Software BioScout-Calibrator June 2013
|
|
- Joan Hancock
- 5 years ago
- Views:
Transcription
1 SARAD GmbH BioScout -Calibrator 1 Manual Software BioScout-Calibrator June 2013 SARAD GmbH Tel.: ++49 (0)351 / Wiesbadener Straße 10 FAX: ++49 (0)351 / D Dresden support@sarad.de GERMANY Internet: Content INTRODUCTION... 3 OVERVIEW... 3 SYSTEM REQUIREMENTS... 3 USE CASES... 4 EXCHANGING NUCLIDES... 4 Goal... 4 Steps... 4 Results... 5 MEASUREMENT PARAMETERS... 6 Goal... 6 Steps... 6 Results... 7 MANAGE THE BIOSCOUT-VIEW SPECTRA COLLECTION... 8 Goal... 8 Steps... 8 Result RECALIBRATE Goal Steps Result FURTHER HELP... 14
2 SARAD GmbH BioScout -Calibrator 2 List of abbreviations RS232 USB kev Standard for serial communication Universal Serial Bus kiloelectron volt
3 SARAD GmbH BioScout -Calibrator 3 Introduction The software BioScout-Calibrator is the expert tool for the BioScout instrument. This document describes the basic usage of this software. It is recommended to read the BioScout instrument manual before starting to read this document. There is a non expert tool (BioScout-View) which enables the costumer to read out spectra from the BioScout. In contrast the BioScout-Calibrator is also capable to calibrate the BioScout and to set the nuclides the BioScout should detect. Both use cases are described in this manual. Additionally the management of spectra is described. Overview The BioScout is a typical windows application. It has a menu bar which enables you to control the application, an area where all the dialogs are placed and a status bar where status information is displayed. System requirements Windows XP / VISTA / 7 (32/64 bit) Serial interface (RS232) or USB port Monitor with at least 1024 x 768 pixels Mouse
4 SARAD GmbH BioScout -Calibrator 4 Use cases In the following it is described how to choose the nuclides the BioScout is supposed to detect and how to calibrate or manage the spectra collection of the BioScout-View software using the BioScout-Calibrator software. Exchanging Nuclides Goal The BioScout is by default configured to detect the following nuclides: Sr-90, Am-241, Cs- 134, Cs-137, I-131, K-40. To be able to detect other nuclides the preset nuclides need to be exchanged. Steps 1. Attach the BioScout to the computer either via RS232 or USB 2. Start the software application 3. Enter the provided password 4. Select a BioScout from the list of auto detected BioScouts 5. Open configuration
5 SARAD GmbH BioScout -Calibrator 5 6. Select a nuclide slot on the left 7. Double click on the wished nuclide in the list on the right 8. Close the dialog 9. Confirm configuration download 10. Switch off the BioScout and on again to let it perform the background check once more Note: The BioScout uses a special mechanism to monitor the background. If a nuclide with a higher energy than 796 kev is selected, this background mechanism needs to be changed. In this case please contact the manufacturer. Results After an updated background check the BioScout is capable to detect the chosen nuclides. Note: The new configuration is stored in the device. It will overwrite the default setting and therefore remains after switching the power off.
6 SARAD GmbH BioScout -Calibrator 6 Measurement Parameters Goal When detecting different levels of activity concentrations with the BioScout it will be necessary to vary parameters like the measurement time or the time for the background measurement to alter the accuracy of the results and the BioScout s detection limit. Steps 1. Attach the BioScout to the computer either via RS232 or USB 2. Start the software application 3. Enter the provided password 4. Select a BioScout from the list of auto detected BioScouts 5. Open configuration
7 SARAD GmbH BioScout -Calibrator 7 6. Change the measurement parameters by moving the slide control 7. Close the dialog 8. Confirm configuration download 9. Switch off the BioScout and on again to let it perform the background check once more Results Users are capable to select measurement parameters of the BioScout according to they measurement goal.
