GC-CI-MS analysis of TMS derivatives

Size: px
Start display at page:

Download "GC-CI-MS analysis of TMS derivatives"

Transcription

1 GC-CI-MS analysis of TMS derivatives This method describes analysis of TMS derivatives with methane chemical ionisation (CI) rather than the more normal electron impact (EI) ionisation. Methane CI is a soft ionisation technique that leads to less fragmentation than EI. Thus electron impact rarely gives molecular ions, whereas methane CI gives molecular ions (MH + ) for most small analytes. While CI is often touted as a means of obtaining molecular weight, this is not its only advantage. Its other advantage is that it gives different fragmentation patterns than EI, and thus CI may be able to distinguish between compounds that are essentially identical in EI (and the reverse is also true). Thus methane CI is useful for confirming tentative IDs made with EI. Unfortunately there are some drawbacks to methane CI. There are no large libraries of spectra like there are for EI. Unlike EI, CI conditions are not easily standardised and thus not so readily transferable among instruments. The other problem is that to obtain abundant (pseudo)molecular ions it is necessary to use a rather low ion source temperature. This low temperature leads to tailing of high-boiling compounds and causes fouling of the ion source. For these reasons methane CI of TMS derivatives will not replace EI for routine analysis methane CI s use is in confirming IDs, not routine analysis of hundreds of compounds. Hardware GC liner: splitless Column: RXi-5Sil-MS (30 metre x 0.25 mm ID x 0.25 µm film with 10 m guard column) MS: CI ion source (don t confuse it with the similar looking NCI) GC Carrier gas: 1 ml min -1 He (constant flow) Purge flow: 5 ml min -1 Injection: µl splitless (30 second sampling time, 15:1 split ratio) Injection temp: 270 C (250 can be used if worried about degradation of analytes) Oven program: 2 min at 70 C; 6 or 8 C min -1 ramp to 330 C; 8 minute hold at 330 C MS

2 FWHM 1 : 0.4 u Reagent gas: methane ( %) Reagent gas pressure 2 : 300 kpa (regulator pressure) Ion source 3 : 150 C Interface: 280 C Solvent delay: 6.0 min MS program: 6.2 to 41 min, scan from 65 to 1000 amu at 0.33 sec per scan MS program: turn off reagent gas at 40 minutes Threshold: 150 Detector: 0.15 higher than tune u is the default, but 0.4 u gives better resolution and only decreases sensitivity by a small amount kpa gives reasonably abundant MH + for most small molecules (e.g. MH + for ribitol is 20% of base peak). Higher pressure gives more abundant MH + but decreases sensitivity 3 This temperature is rather low and leads to peak tailing for compounds with boiling points > 200 C. However, using higher temperatures leads to a dramatic drop in MH + (e.g. at 200 C MH + for ribitol is 4.5%, whereas at 150 C it is 20%)

3

4

5 Samples and standards Each batch of samples should contain at least one blank and one retention index standard Retention index standards: even n-alkanes from C10 to C40 Run an alkane mix with every batch of samples. Update retention index file if necessary Internal standard: 0.2 mg/ml ribitol in MeOH:H2O (50:50) Generally add 5 µl to each sample prior to derivatization Amino acid, organic acid, sugar and sugar alcohol solutions: Make stock solutions at 5 or 1 mg/ml in MeOH:H2O (50:50). Remember to account for HCl, HCl.H2O and Na salts Asp, Glu and Asn should be dissolved in 0.1 M HCl (MeOH:H2O)

6 Tyr should be dissolved in 0.1 M NaOH (MeOH:H2O) Take a subsample of the stock solution and dilute to 0.2 mg/ml For peak ID confirmation, 1 or 2 µg of standard is appropriate mass of analyte (µg) conc (mg/ml) vol (µl)

