USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns
|
|
- Rudolf Hunt
- 6 years ago
- Views:
Transcription
1 USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns Richard Whitney, Laura Chambers, Clifford Taylor Shimadzu Scientific Instruments, Inc. Columbia, MD Spotlight on Method 8270 Instrumentation
2 Venting the MS to... USEPA Methods 8270 and 8260 on a Single GCMS Without Change Columns Richard Whitney, Laura Chambers, Clifford Taylor Shimadzu Scientific Instruments, Inc. Columbia, MD Spotlight on Method 8270 Instrumentation
3 Problem Small laboratory Limited budget Multiple applications No time The Big Bang Theory CBS.com 3
4 Solution 8260 and 8270 Single GCMS Two inlets, two columns No venting The Big Bang Theory CBS.com 4
5 What We Did Instrument configuration Optimize method parameters Analytical results (aka data) 5
6 Instrument Configuration 6 GCMS-QP2010 SE Production Workhorse
7 Instrument Configuration 7 GCMS-QP2010 SE Production Workhorse
8 Instrument Configuration (cont.) GCMS Twin Line Kit 8
9 Instrument Configuration (cont.) 9
10 Instrument Configuration (cont.) 10
11 Instrument Configuration (cont.) 58 L/sec (He) Pfeiffer TMP (Rotary pump 30 L/min) 11
12 Instrument Configuration (cont.) 12
13 Instrument Configuration (cont.) Ecology Mode saves ~26% 13
14 Instrument Configuration (cont.) Parameter EPA Method 8260 EPA Method 8270 Sample introduction Purge-and-trap Liquid syringe Liner Shimadzu multi-purpose split liner, no glass wool Shimadzu multi-purpose split liner, no glass wool Injection conditions Split 75:1 Split 10:1 Column Maximum column temperature Carrier gas MS pump Rxi-624Sil MS 20 m x 0.18 mm x 1.4 µm Rxi-5Sil MS 20 m x 0.18 mm x 0.18 µm Program = 320 C Program = 350 C Helium, constant linear velocity mode Single stage TMP, max flow 4 ml/min (He) MS interface 225 C Ion source EI, 200 C 14
15 Instrument Configuration (cont.) Born in Kyoto Made in the USA 15
16 Method Optimization Primary objectives 1. Keep total column flow rate below 4 ml/minute 2. Keep maximum oven temperature below 320 C 3. Keep GC run time as short as practical The Big Bang Theory CBS.com 16
17 Method Optimization (cont.) Parameter Column constant linear velocity Method 8260 Primary 45 cm/second ~0.8 ml/min Method 8270 Inactive 30 cm/second ~0.5 ml/min Maximum total column flow Oven program 35 C (5 min) 20 C/min to 220 C (hold 2.5 min) ~1.3 ml/minute NA GC run time 12.5 minutes NA 17
18 Method Optimization (cont.) Parameter Column constant linear velocity Method 8260 Inactive 30 cm/second ~0.5 ml/min Method 8270 Primary 40 cm/second ~0.7 ml/min Maximum total column flow Oven program NA ~1.2 ml/minute 45 C (0.5 min) 25 C/min to 315 C (hold 4.2 min) GC run time NA 16 minutes 18
19 Why Constant Linear Velocity? Van Deemter Plot Optimal LV for each carrier Gas Linear Velocity Isothermal Oven 19
20 Why Constant Linear Velocity? (cont.) Head Pressure (kpa) Flow (ml/min) Linear Velocity (cm/sec) Oven Temp 20 With temperature programming: 1. Viscosity of gas increases 2. Gas expands 3. Flow rate and LV drop
21 Why Constant Linear Velocity? (cont.) Constant Pressure Linear Velocity Head Pressure Flow Linear Velocity Linear Velocity Oven Temp 21
22 Why Constant Linear Velocity? (cont.) Constant Flow Linear Velocity Head Pressure Linear Velocity Oven Temp Flow Linear Velocity 22
23 Why Constant Linear Velocity? (cont.) Constant Linear Velocity Linear Velocity Head Pressure Flow Linear Velocity Oven Temp 23
