Commissioning of the Beta Secondary Standard (BSS2)
|
|
- Colleen Jefferson
- 5 years ago
- Views:
Transcription
1 Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria, 0400, South Africa wthoka@nmisa.org, Phone: , Fax: Abstract The National Metrology Institute of South Africa (NMISA) Dosimetry Section through NMISA s recapitalization procured a beta secondary standard irradiation facility type 2 (BSS 2). Beta radiation consists of electrons generated in the nucleus by the decay of a neutron into a proton (or vice versa), an electron and a neutrino. The calibration quantities are H'(0.07) and Hp (0.07) as operational quantities to estimate the local skin dose and H'(3) and Hp (3) as operational quantities to estimate the dose to the lens of the eye. The system was commissioned by verification of the personal dose equivalent (Hp (0.07), and software calculated time was checked for correctness using a calibrated stopwatch. An electronic personal dosimeter (EPD) was used to measure the personal dose equivalent (PDE) for 85 Kr and 90 Sr + 90 Y using polymethylmethacrylate (PPMA) and water phantoms with the results for both irradiation medium presented on this paper. Keywords: Beta radiation, Beta Secondary Standard, Dose Equivalent. 1. Introduction The dosimetry section has been using the irradiation system developed by National Physical Laboratory (NPL) with two beta emitting radioactive sources, 85 Kr and 90 Sr + 90 Y. The system has been mainly used for irradiation of Thermoluminiscent Dosimeters (TLDs) for the industry. TLDs irradiated by the laboratory are mostly used for monitoring radiation exposures to radiation workers. The TLDs are irradiated in a tissue equivalent slab phantom in a field of 85 Kr and 90 Sr + 90 Y beta radiation sources. The system has been maintained through the calibration of the beta sources at 5 yearly intervals at the primary standard laboratory and also use of EPDs for checks in between calibrations. As the irradiator was already old and some of the safety features were no longer functioning, posing a huge safety risk to employees, a decision was made to replace the system. This resulted in a purchase in 2014 of the Beta Secondary Standard 2 (BSS 2) system traceable to the National Metrology Institute of Germany, Physikalisch-Technische Bundesanstalt (PTB). The system offers capability to measure absorbed dose to tissue using beta standards. The beta secondary standard irradiation facility type 2, BSS 2 is an example of the concept of a calibrated secondary standards built as complete irradiation facilities including the instruments necessary to perform air pressure, temperature and humidity corrections. It uses long lived and high-purity radiation sources to produce a secondary reference radiation field and includes a traceable calibration. It consists of three sources of the nuclides 147 Pm, 85 Kr and 90 Sr + 90 Y according to series 1 of ISO [1]. The measuring quantities are Hp (0.07), Hp (3) and H'(0.07). The physical reason for the inclusion of Hp (3) without any additional calibration is that the depth-dose curves are nearly identical for the sources of the BSS 2 [2].
2 The new software also gives detailed information on the photon contribution from the source and due to bremsstrahlung generation. The complete irradiation facility provides, after traceable calibration, the actual dose or doserate value corrected for the influence of air pressure, temperature and humidity [1]. 2. Verification Requirements Once a BSS 2 system with sources traceable to primary standard laboratory together with the recommended beta secondary standard chamber has been purchased, an SSDL is ready for calibrating beta instruments. However before any irradiation facility is used, verification, using a suitable survey instrument or even an electronic personal dosimeter with the capability of beta radiation measurement is recommended. The aim of this work is not to perform a validation but only to assure the constancy of the field as well as detecting defects in the sealing of the beta sources. Since in our case we do not have a beta secondary standard, an EPD was used for verification of the dose and/or dose rate produced by the beta sources purchased with the system. The recommended reference instrument to be used for calibration of beta detecting instrument by secondary standard laboratories is a secondary standard ionisation chambers. This chamber fulfils the requirement of ISO on secondary standard instruments for beta radiation [3]. The calibration setup should strictly follow the terms given for the calibration facility as recommended by ISO The BSS 2 is equipped with a turntable and necessary distance adjustment tools to perform the required mechanical set up to ensure correct positioning. Reference radiation fields are described in ISO 6980 parts 1 to 3 [4, 5, 6] and series 1 secondary reference radiation field is used in this report as recommended on the ISO Series 1 consists of three sources of the nuclides 147 Pm, 85 Kr and 90 Sr + 90 Y and beam flattening filters. The measuring quantities are Hp(0.07), Hp(3) and H'(0.07). Hp(0.07) is the only quantity measured in this paper. 