7 Peak ID and quantification Peak ID and quantification is based on retention index (RI), a private mass spectral library (CRW_METAB_CI) and monoisotopic masses of target compounds. For TMS derivatives, methane CI commonly gives the pseudomolecular ion (MH + ) which may then be compared with monoisotopic masses (M) of different compounds. Beware that identifying MH + requires some intelligence and MH + may not be present for some analytes (e.g. di- and tri-saccharides). As a starting point for simple library comparisons for the private library use RI ± 5 and mass spectral match >80%. Note that RI of Golm and Fiehn libraries do not match exactly those determined here (RXi5-Sil-MS with 8 C/min ramp) because they used different columns and temperature ramps. However, there is a very strong and reproducible relationship between experimentally determined RI (RI_CRW) and those quoted in Golm and Fiehn (R 2 >0.999): RI Golm Quad = RI_CRW ; R² = RI Fiehn = RI_CRW ; R² = NIST RIs are normally quoted for isothermal analyses with standard non-polar columns and thus don t match particularly well with conditions used here. However, in some cases the NIST library has multiple RI entries for the one compound, and thus it may be possible to find a RI determined with a 5% column and similar temperature ramp. Note that the instrument s software displays only a single entry for NIST RIs. To check for multiple RIs you ll need to use NIST MS Search V2.0. Note that the RIs reported here and in the private library are only appropriate for these chromatographic conditions. They will change if you use a different column or temperature ramp or injection method. For example, RIs determined with 1mL/min flow and 8 C/min ramp are up to 10 units different to RIs determined with 2mL/min flow and 15 C/min ramp

ANALYTICAL REPORT. 25B-NBOH (C17H20BrNO3) 2-({[2-(4-bromo-2,5-dimethoxyphenyl)ethyl]amino}methyl)phenol RSUNJYKZRKIBNB-UHFFFAOYSA-N

ANALYTICAL REPORT. 25B-NBOH (C17H20BrNO3) 2-({[2-(4-bromo-2,5-dimethoxyphenyl)ethyl]amino}methyl)phenol RSUNJYKZRKIBNB-UHFFFAOYSA-N Vodovodna 95 1000 Ljubljana SLOVENIJA T: +386 (0)1 428 44 93 Remark other active cpd. detected: none Sample ID: 1700-16 Sample description: Sample type: ANALYTICAL REPORT 25B-NBOH (C17H20BrNO3) 2-({[2-(4-bromo-2,5-dimethoxyphenyl)ethyl]amino}methyl)phenol

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Method of determination of phtalates in spirituous beverages by gaschromatography/mass OIV-MA-BS-33 Method of determination of phtalates in spirituous beverages by gas-chromatography/mass Type IV method 1. SCOPE OF APPLICATION This method applies to the detection and assay of some phthalates

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

Determination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S.

Determination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S. Determination of Elemental Sulfur in Explosives and Explosive Residues by Gas Chromatography-Mass Spectrometry Kimberly S. Bradley Illinois State Police, Springfield Forensic Science Laboratory 24 Hill

More information

Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry

Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry 1 EM and KDE 1.1 E-step In the E-step, we deal with

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Introduction to Capillary GC

Introduction to Capillary GC ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs) Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs) The following 63 mass spectra were measured by gas chromatography (GC) (6890N, Agilent

More information

2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005

2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005 2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005 1. SYNONYMS CFR: CAS #: Other Names: Not Available Base: Not Available Hydrochloride: 69587-11-7 4-Iodo-2,5-dimethoxyphenethylamine 4-Iodo-2,5-dimethoxybenzeneethanamine

More information

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins TECHNICAL NOTE 39 A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins Author Cristian Cojocariu Thermo Fisher Scientific, Runcorn,

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist;

More information

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure

More information

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions GAS CHROMATOGRAPHY MASS SPECTROMETRY Pre-Lab Questions Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions are

More information

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System DiscovIR-GPC DiscovIR-GC Deposition and Detection System Application Note 41 Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System ABSTRACT The DiscovIR-GC

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

Identifying Disinfection Byproducts in Treated Water

Identifying Disinfection Byproducts in Treated Water Identifying Disinfection Byproducts in Treated Water Jonathan D. Byer, LEC Corporation; Saint Joseph, Michigan USA Susan D. Richardson, University of South Carolina; Columbia, South Carolina USA 1. Introduction

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length

More information

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD) Introduction There has been an increase in concern for the detection and removal of organic chloride species from crude aromatic, naphtha, and other hydrocarbon streams. One process called catalytic reforming