8 SARAD GmbH BioScout -Calibrator 8 Manage the BioScout-View Spectra Collection The BioScout-View software is able to read out the BioScout s spectra storage. The read out spectra are saved on the computer in a spectra collection. Because this data is intended to be used as evidence it is not intended to remove any spectrum from the growing collection by using the BioScout-View software. Managing the spectra can only be done by using the BioScout Calibrator software. Goal Remove or add spectra from the BioScout-View s spectrum collection. Steps 1. Attach the BioScout to the computer either via RS232 or USB 2. Start the software application 3. Enter the provided password 4. Select a BioScout from the list of auto detected BioScouts
9 SARAD GmbH BioScout -Calibrator 9 5. Open spectrum collection by clicking on Spectrum menu and afterwards on open 6. Select Energy Spectra Collection... from the drop down box on the lower right, provide %ALLUSERSPROFILE%\SARAD GmbH\BioScout\ as file name and hit enter. Afterwards select store.especc and confirm the dialog.
10 SARAD GmbH BioScout -Calibrator Add or remove spectra from the collection a. Remove spectra from the list by using the context menu (click with right mouse button on the list) b. Add spectra by opening other collections or single spectra or by reading out spectra from the BioScout 8. Save the spectra collection by clicking on the Spectrum menu and afterwards on Save all. Result The BioScout-View s collection of spectra is updated.
11 SARAD GmbH BioScout -Calibrator 11 Recalibrate Note: This is a complex operation. In order to calibrate the device a calibration standard with a definite activity is needed to be able to compare the output result with it. This is called the efficiency calibration. Additionally, well known gamma radiation sources are needed which cover a broad part of the gamma spectrum that is observed by the device. This is called the energy calibration. It is recommended to use the Excel Template (Calibration.xltx, to be requested from SARAD) to calculate the calibration coefficients mentioned below. Goal The accuracy of the results (weight specific activity) provided by the BioScout are dependent on its calibration. If the results don t match the expected values the BioScout needs to be recalibrated. Steps 1. Attach the BioScout to the computer either via RS232 or USB 2. Start the software application 3. Enter the provided password 4. Select a BioScout from the list of auto detected BioScouts
12 SARAD GmbH BioScout -Calibrator Follow the calibration procedure, which is described in the calibration Excel Template by utilizing the command Current from the Spectrum menu. 6. Open configuration by clicking in the upper left corner of the configuration window 7. Modify coefficients by clicking on them 8. Close the coefficient dialog 9. Close the configuration dialog
13 SARAD GmbH BioScout -Calibrator Confirm configuration download. 10. Switch off the BioScout and on again to let it perform the background check once more Result The BioScout shows correct results within the displayed error range.
14 SARAD GmbH BioScout -Calibrator 14 Further help internet:
ON SITE SYSTEMS Chemical Safety Assistant
ON SITE SYSTEMS Chemical Safety Assistant CS ASSISTANT WEB USERS MANUAL On Site Systems 23 N. Gore Ave. Suite 200 St. Louis, MO 63119 Phone 314-963-9934 Fax 314-963-9281 Table of Contents INTRODUCTION
More informationPiCUS Calliper Version 3
Technology PiCUS Calliper Version 3 Digital calliper for precise geometry detection argus electronic GmbH Erich-Schlesinger-Straße 49d 18059 Rostock Germany Tel.: +49 (0) 381/49 68 14 40 www.argus-electronic.de
More informationK D A A M P L I F I E R S F I R M W A R E U S E R G U I D E
K D A A M P L I F I E R S F I R M W A R E U S E R G U I D E T A B L E O F C O N T E N T S S E C T I O N 1 : P R E PA R I N G Y O U R F I L E S Via Network Router 3 S E C T I O N 2 : A C C E S S I N G T
More informationOcean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)
Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette
More informationFRAM V5.2. Plutonium and Uranium Isotopic Analysis Software
V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,
More informationUser's Guide. DISTO online. Leica Geosystems
User's Guide DISTO online Leica Geosystems Copyright 2001 by PMS Photo Mess Systeme AG. All rights reserved. This manual describes the versions 2.x of the program DISTO online. PMS PHOTO-MESS-SYSTEME AG
More informationBloomsburg University Weather Viewer Quick Start Guide. Software Version 1.2 Date 4/7/2014
Bloomsburg University Weather Viewer Quick Start Guide Software Version 1.2 Date 4/7/2014 Program Background / Objectives: The Bloomsburg Weather Viewer is a weather visualization program that is designed
More informationST-Links. SpatialKit. Version 3.0.x. For ArcMap. ArcMap Extension for Directly Connecting to Spatial Databases. ST-Links Corporation.