3. Equipment and Methods 3.1 BSS 2 System Components The following standards and equipment were used during the measurements with the BSS 2 system: 85 Kr source, 90 Sr + 90 Y source and 147 Pm source, each source is mounted on a separate shielded holder with a shutter to fit in the source stand. Three beam flattening filters, one for each of the three sources with filter identity cards according to their matching sources. The filters are held in place by means of four wires with metal tags that fit into grooves in the metal support rods. Irradiating device; a source stand mounted on a linear guiding. Fixed distances (11, 20, 30 and 50 cm) can be seen on a linear guiding. Beam axis pointer that can be screwed into the source position (in the absence of the source). The pointer is 300 mm long and 4 mm in diameter steel rod that passes through the centre of the guide. Control unit; the control unit contains all electronic components necessary to operate the BSS2 system including the sensor for atmospheric pressure. A separate sensor for measuring temperature and humidity is connected to the control unit using a cable. The
3 control unit is connected to the linear guiding or source stand by a special cable (maximum length 5 m). Built-in software to calculate the desired dose or dose rate. A Perspex phantom measuring 30 cm x 30 cm x 15 cm. A water phanthom. 3.2 Methods A calibrated Thermo Scientific, Mk 2.3 EPD with energy response from 250 kev to 1,5 MeV and an accuracy of 20% for 90 Sr + 90 Y and 85 Kr was used. This type of EPD was unable to detect radiation from 147 Pm as this radionuclide has an energy of KeV so its use was only limited to 90 Sr and 85 Kr. The time calculated by the software was verified using a hand calculated time from the calibration certificate of the radioactive sources. A hand timer was simultaneously used with the system time to check for correctness. Once the operation correctness was verified, an EPD was irradiated with a personal dose equivalent Hp(0.07) of 3 msv, 5 msv and 10 msv. The irradiation were performed at different mediums, a 30 cm x 30 cm x 15 cm PPMA and the 30 cm x 30 cm x 15 cm ISO calibration water phantom which complies with the standard ISO 4037 part 3, which is suitable for calibrations and type tests of personal dosimeters, because they measure the quantities Hp(0.07) and Hp(10). The irradiations were performed at 30 cm from the source at a 0º incident angle. The irradiation was repeated 10 times to check repeatability and reproducibility. 4. Uncertainty of Measurements on the BSS 2 System The complete irradiation facility provides, after traceable calibration of the radioactive sources, the actual dose or doserate value corrected for the influence of air pressure, temperature and humidity. It also gives the associated uncertainty determined according to the Guide to the Uncertainty in Measurement (GUM) and includes extensive quality control to assure every parameter used for the determination of the dose or doserate value [7]. This however depends on the level the facility will operate. To determine the absolute dose rate, the user needs calibrated thermometer, barometer and the humidity measuring device. These must then not be part of the facility if they cannot be independently calibrated. For this work, attention has only been paid on the acceptance of the measured personal dose equivalent in conforming to the 20% accuracy of the EPD. 5. Results Tables 1 and 2, show personal dose equivalent results for Kr-85 and Sr-90 in two different mediums. Column 1 shows the given Hp (0.07), columns 2 and 4 show the measured Hp (0.07) and columns 3 and 5 show the percentage difference. Figures 1 to 4 show the repeatability and/or reproducibility of the measurements taken. The verification of the system timer with a model 500 Stopwatch manufactured by Sanji also yielded satisfactory results.