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

Analyzing Semi-Volatiles by GC-MS

Analyzing Semi-Volatiles by GC-MS Analyzing Semi-Volatiles by GC-MS Jonathan A. Karty July 31, 2008 Overview GC-MS Instrumentation Description GC-MS Method Sections In Detail Sequence Programming Data Analysis Tips and Tricks 1 Quick Description

More information

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at

More information

U.S. EPA Method 8270 for multicomponent analyte determination

U.S. EPA Method 8270 for multicomponent analyte determination ENVIRONMENTAL application note U.S. EPA Method 8270 for multicomponent analyte determination Elaine A. LeMoine and Herman Hoberecht Introduction Multicomponent analytes are compounds that yield several

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS, GC-MS, Pegasus BT, HS-SPME,

More information

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands Chem F413W Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands Introduction Gasoline is a complex mixture of volatile

More information

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Application Note Authors Angela Smith Henry and Bruce Quimby Agilent

More information

GAS CHROMATOGRAPHY (GC)

GAS CHROMATOGRAPHY (GC) GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities

More information

Gas Chromatography. Introduction

Gas Chromatography. Introduction Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be

More information

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns Richard Whitney, Laura Chambers, Clifford Taylor Shimadzu Scientific Instruments, Inc. Columbia, MD Spotlight on Method 8270 Instrumentation

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

BENZYLPIPERAZINE Latest Revision: June 1, 2005

BENZYLPIPERAZINE Latest Revision: June 1, 2005 BENZYLPIPERAZINE Latest Revision: June 1, 2005 HN N 1. SYNONYMS CFR: Schedule I CAS #: 2759-28-6 Stride II: Other Names: N-Benzylpiperazine 1-Benzylpiperazine N-Benzylpiperazine 1-(phenylmethyl)piperazine

More information

Theory and Instrumentation of GC. Chromatographic Parameters

Theory and Instrumentation of GC. Chromatographic Parameters Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

Analysis of complex petrochemicals by GC GC TOF MS with Select-eV variable-energy electron ionisation

Analysis of complex petrochemicals by GC GC TOF MS with Select-eV variable-energy electron ionisation JS is an authorised partner of pplication Note 54 nalysis of complex petrochemicals by GC GC TOF MS with Select-eV variable-energy electron ionisation Summary This pplication Note shows that enchtof instruments,

More information

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA

More information

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe A New Web-Based Application for Modeling Gas Chromatographic Separations Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe The Method Development Process in GC $ SUCCESS! Developing

More information

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds

More information

Pittcon P

Pittcon P PO-CON1701E 76% Increase in Throughput for Determination of Semi-Volatiles Using Narrow-Bore GC Columns and Rapid Data Acquisition with a Highly Sensitive Quadrupole GCMS System Pittcon 2017 1440-14P Brahm

More information

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation Electronic Supplementary Material (ESI) for Chemical Communications. This journal is The Royal Society of Chemistry 2015 Content: Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Foods, Flavors & Fragrances Applications A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates By Corby Hilliard; Amanda Rigdon; William Schroeder*,

More information

Introduction to Capillary GC

Introduction to Capillary GC Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2

More information

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005 1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005 HN O N 1. SYNONYMS CFR: N/A CAS #: Base: 35386-24-4 Hydrochloride: 5464-78-8 Other Names: 2-MeOPP 2-Methoxyphenylpiperazine 2. CHEMICAL AND

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation Application Released: January 214 Application Note 522 Characterisation of allergens in cosmetics by GC GC TF MS with Select-eV variable-energy electron ionisation Summary This Application Note shows that

More information

TO-15 Checklist Determination of VOCs in Air by GC-MS

TO-15 Checklist Determination of VOCs in Air by GC-MS LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision

More information

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques Application Note No. 085 The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques Diane Nicholas Multiple techniques using one instrument Simple

More information

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Application Note No. 033 Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Ryoichi Sasano*, Takayuki

More information

Method - Determination of aromatic amines in mainstream cigarette smoke

Method - Determination of aromatic amines in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of aromatic amines in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicaple to quantitative determination

More information

ANALYTICAL REPORT 1. 4-Me-TMP (C15H21NO2) Remark other NPS detected: none. sample contains some impurities (benzeneacetic acid + others)

