Manual X-ray fluorescence Manual remote experiment Project e-xperimenteren+

Size: px
Start display at page:

Download "Manual X-ray fluorescence Manual remote experiment Project e-xperimenteren+"

Transcription

1 Manual X-ray fluorescence Manual remote experiment Project e-xperimenteren+ J.M.Mulder, J. Snellenburg

2 Colofon Manual X-ray fluorescence Manual remote experiment Project e-xperimenteren+ Stichting Digitale Universiteit Oudenoord 340, 3513 EX Utrecht Postbus 182, 3500 AD Utrecht Phone Fax Author(s) J.M.Mulder, J. Snellenburg Copyright Stichting Digitale Universiteit De Creative Commons Naamsvermelding-GeenAfgeleideWerken-NietCommercieel-licentie is van toepassing op dit werk. Ga naar om deze licentie te bekijken. Date page 2 of 15

3 Table of contents 1 Introduction 4 2 Theory 4 3 Setup Schematic view Pictures 6 4 Remote interface Settings Measure Calibrate Analyse Journal 13 5 Constants and parameters 14 6 References 14 page 3 of 15

4 1 Introduction This manual explains the working of a remotely controlled X-ray fluorescence experiment. X-ray fluorescence is a technique that can be used to determine the surface composition of a sample non-destructively. Related to this technique are various processes such as attenuation of x- rays in materials, absorption, fluorescence yields, coherent and incoherent scattering. In this experiment some of these processes can be studied using a 241 Am source. The experimental setup is located at the Vrije Universiteit Amsterdam and is maintained by the Physics student laboratory. The web-experiment can be accessed from practically any location in the world, provided the user s system meets the minimum system requirements as listed in Appendix A. Before starting the experiment it is recommended to formulate a research plan to work as productively as possible. To this end, besides reading this manual, it may help to study the background theory available on the website to find out what the possibilities and limitations are of online experimenting. 2 Theory X-ray fluorescence occurs when an electron from the inner electron shell of an atom (molecule) is ejected by a high energy X-ray photon. The absorption of x-ray photons can be described by the law of Lambert-Beer. The created hole is filled up by an electron from one of the outer shells (an electron from the L, M, state). The energy difference from this transition is emitted as a fluorescence photon. Fig. 2.1: A Bohr-atom with filled K and L shells and one electron in the M shell. These energies are characteristic for each element. A measurement of emitted fluorescence photons coming from a target results in fluorescence spectrum. The composition of the exposed target can be determent by comparing the energies in this spectrum with the known transition energies of all elements. See the table of emission energies. page 4 of 15

5 An electron from the L to the K shell is a Kα transition and from the M to the K shell is a Kβ transition. For an introduction to the theory of fluorescence see refs. 1, 2 and 3. Elements you should be familiar with are (notations of) electron states, characteristic lines, selection rules, fluorescence and intensity yield, absorption and scattering of x-rays. 3 Setup 3.1 Schematic view The most important parts of the experimental setup are the x-ray source, the target materials and the x-ray detector. (Fig. 3.1) X-ray source Fig. 3.1: schematic overview of experimental setup X-ray source The x-ray source is a 241 Am (10 mcu). The x-ray spectrum of this source together with the identification of some peaks is shown in fig The target materials In the target wheel different materials can be selected. For the list of materials that can be chosen see 4.1: Remote interface settings The x-ray detector Specifications of the detector (type XR-100CR) can be found in ref. 4. If an x-ray photon interacts with the silicon plate in the x-ray detector, it produces a number of electron/hole pairs depending on the energy of this photon. This results in a short signal pulse. After amplification these pulses are selected on height, counted and stored by the MCA. The resulting x-ray spectrum is then read by the computer. The detector can be placed at different angles. The angle between the x-ray source and the x-ray detector can be varied from 40 to 180 degrees. page 5 of 15

6 3.2 Pictures Below some pictures of the setup. Legend: A B C D E F G X-ray detector X-ray source Target wheel Power supply x-ray detector and preamplifier Multi channel analyzer Stepper motor control unit web-cams D E A X-ray detector with power supply Multi channel analyzer (MCA) page 6 of 15

7 G A B C D F Overview of the experimental setup A B C Close-up of x-ray detector, target wheel and x-ray source. page 7 of 15