4 Dose Equivalent (msv) Table 1: Personal dose equivalent EPD results on the ISO water phantom for BSS2 system ISO Water Phantom Medium 85 Kr 90 Sr + 90 Y Hp(0.07) % Given Hp(0.07) (msv) Hp(0.07)(mSv) % Difference (msv) Difference 3,0 2,82 6,63 3,35 11,40 5,0 4,73 5,61 5,61 12,05 10,0 9,43 5,87 11,89 18,73 Table 2: Personal dose equivalent EPD results on the PPMA phantom for BSS 2 system PPMA Phantom Medium Given Kr 90 Sr + 90 Y Hp(0.07) Hp(0.07) % Difference Hp(0.07)(mSv) % Difference (msv) (msv) 3,0 2,83 6,12 3,24 7,66 5,0 4,76 5,11 5,32 6,19 10,0 10,29 2,80 10,70 6,92 Repeatability Chart 85 Kr PPMA # of Measurements 3 msv 5 msv 10 msv UL 10 msv UL 5 msv LL 5 msv UL 3 msv LL 3 msv Figure 1. A plot showing three personal equivalent doses Hp(0.07) 3 msv, 5 msv and 10 msv with upper and lower limits of ± 20 % for 85 Kr in PPMA phantom. The maximum standard deviation across the doses was 0,07 msv, achieved at 10 msv.
5 Dose Equivalent (msv) Repeatability Chart 90 Sr + 90 Y PPMA # of Measurements 3 msv 5 msv 10 msv UL 10 msv UL 5 msv UL 3 msv LL 3 msv Figure 2. A plot showing three personal equivalent doses Hp(0.07) 3 msv, 5 msv and 10 msv with upper and lower limits of ± 20 % for 90 Sr + 90 Y in PPMA phantom. The maximum standard deviation across the doses was 0,18 msv, achieved at 10 msv.
6 Dose Equivalent (msv) Repeatability Chart 85 Kr Water # of Measurements 3 msv 5 msv 10 msv UL 10 msv UL 5 msv LL 5 msv UL 3 msv LL 3 msv Figure 3. A plot showing three personal equivalent doses Hp(0.07) 3 msv, 5 msv and 10 msv with upper and lower limits of ± 20 % for Kr-85 in ISO water phantom. The maximum standard deviation across the doses was 0,16 msv, achieved at 10 msv
7 Dose Equivalent (msv) Repeatability Chart 90 Sr + 90 Y Water # of Measurements 3 msv 5 msv 10 msv UL 10 msv UL 5 msv LL 5 msv UL 3 msv LL 3 msv Figure 4. A plot showing three personal equivalent doses Hp(0.07) 3 msv, 5 msv and 10 msv with upper and lower limits of ± 20 % for 90 Sr + 90 Y in ISO water phantom. The maximum standard deviation across the doses was 0,12 msv, achieved at 10 msv.
8 6. Discussions and Conclusions For the purpose of this work, the results achieved were satisfactory. An EPD was used to measure the personal equivalent dose on the two beta sources 85 Kr and 90 Sr + 90 Y, and all measurements taken were within the ± 20% limits which is the most accuracy an EPD used can manage. The results for 85 Kr in both mediums are well within the limits (maximum of 7 %) whereas the 90 Sr + 90 Y results even though still within the 20 % limits, they are closer to the upper limit. The reason for this needs further investigation as the measurement data taken for this work is insufficient to conclude on this. This is more important especially on deciding what phantom to use for customer work. The BSS2 system has thus been verified and can further be used as the previous beta system has been used, that is irradiating customers dosimeters. Output measurements will however be taken at least twice a month until enough data is achieved to make a conclusion on the reason of the performance of the two sources used, with 90 Sr + 90 Y giving results closer to the 20 % limit on the ISO water slab phantom. Measurements will also be taken on a 10 mm PPMA as is a recommendation from ISO This will allow for a better comparison of the suitability of phantoms. Further work will be embarked on to measure absolute dose rate with the extrapolation chamber. This will be in preparation of the participation on the inter-comparison planned for For work that needs to be performed with the system, the system uncertainty with routine quality control suffices. An investigation will also be made into whether there is a need to purchase a survey meter capable of measuring the energy range of the radioactive sources used. 7. Acknowledgements The authors of this paper would like thank NMISA management for providing the technical infrastructure and support.
