for Faster Separations with Conventional Instrumentation Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 243 Key Words TraceGOLD fast GC analysis
More informationMass Spectrometry. Fundamental GC-MS. GC Considerations
Mass Spectrometry Fundamental GC-MS GC Considerations i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
More informationDetermination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)
Method OIV-MA-AS315-16 Type IV method Determination of releasable in wine by cork stoppers (Resolution OIV-Oeno 296/2009) 1 SCOPE: The method of determination of releasable (TCA) by cork stoppers measures
More informationResidual solvents analysis using an Agilent Intuvo 9000 GC system
Residual solvents analysis using an Intuvo GC system Technology advantage: simplified dual column analysis with Flow Chip modularity Introduction USP defines a method for analysis of residual solvents
More informationHighly Sensitive and Rugged GC/MS/MS Tool
Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction
More informationGas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC
Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous
More information