ST-Links SpatialKit For ArcMap Version 3.0.x ArcMap Extension for Directly Connecting to Spatial Databases ST-Links Corporation www.st-links.com 2012 Contents Introduction... 3 Installation... 3 Database
More informationWeatherHub2 Quick Start Guide
WeatherHub2 Quick Start Guide Table of Contents 1 Introduction... 1 2 Packing List... 1 3 Connections... 1 4 IP Addressing... 2 5 Browser Access... 3 6 System Info... 3 7 Weather Station Settings... 4
More informationLast updated: Copyright
Last updated: 2012-08-20 Copyright 2004-2012 plabel (v2.4) User s Manual by Bioinformatics Group, Institute of Computing Technology, Chinese Academy of Sciences Tel: 86-10-62601016 Email: zhangkun01@ict.ac.cn,
More informationQuantumMCA QuantumNaI QuantumGe QuantumGold
QuantumMCA QuantumNaI QuantumGe QuantumGold Berkeley Nucleonics Corporation (San Rafael, CA) and Princeton Gamma Tech (Princeton, NJ) have partnered to offer gamma spectroscopy with either germanium or
More informationSpectrometer User s Guide
Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.
More informationOverlaying GRIB data over NOAA APT weather satellite images using OpenCPN
Overlaying GRIB data over NOAA APT weather satellite images using OpenCPN I receive NOAA weather satellite images which are quite useful when used alone but with GRIB wind and rain data overlaid they can
More informationFree fall with an interface system
Related topics Motion along a straight line subject to constant acceleration, laws governing falling bodies, acceleration due to gravity. Principle and task The fall times t are measured for different
More informationO P E R A T I N G M A N U A L
OPERATING MANUAL WeatherJack OPERATING MANUAL 1-800-645-1061 The baud rate is 2400 ( 8 bits, 1 stop bit, no parity. Flow control = none) To make sure the unit is on line, send an X. the machine will respond
More informationMassHunter Software Overview
MassHunter Software Overview 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks A
More informationOrbit Support Pack for Excel. user manual
Orbit Support Pack for Excel user manual Information in this document is subject to change without notice. Companies, names and data used in examples herein are fictitious unless noted otherwise. No part
More informationCHEMISTRY 170. Radioisotopes
CHEMISTRY 170 Radioisotopes Positron Emission Tomography or PET scans use the radioisotope 18 F to create an image of the brain. DEPARTMENT OF CHEMISTRY UNIVERSITY OF KANSAS Radioisotopes Introduction
More informationTitriSoft 2.5. Content
Content TitriSoft 2.5... 1 Content... 2 General Remarks... 3 Requirements of TitriSoft 2.5... 4 Installation... 5 General Strategy... 7 Hardware Center... 10 Method Center... 13 Titration Center... 28
More informationNMR Predictor. Introduction
NMR Predictor This manual gives a walk-through on how to use the NMR Predictor: Introduction NMR Predictor QuickHelp NMR Predictor Overview Chemical features GUI features Usage Menu system File menu Edit
More informationUser Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: Date: November 04, 2013
User Manual User Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: 1.0.3 Date: November 04, 2013 This document is protected by copyright. Do not copy or publish this document or parts of it without
More informationOcean Optics Educational Spectrometers. Getting Started Manual
Ocean Optics Educational Spectrometers Getting Started Manual We here at Ocean Optics want you to know that we appreciate your business. To speak to a Technical Support representative or an Ocean Optics
More informationSOFTWARE USER MANUAL. Weather Capture Advance WS1640 WM9280
SOFTWARE USER MANUAL Weather Capture Advance WS1640 WM9280 1 TABLE OF CONTENTS 1. Introduction 3 1.1 System requirements 1.2 Connection of the weather station 1.3 Installation of the Weather Capture Advance
More informationJasco V-670 absorption spectrometer
Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions 1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source
More informationBest Pair II User Guide (V1.2)
Best Pair II User Guide (V1.2) Paul Rodman (paul@ilanga.com) and Jim Burrows (burrjaw@earthlink.net) Introduction Best Pair II is a port of Jim Burrows' BestPair DOS program for Macintosh and Windows computers.