8 4 Remote interface Below is a description of the software interface of the experiment. The panel is organized into several tabs: 1. Settings 2. Measure 3. Calibrate 4. Analyse 5. Journal Each tab is described in the next paragraphs. 4.1 Settings Various experimental parameters can be adjusted: The angle between the detector and X-ray source Normally set at an angle of 40 degrees for fluorescence measurements and 180 degrees for absorption measurements. A change of this angle has to be confirmed with the confirmation button which appears below this control. A type of target One can choose targets with known materials such as copper, iron, zirconium for calibration purposes, or targets with a priori unknown compositions like the sand from the shores of the Dutch island of Texel to analyse their content. Also some compounds can be chosen such as brass or a euro coin for this purpose. A pinhole A pinhole is only used with absorption measurements. During fluorescence measurements it should always be set to none. Webcam views Two types of web cam views can be chosen: an upper view of the setup and a side view of the chosen target in front of the source. page 8 of 15

9 Figure 4.1: screenshot of the SETTINGS tab in the remote panel overview. Set angle detector X-ray source: a new angle can be set by changing the value or by dragging the arrow in the picture control to a new position. The actual change to this new position is done by clicking the confirm button, which appears when an change is first initiated Confirm: clicking this button will result in moving the x-ray detector to the new position. Choice target: with this popup button you can select one of the following materials: 0. empty mm Silver mm Copper mm Nickel mm Copper mm Iron mm Copper mm Aluminum mm Copper 12. Euro coin (50 cts) mm Lead 13. Sand (from the sea bottom of Texel) mm Tantalum mm Zirconium mm Tin 15. Brass Choice pinhole: You can choose between the following diameters: 2, 4, 8, and 16 mm. None should always be selected for fluorescence measurements. Large image: when selected the color will change to blue and a large image is displayed. Small image: when selected the color will change to blue and a small image is displayed. Camera upper view: shows an upper view of the experimental setup Camera targets and carrousel: shows a part of the target wheel and the x-ray detector page 9 of 15

10 4.2 Measure The measured energy spectrum of the X-ray detector is displayed. This energy spectrum is measured with a Multi Channel Analyzer (MCA). Controls of this instrument are: start or stop the measurement, clear memory (and graph), number of channels to set the resolution of the energy spectrum and preset time to set the measuring time. When set to zero the measurement can only be stopped manually. The measured time corrected for dead time is shown in the lower indicator. The indicator Measurement settings and variables shows all the settings of the displayed energy spectrum. The data can be saved and previously measured spectra can be read. Read MCA buffer control displays the last measured energy spectrum that is still in the buffer. Figure 4.2: screenshot of the MEASURE tab in the remote panel overview. Energy spectrum: the current energy spectrum is displayed here. Number of photons against channel number (or against energy in kev if the MCA is calibrated). Preset time [sec]: the preset live time to measure the energy spectrum. If zero the measuring time is infinite. Measured time [sec]: the measured live time that has past. Measurement settings and variables: Information about the displayed energy spectrum. Start measurement: Starts the measurement until the preset time is reached or the measurement is stopped manually. Stop measurement: stops the measurement. Empty MCA buffer and graph: clears the buffer of the multi channel analyzer and the graph. Number of channels: determines the resolution of the measurement. Possible values are 256, 512, 1024, 2048, 4096 and 8192 channels. Read MCA buffer: reads the buffer of the MCA. page 10 of 15

11 Write data to file: Saves the displayed energy spectrum in a data file and the information shown in measurement settings and variables indicator in an info file. Read data from file: reads previously saved data- and info-file from disk and displays it in the energy spectrum graph and measurement settings and variables indicator. 4.3 Calibrate This tab gives the possibility to calibrate the x-axis of the measured energy spectrum. The X-ray emission energies of all selected known targets, of which energy spectra are measured, are shown in the upper indicator/control. On the right side of this tab you give input for at least two channel numbers with each a selected X-ray emission energy. After pressing in sequence the buttons Determine straight line fit and calibrate x-axis the x-axis will be calibrated and changed into energy in kev. This is also done for the spectra under the other tabs. The extra controls cursor active will activate an extra cursor. Snap to channel means that the cursor is locked to each point of the energy spectrum. Figure 4.3: screenshot of the CALIBRATE tab in the remote panel overview. X-ray emission energies in kev: here all the emission energies are shown of all the known target materials of which energy spectra are measured. spectrum: displays all the line spectra of the known materials in the upper popup control. page 11 of 15