ANALYTICAL REPORT 1. 4-Me-TMP (C15H21NO2) Remark other NPS detected: none. sample contains some impurities (benzeneacetic acid + others) Vodovodna 95 1000 Ljubljana SLOVENIJA T: +386 (0)1 428 44 93 E: nfl@policija.si www.policija.si Remark other NPS detected: none ANALYTICAL REPORT 1 4-Me-TMP (C15H21NO2) methyl-2-(4-methylphenyl)-2-(piperidin-2-yl)acetate

More information

Selection of a Capillary

Selection of a Capillary Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector

More information

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry Cristian Cojocariu and Paul Silcock Thermo Fisher Scientific, Runcorn,

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

GC-MS Screen for the Presence of Melamine, Ammeline, Ammelide and Cyanuric Acid

GC-MS Screen for the Presence of Melamine, Ammeline, Ammelide and Cyanuric Acid GC-MS Screen for the Presence of Melamine, Ammeline, Ammelide and Cyanuric Acid (Version 2.1) This method for Melamine, Ammeline, Ammelide and Cyanuric Acid should be regarded as interim. Because of the

More information

5-METHOXY- -METHYLTRYPTAMINE Latest revision: June 20, 2005

5-METHOXY- -METHYLTRYPTAMINE Latest revision: June 20, 2005 5-METHOXY- -METHYLTRYPTAMINE Latest revision: June 20, 2005 CH 3 O N H N CH 3 H H 1. SYNONYMS CFR: CAS #: Stride II: Other Names: Not Listed None 5-Methoxy-alpha-methyltryptamine 5-MeO-AMT -O-Dimethylserotonin

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

5-METHOXY-N,N-DIISOPROPYLTRYPTAMINE Latest Revision: June 20, 2005

5-METHOXY-N,N-DIISOPROPYLTRYPTAMINE Latest Revision: June 20, 2005 5-METHOXY-N,N-DIISOPROPYLTRYPTAMINE Latest Revision: June 20, 2005 CH 3 O N CH(CH 3 ) 2 CH(CH 3 ) 2 N 1. SYNONYMS H CFR: CAS #: Other Names: Not Listed None 3-[2-(Diisopropylamino)ethyl]-5-methoxyindole

More information

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TF Application Brief Authors Kevin Rowland 1, Mark Jordi 1, Kai Chen, and Jennifer

More information

Selection of a Capillary GC Column

Selection of a Capillary GC Column Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS POCON1813E Determination of 59 potential Allergens in Perfumes by twinline fast GCMSMS ASMS 2018 ThP 296 Xaver Mönnighoff 1, Christopher Sowa 2, HansUlrich Baier 1 1 Shimadzu Europa GmbH, Duisburg, Germany;

More information

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction Testing for GHB in hair by GC tandem quadrupole MS Marie Bresson, Vincent Cirimele, Pascal Kintz, Marion Villain; Laboratoire Chemtox, Illkirch, France Timothy Jenkins, Waters Corporation, Manchester,

More information

A New Web-Based Application for Modeling Gas Chromatographic Separations

A New Web-Based Application for Modeling Gas Chromatographic Separations A New Web-Based Application for Modeling Gas Chromatographic Separations Jaap de Zeeuw**, Rebecca Stevens*, Amanda Rigdon, Linx Waclaski*and Dan Li* *Restek Corporation, Bellefonte, PA, USA **Restek Corporation,

More information

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option

More information

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009) Method OIV-MA-AS315-16 Type IV method Determination of releasable in wine by cork stoppers (Resolution OIV-Oeno 296/2009) 1 SCOPE: The method of determination of releasable (TCA) by cork stoppers measures

More information

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer) PO-CON9E Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer) ASMS 0 TP7 Sanket Chiplunkar, Prashant Hase, Dheeraj Handique, Ankush Bhone, Durvesh

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Acknowledgement: 4F-PBP (C14H18FNO) T: +386 (0) E: Vodovodna Ljubljana SLOVENIJA

Acknowledgement: 4F-PBP (C14H18FNO) T: +386 (0) E:  Vodovodna Ljubljana SLOVENIJA Vodovodna 95 1000 Ljubljana SLOVENIJA T: +386 (0)1 428 44 93 E: nfl@policija.si www.policija.si 4F-PBP (C14H18FNO) Acknowledgement: Sample was kindly provided by the Laboratório de Polícia Cientifica da