More informationCHEMISTRY 130 General Chemistry I. Radioisotopes
CHEMISTRY 130 General Chemistry I Radioisotopes Positron Emission Tomography or PET scans use the radioisotope 18 F to create an image of the brain. DEPARTMENT OF CHEMISTRY UNIVERSITY OF KANSAS Radioisotopes
More informationPolarSync Quick Start
PolarSync Quick Start Installation and Use In this Quick Start guide, we will cover installing the PolarSync program and using it as a teacher, student or guest. I. Installing PolarSync... 1 II. Teacher
More informationPC Control / Touch Control
PC Control / Touch Control Version 6.0 / 5.840.0150 New Features Manual 8.840.8007EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
More informationMinimum (-31) every 10 s -35 (-31)
Humidity, and Dew Point Data Logger with LCD ORDERING INFORMATION Standard Data Logger (Data Logger, Software on CD, Battery) Replacement Battery EL-USB-2-LCD BAT 3V6 1/2AA FEATURES 0 to 100 measurement
More informationConnect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.
Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. The screen shown in Fig. 1 may be displayed. If status line displays
More informationMASSACHUSETTS INSTITUTE OF TECHNOLOGY Physics Department. Experiment 03: Work and Energy
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Physics Department Physics 8.01 Fall Term 2010 Experiment 03: Work and Energy Purpose of the Experiment: In this experiment you allow a cart to roll down an inclined
More informationUser's Manual altimeter V1.1
User's Manual altimeter V1.1 The altimeter is completely autonomous. It can be installed on any model. It automatically detects the beginning of flights and does not record the period between two consecutive
More informationMagnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor
General Physics Lab Department of PHYSICS YONSEI University Lab Manual (Lite) Magnetic Fields Ver.20181029 NOTICE This LITE version of manual includes only experimental procedures for easier reading on
More informationUniversity of Colorado Denver Anschutz Medical Campus Online Chemical Inventory System User s Manual
University of Colorado Denver Anschutz Medical Campus Online Chemical Inventory System User s Manual Hazardous Materials Division 303-724-0345 chemical.inventory@ucdenver.edu May, 2017 Table of Contents
More informationC101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science
C101-E112 BioSpec-nano Shimadzu Spectrophotometer for Life Science Power of small. BioSpec-nano BioSpec-nano Shimadzu Spectrophotometer for Life Science Quick and Simple Nucleic Acid Quantitation Drop-and-Click
More informationSt art. rp m. Km /h 1: : : : : : : : : : : : :5 2.5.
modified 22/05/14 t 3:5 2.5 3:5 5.0 3:5 7.5 4:0 0.0 4:0 2.5 4:0 5.0 4:0 7.5 4:1 0.0 4:1 2.5 4:1 5.0 4:1 7.5 4:2 0.0 Km /h 0 25 50 75 100 125 150 175 200 225 rp m 0 250 0 500 0 750 0 100 00 125 00 1 2 3
More informationMnova Software for Analyzing Reaction Monitoring NMR Spectra
Mnova Software for Analyzing Reaction Monitoring NMR Spectra Version 10 Chen Peng, PhD, VP of Business Development, US & China Mestrelab Research SL San Diego, CA, USA chen.peng@mestrelab.com 858.736.4563
More informationINTEGRAL. Science Operations Centre. Announcement of Opportunity for Observing Proposals (AO-11) INT/OAG/ /Dc Issue 1.