12 Energy spectrum: the current energy spectrum is displayed here. Number of photons against channel number (or against energy in kev if the MCA is calibrated). Slope: a parameter of the linear fit. Intercept y-as: a parameter of the linear fit. Mean squared deviation: goodness of the linear fit. Cursor active: if selected a cursor is shown in the graph and a active value in the linear fit. Channel number: the position of the cursor in the graph. Energy value kev: here the channel number is set to the emission energy values in kev s. Snap to channel: If selected the cursor is locked to a channel number. Determine straight line fit: A fit is made with the channel numbers and the energy values corresponding to the active cursors. Calibrate x-axis: After a fit is calculated all energy spectra will be shown in kev. 4.4 Analyse On this tab the energy spectrum and a panel with all elements and the corresponding X-ray emission energies is displayed. By selecting elements from this panel it is possible to investigate which elements contribute to the energy spectrum. When an element is selected (clicked on) a line spectrum with the tabulated X-ray energies from this element is shown in the graph. The RESET ALL button will remove all selected elements from the graph. Figure 4.4: screenshot of the ANALYSE tab in the remote panel overview. Energy spectrum: the current energy spectrum is displayed here. Number of photons against channel number (or against energy in kev if the MCA is calibrated). page 12 of 15

13 Elements: here an element can be selected, which results in colored lines in the upper energy spectrum at the specific energy values (kev s) given in this table. All selected elements are also set in the legend on the right hand side of the graph Reset all: clears all colored lines of the chosen elements in the upper graph 4.5 Journal In this tab you can enter items to your lab journal. This is done by selecting a Journal Item and Journal Status from the lists at the top of this tab and entering a text in the field Journal Text. Optionally you can add a reference to an earlier journal entry or log entry. This is done by clicking the entry in the list box at the bottom of the tab or entering its number in the reference field. By clicking the button Submit, the journal entry is added to the log. Note that some actions are automatically logged. Figure 4.5: screenshot of the JOURNAL tab in the remote panel overview. Journal Item: here it is possible to choose a category for your journal entry. Journal Status: here you can choose a qualification of your journal entry. Journal Text: here the actual text of the journal entry can be typed. Reference: You can refer to an earlier journal entry or experiment log entry by clicking in the table of entries or choosing a number in the Reference field. Submit: by clicking the button Submit the journal entry is added to the experiment log. page 13 of 15

14 5 Constants and parameters Name Description Value / range, error, unit X-ray source 241 Am 10 mcu Source/detector The angle between the x-ray source and the 40 to 180 degrees angle x-ray detector. Efficientie Ref. 3 detector Emission energies of all download/tabel.pdf elements Size pinholes Diameter pinhole (lead) 2,4,8 and 16 mm 6 References 1. R. Tertian and F. Claisse, Principles of Quantitative X-ray Fluorescence Analysis 2. B. Dziunikowkski, Energy dispersive X-ray Fluorescence Analysis Website of Amptek ( page 14 of 15

15

Manual Laser Doppler Anemometry Manual remote experiment Project e-xperimenteren+

Manual Laser Doppler Anemometry Manual remote experiment Project e-xperimenteren+ Manual Laser Doppler Anemometry Manual remote experiment Project e-xperimenteren+ J. Snellenburg, J.M.Mulder 19-01-2006 Colofon Manual Laser Doppler Anemometry Manual remote experiment Project e-xperimenteren+

More information

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination Uppsala University Department of Physics and Astronomy Laboratory exercise X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

More information

X-RAY SPECTRA. Theory:

X-RAY SPECTRA. Theory: 12 Oct 18 X-ray.1 X-RAY SPECTRA In this experiment, a number of measurements involving x-rays will be made. The spectrum of x-rays emitted from a molybdenum target will be measured, and the experimental

More information

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P ) Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P2541001) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne

More information

CHEMISTRY SEMESTER ONE

CHEMISTRY SEMESTER ONE BEER-LAMBERT LAW Lab format: this lab is a remote lab activity Relationship to theory: This activity quantitatively relates the concentration of a lightabsorbing substance to the absorbance of light. LEARNING

More information

Lab 1 Uniform Motion - Graphing and Analyzing Motion

Lab 1 Uniform Motion - Graphing and Analyzing Motion Lab 1 Uniform Motion - Graphing and Analyzing Motion Objectives: < To observe the distance-time relation for motion at constant velocity. < To make a straight line fit to the distance-time data. < To interpret

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

Lesson Plan: Introduction to Quantum Mechanics via Wave Theory and the Photoelectric Effect

Lesson Plan: Introduction to Quantum Mechanics via Wave Theory and the Photoelectric Effect Lesson Plan: Introduction to Quantum Mechanics via Wave Theory and the Photoelectric Effect Will Stoll, Norcross High School Problem: To understand the basic principles of Quantum Mechanics through an

More information

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron

More information

This experiment is included in the XRP 4.0 X-ray solid state, XRS 4.0 X-ray structural analysis, and XRC 4.0 X-ray characteristics upgrade sets.