9 8. References 1. P. Ambrosi, G. Buchholz, K. Helmstadter, The PTB Secondary Standard BSS 2 for radiation protection. 2. Brunzendorf, J. Determination of Depth-Dose Curves in Beta Dosimetry, Radiat. Prot. Dosim. 151, (2012) Behrens, et al, International comparison EUROMET.RI(I)-S2 of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation (EUROMET project No 739): Final report, Metrologia 44 (2007), Techn. Suppl (doi: / /44/1A/06003). 4. INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Nuclear Energy- Reference beta particle radiation Part 1: Methods of production, ISO , Geneva (2006). 5. INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Nuclear Energy- Reference beta particle radiation Part 2: Calibration fundamentals related to basic quantities characterizing the radiation field, ISO , Geneva (2004). 6. INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Nuclear Energy- Reference beta particle radiation Part 3: Calibration of area and personal dosimeters and determination of their response as a function of energy and angle of incidence, ISO , Geneva (2006). 7. INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Uncertainty of measurement Part 3: Guide to the expression of uncertainty in measurement, ISO/IEC Guide 98-3, Geneva (2008).
Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements
Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected
More informationExtensions to the Beta Secondary Standard BSS 2 Consolidated version from
Extensions to the Beta Secondary Standard BSS 2 Consolidated version from www.ptb.de R. Behrens 1 and G. Buchholz 2 Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany
More informationHp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015
Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015 Leon de Prez and Frans Bader VSL - The Netherlands www.vsl.nl P 1 Contents - Introduction to VSL and Ionizing
More informationVolume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved
Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah
More informationCONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS
CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS F. SCARLAT 1, A. SCARISOREANU 1, M. OANE 1, E. BADITA 1, E. MITRU 1 1 National Institute for Laser, Plasma and Radiation
More informationReliability and Traceability in Radiation Calibration
Reliability and Traceability in Calibration Sadao Nakashima Teruo Ebato 1. Introduction Fuji Electric has a successful history of delivering radiation measuring equipment used in the controlled areas of
More informationAccuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi
Accuracy, Calibration, Type Testing and Traceability General [Chapter 6 & 7 of RP 160] Peter Ambrosi EURADOS Training Course Lisbon, Portugal, 18-22 Mai Content Introduction - Aim of Measurement & type
More informationAn extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter
An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter Christopher A Perks *a and Stephane Faugoin a a LANDAUER EUROPE, 33 Avenue du General Leclerc, F-92266, Fontenay-aux-Roses
More informationInstructions for Use for the transfer chamber and the electronic measuring device Revision 1
EURAMET supplementary comparison of the personal dose equivalent at 0.07 mm and 3 mm depth, H p (0.07) and H p (3), for beta radiation EURAMET Project No. 1398 BIPM KCDB: EURAMET.R(I)-S16 Instructions
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 6980-2 Première edition 2004-10-15 Nuclear energy Reference beta-particle radiation Part 2: Calibration fundamentals related to basic quantities characterizing the radiation
More informationCalibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co
Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co Pirchio Rosana a*, Lindner Carlos a, Molina Laura a and Vallejos Matías a. a Comisión
More informationDetermination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator
Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,
More informationLaboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)
PROPOSAL FOR EYE-LENS DOSEMETER CALIBRATION AND TYPE TESTING ORAMED WP2 J.-M. Bordy (1), J. Daures (1), M. Denozière (1), G. Gualdrini (2), M. Guijaume (3), E. Carinou (4),F. Vanhavere (5) (1) Laboratoire
More informationHigh-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
More informationIon-Chamber Survey Meter OD-02
Ion-Chamber Survey Meter OD-02 Dose and dose rate meter for measuring the ambient dose equivalent H*(10) and dose rate equivalent dh*(10)/dt as well as the directional dose equivalent H (0,07) and dose
More informationSurvey Meter OD-01 Address: Phone: Fax: URL:
Survey Meter OD-01 Dose meter and dose rate meter for the measurement of the ambient dose and dose rate equivalent H*(10), dh*(10)/dt and the directional dose and dose rate equivalent H'(0.07), dh'(0.07)/dt
More informationTITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137
TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;