Science Operations Centre Announcement of Opportunity for Observing Proposals (AO-11) INT/OAG/13-0383/Dc Issue 1.0 25 February 2013 Prepared by Miguel Arregui, Emilio Salazar Date: 25 February 2013 Page:
More informationNEW HOLLAND IH AUSTRALIA. Machinery Market Information and Forecasting Portal *** Dealer User Guide Released August 2013 ***
NEW HOLLAND IH AUSTRALIA Machinery Market Information and Forecasting Portal *** Dealer User Guide Released August 2013 *** www.cnhportal.agriview.com.au Contents INTRODUCTION... 5 REQUIREMENTS... 6 NAVIGATION...
More informationFlash Point Calculation by UNIFAC
Flash Point Calculation by UNIFAC Short Introduction and Tutorial DDBSP - Dortmund Data Bank Software Package DDBST Software & Separation Technology GmbH Marie-Curie-Straße 10 D-26129 Oldenburg Tel.: +49
More informationAtomic and nuclear physics
Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Moseley s law and determination of the Rydberg constant P6.3.3.6 Objects of the experiment Measuring the K-absorption
More informationForecast User Manual FORECAST. User Manual. Version P a g e
FORECAST Version 1.0 1 P a g e Version Created By Created On Verified By Verified On Description No Draft Mr. Jayendrasinh 22/04/2016 Gohil 1.0 Mr. Jayendrasinh Gohil 23/04/2016 Mr. Hemal Patel 2 P a g
More informationDetermination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )
Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P2541001) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne
More informationAtomic and nuclear physics
Atomic and nuclear physics X-ray physics Attenuation of x-rays LEYBOLD Physics Leaflets P6.3.2.2 Investigating the wavelength dependency of the coefficient of attenuation Objects of the experiment To measure
More informationC2A for Pulsar2 how to control your telescope from C2A
C2A for Pulsar2 how to control your telescope from C2A C2a is a free and powerful planetarium program available from.http://www.astrosurf.com/c2a/english/ that offers native support for Pulsar2 without
More informationHow to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software
How to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software This is a concise guide to setting up and calibrating your Ocean Optics
More informationYou w i ll f ol l ow these st eps : Before opening files, the S c e n e panel is active.
You w i ll f ol l ow these st eps : A. O pen a n i m a g e s t a c k. B. Tr a c e t h e d e n d r i t e w i t h t h e user-guided m ode. C. D e t e c t t h e s p i n e s a u t o m a t i c a l l y. D. C
More informationCHEMICAL INVENTORY ENTRY GUIDE
CHEMICAL INVENTORY ENTRY GUIDE Version Date Comments 1 October 2013 Initial A. SUMMARY All chemicals located in research and instructional laboratories at George Mason University are required to be input
More informationFoodGuard-1 Food and Water Contamination Monitoring System for Microsoft Windows XP Professional SP3 and Windows 7 (FG-1-BW) System User s Manual
ORTEC FoodGuard-1 Food and Water Contamination Monitoring System for Microsoft Windows XP Professional SP3 and Windows 7 (FG-1-BW) System User s Manual Software Version 2 Printed in U.S.A. ORTEC Part No.
More informationStandards-Based Quantification in DTSA-II Part II
Standards-Based Quantification in DTSA-II Part II Nicholas W.M. Ritchie National Institute of Standards and Technology, Gaithersburg, MD 20899-8371 nicholas.ritchie@nist.gov Introduction This article is
More informationAlpha-Energies of different sources with Multi Channel Analyzer
Physical Structure of Matter Radioactivity Alpha-Energies of different sources with Multi Channel Analyzer What you can learn about Decay series Radioactive equilibrium Isotopic properties Decay energy
More informationSAFYR7 KAPPABRIDGE CONTROL SOFTWARE. User Manual
SAFYR7 KAPPABRIDGE CONTROL SOFTWARE User Manual Version 1.1 April 2018 Contents Contents 1 Introduction 1 2 Getting Started 3 2.1 System Requirements................................. 3 2.2 Program Installation..................................