This experiment is included in the XRP 4.0 X-ray solid state, XRS 4.0 X-ray structural analysis, and XRC 4.0 X-ray characteristics upgrade sets. The intensity of characteristic X-rays as a TEP Related topics Characteristic X-radiation, energy levels, Bragg s law, and intensity of characteristic X-rays Principle The X-ray spectrum of an X-ray tube

More information

Photoluminescence Spectrometer (FLS980)

Photoluminescence Spectrometer (FLS980) Photoluminescence Spectrometer (FLS980) Instrument Summary: The Edinburgh Instruments FLS980 photoluminescence spectrometer offers both steady state and time resolved (lifetime) fluorescence spectroscopy

More information

HOW TO ANALYZE SYNCHROTRON DATA

HOW TO ANALYZE SYNCHROTRON DATA HOW TO ANALYZE SYNCHROTRON DATA 1 SYNCHROTRON APPLICATIONS - WHAT Diffraction data are collected on diffractometer lines at the world s synchrotron sources. Most synchrotrons have one or more user facilities

More information

I. Pre-Lab Introduction

I. Pre-Lab Introduction I. Pre-Lab Introduction Please complete the following pages before the lab by filling in the requested items. A. Atomic notation: Atoms are composed of a nucleus containing neutrons and protons surrounded

More information

OneStop Map Viewer Navigation

OneStop Map Viewer Navigation OneStop Map Viewer Navigation» Intended User: Industry Map Viewer users Overview The OneStop Map Viewer is an interactive map tool that helps you find and view information associated with energy development,

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

X-ray Absorption Spectroscopy

X-ray Absorption Spectroscopy X-ray Absorption Spectroscopy Nikki Truss November 26, 2012 Abstract In these experiments, some aspects of x-ray absorption spectroscopy were investigated. The x-ray spectrum of molybdenum was recorded

More information

Lab Manual: Determination of Planck s constant with x-rays

Lab Manual: Determination of Planck s constant with x-rays Lab Manual: Determination of Planck s constant with x-rays 1. Purpose: To obtain a better understanding on the production of X-rays, the bremsstrahlung radiation and the characteristic radiation of a Molybdenum

More information

PHYS 3650L - Modern Physics Laboratory

PHYS 3650L - Modern Physics Laboratory PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Dr. White 11 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To prepare a Beer s

More information

Circular Motion and Centripetal Force

Circular Motion and Centripetal Force [For International Campus Lab ONLY] Objective Measure the centripetal force with the radius, mass, and speed of a particle in uniform circular motion. Theory ----------------------------- Reference --------------------------

More information

Absorption of X-rays

Absorption of X-rays Absorption of X-rays TEP Related topics Bremsstrahlung, characteristic X-radiation, Bragg scattering, law of absorption, mass absorption coefficient, absorption edges, half-value thickness, photoelectric

More information

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

Spectroscopy II Introduction: Spectrophotometry and Fluorometry Introduction: Spectrophotometry and Fluorometry In this experiment you will use the SpectroVis Plus (a small computer controlled Spectrophotometer + Fluorometer) to: Measure the Transmission and Absorbance

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To

More information

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

UV-Vis optical fiber assisted spectroscopy in thin films and solutions UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation

More information

Creating Empirical Calibrations

Creating Empirical Calibrations 030.0023.01.0 Spreadsheet Manual Save Date: December 1, 2010 Table of Contents 1. Overview... 3 2. Enable S1 Calibration Macro... 4 3. Getting Ready... 4 4. Measuring the New Sample... 5 5. Adding New

More information

Calculating Bond Enthalpies of the Hydrides

Calculating Bond Enthalpies of the Hydrides Proposed Exercise for the General Chemistry Section of the Teaching with Cache Workbook: Calculating Bond Enthalpies of the Hydrides Contributed by James Foresman, Rachel Fogle, and Jeremy Beck, York College

More information

Experiment 4 Radiation in the Visible Spectrum

Experiment 4 Radiation in the Visible Spectrum Experiment 4 Radiation in the Visible Spectrum Emission spectra can be a unique fingerprint of an atom or molecule. The photon energies and wavelengths are directly related to the allowed quantum energy

More information

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette

More information

CHEMISTRY SEMESTER ONE

CHEMISTRY SEMESTER ONE EMISSION SPECTROSCOPY Lab format: this lab is a remote lab activity Relationship to theory: This activity covers the relationship between colors and absorbed/emitted light, as well as the relationship

More information

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P ) Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:

More information

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Nuclear Lifetimes. = (Eq. 1) (Eq. 2) Nuclear Lifetimes Theory The measurement of the lifetimes of excited nuclear states constitutes an important experimental technique in nuclear physics. The lifetime of a nuclear state is related to its

