ASSESMENT OF UNCERTAINTIES IN A DAILY-CHECK BEAM ANALYZER. J. Melgar, C. Martín, C. Montes, F. Sáez, E. De Sena, P. Gómez

Size: px
Start display at page:

Download "ASSESMENT OF UNCERTAINTIES IN A DAILY-CHECK BEAM ANALYZER. J. Melgar, C. Martín, C. Montes, F. Sáez, E. De Sena, P. Gómez"

Transcription

1 ASSESMENT OF UNCERTAINTIES IN A DAILY-CHECK BEAM ANALYZER J. Melgar, C. Martín, C. Montes, F. Sáez, E. De Sena, P. Gómez Servicio de Radiofísica y Protección Radiológica. Hospital Universitario de Salamanca. Pº San Vicente, Salamanca (Spain) Phone: mfbordes@usal.es Abstract. An evaluation of the theoretic uncertainties for the measurements of absorbed dose by the central ionization chamber of daily-check beam analyzer QC6 was made, with the aim of establishing an action level for the daily measurements of the cgy/mu factor in a Saturno-42 (GE-CGR) linac for low and high energy photons. Several factors were obtained from technical data reports of electrometer - chamber set. Other factors were evaluated from measurements. All these factors were used to estimate the type A and type B uncertainties. Uncertainties from measurements of thermometer and barometer were estimated too. Using the QC6 beam analyzer for the measurements, a relative measuring uncertainty of 1.1% was obtained, which is expressed at the level of two standard deviation (k=2). The tolerance for the cgy/mu calibration factor in a medical accelerator is established at ±2%, according to the Spanish regulations. So, the action level was set to a value of ± 3.1% to assure reasonably that the linac works properly. In addition, a reference level for the reading of the central ionization chamber of QC6 for both energies was found. This reference value was obtained with a relative standard deviation of about 0.6%. The previous action level has been applied over this reference value in order to set a tolerance range for the daily measurements. In order to verify that this tolerance range was consistent with the daily measurements in the linac, we checked that the measurements of the last year were included in the defined range. 1.Introduction In our department, we make the quality control of radiation beams from a Saturno-42 (GE-CGR) linac every day. We evaluate parameters like homogeneity, symmetry, cgy/mu factor and beam quality with the PTW-QC6 Plus beam analyzer. In this paper, we will only study the constancy of the cgy/mu factor for low and high energy photon beams (6 and 18 MV, respectively), with the aim of establishing an action level, i.e., an upper limit for the change of this factor, from the collected charge in the central ionization chamber of QC6 Plus, corrected by pressure and temperature. If the daily measure exceeds this action level, it is possible to assure that the linac does not work properly, and corrective actions should be taken to assure appropriate working of the linac. Although several authors have studied some kind of daily-check beam analyzer [1-4], only one of them estimates the associated uncertainties of these detectors. None of them sets an action level for daily measurements. 2. Description of the equipment and daily measurements According to technical data from this equipment, QC6 Plus has six ionization chambers of 0.54 cm 3 into a PMMA phantom in different geometrical positions. Five of them are placed at 5 mm depth from the surface; one of these chambers is in the centre of the phantom (D5), and the other four chambers (D1-D4) are in the two principal axis (at 8 cm from D5). The sixth chamber (D6) is in the centre of phantom at 27 mm depth from the surface. Daily positioning of equipment is done in a quick and easy way. QC6 Plus is connected to a PTW - MULTIDOS electrometer, and several PMMA sheets with different thickness (5 mm or 10 mm) are placed over QC6 Plus depending to the quality of the beam (1.5 cm for 6 MV and 3 cm for 18 MV). Also, there is a filter (PMMA or Pb) that can be placed between chambers D5 and D6, in order to evaluate the beam quality from readings of these chambers. In the case of photons, the Pb filter is used. The electrometer has a RS-232 output port, which allows to control its operation by means of a PC computer, with the QCWin application software. This application displays the measurements of 1