More information2. To measure the emission lines in the hydrogen, helium and possibly other elemental spectra, and compare these to know values.
4.1. Purpose 1. To record several elemental emission spectra using arc lamps filled with each element using the Ocean Optics USB650 spectrometer. 2. To measure the emission lines in the hydrogen, helium
More informationClassroom Tips and Techniques: Series Expansions Robert J. Lopez Emeritus Professor of Mathematics and Maple Fellow Maplesoft
Introduction Classroom Tips and Techniques: Series Expansions Robert J. Lopez Emeritus Professor of Mathematics and Maple Fellow Maplesoft Maple has the ability to provide various series expansions and
More informationAssembly and Operation Manual. April 2016
Assembly and Operation Manual April 2016 Table of Contents What is in the OurWeather Box? 3 Step by Step Assembly 13 Building the Weather Sensors 18 Testing the OurWeather Weather Station 28 Power Up OurWeather
More informationWaters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012
Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012 Georgia Institute of Technology School of Chemistry & Biochemistry School of Materials Science
More informationVELA. Getting started with the VELA Versatile Laboratory Aid. Paul Vernon
VELA Getting started with the VELA Versatile Laboratory Aid Paul Vernon Contents Preface... 3 Setting up and using VELA... 4 Introduction... 4 Setting VELA up... 5 Programming VELA... 6 Uses of the Programs...
More informationDOAS measurements of Atmospheric Species
Practical Environmental Measurement Techniques: DOAS measurements of Atmospheric Species Last change of document: April 14, 2014 Supervisor: Dr. Andreas Richter, room U2090, tel 62103, and Dr. Folkard
More informationUniversity of Minnesota Nano Center Standard Operating Procedure
University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Zeta Potential Analyzer Model: Stabino Location: PAN 185 Badger Name: Not on Badger Revision Number: 0-Inital release Revisionist:
More informationFIELD SPECTROMETER QUICK-START GUIDE FOR FIELD DATA COLLECTION (LAST UPDATED 23MAR2011)
FIELD SPECTROMETER QUICK-START GUIDE FOR FIELD DATA COLLECTION (LAST UPDATED 23MAR2011) The ASD Inc FieldSpec Max spectrometer is a precision instrument designed for obtaining high spectral resolution
More informationp tot Mechanics LD Physics Leaflets
GENZ 2014-12 Mechanics Aerodynamics and hydrodynamics Measuring air resistance LD Physics Leaflets Measuring the air resistance as a function of the wind speed Measuring the wind speed with a pressure
More informationTitration of a strong acid with a strong base with Cobra4
Titration of a strong acid with a strong base with Cobra4 TEC Related topics Strong and weak acids and bases, ph value, titration curves, equivalence point, potentiometry. Principle Hydrochloric acid is
More informationOperation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword
More informationOperating Instructions for Digital Manometer. Model: MAN-SD
Operating Instructions for Digital Manometer Model: MAN-SD 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...4 5. Operating Principle...4 6. Mechanical Connection...5
More informationM E R C E R W I N WA L K T H R O U G H
H E A L T H W E A L T H C A R E E R WA L K T H R O U G H C L I E N T S O L U T I O N S T E A M T A B L E O F C O N T E N T 1. Login to the Tool 2 2. Published reports... 7 3. Select Results Criteria...