More information

QuantumMCA QuantumNaI QuantumGe QuantumGold

QuantumMCA QuantumNaI QuantumGe QuantumGold QuantumMCA QuantumNaI QuantumGe QuantumGold Berkeley Nucleonics Corporation (San Rafael, CA) and Princeton Gamma Tech (Princeton, NJ) have partnered to offer gamma spectroscopy with either germanium or

More information

CASSY Lab. Manual ( )

CASSY Lab. Manual ( ) CASSY Lab Manual (524 202) Moseley's law (K-line x-ray fluorescence) CASSY Lab 271 can also be carried out with Pocket-CASSY Load example Safety notes The X-ray apparatus fulfils all regulations on the

More information

LAB 01 X-RAY EMISSION & ABSORPTION

LAB 01 X-RAY EMISSION & ABSORPTION LAB 0 X-RAY EMISSION & ABSORPTION REPORT BY: TEAM MEMBER NAME: Ashley Tsai LAB SECTION No. 05 GROUP 2 EXPERIMENT DATE: Feb., 204 SUBMISSION DATE: Feb. 8, 204 Page of 3 ABSTRACT The goal of this experiment

More information

NMR Predictor. Introduction

NMR Predictor. Introduction NMR Predictor This manual gives a walk-through on how to use the NMR Predictor: Introduction NMR Predictor QuickHelp NMR Predictor Overview Chemical features GUI features Usage Menu system File menu Edit

More information

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440 Jazan University College of Science Physics Department جاهعة جازان كلية العل وم قسن الفيزياء Lab Manual Nuclear Physics (2) 462 Phys 8 th Level Academic Year: 1439/1440 1 Contents No. Name of the Experiment

More information

X-ray Spectroscopy. Danny Bennett and Maeve Madigan. October 12, 2015

X-ray Spectroscopy. Danny Bennett and Maeve Madigan. October 12, 2015 X-ray Spectroscopy Danny Bennett and Maeve Madigan October 12, 2015 Abstract Various X-ray spectra were obtained, and their properties were investigated. The characteristic peaks were identified for a

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Attenuation of x-rays LEYBOLD Physics Leaflets P6.3.2.2 Investigating the wavelength dependency of the coefficient of attenuation Objects of the experiment To measure

More information

AUTOMATIC AND INTERACTIVE ANALYSIS SOFTWARE FOR BETA- GAMMA COINCIDENCE SYSTEMS USED IN CTBT MONITORING

AUTOMATIC AND INTERACTIVE ANALYSIS SOFTWARE FOR BETA- GAMMA COINCIDENCE SYSTEMS USED IN CTBT MONITORING ABSTRACT AUTOMATIC AND INTERACTIVE ANALYSIS SOFTWARE FOR BETA- GAMMA COINCIDENCE SYSTEMS USED IN CTBT MONITORING J. Rynes, K.M.F. Biegalski, P. Donohoe, and S. Biegalski Veridian Pacific-Sierra Research

More information

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013 Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills

More information

Relativistic Electrons

Relativistic Electrons Relativistic Electrons Physics 300 1 Introduction In this experiment you will make independent measurements of the momentum and kinetic energy of electrons emitted from a β source. You will use these data

More information

Scintillation Detector

Scintillation Detector Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous

More information

Overview of X-Ray Fluorescence Analysis

Overview of X-Ray Fluorescence Analysis Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

Computer simulation of radioactive decay

Computer simulation of radioactive decay Computer simulation of radioactive decay y now you should have worked your way through the introduction to Maple, as well as the introduction to data analysis using Excel Now we will explore radioactive

More information

X-RAY SCATTERING AND MOSELEY S LAW. OBJECTIVE: To investigate Moseley s law using X-ray absorption and to observe X- ray scattering.

X-RAY SCATTERING AND MOSELEY S LAW. OBJECTIVE: To investigate Moseley s law using X-ray absorption and to observe X- ray scattering. X-RAY SCATTERING AND MOSELEY S LAW OBJECTIVE: To investigate Moseley s law using X-ray absorption and to observe X- ray scattering. READING: Krane, Section 8.5. BACKGROUND: In 1913, Henry Moseley measured

More information

Standards-Based Quantification in DTSA-II Part II

Standards-Based Quantification in DTSA-II Part II Standards-Based Quantification in DTSA-II Part II Nicholas W.M. Ritchie National Institute of Standards and Technology, Gaithersburg, MD 20899-8371 nicholas.ritchie@nist.gov Introduction This article is