2 each ionization chamber of QC6 Plus in absolute mode or relative to central chamber values. QCWin also calculates symmetry coefficients on both axis, and homogeneity coefficients on the measurement plane from the readings of the five chambers in this plane. For the purpose of this paper, we only take into account the measurements of the central chamber D5, since its reading indicates the daily constancy of cgy/mu factor. QCWin application allows to introduce the values of pressure and temperature, so readings can be corrected by these parameters in order to avoid their influence in the collected charge. Measurement of atmospheric pressure is daily made with a portable precision barometer Lufft hpa, and temperature is measured with a digital thermometer Unitest Beha Full equipment is shown in fig. 1. FIG.1. Full equipment: Lufft barometer Unitest thermometer QC6 Plus Multidos electrometer For daily measurements, QC6 Plus is positioned at 100 cm from radiation source. A field size of 20 x 20 cm 2 at the isocenter is used. PMMA sheets with 1.5 cm of thickness for 6 MV or 3 cm of thickness for 18 MV are placed over detector surface. The equipment is irradiated with 200 MU per beam. 3. Establishment of the action level Action level is defined as the interval which amplitude is equal to the amplitude of the tolerance plus the expanded uncertainty, expressed at the level of two standard deviation (k=2). This uncertainty comes from the instrumentation and the method of measurement employed [5-6]. Therefore, it is important to delimit the uncertainty related to the equipment and the method. In addition, we must establish a reference value on which the action level is applied Analysis of uncertainties in measuring instruments and procedures We can separate the uncertainties associated to each instrument, i.e., QC6 Plus - MULTIDOS set, Lufft barometer and Unitest thermometer, to obtain the uncertainties associated to the following quantities: collected charge by chamber D5 of QC6 Plus (Q), atmospheric pressure (P), and temperature (T), respectively. The purpose is to find a total expanded uncertainty of Q* (being Q corrected by pressure and temperature effects, according to eq. 1) T (º C) Q * = Q (1) P( mb) Uncertainties in beam analyzer electrometer set measurements The main sources of uncertainty associated to QC6 Plus - electrometer set and method of measurement that have been considered are those shown in table I. Furthermore, the type of uncertainty, A or B (type A: uncertainties that can be evaluated from statistical methods; type B: evaluated from other means), the relative standard uncertainty for each one of them and the way of obtaining (experimental or not) are shown in this table too. Table I. Uncertainties associated to QC6 Plus - electrometer set. 2

3 Uncertainty Type of Relative standard uncertainty uncertainty (%) Way of obtaining Reproducibility A 0.07 Experimental Accuracy B 0.29 Reference manual Long term stability per year B 0.29 Reference manual Linearity B 0.29 Reference manual Leakage current B 0.03 Experimental Charge loss B Experimental Daily positioning B 0.15 Experimental All sources of uncertainty included in the reference manual of the equipment have been taken into account. A rectangular distribution have been considered in order to calculate the relative standard uncertainty [6-7]. Obtained values for leakage current and charge loss have been entered as uncertainty values [4]. Daily positioning is related to the uncertainty of position of chamber D5 with respect to the centre of radiation field in the daily placement of QC6 Plus. The value of this uncertainty is estimated placing the detector 2 mm besides and 2 mm underneath of the reference distance from radiation source (100 cm); then, the maximum difference of the measurements is modeled as lower and upper limits of a rectangular distribution. Another type of variations like displacement of detector D5 with respect to the centre of field have not been considered, since their influence is not relevant. From the values of relative standard uncertainty in table I, the expanded uncertainty [6-7] for Q is calculated, being equal to 0.53% Uncertainties in barometer measurements The main sources of uncertainty associated to Lufft barometer that have been considered are those shown in table II. Furthermore, the type of uncertainty, A or B, the relative standard uncertainty for each one of them and the way of obtaining (experimental or not) are shown in this table too. Table II. Uncertainties associated to barometer measurements. Type of Relative standard Uncertainty Way of obtaining uncertainty uncertainty (%) Accuracy B 0.05 Reference manual From the values of relative standard uncertainty in table II, the expanded uncertainty [6-7] for P is calculated, being equal to 0.05% Uncertainties in thermometer measurements The main sources of uncertainty associated to thermometer that have been considered are those shown in table III. Table III. Uncertainties associated to thermometer measurements. Type of Relative standard Uncertainty Way of obtaining uncertainty uncertainty (%) Accuracy B 0.2 Reference manual 3