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Due to rapid growth in the medical cannabis industry, demand is increasing for analysis of residual solvents in cannabis concentrates in order to protect consumer safety. This application note

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,

More information

ANALYTICAL REPORT 1 CUMYL-4CN-BINACA (C22H24N4O) 1-(4-cyanobutyl)-N-(1-methyl-1-phenylethyl)-1H-indazole-3-carboxamide

ANALYTICAL REPORT 1 CUMYL-4CN-BINACA (C22H24N4O) 1-(4-cyanobutyl)-N-(1-methyl-1-phenylethyl)-1H-indazole-3-carboxamide Vodovodna 95 1000 Ljubljana SLOVENIJA T: +386 (0)1 428 44 93 E: nfl@policija.si www.policija.si Remark other NPS detected: none ANALYTICAL REPORT 1 CUMYL-4CN-BINACA (C22H24N4O) 1-(4-cyanobutyl)-N-(1-methyl-1-phenylethyl)-1H-indazole-3-carboxamide

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Margaret Maziarz, Sean M. McCarthy, Michael D. Jones, and Warren B. Potts Waters Corporation, Milford, MA, USA A P

More information

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations OUTLINE OF METHOD This gas chromatography/mass spectrometry (GC/MS) method provides

More information

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

1. GENERAL INFORMATION IUPAC

1. GENERAL INFORMATION IUPAC AM-0 O N H C N. GENERAL INFORMATION IUPAC Name: {-[(-methylpiperidin--yl)methyl]-h-indol--yl}(naphthalen--yl) methanone CFR: Not Scheduled (0/0) CAS#: 76-5-7 Synonyms: Source: Appearance: Kovat's Index:

More information

3-Fluoromethcathinone Latest Revision: March 29, 2013

3-Fluoromethcathinone Latest Revision: March 29, 2013 3-Fluoromethcathinone Latest Revision: March 29, 2013 1. GENERAL INFORMATION IUPAC Name: 1-(3-fluorophenyl)-2-methylaminopropan-1-one CFR: Not Scheduled (3/2013) CAS #: 1049677-77-1 Synonyms: Source: Appearance:

More information

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note Authors Chris Sandy Agilent Technologies Wokingham, Berkshire, UK Jean-Francois Garnier Agilent Technologies

More information

Volatile organic compounds (VOCs):

Volatile organic compounds (VOCs): Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined

More information

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005 Gas Chromatography Chromatography Laboratory Course The laboratory course experiments General Aim: Gain general experience using a GC Constant Injection technique Temperature variations Qualitative and

More information

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK Multi-residue GC-MS analysis Richard Fussell CSL York, UK e-mail: r.fussell@csl.gov.uk Outline of Presentation Background Gas Chromatography Mass Spectrometry - Options Capabilities and Applications Summary

More information

NCFS Sample Preparation and Instrumental Parameters

NCFS Sample Preparation and Instrumental Parameters NCFS Sample Preparation and Instrumental Parameters April 1, 2010 Prepared by Dana-Marie Dennis and Jessica Frisch Addenda by Dana-Marie Dennis and Quashanna Price 1. MICROMETRY A stage micrometer is used

More information

637. Thiamethoxam. HPLC method

637. Thiamethoxam. HPLC method 637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1

More information

Distinct photoproducts of hydroxylated polybromodiphenyl ethers from different photodegradation pathways: A case of 2'-HO-BDE-68

Distinct photoproducts of hydroxylated polybromodiphenyl ethers from different photodegradation pathways: A case of 2'-HO-BDE-68 Electronic Supplementary aterial (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 214 Supplementary Information Distinct photoproducts of hydroxylated

More information

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke British American Tobacco Group Research & Development Method - Determination of phenols in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicable to the quantitative determination of

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Mass Spectral Studies of Polypropylene Chromatographic Well Plates Mass Spectral Studies of Polypropylene Chromatographic Well Plates Brian King, Thermo Fisher Scientific, Runcorn, Cheshire, UK Detlev Lennartz, Thermo Fisher Scientific, Langerwehe, Germany White Paper

More information

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010 METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010 BIOCHEMISTRY The MEK is presented as a colorless liquid odor similar to that of acetone, is volatile and potentially explosive. The

More information