More information

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado Experiment 6 1 Introduction The Compton Effect Physics 2150 Experiment No. 6 University of Colorado In some situations, electromagnetic waves can act like particles, carrying energy and momentum, which

More information

MASS ATTENUATION COEFFICIENT OF LEAD

MASS ATTENUATION COEFFICIENT OF LEAD OBJECTIVE MASS ATTENUATION COEFFICIENT OF LEAD The objective of this experiment is to measure the mass attenuation coefficient of lead by manipulating Beer-Lambert s law of attenuation. INTRODUCTION Background

More information

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report Newton s second law may be written r r F dt = p where F is the force and p is the change in momentum. The area under the force vs. time

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Moseley s law and determination of the Rydberg constant P6.3.3.6 Objects of the experiment Measuring the K-absorption

More information

Kinematics. Become comfortable with the data aquisition hardware and software used in the physics lab.

Kinematics. Become comfortable with the data aquisition hardware and software used in the physics lab. Kinematics Objective Upon completing this experiment you should Become comfortable with the data aquisition hardware and software used in the physics lab. Have a better understanding of the graphical analysis

More information

Developing a Scientific Theory

Developing a Scientific Theory Name Date Developing a Scientific Theory Equipment Needed Qty Equipment Needed Qty Photogate/Pulley System (ME-6838) 1 String (SE-8050) 1 Mass and Hanger Set (ME-8967) 1 Universal Table Clamp (ME-9376B)

More information

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY UV/Vis Spectroscopy is a technique that is widely used to characterize, identify and quantify chemical compounds in all fields of analytical chemistry. The

More information

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values. EXPERIMENT 2.1 GAMMA ENERGY CALIBRATION 1- Turn the power supply on to 900 V. Turn the NIM crate on to power the amplifiers. Turn the Oscilloscope on to check the gamma pulses. The main amplifier should

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

X-ray fluorescence analysis - calibration of the X-ray energy detector

X-ray fluorescence analysis - calibration of the X-ray energy detector X-ray fluorescence analysis - TEP Related topics Bremsstrahlung, characteristic X-radiation, energy levels, fluorescence radiation, conduction processes in semiconductors, doping of semiconductors, pin-diodes,

More information

X-ray Spectroscopy. c David-Alexander Robinson & Pádraig Ó Conbhuí. 14th March 2011

X-ray Spectroscopy. c David-Alexander Robinson & Pádraig Ó Conbhuí. 14th March 2011 X-ray Spectroscopy David-Alexander Robinson; Pádraig Ó Conbhuí; 08332461 14th March 2011 Contents 1 Abstract 2 2 Introduction & Theory 2 2.1 The X-ray Spectrum............................ 2 2.2 X-Ray Absorption

More information

atomic number and mass number. Go over nuclear symbols, such as He-4 and He. Discuss

atomic number and mass number. Go over nuclear symbols, such as He-4 and He. Discuss Nuclear Decay and Chain Reactions ID: 9522 Time required 45 minutes Topic: Nuclear Identify and write equations for the three forms of nuclear decay. Predict decay products. Perform half-life and decay

More information

Experiment 14 It s Snow Big Deal

Experiment 14 It s Snow Big Deal Experiment 14 It s Snow Big Deal OUTCOMES After completing this experiment, the student should be able to: use computer-based data acquisition techniques to measure temperatures. draw appropriate conclusions

More information

Zeeman Effect Physics 481

Zeeman Effect Physics 481 Zeeman Effect Introduction You are familiar with Atomic Spectra, especially the H- atom energy spectrum. Atoms emit or absorb energies in packets, or quanta which are photons. The orbital motion of electrons

More information

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong Introduction: X-ray fluorescence (XRF) spectrometer is an x-ray instrument used for routine, relatively

More information

PHY 192 Compton Effect Spring

PHY 192 Compton Effect Spring PHY 192 Compton Effect Spring 2010 1 The Compton Effect Introduction In this experiment we will study two aspects of the interaction of photons with electrons. The first of these is the Compton effect

More information

ON SITE SYSTEMS Chemical Safety Assistant

ON SITE SYSTEMS Chemical Safety Assistant ON SITE SYSTEMS Chemical Safety Assistant CS ASSISTANT WEB USERS MANUAL On Site Systems 23 N. Gore Ave. Suite 200 St. Louis, MO 63119 Phone 314-963-9934 Fax 314-963-9281 Table of Contents INTRODUCTION

More information

Lab Activity H4 It s Snow Big Deal

Lab Activity H4 It s Snow Big Deal Lab Activity H4 It s Snow Big Deal OUTCOMES After completing this lab activity, the student should be able to use computer-based data acquisition techniques to measure temperatures. draw appropriate conclusions