4 From the values of relative standard uncertainty in table III, the expanded uncertainty [6-7] for T is calculated, being equal to 0.2%. The calculated values for uncertainties associated to the measurements of pressure and temperature are in agreement with the values estimated by other authors [6]. Resolution has not been taken into account as a source of uncertainty in none of these instruments, since we consider that manufacturer has included this parameter into the accuracy of the equipment in the reference manual Total combined uncertainty and expanded uncertainty. From combined uncertainties for parameters Q, P and T, the total combined uncertainty for parameter Q* is obtained, being equal to 0.57%. Therefore, the value of expanded uncertainty is 1.14%, which is expressed at the level of two standard deviation (k=2) Tolerance for cgy/mu factor The tolerance for the cgy/mu calibration factor in a medical accelerator is established at ± 2% according to the Spanish regulations [5] Reference value Date Table IV. Measurements to obtain a reference value. X 6 MV D5 values cgy/mu factor (f) D5 Values corrected for f=1 X 18 MV D5 values cgy/mu factor (f) D5 Values corrected for f= ,000 1,892 1,892 1,008 1,920 1, ,009 1,907 1,890 1,015 1,925 1, ,992 1,904 1,919 1,000 1,886 1, ,997 1,893 1,899 1,005 1,907 1, ,005 1,912 1,902 1,010 1,923 1, ,004 1,888 1,880 1,011 1,907 1, ,998 1,891 1,895 1,003 1,908 1, ,003 1,881 1,875 1,009 1,908 1, ,995 1,882 1,891 1,002 1,902 1, ,995 1,867 1,876 1,001 1,892 1, ,994 1,888 1,899 1,003 1,907 1, ,004 1,895 1,887 1,002 1,885 1, ,007 1,889 1,876 1,003 1,877 1, ,002 1,889 1,885 1,002 1,880 1, ,003 1,901 1,895 0,997 1,888 1, ,002 1,889 1,885 0,996 1,876 1, ,000 1,902 1,902 0,997 1,882 1,888 Average: 1,891 Average: 1,891 S.D.: 0,011 S.D.: 0,010 A reference value for the reading of the central chamber D5 of QC6 Plus for low and high energy photons (6 MV and 18 MV, respectively) was obtained. For this purpose, measurements of the 4

5 cgy/mu factor with an ionization chamber PTW (type Farmer) of 0.6 cm 3 were made, inserting the chamber into a water equivalent material. A relative standard deviation of about 0.6% was obtained. It is important to emphasize that this kind of measurement is made in our linac every Monday. In Table IV, values of cgy/mu factor measured at different days are shown. These values were obtained with an ionization chamber. These measurements are compared with those of QC6 measured the same day. The average value of the readings of the ionization chamber are corrected relative to f= This is taken as a reference value. 4. Results and conclusions From the calculated expanded uncertainty and the established tolerance, we can determine that the action level is ± 3.1%. Applying the action level to the reference value, a tolerance range for daily measurements of cgy/mu factor with QC6 Plus was established. To verify that the tolerance range is coherent with experimental values, measurements made over the last year were analyzed. All the measurements were within the interval. Measured values were out of action level only once. Corrective actions were taken, measuring the cgy/mu factor with a Farmer type ionization chamber for each beam quality and adjusting the gain of the monitor chambers. In Fig. 2, values of the readings of D5 on 2003 for both qualities are shown. Action levels and reference values are displayed in dashed lines. FIG.2. QC6 measurements on 2003 References 1. Lutz, W.R., Maddox, B.J., Kase, K.R., Daily check instrument for photon and electron beam quality assurance of medical linacs. Med. Phys., 12 (4); 462-5, (1985) 5