More informationTOP MARKET SURVEY INSTRUCTION SHEET. Requirements. Overview
INSTRUCTION SHEET TOP SURVEY TOP SURVEY INSTRUCTION SHEET Overview For nearly 40 years, the ACA has surveyed member agencies and conducted the Top Collection Market Survey. This survey provides critical
More informationGamma-Spectrum Generator
1st Advanced Training Course ITCM with NUCLEONICA, Karlsruhe, Germany, 22-24 April, 2009 1 Gamma-Spectrum Generator A.N. Berlizov ITU - Institute for Transuranium Elements Karlsruhe - Germany http://itu.jrc.ec.europa.eu/
More informationISSP User Guide CY3207ISSP. Revision C
CY3207ISSP ISSP User Guide Revision C Cypress Semiconductor 198 Champion Court San Jose, CA 95134-1709 Phone (USA): 800.858.1810 Phone (Intnl): 408.943.2600 http://www.cypress.com Copyrights Copyrights
More informationRADIATION PROCEDURES MANUAL Procedure Cover Sheet
RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Radioactive Material Inventory Procedure Number: TSO-09-16-REV 0 Effective Date: June 18, 2009 Approved By: Date: 11 August, 2009 Technical
More informationLatest generation testo air velocity & IAQ probes with cable handle. Instruction manual
Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of
More informationBecome a Microprobe Power User Part 2: Qualitative & Quantitative Analysis
Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Mike Spilde Spring IOM Seminar February 5, 2008 Qualitative Analysis Why use qualitative scans? Elemental ID (especially trace
More informationRTM2200- RPM EQF3200 EQF A²M4000
Manual RTM2200- RPM2200 - EQF3200 EQF3220 - A²M4000 November 2012 SARAD GmbH Tel.: ++49 (0)351 / 6580712 Wiesbadener Straße 10 FAX: ++49 (0)351 / 6580718 D-01159 Dresden e-mail: support@sarad.de GERMANY
More informationSuperCELL Data Programmer and ACTiSys IR Programmer User s Guide
SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide This page is intentionally left blank. SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide The ACTiSys IR Programmer and SuperCELL
More informationLife Cycle of Stars. Photometry of star clusters with SalsaJ. Authors: Daniel Duggan & Sarah Roberts
Photometry of star clusters with SalsaJ Authors: Daniel Duggan & Sarah Roberts Photometry of star clusters with SalsaJ Introduction Photometry is the measurement of the intensity or brightness of an astronomical
More informationBuilding Inflation Tables and CER Libraries
Building Inflation Tables and CER Libraries January 2007 Presented by James K. Johnson Tecolote Research, Inc. Copyright Tecolote Research, Inc. September 2006 Abstract Building Inflation Tables and CER
More informationLab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls
Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls OBJECTIVES - To perform air flow measurement using the wind tunnel. - To compare measured and theoretical velocities for various
More informationSteelSmart System Cold Formed Steel Design Software Download & Installation Instructions
Step 1 - Login or Create an Account at the ASI Portal: Login: https://portal.appliedscienceint.com/account/login Create Account: https://portal.appliedscienceint.com/account/register 2 0 1 7 A p p l i
More informationɣ-radiochromatography
nuclear instruments Kapitel_y_Graphy.indd 1 19.08.09 10:27 n u c l e a r i n s t r u m e n t s Index HPLC-detectors TLC-detectors GABI* analog output GABI* digital output to PC minigita*, 1 trace scanner
More informationNV-DVR09NET NV-DVR016NET
NV-DVR09NET NV-DVR016NET !,.,. :,.!,,.,!,,, CMOS/MOSFET. : 89/336/EEC, 93/68/EEC, 72/23/EEC,.,,. Novus Security Sp z o.o... 4 1. NV-DVR09NET NV-DVR016NET. 2.,. 3.,... 4... ( ) /. 5..... 6.,,.,. 7.,.. 8.,,.