More information

Concentrations that absorb. Measuring percentage transmittance of solutions at different concentrations

Concentrations that absorb. Measuring percentage transmittance of solutions at different concentrations Measuring percentage transmittance of solutions at different Dimension 2 Cross Cutting Concepts Dimension 1 Science and Engineering Practices FRAMEWORK FOR K-12 SCIENCE EDUCATION 2012 Concentrations that

More information

Multi-element process analyzer

Multi-element process analyzer Multi-element process analyzer Elemental analysis by X-ray fluorescence Compact multi-element process analyzer for liquid streams or f ixed position web applications Featuring advanced third generation

More information

GAMMA RAY SPECTROSCOPY

GAMMA RAY SPECTROSCOPY GAMMA RAY SPECTROSCOPY Gamma Ray Spectroscopy 1 In this experiment you will use a sodium iodide (NaI) detector along with a multichannel analyzer (MCA) to measure gamma ray energies from energy level transitions

More information

Physics Labs with Computers, Vol. 1 P23: Conservation of Angular Momentum A

Physics Labs with Computers, Vol. 1 P23: Conservation of Angular Momentum A Activity P23: Conservation of Angular Momentum (Rotary Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Rotational motion P23 Angular Momentum.DS (See end of activity) (See

More information

mylab: Chemical Safety Module Last Updated: January 19, 2018

mylab: Chemical Safety Module Last Updated: January 19, 2018 : Chemical Safety Module Contents Introduction... 1 Getting started... 1 Login... 1 Receiving Items from MMP Order... 3 Inventory... 4 Show me Chemicals where... 4 Items Received on... 5 All Items... 5

More information

X-Ray Fluorescence and Natural History

X-Ray Fluorescence and Natural History X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

Uniformly Accelerated Motion

Uniformly Accelerated Motion Uniformly Accelerated Motion 2-1 Uniformly Accelerated Motion INTRODUCTION All objects on the earth s surface are being accelerated toward the center of the earth at a rate of 9.81 m/s 2. 1 This means

More information

Experiment P05: Position, Velocity, & Acceleration (Motion Sensor)

Experiment P05: Position, Velocity, & Acceleration (Motion Sensor) PASCO scientific Physics Lab Manual: P05-1 Experiment P05: Position, Velocity, & Acceleration (Motion Sensor) Concept Time SW Interface Macintosh file Windows file linear motion 30 m 500 or 700 P05 Position,

More information

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS by Carly Jean Smith A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment

More information

DP Physics Torque Simulation

DP Physics Torque Simulation DP Physics Torque Simulation Name Go to Phet Simulation: ( http://phet.colorado.edu/simulations/sims.php?sim=torque ) Part I: Torque 1. Click the tab at the top that says torque 2. Set the force equal

More information

Login -the operator screen should be in view when you first sit down at the spectrometer console:

Login -the operator screen should be in view when you first sit down at the spectrometer console: Lab #2 1D 1 H Double Resonance (Selective Decoupling) operation of the 400 MHz instrument using automated sample insertion (robot) and automated locking and shimming collection of 1D 1 H spectra retrieving

More information

Investigating Food Dyes in Sports Beverages. Sample

Investigating Food Dyes in Sports Beverages. Sample Investigating Food Dyes in Sports Beverages Investigation 1 There are many different brands of beverages that fall under the general category of sports drinks. Most of these beverages contain an FD&C food

More information

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219 REX Evaluation Guide American Micro Detection Systems Inc. 2800 March Lane, Suite 200 Stockton, CA 95219 I. INTRODUCTION REX (Real-time Elemental X-ray Fluorescence System) is the only instrument capable

More information

Gamma Ray Spectroscopy

Gamma Ray Spectroscopy Gamma Ray Spectroscopy Uzair Latif, Imran Younus Department of Physics Lahore University of Management Sciences November 4, 2014 1 Objectives 1. To acquaint the students with some of the basic techniques

More information

RADIOACTIVITY MATERIALS: PURPOSE: LEARNING OBJECTIVES: DISCUSSION:

RADIOACTIVITY MATERIALS: PURPOSE: LEARNING OBJECTIVES: DISCUSSION: RADIOACTIVITY This laboratory experiment was largely adapted from an experiment from the United States Naval Academy Chemistry Department MATERIALS: (total amounts per lab) small bottle of KCl; isogenerator

More information

Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate)

Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate) Name Class Date Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Energy P20 Mechanical Energy.DS P23 Cons. Mechanical