More informationEnergy response for high-energy neutrons of multi-functional electronic personal dosemeter
Energy response for high-energy neutrons of multi-functional electronic personal dosemeter T. Nunomiya 1, T. Ishikura 1, O. Ueda 1, N. Tsujimura 2,, M. Sasaki 2,, T. Nakamura 1,2 1 Fuji Electric Systems
More informationM [scale units/s] of the system
APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers
More informationINTERCOMPARISON OF PERFORMANCE CHARACTERISTICS OF OSLDs. AND TLDs USED FOR INDIVIDUAL MONITORING
INTERCOMPARISON OF PERFORMANCE CHARACTERISTICS OF OSLDs AND TLDs USED FOR INDIVIDUAL MONITORING THIS THESIS IS SUBMITTED TO THE DEPARTMENT OF MEDICAL PHYSICS SCHOOL OF NUCLEAR AND ALLIED SCIENCES UNIVERSITY
More informationQuality assurance for sensors at the Deutscher Wetterdienst (DWD)
Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations
More informationImplementing the ISO Reference Radiations at a Brazilian Calibration Laboratory
Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry
More informationTechnical basis of occupational dosimetrygeneral
Technical basis of occupational dosimetrygeneral review Antti Kosunen, Hannu Järvinen STUK 2017 NACP-RPC COURSE Occupational dosimetry in interventional radiology, - cardiology and nuclear medicine 27-29
More informationGLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY
GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad
More informationPERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER
SLAC-PUB-95-6749 (March 1995) PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER J. C. Liu, D. Busick 1, K. R. Kase, R. C. McCall 2, R. Sit and H. Tran 3 Stanford Linear
More informationAkira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency
The 3rd International Symposium on the System of Radiological Protection Seoul, Korea October 20-22, 2015 Akira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency ICRP Publication
More informationABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationTechnical Protocol of the CIPM Key Comparison CCAUV.V-K5
Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative
More information12/18/2016. Radioanalysis Laboratory Capabilities and Issues. Eleventh Annual Radiation Measurement Cross Calibration Workshop RMCC XI
1. Introduction 2. Laboratories Gamma spectrometry laboratory. Radiochemistry laboratory. Analytical chemistry laboratory. Alpha spectrometry laboratory Gross alpha / beta laboratory. Neutron activation
More informationComparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017
Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018
More informationSecondary Particles Produced by Hadron Therapy
Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad
More informationMeasurement of Tritium in Helium
detect and identify Measurement of Tritium in Helium Dr. Alfred Klett Berthold Technologies, Bad Wildbad, Germany 22 nd Annual Air Monitoring Users Group (AMUG) Meeting Palace Station Hotel, Las Vegas,
More informationAbstract. 1. Introduction
The importance of establishing a very-high temperature radiation thermometry measurement capability at the National Metrology Institute of South Africa (NMISA) Speaker: Dr Efrem K Ejigu National Metrology
More informationUse of a radioactive check device for redundancy check of ionization chambers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 13168 First edition 2015-07-01 Water quality Simultaneous determination of tritium and carbon 14 activities Test method using liquid scintillation counting Qualité de l eau Détermination
More informationPersonal Dose Monitoring System
Personal Dose Monitoring System Kei Aoyama Osamu Ueda Takeshi Kawamura 1. Introduction The level of radiation safety control in the nuclear power field and at facilities that utilize radiation has become
More informationDose Rate Levels around Industrial Gamma Sources
Dose Rate Levels around Industrial Gamma Sources Héctor René Vega-Carrillo *, Ricardo Rodríguez-Juárez, Eduardo Manzanares- Acuña, Rubén Hernández-Villasana, Jaime Ramírez-González Universidad Autónoma
More informationDesign, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors
6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator
More informationMonitoring of Ionizing Radiations facilities Experience and challenge
Monitoring of Ionizing Radiations facilities Experience and challenge Lebanese Atomic Energy Commission The LAEC was established in 1996 with the full support and assistance of the IAEA, having the mandate
More informationIntroduction. Neutron Effects NSEU. Neutron Testing Basics User Requirements Conclusions
Introduction Neutron Effects Displacement Damage NSEU Total Ionizing Dose Neutron Testing Basics User Requirements Conclusions 1 Neutron Effects: Displacement Damage Neutrons lose their energy in semiconducting
More informationCalibration of a pencil ionization chamber with and without preamplifier
Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof.