6 2. Muñoz, C., López, A., Gesto, C., Fernández, P., Arregui, G., Solana, V., Determinación de los rangos de variación aceptables de las medidas con el detector Hermes. Libro de Comunicaciones del VIII Congreso Nacional de la Sociedad Española de Protección Radiológica. Maspalomas, (2000) 3. Carmona, V., Lliso, F., Bea, J., Morata, B., Baños, M., Pérez-Calatayud, J., Experiencia con un nuevo analizador rápido del haz. Protocolo de controles asociado. Libro de Comunicaciones del XI Congreso Nacional de Física Médica. Valencia, (1997) 4. De Sena, E., Béjar, M.J., Cutanda, F., Martín J., Gómez, P.L., Herrera, M., Matilla, A., Puesta a punto de un equipo multidetector con cámaras de ionización para la verificación diaria de haces de radiación. Libro de Comunicaciones del XII Congreso de Física Médica. Santander, (1999) 5. Real Decreto 1566/1998, de 17 de Julio, por el que se establecen los criterios de calidad en radioterapia. B.O.E. nº 206 de 28 de Agosto; , (1998) 6. Curso de Metrología y calibración en radiaciones ionizantes y en el área de la Radiofísica Hospitalaria. Organizado por la SEFM. (2002). 7. Granados, C.E., Andreo, P., Brosed, A., Lizuain, M.C., Sáez, C., Gultresa, J., Incertidumbres y tolerancias de la dosimetría en radioterapia. Curso SEFM organizado en el Hospital Central de Asturias, (1997) 6

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo

More information

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Original Article COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Roseli T. Bulla 1, Linda V.E. Caldas 2 * Study developed at Instituto de Pesquisas Energéticas

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius

More information

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento

More information

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Reference Dosimetry for Megavoltage Therapy Beams: Electrons Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of

More information

NEUTOR: Neutrons Monitor for. Project financed by: Radiotherapy

NEUTOR: Neutrons Monitor for. Project financed by: Radiotherapy NEUTOR: Neutrons Monitor for Project financed by: Radiotherapy PURPOSES OF THE PROJECT To develop a Neutrons monitor for its application in radiotherapy, to control, in real time, the dose of neutrons

More information

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:

More information

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel

More information

Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator

Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Christoph Trauernicht Groote Schuur Hospital & University of Cape Town Method is based on: Background A method of

More information

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131 DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131 Autores: J.M. Jiménez 1, A.R. Cortés 2, P.Alcorta 2 E-mail: jmjimenez@hsan.osakidetza.net 1 Unidad de Radiofísica. Hospital Santiago Apóstol,

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting April 17, 2015 Larry DeWerd has partial interest in Standard Imaging Inc. Determination of your uncertainty

More information

A comparison of methods for monitoring photon beam energy constancy

A comparison of methods for monitoring photon beam energy constancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon

More information

8/4/2016. Determination of small field output factors, advantages and limitations of Monte Carlo simulations

8/4/2016. Determination of small field output factors, advantages and limitations of Monte Carlo simulations etermination of small field output factors, advantages and limitations of onte Carlo simulations AAP-APR-SEF Joint Scientific Symposium Josep Puxeu Vaqué Institut Català d Oncologia ICO Edinburgh Cancer

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom

Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom F.Gómez 1, L. Franco 1, A. Iglesias 1, J. Pardo 1, J. Pena 1, A. Rodríguez 1, R.Lobato 2, J. Mosquera

More information

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of a pencil ionization chamber with and without preamplifier Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.