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information

PRELIMINARY ACTIVITY FOR

PRELIMINARY ACTIVITY FOR PRELIMINARY ACTIVITY FOR Beer s Law Investigations Guided Inquiry Version Experiment 11 The primary objective of this Preliminary Activity is to determine the concentration of an unknown copper (II) sulfate

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method

More information

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U Oxford House, 20 Oxford St Newbury, Berkshire. RG14 1JB UK. sales@spectrolab.eu sales@spectrolab.co.uk Offices in UK, EU, Greece, Dubai The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in

More information

OCEAN/ESS 410 Lab 4. Earthquake location

OCEAN/ESS 410 Lab 4. Earthquake location Lab 4. Earthquake location To complete this exercise you will need to (a) Complete the table on page 2. (b) Identify phases on the seismograms on pages 3-6 as requested on page 11. (c) Locate the earthquake

More information

PhET Light Emission and Lasers (27 points available x 2/3 = 18 points max score)

PhET Light Emission and Lasers (27 points available x 2/3 = 18 points max score) IB PHYSICS Name: Period: Date: DEVIL PHYSICS BADDEST CLASS ON CAMPUS GETTING STARTED: PhET Light Emission and Lasers (27 points available x 2/3 = 18 points max score) Go to https://phet.colorado.edu/ Click

More information

Accountability. User Guide

Accountability. User Guide Accountability User Guide The information in this document is subject to change without notice and does not represent a commitment on the part of Horizon. The software described in this document is furnished

More information

THE GEIGER-MULLER TUBE AND THE STATISTICS OF RADIOACTIVITY

THE GEIGER-MULLER TUBE AND THE STATISTICS OF RADIOACTIVITY GMstats. THE GEIGER-MULLER TUBE AN THE STATISTICS OF RAIOACTIVITY This experiment examines the Geiger-Muller counter, a device commonly used for detecting and counting ionizing radiation. Various properties

More information

O P E R A T I N G M A N U A L

O P E R A T I N G M A N U A L OPERATING MANUAL WeatherJack OPERATING MANUAL 1-800-645-1061 The baud rate is 2400 ( 8 bits, 1 stop bit, no parity. Flow control = none) To make sure the unit is on line, send an X. the machine will respond

More information

Student Exploration: Bohr Model: Introduction

Student Exploration: Bohr Model: Introduction Name: Date: Student Exploration: Bohr Model: Introduction Vocabulary: absorption spectrum, Bohr model, electron volt, energy level, laser, orbital, photon Prior Knowledge Questions (Do these BEFORE using

More information

Experiment 10. Zeeman Effect. Introduction. Zeeman Effect Physics 244

Experiment 10. Zeeman Effect. Introduction. Zeeman Effect Physics 244 Experiment 10 Zeeman Effect Introduction You are familiar with Atomic Spectra, especially the H-atom energy spectrum. Atoms emit or absorb energies in packets, or quanta which are photons. The orbital

More information

Software BioScout-Calibrator June 2013

Software BioScout-Calibrator June 2013 SARAD GmbH BioScout -Calibrator 1 Manual Software BioScout-Calibrator June 2013 SARAD GmbH Tel.: ++49 (0)351 / 6580712 Wiesbadener Straße 10 FAX: ++49 (0)351 / 6580718 D-01159 Dresden email: support@sarad.de

More information

DAY LABORATORY EXERCISE: SPECTROSCOPY

DAY LABORATORY EXERCISE: SPECTROSCOPY AS101 - Day Laboratory: Spectroscopy Page 1 DAY LABORATORY EXERCISE: SPECTROSCOPY Goals: To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are

More information

Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods

Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods 1 Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods Junhong Sam Zhou Victor Senior High School LLE Advisor: Christian Stoeckl Laboratory for Laser Energetics University of Rochester

More information

Lasentec Product Group

Lasentec Product Group Document Number: 004-0043 Window Reference Procedure - M500/600 P Probes, version 6.0 and higher FBRM CI Software The Window Reference Procedure is the standard calibration procedure used to reference

More information

IncuCyte ZOOM NeuroTrack Fluorescent Processing

IncuCyte ZOOM NeuroTrack Fluorescent Processing IncuCyte ZOOM NeuroTrack Fluorescent Processing The NeuroTrack TM Software Module (Cat No 9600-0011) is used to measure the processes of neurons in monoculture or with fluorescent labeling in co-culture.

More information

Student Exploration: Bohr Model of Hydrogen

Student Exploration: Bohr Model of Hydrogen Name: Date: Student Exploration: Bohr Model of Hydrogen Vocabulary: absorption spectrum, Bohr model, electron volt, emission spectrum, energy level, ionization energy, laser, orbital, photon [Note to teachers

More information