More informationNeutron Dosimetry with Ion Chamber-Based DIS System
Neutron Dosimetry with Ion Chamber-Based DIS System Christian Wernli 1, Annette Fiechtner 1, Jukka Kahilainen 2 1 Paul Scherrer Institute, Villigen, Switzerland 2 RADOS Technology Oy, Turku, Finland INTRODUCTION
More informationConsistency and confidence: the role of the National Measurement System
Consistency and confidence: the role of the National Measurement System Dr. Martyn Sené Deputy Director, NPL 30 th June, 2011 Welcome to the National Physical Laboratory Consistent measurement - important
More informationMitigation of External Radiation Exposures
Mitigation of External Radiation Exposures The three (3) major principles to assist with maintaining doses ALARA are :- 1) Time Minimizing the time of exposure directly reduces radiation dose. 2) Distance
More informationRadiation Detection. 15 th Annual OSC Readiness Training Program.
Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness
More informationA new neutron monitor for pulsed fields at high-energy accelerators
A new neutron monitor for pulsed fields at high-energy accelerators Marlies Luszik-Bhadra *, Eike Hohmann Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116, Braunschweig, Germany. Abstract.
More informationDosimetry. Sanja Dolanski Babić May, 2018.
Dosimetry Sanja Dolanski Babić May, 2018. What s the difference between radiation and radioactivity? Radiation - the process of emitting energy as waves or particles, and the radiated energy Radioactivity
More informationA Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators
A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,
More informationResearch Physicist Field of Nuclear physics and Detector physics. Developing detector for radiation fields around particle accelerators using:
Christopher Cassell Research Physicist Field of Nuclear physics and Detector physics Developing detector for radiation fields around particle accelerators using: Experimental data Geant4 Monte Carlo Simulations
More informationSimulation of Personal Protective Equipment Exposure to Radioactive Particulates. A Master s Level Submission.
Simulation of Personal Protective Equipment Exposure to Radioactive Particulates M. Roeterink 1*, E.F.G. Dickson 1, P. Bodurtha 1, and E.C. Corcoran 1 1 Royal Military College of Canada, Ontario, Canada
More informationIonizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI
Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio
More informationQuality assurance for sensors at the Deutscher Wetterdienst (DWD)
Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable
More informationInternational comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation
International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation EUROMET project No. 739 BIPM KCDB: EUROMET.RI(I)-S2 Final Report R. Behrens, K. Helmstädter,
More informationCOMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD
COMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD Kirill SKOVORODKO, Birute GRICIENE, Milda PETKELYTE Radiation Protection Division, Vilnius University Hospital Santaros
More informationDepartment of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency
Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency Facility of Radiation Standards It is important that radiation measuring instruments are calibrated by
More informationEmployees Contractors
Attachment Exposure Dose Distribution 1. Effective Dose from External Exposure Table 1 shows the distribution of external exposure dose of workers who were involved in radiation work at the Fukushima Daiichi
More informationTA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1
TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order
More informationComparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv
EUROMET 545: Final report 25/9/28 page 1 of 48 Comparison of national air kerma standards for ISO 437 narrow spectrum series in the range 3 kv to 3 kv L. Büermann (a), M. O Brien (b), D. Butler (c), I.
More informationTHE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*
SLAC-PUB-5122 Rev March 1991 w THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR* N. E. Ipe, R. J. Donahue, and D. D. Busick Stanford Linear Accelerator Center Stanford University,
More informationEnergy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI
Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration s of KAERI B.H. Kim 1, J.L. Kim 1, S.Y. Chang 1, J.K. Chang 1, G. Cho 2 1 Korea
More informationReferensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden
Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties
More informationISO Radiological protection Sealed radioactive sources General requirements and classification
INTERNATIONAL STANDARD ISO 2919 Third edition 2012-02-15 Radiological protection Sealed radioactive sources General requirements and classification Radioprotection Sources radioactives scellées Exigences
More informationEmployees Contractors
Attachment Exposure Dose Distribution 1. Effective Dose from External Exposure Table 1 shows the distribution of external exposure dose of workers who were involved in radiation work at the Fukushima Daiichi
More informationOctober 2017 November Employees Contractors
Attachment Exposure Dose Distribution 1. Effective Dose from External Exposure Table 1 shows the distribution of external exposure dose of workers who were involved in radiation work at the Fukushima Daiichi
More informationAppendix B1. Reports of SMU
Appendix B1 Reports of SMU Euromet Project 600 Comparison of Surface Roughness Standards /13 EUROMET SUPPLEMENTARY COMPARISON SURFACE TEXTURE Project No. 600 Final Report Elaborated by: M. Szmicskova Bratislava,
More informationDetermination of particle size distribution Single particle light interaction methods. Part 4:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne
More informationISO Water quality Measurement of polonium 210 activity concentration in water by alpha spectrometry
INTERNATIONAL STANDARD ISO 13161 First edition 2011-10-01 Water quality Measurement of polonium 210 activity concentration in water by alpha spectrometry Qualité de l eau Mesurage de l activité du polonium
More informationISO INTERNATIONAL STANDARD. Reference radiation fields Simulated workplace neutron fields Part 1: Characteristics and methods of production
INTERNATIONAL STANDARD ISO 12789-1 First edition 2008-03-01 Reference radiation fields Simulated workplace neutron fields Part 1: Characteristics and methods of production Champs de rayonnement de référence
More informationEuropean Project Metrology for Radioactive Waste Management
European Project Metrology for Radioactive Waste Management Petr Kovar Czech Metrology Institute Okruzni 31 638 00, Brno, Czech republic pkovar@cmi.cz Jiri Suran Czech Metrology Institute Okruzni 31 638
More informationRadiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007
Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp. 229 233 Advance Access publication 11 May 2007 doi:10.1093/rpd/ncm047 CHARACTERIZATION AND UTILIZATION OF A BONNER SPHERE SET BASED ON GOLD
More informationInternational Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories
International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED
More informationNuclear Spectroscopy: Radioactivity and Half Life
Particle and Spectroscopy: and Half Life 02/08/2018 My Office Hours: Thursday 1:00-3:00 PM 212 Keen Building Outline 1 2 3 4 5 Some nuclei are unstable and decay spontaneously into two or more particles.
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 14850-1 First edition 2004-05-15 Nuclear energy Waste-packages activity measurement Part 1: High-resolution gamma spectrometry in integral mode with open geometry Énergie nucléaire
More informationINTERNAL RADIATION DOSIMETRY
INTERNAL RADIATION DOSIMETRY Introduction to Internal Dosimetry Importance of External and Internal Exposures by Radiation Type Charged particle radiation (α, p, β) Generally important only for internal
More informationRadiation Protection & Radiation Therapy
Radiation Protection & Radiation Therapy For Medical Students Professor of Medical Physics Radiation Units Activity Number disintegrations per second (Curie, Becquerel) Exposure (Roentgen, C/kg) Absorbed
More informationRecent Primary Standardisations and Traceability in Nuclear Medicine. Lena Johansson, NPL Radioactivity Group
Recent Primary Standardisations and Traceability in Nuclear Medicine Lena Johansson, NPL Radioactivity Group Outline Primary standards Introduction to primary standards of radioactivity and traceability
More informationUpdate on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)
Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS) K. Garrow 1, B.J. Lewis 2, L.G.I. Bennett 2, M.B. Smith, 1 H. Ing, 1 R. Nolte, 3 S. Röttger, R 3 R. Smit 4
More informationGlossary of Terms* BIOASSAY: Assay and measurement procedures used to determine the amount of radioactive material in a biological system.