More information

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Pirchio Rosana a*, Lindner Carlos a, Molina Laura a and Vallejos Matías a. a Comisión

More information

Progress in Nuclear Science and Technology, Volume 6,

Progress in Nuclear Science and Technology, Volume 6, DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi

More information

Measurement of induced radioactivity in air and water for medical accelerators

Measurement of induced radioactivity in air and water for medical accelerators Measurement of induced radioactivity in air and water for medical accelerators K. Masumoto 1, K. Takahashi 1, H. Nakamura 1, A. Toyoda 1, K. Iijima 1, K. Kosako 2, K. Oishi 2, F. Nobuhara 1 High Energy

More information

Visual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-Year Period

Visual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-Year Period International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2015, 4, 290-299 Published Online November 2015 in SciRes. http://www.scirp.org/journal/ijmpcero http://dx.doi.org/10.4236/ijmpcero.2015.44035

More information

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order

More information

Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams

Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams Ramamoorthy Ravichandran,

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

A study on the cost of concrete shielding in a standard radiotherapy facility room

A study on the cost of concrete shielding in a standard radiotherapy facility room BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 06-0 (018) 01-18 A study on the cost of concrete shielding in a standard radiotherapy facility room Eduardo de Paiva a a Instituto de Radioproteção e Dosimetria/Divisão

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

M. A. Navacerrada, C. Díaz, A. Pedrero and L. Iglesias

M. A. Navacerrada, C. Díaz, A. Pedrero and L. Iglesias Acústica 2008 20-22 de Outubro, Coimbra, Portugal Universidade de Coimbra CALCULUS OF THE UNCERTAINTY IN ACOUSTIC FIELD MEASUREMENTS: COMPARATIVE STUDY BETWEEN THE UNCERTAINTY PROPAGATION METHOD AND THE

More information

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Electron therapy Class 2: Review questions

Electron therapy Class 2: Review questions Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a

More information

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.83-87 (2) ARTICLE Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Toshioh FUJIBUCHI,2,*, Satoshi

More information

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA GROUP I SPRING 2014 KEVIN JORDAN GRADUATE STUDENT, RADIOLOGICAL PHYSICS KARMANOS CANCER CENTER WAYNE STATE UNIVERSITY

More information

Chapter 10 Acceptance Tests and Commissioning Measurements

Chapter 10 Acceptance Tests and Commissioning Measurements Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

Optimization of microionization chambers for small-field reference dosimetry

Optimization of microionization chambers for small-field reference dosimetry Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall

More information

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance Michael P. Barnes 1,2 a and Peter B. Greer 1,3

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer SLAC-PUB-8099 July 1999 Air Kerma Calibration Factors and k ch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL) N. E. Ipe*, K. E. Rosser,

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material

Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material Australas Phys Eng Sci Med (2012) 35:485 489 DOI 10.1007/s13246-012-0169-6 SCIENTIFIC NOTE Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material

More information

An introduction to IAEA TRS-483

An introduction to IAEA TRS-483 An introduction to IAEA TRS-483 P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital and Karolinska Institute, Stockholm, Sweden Journées Scientifiques de la SFPM Toulouse 2018

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Radiological impact of rutile covered welding electrodes

Radiological impact of rutile covered welding electrodes LABORATORIO DE MEDIDAS DE BAJA ACTIVIDAD Universidad del País Vasco / Euskal Herriko Unibertsitatea Radiological impact of rutile covered welding electrodes Margarita Herranz 1, Saroa Rozas 1, Carmen Pérez

More information

Inspection of Welds in Small Pipes by NDT Ultrasonic Phased-Array techniques. Detection, positioning, sizing and classification of defects.

Inspection of Welds in Small Pipes by NDT Ultrasonic Phased-Array techniques. Detection, positioning, sizing and classification of defects. More info about this article: http://www.ndt.net/?id=22924 Inspection of Welds in Small Pipes by NDT Ultrasonic Phased-Array techniques. Detection, positioning, sizing and classification of defects. J.