24 Why Constant Linear Velocity? (cont.) Constant Pressure Linear velocity = 15 cm/sec Constant Flow Linear velocity = 20 cm/sec Constant Linear Velocity Linear velocity = 30 cm/sec 24
25 Analytical Results BFB Spectrum EPA Method 8260 BFB Tune Criteria Passed 25
26 Analytical Results (cont.) EPA Method µg/l calibration standard TIC 66 compounds resolved 26
27 Analytical Results (cont.) DFTPP Spectrum EPA Method 8270 DFTPP Tune Criteria Passed 27
28 Analytical Results (cont.) Pentachlorophenol Benzidine Benzidine and Pentachlorophenol Tailing DDT (expanded) DDE DDD DDT Breakdown 28
29 Analytical Results (cont.) Pentachlorophenol Benzidine Benzidine and Pentachlorophenol Tailing DDT (expanded) DDE DDD DDT Breakdown 29
30 Analytical Results (cont.) EPA Method µg/ml calibration standard TIC 66 compounds resolved 30
31 Analytical Results (cont.) EPA Method 8270 Calibration Range 6 Internal Standards Response Factor Linear Curve 8 points, µg/ml Acenaphthene-d10, Cyrysene-d12, 1,4-Dichlorobenzene-d4, Naphthalene-d8, Perylene-d12, Phenanthrene-d10 40 µg/ml each 54 compounds < 15% RSD Min. RF = 0.132, Pentachlorophenol Max. RF = 3.966, bis(2-chloroisopropyl)ether 12 compounds > 15% RSD Min r = 0.997, 2,4-dinitrophenol Max r = 0.999, 8 compounds 31
32 Analytical Results (cont.) r = Nitrophenol (2 µg/ml) 32
33 Analytical Results (cont.) RSD b = 6.1% RSD k = 12.0% 33 Benzo(b&k)fluoranthenes (1 µg/ml)
34 %RSD Analytical Results (cont.) 12.0 % Relative Standard Deviation (0.4 µg/ml*, n = 8) /66 RSD < 8% 39/66 RSD < 4%
35 Average % Recovery Analytical Results (cont.) Average % Recovery (0.4 µg/ml*, n = 8) 57/66 Recovery %
36 IDL (µg/ml) Analytical Results (cont.) 0.80 Statistical Instrument Detection Limit (0.4 µg/ml, n = 8) /66 IDL < 0.2 µg/ml
37 Analytical Results (cont.) DCM carryover < 1.0 ppb 37
38 Analytical Results (cont.) Acetone carryover ~ 1.0 ppb 38
39 Summary Two methods Two intro techniques Two injection ports Two columns One MS All method criteria passed One happy customer Shimadzu application note GCMS-1302 The Big Bang Theory CBS.com 39
40 Acknowledgements Richard R. Whitney, Ph.D. Configuration insight and all the hard work! Julie Kowalski Advice re column and conditions Columns and standards 40
41 Shimadzu Scientific Instruments 41 Products People - Passion
Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS
Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure
More informationDeveloping Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014
Developing Large Volume Injection (LVI) in Split / Splitless Inlets Philip J. Koerner Phenomenex Inc. NEMC 2014 Goal To produce a method that provides the largest response of analytes that will produce
More informationFast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application
Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville
More informationA Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes
A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationDetermination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS
PO-CON1609E Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS Pittcon 2016 1110-1 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole
More informationAnalysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph
Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9 Gas Chromatograph Application Note Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies,
More informationRoger Bardsley, Applications Chemist; Teledyne Tekmar Page 1
Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist;
More informationSystem and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.
ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination
More informationPittcon P
PO-CON1701E 76% Increase in Throughput for Determination of Semi-Volatiles Using Narrow-Bore GC Columns and Rapid Data Acquisition with a Highly Sensitive Quadrupole GCMS System Pittcon 2017 1440-14P Brahm
More informationMaximizing Sample Throughput In Purge And Trap Analysis
Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in
More informationSemivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column
Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column Application Note Environmental Author Doris Smith and Kenneth Lynam Agilent Technologies, Inc. 285 Centerville
More informationAnalysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401
Gas Chromatograph Mass Spectrometer No. GCMS-1401 Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS Introduction Development of methods for analysis of drugs of abuse has become a high
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask
More informationFaster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit
Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit Analysis times for semivolatile compounds can limit sample throughput and decrease overall lab productivity. You can significantly
More informationOptimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s
JSB is an authorised partner of Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s Anne Jurek #120 Introduction: The USEPA developed
More informationDetermination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry
PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole
More informationMeeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS
Application Note # CA 284780 Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Abstract The analysis of semi-volatile organic compounds
More informationINNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical
INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in
More informationAuthor. Abstract. Introduction
Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent
More informationSolid Phase Micro Extraction of Flavor Compounds in Beer
Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental
More informationAgilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings
Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings Technical Overview Introduction Modern GC and GC/MS instruments are important analytical tools for accurate and
More informationDetermination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode
PO-CON1610E Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode Pittcon 2016 1110-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole
More informationMulti-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms
Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms Application Note Author Laura Provoost Agilent Technologies, Inc. Introduction In multi-residue analysis different groups
More informationApplication Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution
Application Note Abstract Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution Page 1 Polycyclic Aromatic Hydrocarbons (PAHs) are a large group of organic compounds
More informationAnalysis of Fast Moving Consumer Goods (FMCG) using GCMS with static and dynamic headspace (HS)
PO-CON1789E Analysis of Fast Moving Consumer Goods (FMCG) using GCMS with static and dynamic headspace (HS) ASMS 2017 TP 280 Dheeraj Handique, Ankush Bhone, Durvesh Sawant, Prashant Hase, Sanket Chiplunkar,
More informationGC-CI-MS analysis of TMS derivatives
GC-CI-MS analysis of TMS derivatives This method describes analysis of TMS derivatives with methane chemical ionisation (CI) rather than the more normal electron impact (EI) ionisation. Methane CI is a
More informationOptimizing of Volatile Organic Compound Determination by Static Headspace Sampling
JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option
More informationDetermination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD
Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road
More informationQuantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor
Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities
More informationAnalysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.
Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection
More informationChemistry Instrumental Analysis Lecture 27. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)
More informationUsing Hydrogen as An Alternative Carrier Gas for US EPA 8260
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative
More information2017 EPA Method Update Rule and EPA Method 624.1
Introduction Method 624 is for the determination of volatile organic compounds in industrial discharges and other liquid environmental samples by gas chromatography combined with mass spectrometry (GC/MS).
More informationAnalysis of Low Level Volatile Organic Compounds in Air Anne Jurek
JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15
More informationAnalysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler
Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in
More informationA Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist
A Single Calibration for Waters and Soil Samples Performing EPA Method 8260 Anne Jurek Applications Chemist Abstract A Single Calibration Method for Water AND Soil Samples The United States Environmental
More informationApplication Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek
Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek
More informationUnderstanding Gas Chromatography
Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector
More informationIdentification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS
Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7B Triple Quadrupole GC/MS Application Note Authors Dale R. Walker and Fred Feyerherm Agilent Technologies, Inc.
More informationDetermination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC
Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,
More informationU.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS
ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.
More informationRtx -CLPesticides Columns Reduce Analysis Time; Reduce System Maintenance
Columns Reduce Analysis Time; Reduce System Maintenance 0 YEARS Turning Visions into Reality www.restek.com 00-- --00 and Fused Silica Columns Detect and confirm 0 common organochlorine pesticides simultaneously.
More informationApplication Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek
Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length
More informationOptimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications
More informationMeasuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS
application Note Gas Chromatography Author Heidi Grecsek PerkinElmer, Inc. Shelton, CT 06484 USA Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS Introduction
More informationDetermination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Kira. Yang PerkinElmer, Inc. Shanghai, China Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS Introduction
More informationValidation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS
Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement
More informationAnalysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70
Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds
More informationAnalysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler
No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing
More informationHelium conservation in volatile organic compound analysis using U.S. EPA Method 8260C
APPLICATION NOTE 10441 Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C Authors Andrea Caruso, 1 Tommaso Albertini, 1 Jacob A. Rebholz 2 ; 1 Thermo Fisher Scientific,
More informationSUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns
Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns L.M. Sidisky, Yizeng Ni, Greg Baney, Rodney George, Katherine K. Stenerson 01-0026 T401044 CFK Discussion
More informationU.S. EPA Method 8270 for multicomponent analyte determination
ENVIRONMENTAL application note U.S. EPA Method 8270 for multicomponent analyte determination Elaine A. LeMoine and Herman Hoberecht Introduction Multicomponent analytes are compounds that yield several
More informationAnalysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404
Gas Chromatograph Mass Spectrometer No. GCMS-1404 Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030 Introduction Contamination of consumer products with pesticides is a growing concern
More informationIntroduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationA New Web-Based Application for Modeling Gas Chromatographic Separations
A New Web-Based Application for Modeling Gas Chromatographic Separations Jaap de Zeeuw**, Rebecca Stevens*, Amanda Rigdon, Linx Waclaski*and Dan Li* *Restek Corporation, Bellefonte, PA, USA **Restek Corporation,
More informationSo Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist
So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist GC Columns Wall Coated Open Tubulars Liquid phase coated capillaries Internal Diameter 0.05 0.53mm Length 5m 100m Porous
More informationAccelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil
Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)
More informationUltra-Inert chemistry for Trace Level Analysis
Ultra-Inert chemistry for Trace Level Analysis Cikui Liang, Ph.D. Challenges and Needs of Today s Laboratories Challenges Qualification/quantification of trace samples Keep instrument up and running Needs
More informationA New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe
A New Web-Based Application for Modeling Gas Chromatographic Separations Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe The Method Development Process in GC $ SUCCESS! Developing
More informationSelection of a Capillary
Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector
More informationSecrets of GC Column Dimensions
Secrets of GC Column Dimensions GC Columns and Consumables Simon Jones Application Engineer May 20, 2008 Slide 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution Equation Changes
More informationA Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
JSB is an authorised partner of A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek #101 Introduction: The United States Environmental Protection Agency (USEPA)
More informationColumn Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010
Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 Page 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution
More informationHydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.
Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System Application Note Chemicals Author Bob Moyer Sr. Research Scientist, R&D Sasol USA, Performance Chemicals
More informationMethanol Extraction of high level soil samples by USEPA Method 8260C
Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed
More informationOptimizing GC Parameters for Faster Separations with Conventional Instrumentation
Optimizing GC Parameters for Faster Separations with Conventional Instrumentation Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 243 Key Words TraceGOLD fast GC analysis
More informationDetection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS
Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)
More informationValidation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler
Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is
More informationImproved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source
Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist
More informationMethod 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8
application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ
More informationSTANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS
PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationApplication News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS
HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationVOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography
VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography Introduction ASTM International publishes a number of different methods for the determination of volatile compounds in water-based coatings.
More informationAnalysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler
Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration
More informationANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.
Determination of Volatile Petroleum Hydrocarbons ANNE JUREK Abstract: Due to current events, the importance of determining volatile petroleum hydrocarbons in both soils and waters has become an issue.
More informationValidation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions
Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationDetermination of Volatile Organic Compounds in Air
JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.
More informationDetermination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface
Application Note No. 033 Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Ryoichi Sasano*, Takayuki
More informationA Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS
PO-CON1518E A Method for Simultaneous Analysis of 174 Pesticides Pittcon 2015 2220-12P Feifei Tian, Xizhi Wang, Shiheng Luo, Guixiang Yang, Taohong Huang, Shin-ichi Kawano, Yuki Hashi, Shimadzu Global
More informationAnalytical Trap Comparison for USEPA Method 8260C
Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component
More informationOptimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS
Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods
More informationActivity in the FID Detection Port: A Big Problem if Underestimated
Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be
More informationAnalysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)
PO-CON9E Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer) ASMS 0 TP7 Sanket Chiplunkar, Prashant Hase, Dheeraj Handique, Ankush Bhone, Durvesh
More informationValidation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the
More informationenvironmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)
environmental #59120 Applications note Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs) Polychlorinated biphenyls (PCBs) are
More informationValidation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System
Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method
More informationSelection of a Capillary GC Column
Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer
More informationOptimizing. Abstract: is standardd. procedures. altered to
Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation
More informationFast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS
PO-CON1731E Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS ASMS 2017 TP-295 Hui Xian Crystal Yeong 1, Stephany Olivia 2, Lai Chin Loo 1 1 Application
More informationEvaluation of a New Analytical Trap for Gasoline Range Organics Analysis
Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible
More informationQuantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS
PO-CON1408E Quantitative Analysis of Opioids Using a Pittcon 2014 1090-8P Laura Chambers, Richard Whitney, Ph.D., Nicole M. Lock, Zhuangzhi Max Wang, Ph.D., Clifford M. Taylor; Shimadzu Scientific Instruments,
More informationAnalysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application
Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies
More informationSensitive Detection of 2-MIB and Geosmin in Drinking Water
Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy
More informationA Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas
A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).
More informationThe Importance of Area and Retention Time Precision in Gas Chromatography Technical Note
The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which
More informationSTANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS
PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationUSP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument
Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these
More information