Glossary of Terms* *With permission from the Manual of Policies and Procedures for Radiation Protection, for the University of Minnesota, Department of Environmental Health and Safety, Radiation Protection
More informationNeutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom
Neutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom Problem P4 Rick J Tanner National Radiological Protection Board Chilton, Didcot, Oxon OX11 0RQ, United Kingdom Intercomparison
More informationFluence-to-Dose Conversion Coefficients for Muons and Pions Calculated Based on ICRP Publication 103 Using the PHITS Code
Progress in NUCLEAR SCIENCE and ECHNOLOGY, Vol. 2, pp.432-436 (20) ARICLE Fluence-to-Dose Conversion Coefficients for Muons and Pions Calculated Based on ICRP Publication 03 Using the PHIS Code atsuhiko
More informationTechnical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1
Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM
More informationLower Bound of Optimization for the Public Considering Dose Distribution of Radiation due to Natural Background Radiation
Lower Bound of Optimization for the Public Considering Dose Distribution of Radiation due to Natural Background Radiation Takatoshi Hattori a* a Central Research Institute of Electric Power Industry, Radiation
More informationRinghals AB. Routines for whole body counting at Ringhals NPP
Routines for whole body counting at Ringhals NPP Structure Equipment - Quick Scan - Whole body counter Marie Carlson M.Sc. Radiology and Dosimetry Radiation Physicist and deputy RP Controller Routines
More informationGCE AS and A Level. Physics A. AS exams 2009 onwards A2 exams 2010 onwards. Unit 5: Approved specimen question paper. Version 1.3
GCE AS and A Level Physics A AS exams 2009 onwards A2 exams 2010 onwards Unit 5: Approved specimen question paper Version 1.3 Surname Other Names Leave blank Centre Number Candidate Number Candidate Signature
More informationRecent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN)
Recent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN) V.Gressier, L. Van Ryckeghem, B. Asselineau, R. Babut, J.F. Guerre-Chaley, V. Lacoste, L.Lebreton,
More informationMeasurement of induced radioactivity in air and water for medical accelerators
Measurement of induced radioactivity in air and water for medical accelerators K. Masumoto 1, K. Takahashi 1, H. Nakamura 1, A. Toyoda 1, K. Iijima 1, K. Kosako 2, K. Oishi 2, F. Nobuhara 1 High Energy
More informationRADIATION SAFETY TRAINING SEALED SOURCES
RADIATION SAFETY TRAINING SEALED SOURCES PLEASE REFER TO THE RADIATION SAFETY HANDBOOK, PARTICULARLY THE SEALED SOURCES CHAPTER, AS A SUPPLEMENT TO THIS PACKET. Sealed source use at CU State and federal
More informationPreparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials
Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials A. Verbruggen, R. Eykens, F. Kehoe, H. Kühn, S. Richter, Y. Aregbe EUR 22944 EN - 2007 The mission of the IRMM
More informationSensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes
Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes B.M. Dantas, E.A. Lucena and A.L.A. Dantas Laboratório de Monitoração In Vivo Divisão de Dosimetria
More information1. RADIOACTIVITY AND RADIATION PROTECTION
1. Radioactivity and radiation protection 1 1. RADIOACTIVITY AND RADIATION PROTECTION Revised August 2011 by S. Roesler and M. Silari (CERN). 1.1. Definitions [1,2] 1.1.1. Physical quantities: Fluence,
More informationIdentification of Naturally Occurring Radioactive Material in Sand
Identification of Naturally Occurring Radioactive Material in Sand Michael Pope 2012 NSF/REU Program Physics Department, University of Notre Dame Advisors: Dr. Ed Stech, Dr. Michael Wiescher Abstract Radionuclides
More informationAtomic and nuclear physics
Atomic and nuclear physics X-ray physics Attenuation of x-rays LD Physics Leaflets P6.3.2.1 Investigating the attenuation of x-rays as a function of the absorber material and absorber thickness Objects
More informationThe Fundamentals of Moisture Calibration
The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity
More informationNeutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace
Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace Gregori, B.N.; Carelli, J.L; Cruzate, J.A.; Papadópulos, S. and Kunst, J.J. Presentado en: Second European of IRPA (International
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 6980-3 First edition 2006-10-01 Nuclear energy Reference beta-particle radiation Part 3: Calibration of area and personal dosemeters and the determination of their response as
More informationRadiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe
Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range
More informationSECTION 8 Part I Typical Questions
SECTION 8 Part I Typical Questions 1. For a narrow beam of photons, the relaxation length is that thickness of absorber that will result in a reduction of in the initial beam intensity. 1. 1/10. 2. 1/2.
More informationSample Examination Questions
Sample Examination Questions Contents NB. Material covered by the AS papers may also appear in A2 papers. Question Question type Question focus number (section A or B) 1 A Ideal transformer 2 A Induced
More informationExternal Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE
External Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE Joseph L Danek - CHP 2018 Fall Meeting FLORIDA CHAPTER HEALTH PHYSICS SOCIETY 1 Part 1 National Voluntary Laboratory Accreditation Program (NVLAP)
More informationImprovements to the UK Primary Standard Therapy Level Electron Beam Calorimeter
Radiotherapy Standards Users Group meeting NPL, 5 th June 2007 Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Mark Bailey, G A Bass, N D Lee, G A Mapp Contents 1. History
More information