More information

Document downloaded from: This paper must be cited as:

Document downloaded from:  This paper must be cited as: Document downloaded from: http://hdl.handle.net/10251/47496 This paper must be cited as: Querol Vives, A.; Gallardo Bermell, S.; Ródenas Diago, J.; Verdú Martín, GJ. (2011). Parametric study of the X-ray

More information

) for Varian TrueBeam high-dose-rate therapy beams

) for Varian TrueBeam high-dose-rate therapy beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

PHYS 5020 Computation and Image Processing

PHYS 5020 Computation and Image Processing PHYS 5020 and Image Processing : Monte Carlo Thursday 2 August 2012 Monte Carlo (MC) is a numerical method that uses random sampling of probability distributions to simulate stochastic processes in nature,

More information

A Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy

A Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy Journal of Magnetics 19(1), 15-19 (2014) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 http://dx.doi.org/10.4283/jmag.2014.19.1.015 A Study on Effective Source-Skin Distance using Phantom in Electron

More information

Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling.

Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. 5 Sergei Zavgorodni 1,2, Eyad Alhakeem 2,1 and Reid Townson 2,1 1 Department of

More information

Manipulation on charged particle beam for RT benefit.

Manipulation on charged particle beam for RT benefit. High Electron Beam Dose Modification using Transverse Magnetic Fields Ion Chamber Response Modification under Strong Magnetic Field Conditions Sion Koren, Radiation Oncology Preface Manipulation on charged

More information

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator SLAC-PUB-7 190 June 1996 Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator K. R. Kase, X. S. Mao, W. R. Nelson, J. C. Liu Stanford Linear Accelerator Center, Stanford

More information

Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET)

Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET) Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET) Christina Jarlskog Department of Radiation Physics Lund University, Malmö University Hospital (UMAS)

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1) Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple

More information

Updating reference dosimetry a decade after TG-51

Updating reference dosimetry a decade after TG-51 Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual

More information

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Transport under magnetic fields with the EGSnrc simulation toolkit

Transport under magnetic fields with the EGSnrc simulation toolkit Transport under magnetic fields with the EGSnrc simulation toolkit Ernesto Mainegra-Hing, Frédéric Tessier, Blake Walters Measurement Science and Standards, National Research Council Canada Hugo Bouchard

More information

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,*

More information

Evaluation of the truebeam machine performance check (MPC): mechanical and collimation checks

Evaluation of the truebeam machine performance check (MPC): mechanical and collimation checks Received: 5 December 2016 Revised: 30 January 2017 Accepted: 14 February 2017 DOI: 10.1002/acm2.12072 RADIATION ONCOLOGY PHYSICS Evaluation of the truebeam machine performance check (MPC): mechanical and

More information

Magnetic removal of electron contamination in radiotherapy x-ray beams

Magnetic removal of electron contamination in radiotherapy x-ray beams University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2006 Magnetic removal of electron contamination in radiotherapy

More information

Composite field dosimetry

Composite field dosimetry Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism

More information

The volume effect of detectors in the dosimetry of small fields used in IMRT

The volume effect of detectors in the dosimetry of small fields used in IMRT The volume effect of detectors in the dosimetry of small fields used in IMRT Wolfram U. Laub a) Department of Medical Physics, University of Tübingen, Germany Tony Wong b) Department of Radiation Oncology,

More information

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017 Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018

More information

The lumped heat capacity method applied to target heating

The lumped heat capacity method applied to target heating INSTRUMENTATION Revista Mexicana de Física 59 (03) 38 334 JULY AUGUST 03 The lumped heat capacity method applied to target heating J. Rickards Instituto de Física, Universidad Nacional Autónoma de México,

More information

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES S. MORATÓ, A.BERNAL, A. QUEROL, A. ABARCA, C. GÓMEZ-ZARZUELA, R.MIRÓ, G.VERDÚ Institute for Industrial, Radiophysical and Environmental

More information

PHITS calculation of the radiation field in HIMAC BIO

PHITS calculation of the radiation field in HIMAC BIO PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,

More information

Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers

Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Klaus Derikum Physikalisch-Technische undesanstalt undesallee 100, 38116 raunschweig, Germany

More information

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry

More information

arxiv: v1 [astro-ph] 30 Jul 2008

arxiv: v1 [astro-ph] 30 Jul 2008 arxiv:0807.4824v1 [astro-ph] 30 Jul 2008 THE AIR-FLUORESCENCE YIELD F. Arqueros, F. Blanco, D. Garcia-Pinto, M. Ortiz and J. Rosado Departmento de Fisica Atomica, Molecular y Nuclear, Facultad de Ciencias

More information

Comparison of two commercial detector arrays for IMRT quality assurance

Comparison of two commercial detector arrays for IMRT quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 2, SPRING 2009 Comparison of two commercial detector arrays for IMRT quality assurance Jonathan G. Li, 1,a Guanghua Yan, 1,2 and Chihray Liu

More information

Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc.

Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc. Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc. Department of Radiation Physics, Radiation Medicine Program Princess Margaret Cancer Centre,

More information

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 COMPARISON OF ABSORBED DOSE TO WATER AT HIGH PHOTON ENERGIES AT A REFERENCE FACILITY BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 Bureau International des Poids et Mesures Pavillon

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron

More information

Monte Carlo uncertainty estimation for wavelength calibration Abstract

Monte Carlo uncertainty estimation for wavelength calibration Abstract Monte Carlo uncertainty estimation for wavelength calibration *Thang H.L., Nguyen D.D., Dung.D.N. Vietnam metrology institute 8 Hoang Quoc Viet street, Cau Giay district, Hanoi city, Vietnam * thanglh@vmi.gov.vn

More information

1 Introduction. A Monte Carlo study

1 Introduction. A Monte Carlo study Current Directions in Biomedical Engineering 2017; 3(2): 281 285 Sebastian Richter*, Stefan Pojtinger, David Mönnich, Oliver S. Dohm, and Daniela Thorwarth Influence of a transverse magnetic field on the

More information

ความร พ นฐานและเทคน คการบ าร งร กษา เคร องโดสคาล เบรเตอร (Dose Calibrator) ส ว ทย ป ณณช ยยะ

ความร พ นฐานและเทคน คการบ าร งร กษา เคร องโดสคาล เบรเตอร (Dose Calibrator) ส ว ทย ป ณณช ยยะ ความร พ นฐานและเทคน คการบ าร งร กษา เคร องโดสคาล เบรเตอร (Dose Calibrator) ส ว ทย ป ณณช ยยะ Principle & Maintenance Technique of Dose Calibrator Contents : 1. General principle 2. Basic system operation

More information

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.162 pissn 2508-4445, eissn 2508-4453 Secondary Neutron Dose Measurement for Proton Line Scanning

More information

Photon-beams monitor-unit calculations

Photon-beams monitor-unit calculations Photon-beams monitor-unit calculations Narayan Sahoo March 24, 2011 The materials included in this lecture notes are from previous lecture notes for this course by Karl Prado, Ph.D. Introduction Standard

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA The Underwater Spectrometric System Based on CZT Detector for Survey of the Bottom of MR Reactor Pool 13461 Victor Potapov, Alexey Safronov, Oleg Ivanov, Sergey Smirnov, Vyacheslav Stepanov National Research

More information

Conformity assessment of a high-performance liquid chromatography calibration.

Conformity assessment of a high-performance liquid chromatography calibration. 16 th International Congress of Metrology, 05003 (2013) DOI: 10.1051/ metrology/201305003 C Owned by the authors, published by EDP Sciences, 2013 Conformity assessment of a high-performance liquid chromatography

More information

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from Extensions to the Beta Secondary Standard BSS 2 Consolidated version from www.ptb.de R. Behrens 1 and G. Buchholz 2 Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany

More information

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Norio Saito, Tadahiro Kurosawa, Masahiro Kato, Yuichiro Morishita, Takahiro Tanaka, and Morihito

More information

IRPA 2 nd European Congress, Paris, May Session: RP in Medicine - Workers. Activation Products In Medical Linear Accelerators

IRPA 2 nd European Congress, Paris, May Session: RP in Medicine - Workers. Activation Products In Medical Linear Accelerators IRPA 2 nd European Congress, Paris, May 15-19 2006 Session: RP in Medicine - Workers Activation Products In Medical Linear Accelerators FISCHER Helmut W.*, TABOT Ben*,+, POPPE, Björn + * University of

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information