More informationinaturalist Training AOP inaturalist Training May 21, 2016 Slide 1
inaturalist Training AOP inaturalist Training May 21, 2016 Slide 1 What is inaturalist? inaturalist is a free tool that allows people to record, share, and discuss their observations inaturalist is designed
More informationThe View Data module
The module Use to examine stored compound data (H, S, C p (T), G, etc.) in Compound type databases and list solutions and their constituents in Solution type databases. Table of contents Section 1 Section
More informationv Prerequisite Tutorials GSSHA WMS Basics Watershed Delineation using DEMs and 2D Grid Generation Time minutes
v. 10.1 WMS 10.1 Tutorial GSSHA WMS Basics Creating Feature Objects and Mapping Attributes to the 2D Grid Populate hydrologic parameters in a GSSHA model using land use and soil data Objectives This tutorial
More informationAlpha-energies of different sources with Multi Channel Analyzer (Item No.: P )
Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P2522015) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne Physik Experiment:
More informationTIMEZERO PROFESIONAL V3.2
TIMEZERO PROFESIONAL V3.2 Main New Features TIMEZERO PROFESIONAL V3.2-1 - WASSP G3 (S3 and F3) Compatibility:... - 3 - Misc. Features:... - 7 - TIMEZERO PROFESIONAL V3.2-2 - TIMEZERO Professional (with
More informationversion ISOL METHOD MASS SEPARATOR MASHA CHARGE STATES...5
version 6.5 Contents: 1. ISOL METHOD... 2 1.1. MASS SEPARATOR MASHA...3 1.2. CHARGE STATES...5 2. OTHER... 6 2.1. ELECTRIC DIPOLE: SEPARATION PLANE...6 2.2. WEDGE ANGLE DIALOG...6 2.3. MSP-144 MODIFICATION...7
More informationComputer simulation of radioactive decay
Computer simulation of radioactive decay y now you should have worked your way through the introduction to Maple, as well as the introduction to data analysis using Excel Now we will explore radioactive
More informationPC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS
XIX IMEKO World Congress Fundamental and Applied Metrology September 611, 009, Lisbon, Portugal PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS Adriana Vâlcu 1, Sterică Baicu 1 National Institute
More informationMDSS Functional Prototype Display System Preview April 2002
MDSS Functional Prototype Display System Preview April 2002 Bill Mahoney National Center for Atmospheric Research Images shown are valid as of 15 April 2002 NCAR MDSS Display System - Overview The MDSS
More informationLocation Intelligence Infrastructure Asset Management. Confirm. Confirm Mapping Link to ArcMap Version v18.00b.am
Location Intelligence Infrastructure Asset Management Confirm Confirm Mapping Link to ArcMap Version v18.00b.am Information in this document is subject to change without notice and does not represent a
More informationElectric Fields and Equipotentials
OBJECTIVE Electric Fields and Equipotentials To study and describe the two-dimensional electric field. To map the location of the equipotential surfaces around charged electrodes. To study the relationship
More informationDewpoint Meter Data Sheet / Instructions
Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment
More informationTechnical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3)
Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) 1.0 Purpose - This technical procedure shall be followed for the operation of the GRIM 3. 2.0 Scope - This procedure applies
More informationPure Component Equations
Pure Component Equations Fitting of Pure Component Equations DDBSP - Dortmund Data Bank Software Package DDBST Software & Separation Technology GmbH Marie-Curie-Straße 10 D-26129 Oldenburg Tel.: +49 (0)
More informationDetermination of the Equivalent Weight and the K a or K b for a Weak Acid or Base
INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.
More informationManual. Low Shear Viscometer LS 300. prorheo GmbH Bahnhofstr Althengstett Tel
Manual Low Shear Viscometer LS 300 prorheo GmbH Bahnhofstr. 38 75382 Althengstett Tel. +49-7051-92489-0 Office@proRheo.de www.prorheo.de 1. Introduction... 3 1.1. Measuring system... 3 1.2. Principle of
More informationMotion on a linear air track
Motion on a linear air track Introduction During the early part of the 17 th century, Galileo experimentally examined the concept of acceleration. One of his goals was to learn more about freely falling
More informationThe CSC Interface to Sky in Google Earth
The CSC Interface to Sky in Google Earth CSC Threads The CSC Interface to Sky in Google Earth 1 Table of Contents The CSC Interface to Sky in Google Earth - CSC Introduction How to access CSC data with
More informationQuantification of JEOL XPS Spectra from SpecSurf
Quantification of JEOL XPS Spectra from SpecSurf The quantification procedure used by the JEOL SpecSurf software involves modifying the Scofield cross-sections to account for both an energy dependency
More informationProviding Electronic Information on Waste Water Discharge Locations
Providing Electronic Information on Waste Water Discharge Locations Application Guidance Notes Environmental Protection Agency PO Box 3000, Johnstown Castle Estate, Co. Wexford Lo Call: 1890 335599 Telephone:
More information